1. 2 Keying Disc Oxygen Nitrous Oxide Medical Air Vacuum WAGD Front Body Cylinder Cylinders between Gemini and Diamond Ill applications Repairs may result in leakage Replacement Components l Gemini and Diamond IlI Outlets Gemini Quick Connect Console Outlet Diamond IlI Quick Connect Console Outlet Item Part Number Description ltem Part Number Description 1 833436 PG Spring pkg of 10 833436 PG Spring pkg of 10 2 4107 2097 44 Piston pkg of 10 z 4107 2097 44 Piston pkg of 10 3 622589 PG O Ring Red pkg of 10 3 622589 PG O Ring Red pkg of 10 4 415038 PG Washer pkg of 10 4 415038 PG Washer pkg of 10 5 833953 PG Lock in Adapter Spring 5 833953 PG Lock in Adapter Spring pkg of 10 pkg of 10 6 831344 PG Eyelet pkg of 50 6 4107 2097 31 Eyelet pkg of 50 7 4107 4005 21 Screw 6 32 x 1 1 2 Z 4107 4005 21 Screw 6 32 x 1 1 2 8 Keying Disc 8 Keying Disc 616318 10 Oxygen 4107 6520 74 Oxygen 616318 11 Nitrous Oxide 4107 6520 73 Nitrous Oxide 616318 12 Medical Air 4107 652075 Medical Air 616318 13 Vacuum 4107 652076 Vacuum 616318 19 WAGD 9 4107 6520 87 Front Body Cylinder Repair Kits Gemini To repair 10 outlets order 1 each of NOTE Do not interchange Keying Discs or Front Body 290804 PG and 1 each of 831344 PG as shown as item 6 Diamond Ili To repair 10 outlets order 1 each of 290804 PG and 1 each of 4107 2097 31 as shown as item 6 Cylinders b
2. French CSA ISO System CSA AGS S Spanish NFPA Instrument Air INA Available in DISS only Example Oxygen NFPA DISS Oxygen O DISCEI U OXY Not avallable for Puritan Bennett Please see note above Amico Corporation 85 Fulton Way Richmond Hill ON L4B 2N4 Canada 71 East Industry Court Deer Park NY 11729 USA Toll Free Tel 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 www amico com Date 2 19 13 Submittal 1 SUBMITTAL Re Submittal Date Project Memroial Hospital Medical Gas Upgrades Project North Campus General Contractor GH Phipps Mechanical Engineer AE and Associates Architect Intergroup Architects Number Taken Medical Gas Outlets Medical Air Systems Architect Engineer Comments The above drawings are returned with action as designated in accordance with the following legend A No Exceptions Taken B Make Corrections Noted Signed C Revise and Resubmit D Rejected see Remarks Olson Plumbing amp Heating n a Kris Damgaard gt Olson Plumbing and Heating Co Mechanical Contractors PO Box 2556 121 W Cucharras St Colorado Springs CO 80901 Phone 719 635 3563 BEACONVIEDES SSB 840 05 4107 9507 40 00 Page 1 of 2 18 May 2012 Series B Medical Gas Outlet Medaes Pin Index Key Style SPECIFICATION Medzs Di nd Medical Gas Wall Outle The pin index medical gas outlets are gas spec
3. in piping system e Check system and or outlet System demand greater than e Check supply source and supply source or system system sizing capacity Outlet pressure or vacuum does e System pressure vacuum not e Reset not meet specification MANO1 049 Part No 847706 00 Rev COO set correctly Check supply source and system sizing System demand exceeds supply source capability or system capacity Restriction in piping system or Check and remove outlet Replacement Components Gemini and Diamond Ill Outlets Concealed Wall Outlet Back Body Console Wall Outlet Back Body ltem Part Number Description _ 4107 2097 29 Spring Bronze for pressure gases pkg of 10 1 i 4107 2097 30 Spring SS for vacuum and KE WAGD pkg of 10 gz Rubber Bushing 2 525344 00 Secondary Check each E A 3 4107209738 O Ring Black pkg of 10 LRS 4 622601 PG Rubber Bushing pkg of 10 WAA AA AN 5 4107 2097 88 Retaining Ring pkg of 10 ii J Repair Kits Below Repair Kits Contain Components g to Repair 10 Backbodies Secondary gt p a E gt E tee Check Towards Towards Finish Part Number Description Rough in Assembly 290802 PG Secondary Check Repair Kit Pressure Service Items 1 3 4 5 290803 PG Secondary Check Repair Kit NOTE Vacuum WAGD Rubber bushing item 4 must be oriented as Items 1 3 4 5 shown for proper operation When replacing rubber bushings item 4 replacement of secondary
4. may be cleaned with a mild detergent solution or wiped with a disinfectant used for patient room surfaces that is compatible with acetal and nylon plastic anodized aluminum and die cast zinc Lubricate elastomer seals sparingly with Krytox GPL206 Refer to Tools and Service Materials DO NOT USE OIL Inspection and Testing Medical gas outlets in high use areas should be inspected at least every six months and more frequently if outlets are used daily All outlets should be inspected at least once a year Follow National Fire Protection Association Standard 99 99C Gas and Vacuum Systems and Canadian Standards Association Standard Z 305 1 Nonflammable Medical Gas Piping Systems requirements and methods for periodic testing and maintenance of medical gas and vacuum systems MANO1 049 Part No 847706 00 Rev COO Outlet Identification Before servicing medical gas outlets it is very important to identify exact type and vintage of outlet Wall Outlets Gemini wall outlets are identified by a Beacon Medical or BeaconMed s logo on right hand side of outlet Diamond IlI DISS Ill wall ceiling outlets may be identified by the words Hill Rom or Med s cast into cover plate on lower right hand side and Model No 300 series or 400 series on back side of cover plate Diamond II wall ceiling outlets are identified by Model A30 series on back side of cover plate Console Outlets Remove console cover
5. plate to expose front body Color Coded Cop a Plastic A 4 Galvanized LAN O Front Body Plate Gemini Console Front Body Gemini console outlets will appear as shown above Front body will have a solid rectangle of color coded plastic approximately 1 7 8 wide x 2 3 4 tall Raised circular keying portion of front body will rise approximately 200 DIET a OEE EROT Diamond til Console Front Body Diamond Ili console outlets will appear as shown above Front body will be easily identified by appearnace of galvanized front body plate visible in corners Diamond ll medical gas outlets must be upgraded to Gemini Medical gas outlets Upgrade of Diamond II Medical Gas Outlets Rough in assembly must be upgraded to accept Gemini front body This requires installation of a new spring secondary check o ring rubber seal and retaining ring Order one Secondary Check Repair Kit P N 290802 PG for pressure service or 290803 PG for vacuum or WAGD service and 10 each of Secondary Check P N 525344 00 for every 10 outlets to be upgraded Order one each of Gemini or DISS finish assembly ref page 6 for each corresponding gas service MANOT 049 5 Part No 847706 00 Rev COO Outlet Assembly Ordering Gemini Diamond Ill Assembly Numbers Outlet Type Gas Service __Rough in Assembly Finish Assembly Gemini Concealed Quick Connect Gemini Concealed DISS Gemini Console Quick Connect DI
6. 0 13 for additional electrical receptacles Note that a typical complete outlet consists of one Rough in one matching Latch Valve and one Trim Plate ama ny ar aa Background Text Wall Outlet Wall Outlet O 233010 0 C 230941 00 C 23094200 C 230943 0 C 230946 00 C 230947 00 Accessories Note good design provides one slide with each vacuum Duplex Electrical Receptacle Gray 15A 125V Wall outlet style Complete Trim Plate Wall Large Style Trim Plate Console Small Style Gas service identification is affixed to the inlet tube and the face of the rough in plate A secondary vaive is installed in the outlet block of the rough in assembly except vacuum and WAGD for both pressure testing and preventing gas flow when the latch valve assembly is removed for service The outlet block contains a double seal to prevent gas leakage between the rough in and latch valve assemblies after the wall is finished Outlets using a single o ring seal are not acceptable c ve Assembl The latch valve assembly includes a captured o ring seal primary valve is gas specific for the labeled service and accepts only hose and apparatus with corresponding pin index adapters The latch valve assembly is indexed to the corresponding rough in assembly to avoid accidental cross connection and self adjusts up to 3 4 1 9 cm to allow for variation in finished wall thickness from 1 2 12 5 mm up to 1 1 4 3 175 cm Trim ion Either a di
7. SS Console Diamond Ill Console Quick Connect Oxygen Nitrous Oxide Air Vacuum WAGD Vacuum Slide Blank Oxygen Nitrous Oxide Air Vacuum Nitrogen WAGD Carbon Dioxide Oxygen Nitrous Oxide Air Vacuum WAGD Oxygen Air Vacuum Oxygen Nitrous Oxide Air Vacuum Finishing plates are not supplied with these assemblies E MANO1 049 Part No 847706 00 Rev COO Complete assemblies for concealed piping and console applications 234510 00 234511 00 234512 00 234513 00 234519 00 234500 00 234500 00 234510 00 234511 00 234512 00 234513 00 234514 00 234519 00 234520 00 234410 00 234411 00 234412 00 234413 00 234419 00 234410 00 234412 00 234413 00 234410 00 234411 00 234412 00 234413 00 234610 00 234611 00 234612 00 234613 00 234619 00 234600 00 234601 00 234710 00 234711 00 234712 00 234713 00 234714 00 234719 00 234720 00 234810 00 234811 00 234812 00 23481 3 00 234819 00 4107 2097 49 4107 2097 51 4107 2097 50 4107 2097 45 4107 2097 46 4107 2097 48 4107 2097 47 Notify appropriate hospital personnel when performing service or maintenance to medical gas systems Consult Troubleshooting Guide to determine required repair Outlets incorporate a primary valve and secondary check Secondary check prevents system gas from escaping when primary valve is removed for service Refer to Replacement Components for identification ofreplacement part numbers Quick Connect
8. and DISS Valve Assemblies v Quick connect and DISS valves may be removed for replacement of seals and some components without interrupting service to other outlets in piping system To check a quick connect or DISS valve for leaks apply leak detector solution at primary valve opening both with and without adapter connected Bubbles indicate pressure leaks Large leaks may be audible NOTE Replace all elastomer seales whenever outlets are disassembled Front Body Disassembly of Gemini Diamond III DISS Ill outlets for servicing and parts replacement is illustrated in exploded parts breakdown pages 10 12 NOTE Gas supply need not be turned off when removing front body secondary check and bushing in back body prevents flow from outlet Adapter locking spring on Gemini Diamond III front body housing and front body o ring can be replaced by dismantling housing as follows 1 Loosen two front body screws NOTE On recessed and surface mount outlet models screws are held in place by rubber washers console models use elastic stop nuts 2 Carefully remove front body assembly from back body assembly 3 Using needle nose pliers crush inside flange of brass eyelets and remove from assembly Remove and replace all worn or damaged parts New brass eyelets must be used when reassembling front body assembly Insert eyelets and flare rear of each with 60 tapered punch A WARNING Be certai
9. check item 2 is also recommended MANO1 049 Ea Part No 847706 00 Rev COO Replacement Components E Gemini and Diamond IlI Outlets Gemini Quick Connect Concealed Wall Outlet Item Part Number Description Spring pkg of 10 Piston pkg of 10 O Ring Red pkg of 10 Washer pkg of 10 Lock in Adapter Spring pkg of 10 Eyelet pkg of 50 Screw 6 32 x 1 1 2 Keying Disc Oxygen Nitrous Oxide Medical Air Vacuum WAGD To repair 10 outlets order 1 each of 290804 PG and 1 each of 831344 PG as shown as item 6 To repair 10 outlets order 1 each of 290804 PG and 1 each of 4107 2097 31 as shown as item 6 1 833436 PG 2 4107 2097 44 J 622589 PG 4 415038 PG 5 833953 PG 6 831344 PG A 4107 4005 21 8 616316 10 616316 11 616316 12 616316 13 616316 19 Repair Kits Gemini Diamond Ill Part Number 290804 PG Description Primary Check Repair Kit Items 1 2 3 4 and 5 10 MANO1 049 Part No 847706 00 Rev COO PNP Diamond Ill Quick Connect Wall Outlet Part Number 833436 PG 4107 2097 44 622589 PG 415038 PG 833953 PG 4107 2097 31 4107 4005 21 4107 4006 08 4107 4006 09 4107 4006 11 4107 4006 10 4107 4006 75 4107 6520 87 NOTE Do not interchange Keying Discs or Front Body Description Spring pkg of 10 Piston pkg of 10 O Ring Red pkg of 10 Washer pkg of 10 Lock in Adapter Spring pkg of 10 Eyelet pkg of 50 Screw 6 32 x 1 1
10. e cast light gray epoxy powder coated trim plate or a smaller plastic trim plate can be provided to trim each outlet The die cast plate is designed specifically to fill the space between adjacent outlets Either trim plate allows latch valves to be individually removed for servicing Rough in Assembly Console Outlet Latch Valve Assembly Wall or Console O 230940 00 L 230948 00 C 120972 00 325161 00 616358 00 BeaconMed s 1800 Overview Drive Rock Hill SC 29730 Phone 803 817 5600 Fax 803 817 5750 www beaconmedaes com WALL CONSOLE OUTLET x lt ui a Z a e a e A iad oS A fea feet n A BEACONVIEDAS SSB 840 05 4107 9507 40 00 Page 2 of 2 18 May 2012 Latch Valve Interlock screws 1 2 OD 4 s Nominal provided ASTM B 819 Type K When wallboard is less than 2 12 5 mm thick as in consoles and headwalls console outlets should be considered Plates may gang together in any length Up to three may be ganged without additional support however top and bottom support is always recommended Total finished length will be 5 127 mm x number of outlets in the gang e g three outlets 5 x 3 15 Wallboard and wall covering 1 2 112 5 mm MINIMUM 1 1 4 31mm MAXIMUM thickness 4 ihis Console Rough in 104 mm oi i onsole 7 87 16 Trim Plate Ya 215 mm 616358 00 Finish Assembly O
11. etween Gemini and Diamond IH applications Repairs may result in leakage Part Number Description 290804 PG Primary Check Repair Kit Items 1 2 3 4 and 5 MANO1 049 E Part No 847706 00 Rev C00 Replacement Components Components DISS Concealed Wall Outlet DISS Console Outlet item Part Number escription 1 Piston Not available as separate item Order appropriate repair kit below PA O ring 622585 PG For Oxygen Vacuum Evacuation and WAGD pkg of 10 622586 PG For N20 Air N2 and CO2 pkg of 10 Refer to page 6 for complete replacement assemblies Repair Kits To repair 10 outlets order 1 repair kit for appropriate gas service Part Number Description NOTE 290805 PG Primary Check Repair Kit Oxygen Repair Kits are not interchangeable 290806 PG Primary Check Repair Kit N20 Air N2 and CO2 between EVACUATION and WAGD services Be certain to order correct 290807 PG Primary Check Repair Kit Vacuum and Evacuation repair kit 290808 PG Primary Check Repair Kit WAGD Part No 847706 00 Rev C00 Pg 12 2012 BeaconMedees All rights reserved Corporate Headquarters For technical support or to place an order call BeaconMedzs BEACONMEDAS 1000 ong aes ee NN rn A Rock Hill SC 29730 1 888 463 3427 Phone 803 817 5600 Fax 866 635 3296 Fax 803 817 5750 www beaconmedaes com Ceiling Outlet DISS or Quick Disconnect tailing Out d M N General Specifications Feature
12. ific for the services indicated and accept only corresponding pin index adapters The outlets are UL listed CSA certified and are fully compliant with the latest edition of NFPA 99 All outlets are 100 tested for flow leaks and connector attachment The outlets are cleaned for oxygen service in the manufacturing process The outlets are made in the U S A Outlet Design A complete medical gas outlet consists of a gas specific rough in assembly for installation before the wall or console is completed A matching gas specific latch valve assembly and trim plate are installed after the finish is complete R h in emb The rough in assembly is of modular design and includes a gas specific 16 gauge steel mounting plate Wall versions permit on site ganging of multiple outlets in any order on 5 12 7cm centerline spacing A 3 8 9 5 mm high metal flange around the outlet opening provides a plaster barrier A temporary cover is provided to keep debris out of the outlet during installation Console rough ins fit standard electrical box cutouts and screw locations A machined brass outlet block is permanently attached to the rough in plate to permit the 1 2 OD 3 8 nominal type K copper inlet tube to swivel 360 for attachment to the piping system Slide Wall outlet style Complete Slide Surface mount style Blank Gas Complete with RI and finish Blank Gas Cover plate only See specification sheet SSB 84
13. n gas supply to outlet is shut off before disassemblying back body Back Body Remove internal parts from back body assembly by lifting inside 1 lock ring from rear housing with small screwdriver as illustrated below Carefully work rubber bushing from back body with screwdriver Remove check unit and spring Replace components as necessary and reassemble outlet Y Lock ring MANO1 049 Part No 847706 00 Rev COO Troubleshooting Guide Symptom Possible Cause Corrective Action Quick Connect or DISS Valve Leaks without quick connect adapter in place e with quick connect adapter in place e without DISS nut and nipple in place e with DISS nut and nipple in place Adapter will not engage and or release Flow rate does not meet specification Faulty primary seal e Replace primary seal Faulty rubber bushing e Replace rubber bushing Faulty primary seal Replace primary seal Faulty rubber bushing e Replace rubber bushing Faulty quick connect stem e Replace quick connect stem Faulty primary seal e Replace primary bushing Faulty rubber bushing e Replace rubber bushing Faulty DISS nipple e Replace DISS nipple Faulty O Ring seal e Replace O Ring seal Damaged adapter e Replace adapter Wrong adapter e Check adapter outlet gas indexing and labels Damaged frontbody assembly Replace frontbody assembly Reset system pressure vacuum Piping system pressure vacuum incorrectly set e Restriction
14. n mounted the outlet shall adjust from 3 8 10mm up to 1 25mm variation in wall thickness The latch valve assembly shall be Onmeda Diamond Quick Disconnect Chemetron Quick Disconnect Puritan Bennett Quick Disconnect or D I S S Diameter Index Safety System depending on the type ordered All M RI outlets are manufactured from nonferrous materials All outlets shall be cleaned and degreased for medical gas ser vice factory assembled and tested The outlets are UL Listed and ETL Listed Amico products comply with NFPA 99 and CSA i O NOTE aa Inch Amico recommends that you have a blank space between outlets and slides mm Maximum 3 outlets may be ganged together without additional support Technical Specifications Wall thickness may vary from 3 8 to 1 10mm to 25mm 5 1 2 140 j Saar 3 NOTE For CSA ISO suction inlets are not supplied with a secondary check valve A pressure plug is provided for testing purposes rated at 150 psi 1 034 kPa Model Numbers O DISCEI L XXX The XXX defines the Gas DIS DISS Oxygen OXY QD Ohmeda Compatible Medical Air AR CH Chemetron Compatible Nitrous Oxide N20 PB Puritan Bennett Nitrogen NIT Carbon Dioxide C02 MedVac VAC Waste Anesthetic The L defines the Language Gas Disposal NFPA WAG U English NFPA Oxy Air Vac and WAG only Anesthetic Gas E English CSA ISO Scavenging F
15. nT GHPhipps CONSTRUCTION COMPANIES PROJECT MHS Medgas Upgrades TO Intergroup Architects 2000 W Littleton Blvd Littleton CO 80120 ATTN Dale Skinner 5995 Greenwood Plaza Blvd Suite 100 Greenwood Village CO 80111 4710 303 571 5377 303 629 7467 fax TRANSMITTAL DATE 02 21 2013 No 13D1315 0001 RE 22 63 13 Gas Piping for Laboratory JOB 13D1315 Ph Fax 303 738 8877 WE ARE SENDING SUBMITTED FOR ACTION TAKEN Shop Drawings Approval Approved as Submitted Letter Your Use Approved as Noted Prints As Requested Returned After Loan Change Order Review and Comment Resubmit Plans Submit Samples SENT VIA Returned Specifications Product Data Attached Returned for Corrections Other Separate Cover Via Due Date 02 26 2013 NOTES Item Package Code Rev Copies Date Description Status Submittal 22 63 13 22 63 13 01 1 1 02 18 2013 Med Gas Outlets Architect Review CC Signed Dan Lane aS GHPhipps CONSTRUCTION COMPANIES Project MH UCH Medgas Upgrades Phipps Job 13D1315 Submittal 22 63 13 01 Submittal Stamp Sheet Intergroup Architects 2000 West Littleton Blvd Littleton CO 80120 Gerald H Phipps Inc SUBMITTED GH Phipps GH Phipps has reviewed approved and herby submits the attached in accordance with the cont
16. piping system When appropriate adapter is inserted into outlet primary valve is opened permitting gas flow Mechanical latch secures adapter in outlet Primary seal closes when adapter is removed Outlets are typically for oxygen nitrous oxide air or vacuum and have been cleaned for indicated service A WARNING Do not use oil on adapters or outlets Gemini Concealed Quick Connect Front Body DISS outlets are typically used in ceilings of operating rooms or in place of quick connect outlets Connection to outlet is by means of a nut and nipple Gas specific indexing to prevent accidental connection to wrong service is accomplished by a series of interference diameters in accordance with Diameter Index Safety System DISS published by Compressed Gas Association CGA in Standard V 5 Nipple opens primary valve as nut is tightened to secure nipple in outlet Primary valve closes when nipple is removed Gemini Concealed DISS Front Body Description Manufacturer MANO1 049 3 Part No 847706 00 Rev C00 Tools and Service Materials and Service Materials Description Manufacturer Lubricant Krytox GPL206 BeaconMedees P N GRSO1 007 No 2 Phillips screwdriver Local source No 2 Flat blade screwdriver Local source Oxygen compatible leak Sherlock or equivalent detector solution Cleaning and Lubrication BeaconMedes outlets are cleaned for oxygen or the indicated service Exposed surfaces of outlets
17. ract documents Note to Subcontractor Material Supplier Subcontractor Material Supplier remains responsible for the confirmation and correlation of dimensions at the jobsite fabrication processes and construction techniques coordination of the work with the work of other trades and satisfactory performance of the work Architect Engineer Other NOTES Signed Dan Lane LEED AP BD C GH Phipps Representative SUBMITTAL Date 2 20 13 Submittal 1 Re Submittal Date Project Memorial Hospital Medical Gas Upgrades Central Campus General Contractor GH Phipps Mechanical Engineer AE Associates Architect Intergroup Architects Section Description Action Number Taken Med Gas Outlets Medical Air Systems Architect Engineer Comments The above drawings are returned with action as designated in accordance with the following legend A No Exceptions Taken C Revise and Resubmit B Make Corrections Noted D Rejected see Remarks Olson Plumbing amp Heating Kris Damga Olson Plumbing an Signed Mechanical Contractors PO Box 2556 121 W Cucharras St Colorado Springs CO 80901 Phone 719 635 3563 wal Orel lead Service Manual Gemini Diamond ll and Diamond lll Medical Gas Outlets for Concealed Piping and Console Applications MANO1 049 Part number 847706 00 Rev C00 BEACONMEDAS Table of Contents General Produc
18. ren 5 127 m 3 3 4 96 mm Sh 4 104 mm ii wa Ne B i 82 mm K ez gi Standard Wall Trim Plate 325161 00 BeaconMedzs 1800 Overview Drive Rock Hill SC 29730 Phone 803 817 5600 Fax 803 817 5750 www beaconmedaes com
19. s Flow rate exceeds the requirements of NFPA and CSA Accepts quick disconnect gas specific adapters or DISS adapters depending on which front assembly is ordered 100 Hydrostatic testing of each outlet and cleaned for medical gas service Indexed to eliminate interchangeability of gas services Up to 1 25mm wall thickness adjustment 360 Swivel Inlet Pipe for easy installation Universal Rough In Assembly to accept Quick Disconnects or DISS front adapters Can be interchanged at any time Outlets are also available for M R I use C GH US LISTED GI The medical gas outlet shall be an Amico Alert 1 series Outlets shall be manufactured with a 6 1 2 165 mm length type K 1 2 12 7 mm outside diameter 3 8 nominal size copper inlet pipe stub which is silver brazed to the outlet body The body shall be of 1 5 16 33 mm diameter one piece brass construction For positive pressure gas services the outlet shall be equipped with a primary and secondary check valve and the secondary check valve shall be rated at a maximum of 200 psi 1 379 kPa in the event the primary check valve is removed for maintenance Outlet bodies shall be gas specific by indexing each gas service to a gas specific dual pin indexing arrangement on the respective identifica tion module A large color coded front plate shall be used for ease of gas identification and aesthetic appeal With the back rough i
20. t Description cscccescecceeteereeees 3 Tools and Service Materials 2 ccsssssesesesseeeseees 4 Cleaning and Lubrication s sssidesscsonsssorcuesssissiseses 4 INSPOCHON BND FESEING rsceseccacssissdsiadivesisniscerenceats eased 4 Outlet lA SrtHCALION lt recceeonvvedsersesnevsecasesovertenseon 5 Diamond Il Medical Gas Outlets c ccsesseeeeeeeeees 5 Outlet Assembly Ordering ssepesssecesnccesessercseceners 6 SOOVICE ARENE aS PEENES E TET 7 Troubleshooting Guide cccecceeseereeeeceeneeeeesensesees 8 Replacement COMPONENLS ccsscseeeceereeseeeseeerees 9 12 TT Introduction This manual provides information needed to service BeaconMedzs Gemini Diamond Il and Diamond Ill Gas Outlets Information contained in this manual is intended for use by technicians or personnel qualified to repair and service medical equipment MANO1 049 Part No 847706 00 Rev COO Definition of Statements Statements in this manual preceded by the following words are of special significance A WARNING Means there is a possibility of injury or death to yourself or others A CAUTION Means there is a possibility of damage to the unit or other property NOTE Indicates points of particular interest for more efficient and convenient operation General Product Description BeaconMedzes medical gas and vacuum quickconnect outlets provide a means of connecting to medical gas