1. 0 05 0 12 0 04 0 011 0 005 50 kHz 100 kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100 kHz 300 kHz 4 00 0 50 4 00 0 50 4 00 0 50 0 20 0 02 Resistance 100 0000 Q 1 mA Current Source 0 0030 0 0030 0 008 0 004 0 010 0 004 0 0006 0 0005 1 000000 kQ 1mA 0 0020 0 0005 0 008 0 001 0 010 0 001 0 0006 0 0001 10 00000 kQ 100 pA 0 0020 0 0005 0 008 0 001 0 010 0 001 0 0006 0 0001 100 0000 kQ 10 pA 0 0020 0 0005 0 008 0 001 0 010 0 001 0 0006 0 0001 1 000000 MQ 5 0 pA 0 002 0 001 0 008 0 001 0 010 0 001 0 0010 0 0002 10 00000 MQ 500 nA 0 015 0 001 0 020 0 001 0 040 0 001 0 0030 0 0004 100 0000 MQ 500 nA 10 MQ 0 300 0 010 0 800 0 010 0 800 0 010 0 1500 0 0002 dc current 10 00000 mA lt 0 1 V Burden Voltage 0 005 0 010 0 030 0 020 0 050 0 020 0 0020 0 0020 100 0000 mA lt 0 6 V 0 010 0 004 0 030 0 005 0 050 0 005 0 0020 0 0005 1 000000 A lt 1 0 V 0 050 0 006 0 080 0 010 0 100 0 010 0 0050 0 0010 3 00000 A lt 2 0 V 0 100 0 020 0 120 0 020 0 120 0 020 0 005 0 0020 True rms 1 000000 A 3 Hz 5 Hz 1 00 0 04 1 00 0 04 1 00 0 04 0 100 0 006 ac current 5 Hz 10 Hz 0 30 0 04 0 30 0 04 0 30 0 04 0 035 0 006 10 Hz 5 kHz 0 10 0 04 0 10 0 04 0 10 0 04 0 015 0 006 3 00000 A 3 Hz 5 Hz 1 10 0 06 1 10 0 06 1 10 0 06 0 100 0 006 5 Hz 10 Hz 0 35 0 06 0 35 0 06 0 35 0 06 0 035 0 006 10 Hz 5 kHz 0 15 0
2. get full documentation a high quality test lead set calibration certificate with test data and a l year warranty all for one low price ea Agilent Technologies Accuracy Specifications of reading of range Function Range Frequency 24 Hour2 90 Day 1 Year Temperature Coefficient etc 23 C 1 C 23 C 5 C 23 C 5 C 0 C to 18 C 28 C to 55 C dc voltage 100 0000 mV 0 0030 0 0030 0 0040 0 0035 0 0050 0 0035 0 0005 0 0005 1 000000 V 0 0020 0 0006 0 0030 0 0007 0 0040 0 0007 0 0005 0 0001 10 00000 V 0 0015 0 0004 0 0020 0 0005 0 0035 0 0005 0 0005 0 0001 100 0000 V 0 0020 0 0006 0 0035 0 0006 0 0045 0 0006 0 0005 0 0001 1000 000 V 0 0020 0 0006 0 0035 0 0010 0 0045 0 0010 0 0005 0 0001 True rms 100 0000 mV 3 Hz 5 Hz 1 00 0 03 1 00 0 04 1 00 0 04 0 100 0 004 ac voltage 5 Hz 10 Hz 0 35 0 03 0 35 0 04 0 35 0 04 0 035 0 004 10 Hz 20 kHz 0 04 0 03 0 05 0 04 0 06 0 04 0 005 0 004 20 kHz 50 kHz 0 10 0 05 0 11 0 05 0 12 0 04 0 011 0 005 50 kHz 100 kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100 kHz 300 kHz6 4 00 0 50 4 00 0 50 4 00 0 50 0 20 0 02 1 000000 V 3 Hz 5 Hz 1 00 0 02 1 00 0 03 1 00 0 03 0 100 0 003 to 750 000 V 5 Hz 10 Hz 0 35 0 02 0 35 0 03 0 35 0 03 0 035 0 003 10 Hz 20 kHz 0 04 0 02 0 05 0 03 0 06 0 03 0 005 0 003 20 kHz 50 kHz 0 10 0 04 0 11
3. software PC standard 1 0 and global support which are combined to more easily integrate test system development LXI www lxistandard org LXI is the LAN based successor to GPIB providing faster more efficient connectivity Agilent is a founding member of the LXI consortium www agilent com For more information on Agilent Technologies products applications or services please contact your local Agilent office The complete list is available at www agilent com find contactus Americas Canada Latin America United States 877 894 4414 305 269 7500 800 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 81 426 56 7832 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Thailand 1 800 226 008 Europe Austria 0820 87 44 11 Belgium 32 0 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 0 10 855 2100 France 0825 010 700 Germany 01805 24 6333 0 14 minute Ireland 1890 924 204 Italy 39 02 92 60 8484 Netherlands 31 0 20 547 2111 Spain 34 91 631 3300 Sweden 0200 88 22 55 Switzerland French Switzerland German United Kingdom 41 21 8113811 Opt 2 0800 80 53 53 Opt 1 44 0 118 9276201 Other European Countries www agilent com find contactus Revised May 7 2007 Product specifications and descriptions in this document subject to change without notice Agilent Technol
4. 06 0 15 0 06 0 15 0 06 0 015 0 006 Frequency 100 mV to 3 Hz 5 Hz 0 10 0 10 0 10 0 005 or period 750 V 5 Hz 10 Hz 0 05 0 05 0 05 0 005 10 Hz 40 Hz 0 03 0 03 0 03 0 001 40 Hz 300 kHz 0 006 0 01 0 01 0 001 Continuity 1000 0 Q 1 mA test current 0 002 0 030 0 008 0 030 0 010 0 030 0 001 0 002 Diode test 1 0000 V 1 mA test current 0 002 0 010 0 008 0 020 0 010 0 020 0 001 0 002 1 Specifications are for 1 hr warm up and 61 digits slow ac filter 2 Relative to calibration standards 3 20 over range on all ranges except 1000 Vdc and 750 Vac ranges 4 For sinewave input gt 5 of range For inputs from 1 to 5 of range and lt 50 kHz add 0 1 of range additional error 5 750 V range limited to 100 kHz or 8 x 107 Volt Hz 8 Typically 30 of reading error at 1 MHz 7 Specifications are for 4 wire ohms function or 2 wire ohms using Math Null Without Math Null add 0 2 Q additional error in 2 wire ohms function i 254 4 mm gt lt 374 0 mm 8 Input gt 100 mV For 10 mV to 100 mV inputs multiply of reading error x10 9 Accuracy specifications are for the voltage measured at the input terminals only 1 mA test current is typical Variation in the current source will create some variation in the voltage drop across a diode junction a o 200022220 Sobecue m 212 6mm i f 0 88 5 mm o
5. e Measure up to 1000 volts with 61 digits resolution e 0 0015 basic dcV accuracy 24 hour e 0 06 basic acV accuracy 1 year e 3 Hz to 300 kHz ac bandwidth e 1000 readings s direct to GPIB Superior Performance The Agilent Technologies 34401A multimeter gives you the perfor mance you need for fast accurate bench and system testing The 34401A provides a combination of resolution accuracy and speed that rivals DMMs costing many times more 61 4 digits of resolution 0 0015 basic 24 hr dcV accuracy and 1 000 readings s direct to GPIB assure you of results that are accurate fast and repeatable Use It on Your Benchtop The 34401A was designed with your bench needs in mind Functions commonly associated with bench operation like continu ity and diode test are built in A Null feature allows you to remove lead resistance and other fixed off sets in your measurements Other capabilities like min max avg Agilent 34401A Multimeter Uncompromising Performance for Benchtop and System Testing Product Overview readouts and direct dB and dBm measurements make checkout with the 34401A faster and easier The 34401A gives you the ability to store up to 512 readings in internal memory For trouble shooting a reading hold feature lets you concentrate on placing your test leads without having to constantly glance at the display Use It for Systems Testing For systems use the 34401A gives you faster bus through
6. er Delay lt 1 ms External Trigger Jitter lt 500 us Memory 512 readings Math Functions NULL min max average dBm dB limit test with TTL output Standard Programming Languages SCPI IEEE 488 2 Agilent 3478A Fluke 8840A 42A Accessories Included Test lead kit with probe alligator and grabber attachments Operating manual service manual test report and power cord General Specifications Power Supply 100 V 120 V 220 V 240 V 10 Power Line Frequency 45 Hz to 66 Hz and 360 Hz to 440 Hz Automatically sensed at power on Power Consumption 25 VA peak 10 W average Operating Environment Full accuracy for 0 C to 55 C Full accuracy to 80 R H at 40 C Storage Temperature 40 C to 70 C Weight 3 6 kg 8 0 Ibs Safety Designed to CSA UL 1244 IEC 348 RFI and ESD MIL 461C FTZ 1046 FCC Vibration amp Shock MIL T 28800E Type III Class 5 sine only Warranty 1 year 1 For 1 KQ unbalanced in LO lead 500 V peak maximum 2 For power line frequency 0 1 3 For power line frequency 0 1 use 40 dB or 3 use 30 dB 4 Reading speeds for 60 Hz and 50 Hz operation 5 Maximum useful limit with default settling delays defeated 8 Speeds are for 41 digits delay 0 auto zero and display OFF Ordering Information Agilent 34410A multimeter accessories included Test lead kit with probe alligator and grabber attachments operating manual service manual calibra
7. led to the fuse and shunt ac coupled true rms measurement measures the ac component only Shunt Resistance 0 1 Q for 1 A and 3 A ranges Input Protection Externally accessible 3 A 250 V fuse Internal 7 A 250 V fuse Frequency and Period Measurement Method Reciprocal counting technique Voltage Ranges Same as ac voltage function Gate Time 1 s 100 ms or 10 ms Continuity Diode Response Time 300 samples s with audible tone Continuity Threshold Selectable from 1 Q to 1000 Q Measurement Noise Rejection 60 50 Hz dc CMRR 140 dB ac CMRR 70 dB Integration Time and Normal Mode Rejection 100 pic 1 67 s 2 s 60 dB 10 plc 167 ms 200 ms 60 dB 1 plc 16 7 ms 20 ms 60 dB lt 1 plc 3 ms or 800 ps 0 dB Operating Characteristics Function Digits Reading s dcV del and 61 0 6 0 5 Resistance 61 2 6 5 51 60 50 51 2 300 41 1000 acV acl 61 0 15 slow 3 Hz 61 1 medium 20 Hz 61 10 fast 200 Hz 61 50 Frequency 61 1 or Period 514 9 8 4h 80 Frequency and Period Configuration rates 26 s to 50 s Autorange rate dc Volts gt 30 s ASCII readings to RS 232 55 s ASCII readings to RS 232 1000 s Maximum internal trig rate 1000 s Max ext trig rate to memory 1000 s Reading HOLD Sensitivity 10 1 0 1 or 0 01 of range Samples Trigger 1 to 50 000 Trigger Delay 0 to 3600 s 10 us step size External Trigg
8. lo 348 3 mm Measurement Characteristics dc Voltage True RMS ac Current Triggering and Memory Measurement Method Continuously integrating multi slope III A D converter A D Linearity 0 0002 of reading 0 0001 of range Input Resistance 10 MQ or 0 1 V 1 V 10 V ranges Selectable gt 10 000 MQ 100 V 1000 V ranges 10 MQ 1 Input Bias Current lt 30 pA at 25 C Input Protection 1000 V all ranges dcV dcV ratio accuracy Vinput Accuracy Vrelevance Accuracy True RMS ac Voltage Measurement Method AC coupled true rms measures the ac component of the input with up to 400 Vdc of bias on any range Crest Factor Maximum of 5 1 at full scale Additional Crest Factor errors non sinewave Crest factor 1 2 0 05 of reading Crest factor 2 3 0 15 of reading Crest factor 3 4 0 30 of reading Crest factor 4 5 0 40 of reading Input Impedance 1 MQ 2 in parallel with 100 pF Input Protection 750 Vrms all ranges Resistance Measurement Method Selectable 4 wire or 2 wire Ohms Current source referenced to LO input Maximum Lead Resistance 4 wire 10 of range per lead for 100 Q 1 kQ ranges 1 kQ per lead on all other ranges Input Protection 1000 V all ranges dc Current Shunt Resistance 5 Q for 10 mA 100 mA 0 1 Qfor1A 3A Input Protection Externally accessible 3 A 250 V fuse Internal 7 A 250 V fuse Measurement Method Directly coup
9. ogies Inc 2007 Printed in USA August 28 2007 5968 0162EN Ee Agilent Technologies
10. put than any other DMM in its class The 34401A can send up to 1 000 readings s directly across GPIB in user friendly ASCII format You also get both GPIB and RS 232 interfaces as standard features Voltmeter Complete and External Trigger signals are provided so you can synchronize to other instruments in your test system In addition a TTL output indicates Pass Fail results when limit testing is used To ensure both forward and back ward compatibility the 34401A includes three command languages SCPI Agilent 3478A and Fluke 8840A 42A so you don t have to rewrite your existing test software An optional rack mount kit is available Easy to Use Commonly accessed attributes such as functions ranges and resolution are selected with a single button press Advanced features are available using menu functions that let you optimize the 34401A for your applications The included Agilent IntuiLink software allows you to put your captured data to work easily using PC applications such as Microsoft Excel or Word to analyze inter pret display print and document the data you get from the 34401A You can specify the meter setup and take a single reading or log data to the Excel spreadsheet in specified time intervals Program mers can use ActiveX components to control the DMM using SCPI commands To find out more about IntuiLink visit www agilent com find intuilink 1 Year Warranty With your 34401A you
11. tion certificate test report and power cord Options 34401A 1CM Rack mount kit P N 5063 9240 34401A 0B0 DMM without manuals 34401A A6J ANSI Z540 compliant calibration Manual Options Please specify one 34401A ABA US English 34401A ABD German 34401A ABE 34401A ABF 34401A ABJ 34401A ABZ 34401A AB0 34401A AB1 34401A AB2 34401A AKT Spanish French Japanese Italian Taiwan Chinese Korean Chinese Russian Agilent Accessories 11059A 11060A Kelvin probe set Surface mount device SMD test probes Kelvin clip set Hard transit case Accessory pouch Input terminal connector sold in pairs Input calibration short sold in pairs 30 A current shunt 5 k thermistor probe 11062A 34131 34161A 34171B 34172B 34330A E2308A For racking two side by side order both items below Lock link kit P N 5061 9694 Flange kit P N 5063 9212 al Agilent Email Updates www agilent com find emailupdates Get the latest information on the products and applications you select Agilent Direct www agilent com find agilentdirect Quickly choose and use your test equipment solutions with confidence Agilent f Open aa www agilent com find open Agilent Open simplifies the process of connecting and programming test systems to help engineers design validate and manufacture electronic products Agilent offers open connectivity for a broad range of system ready instruments open industry