T11024 - PC Pump Test/Calibration
Contents
1. GAMBRO Gambro Renal Products Technical Assistance Services http tech usa gambro com 800 525 2623 303 232 6800 Technical Update PH TR 11024 12 2005 TROUBLESHOOTING E qu i pmen t Ty pe Phoenix Dialysis System Su bj ect PC Pump Test Calibration From Gary Willis Purpose The purpose of this technical update is to clarify the existing Service Manual Calibration and Test procedures to insure the specified PC Pump pressure for the local altitude can be successfully met NOTE The PC Pump pressure is measured verified only in the PC PUMP TEST mode Pressure measurements observed in the PC PUMP CAL mode are INVALID and should be ignored The PC PUMP CAL mode is ONLY for entering the desired PC Pump speed change to achieve the specified pressure The following procedures assume that the R1 and R2 calibrations have been verified and that the PC Pump calibration test tool is properly installed PC Pump Test Procedure Summary 1 From the Service Manual Section 8 Table 8 1 record the PC pressure for your altitude 2 Select the PC PUMP TEST key Allow the Hydraulic value to stabilize at its maximum value and record the external pressure meter measurement 3 Using the example provided in the PC Pump Calibration Section 8 of the Service Manual calculate the percentage difference that will be entered later in the PC PUMP CAL mode Note A positive percentage will cause the PC Pump to speed up and provide a great
2. er negative pressure at the external pressure meter A negative percentage will cause the PC Pump to slow down and provide a lesser negative pressure at the external pressure meter 4 Select the PC PUMP CAL key and allow the Hydraulic value different from the TEST mode value to stabilize at its maximum Enter the percentage change calculated in the previous step and save the calibration NOTE Do not use any external pressure measurements observed while in the PC PUMP CAL mode 5 Select the PC PUMP TEST key and allow the Hydraulic value to stabilize at its maximum Record the external pressure measurement 6 Repeat steps 3 through 5 until the desired external pressure is within specification allowed NOTE Make pressure measurements and enter speed changes only after the PC Pump has stabilized at its maximum speed
Download Pdf Manuals
Related Search
Related Contents
Les nouveaux ESF - Vos informations personnelles 操作編 - SoftBank SELECTION Table of Contents Operating Instructions Installation Manual: SR90, CF90, TR90, SR90+ manual de instrucciones instruction manual VGN-N series - Sony Europe KOBIL Smart Token V1.0 User's Guide Philips AJ301DB/79 MP3 Docking Station User Manual Copyright © All rights reserved.
Failed to retrieve file