Home

CELL-DYN 1800

image

Contents

1. 07H80 01 Operator s Manual 1 07H79 01 Interface Specifications 1 20821 01 Printer 120V 1 20822 01 Printer 220V 1 03H56 01 Keyboard 1 92274 01 Aperture Plate WBC 100 um 92264 01 Aperture Plate RBC 60 um 28561 01 Lyse Syringe 2 5 mL 04H36 01 Diluent Syringe 10 mL 28514 01 Sample Syringe 100 uL 21704 01 Waste Dummy Plug l 93164 01 Sample Probe 1 54305 01 Aperture Brush 1 03H92 01 Detergent Line Inlet Assy l 20005 01 Printer Cable 1 91012 01 Teal Line Filter Isolation 1 07H81 01 Bar Code Scanner 1 03H96 01 Pull Ring Solenoid 1 99650 01 Code 39 Bar Code Labels 1000 pkg 07H67 01 Lyse Cap for 960 mL container 5 07H67 02 Lyse Cap for 3 8 L or 20 L container 5 For Abbott Internal Use Only Not for Customer Distribution Part 96919 101 CELL DYN 1800 Field Service Training Workbook Part 96919 101 Parts and Accessories CELL DYN 1800 Accessory Kit L N 03H54 01 1403204 Keyboard Cover 1 5100161 il Fuse SB 2 5 amps 250 V 2 60160 01 Fuse SB 5 0 amps 220 240 V 2 93501 01 Power Cord 1 5406753 Allen Wrench 3 32 1 5406754 Allen Wrench 7 64 1 54305 01 Aperture Brush 1 91072 01 Reagent Line Kit 1 20005 01 Printer Cable 1 9150143 Instructions for Mixing and Handling 1 93476 01 Silicon Tubing S2 1 24 a For CELL DYN 1700 System use only CELL DYN Reagent Line Kit L N 91072 01 03H82 01 Reagent Line Assembl
2. DOB 4 4 DIF PRESS HELP ERROR KEY Sequence 15 Limit 1 190 288 388 488 5H 18H 15H zHH 25H 235 16 15 z Z5 34 35 CLEAR PRE SPECIMEN PARAMETER PRINT HELP MAIN ORIFICE DILUTE TYPE SELECT REPORT ERROR RUN PRIME Menu Differences e Upto 16 alphanumeric characters can be entered into the Specimen ID field For Abbott Internal Use Only 17 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 DATA LOG MENU DATA LOG Dec B amp 2a2 1 13 pn Operator ID DIF PRESS HELP ERROR KEY Sequence 276 Specimen ID HEC Date BACKGRD 14 19 82 BACKGRD 1421 82 BACKGROD BACKGRD BACKGRO BACKGRD for X B 1234567898123456 34553788 5hkg R jhguyt97 1arz3Bz 1H 23 Hz7 1423 82 18 23 82 co co m co Co C C CO m 21 0C m tmo ON c c Cc m c cm Gm c cl lc t uriurutulolse Seg Specimen ID EDIT DISPLAY FIND REJECT TRANSMIT PRINT HELP MAIN In SPECIMEN SPECIMEN FROM X B DATA DATALOG ERROR DATA LOG Menu Differences e The Data Log stores the last 10 000 run cycles including all patient demographics numeric data and histograms e The mode letter C indicating the closed mode has been removed since this mode does not exist in the CELL DYN 1800 For Abbott Internal Use Only 18 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CALIBRATION MENU CALIBRATION Sep H1 ZBBz 11 85 am Opera
3. C Both rear and front screws D Not attached to Right Side Cover An LCD type of screen is used on the CELL DYN 1800 True or False The floppy disk drive is located on the right side of the Bezel True or False Which printer format s are available on the CELL DYN 1800 A ESC P B PCL C None of the above D Both ESC P and PCL Where do you change the printer formats A SETUP menu B RUN menu C DIAGNOSTICS menu D None of the above The Specimen ID field on the CELL DYN 1800 supports up to 16 alphanumeric characters True or False How many run cycles can be stored in the CELL DYN 1800 Datalog A 2 500 B 5 000 C 7 500 D 10 000 For Abbott Internal Use Only 48 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 11 12 13 14 15 16 17 18 19 20 21 A CN Free Diff Lyse is used on the CELL DYN 1800 True or False WBC Differential Channel Size ranges and Region Alerts were modified on the CELL DYN 1800 True or False How many fans on the CELL DYN 1800 A 3 B 4 C 2 D 1 The 28vde Switching Power Supply fan is turned off when the internal temperature is between 50 C and 70 C True or False The two 2 instrument fans operate at half speed when the internal temperature is below 25 C True or False What voltage is used for the two 2 instrument cooling fans A 28vdc B 14vdc C 12vdc digital
4. PA 25 38 HELP RETURN SPECIMEN SPECIMEN DEMOGRAPH SPECIMEN REPORT ERROR 26 Part 96919 101 CELL DYN 1800 Field Service Training Workbook Part 96919 101 WBC Differential Channel Size Ranges The WBC differential on both the CELL DYN 1700 and the CELL DYN 1800 is determined by counting the number of cells that fall into three 3 general areas of the histogram representing three 3 WBC sub populations These three 3 areas are based on the size of cells as determined by impedance in their lysate modified form and correlate to Lymphocytes Mid sized Cells and Granulocytes The WBC differential channel size range differences between the CELL DYN 1700 and the CELL DYN 1800 are detailed in the table below Electronic Differential Thresholds Comparison Chart CELL DYN 1700 CELL DYN 1800 Cell Group Channels Volume Channels Volume GO 1 15 0 27 fL 1 11 0 20 fL Gl 16 19 28 34 fL 12 19 2 34 fL Lymphs G2 20 25 35 44 fL 20 30 35 53 fL G3 26 49 45 87 fL 31 42 54 74 fL G4 50 62 88 109 fL 43 55 75 97 fL G6 84 96 147 169 fL 70 89 122 156 fL Grans G7 97 109 170 192 fL 90 109 157 192 fL G8 110 228 193 401 fL 110 228 193 401 fL G9 229 253 402 445 fL 229 253 402 445 fL G10 254 255 446 449 fL 254 255 446 449 fL G11 256 450 fL 256 450 fL Fixed Groups and Thresholds The X axis of the WBC histogram for both the CELL
5. CPU Hardware Software Configuration Cell Count Module CCM Signal Processor Module SPM Card Cage Backplane Pre Amplifier Module PAM Power Supply Module Fans For Abbott Internal Use Only 29 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Learning Objectives Learning Objectives Section 4 will explain the hardware changes made to the CELL DYN 1800 After reviewing Section 4 of this workbook you should be able to do the following Describe the User Interface Computer along with the Single Board Computer SBC Describe the changes made to the Signal Processor Module SPM PCB List the adjustments on the SPM PCB and the order they should be adjusted in Describe the changes made to the Cell Count Module CCM PCB Describe the changes made to the Power Supply Assembly List all voltages generated by the Power Supply Assembly and their function List all fans that are used in the CELL DYN 1800 and their function Describe how the power supply and instrument cooling fans operate when the ambient temperature increases and decreases List the connections on the SBC PCB Describe the function of the Backplane PCB List the PCBs that connect to the Backplane PCB Describe the changes made to the PAM PCB Describe the changes made to the Metering PCB Describe the purpose of the LCD Interconnect PCB Describe the purpose of the Backlight Inverter PCB For Abbott Interna
6. Group 4 2 1 75 MID R3 Group 7 Group 6 2 2 20 GRANR3 Group 7 GRAN 0 061 R4 Group 10 Group 9 2 0 110 or Group 10 Group 11 Group 9 2 1 64 CELL DYN 1800 Region Alert Algorithms RO Group 1 2 300 LYMRI Group 2 LYM 2 0 35 LYM R2 Group 4 LYM 2 0 40 MID R2 Group 5 Group 4 2 1 3 MID R3 Group 7 Group 6 2 2 5 GRAN R3_ Group 7 GRAN 2 0 18 R4 Group 10 Group 9 0 14 or Group 10 Group 11 Group 9 2 1 8 For example refer to the LYM R1 alert in the CELL DYN 1800 Region Alert Algorithms table above Note that the algorithm is as follows Group 2 LYM 2 0 35 This means that the LYM R1 Alert will be set if the number of cells in Group 2 channels 20 through 30 divided by the number of cells in the LYM groups 2 3 and 4 channels 20 through 55 is equal to or greater than 0 35 Each of the remaining region alerts for both the CELL DYN 1700 and CELL DYN 1800 are determined in a similar manner Note Detailed information regarding Suspect Parameters or Suspect Populations may be found in the CELL DYN 1800 System Operator s Manual For Abbott Internal Use Only 28 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Section 4 Hardware Learning Objectives CELL DYN 1700 and CELL DYN 1800 Comparison Chart Hardware Overview of Hardware Changes User Interface Computer UIC
7. hard disk drive is the same between the CELL DYN 1700 and CELL DYN 1800 instruments Note Drive size will depend on availability Floppy Disk Drive The 1 44MB 3 5 floppy disk drive is the same between the CELL DYN 1700 and CELL DYN 1800 instruments Data Link Adapter DLA The Data Link Adapter serves much the same purpose in the CELL DYN 1800 as it did in the CELL DYN 1700 It provides interfaces from the Single Board Computer SBC to both the Cell Count Module CCM and the instrument s membrane keyboard These two interfaces function independently under application software However some of the circuits on the DLA PCB are shared LCD Display Screen A 10 4 inch color Liquid Crystal Display LCD screen replaces the 14 inch Cathode Ray Tube CRT used on the CELL DYN 1700 The LCD display screen is driven via a video adapter mounted directly on the SBC The For Abbott Internal Use Only 37 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 video adapter supports SVGA 640 x 480 and 800 x 600 graphics modes A Low Voltage Differential Signaling LVDS cable connects between the LCD display screen and the LCD Interconnect PCB to transmit imaging signals to the screen The LCD Display Screen uses a Backlight Inverter PCB to illuminate the panel and project the image on the screen The SBC provides 12vdc digital to the LCD Interconnect PCB which is then stepped up by the Backli
8. sc z o L T B t 7 zc a Floppy Keyboard COM2 Disk Drive COMI LPTI For Abbott Internal Use Only 9 Not for Customer Distribution vice Training Workbook CELL DYN 1800 Rear View UI UU LA TM ML Drives CELL DYN 1800 Field Service Training Workbook Physical Specifications Physical Dimensions Dimension Instrument Height 17 5 8 45 cm Width 26 66 cm Depth 21 53 cm Weight 122 Ibs 55 kg Dimensions after Packaging for Shipment Dimension Instrument Height 32 81 cm Width 38 1 2 98 cm Depth 28 1 2 72 cm Weight 230 Ibs 104 kg Part 96919 101 Note For complete listing of instrument specifications refer to CELL DYN 1800 System Operator s Manual Section 4 Performance Characteristics and Specifications For Abbott Internal Use Only Not for Customer Distribution 11 CELL DYN 1800 Field Service Training Workbook Part 96919 101 Section 2 User Interface e Learning Objectives e CELL DYN 1700 and CELL DYN 1800 Comparison Chart User Interface e Operator Screens For Abbott Internal Use Only 12 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Learning Objectives Learning Objectives Section 2 will explain the user interface changes made to the CELL DYN 1800 After reviewing Se
9. 2x 50 mL 30005 01 Okidata Graphics Paper 3000 sheets pkg 13401 01 Ribbon Okidata 320 I 03H10 03 HP Inkjet Printer 1 30010 06 Printer Paper 8 x 11 93601 14 Printer Stand Okidata 320 03H10 04 HP Printer Cartridge Black 30010 08 HP Printer Paper 99605 01 CELL DYN Counting Cups 500 pkg For Abbott Internal Use Only 55 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook CELL DYN 1700 and CELL DYN 1800 Comparison Chart Name Covers Upper Front Lower Front Bezel Top Left Right System Performance Parameters WEC Differential Histograms Technology Closed Tube Sampler Sample Volume Open Closed Pre Dilute Reagents CELL DYN 1700 1700CS CELL DYN 1800 Yes Different hinged cover that opens to the left and contains Status Indicator PCB Yes Different secured to bezel Yes Different contains LCD display screen Touch Pad membrane keyboard and floppy disk drive Yes Different dimensions Yes Different dimensions Yes Different contains cut out to connect PC 2 keyboard LIS and printer 18 PCT and PDW not reportable in 18 PCT and PDW not reportable in US 3 Part Diff Lym Mid Gran Same WBC RBC PLT display amp Same WBC RBC PLT Impedance Same HGB Modified Cyanmethemoglobin Modified Methemoglobin Method Method cyanide free T iem Hd gt o c c No Yes Z o Same N A Same Lytic Agent Diff Scre
10. CB automatically accommodates line voltages of 90 130VAC and 200 260V AC by sensing the input voltage and utilizing an internal comparator bank and power transformer to produce the 120V AC necessary for the subsystem s function The Power Supply Assembly then provides an AC and DC voltage source to various components on the CELL DYN 1800 System The Power Supply Assembly used on the CELL DYN 1700 is not compatible with the CELL DYN 1800 The 24v and PC Power supplies have been eliminated and those supplied voltages moved to the CELL DYN 1800 instrument s Power Supply Assembly For Abbott Internal Use Only 42 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 AC Regulator 28vdc Switching Power Supply Power DC Transformer Regulator PCB 28 volt Switching Power Supply e Supplies 28 volts to the DC Regulator PCB AC Regulator PCB e Supplies 12vdc Analog to the Backplane PCB for use on SPM and CCM PCBs These voltages replace the 15vdc found on the CELL DYN 1700 e Supplies 12vdc Analog to the Pre Amplifier Module PAM via the Pre Amp Filter Assembly Cable Distribution Module and Motor Processor Module PCBs e Supplies 100 volts to the Pre Amplifier Module PAM for the constant current supply circuit to the transducers e Supplies 120vac to Pump Relay Module PRM e Supplies input voltage wall voltage to the power transformer In return the powe
11. CELL DYN 1800 Field Service Training Workbook Abbott Laboratories 2003 Abbott Laboratories Abbott Park IL 60064 96919 101 04 09 03 CELL DYN 1800 Field Service Training Workbook Part 96919 101 Proprietary Statement Copyright 2003 Abbott Laboratories Diagnostics Division All rights reserved Printed in the United States of America No part of this document may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopied recorded or otherwise without the prior written permission of Abbott Laboratories Pictorial Disclaimer Sample graphic illustrations contained in this manual are for information and illustration purposes only Abbott Laboratories makes no representations or warranties about the accuracy and reliability of information in the graphics This information should not be used for clinical or maintenance evaluation Field Service Training Workbook Usage Disclaimer Information contained in this document was developed for use by trained Abbott Laboratories Field Service personnel or by other persons knowledgeable or experienced with operation and service of the CELL DYN Analyzer The information should be used under direct supervision and cooperation with Abbott Laboratories technical sales and service representatives In no event shall Abbott Laboratories or its affiliates be liable for any damages or losses incurred in connection with or arising fr
12. D 12vdc analog There are two 2 fan filters located at the rear of the CELL DYN 1800 True or False The SBC PCB on the CELL DYN 1800 replaced the CPU Assembly on the CELL DYN 1700 True or False What connects to the SBC PCB PS 2 101 key Keyboard Input Output Ports LCD Display Screen Hard and Floppy Disk Drives All of the above monos If the LCD Display goes blank an SVGA monitor can be connected to the top of the SBC PCB to allow for CMOS Setup verification in order to troubleshoot the problem True or False The DCM PCB on the CELL DYN 1700 was incorporated into which PCB on the CELL DYN 1800 A SPM B MAM C CCM D DLA For Abbott Internal Use Only 49 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 22 The MAM PCB on the CELL DYN 1700 was incorporated into which PCB on the CELL DYN 1800 A SPM B MAM C CCM D DLA 23 The following adjustments are made on the CELL DYN 1800 SPM PCB RBC gain RBC Editing Ratio WBC gain and PLT gain True or False 24 Match the order in which the adjustments are made on the SPM PCB I e A RBC Editing Ratio Dert a B PLT gain I RN C WBC gain 4 n D RBC gain 25 What is the purpose of the SPM PCB on the CELL DYN 1800 Receives WBC and RBC PLT signal from PAM PCB Fixed amplification gain of signals Applies thresholds to signals Contains cell editing circuitry for RBC All of the abov
13. DYN 1700 and CELL DYN 1800 Systems depicts a size range of uP to 450 fL divided into 256 channels These 256 channels are broken into 12 groups labeled GO through G11 e Each group encompasses a range of cell sizes measured in femtoliters e Each channel represents 1 76 fL and therefore a 5 0 micron diameter polystyrene microsphere will fall into channel 56 on both systems Referring to the table above compare the range of volumes to the range of channels within each group for both instruments Note that not only does the range of volumes for each group differ between the two systems but the channels have also changed For Abbott Internal Use Only 27 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Region Alerts Part 96919 101 In addition to the WBC histogram algorithm and differential channel size range enhancements the region alerts were reassessed and optimized The twelve groups of channels discussed in the Fixed Groups and Fixed Threshold section of this training workbook are also used to determine the criteria for the region alerts Region alerts are set by mathematically comparing the number of cells contained in a certain group to the number of cells in one or more other groups as determined by the Region Alert Algorithms CELL DYN 1700 Region Alert Algorithms RO Group 1 2 135 LYMRI Group 2 LYM 2 0 20 LYM R2 Group 4 LYM 2 0 192 MID R2 Group 5
14. ELL DYN 1700 and CELL DYN 1800 Comparison Chart Covers Name CELL DYN 1700 1700CS Upper Front Yes Bezel Yes Left Yes Right Yes CELL DYN 1800 Different hinged cover that opens to the left and contains Status Indicator PCB Different secured to bezel Different contains LCD display screen Touch Pad membrane keyboard and floppy disk drive Different dimensions Different dimensions Different contains cut out to connect PC 2 keyboard LIS and printer For Abbott Internal Use Only Not for Customer Distribution l CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1800 Front View Hinged Status Upper Bezel Indicator Front LCD Lov LEDs Cover Display l Screen i fo Bo Ho J CELL DYN 1800 alli A SSS ft SS lt Lower Front Cover For Abbott Internal Use Only 8 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1800 Right Side View Power Switch imm TIR RU MM a OU MM I LLL o LLL eC
15. N 1700 the CELL DYN 1800 WBC algorithms which include differential channel size ranges and region alerts have been revised to enhance the System s performance Refer to the histogram s on the following page Both instruments display cell size range to 450 femtoliters In addition the CELL DYN 1700 and CELL DYN 1800 begin to count cells for the differential in channel 20 e The CELL DYN 1700 begins to plot the histogram from channel 16 e The CELL DYN 1800 begins to plot the histogram at channel 12 For Abbott Internal Use Only 25 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Spec HGB HCT MCU MCH CHC RDU PLT MPU PCT PDU PRE SPE HGB HCT MCU MCH CHC RDH PLT MPU PCT PDH PREVIOUS NEXT EDIT TRANSMIT PRINT CELL DYN 1700 Normal Histogram Example imen ID 24602 DISPLAY SPECIMEN DATUR g dL p fL pg g dL 435 Kul 7 0 fL 16 1 10 GSD Mar 31 2003 4 33 pn Operator ID 00 Sequence 2125 open Limit 1 0 30 25 10 15 20 2 30 VIOUS NEXT EDIT TRANSMIT PRINT PRINT HELP RETURN CIMEN SPECIMEN DEMOGRAPH SPECIMEN TICKET REPORT ERROR CELL DYN 1800 Normal Histogram Example DISPLAY SPECIMEN DIF PRESS HELP ERROR KEY M uL g dL fL pg g dL K uL fL 18CGSD For Abbott Internal Use Only Not for Customer Distribution x 2 5 1418 15 Jan 28 2683 1 36 pm Operator ID 882 Sequence 4236 Limit 1
16. Same WBC RBC PLT display amp Same WBC RBC PLT Impedance Same HGB Modified Cyanmethemoglobin Modified Methemoglobin Method Method cyanide free Flo 5 a d Z am T I we TF lt S 9 4 Z o Open Same Closed N A Pre Dilute 40 uL Same Reagents Lytic Agent Diff Screen Cyanide Free CN Free Diff Lyse Diluent Diff Screen Same Detergent Diff Screen Same For Abbott Internal Use Only 24 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Overview of Changes to WBC Differential Minor changes were made to the CELL DYN 1800 WBC differential e The conversion from lyse containing Potassium Cyanide to a CN Free Diff Lyse reagent e Changes made to the WBC algorithms and region alerts to accommodate the CN Free Diff Lyse reagent Cyanide Free Diff Lyse CN Free Diff Lyse Lytic Agent CELL DYN 1700 vs Cyanide Free Diff Lyse CELL DYN 1800 e Reduces the hazardous element of the lyse reagent e Benefits customers when complying with local environmental regulations The CN Free Diff Lyse converts hemoglobin to a modified hemoglobin complex that is measurable at 540 nanometers the quaternary ammonium lysate participates to form a chromagen for hemoglobin measurement Differential Algorithms Although the technology and methodology used to produce the WBC count differential and histogram in the CELL DYN 1800 is generally the same as in the CELL DY
17. YN 1800 include the following e Receives two 2 signals from the PAM Preamp Module WBC white blood cell and RBC PLT red blood cell platelet Two distinct signal paths handle the two signals e Processes WBC data by converting validated signal heights to a digital value Signals too small or too large are excluded according to the thresholds set within the SPM e Splits the RBC signal received from the PAM into two 2 data acquisition paths The two paths include the Platelet PLT data and RBC data paths Due to the size differentiation between the two signals internal thresholds separate the RBC s from the PLT s e Institutes the cell editing circuitry on the RBC data acquisition path During this function the area of the pulse is compared to the pulse height Pulses which have too much area compared to the pulse height are excluded Note The CELL DYN 1800 SPM PCB does not require channel offset and threshold adjustments However the RBC gain RBC Editing Ratio WBC and PLT gains can and need to be adjusted in that order Refer to the CELL DYN 1800 System Service Manual Section 5 VP 09 Signal Processor Module SPM Verification Adjustment For Abbott Internal Use Only 40 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Card Cage Backplane The Card Cage Backplane PCB referred to as the Motherboard Module on the CELL DYN 1700 has been totally redesigned to accommodate t
18. ans located in the CELL DYN 1800 System two 2 fans are used for instrument cooling one 1 fan is used to cool the microprocessor on the SBC PCB and one 1 fan is used to cool the 28vdc Switching Power Supply Instrument Cooling Fans The left side exhaust fan blows out located above the Sample Syringe Assembly functions similarly to the CELL DYN 1700 instrument In addition an intake fan was placed on the sidewall between the Fluid Power Supply compartment and the right side compartment behind the Card Cage Assembly The orientation of the two 2 fans allows air to enter through the right side panel drawing air through the Fluid Power Supply and backside of the flow panel and then exits out through the left side panel Refer to figure on next page The two 2 rear panel fans found on the CELL DYN 1700 were removed on the CELL DYN 1800 Note The two instrument cooling fans are thermistor controlled which means that when the ambient temperature inside the instrument drops below 25 C the fans will operate at half speed Once the temperature rises above 25 C the speed is increased linearly until it reaches 35 C when the fans operate at full speed 28vdc Switching Power Supply Fan The fan is mounted inside the power supply s casing and can be viewed from the top of the unit Note The power supply s cooling fan is thermistor controlled which means that when the internal temperature rises above 70 C the fan operates at ful
19. ction 2 of this workbook you should be able to do the following List all of the changes made to the User Interface Software Identify the printer formats available on the CELL DYN 1800 Describe where to modify the printer formats Describe the change made to the Specimen ID field in the Run Screen Describe how many run cycles can be stored in the CELL DYN 1800 Data Log For Abbott Internal Use Only 13 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1700 and CELL DYN 1800 Comparison Chart User Interface Name User Interface Quality Control Files Replicate Files QC Upload Download Data Log Storage Data Log Storage Type RS232 Interface Computer Interface Configuration Report Demographics Patient Limit Sets Reference Range Display Limit Sets Patient Limit Flag L H Panic Limit Flags LL HH WBC Histogram Range Customized Header Optional Manual Differential Grid CELL DYN 1700 1700CS CELL DYN 1800 Twelve 12 QC Files four 4 each Same for L N H controls Nine 9 Replicate Files 120 data Same points per file 10 000 Run Cycles Done in software Same SETUP menu COMPUTER SETUP Specimen ID nine 9 numeric Specimen ID 16 alphanumeric digits characters Yes Screen Printout Same 0 to 450 fL Same Yes For Alerted and Non Alerted Same Specimens On Screen Maintenance Y
20. e mons 26 What is the purpose of the CCM PCB on the CELL DYN 1800 A Flow System Control and functions B Measurement and analog control C Interfacing with SBC PCB D Receives data from SPM PCB for CCM Histogram Memory E All of the above 27 What are the two 2 PCBs on the power supply assembly A DC Regulator PCB B AC Regulator PCB C Both A and B D None of the above 28 The LEDs on the Cardcage Backplane PCB indicate system status True or False 29 Which PCB s connect into the Cardcage Backplane PCB MPM CDM SBC and DLA SPM CCM DLA and SBC MAM SPM DCM and CCM MPM MAM CDM and CCM onw For Abbott Internal Use Only 50 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 30 The AC Regulator PCB automatically senses the input voltage and produces the 120vac necessary for the system s function True or False 3 Which voltages does the AC Regulator PCB supply 12vdce analog and 120vac 12vdc analog 12vdce digital and 120vac 5vdc 12vdce digital 28vdc and 14vdc None of the above onw 32 Which voltages does the DC Regulator PCB supply A H 12vdc analog and 120vac B 4 12vdc analog 12vdc digital and 120vac C 5vdc 12vde digital 28vdc and 14vdc D None of the above 33 What is the voltage supplied by the 28 volt switching power supply used for A Instrument cooling fans B DC mo
21. en Cyanide Free CN Free Diff Lyse 30 uL 450 uL 40 uL Diluent Diff Screen Same Detergent Diff Screen Same For Abbott Internal Use Only 56 Not for Customer Distribution Part 96919 101 CELL DYN 1800 Field Service Training Workbook Name Hardware Computer Hardware Monitor Display Monitor Controls Power Switch Location Serial Number Location Lyse Delivery Closed Sampler Solenoids CELL DYN 1700 1700CS 386 or 486 Microprocessor depends Celeron 850Mhz Microprocessor on when the instrument was built CELL DYN 1800 All I O video floppy hard drive control circuitry is contained on a single board connected to the cardcage 3 Floppy Drive front panel below Same right side on bezel R 14 Color SVGA Monitor Color LCD Display Screen 10 4 inch Brightness Contrast Knobs One 1 brightness control knob located on left side of bezel 7 dial version 2 dials on right N A 5 dial version 2 dials on left N A Same Same Lyse Syringe Same N A AI Z 7 A gq ga sis ae oes e e a o amp 5 5 eig Fans 4 two 2 instrument thermistor controlled cooling fans one 1 thermistor controlled fan on 28v switching power supply and one 1 CPU fan DLA PCB Yes Same CCM PCB Yes Different PCB combined CCM and DCM circuitry DCM PCB Yes No combined onto CCM SPM PCB Yes Different PCB combined SPM and MAM circuitry MAM PCB Yes No combined ont
22. es Graphic Color Directions Same Procedures For Abbott Internal Use Only 14 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Accessories Handheld Barcode Reader Yes optional Printer Okidata 320 dot matrix printer Okidata 320 dot matrix printer and HP Inkjet Printer Printer formats ESC P and PCL3 selectable in SETUP menu model depends on market availability For Abbott Internal Use Only 15 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Operator Screens SETUP MENU SETUP Sep H1 ZBHHZ 18 15 an Operator ID 123 DIF PRESS HELP ERROR KEY Sequence 15 X B Moving Average Program Automatic Increment nf Specimen I I Humber Print Histograms Print PCT PIH Print ALERTED LYM L MID M GRAN G Results Print ALERTED PLT Results Automatic Graphics Printout Print Manual Differential Grid for ALERTED specimens Pu 1 mfferemtia pid for H N RLERTED specimens Please enter Print current Date Time and Software Version YN DATE PATIENT REAGENT OC COMPUTER UNITS HELP MAIN TIME LIMITS LOG SETUP SETUP SELECTION ERROR SETUP Menu Differences e Printer Type 1 ESC P 2 PCL For Abbott Internal Use Only 16 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 RUN PRIME MENU RUN Sep H3 ZBHz 2 56 pm Ready Operator ID 123 Sex M F
23. ganizations for certification Revision History Pages Revised and Added 96919 101 4 2003 First Release For Abbott Internal Use Only 3 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1800 FSR Training Workbook Index Section 1 Overview Section 2 User Interface Section 3 WBC Differential Changes Section 4 Hardware Section 5 Appendix For Abbott Internal Use Only 4 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Section 1 Overview Learning Objectives CELL DYN 1700 and CELL DYN 1800 Comparison Chart Covers Front View Right Side View Rear View Left Side View Physical Specifications For Abbott Internal Use Only 5 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Learning Objectives Learning Objectives Section 1 will provide an overview of the System s covers and physical dimensions of the CELL DYN 1800 After reviewing Section 1 of this workbook you should be able to do the following e List the differences between the CELL DYN 1700 and CELL DYN 1800 covers e Identify the display screen that is used on the CELL DYN 1800 e Identify the connections located on the right side of the CELL DYN 1800 For Abbott Internal Use Only 6 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 C
24. ght Inverter PCB The backlight is controlled through the software and employs a screen saver that turns the screen on off Note Measuring voltages directly with a digital multi meter on the Backlight Inverter PCB may damage the board For Abbott Internal Use Only 38 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CPU Hardware Software Configuration CMOS Setup The CMOS Setup contains all the information needed by the BIOS system to establish proper communications between the Single Board Computer SBC and various computer hardware devices Note Refer to the CELL DYN 1800 System Service Manual Section 5 VP 05 CMOS Setup Verification Adjustment for instructions on accessing and modifying CMOS settings Troubleshooting In a case where the LCD Display Screen goes blank an SVGA CRT monitor can be connected to the external monitor connector 15 pin on the SBC PCB Refer to the figure in the Single Board Computer SBC description earlier in this section to locate the external monitor connector Once the instrument s image can be viewed refer to the CELL DYN 1800 System Service Manual Section 5 VP 05 CMOS Setup Verification Adjustment Verify the LCD CRT Selection as well as the LCD Type settings in the Advanced ChipSet Setup screen Itis possible that the settings may have been reset to default Follow the instructions in the verification procedure to establish the correct set
25. gital and 12v digital from the DC Regulator located on the Power Supply Assembly Power is supplied through the Card Cage Backplane PCB It receives status and measurement data from the CCM PCB via the DLA PCB The User Interface Computer consists of the following modules e Single Board Computer SBC e Hard Disk Drive e Floppy Disk Drive e Data Link Adapter e LCD Display Screen Single Board Computer SBC The Single Board Computer SBC is an Intel compatible Celeron 850 MHz Microprocessor It connects directly into the Card Cage Backplane PCB where a voltage source of 5v 12v digital and 12v digital is supplied for the internal circuits Note Processor speed will vary according to market availability Floppy Drive COMI COM2 External Monitor For Abbott Internal Use Only 36 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Several I O functions commonly found as individual PCBs on other platforms were consolidated into the SBC They are e Hard Disk Drive IDE Controller e Floppy Disk Drive Controller e Input Output Controller e Video In line connectors exist for the 101 key keyboard printer port LPT1 LIS RS 232 port COM1 Note The external monitor connector located on the SBC PCB refer to figure on previous page can be used to connect an SVGA CRT monitor in case of LCD Display Screen failure during troubleshooting Hard Disk Drive The
26. he SPM CCM DLA and SBC PCBs The board mounted at the rear of the card cage receives supply voltages from the AC and DC Regulator PCBs to power the various PCBs via the Backplane s internal bus The data communication between the PCBs in the card cage is performed through the different flat ribbon cables The Backplane PCB also receives measurement data from the PAM PCB A set of five 5 LEDs are mounted on the front side of the board to indicate 5vdc 12vdc analog and 12vdc digital are being supplied to the Backplane PCB For Abbott Internal Use Only A Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Pre Amplifier Module PAM The Pre Amplifier Module PAM functions similarly between the CELL DYN 1700 and CELL DYN 1800 instruments However the only change made was the addition of a high frequency filter circuit to the PAM This filter circuit will help eliminate high frequency noise that can interfere with PLT signal data PRE AMP ASSY 9602750 REV e Power Supply Module The Power Supply Assembly is comprised of four components AC Regulator PCB DC Regulator PCB 28VDC Switching Power Supply and Power Transformer These components are mounted together as an assembly and are located at the right rear side of the instrument When the system is turned on the AC line voltage is supplied to the AC Regulator PCB and 28VDC Switching Power Supply The AC Regulator P
27. ilter circuit was added to the Pre Amplifier Module PAM to eliminate high frequency noise that may interfere with Platelets For Abbott Internal Use Only 33 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Overhead View of Instrument Power Supply Assembly Card Cage Assembly LCD Display Screen Printed Circuit Boards The table below is a quick reference guide which lists PCB compatibility between the CELL DYN 1700 and CELL DYN 1800 Component CELL DYN CELL DYN Note 1700 1800 Backlight Inverter PCB N A X New Used for LCD Display Screen 8405627701 Cable Dist Module X X Interchangeable CDM Cell Count Module X X Not Interchangeable CCM Revised PCB has DCM circuitry incorporated 8960273001 Chopper Driver Module X X Interchangeable Data Link Adapter X X Interchangeable DLA Device Control Module X N A Eliminated see CCM DCM Indicator status PCB N A X New Located on Upper Front Cover 8960280001 In Line Reagent PCB X X Interchangeable For Abbott Internal Use Only 34 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Component CELL DYN CELL DYN Note 1700 1800 Main Amp Module X N A Eliminated see SPM MAM Metering Board PCB X X Not Interchangeable WBC and RBC Different ca
28. l Use Only 30 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1700 and CELL DYN 1800 Comparison Chart Hardware Name Hardware Computer Hardware Monitor Display Monitor Controls Power Switch Location Serial Number Location Lyse Delivery Closed Sampler Solenoids CELL DYN 1700 1700CS 386 or 486 Microprocessor depends Celeron 850Mhz Microprocessor on when the instrument was built CELL DYN 1800 All I O video floppy hard drive control circuitry is contained on a single board connected to the cardcage 3 Floppy Drive front panel below Same right side on bezel 14 Color SVGA Monitor Color LCD Display Screen 10 4 inch Brightness Contrast Knobs One 1 brightness control knob located on left side of bezel 7 dial version 2 dials on right N A 5 dial version 2 dials on left N A Right Panel Same Right Panel Same Lyse Syringe Same N A AJN Q Ww Jd s Fans 4 two 2 instrument thermistor controlled cooling fans one 1 thermistor controlled fan on 28v switching power supply and one 1 CPU fan DLA PCB Yes Same CCM PCB Yes Different PCB combined CCM and DCM circuitry DCM PCB Yes No combined onto CCM SPM PCB Yes Different PCB combined SPM and MAM circuitry PAM PCB Yes Different PCB new PCB includes a high frequency reduction filter PC Power Supply Yes 5v 12v and No 5v 12v digital 12
29. l speed The fan is then turned off at 50 C and kept off until the temperature rises above 70 C CPU Fan The CPU fan is used to cool the Celeron 850 MHz microprocessor and is mounted directly on the SBC PCB The 12vdc digital used to power the fan is supplied from the DC Regulator PCB through the Backplane PCB For Abbott Internal Use Only 45 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1800 System Air Flow Left Side Right Side Flow Panel Front For Abbott Internal Use Only 46 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Section 5 Appendix e CELL DYN 1800 Field Service Training Final Test e Parts amp Accessories e CELL DYN 1700 and CELL DYN 1800 Comparison Chart For Abbott Internal Use Only 47 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1800 Field Service Training Final Test 1 10 What PCB was added to the Upper Front Cover A Backlight Inverter PCB B LCD Interconnect PCB C Status Indicator PCB D None of the above The Upper Front Cover opens to the left True or False Which cover is the Lower Front Cover secured to A Upper Front Cover B Bezel C Right Side Cover D All of the above In order to remove the Bezel which screw on the Right Side Cover must be removed A Rear screw B Front screw
30. o SPM PAM PCB Yes Different PCB new PCB includes a high frequency reduction filter PC Power Supply Yes 45v 12v and No 5v 12v digital 12v digital 2v generated on DC Regulator PCB 24v Power Supply Yes No replaced by 28v switching power supply For Abbott Internal Use Only 57 Not for Customer Distribution Part 96919 101 CELL DYN 1800 Field Service Training Workbook Name Hardware Cont Linear Power Supply Pre Amp Filter Input Voltage Select Switches User Interface Quality Control Files Replicate Files QC Upload Download Data Log Storage Data Log Storage Type RS232 Interface Computer Interface Configuration Report Demographics Patient Limit Sets Reference Range Display Limit Sets Patient Limit Flag L H Panic Limit Flags LL HH CELL DYN 1700 1700CS Yes 15v 15v 12v unreg 24v unreg and 100v Yes found at rear of instrument on Linear Power Supply Twelve 12 QC Files four 4 each for L N H controls Nine 9 Replicate Files 120 data points per file o Yes 5 000 Run Cycles Numeric amp Graphic data Yes Done in software SETUP menu COMPUTER SETUP Yes Specimen ID nine 9 numeric Patient Name Male Female Date of Birth Physician Name Date Time Collected Comments 4 User Definable Yes Screen Printout Yes CELL DYN 1800 No Power Supply Assembly is made up of three 3 component
31. om use of information in this document by persons not fully trained by Abbott Laboratories as technical sales and service representatives This limitation shall not apply to those persons knowledgeable or experienced with operation and service of the CELL DYN Analyzer who use this information under direct supervision of Abbott Laboratories technical sales and service representatives Trademark Statements CELL DYN is a registered trademark of Abbott Laboratories For Abbott Internal Use Only 2 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Intended Use of Training Workbook The CELL DYN 1800 Field Service Training Workbook is recommended as a self paced training guide to familiarize currently trained CELL DYN 1700 Abbott Laboratories Field Service personnel on design enhancements made to the CELL DYN 1800 instrument The CELL DYN 1800 Field Service Training Workbook is not designed to replace any CELL DYN 1000 Series instrument training program s The prerequisite for taking this course is that the trainee must have completed a CELL DYN 1000 Series Field Service training class or equivalent at one of the Area Service and Support training centers worldwide It is the responsibility of the Area Service and Support Organizations to implement monitor and verify satisfactory completion of this self paced course A final test is provided in Section 5 and can be used by Area Service and Support Or
32. r transformer supplies l6vac and 120vac regardless of the input voltage DC Regulator PCB e Supplies 5vdc to the Card Cage Backplane PCB for use on SPM CCM DLA and SBC PCBs This voltage is also used by the CDM MPM Hard and Floppy Disk Drives e Supplies 12vdc digital is provided to the Backplane PCB for use on the CPU fan The Hard and Floppy Disk Drives receive 12vdc for their operation e Supplies 14vdc to Cable Distribution Module to be used by the Solenoid Driver Modules to maintain solenoids energized hold voltage For Abbott Internal Use Only 43 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 e Supplies 28vdc to Cable Distribution Module Motor Processor Module and instrument cooling fans The 28vdc is used by the Motor Processor Module to supply power to the DC motors via the Chopper Driver Modules Also 28vdc is supplied to the Cable Distribution Module and used by the Solenoid Driver Module to initially energize solenoids e Supplies 5vdc and 14vdc to the AC Regulator PCB Power Supply Assembly Diagram AC In AC DC 428vdc 28vdc Regulator Switching Regulator PCB Pwr Supply rt IH 12vdc A 120vac 5vde 28vde 5vdc 14vde 12vde D 14vdc For Abbott Internal Use Only 44 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Fans Overview There are four 4 f
33. s 28v switching power supply 428v DC Regulator PCB 5v 12v digital 12v digital 14v and AC Regulator PCB 12v analog 12v analog 100v Yes used to filter the 12v analog between the AC Regulator PCB and PAM PCB No AC Regulator PCB senses input voltage and adjusts automatically Same Same Yes 10 000 Run Cycles Same Yes Same Specimen ID 16 alphanumeric characters Same Same Same Same Same Same Same Same Same Same For Abbott Internal Use Only Not for Customer Distribution CA oo Part 96919 101 CELL DYN 1800 Field Service Training Workbook Name CELL DYN 1700 1700CS CELL DYN 1800 WBC Histogram Range 0 to 450 fL Same Optional Manual Yes For Alerted and Non Alerted Same Differential Grid Specimens On Screen Maintenance Yes Graphic Color Directions Same Procedures Accessories fo Handheld Barcode Reader Noes optional Printer Okidata 320 dot matrix printer Okidata 320 dot matrix printer and HP Inkjet Printer Printer formats ESC P ESC P and PCL3 selectable in SETUP menu model depends on market availability For Abbott Internal Use Only 59 Not for Customer Distribution Part 96919 101
34. talog number due to shorter metering tubes RBC 8921027701 WBC 8921027801 Card Cage Backplane X X Not Interchangeable New PCB referred to as PCB accommodates SPM Motherboard Module on CCM DLA and SBC CELL DYN 1700 8960274001 Motor Processor Module X X Interchangeable MPM Power Dist Module X N A Eliminated PDM Power Supply Assembly X X Not Interchangeable Completely redesigned 8921029201 Pre Amp Filter PCB N A X New PCB Used to filter 12vdc analog to PAM 8960281001 Pre Amp Module X X Not Interchangeable PAM Revised PCB has new high frequency filter circuit 8960275001 Pressure Regulator X X Interchangeable Pump Relay Module X X Interchangeable PRM Signal Processor X X Not Interchangeable Module SPM Revised PCB has MAM circuitry incorporated 8960272001 Single Board Computer N A X New PCB Uses Celeron SBC 850 MHz Processor 8921029501 Solenoid Drive Module X X Interchangeable SDM Touch Pad Interconnect N A X New PCB Used for PCB membrane keypad 8960279001 Waste Overflow PCB X X Interchangeable For Abbott Internal Use Only 35 Not for Customer Distribution Part 96919 101 CELL DYN 1800 Field Service Training Workbook Part 96919 101 User Interface Computer UIC Discussion The User Interface Computer UIC is designed around an Intel compatible Celeron 850 MHz Microprocessor and is consolidated into a Single Board Computer design The SBC receives 5v 412v di
35. tings and save them to memory If the symptom returns after the power is turned off on then the SBC PCB will need to be replaced Cell Count Module CCM The Cell Count Module CCM is a dedicated single board computer based on the Motorola MC68HC11KW1 processor In the CELL DYN 1800 the CCM inherited the DCM circuitry formally on the CELL DYN 1700 The CCM functions include the following e Flow system control and functions i e reagent handling waste handling sample handling volume metering etc e Measurement WBC RBC PLT and HGB and analog control e Receiving process commands from the SBC and executing related Flow script s e Extracting MPM related commands from Flow scripts and passing those commands to the MPM e The data from the SPM will directly address the CCM Histogram Memory Four histograms will be built WBC RBC for MCV RBC Count and PLT Note The CCM does not have any adjustments For Abbott Internal Use Only 39 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Signal Processor Module SPM The SPM receives cell pulse input signals from the Pre Amplifier Module PAM and handles the dynamic cell by cell analysis The cells are analyzed for size by the SPM The size information is handed off to the CCM that enumerates the data In the CELL DYN 1800 the SPM inherited the MAM circuitry formally on the CELL DYN 1700 instrument The SPM functions on the CELL D
36. tion CELL DYN 1800 Field Service Training Workbook Part 96919 101 Section 3 WBC Differential Changes e Learning Objectives e CELL DYN 1700 and CELL DYN 1800 Comparison Chart System Performance e Overview of Changes to WBC Differential o Cyanide Free Diff Lyse o Differential Algorithms Differential Channel Size Ranges Fixed Groups and Thresholds Region Alerts For Abbott Internal Use Only 22 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Learning Objectives Learning Objectives Section 3 will explain changes made to the WBC differential as a result of the use of Cyanide Free CN Free Diff Lyse on the CELL DYN 1800 After reviewing Section 3 of this workbook you should be able to do the following e Describe the changes made to the CELL DYN 1800 algorithms to accommodate the CN Free Diff Lyse e Describe changes made to the region alerts as a result of the CN Free Diff Lyse For Abbott Internal Use Only 23 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 CELL DYN 1700 and CELL DYN 1800 Comparison Chart System Performance Name System Performance Parameters WEC Differential Histograms Technology Closed Tube Sampler Sample Volume CELL DYN 1700 1700CS CELL DYN 1800 18 PCT and PDW not reportable in 18 PCT and PDW not reportable in US 3 Part Diff Lym Mid Gran
37. to the Card Cage Assembly Overview of Changes The main changes that have been made to the right side of the instrument are as follows e The 486 CPU Assembly with power supply has been replaced by a Single Board Computer SBC design which includes LCD driver circuitry e The Data Link Adapter DLA and CPU Assembly known as SBC or Single Board Computer connect directly into the Card Cage Backplane PCB e The 14 CRT Monitor was eliminated and replaced by a 10 4 inch LCD Display Screen e The Main Amplifier Module MAM circuitry was incorporated into the Signal Processor Module SPM PCB creating a single PCB for gain pulse editing and threshold responsibilities e The Device Control Module DCM circuitry was incorporated into the Cell Count Module CCM PCB creating a single PCB for Analog to Digital conversion and instrument control functions e The PC Power Supply 24v Power Supply and Linear Power Supply has been replaced with a single power supply assembly that controls and supplies all AC and DC voltages to the instrument e The voltage and frequency selection switches on the Linear Power Supply were eliminated and replaced with an AC Sense Regulator PCB that automatically adjusts for different voltages and frequencies e The Pre Amp Filter PCB was placed between the AC Regulator PCB and the Pre Amplifier Module PAM to eliminate noise that may be generated on the 12vdc analog line e A High Frequency Reduction F
38. tor ID 123 Sequence 15 E Hhole Blood Open Sample Factors Parameter Method Time Operator HBC Factory REC Factory HGB Factory MCW Factory PLT Factory HPU Factory PRE AUTO CAL ENTER HELP DILUTE SELECT FACTOR ERROR CALIBRATION Menu Differences e All references related to the closed mode have been eliminated since this mode does not exist in the CELL DYN 1800 For Abbott Internal Use Only 19 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 DIAGNOSTICS MENU Printer output OFF DIAGNOSTICS Feb 13 2883 14 59 am Operator ID 123 DIF PRESS HELP ERROR KEY Sequence 5383 INITIAL RAH COUNT MORE PRINTER HELP MAIN IZATION DATA TEST OUTPUT ERROR DIAGNOSTICS Menu Differences e All references related to the closed mode have been eliminated since this mode does not exist in the CELL DYN 1800 For Abbott Internal Use Only 20 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 SPECIAL PROTOCOLS MENU SPECIAL PROTOCOL Feb 13 2483 18 59 an Operator ID 123 DIF PRESS HELP ERROR KEY Sequence 5383 DAILY LYSE REAGENT AUTO MORE HELP MAIN SHUTDOWN PRIME PRIME CLEAN ERROR SPECIAL PROTOCOLS Menu Differences e All references related to the closed mode have been eliminated since this mode does not exist in the CELL DYN 1800 For Abbott Internal Use Only 21 Not for Customer Distribu
39. tors C Solenoids D All of the above 34 The Pre Amp Filter PCB is located between the AC Regulator PCB and the PAM PCB True or False 35 The same PAM PCB is used for both the CELL DYN 1700 and CELL DYN 1800 True or False 36 The only physical difference between the CELL DYN 1700 and CELL DYN 1800 Metering PCBs are the metering tubes True or False 37 Which PCB illuminates the LCD Display Screen A Status Indicator PCB B LCD Interconnect PCB C Backlight Inverter PCB D None of the above 38 What happens when the Backlight Inverter PCB is defective A Vertical lines B Horizontal lines C Diagonal lines D Blank screen For Abbott Internal Use Only 51 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Answer Sheet l TEES e o 9 21 _ 3 NEN NE 4 a 2 ur 5 C 24 2 3 4 6 NEN 25s ux 7 a 26 _ 8 T 5 9 qM 28 ps 10 29 gt ll 305 ga t J2 oo elo ut 135 aX 14 33s 1 onn DX 0 34 l6 SX EI 17 36 1 18 NEN 37 2 95 38 GEL For Abbott Internal Use Only 52 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Parts and Accessories CELL DYN 1800 Equipment Parts and Accessories Abbott List Number Description of Part Configuration 03H54 01 Accessory Kit 1 NOTE For a list of items in the Kit see the Accessory Kit Table
40. v digital 2v generated on DC Regulator PCB For Abbott Internal Use Only 31 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Name CELL DYN 1700 1700CS CELL DYN 1800 Hardware Continued 24v Power Supply No replaced by 28v switching power supply Linear Power Supply Yes 15v 15v 12v unreg 424v No Power Supply Assembly is unreg and 100v made up of three 3 components 28v switching power supply 428v DC Regulator PCB 5v 12v digital 12v digital 14v and AC Regulator PCB 12v analog 12v analog 100v Pre Amp Filter Yes used to filter the 12v analog between the AC Regulator PCB and PAM PCB Input Voltage Select Yes found at rear of instrument on No AC Regulator PCB senses input Switches Linear Power Supply voltage and adjusts automatically For Abbott Internal Use Only 32 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Overview of Hardware Changes Repackaging The changes made to the Power Supply Assembly in the CELL DYN 1800 necessitated the relocation of the Card Cage Assembly to the front of the instrument behind the Bezel The CRT Display Monitor was replaced by a smaller and more compact LCD Display Screen to allow for the relocation of the Card Cage Assembly In addition the CPU Assembly was removed and replaced by a Single Board Computer PCB which connects in
41. y Lyse 1L I 92161 02 Waste Line Assembly 1 92163 01 Diluent Line Inlet Assembly 1 92178 01 Lyse Line Inlet Tube 1 CELL DYN Reagents 07H84 01 CN Free Diff Lyse 3 8L 3 8 liter cube 07H84 02 CN Free Diff Lyse 960mL 1 x 960 mL Bottle 99320 01 Detergent 20 liter cube 99326 01 Detergent 4 x 3 8 liter bottles 98329 01 Detergent 1 x 3 8 liter bottle 99220 01 Diluent 20 liter cube 99226 01 Diluent 4 x 3 8 liter bottles 99229 01 Diluent 1 x 3 8 liter bottle Same as CELL DYN 1700 Accessory Kit For Abbott Internal Use Only 54 Not for Customer Distribution CELL DYN 1800 Field Service Training Workbook Part 96919 101 Parts and Accessories CELL DYN Controls and Calibrators 99109 01 CELL DYN 16 Tri Level Control 12 x 2 5 mL 4 of each level 99105 01 CELL DYN 16 Tri Level Control Half 6 x 2 5 mL 2 of Pack each level 02H40 01 CELL DYN 16 Normal Control 6x 2 5 mL 99110 01 CELL DYN Calibrator 2x 2 5 mL 01H92 01 CELL DYN 16 Control Assay Disk 1 93111 01 CELL DYN 22 Tri Level Control 12 x 2 5 mL 4 of each level 99106 01 CELL DYN 22 Tri Level Control Half 6 x 2 5 mL 2 of Pack each level 99103 01 CELL DYN 22 Normal Control 6 x 2 5 mL of normal control 99120 01 CELL DYN 22 Calibrator 2x 2 5 mL 01H91 01 CELL DYN 22 Control Assay Disk 1 CELL DYN Consumables 99644 01 Enzymatic Cleaner Concentrate