Improving Throughput in Network Analyzer Applications
Contents
1. Figure 8 Adapter Removal Calibration Procedure A more accurate way to perform calibrations for noninsertable devices is to use adapter removal calibration Figure 8 outlines the main steps for performing this procedure The electrical length of the adapter must be specified within one quarter wavelength entered as a time value similar to electrical delay Type N 3 5 mm and 2 4 mm cali bration kits for the Agilent 8510 and 8720 family network analyzers contain adapters that are speci fied for this purpose Refer to the operating manual or on screen help text for the 8753 and 8720 net work analyzers for more information 18 Device Connection Time The time needed to disconnect a device and con nect a new one can be a significant portion of the total test process time especially for multiport test devices Most test processes will include one or more of the following e Manual connections e Part handlers e Multiport devices For two port devices that are measured with a transmission reflection test set the user must manually turn the device around in order to make measurements in the reverse direction Users can save time by using switching test sets which can switch the output power to either port so both forward and reverse measurements can be made with a single connection Note that switching test sets generally do not offer additional error correction capability so these measurements are still less accur
2. Instrument State Recall Speed Automation and Data Transfer Speed Measurement Accuracy Device Connection Time Instrument Uptime Product Quality Conclusion Checklist for Increasing Measurement Throughput Appendix Third Party Companies throughput issues and how they might affect differ ent applications It suggests ways to improve net work analyzer performance for better throughput in specific situations and how to get an accurate picture of how an analyzer s performance might impact overall throughput This application note broadly covers network ana lyzer throughput and applies to many different models of Agilent Technologies network analyzers Therefore for specific details on how to use certain features with the Agilent 8753 8711 or 8720 fami lies of network analyzers please refer to the oper ating and programming manuals for these products The level of information presented here assumes that the reader has some familiarity with network analyzers and their usage If you need basic infor mation please refer to the references listed in the appendix Throughput Considerations When considering throughput it is important to examine the overall measurement process which is likely to consist of a number of steps For example a production line might have a process where oper ators use network analyzers to perform manual tuning on filters The process might include connect ing a filter recalling a test setup o
3. continuous process improvement CPI to improve the quality of their products and to reduce waste and lower costs by finding problems earlier in the manufacturing process Network analyzers can help with this task by providing an efficient inter face for data collection Examples include provid ing hardcopy printouts saving data to disk files in easy to use formats or fast transfers of data to an external controller Traceability of a test instrument s performance is important to ensure quality especially if your company is ISO 9000 compliant If you are relying on a certain level of performance from the analyzer in order to make your measurements it is impor tant to note whether these are guaranteed instru ment specifications or only typical instrument performance values that might vary from one ana lyzer to another Variation in instrument perform ance will affect the consistency and repeatability of measurements made on different production lines 20 Another issue is whether the network analyzer s specifications are sufficiently complete to deter mine the accuracy of your measurements Some network analyzers only specify the dynamic accu racy the uncorrected systematic errors or the residual errors after a calibration However having only one of these specifications is not enough to determine total measurement uncertainty To get the total measurement uncertainty for a calibrated transmission or reflection m
4. 10 dBm 300 Hz Or Segment 5 129 ms 38 points 10 dBm 300 Hz Ne Segments 2 4 52 ms START ya 000 MHz STOP 1 Kis MHz No specs here so no points tere No specs here so no points measured in this span 15 points 10 dBm 300 Hz measured in this span Figure 6 Linear Sweep vs Swept List Frequency Filter Measurement Averaging Use the minimum number of averages necessary for the measurement Averaging can be useful for reducing noise and improving dynamic range But it might also be helpful to compare the effects of using a narrower IF bandwidth versus averaging to achieve the same noise reduction to see which yields a faster measurement Type of calibration Choose the fastest type of calibration for the required level of accuracy For most network analyzers sweep speed is about the same for uncorrected measurements and measurements done using a response calibration enhanced response calibration or one port calibra tion However sweep speed might be at least twice as slow for a full two port calibration A full two port calibration requires both forward and reverse sweeps to update all four S parameters for error correction even when only a single S parameter is displayed So use the calibration that yields the fastest sweep speed for the desired level of accuracy See the section on Measurement Accuracy for more details 10 Fast two port mo
5. Ant ports to analyzer Recall Tx gt Ant state and cal NO Measurement 1 YES Lock adjustment screws y Verify Tx gt Ant measurements i Recall instrument state and cal for Ant gt Rx T Verify Ant gt Rx measurements Disconnect filter and adjustments 90 sec meas time 210 sec adjust time 15 sec connection 3 sec recall Total for all measurements and adjustments 90 sec meas time 210 sec adjust time 15 sec connection 3 sec 60 sec Total for all tests 20 sec 3 sec Total for all tests 20 sec 15 sec antenna Ant ports and the one between the Ant and receive Rx ports With a standard two port network analyzer two filter ports are measured at a time with a load termination connected to the unused port Some steps have been left out for simplicity Measurement time 28 Recall 2 8 Connection Adjustment 62 This pie chart shows that with one pass of tuning measurement time is only about one quarter of the total throughput time Figure 1 Example Manual Tuning Process for a Base Station Filter These two examples demonstrate how throughput issues might be different in various applications For manual tuning faster sweep speed is impor tant However once the user perceives a continu ous update of data approximately 30 updates per second any faster upda
6. fixtures If you already use a part handler the easiest way to speed up connection time is to use a faster part handler although that will probably be more expen sive Another improvement is to consider using the analyzer s internal automation capability to control the part handler instead of relying on an external controller as described in the section on Automation and Data Transfer Speed Using internal automation can be faster since no data has to be transferred outside of the analyzer before a decision can be made and a command sent to the part handler For multiport devices some operators use the analyzer to test two ports at a time with termina tions at the unused ports and then switch the cable connections around to make the other necessary measurements For devices with many ports this process can be very tedious and time consuming and it can also contribute to operator fatigue One alternative is to use a multiport test set that allows the operator to make connections to all of the ports once and then have the analyzer make all necessary tests without changing connections Agilent provides a variety of test sets for its net work analyzers including the 8753D Option K36 three port duplexer test set also available for the 8711 8752C and 8720 families the 87075C 75 ohm multiport test sets with the innovative SelfCal feature for the 8711C family and the 87050A B series of 50 ohm test sets for the
7. measurement does not require a low noise floor This type of calibration provides the best accuracy but it does require the analyzer to take both a forward and reverse sweep to update all four S parameters for each updated measurement display Two port calibration will slow down the perceived sweep speed since it effectively takes two sweeps for every trace update instead of one Another form of two port calibration uses through reflect line TRL standards This method is pri marily used in noncoaxial environments such as measurements in test fixtures or on wafer It requires a network analyzer with four receivers such as the Agilent 8720D with Option 400 or 8510C There are a number of variations of TRL calibration includ ing TRL for network analyzers with only three receivers LRM using line reflect match standards or TRM using through reflect match standards From a throughput standpoint TRL calibration and its variations have the same effect on sweep speed as a full two port calibration because it also requires measurement of all four S parameters to calculate corrected data for each displayed sweep update When performing two port calibrations you may need to perform the isolation portion of the cali bration for the best dynamic range For isolation at least 16 averages are recommended Turn aver aging on only during the isolation portion if you do not need it for other calibration standards Turn averaging off prior
8. source and receiver to take a data sweep If the analyzer is in hold mode while the recall is 12 being done it usually won t take the time to set up for a new sweep However as soon as you trig ger the analyzer to take a sweep the analyzer has to do the setup so the time for a recall with single sweep is often significantly longer than the time for just a recall Realistically you will need to know the time for recall with single sweep to approxi mate your real measurement conditions One way to reduce recall time in some network analyzers is to turn off spur avoidance before stor ing the instrument state This is a feature in many network analyzers to reduce low level spurious signals You can check if this is needed for your measurement by seeing if your data changes with spur avoidance on or off Turning spur avoidance off allows the analyzer to bypass the calculations and setup that are needed during an instrument state recall making the recall faster Similarly you can turn off other hardware corrections such as sampler correction If you do you should calibrate and make measurements under the same condi tions so that the calibration can compensate for the lack of hardware correction On some newer network analyzers a very effective way to reduce recall time is to turn the display off since the analyzer does not spend processing time to display the new instrument state For example a typical simple instrument
9. state that takes an 8753E about 0 4 seconds to recall with the display on takes only 0 2 seconds to recall with the display turned off The amount of speed improvement will vary depending on the instrument state conditions If an external controller is being used to control the test it might be faster or more convenient to use the analyzer s learn string to quickly save the current instrument state or restore a previous state The learn string is a compact data string that includes the front panel settings but not calibra tion or memory trace data Learn strings might not be compatible between different models of network analyzers so you need to be careful if your envi ronment includes a mix of network analyzers For more details consult the programming manual for your network analyzer Recalling an instrument state might not be the fastest way to set up and make a new measure ment For example with a two channel network analyzer you can uncouple the channels and set up two different instrument states on the two channels such as two frequency ranges or differ ent numbers of points You will need to check whether it is faster to switch from one channel to the other or to do an instrument state recall to obtain the second instrument state Another example is when two instrument states only differ slightly from each other for example when you only need to change a few settings from the factory preset state It might be faste
10. 1997 Calibration for PC Board Fixtures and Probes Joel Dunsmore 45th ARFTG Conference Digest Spring 1995 Techniques Optimize Calibration of PCB Fixtures and Probes Joel Dunsmore Microwaves amp RF October 1995 pp 96 108 November 1995 pp 93 98 Improving TRL Calibrations of Vector Network Analyzers Don Metzger Microwave Journal May 1995 pp 56 68 The Effect of Adapters on Vector Network Analyzer Calibrations Doug Olney Microwave Journal November 1994 Third Party Companies Test Fixtures Inter Continental Microwave 1515 Wyatt Drive Santa Clara CA 95054 1524 USA E mail icmfixture aol com Phone 408 727 1596 Fax 408 727 0105 Multiport Test Sets ATN Microwave Inc 85 Rangeway Road North Billerica MA 01862 2105 USA E mail info atn microwave com Phone 978 667 4200 Fax 978 667 8548 www atn microwave com Wafer Probes and Stations Cascade Microtech 14255 SW Brigadoon Court Beaverton OR 97005 USA E mail sales cmicro com Phone 503 626 8245 Fax 503 626 6023 23 By internet phone or fax get assistance with all your test and measurement needs Online Assistance www agilent com find assist Phone or Fax United States tel 1 800 452 4844 Canada tel 1 877 894 4414 fax 905 206 4120 Europe tel 31 20 547 2323 fax 31 20 547 2390 Japan tel 81 426 56 7832 fax 81 426 56 7840 Latin America tel 305 2
11. 69 7500 fax 305 269 7599 Australia tel 1 800 629 485 fax 61 3 9272 0749 New Zealand tel 0 800 738 378 fax 64 4 495 8950 Asia Pacific tel 852 3197 7777 fax 852 2506 9284 Product specifications and descriptions in this document subject to change without notice Copyright 1998 2000 Agilent Technologies Printed in U S A 5 00 5966 3317E Agilent Technologies Innovating the HP Way
12. 720D family provides this capability with Option 012 direct sampler access As shown in Figure 4 you can connect the output of your device under test directly into the B sampler instead of to port 2 This direct connec tion increases your dynamic range by about 20 dB which is the amount of the coupling loss Sweeps per second Agilent 8753E Full 2 port Cal Sweep Update Time 201 points 8 75 L 7L 6 5 L 6 L 3 7 kHz IF BW 5 5 L i oF 3 kHz a5 L IF BW L E 6 kHz IF BW 4 1 1 1 1 1 0 015 0 02 0 025 0 03 0 035 0 04 Typical Trace Noise dB peak to peak 0 045 Figure 3 IF Bandwidth vs Trace Noise and Sweep Speed For analyzers such as the 8753 you can get a special version of the test set with the coupler reversed on port 2 see Figure 5 The reversed coupler will improve the sensitivity because the signal entering port 2 is routed to the sampler via the through arm of the coupler with a few dB of loss rather than the coupled arm which has a loss equal to the coupling factor typically 15 to 20 dB The output power from port 2 will now be reduced by the coupling factor so reverse direction measurements will have less dynamic range than normal which is why this configuration is only recommended if high dynamic range is needed in one direction The same noise floor improvement can be obtained for measurements in the reverse direction by reversing the port 1 coupler with the cor
13. 8711 8753 or 8720 families Other test set designs provide two sets of test ports for one network analyzer so that the operator can connect a new device while another device is being tested Some vendors offer multiport solutions for Agilent network analyzers For example the SPTS 4 four port S parameter test system from ATN Microwave see Appendix provides full four port error corrected measurements with an 8753 network analyzer Users can also build their own test sets to switch signals to and from the network analyzer to the proper ports of the test device Instrument Uptime Many people overlook an instrument s availability when considering throughput but it is actually a very important part A fast network analyzer has little value if the analyzer breaks or has to be sent out for service or maintenance frequently causing a production line to shut down or requiring arrange ments for spare instruments When making a pur chase consider instrument quality expected fail ure rate calibration interval and turn around time for repair or calibration all of which add to the maintenance cost The location of the nearest serv ice center and the availability of on site repair are some of the factors in the turn around time Product Quality Making measurements faster and increasing throughput is not the only concern for manufac turing companies Many companies are also using techniques like statistical quality control SQC or
14. Agilent AN 1287 5 Improving Throughput in Network Analyzer Applications Application Note ss Agilent Technologies f Innovating the HP Way 12 13 15 18 20 20 20 21 22 23 Introduction In today s competitive environment prices for electronic components are continually decreasing Reducing manufacturing cost by improving through put while maintaining product quality is an impor tant goal for many production test engineers and managers The topic of improving throughput is very broad and it can span methods from how to minimize testing and the number of specifications to using just in time JIT manufacturing with a Kanban inventory control system This note will not cover broad throughput issues such as whether distributed testing versus centralized testing is more efficient or cost effective Instead this appli cation note will focus only on test processes that include network analyzers Even within the relatively focused topic of network analyzer applications many factors need to be con sidered when deciding how to improve throughput It isn t always as simple as analyzing instrument specifications and choosing a network analyzer with the best measurement speed per data point It is also important to consider all the aspects of throughput that may be applicable for your situa tion This application note explores a variety of Table of Contents Introduction Throughput Considerations Sweep Speed
15. a specific measurement It includes the frequency range number of points IF band width power level and other front panel settings It may also include calibration data and memory traces Recalling an instrument state is a quick way to set up an instrument for a particular meas urement The fastest recalls are done from the ana lyzer s internal memory but recalls can also be done from a floppy or hard disk file or from an external controller Recall speed depends greatly on the content of the memory register or instrument state that s being recalled More complicated states will take longer For example a simple instrument state with a measurement on one channel only and no calibration can be recalled much faster than one with measurements set up on both channels with full two port calibration and limit lines and limit testing turned on For the Agilent 8753E network analyzer the recall times for these two states are about 0 5 seconds and 0 9 seconds respectively with mostly preset conditions and no optimiza tion Therefore it is very difficult to specify a single number for instrument state recall speed It is best to examine the recall time for the instru ment state that is needed for the application In many cases you might see times given for just recall rather than recall with single sweep These times may be quite different because at some point the analyzer needs to take time to actually set up the
16. alibration to several hours for a series of two port calibrations for testing a high dynamic range multiport device For this application note throughput considera tions are divided into the following topics e Sweep speed e Instrument state recall speed e Automation and data transfer speed e Measurement accuracy e Device connection time e Instrument uptime e Product quality Each topic will be described in greater detail and suggestions for improving throughput in each area will be provided Sweep Speed Sweep speed also referred to as sweep time can be a confusing term not everyone means the same thing when referring to sweep speed In general sweep speed refers to the amount of time needed for the analyzer to take one sweep of the source and acquire data over the defined range Many analyzers technical specifications report a num ber in the form of time per data point which one might assume would yield the sweep time when multiplied by the number of points in a trace Many instruments also have a function that reports a value for hardware sweep time However users may never get this sweep time in their measure ments because in reality what they will get is the cycle time This cycle time includes sweep hardware set up time band switch times when the source or receiver crosses frequency bands data acquisition time retrace time for the source to move from the end of one sweep to the start of the n
17. ate than those made with an S parameter test set If manual connections are being used for a two port device it is possible to speed up the connec tion time using a part handler to automate the process Gravity feed part handlers tend to be faster and less expensive than pick and place part handlers but they are more limited in the types of device packages they can handle Using a part handler generally requires a custom test fixture which adds development time and can also add calibration time so this might not be helpful in all situations For devices that require coaxial connections push on connectors can make the connection faster but they will be less repeatable and can result in more measurement uncertainty Special test fixtures that allow a user to connect devices more quickly can be useful with or with out a part handler Many companies make their own fixtures since they have many custom packages for their devices When designing a test fixture it is important to have good RF performance low loss and low parasitics and ease of calibration within the fixture should be considered Because of the difficulty in making good RF fixtures full two port calibration is generally required requiring a set of in fixture calibration standards There are some vendors who specialize in making test fixtures and calibration standards The appendix contains some references for vendors and more information on designing and calibrating
18. certainty so you can reduce guard bands and still have less likelihood of incorrect pass fail results Tighter guard bands improve throughput by allowing more devices to pass without sacrificing quality These same devices might have failed test limits based on wider guard bands when they were actually good devices For applications that do not require the highest accuracy analyzers with transmission reflection test sets such as the 8711C family or the 8752C can be an economical solution These analyzers offer the types of calibration methods listed in the table below The time required for performing the calibration will vary depending on the number of calibration standards that need to be measured Calibration Type Errors Corrected Number of Standards Required Response Reflection OR transmission 1 Frequency response tracking Response and Isolation Reflection Tracking and directivity 2 OR Transmission Tracking and crosstalk One port Reflection only directivity source match 3 and reflection tracking Enhanced Response Transmission tracking and source match 4 Available only in newer AND Agilent network analyzers such as the 8711C family Reflection directivity source match and reflection tracking These calibration methods are good for improving throughput because they have almost no impact on sweep speed and the calibrations themselves are quick and easy to perform Since these meth ods only corr
19. de For faster tuning with full two port calibration minimize reverse sweeps If a full two port calibration is used for a tuning application the sweep speed and trace update time can be improved by using a feature in some Agilent network analyzers called fast two port mode Normally the analyzer will switch the out put power sequentially between port one and port two in order to measure all four S parameters which is necessary for calculating the corrected results with two port calibration This means it takes the analyzer two sweeps one forward one reverse before it can update the trace With fast two port mode you can specify how many forward sweeps the analyzer should take before it switches the power to port two to take the reverse sweep The analyzer will then update the trace on every forward sweep using data from the last reverse sweep until it takes the next reverse sweep This makes it twice as fast until the reverse sweep is taken Fast two port mode can also be used to tune reverse parameters by specifying the number of reverse sweeps to take before the analyzer takes a single forward sweep Fast two port mode provides a more real time response for tuning It gives good results because the reverse S parameters only have a secondary effect on the corrected forward S parameters Generally updating the reverse parameters less often will not cause large errors on the forward parameters All data is fully error corrected i
20. e network analyzer sweeps a segment instead of stepping the source result ing in a faster measurement b Ability to change IF bandwidth and power level for each segment For regular list frequency mode the same IF bandwidth and power level are used for all segments in the sweep Swept list mode includes a feature that allows you to choose a higher power level and smaller IF bandwidth in segments where better dynamic range is needed such as in the reject bands of a filter You can use a wider IF bandwidth and lower power for faster measurements in segments with high level Cow loss signals such as in the passband of a filter The ability to change power levels can be especially helpful for a device such as a filter combined with a low noise amplifier where high power is desired for measuring the reject bands but lower power is needed in the passband to avoid damaging the amplifier or the analyzer s receiver When the best dynamic range is not needed you can also use higher power with a wider IF bandwidth for measuring filter stop bands to provide adequate dynamic range while sweeping more quickly Segment 3 29 ms 108 points 10 dBm 6000 Hz CH1 S21 log MAG 12dB REF 0dB tal PRm a Linear sweep 676 ms 201 pts 300 Hz 10 dBm SS ae 201 Swept list sweep 349 ms pts variable BW s amp power PASS Segment 1 87 ms 25 points
21. easurement you need to combine the effects of dynamic accuracy with other system errors such as the residual systematic errors Most Agilent network analyzers provide graphs of the total measurement uncertainties for a test system based on particular test port connectors Conclusion The task of improving throughput while maintain ing product quality requires the consideration of many different aspects of the network analyzer and the test device besides the most obvious data sheet items such as the microseconds per point sweep speed Optimizing the important areas for each application can provide a more thorough and effective way to improve overall throughput The following checklist is a brief summary of the key points presented in this application note A Checklist for Increasing Measurement Throughput Sweep Speed Use the widest IF bandwidth with acceptable dynamic range and trace noise For better dynamic range change setup to bypass coupler loss Use the highest source power that does not overload the device or network analyzer Choose smaller frequency spans to minimize band switches Use the minimum number of points Use swept list mode including setting IF bandwidth and power for each segment Minimize use of averaging and compare speed of averaging vs using smaller IF bandwidth Choose the fastest type of calibration for required level of accuracy For tuning while using full two port calibra tion try fast two
22. ect for some of the errors that might be present in a measurement they are best suited to certain types of devices For example devices that have very good input and output match will be less affected by source and load match errors so response or one port calibrations can yield good results One port calibrations can also yield good results for devices with high loss in the transmis sion path or high isolation between ports However a device that has low insertion loss will have its 16 measurements affected by source and load match errors For example a filter that has low insertion loss in its passband will show ripples in the meas urement due to these errors Some of these ripples might have peaks with magnitudes greater than 0 dB indicating gain in a passive device which is clearly an error For better accuracy a network analyzer with an S parameter test set is needed such as the 8753 or 8720 families These systems can provide full two port error correction such as short open load thru SOLT calibration SOLT calibration corrects for twelve errors reflection tracking directivity source match transmission tracking load match and crosstalk in both the forward and reverse directions Twelve measurements need to be made using four known standards to correct for all of these errors so it takes longer to perform a cali bration although the two measurements required for the crosstalk correction can be omitted if the
23. ernal controller In newer analyzers with faster CPUs internal calculation time can be faster than the time needed to transfer data out and do the cal culation in an external controller However in some cases it might be better to do the error correction externally One example is for multi port applications where many different calibra tions are required for each test device and there might not be enough room in the analyzer s mem ory to store all the required correction arrays Measurement Accuracy This application note assumes that the reader has some familiarity with the concepts of measurement errors and error correction or calibration in network analyzers For more details refer to Application Note 1287 3 Applying Error Correction to Network Analyzer Measurements The appendix also lists other references on calibration To review briefly measurement accuracy or uncertainty can be thought of as how close a measurement is to the true or correct value you are trying to measure No network analyzer is per fect The factors that contribute to measurement uncertainty can be grouped into the following types of errors e Systematic Caused by imperfections in the test equipment and test setup These are gener ally repeatable and can be characterized and removed through calibration also called error correction e Random Errors that vary randomly as a func tion of time including connector repeatability and chan
24. ext one data calculation and formatting time and display update time see Figure 2 Also error correction time might not be included and if two port calibration is used the analyzer might need to take two sweeps instead of one for each display update see the Measurement Accuracy section for more details on calibration So for the purpose of consistency in this application note sweep speed refers to cycle time unless otherwise stated Also it is important not to assume that an analyzer s sweep speed under actual test condi tions will be the same as the time per point num ber published in the technical specifications In most technical specifications the value reported is a best case number Often it is measured at the instrument s widest IF bandwidth which might have too much trace noise and too little dynamic range to be useful with a single band sweep to avoid band switch delays and with the highest number of points to spread out the effects of over head items such as sweep set up time and obtain the smallest time value per point Actual sweep speed is closely tied to an instrument s set up parameters including the number of points and frequency range and the degree of accuracy and amount of dynamic range required which also impact the type of calibration necessary Data calculation and formatting Y Display update N E t t Sweep and da
25. ges from movements of cables These cannot be removed by calibration e Drift Errors due to temperature changes or drift over time These errors can be removed by repeating the calibration Network analyzers offer a variety of calibration methods that remove some or all of the systematic errors Calibration methods that correct more errors also take more time to perform since more calibration standards need to be measured More accurate calibrations can also slow down the meas urement time so the user needs to compromise between the desired measurement accuracy and the test process speed including calibration time How often to calibrate is another issue Recalibration will correct for drift errors which may be caused by changes in the hardware over time temperature changes in the environment or changes in the test setup such as movement of cables How often a new calibration is required will depend mostly on the environment and the desired level of accuracy Many users perform validation checks by measur ing a verification device If the measurement falls within acceptable limits the previous calibration is still considered good Agilent provides verification kits that contain devices with factory measured data that can be used for this purpose The level of accuracy that is required depends on the application tuning vs final test and the specifications of the device under test Better accu racy means lower measurement un
26. he analyzer to sweep as quickly as possible for the current set tings Some analyzers also allow you to specifi cally select either swept mode or stepped mode Use swept mode when possible since this will be faster However some measurements might require slower sweep time especially measure ments through devices with long electrical delay such as cables or surface acoustic wave SAW devices The slower sweep time can be set either by selecting stepped mode or by entering a longer sweep time value You can verify if the device needs a slower sweep time by examining the measurement results using both the faster and slower sweep speeds If there is no signifi cant difference then it is acceptable to use the faster setting for that measurement Unnecessary functions Turn off unnecessary functions to reduce sweep time Sometimes you turn on a feature when designing a test but later on you might forget to turn it off when the feature is no longer needed This might cause the analyzer to take extra time to update infor mation that s not being used For example turn off unused markers averaging smoothing limit tests or measurements of other parameters if they are not needed For some analyzers turning off the display in an automated environment might result in faster measurements Instrument State Recall Speed An instrument state is a particular set of stimulus and response parameters that controls how an analyzer makes
27. ion and data transfer speed 1 Use the analyzer s single sweep mode to ensure that a measurement is complete before starting data transfer Otherwise the analyzer might send data to the PC in the middle of a sweep so the data received by the PC is a mixture of data from the old sweep and the new one 14 2 Pick a data format and the associated commands that provide the fastest transfer speeds for the application The number of bytes per data point that need to be transferred depends on the for mat The analyzer s internal format is usually the fastest but it requires reformatting in a PC to be interpreted ASCII data transfers are the slowest Use the fastest data transfer method available Many Agilent network analyzers have fast data transfer commands that may be helpful in cer tain cases because they transfer an array as a block compared to the usual byte by byte transfer Transfer the minimum amount of data needed Users should try different methods to see what yields the fastest results in their application For example it might be faster to transfer a trace with a only a few points in it possibly using a frequency list instead of using markers to read out data Some analyzers also have a command for obtaining the maximum and mini mum values within each limit line segment which can yield sufficient data Consider whether error correction should be done internally in the analyzer or in an ext
28. mme diately after the reverse sweep is taken Sweep modes Chopped vs alternate mode Use alternate sweep instead of chopped mode for better dynamic range The default sweep mode in most Agilent net work analyzers is chopped mode in which both input ports are measured when active using their corresponding samplers during one sweep by measuring on one sampler and then switch ing to the other at each point Chopped mode provides the fastest measurements but it might not be the best mode in all situations There is also a mode called alternate sweep in which only one sampler is measured during a sweep The analyzer measures the other sampler during the next sweep Alternate mode is slower but it provides the best dynamic range by turning off the unused sampler to reduce crosstalk It is also selected automatically when the measure ment channels are uncoupled so two different instrument states can be measured on the two channels sequentially Using alternate mode can yield faster results than using a lower IF band width with chopped mode to get better dynam ic range or recalling an additional instrument state to make another measurement Swept vs stepped sweep Use swept mode to minimize sweep time when possible Many analyzers can also do a frequency sweep in swept mode stepped mode or a combination of both depending on the instrument state settings Setting the sweep time to auto mode usually the default causes t
29. ng the number of points will not further reduce the sweep time List frequency sweep Use list mode to focus test data where you want it List frequency sweep allows you to define an arbi trary list of frequency points at which the analyzer makes measurements This capability is useful for optimizing sweep time because you can choose a larger number of sweep points in frequency ranges of interest while minimizing the number of points for ranges that are not as important Port2 R Channel Jumper Typical standard test set configuration R Channel Jumper Test set with port 2 coupler reversed Figure 5 Improving Dynamic Range with a Reversed Port 2 Coupler For example in a filter measurement you might choose to measure many points in the reject bands and in the passband but very few points on the skirts of the filter The frequency list can even skip over frequency ranges where no data is needed This will enable you to get the detail you want with fewer total points measured Also you can choose to sweep a single segment in the list with out losing calibration or needing to interpolate the calibration data Some network analyzers such as the Agilent 8753E also offer an enhanced version of this mode that provides two additional features a Swept list mode Many network analyzers normally default to a stepped sweep mode when list fre quency is used which slows the analyzer down In swept list mode th
30. nnection time but will probably require test fixtures Be aware of fixture design and calibration considerations Consider using multiport test sets to simplify connections Instrument Uptime Choose analyzer with low failure rate fast turn around time and reasonable cost for repairs and calibrations Maintaining Product Quality Use easiest ways to collect necessary data from the analyzer printouts data transfers to PCs etc Make sure analyzer has necessary specifica tions to guarantee the desired level of accuracy 21 Appendix Related Agilent Application and Product Notes Understanding the Fundamental Principles of Vector Network Analysis Application Note 1287 1 Exploring the Architectures of Network Analyzers Application Note 1287 2 Applying Error Correction to Network Analyzer Measurements Application Note 1287 3 Network Analyzer Measurements Filter and Amplifier Examples Application Note 1287 4 In fixture Microstrip Device Measurements Using TRL Calibration Product Note 8720 2 Specifying Calibration Standards for the Agilent 8510 Network Analyzer Product Note 8510 5A Applying the Agilent 8510 TRL Calibration for Non Coaxial Measurements Product Note 8510 8A Measuring Noninsertable Devices Product Note 8510 13 22 Suggested Reading Design of an Enhanced Vector Network Analyzer Frank David et al Hewlett Packard Journal April
31. port mode to minimize reverse sweeps Try using alternate sweep instead of chopped mode for improving dynamic range Use swept mode instead of stepped mode and minimize sweep time when possible Turn off unnecessary functions like markers averaging smoothing limit tests unused parameters Instrument State Recall Turn off spur avoidance and hardware corrections Turn off display for newer network analyzers For automated test try using learn strings instead of recalling instrument states Consider using uncoupled channels to set up two instrument states instead of using a recall Consider using list frequency mode instead of recalling instrument states with different frequency ranges Automation and Data Transfer Consider using internal automation where possible Use fastest data format for data transfers Use any available fast data transfer commands Transfer minimum amount of data needed Consider whether to use internal error cor rection or to use an external computer for calculations Measurement Accuracy Use calibration type that gives you the best compromise between measurement speed and accuracy Calibrate as close to the device under test as possible Use adapter removal calibration where appropriate Device Connection Part handlers may speed co
32. r instrument state on the analyzer tuning and watching for a desired result indicated perhaps by markers that display values or limit lines that display whether a device passes or fails More tuning might be nec essary then the operator might move on to a dif ferent instrument state to tune another parameter and so on See Figure 1 Another example is an automated final test that uses a part handler Here the process might include the part handler placing a part in a test fixture setting up an instrument state taking data trans ferring data to a PC having a PC perform calcula tions or store data to a file and then setting up an analyzer for the next test The instrument state setup might be done by recalling an instrument state file that had previously been set up and stored or the PC could issue commands to the analyzer to set up the desired conditions Connect Tx Ant filter ports to analyzer 15 sec Recall instrument state and calibration 3 sec roo Total for multiple measurements Adjust screws to This is a simplified example of a manual tuning process for a base station duplexer filter The filter is a 3 port device with two main signal paths of interest the one between the transmit Tx and tune Tx gt Ant path j Connect Ant Rx ports to analyzer Recall new state and calibration i Adjust screws to tune Ant gt Rx path i Connect Tx
33. r to just change those settings instead of recalling a new instrument state Automating these changes with remote commands via GPIB or built in automation features such as test sequencing can help make changes easy and repeatable If an application requires measurements over a series of different frequency ranges consider using list frequency mode instead of instrument state recalls Each desired frequency range can be set up as a segment in the frequency list All of the segments can be calibrated at once and afterwards you can choose to sweep any one of the segments individually without losing the calibration instead of having to recall a series of different instrument states Automation and Data Transfer Speed Sooner or later most production managers consider automating part or all of their test processes to improve throughput An important part of test process development is deciding what and how much to automate and deciding on the method of automation The first decision is whether to use some form of automation internal to the network analyzer or to use some type of external controller The main choices are 1 External controller for example a PC or workstation 2 Internal programming language for example built in IBASIC in the Agilent 8711 family of network analyzers 3 Other internal automation for example test sequencing in the Agilent 8752 8753 or 8720 network analyzer families External automation
34. responding loss of dynamic range for forward measurements Using one of these special test set configurations allows you to use a higher and faster IF bandwidth to achieve the same dynamic range compared to a standard test set so you can use these configura tions to get faster measurements even if you don t need the improved dynamic range to test your device s specifications Agilent 8720D Option 012 Test Set Configuration R Channel Jumper Measure filter rejection to 120 dB by connecting directly to B sampler Figure 4 Improving Dynamic Range with Direct Sampler Access Source power Use the highest source power that does not overload the device or network analyzer To extend the upper limit on dynamic range use the highest source power from the network analyzer that will not overload the device under test or cause the analyzer s receiver to overload Frequency span Choose smaller frequency spans that minimize the number of band switches Test only the frequency spans that are necessary for your device Information on the band switch frequencies for each network analyzer can usually be found in the operating or service manual Number of points Use the minimum number of points required for the measurement For most analyzers sweeping fewer points results in less time per sweep However network analyzer sources have a maximum sweep rate limited by the hardware Once this limit is reached reduci
35. ta acquisition Band switches Diagram notto scale Retrace Figure 2 Components of Cycle Time Here are some ideas to optimize sweep speed and cycle time IF Bandwidth Use the widest IF bandwidth with acceptable dynamic range and trace noise Wider IF bandwidths result in faster measure ments but they also give you more trace noise ripples in high power level measurements and higher noise floor less dynamic range Typically a ten fold reduction in IF bandwidth will give you a 10 dB reduction in the noise floor Use the widest IF bandwidth that will give you reasonable results especially with regard to trace noise and dynamic range Figure 3 shows an example of some typical relationships between IF bandwidth trace noise and sweep speed for the Agilent 8753E RF network analyzer Note that narrowing the IF bandwidth in some Agilent network analyzers such as those in the 8753 and 8720 families has the same effect as increasing point by point averaging in other ana lyzers such as the 8510 In the 8753 and 8720 families the averaging feature performs a trace by trace average Refer to the operating manuals for these analyzers for more details Test Set Changes Consider special test set configurations for higher dynamic range If a lower noise floor is required only in the for ward direction you can configure the test set to bypass the usual coupler loss on port 2 for trans mitted signals The Agilent 8
36. te speed would not be noticeable or result in a increase in filter through put Conversely faster sweep speed could be useful in automated testing where a computer is faster than an analyzer Part handler speed and data transfer speed are not relevant to the manual tun ing application but the time needed to manually connect the test device is relevant The time required to recall or set up an instrument state is important in both applications It is also important to consider the relative impor tance and value of improving each part of the process Many people focus on the sweep time of a network analyzer when trying to improve through put but improving sweep time alone does not always provide the best throughput improvement For example in a multiport test application if it takes the operator 1 5 minutes to connect a new device into place while the analyzer takes 10 sec onds to perform the test then cutting the analyzer test time in half only reduces the total test time from 100 seconds to 95 seconds for a 5 improve ment However reducing the device connection time to 1 minute will reduce the total test time to 70 seconds which is a 30 improvement The complete test process might include additional items such as calibration time that are not part of testing every device but they might need to be done occasionally and will affect the overall throughput Calibration time can range from a few minutes for a simple one port c
37. tion Transmission Tracking ZF e Crosstalk Va Source match e Load match Test Set cal type 4444 S parameter two port Figure 7 Correctable Errors for Different Test Sets and Calibration Types Some devices have connectors that make them noninsertable meaning that the connectors are the wrong type to fit in place of a zero length through connection between the test port cables The connectors can have the same type and sex on each port or they can be different types such as type N on one side and 3 5 mm on the other One way to calibrate in this situation is the swap equal adapters method which uses one adapter to perform the transmission calibration but a dif ferent adapter for the reflection calibration and actual measurement The two adapters need to be as equal as possible especially in loss electrical length and match Pot D D DUT Port2 Use cal adapter with the same connectors as the device under test DUT Port cal Port2 1 Perform 2 port cal with adapter on port 2 P 2 oe Save in cal set 1 Cal Set1 Port 1 5 seer Port 2 2 Perform 2 port cal with adapter on port 1 CalSet2 Save in cal set 2 CAL MORE ADAPTER REMOVAL 3 REMOVE ADAPTER Perform ADAPTER REMOVAL o generate new cal set Portl DUT Port2 4 Measure DUT without cal adapter Measurement
38. to finishing the calibration This will allow you to make measurements at the faster conditions with no averaging without the analyzer indicating that conditions have been changed after the calibration was complete A quick check on whether accuracy is being com promised for sweep speed can be done by making a measurement with the settings for the best accu racy saving the results as a memory trace then changing the settings or calibration type and com paring the new results with the memory trace Note that for the best accuracy it is necessary to perform a calibration as close to the actual meas urement plane as possible For example for an on wafer measurement the setup might include an S parameter test set test port cables and a wafer probe station The calibration should be performed on wafer using a calibration substrate in order to remove systematic errors caused by all the compo nents between the network analyzer and the wafer probe tips Calibration Summary Test Set cal type T R S parameter one port two port Reflection SHORT ju Reflection tracking Va Va Directivity va Ta Source match SF VY a Load match lt gt OPEN a LOAD a r y error can be corrected OO error cannot be corrected A 8711C enhanced response cal can correct for source match during transmission measurements Transmission T R response isola
39. with a controller is probably best if data manipulation or storage is required In this case there are additional considerations such as the operating system to use programming language or software package and type of GPIB card to install to communicate with the network analyzer You can use programs such as Agilent VEE that help you write test software quickly or design your own software in your preferred pro gramming language This can require training or experience in programming or software Internal automation might be easier than external automation in some situations Often internal automation is easier to learn if there is a keystroke recording mode that lets a user quickly duplicate a test Both the 8711 s IBASIC and the test sequenc ing feature in other Agilent network analyzers pro vide this capability An internal programming lan guage like IBASIC can be quite powerful but it does require some programming expertise to use it effectively and go beyond simple keystroke record ing Test sequencing is simpler but also less exten sive and it is not suitable for data manipulation However both forms of internal automation can be quite powerful For example you can use either method to program the analyzer s parallel port to control an external test set read a limit test result and send an external trigger signal to control a part handler Here are some general ideas for improving automa t