ALS life support on the run
Contents
1. 10 electrode Cable Set Extremities Grabber use with M1976A AAMI 10 electrode Cable Set Extremities Grabber use with M1978A IEC 10 electrode Cable Set Chest Grabber use with M1968A AAMI 10 electrode Cable Set Chest Grabber use with M1971A IEC 5 Lead ECG Lead Set Limb Leads Snaps Shielded Electrode AAMI 5 Lead ECG Lead Set Chest Leads Snaps Shielded Electrode AAMI 5 Lead ECG Lead Set Limb Leads Snaps Shielded Electrode IEC 5 Lead ECG Lead Set Chest Leads Snaps Shielded Electrode IEC Some options upgrades and accessories are not available in all countries Contact your local Philips Sales Representative for specific information ao 3 Lead ECG Cables M1500A M1510A M1605A M1615A M1669A M1671A M1672A M1673A M1674A M1675A M1678A 989803173121 989803173141 989803174201 3 Lead ECG Trunk Cable AAMI 3 Lead ECG Trunk Cable IEC 3 Lead ECG Set with Snaps AAMI 3 Lead ECG Set with Snaps IEC 3 Lead ECG Trunk Cable AAMI IEC 3 Lead ECG set grabber AAMI ICU 3 Lead ECG set ICU grabber IEC 3 Lead ECG set with snaps ICU 3 Lead ECG set with snaps IEC ICU 3 Lead ECG set with grabbers OR 3 Lead ECG set grabber IEC OR 3 Lead ECG disposable bedside AAMI 3 Lead ECG disposable telemetry AAMI 3 Lead ECG disposable bedside IEC 5 Lead ECG Cables M1520A M1530A M1602A M1604A M1625A M1635A M1644A M1645A M1668A M1949A M1968A M1971A M1973A M1974A M1976
2. ALS life support on the run Philips M3536A HeartStart MRx monitor defibrillator PHILIPS sense and simplicity The first thing you ll notice about the HeartStart MRx M3536A monitor defibrillator is its large color display Look further and you ll see that it has much more to offer today s EMS professionals Designed with the EMS market specifically in mind the HeartStart MRx M3536A combines a multi parameter monitor with 12 Lead ECG acquisition transmission capability manual and semi automated defibrillation pacing and EMS specific configuration options in one device The M3536A unites Philips industry leading monitoring technologies with superior diagnostic measurements predictive instruments Vital Signs Trending reports Event Summaries an open systems approach to data transmission and our patented resuscitation therapies Monitoring starts once a patient cable is connected to the device The HeartStart MRx is equipped for 3 and 5 Lead ECG monitoring with arrhythmia detection Options include the advanced DXL 12 Lead ECG algorithm pulse oximetry noninvasive blood pressure invasive pressures temperature and end tidal CO The HeartStart MRx is prepared for today s needs and upgradeable to meet tomorrow s as your needs evolve so does your HeartStart MRx Its therapies manual and semi automatic defibrillation and synchronized cardioversion feature Philips patented low energy SMART Biphasic w
3. C to 60 C 4 F to 140 F Temperature Operating 0 C to 50 C 32 F to 122 F Temperature Storage Relative 0 to 75 Humidity Operating 0 to 95 Relative Humidity Solids Water IP55 Meets ISO IEC 60529 Resistance EMC Meets IEC 60601 1 2 and RTCA DO 160E CPR Meter Adhesive Pads Dimensions 39 mm x 90 mm Storage 20 C to 60 C 4 F to 140 F Temperature Operating 0 C to 50 C 32 F to 122 F Temperature Storage Relative 0 to 75 Humidity Operating 0 to 95 Relative Humidity Material Foam pad with biocompatible adhesive on both sides Shelf Life 2 years when applied to the CPR meter or 4 years in an unopened package Bluetooth 100 meters approximately 300 feet Class maximum transmission range Dependent upon transmission range of lowest class Bluetooth device Most Bluetooth devices are Class Il which transmit at maximum ranges of up to 10 meters 33 feet Bluetooth Tested with Toshiba 4 20 11 IVT Stacks 2 1 2 0 Product 05 04 11 20060301 stack Widcomm 4 0 1 2400 Bluetooth 1 1 or greater Version Bluetooth devices used with the MRx must support the Bluetooth Dialup Networking Profile DUN or the File Transfer Profile FTP DUN devices must also have a data transfer plan that supports packet data transmission Event summaries can only be transmitted via Bluetooth File Transfer Profile not DUN AC Power Module Input 100 240 VAC 5
4. M4761A Compression Sensor M4762A Sensor Adhesive Pads 10 pack M4763A Compression Sensor Pads CPR cable Defibrillator Pads Hands Free Cable barrel style 2 2 m 7 ft Defibrillator Pads Hands Free Cable plug style 2 2 m 7 ft Pads Adapter use with M3507A Defibrillator Pads Hands Free Cable HeartStart pads CPR meter cable and connector ECG Monitoring Electrodes M2202A M4612A M4613A High Tack Foam 5 electrodes pack 60 packs case Solid Gel Electrodes 5 electrodes pack 60 packs case Solid Gel Electrodes 30 electrodes pack 10 packs case External Paddles M3543A Water Resistant External Paddles 12 Lead ECG Cables M3525A 989803147691 M3526A M3527A M3528A M3529A M5530A M1663A M1949A M1968A M1971A M1976A M1978A 989803158061 989803158071 989803158081 989803158091 2 7 meter 10 Lead ECG Trunk Cable 12 pin Connector for 3 Lead 5 Lead and 12 Lead use 1 3 meter 10 Lead ECG Trunk Cable 12 pin Connector for 3 Lead 5 Lead and 12 Lead use 3 Lead ECG Set and Plug with Snap AAMI Add 7 Lead ECG set for 12 Lead Use AAMI 3 Lead ECG Set and Plug with Snap IEC Add 7 Lead ECG Set for 12 Lead use IEC Combiner Plug for 3 wire Lead Set for use with M3526A M3528A 10 Lead ECG Patient Trunk Cable 12 pin ECG Input Connector for 5 Lead and 12 Lead use 10 Lead ECG Patient Trunk Cable 12 pin ECG Input Connector for 5 Lead and 12 Lead use
5. CO O M2520A Smart CapnoLine Pediatric M2522A Smart CapnoLine Adult Non Intubated Single Purpose Circuits CO M2524A Smart CapnoLine Pediatric M2526A Smart CapnoLine Adult M3538A Lithium lon Battery with fuel gauge M3539A AC Power Module M5528A Vehicle Wall Mount M5529A DC Power Module 989803135301 2 Bay Battery Support System for Lithium lon Batteries 989803135331 4 Bay Battery Support System for Lithium lon Batteries 989803135341 4 Bay Battery Support System for Sealed Lead Acid and Lithium lon Batteries Paper 40457C 40457D 989803138171 989803138181 Sync Cables M1783A M5526A 50 mm Chemical Thermal Gray Grid 10 rolls 50 mm Chemical Thermal Gray Grid 80 rolls 75 mm Chemical Thermal Red Grid 10 rolls 75 mm Chemical Thermal Red Grid 80 rolls Sync Cable 2 5m 8 ft Sync Cable 7 6m 25 ft Miscellaneous M1781A M3537A M3541A M3549A M3725A M4737A M4759A M5525A 453564063841 453564063851 989803180111 M5528A 989803146981 989803174901 989803176411 989803179151 989803172501 989803174261 989803180871 Test Load for use with M3507A Pad Cable Bedrail Hook mount Red Carrying Case includes 3 accessory pouches and shoulder strap Wide Bedrail Hook mount Test Load for use with M3508A Pad Cable Display cover Replacement adult paddle Handle Light gray for white M3536A Calibration Kit NBP Calibration Kit EtCO H
6. M4554B Easy Care Reusable Cuff Small Adult M4554B5 Easy Care Reusable Cuff Small Adult 5 M4555B Easy Care Reusable Cuff Adult M4555B5 Easy Care Reusable Cuff Adult 5 M4556B Easy Care Reusable Cuff Adult Long M4556B5 Easy Care Reusable Cuff Adult Long 5 M4557B Easy Care Reusable Cuff Large Adult M4557B5 Easy Care Reusable Cuff Large Adult 5 M4558B Easy Care Reusable Cuff Large Adult X Long M4558B5 Easy Care Reusable Cuff Large Adult X Long 5 M4559B Easy Care Reusable Cuff Thigh M4559B5 Easy Care Reusable Cuff Thigh 5 M1572A Multi Patient Comfort Cuffs Pediatric M1573A Multi Patient Comfort Cuffs Small Adult M1574A Multi Patient Comfort Cuffs Adult M1575A Multi Patient Comfort Cuffs Large Adult Disposable Blood Pressure Cuffs M4572B Soft Single Patient Disposable Cuff Infant M4573B Soft Single Patient Disposable Cuff Pediatric M4574B Soft Single Patient Disposable Cuff Small Adult M4575B Soft Single Patient Disposable Cuff Adult M4576B Soft Single Patient Disposable Cuff Adult X Long M4577B Soft Single Patient Disposable Cuff Large Adult M4578B Soft Single Patient Disposable Cuff Large Adult X Long M4579B Soft Single Patient Disposable Cuff Thigh Invasive Pressures CPJ840 6 Reusable Pressure Transducer CPJ84022 Sterile disposable pressure dome for use with CPJ840J6 CPJ84046 Transducer holder for CPJ840 6 M1567A Sin
7. requirements of standard EN 60601 1 2 2001 Meets the UL 2601 1 CSA C22 2 No 601 1 EN 60601 1 and 60601 2 4 standards Device not suitable for use in the presence of concentrated oxygen or a flammable anesthetic mixture with air oxygen or nitrous oxide Continuous Philips Healthcare is part of Royal Philips Electronics How to reach us www philips com healthcare healthcare philips com Asia 49 7031 463 2254 Europe Middle East and Africa 49 7031 463 2254 Latin America 55 11 2125 0744 North America 1 425 487 7000 800 285 5585 toll free US only Trademarks and trade names are those of their respective owners Please visit www philips com heartstart 2012 Koninklijke Philips Electronics N V PHILIPS All rights are reserved Philips Healthcare reserves the right to make changes in specifications or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication Printed in The Netherlands 4522 962 52011 FEB 2012
8. set Red Carry Case detachable pouches Black soft carry case pads only Instructions for Use printed copy User Training Video English only User Training DVD English only Label for AED emphasis Service Manual English only Service Training Workbook Service Training DVD English only One Year On Site Warranty Three Year Biomed Warranty U S Canada and Australia only Two Year Bench Warranty with Loaner U S and Canada only Some options upgrades and accessories are not available in all countries Contact your local Philips Sales Representative for specific information 4 Upgrades Some options upgrades and accessories are not available in all countries Contact your local Philips Sales Representative for specific information Supplies Multifunction Electrode Pads White Barrel Connector M3501A M3504A Adult Child Pads barrel connector Infant Pads barrel connector Gray Plug Connector M3713A M3716A M3717A M3718A M3719A 989803166021 HeartStart Adult Child Plus Pads HeartStart Adult Child Radiolucent Pads HeartStart Infant Plus Pads HeartStart Adult Child Radiotransparent Reduced Skin Irritation Pads HeartStart Infant Radiotransparent Reduced Skin Irritation Pads HeartStart Adult Child Preconnect Pads Hands Free Pads Therapy Cables M3507A M3508A 05 10200 989803158661 989803162401 CPR meter 989803163291 CPR meter Adhesive Pads 989803158661 Pads CPR meter Cable
9. to the TraceMaster ECG Management System or other ECG management systems via the DatamedFT Patient Data Storage Internal Event The internal Event Summary stores up Summary to 12 hours of 2 continuous ECG waves 1 CO wave and 2 invasive pressure waves events and trending per event summary There is a maximum capacity of 55 Event Summaries or 240 megabytes 62 megabytes if you have a 64 megabyte card installed of patient data whichever comes first Data Card Event The Data Card has a maximum capacity of 60 Event Summaries or 240 Summary megabytes 62 megabytes if you have a 64 megabyte card installed of patient data whichever comes first Q CPR Measurements Compressions Depth rate release complete or incomplete and duty cycle Ventilations Volume rate and inflation time Feedback Type Verbal Prioritized corrective verbal feedback for all measurements Numerical Measurement values for compression rate ventilation rate and no flow time Graphical Compression wave with correct depth target zone Lung icon for ventilation volume Integrated into Code ALS resuscitation and AED BLS resuscitation views User Interface CPR Meter Dimensions 154mm x 64mm x 28mm with a 91m integrated cable Weight 8 3 oz 235 g Input voltage 4 0 6 0V at 170mA The CPR meter is electrically and galvanically isolated from the defibrillator power and communication sources Storage 20
10. 0 60 Hz 1 0 46 A Class 1 Output 18 V 5 A 90 W Battery Minimum 14 4 V Rechargeable Lithium lon DC Power Module 10 32 VDC 11 A 18V 5 A 90W Input Output Temperature Humidity Atmospheric Pressure Range Shock Operating Impact Shock Non Operating Bump Free fall Vibration Operating Impact Vibration Non Operating Solids Water Resistance EMC Safety Other Considerations Mode of Operation 0 C to 45 C operating 20 to 70 C storage Up to 95 relative humidity Operating and Storage 1014 hPa to 572 hPa 0 to 15 000 ft 0 to 4 500 m Half sine waveform duration lt 3 ms acceleration gt 145 g 1 time on all six faces Trapezoidal waveform acceleration gt 30 g velocity change 742 cm s 10 on all six faces EN60068 2 29 Bump Half sine 40 g peak 6 msec duration 1 000 bumps x 3 axes EC 68 2 32 Free fall Drops on all faces onto a steel surface excluding bed rail hook 30 in 76 2 cm with carrying case 16 in 40 6 cm without carrying case Operating MIL STD 810E 514 4 Category 6 Helicopter General Storage UH60 IEC 68 2 6 Vibration sinusoidal 10 57 Hz 0 15mm 58 150 Hz 2g 20 sweeps x 3 axes IEC 68 2 64 Vibration broad band random 10 20 Hz 0 05 g2 Hz 20 150 Hz 3 dB octave 150Hz 0 0065g2 Hz 1 5 hours x 3 axes IP24 Water testing performed with cables connected to the device Complies with the
11. 21 A22 A23 A24 A25 A26 B01 B02 B04 B05 B06 B07 B08 B09 B10 B11 B12 B14 SpO white SpO and NBP white SpO NBP and EtCO white EtCO white SpO NBP EtCO and Temperature white SpO NBP EtCO Invasive Pressures and Temperature white SpO NBP Invasive Pressures and Temperature white EtCO and SpO white Base Unit gray SpO gray SpO and NBP gray SpO NBP and EtCO gray EtCO gray SpO NBP EtCO and Temperature gray SpO NBP EtCO Invasive Pressures and Temperature gray External Pacing 12 Lead ECG Acquisition 75 mm Printer Asian 75mm Printer 12 Lead ECG Transmission Bluetooth wireless technology 12 Lead ECG Transmission RS 232 and Bluetooth Q CPR Q CPR Data Capture Event Summary Bluetooth 12 Lead Transmission Rosetta Lt Interface Available in the U S only Batch LAN Data Transfer Audio Recording all modes B17 B18 C01 C02 C03 C05 C06 C07 C10 C11 C12 C15 C16 C20 C21 LP1 LP2 LP3 LPK SM1 SM2 SM3 Wo1 W22 WA2 ACI TIPI and TPI Periodic Clinical Data Transmission External Paddles China only Standard External Paddles water resistant for EMS use Data Card Additional Battery AC Power Module Barrel style Pad Cable replacement for Standard Pad Cable 5 5 ECG lead set with grabbers Long 2 7m ECG Trunk Cable 3 7 Snap Lead set 5 Lead ECG Cable Shielded 12 Lead ECG Cable
12. A M1978A M1979A M1984A 989803173131 989803173151 989803174211 989803176161 989803176171 989803176181 989803176191 5 Lead ECG Trunk Cable AAMI 5 Lead ECG Trunk Cable IEC 5 Lead Chest ICU Snaps AAMI 5 Lead Chest ICU Snaps IEC 5 Lead ECG set with snaps AAMI 5 Lead ECG set with snaps IEC 5 Lead ICU snaps AAMI 5 Lead ICU snaps IEC 5 Lead ECG Trunk Cable AAMI IEC 5 plus 5 ECG Trunk Cable AAMI IEC 5 Lead ICU grabber AAMI 5 Lead ICU grabber IEC 5 Lead OR grabber AAMI 5 Lead OR grabber IEC 5 Lead OR Grabber IEC 5 Lead Chest ICU Grabber AAMI 5 Lead Chest ICU Grabber IEC 5 Lead Chest OR Grabber IEC 5 Lead ECG disposable bedside AAMI 5 Lead ECG disposable telemetry IEC 5 Lead ECG disposable bedside IEC 5 Lead Shielded Limb Snap AAMI 5 Lead Shielded Chest Snap AAMI 5 Lead Shielded Limb Snap IEC 5 Lead Shielded Chest Snap IEC M1131A M1191A M1191B M1191AL M1191BL M1191T M1192A M1192T M1193A M1194A M1195A M1196A M1196T M1903B M1904B M1941A M1943A 989803164571 989803164581 Disposable SpO Sensor Adult Pediatric Finger Reusable SpO Sensor Adult finger Reusable SpO Sensor Adult finger Reusable SpO Sensor Adult finger 3m cable Reusable SpO Sensor Adult finger 3m cable Reusable SpO Sensor Adult finger 9 pin connector Reusable SpO Sensor Pediatric Small Adult Reusable SpO Sensor Pediat
13. Sweep Speed 25 mm s nominal stationary trace sweeping erase bar for ECG Invasive Pressures and SpO 6 25 mm s for CO Type Rechargeable Lithium lon minimum 6 30 Ah 14 4 V 91 Wh Dimensions 6 5 in H x 3 8 in W x 1 6 in D 165 mm x 95 mm x 42 mm Weight Less than 1 8 Ib 0 82 kg Charge Time Approximately 3 hours to 100 Approximately 2 hours to 80 indicated by battery fuel gauge Charging the battery at temperatures above 45 C may degrade battery life Battery Fuel gauge on battery capacity indicator Indicators on display flashing RFU indicator chirp and LOW BATTERY message appears on display for low battery condition Storage Storing the battery for extended periods at temperatures above 40 C will reduce battery capacity and degrade battery life Low battery condition triggered with at least 10 minutes of monitoring time and 6 maximum energy discharges remain with a new battery at room temperature 25 C Battery Capacity Shocks At least 50 200 charge shock or disarm cycles Monitoring At least 9 hours of ECG monitoring only with no other options installed Monitoring and Shocks At least 5 hours of monitoring with ECG SpO CO temperature and 2 invasive pressures monitored continuously NBP measured every 15 minutes and 20 200 joule discharges Monitoring and At least 3 5 hours while pacing at 180 Pacing ppm at 160 mA and monitoring as described above With a new
14. andle Dark gray for gray M3536A Vehicle Wall Mount Data Card and Tray Green Hard Transport Case pads only Paddle Tray Kit Night Vision Goggle Compatible Display Cover Black Soft Carry Bag Black Soft Carry bag Straps only Black Carrying Case includes 3 accessory pouches and shoulder strap Some options upgrades and accessories are not available in all countries Contact your local Philips Sales Representative for specific information 9 Specifications Delivered Energy Accuracy Nominal Delivered Energy vs Patient Impedance Energy Load Impedance ohms 2 25 50 75 100 125 150 175 1J Z ce 10 09 08 2 1 8 2 0 2 0 1 9 1 7 1 6 1 5 3J 2 8 3 0 3 0 3 1 3022927 4 3 7 40 40 41 42 42 1 40 5J 46 5 0 51 IS S2 S2 5 0 6 5 5 6 0 61 62 63 63 61 7 6 4 17 0 17 1 72 7 3 7 3 17 1 8 7 4 80 81 82 84 83 81 9 83 9 0 9 1 93 194 94 19 1 10 J 9 2 10 10 10 10 10 10 15J d MiS 1S Iis 16 16 15 20J 18 20 20 21 21 21 20 30 28 30 30 31 31 31 30 50 46 50 51 51 52 52 50 70 J 64 70 71 72 ma za Za 100 J 92 100 101 103 104 104 101 120 J 110 120 121 123 125 125 121 150 J 138 150 152 154 157 156 151 170 J 156 170 172 175 177 177 172 200 J 184 200 202 206 209 209 202 Accuracy 2 J for 1 10 energy levels 15 for all other energy levels Defibrillator Waveform Biphasic Truncated Exponential Wave
15. aveform which is proven effective in emergency resuscitation and for minimizing post resuscitation heart dysfunction No other external defibrillation waveform is supported by more peer reviewed clinical data Transcutaneous pacing can be added and the MRx will pace in either demand or fixed mode Leading the way with meaningful innovations in emergency care Clinical decision support that positively impacts patient care Predictive instruments designed to help support confident decision making Enhanced resuscitation therapies Rugged durable and reliable Intuitive and easy to use This is the HeartStart MRx M3536A ALS Monitor Defibrillator for emergency care To help caregivers perform high quality CPR the Q CPR option is available It offers real time measurement and corrective feedback on the rate depth and duration of compressions as well as the frequency of ventilations and also provides notification of lack of CPR activity Q CPR supports AHA ERC 2010 Guidelines and protocols The CPR meter provides feedback on a graphical display right in the line of site of the caregiver performing CPR HeartStart MRx displays measurements and patient care data on an easy to read backlit 8 4 inch screen and also comes in your choice of colors white or gray Numerics and waveforms can be reconfigured and the screen reorganized enabling you to quickly locate the information you need most With wide viewing angles it displays an even
16. cond maximum pediatric or Time adult Pressure 3 mmHg Transducer Accuracy Alarm Range Systolic high limit 35 270 Adult 35 180 Pediatric Systolic low limit 30 265 Adult 30 175 Pediatric Diastolic high limit 15 245 Adult 15 150 Pediatric 10 240 Adult 10 145 Pediatric 25 255 Adult 25 160 Pediatric Mean low limit 20 250 Adult 20 155 Pediatric Diastolic low limit Mean high limit Calibration yearly or every 10 000 cycles schedule Auto Mode 1 2 5 5 10 15 30 60 or 120 Repetition Time minutes Measurement Auto manual mode 30 seconds Time average HR gt 60 bpm 170 seconds maximum Interconnect Tube M1598B Connect tubing 5 ft 1 5 m Length M1599B Connect tubing 10 ft 3 m End Tidal Carbon Dioxide Range 0 to 99 mmHg at sea level Resolution 1mmHg 0 1 kPa Accuracy For values between 0 and 38 mmHg 2 mmHg For values between 39 and 99 mmHg 5 of reading 0 08 for every 1 mmHg above 40 mmHg For breath rates above 80 and EtCO values gt 18 mmHg accuracy is 4 mmHg or 12 of reading whichever is greater Alarm Range Low Limit 10 to 94 mmHg Adult Pediatric High Limit 20 to 95 mmHg Adult Pediatric Calibration yearly or every 4 000 hours schedule Sample Size 50 ml per min Drift of Over a 24 hour period accuracy claims Measurement above are maintained Accuracy Range 0 to 150 rpm Resolution 1 rpm Ac
17. curacy 0 to 40 rpm 1 rpm 41 to 70 rpm 2 rpm 71 to 100 rpm 3 rpm 101 to 150 rpm 5 rpm Alarm Low Limit 0 to 99 rpm Adult Pediatric Range High Limit 10 to 100 rpm Adult Pediatric Calibration Gas for CO2 5 Carbon Dioxide 21 Oxygen 74 Nitrogen Cylinder Size BD Ingredients Method of Preparation Gravimetric Blend Tolerance 0 03 Accuracy 0 03 absolute Moisture 10 PPM Maximum Expiration Period 2 years Pressure 144 PSIG Volume 10L Invasive Pressures Transducer Sensitivity 5uV V mmHg 37 5yV V kPa Sensitivity Adjustment 10 Range Transducer Load 195 to 2200 ohms Resistance Transducer Output 0 to 3000 ohms Resistance Frequency Response 0 12 Hz or 0 40 Hz Zero Adjustment Range 200 mmHg 26 7 kPa Zero Adjustment 1 0 mmHg 0 1 kPa Accuracy Zero Setting Drift lt 0 1 mmHg C 0 013 kPa C Gain accuracy excluding 1 of reading or 1 mmHg 0 1 transducers Gain Drift Overall Accuracy included listed transducers kPa whichever is greater less than 0 05 C 4 of reading or 4 mmHg 0 5 kPa whichever is greater Measurement Range 40 to 361 mmHg 5 3 to 48 1 kPa Measurement 1 mmHg 0 1 kPa Resolution Noise lt 1 mmHg 0 1 kPa Transducer Dome Refer to the specific device s Volume Displacement specifications Additional Noise from lt 3 mmHg EMI if operating under conditions according to EMC standard EN60601 1 2 Radiated Immunity 3 V m or Conduc
18. eart Rate Lob sue Naess ventricular flutter and polymorphic Arrhythmia Alarms VTACH Extreme Tachy Extreme ventricular tachycardia Brady PVC rate Pacer not capture P Shock Advisory Meets AAMI DF 39 acer not pacing Algorithm Pacemaker Pulse Amplitude from 2 mV to 700 mV Sensitivity and Specificity Rejection Capability ECG Cable Length Common Mode width from 0 1 ms to 2 0 ms as per ANSI AAMI EC13 1992 3 1 4 1 9 ft 2 7 m Greater than 90 dB measured per Rejection AAMI standard for cardiac monitors EC 13 ECG Size 2 5 5 10 20 40 mm mV autogain Frequency Response Bandwidth AC Line 60 Hz or 50 Hz Filter 3 Lead 5 Pads ECG for Display Monitor 0 15 40 Lead and Hz or EMS 1 30 Hz Pads ECG for Pads Printer Monitor 0 15 40 Hz or EMS 1 30 Hz Leads ECG for Display Monitor 0 15 40 Hz or EMS 1 30 Hz Leads ECG for Printer Diagnostic 0 05 150 Hz or Monitor 0 15 40 Hz or EMS 1 30 Hz 12 Lead ECG for Display 0 05 150 Hz 0 05 40 Hz 0 15 40 Hz ECG for Report 0 05 150 Hz 0 05 40 Hz 0 15 40 Hz 0 05 150 Hz Patient Isolation defibrillation proof ECG Type CF SpO Type CF EtCO Type CF NBP Type CF Invasive Type CF Pressures Temperature Type CF External Defib Type BF Internal Defib Type CF Size 8 4 in diagonal 128 mm x 171 mm Type TFT Color LCD Resolution 640 x 480 pixels VGA Wave Viewing 5 seconds ECG Time
19. ensor M1191A 1 standard deviation 70 to 100 2 0 M1191AL 1 standard deviation 70 to 100 2 0 M1191B 1 standard deviation 70 to 100 2 0 M1191BL 1 standard deviation 70 to 100 2 0 M1191T 1 standard deviation 70 to 100 2 0 M1192A 1 standard deviation 70 to 100 2 0 M1192T 1 standard deviation 70 to 100 2 0 M1194A 1 standard deviation 70 to 100 3 0 M1195A 1 standard deviation 70 to 100 3 0 M1196A 1 standard deviation 70 to 100 3 0 M1196T 1 standard deviation 70 to 100 3 0 M1131A 1 standard deviation 70 to 100 3 0 Pulse Rate Accuracy 2 or 1 bpm whichever is greater Pulse Alarm Range Low Limit 30 to 195 Adults 30 to 235 Pediatric High Limit 35 to 200 Adult 35 to 240 Pediatric SpO Alarm Range Low Limit 50 to 99 Adult Pediatric High Limit 51 to 100 Adult Pediatric SpO and Pulse High Low 10 seconds Alarm Signal Generation Delay Note The above referenced sensors were validated for use with the HeartStart MRx using the Philips picoSAT II SpO module with Fourier Artifact Suppression Technology FAST This module is not available as a stand alone device Non Invasive Blood Pressure Pressure Range Systolic 40 260 mmHg Diastolic 20 200 mmHg Initial Pressure 160 mmHg Adult 120 mmHg Pediatric Maximum 280 mmHg Pressure Overpressure Maximum of 300 mmHg Safety Limits Cuff Inflation 75 se
20. form parameters adjusted as a function of patient impedance Shock Via multifunction electrode pads or Delivery paddles Charge Less than 5 seconds to 200 joules with a Time new fully charged Lithium lon battery pack at 25 C Patient Impedance Range Minimum 15 ohm internal defibrillation 25 ohm external defibrillation Maximum 180 ohm Note Actual functional range may exceed the above values Dimensions 12 4 in W x 8 3 in D x 11 7 in H with pads 31 5 cm x 21 0 cm x 29 5 cm Dimensions 13 4 in WV x 8 3 in D x13 6 in H with paddles 34 0 cm x 21 0 cm x 34 5 cm Weight 13 9 Ibs 6 3 kg including pads pads cable full roll of paper and battery Incremental weight of external standard paddles and paddle tray is 2 5 Ibs 1 1 kg Additional battery weighs less than 1 8 Ibs 0 82 kg Manual Mode Manual Output 1 10 15 20 30 50 70 100 120 Energy Selected 150 170 200 joules Controls On Off Therapy Knob Charge Shock Sync Print Mark Event ECG Lead Select Alarm Pause Event Review Disarm Energy Selection Front panel Therapy Knob Charge Control Front panel button button on external paddles Shock Control Front panel button buttons on external or switched internal paddles Synchronized Front panel Sync button Control Indicators Text prompts audio alerts QRS beeper battery status Ready For Use external power Sync mode Armed Indicators Armed Indica
21. full charged battery at 25 C 77 F Thermal Array Printer Continuous The Print key starts and stops the strip ECG Strip The printer can be configured to run real time or with a 10 second delay The strip prints the primary ECG lead with event annotations and measurements Auto Printing The printer can be configured to automatically print on Marked Events Charge Shock and Alarm When an alarm condition occurs the unit prints the primary ECG wave and the alarming wave if configured Reports The following reports can be printed Event Summary Vital Signs Trending 12 Lead Operational Check Configuration Status Log and Device Information Speed 25 or 50 mm s with an accuracy of 5 Amplitude 5 or 40 uV whichever is greater Accuracy Paper Size 50 mm W by 30 m 100 ft L 75 mm W by 30 m 100 ft L Non Invasive Pacing Waveform Monophasic Truncated Exponential Current 10 mA to 175 mA 5 mA increments Pulse accuracy 10 or 5 mA whichever is greater Amplitude Pulse 40 ms with 10 accuracy Width Rate 30 ppm to 180 ppm 10 ppm increments accuracy 1 5 Modes Demand or Fixed Rate Refractory 340 msec 30 to 80 ppm 240 msec 90 to Period 180 ppm SpO Pulse Oximetry SpO Range 0 100 Pulse rate 30 to 300 bpm Maximum Power Output lt 15 mW Wavelength Range 500 1000 nm Resolution 1 Display Update Period 1 sec typical numeric update rate Accuracy with s
22. gle channel disposable blood pressure kit Available in Europe and Asia only M1568A Dual Line blood pressure kit for measuring CVP ABP and other pressure measurements available in Europe and Asia only M1634A Reusable adapter cable available in Europe and Asia only Disposable Transducers TransPac IV ICU Medical Inc TruWave PX212 Edwards Lifescience DTX Plus DT 4812 Becton Dickinson and Co Available for purchase service from the respective manufacturers Disposable Temperature Probes 21090A Esophageal rectal 21091A Skin surface 21093A Esophageal stethoscope 21094A Esophageal stethoscope 21095A Esophageal stethoscope 21096A Foley Catheter 21097A Foley Catheter M1837A Esophageal rectal M2255A Foley Catheter Some options upgrades and accessories are not available in all countries Contact your local Philips Sales Representative for specific information 8 Reusable Temperature Probes 21075A Esophageal rectal adult 21076A Esophageal rectal pediatric 21078A Skin surface Reusable Temp Probe Extension Cables 21082A 3 0 m 2 pin plug extension cable for mini phone plug 21082B 1 5 m 2 pin plug extension cable for mini phone plug EtCO Intubated Circuits M1920A FilterLine Set Adult Pediatric 25 sets case M1921A Filter H Set Humidified Adult Pediatric 25 sets case M1923A Filter H Set Humidified Infant Neonatal yellow 25 sets case Non Intubated Dual Purpose Circuits
23. ric Finger 9 pin connector Reusable SpO Sensor Neonatal Hand Foot Reusable SpO Sensor Adult Pediatric ear clip Reusable SpO Sensor Infant Reusable SpO Sensor Adult clip Reusable SpO Sensor Adult clip 9 pin connector Disposable SpO Sensor Pediatric Finger available outside the U S only Disposable SpO Sensor Adult Finger available outside the U S only SpO Extension Cable 2m Reusable SpO Sensor Adapter Cable 1m Use with M1903B M1904B Cardinal Reusable SpO Clip Sensor Cardinal Disposable Adult Pediatric SpO Sensor NBP Interconnect Tubing M1598B M1599B Adult Pressure Interconnect Cable 1 5m Adult Pressure Interconnect Cable 3m Some options upgrades and accessories are not available in all countries Contact your local Philips Sales Representative for specific information 7 Reusable Blood Pressure Cuffs 40400A Reusable Cuff Kit 3 sizes Pediatric Adult Large Adult 40400B Reusable Cuff Kit 5 sizes Infant Pediatric Adult Large Adult Thigh 40401A Traditional Reusable Cuff Infant 40401B Traditional Reusable Cuff Pediatric 40401C Traditional Reusable Cuff Adult 40401D Traditional Reusable Cuff Large Adult 40401E Traditional Reusable Cuff Thigh M4552B Easy Care Reusable Cuff Infant M4552B5 Easy Care Reusable Cuff Infant 5 M4553B Easy Care Reusable Cuff Pediatric M4553B5 Easy Care Reusable Cuff Pediatric 5
24. t timer event markers numeric vital signs and up to four waves as well as text prompts alarms and battery status indicators On screen menus simplify navigation for configuring data setting and responding to alarms and accessing additional functionality Automated self tests straightforward ready for use checks data collection and two long life batteries make the device easy to operate For whatever situation faced during the EMS work day the HeartStart MRx is built to be tough and ready for action The M3536A model was subjected to extensive testing for rigorous and demanding environments and has received an Airworthiness Certificate from the United States Army All of these features measurements and therapies plus its compact size low weight 13 2 Ibs 5 9 kg and balanced shape mean that HeartStart MRx has the capabilities you need and the performance you demand for rapid intervention thorough care and positive patient outcomes Features Options Upgrades Standard Features Standard Accessories e ST AR Basic algorithm for arrhythmia detection e Lithium ion battery with fuel gauge ECG monitoring through monitoring electrodes and e Hands free multifunction electrode cable defibrillation pads 5 Lead ECG cable e Synchronized cardioversion e Disposable monitoring electrodes e Adjustable ECG size and autogain e Printer paper e Manual and AED operation Carrying case e SMART Biphasic waveform for defibrillation
25. ted Immunity 3 VRMS Pulse Rate Range 25 350 bpm Pulse Rate Accuracy 1 of full range Pulse Rate Resolution 1 bpm Range 0 45 C 32 113 F Resolution 0 1 C 0 2 F Accuracy 0 1 C from 25 C to 45 C 0 3 C from 0 C to 24 9 C excluding any adapter cable Settling Time lt 10 sec Constant Averaging 1 sec Time Minimum See the probe s Instructions for Use to measurement obtain minimum measurement times for time accurate readings The HeartStart MRx does not add any clinically significant time to obtain accurate readings 12 Lead ECG Inputs With a 10 Lead cable leads Il III aVR aVL aVF V C1 V C6 can be obtained All 12 Lead ECG waves can be viewed on the display simultaneously All 12 leads can be printed on the strip chart printer in 3x4 format ECG 0 15 40 Hz 0 05 40 Hz Bandwidth 0 05 150 Hz Filters Cellular transmission via a device with Bluetooth wireless technology or a cell phone with an RS 232 connection 12 Lead ECGs are transmitted through an ISP to the 12 Lead Transfer Station Bluetooth wireless transmission to an external computer which supports File Transfer Profile Server 1 1 Two way radio transmission of 12 Lead ECGs in conjunction with General Devices Rosetta Lt device Destinations Once a 12 Lead reaches the Telemedicine System it can be displayed printed faxed emailed or forwarded to another Telemedicine System It can also be forwarded
26. therapy e Defibrillator test load e Large 4 wave color display e Documentation CD containing Instructions for Use e 50mm printer white devices User training materials and Application notes e 75 mm printer gray devices e Quick reference cards e Individual adjustable volume of QRS beeper voice product Ordering nGenanen prompts and alerts M3536A HeartStart MRx ALS Monitor Defibrillator e Event summary e Vital Signs Trending Report e Configuration mode Service mode e Operational checks e Automated self tests with ready for use indicator e Lithium ion battery with fuel gauge e One Second Vitals e Static IP address capability Optional Features e SpO with Fourier Artifact Suppression Technology FAST e Noninvasive Blood Pressure e Invasive Pressures 2 channels e Temperature e Microstream EtCO e Noninvasive Pacing e 12 Lead ECG with Philips DXL algorithm e 12 Lead ECG Transmission e 75 mm printer white devices only e Q CPR CPR measurement and feedback e Q CPR Data Capture e ACI TIPI and TPI analysis e Periodic Clinical Data Transmission e Batch LAN Data Transfer via LAN cable e Audio Recording e Event Summary Transfer via FTP Some options upgrades and accessories are not available in all countries Contact your local Philips Sales Representative for specific information Option Ordering Information Option Ordering Information A01 A02 A03 A04 A05 A06 A07 A11 A20 A
27. tors Charging tone charged tone flashing Shock button and energy level indicated on display AED Mode ECG and Arrhythmia Monitoring AED Energy 150 joules nominal into a 50 ohm Inputs Up to four 4 ECG waves may be Profile test load viewed on display and up to 2 waves Text and Voice Extensive text audible messages printed simultaneously Lead Il or Prompts guide user through configured Ill is obtained through the 3 Lead protocol ECG cable and separate monitoring AED Controls On Off Shock electrodes With a 5 Lead cable 3 z z leads aVR aVL aVF and V can also Indicators Monitor display messages and be obtained Pads ECG is obtained prompts voice prompts battery through 2 multifunction pads status Ready For Use external power Lead Fault Device automatically switches to a hat h d signal wh Armed Indicators Charging tone charged tone flashing oe eae tan ior the current signal is unavailable Shock button energy level indicated Target ECG waveform can be on display and voice prompts configured ECG Analysis Evaluates patient ECG and signal Pale ect eena ifa hoki Pad Fault Dashed line appears on the display if oe a pad becomes disconnected appropriate and evaluates connection impedance for proper Heart Rate Display Digital readout on display from 15 defibrillation pad contact 0 300 bpm withianjacciracy of 10 Shockable Ventricular fibrillation and certain Oe Rhythms ventricular tachycardias including H
Download Pdf Manuals
Related Search
Related Contents
MANUEL D`UTILISATION DES OUTILS DU « Gis Portal » GS "取扱説明書" A4 deckblatt f hand liquid de.cdr Compaq AlphaServer DS20E Page 1 Page 2 《安全作業の確保のために》 西己管施工および接合作業 第2節 交通安全思想の普及徹底 Avalon Stoves 21 TRV User's Manual User`s Manual HI 38050-200 - Hanna Instruments Portugal 1 - Probst Maveg Copyright © All rights reserved.