Home
- Transcat
Contents
1. USE STACKABLE TEST LEADS AS SUPPLIED WITH CALIBRATOR 2 CHANN METER FLUKE 5502A CALIBRATOR Perf Ver X2 eps Figure 5 9 5502A NORMAL Output to Test Tool Banana Input 3 Set the 5502A to supply a DC voltage to the first calibration point in Table 5 7 Set the 5502A to operate OPR Press Fs to start the calibration Wait until the display shows calibration status READY N Oa A Press SS to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points of Table 5 7 8 Set the 5502A to Standby and continue at the Multimeter DMM Ohm Gain section Table 5 7 DMM Volt Gain Calibration Points Cal step UUT input value 5502A NORMAL dc output CL 0840 500 mV CL 0849 2 5 V CL 0841 5V CL 0842 50 V set 5502A to OPR CL 0843 500 V CL 0844 1000 V 5 25 Fluke 190 062 102 104 202 204 502 504 Service Manual Multimeter Ohm Gain Note The adjustment steps for the meter section are only for the models 190 062 190 102 190 202 and 190 502 To do the DMM Ohm Gain calibration 1 Press to select the first calibration adjustment step in Table 5 8 2 Connect the Test Tool to the 5502A as shown in Figure 5 10 Notice that the sense leads must be connected directly to the Test Tool inputs USE STACKABLE TEST LEADS AS SUPPLIED WITH CALIBRATOR Sk e A B METER 2 CHANN METER 0 0
2. FLUKE 5502A CALIBRATOR Perf Ver E2 eps Figure 5 4 5502A SCOPE Output to Test Tool Input B Cal step Calibration Adjustment Final Calibration For v11 10 and later Table 5 2 Input B LF HF Gain Calibration Points UUT input signal 5502A Setting CL 0674 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0435 100 mVpp square wave 1 kHz SCOPE edge 100 mVpp 1 kHz CL 0530 2 5 Vpp square wave 1 kHz SCOPE edge 2 5 Vpp 1 kHz CL 0582 2 5 Vpp sine wave 50 kHz SCOPE levsine 2 5 Vpp 50 kHz CL 0583 2 5 Vpp sine wave SCOPE levsine 2 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz CL 0482 500 mVpp sine wave 50 kHz SCOPE levsine 500 mVpp 50 kHz CL 0483 0 5 Vpp sine wave SCOPE levsine 0 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz Fluke 190 202 204 221 MHz 221 MHz Fluke 190 102 104 141 MHz 141 MHz Fluke 190 062 91 MHz 91 MHz CL 0462 100 mVpp sine wave 50 kHz SCOPE levsine 100 mVpp 50 kHz CL 0463 100 mVpp sine wave SCOPE levsine 100 mVpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz Fluke 190 202 204 381 MHz Fluke 190 102 104 151 MHz Fluke 190 062 111 MHz Fluke 190 202 204 221
3. REE e e E aici aratza 500 0 mV 5 000 V 50 00 V 500 0 V 1100 V Full Scale Reading 5000 counts ACCURACY a kee eae 0 5 6 counts Normal Mode AC Rejection at 50 0060 HZ Bunnen gt 60 dB AC Voltage VAC EEA IEA 500 0 mV 5 000 V 50 00 V 500 0 V 1100 V Full Scale Reading rra 5000 counts Accuracy 15 HZ10 60 aS GUI 10 counts 60 Hzto Aa 2 5 15 counts For higher frequencies the frequency roll off of the Meter input starts affecting accuracy Normal Mode DC Rejection rra gt 50 dB AC DC Voltage True RMS RANGES a e E e agerri 500 0 mV 5 000 V 50 00 V 500 0 V 1100 V Full Scale Reading rra 5000 counts Accuracy DO 10 60 EA 1 10 counts GONZO BA a 2 5 15 counts For higher frequencies the frequency roll off of the Meter input starts affecting accuracy All accuracies are valid if the waveform amplitude is larger than 5 of full scale Amperes AMP With Optional Current Probe or Current Shunt RANQCS 2 a E E aa adarrari same as VDC VAC VAC DC Probe EUNA 100 uV A 1 mV A 10 mV A 100 mV A 1 V A 10 V A and 100 V A OGUETA died Janeen same as VDC VAC VAC DC add current probe or current shunt accuracy Recorder TrendPlot Meter or Scope Chart recorder that plots a graph of min and max values of Meter or Scope measurements over time Measurement Speed gt 5 measurements s DIBI ak Eee 5 s div to 30 min div Record Size min max average 19200 points Recorded Time Span
4. Set the 5502A in operate OPR or standby STBY as indicated Press Fs to start the calibration Wait until the display shows calibration status READY Press SS to select the next calibration step set the 5502A to the next calibration point signal and start the calibration Continue through all calibration points of Table 5 13 and Table 5 14 When you are finished set the 5502A to Standby Continue at the nput D LF HF Gain section 5 33 Fluke 190 062 102 104 202 204 502 504 Service Manual Table 5 13 Input C LF HF Gain Calibration Points Cal step UUT input signal 5502A Setting CL 0656 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0484 0 62 Vpp sine wave 50 kHz SCOPE levsine 0 62 Vpp 50 kHz CL 0485 0 62 Vpp sine wave Fluke 190 204 Fluke 190 104 221 MHz 121 MHz SCOPE levsine 0 62 Vpp 221 MHz 121 MHz Table 5 14 Input C LF HF Gain Calibration Points Extra steps for 190 104 204 V10 4x firmware Cal step UUT input signal 5502A Setting CL 0464 124 mVpp sine wave 50 kHz SCOPE levsine 124 mVpp 50 kHz CL 0465 124 mVpp sine wave SCOPE levsine 124 mVpp Fluke 190 204 Fluke 190 104 221 MHz 121 MHz 221 MHz 121 MHz Four channel instruments 190 104 204 with firmware V10 4x require the extra steps CL 464 and CL 465 Installed firmware version can be checked via key sequence USER key F3 VERSION amp CAL Software Ve
5. 1 O MO N O 0 FW N Select the Test Tool setup Press METER Press EJ to open the Measurement menu and select Continuity Connect the Test Tool to the 5502A as shown in Figure 4 13 Set the 5502A to 20 Q Use the 5502A COMP OFF mode Listen to hear that the beeper is on Set the 5502A to 80 Q Listen to hear that the beeper is off When you are finished set the 5502A to Standby Diode Test Function Test To do the diode test function 1 w Select the Test Tool setup Press METER Press RIINI to open the Measurement menu and select NI Diode Connect the Test Tool to the 5502A as shown in Figure 4 13 4 37 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 38 O ONO Set the 5502A to 1 kQ Use the 5502A COMP OFF mode Observe the main reading and check that it is within 0 4 V and 0 6 V Set the 5502A to 1 V de Observe the main reading and check that it is within 0 975 V and 1 025 V When you are finished set the 5502A to Standby Ohms Measurements Test To test the Ohms measurement accuracy 1 Connect the Test Tool to the 5502A as shown in Figure 4 14 2 CHANN METER USE STACKABLE TEST LEADS AS SUPPLIED WITH CALIBRATOR FLUKE 5502A CALIBRATOR N A o o a 9 perf ver z2 eps Figure 4 14 Test Meter Tool Input to 5502A Normal Output 4 Wire Select the Test Tool setup Press METER Press KIIN to open the Me
6. Fluke 062 102 104 202 204 502 504 Service Manual viii Chapter 1 Safety Instructions daua eela Safety Precautions rrura Caution and Warning Statement rra balea Li ion Battery ESk rss Safe Storage of Battery ESk Safe Use of Battery PacK rss Safe Transport of Battery Pack Safe Disposal of Battery Pack sa Fluke 062 102 104 202 204 502 504 Service Manual Safety Instructions Introduction Introduction Read these pages carefully before beginning to install and use the Test Tool This section contains information cautions and warnings that must be followed to ensure safe operation and to keep the Test Tool in a safe condition AA Warning To prevent possible electrical shock fire or personal injury do not service the Test Tool unless you are qualified to do so Service described in this manual is to be done only by qualified service personnel Safety Precautions For the correct and safe use of this Test Tool it is essential that both operating and service personnel follow generally accepted safety procedures in addition to the safety precautions specified in this manual Specific warning and caution statements where they apply will be found throughout the manual Where necessary the warning and caution statements and or symbols are marked on the Test Tool Caution and Warning Statements A Warning identifies hazardous conditions and procedures that are dangerous to the user A Caution
7. Specifications Safety 2 A Max Floating Voltage Fluke 190 xxx test tool or test tool VPS410 From any terminal to earth ground Between any terminal Fluke 190 xxx VPS510 From any terminal to earth ground Between any terminal Voltage ratings are given as working and as Vdc for DC applications 1000 V CAT III 600 V CAT IV 1000 V CAT III 600 V CAT IV 300 V CAT III arara 300 V CAT III voltage They should be read as Vac rms 50 60 Hz for AC sinewave applications MAX INPUT E Vrms VOLTAG 500 200 100 50 20 0 01 0 02 0 05 0 1 Safe Handling VOLTAGE Vrms 1000 500 200 100 50 30 20 0 2 0 5 1 2 5 10 20 50 100 FREQUENCY MHz hpp049 eps Max Input Voltage vs Frequency r gi CAT Ill CAT IV 0 010 02 0 05 01 0205 1 2 5 10 20 50 100 200 FREQUENCY kHz hpp50 Max Voltage between Scope References and between Scope References and Earth Ground 2 15 Fluke 190 062 102 104 202 204 502 504 Service Manual 10 1 Probe VPS410 Accuracy Probe accuracy when adjusted on the test tool DE 10 20 aa Se 1 GRA EAN Ala EA 2 1 MHZ 10 25 MA E 3 For higher frequencies the probe s roll off starts affecting the accuracy For further probe specifications see the instruction sheet supplied with the VPS4
8. The final calibration requires input conditions that are described in each step After a step starts steps that require the same input conditions are done automatically For example if you start calibration step CL 0850 the calibration can include step CL 0869 and at the end the display shows CL 0799 READY Note You must always start the Final Calibration at the first step See the Warming Up 2 Warm Up Final and ADC Timing section Starting at another step will make the calibration invalid If you do calibration step N for example step CL 0581 then return to a previous step for example step CL 0580 and then calibrate this step the complete final calibration becomes invalid then you must repeat the calibration starting at the Warming Up 2 Warm Up Final and ADC Timing section It is allowed to repeat a step that shows the status READY by pressing F3 o again Error messages Proceed as follows if an error message ERROR nnnn shows on the display during calibration e if nnnn lt 5000 check input signal and test leads and press 2 to repeat the current step e if nnnn 25000 check input signal and test leads and repeat the final calibration in the Warming Up 2 Warm Up Final and ADC Timing section If the error persists the Test Tool is defective Calibration Adjustment Final Calibration For v11 10 and later Warming Up 2 Warm Up Final and ADC Timing The Warming Up 2 step CL 0500 must be done with open in
9. USE STACKABLE TEST LEADS AS SUPPLIED WITH CALIBRATOR A METER B 2 CHANN METER 0 0 lt gt lt gt FLUKE 5502A CALIBRATOR Perf Ver X2 eps Figure 5 17 5502A NORMAL Output to Test Tool Banana Input 3 Set the 5502A to supply a DC voltage to the first calibration point in Table 5 19 Set the 5502A to operate OPR Press E to start the calibration Wait until the display shows calibration status READY N Oa gt Press Z to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points in Table 5 19 8 Set the 5502A to Standby and continue at the Multimeter DMM Numeric Zero section 5 40 CL 0840 CL 0849 CL 0841 CL 0842 CL 0843 CL 0844 Calibration Adjustment Final Calibration Firmware V10 9 and Lower Table 5 19 DMM Gain Calibration Points Cal step UUT input value 55024 NORMAL 500 mV 2 5 V 5V 50V set 5502A to OPR 500 V 1000 V Multimeter Numeric Zero Note The adjustment steps for the meter section are only for the models 190 062 190 102 and 190 202 To do the Multimeter DMM Zero calibration 1 2 on s o Press SS to select calibration adjustment step CL 0890 Short circuit the banana jack Meter inputs properly calibration includes Ohms zero Use a test lead as short as possible Press Fs to start the zero calibration Wait until the display s
10. test leads and accessories for mechanical damage and replace when damaged Remove all probes test leads and accessories that are not in use Always connect the power adapter first to the ac outlet before connecting it to the product Do not touch voltages gt 30 V ac rms 42 V ac peak or 60 V dc Do not connect the ground spring to voltages higher than 42 V peak 30 Vrms from earth ground When using the ground reference lead with any of the probes make sure that the black isolation sleeve is over the probe tip Do not apply more than the rated voltage between the terminals or between each terminal and earth ground Do not apply input voltages above the rating of the instrument Use caution when using 1 1 test leads because the probe tip voltage will be directly transmitted to the product Do not use exposed metal BNC or banana plug connectors Fluke offers cables with plastic safety designed BNC connectors suitable for the ScopeMeter Test Tool product Do not insert metal objects into connectors Fluke 062 102 104 202 204 502 504 Service Manual 1 6 Use the product only as specified or the protection supplied by the product can be compromised Carefully read all instructions Do not use the product if it operates incorrectly Do not use the product or its accessories in case of any damage Disable the product or its accessories in case of any damage Keep fingers behind the finger
11. 1 1 c Use to change the time base to select manual time base ranging and lock the time base on 10 ms div d Use d we D to move the trace to the left Once the trigger point is shifted across the left hand border of the display going off display the trigger delay time with respect to the first vertical grid line will be indicated in the lower right of the display see Figure 4 7 Adjust the trigger delay time to 8 000 ms AJ 8 00 ms e Use to set the time base on 10 us div Use d we b to move the trace to the right until the indicated trigger delay is 7 940 ms Examine the rising edge of the time marker pulse at the height of the trigger level indicator top Verify that the rising edge is at the center grid line The allowed deviation is 3 pixels see Figure 4 7 172 AUTO INPUT A COUPLING PROBE A INPUT A OFF AC 1 1 OPTIONS 190c tb3 bmp Figure 4 7 Time Base Verification Input A Trigger Sensitivity Test Proceed as follows to test the Input A trigger sensitivity 1 2 3 Connect the Test Tool to the 5502A as shown in Figure 4 6 Reset the Test Tool to select the Test Tool setup Use EN and es to change the sensitivity range to select manual sensitivity ranging and lock the Input A sensitivity range on 2 V div Use me to select the time base indicated under the second column of Table 4 9 4 21 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 Set the 5502A to source
12. 102 104 202 204 502 504 Service Manual 6 Wait until the display shows calibration status READY 7 Press Eri to select the next calibration step set the 5502A to the next calibration point signal and start the calibration Continue through all calibration points of Table 5 9 and Table 5 10 When you are finished set the 5502A to Standby Continue at the nput B LF HF Gain section Table 5 9 Input A LF HF Gain Calibration Points Cal step UUT input signal 5502A Setting CL 0654 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0480 0 62 Vpp sine wave 50 kHz SCOPE levsine 0 62 Vpp 50 kHz CL 0481 0 62 Vpp sine wave SCOPE levsine 0 62 Vpp Fluke 190 202 204 221 MHz 221 MHz Fluke 190 102 151 MHz 151 MHz Fluke 190 104 121 MHz 121 MHz Fluke 190 062 81 MHz 81 MHz Table 5 10 Input A LF HF Gain Calibration Points Extra steps for 190 104 204 V10 4x firmware Cal step UUT input signal 5502A Setting CL 0460 124 mVpp sine wave 50 kHz SCOPE levsine 124 mVpp 50 kHz CL 0461 124 mVpp sine wave SCOPE levsine 124 mVpp Fluke 190 204 221 MHz 221 MHz Fluke 190 104 121 MHz 121 MHz Four channel instruments 190 104 204 with firmware V10 4x require the extra steps CL 460 and CL 461 Installed firmware version can be checked via key sequence USER key F3 VERSION amp CAL Software Version 5 30 Calibration Adjustment Final Calibration Firmware V10 9 and Lower Input B LF
13. 190 204 and 190 504 Proceed as follows to test the Input D high frequency automatic scope ac voltage measurement accuracy and the bandwidth Connect the Test Tool to the 5502A as shown in Figure 4 10 Recall the setup see Standard Test Tool Setup to select the Test Tool setup Press ane IN RECALL and select SETUP press GH select the setup name and press to recall the setup Press 888 then press EM READING and select READINGS 4 on D V ac MANUAL AUTO Press Use l 4 and EX to select autoranging AUTO in upper right LCD edge to change the sensitivity range to select manual sensitivity ranging and lock the Input D sensitivity range on 500 mV div Use it R Press Lil to move the Input D trace zero to the center grid line J and use EINI to select Input D as trigger source Set the 5502A to source a sine wave and to the first test point in Table 4 16 Observe the Input D reading and check that it is within the range shown under the appropriate column of Table 4 16 11 Continue through the test points 12 When you are finished set the 5502A to Standby Performance Verification 4 Scope Input A B C D Tests Table 4 16 HF AC Voltage Verification Points 5500A SC MODE levsine UUT 2 545 Vpp 1 MHz 835 mV to 965 mV 2 545 Vpp 25 MHz 790 mV to 1 010 V 190 104 2 545 Vpp 100 MHz gt 630 mV 190 204 2 545 Vpp 200 MHz gt 630 mV 190 504 2
14. 8 Press 2 again to do the next Contrast CL 0110 9 Press ENIN CALIBRATE The Test Tool shows the display test pattern at default contrast see Figure 4 2 Observe the display closely Verify that the display shows no abnormalities and that the contrast of the upper left and upper right sguare of the test pattern is egual Fluke 190 062 102 104 202 204 502 504 Service Manual 10 Press Foe The test pattern is removed and the Test Tool shows Contrast CL 0110 11 Press F2 again to do the next step Contrast CL 0120 12 Press EZE CALIBRATE The Test Tool shows a light display The test pattern shown in Figure 4 2 may not be visible or hardly visible Observe the display closely and verify that the display shows no abnormalities 13 Turn off and turn on the Test Tool to exit the calibration menu and return to the normal operating mode If the maximum minimum or default display contrast is not OK then you can adjust these items without performing a complete calibration adjustment refer to Section 5 for detailed information Scope Input A B C D Tests Input A B C D Vertical Accuracy Test 4 8 AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Note
15. 8 When you are finished set the 5502A to O zero Volt and Standby Table 4 7 Input A B C D Frequency Measurement Accuracy Test 190 062 20 ns div levsine 600 mVpp 60 MHz 59 68 to 60 32 190 104 190 102 190 204 190 202 20 ns div 600 mVpp 200 MHz 198 8 to 201 2 190 502 190 504 20 ns div 600 mVpp 500MHz 497 3 to 502 7 Note Because Duty Cycle and Pulse Width measurements are based on the same principles as Frequency measurements these measurement functions will not be verified separately 20 ns div levsine 600 mVpp 100 MHz 99 3 to 100 7 Input A amp B C amp D Phase Measurements Test Note The test steps for channel C and D are only for the models 190 104 190 204 and 190 504 Proceed as follows to test the phase measurement accuracy 1 Connect the Test Tool to the 5502A as shown in Figure 4 5 2 Select the Test Tool setup d Recall the created setup see Standard Test Tool Setup Press ane RECALL and select SETUP press Gi select the setup name and press to recall the setup e Press score f Press READING and select with KIIN READINGS and EC Reading 1 on A Phase Reading 2 on B Phase Reading 3 on C Phase Reading 4 on D Phase 3 Use kt N to select range 100 mV div for A B C and D 4 Use Ito select the reguired time base setting 4 19 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 Set the 5502A to source a s
16. AC Coupled MR Nela lt 2 Hz 3 dB direct T EEE lt 5 Hz 3 dB Rise Time Fluke 19080 rrera ra 0 7 ns Fluke 190 2808080888 1 7 ns Fluke 19030000 3 5 ns Fluk 190 062 agu gr EE e 5 8 ns Analog Bandwidth Limiters 20 MHz 20 kHz and 10 kHz varies according to version Input COUpIING gn EEk AC DC BASEA Normal Inverted Sensitivity Ranges with 10 dusu 20 mV to 1000 V div direct A BAA 2 mV to 100 V div Dynamic Range gt 8 div lt 10 MHz gt 4 div gt 10 MHz Waveform Positioning Ranqo rr 4 divisions Input Impedance on BNC DC Coupled 4 channel models rss 1 MQ 41 14 pF 2 25 pF 2 channel rmen 1 MQ 1 15 pF 2 25 pF Vertical MOGUTZEN rra 2 1 0 04 range div 2 mV div 2 9 0 08 range div Digitizer Resolution rra 8 bits separate digitizer for each input Fluke 190 062 102 104 202 204 502 504 Service Manual Horizontal Minimum Time Base Speed Scope RecorJJ 2 min div Real Time Sampling Rate Fluke 190 50x 5 ns to 4 us div 3 or 4 channels up to 1 25 GS s 2 ns to 4 us div 2 channels rrua up to 2 5 GS s 1 ns to 4 us div 1 channel Up to 5 GS s 10 15 10 120 S dV minien 125 MS s Fluke 190 202 204 2 ns to 4 us div 1 or 2 channels up to 2 5 GS s 5 ns to 4 us div 3 or 4 channels up to 1 25 GS s 10 us to KEAK GA 125 MS s Fluke 190 102 104 5 ns to 4 us d
17. EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA ARISING FROM ANY CAUSE OR THEORY Since some countries or states do not allow limitation of the term of an implied warranty or exclusion or limitation of incidental or consequential damages the limitations and exclusions of this warranty may not apply to every buyer If any provision of this Warranty is held invalid or unenforceable by a court or other decision maker of competent jurisdiction such holding will not affect the validity or enforceability of any other provision Fluke Corporation Fluke Europe B V P O Box 9090 P O Box 1186 Everett WA 98206 9090 5602 BD Eindhoven U S A The Netherlands 11 99 Table of Contents Chapter Title 1 Safety Instructions uses indiske HMTFOCUCHON AAA Safety Precautions rra Caution and Warning Statements rss Li ion Battery ESk rss Safe Storage of Battery Pack rss Safe Use of Battery Pack rss Safe Transport of Battery ESk Safe Disposal of Battery Pack rrua 2 iee baa Ta 11010 U 0 0 EE Oscilloseggg Automatic Scope Measurements rra Meter Measurements for Fluke IGOR Meter Measurements for Fluke IGOR ra IRGCOFAOF A Zoom Replay and Cursors rra Julie d Elaia UEA Environmental EAEE Certifications rra ZUALA 10 1 Probe VPS410
18. Error xxxx Calibration adjustment failed due to wrong input signal s or because the Test Tool is defective If the error code is lt 5000 you can repeat the failed step If the error code is 25000 you must repeat the complete final calibration start at Warming Up 2 Warm Up Final and ADC Timing The functions of the keys are Fi PREVIOUS select the previous step A NEXT select the next step Ea CALIBRATE start the calibration adjustment of the actual step ra EXIT leave the calibration mode Contrast Calibration Adjustment After you enter the calibration mode the display shows WarmingUp CL 0200 IDLE valid Do not press Fs If you do turn off and turn on the Test Tool and enter the calibration mode again To adjust the maximum display darkness CL 0100 the default contrast CL 0110 and the maximum display brightness CL 0120 5 7 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 8 10 11 12 13 Press Era three times to select maximum darkness calibration Contrast CL 0100 Press Er CALIBRATE The display shows a dark test pattern see Figure 5 2 Use d to adjust the display to the maximum darkness where the test pattern is only just visible Press F3 to return to the softkey menu Press F2 to select default contrast calibration Contrast CL 0110 Press E CALIBRATE The test pattern shows on the display at default contrast Use dB to set the display to optimal this s
19. Fluke 190 062 91 MHz 91 MHz CL 0460 100 mVpp sine wave 50 kHz SCOPE levsine 100 mVpp 50 kHz CL 0461 100 mVpp sine wave SCOPE levsine 100 mVpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz Fluke 190 202 204 381 MHz Fluke 190 102 104 151 MHz Fluke 190 062 111 MHz Fluke 190 202 204 221 MHz Fluke 190 102 104 141 MHz Fluke 190 062 91 MHz 501 MHz 381 MHz 151 MHz 111 MHz 221 MHz 141 MHz 91 MHz Fluke 190 062 102 104 202 204 502 504 Service Manual 5 14 Input B LF HF Gain To do the Input B LF HF Gain calibration w N Oo gt Press to select the first calibration step in Table 5 2 Connect Ch B of the Test Tool to the 5502A as shown in Figure 5 4 Set the 5502A SCOPE output to source the signal reguired for the first calibration point in Table 5 2 CL 0674 Pos B Fast Set the 5502A to operate OPR or standby STBY as indicated Press to start the calibration Wait until the display shows calibration status READY Press to select the next calibration step set the 5502A to the next calibration point signal and start the calibration Continue through all calibration points of Table 5 2 When you are finished set the 5502A to Standby Continue at the Input C LF HF Gain section 50 OHM e g PM9585 or TRM50 For 190 502 and 504 use only TRM50 PM9091 PM9091 0 OHM e g PM9585 or M50 USE 50 OHM TERMINATION
20. HF Gain To do the Input B LF HF Gain calibration 1 Press ETT to select the first calibration step in Table 5 11 2 Connect the Test Tool to the 5502A as shown in Figure 5 12 PM9091 PM9091 50 OHM e g PM9585 or C TRM50 m ad For 190 502 and 504 B 0 OHM e g PM9585 or RM50 use only TRM50 USE 50 OHM TERMINATION A METER 2 CHANN METER 4 CHANNELS FLUKE 5502A CALIBRATOR Perf Ver E2 eps Figure 5 12 5502A SCOPE Output to Test Tool Input B 3 Setthe 5502A SCOPE output to source the signal reguired for the first calibration point in Table 5 11 Set the 5502A to operate OPR or standby STBY as indicated Press to start the calibration Wait until the display shows calibration status READY N Oa gt Press ENIN to select the next calibration step set the 5502A to the next calibration point signal and start the calibration Continue through all calibration points of Table 5 11 and Table 5 12 When you are finished set the 5502A to Standby Continue at the Input C LF HF Gain section 5 31 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 32 Table 5 11 Input B LF HF Gain Calibration Points Cal step UUT input signal 5502A Setting CL 0674 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0482 0 62 Vpp sine wave 50 kHz SCOPE levsine 0 62 Vpp 50 kHz CL 0483 0 62 Vpp sine wave SCOPE levsine 0 62
21. Input A essa Time Base Verification ss 5502A Scope Output to Test Tool Input B rss 5502A Scope Output to Test Tool Input C rrua 5502A Scope Output to Test Tool Input D sa Test Tool Input A to TV Signal Generator rra SC600 Marker POE Test Tool Meter Ext Input to 5502A Normal Output rss Test Meter Tool Input to 5502A Normal Output 4 Wire Prob Calibration EEA Version and Calibration Data example rrrrnrrnnnrnnrrnnnnrrnnrnnnrrnnnnnrrnnnnenrrnn nen 5 4 Display Test Pattern rrura 5502A SCOPE Output to Test Tool Input A rra 5502A SCOPE Output to Test Tool Input B rss 5502A SCOPE Output to Test Tool TGG 5502A SCOPE Output to Test Tool Input D rss Test Tool Input ABCD to 5502A SCOPE OODDUL rrura Test Tool Input AB to 5502A Normal Output rss 5502A NORMAL Output to Test Tool Banana Input rrura Four wire Ohms Calibration Connections rra 5502A SCOPE Output to Test Tool Input A rrura 5502A SCOPE Output to Test Tool Input B ss 5502A SCOPE Output to Test Tool Input C su 15 33 5502A SCOPE Output to Test Tool Input D rss Test Tool Input ABCD to 5502A SCOPE Output rra Test Tool Input ABCD to 5502A Normal Output rrura 5502A NORMAL Output to Test Tool Banana Input rrura Four wire Ohms Calibration Connections rra 10 1 Probe Calibration Connection rs 10 1 Probe Calibration rr Flat Cable Connector sa Opened Case and SCrews rrua Screening Plate Removed and Screws ss 6 8 PCA removed from Chassis Bottom Side ulie rss
22. MHz Fluke 190 102 104 141 MHz Fluke 190 062 91 MHz 501 MHz 381 MHz 151 MHz 111 MHz 221 MHz 141 MHz 91 MHz Fluke 190 062 102 104 202 204 502 504 Service Manual Input C LF HF Gain Sections nput C LF HF Gain and Input D LF HF Gain are for 4 channel ScopeMeters 190 104 190 204 and 190 504 models For 2 channel models proceed to the nput AB Position section To do the Input C LF HF Gain calibration on 190 104 190 204 and 190 504 models 1 Connect Ch C of the Test Tool to the 5502A as shown in Figure 5 5 2 The display must show step CL 0694 Pos C Fast If it does not then press or to select the first calibration step in Table 5 3 PM9585 50 OHM For 190 502 and 504 use only TRM50 USE 50 OHM TERMINATION FLUKE 5502A CALIBRATOR Perf Ver F2 eps Figure 5 5 5502A SCOPE Output to Test Tool Input C 3 Setthe 5502A SCOPE output to source the signal reguired for the first calibration point in Table 5 3 Set the 5502A to operate OPR or standby STBY as indicated Press to start the calibration Wait until the display shows calibration status READY Press ENIN to select the next calibration step set the 5502A to the next calibration point signal and start the calibration N Oa gt Continue through all calibration points of Table 5 3 When you are finished set the 5502A to Standby Continue at the nput D LF HF Gain section Calibration A
23. Multi Product Calibrator including SC600 Oscilloscope Calibration Option Stackable test leads 4x as supplied with the 5502A reguired for Test Tools with banana jacks and 2 BNC oscilloscope inputs 50 Q Coax Cables 4x use Fluke PM9091 1 5 m 3 pcs and PM9092 0 5 m 3 pcs For Test Tools with banana jacks and 2 BNC oscilloscope inputs 2 Coax Cables are sufficient 50 Q feed through termination Fluke TRM50 4x for Test Tools with 4 BNC oscilloscope inputs 2x for Test Tools with banana jacks and 2 BNC oscilloscope inputs The use of model TRM50 is mandatory for Fluke 190 502 Male BNC to Dual Female BNC adapter 3x Fluke PM9093 001 Dual Banana Plug to Female BNC Adapter 1x Fluke PM9081 001 Calibration Procedure Steps To do a complete calibration adjustment you must do all steps 1 5 6 Select the Calibration Mode 2 Do the Contrast Calibration Adjustment 3 4 Do the Final Calibration for V11 10 and Later section or Final Calibration Do the Warming Up and Pre Calibration Firmware V10 9 and Lower If the installed firmware is V09 00 V10 00 or V10 4x you must do the steps in the Final Calibration Firmware V10 9 And Lower section If the installed firmware is V11 10 you must do the steps in the Final Calibration For V11 10 and later section Save the Calibration Data and Exit the calibration mode Do the probe calibration The following partial calibrations are allowed Contrast calib
24. SCOPE levsine 2 5 Vpp 50 kHz CL 0587 2 5 Vpp sine wave SCOPE levsine 2 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz CL 0486 500 mVpp sine wave 50 kHz SCOPE levsine 500 mVpp 50 kHz CL 0487 0 5 Vpp sine wave SCOPE levsine 0 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz Fluke 190 202 204 221 MHz 221 MHz Fluke 190 102 104 141 MHz 141 MHz Fluke 190 062 91 MHz 91 MHz CL 0466 100 mVpp sine wave 50 kHz SCOPE levsine 100 mVpp 50 kHz CL 0467 100 mVpp sine wave SCOPE levsine 100 mVpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz Fluke 190 202 204 381 MHz Fluke 190 102 104 151 MHz Fluke 190 062 111 MHz Fluke 190 202 204 221 MHz Fluke 190 102 104 141 MHz Fluke 190 062 91 MHz 501 MHz 381 MHz 151 MHz 111 MHz 221 MHz 141 MHz 91 MHz Fluke 190 062 102 Service Manual 104 202 204 502 504 Input AB Position To do the Input AB Position calibration E d oO N gt Press to select calibration adjustment step CL 0637 Pos AB Remove all Input A B connections open inputs Press to start the calibration Wait until the display shows calibration status READY Continue at the Input AB LF HF Gain
25. Screw 6x 10 5 mm to 3 3112 fix Main PCA Module to Front Case Side Strap 3945370 Can be fixed on Left or Right side Hang Strap 946769 Can be fixed on Top Side of EA Instrument Bottom holster assy 3 1 7 Bottom Holster Set 3981867 Cover for USB 3 1 9 Cover for DC adapter power 3 1 10 Keypad 4 channels 3942805 3 3 6 Keypad 2 ch meter 4035336 3 3 6 Keypad Foil all models 3942810 Incl Flat Cable 3 3 9 USB cable 3945381 USB A to mini USB B for PC connection BNC Connector Red 1 500 MHz 4306959 X1100 BNC Connector Blue 1 500 MHz 4306967 X1300 BNC Connector Gray 1 500 MHz 4306971 X1400 BNC Connector Green AI 500 MHz 4306980 X1200 BNC Connector Black 500 MHz 4306998 BNC Connector Red 2 200 MHz 3945031 X1100 BNC Connector Green 2 200 MHz 3945068 X1200 BNC Connector Blue 2 200 MHz 3945046 X1300 BNC Connector Gray 2 200 MHz 3945054 X1400 Banana Jack Black 4035403 X 1501 Banana Jack Red 4035415 X 1500 DC Power Input Socket 215785 X9100 1 Later PCA sub code 5 2 Early PCA sub code 2 or firmware version lt 10 Check that the serial number is 225375604 If yes use the later connectors If lt 25375604 evaluate the firmware and subversion To find the version press in sequence USER and F3 VERSION amp CAL If the firmware version is lt V11 00 use the early connectors If the firmware is gt 11 10 check under subversions that the last datablock is x5x
26. The test steps for channels C and D are only for the models 190 104 190 204 or 190 504 Proceed as follows 1 Connect the Test Tool to the 5502A as shown in Figure 4 3 The vertical channels A B C and D are checked in succession so that there is one waveform on the display at a time to facilitate amplitude adjustment Performance Verification 4 Scope Input A B C D Tests PM9091 PM9091 CONNECT TO CHANNEL A B C D IN SUCCESSION NO 50 OHM TERMINATION ra A METER B 2 CHANN METER 4 CHANNELS FLUKE 5502A CALIBRATOR perf ver a2 eps Figure 4 3 Test Tool Inputs A B to 5502A Normal Output Select the following Test Tool setup SAVE a Recall the created setup see Standard Test Tool Setup Press RECALL and select SETUP press Gi select the setup name and press to recall the setup b Press HIN press ETZ INPUT A OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the Test Tool c Press MIEN press ETZ INPUT B OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the Test Tool d Press C press Fa INPUT C OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the Test Tool e Press D press Fa INPUT D OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the availa
27. affecting accuracy Normal Mode DC Roejection rra gt 50 dB All accuracies are valid if e The waveform amplitude is larger than one division e Atleast 1 5 waveform period is on the screen AC DC Voltage True RMS Maximum Voltage With 10 1 probe sisien 1000 V direct Milan 300 V Maximum Resolution with 10 1 duel 1 MV direct Mil a NS 100 uV Full Scale Reading rr 1100 counts Accuracy Fluke 190 xx2 DC 10 60 HZ en babe 1 5 10 counts 60 Hz to 20 kHz 2 5 15 counts 20 KHZ 101 5 20 counts ra 10 20 counts For higher frequencies the instrument s frequency roll off starts affecting accuracy Accuracy Fluke 190 xx4 DC to 60 Hz GIO 10 counts 60 HZ 10 20 Aa 4 15 counts Sage Bauke 6 20 counts 1 MH2 10 25 A a 10 20 counts For higher frequencies the instrument s frequency roll off starts affecting accuracy Amperes AMP With Optional Current Probe or Current Shunt RANGES ea E een Seed same as VDC VAC VAC DC Probe Sensitivity sr 100 uV A 1 mV A 10 mV A 100 mV A 400 mV A 1 V A 10 V A and 100 V A OORT i tetas arched iene terrera same as VDC VAC VAC DC add current probe or current shunt accuracy 2 2 7 Fluke 190 062 102 104 202 204 502 504 Service Manual 2 8 Peak MOOS aka ara ta E ede Max peak Min peak or peak to peak Maximum Voltage with 10 71 beie 1000 V direct I eurae E E eds ered 300 V Maximum Resolution w
28. and Position Calibration Points Cal step UUT input value 5502A SCOPE MODE volt SCOPE Z 1 MQ 1 kHz CL 0504 500 mV 1 kHz MODE VOLT CL 0624 Open inputs Calibrator STBY CL 0673 50 V 1 kHz MODE volt CL 0660 300 mV CL 0604 500 mV CL 0637 none 5500 standby CL 0504 500 mV CL 0624 none 5500 standby CL 0599 10 mV CL 0600 25 mV CL 0601 50 mV CL 0602 100 mV CL 0622 none 5500 standby CL 0603 250 mV CL 0662 2V CL 0605 1V CL 0606 2 5 V CL 0607 5y CL 0664 20 V CL 0608 10V CL 0609 25V CL 0610 50 V set 5502A to OPR 5 21 Fluke 190 062 102 104 202 204 502 504 Service Manual Input Pos ABCD AB Calibration To do the Input Pos AB calibration 1 Press to select calibration adjustment step CL 0619 Remove all Input A B connections Calibrator STBY Press to start the calibration Wait until the display shows calibration status READY Continue at the Input ABCD AB NoiseF FBW Calibration section ARE N Input ABCD AB Noise E FBW Calibration To do the Input AB NoiseF FBW Calibration Press to select calibration adjustment step CL 0850 Connect 50 Q feed through terminations to all BNC Inputs A B C and D Press to start the calibration Wait until the display shows calibration status READY Continue at the nput AB Volt Gain section E PE N gt Input AB Volt Gain AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is
29. case and move the stand away from the housing There is no need to remove screws or other fixing devices Before opening the Test Tool you must remove the hang strap and the side strap How to remove and install the hang strap is explained in the Users Manual in the Tips Chapter The grip of the side strap consists of two halves kept together with Velcro tape After having opened it the straps can be taken apart and be removed from their fixing dowels in the side of the Test Tool Before you do this take careful notice on the correct position of the strap To install work in reverse order Fluke 190 062 102 104 202 204 502 504 Service Manual Open the Test Tool Remove the Battery Pack Proceed as follows Turn the plastic battery door screws one quarter turn counterclockwise with a standard blade screwdriver 2 Remove the battery access door 3 Remove the battery from the instrument Note Do not short circuit the battery contacts Do not open or damage the battery housing If attached remove the hang strap and the side strap see the Removing the Tilt Stand Hang Strap and Side Strap section Loosen the two black self tapping screws that fasten the grey yellow input cover around the BNC input and banana sockets Remove the cover Note When reinstalling the input cover do not forget to reinstall the flexible sealing strip around the input sockets The holes in this strip have a flat side that mus
30. coupled low frequencies 50 Hz ditet T A 60 Hz direct 1 1 ra A B C and D gt 100 dB gt 60 dB 1000 V 300 V 1 MV 100 uV 999 counts 1 5 10 counts 1 5 6 counts 3 10 counts 3 6 counts gt 60 dB 1000 V 300 V 1 MV 100 uV 999 counts 1 5 10 counts 1 5 10 counts 0 6 1 5 10 counts 0 4 With the 10 1 probe the low freguency roll off point will be lowered to 2 Hz which improves the AC accuracy for low freguencies When possible use DC coupling for maximum accuracy AC or DC coupled high freguencies GO HZ 1020 Aa EA 20 KHZ to 1 MHZ sr 1 MHz to 25 MEZ EO 2 5 15 counts rra 5 20 counts ee 10 20 counts For higher frequencies the instrument s frequency roll off starts affecting accuracy Accuracy Fluke 190 xx4 DC coupled DC 10 60 Ea EAEE 3 10 counts AC coupled low frequencies Ga Palak BAN 3 10 counts 0 6 60 Hz direct GIE GIO 10 counts 0 4 Specifications Automatic Scope Measurements With the 10 1 probe the low frequency roll off point will be lowered to 2 Hz which improves the AC accuracy for low frequencies When possible use DC coupling for maximum accuracy AC or DC coupled high frequencies 60 HZ 10 20 Aa EAE 4 15 counts 20 KH EA to BAL Ea 6 20 counts 1 MHz to 25 MHz 10 20 counts For higher freguencies the instrument s freguency roll off starts
31. fluke com In the event that the part ordered has been replaced by a new or improved part the replacement will be accompanied by an explanatory note and installation instructions if necessary To ensure prompt delivery of the correct part include the following information when you place an order e Instrument model for example Fluke 190 502 and serial number 25530001 that is printed on the type plate on the bottom cover e Ordering code e Item number Reference designator e Description e Quantity 3 3 Fluke 190 062 102 104 202 204 502 504 Service Manual Final Assembly Parts See Table 3 1 and Figure 3 1 3 2 and 3 3 for the Final Assembly parts A Caution The Test Tool contains a Li ion battery Do not mix with the solid waste stream Spent batteries should be disposed of by a qualified recycler or hazardous materials handler Table 3 1 Final Assembly Parts and Kits Part or Kit Ordering Code Consists of Following Parts Front case excluding lens decal 4 Figure ltem nr 3 3 5 channel Dust seal long 2x 3 3 3 Dust seal short 2x 3 3 4 Case seal 3 3 13 Case Set 4 channel 3981815 Bottom case assy 3 13 Battery door 3 1 14 Ouarter turn screw 2x 3 1 15 Adhesive foam for battery door Standup bracket 3 1 16 Front case Excl lens decal 2 3 3 5 channel Dustseal long 2x 3 3 3 Dustsea
32. guards on the probes Use only correct measurement category CAT voltage and current rated probes test leads and adapters for the measurement Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a product probe or accessory Do not use the product around explosive gas vapor or in damp or wet environments Measure a known voltage first to make sure that the product operates correctly Examine the case before you use the product Look for cracks or missing plastic Carefully look at the insulation around the terminals Do not work alone Comply with local and national safety codes Use personal protective equipment approved rubber gloves face protection and flame resistant clothes to prevent shock and arc blast injury where hazardous live conductors are exposed The battery door must be closed and locked before you operate the product Do not operate the product with covers removed or the case open Hazardous voltage exposure is possible Remove the input signals before you clean the product Use only specified replacement parts Use of the product in a manner not specified may impair the protection provided by the equipment Do not use test leads if they are damaged Examine the test leads for damaged insulation exposed metal or if the wear indicator shows Safety Instructions Li ion Battery Pack Voltage ratings that are mentioned in the warnings are given a
33. in Table 5 1 Set the 5502A in operate OPR or standby STBY as indicated Press I to start the calibration Wait until the display shows calibration status READY Press EH to select the next calibration step set the 5502A to the next calibration point signal and start the calibration Continue through all calibration points of Table 5 1 When you are finished set the 5502A to Standby Continue at the nput B LF HF Gain section Calibration Adjustment Final Calibration For v11 10 and later Table 5 1 Input A LF HF Gain Calibration Points UUT input signal 5502A Setting CL 0654 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0415 100 mVpp square wave 1 kHz SCOPE edge 100 mVpp 1 kHz CL 0510 2 5 Vpp square wave 1 kHz SCOPE edge 2 5 Vpp 1 kHz CL 0580 2 5 Vpp sine wave 50 kHz SCOPE levsine 2 5 Vpp 50 kHz CL 0581 2 5 Vpp sine wave SCOPE levsine 2 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz CL 0480 500 mVpp sine wave 50 kHz SCOPE levsine 500 mVpp 50 kHz CL 0481 0 5 Vpp sine wave SCOPE levsine 0 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz Fluke 190 202 204 221 MHz 221 MHz Fluke 190 102 104 141 MHz 141 MHz
34. in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps To do the Input AB Volt Gain calibration 1 Press to select the first calibration step in Table 5 6 2 Connect Ch A and B of the Test Tool to the 5502A NORMAL output as shown in Figure 5 8 PM9093 PM9092 CONNECT TO CHANNEL A B C D IN PARALLEL NO 50 OHM TERMINATION i A METER B 2 CHANN METER 070 FLUKE 5502A CALIBRATOR Perf Ver B2 eps Figure 5 8 Test Tool Input AB to 5502A Normal Output 3 Set the 5502A to supply a DC voltage NORMAL output to the first calibration point in Table 5 6 5 22 Calibration Adjustment 4 Set the 5502A to operate OPR 5 Press ENIN to start the calibration 6 Wait until the display shows calibration status READY 7 Press F2 to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points of Table 5 6 8 Set the 5502A to Standby and continue at the Multimeter DMM Meter Zero section Table 5 6 Input ABCD Gain Calibration Points Cal step UUT input value 5502A NORMAL dc output CL 0799 5 mV CL 0800 12 5 mV CL 0801 25 mV CL 0802 50 mV CL 0803 125 mV CL 0804 250 mV CL 0805 500 mV CL 0806 1 25 V CL 0807 2 5 V CL 0808 5V CL 0809 12 5 V CL 0810 2
35. levsine 0 62 Vpp 50 kHz CL 0487 0 62 Vpp sine wave SCOPE levsine 0 62 Vpp Fluke 190 204 221 MHz 221 MHz Fluke 190 104 121 MHz 121 MHz Table 5 16 Input D LF HF Gain Calibration Points Extra steps for 190 104 204 V10 4x firmware Cal step UUT input signal 5502A Setting CL 0466 124 mVpp sine wave 50 kHz SCOPE levsine 124 mVpp 50 kHz CL 0467 124 mVpp sine wave SCOPE levsine 124 mVpp Fluke 190 204 Fluke 190 104 221 MHz 121 MHz 221 MHz 121 MHz Four channel instruments 190 104 204 with firmware V10 4x require the extra steps CL 466 and CL 467 Installed firmware version can be checked via key sequence USER key F3 VERSION amp CAL Software Version Input ABCD AB LF HF Gain Note The adjustment steps for channel C and D are only for the models 190 104 and 190 204 To do the Input ABCD LF HF Gain calibration 1 Press ETT to select the first calibration step in Table 5 17 2 Connect the Test Tool to the 5502A as shown in Figure 5 15 PM9091 PM9092 CONNECT TO CHANNEL A B C D IN PARALLEL PM9093 NO 50 OHM TERMINATION E METER A B 2 CHANN METER 4 CHANNELS Gu 050 FLUKE 5502A CALIBRATOR Perf Ver C2 eps Figure 5 15 Test Tool Input ABCD to 5502A SCOPE Output 3 Set the 5502A to supply a 1 kHz square wave SCOPE MODE volt SCOPE Z 1 MQ to the first calibrat
36. of three battery packs may be shipped by mail The package must be marked as follows PACKAGE CONTAINS LITHIUM ION BATTERIES NO LITHIUM METAL Safe Disposal of Battery Pack e A failed battery pack shall be properly disposed of in accordance with local regulations e Properly dispose of the battery pack Do not dispose of the battery as unsorted municipal waste e Dispose in discharged condition and cover the battery terminals with isolation tape Fluke 062 102 104 202 204 502 504 Service Manual Chapter 2 Specifications Title Page IRTOA CTI OM EEA 2 3 OSGCIIIOSCOPE ua 2 3 Automatic Scope Measurements rra 2 6 Meter Measurements for Fluke 190 XX4 rss 2 9 Meter Measurements for Fluke 190 Xx2 rra 2 9 FRECONG ER sakea AA eh aes 2 11 Zoom Replay nd e edale 2 12 AUS Elaia Ea 2 12 Environmental rra 2 14 Os ties st erraitera 2 14 BALA 2 14 10 1 Probe VPS410 surra 2 16 Electromagnetic Immunity ra 2 16 Fluke 190 062 102 104 202 204 502 504 Service Manual Specifications 2 Introduction Introduction Specifications are subject to change without notification Oscilloscope Isolated Inputs A B C and D Vertical Number of Channels Fluke 05 EN t eei Fluke 190 xx4 Bandwidth DC Coupled Fluke 190 50 xX ut ons ante usta sve cents 500 MHz 3 dB SU ASA 200 MHz 3 dB Fluke 190 EA 100 MHz 3 dB SU aS BA RA 60 MHz 3 dB Lower Frequency Limit
37. section E d O N gt Input ABCD AB Volt Gain AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Note The adjustment steps for channel C and D are only for the models 190 104 and 190 204 To do the Input ABCD AB Volt Gain calibration 1 Press to select the first calibration step in Table 5 18 2 Connect the Test Tool to the 5502A as shown in Figure 5 16 LIL PM9093 PM9092 CONNECT TO CHANNELA B C D IN PARALLEL NO 50 OHM TERMINATION FLUKE 5502A CALIBRATOR Figure 5 16 Test Tool Input ABCD to 5502A Normal Output 3 Set the 5502A to supply a DC voltage NORMAL output to the first calibration point in Table 5 18 4 Set the 5502A to operate OPR Press to start the calibration 6 Wait until the display shows calibration status READY 5 38 Calibration Adjustment Final Calibration Firmware V10 9 and Lower 7 Press a to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points of Table 5 18 8 Set the 5502A to Standby and continue at the Input ABCD AB Zero section Table 5 18 Input ABCD Gain Calibration Points C
38. the calibration source and connection cables during these steps Note The test steps for channel C and D are only for the models 190 104 190 204 and 190 504 To test the peak measurement accuracy 1 2 Connect the Test Tool to the 5502A as shown Figure 4 4 Select the Test Tool setup a Recall the setup see Standard Test Tool Setup Press ane F2 RECALL and select SETUP press GEO select the setup name and press to recall the setup b Press EE c Press ENIN READING and select with ETZE READINGS and with Reading 1 on A Peak and next Peak Peak Reading 2 on B Peak and next Peak Peak Reading 3 on C Peak and next Peak Peak Reading 4 on D Peak and next Peak Peak d Press to clear the softkey menu and to see the full display Ito change the time base and lock the time base on 1 ms div E to set the Input A B C and D sensitivity ranges to Set the 5502A to source the appropriate ac voltage and frequency as listed in Table 4 6 Observe readings A B C and D and check that they are within the range shown under the appropriate column When you are finished set the 5502A to 0 zero Volt and Standby Table 4 6 Volts Peak Measurement Verification Points 5502A output Vrms sine 5502A Frequency Reading A B Fluke 190 062 102 104 202 204 502 504 Service Manual Input A B C D Frequency Measurement Accuracy Test Note The test steps
39. the leveled sine wave for the appropriate Test Tool model 6 Adjust the 5502A output voltage until the displayed trace has the trigger amplitude indicated under the last column of Table 4 9 7 Verify that the signal is well triggered If not press Lia J and use ENIN to enable EO for manual Trigger Level adjustment Adjust the trigger level and verify that the signal is triggered The trigger icon T indicates the trigger level 8 Continue through the test points 9 When you are finished set the 5502A to Standby Table 4 9 Input A Trigger Sensitivity Test Points 5502A SC MODE levsine UUT Model UUT Time base UUT Trigger Amplitude Initial Input Voltage All 200 ns div 100 mV pp 5 MHz 0 5 div 10 ns div 400 mV pp 60 MHz 1 div 190 062 pe 300 mV pp 100 MHz s ee 400 mV pp 100 MHz 190 104 10 ns div 800 mV pp 150 MHz 190 202 10 ns div 400 mV pp 200 MHz 190 204 10 ns div 800 mV pp 250 MHz 190 502 2 ns div 400 mV pp 500 MHz 1 div 190 504 2 ns div 800 mV pp 600 MHz 2 div 4 22 Performance Verification 4 Scope Input A B C D Tests Input A AC Voltage Accuracy HF and Bandwidth Test Proceed as follows to test the Input A high frequency automatic scope ac voltage measurement accuracy and the bandwidth 1 Connect the Test Tool to the 5502A as shown in Figure 4 6 2 Select the Test Tool setup a Recall the setup see Standard Test Tool Setup Press ane SS RECALL and select SETUP press Gi
40. the version and calibration data Figure 5 1 2 Press to close the version and calibration menu VERSION amp CALIBRATION BATTERY INFO Figure 5 1 Version and Calibration Data example Note The calibration date and calibration number do not change if you do only the Contrast Calibration Adjustment and or the Probe Calibration version calibration 3 bmp General Instructions Follow these general instructions for all calibration steps e Allow the specified warm up period for the 5502A For each calibration point wait for the 5502A to settle e The required warm up period for the Test Tool is included in the Warming Up and Pre Calibration step e Ensure that the Test Tool battery is charged sufficiently e Power the Test Tool with the BC190 Power Adapter e This procedure is for all models Test steps that are not applicable to the Test Tool to be adjusted can be skipped For example the adjustment of the meter with banana jacks can be skipped in instruments with four scope BNC inputs e The figures that show how to interconnect Signal Source and Test Tool are for 2 Scope Inputs Meter Input and for 4 Scope Inputs Calibration Adjustment Calibration Procedure Steps Equipment Required For Calibration The primary source instrument used in the calibration procedures is the Fluke 5502A If a 5502A is not available you can substitute another calibrator that meets the minimum test reguirements Fluke 5502A
41. to 158 139 to 161 50 V div 150 V 158 to 142 161 to 139 300 V 289 to 311 285 to 315 100 V div 300 V 311 to 289 315 to 285 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 14 Input A B C D AC Voltage Accuracy Test LF Note The test steps for channel C and D are only for the models 190 104 190 204 and 190 504 This procedure tests the Volts ac accuracy with dc coupled inputs up to 50 kHz The high frequencies are tested in sections Input A AC Voltage Accuracy HF amp Bandwidth Test and Input B AC Voltage Accuracy HF amp Bandwidth Test AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Proceed as follows to test the Input A B C and D automatic scope ac Voltage measurement accuracy 1 Connect the Test Tool to the 5502A as shown in Figure 4 4 2 Select the Test Tool setup a Recall the created setup Creating a Standard Test Tool Setup Press E N RECALL and select SETUP press GEO select the setup name and press to recall the setup b Press A then press Fa INPUT A OPTIONS c Select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the ScopeMeter 2 mV d and 5 mV d or Bandwid
42. until the displayed trace has the amplitude indicated under the appropriate column of Table 4 13 10 Verify that the bd is well triggered If not press Masses and use KNN to enable the O for manual Trigger Level Adjust the trigger level and verify that the signal is triggered The trigger icon 1 indicates the trigger level 11 Continue through the test points 12 When you are finished set the 5502A to Standby Table 4 13 Input C Trigger Sensitivity Test Points Time base Trigger Amplitude 190 104 10 ns div 800 mV pp 150 MHz 2 div 10 ns div 400 mV pp 200 MHz 1 div 190 204 10 ns div 800 mV pp 250 MHz 2 div 2 ns div 400 mV pp 500 MHz 1 div 190 504 2 ns div 800 mV pp 600 MHz 2 div Input C AC Voltage Accuracy HF and Bandwidth Test Note The test steps for channel C are only for the models 190 104 190 204 and 190 504 Proceed as follows to test the Input C high frequency automatic scope ac voltage measurement accuracy and bandwidth 1 Connect the Test Tool to the 5502A as shown in Figure 4 9 4 27 Fluke 190 062 102 104 202 204 502 504 Service Manual 2 Select the Test Tool setup a Recall the setup see Standard Test Tool Setup Press EO EN RECALL and select SETUP press D select the setup name and press to recall the setup b Press score then press F2 READING and select READINGS 3 on C Vac MANUAL AUTO c Press to select autorangi
43. 04 Four 100 MHz Scope Inputs BNC 190 202 Two 200 MHz Scope Inputs BNC one Meter Input banana jacks 190 204 Four 200 MHz Scope Inputs BNC 190 502 Two 500 MHz Scope Inputs BNC one Meter Input banana jacks 190 504 Four 500 MHz Scope Inputs BNC The Test Tool should be calibrated and in operating condition on arrival The following performance tests are provided to ensure that the Test Tool is in proper operating condition If the Test Tool fails any of the performance tests calibration adjustment see Chapter 5 and or repair is necessary The Performance Verification Procedure is based on the specification see Chapter 2 The values given here are valid for ambient temperatures between 18 C and 28 C The Performance Verification Procedure is a quick and efficient way to check all main specifications of the Test Tool Accuracy of Test Tool specifications not tested is linked to those tested in this verification procedure and is embedded in the Test Tool s software This link is tested extensively for each new software release 4 3 Fluke 190 062 102 104 202 204 502 504 Service Manual Equipment Requirements for Verification The primary source instrument used in the verification procedures is the Fluke 5502A If a 5502A is not available you can substitute another calibrator as long as it meets the minimum test requirements Fluke 5502A Multi Product Calibrator incl
44. 10 probe set Electromagnetic Immunity The Fluke 190 Series II test tools including standard accessories are compliant to EN 61326 1 with the addition of the table that follows Scope Mode 10 ms div Waveform disturbance with VPS410 voltage probe shorted see table below E 3V m i o i o Freguency No Disturbance Disturbance lt 10 of Disturbance gt 10 of full scale full scale 80 MHz 450 MHz 2500 mV d 100 200 mV div 2 5 10 20 50 mV div 1 4 GHz 2 GHz Alranges 1 2 GHz 27 GHz ivim AllRanges O oo S Chapter 3 Parts List Title Page INTFOGUCUION AAE 3 3 How to Obtain Pertza 3 3 Final Assembly Egita 3 4 Accessory EE AA 3 8 3 1 Fluke 190 062 102 104 202 204 502 504 Service Manual Parts List Introduction Introduction This section contains a list of replaceable parts for all the models of the Test Tool Parts are listed by assembly and alphabetized by item number or reference designator The figures show the location of each part and the item number or reference designator The parts list shows e Description e Ordering code A Caution Electrical components and in particular active components such as ICs transistors and diodes may be damaged by static discharge Only qualified personnel at a static free workstation should handle and service static sensitive components and assemblies How to Obtain Parts To locate an authorized service center go to www
45. 102 190 202 and 190 502 To test the external trigger level 1 Connect the Test Tool to the 5502A as shown in Figure 4 13 2 CHANN METER USE STACKABLE TEST LEADS AS SUPPLIED WITH CALIBRATOR FLUKE 5502A CALIBRATOR NQ ar amp N 8 9 10 11 12 13 14 15 perf ver x2 eps Figure 4 13 Test Tool Meter Ext Input to 5502A Normal Output Reset the Test Tool to select the Test Tool setup Press Use E to select the TRIGGER OPTIONS menu Select On Edges from the TRIGGER OPTIONS menu Press d Select Update Single Shot GD Trigger Filter Noise Reject MD NCycle of GD Use HIN EDGE TRIG to select Ext Use N SLOPE to select positive slope triggering trigger icon J Use EEE Ext LEVEL to select 1 2 V Set the 5502A to source 0 4 V dc Verify that no trace is shown on the Test Tool display and that the status line at the display top shows SINGLE MANUAL or SINGLE WAITING HOLD If the display shows the trace and status as SINGLE HOLD press Wibi to re arm the Test Tool for a trigger Set the 5502A to source 1 7 V To verify that the Test Tool is triggered check that the trace becomes visible To repeat the test start at step 3 Set the 5502A to Standby Performance Verification 4 Meter Tests Meter Tests Note The following tests are for the models 190 062 190 102 190 202 and 190 502 Meter DC Voltage Accuracy Test AA Warning To preve
46. 18 C to 28 C Add 0 1x specific accuracy for each C below 18 C or above 28 C Fluke 190 062 102 104 202 204 502 504 Service Manual Meter Input Banana Jacks Input Coupling rr DC Frequency Response DC to 10 kHz 3 dB lua 1 MQ 41 14 pF 41 5 pF Max Input Voltage ra 1000 V CAT III 600 V CAT IV For detailed specifications see Safety Meter Functions PREM GUN AA Auto Manual Aeta EAE Normal Relative General DC Common Mode Rejection CMRR gt 100 dB AC Common Mode Rejection at 50 60 Co ARNa EAE gt 60 dB Ohms Q PSOE a geen TE E EEEa 500 0 Q 5 000 KQ 50 00 kQ 500 0 KQ 5 000 MQ 30 00 MQ Full Scale Reading 500 O40 5 MQ ie cscs raa 5000 counts S0 MO SE a 3000 counts ACCURACY EEE 0 6 6 counts Measurement DUT 0 5 mA to 50 nA 20 decreases with increasing ranges Open Circuit Voltage rra lt 4V Continuity CONT B CD arketa rre d ate sarees lt 50 Q 30 Q Measurement Current rrua 0 5 mA 20 Detection of shorts Oir gt 1 ms Diode Maximum Voltage EGO 2 8V Open Circuit Voltage rra lt 4 V CCU ACY a E Sees tates E 2 5 counts Measurement Current rrua 0 5 MA 20 Temperature TEMP With Optional Temperature Probe Ranges ga la EA 40 0 to 100 0 100 0 to 250 0 100 0 to 500 0 100 to 1000 100 to 2500 BulaG ALE 1 mV C and 1 mV F Specifications 2 Recorder DC Voltage VDC
47. 200 div Max Delay ssticccscactincnsienndincandnancn 48 s at 4 s div Automatic Connect and View Trigger SOW da A B C D EXT 190 xx2 SPE Positive Negative Dual Specifications 2 Automatic Scope Measurements Edge Trigger Screen e ESO Free Run On Trigger Single Shot Source A B C D EXT 190 xx2 SIOPE mama E arab a Positive Negative Dual Trigger Level Control Range rr 4 divisions Trigger Sensitivity DC to 5 MHz at gt 5 V diV rra 0 5 division DC to 5 MHz at 2 mV div and OKO a kor rape ertai 1 division 500 MHz Fluke 190 500 rr 1 division 200 MHz Fluke 190 2XX rra 1 division 100 MHz Fluke 190 1XX rra 1 division 60 MHz Fluke 190 062 eee ences 1 division Isolated External Trigger 190 xx2 Bandwidth AAA 10 kHz 1610 1 EE Automatic Edge Trigger Levels DC to 10 KEZ rr 120 mV 1 2 V Video Trigger Standards e EEEE ASE NET PAL PAL NTSC SECAM Non interlaced Modes Lines Line Select Field 1 or Field 2 Sougffffeeee dike A Polarity e Etura E darrek Positive Negative 0 7 division sync level Sensitivity Pulse Width Trigger Screen eka BA On Trigger Single Shot Trigger Conditions rra lt T gt T T 410 T 10 Souffe arata A Polarity Positive or negative pulse Pulse Time Adjustment Range rra 0 01 div to 655 div with a minimum of 300 ns lt T gt T or 500 ns T T a maximum of 10 s and a resolution of 0 01 div with a minimum of 50 ns
48. 5 V CL 0811 50 V set 5502A to OPR CL 0812 125 V CL 0813 250 V Final Calibration For v11 10 and later 5 23 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 24 Multimeter Meter Zero Note The adjustment steps for the meter section are only for the models 190 062 190 102 and 190 202 For 4 Channel ScopeMeters go to the Save Calibration Data and Exit section Proceed as follows to do the Multimeter DMM Zero calibration Press to select calibration adjustment step CL 0890 Short circuit interconnect the banana jack Meter inputs Use a test lead as short as possible Press to start the zero calibration Wait until the display shows the status READY Remove the input terminations Continue at the Multimeter DMM Volt Gain section oa Fw Multimeter Volt Gain AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Note The adjustment steps for the meter section are only for the models 190 062 190 102 190 202 and 190 502 To do the DMM Volt Gain calibration 1 Press to select the first calibration step in Table 5 7 2 Connect the Test Tool to the 5502A as shown in Figure 5 9 Calibration Adjustment Final Calibration For v11 10 and later
49. 545 Vpp 500 MHz gt 630 mV Video Test with SC600 Scope Calibration Option Only one of the video systems NTSC PAL PALplus or SECAM has to be verified To verify 1 Connect the Test Tool to the calibrator as shown in Figure 4 11 PM9091 PM9091 50 OHM e g PM9585 or M50 d 4 Es For 190 502 and 504 only use TRM50 METER USE 50 OHM TERMINATION B A 2 CHANN METER 4 CHANNELS FLUKE 5502A CALIBRATOR perf ver d2 eps Figure 4 11 Test Tool Input A to TV Signal Generator 2 Resetthe Test Tool to select the Test Tool setup Press li J and then press HEO to open the Trigger Options menu Choose VIDEO on A and choose from the shown menu Polarity POSITIVE PAL or NTSC or PALplus or SECAM 5 Press F2 to select ALL LINES 6 Press F3 to enable to select the video line number 7 Use EO to select line number e 622 for PAL PALplus or SECAM e 525 for NTSC to move the Input A trace zero to the center grid line 4 31 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 32 10 11 12 13 14 Use TZ and E setting is 10 1 to set the Input A sensitivity to 2 V div the actual probe Use J to select the time base to 20 us div Set the calibrator to video mode with amplitude 100 Set format and marker line number to e PAL 622 even for PAL and PALpl
50. 90 062 102 104 202 204 502 504 Service Manual Input D LF HF Gain To do the Input D LF HF Gain calibration Press to select the first calibration step in Table 5 4 2 Connect Ch D of the Test Tool to the 5502A as shown in Figure 5 6 Set the 5502A SCOPE output to source the signal required for the first calibration point in Table 5 4 CL 0675 Pos D Fast Set the 5502A to operate OPR or standby STBY as indicated Press to start the calibration Wait until the display shows calibration status READY Press to select the next calibration step set the 5502A to the next calibration point signal and start the calibration w N Oa gt Continue through all calibration points of Table 5 4 When you are finished set the 5502A to Standby Continue at the nput AB Position section PM9091 Pm9585 50 OHM For 190 502 and 504 USE 50 OHM use only TRM50 TERMINATION 4 CHANNELS FLUKE 5502A CALIBRATOR Perf Ver G2 eps Figure 5 6 5502A SCOPE Output to Test Tool Input D 5 18 Cal step Calibration Adjustment Final Calibration For v11 10 and later Table 5 4 Input D LF HF Gain Calibration Points UUT input signal 5502A Setting CL 0675 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0475 100 mVpp square wave 1 kHz SCOPE edge 100 mVpp 1 kHz CL 0590 2 5 Vpp square wave 1 kHz SCOPE edge 2 5 Vpp 1 kHz CL 0586 2 5 Vpp sine wave 50 kHz
51. Base UA 4 20 Input A Trigger Sensitivity TGG 4 21 Input A AC Voltage Accuracy HF and Bandwidth Test 4 23 Input B Trigger Sensitivity TS 4 24 Input B AC Voltage Accuracy HF and Bandwidth Test 4 25 Input C Trigger Sensitivity Test cccooomusmmss inka annan maanneen 4 26 Input C AC Voltage Accuracy HF and Bandwidth Test 4 27 Input D Trigger Sensitivity TS 4 28 Input D AC Voltage Accuracy HF and Bandwidth Test 4 30 Video Test with SC600 Scope Calibration Option rss 4 31 Fluke 190 062 102 104 202 204 502 504 Service Manual External Trigger Level Test rss Meter DMM Tests rrua Meter DC Voltage Accuracy TGG Meter AC Voltage Accuracy and Frequency Response Test Continuity Function TGG Diode Test Function Tegi Ohms Measurements TS Probe Calibration Generator TS Performance Verification 4 Introduction Introduction AA Warning To prevent possible electrical shock fire or personal injury do not service the Test Tool unless you are qualified to do so Service described in this manual is to be done only by qualified service personnel Table 4 1 lists the available models for the Test Tools Table 4 1 Model Descriptions Model Description of Main Features 190 062 Two 60 MHz Scope Inputs BNC one Meter Input banana jacks 190 102 Two 100 MHz Scope Inputs BNC one Meter Input banana jacks 190 1
52. Continuous Auto Set Autoranging attenuators and time base automatic Connect and View triggering with automatic source selection Modes Alu 15 Hz to max bandwidth LOW Sek EAE 1 Hz to max bandwidth Minimum Amplitude A B C D DE HO Ale EA 10 mV 1 MHz to max Pangua 20 mV Automatic Capturing Scope Screens Capacity ccc sanna s ana erare 100 Scope Screens For viewing screens see Replay function Fluke 190 062 102 104 202 204 502 504 Service Manual Automatic Scope Measurements The accuracy of all readings is within of reading number of counts from 18 C to 28 C Add 0 1x specific accuracy for each C below 18 C or above 28 C For voltage measurements with 10 1 probe add probe accuracy At least 1 5 waveform period must be visible on the screen General DC Common Mode Rejection CMRR AC Common Mode Rejection at 50 60 0r 400 HZ aaa DC Voltage VDC Maximum Voltage With 10 1 DOE direct Til ME EEE Maximum Resolution with 10 1 probe rra lia Full Scale Reading rr Accuracy at 4 s to 10 us div Fluke 190 xx2 EGONA 5 mV div to 100 NIN Accuracy at 4 s to 10 us div Fluke 190 xx4 EON 5 mV div to 100 NIN Normal Mode AC Rejection at 50 06 60 HZ AC Voltage VAC Maximum Voltage with 10 1 probe rra ise J J J Maximum Resolution With 10 1 probe rra direct ABAR Full Scale Reading rr Accuracy Fluke 190 xx2 DC coupled DONO dae EO AC
53. EEE Directive 2002 96 EC marking reguirements The affixed label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the eguipment types in the WEEE Directive Annex I this product is classed as category 9 Monitoring and Control Instrumentation product Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Safety Instructions 1 Symbols AA Warning To prevent electrical shock or fire Use only the Fluke power supply Model BC190 Power Adapter Before use check that the selected indicated range on the BC190 matches the local line power voltage and frequency For the BC190 808 and BC190 820 universal Power Adapters use line cords that comply with the local safety regulations Note The BC190 808 and BC 190 820 universal Power Adapters are equipped with a male plug that must be connected to a line cord appropriate for local use The adapter is isolated so the line cord does not need to be equipped with a terminal for connection to protective ground AA Warning To prevent electrical shock or fire if a product input is connected to more than 42 V peak 30 Vrms or 60 V dc Use only insulated voltage probes test leads and adapters supplied with the product or indicated by Fluke as suitable for the Fluke 190 Series II ScopeMeter Test Tool series Before use inspect voltage probes
54. FLUKE 5502A CALIBRATOR Perf Ver Z2 eps Figure 5 10 Four wire Ohms Calibration Connections 3 Set the 5502A to the first test point in Table 5 8 Use the 5502A COMP 2 wire mode for the calibration adjustments up to and including 100 kQ For the higher values the 5502A will turn off the COMP 2 wire mode 4 Set the 5502A to operate OPR 5 26 Calibration Adjustment Final Calibration For v11 10 and later Press Fs to start the calibration Wait until the display shows the calibration status READY 7 Press N to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points When you are finished set the 5502A to Standby Continue at the Save Calibration Data and Exit section Table 5 8 Ohm Gain Calibration Points UUT input Value 5500 NORMAL CL 0913 100 kQ CL 0914 1M2 CL 0915 10 MQ 5 27 Fluke 190 062 102 104 202 204 502 504 Service Manual Final Calibration Firmware V10 9 and Lower The Final Calibration steps depend on the software that is installed in your Test Tool To view the software version installed on your Test Tool 1 Press ED 2 Press and VERSION amp CAL If the installed firmware is V09 00 V10 00 or V10 4x do the steps in this Final Calibration Firmware V10 9 and Lower section For firmware V11 10 do the steps in Final Calibration For V11 10 and late
55. Gain Calibration Points Cal Step UUT input Value 5500 NORMAL CL 0910 100 Q CL 0911 1 ko 10 kQ CL 0913 100 kQ CL 0914 1 MQ CL 0915 10 MQ Calibration Adjustment Save Calibration Data and Exit Save Calibration Data and Exit To save the calibration data and exit the Maintenance mode 1 Remove all test leads from the Test Tool inputs 2 Press Fa EXIT The Test Tool shows on the display Calibration data valid Save data and exit maintenance mode Note Calibration data valid indicates that the calibration adjustment procedure is performed correctly It does not necessarily mean that the Test Tool meets the specifications listed in Chapter 2 3 Press SA YES to save and exit Note After saving the calibration data the calibration number and date updates if the calibration data changes and the data are valid The calibration number and date do not change if e the calibration mode is entered and left without doing a calibration adjustment e only the contrast calibration adjustment and or the probe calibration are done If you press ET NO the Test Tool returns to the calibration mode You can either calibrate the Test Tool again or press SA EXIT YES to save and exit Error messages WARNING Calibration data not valid Save data and exit maintenance mode If you did the Warming Up and Pre Calibration successfully and you want to store the Pre Calibration data before continuing wi
56. M TERMINATION 2 CHANN METER 4 CHANNELS FLUKE 5502A CALIBRATOR B ES perf ver e2 eps Figure 4 8 5502A Scope Output to Test Tool Input B Reset the Test Tool to select the Test Tool setup Press LR to turn Input B on Press and use Hd to select Input B as trigger source a Fw N ee DO Use and J to change the sensitivity range to select manual sensitivity ranging and lock the Input B sensitivity range on 2 V div D Use to select the time base listed in Table 4 11 Set the 5502A to source the leveled sine wave given in the first row of Table 4 10 8 Adjust the 5502A output voltage until the displayed trace has the amplitude indicated under the appropriate column of Table 4 11 9 Verify that the signal is well triggered If not press and use ENIN to enable EO for manual Trigger Level adjustment Adjust the trigger level and verify that the signal will be triggered now The trigger icon T indicates the trigger level 10 Continue through the test points 11 When you are finished set the 5502A to Standby 4 24 UUT UUT 5502A SC MODE levsin Model Time base Initial Input Voltage Frequency 190 502 200 ns div 100 mV pp 5 MHz 190 062 190 102 190 104 190 202 190 204 190 502 190 504 Performance Verification Scope Input A B C D Tests Table 4 11 Input B Trigger Sensitivity Test Points UUT Trigger Amplitude 0 5 div 400 mV pp 60 MHz 800 mV pp 100 MH
57. N erregaia 50 Hz and 60 Hz Probe Calibration Manual pulse adjustment and automatic DC adjustment with probe check Generator Output a 1 225 Vpp 500 Hz square wave Internal Memory Number of Scope Memories nenne up to 30 varies according to version Each memory can contain 2 4 waveforms plus corresponding setups Number of Recorder Memories rra 10 Each memory can contain e 2 4 channel input TrendPlot e 2 4 channel input Scope Record e 100 2 4 channel input Scope screens Replay Number of Screen Image memories 9 Each memory can contain one screen image External Memory USB stick 2GB max Mechanical SIZE ruarena 265 mm x 190 mm x 70 mm 10 5 in x 7 5 in x 2 8 in Weight Fluke IBU 2 2 kg 4 8 Ib including battery BID au E AE 2 2 kg 4 8 Ib including battery Fluke 190562 a draia E ek gr EE 2 1 kg 4 6 Ib including battery Fluke 190 062 102 104 202 204 502 504 Service Manual Interface Ports Two USB ports provided Ports are fully insulated from instrument s floating measurement circuitry e A USB host port directly connects to external flash memory drive USB stick lt 2 GB for storage of waveform data measurement results instrument settings and screen copies e Amini USB B is provided which allows for interconnection to PC for remote control and data transfer using SW90W FlukeView software for Windows O One port can be active at the same time so remote contro
58. ScopeMeter 190 Series II Fluke 190 062 102 104 202 204 502 504 Service Manual February 2013 Rev 2 2 15 2013 2015 Fluke Corporation All rights reserved All product names are trademarks of their respective companies LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service The warranty period is three years and begins on the date of shipment Parts product repairs and services are warranted for 90 days This warranty extends only to the original buyer or end user customer of a Fluke authorized reseller and does not apply to fuses disposable batteries or to any product which in Fluke s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non defective media Fluke does not warrant that software will be error free or operate without interruption Fluke authorized resellers shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of Fluke Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price Fl
59. Tool Input D Reset the Test Tool to select the Test Tool setup Press MIEN to turn Input D on VAN Use L3H to move the Input D trace zero to the center grid line Press and use a to select Input D as trigger source mV RANGE Use it and E to change the sensitivity range to select manual sensitivity ranging and lock the Input D sensitivity range on 2 V div TIME Use 4 to select the time base indicated in Table 4 15 Set the 5502A to source the leveled sine wave given in the first row of Table 4 14 Adjust the 5502A output voltage until the displayed trace has the amplitude indicated under the appropriate column of Table 4 15 Verify that the signal is well triggered If not press EN and use to enable the O for manual Trigger Level adjustment Adjust the trigger level and verify that the signal is triggered The trigger icon T indicates the trigger level Continue through the test points 4 29 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 30 12 When you are finished set the 5502A to Standby Table 4 15 Input D Trigger Sensitivity Test Points Model Time base Initial Input Voltage Trigger Amplitude ALL 200 ns div 100 mV pp 5 MHz 0 5 div 190 104 10 ns div 400 mV pp 100 MHz 1 div 190208 reo 100 mv pp 190504 2nsdiv 200 mv pp Input D AC Voltage Accuracy HF and Bandwidth Test 1 2 10 Note The test steps for channel C are only for the models 190 104
60. Vpp Fluke 190 202 204 221 MHz 221 MHz Fluke 190 102 151 MHz 151 MHz Fluke 190 104 121 MHz 121 MHz Fluke 190 062 81 MHz 81 MHz Table 5 12 Input B LF HF Gain Calibration Points Extra steps for 190 104 204 V10 4x firmware Cal step CL 0462 CL 0463 UUT input signal 124 mVpp sine wave 50 kHz 124 mVpp sine wave 5502A Setting SCOPE levsine 124 mVpp 50 kHz SCOPE levsine 124 mVpp Fluke 190 204 Fluke 190 104 221 MHz 121 MHz 221 MHz 121 MHz Four channel instruments 190 104 204 with firmware V10 4x require the extra steps CL 462 and CL 463 Installed firmware version can be checked via key sequence USER key F3 VERSION amp CAL Software Version Calibration Adjustment Final Calibration Firmware V10 9 and Lower Input C LF HF Gain Note The adjustment steps for channel C are only for the models 190 104 and 190 204 Proceed as follows to do the Input C LF HF Gain calibration k 2 Press ENIN to select the first calibration step in Table 5 13 Connect the Test Tool to the 5502A as shown in Figure 5 13 4 CHANNELS PM9585 50 OHM For 190 502 and 504 use only TRM50 USE 50 OHM TERMINATION FLUKE 5502A CALIBRATOR N Oa gt Perf Ver F2 eps Figure 5 13 5502A SCOPE Output to Test Tool Input C Set the 5502A SCOPE output to source the signal reguired for the first calibration point in Table 5 13
61. al step UUT input value 55024 NORMAL CL 0799 5 mV CL 0800 12 5 mV CL 0801 25 mV CL 0802 50 mV CL 0803 125 mV CL 0804 250 mV CL 0805 500 mV CL 0806 1 25 V CL 0807 2 5V CL 0808 EN CL 0809 12 5V CL 0810 25V CL 0811 50 V set 5502A to OPR CL 0812 125 V CL 0813 250 V Input ABCD AB Zero To do the Input ABCD AB Zero calibration oa FW N gt Press ENIN to select calibration adjustment step CL 0852 Short circuit Input A B C D A B with 50 Q feed through terminations Press Fs to start the zero calibration Wait until the display shows the status READY Remove the input terminations For Fluke 190 062 190 102 and 190 202 continue at the Multimeter DMM Volt Gain section For Fluke 190 104 and 190 204 continue at the Save Calibration Data and Exit section 5 39 Fluke 190 062 102 104 202 204 502 504 Service Manual Multimeter Volt Gain AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Note The adjustment steps for the meter section are only for the models 190 062 190 102 and 190 202 To do the DMM Volt Gain calibration 1 Press to select the first calibration step in Table 5 19 2 Connect the Test Tool to the 5502A as shown in Figure 5 17
62. and Position section Input AB LF HF Gain and Position To do the Input AB LF HF Gain calibration 1 Press to select the first calibration step in Table 5 5 for the subversion 2 Connect Ch A and B of the Test Tool to the 5502A as shown in Figure 5 7 LIL Pmsoss page PM9091 CONNECT TO CHANNEL A B C D IN PARALLEL NO 50 OHM TERMINATION 4 CHANNELS O O O O Ole FLUKE 5502A CALIBRATOR 5 20 perf ver c2 eps Figure 5 7 Test Tool Input ABCD to 5502A SCOPE Output 3 Set the 5502A to supply a 1 kHz square wave SCOPE MODE volt SCOPE Z 1 MQ of 500 mV to channel A and B AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps 4 Set the 5502A to operate OPR 5 Press E to start the calibration 6 Wait until the display shows calibration status READY Calibration Adjustment Final Calibration For v11 10 and later 7 Press El to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points of Table 5 5 8 Set the 5502A to Standby and continue at the nput Pos ABCD AB Calibration section Table 5 5 Input AB Gain
63. areful notice of correct placement of the battery in the product or the External Battery Charger Do not short circuit a battery pack Do not keep battery packs in a place where the terminals can be shorted by metal objects for example coins paperclips pens or other Never use a battery pack or charger showing visible damage Batteries contain hazardous chemicals that can cause burns or explode If exposure to chemicals occurs clean with water and get medical aid Repair the product before use if the battery leaks Do not open modify reform or repair a battery pack that appears to malfunction or has been physically damaged Do not disassemble or crush battery packs Use the battery only as intended for the application Retain the original product information for future reference Safety Instructions Li ion Battery Pack Safe Transport of Battery Pack e The battery pack must be protected against short circuit or damage during transport e Always consult the IATA guidelines describing safe air transport of Li ion batteries Refer also to the section in the beginning of this paragraph on safe use of the battery pack e Check in luggage battery packs are only allowed when installed in the Product e Hand carry luggage a number of battery packs as required for normal and individual use is allowed e Always consult national local guidelines that are applicable for shipment by mail or other transporters e Amaximum
64. as indicated in the Calibration Procedure Steps section You can exit the calibration mode without changing the calibration data by turning the Test Tool off Calibration Adjustment Contrast Calibration Adjustment Explanation of Display Messages and Key Functions When the Test Tool is in the calibration mode only the EN to KNN softkeys the 0 key and the key are active unless otherwise stated The calibration adjustment menu shows the actual calibration step name and number and its status Cal Name CL nnnn Status Cal Name Name of the selected calibration step for example WarmingUp CL nnnn Number of the calibration step Status can be IDLE valid After re entering this step the calibration process is not started The calibration data of this step are valid This means that the last time this step was done the calibration was successful It does not necessarily mean that the unit meets the specifications related to this step IDLE invalid After re entering this step the calibration process is not started The calibration data are invalid This means that the last time this step was done the calibration was not successful Most probably the unit will not meet the specifications if the actual calibration data are saved BUSY aaa bbb Calibration adjustment step in progress progress for Input A and Input B During Warming Up the elapsed time is shown READY Calibration adjustment step finished
65. asurement menu and select Ohms MANUAL Press Haas to select AUTO ranging Set the 5502A to source the appropriate resistance value for the first test point in Table 4 19 Use the 5502A COMP 2 wire mode for the verifications up to and including 50 kQ For the higher values the 5502A will turn off the COMP 2 wire mode Observe the reading and check that it is within the range shown under the appropriate column Continue through the test points 10 When you are finished set the 5502A to Standby Performance Verification 4 5502A output Meter Reading COMP 2 wire 400 Q 4 KQ 40 kQ 400 KQ 4 MQ Meter Tests Table 4 19 Resistance Measurement Verification Points 0 0 to 0 5 COMP 2 wire 397 1 to 402 9 COMP 2 wire 3 971 to 4 029 COMP 2 wire 39 71 to 40 29 COMP 2 wire 397 1 to 402 9 off 3 971 to 4 029 off 30 MQ o e 29 77 to 30 23 off Probe Calibration Generator Test To calibrate connect a 10 1 probe as supplied with the Test Tool to input A red probe Connect the probe tip and the probe ground lead with the probe cal terminals on the lower left side of the Test Tool as shown in Figure 4 15 the figure is universal and shows a Test Tool with four oscilloscope channels such as the model 190 204 1 2 3 4 5 Reset the Test Tool Press to show the input A key labels Press Z PROBE A 10 1 Press E PROBE CAL and follow the instructions on the d
66. at fixes the top and bottom screening of the suspected channel Reinsert the flat cables if you want to measure the bottom side of the PCA under working condition See Figures 6 2 6 3 and 6 4 at the end of this chapter Note Before you attach the PCA again to the lower chassis plate it is advised to place the isolation foils around the channels Access to LCD Keypad Foil and Keypad Proceed as follows 1 Unlock both flat cables by shifting the connector latch at the left and right edge using a small screwdriver The latch is an integral part of the connector body Remove the flat cables from connector X9303 to LCD J9414 to keyboard J9415 to LCD backlight Remove 6 self tapping screws 10 mm long total length that fix the Main PCA module to the top case assembly Separate the Main PCA module from the top case Now you have access to LCD module keypad foil and keypad They can be separated from the top case without the removal of screws or clamps To prevent contamination do not touch contact areas with your hands or wear cotton gloves Note When installing the LCD module into the top case take care that no dust or dirt is present between module and the window decal Before reinstalling the Main PCA module on to the top case place the grey plastic strip around the BNC inputs Disassembly Steps See Figures 6 2 6 3 and 6 4 for guidance on disassembly Note Pictures may be subject to minor changes wit
67. ble setting in the Test Tool f Press to clear the softkey menu and see the full display Note The 10 kHz or 20 kHz bandwidth limiter rejects calibrator noise It does not affect the gain accuracy at a 50 Hz input signal RANGE Press and use a and to set the Input A sensitivity range to the first test point in Table 4 2 Set the 5502A to source the appropriate initial ac voltage Adjust the 5502A output voltage until the displayed Input A trace amplitude is 6 divisions Observe the 5502A output voltage and check to see if it is within the range shown under the appropriate column Continue through the test points 4 9 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 10 8 Check channel B C and D in succession Connect channel B C or D to 5502A when appropriate 9 Press W and select B as trigger source with Emo 10 Press B C or D to assign vertical range to channel B C or D 11 Observe the 5502A output voltage and check to see if it is within range 12 When you are finished set the 5502A to 0 zero Volt and Standby Table 4 2 Vertical Accuracy Verification Points V ac sine 50 Hz 6 divisions 50 mV div 106 06 mV 101 83 to 110 29 100 mV div 212 13 mV 203 67 to 220 58 20 V div 42 426 V 40 735 to 44 117 50 V div 106 06 V 101 83 to 110 29 100 V div 212 13 V 203 67 to 220 58 The vertical accuracy test can be done with dc voltage This method is advised
68. bly removed from the Test Tool must be adequately protected against damage The Test Tool contains static sensitive components Handling and servicing these components should be done only at a static free workstation by qualified personnel The Test Tool contains a Li ion battery pack See Section 1 for instructions on how to safely handle and use this battery pack The Users Manual is available at www fluke com The Test Tool uses self tapping screws For longer life use a hand operated screwdriver to reinsert the screws into the same screw hole threads on the case At the end of this chapter see Figures 6 2 6 3 and 6 4 for disassembly AAWarning To prevent electric shock disconnect test leads probes and power supply from any live source and from the Test Tool itself Always remove the battery pack before completely disassembling the Test Tool Only qualified personnel using customary precautions against electric shock should work on a disassembled unit with power on Disassembly and Reassembly Procedures Required Tools To access all the assemblies you need e Static free work surface and anti static wrist wrap e 10 Torx screwdriver e Asmall screwdriver or pair of tweezers to unlock flat cables from their connector e Cotton gloves to avoid contaminating the lens and the PCA Remove the Tilt Stand Hang Strap and Side Strap To separate the tilt stand from the rear case gently bend one rotation point away from the rear
69. crews M3x6 5 total length with a spring washer left side right side and bottom side Remove the four screws M3x10 total length that are grouped in a square around the sampling chip N2000 Observe how the screening plate fits onto the lower chassis before you remove this plate to access the top side of the PCA How to Access the Bottom Side of PCA To avoid contaminating the flex cable contacts with grease from your fingers do not touch these contacts or wear cotton gloves Contaminated contacts may not cause immediate instrument failure Failures typically show up when contaminated instruments are operated in humid areas 1 Unlock both flat cables by shifting the connector latch at the left and right edge with a small screwdriver The latch is an integral part of the connector body See Figure 6 1 st8682 eps Figure 6 1 Flat Cable Connector Remove the flat cables from connector X9303 to LCD J9414 to keyboard J9415 to LCD backlight Remove the four screws M3x10 that fix the PCA to the lower chassis shielding assembly Carefully slide the PCA out of the holes for the BNCs and Banana Jacks 2 ch Test Tools The A B and Meter input circuits are covered with an isolation foil Take careful notice on how the foil is positioned around the PCA before you remove the foil as far as required to repair a defective channel Fluke 190 062 102 104 202 204 502 504 Service Manual Remove a screw M3x22 th
70. d level indicated by the zero icon Tin the left margin to the center grid line Do this for all channels SAVE Press Press Hed SAVE Use OO to select SCREEN SETUP Press GD Use ESS to select OK SAVE Remember the name under which the settings are saved for instance SCOPE 1 Press to save the settings HOLD RUN Press to leave the Hold mode Performance Verification Display and Backlight Test Display and Backlight Test Proceed as follows to test the display and the backlight Press fo to turn the Test Tool on Remove the BC190 power adapter and verify that the backlight is dimmed 3 Apply the BC190 power adapter and verify that the backlight brightness increases 4 Press and hold E USER then press and release amp 9 CLEAR MENU The Test Tool shows the calibration menu in the bottom of the display 5 Do not press E now If you do turn off and turn on the Test Tool and start at Step 4 6 Press to toggle on and off the menu 5 Press Eg PREVIOUS three times The Test Tool shows Contrast CL 0100 6 Press ENIN CALIBRATE The Test Tool shows a dark display The test pattern shown in Figure 4 2 may be not visible or hardly visible Observe the display closely and verify that the display shows no abnormalities such as very light pixels or lines hpp204 eps Figure 4 2 Display Test Pattern 7 Press FE The test pattern is removed and the Test Tool shows Contrast CL 0100
71. djustment Final Calibration For v11 10 and later Table 5 3 Input C LF HF Gain Calibration Points UUT input signal 5502A Setting CL 0694 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0455 100 mVpp square wave 1 kHz SCOPE edge 100 mVpp 1 kHz CL 0550 2 5 Vpp square wave 1 kHz SCOPE edge 2 5 Vpp 1 kHz CL 0584 2 5 Vpp sine wave 50 kHz SCOPE levsine 2 5 Vpp 50 kHz CL 0585 2 5 Vpp sine wave SCOPE levsine 2 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz CL 0484 500 mVpp sine wave 50 kHz SCOPE levsine 500 mVpp 50 kHz CL 0485 0 5 Vpp sine wave SCOPE levsine 0 5 Vpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz 501 MHz Fluke 190 202 204 381 MHz 381 MHz Fluke 190 102 104 151 MHz 151 MHz Fluke 190 062 111 MHz 111 MHz Fluke 190 202 204 221 MHz 221 MHz Fluke 190 102 104 141 MHz 141 MHz Fluke 190 062 91 MHz 91 MHz CL 0464 100 mVpp sine wave 50 kHz SCOPE levsine 100 mVpp 50 kHz CL 0465 100 mVpp sine wave SCOPE levsine 100 mVpp Red subversion 5 Blue Subversion 2 Fluke 190 502 504 501 MHz Fluke 190 202 204 381 MHz Fluke 190 102 104 151 MHz Fluke 190 062 111 MHz Fluke 190 202 204 221 MHz Fluke 190 102 104 141 MHz Fluke 190 062 91 MHz 501 MHz 381 MHz 151 MHz 111 MHz 221 MHz 141 MHz 91 MHz Fluke 1
72. ed set the 5502A to Standby Table 4 12 HF AC Voltage Verification Points UUT 5502A SC MODE levsine UUT Model Voltage Frequency Reading B All 2 545 Vpp 835 mV to 965 mV All 2 545 Vpp 790 mV to 1 010 V 190 104 102 2 545 Vpp 100 MHz gt 630 mV 190 204 202 2 545 Vpp 200 MHz gt 630 mV Input C Trigger Sensitivity Test Note The test steps for channel C are only for the models 190 104 190 204 and 190 504 Proceed as follows to test the Input C trigger sensitivity 1 Connect the Test Tool to the 5502A as shown in Figure 4 9 PM9091 PM9585 50 OHM For 190 502 and 504 use only TRM50 USE 50 OHM TERMINATION 4 CHANNELS FLUKE 5502A CALIBRATOR perf ver f2 eps Figure 4 9 5502A Scope Output to Test Tool Input C 4 26 Performance Verification 4 Scope Input A B C D Tests 2 Reset the Test Tool to select the Test Tool setup 3 Press J to turn Input C on 4 Press Hd and use EH to move the Input C trace zero to the center grid line 5 Press ang use EINI to select Input C as trigger source 6 Use li cA and j to change the sensitivity range to select manual sensitivity ranging and lock the Input C sensitivity range on 2 V div Use E W to select the time base in Table 4 13 Set the 5502A to source the leveled sine wave given in the first row of Table 4 13 9 Adjust the 5502A output voltage
73. etting becomes the default contrast Press to return to the softkey menu Press F2 to select maximum brightness calibration Contrast CL 0120 Press CALIBRATE The display shows a bright test pattern Use ES to adjust the display to the maximum brightness where the test pattern is only just visible Press to return to the softkey menu Now you can either e Exit if only the Contrast had to be adjusted Continue at the Save Calibration Data and Exit section or e Dothe complete calibration Press ka to select the next step Warming Up and continue hpp204 eps Figure 5 2 Display Test Pattern Calibration Adjustment Warming Up and Pre Calibration Warming Up and Pre Calibration The Warming Up and Pre Calibration state is entered after the calibration mode or after selecting the next step if you have done the Contrast Calibration step CL 120 The display will show WarmingUp CL 0200 IDLE valid or invalid Unless you want to calibrate the display contrast only you must always start the calibration adjustment at the WarmingUp CL 0200 step Starting at another step will make the calibration invalid The Warming Up and Pre Calibration consists of a 30 minutes warming up period followed by several internal calibration adjustment steps that do not require input signals The total process takes about 75 minutes To do the Warming Up and Pre Calibration 1 Remove all input connections from the Test Tool 2 P
74. eypad Foil and Keypad Disassembly Steps iii Contents continued D H EN Fluke 062 102 104 202 204 502 504 Service Manual eU ulea EEA Final Assembly Parts and Kits Model Descriptions rss Vertical Accuracy Verification Points Volts DC Measurement Verification Points Volts AC Measurement Verification Points List of Tables Input A B AC Input Coupling Verification EOS Volts Peak Measurement Verification Points Input A B C D Frequency Measurement Accuracy Test Phase Measurement Verification Points Input A Trigger Sensitivity Test Points HF AC Voltage Verification Points Input B Trigger Sensitivity Test Points HF AC Voltage Verification Points Input C Trigger Sensitivity Test Points HF AC Voltage Verification Points Input D Trigger Sensitivity Test Points HF AC Voltage Verification Points Meter Volts dc Measurement Verification Points rrua Meter Volts AC Measurement Verification EGOITZ rss Resistance Measurement Verification Points rrua Input A LF HF Gain Calibration Points Input B LF HF Gain Calibration Points Input C LF HF Gain Calibration Points Input D LF HF Gain Calibration Points Input AB Gain and Position Calibration Points sr Input ABCD Gain Calibrati
75. for automatic verification that uses the Fluke Met Cal Metrology Software For each sensitivity range you must proceed as follows 1 Apply a 3 division voltage and adjust the voltage until the trace is at 3 divisions Write down the applied voltage V1 2 Apply a 3 division voltage and adjust the voltage until the trace is at 3 divisions Write down the applied voltage V2 3 Verify that V1 V2 6 x range 2 1 0 04 x range Example for range 10 mV div range div figure doubles because 2 measurements V1 and V2 are done for one accuracy check the allowed V1 V2 60 mV 0 021 x 60 0 08 x 10 60 mV 1 26 0 8 60 mV 2 06 mV Performance Verification 4 Scope Input A B C D Tests Input A B C D DC Voltage Accuracy Test AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Note The test steps for channel C and D are only for the models 190 104 190 204 and 190 504 Proceed as follows to verify the automatic dc voltage scope measurement 1 Connect the Test Tool to the 5502A as shown in see Figure 4 4 PM9091 CONNECT TO CHANNELA B C D IN PARALLEL PM9093 NO 50 OHM TERMINATION 4 CHANNELS INORMAL SCOPE FLUKE 5502A CALIBRATOR p EA
76. for channel C and D are only for the models 190 104 190 204 and 190 504 Proceed as follows to test the frequency measurement accuracy 1 Connect the Test Tool to the 5502A as shown in Figure 4 5 Do not use 50 Q terminations PM9091 CONNECT TO CHANNEL A B C D IN PARALLEL 1 pmooo3 PM3092 NO 50 OHM TERMINATION 4 CHANNELS O oO O O Ole perf ver c2 eps FLUKE 5502A CALIBRATOR Figure 4 5 5502A Scope Output to Test Tool Input A B C D 2 Select the following Test Tool setup SAVE a Recall the created setup see Standard Test Tool Setup Press RECALL and select SETUP press cd select the setup name and press to recall the setup b Press score c Press ZN READING and select with ENIN READINGS and with OG Reading 1 on A Hz Reading 2 on B Hz Reading 3 on C Hz Reading 4 on D Hz 3 Use Ka and EM to select range 100 mV div for A B C and D Use EE to select the required time base setting 5 Set the 5502A to source a sine wave according to the first test point in Table 4 7 Because no 50 Q termination is applied the 5502A leveled sine wave output amplitude is twice the set value 6 Observe reading A B C and D and check that it is within the range shown under the appropriate column 4 18 Performance Verification 4 Scope Input A B C D Tests 7 Continue through the test points
77. h the test points When you are finished set the 5502A to 0 zero Volt and Standby Performance Verification 4 Scope Input A B C D Tests Table 4 3 Volts DC Measurement Verification Points 5502A Output V dc 2 Channel Input Reading 4 Channel Input Reading 6 0 mV 4 9 to 7 1 4 8 to 7 2 2 mV div 6 0 mV 7 1 to 4 9 7 2 to 4 8 15 0 mV 14 2 to 15 8 13 9 to 16 1 5 mV div 15 0 mV 15 8 to 14 2 16 1 to 13 9 30 0 mV 28 9 to 31 1 28 5 to 31 5 10 mV div rca aa 30 0 mV 31 1 to 28 9 31 5 to 28 5 60 0 mV 58 5 to 61 5 57 6 to 62 4 20 mV div 60 0 mV 61 5 to 58 5 62 4 to 57 6 150 mV 142 to 158 139 to 161 50 mV div i eT 150 mV 158 to 142 161 to 139 300 mV 289 to 311 285 to 315 100 mV div K SS 300 mV 311 to 289 315 to 285 600 mV 585 to 615 576 to 624 200 mV div SS 600 mV 615 to 585 624 to 576 1 50 V 1 42 to 1 58 1 39 to 1 61 500 mV div 1 50 V 1 58 to 1 42 1 61 to 1 39 3 00 V 2 89 to 3 11 2 85 to 3 15 1 V div 3537 3 00 V 3 11 to 2 89 3 15 to 2 85 6 00 V 5 85 to 6 15 5 76 to 6 24 2 Vidiv 6 00 V 6 15 to 5 85 6 24 to 5 76 15 0 V 14 2 to 15 8 13 9 to 16 1 5 V div i 15 0 V 15 8 to 14 2 16 1 to 13 9 30 0 V 28 9 to 31 1 28 5 to 31 5 10 V div 30 0 V 31 1 to 28 9 31 5 to 28 5 60 0 V 58 5 to 61 5 57 6 to 62 4 20 V div 60 0 V 61 5 to 58 5 62 4 to 57 6 150 V 142
78. hout prior notice 6 Disassembly Disassembly and Reassembly Procedures Figure 6 2 Opened Case and Screws 6 7 Fluke 190 062 102 104 202 204 502 504 Service Manual hpp202 eps Figure 6 3 Screening Plate Removed and Screws 6 8 6 Disassembly Disassembly and Reassembly Procedures hpp203 eps Figure 6 4 PCA Removed from Chassis Bottom Side Visible 6 9 Fluke 190 062 102 104 202 204 502 504 Service Manual
79. hows the status READY Remove the input terminations Continue at the Multimeter DMM Ohm Gain section Multimeter Ohm Gain Note The adjustment steps for the meter section are only for the models 190 062 190 102 and 190 202 To do the DMM Ohm Gain calibration 1 Press to select the first calibration adjustment step in Table 5 20 2 Connect the Test Tool to the 5502A as shown in Figure 5 18 Notice that the sense leads must be connected directly to the Test Tool 5 41 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 42 0 OD ER ZE A METER B 2 CHANN METER I USE STACKABLE TEST LEADS i AS SUPPLIED WITH CALIBRATOR 0 FLUKE 5502A CALIBRATOR N Oa gt CL 0912 Perf Ver Z2 eps Figure 5 18 Four wire Ohms Calibration Connections Set the 5502A to the first test point in Table 5 20 Use the 5502A COMP 2 wire mode for the calibration adjustments up to and including 100 kQ For the higher values the 5502A will turn off the COMP 2 wire mode Set the 5502A to operate OPR Press to start the calibration Wait until the display shows the calibration status READY Press to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points When you are finished set the 5502A to Standby Continue at the Save Calibration Data and Exit section Table 5 20 Ohm
80. identifies conditions and procedures that can cause damage to the Product or the equipment under test 1 Fluke 062 102 104 202 204 502 504 Service Manual Symbols Table 1 is a list of symbols that are used on the Test Tool in the Users Manual in this Service Information or on spare parts for this Test Tool Table 1 1 Symbols Symbol Description Risk of Danger Important information See manual Hazardous voltage Live Voltage Earth Ground jib eg i al z Q 5 8 DC Direct Current AC or DC Alternating or Direct Current Conforms to relevant Australian EMC standards Bla E Conforms to relevant North American Safety Standards Conforms to European Union directives Conforms to relevant South Korean EMC Standards Double Insulation Measurement Category Ill is applicable to test and measuring circuits connected to the CAT II distribution part of the building s low voltage MAINS installation CAT E Measurement Category IV is applicable to test and measuring circuits connected at the source of the building s low voltage MAINS installation ja This product contains a Lithium ion battery Do not mix with the solid waste stream Spent ta batteries should be disposed of by a gualified recycler or hazardous materials handler per Li ion x local regulations Contact your authorized Fluke Service Center for recycling information This product complies with the W
81. ifications Meter Measurements for Fluke 190 xx4 Power A and B C and D ratio between Watts and VA 0 00 to 1 00 RMS reading of multiplication corresponding samples of input A or C volts and Input B or D amperes Full Scale Reading rra 999 counts VA tienen Vrms x Arms Full Scale Reading rra 999 counts VA Reactive NEI N VAY W Full Scale Reading rra 999 counts Phase A and B C and D KOOR arta eearri 180 to 180 degrees Resolution E a E a aa ee Ed 1 degree Accuracy 0 41 AEA 2 degrees 1 MHz to 10 MEZ rr 3 degrees Temperature TEMP With Optional Temperature Probe F not for Japan Ranges CC 0r SF nissa ansat TE de 40 0 to 100 0 100 to 250 100 to 500 100 to 1000 100 to 2500 Probe Sensitivity rra 1 mV C and 1 mV F DOUEN EEN eaa v n 1 5 5 counts add temperature probe accuracy for overall accuracy Decibel dB UEN dB relative to one volt GELARA dB relative to one mW in 50 Q or 600 Q BION Ee VDC VAC or VAC DC GGI CY ceciren aa e aia same as VDC VAC VAC DC Meter Measurements for Fluke 190 xx4 Four of the Automatic Scope Measurements as defined above may be displayed at the same time using larger screen area for convenient reading suppressing the scope waveform information For specifications see Automatic Scope Measurements above Meter Measurements for Fluke 190 xx2 The accuracy of all measurements is within of reading number of counts from
82. ine wave according to the first test point in Table 4 8 Because no 50 termination is applied the 5502A leveled sine wave output amplitude will be twice the set value 6 Observe the readings A B C and D and check that they are not outside the range shown under the appropriate column Continue through the test points When you are finished set the 5502A to 0 zero Volt and Standby Table 4 8 Phase Measurement Verification Points Time base 5502A SC MODE Frequency Voltage Input A B D 2 lt 20 ms div wavegen sine 1 MQ 600 mVpp Time Base Test Proceed as follows to test the time base accuracy 1 Connect the Test Tool to the 5502A as shown in Figure 4 6 For the Fluke 190 502 and 190 504 you must use the Fluke TRM50 50 Q terminator PM9091 D 50 OHM e g PM9585 or tRM50 For 190 502 and 504 only use dak METER B USE 50 OHM TERMINATION 2 CHANN METER 4 CHANNELS 020 050 FLUKE 5502A CALIBRATOR perf ver d2 eps Figure 4 6 5502A Scope Output to Test Tool Input A 2 Setthe 5502A to source an 8 ms time marker MODE marker 4 20 Performance Verification 4 Scope Input A B C D Tests 3 Select the Test Tool setup a Reset the Test Tool E RANGE b Use and 22 to select manual vertical ranging and set the Input A sensitivity range to 5 V div 10 1 probe or 500 mV div probe A factor is
83. ion point in Table 5 17 5 36 Calibration Adjustment Final Calibration Firmware V10 9 and Lower AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Set the 5502A to operate OPR Press Fs to start the calibration Wait until the display shows calibration status READY N Oa gt Press SS to select the next calibration step set the 5502A to the next calibration point and start the calibration Continue through all calibration points of Table 5 17 8 Set the 5502A to Standby and continue at the nput ABCD AB Position section Table 5 17 Input ABCD Gain Calibration Points UUT input value 5502A SCOPE MODE volt SCOPE Z 1 MQ 1 kHz none 5500 standby CL 0504 500 mV CL 0624 none 5500 standby CL 0600 25 mV CL 0601 50 mV none 5500 standby a N sv 50V set 5502A to OPR 5 37 Fluke 190 062 102 104 202 204 502 504 Service Manual Input ABCD AB Position To do the Input ABCD AB Position calibration Press to select calibration adjustment step CL 0619 Remove all Input A B C D A B connections open inputs Press Fen to start the calibration Wait until the display shows calibration status READY Continue at the Input ABCD AB Volt Gain
84. isplay Press KNININ to start the probe calibration The first step is to manually adjust the square wave response to a pure square wave pulse top must be straight The trimmer is located in the probe housing and can be reached by rotating the center part of the housing For further information refer to the probe instruction sheet When done press E to start the DC calibration automatically The Probe Calibration is OK if all instructions shown on the display are finished successfully Close the hole of the trimmer by rotating the center part of the housing This is important for safe use of the probe at high input voltages Repeat the procedure for channel B blue probe For the 4 Channel Test Tools 190 104 190 204 190 504 repeat the procedure for channel C gray probe and channel D green probe 4 39 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 40 Probe Cal Probe Cal Ref Figure 4 15 Probe Calibration This is the end of the Performance Verification Procedure st8416 t eps Chapter 5 Calibration Adjustment Title Page dauala 5 3 General dabble A A 5 4 Calibration Number and Date rrura 5 4 General Instructions rra 5 4 Equipment Required For Calibration 5 5 Calibration Procedure Steps 5 5 How to Start the Calibration rss 5 6 Contrast Calibration Adjustment rss 5 7 Warming Up and Pre Calibration rss 5 9 Final Calibration For v11 10 erdie Warming Up 2 Warm U
85. ith 10 1 probe rra 10 mV direct BAN 1 mV Full Scale Reading rra 800 counts Accuracy Max peak or Min peak 0 2 division E A a ELA 0 4 division Frequency Hz ELLE 1 000 Hz to full bandwidth Full Scale Reading c ccecceeceseeeeereeeeeeereeneeneees 999 counts Accuracy 1 Hz to full bandwidth rra 0 5 2 counts 4 s div to 10 ns div and 10 periods on the screen Duty Cycle DUTY FRAN QC wes 4 0 to 98 0 PSE a EK a Eee 0 1 when period gt 2 div Full Scale Reading rra 999 counts 3 digit display Accuracy logic or pulse rro 0 5 2 counts Pulse Width PULSE Resolution with GLITCH Off rra 1 100 division Full Scale Reading rra 999 counts Accuracy 1 Hz to full bandwidth eee 0 5 2 counts Vpwm Purpose to measure on pulse width modulated signals like motor drive inverter outputs Principle readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental freguency Accuracy aS Vims for sinewave signals V Hz Purpose to show the measured Vpwm value see Vpwm divided by the fundamental frequency on Variable AC Motor Speed drives Accuracy Vrms Hz Note AC motors are designed for use with a rotating magnetic field of constant strength This strength depends on the applied voltage Vpwm divided by the fundamental frequency of the applied voltage Hz The nominal Volt and Hz values are shown on the motor type plate Spec
86. its original packaging until ready to use e When possible remove the battery pack from the equipment when not in use e Fully charge the battery pack before storing it for an extended period to avoid a defect e After extended periods of storage it may be necessary to charge and discharge the battery packs several times to obtain maximum performance e Keep the battery pack out of the reach of children and animals e Seek medical advice if a battery or part of it has been swallowed Fluke 062 102 104 202 204 502 504 Service Manual Safe Use of Battery Pack AA Warning To prevent electrical shock or fire The battery pack needs to be charged before use Use only Fluke approved power adapters to charge the battery pack Refer to Fluke s safety instructions and Users Manual for proper charging instructions Do not leave a battery on prolonged charge when not in use The battery pack gives the best performance when operated at normal room temperature 20 C 5 C 68 F 9 F Do not put battery packs near heat or fire Do not put in sunlight Do not subject battery packs to severe impacts such as mechanical shock Keep the battery pack clean and dry Clean dirty connectors with a dry clean cloth Do not use any charger other than that specifically provided for use with this equipment Do not use any battery which is not designed or recommended by Fluke for use with the Product Take c
87. iv all channels Up to 1 25 GS s 10 US AREAKO RN 125 MS s Fluke 190 062 10 ns to 4 us div all channels up to 625 MS s 10 us to 120 RG UG 125 MS s Glitch Detection 4 us to 120 AAE displays glitches as fast as 8 ns Waveform Display rra A B C D Math x X Y mode spectrum Normal Average Persistence Reference Time Base Accuracy rra 100 ppm 0 04 div Record Length all models see table that follows Record Length all models Samples points per input Mode Glitch Detect On Glitch Detect Off Max Sample Rate Scope Normal 300 min max pairs 3 k true samples compressed into 1 screen 300 samples per screen Scope Fast 300 min max pairs Scope Full 300 min max pairs 10 k true samples compressed into 1 screen Use Zoom and Scroll to see waveform details 190 062 625 MS s 190 102 104 1 25 GS s 190 202 204 2 5 GS s 1 or 2 channels on 190 204 1 25 GSf s 3 or 4 channels on 190 50x 5 GS s 1 channel on 190 50x 2 5 GS s 2 channels on 190 504 1 25 GS s 3 or 4 channels on Scope Record Roll 30 k samples 4x 125 MS s Trend Plot gt 18 k min max average Up to 5 measurements second values measurement Trigger and Delay Tigger U o g eee Automatic Edge Video Pulse Width N Cycle External 190 xx2 Trigger ENA up to 1200 divisions Pre Trigger View rra one full screen length Delay 12 div to 1
88. l and data transfer via mini USB is not possible when saving or recalling data to or from the USB stick Environmental Ewironnent MIL PRF 28800F Class 2 Temperature Operating auker do estes cerns 0 C to 50 C 32 F to 122 F Operating and charging rra 0 C to 40 C 32 F to 104 F Storage 20 C to 60 C 4 F to 140 F Humidity Maximum Relative Operating 0 C to 10 C 32 F to 50 F rra noncondensing 10 C to 30 C 50 F to 86 F oe 95 5 30 C to 40 C 86 F to 104 F raa 75 5 40 C to 50 C 104 F to 122 F ra 45 5 Storage 20 C to 60 C 4 F to 140 F noncondensing Altitude Operating CATIII 600 V CATII 1000 V rrera 3 km 10 000 feet CATIV 600 V CATIII 1000 V rrea 2 km 6 600 feet Sto sasggggg 12 km 40 000 feet Vibration Sinusoidal rr max 3 g Vibration Random a 0 03 g Hz euela max 30g Electromagnetic Environment rrua EN IEC61326 1 Portable Equipment Enclosure Protection uuu IP51 ref IEC60529 Certifications EGOI SO teta artEa o teren EK bed Safety Designed for 1000 V Measurement Category III 600 V Measurement Category IV with supplied 10 1 probes in accordance with e EN IEC 61010 1 Pollution Degree 2 e EN IEC 61010 2 030 e IEC 61010 031 A Max Input Voltages BNC Input A B C D directly rra 300 V CAT IV Vis giza 1000 V CAT III 600 V CAT IV METER EXT banana input rru 1000 V CAT III 600 V CAT IV
89. l short 2x 3 3 4 Case seal 3 3 13 Case Set 2 channel 4035349 Bottom case assy 3173 Battery door 3 1 14 Quarter turn screw 2x 3 1 15 Adhesive foam for battery door Standup bracket 3 1 16 Quarter turn screw 948609 For battery door 1x 3 1 15 Li ion Battery Pack BP290 26 Wh 10 8 V Li ion Battery Pack BP291 52 Wh 10 8 V Lens decal 190 062 4035360 3 3 14 Lens decal 190 102 4035372 3 3 14 Lens decal 190 104 3981826 3 3 14 Lens decal 190 202 4035324 3 3 14 Lens decal 190 204 3981832 3 3 14 Lens decal 190 502 4035385 3 3 14 Lens decal 190 504 4534710 3 3 14 LCD module 3 3 1 LCD assy Flk 190 II 3981844 LCD fixation foam 3 3 2 Flat cable 3 3 3 Top holster Input Cover 4035397 NN 31 1 2 channels meter Top holster Input Cover 3045328 3 1 1 4 channels Sealing strip flexible 3945319 Set of 2 1 pce for 4 channel 1 31 11 around inputs pce for 2 channel instruments Table 3 1 Final Assembly Parts and Kits cont Parts List Final Assembly Parts Part or Kit Ordering Code Consists of Following Parts Self tapping Screw 10 mm 2x to fix Figure ltem nr i 3 1 2 input cover Dowel 6x to fix straps 3 1 4 Steel Plate for Lock 3 1 5 Mounting Material Set 3981859 Self tapping Screw 16 5 mm 4x to 3 1 6 fix Rear Case Screw M3x6 2x to fix bottom 3 1 8 holster Self tapping
90. nction TGG Ohms Measurements Tegi Probe Calibration Generator TS 5 Calibration Adjustment rss ITO ACTION EZE Calibration Number and DSen General Instructiopss sa 5 4 Equipment Required For Calibration rr Calibration Procedure Steps rra How to Start the Calibration rra Contrast Calibration Adjustment rrura Warming Up and Pre Calibration rrua Final Calibration For v11 10 and later rra Warming Up 2 Warm Up Final and ADC Timingq rss Input A LF HF EUA Input B LP AF Gain eztiote gae Input C LF HF GOI IRABA Ee ELSENAAR Input AB Position rss Input AB LF HF Gain and Position rss Input Pos ABCD AB Calibration rrura Input ABCD AB Noise F FBW Calibration rss IGA El aga Multimeter Meter Zero sr Multimeter Volt EUA Multimeter Ohm GOI Final Calibration Firmware V10 9 and Lower Error Messages rra Warm Up Final and ADC Timing Input A LF HF Gain Input B LF HF Gain Input C LF HF Gain Input D LF HF Gain Input ABCD AB LF HF Gain Input ABCD AB POSItiON rss Input ABCD AB Volt Gain Input ABCD AB Zero ARIZALA Multimeter Numeric Zero Multimeter Ohm Gain ss Save Calibration Data and Exit Probe Calibration Disassembly IMO AWCTION A Disassembly and Reassembly Procedures Reguired Ke 0 aa aita paiva matoa tatan omaat maks tad azEi atarteak Remove the Tilt Stand Hang Strap and Side Strap Open the Test Tool Remove the Battery Pack How to Access the Top Side of PCA rss How to Access the Bottom Side of EGAra Access to LCD K
91. nd press to recall the setup b Press c Press ENIN READING and select with B READINGS and E Reading 1 on A V ac Reading 2 on B V ac Reading 3 on C V ac Reading 4 on D V ac d Press MINAN and use I to select COUPLING AC e Press MENN and use to select COUPLING AC f Press au and use Fz to select COUPLING AC g Press EZOZI and use 2 to select COUPLING AC h Press to clear the softkey menu and see the full display 3 Use EE to change the time base to lock the time base on 40 ms div DE 4 Use Hit and E to set the Input A B C and D sensitivity range to 500 mV 5 Set the 5502A to source the appropriate ac voltage and frequency listed in Table 4 5 6 Observe the reading A B C and D and check that they are within the range shown under the appropriate column Continue through the test points When you are finished set the 5502A to O zero Volt and Standby Table 4 5 Input A B AC Input Coupling Verification Points 5502A output V rms 5502A Frequency 2 Ch Reading 4 Ch Reading 60 Hz 873 mV to 920 mV 859 mV to 933 mV 5 Hz gt 630 mV gt 630 mV 4 16 Input A Performance Verification 4 Scope Input A B C D Tests B C D Peak Measurements Test AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on
92. ng AUTO in upper right LCD edge d PI and EN to change the sensitivity range to select manual sensitivity ranging and lock the Input C sensitivity range on 500 mV div e 4 to move the Input C trace zero to the center grid line f Press HAGN and use KIN to select Input C as trigger source 3 Setthe 5502A to source a sine wave and to the first test point in Table 4 14 Observe the Input C reading and check that it is within the range shown under the appropriate column of Table 4 14 Continue through the test points When you are finished set the 5502A to Standby Table 4 14 HF AC Voltage Verification Points UUT 5500A SC MODE levsine UUT Model Voltage Freguency Reading A all 2 545 Vpp 1 MHz 835 mV to 965 mV all 2 545 Vpp 25 MHz 790 mV to 1 010 V 190 104 2 545 Vpp 100 MHz gt 630 mV 190 204 2 545 Vpp 200 MHz gt 630 mV 190 504 2 545 Vpp 500 MHz gt 630 mV Input D Trigger Sensitivity Test Note The test steps for channel D are only for the models 190 104 190 204 and 190 504 Proceed as follows to test the Input D trigger sensitivity 1 Connect the Test Tool to the 5502A as shown in Figure 4 10 4 28 Performance Verification 4 Scope Input A B C D Tests 4 CHANNELS _ PM9585 50 OHM For 190 502 and 504 USE 50 OHM use only TRM50 TERMINATION FLUKE 5502A CALIBRATOR 10 11 perf ver g2 eps Figure 4 10 5502A Scope Output to Test
93. nt possible electrical shock fire or personal injury ensure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Proceed as follows to test the meter dc voltage measurement accuracy 1 2 D a fw N Connect the Test Tool to the 5502A as shown in Figure 4 13 Select the Test Tool setup Press METER Press ET to open the Measurement menu and select V de Press d MANUAL Press K to select MANUAL ranging Use ranges Set the range to the first test point in Table 4 17 and E2 to select the Set the 5502A to source the appropriate dc voltage Observe the reading and check to see if it is within the range shown under the appropriate column 10 Continue through the test points 11 When you are finished set the 5502A to O zero Volt and Standby 4 35 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 36 500 0 mV 5 000 V 50 00 V 500 0 V 1100 V Table 4 17 Meter Volts dc Measurement Verification Points 50 00 V 49 70 to 50 30 50 00 V 49 70 to 50 30 500 0 V 497 0 to 503 0 1000 V 0 990 to 1 010 1000 V 0 990 to 1 010 Meter AC Voltage Accuracy and Frequency Response Test AA Warning To prevent possible electrical shock fire or personal injury ensure that the calibrator is in s
94. on Points DMM Volt Gain Calibration Points Ohm Gain Calibration Points Input A LF HF Gain Calibration Points Input A LF HF Gain Calibration Points Input B LF HF Gain Calibration Points Input B LF HF Gain Calibration Points Input C LF HF Gain Calibration Points Input C LF HF Gain Calibration Points Input D LF HF Gain Calibration Points Input D LF HF Gain Calibration Points Input ABCD Gain Calibration Points Input ABCD Gain Calibration Points DMM Gain Calibration Points Ohm Gain Calibration Points Fluke 062 102 104 202 204 502 504 Service Manual vi Figure 3 1 3 2 3 3 4 1 4 2 4 3 4 4 4 5 4 6 4 7 4 8 4 9 4 10 4 11 4 12 4 13 4 14 4 15 5 1 5 2 5 3 5 4 5 5 5 6 5 7 5 8 5 9 5 10 5 11 5 12 5 13 5 14 5 15 5 16 5 17 5 18 5 19 5 20 6 1 6 2 6 3 6 4 List of Figures Title Page Open Case and Ke Screening Plate Removed and SCrews rss 3 7 PCA removed from Chassis Bottom Side utiles Menu Item Beia ela 4 5 Display Test P tter ss Test Tool Inputs A B to 5502A Normal Output rss Test Tool Inputs A B C D to 5502A Normal Output surra 5502A Scope Output to Test Tool Input A B C Dra 5502A Scope Output to Test Tool
95. onse to a pure square wave pulse top must be straight see Figure 5 20 The trimmer is located in the probe housing and is reached by rotating the center part of the housing For more information see the instruction sheet for the probe H ST7991 WMF Figure 5 20 10 1 Probe Calibration When done press EA o to start the DC calibration automatically The Probe Calibration is OK if all instructions shown on the display are finished successfully Close the hole of the trimmer by rotating the center part of the housing this is important for safe use of the probe at high input voltages 10 Repeat the procedure for the blue VPS410 II B probe that is connected between the blue Input B BNC and the probe calibration terminals on the left side of the instrument 5 45 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 46 Chapter 6 Disassembly ITO Dieta BAA Disassembly and Reassembly Procedures rra Required ala EAEE Remove the Tilt Stand Hang Strap and Side Strap Open the Test Tool Remove the Battery Pack rra How to Access the Top Side of EGA rru How to Access the Bottom Side Of EGA rru Access to LCD Keypad Foil and Keypad eseceeeeeeeeeeteeeeeteeeseees Disassembly Steps rru 6 1 Fluke 190 062 102 104 202 204 502 504 Service Manual Disassembly Introduction Introduction This section provides the required disassembly procedures The printed circuit assem
96. or if it fails the Performance Verification The Test Tool has a normal calibration cycle of one year Because the hardware and firmware has changed over the life of the Product development of a second version of the adjust procedure has been necessary This section has both adjust procedures To identify the procedure version you need 1 Check that the serial number is 25375604 or higher If yes use the first procedure 2 Ifno evaluate the firmware and Subversion Press in sequence USER F3 VERSION amp CAL If the firmware version is V11 00 or lower use the second procedure If the firmware is V11 10 or higher 3 In the same popup menu check under Subversions that the last datablock is x5xx for instance 2516 This value determines the higher frequency adjust point that should be used In the tables when red or blue values are listed e 5 The Final Calibration for V11 10 and later procedure should be done Use the levels in Red e 2 The Final Calibration for V11 10 and later procedure should be done Use the levels in Blue 5 3 Fluke 190 062 102 104 202 204 502 504 Service Manual General Calibration Number and Date When storing valid calibration data in FlashROM after the calibration adjustment procedure is complete the calibration date is set to the actual Test Tool date and the calibration number increments by one To show the calibration date and number 1 Press E then press GN to see
97. ote The setup steps for channel C and D are only for the models 190 104 190 204 and 190 504 Proceed as follows to create a setup for instance SCOPE 1 Reset the Test Tool Input A is ON and other inputs are OFF 2 Press B INPUT B ON The black text with yellow background indicates the actual settings 4 5 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 6 Oo NO E Fw 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 Press ENIN to change the PROBE B setting Select Probe Type Voltage Attenuation 1 1 Press IG INPUT C ON Press NNN to change the PROBE C setting Select Probe Type Voltage Attenuation 1 1 Press eN input D ON Press EN to change the PROBE D setting Select Probe Type Voltage Attenuation 1 1 Press Af The inverse text indicates the actual settings Press ETE to change the PROBE A setting Select Probe Type Voltage Attenuation 1 1 Press score Press Lm READINGS ON Press F2 O READING and select with READINGS and with e a Reading 1 on A V dc b Reading 2 on B V dc c Reading 3 on C V dc d Reading 4 on D V dc Press ENIN WAVEFORM OPTIONS and select Glitch Off Acquisition Normal Average Off Waveform Normal MANUAL AUTO Press to select MANUAL ranging MANUAL in upper right of display ZS Press HENNI Use and Ei to move the Input A groun
98. p Final and ADC Timing 5 11 Input A LF HF GOI Input B LF HF OK rarte Input C LF MF Sa toh et toa me Input AB Position rss 5 20 Input AB LF HF Gain and Position ees Input Pos ABCD AB Calibration Input ABCD AB Noise F FBW Calibration rss 5 22 Input AB Volt Cale cetera aketa Multimeter Meter ZESIO 5 24 Multimeter Volt Gain a 5 24 Multimeter Ohm Gam EA 5 26 Fluke 190 062 102 104 202 204 502 504 Service Manual Final Calibration Firmware V1 Error Messages 0 0 RSSRA SAK Warm Up Final and ADC Timing essa Input A LF HF Gain Input B LF HF Gain Input C LF HF Gain Input D LF HF Gain Input ABCD AB LF HF DG Input ABCD AB Position Input ABCD AB Volt Gain Input ABCD AB Zero Multimeter Volt Gain Multimeter Numeric Zero Multimeter Ohm Gain Save Calibration Data and Exit Probe Calibration Calibration Adjustment Introduction Introduction This section provides the complete Calibration Adjustment procedure for the Test Tool The Test Tool allows closed case calibration with known reference sources It measures the reference signals calculates the correction factors and stores the correction factors in RAM When the calibration is complete the correction factors can be stored in FlashROM The Test Tool should be calibrated after repair
99. perf ver b2 eps Figure 4 4 Test Tool Inputs A B C D to 5502A Normal Output Fluke 190 062 102 104 202 204 502 504 Service Manual 4 12 2 Select the Test Tool setup a Recall the setup see Standard Test Tool Setup Press E Fz o RECALL and select SETUP Press Gi select the setup name and press to recall the setup b Press MINIMI press ETT INPUT A OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the ScopeMeter c Press B press Fa INPUT B OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the ScopeMeter d Press Petri press Fo INPUT C OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the ScopeMeter e Press D press Kaa INPUT D OPTIONS and select Attenuator Normal Bandwidth 10 kHz or 20 kHz for available setting in the ScopeMeter f Press to clear the softkey menu and the full 8 divisions display RANGE 3 Press MININ and use EO and Will to set the Input A sensitivity range to the first test point in Table 4 3 Do this also for channel B C and D Set the 5502A to source the appropriate dc voltage 5 Observe readings A B C and D and check they are within the range shown under the appropriate column Due to calibrator noise occasionally OL overload can be shown Continue throug
100. puts 8 9 Press Fs to start the calibration 1 2 Wait until the display shows calibration status End Precal READY 3 4 5 Press ETZE to select the next calibration step CL 201 WarmUpFinal Press Fs to start the calibration Wait until the display shows calibration ready Press ES to select the next calibration step CL 0570 ADC Timing Connect Ch A of the Test Tool to the 5502A SCOPE output as shown in Figure 5 3 Use a 50 Q termination Set the 5502A to generate a sine wave 50 MHz 0 5 V pp mode LEVSINE at the SCOPE output Set the 5502A in operate OPR Press Fs to start the calibration 10 Wait until the display shows calibration status READY 11 Set the 5502A in standby STBY 12 Continue at the Input A LF HF Gain section TRM 50 50 OHM H For 190 502 and 504 use only TRM50 USE 50 OHM TERMINATION METER 2 CHANN METER FLUKE 5502A CALIBRATOR perf ver d2 500 eps Figure 5 3 5502A SCOPE Output to Test Tool Input A 5 11 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 12 Input A LF HF Gain To do the Input A LF HF Gain calibration 1 2 Connect Ch A of the Test Tool to the 5502A as shown in Figure 5 3 The display must show step CL 0654 Pos A Fast If it does not then press or to select the first calibration step in Table 5 1 Set the 5502A SCOPE output to source the signal required for the first calibration point
101. r section Before you start the final calibration do the steps in the Warming Up and Pre Calibration section The final calibration reguires input conditions that will be described in each step After starting a step several steps that reguire the same input conditions are done automatically For example if you start calibration step CL 0852 the calibration can include step CL 0929 and at the end the display shows CL 0929 READY Note You must always start the Final Calibration at the first step See the Warm Up Final and ADC Timing section Starting at any other step will make the calibration invalid If you proceed to calibration step N for example step CL 0481 then return to a previous step for example step CL 0480 and then calibrate this step the complete final calibration becomes invalid You must repeat the calibration starting at Warm Up Final and ADC Timing It is allowed to repeat a step that shows the status READY by pressing F3 again Error Messages Proceed as follows if an error message ERROR nnnn is displayed during calibration e if nnnn lt 5000 check the input signal and test leads and press Fz again to repeat the current step e if nnnn 25000 check the input signal and test leads and repeat the final calibration at the Warm Up Final and ADC Timing section If the error persists the Test Tool is defective Warm Up Final and ADC Timing The Warm Up Final step CL 0201 must be done with open inp
102. ration do the above mentioned steps 1 2 and 5 If during normal operation the display cannot be made dark or light enough or if the display after a Test Tool reset is too light or too dark you can do this calibration Probe calibration do the above mentioned step 6 The probe calibration matches the probe to the used input channel Fluke 190 062 102 104 202 204 502 504 Service Manual 5 6 How to Start the Calibration To start the calibration 1 Power the Test Tool with the power adapter input and the BC190 power adapter Check the actual Test Tool date and adjust the date if necessary the calibration date will become the Test Tool date when saving the calibration data Press EN toggles the menu bar on and off a b Press EI to open the OPTIONS menu c Use to select the DATE ADJUST option d Press to open the DATE ADJUST menu e If necessary adjust the date with amp S Press to activate all selections and leave the menu Select the calibration mode The Calibration Adjustment Procedure uses built in calibration setups that can be accessed in the calibration mode To enter the calibration mode proceed as follows e Press and hold user press and release and release EZ The display shows the CAL MODE Calibration Adjustment screen The display shows the calibration step WarmingUp CL 0200 the calibration status IDLE valid or IDLE invalid and the softkey menu Continue
103. ress Fs to start the Warming Up and Pre Calibration The display shows the calibration step in progress and its status The first step is WarmingUp CL 0200 BUSY 00 29 59 or WarmingUp1 CL 0200 BUSY 00 09 59 The warming up period is counted down to 00 00 00 Then the remaining pre calibration steps are performed automatically The entire procedure takes about 60 minutes 3 Wait until the display shows End Precal READY The PreCal data have now been stored in FlashROM If you turn off the Test Tool now by accident turn it on again immediately and select the calibration mode Continue with step 5 below 4 Press aN NEXT several times see Final Calibration If you turn off the Test Tool now and you do not turn on immediately the Test Tool has cooled down and you must repeat the Warming Up and Pre Calibration select the calibration mode and start at CL 0200 5 Press EN NEXT and continue at the Final Calibration section Error Messages If error message 1000 is displayed during Warming Up or Pre Calibration step CL 0215 the Main PCA hardware version is not suitable for the installed software version Other error messages during Warming Up or Pre Calibration indicate that the Test Tool is defective and should be repaired 5 9 Fluke 190 062 102 104 202 204 502 504 Service Manual 5 10 Final Calibration For v11 10 and later Before you start the final calibration do the Warming Up and Pre Calibration
104. riate column 9 Continue through the test points 10 When you are finished set the 5502A to O zero Volt and Standby Table 4 4 Volts AC Measurement Verification Points 5502A output 2 Ch Input Reading 4 Ch Input Reading 2 mV div Select 10 ms div 60 Hz 2 9 mV to 5 1 mV 2 9 mV to 5 1 mV Note Set input A B Bandwidth 10 kHz or 20 kHz to prevent OL due to calibrator noise see step 2 T omv em ssmvenzmv ETT 10 mV div Select 20 us div 20 kHz 18 0 mV to 22 0 mV 17 7 mV to 22 3 mV Note Set input A B Bandwidth 20 MHz 20 mV div 20 kHz 37 5 mV to 42 5 mV 36 9 mV to 43 1 mV zar Sr oom oore rareza 500 mV div Select 20 us div 900 mV 20 kHz 862 mV to 938 mV 849 mV to 951 mV 1 V div 2V 20 kHz 1 80 V to 2 20 V 1 77 V to 2 23 V 20 V div 40 V 20 kHz 37 5 V to 42 5 V 36 9 V to 43 1 V 50 V div 90 V 20 kHz 86 2 V to 93 8 V 84 9 V to 95 1 V 100 V div 200 V 20 kHz 180 V to 220 V 177 V to 223 V Fluke 190 062 102 104 202 204 502 504 Service Manual Input A B C D AC Coupled Lower Frequency Test Note The test steps for channel C and D are only for the models 190 104 190 204 and 190 504 To test the ac coupled input low freguency accuracy 1 Connect the Test Tool to the 5502A as for the previous test see Figure 4 4 2 Selectthe Test Tool setup a Recall the setup Standard Test Tool Setup Press E F2 RECALL and select SETUP press ENTER J select the setup name a
105. rrua 64 min to 546 hours Time Reference ru time from start time of day Fluke 190 062 102 104 202 204 502 504 Service Manual 2 12 Scope Record Records scope waveforms in deep memory while displaying the waveform in Roll mode Gea dai Input A B C D Max Sample Speed 4 ms div to 1 mirra 125 MS s Glitch capture 4 ms div to 2 min div 8 ns Time Div in normal mode 4 ms div to 2 min div RECOM GEO 30 k points per waveform Recorded Time Span rra 4 8 s to 40 hours Acquisition Modes rra Single Sweep Continuous Roll Start Stop on Trigger Time Reference a time from start time of day Zoom Replay and Cursors Zoom Zoom ranges from full record overview to detailed view of individual samples Replay Displays a maximum of 100 captured quad input Scope screens Replay ua AEA Step by Step Replay as Animation Cursor Measurements Cursor Modes dosari aon aena aak single vertical cursor dual vertical cursors dual horizontal cursors Scope mode AEA ea automatic markers at cross points Measurements e value at cursor 1 e value at cursor 2 e difference between values at cursor 1 and 2 e time between cursors e RMS between cursors e Time of Day Recorder modes e Time from Start Recorder modes e Rise Time fall time e Axs current over time between cursors e Vxs voltage over time between cursors e W xs power over time between cursors using powerwa
106. rsion 5 34 Calibration Adjustment Final Calibration Firmware V10 9 and Lower Input D LF HF Gain Note The adjustment steps for channel D are only for the models 190 104 and 190 204 To do the Input D LF HF Gain calibration 1 2 Press ENIN to select the first calibration step in Table 5 15 Connect the Test Tool to the 5502A as shown in Figure 5 14 4 CHANNELS 0 0 Pm9585 50 OHM For 190 502 and 504 USE 50 OHM use only TRM50 TERMINATION 0 0 FLUKE 5502A CALIBRATOR N Oa gt Perf Ver G2 eps Figure 5 14 5502A SCOPE Output to Test Tool Input D Set the 5502A SCOPE output to source the signal reguired for the first calibration point in Table 5 15 Set the 5502A in operate OPR or standby STBY as indicated Press to start the calibration Wait until the display shows calibration status READY Press Zi to select the next calibration step set the 5502A to the next calibration point signal and start the calibration Continue through all calibration points of Table 5 15 and Table 5 16 When you are finished set the 5502A to Standby Continue at the nput ABCD AB LF HF Gain section 5 35 Fluke 190 062 102 104 202 204 502 504 Service Manual Table 5 15 Input D LF HF Gain Calibration Points Cal step UUT input signal 5502A Setting CL 0675 0 5 Vpp square wave 1 kHz SCOPE edge 0 5 Vpp 1 kHz CL 0486 0 62 Vpp sine wave 50 kHz SCOPE
107. s follows to reset the Test Tool 1 2 3 4 Wait until the Test Tool beeps twice and then release E Two beeps Press 0 to turn off the Test Tool Press and hold EI Press and release OD to turn on the Test Tool indicate a successful reset Performance Verification 4 Operation Instructions Menu Navigation During verification you must open menus and choose items from the menu Proceed as follows to make choices in a menu 1 Reset the Test Tool 2 Open a menu for example press EN and press Hed READING The menu shown in Figure 4 1 opens A yellow background or yellow characters mark the active functions If more than one menu group is available they are separated by a vertical line The menu shown indicates that READING 1 that is the upper left reading shows the result of a V rms measurement V ac dc on Input A on A 3 Press OD or to highlight the function to be selected Press to confirm the selection The active function in the next menu group is highlighted If the confirmation is made in the last most right menu group the menu will close READING 23 ws Read2 BMP Figure 4 1 Menu Item Selection Standard Test Tool Setup Before you start the verification procedure you must define a standard Test Tool setup for example SCOPE 1 During verification you will be asked to recall this setup This defines the initial Test Tool setup for each verification Press to confirm each setting N
108. s limits for working voltage They represent V ac rms 50 60 Hz for ac sinewave applications and as Vdc for dc applications The terms Isolated or Electrically floating are used in this manual to indicate a measurement in which the product input BNC is connected to a voltage different from earth ground The isolated input connectors have no exposed metal and are fully insulated to protect against electrical shock The BNC jacks can independently be connected to a voltage above earth ground for isolated electrically floating measurements and are rated up to 1000 Vrms CAT III and 600 Vrms CAT IV above earth ground Whenever it is likely that safety has been impaired the product must be turned off and disconnected from the line power The matter should then be referred to qualified personnel Safety is likely to be impaired if for example the product fails to perform the intended measurements or shows visible damage Li ion Battery Pack The battery pack has been tested in accordance with the UN Manual of Tests and Criteria Part III Subsection 38 3 ST SG AC 10 11 Rev 3 more commonly known as the UN T1 T8 tests and have been found to comply with the stated criteria The battery pack has been tested to EN IEC62133 Safe Storage of Battery Pack AA Warning To prevent electrical shock or fire e Do not store battery packs near heat or fire Do not store in sunlight e Do not remove a battery pack from
109. saso Electromagnetic Immunity aaa 3 Parts BO EEE WTFOCUGCTION A How to Obtain ac ffffffee Final Assembly PariS siiis ane ea aaa ea aoa n a EE TEER ACCOSSONV EN ara a dakoa irai gd darea 4 Performance Verification rra Ta Lig 010 0 03 0 g EE EE akeg ariak Equipment Requirements for Verification Gener l IMStFUCHONS AE Operation Instructions a Reset the Test TOOL rss 4 4 Menu Navigation ss Standard Test Tool Setup rra Display and Backlight Test sr Fluke 062 102 104 202 204 502 504 Service Manual scope Input A B C D REA Input A B C D Vertical Accuracy Tegi Input A B C D DC Voltage Accuracy Test rrura Input A B C D AC Voltage Accuracy Test F ss Input A B C D AC Coupled Lower Frequency TS Input A B C D Peak Measurements gisua Input A B C D Frequency Measurement Accuracy TGG Input A amp B C amp D Phase Measurements TGG Time Base TOSI AEE Input A Trigger Sensitivity TGG rrura Input A AC Voltage Accuracy HF and Bandwidth TGG Input B Trigger Sensitivity TS Input B AC Voltage Accuracy HF and Bandwidth TG Input C Trigger Sensitivity Test rrua Input C AC Voltage Accuracy HF and Bandwidth TS Input D Trigger Sensitivity TS Input D AC Voltage Accuracy HF and Bandwidth TS Video Test with SC600 Scope Calibration OOG External Trigger Level Seis Meter DC Voltage Accuracy Test rss Meter AC Voltage Accuracy and Frequency Response Test Continuity Function TS Diode Test Fu
110. select the setup name and press to recall the setup b Press score then press Fe READING and select READINGS on A V ac c Press W to select autoranging AUTO in upper right LCD edge d Use and J to change the sensitivity range to select manual sensitivity ranging and lock the Input A sensitivity range on 500 mV div AUTO in upper right LCD edge becomes AUTO e Use to move the Input A trace zero to the center grid line 3 Set the 5502A to source a sine wave and to the first test point in Table 4 10 Observe the Input A reading and check that it is within the range shown under the appropriate column Continue through the test points When you are finished set the 5502A to Standby Table 4 10 HF AC Verification Points 5502A 55024 SC MODElevsine MODE levsine UUT 2 545 Vpp 1 MHz 835 mV to 965 mV 2 545 Vpp 25 MHz 790 mV to 1 010 V 190 062 2 545 Vpp 60 MHz gt 630 mV KN 104 102 2 545 MHz mV 190 204 202 202 2 545 2 545 vop 200MHz MHz egan MV 190 502 504 2 545 Vpp 500 MHz gt 630 mV 4 23 Fluke 190 062 102 104 202 204 502 504 Service Manual Input B Trigger Sensitivity Test Proceed as follows to test the Input B trigger sensitivity 1 Connect the Test Tool to the 5502A as shown in Figure 4 8 PM9091 PM9091 50 OHM e g PM9585 or TRM50 For 190 502 and 504 use only TRM50 USE 50 OH
111. t align with the flat side of the BNC input sockets The strip has six holes When reinstalling the input cover reinstall the four steel pins 2x17 mm in the left side and right side of the Test Tool The pins are used to attach the hang strap and the side strap Remove the two screws M3x10 total length from the bottom holster The screws fit into square nuts that fit into the rear case Remove the bottom holster Note When reinstalling the holster reinstall the two steel pins 2x17 mm in the left side and right side of the instrument The pins are used to attach the side strap Take care that the yellow covers for the DC input and USB inputs are in place correctly 9 Remove the four self tapping screws 16 mm long total length that attach the rear case Two of these screws are located in the battery compartment 10 Remove the rear case Note When reinstalling the rear case do not forget to put the steel plate 16x17 mm in place again This plate is in the cavity on the right hand side of the Test Tool and can be used to attach a Kensington Lock When reinstalling the bottom case take care that the flat cables to the LCD and keyboard are not damaged between the case parts Disassembly Disassembly and Reassembly Procedures How to Access the Top Side of PCA Most of the measurement points are located on the top side of the PCA For access to this side remove the upper plate shielding lid 1 Remove the four s
112. tandby mode before making any connection between the calibrator and the Test Tool Dangerous voltages are present on the calibration source and connection cables during these steps Proceed as follows to test the ac voltage measurement accuracy 1 2 oa b E N Connect the Test Tool to the 5502A as shown in Figure 4 13 Select the Test Tool setup Press METER Press ca to open the Measurement menu and select V ac Press GD MANUAL MENGE Press to select MANUAL ranging Use cl and E to select the ranges Set the range to the first test point in Table 4 18 Set the 5502A to source the appropriate ac voltage Observe the reading and check that it is within the range shown under the appropriate column 10 Continue through the test points 11 When you are finished set the 5502A to O zero Volt and Standby Performance Verification 4 Meter Tests Table 4 18 Meter Volts AC Measurement Verification Points Range 5502A output V ac Frequency Meter Reading 60 Hz 494 0 to 506 0 500 0 mV 500 0 mV 1 kHz 486 0 to 514 0 3 kHz gt 350 0 60 Hz 4 940 to 5 060 5 000 V 5 000 V 1 kHz 4 860 to 5 140 3 kHz gt 3 500 60 Hz 49 40 to 50 60 50 00 V 50 00 V 1 kHz 48 60 to 51 40 3 kHz gt 35 00 60 Hz 494 0 to 506 0 500 0 V 500 0 V 1 kHz 486 0 to 514 0 3 kHz gt 350 0 60 Hz 0 980 to 1 020 1100 V 1 1 kV 1000 V 1 kHz 0 960 to 1 040 3 kHz gt 0 700 Continuity Function Test To test the continuity function
113. th 20 MHz other ranges d Press ENEM then press ENIN INPUT B OPTIONS e Select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the ScopeMeter 2 mV d and 5 mV d or Bandwidth 20 MHz other ranges f Press PTE and then press INPUT C OPTIONS g Select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the ScopeMeter 2 mV d and 5 mV d or Bandwidth 20 MHz other ranges h Press D then press SA INPUT D OPTIONS i Select Attenuator Normal Bandwidth 10 kHz or 20 kHz for the available setting in the ScopeMeter 2 mV d and 5 mV d or Bandwidth 20 MHz other ranges j Press to clear the softkey menu and see the full 8 divisions display 3 Press 4 Press N READING and select with Emu READINGS and the E Reading 1 on A V ac Reading 2 on B V ac Reading 3 on C V ac Reading 4 on D V ac Performance Verification 4 Scope Input A B C D Tests 5 Use to change the time base and lock on 20 us div for the 20 kHz signal and on 10 ms div for the 60 Hz signal 6 Use and G to select the manual vertical ranging Set the input A and B sensitivity range to the first test point in Table 4 4 The sensitivity ranges are indicated in the lower display edge Set the 5502A to source the appropriate ac voltage Observe readings A B C and D and check to see if they are within the range shown under the approp
114. th the Final Calibration 1 Press ENIN YES When you turn off and turn on the Test Tool again it will show the message The instrument needs calibration Please contact your service center The calibration date and number do not update You must continue with the Final Calibration To return to the Maintenance mode and repeat the complete calibration 1 Press NO 2 Press Ea until the display shows WarmingUp CL 0200 IDLE and calibrate the Test Tool starting at Warming Up and Pre Calibration If you want to exit and maintain the old calibration data 1 Turn the Test Tool off 5 43 Fluke 190 062 102 104 202 204 502 504 Service Manual Probe Calibration To meet full user specifications you need to adjust the supplied red R blue B gray G and green V VPS410 II 10 1 voltage probes for optimal response To adjust the VPS410 II probes 1 Connect the red probe from the red Input A BNC to the banana jack See Figure 5 19 Probe Cal Ref st8416_t eps Figure 5 19 10 1 Probe Calibration Connection Press and then HEM to open the Probe on A menu Select Probe Type W Voltage Attenuation W 10 1 Press DD Press ETT PROBE A Press PROBE CAL and follow the instructions shown on the display O a O N 5 44 Calibration Adjustment Probe Calibration Press E to start the probe calibration The first step is to manually adjust the square wave resp
115. uding SC600 Oscilloscope Calibration Option Stackable test leads 4x as supplied with the 5502A 50 Q Coax Cables 2x use Fluke PM9091 1 5 m 3 pcs set and PM9092 0 5 m 3 pcs set Male BNC to Dual Female BNC adapter 1x Fluke PM9093 001 50 Q feed through termination always use Fluke TRM50 for Fluke 190 502 and 190 504 Dual Banana Plug to Female BNC Adapter 1x Fluke PM9081 001 Dual Banana Jack to Male BNC Adapter 1x Fluke PM9082 001 TV Signal Generator part of SC600 600 MHz Oscilloscope Calibration Option 10 1 Attenuator Probes as supplied with Test Tool General Instructions Follow these general instructions for all tests Power the Test Tool with the BC190 power adapter The battery pack must be installed Allow the 5502A to satisfy its specified warm up period For each test point wait for the 5502A to settle Allow the Test Tool a minimum of 30 minutes to warm up One division on the LCD consists of 25 pixels 1 pixel 0 04 division This procedure is set up for all models of the Test Tool These have either two oscilloscope channels A and B with BNC inputs and a multimeter channel with banana inputs or four oscilloscope channels The figures that show the connection between calibrator and Test Tool are universal and also show the connection between calibrator and a Test Tool with four oscilloscope channels for instance model 190 204 Operation Instructions Reset the Test Tool Proceed a
116. uke reserves the right to invoice Buyer for importation costs of repair replacement parts when product purchased in one country is submitted for repair in another country Fluke s warranty obligation is limited at Fluke s option to refund of the purchase price free of charge repair or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period To obtain warranty service contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no risk for damage in transit Following warranty repair the product will be returned to Buyer transportation prepaid FOB Destination If Fluke determines that failure was caused by neglect misuse contamination alteration accident or abnormal condition of operation or handling including overvoltage failures caused by use outside the product s specified rating or normal wear and tear of mechanical components Fluke will provide an estimate of repair costs and obtain authorization before commencing the work Following repair the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges FOB Shipping Point THIS WARRANTY IS BUYER S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES
117. us e SECAM 622 even for SECAM e NTSC 262 even for NTSC Observe the trace and check if the Test Tool triggers on the negative pulse before the marker pulse see Figure 4 12 Use EOO to select Test Tool line number e 310forPAL PALplus or SECAM e 262for NTSC Set the calibrator format and marker line number to e PAL 310 odd for PAL and PALplus e SECAM 310 odd for SECAM e NTSC 262 odd for NTSC 15 16 17 18 19 20 21 Performance Verification 4 Scope Input A B C D Tests Observe the trace and check if the Test Tool triggers on the negative pulse before the marker Select the Test Tool setup and press Fa open the Trigger Options menu Choose VIDEO on A and from the opened menu choose Polarity NEGATIVE PAL or NTSC or PALplus or SECAM Set the calibrator video trigger output signal to 100 Use EOO to select line number 310 PAL PALplus or SECAM or 262 NTSC Set the calibrator format and marker line number to e PAL 310 odd for PAL and PALplus e SECAM 310 odd for SECAM e NTSC 262 odd for NTSC Observe the trace and check if the Test Tool triggers on the positive pulse before the marker ELSI NK teen eee eens video sc600 bmp Figure 4 12 SC600 Marker Pulse 4 33 Fluke 190 062 102 104 202 204 502 504 Service Manual 4 34 External Trigger Level Test Note The external trigger level test is for the models 190 062 190
118. uts 1 Press Fs to start the calibration 2 Wait until the display shows calibration status READY 3 Press F2 to select the next calibration step CL 0570 5 28 a GE e EO Calibration Adjustment Final Calibration Firmware V10 9 and Lower Connect the Test Tool to the 5502A SCOPE output as shown in Figure 5 11 Use the 50 Q termination Set the 5502A to generate a sine wave 50 MHz 0 5 V pp mode LEVSINE at the SCOPE output Set the 5502A to operate OPR Press ENIN to start the calibration Wait until the display shows calibration status READY Set the 5502A to standby STBY 0 Continue at the nput A LF HF Gain section 50 OHM e g PM9585 or trmso a For 190 502 and 504 only use Mg A METER B 2 CHANN METER 4 CHANNELS PM9091 USE 50 OHM TERMINATION 00 00 FLUKE 5502A CALIBRATOR Perf Ver D2 eps Figure 5 11 5502A SCOPE Output to Test Tool Input A Input A LF HF Gain To do the Input A LF HF Gain calibration 1 2 Connect the Test Tool to the 5502A as shown in Figure 5 11 The display must show step CL 0654 If it does not press Fo or to select the first calibration step in Table 5 9 Set the 5502A SCOPE output to source the signal required for the first calibration point in Table 5 9 Set the 5502A to operate OPR or standby STBY as indicated Press ENIN to start the calibration 5 29 Fluke 190 062
119. veform AxB or CxD Miscellaneous Display View Area 126 8 mm x 88 4 mm 4 99 in x 3 48 in Resolution 320 pixels x 240 pixels Ee Alea LED Temperature compensated elei u EEE Power Adapter 200 cd m Battery Power 90 cd m Display Auto OFF time battery saving 30 seconds 5 minutes or disabled Specifications D Miscellaneous A Power Fluke 190 xx4 50x Rechargeable Li ion Battery model BP291 Operating TIME ein eo semestre up to 7 hours Low Intensity Charging Vie vice crete 5 hours Capacity Voltage uarrornnrrnnvnnvnnvnnvnnennennvenrennrnnn 52 Wh 10 8 V Fluke 190 062 102 202 Rechargeable Li ion Battery model BP290 Operating TES rra up to 4 hours Low Intensity Charging TES rra 2 5 hours Capacity Voltage 26 Wh 10 8 V Rechargeable Li ion Battery model BP 290 and BP291 Life Time gt 80 capacity rra 300 x charge discharge Allowable ambient temperature during charging rr 0 C to 40 C 32 F to 104 F Auto power down time eue 5 min 30 min or disabled Power Adapter BC 190 e BC190 801 European line plug 230 V 10 e BC190 813 North American line plug 120 V 10 e BC190 804 United Kingdom line plug 230 V 10 e BC190 806 Japanese line plug 100 V 10 e BC190 807 Australian line plug 230 V 10 e BC190 808 Universal switchable adapter 115 V 10 or 230 V 10 with plug EN60320 2 2G e BC190 820 Universal adapter 100 240 V 10 with plug EN60320 2 2G Line EGO UE
120. x for instance 2516 This value determines that the higher frequency adjust point should be used e for x5xx use the later connectors e for x2xx use the early connectors 3 5 Fluke 190 062 102 104 202 204 502 504 Service Manual hpp201 eps Figure 3 1 Open Case and Screws 3 6 3 Te DD oo E Zu D O Qe D n n lt F S hpp202 eps Figure 3 2 Screening Plate Removed and Screws 3 7 Fluke 190 062 102 104 202 204 502 504 Service Manual hpp203 eps Figure 3 3 PCA Removed from Chassis Bottom Side Visible Accessory List For an Accessory list see Maintaining the Test Toolin the Fluke ScopeMeter 190 Series II Users Manual Go to www fluke com to download the users manual 3 8 Chapter 4 Performance Verification Title Page Jausi ee A 4 3 Equipment Requirements for Verification rss General atua ena E 4 4 Operation Instructions ra 4 4 Reset the Test a Menu Navigation aa 4 5 Standard Test Tool GGI 4 5 Display and Backlight TS 4 7 Scope Input A B C D Tests 4 8 Input A B C D Vertical Accuracy TGG 4 8 Input A B C D DC Voltage Accuracy TGG 4 11 Input A B C D AC Voltage Accuracy Test LF rss 4 14 Input A B C D AC Coupled Lower Frequency TGG 4 16 Input A B C D Peak Measurements TGG 4 17 Input A B C D Frequency Measurement Accuracy Test 4 18 Input A amp B C amp D Phase Measurements TGG 4 19 Tim
121. z 400 mV pp 100 MHz 800 mV pp 150 MHz 400 mV pp 200 MHz 800 mV pp 250 MHz 400 mV pp 500 MHz 800 mV pp 600 MHz 10 ns div 1 div 10 ns div 2 div 10 ns div 1 div 10 ns div 2 div 10 ns div 1 div 10 ns div 2 div 2 ns div 1 div 2 ns div 2 div Input B AC Voltage Accuracy HF and Bandwidth Test Proceed as follows to test the Input B high frequency automatic scope ac voltage measurement accuracy and the bandwidth 1 Connect the Test Tool to the 5502A as shown in Figure 4 8 2 Select the Test Tool setup a b c Recall the setup see Standard Test Tool Setup Press bel ES RECALL and select SETUP press Gi select the setup name and press to recall the setup Press score then press E READING select E READINGS 2 and select on B V ac MANUAL AUTO d e Press Use I A and l to select autoranging AUTO in upper right LCD edge to change the sensitivity range to select manual sensitivity ranging and lock the Input B sensitivity range on 500 mV div Press SZ and use Ea to select Input B as trigger source 4 3 Setthe 5502A to source a sine wave at the first test point in Table 4 12 4 Observe the Input B reading and check that it is within the range shown under the appropriate column of Table 4 12 5 Continue through the test points 4 25 Fluke 190 062 102 104 202 204 502 504 Service Manual 6 When you are finish
Download Pdf Manuals
Related Search
Related Contents
6 - unipac Sharp MX-M1100 MSDS Radlink GPS Hardware User Guide Copyright © All rights reserved.
Failed to retrieve file