image

Contents

1. S 3 2 CAUTION amp amp 3 3 Checking for Confirmation eoccccccccccccccccccccccccccccscce Features of product ecce eee ee eee eee eee eee eeoeoecococcccccceeeeeeeeeeece SPECIFICATIONS 7 Product CONSIST 5 Product Description 7 1 Features G 7 2 LCD Window and functions cocccccccccccccccccccccccccccsscsss 7 3 Each part of the button and functions 0 7 4 X ray Exposure using Remote amp 2G 7 5 Replacement and charge of the battery packeeeseesseesse 2 7 6 Battery pack charge OD Operation 8 1 Starting and ending of 5 8 2 X ray generating time Control 8 9 Relation between X ray g
2. Increase time set up Y Decrease time set up 8 3 Relation between X ray generating time and part of teeth Table Adult Mode Film 0 10 0 15 gt 0 25 gt 030 100405 0452 6 55 0 60 vare 0 15 0 25 0 30 0 40 0 45 0 55 0 60 Digital 0 05 40022 5 905 19258 1650 040 1045 0552 0 10 0 15 0 25 0 30 0 40 0 45 0 50 Indication af 3 step time increase level by of teeth SSS 24 8 4 X ray EXPOSURE Using Hand Held button Exposing of X ray with buzzer sound only while the button 15 being pressed X ray is generated only while C button is pressed So even though the buzzer sound is being made the generation of X Ray stops when taking the hand off the button Setting time 15 stored to product that do exposure on last 1me 15 not stored that do not exposure The status indication changes from READY to EXPOSURE while the X Ray is being generated READY EXPOSURE State Change 6 60 e 90 8 5 User Set Up Mode At the status of power on pressing the A together and then releasing you will enter into User Set up Mode MENU CHANGE V B Mode Film Power save 5 min O Back light 1 sec Bat 15 5V Ver 1 01 Exp 000000 Move between menu button Change the set up value button Exit the user set up mode After turning off the unit by pressing it turn on the unit again then unit starts with changed mode and
3. OG 3 1 Warning In use Not to exceed the dose of X ray that is necessary for image diagnosis Be handled or operated only by qualified personnel something wrong with the patient during using of the product Proper actions have to be taken such as stop the using product stop operating the unit when there is problem with the unit Please contact the authorized A S center Use the product avoiding the follow places Where there is chemical material or explosive gas Where it is damp Where there is direct ray of light Where ventilation is poor Dusty place Humid place salty place keep the following temperature and humidity for normal operation of the unit For safe keeping e temperature 40 70 C e humidity 10 90 RH During using e temperature 10 40 C e humidity 30 70 RH For optimum condition e temperature 10 80 C e humidity 40 60 RH safekeeping not use Do not keep the unit where it is damp salty and dusty which might cause the bad effects for operating of the unit Do not place the unit long at the place where the temperature changes steeply or in the direct ray of light Do not keep the unit where there is explosive gas or chemical materials Do not keep the battery pack long in the charger while connecting with the electricity Keep the charged battery pack in the separate keeping case Other Contact the authorized A S center in case of the unit is out of order Discard in the
4. REP SERIAL NUMBER AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY The design of the product is subject to change without notice to enhance the performance 17 Product Description 7 2 LCD Window and functions CD 2 eg p gt RADY T 1 0 10 9 CD Adult mode Adult mode indicates 2 Teeth part of adult Adult teeth part indicates 3 Child mode Child mode indicates 4 Teeth part of child Child teeth part indicates 5 Indication of status Indication of the present operation of the unit 6 Status of battery Indicate the remaining of battery by 3 steps D Digital mode icon Indicates digital sensor is setup and disappears when film mode is setup default is film mode Exposure time display Indicates irradiation time at each part of teeth mode 9 3 step time increase level by part of teeth The specifications and functions in this manual are subject to change without notice to enhance the performance 18 Product description 7 3 Each part of the buttons and functions Top side button Front side button EXP button Power button Time mode control button e button Turn on off the unit Time Mode control button Time setup by each teeth mode 24 steps in each mode gt button Time increase gt button Time decrease e td button Exposing of X ray with buzzer sound only while the button is being pressed and irradiation
5. EU this Symbol indicates waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately To minimize the risks involved in disposing of these waste products please contact an authorized representative of the manufacturer or an authorized waste management company for information concerning the decommissioning of your equipment 9 A 3 2 Caution Do not expose the unit to the extreme shock or vibration Do not use the unit placing it at dusty dirty and Inclined place Do not jerk the connection cable between the charging desk device and the charger Do not allow the patient touch the unit and do not leave the unit within the reach of the patient 3 3 Confirmation check Check the condition of the battery pack and the unit before using make sure the proper operation state of the unit Check again the parts of the unit that is touching directly the patient Use the unit according to the procedure for operation Use the unit in the comfortable place where ventilation is good After using of the unit check indicating lamp for remaining battery without fail and charge the battery if the remaining of the battery is not sufficient For charging of battery pack use only the exclusive charger without exception Keep the unit clean by using the clean dry piece of cloth to prevent the terminal of battery pack from touch of metal and other different materials The
6. IFU 01 DEXCOWIN Creative Innovation and Technology Serial No Thank you for purchasing DX3000 Please read this manual carefully before using DX3000 The quality of this product guarantee is contained in this manual C 0086 Portable X ray System DX3000 Super Small Super Light Convenience of cordless Use the charging battery pack take off type Great reduction of X ray exposure Up to 1 7 of existing X ray systems Adopting high frequency circuit of inverter type The most suitable to diverse digital sensors Simultaneous using of digital sensor and film Auto set Timer Automatic exposure time set up by adult child and location of teeth gt More than 300 radiography is possible with one time charging of battery More compact weight than existing portable dental X ray 1 8Kg User Manual 1 Contents User Manual CONTENTS amp G Safety Marks 7 3 1 Warning
7. On it button Method How to set up the mode of radiography CD User set up mode 4l 2 Move the menu Change the mode Select the time CD Increase set up time 2 Decrease set up time EXPOSURE of X Ray C Button service Manual specification Manufacturer DEXCOWIN Co Ltd Name of Medical equipment Portable X ray System Model DX3000 Device Classification Class MDD 93 42 EEC as amended by 2007 47 EC Classification Type of protection against electric shock Internal powered equipment Degree of protection against electric shock TYPE Applied part Voltage range 12V 16 8V DC Power consumption 240W Max Applications This product has to be used only for dental diagnostic X ray system for dental or extra oral X ray radiograph Battery pack Material Lithium Polymer Charge Condition e Current 1 0 A e Voltage 16 8 VDC e Temperature 0 C 40 C Use condition e Max current 20 A e Voltage range 12 min 16 8 MAX V e Temperature 20 60 C Charger Model JBL910A 1600 Inout Voltage 100 250 VAC Output Voltage 16 8 VDC Current 1 0A High Voltage Tank 1 X ray tube D 081B Model D 081B Manufacture TOSHIBA Electrical Data Circuit Self rectified Center grounded Operating Tube Voltage 50kV 65kV A Focal Spot 0 8 mm Input Energy at 1 0s 600W refer to rating charts Mec
8. as stove and heating device during using Please do not remodel device arbitrary For safe use it is recommended to have the unit checked regularly at the designated A S Center For using and keeping avoid as following places Damp places Places in the direct ray of light Dusty places Moist places Places in poor ventilation oalty places Where there is explosive gas or with chemicals For normal operation following temperature and humidity have to be kept For safe keeping e Temperature 40 70 C e Humidity 10 90 RH During using e Temperature 10 40 C e Humidity 30 0 RH For optimum condition e Temperature 10 30 C e Humidity 40 60 30 Maintenance and Keeping Safe keeping DX3000 Keep the unit avoiding the place where it is damp and in the direct ray of light Do not keep the unit where it is dusty dirty and sloping Do not keep the unit where it is so high temperature or so low temperature When the unit is not in use for long remove the battery pack and keep it separately Safe keeping of the charger and the table top device for charging Care has to be taken not to let the foreign material be stick at the terminal of the charging device It has to be kept after lightly cleaning with cotton bud or soft cloth When not in use take the plug off the outlet and when it will not be used for long keep it in the case Be careful of metal material such as coins key and clip etc not
9. e of Portable X ray Product System Place of Purchase Date of Purchase Address The product of DEXCOWIN Co Ltd observes regulation for protection of consumers notified by The Ministry of Economy DEXC OWIN Creative Innovation and Technology DEXCOWIN Co Ltd 606 Woolim Lions Valley II 680 Gasan dong Geumcheon gu Seoul Korea For service contact with phone 82 2 202 2880 htto www dexcowin com 35 13 Instruction for Quick Start Quick Start summarized tables which are explained in the user manual This explanation can not take whole explanation of product use method This is just for user convenience to help understand the unit prior to actual use Based on this summarized Quick Start use the unit after having read the manual completely without fail If you use this unit after familiarized with it you can easily use with Quick Start Instruction for Quick Start Ready Insert the battery pack into the unit Power ON Using it Button turn the unit on and LCD window is lit with sound Check Check the balance of battery pack and make sure the conditions of the product Mode selection Select the method of radiography such as film or digital sensor Exposure time Select the exposure time by location of teeth for radiography Radiography Radiography the area of patient s teeth using film Digital sensor Power Off After press W button LCD screen turns off with OFF sound Operating Power
10. enerating time and part of teeth table eee eee ee ee 24 8 4 X ray EXPOSURE DF 8 5 User Set Up MOde amp ee eee eee eee e eee eoo oo oossssscsccceeeeeeeee OF 8 6 Replacing and charging of battery pack ee eee ee ee eee e 2O 4 Contents Maintenance and Keeping 10 Instruction and Troubleshooting guidelines 33 11 Quality Guaranteed J T2 Warranty 000000000000000000000000000000000000000000000000000000000000006 IN 13 Instruction for Quick Start 00000000000000000000000000000000000000000000 36 Service Manual Specifications C High Voltage Tank Electro Magnetic Compatibility 000000000000000000000000000000000000000006 Q Radiation safety 2 Preface Thank you for purchasing the portable X ray system of D
11. er convenience 4 Remaining voltage of battery Display value of voltage that is remained in battery pack 5 Numbers of times of exposure of the unit Display the number of exposure times of the unit up to now Version of the unit Displays the version of the present unit 8 6 Replacing and charging of battery pack Change with the extra battery pack when the LOW battery direction lamp starts flicking Turn off if LOW battery sign appears Please open battery pack cover side of product and change with extra battery pack Battery pack cover opens by pulling forward after press fixing lever down e Put homologizing correctly battery pack insertion slot and guide line Please close cover after making sure Low battery screen battery pack came to end certainly Auto power off Power lamp flicking or LOW battery indication appears on LCD screen stop using and replace with the BATTERY charged battery pack EMPTY Please charge battery f use continuously without replace of battery after LOW battery patent appears the power will be shut off automatically Auto power off screen 29 9 Maintenance and Keeping Keep the unit and the battery pack in the place where only authorized personnel can reach Please clean often the product by using piece of cloth and to keep radiation aperture s clean because patients are contacted directly Do not put the unit close to the heat such
12. ethod Oil immersed and convection oil cooling x C C CU O gt c io L Rating Characteristics V Filament Characteristic VI High Voltage Tank Vil Anode Heating amp Cooling Characteristics VII Cathode Emission Characteristic ic iy p X E JR jJ J High Voltage Tank Distance from focal spot to radiation aperture and the diameter of X ray I Electro Magnetic Compatibility EMC As this product is used without connection of a cable there is no EMC risk This product has EMI characteristics less than 4 9dB at 1000MHz which may make an interference with another device having similar characteristics Avoid to use together with these devices Refer to table 1 Keep away from devices which make serious radiation noise strong electromagnetic field Refer to table 2 As this product is not used with cable there is no risk of EMC interference with another device through the cable If this product is remade for connection with another device or part intentionally it will make an interference with another device the manufacture don t take responsibility for risk of EMC Refer to EMC part of Service manual Electro Magnetic Compatibility EMC Electro Magnetic Interference Radiated emission electric field 30Mhz 1000MHz Min limit margin 4 9 dB at 1000 MHz Radiated emission data Table 1 Anten
13. excowin Co Ltd This product has been so developed as to not only take radiographs in analog way using conventional film but also be used in Dental Digital Radiograph DDR using digital sensor as well Please use following requirement for application Applications This product has to be used only for dental diagnosis for taking dental or extra oral X ray radiograph TYPE B Applied Part Type B Applied Part provide the lowest degree of patient protection of all the types of applied part and are not suitable for direct cardiac application For proper and safe use of this product this device has to be used only authorized and qualified personal as doctors and diagnosis and treatment X ray engineers and has to be used after fully familiarized with user manual 3 Satety Marks Warning caution and acknowledgement indication in this manual are provided for assuring of proper use of the product to protect the user s safety and to prevent the loss of the property Therefore the users have to observe these safety requirements Warning marks indicate and explain for the Warning cases which might be harm to the user cause death or property damage Caution marks indicate and explain for the cases Caution which might cause harm to the body due to improper use of the product The check marks indicate and explain for The cases which the user must follow in installation operation and maintenance repair of this product
14. hanical Data Dimensions Overall length 69 mm Max Diameter 31 mm Target Angle 20 degrees Inherent Filtration 0 8 mm Al Weight Approx 96g Cooling Method Oil cooling 41 High Voltage Tank Maximum and Minimum Ratings Maximum Tube Voltage 65kV Maximum Inverse Tube Voltage 5kV Maximum Tube Current 19mA see rating charts Maximum Filament Current 2 0A Filament Voltage 2 9 4 0 V At max filament current Temperature Characteristics Anode Heat Storage Capacity 6 0 kJ 8 5KHU Max Anode Heat Dissipation Rate 128 W 180 HU s Maximum Rating Charts Absolute Maximum Rating Charts SELF RECTIFIED FOCAL SPOT 0 8 mm TUBE CURRENT mA EXPOSURE TIME s 42 High Voltage Tank Emission amp Filament Characteristics SELF RECTIFIED VOLTAGE V TUBE CURRENT mA FILAMENT FILAMENT CURRENT A Anode Thermal Characteristics HEAT STORAGE kJ HEATING TIME min 43 High Voltage Tank 2 X ray Tube XD12 0 56 70 Model XD12 0 56 70 Manufacture WANRAY Electrical Data Circuit Self rectified Center grounded Operating Tube Voltage 40kV 100kV Max Focal Spot 0 8 mm Power Is W 560 W Max Continuous Power W 110W Mechanical Data Dimensions Overall length 73 mm Max Diameter 30 5 mm Target Angle 19 degrees Inherent Filtration 0 5 mm Al Weight Approx 120g Cooling M
15. is inserted to battery pack holder to direction that is correct on slot of device for charger Battery pack insert Battery holder direction exercise direction If the charge is completed charge lamp changes from red color to green When segregate battery pack in device for charger after charge is completed please pull backward and segregate battery pack holder Keep the battery pack separately or unplugged after charge lamp changes green or not in use Product description Please keep the following rules whenever charging the battery pack Use only appointed charger and battery pack Keep apart the battery pack in the special storage case after charging completed Avoid the recharge for charge completed pack Take care not to contact the metal goods of coin key clio etc on the battery terminal plate Battery pack is consumer goods Duration of battery pack is shorted slowly than first purchase Please buy new battery pack if duration decreases by half than when use first Purchase inquires in the A S center Exhaust of battery pack Used battery pack should not throw away in trash but please put that in the recycling box or exclusive case 8 Operation Starting and ending of the system Press the power button during 1 second then logo will appear and become initial stage 02 Press the power button during 1 second in active state power turns off 8 2 X ray generating time control e
16. na Reading Correctio Emission Pol Level n Level H V dBuV AF CL dB m Electro Magnetic Susceptibility Table 2 Frequency Field strength Modulation Frequency range step Radiated Field 80 2500 MHz 3 V m 10 V m AM 80 196 3s 1KHz sine Wave Test Field level Duration Axis of frequency EMF orientation Magnetic Field 50 Hz 3 A m 60 seconds X Y Z axis each axis Radiation safety Leakage radiation gt The following is leakage radiation at 1m from focal spot Unit mR h mGy h AD 23K ND Not Detected gt The following is leakage radiation at 1cm from case surface Unit mR h mGy h 3 93 0 0893 1 43 0 0143 7 8 0 078 jw i88 gt 4 002 3 08 Fi E 0 0002 0 0308 5 85 0 0585 A 0 04 0 0004 5 23 0 0523 7 19 0 0719 Leakage radiation at 1 from case surface is less than 8mR h 0 08mGy h 50 Radiation Safety Strayed Radiation gt The following designate significant zone of occupancy gt Please put private protection devices within this area 2 1 Set the significant zone of occupancy significant zone of occupancy sl Radiation Safety 2 2 Test method and result of stray radiation 60 cm Test Method of stray radiation I Stray Radiation 0 01 Air kerma mR h Absorbed Dose mGy h Test result of stray radiation 52 DEXCOWIN Creative Innovation and Technology DEXCOWIN Co L
17. replace the On or LCD window battery pack does not work when If the battery will not be used for long time the unit is turned on separate the battery from the unit for prevention discharge LCD window is not Replace the battery pack lit after the sound If the status remain same even after having of Power On is replaced the battery contact with the designated heard pr A S Center for repair The power turns off Check the remaining battery itself while in use This unit features automatic power disconnection to save the unnecessary consumption of the battery pack Check the set up of the automatic power disconnection 11 Quality Guaranteed This product is warranted its quality from the date of purchase to one year It is hard to recognize the warranty period the warranty Is applicable three months after it was manufactured Products defects by itself or Improper workmanship the product will be either repaired free of charge or exchanged according to the status of the product Even within the warranty period if the damage is caused by failure of the user free of charge is not applied The attached level for sealing is broken free of charge and warranty are not applied The repair at the place any other than designated A S Center is excluded from warranty The service period will not be extended for the product which was repaired with charge or free of charge 12 Warranty Name of DX3000 Model Nam
18. set up value Menu explanation Mode Film film digital mode indicates Power save automatic power disconnection time default 5 min Back light activating of back light default 1sec Bat remaining voltage of battery Exp the number of time of EXPOSURE of the unit Ver 1 01 version of the unit 26 D Film Digital mode Using locate the menu in Set up Mode position and by pressing the v utton film mode or digital mode can be selected When digital mode is selected the digital mode icon is displayed as followings MENU amp CHANGE V gt READY m B Mode Digital Power save 5 min Back light 1 sec Bat 155V Ver 1 01 Exp 000000 User set up mode Digital Mode Icon For film mode exposure time is from 0 28 up to 1 355 for digital mode exposure time is from 0 055 up to 1 205 each mode has 24 steps of ranges 2 Automatic power disconnection time While the unit is not in use the power is automatically disconnected following by the set up time in order to save unnecessary power consumption Default is set up as 5 minute and it could be adjusted by 1minute According to the convenience of the user interval from 1 minute up to 10 minutes at 3 Back light activating time g READY T 740 20 The light illuminating time can be set for LCD window Basically setting value of 1 second is given and can be adjusted by 1 second up to 10 second for us
19. td oo 606 Woolim Lions Valley 680 Gasan dong Geumcheon gu Seoul Korea For service contact with phone 82 2 2027 2880 http www dexcowin com Paul Sweeney Boston Dental Scientific C Isabel Colbrand 10 Local 68 28050 Madrid Spain Tel 0034 91 358 8006 Fax 0034 91 358 8018 53
20. to be contacted with terminal plate of battery pack or charger Do not attempt to use charger with other makers battery Maintenance and Keeping Daily Monthly tests and Annual maintenance 1 Daily tests When turned on power of product please confirm whether on sound is heard and normal early screen appears Change establishment investigation time using manufacturing button and time confirms whether is changed normally in display Please confirm whether buzzer sound is heard and EXPOSURE phrase and investigation lamp of LCD screen are seen normally at X ray exposure Please confirm charging of battery pack and normal operation existence and nonexistence of charger 2 Monthly tests Please confirm whether duration of use of used battery pack is suitable Buy new battery pack if duration of use decreases comparing with first purchase by half Battery pack is consumer goods duration of use is shorted slowly than bought 3 Annual maintenance Every year for safe use of product safety check up is necessary 10 Instruction amp Troubleshooting guidelines User should check below symptoms list before asking repair Battery indication Check the remaining battery and replace with lamp is flicking charged battery Ensure if the battery pack cover is properly closed and if the cover open close the cover completely to the end The sound of Power Check if the battery pack is loaded or
21. unit and parts should have the regular check without fail 10 4 Features of product Name of product Portable X ray System Model DX3000 Real meaning portable system by using replaceable battery pack Graphic LCD window user friendly interface by simple button touch Life time is 8 years LCD window indicates Exposure time setup by adult and child Film and digital mode Operation status of the unit Remaining of battery Exposure time setup by part of the teeth Function for user Simultaneous use of digital and film Automatic power disconnecting time setup Continuation of back light time setup Remaining voltage of battery indication 3 steps time increase level by part of teeth 1 Step 2 Step 11 5 Specifications Name Mobile X ray System Item Classification 2 Code A11080 02 in Korea Grade Class 16 MDD 93 42 EEC as amended by 2007 47 Classification B Type applied part Tube Voltage 60kV fixed Tube Current 2mA fixed X Ray Tube Focal Size 0 8mm Cooling system Oil cooling Total filtration 1 5 mm AI Inherent filtration 0 8mm fixed added filter 0 7 mm Aluminum Target angle 19 20 Time Set up function 0 05 1 35 sec Film Digital included Source to image receptor distance 10cm fixed by cone X ray field 50 mm round style Size of main body 139 W x 163 5 H x 66 5mm D Weight of body About 1 8 kg except the batter
22. will stop when button is released DMS method Dead Man Switch 19 Descriptions 7 4 X ray Exposure using Remote control Remote Control Jack i 1 Allows user to operate X ray by way of approved remote control m Use only approved Remote Control Approved remote control is not included with the X ray but can be purchased as an option from distributors Product description 7 5 Replacement and charge of the battery pack PREADY v pepe zem nl LCD Screen when low battery The battery pack replacement If LOW battery direction ramp is blinking or LOW battery sign appears on battery pack residual quantity expression 3 Turn off the power when the LOW battery sign appears Please open the battery pack cover and change with extra battery pack Fixing lever down and then pull the cover forward for opening the battery pack cover Put homologizing correctly guide line of the pack top portion and battery pack insertion slot of device Please close cover after making sure whether battery pack came into end certainly Please change battery pack when LOW battery sign of LCD screen blinking for expend the length of using battery pack When change battery pack please take care not to drop battery pack Product description 7 6 Battery pack charge Please power plug of AC code to concert after connect device for charger and charger Please thread battery pack when battery pack
23. y pack 12 specifications Battery pack es Material Lithium Polymer Charge Condition e Current 1 0A e Voltage 16 8 VDC e Temperature 0 40 C Use condition e Max current 20A e Voltage range 12 min 16 8 MAX V e Temperature 20 60 C Charger Model JBL910A 1600 Input Voltage 100 250 VAC Output Voltage 16 8 VDC Current 1 0A specifications Distance from X ray focal source to X ray field and angle Please use this product by Hand Held like below picture Don t put any material that can attenuate intensity of X ray between X ray field and subject at X ray exposure 6 Product consist Portable X ray System DX3000 Components gt 2EA gt Charger TEA gt Device for charger TEA Accompanying document gt User Manual CD gt Quality Assurance Keeping case 7 Product description 7 1 Features Hemote control LOW battery indicating lamp connector E Pu EXPOSURE lamp E m un JI ee T 7 tl X ray Field Battery pack cover Skid proof grip _ Power button amp lamp qu M Mode control button Handle The design of the product is subject to change without notice to enhance the performance 7 Product description 7 1 Features CAUTION CONSULT ACCOMPANYING DOCUMENTS 7 DATE OF MANUFACTURE PRODUCT LABEL STICKER SN