Home

Frame Tracer HFR-8000 Huvitz

image

Contents

1. e DCS Parameters Machine Id 00 99 TIMEOUT Timeout interval for communication 10 90 seconds DATA PT Data points for DCS output FORMAT Data format for DCS output ASCII ABS HFR 8000 Operator s Manual for Remote Tracing 27 ASCII ABS ASCII Absolute BINABS BIN ABS Binary Absolute BIN DIF BIN DIF Binary Differential PACKED BIN PACKED BIN Packed Binary ANGLE Angle type for DCS output EVEN UNEVEN FR SIZE Size offset of frame data for DCS l l 9 99 9 99 output in terms of circumference PT SIZE Size offset of pattern data for DCS l l 9 99 9 99 output in terms of circumference MIRROR Mirroring option for DCS output when YES single side is traced NO FRAME Default frame material for a job for DCS output METAL METAL Metal PL HARD PL HARD Hard Plastic PL SOFT PL SOFT Soft Plastic UNDEFIN UNDEFIN Undefined OFF OFF Do not output frame material info ZTILT Unit of ZTILT value DEGREE RADIAN 7 4 Using the Barcode Interface for Inputting Job Number HFR 8000 supports barcode scanning function through the BARCODE port by setting the BARCODE option to ON This allows it to assign a job number for a new tracing job Go to the CONFIG menu and set the options like below Options for the barcode interface BARCODE ON BPS BC 9600 depends on the setting of a barcode scanner 28 HFR 8000 Operator s Manual f
2. 6 Menu Menu Selection CONFIG System Configuration lt Refer to the section 6 1 Configuration of Frame Reader CALIB Automatic Calibration Refer to the section 7 1 Automatic Calibration of Frame Reader ETC User Preferences lt Refer to the section 6 2 User Preferences of Frame Reader 6 Frame Material Selection of Frame Material METAL PL HARD Hard Plastic PL SOFT Soft or flexible Plastic Before tracing a frame choose one of the material types that is used by the edger for such as predefined size adjustment Selection of Pattern Tracing Mode LO CURV 2D Tracing e HI CURV 3D Tracing When the Pattern Holder is loaded in the Frame Reader it automatically recognizes the pattern tracing mode When tracing a demo lens of higher curve using the HI CURV mode could give more precision in the size HFR 8000 Operator s Manual for Remote Tracing 19 0 L BOTH R STOP Tracing Buttons Stop Button BOTH Both sides L Left only R Right only It stops any ongoing operation or cancels the execution of a menu function When it s for a frame holding down one of the The system will halt when the tracing buttons for 2 seconds will executes the D stop button is pressed during SEMI AUTO tracing mode which is to aid tracing a o its initialization process frame with its frame groove lopsided by allowing the operator position the
3. potential hazard The classifications and symbol s are shown below Save these instructions and O on power switch represent ON and OFF respectively This symbol identifies a safety note Ensure you understand the function of this control before using it Control function is described in the appropriate User s or Service Manual Identifies the point where the system safety ground is fastened to the chassis Protective earth connected to conductive parts of Class equipment for safety purposes Disposal of your old appliance 1 When this crossed out wheeled bin symbol is attached to a product it means the product is covered by the European Directive 2002 96 EC 2 All electrical and electronic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities 3 The correct disposal of your old appliance will help prevent potential negative consequences for the environment and human health 4 For more detailed information about disposal of your old appliance please contact your city office waste disposal service or the shop where you purchased the product 8 HFR 8000 Operator s Manual for Remote Tracing 2 3 Environment factors Avoid the following environments for operation or storage Where the machine is exposed to water vapor Don t operate the machine with wet hands Indoor use only Where the machine is
4. HFR 8000 Operator s Manual for Remote Tracing 33 8 2 Cleaning D Sweep the outer surface with soft fabric material 2 When you do not use this device wrap it with dust cover Before cleaning your product disconnect the power cord e Since the exterior of the product is easily scratched be sure to use soft cleaning cloth When cleaning the product avoid spraying cleansing agent directly onto the product This may result in the exterior being deformed or the print being removed Since using a strong chemicals may result in the discoloration or cracking of the product exterior be sure to use the clean water only 34 HFR 8000 Operator s Manual for Remote Tracing Standard Accessory Standard Frame for Frame Reader Calibration E E Standard Pattern for Frame Reader Calibration Zim aia am Rial a MUR a mia RialR A Rin RUN A TRLR ee B B CR CR RIS RUM NR A C8 Pattern Holder 253i n a ninm dd EE a RIA ete esa sav cel RR RR B E RR B E S B MPa S M a UR NUR a RC RcR CR RR R A RN Fuse 3 15A a E E Power Cable Frame Reader A NUR RON E A BOR RC RO Te Cee CR RR RR M ee ee a Option Barcode Scanner A AAAA E ee eee A AA EE ee ee eC AARAA EAA EAE 9 pin D Sub Serial Cable Crossed AEA eee ee E ee eee oe ee ee ee ee Lens Adapter wate anatase Ra NUS Died a ene eee Wao ER RE EUR NIC wee GST a eS RORIS 7A Nw Wh ee Dae See RS Specification Frame Reader Dimension 284 W x 319 D x 19
5. as PROTO 1 EDGER 2 port BPS 2 Baud rate for EDGER 2 port Same as BPS 1 BARCODE Barcode Reader port ON OFF BPS BC Baud rate for BARCODE port Same as BPS 1 HFR 8000 Operator s Manual for Remote Tracing 23 6 2 Setting User Preferences Refer to 6 1 for operating the Menu Mode B Execute ETC 1 IDINFO User Preference Parameters ID INFO Product identification VERSION Software version number _ as DBL Bridge size input option for tracing YES A SEMI AT Prompt option SEMI AUTO tracing YES Mmm we Se PS MINI FR Low speed tracing option for frames YES mar with minimal vertical size FPD OFF FPD offset 2 99 2 99 JOB NO Job number reset option Turning on the option keep the last job number even after turning off the power and the last job number is continued next time it s turned on TRANSMT Data transmission option for when both the edger 1 and 2 ports are turned on AUTO option transmits automatically USER option transmits through user selecting output port EXIT Exits the menu mode 24 HFR 8000 Operator s Manual for Remote Tracing 7 Using the Remote Tracing Function 7 1 Interface Specification Version Information Compatible with DCS V3 07 or its compliances RS232C Interface The basic specification of the communication port of the Frame Reader for interfacing PC follows the information below Baud Rate 115200 57600 38400 19200 1
6. exposed to direct sunlight Where there are big changes in temperature Optimal temperature range for normal operation is from 10 C to 40 C Humidity 30 80 Where there is a hot equipment nearby Where the humidity is extremely high or there is a ventilation problem Where the machine is exposed to excessive shocks or vibrations Where the machine is exposed to chemical material or explosive gas HFR 8000 Operator s Manual for Remote Tracing 9 Be cautious so that things like dust and metal do not fall inside the machine Don t disassemble or open the product HUVITZ does not take responsibility for the possible problems Be careful not to block the fan located on the backside of the machine Don t plug the AC power cord into the outlet unless all parts of the machine are completely connected Otherwise it will cause severe damage on the machine Pull out the power cord with holding the plug not the cord To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth This instrument must be followed by these following conditions 1 Operation An ambient temperature range of 10 C 40 C 50 F 104 F A relative humidity range of 30 80 An atmospheric pressure range of 800 1060hpa An indoor use Altitude up to 2000 m 2 Transportation An ambient temperature range of 25C 70C 13 F 158 F A relative humidity range of 10
7. in humid rooms Humidity should be maintained between 50 and 80 for normal operation Do not expose the device to water splashes dripping water or sprayed water Do not place containers containing fluids liquids or gases on top of any electrical equipment or devices 3 The equipment must be operated only by or under direct supervision of a properly trained and qualified person 4 The service technician of Huvitz or the technician authorized by Huvitz may only carry out modifications of this equipment Customer maintenance of this equipment may only be performed as stated in the User s Manual and Service Manual Any additional maintenance may only be performed by the service technician of Huvitz or the technician authorized by Huvitz The manufacturer is only responsible for effects on safety reliability and performance of this equipment when the following requirements are fulfilled 1 The electrical installation in the respective room corresponds to the specifications 12 HFR 8000 Operator s Manual for Remote Tracing 13 stated in this manual and 2 This equipment is used operated and maintained according to this manual and Service Manual The manufacturer is not liable for damage caused by unauthorized tampering with the device s Such tampering will forfeit any right to claim under warranty This equipment may only be used together with accessories supplied by Huvitz If the customer makes use of ot
8. stylus tip in the frame groove center When it s for a demo lens holding down one of the tracing buttons for 2 seconds will executes the SEMI AUTO tracing mode which is to allow recognizing concave shape while tracing Otherwise it doesn t recognize concave shape and reform it to make convex shape in order to avoid unintended dents on the final lens This is because the final lens could be uneven on the edge if it recognizes an unintended notch which is used for pulling out the wire of a half rim frame 20 HFR 8000 Operator s Manual for Remote Tracing 5 2 Tracing Frame Tracing Place the upper rim of a frame between the grippers on the rear gripping bar first and then the lower rim between the grippers on the front gripping bar 2 Move the frame so that the middle of it is aligned with the center mark on the gripping bar 3 Select the frame material by the material selection button 4 Press one of the tracing buttons BOTH for both sides R for right or L for left Pattern Pattern Tracing Demo Lens Put a pattern or a blocked demo lens in the pattern holder The demo lens should be blocked on Pattern the horizontal axis Lens Adapter Fixing Screw 2 Press the FRAME button to select a tracing mode between LO CURV and HI CURV 3 Press one of the tracing buttons R for right or L for left Loading the Pattern Holder HFR 8000 Operator s Manual for Remote Tracing 21 6 U
9. 28 M Introd ctioN er DD SALCCY Symbols dia 2 9 Environment LACCOTS sivsiecsesisccducsecsccceseccavscdectecccsdscadassessscne 2 4 Salety Precautions 4er oces es eec tt veu H epe UN Nee aeo ec eeu Eee a E Ven UR System CORfIP UTILIOR NH RR EN RES S Frame Reader asador E 9E CE C00 2090023 35 2 Standard AGCOSSOEY inca Tene gU OPA 5 1 Command Buttons ssssiccsasssisssaissincscssssasdatsscacsaassisssatessctaasss 5 2 A sensven vacsececoverssesoveaeseesvecsuesees US the Me eli ieee hase eT 6 1 Configuring the Frame Reader 6 2 Setting User Preferences cis HFR 8000 Operator s Manual for Remote Tracing 3 7 Using the Remote Tracing Function eee ee eee ee eese ee eee eee ee eene nnn nun 24 7 1 Interface Specification ooooeree rene e esae etes a ripae aaa ne ena haa erasa aant ooa 24 Tia Connechines toa DCS DOS 2 0 praeco haa Peu Ea nario rea aiaa ais 25 Ti DC S ODUOFHS sucses serit Ee Uepi cuve vv esI do Tee VER S PEU EET Pp suu Vk E E o No e FECE UPPER EUR da 26 7 4 Using the Barcode Interface for Inputting Job Number 27 8 Systent Mamtenulice 2 m octetie A O 29 8 1 Automatic Calibration of Frame Reader 29 3 2 C Id Ig eee tr REOR DENITRO ERI 33 O A 34 0 A v M 34 ooo s vo EM 35 4 HFR 8000 Operator s Manual for Remote Tracing 1 Introduction 1 1 Syste
10. 3 H mm Tracing System 3D Binocular Weight 8 kg Power Supply AC 100 240V 50 60Hz Power Consumption 32W Maximum HFR 8000 Operator s Manual for Remote Tracing 35 System Dimension 284 319
11. 4400 or 9600 Data Bit 8 e Stop Bit 1 Parity None Flow control None Data Format Format ASCII Absolute Binary Absolute Binary Differential Packed Binary Data Points 1440 1200 1000 800 500 400 Radius mode Even Uneven Side Both Right Left e Tracing Type Frame Pattern HFR 8000 Operator s Manual for Remote Tracing 25 7 2 Connecting to a DCS host Before connecting the Frame Reader to any PC which is running a lab management software or a host program on it make it sure to turn the instrument off Procedure D Connect PC to EDGER 1 port with a 9 pin DSUB crossed serial cable 2 Connect a barcode reader to the BARCODE port 3 Turn on the Frame Reader 4 Set the DCS options for the DCS host Cable connection CONSOLE BARCODE EDGER 2 EDGER 1 F d dl ED 13204854 NOTE BARCODE port delivers power through it 26 HFR 8000 Operator s Manual for Remote Tracing 7 3 DCS Options To access the DCS options in the menu mode at least one of the EDGER 1 and EDGER 2 port should be set as DCS Procedure for Menu Execution Press MENU button and select the CONFIG menu Refer to the section 6 1 for menu operation 2 Set at least one of EDGER 1 and EDGER 2 port like bellows EDGER 1 2 PC PROTO 1 2 DCS or DCS Z 3 Change the menu category to DCS DCS menu category is activated only if one of the edger ports is set as DCS B Execute DCS
12. 95 with non condensing An atmospheric pressure range of 500 1060hpa 10 HFR 8000 Operator s Manual for Remote Tracing 3 Storage An ambient temperature range of 25 C 70C 13 F 158 F A relative humidity range of 10 95 with non condensing An atmospheric pressure range of 700 1060hpa 4 Temporary Overvoltage Short term temporary overvoltage 1440V Long term temporary overvoltage 490V Please avoid where the equipment is exposed to excessive shocks or vibration Operation time Max 6 minutes for short term operation Rest time Min 2 minutes HFR 8000 Operator s Manual for Remote Tracing 11 2 4 Safety Precautions This equipment has been developed and tested according to safety standards as well as national and international standards This guarantees a very high degree of safety for this device The legislator expects us inform the user expressively about the safety aspects in dealing with the device The correct handling of this equipment is imperative for its safe operation Therefore please read carefully all instructions before switching on this device For more detailed information please contact our Customer Service Department or one of our authorized representatives This equipment must not be used a in an area that is in danger of explosions and b in the presence of flammable explosive or volatile solvent such as alcohol benzene or similar chemicals Do not put or use this device
13. Operator s Manual HFR 8000 For Remote Tracing Pacing Progress toward People HFR 8000 Operator s Manual for Remote Tracing 1 IMPORTANT NOTICE This product may malfunction due to electromagnetic waves caused by portable personal telephones transceivers radio controlled toys etc Be sure to avoid having the above objects which affect the normal operation of the product brought near the product The information in this publication has been carefully checked and is believed to be entirely accurate at the time of publication HUVITZ assumes no responsibility however for possible errors or omissions or for any consequences resulting from the use of the information contained herein HUVITZ reserves the right to make changes in its products or product specifications at any time without prior notice and is not required to update this documentation to reflect such changes 2012 2014 HUVITZ Co Ltd 689 3 Geumjeong dong Gunpo si Gyeonggi do 435 862 Republic of Korea All rights are reserved Under copyright laws this manual may not be copied in whole or in part without the prior written consent of HUVITZ Co Ltd 2 IMPORTANT NOTICE 1 Introduction 4 Installation Procedure 5 HFR 8000 Operator s Manual for Remote Tracing Contents 1 1 System Configurration cccccccccccsssssssscccssssssssssscccsscsees 1 2 Main Features m AAN aerias r a E sod CLV Information uu ooo e E ies
14. e instructions on this manual Do not disassemble the product without the proper training It may cause the electric shock serious hurt during the operation or cause the malfunction of the product The product should be properly installed and maintained at the horizontally leveled place Otherwise it may affect the normal operation of the product Do not force on the stylus of frame reader It may cause the malfunction of the product Be sure that the stylus of frame reader is not interfered with the human body or other objects during the operation Otherwise it may cause the hurt or malfunction of the product 14 HFR 8000 Operator s Manual for Remote Tracing When moving the equipment first fix the stage and check whether the power supply is off In case of setting down the equipment make sure not to be interfered with the obstacles Setting down the product slowly in order to prevent the hurt of human body or damage to the product e When moving the equipment make sure the locking device and the screw for the cover are properly tightened In case of packaging the equipment use the recommended packaging material and method in order to prevent the damage during the transportation Be sure to use the standard accessories or tools provided together with the product for the maintenance Otherwise it may cause the malfunction of the product Be sure to use the standard accessories or t
15. her accessories use them only if their safe usability under technical aspect has been proved and confirmed by Huvitz or the manufacturer of the accessory Only person who has undergone proper training and instructions are authorized to install use operate and maintain this equipment Keep accompanied manual s or technical materials such as the User s Manual and Service Manual in a place easily accessible at all times for persons operating and maintaining the equipment Do not force cable connections If a cable does not connect easily be sure that the connector plug is appropriate for the receptacle socket If you cause any damage to a cable connector s or receptacle s let the damage s be repaired by an authorized service technician Please do not pull on any cable Always hold on to the plug when disconnecting cables Before every operation visually check the equipment for exterior mechanical damage s and for proper function Do not cover any ventilation grids or slits Immediately turn off and unplug any equipment that gives off smoke sparks strange noises or odors HFR 8000 Operator s Manual for Remote Tracing 13 5 Risk of explosion if battery is replaced by an incorrect type Dispose of used batteries according to the instructions 8 This machine might be impaired if used in a manner not specified by manufacturer and this manual The equipment should be properly installed and operated based on th
16. ic calibration for frame reader consists of three parts Stroke Frame Pattern In case of any change happened in the stylus do the full calibration in the following sequence Stroke calibration Frame calibration e Pattern calibration Otherwise just perform frame or pattern calibration correspondingly Frame Calibration Purpose Detecting the frame reading position axis and calibrating the size Procedure D Load the standard frame which is provided as a standard accessory 2 Press MENU button and select the CALIB gt gt FRAME menu Refer to the section 6 1 for menu operation 3 Press the BOTH button two times to trigger the automatic calibration gt It traces the standard frame on both sides twice 30 HFR 8000 Operator s Manual for Remote Tracing 4 If it result in any error check if the standard frame is loaded in correct position and try again gt If it fails again contact proper service engineer for maintenance Preparation for the FRAME calibration Standard Frame B Execute CALIB FRAME Calibration results are stored in its permanent memory automatically right after the completion of each calibration Be sure to check the related information before executing the calibration work because you cannot get back the machine to the previous status When a calibration seems not to be done properly try executing the calibration again During the calibration you may press the STOP butto
17. m Configuration Frame Reader HFR 8000 DCS Host PC 1 2 Main Features Binocular 3 dimensional tracing of frames guaranteeing more precision in frame fitting Tracing of a demo lens or a pattern executes in two different modes HI and LO Especially the HI mode compensates the size by the tilting angle of the demo lens or the pattern SEMI AUTO frame tracing mode is additionally provided for manual handling of th e stylus position into the frame groove especially when the groove is lopsided SEMI AUTO demo lens tracing mode is additionally provided for recognizing and processing concave shape Barcode scanner interface is supported by default to provide a means to input job number Variety of DCS communication options are provided to customize it for being com pliant with different types of Lab Management Software such as data format points of data angle type communication speed size offset and more Easy to use sliding door for protecting the frame reader from dusts being introduced inside of it Powerful cooling fan for stabilizing the performance of the frame reader especially HFR 8000 Operator s Manual for Remote Tracing 5 in an environment with continuous operation over a long period of time 1 3 Classification Protection against electric shock Class l earthed Installation Category II Pollution Degree 2 6 HFR 8000 Operator s Manual for Remote Tracing 2 Safety Information 2 1 Intr
18. n to abort the execution HFR 8000 Operator s Manual for Remote Tracing 31 Pattern Calibration Purpose Detecting the pattern reading position axis and calibrating the size Procedure D Load the pattern holder with the standard pattern attached 2 Press MENU button and select the CALIB gt gt PATTERN menu gt Refer to the section 6 1 for menu operation 3 Press the BOTH button two times to trigger the automatic calibration gt It traces the standard pattern twice 4 If it result in any error check if the standard pattern is loaded in correct position and try again gt If it fails again contact proper service engineer for maintenance Preparation for the PATTERN calibration Pattern Holder Standard Pattern PATTERN calibration menu 32 HFR 8000 Operator s Manual for Remote Tracing Stroke Calibration Purpose Detecting the maximum traceable dimension for frames and patterns Procedure D Load the standard frame which is provided as a standard accessory 2 Press MENU button and select the CALIB gt gt STROKE menu Refer to the section 6 1 for menu operation 3 Press the BOTH button two times to trigger the automatic calibration 4 If it result in any error check the movement of the stylus in every direction and try again gt If it fails again contact proper service engineer for maintenance Preparation for the STROKE calibration Standard Frame
19. oduction Safety is everyone s responsibility The safe use of this equipment is largely dependent upon the installer user operator and maintainer It is imperative that personnel study and become familiar with this entire manual before attempting to install use clean service or adjust this equipment and any associated accessories It is paramount that the instructions contained in this manual are fully under stood and followed to enhance safety to the patient and the user operator It is for this reason that the following safety notices have been placed appropriately within the text of this manual to highlight safety related information or information requiring special emphasis All user operators and maintainers must be familiar with and pay particular attention to all warning and cautions incorporated herein Warning indicates the presence of a hazard that could result in severe personal injury death or substantial property damage if ignored Note describes information for the installation operation or maintenance of which is important but hazard related if ignored Caution indicates the presence of a hazard that could result in minor injury or property damaged if ignored HFR 8000 Operator s Manual for Remote Tracing 7 2 2 Safety Symbols The International Electro technical Commission IEC has established a set of symbols for medical electronic equipment which classify a connection or warn of any
20. ools provided together with the product for the maintenance Otherwise it may cause the malfunction of the product HFR 8000 Operator s Manual for Remote Tracing 15 3 System Configuration 3 1 Frame Reader Front View Protection Door Work Space for Control Panel Stylus Display Window Gripper Gripping Bar Rear View CONSOLE BARCODE EDGER 2 EDGER 1 Main Switch Power Inlet amp Interface Terminal Fuse Holder EDGER 1 EDGER 2 Barcode Reader S W Upgrade 16 HFR 8000 Operator s Manual for Remote Tracing 3 2 Standard Accessory P dis D gt eo X a Y AR L y ww w Standard Frame Standard Pattern Pattern Holder S HFR 8000 Operator s Manual for Remote Tracing 17 4 Installation Procedure Install a Frame Reader in a sequence below 1 Remove the shock absorbing material from the package box and pull out the Frame Reader carefully 2 Remove the shock absorbing material from the workspace of Stylus 3 Plug the Power Cable into the power inlet at the back of the product and turn on the power to check the initial operation 4 If the initial operation works properly turn off the instrument 5 Connect Edger and Frame Reader by using a 9 pin D sub Crossed Interface Cable and make sure the connectors are fixed firmly by tightening the scre ws in the cable connector 18 HFR 8000 Operator s Manual for Remote Tracing 5 Operation 5 1 Command Buttons
21. or Remote Tracing e Setup for Barcode Scanner Baud rate 9600 or what else bps that can be matched with the BPS BC option in the Frame Reader Termination code CR lt LF gt or lt CR gt lt LF gt Inputting sequence of Job Number with a barcode scanner Press a button to execute pe Scanner ready frame or pattern tracing 1 METAL Scan job no Tracer beeps and prompts to scan a barcode STOP Cancel Job no scanned Scan a barcode and it ll display the number for a gt 123456789012 second on the screen Frame tracing is automatically Scanning right AUTO1 METAL started after the job number input The job number is transmitted Transmit Edger1 with the shape data through j PAUTO METAL the EDGER 1 port The job number scanned by the barcode reader is recognized by the target device which is connected to the EDGER 1 port with the protocol DCS or DCS Z HFR 8000 Operator s Manual for Remote Tracing 29 8 System Maintenance 8 1 Automatic Calibration of Frame Reader The Frame Reader requires calibration in following cases e Traced shape is off axis e Traced shape has smaller or bigger size Stylus tip is not locating the frame groove properly Frame Reader issues height or dimension error for frames or patterns especially when it seems that the mechanical stroke is enough for a frame or a pattern Automat
22. sing the Menu 6 1 Configuring the Frame Reader Operating the Menu Mode Scanner ready 1 METAL Press the MENU button and then BOTH button to enter the menu mode Press the STOP button over 2 seconds to exit the menu mode Parameter Name B Execute Menu Category AAA CONFIG BE Name E E MENU button changing the selection of menu category gt Holding down the button for 2 seconds reverses the selection FRAME button selecting a configuration parameter in a menu category gt Holding down the button for 2 seconds reverse the selection BOTH button executing a selected parameter for change Parameter Name Executed Parameter Name R L button changing the value of the executed parameter BOTH button saving the change for the executed parameter and return to the menu STOP button canceling the execution and return to the menu 22 HFR 8000 Operator s Manual for Remote Tracing Configuration Parameters BEEP Buzzer activation ON pee NI FRAME Default frame material METAL PL HARD PL SOFT EDGER 1 Target device for EDGER 1 port i mE Selecting NONE disable the port PROTO 1 Data communication protocol for HUVITZ EDGER 1 port DCS mE DCS_Z BPS 1 Baud rate for EDGER 1 port 9600 14400 19200 57600 115200 38400 EDGER 2 Target device for EDGER 2 port Selecting NONE disable the port PROTO 2 Data communication protocol for Same

Download Pdf Manuals

image

Related Search

Related Contents

  Manual de Instrucciones Shearwater Predator  Pwn Plug Release 1.1 User Manual  HD OMNI Troubleshooting Read Me  Istruzioni d`uso / Istruzioni d`installazione  NEC Express5800/320Lb Linux Getting Started Guide  取扱説明書 [衛星可搬端末 01] - 【AKTIO】アクティオエンジニアリング  Philips AZ6812 User's Manual  Viti a rulli  manuel d`utilisation et d`entretien  