image

Contents

1. SM DO M GB 12_02 2007 16 SERVICE MANUAL ASPINA DO M PIC B L bonded joints anaerobic adhesives Loctite 620 Loctite 670 603011194 COVER OF THE CONSOLE COMPLETE PIC A POS 3 PIC C 602011192 BOX COMPLETE PIC A POS 5 SM DO M GB 12_02 2007 17 SERVICE MANUAL ASPINA DO M PIC D N 14 q k Ng G Chassis for motor LUZZANI AIRMATIC 601011191 CHASSIS PIC A POS 6 PIG E a N yr XS pet PE N Chassis for motor CATTANI UNI JET 603011378 CHASSIS PIC A POS 6 SM DO M GB 12_02 2007 18 SERVICE MANUAL ASPINA DO M PIC F ra BSS FN X Ny p SS Se 24 LO A G S MN 7 V y v y ED gt s SS L j P Ol x 603011193 COMPLETE DOOR PIC A POS 4 PIC G 603021056 FAN HOLDER COMPLETE PIC C POS 1 SM DO M GB 12 02 2007 19 SERVICE MANUAL ASPINA DO M PIC H Base for motor LUZZANI AIRMATIC 602021051 BASE COMPLETE PIC D POS 3 PIC I Base for motor CATTANI UNI JET 603021156 BASE COMPLETE PIC E POS 1 SM DO M GB 12_ 02 2007 20 SERVICE MANUAL A
2. SERVICE MANUAL EKOM spol s r o Priemyseln 5031 18 921 01 PIESTANY Slovak republic tel 421 33 7967255 fax 421 33 7967223 e mail ekom ekom sk SERVICE MANUAL ASPINA DO M IMPORTANT INFORMATION i ii bb o g 3 l VARNN AI 3 11 CEManor COMI ANC use 3 LA General War O Sa A 3 1 3 General safety VATN ee 3 1 4 Electrical system safety warnings maa idilio 3 Ia TTS 4 2 PRODUGT INFORMATION ve 5 2 APNOE A eee 5 2 PM D 3 COMPONENTS a ric Sone 5 sl MTESCANICALDA TA ee 6 5 DIAGRAM OF THE ASPIRATOR 2 ps 7 6 FUNCTION il taa a e o 8 Ota DESCUIDO JR 8 e 8 ASSEMBLY see ee 8 7 STORAGE AND TRANSPORT cui a Geek 8 8 INSTALLATION AND FIRST OPERATION nn KK KK KR ARR KRK n kn 9 Sali Environmental ONCINO MS sadel 9 2 Blectmical CONNECCION uvenn 9 ier A A a a i 9 Di WIRING DAN 10 ME 11 10 PENN 11 10 1 SWITCHING on the aspi ator EG 11 11 MAINTENANCE INTERVALS USER TECHNICIAN eee eee nen nen 11 12 MAINTENANCE CLEANING AND DISINFECTING eee ne nen nen 12 121 Clean the inlet SIVO ii si lee ieee a eee b 12 12 2 Disinfect the tubing and separation VesSSel o ooooncccccooccccccoccnnononcnnnonanonconononnnnnaronnonnnrnnonanrnnnnarennns 12 12 3 Replace the output pre filter ccooocccoocncccocnnococncocononononnnonnnnononnnonannnnannnonannnonanononanenenanenes 12 12 4 Replace the output filter cooocccooccnccccnncconnnoconononononononnnononnnonnn
3. G10 Nut M4 042000002 L2 Magnet washer 062000005 G11 Screw M4x20 041000129 L3 Washer 3 2 043000015 G12 Screw M3x6 041000042 L4 Nut rivet M5 SFL RK 042000058 G13 Fan 035300001 L5 Nut M3 042000001 G14 Thermoswitch 033510004 L6 Magnet unit 029000001 G15 Holder of the fan 023000102 Picture M COVER COMPLETE Picture H BASE COMPLETE M1 Isolation washer compl 604031033 H1 Filter cover compl 603031024 M2 Console complete 603031030 H2 Gasket short 074000012 M3 Cover complete raw 603031026 H3 Gasket long 074000008 M4 Label of fuse 111000050 H4 Base enclosure 061000001 M5 Duct 073000022 H5 Washer 5 3 043000002 M6 Duct 073000021 H6 Nut rivet M5 STC R 042000081 M7 Duct 073000020 H7 Nut rivet M5 M5 SFL 042000058 M8 Gasket of the cover 062000119 H8 Screw M5x12 041000124 M9 Lower gasket 062000118 H9 Wheel fixed 029000009 M10 Washer 4 3 043000019 H10 Wheel turning 029000008 M11 Nut rivet M4 SFL RK 042000061 H11 Base 023000092 M12 Screw M6x120 041000099 M13 Screw M4x20 041000072 Picture l BASE COMPLETE M13 Screw M3x16 041000003 11 Filter cover compl 603031039 M14 Fuse 038100004 I2 Gasket short 074000012 M15 Terminal board 033100002 I3 Gasket long 074000008 M16 Terminal board 033100001 I4 Base enclosure 061000001 I5 Washer 5 3 043000002 Picture N CANNULAE HOLDER LUZZANI IG Nut rivet M5 STC R 042000081 N601 Plastic pin 062000052 I7 Nut rivet M5 M5 SFL 042000058 N602 Arm I8 Screw M5x12 041000124 N603 Screw 19 Wheel fixed 029000009 N604 Microswitch 0620000
4. ible ALA SE pil ki J Complete bracket T seek ST PEAN belt Fa us h M jek g No Y Ao EV T M HL A J J Au s Z Complete cover ight Ea on Plastic of box gt niu X1 T6 3A ais vin y Main board 1 Doors of box a I 230V 50Hz SM DO M GB 12 02 2007 10 SERVICE MANUAL ASPINA DOM USE 10 OPERATION N Users of the aspirator must be trained in its operation In case of emergencv disconnect the apparatus from the mains unplug The motor is hot Do not touch danger of burn 10 1 Switching on the aspirator Switch on the dental aspirator by turning the power supply switch 5 to position T The aspirator turns on automatically when a suction tube 6 with a mouthpiece 1 is removed from its holder 2 The aspirator runs until both tubes are replaced in the holders or until the separation vessel 11 is full With continuous use the vessel fills in 6 to 10 hours The holder for suction tips is equipped with a vacuum regulator to control the amount of suction according to the dentist s needs When the separation vessel is full sensors signal the motor 9 to pause and the indicator light 4 comes on to signal that the vessel is full At this point it is necessary to replace both tubes in the holder and switch the power supply switch off Next open the door by pulling the handles on the sides uncouple the rubber lugs from the separator s cover 10 and rem
5. Picture O CANNULAE HOLDER CATTANI SM DO M GB 12_02 2007 26
6. Section 11 Draw warm water containing a non foaming disinfectant through each suction tube 6 The disinfectant used must be in compliance with national regulations governing such use Follow the disinfectant manufacturer s instructions for use To ensure correct function of the aspirator perform the following tasks at the recommended maintenance intervals 12 3 Replace the output pre filter The output pre filter 15 must be replaced once a year see Section 11 Before replacing the filter turn the aspirator s power supply switch 5 off The pre filter and the filter 14 are located on the bottom left of the aspirator s base near the symbol With one hand turn the fixing lugs 16 90 while simultaneously pulling down Catch the filter holder 13 in the other hand Remove the filter holder with the output filter Inside the opening remove the output pre filter Also examine the output filter and replace it if necessary To reassemble set the output filter in its holder place the output pre filter onto the filter so that the side with the gummed cloth faces toward the filter and insert this assembly into the opening from below Rotate the fixing lugs back 90 so they engage fixing the filter holder in place 12 4 Replace the output filter When the equipment is used regularly the output filter 14 must be replaced once a year see Section 11 This procedure is the same as that described in Section 12 3 for the replacement of the ou
7. consumption Defective starting capacitor replace the capacitor Aspirator switchs without Defective automatic control Replace the defective part Eo MIO suction MANGE Defective microswitch in the suction from holder Replace the defective part tubing holder Aspirator is noisv Box door improperly closed Close the box door Defective bearing in the motor Replace the defective bearing Incorrectly mounted filter holder mount Me S CLO S Le CO MSC position Incorrectly positioned separation Reposition the separation vessel cover vessel cover correctly Aspirator efficiency Clogged output pre filter Replace the pre filter see Section 11 reduced Clogged output filter Replace the filter see Section 11 Leak in the suction system Check connections tighten or seal loose connections Defective control valve in the Replace the defective part suction tubing holder Clogged inlet sieve Clean the inlet sieve SM DO M GB 12_02 2007 14 SERVICE MANUAL ASPINA DO M LIST OF ENCLOSURES Enclosure No 1 FIGURES OF THE MOBILE DENTAL ASPIRATOR Pic 1 o Pic 2 13 11 15 Pic 3 Pic 4 Pic 5 SM DO M GB 12 02 2007 15 SERVICE MANUAL ASPINA DO M PIC A
8. the apparatus is plugged into an electrical socket the socket must be easily accessible for safety reasons allowing the apparatus to be safely unplugged in case of emergency e The distribution network must be protected by maximum 10 A 8 3 Initial start up After unpacking place the aspirator on the floor open the door and ensure that the separator s cover 10 with its sensing electrodes is attached to the separation vessel 11 If necessary attach the cover to the vessel and follow the instructions in Chapter 10 Close the door and connect the aspirator s power cord to the socket Turn the power supply switch 5 to position I The indicator lamp for the network 3 goes on and the aspirator is ready for use use See the caution attached to the product regarding sterilization 135 C The tips are not sterile when shipped from the manufacturer and must be sterilized before The sterilization process must be done according to law SM DO M GB 12 02 2007 9 SERVICE MANUAL ASPINA 9 WIRING DIAGRAM DO M Model 1 N PE AC230V List of Devices M Motor 230 V 50 Hz PS Control electronics EV Fan 230 V 50 60 Hz ST Temperature switch HL1 2 Indicator glow lamps Q Power switch PS LD50 2 LD50 3 M N LP T sija 25 J2 1 2 3 4 5 6 7 8 9 11 12 13 14 15 16 17 JI E EEG Cover of bracket lite
9. 20 Hose 230 A21 Hose 400 A22 Hose 700 A23 Holder with hoses A24 Stopper A25 Elbow A26 Separator bin A27 Cover of the separ mod A28 Tray A29 Covering foil A30 Washer 4 3 A31 Washer 4 A32 Nut M4 A33 Screw M4x10 A34 Screw M5x16 A35 Screw M5x45 A36 Screw M4x8 A37 Neon indicator GR A38 Neon indicator TR A39 Switch A40 Fixing lug A41 Fixing lug A42 Fixing lug A43 Separator electronics 603011194 603011193 602011192 601011191 603011378 111000092 111000042 111000032 111000031 111000030 111000029 111000028 111000025 111000004 074000016 073000022 062000121 062000112 062000111 062000110 062000090 062000056 062000053 062000039 062000038 062000034 062000017 043000019 043000003 042000002 041000110 041000087 041000065 041000006 033600008 033600007 033500001 033400011 033400005 033400001 032800001 Picture B COVER OF THE CONSOLE COMPLETE B1 Cover of the console raw B2 Label of electric drawing B3 Protective earth sticker B6 Washer 6 4 B7 Washer 5 3 B8 Nut M6 BY Screw M5x16 B10 Column B11 Pin Picture C BOX COMPLETE C1 Fan holder complete C2 Cover complete SM DO M GB 12_02 2007 604021058 111000016 111000004 043000010 043000002 042000005 041000041 024000216 024000108 603021056 602021055 C3 Case of the box complete C4 Protective earth sticker C5 Enclosure VI C6 Enclosure V C7 Enclosure Ill C8 Enclosure ll C9 Washer 4 3 C10 Washer 4 C11 Nut M
10. 30 50 60 V Hz Nominal current 520 VA Maximum flow rate 4100 I min m S 1 Dimensions of the aspirator 565x350x860 565x350x860 350 x 860 Dimensions of the aspirator with carton 950 x 560 x 595 H x W xD EE of the apirator Weight ta the aspirator with carton Classification STN EN 60 601 1 IEC 601 1 pee ess Climatic conditions for storage and transport Temperature 25 C 55 C 24 hours up to 70 C Relative air humidity 10 90 without condensation Climatic conditions for operation Temperature 5 C 40 C Relative air humidity under 70 SM DO M GB 12_02 2007 6 SERVICE MANUAL 5 DIAGRAM OF THE ASPIRATOR Suction cannulas Vacuum pressure valve Inlet sieve Separation vessel SINS SM DO M GB 12_02 2007 190 Motor Output pre filter Output filter Silencer ASPINA DO M SERVICE MANUAL ASPINA DO M 6 FUNCTION 6 1 Description When the power supply switch 5 is turned on position l the network indicator 3 goes on When the suction tube 6 is removed from its holder 2 the motor 9 comes on creating suction in the mouthpiece 1 After replacing the suction tube in the holder the motor turns off When the separation vessel 11 is filled with waste the motor turns off and the indicator light 4 goes on to signal that separation vessel is full Return the suction tube to its holder and empty the separation vessel Du
11. 4 C12 Screw M4x10 C13 Listel C14 Holder of the separat hood Picture D CHASSIS D1 Filter holder compl D2 Base board complete D3 Base complete D4 Screw M5x40 D5 Screw M5x12 D6 Outlet filter D7 Outlet pre filter D8 Fixing lug DY Spring Picture E CHASSIS E1 Base complete E2 Holder of the filter compl E3 Base board complete E4 Screw M5x40 E5 Screw M5x12 E6 Outlet filter E7 Outlet pre filter E8 Fixing lug E9 Spring Picture F COMPLETE DOOR F1 Door of the box raw F2 Protective earth sticker F4 Clamp Fo Inclosure 15 F6 Washer 3 2 F7 Washer 6 4 F8 Nut M6 F9 Nut M3 F10 Screw M3x12 F11 Magnet clip F12 Pin DO M ASPINA 602021054 111000004 061000012 061000011 061000009 061000008 043000019 043000003 042000002 041000072 034800025 023000068 603021053 602021052 602021051 041000016 041000010 025200028 025000001 604031091 022000008 603021156 603021053 602021157 041000016 041000010 025200028 025000001 604031091 022000008 604021072 111000004 062000057 061000002 043000015 043000010 042000005 042000001 041000002 029000002 024000108 Picture G HOLDER OF THE FAN COMPL G1 Duct G2 Cover of the thermoswitch G3 Cover of the fan GA Fan inelosure 5 G5 Fan enclosure 2 G6 Fan enclosure 1 G7 Washer 4 3 G8 Washer 3 2 G9 Washer 4 25 073000021 062000104 062000009 061000007 061000004 061000003 043000019 043000015 043000003 SERVICE MANUAL DO M ASPINA
12. 59 110 Wheel turning 029000008 N605 Plastic part for upper left 111 Base 023000092 membrane 062000074 N606 Valve 609000005 Picture J BASIC BOARD COMPLETE N607 Rubber membrane 074000014 J1 Basic board complete raw 603031031 N608 Plastic part for lower left J2 Silentblock M4 074000035 membrane J3 Outlet element 073000023 N609 Rubber gasket 074000015 J4 Hose mouthpiece 062000011 N610 Stopper J5 Washer 8 043000017 N611 Support J6 Nut rivet M5 SFL RK 042000058 N612 Screws J7 Screw M8x20 041000067 N613 Plastic connection 430 062000011 J8 Motor 035400001 N615 K616 K617 Filter J9 Suction pump column ll 024000264 N618 Microswitch 062000060 J10 Suction pump column I 024000263 N619 Plastic part for upper right membrane 062000073 Picture K BASIC BOARD COMPLETE N620 Plastic part for lower right K1 Basic board complete raw 603031031 membrane K2 Silentblock 074000009 N621 Main body K3 Outlet element 073000023 N622 Little cannula holder 062000047 K4 Hose mouthpiece 062000011 N623 Large cannula holder 062000048 Ko Washer 8 043000017 N624 K625 K627 Kit of suction K6 Washer 8 4 043000009 cannulas 062000042 K7 Nut rivet M5 SFL RK 042000058 N626 Adapter K8 Nut M8 042000006 N628 Screw K9 Screw M8x20 041000067 N629 External body K10 Motor 035400003 N630 Filter fastener K11 Suction pump column V 024000049 N631 Suction tube diam 15 072000035 K12 Holder of the aspirator 023000067 N632 O ring 073000012 Picture L CASE OF THE BOX COMPLETE L1 Box case raw 603031025
13. ALLATION AND FIRST OPERATION A Before initial start up remove all transport stabilizers A Initial start up must be done by a qualified professional 8 1 Environmental conditions e The area where the apparatus is installed and operated must be dry well ventilated and dust free e Install the aspirator where it is easily accessible for operation and maintenance The label must be visible e Position the aspirator on a level stable base that is capable of supporting its weight see Item 4 in the Technical data A Never stretch or kink the electrical cord or the suction tubing e To assure trouble free operation of the aspirator room temperature should never drop below 5 C nor exceed 40 C Ideal ambient temperature is between 10 C and 25 C e Most of the electrical energy used by the motor 9 is converted to heat During extended use especially with the saliva ejector the temperature inside the box rises above 40 C which automatically activates the cooling fan After the interior of the box cools to approximately 32 C the fan turns off 8 2 Electrical connection The dental aspirator model 230V is equipped with a grounded safety plug Observe all electrical regulations when connecting the appliance to the mains The network voltage and frequency must be in compliance with the data on the appliance label e f the apparatus is hard wired to the electrical supply a shutoff switch must be situated near the apparatus e lif
14. SPINA DO M PIC J Basic board for motor LUZZANI AIRMATIC 602021052 BASIC BOARD COMPLETE PIC D POS 2 PIC K Basic board for CATTANI UNI JET 602021157 BASIC BOARD COMPLETE PIC E POS 3 SM DO M GB 12_02 2007 21 ASPINA SERVICE MANUAL DO M PIC L PIC C POS 3 CASE OF THE BOX COMPLETE 602021054 PIC M PIC C POS 2 COVER COMPLETE 602021055 ej SM DO M GB 12 02 2007 SERVICE MANUAL ASPINA DO M PIC N o OSOS o o A Valid for products till serial number O 4085 3 05 023000116 CANNULAE HOLDER LUZZANI PIC A POS 23 SM DO M GB 12_02 2007 23 SERVICE MANUAL ASPINA DO M PIC O 5502 203260 210901 TO 183010 202452 203336 201034 203335 201092 191852 210820 5517 210911 201155 201156 201157 201515 5516 Valid for products from serial number O 4086 3 05 062000090 CANNULAE HOLDER CATTANI PIC A POS 23 SM DO M GB 12_02 2007 24 SERVICE MANUAL LIST OF SPARE PARTS Picture A ASPINA DO M A3 Cover of the console complete A4 Door complete A5 Box complete A6 Chassis A6 Chassis A8 Apparatus label A9 Label IV A10 Foil Il A11 Foil A12 Foil of sliding P A13 Foil of sliding L A14 Label Ill A15 Label I A16 Protective earth sticker A17 Rubber clamp A18 Duct A19 Fixing tape A
15. ician or by the maufacturer s service department Use only spare parts and accessories approved by the maufacturer Before beginning all maintenance repairs and cleaning disconnect the apparatus from the JN network unplug To maintain good hygiene and ensure that the aspirator functions correctly the following maintenance must be performed 12 1 Clean the inlet sieve Solid particles that are drawn in with air and liquid are entrapped in the inlet sieve 7 see Enclosure No 1 Fig 1 which must be cleaned at the end of each shift as described in Section 11 The inlet sieve cannot be cleaned until the moisture inside the suction tubing 6 and the sieve is removed Remove the tubing from the holder 2 and allow air to flow through it for a few seconds Turn the power supply switch 5 off position O and slightly lift the cover of the inlet sieve Pull the sieve out of its holder remove solid particles and clean the sieve Replace the sieve in its the holder and reassemble the unit If the inlet sieve contains amalgam particles empty the sieve contents into a sealed container and forward it to the collection centre 12 2 Disinfect the tubing and separation vessel To disinfect the separation vessel 11 empty it rinse with warm water and clean it mechanically with warm water and a disinfectant Disinfection of the internal suction tubing and the separation vessel should be performed daily at the end of each shift as described in
16. iene for working with contaminated material Clean the device as described in Section 12 Separate label pack and decontaminate anv contaminated parts observing all health regulations Perform any necessary repairs The manufacturer reserves the right to modify the appliance in any way that will not alter the function or the operation of the appliance SM DO M GB 12_02 2007 13 SERVICE MANUAL ASPINA DO M 16 SOLVING COMMON PROBLEMS Disconnect the appliance from the network before performing all maintenance and repairs Repairs must be performed by a qualified service technician Failure Problem and possible reasons Aspirator does not work No electrical power Check the mains voltage in the socket Restore power by replacing a fuse or via the circuit breaker Interruption in electrical system Power supply switch is off Ensure that the power supply switch is in position l and the network indicator light is on Damaged power cord Replace the defective part Loose clamp at terminal board Tighten the clamp Damaged motor wiring Defective thermal protection Replace the motor Aspirator s mains fuse is defective Check the mains fuse replace defective fuse fuses 2 x T 6 3 A are situated in the upper narrowed part of the aspirator Full separation vessel Full separation vessel vessel Empty the separation vessel contents Empty the separation vessel contents separation vessel contents High current
17. nnnnnnnnnnnonannnnnonnnonananenanenes 12 13 SHU OFF 12 DISPOSAL mataia a a dla 13 14 DISPOSALOE THE APPEANGE ii i MA v 13 SOLVING PROBLEMS tabu ii Di la A a t a katal aka 13 15 REPAIR SERVICE 0 ios 13 16 SOLVING COMMON PROBLEMS usa e 14 LIST OF ENCLOSURE ia a e ani 15 FIGURES OF THE MOBILE DENTAL ASPIRAT OR vaere 15 OP INA OI inc ia Bs eS Sisco tect d A AN one 16 COVER OF THE CONSOLE COMPLETE 445 17 BOXCOMPEE NE NN 17 Gp EP AT 17 EM DN 17 COMPLETE PR 17 FAN OLDER GONE PE NNN 17 BASE COMPETE A i TA ia 17 BASE COMPLETE NN 17 OA A et 17 DAI BOARD C ONPE TE S a a L ta taa ae 17 CASE OFTHE BOX COMPLETE Luse 17 GOVER COMPLETE aiian e a o oso 17 GANNULAE HOLDER LUZZAN ve 17 GANNULAE HOLDER GATTAN be 17 HSTOE SPARE PARIS eaaa aE Sie lie bob a bono oo k o Doo o do 17 SM DO M GB 12 02 2007 2 SERVICE MANUAL ASPINA DO M IMPORTANT INFORMATION 1 WARNINGS 1 1 CE Mark of compliance Products marked with the CE mark of compliance meet the safety guidelines of the European Union 93 42 EEC 1 2 General warnings e The Installation Operation and Maintenance Manual is an integral part of the appliance and must be kept with the aspirator Careful review of this manual will provide the information necessary for correct operation of the appliance e The safety of operating personnel and trouble free operation of the appliance are ensured only if original parts are used Only accessories and parts specified in the technical d
18. ocumentation or expressly approved by the manufacturer can be used e lf any other accessories or consumable materials are used the manufacturer cannot be held responsible for the safe operation and functionality of the device e The manufacturer s warranty does not cover damages originating from the use of accessories or consumable material other than those specified or recommended by the manufacturer e The manufacturer guarantees the safety reliability and function of the appliance only if installation calibration amendments extensions and repairs are performed by the manufacturer or its representative or a service provider authorized by the manufacturer the appliance is used in accordance with the Installation Operation and Maintenance Manual e The Installation Operation and Maintenance Manual describes the appliance and all relevant safety and technical standards for its use All wiring schemes methods and names in the manual are copyrighted and may not be reproduced without permission from the manufacturer 1 3 General safety warnings The appliance is designed to operate safely when used correctly Please note the following safety measures to avoid damage and risk e Operation of the appliance must be in compliance with all local codes and regulations e The original packaging should be kept for the possible return of the appliance Only the original packaging ensures optimal protection of the appliance during transport If i
19. ove the vessel Hang the cover on the holder on the left side 8 Drain the contents of the separation vessel into the sewer rinse the vessel with water and refasten the separator s cover Replace the vessel in its holder 12 in the aspirator box the holder is detachable Ensure that the cover is tight and the vessel is seated in its holder and close the door The suction mouthpieces are not sterile when shipped from the manufacturer Replace the used mouthpiece after each patient Tips that are designed for multiple must be sterilized after each patient The cannulas must be replaced after a maximum of 400 cvcles of sterilization or after one vear Whichever comes first Never cover the air outlets on the upper sides of the unit The aspirator does not contain a backup power suppiv Before turning the unit on the door of the box must be closed gt pbe b 11 MAINTENANCE INTERVALS USER TECHNICIAN Replacef output filter 0 output Replacef output filter 0 124 4 ser m a maximum of 400 Replace suction cannula cycles of sterilization or user once a year Regular inspection inspection every 2 years service technician service technician Note the suction end pieces are also at this time whichever occurs first SM DO M GB 12 02 2007 11 SERVICE MANUAL ASPINA DO M 12 MAINTENANCE CLEANING AND DISINFECTING Repairs that go beyond general maintenance must be performed by a qualified techn
20. ring extended use of the saliva ejector the temperature inside the box may increase which automatically activates the cooling fan The fan turns off automatically when the temperature inside the box falls 6 2 Vacuum function Suction in the tubing system connected to the mouthpiece 1 draws waste from the mouth through the inlet sieve 7 where solids are trapped Air and liquid free of solids larger than 2 mm flow into the separation vessel 11 which holds the liquid waste The air flows through the motor 9 the silencer the output pre filter 15 and the bacteria trapping output filter 14 Clean air free of impurities is blown out through an opening underneath the aspirator ASSEMBLY 7 STORAGE AND TRANSPORT The dental aspirator is delivered in a transport carton that protects the apparatus from damage during transport For transport always use the original product packaging if possible Transport the dental aspirator in the upright position Protect the aspirator from moisture dust and extreme temperatures The aspirator in its original packaging can be stored in a warm dry and dust free area dispose of the packing material in an environmentally friendly way Cardboard can be recycled Always empty the separation vessel before transporting the aspirator B Keep the packing material for possible future return of the unit If this is not possible SM DO M GB 12_02 2007 8 SERVICE MANUAL ASPINA DO M 8 INST
21. sed for important information Information instructions and cautions for the prevention of damage to health and equipment Caution Dangerous electrical voltage Information regarding correct use of the appliance and other warnings APP CE mark of compliance AN mn Caution Hot surface Handling mark on packaging Fragile handle with care Handling mark on packaging This way up vertical position of freight rap Handling mark on packaging Protect against moisture danos d if Handling mark on packaging Temperature during storage and transport Handling mark on packaging Limited stacking Mark on packaging Recyclable material Earth ground connection Fuse Electric appliance Tvpe B Alternating current Biological hazard R gt gt Bj SM DO M GB 12 02 2007 4 SERVICE MANUAL ASPINA DO M 2 PRODUCT INFORMATION 2 1 Application The portable dental aspirator ASPINA DO M gives the user great flexibility in location and placement and allows for the design of an ergonomic workspace It is ideal for dental units that are not equipped with a suction apparatus and separator The appliance aspirates separates and entraps waste in its built in separation vessel lts mobility makes it easy to move between workstations and it can be used in areas that lack access to a sewer system e The dental aspirator is designed for operation in a dry ventilated room with an ambient temperat
22. t becomes necessary to return the appliance during the warranty period the manufacturer is not liable for damages caused by improper packaging e Each time the appliance is used the operator must ensure that the appliance is functioning safely and correctly e The user must fully understand the operation of the appliance e lf any problem occurs during use of the appliance the user must inform his supplier immediately e Do not use the appliance in the presence of explosive gases or flammable liquids 1 4 Electrical svstem safetv warnings e The appliance must be connected to earth grounded e Before the appliance is plugged in ensure that the voltage and frequency specified on the apparatus are the same as the power supply e Before using check for possible damage to the appliance and the air connections Damaged parts or wires must be replaced immediately e In case of emergency immediately disconnect the appliance from the power supply pull out the plug e When performing all repairs and maintenance ensure that the mains plug is removed from the power socket pressure pipes must be air vented pressure must be released from the air tank e The appliance must be installed by an approved qualified technician SM DO M GB 12_02 2007 3 SERVICE MANUAL ASPINA DO M 1 5 Cautions and symbols In the Installation Operation and Maintenance Manual and on the appliance and its packaging the following labels or symbols are u
23. tput pre filter 15 13 SHUT OFF If the dental aspirator is not to be used for a long period of time clean and disinfect it as described in Sections 12 1 and 12 2 Turn the power supply switch 5 on position l remove the suction tubes 6 from the holder 2 and let air flow through them for 15 20 minutes or until the suction system is fully dried out Replace the tubes onto the holder and turn off the power supply switch position O Disconnect the apparatus from the network unplug and remove the mouthpieces 1 from the tubing SM DO M GB 12_02 2007 12 SERVICE MANUAL ASPINA DO M DISPOSAL 14 DISPOSAL OF THE APPLIANCE Disconnect the appliance from the mains Follow all rules of hygiene for working with contaminated material Clean the device as described in Section 12 Separate label pack and decontaminate any contaminated parts observing all health regulations Dispose of the unit as prescribed by law A Inner components of the unit may be contaminated with biological material Ensure that AN the unit is fully decontaminated before disposing of it SOLVING PROBLEMS 15 REPAIR SERVICE Guarantee and post guarantee repairs must be performed by the manufacturer its authorized representative or service personnel approved by the supplier When servicing components of the device that might be contaminated follow the instructions below Disconnect the appliance from the mains Follow all rules of hvg
24. ure between 5 C and 40 C and relative humidity less than 70 e The dental aspirator must not be exposed to moisture and cannot be operated in an excessively humid or wet environment Do not use the unit near gases or combustible liquids e lf the appliance is used in any manner or for any purpose not described in this manual the manufacturer is not responsible for damage that may result Such risk will be assumed exclusively by the operator user 2 2 Product description The mobile dental aspirator ASPINA DO M is enclosed in a soundproof box mounted on wheels Inside the box is a fan cooled motor 9 electrical wiring and a waste trapping separation vessel 11 Under the box is a silencer with an output filter 14 and a pre filter 15 that filters the air from the motor The upper part of the aspirator contains a holder for suction tubes 2 equipped with mouthpieces 1 automatic separator and fuse box On the side is the mains switch 5 and illuminated indicators for the network 3 and the fill level of the separation vessel 4 see Enclosure 1 Fig 1 3 COMPONENTS e Dental aspirator DO M e Installation Operation and Maintenance Manual NP DO M e Output pre filter 2 pcs No 025000001 e Fuse T6 3A 2 pcs No 038100004 e Antifoam tablets 2 pcs No 099000003 e Setof tips 1 pc No 604011371 e Saliva ejector 1 pc No 062000241 SM DO M GB 12_02 2007 5 SERVICE MANUAL ASPINA DO M 4 TECHNICAL DATA Voltage frequency 2

Download Pdf Manuals