Orphee Mystic 22 Analyzer - Service manual

image

Contents

1. Rep Part Number Designation 1 002 1002 0305 M22 Syringes body 1 1 10 10 1 2 010 2501 0306 ooo o 4 005 1001 0303 Piston 220 5 010 2501 0301 Piston 6 5 6 005 1001 0309 Syringe Piston 1 6 8 312 0501 2189 262 O ring 221 89x2 62 Silicone 60SH 9 312 0501 630240 26 3x2 4 Silicone 60SH 10 950951012 O ring 1 4x1 25 Fluocarbon 80SH_ ERI r O ring spacer 3630201030 100 Serewrk 30x10 Copyright Orph e SA All Rights Reserved MYTHIC 22 REF M22 SM 000 Rev 01 MYTHIC 22 Part Number 3 4 360 0101 040 200 Screw CZX M4 x 20 5 885 1001 0430 Silent bloc M4 6 005 1002 82 8803 M22 Syringe Optical Sensor 7 863 0201 030 080 Screw K 30 x 8 Rep 1 005 1001 0310 Syringes Motor KE 3 Copyright Orph e SA All Rights Reserved 6 PARTS LISTE amp BLOW UP VIEW Page 43 77 6 PARTS LISTE amp BLOW UP VIEW Page 44 77 Rep Part Number Designation 1 005 102 0307 M22 Syringesmotorholder 2 002 1002 0305 M22 Syingesbod pk 1 1 1 10 4 360 0101 030 080 Screw CZXM3x8 0000 0000 5 005 10
2. amp E E E E 8 Slagags a 8 S 8 w Part Number REV Designation 4 005 1001 90 05 A Tubing 5 210mm Tubing 6 mnm 005 1001 90 08 A Tubing8 8mm 005 1001 90 12 A Tubingi2 mm J J 005 1001 90 13 A Tubng13 005 1001 90 14 B Tubngl4 250mm 005 1001 90 15 A Tubingl5 Jmm 005 1001 9017 A Tubingi7 m J num mm nm ma a pe 150mm 005 1001 90 26 A Tubing26 9mm 005 1002 90 03 A Tubings 35 20mm 005 1002 90 03 A Tubing 3 135 20 mm 005 1002 90 09 A Tubing 9 40mm Tubing 10 500mm exom 005 1002 90 30 A Tubing3O 0 m 005 1002 90 31 A Tubinp31 sm 5mm 005 1002 90 32 Tubing32 om 005 1002 90 33 B Tubing 33 240mm 005 1002 90 34 A Tubing34 005 1002 90 35 A Tubing35 150mm 005 1002 90 36 A Tubing36 260m j 005 1002 90 37 A Tubing37 m yi 005 1002 90 38 A Tubi
3. NO INTERPRETATION REF M22 SM 000 Rev O1 5 OUTPUT FORMAT 5 2 2 2 3 RBC pathologic messages ANE ERY gt MICRO MACRO MICR MICR gt gt MICR gt gt gt MACR gt MACR gt gt MACR gt gt gt ANIS gt ANIS gt ANIS HYPOCR COLDAGG NO INTERPRETATION 5 2 2 2 4 PLT pathologic messages THR THR MACROP PLTAGGR MICROC SCHIZ CELLD NO INTERPRETATION 52 2 3 CALIBRATION MYTHIC X Y CALIBRATION Y Calibration date calibration time Expiry date Creation date Creation time X A B C D F Number of results CR WBC target value tolerance CR RBC target value tolerance CR HGB target value tolerance CR HCT target value tolerance CR PLT target value tolerance CR END CALI Checksum CR Where X User who creates the lot Y User who performs the calibration A B C D and F are the coefficient of the calibration for respectively WBC RBC HGB HCT and PLT Follow N results which served for calibration In case of manual input of the coefficients information of the lot are empty and the number of results is null Acknowledgement of the calibration is done by the frame ACK CALI Lot A CR Where A OK or Error Code MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 31 77 5 OUTPUT FORMAT 5 2 2 3 1 Results MYTHIC X Y RESULT CR DATE dd mm yyyy CR TIME hh mm ss CR MODE CALIBRATION CR UNIT Unit CR SEG Number of sequence O CR TEST LMG CR
4. Ji An banu 60 24 3 Complete syringe module replacement nat d am idt lieb pu i ei deca iuda tet ata dope dus 61 TIPARO REPLACEMENT o m eco hof cute E coL ILC nod ALL uA bts et 62 oi GUT board replacement LU LL M E 62 72 2 Main Board gl 7 LI tL LU M LE 64 5 2 1 ebla emeM caisse dante Decr pede ukasa E aa DELE 66 9 9 2 Battery repla e MENI Siu UE tiet c Gan DAE unis 66 2 9 peambplirier board hep cease ucu uuu d eter a Ne Css ie tae tie te RIA E A LN E has 68 IO OPAICAL PARTS Sur uu nanan ee NE ul eue tI DM kaa 69 Optical bench replacembelit us e coe buie ee del si sU LL HELLE HELME 69 Optical manifold replacement euch tence vad aken 70 46 3 Heating module replacement nee u D NA g a d be s s 71 10 TROUBLESHOOTING 74 TOT ANALYTICAL PROBLEMS aa ie in en ii 74 2 OTHER PROBLEM SSR NE NN A aa amun 75 IO S TROUBLESHOOTING ME SAG Eras cdit ee di ie dec Na ne m o as 76 MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 7 77 JA INTRODUCTION REF M22 SM 000 REVOL 1 INTRODUCTION BEFORE TO INSTALL OR TO MAINTAIN A MYTHIC 22 YOU HAVE TO READ AND TO REFER TO THE USER S MANUAL The following chapters describe the different technical parts
5. from the mono or polynucleated cells line Presence of atypical lymphocytes or basophiles Presence of basophiles small neutrophils without granulations or few segmented band cells Presence of small neutrophils without granulations or few segmented band cells or hyper basophil monocytes Presence of giant neutrophils hyper segmented neutrophils eosinophils with few granulations or damaged eosinophils Percentage of cells in the zone in regard of the total number of leucocytes in the scattergram Percentage of cells in the zone in regard of the number of lymphocytes Percentage of cells in the zone in regard of the total number of leucocytes in the scattergram Percentage of cells in the zone in regard of the number of lymphocytes and monocytes Percentage of cells in the zone in regard of the number of lymphocytes and neutrophils Percentage of cells in the zone in regard of the number of monocytes and neutrophils Percentage of cells in the zone in regard of the number of eosinophils and neutrophils Copyright Orph e SA All Rights Reserved Number of cells in the zone Number of cells in the zone Number of cells in the zone Number of cells in the zone Number of cells in the zone Number of cells in the zone Number of cells in the zone MYTHIC 18 REF M22 SM 000 Rev O1 FLAGS ALARMS 4 2 3 DIF parameters display rules DIF VALUE DIF VALUE
6. 22 the CPU main board the preamplifier board and the GUI board In the case of replacement of the main board or preamplifier board be aware of the sensibility of this board to the electrical static discharge ESD Before operating check the earth connection and use the antistatic bag supply with the board Please contact your Orph e for more information Be careful with the edge of the metal sheets after having dismantled the covers 3 2 ELECTRICS PARTS VIEW 3 2 1 Main board 205 1002 01 an UNIVERSAL E SIG Bin ia MICROELECTRONICS a ad UE pete za ug r npe iir i sd zi x i ur a i T 1 HE PR 3 i E E qii T s Fill JM Ed Il d UNIVERSAL MICROELECTROMICS hi i Hi TE ep Ju E Boo PRE Hei ILE at i 8 E THAT d aga d Ti 71 UNIVERSAL g 1 i MICROELECTRONICS zi Pi The mono electronic board is located between the hydraulic part and the reagent tray The board driven by a 32 bit processor manages the following parts Sample needle rocker syringe block motors Display and keyboard Connexion mode RS232 Ethernet Printer Measurement Counting hemoglobin measurement Data processing External barcode reader MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 19 77 3 ELECTRIC amp ELECTRONICS REF M22 SM
7. 6 21 Syringe down to perform the vacuum counting valve 8110 22 Second measurement sequences CBC measurements 8 second valve 10 Syringe up for DIF measurements 12 seconds valve 4 9 12 13 16 17 DRAIN AND RINSE BATH 22 Cleaning Cleaning the apertures valve 11 Back pressure in the injector valve 4413114 23 Syringe down WBC bath drain valve 1 Tubing transfer drain valve 11414 RBC bath drain valve 2 24 Syringe up RBC bath rinse diluent 2ml valve 4 Waste syringe drain valve 7 24 Needle up and syringes up for back flush valve 4 6 25 Rocker move above the WBC bath 26 Syringe up WBC bath rinse diluent 2 5ml valve 4 Waste syringe drain valve 7 NEEDLE BACK AND WASTE DRAIN 27 Syringe down Transfer tubing rinsing with 0 5ml valve 11 14 Dry the outside of the needle valve 3 _REF M22 SM 000 Rev 01 4 4 4 f FLUIDICS 28 Rocker move in the sampling position 26 Syringe up Waste syringe drain valve 7 30 Needle down in sampling position The MYTHIC 18 is ready to perform a new analyze 2 3 FLUIDICS VIEW All the fluidics part is on the right side of the instrument and consists of only three modules Sampling module o Rocker patent pending Manages the rise and descent of the needle Syringe module patent pending consists of one block o Reagent syringes Diluent lysis sampling and air syringes o Liquid valve manifold assembly and tubing Counting chambers W
8. A B Target tolerance Target tolerance CR Value A B Target tolerance Target tolerance CR GRA Value A B Target tolerance Target tolerance CR Value A B Target tolerance Target tolerance CR MON Value A B Target tolerance Target tolerance CR Page 32 77 Copyright Orph e SA All Rights Reserved REF M22 SM 000 Rev 01 MYTHIC 18 REF M22 SM 000 Rev O1 5 OUTPUT FORMAT GRA Value A B Target tolerance Target tolerance CR END RESULT Cheksum value CR A and B like routine results 5 2 2 5 REPEATABILITY 5 2 2 5 1 Results MYTHIC X Y RESULT CR DATE dd mm yyyy CR TIME hh mm ss CR MODE REPEATABILITY CR UNIT Unit CR SEG Number of sequence CR TEST LMG CR OPERATOR login Mythic when analysis is performed CR WBC Value A B CR RBC Value A B CR HGB Value CR HCT Value CR MCV Value CR MCH Value CR MCHC Value A B CR RDW Value A B CR PLT Value CR MPV Value A B CR THT Value A B CR PDW Value CR LYM Value A B CR MON Value CR GRA Value A B CR LYM Value CR MON Value A B CR GRA Value CR END RESULT Checksum value CR A and B like routine results 5 2 3 CHECKSUM For the frames with a few number of lines the block is finished with an end l
9. Do 2 51 RESUS uc uncia dan dd A MEE LI A CM a aa Q On 33 92 S O el UA O sr ODRE ST RE NE 33 6 PARTS LIST 4 BLOW UP VIEW 35 O PARIS LIST Tm MX 35 Ole Two wears spare pan Ski tante ctn eu totu es RA Loa 35 MAIN ENA NEO sete doute a uet iA eui e lia AA LL oin ee 35 REF M22 SM 000 Rev 01 G lo esed T S 36 TE O OE O O TE RN 36 OURS DORE TR A t O a M n O O E 37 OL BEOW UP VIEW tu ua RAN SA aaa RO eal cg 40 ng MO dle detente one Le 40 Ge 2 Sampling Module naue ANM ite LM E I ae 41 Oc JS T emo A EI u m MOTENO MI E MIL LULA RE 42 Ge Droni COVER oce seno uius uua NS IU x i niu ieri 45 6 2 5 Oprical CMC beo iL eub i hae aah Sedat tne LIH D Sci Lo pud ertet 46 A AE 47 0 2 e Heating NOS uuu u 48 7 SOFTWARE 49 rrr RODVE pK FE 49 JE SPECIFIC MECHNECEAN DE SP GAV cio canoe ee to AL cer MM MEDI D UM AM DEA 49 Technicidh DOS SW aD I UC C Lu LM I LM 49 Te ecc I dL cl iU uu E wee REDS 49 7 2 NE Ioui MERE NR 49 2 2 Counting
10. OPERATOR login Mythic when analysis is performed CR WBC Numerical Value CR RBC Numerical Value A B CR HGB Numerical Value A B CR HCT Numerical Value A B CR PLT Numerical Value A B CR END RESULT Checksum value CR A and B like routine results 2224 QC 5 2 2 4 1 Results MYTHIC X Y RESULT CR DATE dd mm yyyy CR TIME hh mm ss CR MODE QC CR UNIT Unit CR SEG Number of sequence O CR LOT Number of lot CR Maximum 10 characters LOT DATE Creation Dateof the lot time of creation CR EXPIRY DATE Expiry date CR USER User login who created the lot CR TEST LMG CR OPERATOR login Mythic when analysis is performed CR WBC Value A B Target tolerance Target tolerance CR RBC Value A B Target tolerance Target tolerance CR HGB Value A B Target tolerance Target tolerance CR HCT Value A B Target tolerance Target tolerance CR MCV Value A B Target tolerance Target tolerance CR Value A B Target tolerance Target tolerance CR MCHC Value A B Target tolerance Target tolerance CR RDW Value A B Target tolerance Target tolerance CR PLT Value A B Target tolerance Target tolerance CR MPV Value A B Target tolerance Target tolerance CR THT Value A B Target tolerance Target tolerance CR Value A B Target tolerance Target tolerance CR LYM Value
11. Slo Blo fuse model 0454 004 to replace the old one It s strictly forbidden to use an another type of fuse I This battery is only used to supply the real time clock With a screwdriver lift the clips maintaining the battery Slip down the battery to remove it and replace it by a new one Use a Maxell battery model CR2016 to replace the old one It s strictly forbidden to use an another type of battery I DN ij P Li I i i 5 ms IL NI H L a JF RI i t pf cere lia C P 1 5 li READ CAREFULLY THE FOLLOWING RECOMMENDATIONS Handling storage Never expose the battery to ultrasonic sound Exposing the battery to ultrasonic sound may cause short circuiting because the inside material is broken into pieces leading to distortion leakage overheating explosion or fire Never subject the battery to severe shock Dropping throwing or stomping on the battery may cause distortion leakage overheating explosion or fire Never short circuit the battery while installing into equipment REF M22 SM 000 Rev O1 9 REPARING Please be careful when installing the battery not to short circuit it with metal portions of the equipment Use the correct battery suitable for the equipment The battery may not be suitable for the specific equipment due to the using conditions or ty
12. if unsuccessful Check the level bubble flow in the WBC bath during the run cycle Check the reagent and enclosure temperature 5 DIFF No result No HGB Check the lysis level amp WBC Check the right lysis tubing connection to the WBC bath Check the valve Nb 9 HGB amp WBC OK Perform a Back flush Cycle and then a Bleach cycle if unsuccessful Check the reagent and enclosure temperature Check the preampli connection wires RBC No results No HCT amp PLT Check the bath wires Perform a Cleaning Cycle and then a Bleach cycle if unsuccessful Bad stability HCT amp PLT too Perform a Back flush and a Cleaning Cycle and then a Bleach cycle if unsuccessful Check the level bubble flow in the RBC bath during the run cycle Check the level bubble flow in the WBC bath during the first dilution HGB No results P Check the led light on Bad stability Check if no bubble in the lysis tubing Check the reagent and enclosure temperature Check the level bubble flow in the WBC bath during the run cycle In any case if a problem is not solved call Orph e s REF M22 SM 000 Rev 0110 TROUBLESHOOTING 10 2 OTHER PROBLEMS needle during the run cycle it if necessary see section 9 3 2 All results bad Check the level diluent and if the tubing is pinched Check if the diluent is at the same level than the analyzer No display Check the flat cable Machine switch off after See section 5 4 2 few second following the Switch
13. tal 14 25 A isdem tI MN DOM Ct Le Ci CLE oM Lu A in marsu 14 2 5 2 Manirola Tubing ana valves DO SITIO a uy ted eia t nite ton ne edet teet esie te de 15 2 5 3 Optical Manifold tubing and valves position sn 17 LENET WAU rins Cu uy IURE D 18 3 ELECTRIC amp ELECTRONIC 19 BP GENERALE E E e tune Ua i iu Rl E Cu LI uu un Ni uM e NL stabelen 19 Se ELECTRIC PART SNIE e 19 Se LMain board C 05 1002 uD ER CN LEM E DE M 19 3220UIboaPd 205 1002 02 e eerte etat irte Ar E M 20 32 9 PreamplifierDoard OS TOO TOBY a upon NC S EL iit 20 3 2 4 Power SUDDIY nn NS Nj u au au nu ma NAJ ED KELLER ja 20 S ELECTRONIC DIANA vC V 21 Sd WIRING ee A NA ENA NE OO LE RE Ne Ll s ARE 22 4 FLAGS amp ALARMS 23 4 DOENERAE T ADS cal ena set 23 EVO AC S LI Lu ED ua a MEI LEE 23 4 2 1 WBECUTVE TAGS ER 23 4 2 256 GOS y y op dean ttti bot 24 qo DEF parameters display rales cus u Sam tat Arc ar 25 ERYT ROOSTE AND RAGE de 25 FAL PLATELET FLAC
14. 0 O ring 1 4x1 25 Fluocarbon 80SH 2 312 0505 500100 _ jO ringg5xiFluocarbonBOSH 6 1 2 Tubing kit Part Number 002 1002 50 11 Designation Mythic 22 Tubing kit A A A 005 2 005 z 005 1001 9018 Tubing18 005 1001 90 21 Tubing 21 005 1002 90 24 M22 WASTE Tubing A 1 A 1 A 1 A 1 A 1 1 1 A 1 A A A 005 1002 90 38 Tubing 38 005 1002 90 39 Tubing 39 005 1002 90 40 Tubing 40 005 1002 90 41 Tubing 41 005 1002 90 42 Tubing 42 005 1002 90 43 Tubing 43 REF M22 SM 000Revol U U 6 PARTS LISTE amp BLOW UP VIEW 005 1002 90 44 1 A Tubing 44 1 A 005 1002 9045 Tubing 45 1 A 005 1002 90 46 Tubing 46 1 A 005 1002 90 47 Tubing 47 1 A 005 1002 9048 Tubing 48 1 A 005 1002 90 49 Tubing 49 1 A 005 1002 9050 Tubing 50 1 A 005 1002 9051 Tubing 51 1 A 005 1002 9052 Tubing 52 1 A 005 1002 90 53 Tubing 53 1 A 05 1002 90 54 Tubing 54 1 A 005 1002 9055 Tubing 55 1 A 05 1002 90 56 Tubing 56 1 A 005 1002 9057 Tubing 57 1 B 005 1002 9058 Tubing 58 6 1 2 Spare parts list Designation PartNumber Oty Pkg Unit 500mL Straw 005 1002 06 05 1 50P Flat Cable 170 5002 06 221 1 UNIT ALL coaxial cable 005 1002 84 09200 1 UNIT ALL Power Cable 005 1002 82 02 1 UNIT UNIT Bottle cap filter 010 2501 06 03 10 UNIT UNIT Counting Manifold 010 2501 01 UNIT Door Flat Screwdrver 700 7101 02 09 0
15. 000 Rev 01 MYTHIC 22 is supplied with an external power supply block In the case of replacement of the main power wire supplied with the MYTHIC 22 the new one must comply with the local regulation The MYTHIC 22 has been certified with the power supply adapter provided with the machine The use of another external power supply is not guaranteed Please contact your Orph e s representative REF M22 SM 000 Rev 01 3 ELECTRIC ELECTRONICS 3 3 ELECTRONIC DIAGRAM MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 21 77 3 ELECTRIC 8 ELECTRONICS CREF M22 SM 000 Rev 01 3 4 WIRING DIAGRAM REF M22 SM 000 Rev O1 FLAGS ALARMS 4 FLAGS amp ALARMS MYTHIC 22 manages 22 different flags These flags allow the user to be alerted if there is a mistake which can affect the quality of the results All of these flags appear on the right of the result In presence of one or more flags it is recommended to check the result by a conventional measure or on blood smear NOTA Most of these alarms can be modified by the user see section 3 of the User s Manual 4 1 GENERAL FLAGS The following alarms are common for all parameters XXX Counting or measure rejection Can appear with WBC RBC PLT and HGB see section 8 5 3 for HGB and section 8 5 2 for DIF parameters XXXD Higher than linearity limits but lower than the reportable limits WBC RBC PLT HCT HGB D Higher than the reportabl
16. 01 9026 Tubing2e 0 000 Copyright Orph e SA All Rights Reserved REF M22 SM 000 Rev 01 MYTHIC 22 REF M22 SM 000Revol U U 6 PARTS LISTE amp BLOW UP VIEW 6 2 5 Front cover Module Rep Part Number 4 1005 1002 04 02 M22 Front cover 2 1005 1001 84 06 170 Start key Switch MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 45 77 6 PARTS LISTE amp BLOW UP VIEW REF M22 SM 000 Rev 01 6 2 6 Optical bench 2 Optical Flow Cell 005 1002 02 01 1 Optical Bench 005 1002 01 01 Page 46 77 Copyright Orph e SA All Rights Reserved MYTHIC 22 REF M22 SM 000Revol U U 6 PARTS LISTE amp BLOW UP VIEW 6 2 7 Optical manifold Optical Manifold 005 1002 02 02 MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 47 77 6 PARTS LISTE 4 BLOW UP VIEW REF M22 SM 000 Rev 01_ 6 2 8 Heating module Heating Module 005 1002 04 0 2 REF M22 sm o0ORevOl 7 SOTWARE 7 SOFTWARE 7 1 INTRODUCTION Please refer to the User s Manuel for explanation of the software operation 7 2 SPECIFIC TECHNICIAN DISPLAY The following section describe the software operation which are only accessible with the technician and super technician passwords 7 2 1 Technician password The unmodified technician password is 8SUP It is strictly forbidden to disclose the technician password to the final customer In such case
17. 24 Thermal The modification of one of these parameters must be done in compliance with Orph e after having performed a specific training In this case of modification without authorization Orph e guarantee will not be valid PREVIOUS E SCREEN TOOLS ins REAGENT SET POINT PIO GAF 1 98 UNLOCK HEAT IL Press Er to unlock the heating system see the possible occurrence in section 4 5 ENCLOSURE SET POINT PIO GAP F 15688 UNLOCK HEAT 7 2 2 5 Optical bench adjust The use of this display can be done only after having performed a specific Orph e training PREWIOUS sa COUNTING READING PULSES ADJUSTMENT GRAPH This display allows to adjust the optical beam with the use of special latex particles MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 51 77 J SOTWARE REF M22 5M 000 Rev 01 7 2 2 6 Optical gain adjust The modification of one of these parameters must be done in compliance with Orph e after having performed a specific training In this case of modification without authorization Orph e guarantee will not be valid PREVIOUS SCREEH This display allows to adjust the FSC and ALL gain with the use of special latex particles FSC GAIN a ALL GAIN B Page 52 77 Copyright Orph e SA All Rights Reserved MYTHIC 22 REF 22 5 000 SOTWARE 7 2 SOFTWARE ARBORESCENCE y M C Launch new manu an E Launc
18. 501 152401 09 0501 152402 09 0501 01 016 05 1002 85 01 10 05 1002 84 09 260 CIC zz gJ C Z C 2 Z Z 159 C lt 05 1002 82 01 05 1002 85 02 05 1002 04 02 85 2501 17 11 65 0110 05 2 x 500 ubriplate DS ES 10 0502 01 55 6101 01 300 05 1002 86 02 yse Spring 3 05 1002 86 03 05 1002 90 23 02 1002 84 06 270 05 1002 80 01 2 Ce Z gt C lt c Z 2 Z dj A C Z gt C Z C Z 05 1002 80 01 1 05 1002 82 880 1 M22 Sampling module sub assembly 005 1001 02 01 22 02 1002 03 05 CCC ee Z Z Z Z Z Z Z 2 IPIE IPIE IEEE O O O O I O o jojo O O O O C2 M22 Syringe motor cable 05 1002 80 01 3 M22 Syringe Optical Sensor 005 1002 82 880 3 22 Copyright Orph e SA All Rights Reserved Page 37 77 6 PARTS LISTE amp BLOW UP VIEW M22 Cleaner Straw 005 1002 06 06 1 UNIT M22 Counting Valves Cable 005 1002 84 04 1 UNIT UNIT MYTHIC barcode reader 005 1099 87 01 1 UNIT UNIT Mythic 22 2 years spare part kit 002 1002 50 20 02 1 UNIT Mythic 22 CPU Board 205 1002 01 1 JUNIT Mythic 22 IHM Board 205 1002 02 1 UNIT UNIT Mythic 22 Maintenance Kit 002 1002 50 02 1 UNIT Mythic 22 Tubings kit 002 1002 50 11 1 UNIT Designation PartNumber Qty Pkg Unit MYTHIC 22 User manual 605 1202 02 1 UNIT UNIT Mythic22 LCD Display 555 0102 01 02 1 UNIT Needle Belt 353 0301 08 120 19 1 UNIT Needle Optic
19. 9 1000 mm Q Vacuum Tank 005 1002 03 01 1 UNIT UNIT MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 39 77 6 PARTS LISTE amp BLOW UP VIEW REF M22 SM 000 Rev OL 6 2 BLOW UP VIEW 6 2 1 Counting Module Rep Part Number Designation 4 005 1001 01 01 Counting chambers 2 010 2501 0103 Counting manifold 3 005 1001 01 0201 WBC Counting Head 4 005 1001 010202 RBC Counting Head 5 312 0505 500 100 O ring 5x1 Silicone 80SH 6 360 0101 030060 Screw CZX M3 x 6 REF M22 SM 000 Rev 01 6 2 2 Sampling Module MYTHIC 22 Rep Part Number Designation 1 005 1001 0201 Sampling module 2 005 1001 0203 fRockerMotor 4 005 1001 0204 fRinsingheed 0 0 5 010 2501 0205 Sampling Neede 6 312 0505 140 125 O ring 1 4x1 25 Fluocarbon 80SH 360 0102 030 060 Screw M3 x 6 360 0101 030 080 Screw CZX M3 x 8 13 005 1002 82 880 1 M22 Rocker Optical Sensor 44 370 4050 030 10 lAnti loss washer 23 x 1 Ex EX 10 363 0201 030080 ScrewK30x8 1 10 0 ELE _12_ Copyright Orph e SA All Rights Reserved 6 PARTS LISTE amp BLOW UP VIEW Page 41 77 6 PARTS LISTE amp BLOW UP VIEW REF M22 SM 000 Rev OL 6 2 3 Syringe Module Page 42 77
20. BC and RBC counting chambers and hemoglobin measurement o Liquid valve manifold assembly and tubing 2 E A i MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 11 77 2 FLUTDICS REF M22 SM 000 Rev 01 2 4 FLUIDICS PARTS DESCRIPTION The hydraulic part of the MYTHIC 22 is very simple and made of only 5 modules Sampling module Counting bath module Syringes module Optical bench Optical manifold The modules are connected together by semi rigid tubing 2 4 1 Sampling module This module patent pending enables to draw the sample and to perform the WBC and RBC PLT dilutions It is assembled with a rotating rocker moving around a support which maintains the system to move up and down the sampling needle A very reliable system of rack gear moves the rocker The cleaning system of the sampling needle can be removed without tool see section9 3 2 in the user manual The o ring of the needle included in the cleaning system can also be removed without tool see section 9 3 2 in the user manual The maintenance of these parts is very easy to perform 2 4 2 Counting bath module This module allows to count the WBC and RBC PLA and to measure the HGB It is made with a manifold maintaining the reagent commutation valves and the counting bath block with their measurement block including the apertures The counting bath block assembly and these apertures can be removed without t
21. C 22 Copyright Orph e SA All Rights Reserved Page 57 77 3 REPARING REF M22 SM 000 Rev 01 Unscrew and remove the screw which maintain the IR barrier Unscrew the rocker fixation screw but DO NOT remove it Disconnect the three tubing and cut the ties Unscrew the two guide screw but DO NOT remove it Then slide down the guide Remove the complete rocker module REASSEMBLY PROCEDURE To reassemble the unit carry out the various operations in the opposite direction _REF M22 SM 000 Rev O1 9 REPARING 9 3 COUNTING BATHS MODULE 9 3 1 Baths dismantling Perform baths dismantling as described in the User s Manual 9 3 2 Baths o ring replacement Perform baths o ring replacement as described in the User s Manual 9 3 3 Aperture block replacement Perform aperture block replacement as described in the User s Manual 9 3 4 Complete baths module replacement Before to replace the complete syringe module drain all the tubing Remove the diluent lysis and cleaner straw from their bottle Perform a PRIME ALL MAIN MENU REAGENT PRIME ALL Switch off the MYTHIC and disconnect the power supply Remove the Lysis and Cleaner bottle from their location Then put the MYTHIC on its left side Remove the baths connection HGB WBC coaxial RBC coaxial and thermal captor Which maintain the baths module the third screw is located behind the over flow tubing Cut the fix
22. CS DIAGRAM 2 5 1 General Diagram CLEANER LYSE HAS 318VM REF M22 sm ooORevot I 4 4 4 8 FLUIDICS 2 5 2 Manifold tubing and valves position Each figure corresponds to a tubing or a valve number see the Fluidic diagram section 2 5 1 Counting manifold tubing position S t 6 45cNm 2 2 PIN 309 0501 152401 3 2 P N 309 0501 152402 MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 15 77 2 FLUIDICS REF M22 SM 000 Rev 01 Counting manifold valves position Syringe tubing position P N 309 0501 01 016 O 2 TORX 6 25cNm P N 309 0501 152401 PIN 309 0501 152402 AR Syringe manifold valves position REF M22 SM O00RevO e FLUIDICS 2 5 3 Optical Manifold tubing and valves position EV 2 2 309 0501 152401 MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 17 77 2 FLUTDICS REF M22 SM 000 Rev 01 2 6 SIZE amp LENGTH TUBING To ensure perfect results it is mandatory to respect the size and the length of the tubing 005 1001 9018 A Tubingl8 80mm 005 1001 90 21 B Tubing21 600mm 005 1001 9022 A Tubing22 005 1001 90 23 A Tubing 23 DILUENT 500mm 005 1001 90 24 A Tubing 24 WASTE om 005 1001 90 25 A Tubing25 omm 45
23. ORPHEE 19 chemin du champ des filles CH 1228 Geneva Plan les Ouates SWITZERLAND Orph e Tel 41 0 22 706 1840 Fax 41 0 22 794 4391 http www orphee medical com MYTHIC 22 SERVICE MANUAL REF M22 SM 000 Rev O1 MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 1 77 REF M22 SM 000 Rev 01 REVISIONS i i CONTACT ADDRESS MANUFACTURER C2 Diagnostics 2214 Bd de la Lironde 34397 Montpellier Cedex 5 FRANCE AUTHORIZED REPRESENTATIVE ORPHEE SA 19 chemin du champ des filles CH 1228 Geneva Plan les Ouates WORLDWIDE SWITZERLAND Tel 41 0 22 706 1840 Fax 441 0 22 794 4391 http www orphee medical com LOCAL REPRESENTATIVE REF M22 SM 000 Rev O1 IMPORTANT SAFETY INSTRUCTIONS RISK OF DANGER Indicates a procedure to be strictly respected in order to avoid any risks for the operator user or damages on the instrument or on the quality of results Indicates that wearing gloves is mandatory before performing the described operation due to risk of contact with materials that may be infectious NOTA Indicates important additional information BEFORE TO INSTALL OR TO MAINTAIN A MYTHIC 22 YOU HAVE TO READ AND TO REFER TO THE USER S MANUAL OPERATORS MUST HAVE RECEIVED ADEQUATE TRAINING BE CAREFUL WITH THE EDGE OF THE METAL SHEETS AFTER HAVING DISMANTLED THE COVERS MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 3 77
24. Orph e guarantee will not be valid In the technician log mode the most part of the control of the MYTHIC 22 are invalided perform a specific Orph e training before repairing or maintaining 7 2 2 Technician display The technician display are accessible from MAIN SCREEN then SERVICE then TECHNICIAN 7 2 2 1 Technician alows access to the adjustment of the PREBEN THA counting parameters and lysis quantity These adjustments are possible with the super technician code only CO OFF Dr allows to test the aperture current FROD ADJUST For an access to the latex adjustment and HGB blanc adjustment press CLEAN OUT PROO ADJUST THERMAL OPTICAL BENCH ADJUST CLEAN OUT see section 8 3 OPTICAL GAIN ADJUST LOG INTERVENTION LOG INTERVENTION BMC T COUDE allows to fill up the log accessible from the MAIN MENU TERA ce section7 2 2 4 OPTICAL BENCH ADJUST OPTICAL BENCH ADJUST see section 7 2 2 5 OPTICAL GAIN ADJUST OPTICAL GAIN see section 7 2 2 6 MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 49 77 7 SOTWARE REF M22 SM 000 Rev 01 7 2 2 2 Counting parameters The modification of one of these parameters must be done in compliance with Orph e after having performed a specific training In this case of modification without authorization Orph e guarantee will not be valid PREVIOUS SCREEH The access of this
25. RE T 9 ec RT 25 D MACHINE SE a S 26 4 511 Pressure Tag sce ae v itam aite asa ni 26 go 2 MA usu sa a 26 5 OUTPUT FORMAT 28 TOP 6 M tie 28 5 2 1 General principle M 28 D 2 2 FRAMES utet O 28 32 21 REQUEST TO SEND RESULTS u uu et usa Qua ha ate tea bte een Me et 28 92 24 T REQUEST een Rp Dot pne dix ma s IDA FOL EE 28 5 2 2 1 2 ACKNOWLEDGEMENT REQUEST TO SEND RESULTS 28 5 2 2 LS ACKNOWLEDGEMENT RESULTS nan Rei E ee o sime sedie nate e a acte a ues ble Ub A et 28 De ROUTINERESULTS rouen data CLE ELI Cn E t CA Lin re 29 ni MU d D MEE Ce MEME MM LM Mau M CREE UM duce a 30 WBC pathologie alas bad einen 30 DR RDC pathologic Message 6 aeta ceto ASERNE ET MUI LIBET Ve DM DNE DUE 31 Scc PLNpathologiemessdgeB ua ul REE M uL E M A I A 31 ood SCALIBRA TEON S Senecio Saunas A M SR E sd 31 920 PEUS A MEAM EL ME ME MEA LE M D OI Li TUT E LU 32 OMM T RO cer PEN AN E 32 GU MIC E M 32 Dee D REPEATABILIETVY sin t n tuit a A E LA A rni aM 33
26. REF M22 SM 000 Rev 01 This equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information listed below Guidance and manufacturer s declaration Electromagnetic immunity The MYTHIC 22 is intended for use in the electromagnetic environment specified below The customer or the user of the MYTHIC 22 should assure that it is used in such environment Immunity test IEC 60601 test level Compliance Electromagnetic environment guidance level Portable and mobile RF communications equipment should be used no closer to any part of the MYTHIC 22 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms d 1 2 P IEC 61000 4 6 150Khz to 80Mhz Radiated RF 3 Vrms d 1 2 P 80MHz to 800MHz IEC 61000 4 3 80Mhz to 2 5Ghz d 2 34P 800MHz to 2 5GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strenghts from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol e NOTE 1 At 80Mhz and 800MHz the higher frequency range apllies NOTE 2 Theses guid
27. WBC lt 0 5 10 WBC gt 150 x10 mm Se ma Ii WBC rejet NL amp RL or LH or HL amp A 4 3 ERYTHROCYTE AND HGB FLAGS In any case if one or more DIF parameters LYC MON NEU EOS or BAS are following with a star the sample must be checked on smear CHANNELS MEANS VALUE VALUE O to CR1 Abnormal quantity of Percentage of cells in the Number of cells in the microcytes at the left side zone in regard of the number zone of erythrocyte FR2 CR2 to the Abnormal quantity of Percentage of cells in the Number of cells in the end macrocytes at the right side zone in regard of the number zone of erythrocyte Appear Bad measure of the HGB Not Available Not Available with HGB blank result 4 4 PLATELET FLAGS Presence of abnormal Percentage of cells in the Number of cells in the quantity of debris or small zone in regard of the number zone cells of platelets until P MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 25 77 2 to CP Presence of schizocytes Percentage of cells in the Number of cells in the zone in regard of the number zone of platelets until P FP3 to CP3 2 Presence of microcytes Percentage of cells in the Number of cells in the zone in regard of the number zone of platelets until P 4 5 MACHINE FLAGS This section describe the main machine flags displayed in case of problem for all the other refer to the sect
28. Y and SLOWLY pull the upper side of the connector and then pull the lower side to up the fixing system This part is VERY sensitive In case of brake the complete GUI board have to be changed VERY SLOWLY pull the flat cable outside the connector MYTHIC 22 Copyright9 Orph e SA All Rights Reserved Page 63 77 9 REPARING REF M22 5M 000 Rev 01 9 5 2 Main board replacement NOTA Before to dismantle the main board dismantle the GUI board see section 9 5 1 VERY SLOWLY push the flat cable inside the window Remove the baths connection HGB WBC coaxial RBC coaxial and temperature sensor N Remove this three screws at the back of the instrument 22 5 000 9 REPARING Remove the two screws located at the hydraulics side mE CE a CSCO Lift slowly the board and remove the seventeen connectors located at the top of the board Each connectors is identified with a number to allows the reassembling following their position writing on the sticker Remove the vacuum tubing then remove the board NOTA After having performed the replacement of the board don t forget to reconnect the vacuum tubing MYTHIC 22 Copyright9 Orph e SA All Rights Reserved Page 65 77 3 REPARING REF M22 SM 000 Rev 01 9 5 2 1 Fuse replacement Theses fuses protect the 24v circuit Remove the fuse with a thin pliers and replace it by a new one Use a Littelfuse NANO
29. agent straws from all the reagent and run TWO Prime all to empty the MYTHIC Put the MYTHIC on the left side and remove the six screws from the reagent plate located under the machine three screws stay on the basement Dismantle the connector The reagent module is fixed by five screws located in the fluidics part Remove the two screws located at the left of the counting module MYTHIC 22 Copyright9 Orph e SA All Rights Reserved Page 71 77 9 REPARING REF M22 SM 000 Rev 01 gf Remove the two screws located at the right of the counting module Remove the last screw located above the counting module and which maintain a ground cable Dismantle the four tubing located at the right of the counting module _REF M22 SM 000RevOl 9 REPARING Remove the heating module from the MY THIC 22 To reassemble the part carry out the various operations in the opposite direction MYTHIC 22 Copyright9 Orph e SA All Rights Reserved Page 73 77 10 TROUBLESHOOTING REF M22 SM 000 Rev 01 10 TROUBLESHOOTING A 10 1 ANALY TICAL PROBLEMS PROBLEMS CONDITIONS WBC No results No HGB Check the lysis level Check the right lysis tubing connection to the WBC bath Check the valve Nb 9 HGB OK Check the bath wires Perform a Cleaning Cycle and then a Bleach cycle if unsuccessful Change the aperture block if unsuccessful Bad stability Perform a Back flush and a Cleaning Cycle and then a Bleach cycle
30. al Sensor 005 1001 82 880 300 1 U Z O ring 1 4x1 25 Fluocarbon 80SH 312 0505 140 125010 10 0 O ring 13 1 1 6 Fluocarbon 80SH 312 0505 1310 160 O ring 215 54x2 62 Silicone 60SH 312 0501 1554 262 10 0 C lt lt O ring 18 77x1 78 Silicone 60SH 312 0501 1877 178 O ring 21 89x2 62 Silicone 60SH 312 0501 2189 262 10 JU O ring 247 34x1 78 Silicone 60SH 312 0501 4734 178 C Z Z C Z lt O ring 5x1 Fluocarbon 80SH 312 0505 500100 10 U O ring 6 35 1 78 Silicone 60SH 312 0501 635 178 C Z mr Z O ring 6 3x2 4 Silicone GOSH 312 0501 630240 10 JU lastic grease SPG35 410 0503 01 RBC Counting Head 005 1001 01 02 02 ampling needle 10 Page 38 77 Copyright Orph e SA All Rights Reserved CICICICICICICICICICICICICIC zizizizzizzzzzzzzzzz SIS ISS SSeS Sa a Z gt C Z mr Rinsing head 005 1001 02 04 10 10 10 10 10 c Z gt 22 5 000 01 MYTHIC 22 REF M22 SM 000Revol 6 PARTS LISTE amp BLOW UP VIEW screw K 30 x 8 363 0201 030 080 10 UNIT Screw RLX M3 x 6 360 0102 030 060 10 UNIT UNIT Seal plate 010 2501 03 13 1 UNIT UNIT Short Arm TORX T20 Tool 700 1101 20 1 UNIT UNIT Silicon grease 3gr 410 0501 02 02 1 UNIT Silicon grease tube UNIT Start key Switch 005 1001 84 06 170 1 UNIT UNIT Syringe Motor Holder 005 1001 03 07 1 UNIT UNIT Tygon tubing 0 89mm 300 0502 00
31. art carry out the various operations in the opposite direction After the replacement of the preamplifier board it is mandatory to run a latex cycle to adjust the WBC and RBC PLT gain See section 7 2 2 3 _REF M22 SM 000 Rev O1 9 REPARING 9 5 OPTICAL PARTS REPLACEMENT 9 6 1 Optical bench replacement eS E Unscrew and remove the ground screw ea E x Unscrew but WITHOUT dismantle them the gi three screws which maintain the optical bench Disconnect the four tubing located under the optical bench Disconnect the filth tubing from the optical manifold MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 69 77 9 REPARING REF M22 SM 000 Rev 01 Remove the connectors from FSC and ALL preamplifier boards ALL preamplifier FSC preamplifier To reassemble the part carry out the various operations in the opposite direction and proceed at the gain adjustment see section 7 2 2 9 6 2 Optical manifold replacement Remove all the tubing connected to the optical bench Remove the four screws and remove slowly the optical manifold and then disconnect the valve connectors ODG Lh iag To reassemble the part carry out the various operations in the opposite direction _REF M22 SM 000RevOl 9 REPARING 9 6 3 Heating module replacement Before dismantling the heating module remove the re
32. ation ties of the tubing Disconnect the tubing from the manifold and the bath Slide up and remove the baths module NOTA Take care about the tubing reconnection if necessary cut few millimeters 2 to 3mm of tubing or exchange them REASSEMBLY PROCEDURE To reassemble the unit carry out the various operations in the opposite direction MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 59 77 3 REPARING REF M22 SM 000 Rev 01 9 4 SYRINGE MODULE 9 4 1 Piston greasing Perform the piston greasing as described in the User s Manual 9 4 2 O ring or piston replacement Before replacing the complete syringe module drain all the tubing Remove the diluent lysis and cleaner straw from their bottle Perform a PRIME ALL MAIN MENU REAGENT PRIME ALL Put the syringe in lubrication position MAIN MENU SERVICE MECHANICS LUBRICATE POS Switch off the MYTHIC and disconnect the power supply Remove the Lysis and Cleaner bottle from their location Then put the MYTHIC on its left side pi p Unscrew but WITHOUT dismantle them the four Ff screws which maintain the syringe module Lift the bloc to exit the syringe pistons from the piston carriage Exchange the piston if needed _REF M22 SM 000 Rev O1 8 REPARING Unscrew and remove the eight screws which maintain the o ring late Remove the o ring plate and e
33. crease during emn Enclosure heating at 100 and Heating switch off no temperature increase during omn MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 27 77 OUTPUT FORMAT REF M22 SM 000 Rev 01 5 OUTPUT FORMAT The choice of the setting format speed is explained in page N 24 of the User s Manual 5 2 C2 FORMAT 5 2 1 General principle Mythic begins frame with headline MYTHIC X 2 eventual parameters CR Where X is the number of the Mythic maximum 2 characters Y is the user identification login maximum 10 characters Z frame identification The decimal separator is the dot The field separator is The line separator and the indicator of end is CR The uncut frame are Request of results send with send results Send calibration pack with results of calibration All the lines of a frame must be sent If no information the parameter is not fulfilled All the identifiant frame or parameters are in block letters All editable text field Identification lot are coded in UTF8 non Latin languages management 5 2 2 FRAMES 5 221REQUEST TO SEND RESULTS Mythic sends results after acknowledgement by the host Frames used in Hand Shake mode 5 2 2 1 1 Request to send results MYTHIC X Y RESULT READY Size CR 5 2 2 1 2 ACKNOWLEDGEMENT REQUEST TO SEND RESULTS ACK RESULT READY CR 5 2 2 1 3 ACKNOWLEDGEMENT RESULTS RESULT A B CR Where A OK or Erro
34. display is available with the super technician password BLANK HGE WBC GAIN only REC GRIH PLT GRIH FSC GAIN ALL GAIN LED The twenty types of blood can be changed and selected with the type STANDARD w STANDARD window ka This display enables to modify the HGB WBC RBC and PLT gain located in the upper window and the counting pulse parameters in the lower window THRES WEC WIDTH WEC WIDTH FSC THRES RBC WIOTH RBC TRHES WIOTH PLT FLT OFFSET e CI E m If certain type of blood animals need another quantity of lyse this quantity can be modified in the lowest window located on the right 7 2 2 3 Prod adjust The modification of one of these parameters must be done in compliance with Orph e after having performed a specific training In this case of modification without authorization Orph e guarantee will not be PREVIOUS FEER BURM IH n Not available manufacturing parameters ADJUST HGE ADJUST HGE BLANK HGE 2 allows to adjust the HGB blanc level LATEX WEC LRTEX WBC mE allows to adjust the WBC gain with the use of special latex particles LATEX RBC PLT REC GAIN LATEX REC PLT allows to adjust the RBC and PLT gain with the use of PLT GAIN special latex particles PUT Tag means that the adjustments are automatically done REF M22 sm o0ORevOl I 4 4 4 44 7 SOTWARE 72
35. e limits WBC RBC PLT HCT HGB Rejected value h results higher than normal value b results lower than normal value H results higher than panic value B results lower than panic value 4 2 LEUCOCYTE FLAGS 4 2 1 WBC curve flags This curve is done by the WBC channel measurement and does not appear on the normal results screen The flags L1 and L5 complete the flags N2 and IC CHANNELS MEANS VALUE VALUE Begining of Presence of platelet Percentage of cells in the Number of cells in the the curve to aggregate erythroblast or zone in regard of the number zone L1 small lymphocytes of lymphocytes L5 to the ned of Immature Percentage of cells in the Number of cells in the of the curve Cells from the mono zone in regard of the total zone polynucleated cells line number of leucocytes in the curve MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 23 77 4 FLAGS amp ALARMS 4 2 2 Scattergram flags REF M22 SM 000 Rev 01 All the followings flags level are adjustable in the 20 types of blood see section 3 4 4 of the User s manual MEANS SUSPICION VALUE VALUE En Noise 2 Immature Cells Right Lymphocytes Neutrophils Low Neutrophils High HL High Lymphocytes Page 24 77 Presence of platelet aggregate debris or resistant RBC Presence of platelet aggregate erythroblast or small lymphocytes Presence of Immature Cells
36. elines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strenghts from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM an FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should considered If the measured field strenght in the location in which the MYTHIC 22 is used exceeds the applicable RF compliance level above the MYTHIC 18 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the MYTHIC 22 P Over the frequency range 150KHz to 80MHz field strenghts should be less than 3V m REF M22 SM 000 Rev O1 DECLARATION OF CONFORMITY DECLARATION DE CONFORMITE DIAGNOSTICS WE THE MANUFACTURER NOUS LE FABRICANT NAME NOM ADDRESS ADRESSE DECLARE THAT THE PRODUCT DECLARONS QUE LE PRODUIT NAME OF THE MODEL NOM DU MODEL C2 DIAGNOSTICS Parc Agropolis II Batiment 14 2214 bd de la Lironde 34397 MONTPELLIER CEDEX 5 FRANCE MYTHIC 22 CONFORMS TO THE FOLLOWING DIRECTIVE 4 STANDARDS EST CONFORME A LA DIRECTIVE ET AUX NORMES SUIVANTES DIRECTIVE STANDARDS Henri CHAMPSEIX CEO Pr sident Directeur G
37. eneral FormDeclarM22 Version O1 MYTHIC 22 Copyright Orph e SA All Rights Reserved 98 79 CE IVD Medical devices IEC 60601 1 2 2001 EN 61000 3 2 EN 61000 3 3 EN 61000 4 2 95 A1 98 A2 01 EN 61000 4 3 02 EN 61000 4 4 95 A1 01 EN 61000 4 5 A1 01 EN 61000 4 6 96 A1 01 EN 61000 4 11 94 A1 01 EN 55011 Class B EN 55022 Class B IEC 61010 1 2001 IEC 61010 2 081 2001 IEC 61010 2 101 2002 JIS TO601 1 2 2002 CFR 47 Part 15 Class A DECEMBRE 2005 Page 5 77 REF M22 SM 000 Rev 01 TABLE OF CONTENTS DECLARATION OF CONFORMITY 5 1 INTRODUCTION 8 2 FLUIDICS 9 DOBNE RAI tenenti ue Mba ten iic Mere ele ut asia stan LET teka 9 LP PLUEDICS SEQUENCES DESCRIPTION tne etr tested ad dat ua k este el evi 9 O 3 F UTDICO VI W an 11 2A FLULDICS PARTS DESCRIPATON didt endet td LE M M EU ELM D EIU LIEN ten 12 241 Sampling module tet td eene atium edd ee i d E UL 12 DAR Counting bath 12 2 4 5 SYRINGES 12 Optical Dene deu ERE are Ome Reet net M nM CI LI DEM UE LONE 13 255 Obi cal Ol le dr EIE MuR 13 ES FEUIDICS DIAGRAM kaa ute Loi ut aac Ue Eo MA t M AA vege Pru UE
38. h Mechanical action SHUT DOWN 1 Entry menus End of day Cycle View Results TR NERE KKKKKKKK CHANGE RESULT GRAPHIC Target values Runs for checking View Levey Jennings entry RESULTS View chart results VIEW View full screen Cet MMA Na m FLAGS LEVEL CORRECT FACTORS CONTROL CYCLE MOTOR INIT REPEATABILITY LOGERRORS S SS SEMEN NS General Control Mechank Control Runs cleaning Runs mecha nica Runs for repeatability NA t z EFL CHECK NEEDLE PARAMETERS Counting windows CHECKROCKER adjustment EMEN CHECK SYRINGES CHECK SENSORS WN ANN RTS MXN oe LATEX WRC WY LATEX RBC PLT COE NEBEANERY WASTE __ OPTICAL BENCH ADJUST OPTICAL GAIN ADJUST THERMAL Launch Machanical action LOG INTERVENTION Se L sa MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 53 77 8 MAINTENANCE The quality of the results and the reliability of the MYTHIC 22 are directly linked to the strict respect of the maintenance hereafter described Piston greasing Needle o ring replacement Syringes o ring replacement Motor screw greasing Reagents level control see User s Manual Start up see User s Manual Automatic Cleaning see User s Manual Concentrated cleaning see User s Manual Shut down see User s Manual Cover cleaning see User s Manual Piston grea
39. ine with a checksum This sum is calculated from the beginning of the headline to the end of the line finishing the block The algorithm of the calculation for the sum is CRC OxFFFF For each byte Work on high neeble Indix byte XOR CRC Indix Indix AND OOOF CRC Table Indix XOR CRC divided by 16 Work on low neeble MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 33 77 3 OUTPUT FORMAT REF M22 SM 000 Rev 01 Indix byte divided by 16 Indix Indix XOR CRC Indix Indix AND OOOF CRC Table Indix XOR CRC divided by 16 Example in C language Hash table static const unsigned short ausCrcTab1 0 0000 OxCCO1 0xD801 0x1400 0xF001 0x3C00 0x2800 OxE401 OxA001 0x6C00 0x7800 0xB401 0x5000 0x9C01 0x8801 0x4400 computing unsigned short crc unsigned char pucData long Size 9 9 p 9 unsigned short 1 OxFFFF while ISize gt gestion par quartet du calcul 1 ausCrcTabi pucData usAcc1 amp 15 usAcci gt gt 4 1 ausCrcTabi pucData gt gt 4 usAcc1 amp 15 usAcc1 gt gt 4 pucData ISize return usAcc1 XOR gt Logical shift right d logical AND Page 34 77 Copyright Orph e SA All Rights Reserved MYTHIC 18 REF M22 SM 000Revol U U 6 PARTS LISTE amp BLOW UP VIEW 6 PARTS LIST amp BLOW UP VIEW 6 1 PARTS LIST 6 1 2 Two
40. ion 10 4 5 1 Pressure flag Occurrences conditions Consequences Flag message Pressure default Emergency stop check the tubing connection of the pressure sensor if all the tubing around the waste syringes are connected a leak on the waste syringe piston 4 5 2 Thermal flags Set temperature reagent or Run sample inaccessible enclosure not reached The set temperature are reached Fan failure Heating enclosure stopped REF M22 SM 000 Rev O1 FLAGS ALARMS Occurrences conditions Consequences Flag message Reagent temperature gt 60 C Heating reagent stopped all the fluidic _ cycles are inaccessible except for the technician Reagent temperature gt 70 Machine switch off check the heating bloc connection Only the technician can restart the machine See section 7 2 2 4 Enclosure temperature gt 50 Heating reagent and enclosure stopped all the fluidic cycles are inaccessible except for the technician Enclosure temperature gt 55 C Machine switch off check temperature sensor connection in the fluidics part Only the technician can restart the machine See section 7 2 2 4 Reagent temperature out of Run sample inaccessible range 34 amp gt 45 Enclosure temperature out of range 32 amp gt 40 Room temperature 17 or gt 35 Reagent heating at 100 and no Heating switch off temperature in
41. leaning 1 5ml valve 4 Waste syringe drain valve 7 7 Syringe down WBC bath drain valve 1 8 Syringe up Waste syringe drain valve 7 WBC DILUTION 9 Needle down to the dilution position and the syringes down to prime the diluent valve 8 10 Syringe up for the WBC dilution 2 ml Diluent outside of the needle valve 4 Diluent inside the needle valve 4 5 Bubble to mix the dilution valve 1 RBC DILUTION AND LYSING 11 Syringe down no valve MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 9 77 FLUIDICS REF M22 SM 000 Rev 01 Drawing sample of the first dilution 251 Diluent prime Vacuum prime 12 Needle up and syringe up Rocker move above the RBC bath Needle outside cleaning with diluent 0 5ml and waste vacuum valve 314 13 Needle down to the rinsing position 14 Syringe up Second outside needle cleaning 0 9ml valve 4 8 Lysis in the WBC bath 0 73ml valve 9 Bubble to mix the dilution valve 1 15 Syringe down RBC bath drain valve 2 16 Needle down to the dilution position 17 Syringe up for the RBC dilution 2ml Waste syringe drain valve 7 Diluent outside of the needle valve 4 Diluent inside the needle valve 5 Bubble to mix the dilution and diluent to rinse the counting head valve 611012 MEASUREMENTS 18 Syringe down to process the vacuum counting valve 8110 19 First measurement sequences CBC 2 second valve 10 20 Syringe up Diluent back pressure valve 4
42. ng38 20mm 005 1002 90 39 A Tubing39 180mm 005 1002 90 40 A Tubing4o j o 005 1002 90 41 A Tubinga jm 005 1002 90 42 A Tubing42 m _ 005 1002 90 43 A Tubinp43 250mm 1mm xm 005 1002 90 44 Tubing44 zem 005 1002 90 45 A Tubing45 10mm S 005 1002 90 46 A Tubing46 omm 005 1002 90 47 A Tubing47 2 15mm 2X5m J 005 002 9048 A Tubing48 m 005 1002 90 49 A Tubinp49 mm OF 005 1002 90 50 A Tubing50 2x20mm 005 102 9051 A Tubinp51 005 1002 90 52 A Tubing52 5mm 005 1002 90 53 A Tubing53 6omm 4 005 1002 90 54 A Tubing54 jm 005 1002 90 55 A Tubing55 140mm domm 005 1002 90 56 A Tubing56 imm 005 1002 90 57 A Tubing57 80m REF 22 5 000 Rev 013 ELECTRIC amp ELECTRONICS 3 ELECTRIC amp ELECTRONIC 3 1 GENERALITIES Only three boards compose the MYTHIC
43. nic low normal high normal High panic CR MON Value A B low panic low normal high normal High panic CR GRA Value A B low panic low normal high normal High panic CR WBC CURVE Channel valuet Channel value 128 CR WBC THRESHOLDS S1 S2 S3 CR RBC CURVE Channel valuet Channel value 128 CR RBC THRESHOLDS 51 52 CR PLT CURVE Channel valuet Channel value 128 CR PLT THRESHOLDS 51 CR 5 CR INTERPRETIVE WBC y y y y y y y y y y y y CR INTERPRETIVE_RBC y y y y y y y y y y y y CR INTERPRETIVE PLT y y y y y y y y y y y y CR COMMENT RUO message if necessary END RESULT Checksum value CR MYTHIC 22 Copyright Orph e SA All Rights Reserved 5 OUTPUT FORMAT Page 29 77 OUTPUT FORMAT REF M22 SM 000 Rev 01 A None or Rejection flag None or D or Hor hor or L Out of limit flags x None or Alarm y None or pathologic message S1 S2 S3 thresholds of the curves Between O and 127 This frame is available for LMG test Numeric values should be replaced by 5 plus for a out of range result or by 5 dots donor invalid results If RUO mode is not activated PCT and PDW are not transmitted 5 2 2 2 Alarm list S UP NOT DONE S UP FAIL QC NOT DONE QC FAIL INS M INS T INS P INS R INS H L1 L2 L3 L4 L5 R1 R2 P1 P2 P3 5 2 2 2 2 WBC pathologic alarms LEU LEU LYM gt LYM GRA GRA
44. of the MYTHIC 22 Fluidic Electric amp electronic Technical specification Output format Spare parts list and blowup view Maintenance procedure REF M22 SM 000 Rev 01 T I A FLUDICS 2 FLUIDICS 2 1 GENERALITIES MYTHIC 22 is a fully automated analyzer performing hematological analysis on whole blood The human blood venous sample must be collected in an EDTA K3 Ethylene Diamine Tetracetic Acid tri potassic tube in sufficient quantity The LMG results are available for six hours after the blood draw The use of EDTA K2 leads to a poor quality level of the LMG results It must be correctly homogenized before analysis It is recommended to use a rotary agitator turning between 20 to 30 turns mn during 10 minutes A volume of insufficient blood for the quantity of anticoagulant or a bad mixing may involve an erroneous result Two dilution rate are performed 1 170 for WBC HB 1 10000 for RBC PLT 2 2 FLUIDICS SEQUENCES DESCRIPTION SAMPLING SEQUENCE 1 Syringes down no valve Less than 16 7 pl of blood sample are draw inside the sampling needle Diluent prime Vacuum prime 2 Needle up and syringe up valve 314 Needle outside cleaning with diluent 0 7ml and waste vacuum 3 Rocker move above the WBC bath and the syringes up to drain the waste valve 7 4 Needle down to the rinsing position 5 Syringe down WBC bath drain valve 1 Tubing transfer drain valve 11 14 6 Syringe up Second outside needle c
45. on Printer No printing Check the paper Check the electrical connection Bad printing Check the black ribbon MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 75 77 10 TROUBLESHOOTING REF M22 SM 000 Rev O1 10 3 TROUBLESHOOTING MESSAGE This section allows to know what to do when a troubleshooting message appears on the screen MESSAGE ACTION ANALYSIS MODE UNKNOWN Enter an analysis mode BACKUP BAD FOLDER DUPLICATION Re start the Mythic BACKUP CALIBRATION HISTORY IS FULL Delete the calibration results BACKUP ERROR HISTORY IS FULL Delete the error history BACKUP FOLDER NOTFOUND Re start the Mythic BACKUP MEMORY IS ALMOST FULL PLEASE DELETE RESULTS Delete results BACKUP NO MEMORY AVAILABLE FOR BACKUP REPEATABILITY HISTORY IS FULL Delete the repeatability results ID AND OR PID MANDATORY CHECK SETUP SID ALWAYS MANDATORY Enter an ID and or PID and SID REF M22 SM 000 Rev O1 BO TROUBLESHOOTING MECA HOME NOT FOUND Perform an initialization or a control cycle IMECA MOTOR GAP Nothing IMECA NEEDLE NOT IN TOP POSITION Perform a control cycle MECA MOTOR BUSY Re start the Mythic PRINTER NO PAPER Add some paper O PRINTER RESPONSE Switch on the printer or invalidate the printings O PRINTER SELECTED Switch on the printer or invalidate the printings OUT OF RANGE Modify the value RINTER ERROR Switch on the printer or invalidate the printing
46. ool see section 9 3 3 4 5 in the user manual 2 4 3 Syringes module This module patent pending enables to draw the sample to distribute the reagents to drain the baths to do the vacuum necessary for counting and to push the waste to the waste container It is made with a manifold maintaining the fluid commutation valves and with the syringes bloc including five syringes The sampling syringe The lysis syringe The two waste and vacuum pressure syringes And the diluent syringe Only one motor drives the five syringes The diluent input and the waste output are also included in this manifold REF M22 SM 000 Rev 01 T I 2 FLUIDICS 2 4 4 Optical bench This manifold allows to drive the different reagent toward the counting bath module or toward the optical bench It maintain seven commutation valves 2 4 5 Optical manifold This very important part two pending patents allows to measure the leucocytes sub population It is made of three parts an illumination module which contain the semi conductor light source a beam adjustment glass smear and lens a flow cell made in injected plastic in which flow away the sample and the passive sheath and the detection module which contain lens a beam separation glass smear photodiodes and the amplifier boars for the two measurement axes MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 13 77 2 FLUIDICS REF M22 SM 000 Rev 01 2 5 FLUIDI
47. parameters nc d Man MR des Me UE MA i DUM EA nn eA 50 72 2 Od aa nct uuu aun ua EN eite EER OE NEN RE RES 50 Ze MI 1 6 RS NU aa PM 51 2 9 beg adustus toad c Dd LU eL spa uae Ree 51 Ze 26 OPTICG gam adjust si asan titel de addled LE nb 52 ARG ARC ORE SCENIC eaters et o Me EDU dto LL C S aA e E 53 8 MAINTENANCE 54 OL rung z P Nac I cem X M 54 02 CONCENTRATED CLEANING nr 55 9 3 CLEAN OUT etn a ten Ga ua ana a oon ML de CIEL NI Acne IA D 26 9 REPAIRING 57 SYRE NER GENE Y coctus de EN MUI 57 ge SAMPLING MODULE Se ER aaa LH M enc RNA IL Ma A Md E Lu 57 9 2 Needle replace men sni sta u en UA LLLI nsn tu ulat Da Un eee hdd or 57 9 2 exComplete rocker module replacetells oi od een lois oit al k aide a a 57 2 S COUNTING BATHS MODULE cc asd EM LEE SEDIS MM LIED IU M EEUU MM uM EI US UE 29 2o eue doti trt AURI MIENTO MEN DUERME MM MEUS IMMER 59 S 3 2 B lhs o nte ciat M M E E E M akuy kaa M 29 92 3 3 Aperture block replacement teo HAR obe D n eA DM LA E de 29 9 3 4 Complete baths module replacement inner 99 OVBINGE MODULE s kS AN NA td tI I OP MM A MILI M 60 SAT PIS TOO ESA EME A OE css A AE 60 94 2 Osring or piston replacement dataset at
48. pe of equipment Please select the suitable battery according to the handling instructions of the equipment Never use or leave the battery in a hot place such as under the direct rays of the sun or in a car in hot weather If you do this may cause distortion leakage overheating explosion or fire Never allow the battery to come in contact with water If it does this may cause the battery to rust or lead to distortion leakage overheating explosion or fire Never store the battery in a hot and highly humid environment Doing so may cause the performance of the battery to deteriorate In certain environments this may lead to distortion leakage overheating explosion or fire Disposal The battery may be regulated by national or local regulation Please follow the instructions of proper regulation As electric capacity is left in a discarded battery and it comes into contact with other metals it could lead to distortion leakage overheating or explosion so make sure to cover the and terminals with friction tape or some other insulator before disposal MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 67 77 3 REPARING REF M22 SM 000 Rev 01 9 5 3 Preamplifier board replacement NOTA Before to dismantle the preamplifier board dismantle the Main board see section 9 5 2 Remove the two screws located on the copper side Remove the preamplifier box from the main board To reassemble the p
49. r Code B Reserved REF M22 SM 000 Rev 01 5222 ROUTINE RESULTS MYTHIC X RESULT CR DATE jj mm aaaa CR TIME hh mm ss CR MODE NORMAL CR UNIT Unit CR 0 1 ou 2 SEQ Sequence number CR SID Sample Identification CR PID Patient Identification CR ID Identification CR TYPE Blood Type CR TEST Run Test CR OPERATOR login Mythic when analysis is performed CR WBC Value A B low panic low normal high normal High panic CR RBC Value A B low panic low normal high normal High panic CR HGB Value A B low panic low normal high normal High panic CR HCT Value A B low panic low normal high normal High panic CR MCV Value A B low panic low normal high normal High panic CR MCH Value A B low panic low normal high normal High panic CR MCHC Value A B low panic low normal high normal High panic CR RDW Value A B low panic low normal high normal High panic CR PLT Value A B low panic low normal high normal High panic CR MPV Value A B low panic low normal high normal High panic CR THT Value A B low panic low normal high normal High panic CR PDW low panic low normal high normal High panic CR LYM Value A B low panic low normal high normal High panic CR MON Value A B low panic low normal high normal High panic CR GRA Value A B low panic low normal high normal High panic CR LYM Value A B low pa
50. s PRINTER IS BUSY Switch on the printer or invalidate the printings RINTER IS OFF Switch on the printer or invalidate the printings INSE NOT DONE Perform a rinse cycle RS232 BUFFER FULL S232 ACK ERROR Re start the Mythic S232 INTERNAL ERROR Re start the Mythic RS232 SYNCHRO ERROR Re start the Mythic 5232 TIME OUT Re start the Mythic TART UP CYCLE NOT DONE VM BAD VERSION VM COM TIME OUT VM ILLEGAL SERIAL NUMBER VM UNMATCH VM WG VM XB YSTEM EEPROM COM ERROR YSTEM INTERNAL TIME OUT YSTEM FATAL ERROR YSTEM UNKNOWN ERROR w 55 Replace the bottle and perform a prime Lysis Replace the waste container MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 77 77
51. sing see User s Manual VON Needle o ring replacement see section 9 2 1 Syringes o ring replacement see section 9 4 2 Motor screw greasing This operation is described during the ORPHEE training _REF M22 SM 000 Rev O1 8 MAINTENANCE 8 2 CONCENTRATED CLEANING Such cleaning is necessary when MYTHIC 22 is in permanent rejection for one measured parameter Prepare a Sodium Hypochlorite solution at 12 in chloride diluted 4 times with distilled water if sodium hypochlorite is concentrated at 48 in chloride HERLI Return to the main menu by pressing EN Then press START SET UP OPERATOR LOG OUT REAGENTS LOGS FFF HYDRAULICS REUIOLS Qe JE SCREEN Vis nee HU ED Press Press to start the cycle drain the counting chamber BLEACH INIT PRESS MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 55 77 Wait for the information window Open the door on the right side see section 1 1 3 Put 3 ml of sodium hypochlorite solution in each counting chamber Press the button in the window located in the middle of the screen MYTHIC 22 performs a maintenance cycle of the apertures followed by a stand by mode during 2 min After 2 min MYTHIC 22 rinses all of the elements Operator can perform an analytic cycle Wear rubber gloves and wash hands with a disinfectant after completion of work 8 3 CLEAN OUT This procedure has
52. to be used when the MYTHIC 22 must be transported for a long distance by car train or plane This procedure consists of cleaning the MYTHIC 22 with sodium hypochloride then rinsing it with distilled water and drying it before the transportation PREVIOUS SCREEN _REF M22 SM 000 Rev O1 9 REPARING 9 REPAIRING For all the following operations wear rubber gloves and wash hands with a disinfectant after completion of work Be careful with the edge of the metal sheets after having dismantled of the covers 9 1 EMERGENCY STOP In case of mechanical or hydraulic problem immediately press e the MYTHIC 22 will make a emergency stop After having identified the problem it is necessary to perform a Control cycle SERIJICE For an access to this cycle press from the MAIN MENU then press MOTOR INIT HDRRULICS MECHANICS REPEATABILITY LOG ERRORS TECHHICIBH 9 2 SAMPLING MODULE 9 2 1 Needle replacement Perform needle replacement as is describe in the User s Manual 9 2 2 Complete rocker module replacement Before to replace the complete syringe module drain all the tubing Remove the diluent lysis and cleaner straw from their bottle e Perform a PRIME ALL MAIN MENU REAGENT PRIME ALL Switch off the MY THIC and disconnect the power supply Remove the Lysis and Cleaner bottle from their location Remove the upper cover three screws from the MY THIC MYTHI
53. xchange the o rings REASSEMBLY PROCEDURE To reassemble the unit carry out the various operations in the opposite direction 9 4 3 Complete syringe module replacement Before exchanging the complete syringe module put the MYTHIC on its left side as describe above section 9 4 2 Disconnect all the valves connection Unscrew but WITHOUT dismantle them the four screws which maintain the syringe module 7 lt Remove the waste and diluent tubing Cut the fixation ties of the tubing rs Unscrew and remove the screws which maintain the ground Y w Disconnect the tubing Remove the complete syringe module NOTA Take care about the tubing reconnection if necessary cut few millimeters 2 to 4mm of tubing or exchange them MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 61 77 9 REPARING REF M22 SM 000 Rev 01 9 5 BOARDS REPLACEMENT THE THREE BOARDS OF THE MYTHIC 22 ARE VERY SENSITIVE TO ELECTROSTATIC DISCHARGE YOU HAVE TO USE A BODY EARTH CONNECTION TO HANDLE THE BOARDS FOR THEIR REPLACEMENT OR DISMANTLING 9 5 1 GUI board replacement Unscrew but WITHOUT dismantle them the two upper screws which maintain the front panel Unscrew but WITHOUT dismantle it the lower screw which maintain the front panel _REF M22 SM 000RevOl 9 REPARING Lift the front panel and then pull it toward the front of the instrument VERY CAREFULL
54. years spare parts kit Part Number 002 1002 50 20 02 Designation Mythic 22 2 years spare parts kit Rev Part Number 1 A 002 1002 50 11 Mythic 22 Tubings kit 2 A 005 1001 03 03 Piston 22 A lo10 2501 0205 Sampling needle 1 312 05011554262 O ring 21554x262 Silicone 60SH 2 ste osor 21e9 262 O21 89x2 62 Silicone 6054 1 312 0501 630 240 O ring 6 3x2 4 Silicone 60SH 2 jpuzososisi160 lo ring 0132916 Fluocarbon 80SH 3 05054012500 O ring L4x1 25 Fluocarbon 8084 5 2 312 0505 500 100 O ring Q5x1 Fluocarbon 80SH 6 1 2 Maintenance kit Part Number 002 1002 50 02 Designation Mythic 22 Maintenance Kit Bottle cap filter 5 153 0101 0025 Cables Ties O OOS 1 812 0505 140 125010 O ring 1 4x1 25 Fluocarbon 80SH 1 4t0 0501 0202 Silicon grease 3gr 1 700 1101 10 Short TORX T10 Tool 1 1 1700 1101 20 Short Arm TORX T20 Tool MYTHIC 22 Copyright Orph e SA All Rights Reserved Page 35 77 6 PARTS LISTE amp BLOW UP VIEW REF M22 SM 000 Rev OL 6 1 2 O rings kit Part Number 002 1002 50 10 Designation Mythic 22 O Rings kit Part Number Designation 1 312 0501 1554262 O ring 15 54x2 62 Silicone 2 312 0501 2189262 O ring 21 89x2 62 Silicone 6OSH 1 312 0501 630240 lO ring 6 3x24 Silicone60SH 2 312 0505 1310160 O ring 13 1x1 6 Fluocarbon 812 0505 40 12501