1. Remedy Turn the unit OFF wait 1 minute and then ON again Retake temperature using proper technique If still reads LO this indicates that the patient has a low body temperature r ca Er 1 If the thermometer is too cold or too hot to operate The device will beep 5 times and display Er 1 Remedy 1 Verify room temperature within product range 60 8 F 104 F 16 C 40 C 2 Wait 1 minute and then retake temperature using proper technique mo D Er 2 If the temperature did not register because of poor contact patient movement or poor placement the device will beep 5 times and display Er 2 Remedy Turn the unit OFF wait 1 minute turn it ON again and retake temperature using proper technique Er 3 If the unit malfunctioned the device will beep 5 times and display Er 3 Remedy Do not use the thermometer Call 1 800 232 2670 for service Low Battery Battery icon is on continuously 0 Remedy The battery has enough remaining life for approximately 10 more measurements Drained Battery Battery icon appears as unit is turned on and the device turns itself OFF Remedy Replace battery Standards Complies with EN 12470 3 and ASTEM E 1112 00 How to care for this product e Wipe the tip and probe with a cotton swab moistened with 70 isopropyl alcohol before amp after each use After cleaning wait at least 2 minutes before taking a measurement for the unit to re adjust to
2. claim under this warranty send item s postage paid to Service Center Attn Repair Dept 55 Commerce Dr Hauppauge NY 11788 Please include your name and address phone no proof of purchase and a brief note explaining the problem your return ad dress and 2 00 to cover shipping and handling To register your product visit us at www adctoday com register This manual is available online in a variety of languages follow the links for language options www adctoday com care tel 631 273 9600 1 800 232 2670 fax 631 273 9659 www adctoday com email info adctoday com Patent 7597668 Printed in China Adtemp 427 TempleTouch Thermometer O USE CARE amp MAINTENANCE S ee TT E AMERICAN DIAGNOSTIC CORPORATION Intended Use ntended Use gt A Temple Touch thermometer is a device which enables accurate body tem gt perature measurements obtained by using a unique probe which measures gt the heat flow from the blood vessels to the skin and converts it to body temper a x gt ature from the temple area in approximately 6 8 seconds Vn Safety Information e High prolonged fever requires immediate medical attention especially for young children e Do not allow children to take their temperatures unattended e Read the enclosed instructions carefully to ensure accurate temperature readings Temperature readings are affected by many factors including exercise drin
3. room temperature e Use only a soft dry cloth to clean the body of the thermometer if necessary Never use an abrasive cleanser or submerge the thermometer in water or other liquids
4. of the eyebrow and hairline 2 Do not place the thermometer on scarred tissue compromised by skin disorder open sores or abrasions and patients in trauma 3 Temperature readings are affected by many factors including exercise and drinking hot or cold beverages prior to taking a reading 4 Consult your physician when using drug therapies that may raise the local skin temperature in the forehead this may lead to an incorrect reading 5 Do not place the thermometer on skin exposed to direct sunlight fireplace heat air conditioner flow and cold warm compress therapies 6 Wait at least one minute between consecutive measurements for the unit to readjust to room temperature 7 If the thermometer has been stored at a cold temperature allow the unit to warm naturally to room temperature before using 8 Performance may be degraded if operated or stored outside stated temperature humidity ranges Taking Temperature OY Use this thermometer on a clean dry forehead Wipe away any perspiration aye move hair aside from the temporal area and apply the probe to bare skin 1 Press and release the Multi Function button all segments will be displayed along with a a single audible beep Figure 1 2 The previous result will be displayed for 2 seconds Figure 2 g z 3 Wait for the hourglass icon to disappear Figure 3 Do not try to take a measurement before the hourglass icon disappears as this may compromise the results 4 Plac
5. Specifications Feature Performance Data Measurement location Measurement range Measurement time 6 8 seconds Temple area of the forehead 95 F to 107 6 F 35 C to 42 C Operation environment 60 8 F 104 F 16 C 40 C Humidity 15 95 non condensing Storage environment 4 F to 122 F 20 C 50 C Humidity 15 95 non condensing Measurement readout Celsius or Fahrenheit Accuracy Complies with ASTM E 1112 00 standard requirements Display TN LCD 0 63 X 0 44 16 mm X 11 mm Battery 2 AAA 1 5volt batteries Battery life Minimum 1000 full duty cycles over 24 month period Weight 1 302 37gr including batteries Dimensions 47 x 1 17 x 1 12 cm x 3 0 cm x 2 4 cm Memory Last reading display when turned on Auto shut off After 60 seconds Case material Medical grade ABS plastic Probe material Medical grade Stainless steel medical grade ABS plastic Warranty 2 year Parts and Assembly sensor Display Window Multi Function Button AMERICAN DIAGNOSTIC CORPORATION ADC 55 Commerce Drive Hauppauge NY 11788 Made in China IB p n 93 427 00 rev 1 Non invasive WARRANTY This Thermometer is guaranteed for a period of 24 months from date of purchase This warranty does not cover batteries or damage resulting from improper handling or use Specific legal rights stipulated in your State may guide this warranty If you have a legitimate
6. e the thermometer s probe in the center of the temple area between the end of the eyebrow and the hairline Gently press and hold it against the temple to ensure good contact with the skin is maintained Figures 4a 4b Correct placement and good contact with the skin ensure reliable readings 5 As soon as the thermometer is in place the measurement will start automatically A running circle will appear indicating that the measurement is progressing successfully Figure 5 Make sure that the thermometer and the patient do not move while measurement is in progress 6 A temperature reading will be displayed within 6 8 seconds One long beep will be heard to indicate that measurement is complete and temperature will be displayed for about 1 minute 7 Press and release the Multi Function button to turn the thermometer off or it will shut off automatically after 1 minute If you need to take another reading wait at least a minute after turning the thermometer off and begin again If you take a measurement before 1 minute has elapsed the device will perform a 10 to 0 countdown to allow the device to re adjust DO NOT TRY TO MEASURE DURING THE COUNTDOWN If you attempt to do so the device will display the previous result beep amp times and turn itself off automatically Operation of the F C feature The device setting is by default Fahrenheit F o To change the scale from the OFF position press and hold the Multi Function button f
7. king hot or cold beverages prior to taking a reading as well as technique e WARNING Keep batteries out of children s reach Swallowing a battery could be harmful Batteries should not be recharged or placed into extreme heat as they may explode e WARNING It is recommended to remove the batteries from the thermometer when it is not used for a long period Do not dispose of batteries in fire recharge put in backwards disassemble mix with used or other battery types they may explode or leak and cause injury Waste batteries should be deposited in a batteries recycling trash only e CAUTION Do not drop the thermometer or expose to heavy shock e CAUTION Do not immerse thermometer in water or any other liquid e CAUTION Do not store the unit in direct sunlight at a high temperature in high humidity or dust e CAUTION Stop using the thermometer if it operates erratically or if display malfunctions e CAUTION Clean the thermometer probe before storing e CAUTION Do not attempt to disassemble the unit except to replace the batteries e CAUTION Do not autoclave Please note the cleaning procedures described in this manual Quick Start 1 Press and release the MULTI FUNCTION button 2 Wait 2 seconds 3 Hold firmly to temple wait for beep 4 Read results 5 Before re measuring wait at least 1 minute Measuring Guidelines 1 Measurement body location is the temple in proximity to the temporal artery between the end
8. or wets 5 seconds On the screen you will see the hourglass icon disappear all segments will be J displayed the last reading will be displayed and then the scale C F will change Continuously pressing the button will allow you to toggle between scales Release the button and the scale will be selected after 5 seconds then the thermometer is ready for use Changing the battery To install replace first switch off the device Remove the battery cover by sliding it out of its housing Firmly attach the batteries to their place making sure polarity is correct see engraving on the bottom of battery compartment Slide the battery cover back into place Troubleshooting and special indications The thermometer has been designed to display feedback warnings to ensure proper usage and accurate temperature measurement When an error appears the device will beep 5 times and then the thermometer will turn off automatically In the event of a malfunction or incorrect temperature measurement the following warnings will be displayed HI If the measured temperature is above 107 6 F 42 C the device will beep and display HI Remedy Turn the unit OFF wait 1 minute and then ON again Retake temperature using me I i proper technique If still reads HI this indicates that the patient has a high fever Seek immediate H medical attention LO If the measured temperature is below 95 F 35 C the device will beep and display LO