LIGHT SOURCE LX 2300 (Xenon 300W)
Contents
1. serving healthcare worldwide lo rudolf MEDICAL LIGHT SOURCE LX 2300 Xenon 300W IM 200 300 Servicing Manual WA001 GB 2006 01 11 MI rudolf MEDICAL Servicing Manual ly LIGHT SOURCE LX2300 Xenon 300W Summary 1 Warnings and Precautions 2 2 Technical Features 3 3 Principle Plan 4 4 Comprehensive view Setting up Plan 5 4 1 Filter board 6 4 2 Push button board 7 4 3 Standby board 8 4 4 Power supply 9 4 5 Servomotors IRIS 10 4 6 Time meter 11 4 7 Rear face 12 4 8 Connector for power supply outline 13 5 IRIS Regulating 14 6 Proceedings for the Change of the Lamp 15 7 In Case of Breakdown 18 8 Pieces List 19 Page 1 of 23 2006 01 11 MI WAoo1 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W A 1 Warnings and precautions Only an authorized staff can use this kind of equipment after a training Please read carefully the instructions manual before setting up and using the light source Before plugging the light source into the mains make sure that the electric features of your wiring are quite the same as those specified on the equipment The identification label which mention the service voltage the electric consumption etc is on the rear panel of the appliance Be careful HIGH VOLTAGE inside the appliance Don t open the rear panel of the equipment Maintenance only by the authorized staff Never have your eyes neither in the
2. TUA BLOC ALIMENTATION HATCH serving healthcare worldwide Page 19 of 23 WA001 GB 2006 01 11 MI lo rudolf MEDICAL LIGHT SOURCE LX2300 Xenon 300W Description Photographs MO XENON IRIS Servomotor IRIS XENON JACKET XENON FRAME serving healthcare worldwide Page 20 of 23 Servicing Manual WA001 GB 2006 01 17 MI serving healthcare worldwide lo rudolf MEDICAL LIGHT SOURCE LX2300 Xenon 300W Description Photographs SHORT STORZ CONNECTOR WOLF CONNECTOR SHORT OLYMPUS CONNECTOR Page 21 of 23 Servicing Manual WA001 GB 2006 01 17 MI serving healthcare worldwide lo rudolf MEDICAL LIGHT SOURCE LX2300 Xenon 300W Description Photographs REAR FAN ELECTRIC TIME METER CURTIS FILTER BOARD STANDBY BOARD Page 22 of 23 Servicing Manual WA001 GB 2006 01 17 MI rudolf MEDICAL Servicing Manual ly LIGHT SOURCE LX2300 Xenon 300W Description Photographs PUSU BUTTON BOARD S300 POWER SUPPLY 300W CENTRAL FAN serving healthcare RUDOLF MEDICAL GmbH Co KG mail rudolf med com worldwide ZollerstraRe 1 78567 Fridingen Germany Phone 49 74 63 99 56 0 Fax 49 74 63 99 56 56 Page 23 of 23 CE CE conformity label in compliance with the req
3. 23 2006 01 11 MI WAoo1 GB serving healthcare worldwide g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 6 Proceedings for the change of the lamp A Warnings and precautions A change of lamp can be made only by an authorized staff who must take safety rules into consideration Before any handling wait for the cooling of the lamp The powers and electric currents involved by the power supply of the source require a perfect knowledge of the equipment Send back your source to your supplier who will seize to make a servicing of your equipment If you have to change the lamp whereas the source is still tepid always use a pair of gloves and protective glasses After removing the lamp put it in its original packaging Never let it without any protection gas under high tension Don t bring any pressure nor tension on the lamp Send it back to the supplier in its original packaging Switch off the source and unplug it from the mains Put the source on a large and stable working area Position on 0 With a cruciform screwdriver unscrew the 4 screws on the left side of the casing of the source to have access to the lamp 4 screws Oe aa Page 15 of 23 2006 01 11 MI WAoo1 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W Unplug the white connector by a simultaneous pressure on both of the sides of it Re
4. Page 17 of 23 2006 01 11 MI WAoo1 GB serving healthcare worldwide g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 7 In case of breakdown No particular maintenance is necessary for this light source Actually it s enough to control the brightness intensity produced by the source and to change the lamp if necessary Nevertheless you might be confronted with a minor incident during its utilization Here above the causes and consequences of incidents which you could easily cure are listed In all the other situations please contact our After Sales Service that will be able to help you as soon as possible No signal of the front panel lights up when you switch the source on Control if the mains outlet on the rear face of the light source is rightly connected up to the network Control the good working order of the fuses use only fuses as specified on the rear face T4A The front panel of the light source lights up but the brightness flow is insufficient Control if the iris isn t totally closed by pressing the right button Control the working order of your light cable and of your optics Change your lamp The images produced are too clear Control if the iris isn t too opened There is no more light but the fan works and the led of the 2 brightness buttons flash The light source has a safety system which stops the power supply of the lamp if the inside temperature of the source is
5. board E tvozoLo i Microcontroller LI i a i a i a 4 i ti i i SBESBESBS S865 85554 To the power supply Potentiometer for the regulation of the servomotor iris position CN3 Molex bracket 2 pts 2 54 cf connectors for 300W power supply 1 2 Switch on control of the lamp continuity between 1 and 2 switched on lamp uncontinuity between 1 and 2 switched off lamp CN5 Molex bracket 4 pts 2 54 to the time meter 1 orange reset of the time meter 5V during the operation 5ms 2 blue counting 5V 3 red 5V 4 black Ground CN10 Autocom bracket 6pts 2 54 to the standby board Ground Standby button 5V switched on lamp Led of the Standby button oV switched on lamp switched off led 0 5V cyclic connection when winking led CTN voltage which depends on the temperature switching off voltage of the lamp gt 2V 5V Servomotor control square cyclic signal OV 5V oV duration which gives the iris position whatever the condition of the source switched on or in standby Au FWN A CN1 Autocom bracket 3pts 2 54 to the filter board 1 Unregulated 24V input 16V to 30V 2 Ground 3 Starting control of the main power supply with a relay unit flow from o gt nearly 3V 2 8 to 3 5V for starting releasing of the relay from 2 5V serving healthcare worldwide Page 7 of 23 2006 01 11 MI WA001 GB serv
6. burns serving healthcare worldwide Page 2 of 23 2006 01 11_MI WAoo1 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 2 Technical Features Xenon lamp Nominal power 300 W Color temperature 6 200 K Usual life of the lamp 500 hours for classical utilization cycles 30 min ON 15 min OFF Compatible light cables Storz Wolf and Olympus with other adaptaters Mechanical IRIS Adjustment of the light intensity with the push buttons Indication of the minimum and maximum light levels Litter L 300 mm D 311 mm H 125 mm Weight 5 kg Power supply voltage 115 230 V 50 60 Hz Electric Consumption 350 VA 2 THA Fuses Working temperature 10 a 40 C Moisture working 85 Storage temperature 20 a 45 C Storage moisture 85 Continuous service Unprotected against water falls IPXo Unadapted to an utilization in the presence of an anaesthetic mixture flammable with air with oxygen or nitrous oxide serving healthcare worldwide Page 3 of 23 2006 01 11 MI WAoo1 GB serving healthcare worldwide lo rudolf MEDICAL LIGHT SOURCE LX2300 Xenon 300W 3 Principle Plan Filter board Power supply 115 230 V S4A 115 230 V Sieh Standby board Push button board 84 3 84 2 Servomotor IRIS Servomotor IRIS 84 5 84 5 Page 4 of 23 Fans 84 7 Servicing Manual WA001
7. to the BP board Connector 9pts Unused Unused Connected on 6 12V Switch on control of the lamp return of the signal from the pin 8 oV switched off lamp or 12V switched on lamp Unused Power circuit lamp Unused Switch on circuit lamp output signal to optocoupler on BP board continually 12V Unused O ON nw FWN Molexconnector 2pts Switch on circuit of the lamp 1 Going signal from 8 to the power supply From the optocoupler to the BP board 2 12V return signal to switch on the lamp to 4 of the power supply Page 13 of 23 2006 01 11 MI WAoo1 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 5 IRIS regulation The iris position can be finely adjusted with the potentio meter of the BP board Connect the source to the mains Rock the switch backwards on I The green led on the front panel flashes Switch on the source with the Standby button Press until it lights up to have the maximum light Keep on pressing the button and regulate the potentio meter to put the IRIS at 2 or 3 mm from the brace put the IRIS at 2 or 3 ie mm from the brace Lessen the light to the minimum by pressing until it lights up Control that the beam of light is directed on the IRIS steel sheet and that the IRIS doesn t touch the fan beam of light somemm from the fan Adjust again from the necessary serving healthcare worldwide Page 14 of
8. too high When the light source has finally cooled it can be switched on again Control if there s enough space all around the light source to allow a good cooling 15 cm on each side Control that there isn t any obstacle to the ventilation on the rear face of the source lf the defect is persistent and if you have to send us back the light source take care of sending it in its original packing Then it s better to add your light cable to the packing so that we can examine your outline Insucha case it would be kind of you to add to your delivery note a little explanation regarding the defect you have noticed When you will receive your equipment it will be advisable to control its working order and enter reservations in respect of the transport note if necessary Then you will spare 48 hours to confirm them by registered letter to the carrier After this delay the carrier will be able to refuse those reservations If an equipment sent by us is damaged during its transport the amount of the repair will be charged either to the carrier if the reservations have been entered within the required time or to the consignee in the opposite case So remember to control as soon as possible the good working order of the equipment which has just been carried Page 18 of 23 2006 01 11_MI WAoo1 GB lo rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 8 Pieces list Description Photographs FUSE
9. GB 2006 01 11 MI g rudolf LIGHT SOURCE LX2300 Xenon 300W 4 Comprehensive view Setting up Plan Rear panel MEDICAL Servicing Manual RIA sa RTT Va MII x lt red 0 22mm2 Filter board Wire lampi E Ventilateur black 0 22mm2 Servomotor Graving piece 3 pts 2 54 Graving piece 6 pts 2 54 Front panel ll inside view black 0 22mm2 BP Board E orange 0 22mm2 Time meter orange 0 22mm2 7 uli SERIO red 0 22mm2 healthcare worldwide Page 5 of 23 red 0 22mm2 red orange or yellow orange Twisted cord red or yellow 0 22mm2 WA001 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 4 1 Filter board VI a Lc Oo 7 eV28 4 0 02 2M3 Funu E4 0 02 23 DALA Ak blue to the power supply brown to the power supply blue from mains socket brown from mains socket i gt w 4 E ia un Cc CN2 Autocom bracket 3pts 2 54 to BP board a 1 Unregulated 24V output 16V to 30V 5 2 Ground 3 Starting control of the main power supply with a relay unit flow from o gt nearly 3V 2 8 to 3 5V for the x starting setting off of the relay unit on and after 2 5V a serving D healthcare z lt worldwide Page 6 of 23 g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 4 2 Push button
10. axis of the light output nor at the light cable output In order to avoid every risk of burn or electrocution don t expose this equipment to rain or to an excessive moisture If any fluid infiltrate into the appliance stop it immediately and remove the plug from the mains Never let the mains flex fall and never put heavy things on it If the mains flex is damaged unplug immediately the light source from the mains Never start it with a damaged mains flex To remove the mains flex without damaging it take it by the end Never pull on the cable itself The mains plug must be near to the light source and easily accessible Check the ventilation to avoid any overheating of the equipment fasten a minimum space of 15 cm on the left on the right and backwards Be careful Never use the appliance next to a flammable gas Never put the equipment next to a heat source or in a place exposed to vibrations or shocks Don t use abrasive corrosive products to clean the surface of the light source use only cleaning products and disinfectants generally used in hospitals Clean the light source as recommended by the standard cleaning proceedings of your hospital This light source and its accessories require no special treatment at the end of their life Be careful Don t put the end of the light cable directly on the patient nor on every flammable material paper gauzes operative fields because it may be very hot risks of
11. ing healthcare worldwide g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 4 3 Standby board CN4 socket Molex 3pts 2 54 towards servomotor 1 black Ground 2 red 5V 3 white Servomotor control cycle square signal o 5V time at oV indicating to the servomoto the IRIS location regardless of state of the light state ON or stand by CN9 socket Autocom 6pts 2 54 towards Standby board Ground Standby Pushbutton 0 5V Led of Standby Pushbutton oV when light ON led is OFF o 5V cycle when led is flickering CTN voltage depending on temperature switching off voltage forsecurity gt 2V 5V UW FWN Page 8 of 23 2006 01 11 MI WAoo1 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 4 4 Power supply 2 00 110V 220V power supply DC300 connector J1 connector J2 J2 connector connector for power outline J connector connector for fan power supply PTI Control voltage test point PT2 o Volt test point HV HV lamp output Connector for fans power supply K1 4pts 2 54 Molex connector for fans supply 4 1 e n black red 1 Unused 2 Unused 3 12V DC 4 OV The wire of the lamp can be identified thanks to a red sheath serving healthcare worldwide Page 9 of 23 2006 01 11 MI WAoo1 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 4 5 Servomot
12. move carefully the lamp from its stand by pulling it aft Then put the new lamp in the same position as the former one according to the opposite proceedings It s not at all necessary to screw or unscrew the nut behind the wheel serving Warnings and precautions healthcare Don t let any fingers mark on the lamp Use gloves worldwide Page 16 of 23 2006 01 11 MI WAoo1 GB g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W Control the right position of the lamp Always be careful to have at least one replacement lamp in stock Reset of the time meter After changing a lamp you have to reset the time meter control that the light source is on Standby press simultaneously the two brightness buttons for ten seconds The time meter must be reset Remark The usual lasting of the lamp is 500 hours It s an average for 30 min ON and 15 min OFF Shorter utilizations can reduce the lasting in a significant way The technical data is Rated lamp current A 16 Rated lamp wattage W 300 Initial voltage range V 14 19 Ignition voltage cold hot kV Max 20 max 20 Colour temperature K Approx 6 200 Colour rendering index Ra gt 95 Beam to axis angle deg Max 30 Length mm Max 80 Length p mm Max 30 Length r mm Max 62 Reflector diameter D mm Max diam 82 5 Diameter d1 mm Diam 31 serving healthcare worldwide
13. or Iris The connector has got three wires and allows to control the servomotor There are the earth black wire the 5V red wire and the control white wire a black wire lt _______ red wire lt lt white wire Adjustment of the iris see 5 IRIS Regulation serving healthcare worldwide Page 10 of 23 2006 01 11 MI WAoo1 GB lo rudolf LIGHT SOURCE LX2300 Xenon 300W 4 6 Time meter Technical specification Recording and display Precision Consumption Utilization temperature Moisture CURTIS E ral Connection on faston type hull Service voltage red black orange blue Molex 2 54 4 points black red blue orange serving healthcare worldwide Page 11 of 23 MEDICAL Servicing Manual 99 999 9 hours 0 04 1 MA max 40 C to 85 C 95 HR uncondensed when 38 C 4 76 mm 5 to 12 Vcc 2006 01 11 MI WAoo1 GB lo rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 4 7 Rear face outcoming air incoming air SUNON KDE 1208PTS3 fan SUNON KDE 1208PTS3 fan switch SUNON fan KDE 1208PTS3 DC 12V 1 4W serving healthcare worldwide Page 12 of 23 WA001_GB 2006 01 1_MI serving healthcare worldwide g rudolf MEDICAL Servicing Manual LIGHT SOURCE LX2300 Xenon 300W 4 8 Connector for power supply outline Power supply connectors for 300W lamp to the power supply
14. uirements of the guideline 93 42 EEC 2006 01 11 MI WAoo1 GB
Download Pdf Manuals
Related Search
Related Contents
1771-2.109, Loop Controller Interface Module, Product Data ② ① - ECHOWELL ロータリ真空ポンプRPV064 GUIA DE INSTALAÇÃO Série de câmeras de rede AXIS M31 USER`S MANUAL - Green Fit Co. Volkshochschule Celle - Deutsches Institut für Erwachsenenbildung Hier beginnen HL-2135W / PC CHIPS P23G (V1.0) 簡易設置形エアーシャワー Copyright © All rights reserved.