Home
91xx Service Manual - Frank`s Hospital Workshop
Contents
1. m ano 0 an9a Im Jo 6 ag 6661 0 Aew Aepuow exeg 9 610842 x e Jaquinn Juswnoog azg 09 1816 0816 OLLVINSHOS 390v 1 2 namay AUT XTIHLVWIXONddV GST T Z S vas Q 6012191 ely 2 E 1 EOFLON VAs p ns 1005801 zanese aos aus o nooso44na x NS ov 135280 ho v anas c vowso H naan 0 G en Ee im SN 1 0 Nn 001 snivisav gt 50015 ini 4 094 994 ar ze igo emia Xo uoge i auos T 9 z13sav 8 1 13sqv 1n043H 989 989 SS V o1asaVv 9 62 Phi L HNI 68 dvo43u vas o 33 88 189 2 x o Sivadv 2 Nas 3ni Mash amp Ssivdaav m NIJ38 x TIVdOv x eg X zg OTNI son x n L 8 IHNI x ox olvaav SE z t e gm 5 G 2 90 ES 69H 198 ho T ho 9 1 1 59 9 nom WHINHSANOO G V HWOLONTUIS ISNNWHD April 2007 Version 2 0 Service manual SBC PCB analog selector and A
2. o an9a gt gt Nadwna 0310 lk Ley 99 C Ol ecu 99 anos L L ges ses anos anos angs angs T ans ho 38621 ano SA 07 voornt Wr S OF vid era zia Sans SSK vasnas vino St easnas ana St et sino T asns 4 ano 5 VSSN3S k k uno INI em WA N os s za o 5 5 5 anos anos S S S SA hs hs anro 8 4 00 4 990 anyo Svo f 4 gt SA um p 5 iesus 6661 E0 KEN Aepuow sie 5 510587 lt weunoog eig L816 08L6 GHVOd NIYIN OLLVIWW3HOS enu gt gt angs auos auos 299 zes ASL INOL wo L OL vou anos n gt quee 64 is 1621 ano SA 968 33A idi zasNas oso 135 35 ali 8 ONAS ZHNI OL iE 1553 mgawa oz LHNI dais St g k auos anto zo SL 5 gt gt daisdwnd SBC PCB pump motor driver door detect 9180 9181 Figure 24 April 2007 Roche Diagnostics A 59 Version 2 0 Service manual 91xx pH El
3. D 13 ERROR LOWER NEEDLE 5 D 13 Roche Diagnostics April 2007 Service manual Version 2 0 D 3 13 Troubleshooting 91xx pH Electrolyte Analyzers Table of contents WHEEL MISSING OR SAMPLER DEFECT D 13 SAMPLER JAMMED OR DEFECTIVE D 14 Roche Diagnostics April 2007 D 4 Service manual Version 2 0 91xx pH Electrolyte Analyzers 13 Troubleshooting Error messages and troubleshooting Error messages and troubleshooting Status LEDs The 5 status LEDs at the bottom of the chassis back panel are lit simultaneously during normal operation See below if any LED is blank ya gings Figure D 1 Status LEDs back panel bottom Possible causes and remedies One ore more supply voltages are out of range or missing Check the voltage test points amp For details see Power circuits on page A 33 If the voltages are correct Check subsequent cicuits and cabling for obvious damage and replace if necessary Check and replace the corresponding fuse Replace the SBC board Replace the power supply STATUS NOT CALIBRATED This message is only displayed when a calibration has been interrupted by pressing NO or opening the sam
4. LanguageSwitch English English Deutsch WARN Polski Fran ais 0 Portugu s d Espafo WARE Spare 3 i Spare 4 Figure B 46 Example 9180 language label TEST SAMPLE DOOR except 9181 Example display This test verifies the operation of the sample door light gate and its associated circuitry The display indicates an O or C depending on the position of the door SampleDoor C C Closed O Open TEST PROBE 9181 only Example display This test verifies the operation of the sample probe motor position sensor and associated circuitry The U represents the output of the upper position sensor the represents the output of the lower position sensor U 0 L 1 Sample Position TEST AUTOSAMPLER 9181 only Example display TEST PUMP Roche Diagnostics This test verifies the operation of the Autosampler turntable assembly The display indicates the output of both the position and home sensors Note If the Autosampler is not plugged in or the TABLE PRESENT bit is not activated this test will not appear under the SERVICE FUNCTIONS menu Home 0 Pos 1 This test turns on the peristaltic pump at each of the four pump speeds The display indicates for two seconds each pump speed while each pump speed is being tested Observe that the peristaltic pump changes spe
5. SBC Board Display Board Preamplifier Display Driver Display Module YES NO Switches Ph Elect Ph Amp n not used not used Reference Temp Sensor Sample Contacts Electrode Temperature Control l Sample Contact1 Preheater Printer Driver Analog Selector and A D Converter JP1 Printer Assembly Microprocessor Power Supply 22 VDC 10 VDC 8 VDC 8 VDC 22 VDC 16 VDC Peristaltic Pump Driver P Pump Motor JP8 JP6 Lamp Board Sample Door v JP7 Interface Driver Solenoid Valve Drivers Valve A JP1 Valve B JP2 JP3 Valve C JP3 JP7 v Valve V JP4 rd Valve R JP5 JP1 90 250 V Power Receptacle Power Supply Module MAIN 1 6 AT y 50 60H 16 AT Fuse F1 2 5 A 250 V 5 VDC 24 VDC Saare 12 VDC 12 VDC Figure 4 9110 analyzer block diagram Roche Diagnostics April 2007 Service manual Version 2 0 A 39 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams 9120 9130 9140 9180 analyzer block diagram SBC Board Preamplifier Fluid Pack Detector Sensor 9180 only Na Electrode s s s Na Amp K Display Board Cl Ca Li Amp Reference Display Driver JP
6. Aqpueis BSuluonipuoo ldues 1591 0 09 Area SO Wd s pono la YYY 4eqdeus Buiue l5 ldues Hodey 5 1591 poO 19113 eBueuo euo Area L jug SHA 9120 9130 9140 program flow Figure B 11 April 2007 Roche Diagnostics Version 2 0 Service manual B 20 91xx pH Electrolyte Analyzers 7 Software operation 9120 9130 9140 Software operation 9120 9130 9140 Main menu READY 12 45 NO PRINT FUNCTIONS NO QC STANDARD URINE SAMPLE NO DAILY MAINTENANCE NO Legend Display prompt Action performed by analyzer 2 second display prompt Decision making by analyzer QUQL OPERATOR FUNCTIONS NO PROGRAM INSTRUMENT NO SERVICE FUNCTIONS NO NO CALIBRATION Figure B 12 9120 9130 9140 main menu Roche Diagnostics April 2007 Service manual Version 2 0 B 21 7 Software operation 9120 9130 9140 91xx pH Electrolyte Analyzers Software operation 9120 9130 9140 Measurement sequence READY 12 45 OPEN PRINT FUNCTIONS Open Sample Door Introduce Sample DOOR OPENED Please Introduce Sample lt air voltage lt air voltage Remove Sample and Wipe Probe 2s Please Close Sample Door FLAP CLOSED Serum Sample in Progress 65 Remove Sample and Wipe Probe Please Close Sample Door FLAP
7. ius o Z064 GNOV T 10 106 Se o mor 10 ATIVOISAHA LON OG 06H 006H IHILON System interconnect 9180 9181 Figure A 21 April 2007 Roche Diagnostics Version 2 0 Service manual A 56 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams SBC PCB power supply 9180 9181 m P Y 15945 6661 E0 AEN Repuop E F T T 5 910842 v I on oqunn wewnoog ws 1816 0816 GHVOS NIYIN DLIVIWBHOS 4 1 OW AG anu 5 T T 5 T anyo anyo anyo anyo anyo anro gio gt lt sol sol ro t T T T F T anto anto anro anro anro ano anro anro anro anro anto anro anro anro anro anto anro so 529 zuo 129 012 695 895 199 999 599 zo uo oo LL 69 so 19 I l l l l l l sas F L1 ho 1547 anto wo 9 m in B S 5 VAS O g 00 NIA A uagioHasna SOSZWT LJ en PATH cua FH
8. 38 Snivisav 35 113SaV 6 p 2 19909 SE 3 500502 v Jequinn XX16 GYVOd OLLVINSHOS enu qos Loar AUT ATHLIVHWIXONddV T ZN ZX MEG V 6012121 5 T 028 0PLON 1nooso gus amp ano Nioso 66 ZHW8S 0086 2 o A 1noosoJ4n8 x T 100 NIA Hons 195280 x rem anas GIOH NNY 82 2 ls NS8H NEST 001 enivis p 091 99H z 190 5 Sivadv 1 TH N 2 Tlvaav zia 3 0 S 999 7 99 See 0 v oig 62 521 m m L 68 dvO nN o 88 dvo4ay H amp E 259 x 18 4 m AS xx z 98 zu x 2 Y1VQ0V RE NOWNOS Hg siot As0 t Ht E x Es Ig HE z 7 ox den om zur ML WEINSANOO G V T OL 8 au S no i lo gt gt LNOdWAL gt gt 1nor ono gt gt gt gt indiden SBC PCB temperature circuit 9110 9120 9130 9140 Figure A 12 April 2007 Roche Diagn
9. cL OL 100 E vr 2 k ZL 8L L TU T een 29 v6SDHPL TULOYA Nadsid gt 1SN 1VO WudsSIq gt gt gt gt YLSA SN3dO FN3dO N3dO ZN3dO azy IN3dO aon e L 6264 ah 99 Z 3oov gt gt m z T HESOHPZ 8z6H 99 Nnuav ber 13931 Na8HQv gt gt Y Nass Qo Nous zasav 8855 E h 013807 L I gt gt ISON v SBC PCB microprocessor 9180 9181 Figure A 29 April 2007 Roche Diagnostics Service manual Version 2 0 A 64 Electronic diagrams 5 Electronic diagrams SBC PCB Fluid pack detector 9180 9181 91xx pH Electrolyte Analyzers LL 30 says 6661 0 Ae Aepuoyy eje d 9 910802 v sequiny wewnsog zs 1816 0816 NIYIN OLLVINSHOS ho ho ho 310001 yi 819 ano 3109 yo g 091 gt L 189 Z T 2901099 26 E NB I a Z11840 y w z J3nio o Re 1343 r 2 J Te T T 089 TOST ho Ineo ho ho ern WH Ki 002 091 a i 1168 8 NLoszwI 99 ven e le mu 9 99 99 A 029
10. New SnapPak Reset installed counter to 00 Background check in all menus except Take out of Operation and Service Functions PRINT FUNCTIONS NO SnapPak NAK CA READY CALIBRATION Figure B 32 9180 power up QC STD DIALYSATE URINE SAMPLE DAILY MAINTENANCE NO NO SERVICE PROGRAM OPERATOR FUNCTIONS INSTRUMENT FUNCTIONS Roche Diagnostics April 2007 Service manual Version 2 0 B 45 8 Software operation 9180 91xx pH Electrolyte Analyzers Software operation 9180 Roche Diagnostics April 2007 B 46 Service manual Version 2 0 91xx pH Electrolyte Analyzers 9 Software operation 9181 Table of contents Software operation 9181 In this chapter Chapter Software operation 9181 49 OVERVIEW 49 Overall program operting flow B 50 51 Manual measurement sequence B 52 Automated measurement sequence B 53 Print functions MENU JN clip codes AS B 54 QC std dialysate urine sample menu B 55 Daily maintenance menu B 56 Operator functions 57 Program instrument
11. eitis 15 PLC E 15 Roche Diagnostics April 2007 Service manual Version 2 0 A 11 2 Specifications 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 A 12 Service manual Version 2 0 91xx pH Electrolyte Analyzers 2 Specifications Specifications Specifications Specifications 9110 Sample types Whole blood Serum Plasma QC solutions Reported parameters Measuring range Resolution pH 6 00 8 00 0 001 pH units Table A 1 9110 reported parameters 9120 9130 9140 9180 9181 Sample types Whole blood Serum Plasma Reported parameters Measuring range Resolution Sodium Na 40 205 mmol L 0 1 mmol L Potassium K 1 5 15 mmol L 0 01 mmol L 0 8 15 mmol L dialysate Chloride 50 200 mmol L 0 1 mmol L Calcium Ca 0 2 5 0 mmol L 0 01 mmol L Lithium Li 0 1 6 0 mmol L 0 001 mmol L Table A 2 9120 9130 9140 9180 9181 reported parameters 1 Sample type Urine Reported parameters Measuring range Resolution Sodium Na 1 300 mmol L 1 0 mmol L Potassium K 4 5 120 mmol L 0 1 mmol L 60 120 mmol L with add dilution Chloride 1 300 mmol L 1 0 mmol L Table A 3 9120 9130 9140 9180 9181 reported parameters 2 a Calcium and Lithium are not measured in urine samples Lithium is not measured in dialysate samples Operating Parameters 9110 Sample types Whole blood serum plasma aqu
12. 58 Service functions menu nad gutter e based Det bete eat bein B 59 Calibration sequence 1 example Na K Li calibration B 60 Calibration sequence 2 example Na K Li calibration B 61 Power up sequence tnnt B 62 Roche Diagnostics April 2007 Service manual Version 2 0 47 9 Software operation 9181 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 B 48 Service manual Version 2 0 91xx pH Electrolyte Analyzers 9 Software operation 9181 Software operation 9181 Overview Roche Diagnostics Overall Program Operating Flow Main Menu Measurement Sequence Print Functions Menu QC Std Dialysate Urine Sample Menu Daily Maintenance Menu Operator Functions Menu Program Instrument Menu co a G Ch G N ervice Functions Menu 10 Calibration Sequence 11 Power up Sequence Software operation 9181 April 2007 Service manual Version 2 0 B 49 91xx pH Electrolyte Analyzers 9 Software operation 9181 Software operation 9181 Overall program operting flow waung ul apog 1eju3 adwy 1991 3 Z suiq dune 1591 oN 5 1891 oN dunq 1591 ON 1591 ON 1 1591 oN yamg 6en6ue1 1891 1osu s 1591 1osu s ldues
13. A 50 Printer daughter PCB 9110 9120 9130 9140 9180 9181 A 51 Lamp PCB 9110 9120 9130 9140 Listen nne 52 Display daughter 53 Preheater 54 Display daughter PCB 9120 9130 9140 55 System mterconnect 91509181 uu een 56 SBC power supply 9180 9181 A 57 SBC PCB valve drivers 9180 9181 A 58 SBC pump motor driver door detect 9180 9181 59 SBC PCB input amplifier 9180 9181 A 60 SBC PCB temperature circuit 9180 9181 A 61 SBC PCB analog selector and A D converter 9180 9181 62 SBC PCB microprocessor 9180 9181 63 SBC microprocessor 9180 9181 64 Roche Diagnostics April 2007 Service manual Version 2 0 A 37 5 Electronic diagrams 91xx pH Electrolyte Analyzers Table of contents SBC PCB Fluid pack detector 9180 9181 eee A 65 Lamp PCB 29180 9181 i ENDE IE A 66 Display daughter a a a O A A 67 Roche Diagnostics April 2007 A 38 Service manual Version 2 0 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams Electronic diagrams 9110 analyzer block diagram
14. Roche Diagnostics April 2007 A 6 Service manual Version 2 0 91xx pH Electrolyte Analyzers 1 Safety information 9 5 Disinfectants A Deproteinizer Composition Hazards identification First aid measures Other disinfectants 9 5 Roche Diagnostics Disinfectants e Service and repair work must be done only as specified in this manual Unqualified service or repair work may result in warranty claims not being granted e only suitable tools and test instruments for service and repair work e Do not allow fluids to enter the interior of the instrument because this may damage the electronics e Use only slightly moistened cloths or cotton sticks to clean the instrument Use only liquid disinfectant such as protein remover Roche deproteinizer or an alcohol based about 70 surface disinfectant Do not spray disinfectant directly onto the instrument because this could cause malfunctions in the electronics Do not use any type of bleaching agent Exception Roche Deproteinizer Do not attempt to clean decontaminate any part of the instrument before shutting it down and unplugging it from the power source Before plugging in the instrument again and switching it on always wait for 15 minutes to allow the disinfectant to evaporate Aqueous NaOCI solution with active chlorine x 2 Due to the basic and oxidizing character of the reagent Deproteinizer local irritat
15. ae ENIA OL Zn wsamne noni 189 6 en re I jodie Le zoas 204 90 2 ELELOHbL E Dux eoa eay Sour zodizav 909 1 ne oda oav 528 13931 20 220 a Svd i20 20 wx Eg Pal 90 00 vd 2O S wor zune a vi amp 3922 T 5 k I LI 9 8 Si Ten 690 SSuudav 9110 9120 9130 9140 microprocessor SBC PCB Figure 14 April 2007 Roche Diagnostics A 49 Service manual Version 2 0 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams SBC PCB printer display drivers 9110 9120 9130 9140 75 557 SI 3 500502 x sequinn wewnsog XXL6 AHVO OLLVINSHOS awy Wes OTT6 anys gt gt Favs S9LOHPL FS 55 Uns uisiav HNI smog a gu WLvadw
16. Display board 9110 only 9181 only Roche Diagnostics Description of modules The display board contains the LCD display sample contact circuit 9110 only or sample sensor connector the preheater 9110 only and the YES and NO switches used for operation of the analyzer Power and signal interconnection is accomplished via a ribbon cable from the SBC board through JP2 on the display board The sample contact circuit consists of a square wave generator U1B which generates a 500 Hz signal U1A amplifies the signal D2 and C3 rectify the square wave The gain can be adjusted with R10 The sample contact output provides a DC voltage which is proportional to the square wave amplitude During sample aspiration the preheater tube and the sample contact 1 tube are connected to the square wave via relay K1 The heater control circuit controls two transistors which are placed in the preheater block A thermistor provides temperature feedback to the setting of potentiometer R20 The signal difference drives the base of the heating transistors via buffer U2C U2D shuts the system off in case the thermistor opens U2B limits the maximum heating current during warm up U1C and attached circuit is used as an AC DC converter to create negative voltage Vss using square wave generator U1B Display boards utilized in the 9181 analyzer also contain circuitry for both the needle mechanism motor control and position sensor The autosampler Table
17. Fluidic Module A 27 Cleaning solution 9110 only A 28 Roche Diagnostics Main tube set all models Intake tube set 9110 only 28 Reference solution 28 Sample Contacts 9110 only A 29 Standard A A 27 Standard B A 27 Standard C not on 9110 A 27 Waste liquids A 28 Front door assembly A 23 How to Use This Manual 9 Important information A 5 Important notes and warnings A 6 Intended use 3 Interface information A 19 Interface specifications Data Link Information 9180 9181 only A 20 Example data string information A 20 Interface information A 19 ISE SBC card A 33 Lamp board A 27 A 34 Main tube set all models Intake tube set 9110 only A 28 Maintenance C 9 Annual preventive C 9 Daily 0 9 Monthly C 9 Six month C 9 Weekly C 9 Mechanical assemblies Display board A 27 Electrode holder A 24 Front door A 23 Lamp board A 27 Needle unit A 23 Peristaltic pump A 24 Printer A 25 Rear panel A 25 SBC board A 26 Solenoid A 24 April 2007 F 4 Service manual Version 2 0 91xx pH Electrolyte Analyzers Needle unit assembly A 23 Operating Parameters 9110 A 13 Operating Parameters 9120 9130 9140 9180 9181 14 Operating safety information A 6 Other disinfectants A 7 Peristaltic pump assembly A 24 P
18. VAS z Lou 95 65 MSYN AWATO AUOOZZ 195440 08 4n sro NI ASYN ON ALON sHOHd DNIN LON SUAIAI IdWNY SBC PCB input amplifier 9180 9181 Figure A 25 April 2007 Roche Diagnostics Version 2 0 Service manual A 60 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams SBC PCB temperature circuit 9180 9181 lt lt TEMPOUT R62 1K R110 100K TP12 E a 3 EI 855 22 D x e R51 10K 1 R53 274K 1 R52 10K 1 R55 68 1K 1 lt z So UU S lt lt 7 T gt 8 o bu TP13 40 4 R54 2 ADJ 10KPOT 30 x N K x g 2 i 8 En M ou 5 oz T D 85 x 5 SS s sk 5 b g z gi 4 s M is lt ak Y g lt ID ol B Figure A 26 SBC PCB temperature circuit 9180 9181 Roche Diagnostics April 2007 Service manual Version 2 0 A 61 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams
19. uagioHasna F zu FH anto de m anyo so 820 T T uagioHasna ho he o no vas z 100 NIA Fg S I mut zn SdL ar ZS Sa AET RS 095 SATA r asz anozy Le 1 w z 00 gt gt morsiisv4 Seo F sa ou 1 0 iH No Io 53 2 095 AET u Aszanoze _ S20 z k wEWO0E 0 E m 2H 1 0 MNISLV3H WSN oy ONZI d T we LSH EA ADE hs Bel gt anyo vH gt anyo m 5 to 05 1 005 Z z INGE Ta F F 66 T a 9H 19871 S z 00 NIA Fe onvz Avze 1 gt T B 4 wACT AZC u ozz Ast Aewdnoogz SI saol T z M z T TEM zu oso F su gt gt Nadwv1 uber w 3 Ta AS ozz lt t99 F m CN DIES ive o 6 id gt T O09ALNIHd su SBC PCB power supply 9180 9181 Figure 22 April 2007 Roche Diagnostics A 57 Version 2 0 Service manual 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams SBC PCB valve drivers 91
20. AINT V 5 aima 5 13 1 3 dWNd 434 38 334 3808 JIdWYS IN3AYIV Z 3 3 ASSY 300412313 X YOLDANNOD aud JINI3Y343Y A18W3SSV 38OHUd 31dAVS 100 MOTA 10d MOTI3A 3 0116 9110 fluidic diagram Figure A 1 April 2007 Roche Diagnostics Service manual Version 2 0 A 30 4 Description of modules 91xx pH Electrolyte Analyzers Description of modules YOLDINNOD AINT 38 V GION31OS aim3 1dOdTIl4 2 435 ALSVM 338 dWNd 434 38 334 3308 31dWVS P uqloN31OS YOLIANNOD 33 383338 Al8W3SSV 38084 JIdWVS 10 AOTI3A TONNES WVYSVIA DIGINIA Ot L6 0 16 0216 9120 9130 9140 fluidic diagram Figure A 2 April 2007 Roche Diagnostics A 31 Service manual Version 2 0 91xx pH Electrolyte Analyzers 4 Description of modules Description of modules YOLDINNOD GIN V d aima g Reb dc 2 38 1804 dWNd 434 38 334 3808 JIdWYS 1 3 ASSY 3408132313 DI gt 438 MOTI3A 35 383138 A18W3SSV 38084 JIdWVS MOTI3A INVu5VIG DIGI 113 1816 0816 9180 9181 fluidic diagram Figure A 3 April 2007 Roche Di
21. NO Calibration due NO Figure B 16 9120 9130 9140 daily maintenance menu Roche Diagnostics April 2007 Service manual Version 2 0 B 25 7 Software operation 9120 9130 9140 91xx pH Electrolyte Analyzers Software operation 9120 9130 9140 Operator functions menu OPERATOR NO PROGRAM FUNCTIONS INSTRUMENT YES Check Change YES 37 SnapPak remaining NO Go into STANDBY Standby Mode YES gt READY YES NO Leave Standby Mode YES Change YES New SnapPak YES Fluid Counter SnapPak installed reset to 100 CALIBRATE IS NO READY 12 45 CALIBRATION gt Daily Remain YES Daily i in Daily Cleaning Conditioning Maintenance 1min NO IS CALIBRATION DUEZ 30 SECOND STABILIZATION READY 12 45 CALIBRATE Reset Sample A 2 SET SAMPLE Number you Sure NUMBER TO 0 YES Take out of Shutdown Plug YES Transp Housing YES Remove Water YES All Electrodes amp Water installed Ref installed and Press YES Operation Plug Removed Remain in Com Do You Wish Pump Windings Operator Func ee Off DELETE all Data Relieved YES Set Date 01 FEB 07 Time Date Time 12 00 ok Date 01 FEB 07 Time 12 00 ok 4 Relief Clamps Inserted Figure B 17 9120 9130 9140 operator functions menu Roche Diagnostics April 2007 B 26 Service manual Version 2 0 91xx pH Electrolyte An
22. gt adjust the sample sensor 1 Select TEST SAMPLE SENSOR in the SERVICE FUNCTIONS menu which will display the current sample sensor setting 2 Adjust R93 Sample Sensor Adjust located at the top of the ISE SBC card to a value of 100 3 Press YES to turn on peristaltic pump and ensure displayed value remains constant Raise sample door and aspirate clear fluid e g water Displayed reading should now be 150 or greater April 2007 Service manual Version 2 0 C 5 11 Adjustments 91xx pH Electrolyte Analyzers Adjustments Q Note R93 can be accessed through the rear cover vent slot and is located at the top center on the circuit card Temperature Adjustment 9110 9140 9180 9181 only adjust the temperature amplifier 9110 9140 9180 9181 Roche Diagnostics 1 Select AMPLIFIER TEST in the SERVICE FUNCTIONS menu which will display the current amplifier voltages 2 Press NO twice to display the temperature voltage and actual temperature display 3 Unplug the preheater 9110 only or the sample sensor and fit the Temperature Adjustment Test Plug in place of the preheater or sample sensor plug 4 Adjust R54 located on the SBC Board for a display of 3000 mV 5 Unplug the Temperature Adjustment Test Plug and fit the preheater or the sample sensor connector to display the actual temperature Wait 5 minutes for the preheater to stabilize The display should now read 37 0 0 2
23. z z m T T qWixa ISSA 98 ISSA WOL 09 3SS HnOL 529 g Nasaan ISSA gt gt OHOVL sevi DOA Og AA pe 35 ISSA 1 Pa 7200 ISSA 1N04919 3SS Ko Siu 5 HOL SH 209 950 204 ISSA LOd PSO L94 ISSA T PLE oy LEE 02d S5 024 3SS SEL ZSO 2vos ISSA XOL ZEL 1595 59 VSSA o r sH aa AALSA YN 166955 DLO gt gt Us SOd 5 SsSuawusr Zsod St N310W8L X gt St HISVA N3LOWON 9 25 0559 1 MOIMI X 89 5 ISON 28 1osauaxa OSIN 59 Sos gt soda m Laa ADVS SOdaldn 38zuammw _ SOJANMT WISIS 58 S3d SV 0999 1021 I 013891 SOdANdN 98 739 59 999 221 5 113591 93d ozis 8 213591 234 25 edodnoo 13897 814 rd9d4 L 90 290 daisdWna a SdDd LD0 90 aA WA gt gt T o saodnoo oo EEL gt gt 41044 MO1SASVA d94 190 590 vOl NadWna gt sr wa Naawv12 zia eto anna i gt gt wazzna za 21 v6 liia epe XOL 56 1asau2 gt Da er gun T z 001 Sd 66 sd EOS ar 9d 0 za savarvav ES HH savasvav 22 gt 91 sa SE T 55 S 93H9 9 savarvav 56 Dy zavd evav Ir gt gt Snuvis
24. CHECK ELECTRODES Possible causes and remedies NOT Roche Diagnostics This message is displayed when the printer head is jammed and tries to print It is a temporary message that is displayed for 2 seconds and then sample results are displayed To clear paper jam remove printer from the analyzer and remove jammed paper Replace printer and retry The printer can be removed by grasping the rear edge of the printer assembly and pulling it to the front The printer can be removed with the Analyzer power on To free up jammed paper turn the small gears on the left side of the printer module This message is displayed when either of the following conditions are present 1 More than six aspirations of Standard A are required for all electrode channels 2 The difference between voltage A and B is out of specification for all channels Check the level of the filling solution in the pH electrode 9110 only Electrodes are not plugged in Check the slide mechanism to ensure it is in the correct position and firmly pushed into electrode connectors Perform Electrode Cleaning 9110 only or Daily Maintenance Check or replace reference electrode Replace ISE SnapPakTM NA NOT CALD all except 9110 KNOTCALD CLNOT CALD April 2007 Service manual Version 2 0 D 9 13 Troubleshooting 91xx pH Electrolyte Analyzers Error messages and troubleshooting CA NOT CA D LINOTCALD These me
25. Daily maintenance menu B 25 Main menu B 21 Measurement sequence B 22 Operator functions menu B 26 Overall program operating flow B 20 Print functions menu B 23 Program instrument menu B 27 QC standard urine sample menu B 24 Service functions menu B 28 Software operation 9180 B 33 Calibration sequence Example Na K Li calibration B 43 B 44 Daily maintenance menu B 39 Main menu B 35 Measurement sequence B 36 Operator functions menu B 40 Overall program operation flow B 34 Power up 45 Print functions menu 37 Program instrument menu B 41 QC standard dialysate urine sample menu B 38 Service functions menu B 42 Software operation 9181 B 49 Automated measurement sequence B 53 Calibration sequence example Na K Li calibration B 60 B 61 Daily maintenance menu B 56 Main menu B 51 Manual measurement sequence B 52 Operator functions menu B 57 Overall program operting flow B 50 Power up sequence B 62 Print functions menu B 54 Program instrument menu B 58 QC std dialysate urine sample menu B 55 April 2007 Service manual Version 2 0 F 5 91xx pH Electrolyte Analyzers Service functions menu B 59 Solenoid assembly A 24 Specifications 9110 A 13 9120 9130 9140 9180 9181 A 13 System interconnect 9110 9120 9130 9140 A 4
26. Software operation 9181 Daily maintenance menu DAILY OPERATOR MAINTENANCE FUNCTIONS Perform Daily Probe moves to upper position Cleaning Sol detected Introduce Sample Cleaning Sol not detected Introduce Sample Condition Sol not detected Cleaning Solution Perform Daily Probe moves to Conditioning upper position Remain in Daily Maintenance SS ram NAKCA READY RERFORMED STABILIZATION CALIBRATE s Calibration due Figure B 39 9181 daily maintenance menu Roche Diagnostics April 2007 B 56 Service manual Version 2 0 91xx pH Electrolyte Analyzers 9 Software operation 9181 Software operation 9181 Operator functions menu STATUS NO SnapPak Pack OPERATOR NO PROGRAM detected FUNCTIONS INSTRUMENT LOBAL Background check in all menus YES except Take out of Operation Check Change YES 3795 SnapPak remaining NO Go to STANDBY Standby Mode YES READY YES NO YES Change YES New SnapPak YES Fluid Counter SnapPak installed d reset to 100 eave Standby Mode Na K READY M mE Daily T Daily i Remain in Daily Cleaning Conditioning Maintenance Set Time Date Date 01 FEB 07 Time 12 00 ok Parameter added YES Electrode in Place YES Sel Parameter Na K Li NO Reset Sample ves
27. Standard A STANDARD A detected NOT DETECTED YES Aspir Std C into Meas Chamber Standard C detected YES Obtain Electrode Voltage Std C Perform Wash Cycle Standard A detected Obtain Electrode Voltage Std A Calc of Std A Std B slope NO STANDARD C NOT DETECTED STANDARD A NOT DETECTED At least 1 slope in Range CHECK ELECTRODES Slope A B and A C ok for all Electrodes NAKLI READY XX XX NOT CAL D All Std A volt Reprod NAKLI READY XX XX NOT CAL D YES NAKLI READY Figure B 31 9180 calibration sequence 2 Example Na K Li calibration Roche Diagnostics 44 Service manual Version 2 0 91xx pH Electrolyte Analyzers 8 Software operation 9180 Software operation 9180 Power up 9180 A80X2 AA WARM UP All dots lit for sec Date 21 FEB 07 Time 12 00 Set Time Date Power Fail 20 FEB 07 1 12 P Background check in all menus k lace er except Take out of Operation and service Functions YES Tn Perform Daily Perform Daily Remain in Daily S Calibration Wash Cleaning Conditioning Maintenance due YES E Stabilzi lean Cond iti abilzing ceded f 5 9180 A80X2 AA Clean Cond WARM UP eeded Calibration Pack inserted Global STATUS Pack in place Global YES
28. uAZT AZZ u aa ozz nAG u Q Lro sel 7 z 2 T 9 Ta m T su 1V8 l gt gt N3dWv1 4 zo NOT 24 uTd AStu qa 022 eola e v99 F L t Za 5 o 1 9 Nor MO0L 6 13 Dan su SBC PCB power supply 9110 9120 9130 9140 Figure 8 April 2007 Roche Diagnostics 43 Version 2 0 Service manual 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams SBC PCB valve drivers 9110 9120 9130 9140 eag XX16 NIYIN OLLVIWSHOS enu 218 qos equ a gt gt D dWalua o JNOS gt gt anos a ae gt gt sNado sta Aos anz z T K z or am gt gt Aos anz z gt gt 4 n 46 z mL IN gt gt N3do ela Aos anz z gt gt 4 i 46 z se ML om gt gt zN3do ziH 9 3 2 2 gt T 2 5 sera gt gt N3dO 0S Jnz z P V3ATVA gt T z ho Nur 5 DIANIA 028 iza 25 JH LLO SELN D
29. 1961 s pono l3 1591 SNOLLONNA ON 3OIAH3S NOLIVHSITVO duny ul oel lul dnas weiBold 29101024 4402 euuf 29101024 4402 ejejeoy 290 0 4 1109 qie91g 198 1102 jeuuoN s Bueu 3497 90 sebuey jene OO 8197 OO 18113 LNANNYLSNI WvH9OHd duny 10je1edo uonei do Jo no exer 1 quunN ajdwes 19594 Jejaweieg 109 95 eyeq eui p 195 apon qpuels 0105 amp xyegdeug oDueu2 oeu5 SNOLLONNA HOLVU34O 8 1 9 oueu lu eW ur ureway Suluonipuoo Areg wiopad Suliue lo wiopad amp 3ONVN3LNIVIA April 2007 lduies euunypis OO suonouna juug ejeuogieoig ldues Sonsneis pue piepuejs senjeA 20 z 9 e dwes uud 9 5 D 3 2 dues uny 6 ise 2 5 uod y ajdwes z 18497 OD jse 2 lt lt e gt SOA es 22 m a 3 1d NVS SNOILONNA LNAWA
30. 80 and 81 Analyzers have been designed to require minimal electronic adjustment The sample contact circuit may require adjustment following a change of the sample contact components or during a routine troubleshooting 9110 only or the sample sensor circuit may require adjustment following a change of the sample sensor or during routine troubleshooting The temperature adjustment is active only for models 9110 9140 9180 9181 and requires the temperature adjustment test plug for correct calibration of the temperature circuit Sample contacts adjustment 9110 only 0 In preparation for correct electronic adjustment several steps must performed To adjust the sample contacts 1 Slide the electrode holder forward and remove the electrode from the analyzer 2 Disconnect the intake tube from the preheater 3 Remove the trim panel located below the needle mechanism and disconnect the plug for sample contact 1 4 Select TEST SAMPLE CONTACTS in the SERVICE FUNCTIONS menu and adjust R10 Display Board for a reading of 140 5 Reconnect sample contact 1 plug intake tubing and electrode 6 Check to ensure the value 140 remains the same Note If the value drops after reconnecting each of the three sample contacts follow the troubleshooting guidelines for fault isolation lt For details see Error messages and troubleshooting on page D 5 Sample sensor adjustment 9120 9130 9140 9180 9181 only Roche Diagnostics
31. 9140 LCD DISPLAY Electronic diagrams SAMPLE SENSOR JP1 5VA 9 SAMPLELED DD1 9 SAMPLEPT DD2 pi SAMPLEGND DD3 i TEMP DD4 9 DDS RJ12 DD6 a DD7 9 15 43 8566 HEADER 14 ALTERNATE LCD DISPLAY 3 CONTRAST NOT STUFFED 10 DATAJINST ji 4 DISPR W DISPEN HEADER 26 1P1241 1082 3 SCHURTER 52 o o 1 1241 1082 3 SCHURTER NO MAINDOOR display sch ZPWB1 PWB NOT STUFFED SCHEMATIC DISPLAY DAUGHTER BOARD ize Document Number B LCISE ate Thursday February 27 2003 heet Figure A 20 Display daughter PCB 9120 9130 9140 Roche Diagnostics April 2007 Service manual Version 2 0 A 55 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams System interconnect 9180 9181 SvLrNL SANIT NSINYVHIEN HHL LV ATIVOGIAIGNI SI GNOS IdWVS Juswnoog sz y z T 1819 0816 QUVOS NIV OUWWSHOS DA en ees
32. C Adjust R20 on the display board clockwise to increase for 37 0 0 1 C Due to the elevated temperature inside the analyzer it is normal for the displayed temperature to be several degrees C higher than the actual room temperature If no Temperature Adjustment Test Plug is available place a quality thermometer preferably mercury next to the right side electrode holder With the front door open wait approx 10 min Then adjust the displayed temperature to match the reading on the thermometer April 2007 C 6 Service manual Version 2 0 91xx pH Electrolyte Analyzers 12 Maintenance Maintenance Insert introducing text here In this chapter Table of contents Chapter 12 Maintenance Daily maintenance Weekly maintenance Monthly maintenance Six month maintenance Annual preventive maintenance Roche Diagnostics April 2007 Service manual Version 2 0 7 12 Maintenance 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 C 8 Service manual Version 2 0 91xx pH Electrolyte Analyzers 12 Maintenance Maintenance Maintenance Maintenance procedures for the 9110 20 30 40 80 81 Analyzers require minimal time by the operator to perform The procedures outlined below should be performed by the schedule indicated Detailed in
33. ES 9 3 3 t06 NZ a 2 ne gt gt u3zzna v u3s 3 ol vaav S6SOHr VLvqasq gt gt ausia L o TaLoud g YOLOW gt HO 5 90 sos Jod 30 gt gt anons 100 20 8 1100 vo aas OOALNIHd DD 30 PEE gt gt SN3dO TN3dO do anys la N3dO 29 nS n ZN3dO LN3dO NS ws vL Rn doen iT zT IK PESOHrZ Is DELOGV HO gt 90 3o uLsoayv see 3 do M1ous xm Nass 90 4 oL gt gt 8 aas Ar gt gt son SBC PCB printer display drivers 9110 9120 9130 9140 Figure 15 April 2007 Roche Diagnostics Version 2 0 Service manual A 50 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams Printer daughter PCB 9110 9120 9130 9140 9180 9181 Lp p T 19949 sser UL ASW epugw a 005 v Jaquinn Juawnoog ezig HIHOR VG OLLVWI
34. Roche Diagnostics April 2007 B 14 Service manual Version 2 0 91xx pH Electrolyte Analyzers 6 Software operation 9110 Calibration sequence CALIBRATION YES Calibration in Process 124 Clear Sample Path Calibrate Sample Contacts Sample Contact Voltage ok CHECK SAMPLE CONTACTS Prime Standard B Standard B detected STANDARD B NOT DETECTED Prime Standard B StandardA detected STANDARD A NOT DETECTED Pressurize Ref Asp Std A into Meas Chamber Reference Housing ok CHECK REF HOUSING Software operation 9110 Asp Std A into Meas Chamber STANDARD A NOT DETECTED Standard A detected Obtain Electrode Voltage Std A 3 Std A Values obtained Are Values 6 Aspir Of Std A reproducible performed YES YES Asp Std B into Meas Chamber CHECK ELECTRODES Standard B detected NO STANDARD B NOT DETECTED YES Obtain Electrode Voltage Std B Perform Wash Cycle Asp Std A into Meas Chamber Standard A detected NO STANDARD A NOT DETECTED YES Calculation of Std A Std B NO At least 1 slope in Electrodes in Range Range NO CHECK ELECTRODES YES YES READY 12 45 READY 12 5 XX NOT CALD Figure B 10 Roche
35. Service Code NO Code Error ini NO se s Calibration Calibrate toe Figure B 19 9120 9130 9140 service functions menu Test Amplifier YES Na 1030 1200 NO Ground 2 1093 2500mV 2496 Roche Diagnostics April 2007 28 Service manual Version 2 0 91xx pH Electrolyte Analyzers 7 Software operation 9120 9130 9140 Calibration sequence CALIBRATION YES Calibration in Process 124 Clear Sample Path Calibrate Sample Contacts Sample Contact Voltage ok Prime Standard B Standard B detected Prime Standard B Standard A detected CHECK SAMPLE CONTACTS STANDARD B NOT DETECTED STANDARD A NOT DETECTED Pressurize Ref Asp Std A into Meas Chambe Reference Housing CHECK REF HOUSING Software operation 9120 9130 9140 Asp Std A into Meas Chamber Standard A detected Obtain Electrode Voltage Std A 3 Std A Values obtained STANDARD A NOT DETECTED Are Values reproducible 6 Aspir Of Std A performed YES YES CHECK ELECTRODES Asp Std B into Meas Chamber Standard B detected NO STANDARD B NOT DETECTED YES Obtain Electrode Voltage Std B Perform Wash Cycle Asp Std A into Meas Chamber Standard A detected NO STANDARD A NOT DETECTED YES alculation
36. ov ay z m 2 x d 9 r4 23 a 20 vice a A an 981 90 9L ER 90 _8r ER si y b P ER L fa ER sav L 20 L L 1 toy sr Eg over 59 av sr 0 5 t av r 8 Fara idv oq Ww ei oq PA 2149 w lo Stuy p tr Ov i Ov it Ov 5 En sen Y ese u gare gt gt Lrolav MDOTO 3 NWA TON Lnozezsy gt gt WVH d S ILLYIOA NON WwwuDowd Im 5 voen gt TA gt gt Nizezsu 13534 gt L OS gt ISOW gt gt gt MO1s isv4 gt gt 9 dalsdWNd gt gt 38dsia ae Mazzna OSIN OX wsawne SINN gt gt z T gt gt Bons zs0014 a gt gt 0334 ee URS z T z T isla ks a LOH899A a az 904 58 DIE s aa ISON e reaia SEN gt gt S3A T 999 Suis NH vauls Sv 5 ads 98d riv 2 52 SAd ELV sagisnv T SONNE l taan 27 j Eu NAWONd
37. serum measuement CHP Display results in concentratio of cH ions X CDC Clears all Service Codes X X X X X X DEC Additional display resolution X X X X X X ECO Economy mode X X X X X X EXW Fxtended wash following serum blood X samples FIF Cal voltages on display X X X X X LEM Last 20 error messges printed MGL Display mg dL x X X NOB No beep X X X NTD Prevents temperature from being reset X following each sample QCC QC values direct ISE measurement X X X X X X SFC SnapPak counter input X X SLC Set Leasing counter X X SSC Small sample cups X TMP Allows temperature correction for X measured results Table B 2 Instrument service code matrix Roche Diagnostics April 2007 Service manual Version 2 0 B 69 10 Service functions 91xx pH Electrolyte Analyzers Service functions Miscellaneous functions Deleting data Clearing sample count gt Toreset the sample count to 0 1 At the READY display and press NO until the display OPERATOR FUNCTIONS appears Confirm with YES 2 Press NO until Reset Sample Number appears Press YES 3 When the prompt Are you sure appears press YES again Clearing all data gt To reset the analyzer software to the default values 1 Press the YES and NO keys simultaneously and switch the analyzer off and on again This function can be performed optionally upon completion of decommissioning CAUTION 9 5 All QC data and instrument programming will be lost All Q
38. 12 Maintenance Calibration sequence 1 Example Na K Li Maintenance 9 calibration B 43 Daily maintenance C 9 Calibration sequence 2 Example Na K Li Weekly maintenance C 9 calibration B 44 Monthly maintenance C 9 Power up B 45 Six month maintenance C 9 Annual preventive maintenance C 9 9 Software operation 9181 zn m Troubleshooting Part D Overall program operting flow B 50 Main menu B 51 13 Troubleshooting Manual measurement sequence B 52 Error messages and troubleshooting D 5 Automated measurement sequence B 53 Status LEDs D 5 Print functions menu B 54 STATUS NOT CALIBRATED D 5 QC std dialysate urine sample menu B 55 STANDARD A NOT DETECTED D 5 Daily maintenance menu B 56 STANDARD B NOT DETECTED STANDARD C Operator functions menu B 57 NOT DETECTED 9180 9181 only D 6 Program instrument menu B 58 CHECK SAMPLE CONTACTS 9110 CHECK Service functions menu B 59 SAMPLE SENSOR 9120 30 40 80 81 D 6 Calibration sequence 1 example Na K Li CHECK REFERENCE HOUSING D 7 calibration B 60 PLEASE CLOSE SAMPLE DOOR except 9181 D 7 Calibration sequence 2 example Na K Li NO SAMPLE D 8 calibration B 61 CLEANING FLUID NOT DETECTED D 8 Power up sequence B 62 CONDITIONING FLUID NOT DETECTED except 9110 D 8 INTERFACE ERROR D 9 Roche Diagnostics April 2007 6 Service manual Version 2 0 91xx pH Electrolyte Analyzers PAPER JAM OR PRINTER DEFECT D 9 CHECK ELECTRODES D 9 NOT CALD D 9 CLOG IN SAMPLE PATH 9110 only CLOG
39. 5 82 1 250 1 NO YES 5s SAMPLE Urine pen Door and OPEN Wipe Probe Close CLOSED Urine Sample Na K Cl IN Introduce Sample Introduce Sample Sample Door inProgress 49 238 18 8 198 re NO YES Remain in QC Std 3 NAKCA READY Dialysate Urine Sample JNO 2 min 1 j Figure B 25 9180 QC standard dialysate urine sample menu Roche Diagnostics April 2007 B 38 Service manual Version 2 0 91xx pH Electrolyte Analyzers 8 Software operation 9180 Software operation 9180 Daily maintenance menu DAILY NO OPERATOR MAINTENANCE FUNCTIONS Perform Daily YES Open Sample Door OPEN Cleaning Introduce Sample Wipe Probe Close CLOSE Srey JE Sample Door detected leaning Solutio not detected Perform Daily YES Open Sample Door OPEN wipe Probe Close CLOS a Conditioning Introduce Sample Sample Door detected Cond Solution not detected Remain in Daily Maintenance NO 1min Calibration YES due NO Figure B 26 9180 daily maintenance menu Roche Diagnostics April 2007 Service manual Version 2 0 B 39 8 Software operation 9180 91xx pH Electrolyte Analyzers Software operation 9180 Operator functions menu STATUS LOBAL Background check in all menus except Take out of Operation NO SnapPak Pack OPERATOR NO PROGRAM detected FUNCTIONS INSTRUMENT 91601067 YES YES Check Change YES 37 Change YES New SnapPak YES
40. 9130 9140 A 44 SBC PCB pump motor driver door detect 9110 9120 9130 9140 A 45 SBC PCB input amplifiers 9110 9120 9130 9140 A 46 SBC PCB temperature circuit 9110 9120 9130 9140 A 47 SBC PCB analog selector and A D converter 9110 9120 9130 9140 A 48 SBC PCB microprocessor 9110 9120 9130 9140 A 49 SBC PCB printer display drivers 9110 9120 9130 9140 A 50 Printer daughter PCB 9110 9120 9130 9140 9180 9181 A 51 Lamp PCB 9110 9120 9130 9140 A 52 Display daughter PCB 9110 A 53 Preheater PCB 9110 A 54 Display daughter PCB 9120 9130 9140 A 55 System interconnect 9180 9181 A 56 SBC PCB power supply 9180 9181 A 57 SBC PCB valve drivers 9180 9181 A 58 SBC PCB pump motor driver door detect 9180 9181 A 59 SBC PCB input amplifier 9180 9181 A 60 SBC PCB temperature circuit 9180 9181 A 61 SBC PCB analog selector and A D converter 9180 9181 A 62 SBC PCB microprocessor 9180 9181 A 63 SBC PCB microprocessor 9180 9181 A 64 SBC PCB Fluid pack detector 9180 9181 A 65 Lamp PCB 9180 9181 A 66 Display daughter PCB 9180 A 67 Software operation Part 6 Software operation 9110 Software operation 9110 B 5 Overview B 5 Overall program operation flow B 6 Main menu B 7 Measurement sequence B 8 Print functions menu B 9 QC sample menu B 10 Daily maintenance menu B 11 Operator functions menu B 12 Program instrument menu B 13 Ap
41. Analyzers 9 Software operation 9181 Software operation 9181 Automated measurement sequence NA K CL READY NO AUTOMATED MEASUREMENT Manual Probe moves to Measurement upper position Probe moves to Tray upper position applicable oup position Probe moves down to cup position Probe moves to upper position Turntable moves to mid position Probe moves to lower position No sample lt 2 No sample 2 NO SAMPLE o End of Run Sample in Process Na K CI Na KCI READY 151 54 114 applicable sample number Figure B 36 9181 automated measurement sequence Roche Diagnostics April 2007 Service manual Version 2 0 B 53 9 Software operation 9181 91xx pH Electrolyte Analyzers Software operation 9181 Print functions menu PRINT FUNCTIONS NO Print Last Sample Report Print Last Sample Report Print Cal Report Print QC Values and Statistics NO QC STANDARD URINE SAMPLE Na K Cl READY Remain in Print Functions Na K Cl READY Figure B 37 9181 print functions menu Roche Diagnostics April 2007 B 54 Service manual Version 2 0 91xx pH Electrolyte Analyzers 9 Software operation 9181 Software operation 9181 QC std dialysate urine sample menu DAILY URINE SAMPLE MAINTENANCE YES QC Level YES Probe moves to introduci Sasi Remove Sample Retract Probe gt Probe moves to Sample upp
42. B is identical to that of Standard A with the exception of the operation of solenoid valve B This solenoid valve is operated in reverse order of Standard A aspiration The process for Standard C is identical to that of Standard A with the exception of the operation of solenoid valve C This solenoid valve is operated in reverse order of Standard A aspiration April 2007 Service manual Version 2 0 A 27 4 Description of modules 91xx pH Electrolyte Analyzers Description of modules Cleaning solution 9110 only Reference solution Waste liquids The cleaning routine within the 9110 pH Analyzer is automatic with cleaning solution being aspirated through the system every 40th sample or every 24 hours Automatic operator initiated cleaning can also be performed by selecting ELECTRODE CLEANING from the main menu The cleaning routine is very similar to the aspiration of Standard or B with solenoid valve C being opened to allow cleaning solution to be aspirated through the fluidic path The timing of the cleaning solution routine allows for a larger volume of cleaning solution to be aspirated than the Standard or B cycles to ensure that cleaning solution covers the electrodes wash port fill port sample needle and preheater during the cleaning cycle The reference housing is filled automatically using the second winding of the peristaltic pump and solenoid valve R The reference solution connector allows for the refere
43. CHECK FLUID PATH except 9110 D 11 1111 LA Out of range D 11 7 Temparature out of range D 11 ERR D 12 PERFORM DAILY MAINTENANCE except 9110 D 12 REPLACE FLUID PACK D 12 CHECK TEMPERATURE 9140 9180 9181 only D 12 VALVE OVERTEMP CHECK VALVES D 13 ERROR UPPER NEEDLE SENSOR D 13 ERROR LOWER NEEDLE SENSOR D 13 WHEEL MISSING OR SAMPLER DEFECT D 13 SAMPLER JAMMED OR DEFECTIVE D 14 Glossary Part Glossary E 3 Index Part n 15 Index Versions Part 16 Versions Software G 5 Service Manual G 6 Version 1 0 G 6 Version 2 0 G 6 Roche Diagnostics April 2007 Service manual Version 2 0 7 91xx pH Electrolyte Analyzers Roche Diagnostics April 2007 8 Service manual Version 2 0 91xx pH Electrolyte Analyzers Preface This manual contains all of the information required for the maintenance and repair of the 91xx pH Electrolyte Analyzers The user must be familiar with the function and operation of the instrument to fully understand the processes described here amp For more information about instructions refer to 9180 Electrolyte Analyzers Instructions for Use Observe the service and repair procedures described in this manual and use only genuine Roche replacement parts and Roche approved materials to guarantee the full functionality of the 91xx pH Electrolyte Analyzers For the order numbers of replacement parts refer to 91xx pH Electrolyte Analyzers Spare Part List lt
44. Diagnostics 9110 calibration sequence April 2007 Service manual Version 2 0 B 15 6 Software operation 9110 91xx pH Electrolyte Analyzers Software operation 9110 Roche Diagnostics April 2007 B 16 Service manual Version 2 0 91xx pH Electrolyte Analyzers 7 Software operation 9120 9130 9140 Table of contents Software operation 9120 9130 9140 In this chapter Chapter Software operation 9120 9130 9140 a B 19 dau 19 Overall program operating flow sentent B 20 Main menu E M HS B 21 Measurement sequence B 22 Print functions menu oett 23 QC Standard urine sample menu eene B 24 Daily maintenance menu B 25 Operator functions menu enrian aa E N B 26 Program instrument 27 Service functions MENU 28 sequence titii itte B 29 Roche Diagnostics April 2007 Service manual Version 2 0 B 17 7 Software operation 9120 9130 9140 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 B 18 Service manual Version 2 0 91xx pH Electrolyte Analyzers 7 Software operation 9120 9130 9140 Software operation 9120 9130 9140 Overview Roche Diagnostics Overall Program Operating Flow Main Menu Measurement Sequence
45. Electrolyte Analyzers The user must be familiar with the function and operation of the instrument to fully understand the processes described here For more information about 91xx pH Electrolyte Analyzers refer to corresponding Instructions for Use Observe the service and repair procedures described in this manual and use only genuine Roche replacement parts and Roche approved materials to guarantee the full functionality of the 91xx pH Electrolyte Analyzers For the order numbers of replacement parts refer to 91xx pH Electrolyte Analyzers Spare Part List April 2007 Service Manual Version 2 0 3 91xx pH Electrolyte Analyzers amp For an overview of possible revisions and available software versions see Software on page G 5 Service Manual on page G 6 Copyrights 2007 Roche Diagnostics GmbH All rights reserved The contents of this document may not be reproduced in any form or communicated to any third party without the prior written consent of Roche Diagnostics Every effort is made to ensure its correctness Subject to change without notice Trademarks COBAS LIFE NEEDS ANSWERS ROCHE MICROSAMPLER SNAPPAK ISETROL are trademarks of Roche All other trademarks are the property of their respective owners Contact addresses Manufacturer Roche Diagnostics GmbH D 68298 Mannheim Germany www roche com Roche Diagnostics April 2007 4 Service Manual Version 2 0 91xx pH Electrol
46. For an overview of possible revisions and available software versions see Software on page G 5 Service Manual on page G 6 How to use this manual Keep this Service manual in a safe place to ensure that it is not damaged and remains available for use This Service manual should be easily accessible at all times To help finding information quickly there is a table of contents at the beginning of the book and each chapter In addition a complete index can be found at the end Where to Find Information In addition to the Service manual the following documents are also provided to assist in finding desired information quickly 91xx pH Electrolyte Analyzers Instructions for Use 9180 Electrolyte Analyzer Short Instruction 91xx pH Electrolyte Analyzers Service manual PDF or iSDoc version in GRIPS 91xx pH Electrolyte Analyzers Spare Part List PDF or iSDoc version in GRIPS Roche Diagnostics April 2007 Service manual Version 2 0 9 91xx pH Electrolyte Analyzers Conventions used in this manual Roche Diagnostics Symbols Visual cues are used to help locate and interpret information in this manual quickly This section explains formatting conventions used in this manual The following symbols are used Symbol Used for gt Start of procedure List item Cross reference Gn Call up software reference 1 Note All sections or text locations marked with NOTE de
47. Menu 6 Electrode Cleaning Menu 7 Operator Functions Menu 8 Program Instrument Menu 9 Service Functions Menu 10 Calibration Sequence Roche Diagnostics April 2007 Service manual Version 2 0 B 5 91xx pH Electrolyte Analyzers 6 Software operation 9110 Software operation 9110 Overall program operation flow U31vsugrvo SNOILONNA LNAWNNYLSNI SNOILONNA amp ONINV3TIO 80 35 ON HOl1VH3dO JON 3GOHl19313 ounj aS S3A ulewoy apop oung SIA Jeju3 ul dnjas Jejuud 1591 weu old 9 zsuld 51 198 AUOD 9unJ 1591 5 jewuon 1591 Jo 1no exer ON seDuey e 3497 ejduies ldues SO suoljouny 1959 1ooq ejduies eseBuey 9497 83A sonses pue 1591 OO 195 06 3 UIEWON 134971 5 20 9497 lduies 2 1591 o oo Z 19491 SD Ud s pono l3 1591 Bulue
48. Needed PLEASE WAIT Wash YE Clean Cont gt Needed 9180 A80X2 AA WARM UP ibrati Pack insertets New SnapPak Calibration Global installed a to Background check in all menus except Take out of Operation and Service Functions PRINT FUNCTIONS STATUS NO SnapPak CA READY CALIBRATION Figure B 45 9181 power up QC STD DIALYSATE URINE SAMPLE Pack in M DAILY Global MAINTENANCE YES NO NO SERVICE PROGRAM OPERATOR FUNCTIONS INSTRUMENT FUNCTIONS Roche Diagnostics April 2007 B 62 Service manual Version 2 0 91xx pH Electrolyte Analyzers 10 Service functions Table of contents Service functions Insert introducing text here In this chapter Chapter 10 Service 65 LEST ELECTRODES s Aash B 65 TEST SAMPLE SENSOR 4 222222 eni ea ea rE A E T a B 65 TEST SnapPak SENSOR 9180 and 9181 only B 65 TEST LANGUAGE SWITCH 9180 and 9181 only B 65 TEST SAMPLE DOOR except 9181 B 66 TEST PROBE 9181 only ine top ie B 66 TEST AUTOSAMPLER 9181 only B 66 TEST PUMP un RE ER RENI B 66 TEST PINCH VALIV BS eiie ca
49. Print Functions Menu QC Standard Urine Sample Menu Daily Maintenance Menu Operator Functions Menu Program Instrument Menu co N G Ch G N Service Functions Menu 10 Calibration Sequence Software operation 9120 9130 9140 April 2007 Service manual Version 2 0 B 19 91xx pH Electrolyte Analyzers 7 Software operation 9120 9130 9140 Software operation 9120 9130 9140 Overall program operating flow Ava al Ivo SNOILONNA LNAWNYLSNI SNOILONNA amp 3ONVN3LNIVIA d dl AvS SNOILONNA ON NOILVHSITVO ON 35 NVUDO0Ud oN JON ON 1 5 JON ONISF ZL Sunq 1 uiei amp o4d 19111 adwy amp dnjes Jejuug 1591 9 Z Suid dune 51 198 4102 Sund Joyesodo lduies uun p1s 1s L JOO u ESONIEN 5 jJeuuoyN uonei do ldues 1591 Jo 1no exer euufi sebuey jene ejduies ldues suoyoun4 198593 u uUleway c 9 97 91eq euir 193 9 dues ESONSNEIS pue 1591 195 Aeq u ureway 19491 20 Wd JOSUaS lduuesj sobuey
50. Roche Diagnostics This Service Manual contains vital warning and safety information This instruments are intended to be used only for the specialized purpose described in the instructions The most important prerequisites for use operation and safety are explained to ensure smooth operation No warranty or liability claims will be covered if the instrument is used in ways other than those described or if the necessary prerequisites and safety measures are not observed The instrument may be operated only by persons whose qualifications enable them to comply with the safety measures that are necessary during operation of the instrument Suitable protective equipment like laboratory clothing protective gloves protective goggles and if necessary mouth protectors must be worn to prevent direct contact with biological working materials In addition a face mask is required if there is a risk Adjustments and maintenance performed with removed covers and connected power may be attempted only by a qualified technician who is aware of the associated dangers Instrument repairs are to be performed only by the manufacturer or qualified service personnel Only accessories and supplies either delivered by or approved by Roche are to be used with the instrument These items are manufactured especially for use with this instrument and meet the highest quality requirements Operation of the instrument with solutions whose composition is not cons
51. Step Pulses enable IC U10 to provide the necessary control to driver IC U9 which supplies pump drive signals PHASE A through PHASE D to control peristaltic pump operation IC 11 operates with signal DCLOSED from the sample door light gate to provide signal DOOR which detects the status of the sample door open or closed IC U13 U14 and U15 9110 9180 9181 U13 only provide high to low input impedance matching for each Electrode input Offset generator IC U12D and amplifiers IC U12A U12B and U12C provide the necessary gain and offset for each electrode channel TP9 TP10 and TP11 provide test points for the Na TP9 pH on the 9110 Kt and Cl Ca Li channels respectively 9110 only IC U300 is used as the input amplifier for the reference electrode IC U12A is used for the subtraction of the reference electrode signal from the pH electrode signal For sample detection the reference electrode is switched to ground relay K300 during aspiration This circuit is used in the 9110 pH Analyzer to amplify the signal of the temperature sensor in the preheater This is an optional circuit which is used in the 9140 9180 and 9181 The circuit creates a temperature dependent voltage which is converted by software to degrees Celsius R54 can be adjusted for correct temperature display Analog inputs from the input amplifiers are selected via signals ADSELO through ADSEL2 and are multiplexed by IC U19 This multiplexed signal is then fed to the
52. The calculation of the Install before date is based on the production date of the elctrode Danger symbol T Toxic on the packaging of the reference electrode Rating Inhalation swallowing or skin contact with even small quantities can lead to serious health risks including fatal risks Characteristics of this type of item are severe possible irreversible T R23 33 50 53 damage to health through repeated or prolonged contact particular with carcinogenic genetic or reproductive danger to reproductive capabilities effects Caution Avoid any contact with the human body If you feel unwell contact a doctor immediately Any substances with carcinogenic genetic or reproductive dangers are indicated appropriately Always observe the regulations when handling such substances Danger symbol N Dangerous to the environment on the packaging of the reference electrode Rating If released into aquatic and non aquatic environments can cause immediate or delayed damage to ecosystems through a change N 45 60 61 7 in environmental conditions These substances or their by products can cause simultaneous damage to sensitive environmental areas Caution Depending on the potential for damage do not allow the substance to enter sewers soils or the environment Observe the specific disposal regulations Danger symbol Xi Irritant Rating Although not corrosive momentary longer lasting or repeated contact with skin or mucous m
53. continuous operation IP 20 the device is not designed for operation in explosive environments CSA CE FCC Class B CLIA Complexity Category Moderate Test System Code 04739 April 2007 Service manual Version 2 0 A 15 2 Specifications 91xx pH Electrolyte Analyzers Specifications FDA 510 k numbers K932642 9110 K961458 9180 K972673 9181 Analyte Codes 5805 Kt 4910 Cl 1018 Ca 1004 Li 3712 Special requirements for IVD 61010 2 2001 12 products 61010 2 2002 01 9180 SN gt 12001 only Y Data subject to change without notice Technical information is supplied for general informational e purposes only Roche Diagnostics April 2007 A 16 Service manual Version 2 0 91xx pH Electrolyte Analyzers 3 Interface specifications Table of contents Interface specifications In this chapter Chapter Interf ce specihieations asian a u qq 19 Interface information sieisen N n a u aan n a A 19 SONWALE SS L a ends 19 Example data string information s ss sansa aaa sasa a es A 20 Automatic Calibration Report A 20 Serumiisample report zoe oe eO II Ie etta A 20 Data Link Information 9180 9181 only A 20 Roche Diagnostics April 2007 Service manual Version 2 0 A 17 3 Interface specifications 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 A 18 Service
54. gt K 5 4 mmol L cr crlf Cl 122soh mmol L lt cr gt lt crlf gt lt cr gt lt crlf gt PERFORM DAILY cr crlf MAINTENANCE lt cr gt lt crlf gt lt cr gt lt crlf gt lt ex gt Data Link Information 9180 9181 only Roche Diagnostics The data link with the COMPACT 2 3 blood gas analyzer allows to combine ISE results with pH blood gas results on one printout If Ca is activated on the 9180 9181 a pH corrected Ca value will be calculated and printed on the combined sample report For connection ofthe 9180 9181 analyzer to the COMPACT 2 3 analyzer the optional Interface Kit is required To install the kit first turn both instruments off Connect the interface filter provided in the kit to the RS232 port on the 9180 9181 Then connect one end of the cable to the interface filter the other end to the COM 2 port on the COMPACT 2 3 On the COMPACT 2 3 select 9180 under the COM 2 interface options amp See COMPACT 2 3 Operator s Manual for details April 2007 A 20 Service manual Version 2 0 91xx pH Electrolyte Analyzers 4 Description of modules Table of contents Description of modules Insert introducing text here In this chapter Chapter n Descriptiomnoft modules eee A 23 Mechanical assemblies aan diete p Feri operi tede te eieo A 23 Front door assembly 23 Needleunit assembly A 23 Electrode holder assembly A 24 P
55. lt See the 91xx Spare Parts List for a detailed reference of the parts of this assembly remove the 9110 fill port for replacement or cleaning grasp the white tab on the fill port and rotate the bayonet style connector counter clockwise Gently pull the fill port forward to remove Replacement can be accomplished in the reverse order During removal be careful not to bend the needle The 9110 fill port assembly is desinged for easy removal without the use of tools for cleaning changing tubing or replacement First disconnect the sample intake tube on the left side of the fill port assembly from the sample preheater Grasp the fill port assembly and pull it forward carefully Disconnect the wire from the needle arm To reassemble reconnect the wire press the fill port assembly into place and reconnect the tubing The 9110 wash port located on the underside of the flap can be removed by appying pressure to one side to overcome the snap fit To reinsert align wash port and apply pressure to snap into place The 9120 30 40 80 and 81 sample needle and fill port are both designed for easy replacement by the user To remove the fill port assembly for replacement pull the fill April 2007 Service manual Version 2 0 A 23 4 Description of modules 91xx pH Electrolyte Analyzers Description of modules Electrode holder assembly Peristaltic pump assembly Roche Diagnostics port holder towards the front for easy ac
56. pee 63 April 2007 91xx pH Electrolyte Analyzers 6 Software operation 9110 Table of contents Software operation 9110 In this chapter Chapter 6 Software operation 9110 B 5 OVeEVIEW 3 5 Overall program operation B 6 Main men anaiei e B 7 Measurement sequence B 8 Print furicti ns menu uA B 9 QC sample menu nn Reese B 10 Daily maintenance menu yp uta Sana hka a saus B 11 Operator functions menu onrars ane a AN B 12 Prosram instrument e e e B 13 Service functions MENU sein Bee B 14 Galibration sequence titii etie B 15 Roche Diagnostics April 2007 Service manual Version 2 0 B 3 6 Software operation 9110 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 4 Service manual Version 2 0 91xx pH Electrolyte Analyzers 6 Software operation 9110 Software operation 9110 Software operation 9110 Overview Software operation of the 9110 pH Analyzers can be accessed by YES NO selection on the analyzer front panel The following flow chart diagrams are provided to identify operating sequences of the 9110 pH Analyzers 1 Overall Program Operation Flow 2 Main Menu 3 Measurement Sequence 4 Print Functions Menu 5 QC Standard Urine Sample
57. qpe 1 1 asns e 2 13931 77 X 24 manawa vasvHd gt ks Fino tnt EE et 7711 77 m voorn C 07 s sa sa anos 5 anos anos anro anto SA SA SA SA Sto evo WO ao 1 90 gt gt daisdwna SBC PCB pump motor driver door detect 9110 9120 9130 9140 Figure A 10 April 2007 Roche Diagnostics A 45 Version 2 0 Service manual 91xx pH Electrolyte Analyzers SBC PCB input amplifiers 9110 9120 9130 9140 5 Electronic diagrams Electronic diagrams ls s 2002 Zz eunr Aeps ny ed XX 6 GVO NIYIN NN 1u tunooq WHOLWVHSHNHD LASAAO 4Z8NL NOBLZOUL sia vOGENZ 0080 eni AINO 6 9 SGNANO4WOD SHINSS 00 qos gaq z uaavaH ODA SI T Hd NEGO 1545 diS S gt ooe anro azo a S meme ne AQUA T lo 0069 Sd zoe m 5 Yl AL c z T S Jeu 00 i Loew ee NO64201L Lo i gt z T 5 kaa
58. respective DC supply voltage Test point Voltage Fuse Rating LED 1 24 VDC F3 1 25 AT D3 2 5 VDC F1 F2 2 0 AT 1 6 AT D1 D2 TP3 12 VDC F4 0 3 AT D4 TP4 12 VDC F5 0 3 AT D5 5 Ground Table A 5 Voltage test points LEDs Test point Voltage 6 8 VDC 9180 9181 5 VDC 7 122 12 VDC TP8 8 VDC 9180 9181 5 VDC TP9 Na channel pH channel on the 9110 TP10 K channel not used on the 9110 Ca CI channel 141 9180 9181 only not used the 9110 TP12 Temperature TP13 10 000 V ref voltage 14 2 500 V ref voltage 2 500 V the 9110 15 Sample sensor sample contact voltage the 9110 TP16 22 12 5 VDC 22 16 VDC on the 9110 Table A 6 Voltage test points April 2007 Service manual Version 2 0 A 33 4 Description of modules 91xx pH Electrolyte Analyzers Description of modules Valve drivers Pump motor driver cover sensor Input amplifiers Temperature sensor Analog channel selector and A D SnapPak Sen Lamp board Roche Diagnostics converter Microprocessor sor 9180 9181 only The solenoids are activated by transistors Q9 Q13 To open the solenoid a 0 5 second pulse of 22 V is applied then the voltage drops down to 12 5 V 16 V on the 9110 to hold the solenoid open The pump driver circuit IC U9 and IC U10 provide the drive and control for the peristaltic pump Signals PUMPEN Pump Enable and PUMPSTEP Pump
59. screws located on the front of the housing near the peristaltic pump April 2007 A 24 Service manual Version 2 0 91xx pH Electrolyte Analyzers 4 Description of modules Solenoid assembly Printer assembly 9 5 Rear panel assembly Roche Diagnostics Description of modules Removal of each solenoid is identical Each solenoid valve has a removable pressure piece which is held in place by a solenoid cap To remove the pressure piece locate the arrow on the solenoid cap and remove the solenoid cap by sliding it off the solenoid shaft in the direction of the arrow This exposes the pressure piece which can now be removed from the solenoid shaft To replace pressure piece and cap energize the solenoid as described in TEST PINCH VALVES on page B 67 This extends the solenoid shaft to the outermost position for ease in replacing pressure piece and cap Removal of the pressure piece and cap is required prior to removing the solenoid and allows for easier replacement of the tubing under each solenoid remove the solenoid assembly the rear panel and the SBC Board described in Rear panel assembly on page A 25 and SBC board on page A 26 must be removed first lt For details see Rear panel assembly on page A 25 SBC board on page A 26 Each solenoid has an electrical connection to the lamp board which must be unplugged prior to removal of the solenoid At this point remove the two screws on the front p
60. see Rear panel assembly on page A 25 SBC board on page A 26 Pull the two pump windings at the peristaltic pump tube set off the pump roller and remove the white reference solution tube connector Remove the white TPR reference solution block by grasping the front of the block and pulling forward The three tube connections are labeled with green blue and red tags All remaining tubes can now be removed and the entire tubing harness can be discarded Fit the replacement tubing harness and assemble in reverse order as disassembly Reference the Tube Diagram on the next page for correct tube locations The intake tube set comprises three pieces of tubing A short piece red tag fits between the white T piece and the standard contact The long piece can be positioned between the standard contact and the wash port The remaining tube can be fitted between the sample needle and the preheater tube April 2007 A 28 Service manual Version 2 0 91xx pH Electrolyte Analyzers 4 Description of modules Description of modules Sample Contacts 9110 only The 9110 fluid detection is based on the principle of measuring the impedance in a sample contact path There are three sample contacts in the 9110 pH Analyzer Sample Contact 1 senses standard solution draws Roche Diagnostics April 2007 Service manual Version 2 0 A 29 91xx pH Electrolyte Analyzers 4 Description of modules Description of modules YOLDINNOD
61. sensor response When clear fluid e g water is aspirated the sample sensor display must indicate an increase of of at least 40 units For blood samples not transparent the reading should decrease by at least 40 units The sample contacts can be adjusted see cross reference below the sample sensor is calibrated with air during each calibration For details see Sample contacts adjustment 9110 only on page C 5 Perform a calibration to remove the error message after troubleshooting Roche Diagnostics April 2007 D 6 Service manual Version 2 0 91xx pH Electrolyte Analyzers 13 Troubleshooting Error messages and troubleshooting Possible causes and remedies Check for correct drying of measuring chamber during the wash cycle Replace peristaltic pump tube set 9110 Check for clots and leaks in the sample path especially in the fill port area Clean the sample contacts using Electrode Cleaning procedure Test and adjust sample contacts to correct adjustment Replace sample contacts as needed Note Q Sample contacts can be tested by invoking TEST SAMPLE CONTACTS within SERVICE FUNCTIONS If reading is low disconnect each sample contact and notice a change in the displayed result to isolate a specific contact fault 9120 30 40 80 81 Clean sample sensor using Daily Cleaning procedure Test and adjust sample sensor to correct adjustment Replace sample sensor asse
62. shielding The total resistance should not exceed 0 5 ohms Be sure to consider the resistance of the ohmmeter s connecting cables Note Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance the printouts display messages and data links sequences of the 9180 Electrolyte Analyzers serial number gt 12001 will show Roche instead of AVL This also applies to 9180 Electrolyte Analyzers serial number lt 11999 in case of replacing the main board MAINBOARD 9180 gt SN 12001 or the program set PROGRAM SET 9180 April 2007 A 26 Service manual Version 2 0 91xx pH Electrolyte Analyzers 4 Description of modules Lamp board Display board Fluidic Module Standard A 9110 pH Analyzers 9120 30 40 80 81 Electrolyte Analyzers Standard B Standard C not on 9110 Roche Diagnostics Description of modules The rear panel and SBC board must be removed to gain access to the lamp board Remove one screw and washer which hold the board in place and remove the electrical connectors to remove this board The rear panel and SBC board must be removed to gain access to the display board Remove three screws and 50 degree spacers and unplug the preheater or sample sensor cable from the front and the ribbon cable to the SBC board to remove the display board Standard A is drawn to the electrode module by vacuum provided by the peristaltic pump When Standard A is to be aspirated i
63. the operator to run the instrument in the economy mode The instrument will enter the STANDBY MODE This occurs at the time of the next 2 point calibration when no samples have been analyzed since the last 2 point calibration FIF IFI Allows printing of electrode voltages during printout of measurements and calibrations not available for 9140 MEL 9180 9181 only This service code will print the last 20 error messages MGL LGM This service code is available only in the model 9140 9180 9181 analyzer and enables values to be displayed in mg dL The normal mode displays values in mmol L QC and Standard results are always displayed in mmol L NOB BON 9180 9181 only This service code disables the beep acc cca When this code is enabled QC values will be reported in direct ISE measurement instead of flame equivalent values In this mode correlation factors are NOT applied to the measured values SFC 9180 9181 only This code allows the input of a percentage for the ISE SnapPak counter SLC CLS 9110 9180 9181 only When this code is set the current setting of the leasing counter which cannot be reset by the operator is printed on the calibration report SSC css 9181 only This code toggles needle height to accommodate smaller 9181 only sample cups In the default setting the analyzer works with 2 0 mL cups To select 0 25 mL sample cups SSC has to be activated To return
64. 2 Temperature Adjustment 9110 9140 9180 9181 only C 6 Trademarks 4 Troubleshooting D 5 Status LEDs D 5 Waste liquids A 28 Roche Diagnostics April 2007 F 6 Service manual Version 2 0 Versions G 16 7 VerSIOHST De PA REE ae G 3 April 2007 91xx pH Electrolyte Analyzers 16 Versions Table of contents Versions This chapter provides information about the released software versions and the released service manuals In this chapter Chapter 16 SOR G 5 Service B d UO EORR EGG G 6 G 6 20 G 6 Roche Diagnostics April 2007 Service manual Version 2 0 G 3 16 Versions 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 G 4 Service manual Version 2 0 91xx pH Electrolyte Analyzers 16 Versions Software Software Version Nr Release Date Remark A10X1 18 _ 9110 only A20X3 18 _ 9120 only A30X2 18 _ 9130 only A30X2 24 _ 9140 only Table G 1 9110 9120 9130 and 9140 software releases For the 9110 9120 9130 and 9140 software substitute X by one of the following language codes English E French F German D Italian 1 Japanese J and Spanish 5 Version Nr Release Date Remark A80X1 00 _ initial version for 9180 only A80X2 00 _ 9181 software included A80X2 04 _ minor bug fixes and text changes
65. 2 SET SAMPLE Sel Parameter Sel Parameter re you sure NUMBER TO 0 NO Number Na Li ok NO INal Li ok YES Take out of Shutdown Plug YES Transp Housing YES Remove Water YES Operation Water installed Ref installed and Press YES Remain in en Do You Wish to Pump Windings 5 Relief Clamps Operator Func Pewee d YES DELETE all Data YES Relieved YES Inserted YES NAK Sel Parameter i Sel Parameter s Sel Parameter Na K 1 ok Na CI ok Na K Ca ok Select Phrameter Configuration All Electrodes amp Plug Removed Figure 40 9181 operator functions menu Roche Diagnostics April 2007 Service manual Version 2 0 B 57 9 Software operation 9181 91xx pH Electrolyte Analyzers Software operation 9181 Program instrument menu PROGRAM NO SERVICE INSTRUMENT FUNCTIONS YES Enter Code TCODE ERROR AAA RETRY NO Program QC m Current Lot 0000 Print old Values me New Lot YES Enter last 4 Dig Level 1 Ranges Change Lot and Statistics Delete old Data of Lot 0000 Additional Na low 040 YES Enter last 4 Dig Parameters Na high 205 ok of Lot 0000 NO YES Calow 050 YES high 200 Lilow 04 YES Li high 6 0 ok Program QC YES Level 2 Ranges See Level 1 NO YES Level 2 Ranges See Level 1 NO Program YES Nal
66. 2 Temperature Control 9140 9180 only Electrode T 5 Sample Sensor 9140 9180 only peux Analog Selector and Printer Driver Sample Sensor Temp Sensor A D Converter JP1 Printer Assembly Microprocessor Power Supply 22VDC 10VDC 5VDC 5VDC 9180 8VDC 8VDC 22VDC 12 5VDC Peristaltic Pump Driver Pump Motor 6 Lamp Board Sample Door JP7 Solenoid Valve Drivers Interface Driver Valve A JP1 Valve B JP2 gt JP3 Valve C JP3 gt Valve V JP4 gt Valve R JP5 gt JP1 100 240 V Power Receptacle Power Supply Module u pl Fuse F1 2 5A 250V 5VDC 24VDC Sil 12VDC 12VDC Figure A 5 9120 9130 9140 9180 analyzer block diagram Roche Diagnostics April 2007 40 Service manual Version 2 0 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams 9181 analyzer block diagram SBC Board Preamplifier Na Electrofle Cl Ca Li Amp Reference Electrode Analog Selector and A D Converter Fluid Pack Detector Sensor Display Driver Temperature Control Printer Driver Needle Ctrl Lines Microprocessor Turntable Ctrl Lines JP1 Display Boa Needle Mechanism Assembly L JP1 Sample Sensor Temp Sensor Turntable Assembl Print
67. 2670 Cl 1093 2500mV 2496 Temp C 26 2 NO NO NO Enter Service Code NO YES Enter Code Is Code w Code uw Enter Other NO Accepted ok Ea valid Service Code Code Error Retry ini NO YES kusasqa Figure 29 9180 service functions menu NAKCA READY Roche Diagnostics April 2007 B 42 Service manual Version 2 0 91xx pH Electrolyte Analyzers 8 Software operation 9180 Software operation 9180 Calibration sequence 1 Example Na K Li calibration CALIBRATION YES Calibration in Process 124 Clear Sample Path Calibrate Sample Contacts Asp Std A into Meas Chamber Standard A detected Obtain Electrode Voltage Std A 3 Std A Values obtained STANDARD A NOT DETECTED CHECK SAMPLE CONTACTS Are Values reproducible 6 Aspir Of Std A performed Standard B detected STANDARD B NOT DETECTED CHECK ELECTRODES Calc of Std A Std B slope CHECK ELECTRODES Pressurize Ref Aspl Std A into Meas Chamber Figure B 30 9180 calibration sequence 1 Example Na K Li calibration Roche Diagnostics April 2007 Service manual Version 2 0 B 43 8 Software operation 9180 91xx pH Electrolyte Analyzers Software operation 9180 Calibration sequence 2 Example Na K Li calibration Asp Std A into Meas Chamber
68. 34 wauw gt gt N3asia HHL LV qvos asia ATIVOGIAIGNI Now SI qNSSTaNVS gt gt Sniviva 55 1SVHINOO gt gt Das gt gt a3481awvs 5 amv Nizezsu sts Lnozezsu 515 GaMOTTY S VIA ON HOVNL amp LUASNI sOVuL LND AINO TAY HOA L08 0S 60 X IOW HLIM SALUm MIL oz anos gt gt anss z 0202 09 92 bas gt rt 1 09929 anov z T 0 1068 3a H2S 68W GNOOND WALSAS Tun HDS 9mm 0068 HOS L8W HOS Saw HOS Saw OL SANNOND MOTIV OL MAY HOS pan LSIXA ATTWOISAH LON E064 0064 HOS HOS Zan HOS Taw NNIT System interconnect 9110 9120 9130 9140 Figure A 7 April 2007 Roche Diagnostics Version 2 0 Service manual A 42 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams SBC PCB power supply 9110 9120 9130 9140 s p z SE E Jnio 500542 x TT 619 wawnsog E K R XX L
69. 4 5 Yo i Y a yo vo 2 T Figure A 17 Lamp PCB 9110 9120 9130 9140 Roche Diagnostics April 2007 A 52 Service manual Version 2 0 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams ALDIBISV33 304 NOISH3A qQuvosqvaus Z66L 834 119 dWalHd ISY34 HO4 NOISH3A aut Fe eu AV 1dSIQ A01 rama z pe z wg EOE COUSIN a 29 tequinn weumsoa 0116 AV1dSIq Hd ava 330108 gt anro anro T 29 50 u ava uaa1os mez vs wASt 99 T z T ozz ven oey m oseng T z Ur ieu D as zeu T z een azn izy I oe 5 g z T DES XION Mz Vs z ra Is 9 N ZUOISWHSHI 5 zi PERPE LSOISINSSHL 5 PPR NOLON L3 2 azn z T 2 T LE an D ore ses b n 7 Ws 005 89 GA tzen z T 2 019 01 any BERN or ES 5 T zs z
70. 5 T T OL 8 ou S du ho i ho S L gt gt LNOdWAL gt gt gt gt LNON April 2007 Version 2 0 Service manual SBC PCB analog selector and A D converter 9110 9120 9130 9140 Roche Diagnostics Figure A 13 A 48 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams 9110 9120 9130 9140 microprocessor 5 5 30 g 19905 2002 2 sed 3 006 2 v XXL6 GHVOS OLLVIN3HOS 0N3NV um gt gt Na oud A 55 rs ramal sar 21522 vozo aoen Ms pon n ok yi NO 29 29 39 poe er o z 6 255 stv 5 155 155 viv sen zw 0008 Liv 217 uv 217 uv LE 21 ez l 5 en ANI ov
71. 6 NIVIN OLLVINSHOS oe anu H T 15 Sno 819 4 49 x so x no EI Li L L L L m ERAS 4 4 4 Age s T F I F F wo L 02 695 890 199 995 592 zo 65 L 10 I l In RS AQL ANOL diioasna diToasna 429 9 H H L 01179 NU QUA anoasna drossna Sogn en sano wAS u dnoasna dnoasna LLO SEL 84 is sano dnoasna dnoassna r sano anyo A9LIANOL anyo 59 829 eo mare A zum tano ATI z FN W zn 4V0 SELI uA8 u Sal 4V0 SELI N ZL0 SELN 9 em E g ES ZN sSa AZT a q31 095 E asz anozy i z4 z L 21006 0 E zu DRS so FT sa 1vO0 0 O 110 6 11 ES 2 jp sPaA AZT a 095 sAZT u e ASZ AN0Lr Ir 569 m z L L sLVO008 05 T u f 28 1VO0E 0 Ao t Es EN vd 7 5 18 gt 08 e F Q31 Atu O vH anyo 99 m 19 Aosianoozz INSZ Ta im 620 124 S o z 1 NIA Over zo Tosh ONvE d ZEW 4 in
72. 6168 p were 0264 April 2007 A 65 SBC PCB Fluid pack detector 9180 9181 Roche Diagnostics Service manual Version 2 0 Figure A 30 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams Lamp PCB 9180 9181 S88zd NIH T s T n seezdWiH t PH AWAUE AAT sr AWAUE QAT sr vi t 6 8 9 5 v L Jo i 19909 2661 Te Aenuer Aepuj vLosaz juewnoog 1816 0816 GYVOd OLLVW3HOS enu Was Auer 6 0 9 01 div 6 8 T 9 WSINVHOSNW SIGN L 5 wood rama dasoiod 0310 GasvHd S3SVHd V3SVHd 1 3 BSN YANA 2 0 9 01 dW 2 0 9 01 dV Z 0L9 01 dW 2 0 9 01 diNv Z 0L9 01 diNv diNSLHSAO g z 2 z z z L L L L L Sdf tar FATWA A FATWA FATWA FATWA V FATWA Lamp PCB 9180 9181 Figure A 31 April 2007 Roche Diagnostics Version 2 0 Service manual A 66 91xx pH Electrolyte Analyzers 5 Electr
73. 80 9181 LL 3o S jesus 6661 EO Kew Aepuow ewed 510547 v Jeqwnn 1816 0816 GYVOd NIYIN OLLVINSHOS OSZNL 5 gt z T gt gt anne gt gt anos 1 5 T gt gt SN3dO Aos danz z i n or A 0083rW am gt gt A0S dnz z T J z gt gt N3dO ela AOS ANSS gt gt B 16 96 E oosarw g T gt gt zN3dO S3ATVA gt T 4 z 265 RN gt gt IN3dO AOS ANZ 2 vaN X T 16 z 9 SHHAIHG FATWA d ho vOGENZ lt 80 22 9 wAS 2T AZT u 002 anvo 8 4 3ATVA gt vC SBC PCB valve drivers 9180 9181 Figure A 23 April 2007 Roche Diagnostics A 58 Version 2 0 Service manual 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams door detect 9180 9181 iver SBC PCB pump motor dr gt O3SVHd gt 3SVHd gt vasvHd gt MEE 99 LoaLad HOLIMS wood
74. 91xx pH Electrolyte Analyzers Service Manual 5 is trademark of Roche 2007 Roche Diagnostics Roche Diagnostics GmbH D 68298 Mannheim Germany www roche diagnostics com 91xx pH Electrolyte Analyzers Version History Edition notice 91xx pH Electrolyte Analyzers Service Manual 91xx pH Electrolyte Analyzers Intended use Roche Diagnostics Manual Version Software Version Version date Changes 1 0 n a July 2004 First edition 2 0 n a April 2007 New template Language Order Number English 03077063001 amp For more information see Software on page G 5 Service Manual on page G 6 This Service Manual is for the maintenance and repair of 91xx pH Electrolyte Analyzers 9110 pH Analyzers and 9120 9130 9140 9180 and 9181 Electrolyte Analyzers Every effort has been made to ensure that all the information contained in this manual is correct at the time of printing However Roche Diagnostics GmbH reserves the right to make any changes necessary without notice as part of ongoing product development Any customer modification to the instrument will render the warranty or service agreement null and void Software updates are done by Roche Service representatives 9180 Electrolyte Analyzers SN gt 12001 are IVD compliant This manual contains all of the information required for maintenance and repair of 91xx pH Electrolyte Analyzers 9110 pH Analyzers and 9120 9130 9140 9180 and 9181
75. A D Converter IC U20 IC U21 supplies the necessary reference voltage 2500mV for the A D Converter and can be measured at Microprocessor IC U31 EPROM IC U25 9180 9181 U40 and MOSTEK Real Time Clock RAM IC U22 form the components of the central processing system IC U23 Address Demultiplexer IC U24 Programmable Array Logic PAL and IC U28 provide serial data buses 9180 9181 U23 U24 and U28 not present IC U34 through IC U38 convert serial data to parallel data and provide printer data display data solenoid valve control and peristaltic pump control 9110 U34 not present This fashion of serial to parallel data conversion is used to prevent analyzer lock up by peripheral devices The RS232 serial interface is a direct input output from the microprocessor utilizing interface driver and receiver U29 and U30 Optical detector 1501 provides signal indicating if the ISE SnapPak is in place This signal is fed to U41 which converts the light gate input to a digital logic output indicating SnapPak position IN or OUT The lamp board contains the LED array used to illuminate the measuring chamber and the interconnectors used for all solenoids the peristaltic pump door sensor and sample turntable 9181 only A ribbon cable from the SBC Board supplies the electrical connection for each signal through connector JP6 April 2007 A 34 Service manual Version 2 0 91xx pH Electrolyte Analyzers 4 Description of modules
76. A80X2 05 17 May 2004 9180 81 SN gt 12001 only AVL replaced by Roche minor text changes Table G 2 9180 9181 software releases The 9180 and 9181 software was published as multi language version Roche Diagnostics April 2007 Service manual Version 2 0 G 5 16 Versions 91xx pH Electrolyte Analyzers Service Manual Service Manual Version 1 0 Version Nr Date of Release Modified Chapters Applicable from Software Version 1 0 July 2004 First edition A80X2 05 Table G 3 Version 1 0 release information Version 2 0 Version Nr Date of Release Modified Chapters Applicable from Software Version 2 0 April 2007 all n a Table G 4 Version 2 0 release information Chapter Page Modification Addition all FrameMaker Templates applied for usage with iSDoc Preface 10 symbols updated A 5 important information protective equipment updated A 15 nominal power consumption modified A 20 data link information 9180 81 only new B 69 deleting data new D 5 Status LEDs new Glossary Versions new Table G 5 Version 2 0 modifications Roche Diagnostics April 2007 G 6 Service manual Version 2 0
77. AA RETRY NO Program QC mE Current Lot 0000 Print old Values E New Lot YES Enter last 4 Dig and Statistics Level 1 Ranges Change Lot Delete old Data of Lot 0000 Na low 040 Enter last 4 Dig Na high 205 ok of Lot 0000 NO low 050 YES Lilow 0 1 Ca high 200 ok Lihigh 6 0 ok Additional Parameters YES YES Cllow oso YES Cl high 200 ok Progam YES Level 2 Ranges See Level 1 Program YES Level 2 Ranges See Level 1 NO Program YES Na low 135 0 YES 3 50 YES Normal Ranges Na high 150 0 Khigh 8 00 NO YES Program YES Reset Corr Factor Input Verify Remain in Na b 00 0 YES K b 00 0 Corr Factors default Corr Factors Corr Factors Na m 1 0000 1 0000 NO NO YES YES Program Bicarb YES Reset Bicarb Input Verify Remain in Na b 00 0 YES K b 00 0 Corr Factors Factor default Bicarb Factors 22 Na m 1 0000 K m 1 0000 NO NO YES YES Program Acetate See Bicarbonate Corr Factors NO Program Urine MES See Bicarbonate Corr Factors NO YES YES Program Printer YES Patient Report NO Print two NO Automatic NO Enter Comment YES Enter Comment Setup off EM Patient Reports NE Cal Report Line 5 Program YES Activate Data Interface Link NO YES NO Remain in NO READY 12 45 YES Program Fu
78. AGE RANGES 9120 9130 9140 Electrode Standard A Standard B Allowable Difference A B 9120 9130 9140 Nat 600 to 2400 1600 to 2200 720 to 2000 250 to 680 700 to 1000 2500 to 500 520 to 1800 470 to 120 3100 to 100 1100 1800 520 to 1200 not used Ca 2600 to 1400 2100 to 2700 not used 480 to 900 Table D 2 9120 9130 9140 electrode voltage ranges ELECTRODE VOLTAGE RANGES 9180 9181 Electrode Standard A Standard B Standard C Allowable Difference A B A C Nat 600 to 2400 1600 to 2000 600 to 2400 250 to 680 50 to 50 700 to 1000 2500 to 500 700 to 1000 470 to 1200 40 to 40 Cr 3100 to 100 1000 to 3000 3100 to 100 370 to 860 not used Ca 3100 to 1000 2300 to 2500 3100 to 1000 350 to 660 150 to 150 Lit 3100 to 1900 3600 to 1400 2600 to 3400 1 to 760 1730 to 285 Table D 3 9180 9181 electrode voltage ranges Roche Diagnostics April 2007 D 10 Service manual Version 2 0 91xx pH Electrolyte Analyzers 13 Troubleshooting Error messages and troubleshooting CLOG IN SAMPLE PATH 9110 only CLOG CHECK FLUID PATH except 9110 After completion of sample measurement or calibration the analyzer will monitor the sample sensor to determine if the sample path has been cleared This message will be displayed if the sample path has not been cleared Possible causes and remedies Check for leaks in the fill port and electrode area Check for blockages i
79. C values and statistics will be deleted and the sample number reset to 0 QC and normal values as well as correlation factors printer settings date time and liquid content of the SnapPak will be reset to default values All service codes will be deactivated It is recommended that all data be deleted if the analyzer is to be used by other persons at a later date Warm up bypass bypass the instrument warm up 1 Press YES and NO keys simultaneously during instrument warm up Roche Diagnostics April 2007 B 70 Service manual Version 2 0 Maintenance LI MDUSUHPUS es apu C 3 12 MdaInteHaHnce 7 April 2007 11 Adjustments 91xx pH Electrolyte Analyzers Table of contents Adjustments Insert introducing text here In this chapter Chapter 1 5 Sample contacts adjustment 9110 only C 5 Sample sensor adjustment 9120 9130 9140 9180 9181 only C 5 Temperature Adjustment 9110 9140 9180 9181 only C 6 Roche Diagnostics April 2007 C 3 Service manual Version 2 0 11 Adjustments 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 4 Service manual Version 2 0 91xx pH Electrolyte Analyzers 11 Adjustments Adjustments Adjustments The 9110 20 30 40
80. CLOSED Blood Sample in Progress 65 Na K 151 54 114 YESINO 1 min 1 min Figure B 13 9120 9130 9140 measurement sequence Roche Diagnostics April 2007 B 22 Service manual Version 2 0 91xx pH Electrolyte Analyzers 7 Software operation 9120 9130 9140 Software operation 9120 9130 9140 Print functions menu PRINT FUNCTIONS NO Print Last Sample Report Print Cal Report NO Print QC Values and Statistics NO Remain in Print Functions Figure B 14 9120 9130 9140 print functions menu QC STANDARD URINE SAMPLE READY 12 45 READY 12 45 Roche Diagnostics April 2007 Service manual Version 2 0 B 23 7 Software operation 9120 9130 9140 91xx pH Electrolyte Analyzers Software operation 9120 9130 9140 QC standard urine sample menu QC STANDARD NO DAILY URINE SAMPLE MAINTENANCE YES QC Level 1 YES Open Door and OPEN PAMPLH Wipe Probe Close CLOSED Level 1 in Introduce Sample Introduce Sample Sample Door Progress 49 NO VALUES Store pH REJECTED Values 7 380 2 min YES 28 VALUES ELECTRODE CLEANING QC Level 2 YES Open Doorand OPEN Sania SAMPLE wipe Probe Close CLOSED QC Level 2 in Sample Introduce Sample SS Sample Sample Door Progress 49 2 min NO VALUES Store ar pH REJECTED Values 7 380 2 min YES 25 VALUES ELECTRODE ACCEPTED CLEANING QC Level 3 YES Open Door and OPEN introduce Sarnia PAMPLH Wi
81. D converter 9180 9181 Roche Diagnostics Figure A 27 A 62 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams 9180 9181 microprocessor SBC PCB ZIILZVIOA NON Iu E Tssus 6661 E0 ACW Repuon SIE sLosaz v ro Jaquinn S anal gt gt anoa 1816 0816 GHVOGS NIYIN OLLVWSHOS anuj goen abba 999 gt gt s19 T z T aan AN 5561 ET iT m 68 voen z TTA 59 gt gt Nizezsu E je A 4 gdz 099 859 cov I x HABIL Tm z Si oh 1nozezsu Wax 3SS 2
82. EN ISO 780 1997 Symbol Description 8 30 2 e E Q s 8 31 P 8 This product complies with the requirements in the directive 98 79 EC on in vitro diagnostic medical devices Lot designation Consumables use by expiry date The consumables must be completely consumed by the indicated date If a day is not indicated apply the last day of the respective month Store at The conditions necessary to preserve the product s shelf life before opening For in vitro diagnostic use Gr ner Punkt in Germany Manufacturer according to directive 98 79 EC on in vitro diagnostic medical devices Store upright Risk of infection according to the standard DIN ISO 15223 1 2005 Consumables Catalogue number Caution refer to accompanying documents Please refer to safety related notes in the manual accompanying this instrument Please consult instructions for use Serial number model plate April 2007 Service manual Version 2 0 11 91xx pH Electrolyte Analyzers Other symbols The following symbols are listed as additional information Symbol Description Electrodes This date indicates the limit of the maximum storage time of an electrode The electrode must be installed in the instrument no later than the imprinted date If the installation takes place on the imprinted date it still falls within the specifications
83. ES 0000 00 0000 r 008 NMO SOdGNM1 SOdandn LE n NALONAN soda oe N1OLOY 5 PE NAE 62 MOIMI Zz 5 olaa 999 gt gt Nass z T P gt gt z 5 S3A iz iz fo oz gt gt fo 6L 6L lo sad 5 Sauna at bp saa 2 2 fp A ap Cea 9 brag YJ st v3SvHd st le rt H3ATVA le er 3ANA le gt gt Nadsia zt l gt sds 1a m m gt gt dE NINLVG or gt gt IINA or gt L AS 6 6 H Eq TTA 2 1le dvs Be 9 THEA 9 gt as yoLoara uoLosra E i sar ra 2D toT L var eaird i Na v T90SPL auvos MIYOS MIYOS dirii SLY 1nozezsu S19 zesu c 99 GSITIVLSNI ATIWOISAHa LON SI SNI TINS oza zuaNalsvd L tar isa ASHALNIAU4I Zar var AWV IdSIG 0L 90684 4 4 2064 9064 5064 Aos ang a 213891 113591 013591 929 WALSAS LOTTAS SOVODONV I ad L08 0S 60 HLIM SHINW assie as Z 0205 09 92 FE
84. Fluid Counter SnapPak remaining SnapPak installed r reset to 100 Go to STANDBY YES Standby Mode YES gt READY YES NO New Pack Prime CALIBRATE Na K READY Daily m Daily i Remain in Daily Cleaning L Maintenance Parameter added eave Standby Mode Set Time Date Date 01 FEB 07 Time 12 00 ok YES Electrode in Place YES Sel Parameter Sel Parameter Sel Parameter Sel Parameter Na K Li ok NAK Select Phrameter Configuration Na K ok Na K CI ok Na K Ca NO Reset Sample ves Ardyauis re SET SAMPLE Sel Parameter ui Sel Parameter y No NUMBER TO 0 Na Li ok NO Na ILi ok Number Take out of Transp Housing YES Remove Water YES All Electrodes amp Operation Ref installed and Press YES Plug Removed YES Remain in Do You Wish to Pump Windings 5 Relief Clamps Operator Func ae YES DELETE all Data YES Relieved YES Inserted YES YES Shutdown Plug YES Water installed Figure B 27 9180 operator functions menu Roche Diagnostics April 2007 B 40 Service manual Version 2 0 91xx pH Electrolyte Analyzers 8 Software operation 9180 Software operation 9180 Program instrument menu PROGRAM NO SERVICE INSTRUMENT FUNCTIONS YES Enter Code INCORRE TCODE ERROR A
85. INT FUNCTIONS NO Print Last Sample Report Print Cal Report NO Print QC Values and Statistics NO Remain in Print Functions READY 12 45 READY 12 45 Software operation 9110 Figure 4 9110 print functions menu Roche Diagnostics April 2007 Service manual Version 2 0 B 9 6 Software operation 9110 91xx pH Electrolyte Analyzers Software operation 9110 sample menu QC STANDARD No URINE SAMPLE YES ELECTRODE CLEANING Open Door and Introduce Sample Introduce Sample 2 min No 25 VALUES 2s VALUES DE m o Introduce Sample 2 min 25 VALUES 2s VALUES gt m Ya Open Doorand Introduce Sample Introduce Sample YES 2s VALUES Remain in QC Std Urine Sampl READY 12 45 N Store REJECTED Values RS Wipe Probe Close cLosED Sample Door No Store REJECTED Values un Wipe Probe Close cLosED Sample Door 25 VALUES No REJECTED Wipe Probe Close CLOSED Sample Door QC Level 1 in Progress 49 2s 2 ELECTRODE CLEANING QC Level 2 in Progress 49 2s 2 min ELECTRODE CLEANING QC Level 3 in Progress 49 2s Store Values 2min ELECTRODE CLEANING Figure B 5 9110 QC sample menu Roc
86. MAINTENANCE YES QC Level 1 YES Open Door and iniroduce Sample PANPE Wipe Probe Close CLOSED QC Level 1 in Sample Introduce Sample P Sample Door Progress 49 NO VALUES Store y Na K Ca REJECTED Values 99 6 5 34 1 250 2 min YES 2s VALUES NAKCA READY ACCEPTED QC Level 2 YES Open Door and OPEN introduce Sample PAMPLH Wipe Probe Close CLOSE QC Level 2 in Sample Introduce Sample u Sample Door Progress 49 2 min VALUES Store Na K Ca REJECTED Values 99 6 5 34 1 250 2 min YES 25 VALUES READY ACCEPTED QC Level 3 YES Open Doorand OPEN introduce Sample pAMPLE Wipe Probe Close CLOSE QC Level 3 in Sample Introduce Sample P Sample Door Progress 49 NO I VALUES Store Na K Ca REJECTED Values 99 6 5 34 1 250 2 min NAKCA READY VALUES ACCEPTED 5s Standard ies Door and OPEN Wipe Probe Close CLOSED Standard Sample Na K Ca Introduce Sample Sample Introduce Sample P Sample Door in Progress 49 158 8 5 82 1 250 Bicarl NO YES d icarbonate Open Door Introduce Sample Wipe Probe Close CLOSED Bicarb Sample Na K Ca Sample Introduce Sample P Sample Door in Progress 49 158 8 5 82 1 250 NO YES T d pampig Acetate Door and Introd 5 Wipe Probe Close CLOSED Acetate Sample Na K Ca Sample E Introduce Sample min Sample Sample Door Progress 49 158 8
87. NO Print two YES NO Automatic YES NO Setup off a Patient Reports mE Cal Report Remain in YES Program Func No READY 12 45 Figure B 18 9120 9130 9140 program instrument menu Roche Diagnostics April 2007 Service manual Version 2 0 B 27 7 Software operation 9120 9130 9140 91xx pH Electrolyte Analyzers Software operation 9120 9130 9140 Service functions menu NO CALIBRATION Test YES Test YES Std Na 0000 Electrodes Standard A K 0000 0000 NO Test Std Na 0000 Standard B K 0000 0000 NO Test YES Open Door and OPEN introduce Sample External Sample Introduce Sample Remain in Ext Na 0000 Wipe Probe and Test Electrodes K 0000 CI 0000 Close Sample Doqr NO Test YES _ DRY 80 120 102 External 4102 BB m n z Sample Sensor YES Pump NO Exit Pump NO OFF NO Test YES Sample Door Sample Door C Closed O Open NO Test Pump YES Test Pump 2s Test Pump N 25 Test Pump 2s Test Pump Very Slow Speed Slow Speed edium Speed Fast Speed NO 2s Test Valves YES Test Valve A NO Test Valve NO Test Valve V NO Test Valve R YES NO ON OFF YES NO ON OFF YES NO ON OFF YES NO ON OFF NO NO YES Test Interface YES Test a Interface Rd Remain in Test Jump Pins 2 amp 3 Interface ok Interface NO NO NO YES Enter YES Enter Code m Is Code c Code Enter Other NO Service Code AAA ok valid Accepted
88. NSHOS enu EOOSW z HOLIMS HXHG0I2ZS SNITINS HLIM 5 GIS 5 NO 001 TV 5 LNuNOdWOO WHLHOHOS 1545 ts i i Zat 9 516026 ANY f 023903 2 ee sz qurzd 3 Au pue peggnas ATUO dam A E d oiv r D d o dq 88 LS o d In OSHS IWW T 4 T 34 E q 9v OHOVL d 86 YOLON d sv 56 d ne 9100 Q by SIOG d d LOG d za ZLOG zv LLOG i8 ir 99 pd Printer daughter PCB 9110 9120 9130 9140 9180 9181 Figure 16 April 2007 Roche Diagnostics A 51 Service manual Version 2 0 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams Lamp PCB 9110 9120 9130 9140 x 5 gt am 5 5 m Py 5 E q 5 PWB1 PWB lamp sch Rev SCHEMATIC LAMP BOARD Document Number ZD5007 Bs 8 Pi Pi E E E d d d a a a B B fy ww i
89. Present signal is also buffered through U1 on this board April 2007 Service manual Version 2 0 A 35 4 Description of modules 91xx pH Electrolyte Analyzers Description of modules Roche Diagnostics April 2007 A 36 Service manual Version 2 0 91xx pH Electrolyte Analyzers 5 Electronic diagrams Table of contents Electronic diagrams In this chapter Chapter Electronic Gia Stars i au Aa ee Res 39 9110 analyzer block diagram netter tte S 39 9120 9130 9140 9180 analyzer block diagram eee 40 9181 analyzer block diagram seen 41 System interconnect 9110 9120 9130 9140 a 42 SBC power supply 9110 9120 9130 9140 43 SBC PCB valve drivers 9110 9120 9130 9140 A 44 SBC PCB pump motor driver door detect 9110 9120 9130 9140 A 45 SBC PCB input amplifiers 9110 9120 9130 9140 A 46 SBC PCB temperature circuit 9110 9120 9130 9140 A 47 SBC PCB analog selector and A D converter 9110 9120 9130 9140 A 48 SBC PCB microprocessor 9110 9120 9130 9140 A 49 SBC PCB printer display drivers 9110 9120 9130 9140
90. T CALIBRATED D 5 STANDARD A NOT DETECTED D 5 STANDARD B NOT DETECTED STANDARD C NOT DETECTED 9180 9181 QT 0 6 CHECK SAMPLE 5 9110 CHECK SAMPLE SENSOR 9120 30 40 E D 6 CHECK REFERENCE HOUSING D 7 PLEASE CLOSE SAMPLE DOOR except 9181 D 7 NOSAMDEE e tr D 8 CLEANING FLUID NOT D 8 CONDITIONING FLUID NOT DETECTED except 9110 D 8 INTERFACE ERROR ehe RR URS D 9 PAPER JAM OR PRINTER DEFEGT tato oss D 9 D 9 HH NOT D 9 CLOG IN SAMPLE PATH 9110 only CLOG CHECK FLUID PATH except D 11 T1117 4144 Out of D 11 f Temparature out of range astra Shires hacia ieee Gills D 11 PRR D 12 PERFORM DAILY MAINTENANCE except 9110 D 12 REPLACE FLUID PACE 3 55 2 n q biuk de anne D 12 CHECK TEMPERATURE 9140 9180 9181 only D 12 VALVE OVERTEMP CHECK 5 D 13 ERROR UPPER 5
91. T T L 39018 f ven T T Tavs gt gt In i var F we 3514606 26 06 T vra tat G 55 awan sh Hle gt Lads dia e pie anor baj B so E NE 2 anor 5 NaLUNHos ear 8 E 2801102141 8 on or eza ie sn ON 1 vzem Z er Yr ON ea azn zr uarunuos 2801 5251 5 5 Fr SSA Saar AOL vi im e Scu os ay e Wi z Jnzz SZ gs 9 L ISNUVIVG 1SVHLNOO CIEN sos INODANVS 128 E T z z was gt L ezon vu ou e mE TNODGAWS z T T 5 D T T S SEL 5 23 sey 48 an Display daughter PCB 9110 Figure A 18 April 2007 Roche Diagnostics A 53 Service manual Version 2 0 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams Preheater PCB 9110 R T 10K 25C 1 2 MJE800 R1 1 0 MODULAR PLUG BLOCK TP1 CONNECT TO BLOCK Q2 USE INSULATOR PADS MJE800 FOR Ql AND 02 R2 1 0 SCHEMATIC SAMPLE HEATER BL PHb sch Figure A 19 Preheater PCB 9110 Roche Diagnostics April 2007 A 54 Service manual Version 2 0 91xx pH Electrolyte Analyzers 5 Electronic diagrams Display daughter PCB 9120 9130
92. This message is printed only Replace the ISE SnapPakTM following the instructions outlined in the Operator Manual To maximize the life of the ISE SnapPakTM continue running the analyzer until a STANDARD A NOT DETECTED message is displayed This will ensure that the ISE SnapPakTM is completely depleted before replacement CHECK TEMPERATURE 9140 9180 9181 only On the 9140 80 81 models the analyzer will display CHECK TEMPERATURE at the end of each measurement if the temperature measured is out of range 10 0 C to 40 C The temperature sensor is located in the right side electrode holder On the 9110 this message will be displayed if the temperature of the preheater is out of range 36 0 C to 38 0 C Possible causes and remedies Ensure that the preheater cable 9110 only or sample sensor cable is securely plugged in Check whether room temperature is within specified limits 15 C to 32 C 60 F to 90 F e Check fuse LED D3 Perform temperature adjustment procedure Replace SBC board or preheater 9110 only or right side electrode holder Roche Diagnostics April 2007 D 12 Service manual Version 2 0 91xx pH Electrolyte Analyzers 13 Troubleshooting Error messages and troubleshooting VALVE OVERTEMP CHECK VALVES In case one of the valve solenoids overheats the message VALVE OVERTEMP CHECK VALVES will be displayed Each solenoid incorporates a thermo fuse which will reset automat
93. YNSVAN gt lt o 31VSAT1vIG G1S OO ON S 5 5 SO Version 2 0 Service manual B 50 91xx pH Electrolyte Analyzers 9 Software operation 9181 Software operation 9181 Main menu NA K CL READY NO Legend Display prompt PRINT FUNCTIONS NO QC STD DIALYSATE URINE SAMPLE NO DAILY MAINTENANCE NO OPERATOR FUNCTIONS NO PROGRAM INSTRUMENT NO Action performed by analyzer 2 second display prompt Decision making by analyzer QUQL SERVICE FUNCTIONS NO CALIBRATION Figure B 34 9181 main menu Roche Diagnostics April 2007 Service manual Version 2 0 B 51 9 Software operation 9181 91xx pH Electrolyte Analyzers Software operation 9181 Manual measurement sequence K CL READY NO AUTOMATED MEASUREMENT YES Probe moves to upper position Please Wait Introduce Sample lt air voltage Remove Sample and Wipe Probe Retract Probe gt YES lt air voltage Remove Sample and Wipe Probe Retract Probe gt YES Probe moves to upper position lt gt Probe retracts automatically in 30 sec Serum Sample Blood Sample in Progress 49 in Progress 49 Probe moves to upper position Na Na K Cl YES NO 451 5 151 54 114 1 min YES NO 1 min Figure B 35 9181 manual measurement sequence Roche Diagnostics April 2007 B 52 Service manual Version 2 0 91xx pH Electrolyte
94. _ FS sAS ZT AZZ 002 anyo A 818 gt L gt L gt gt z 109 NIA sn SBC PCB valve drivers 9110 9120 9130 9140 Figure A 9 April 2007 Roche Diagnostics Version 2 0 Service manual A 44 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams SBC PCB pump motor driver door detect 9110 9120 9130 9140 auos auge ecu 99 auos lo anos gt gt Naawna e p p wa E 38 soosaz v XX L6 NIYIN OLLVINHHOS qos pour ana gt gt MOL T s iooq gt SH gt gt sa l BEN LOSLLSG HOLIMS UOOA 5 Ed ud T uS zes m In ASL SNOL i aise em oo n a 28 Bd chee anos ange anas ss anise ish M ho aseza ozara ji 2621 n ES 9r ano SA es lt e WF worn oov pr sr wie nu tu ano SSA LaSNaS 1 L L s S 1 vasuas 5 aasvHa 4 t TN 318vN3 ir TL 3SN3S TE 8 HNI TOULNOO X 1 ENI
95. agnostics Service manual Version 2 0 A 32 91xx pH Electrolyte Analyzers 4 Description of modules Electronics ISE SBC card Power circuits Roche Diagnostics Description of modules Power is supplied to this circuit board through connector JP3 from the power supply assembly Test points are provided to measure supply voltages and light emitting diodes D1 through D5 are turned on to indicate each supply voltage present on the card Since each LED is connected in series after the fuse a blown fuse results in the respective LED turned off Four additional voltage regulator circuits are mounted on the ISE SBC II Card IC U2 and IC U3 provide 8 VDC and 8 VDC 9180 9181 5 VDC and 5 VDC and are used to provide the supply voltages for the analog circuitry IC U1 develops the supply voltage used to operate the peristaltic pump and provide either 10 VDC or 22 VDC The 22 VDC is supplied only for the FAST speed and 10 VDC is used for all other pump speeds Signal FAST SLOW determines which supply voltage is selected U5 switches the voltage between 22 VDC and 12 5 VDC 16 VDC on the 9110 which supplies the solenoids When a solenoid is turned on the voltage switches to 22 VDC for 0 5 sec then it returns to 12 5 VDC 16 VDC on the 9110 The circuit which includes Q7 and R5 provides necessary switching for the measuring chamber illuminator LEDs The following list identifies the test points fuses and LEDs for each
96. alyzers 7 Software operation 9120 9130 9140 Software operation 9120 9130 9140 Program instrument menu PROGRAM NO SERVICE INSTRUMENT FUNCTIONS YES Enter Code CODE ERROR AAA RETRY of Lot 0000 Cllow 3 50 Klow 3 50 Nalow 135 0 YES Lot Number Cl high 8 00 K high 8 00 Na high 150 0 NO Current Lot 0000 YES Print old Values NO NewLot YES Enter last 4 Dig Change Lot and Statistics Delete old Data of Lot 0000 NO Program QC T Current Lot 0000 Print old Values m New Lot YES Enter last 4 Dig Level 1 Ranges Change Lot and Statistics Delete old Data i Level 2 Ranges lt n Cllow 3 50 Klow 3 50 Nalow 135 0 _ YES Lot Number Clhigh 8 00 K high 8 00 Na high 150 0 NO Program QC Current Lot 0000 YES Print old Values NO New Lot YES Enter last 4 Dig Level 2 Ranges Change Lot and Statistics Delete old Data of Lot 0000 Cllow 3 Cl high 8 00 Nalow 135 0 YES Na high 150 0 Reset Corr Factor Input Verify E Remain in y default Corr Factors mM YES K b 00 0 K m 1 0000 Nalow 135 0 YES Lot Number Na high 150 0 ok NO fi Cllow 3 50 YES Clhigh 8 00 Program Normal Ranges Program Corr Factors Cl b 00 0 m 1 0000 Na b 00 0 Na m 1 0000 Program Printer YES Patient Report
97. andard B A 27 Standard C not on 9110 A 27 Waste liquids A 28 Mechanical assemblies Display board A 27 Roche Diagnostics Electrode holder Electrode holder assembly A 24 Front door A 23 Lamp board A 27 Needle unit A 23 Peristaltic pump A 24 Printer A 25 Rear panel A 25 SBC board A 26 Solenoid A 24 Disinfectants Deproteinizer A 7 Other disinfectants A 7 Display board A 27 A 35 Document information 3 Edition notice 3 Electronic diagrams A 39 9110 analyzer Display daughter PCB 53 Preheater PCB A 54 9110 9120 9130 9140 analyzer Lamp PCB A 52 Printer daughter PCB A 51 SBC PCB Analog selector and A D converter A 48 Input amplifiers A 46 Microprocessor A 49 Power supply A 43 Printer display drivers A 50 Pump motor driver door detect A 45 Temperature circuit A 47 Valve drivers A 44 9120 9130 9140 analyzer Display daughter PCB A 55 9180 analyzer Display daughter PCB A 67 9180 9181 analyzer Lamp PCB A 66 Printer daughter PCB A 51 SBC PCB Analog selector and A D converter A 62 Fluid Pack detector A 65 Input amplifier A 60 Microprocessor 63 A 64 Power supply A 57 Pump motor driver door detect A 59 Temperature circuit A 61 April 2007 Service manual Version 2 0 F 3 91xx pH Elec
98. anel to remove the solenoid assembly Note The 9110 R solenoid incorporates two heat sinks and is not interchangeable with the other solenoids The printer assembly is designed to allow for easy removal by the user which can be accomplished without removal of the electrical power to the analyzer Slide the paper tray forward to allow access to the printer tear the paper roll and completely remove it together with the paper tray Slide fingers under the printer assembly and pull the printer forward This will disengage the printer from the interconnector and enable removal of the assembly from the front of the analyzer Removal of the printer should be performed for replacement and for removal of a paper jam To replace the printer locate the printer slide and insert the printer assembly Press firmly into place to ensure electrical connection of the printer Note Never attempt to dislodge paper from the printer with a paper clip or similar object to avoid damage to the print head or printer platen Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new printer assembly PRINTER MODULE ISE is available and compatible with 9110 pH Analyzers and with any 9120 30 40 80 and 81 Electrolyte Analyzer 8 Further information see the 91xx Spare Parts List To remove the rear panel assembly ensure that the power cord has been disconnected from the analyzer Remove the four corner screws to e
99. ated in the right side electrode holder Possible causes and remedies Ensure that sample sensor cable is securely plugged in Check whether room temperature is within specified limits 15 C to 32 C 60 to 90 F Perform temperature adjustment procedure Replace SBC board or right side electrode holder Roche Diagnostics April 2007 Service manual Version 2 0 D 11 13 Troubleshooting 91xx pH Electrolyte Analyzers Error messages and troubleshooting ERR The analyzer will display ERR in place of sample results when the analyzer is unable to obtain valid voltage readings from the electrode A D over or underflow Possible causes and remedies Ensure that the electrodes are securely in place and plugged into the analyzer Check to ensure proper filling of the reference electrode has occurred Ensure proper sampling and proper sample preparation Check for air bubbles in the sample Check level of filling solution in the pH electrode and refill if necessary DAILY MAINTENANCE except 9110 This message will be printed at the end of a sample report when cleaning or conditioning have not been performed within the last 24 hours This message is printed only Perform Daily Maintenance to clear this message REPLACE FLUID PACK When the monitored fluid level in the ISE SnapPak TM reaches 5 remaining the analyzer will print REPLACE FLUID PACK at the end of each sample report
100. cess Then press the two plastic tabs on the fill port to allow removal of the fill port from the fill port holder The sample needle can also be removed and is supplied as a one piece assembly including the intake tubing remove the 9120 30 40 80 and 81 sample needle grasp the needle near the white holder block and pull up to unsnap the sample needle from its position To replace the assembly align a replacement sample needle and press the replacement assembly into place Two 9110 or three screws 9120 30 40 80 81 located on the left side of the assembly hold the needle unit assembly in place To access the two rear screws turn power off and remove the rear panel The rear panel can be moved away from the unit for easier access by unplugging the wire harness from the power supply The SBC board can now be removed by removing the screws which secure the SBC board and disconnecting the cables on the SBC board Disconnect the connector to the lamp board which supplies the Door Open signal The two rear screws which secure the needle unit can now be removed using a Philips screwdriver 9110 or an Allen wrench 9120 30 40 80 81 Then remove the front securing screw n a on the 9110 the sample intake tubing and the reagent supply tubing red tag from the fill port The needle unit can now be removed from the front of the analyzer for replacement lt See the 91xx Spare Parts List for a detailed reference of the parts of t
101. d TTL normal transistors etc Today the principle of current logic is virtually the only technology in use in MOS and CMOS components Currents generated by an electrostatic discharge as much as several kV destroy the sensitive component inputs or cause hidden damage This damage is generally not immediately detectable Another effect is that the clearances inside the ICs continuously become smaller The internal wires become thinner and thinner the allowable maximum input voltages become smaller and the effects of any discharges become more and more critical How can ESD protection be guaranteed A continuous discharge is required when working on an electronic assembly Do this as follows Use ESD wrist bands special wrist bands connected to a protective ground Repairs and tests on assemblies must only be conducted on tables with ESD mats connected to grounds or ESD wrist bands Always pick up components at the edge e g like a photo Components must always be transported in ESD packages or appropriate storage or shipping containers use original packaging Shoes with rubber soles or clothing of synthetic fibers must not be worn in workshops where electronic components are repaired Use a humidifier if necessary to maintain optimum humidity in the work area Do not touch assemblies or components with the hand after testing Always transport and ship assemblies or components for repair in ESD pr
102. e Open liquid junction flow through electrode Fully automatic 1 point with each sample 2 point every 3 hours 9180 9181 every 4 hours 10 minutes Suspends calibrations Room temperature 15 32 C 60 90 F April 2007 A 14 Service manual Version 2 0 91xx pH Electrolyte Analyzers 2 Specifications Humidity Data management Diganostic programs Electronics Display Printer Autosampler 9181 only Interface DataLink Electrical requirements Nominal power consumption Dimensions Classifications Product data Safety category Device type Mode of operation Protection classification Explosion protection Approvals Roche Diagnostics Specifications maximum 85 relative humidity non condensing Quality control memory storage 3 levels 35 values calculation of mean standard deviation and coefficien of variation CV user controlled diagnostics YES NO operation via the display Microprocessor controlled LCD dot matrix 2 lines 16 characters per line Integral thermal printer 16 characters width Integral turntable 18 positions 2 mL or 0 25 mL sample cups RS232C serial port Data link to COMAPCT 2 3 9180 9181 100 240 VAC 50 60 Hz 90 250 VAC for 9110 self adjusting 50 W Height 335 mm 12 2 Width 315 mm 12 4 Depth 295 mm 12 0 Weight approx 6 kg 13 lbs Table A 4 Instrument dimensions I B according to OVE MG EN 60601 1 IEC 601 1
103. e conditions for overvoltage category II The instrument meets the conditions for contamination level 2 Do not operate the instrument in an explosive environment or in the vicinity of explosive anesthetic mixtures containing oxygen or nitrous oxide Ifan object or liquid enters the internal areas of the instrument remove the instrument from its power supply and allow an expert to check it thoroughly before using it again The instrument is suitable for long term operation indoors Y e power cord may be plugged into a grounded socket only When using an extension cord e make sure it is properly grounded e Any rupture of the ground lead inside or outside the instrument or a loose ground connection may result in hazardous operating conditions Intentional disconnection of the grounding is not permitted The instrument is not suitable for operation with a direct current power supply Use only the original mains plug delivered with the instrument Important notes and warnings e Never operate the analyzer in the vicinity of volatile or explosive gases e g anesthetic gases etc The instrument must be connected to a 2 grounded socket e The power cable and the plug must be undamaged Damaged power cables and plugs must be replaced immediately Before opening the back panel switch off the instrument and disconnect the power cable Replace damaged fuses with the specified type of fuse
104. e should not exceed 0 5 ohms Be sure to consider the resistance of the ohmmeter s connecting cables After the rear panel has been removed the SBC Board is accessible for removal This circuit board can be removed by removing the seven screws securing the board to the housing and by disconnecting electrical connectors from the power supply module display board and lamp board lt For details see Rear panel assembly on page A 25 The electrode push pins can now be replaced by simply pulling the connecting pin from its socket When installing the board make sure to place the plastic washers under the screw heads Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new SBC board MAINBOARD 9180 gt SN 12001 is available and compatible with 9180 Electrolyte Analyzers serial number gt 12001 only The former SBC board MAINBOARD 9180 lt SN 11999 is still available and has to be used for 9180 Electrolyte Analyzers SN lt 11999 Further information see the 91xx Spare Parts List After changing the main board or the power supply always perform a measurement of protective ground resistance to ensure a safe connection of the system ground Always measure with the power cord attached to the analyzer The resistance measurement has to be done between the ground pin of the power plug and several ground connectors placed on each of the analyzer s electronic components such as power supply or main board and
105. ectrolyte Analyzers 5 Electronic diagrams Electronic diagrams SBC PCB input amplifier 9180 9181 s z 2xas ew1 e 7 595 O Ae AEpUO end 10842 sequinn T z T L8 L6 0816 GHVOS NIYIN OLLVIIHOS NOBLZOUL en zsazais b uaNarsva Pass lr iaqqoragoxa A WHOJLWHHNSO LASAAO gt anro anro eso 159 l INS e r9ogow1 VAS VAS zea NI ALON ML siis N2642O91L E 1non eo103 5 s a NE sou 6 z L seu T1 92 12 MMYA aen im or vasoudZ Yl m 17 T nostanee AYUOOL 135440 VAS NI NYHL SdONd ALON ET ar 1 seu Ei Ab ei vou az n 9 L L ri ova z zer R vein A ud Orar ji vAS Diet see aS Aos ane sro gt VAS NI SHOHd ALON E m oM aumente AARON NOBLZOL I ar 1 00L L ML e 1noeN gt gt M 5 J N cou gt L LH ven 951 99
106. ed as is indicated on the display April 2007 B 66 Service manual Version 2 0 91xx pH Electrolyte Analyzers 10 Service functions Example displays TEST PINCH VALVES Example displays TEST INTERFACE TEST AMPLIFIER Example display Roche Diagnostics Service functions Test Pump Test Pump Very Slow Speed Slow Speed Test Pump Test Pump Medium Speed Fast Speed This test turns on off each solenoid valve as indicated and is useful to verify both electronic control and mechanical operation of each solenoid valve Test Valve A Test Valve B Test Valve C YES NO ON OFF YES NO ON OFF YES NO ON OFF 9180 9181 only Test Valve V Test Valve R YES NO ON OFF YES NO ON OFF The Interface Test can be used to verify correct operation of the RS232C serial port To operate this test jumper pin 2 and pin 3 of the rear interface connector prior to invoking the test During the interface test the analyzer sends out a test string and checks if it is received within a set period of time If correct operation is observed after completion of the test INTERFACE OK will be displayed The Amplifier Test displays ground reference and output voltage of each input amplifier The values displayed for each electrode channel will be dependent on the electrode inputs and may vary Display of the ground
107. embrane may result in inflammation Danger of sensitization during contact with skin XI R36 37 38 8 26 37 when classified with R 43 Caution Avoid contact with eyes and skin do not inhale vapors Do not use content if the packaging is damaged Protective gloves protective goggles and suitable protective clothing must be worn Roche Diagnostics April 2007 12 Service manual Version 2 0 91xx pH Electrolyte Analyzers Abbreviations The following abbreviations are used Abbreviation Definition A ADC ANSI C CMOS E EEPROM e g EC EN ESD HIV HW IEC ISE IVD LCD LED MC MOS MSDS PCB QC REF s SBC board SW T TTL Analogue to digital converter American National Standards Institute Complementary Metal Oxide Semiconductor Electrically erasable programmable read only memory exempli gratia for example European community European standard Electrostatic Discharge Human Immunodeficiency Virus Hardware International Electrical Commission selective electrode In vitro Diagnostics Liquid Cristal Display Light Emitting Diode Measuring chamber Metal Oxide Semiconductor Material safety data sheet Printed Circuit Board Quality control Reference solution Mainboard of the instrument Software Transisitor Transistor Logic Roche Diagnostics April 2007 Service manual Version 2 0 13 91xx pH Electrolyte Analyzers Roche Diagno
108. emedies Check for obstructions preventing the sample wheel from moving freely Replace sampler Roche Diagnostics April 2007 D 14 Service manual Version 2 0 JW S I EN E 3 April 2007 91xx pH Electrolyte Analyzers Glossary Alkaline Basic Arterial blood Blood taken from aorta Autosampler Sample wheel for automatic measurements with 9181 Electrolyte Analyzers Clot Catcher Coagulate trap Dummy electrode The dummy electrode is simply a flow electrode and is only intended as a spacer and it has no additional measuring functions Electrodes The electrodes are flow through electrodes with a visible sample channel Electrode simulator Test device for the Measuring chamber Heparin salts Are the only clot inhibitors that may be used for analyses in the 91xx pH Electrolyte Analyzers Input unit The sample insertion as well as the aspiration of solutions is carried out via input unit which consists of the sample probe mechanism behind the sample door ISE abbr for ion sensitive electrodes Roche Diagnostics ISE measuring chamber The ISE measuring chamber and its sensors are used to measure the hematocrit value and the electrolyte values Nat Kt Ca and Lit ISETROL QC material for checking ISE Locking lever This lever fixes the electrodes in the measuring chamber Measurement evaluation The validity of the test results
109. eous and QC solutions Sample device Syringe sample cup capillary Roche MICROSAMPLER Roche Diagnostics April 2007 Service manual Version 2 0 A 13 2 Specifications 91xx pH Electrolyte Analyzers Specifications Sample size Analysis time Sample rate pH Sensor Reference System Calibration Warm up time 9120 9130 9140 9180 9181 Sample types Sample device Sample size Analysis time Sample rate Sodium Na sensor Potassium K sensor Chloride CH sensor Calcium Ca sensor Lithium Li sensor Reference system Calibration Warm up time Common operating parameter Standby mode Temperature Roche Diagnostics 25 uL 70 seconds 40 per hour with printout 50 per hour without printout pH sensitive flow through glass capillary electrode in transparent acrylic plastic Open liquid junction flow through electrode Fully automatic 1 point with each sample 2 point every 3 hours 10 minutes Whole blood serum plasma urine dialysate acetate or bicarbonate Syringe sample cup capillary Roche MICROSAMPLER 95 uL 50 seconds 45 per hour with printout 60 per hour without printout Ion selective flow through glass capillary electrode Ion selective flow through liquide membrane electrode Ion selective flow through liquide membrane electrode Ion selective flow through liquide membrane electrode Ion selective flow through liquide membrane electrod
110. er Assembly Power Supply BK5014 BK5013 22 VDC 10 VDC 5 VDC 5 VDC 8 VDC 8 VDC Bulkhead 22 VDC 12 5 VDC Interface Driver 100 240 MAIN 50 60 Hz Peristaltic Pump Driver Solenoid Valve Drivers JP3 JP8 Ctrl Lines Turntable Lamp Board P Pump Motor JP6 Sample Door JP7 Valve A JP1 Valve B JP2 Valve C JP3 Valve V JP4 Valve R JP5 www ww Power Receptacle 1 6 AT 1 6 JP1 Power Supply Module Fuse F1 2 5 A 250 V 5 VDC 24 VDC 12 VDC 12 VDC Figure A 6 Roche Diagnostics 9181 anayzer block diagram April 2007 Service manual Version 2 0 A 41 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams System interconnect 9110 9120 9130 9140 AN 99Z4HO gaso1oq gt S S3SVHd vasvHd 3ATVA DY gt gt gt gt tanva dnW I SNITIAS HxtidoLoz3 WSLLNI Hd 5 jo L sung epsiumul srg 500587 x Joquinn eas XXL6 NIYIN OLLVINSHOS n gt gt Lolaa an 99 Nass DE gt gt S
111. er position Wipe Probe YES lower position NO same for Level 1 3 VALUES NO Store Na K QC Level 1 in REJECTED Values 99 6 5 34 1 250 Progress 49 2min YES VALUES NAKCA READY ACCEPTED Standard YES Automated NO Probe moves to Please Remove Sample Retract Probe gt Measurement upper position Wait Introduce Sample Wipe Probe YES Probe moves to lower position Retract Probe gt YES Probe moves to lower position Retract Probe gt YES NO Na K Ca Standard Sample 99 6 5 34 1 250 Progress 49 Please Remove Sample Introduce Sample Wipe Probe Na K Ca Bicarb Sample in 99 6 5 34 1 250 Progress 49 Please Remove Sample Introduce Sample Wipe Probe Na K Ca Acetate Sample in Probe moves to 99 6 5 34 1 250 Progress 49 lower position Please Introd s Remove Sample Retract Probe gt Wait Sample Wipe Probe YES Na K Ca Urine Sample in Probe moves to 99 6 5 34 1 250 Progress 49 lower position Bicarbonate YES Automated NO Probe moves to Sample Measurement upper position NO NO Acetate YES Automated NO Probe moves to Sample Measurement upper position NO YES Urine Automated NO Probe moves to Sample Measurement upper position Figure B 38 9181 QC std dialysate urine sample menu Roche Diagnostics April 2007 Service manual Version 2 0 B 55 9 Software operation 9181 91xx pH Electrolyte Analyzers
112. eristaltic pump assembly A 24 Solenoid assembly ee elei iet 25 Ponterassembiy au u a a pl O kb pa SS RSS A 25 Rear panel assembly A 25 SBE board 26 Lamp pe 27 a 27 Modules 27 e e aa 27 StandardB 27 Standard C not on 9110 A 27 Cleaning solution 9110 only 28 Reference solution est 28 Waste liquids papapa Bu A 28 Main tube set all models Intake tube set 9110 only A 28 Sample Contacts 9110 A 29 Electronics g ees A 33 ISE SBG card A 33 Eamp board 2 2 2222 A 34 Display board erre eR SERRE odi 35 Roche Diagnostics April 2007 Service manual Version 2 0 A 21 4 Description of modules 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 A 22 Service manual Version 2 0 91xx pH Electrolyte Analyzers 4 Description of modules Description of modules Description of modules A After changing any part of the analyzer always perform a calibration of the analyzer as well as a full QC measurement QC measurements must be performed in their entirety i e all three QC levels must be measured Omitting QC measurements or ignoring QC measureme
113. essages and troubleshooting Possible causes and remedies Close sample door within the time allowed Check needle mechanism light gate Check whether preheater 9110 or sample sensor is plugged in NO SAMPLE This message is displayed when any of the following conditions occur During sampling either a bubble in the sample is detected or the sample size is smaller than the minimum sample required The sample door is opened and no sample is fed into the analyzer Sample door is not closed within 20 seconds after sample is detected Possible causes and remedies Check whether the sample sensor is plugged in and perform TEST SAMPLE CONTACTS Gr TEST SAMPLE SENSOR to verify that it operates correctly Check sample draw and look for clots or leaks in the sampling system CLEANING FLUID NOT DETECTED When the cleaning cycle has been initiated the analyzer checks for proper aspiration of the cleaning solution using the sample contacts 9110 or the sample sensor If solution is not detected in the proper time above error message will be displayed Possible causes and remedies Ensure that a sufficient volume of cleaning solution is presented for aspiration and no air bubbles are present as solution is aspirated Check to ensure that the solution is being aspirated through the sampling mechanism correctly The sample path must be free of leaks or clots Verify that sample sensor is properly p
114. etd z T a Sou E 6 z T 8 99H _ T eeu 2 TI 92 l2 P I z T r vagsoud ZLO SELN ru Lidl f H AWOOL 195430 LON F i 6 i LON OSO 5 NI NHOHd ALON 0 WM IE a Y 4 71 gt H 5 1 r z r i rou We z 8 seu d ovy z nidi MMs 2 casos OLdL i su mama D III uu 4 i anzz o 6 LON 3l ner z T Reet MODA MINA eria bcm WM NI FHS E x d CTET 1nOHd EN gt z T T T fen pe ano E Hd 9Nu vein 2 5 Hd L q InoHsS ASL LNG svzHV sHONd ONIN T VANY NI XSVW ON ALON SALON Hd Lasaao Aw00ZZ 15440 z Ir 8v Hd SSI LNHSHSAGIG LVH ANY PTS JO SANTIYA ALON SHNUIAI IdWV LAANI April 2007 Version 2 0 Service manual SBC PCB input amplifiers 9110 9120 9130 9140 Roche Diagnostics Figure A 11 A 46 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams SBC PCB temperature circuit 9110 9120 9130 9140 NNYaV 390V
115. from the instrument must be carefully examined by a clinical medical specialist who will take the patient s clinical condition into consideration before any clinical decisions are reached based on the test results Measuring chamber Underneath the front door is the ISE measuring chamber with the electrodes MC abbr for measuring chamber NIST standards Are precise serums with accredited target values Peristaltic pump see Pump Plasma Plasma samples are obtained by centrifuging heparinized whole blood during which the cel lular components of the blood are removed from plasma PP abbr for peristaltic pump Printer Low noise thermoprinter April 2007 Service manual Version 2 0 E 3 91xx pH Electrolyte Analyzers Pump One peristaltic pump transports the sample and the operating fluids inside the instrument QC abbr for Quality control QC material see ISETROL Quality control A quality control program includes the analysis of sample materials with known ranges of expected values and the comparison of these values with analyzer results Ready mode Top level of the analyzer mode Reference electrode The reference electrode must always produce the same signal regardless of the sample s composition This is achieved by mixing the sample flow with a liquid of a higher KCl concentration reference solution Rubber sealing sealing of SnapPak Sample contact Sample contacts support mo
116. he Diagnostics April 2007 B 10 Service manual Version 2 0 91xx pH Electrolyte Analyzers 6 Software operation 9110 Software operation 9110 Daily maintenance menu ELECTRODE NO OPERATOR CLEANING FUNCTIONS Perform Auto YES Cleaning NO Perform External YES Cleaning NO YES Remain in Electrode Cleaning NO PERFORMED YES 30 SECOND READY 12 45 STABILIZATION 6 9110 daily maintenance menu NO Roche Diagnostics April 2007 Service manual Version 2 0 B 11 6 Software operation 9110 91xx pH Electrolyte Analyzers Software operation 9110 Operator functions menu PRINT NO 5 FUNCTIONS QC SAMPLE YES Check Change YES 37 SnapPak remaining New Reagent Pack installed YES NO Go into STANDBY 5 YES opinas re you sure ounter to Standby Mode YES gt READY YES NO CALIBRATE READY 12 45 YES Perform Auto NO Perform External NO Cleaning 30 SECOND STABILIZATION CALIBRATE Reset Sample N 5 YES SET SAMPLE READY 12 45 Number lene NUMBER TO 0 NO READY 12 45 Take out of YES YES Shutdown Plug YES Transp Housing Are you sure Operation Water installed Ref installed NO Leave Standby Mode YES Set Date 01 FEB 07 Time Date Time 12 00 ok NO NO Date 01 FEB 07 Time 12 00 ok READY 12 45 YES Remain in 5 Relief Clamps YES All Elect
117. he discharges are perceived as small shocks or visible sparks Shoes with rubber soles Friction is caused by walking A person develops an electrical charge different from the ground A discharge occurs when an object e g door handle is contacted Synthetic fiber clothing Discharge is audible and is visible in darkness The dryer the air the greater the risk of electrical discharge caused by friction Eectrostatic charges are less likely to build up in humid air particularly when saturated with water vapor The probability of ESD phenomena is therefore particularly great in a northern hemisphere winter in centrally heated rooms where the humidity is low Influence of electrostatic charges on components Roche Diagnostics If a person with an electrostatic charge touches a component a discharge over a connection of an IC or semiconductor element may occur The resulting voltages may damage the component Critical situations can occur while repairing or testing components if they are lying on a more or less conductive surface e g table top and a person with an electrostatic charge touches them A discharge through a critical component connection can also occur in this case April 2007 A 8 Service manual Version 2 0 91xx pH Electrolyte Analyzers 1 Safety information ESD protection measures Why is ESD protection so important today Formerly current controlling semiconductors were mostly use
118. he needle unit To remove the electrode holder assembly from the analyzer first unplug the sample intake tubing on the right side of the module then the preheater cable 9110 only or the sample sensor cable 9120 30 40 80 and 81 then remove the reference solution tubing connector and last take out the waste tube green tag at the left side of the electrode holder The electrode holder can now be removed by pressing the two plastic tabs right and left side of the module and sliding the module forward Press the plastic tabs again to release the second detent and completely remove the module from the analyzer The sample preheater or the sample sensor and or the left side electrode holder can be removed and replaced by removing the screws on the under side of this module amp See the 91xx Spare Parts List for a detailed reference of the parts of this assembly Replacement of the roller assembly of the peristaltic pump can be accomplished by removing the pump tubing and firmly pulling the roller assembly toward the front To replace the roller assembly align the roller and the flat on the motor shaft and press the replacement roller into position To remove the motor first remove the rear panel assembly and SBC Board lt For details see Rear panel assembly on page A 25 SBC board on page A 26 Now the electrical connector on the lamp board can be accessed to unplug the motor connector The motor can now be removed by removing four
119. ically Possible causes and remedies Turn power off for at least one hour Check fuse 24V Check whether room temperature is within specified limits 15 C to 32 C 60 F to 90 F ERROR UPPER NEEDLE SENSOR If probe movement to the upper position is not detected within a certain time period the above message is displayed Possible remedies Check for foreign material that may prevent the probe from moving freely Contact Roche Diagnostics for technical assistance ERROR LOWER NEEDLE SENSOR If probe movement to the lower position is not detected within a certain time period the above message is displayed Possible remedies Make sure the fill port is installed correctly Check for foreign material that may prevent the probe from moving freely Contact Roche Diagnostics for technical assistance WHEEL MISSING OR SAMPLER DEFECT During the homing sequence the sampler checks for the sample wheel to be present Possible remedies Make sure the wheel is present and seated correctly Replace sample wheel Replace sampler Roche Diagnostics April 2007 Service manual Version 2 0 D 13 13 Troubleshooting 91xx pH Electrolyte Analyzers Error messages and troubleshooting SAMPLER JAMMED OR DEFECTIVE The analyzer checks for correct movement of the sample wheel If the sample wheel does not move at the expected speed the above error message is displayed Possible r
120. independent of the printer settings the data is always sent Note The arrow up e g out of normal range is sent as HEX 18 N the arrow down as HEX19 J and the degree is sent as HEX1A gt April 2007 Service manual Version 2 0 A 19 3 Interface specifications 91xx pH Electrolyte Analyzers Interface specifications Example data string information Automatic Calibration Report Serum sample report lt sx gt AVL 9130 lt cr gt lt crlf gt ELECTROLYTE ISE lt cr gt lt crlf gt 03JAN92 10 51crcrlf CALIBR REPORT cr crlf cr crlf DailyMaintenance lt cr gt lt crlf gt Performed Last cr crlf 02JAN92 10 35 cr crlf cr crlf Standard A lt cr gt lt crlf gt Na 112mV 3 lt cr gt lt crlf gt 1392mV 3 lt cr gt lt crlf gt Cl 106mV 3 cr crlf cr crlf Difference A B lt cr gt lt crlf gt Na 1402mV lt cr gt lt crlf gt K 1032mV lt cr gt lt crlf gt Cl 1006mV lt cr gt lt crlf gt lt cr gt lt crlf gt Fluid Pack lt cr gt lt crlf gt 68 Remaining lt cr gt lt crlf gt lt cr gt lt crlf gt lt ex gt lt sx gt AVL 9130 lt cr gt lt crlf gt ELECTROLYTE ISE lt cr gt lt crlf gt 03JAN92 10 59 lt cr gt lt crlf gt lt cr gt lt crlf gt Name SOURCE IE oris va voro lt cr gt lt crlf gt Sample SERUM lt cr gt lt crlf gt lt cr gt lt crlf gt Sample No 13 cr crlf cr crlf Na 159soh mmol L lt cr gt lt crlf
121. ions after contact with eyes skin or mucous membranes cannot be excluded After inhalation breath fresh air drink large amounts of water After skin contact wash with generous amounts of water remove contaminated clothing After eye contact rinse eyes with generous amounts of water contact an eye specialist After drinking drink large amounts of water avoid vomiting contact a doctor Use standard alcohol based 7096 disinfectants to clean the surface of the instrument Read the product information first Never use standard disinfectant to clean the tubes and tubing paths under any circumstances Do not use any type of bleaching agent Exception Roche Deproteinizer April 2007 Service manual Version 2 0 7 1 Safety information 91xx pH Electrolyte Analyzers ESD protection measures ESD protection measures ESD Electrostatic Sensitive Device Explanation of the phenomenon Examples The most frequent cause of electrostatic discharge is friction of various materials such as plastic synthetic fiber hard rubber or paper It can also be caused by bending or applying pressure to a material Discharges that are dangerous to components occur when charged bodies or components do not have a discharge connection ground to discharge the electrostatic charge The discharges normally do not have a high capacity but there are often voltage differences in the range of several thousand volts T
122. isplays In this mode the sample sensor output is displayed which is useful for electrical adjustment or for troubleshooting sample sensor faults A dry sample sensor should display a value of 100 with a range of 80 to 120 After verifying a dry sensor aspiration ofa clear fluid i e water should increase the displayed value by at least 40 units DRY 80 120 102 External 153 DRY 135 145 141 YES Pump NO Exit Pump NO OFF YES Pump NO Exit 9110 pH Analyzer TEST SnapPak SENSOR 9180 and 9131 only Example display In this test the SnapPak Sensor detects whether the ISE SnapPak is in place I indicates that the ISE SnapPak is in place O indicates that it is either not present or not completely seated SnapPak I I IN O OUT TEST LANGUAGE SWITCH 9180 and 9181 only Roche Diagnostics This mode displays the language selected for display and printout To change the selection set the language switch located behind printer to the desired language The display will indicate the language choice according to the switch setting To activate a new language turn the analyzer off and back on with the switch set to the desired language April 2007 Service manual Version 2 0 B 65 10 Service functions 91xx pH Electrolyte Analyzers Service functions Example display
123. istent with that of the original solutions can negatively affect above all the long term measurement accuracy Deviations in the composition of the solutions can also decrease the service life of the electrodes The quality control requirements must be completed at least once daily for safety reasons Because accurate measurement results depend not only on the proper functioning of the instrument but also on a number of other factors such as preanalytics the results produced by the instrument should be examined by a trained expert before subsequent decisions are reached that are based on the measurement values Explanation Caution refer to accompanying documents April 2007 Service manual Version 2 0 A 5 1 Safety information 91xx pH Electrolyte Analyzers Operating safety information Operating safety information The instrument has been constructed and tested according to the following European Standards e IEC EN 61010 1 2001 IEC EN 61010 2 101 2002 valid for 9180 Electrolyte Analyzers SN gt 12001 only IEC EN 61010 2 081 2002 A1 2003 It was delivered from the factory in flawless condition with regards to safety features In order to preserve this condition and ensure safe operation the user must respect the notices and warnings that are contained in this Service Manual This instrument is classified under the protection class I according to IEC EN 61010 1 The instrument meets th
124. lo ejduies AO 2 07 woad 184971 SD 158 Juayed SNOILONNA 045 sonvHofonlsr zr Aava 9110 Figure 1 April 2007 Roche Diagnostics B 6 Version 2 0 Service manual 91xx pH Electrolyte Analyzers 6 Software operation 9110 Software operation 9110 Main menu READY 12 45 NO Legend CHANGE PATIENT TEMPERATURE NO PRINT FUNCTIONS NO QC STANDARD URINE SAMPLE NO Display prompt Action performed by analyzer 2 second display prompt Decision making by analyzer Qual DAILY MAINTENANCE NO OPERATOR FUNCTIONS NO PROGRAM INSTRUMENT NO SERVICE FUNCTIONS NO CALIBRATION Figure B 2 9110 main menu Roche Diagnostics April 2007 Service manual Version 2 0 B 7 6 Software operation 9110 91xx pH Electrolyte Analyzers Software operation 9110 Measurement sequence READY 12 45 OPEN PRINT FUNCTIONS Open Sample Doo Introduce Sample lt air voltage Remove Sample Close Door FLAP CLOSED pH Measurement in Progress 65 YES NO 1 min Figure B 3 9110 measurement sequence Roche Diagnostics April 2007 B 8 Service manual Version 2 0 91xx pH Electrolyte Analyzers 6 Software operation 9110 Print functions menu PR
125. lugged in and perform TEST SAMPLE CONTACTS or Gr TEST SAMPLE SENSOR to verify the correct operation of the sample sensor Check pump tubing and replace as necessary to ensure a correct pump rate CONDITIONING FLUID NOT DETECTED except 9110 When the conditioning cycle has been initiated the analyzer checks for proper aspiration of the conditioning solution using the sample sensor If the solution is not detected in the proper time this error message will be displayed Possible causes and remedies Ensure that a sufficient volume of conditioning solution is presented for aspiration and no air bubbles are present as solution is aspirated Roche Diagnostics April 2007 D 8 Service manual Version 2 0 91xx pH Electrolyte Analyzers 13 Troubleshooting INTERFACE ERROR Possible cause and remedy Error messages and troubleshooting Check to ensure that the solution is being properly aspirated through the sampling mechanism The sample path must be free of leaks or clots Verify that sample sensor is correctly plugged in and perform TEST SAMPLE SENSOR to verify its proper operation Check pump tubing and replace as necessary to ensure correct pump rate This message is displayed only during the G INTERFACE TEST Pins 2 and 3 must be shorted together to perform the INTERFACE TEST Make sure that the pins are not shorted to chassis ground Replace SBC Card PAPER JAM OR PRINTER DEFECT
126. manual Version 2 0 91xx pH Electrolyte Analyzers 3 Interface specifications Interface specifications Interface specifications Interface information AN Software 9 5 Roche Diagnostics The 9110 9120 9130 9140 9180 and 9181 analyzers are equipped with a standard serial interface output This interface output is intended to be used with standard commercially available computer systems The data transmitted through the serial interface port employs the ASCII code The serial interface is terminated on the rear cover with a 9 pin male DB 9 connector The signal levels are as follows Binary 1 12 V to 3 V Binary0 3 V to 12 V Two stop bits follow the eight data bits to complete the 10 bit word The baud rate is set at 9600 Baud fixed The maximal recommended cable length is 40 feet The pin assignment is as follows pin l sample ground GND pin 2 receive data RxD pin 3 send data TxD pin 4 NC pin 5 signal ground GND pin 6 NC pin 7 RTS not used pin 8 CTS not used pin 9 NC NC Not Connected Higher discharge current A higher discharge current can be expected when using the serial interface This must be checked by suitably qualified personnel depending on the local regulations The patient sample data is sent at the end of each measurement the calibration report is sent at the end of each calibration The interface is always on
127. mbly CHECK REFERENCE HOUSING At the beginning of a calibration pressure is formed in the reference housing while the R valve is closed and the V valve is open After Standard A is detected in the sample chamber the back pressure formed in the reference housing causes the Standard A solution to move back toward the preheater 9110 or the sample sensor When the standard A solution is detected by the preheater 9110 or the sample sensor the R valve is opened to release the pressure If the preheater 9110 or the sample sensor does not detect standard A as it moves back CHECK REFERENCE HOUSING will be displayed Possible causes and remedies Check for clogged reference housing and clean per monthly maintenance procedures Check and ensure that reference tubing is securely connected at the tubing connector Check for proper filling of the reference housing and ensure that reference solution reagent level allows complete filling of the reference housing Check for bubble free aspiration of Standard A PLEASE CLOSE SAMPLE DOOR except 9181 This message occurs and prints NO SAMPLE on the print out when either of two conditions exists Sample door is not closed within 20 seconds after samples are in place Sample door has been opened and no sample is detected within 20 seconds Roche Diagnostics April 2007 Service manual Version 2 0 D 7 13 Troubleshooting 91xx pH Electrolyte Analyzers Error m
128. n the sample path especially in the sample probe the tubing to the preheater or the sample sensor and in the preheater or the sample sensor Perform Electrode Cleaning 9110 only Check sample contacts 9110 only or sample sensor is securely plugged in and check Sample Contact Test 9110 only or Sample Sensor Test to ensure correct operation Check peristaltic pump and tubing are correct and all tubing is properly in place 1111 1444 Out of range In case the analyzer displays arrows up or arrows down instead of sample results the concentration of the sample is outside of the measurement range Possible causes and remedies Ifthe sample is a urine sample arrows up instead of the result indicate that further dilution of the specimen is required except 9110 Check for proper sample preparation Ensure that the sample is correctly aspirated into the measuring chamber and ensure small air bubbles are not present Check for proper aspiration of Standard A Check the programmed conversion factors The sample result has to be within the measurement range after conversion factors have been applied Check the level of Filling Solution in the pH Electrode 7 4 Temparature out of range On the 9140 9180 and 9181 models the calibration report will print an arrow up or down instead of the actual temperature if the temperature measured is out of range 10 0 C 40 0 The temperature sensor is loc
129. nc Figure B 28 9180 program instrument menu Roche Diagnostics April 2007 Service manual Version 2 0 B 41 8 Software operation 9180 91xx pH Electrolyte Analyzers Software operation 9180 Service functions menu N EDICIONES 9 CALIBRATION YES Test YES Test YES Std A Na 0000 YES Test YES Std B Na 0000 Electrodes Standard A K 0000 Ca 0000 Standard B K 0000 Ca 0000 NO NO Test Std C 0000 Standard C K 0000 Ca 0000 NO Test YES External Sample Introduce Sample Remain in Ext 0000 Wipe Probe and Test Electrodes K 0000 0000 Close Sample Dodr fees NO Test YES DRY 80 120 102 External 102 Sample Sensor YES Pump NO Exit Pump NO OFF NO Test SnapPak 1 SnapPak Sensor I In O Out NO Open Door and Introduce Sample z o NO Test YES Language Switch Language Switch English NO NO Test YES Sample Door C Sample Door C Closed O Open NO NO Test Pump YES Test Pump 2s Test Pump N 25 Test Pump N 25 Test Pump Very Slow Speed Slow Speed Medium Speeg Fast Speed NO 2s Test Valves YES Test Valve A NO Test Valve NO Test Valve Test Valve V YES NO ON OFF YES NO ON OFF YES NO ON OFF YES NO ON OFF NO YES NO NO Test Interface YES Test Interface mE Remain in Test Test Valve R Jump Pins 2 amp 3 ok Interface YES NO ON OFF NO NO Test Amplifier YES Na 1030 K 120 NO Ground 2 Temp mV
130. nce housing tubing to be plugged into the reference solution circuit As the peristaltic pump aspirates reference solution solenoid valve R is opened to allow excess solution to be pumped into the reference return line of the ISE SnapPak A check valve 15 placed in the return line to ensure that the output is constant 9110 only Calibration standard and sample waste are pumped out of the left side of the electrode module green banded tubing through the peristaltic pump to the waste line of the ISE SnapPak An internal one way valve is incorporated into the waste line of each ISE SnapPak to prevent any waste products from leaking out of the ISE SnapPak Main tube set all models Intake tube set 9110 only Intake Tube Set 9110 only Roche Diagnostics The main tube set is supplied as a pre cut tubing harness with the fluidic interconnections preassembled Replacement of the tube set should be performed annually For removal of the tubing first remove the ISE SnapPak from the reagent compartment This will allow access for removal of the white TPR block which is the main interconnection between the ISE SnapPak and the tubing This block can be removed by grasping the front surface and pulling forward Next remove tubing from each solenoid valve Follow the procedure outlined in Section 4 1 5 for disassembly of the solenoid valves to enable fluidic tubing to be easily removed at each solenoid lt For details
131. ne Probe Close CLOSED QC Level 3 in Sample Introduce Sample u Sample Door Progress 49 VALUES Store Na K CI REJECTED Values 99 6 5 34 88 2 2 min VALUES READY 12 45 ACCEPTED Standard ves Open Door and OPEN Sample Probe Close CLOSED Standard Sample Na K Cl Sample Introduce Sample P Sample Door in Progress 49 99 6 5 34 88 2 Uri a a NO YES a ine vel Door and OPEN SAMPLE Wipe Probe Close CLOSED Urine Sample in Na K Cl Introduce Sample Introduce Sample Sample Door Progress 49 99 6 5 34 88 2 NO YES Remain in QC Std READY 12 45 Urine Sample j 2 min Figure B 15 9120 9130 9140 QC Standard urine sample menu Roche Diagnostics April 2007 B 24 Service manual Version 2 0 91xx pH Electrolyte Analyzers 7 Software operation 9120 9130 9140 Software operation 9120 9130 9140 Daily maintenance menu DAILY NO OPERATOR MAINTENANCE FUNCTIONS Daily Open Sample Door OPEN Wipe Probe close CLOSE Cleaning Introduce Sample Sample Door Cleaning Solution not detected Open Sample Door OPEN Wipe Probe Close CLOSE Introduce Sample Sample Door Cleaning Solution not detected detected Solution detected Daily Conditioning Remain in Daily Maintenance NO 1min YES 30 SECOND READY 12 45 STABILIZATION CALIBRATE YES AND OR COND PERFORMED
132. nitoring conductivity when filling the measuring chamber with fluid Sample sensor Sample sensors support monitoring optically when filling the measuring chamber with fluid Sample throughput Number of samples per hour SnapPak Calibration solutions for ISE including waste container Volume limitation Is the maximum volume of sample which is aspirated from the container Roche Diagnostics Waste Container Waste water part of the SnapPak April 2007 4 Service manual Version 2 0 Index April 2007 91xx pH Electrolyte Analyzers Index Adjustment Sample contacts 9110 only 5 Sample sensor 9120 9130 9140 9180 9181 only C 5 Temperature 9110 9140 9180 9181 only C 6 Approvals A 15 Block diagram 9110 analyzer A 39 9120 9130 9140 9180 analyzer A 40 9181 analyzer A 41 Classifications Approvals A 15 Product data A 15 Cleaning solution 9110 only A 28 Clear all data B 69 Sample count B 69 Common operating parameter A 14 Contact addresses 4 Copyrights 4 Deproteinizer A 7 Description of modules Electronics Display board A 35 ISE SBC card A 33 Lamp board A 34 Fluidic Module A 27 Cleaning solution 9110 only A 28 Main tube set all models Intake tube set 9110 only A 28 Reference solution A 28 Sample Contacts 9110 only A 29 Standard A 27 St
133. nt results may lead to incorrect patient measurements which may result in incorrect clinical decisions possibly endangering the patient s health Mechanical assemblies Front door assembly A Needle unit assembly Roche Diagnostics For the 9110 20 30 40 and 80 analyzers the front door can be removed by moving the analyzer to the edge of a work surface so the analyzer door when opened will extend past the edge of the work surface With one hand hold the analyzer door near the right side hinge pin and with the other hand gently apply pressure to the middle rear area of the door This will allow the right hinge pin to clear the retaining hole in the main chassis The door can then be removed from the analyzer The 9181 front panel can be removed by tilting it slightly away from the analyzer and lifting it straight up The plastic window can also be replaced in the field by gently pressing from the rear of the window and snapping the plastic window out toward the front of the analyzer door The door magnets cannot be replaced in the field as they are glued in place with conductive adhesive Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new shielded front door FRONT DOOR 9180 is available and compatible with any 9180 Electrolyte Analyzer old type SN lt 11999 and new type SN gt 12001 and with 9120 30 40 Electrolyte Analyzers Further information see the 91xx Spare Parts List
134. nto the electrode module solenoid valve A and R are opened Standard A is then drawn from the ISE SnapPak until Sample Contact 1 detects Standard A bridging the T piece and Sample Contact 1 At this point solenoid valve V is opened and solenoid valve A is closed The standard is pumped up the tubing through the wash portand into the sample needle until Sample Contact 2 detects the standard The peristaltic pump slows down and the standard moves into the electrode module until detected by Sample Contact 3 bridge between sample preheater tube and reference electrode Then the pump stops When Standard A is to be aspirated into the electrode module solenoid valve A is opened and solenoid valves B V and R are closed Standard A is then drawn from the ISE SnapPak to the fill port through the sample needle and is sensed by the sample sensor At this point solenoid valve V is opened and solenoid valve A is closed as the peristaltic pump continues to pump Standard A into the electrode module As the trailing edge of the Standard A sample is sensed by the sample sensor the peristaltic pump stops During the time Standard A is aspirated into the electrode module the reference housing is pressurized due to the peristaltic pump providing pressure to the reference solution line and solenoid valve R being closed An amount of reference solution is forced out through the reference junction to provide contact to the sample The process for Standard
135. o Std A Std B slope YES Slope of all NO Electrodes in ange At least 1 slope in NO CHECK ELECTRODES YES YES READY 12 45 READY 12 45 XX CAL D Figure B 20 Roche Diagnostics 9120 9130 9140 calibration sequence April 2007 Service manual Version 2 0 B 29 7 Software operation 9120 9130 9140 91xx pH Electrolyte Analyzers Software operation 9120 9130 9140 Roche Diagnostics April 2007 B 30 Service manual Version 2 0 91xx pH Electrolyte Analyzers 8 Software operation 9180 Table of contents Software operation 9180 In this chapter Chapter Software operation 9180 33 OVERVIEW ip splen Seda dash B 33 Overall program operation B 34 Main men B 35 Measurement sequence sss nnne nnne B 36 Print functions menu QC standard dialysate urine sample menu B 38 Daily maintenance menu B 39 Operator functions menu 40 PLOSTAMANSEFUM ENE Menu ssn annene PURSE Febre bte iden 41 Service f ncti nis menu iip Pa OO RE ipta Pei B 42 Calibration sequence 1 Example Na K Li calibration B 43 Calibration sequence 2 Example Na K Li calibration B 44 B 45 Roche Diagnostic
136. onic diagrams Display daughter PCB 9180 SAMPLE SENSOR JP1 5VA 5VA SAMPLELED SAMPLEPT SAMPLEGND TEMP RJ12 MOLEX 15 43 8566 Electronic diagrams 2 LCD 1b DISPLAY 2 3b 4p VCC 5b 5VA 6b JP4 sb VCC i o hI 4B CONTRAST ds db DATA INST gt 126 DISPR W ds 135 DDO DDO 26 14 7 00 15 8 b 2 N DD2 16 P 21 DD3 17 O 10 N DD4 18 p O 11 DD5 19 O 12 DD6 20 p DD7 O 13 21 7 N 44 22 p DDIO 7Z 23 p MOLEX 24 p 26 p 24V 27 p PWB 28 p 29 P 5 30 54 31 2b YES 2 1P1241 1082 3 SCHURTER HEADER 34 NO 52 NO i 2 IP1241 1082 3 SCHURTER Title SCHEMATIC DISPLAY BOARD 9180 Size Document Number Rev A ZD5016 Date Monday July 22 2002 heet 1 of 1 Figure A 32 Display daughter PCB 9180 Roche Diagnostics April 2007 Service manual Version 2 0 A 67 5 Electronic diagrams 91xx pH Electrolyte Analyzers Electronic diagrams Roche Diagnostics April 2007 A 68 Service manual Version 2 0 Software operation Software operation 9110 B 3 Software operation 9120 9130 9140 17 Software operation 9180 B 31 Software operation IIS L rennen sen B 47 SEVICE TUNC
137. or clots or crystals formed in the Standard A tubing or electrode chamber Replace peristaltic pump tube set to ensure proper sample aspiration Ensure that the preheater cable 9110 only or the sample sensor is plugged in securely and perform C amp TEST SAMPLE CONTACTS amp TEST SAMPLE SENSOR to ensure correct operation Clean adjust or replace sample sensor to correct sample sensor operation Check the fluid remaining in the ISE SnapPak M If less than 5 remains replace the ISE SnapPak STANDARD B NOT DETECTED STANDARD C NOT DETECTED 9180 9181 only The system for drawing Standard B or C is like that of Standard A If Standard A is drawn properly and Standard B or C is not detected check Standard B tubing for crystallization Perform checks for Standard A detection problems to correct During the calibration sequence Standard B or C is drawn into the measuring chamber prior to Standard A therefore check if Standard A calibrant can be properly detected to isolate possible fluidic faults If neither Standard A B nor C are detected check for air leaks or blockages within the fluidic path CHECK SAMPLE CONTACTS 9110 CHECK SAMPLE SENSOR 9120 30 40 80 81 The sample contacts must provide a reading of 135 145 when air is detected and the sample sensor must provide a reading of 80 120 when air is detected Perform TEST SAMPLE CONTACTS Gr TEST SAMPLE SENSOR to observe sample contact or sample
138. ostics A 47 Version 2 0 Service manual 91xx pH Electrolyte Analyzers SBC PCB analog selector and A D converter 9110 9120 9130 9140 5 Electronic diagrams Electronic diagrams 6 p 7 19919 200210 sunp owg 3 500502 v e Jequinn XXL6 GYVOd OLLVW3HOS enu Nnuav 390v NagHav N3g1av qu AUT ATIHLVHWIXONddV T FALON S S ZN Ls 601 121 ML 52 sii z 028 0PL9N 100950 amp ano Nioso LZ zanas A00S Z ano o y 1noosoJ4n8 T 100 NIA EE 185580 x ve ten anas MOH NAS 82 ls 06 Sa 00 aL 1891 A 999 099 99 SNLvLsav gt z 50115 Jod Lsor L SIvddv IE S m Ep 3 0 30081 5 6 P M v asav Dia 62 521 m m L 68 dvO 438 5 Ag o 88 dvo4ay H amp E 259 x 18 s i 9x 98 zu x vlvddv x NOWNOO see 450 zx Let IR A I8 HE gt i x ox HE ven 193 sm U olivaav WEINSANOO
139. otective packaging only This will prevent further damage that may result in misinterpretation of the original cause of the fault The ESD mat consists of materials that have a very low defined conductivity 1012 ohms The following materials prevent charges caused by friction from building up and protect the component from damage ESD mats ESD packages e Shipping containers Conclusion Of course not all circuit boards and electronic assemblies require such careful handling An electrical board that only carries simple plug connectors does not require ESD packaging In cases of doubt always use ESD packaging Roche Diagnostics April 2007 Service manual Version 2 0 A 9 1 Safety information 91xx pH Electrolyte Analyzers ESD protection measures Roche Diagnostics April 2007 A 10 Service manual Version 2 0 91xx pH Electrolyte Analyzers 2 Specifications Table of contents Specifications In this chapter Chapter 2 SpPecificatiohs m ukqa RHET IIS A 13 Specificationis 13 13 9120 9130 9140 9180 9181 LI assays teet dedans 13 Operating Parameters csscsiscsssvestsivessserssocsncueasecbestvepseoestavevevesevavecsnavsvesevevesevenseanss A 13 13 9120 9130 9140 9180 9181 ott rr ger tgo 14 Common operating parameter A 14 Classifications se a A 15 Product
140. ow 135 0 YES Klow 350 YES Normal Ranges Na high 150 NO YES Program YES Reset Corr Factor Input Verify Remain in Na b 00 0 YES K b 00 0 Corr Factors default Corr Factors ee Na m 1 0000 K m 1 0000 NO YES YES Program Bicarb YES Reset Bicarb Input Verify Remain in Na b 00 0 YES Corr Factors Factor default Bicarb Factors sl Na m 1 0000 NO NO YES See Bicarbonate NO i YES Program Urine See Bicarbonate Corr Factors NO K b 00 0 1 0000 YES YES Program Printer YES Patient Report NO Print two NO Automatic NO Enter Comment YES Enter Comment Setup off EM Patient Reports EM Cal Report Line NO YES YES Program YES Activate Data Interface Link NO YES NO Remain in NO READY 12 45 YES Program Func Figure B 41 9181 program instrument menu Roche Diagnostics April 2007 B 58 Service manual Version 2 0 91xx pH Electrolyte Analyzers 9 Software operation 9181 Software operation 9181 Service functions menu SERVICE NO CALIBRATION YES Test YES Test Electrodes Standard A same for Std A B Test YES introduce Retract Probe gt External Sample YES Remain in Ext 0000 Test Electrodes K 0000 CI 0000 Test Sample Sensor Test SnapPak Sensor Test Language Switch Test Probe YES NO SnapPak 1 I In O
141. ple door Perform a calibration to return to the READY Mode STANDARD A NOT DETECTED This message is displayed when sample contact 1 9110 only or the sample sensor is unable to properly detect the Standard A solution in a programmed time period As the analyzer draws a sample of Standard A the leading edge of the solution is detected by sample contact 1 by bridging solution between both parts of sample contact 1 or the sample sensor The sample sensor must also detect a continuous flow of Standard A Solution into the sample chamber lasting several seconds without encountering air bubbles or air After several seconds the trailing edge of the Roche Diagnostics April 2007 Service manual Version 2 0 D 5 13 Troubleshooting 91xx pH Electrolyte Analyzers Error messages and troubleshooting Standard sample is sensed the peristaltic pump stops and measurement occurs If aspiration fails the analyzer will wash and repeat up to 3 times Possible causes and remedies Check for air leaks preventing the Standard from being drawn into the chamber Ensure that all O rings are in place There must be a continuous draw of standard solution free of air bubbles Check solenoid V for proper sealing of the vent line Check fill port to ensure the absence of air leaks Verify that the sample 15 not bent Check replace fill port and wash port to ensure the absence of air leaks 9110 only Check f
142. qv So za Lavd tvav St Wea HL 0 dvs 0v0242 50 02 2161 sor 23 11 002 usarezv 14 5 24 os dan 35 te ezt s5gzeso zzv saa at se t e as soags5nzv gt gt REN S T m vOd ZSSI 0Zv z gt gt aada BOLBPAN T 7929 sd PME Fra en h sy zz m S az 29 Ez ts ZV Ziv t 3 534 29 eso 25 HE stv 6 sv vaqa i 139 Foz BIN ziv a ON zw ESE ev w 13530 585 Q Q Q ttv tly ttv Le 21 ziv dad veu zou ary der a biy M b ANI ev Div 16 LIE ev Lee DIV IE ev Gv ev ev Lee oy 99 99 P 2v se 2v 56 BV 10 2v 66 sv Sv S2 jv Sig a ay LE ZV Sid er ga SE ZV Sa 21 5 IV 2Y 12 ov Yid et ea Sv via L sa Sv id eo s Sv S ev a 2L ya vv ES Sv Eid 6 sv 6 po vv 6 SV Sv s vv anto Zid eq ev Zid st eq LS vv co 8 wy 8s9 x L ev trast EZ EV c ev 6 9rd et iy 8 av o
143. ra et 9 w F ev 010 581 to tv Por av y fy sa he TY sa 6g Mes LY Sd V IEEE IA 6 ee Sg Sd i Ov sat Ov Sd i ow Ov 06 m Y mm tala E nao tar 01 SBC PCB microprocessor 9180 9181 Figure A 28 April 2007 Roche Diagnostics A 63 Service manual Version 2 0 91xx pH Electrolyte Analyzers 5 Electronic diagrams Electronic diagrams 9180 9181 microprocessor SBC PCB S um pe LL 3eeug 666 EO Kepun W lt 9 510507 x Jequiny dii 1816 0816 QUVOR OLIVW3HOS ids anu aa1adWvs gt gt AL S 66H gt gt laws SOLOHPL Se gt gt uisiav THN GE 5 gt gt 5 3 3 OL 4 Ne 5 gt uazzna 8 16H se gt gt U olivaav 100t dd3c2W d Z S6SOHPL L L vrvqasa 064 E gt gt uisia 18 99 A vr gt gt Br 1 Lolaa PESOHPZ u 992181 TALNA gt gt INS W p Sioa a eu gt gt uisud s100
144. re B 8 9110 program instrument menu Roche Diagnostics April 2007 Service manual Version 2 0 B 13 6 Software operation 9110 91xx pH Electrolyte Analyzers Software operation 9110 Service functions menu NO SERVICE CALIBRATION FUNCTIONS YES Test YES Test YES Electrodes Standard A Test Standard B pH 0000 Test Open Door and External Sample Introduce Sample Remain in Ext Wipe Probe and Test Electrodes pH 0000 Close Sample Door NO Test DRY 135 145 140 External 140 wen Sample Sensor YES Pump NO Exit Pump NO OFF Test YES Sample Door Sample Door C Closed O Open NO Test Pump YES NO Test Valves YES Test Valve A NO Test Valve B NO Test Valve V NO Test Valve R YES NO ON OFF YES NO ON OFF YES NO ON OFF YES NO ON OFF NO NO YES Test Interface Test eS Test Interface ok Introduce Sample Test Pump Very Slow Speed Test Pump Fast Speed Test Pump Medium Speed Test Pump Slow Speed Jump Pins 2 amp 3 Interface Jump Pins 2 amp 3 NO Test Amplifier YES Haia NO Ground 2 Temp mV 3672 pes 2500mV 2496 Temp C 37 1 NO YES Enter Code w Is Code EP Remain in Enter NO AAA ok Accepted Service Code Enter YES Service Code Code Error READY 12 45 Remain in Remain in YES Service Func Service Func Calibrate Figure B 9 9110 service functions menu
145. ril 2007 Service manual Version 2 0 5 91xx pH Electrolyte Analyzers Service functions menu B 14 10 Service functions Calibration sequence 15 Service functions B 65 TEST ELECTRODES B 65 7 Software operation 9120 9130 9140 TEST SAMPLE SENSOR B 65 Software operation 9120 9130 9140 B 19 TEST SnapPak SENSOR 9180 and 9181 only B 65 Overview B 19 TEST LANGUAGE SWITCH 9180 and 9181 only Overall program operating flow B 20 B 65 Main menu B 21 TEST SAMPLE DOOR except 9181 B 66 Measurement sequence B 22 TEST PROBE 9181 only B 66 Print functions menu B 23 TEST AUTOSAMPLER 9181 only B 66 QC standard urine sample menu B 24 TEST PUMP B 66 Daily maintenance menu B 25 TEST PINCH VALVES B 67 Operator functions menu B 26 TEST INTERFACE B 67 Program instrument menu B 27 TEST AMPLIFIER B 67 Service functions menu B 28 ENTER SERVICE CODE B 67 Calibration sequence B 29 Miscellaneous functions B 69 8 Software operation 9180 5 Software operation 9180 B 33 Maintenance Part Overview B 33 Overall program operation flow B 34 11 Adjustments Main menu B 35 Adjustments C 5 Measurement sequence B 36 Sample contacts adjustment 9110 only C 5 Print functions menu B 37 Sample sensor adjustment 9120 9130 9140 9180 QC standard dialysate urine sample menu B 38 9181 only C 5 Daily maintenance menu B 39 Temperature Adjustment 9110 9140 9180 9181 Operator functions menu B 40 only C 6 Program instrument menu B 41 Service functions menu B 42
146. rinter assembly A 25 Product data A 15 Rear panel assembly A 25 Reference solution A 28 Safety information Important information A 5 Operating safety information A 6 Sample Contacts 9110 only A 29 Sample contacts adjustment 9110 only C 5 Sample sensor adjustment 9120 9130 9140 9180 9181 only C 5 SBC board A 26 Service functions B 65 ENTER SERVICE CODE B 67 TEST AMPLIFIER B 67 TEST AUTOSAMPLER 9181 only B 66 TEST ELECTRODES B 65 TEST INTERFACE B 67 TEST LANGUAGE SWITCH 9180 and 9181 only B 65 TEST PINCH VALVES B 67 TEST PROBE 9181 only B 66 TEST PUMP B 66 TEST SAMPLE DOOR except 9181 B 66 TEST SAMPLE SENSOR B 65 TEST SnapPak SENSOR 9180 and 9181 only B 65 Service Manual Abbreviations 13 Conventions used 10 Finding information 9 Roche Diagnostics How to use 9 IVD symbols 11 Other symbols 12 Symbols 10 Version 3 G 6 Software Version 3 G 5 Software operation 9110 B 5 Calibration sequence B 15 Daily maintenance menu B 11 Main menu B 7 Measurement sequence B 8 Operator functions menu B 12 Overall program operation flow B 6 Print functions menu B 9 Program instrument menu B 13 sample menu B 10 Service functions menu B 14 Software operation 9120 9130 9140 B 19 Calibration sequence B 29
147. rodes amp YES Remove Water YES Operator Func Inserted Plug Removed and Press YES NO YES lt m n READY 12 45 Pump Windings Do You Wish to Relieved DELETE all Data Shutdown Complete Turn Power Off DELETE DATA RESET FACTORS Figure B 7 9110 operator functions menu Roche Diagnostics April 2007 B 12 Service manual Version 2 0 91xx pH Electrolyte Analyzers 6 Software operation 9110 Software operation 9110 Program instrument menu PROGRAM SERVICE INSTRUMENT FUNCTIONS YES Enter Code mw CODE ERROR AAA RETRY NO Program QC YES Current Lot 0000 YES Print old Values NO New Lot YES Enter last 4 Dig Level 1 Ranges Change Lot and Statistics Delete old Data of Lot 0000 YES Lot Number ok NO Jt Program QC Current Lot 0000 Print old Values New Lot YES Enter last 4 Dig Level 2 Ranges Change Lot and Statistics Delete old Data of Lot 0000 YES Lot Number ok NO NO Program QC Current Lot 0000 Print old Values New Lot YES Enter last 4 Dig Level 2 Ranges Change Lot and Statistics Delete old Data of Lot 0000 YES Lot Number ok NO Program Normal Ranges YES Program YES Reset Corr Factor Input Verify Remain i Corr Factors default Corr Factors Corr Factors YES pH b 000 pH m 1 0000 Program Printer Patient Report Print two Automatic YES NO Setup off Patient Reports Cal Report Remain in YES Program Func NO READY 12 45 Figu
148. s April 2007 Service manual Version 2 0 B 31 8 Software operation 9180 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 B 32 Service manual Version 2 0 91xx pH Electrolyte Analyzers 8 Software operation 9180 Software operation 9180 Overview Roche Diagnostics Overall Program Operating Flow Main Menu Measurement Sequence Print Functions Menu QC Std Dialysate Urine Sample Menu Daily Maintenance Menu Operator Functions Menu Program Instrument Menu G Ch G N Service Functions Menu 10 Calibration Sequence 11 Power up Sequence Software operation 9180 April 2007 Service manual Version 2 0 B 33 91xx pH Electrolyte Analyzers 8 Software operation 9180 Software operation 9180 U31vugrivo amp NOLIVHSITIVO Overall program operation flow 3und aS ul po 19113 1591 9 Z Suid duinr 1591 5 1891 100g ejdues sel youms eBenfBue s l 1osu s Yeddeus s l Josuas ejdures 1591 s pono la 1591 SNOILONNA 80 35 Dun ul dnyes i luliq 2510 4 1100 amp S1019 4 1109 ayejeoy 81 1102 25103004
149. s 9 Software operation 9181 Software operation 9181 Calibration sequence 2 example Na K Li calibration At least 1 lope in Range CGHECK ELECTRODES Asp Std A into Meas Chamber Standard A STANDARD A detected NOT DETECTED YES Aspir Std C into E a NAKLI READY Meas Chamber ee ond XX XX NOT Electrodes Standard C detected YES Obtain Electrode Voltage Std C Perform Wash Cycle Aspir Std A into Meas Chamber Standard A detected Obtain Electrode Voltage Std A Calc of Std A Std B slope NO STANDARD C NOT DETECTED All Std A volt Reprod NAKLI READY XX XX NOT CAL D YES NAKLI READY STANDARD A NOT DETECTED Figure B 44 9181 calibration sequence 2 example Na K Li calibration Roche Diagnostics April 2007 Service manual Version 2 0 B 61 9 Software operation 9181 91xx pH Electrolyte Analyzers Software operation 9181 Power up sequence 9180 A80X2 AA WARM UP All dots lit for sec Date 21 FEB 07 Time 12 00 Set Time Date Power Fail 20 FEB 07 1 12 Print i Background check in all menus ack in place except Take out of Operation and Service Functions YES Perform Daily Perform Daily Remain in Daily 5 Calibratid due ES Cleaning Conditioning Maintenance see YES lean Cond
150. scribe safe procedures that are intended to provide the user with additional help Caution All sections or text passages that are marked with this symbol describe procedures or indicate conditions or dangers that could damage or lead to malfunctions in the instrument and which therefore should never be attempted Warning Sections marked with this symbol contain information that must be observed for the prevention of personal injury to patients users or third parties Risk of infection according to the standard DIN EN 61010 2 101 2002 Instrument Risk of infection according to the standard DIN ISO 15223 1 2005 Consumables High Voltage Passages that are marked with this symbol warn of an immediate danger in connection with electrical wiring or components ESD protection measures All sections or passages that are marked with this symbol warn of specific dangers in connection with static discharge Packages that are marked with this symbol must only be opened by trained technical staff For more information about ESD protection measures see ESD protection measures on page A 8 April 2007 10 Service manual Version 2 0 91xx pH Electrolyte Analyzers IVD symbols Roche Diagnostics Instruments applicable for 9180 Electrolyte Analyzers SN gt 12001 The IVD symbols are used in accordance with DIN EN 980 2003 ISO 15223 2000 as per the In Vitro Diagnostics Directive 98 79 EC and DIN
151. should read between 10 mV and 10 mV with the reference voltage reading between 2490 mV and 2510 mV With all inputs open the display should read 4095 for each channel On the 9140 9180 9181 an additional display indicating the temperature voltage and temperature in degrees C is available to verify the temperature adjustment On the 9110 the temperature reading should be between 36 8 and 37 2 C Na K GND 0 Temp mV C1 Ca Li 2500mV 2499 Temp C depending on parameter configuration 9180 9181 only pH 9110 only April 2007 Service manual Version 2 0 B 67 10 Service functions 91xx pH Electrolyte Analyzers Service functions ENTER SERVICE CODE Service Codes are available and allow additional programming features Following isa listing of the available Service Codes and functions Set Reset Service code description ALL LLA This service code will enable the value during auto 9181 only matic serum blood measurement Special precautions are necessary to obtain the correct value due to sample exposure to room air CDC Clears all Service Codes 9110 Except SLC DEC CED Provides one additional digit of resolution for each measured parameter for blood and serum samples QC and Standard samples are always displayed with high resolution Urine samples are always displayed in low resolution ECO OCE Allows
152. sobuey jeuuoN s Bueuti e 3497 s Buevi z sobuey 8197 20 9po2 19111 Sund 10 81 40 ul Jo exer amp Jequinw jeseu 199 95 195 poN 05 SNOILONNA oSueu lu elN ul Suliuonipuoo Aeq uuojeg BSulue o Aeq wiopad 3ONVN31NIVIA elduies euunypis OO ldues uun SIEISIY ldues eyeuoqiesig ldues piepuels ldues OO 90 20 3 1dNANVS I1VSA1IVIQ ALS ID suonounj lulid 2525115 pue OO Wud 180 uod y a dwes jud SNOILONNA AQV3H IO 9180 program flow Figure B 21 April 2007 Roche Diagnostics Version 2 0 Service manual B 34 91xx pH Electrolyte Analyzers 8 Software operation 9180 Software operation 9180 Main menu NA K CL READY NO PRINT FUNCTIONS NO QC STD DIALYSATE URINE SAMPLE NO DAILY MAINTENANCE NO Legend Display prompt Action performed by analy
153. ssages will be displayed when a specific electrode does not calibrate properly during a calibration sequence If more than six standard draws are required or the difference between Standard A and Standard B voltages is out of range the NOT CALIBRATED message will be displayed for each channel which exhibits either of these conditions A copy of the calibration report should be printed to provide the electrode output voltages measured during calibration The electrode voltage range table identifies correct voltage ranges for each electrode parameter If the difference between Standard A and Standard C 9180 9181 only is out of range for Li or LI CALD will be displayed Possible causes and remedies e Check to ensure calibration reagents are transported correctly Test for leaks blockages bubbles or improper pump flow rate Check or replace ISE SnapPak Check whether electrodes are plugged in properly Check the slide mechanism to ensure it is in the correct position and firmly pushed into electrode connectors Clean and or replace electrode If A C is out of range for repeat the calibration or replace the ISE SnapPak Check or replace SBC Card ELECTRODE VOLTAGE RANGES 9110 Electrode Standard A Standard B Allowable Difference A B pH 2100 to 100 1600 to 900 430 to 650 Table D 1 9110 electrode voltage ranges ELECTRODE VOLT
154. stics April 2007 14 Service manual Version 2 0 System Description A QUQ Safety seb SDOCITCULIONS Interface specifications Electroni TOTIS 4 e dh t ase April 2007 91xx pH Electrolyte Analyzers 1 Safety information Safety information Roche Diagnostics In this chapter Table of contents Chapter a Important information 5 Operating safety information A 6 Important notes and warnings KEES E A 6 Disinfectants E 7 7 Other 7 ESD pr tectionmimeas resu 8 Explanation of the phenomenon A 8 Influence of electrostatic charges on components 8 Why is ESD protection so important today A 8 How can ESD protection be guaranteed A 9 Conclusion eet ep oH ER RR 9 April 2007 Service manual Version 2 0 A 3 1 Safety information 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 A 4 Service manual Version 2 0 91xx pH Electrolyte Analyzers 1 Safety information Important information Important information
155. structions for these procedures can be found in the Instructions for Use describing the correct method to perform routine analyzer maintenance Daily maintenance Perform cleaning cycle O Perform conditioning cycle Weekly maintenance O Clean sample probe and fill port Clean analyzer surfaces Monthly maintenance O Clean reference electrode housing O Check level of pH Filling Solution 9110 only Six month maintenance O Replace peristaltic pump tubing Annual preventive maintenance Replace complete tubing set Replace peristaltic pump tubing Replace intake tubing 9110 only Replace fill port assembly Replace sample probe Replace sample sensor Quad ring 9120 30 40 80 81 only Replace preheater Quad ring 9110 only Check sample sensor voltage 9120 30 40 80 81 only Check sample contact voltage 9110 only Check sample temperature 9140 80 81 only Roche Diagnostics April 2007 Service manual Version 2 0 9 12 Maintenance 91xx pH Electrolyte Analyzers Maintenance Roche Diagnostics April 2007 C 10 Service manual Version 2 0 Troubleshooting D D 3 April 2007 91xx pH Electrolyte Analyzers 13 Troubleshooting Table of contents Troubleshooting In this chapter Chapter 13 Error messages and troubleshooting esses D 5 Status LEDS D 5 STATUS NO
156. te e e eiit ee ost cbe conde ua B 67 TEST INTERFAGE dae ER RC GEORG EE eH pee B 67 TEST AMPLIFEIER 67 ENTERSERVIGE CODE 3 68 Miscellaneous f nctiOns io oor tete i eoe t ree ee ETE T OUS B 70 Deleting data seen een rrr rre P tinte B 70 e re ee en B 70 Roche Diagnostics April 2007 Service manual Version 2 0 B 63 10 Service functions 91xx pH Electrolyte Analyzers Table of contents Roche Diagnostics April 2007 B 64 Service manual Version 2 0 91xx pH Electrolyte Analyzers 10 Service functions Service functions TEST ELECTRODES Example displays Service functions The Service Functions mode provides a menu of service test functions to assist in failure diagnosis and identification A logic flow chart is provided for reference of the sequences included in the Service Functions mode This mode allows for the measurement and display of electrode voltages Standard A Standard B or an external sample can be measured to determine electrode performance and provide measurement data useful in locating electrode calibration faults Each electrode parameter will be displayed and the electrode voltage associated with the sample type selected can be observed St A Na 1582 St B Ext Na 1321 K 654 C1 904 pH 808 K 598 C1 127 9110 pH Analyzer TEST SAMPLE SENSOR Example d
157. to the default setting CSS has to be activated CHP PHC This code allows for the display and printout of pH results in concentration of H ions The cH will be shown in addition to the pH value cH is not shown on QC values and statistics reports Table B 1 Service code descriptions Roche Diagnostics April 2007 B 68 Service manual Version 2 0 91xx pH Electrolyte Analyzers 10 Service functions Service functions Set Reset Service code description EXW WXE 9110 only This code extends the wash cycle following a serum or blood measurement Normally 10 Standard A segments are used for the wash cycle with EXW set 20 Standard A segments will be used NTD DTN 9110 only When the code TMP is set NTD prevents the temperature from 9110 only being reset after each sample and causes the last temperature value to remain until changed TMP PMT 9110 only When this code is set the CHANGE PATIENT TEMPERATURE menu will appear in the sampling process The analyzer will perform a temperature correction calculation to convert the measured results for the desired patient temperature The temperature will be reset after each sample to 37 C unless NTD is entered Temperature corrections can only be performed in the range of 15 45 C Table B 1 Service code descriptions Code Description 9110 9120 9130 9140 9180 9181 ALL Enables during automat blood X
158. trolyte Analyzers Valve drivers A 58 System interconnect A 56 Block diagram 9110 analyzer A 39 9120 9130 9140 9180 analyzer A 40 9181 analyzer A 41 System interconnect 9110 9120 9130 9140 A 42 Electronics A 33 Display board A 35 ISE SBC A 33 Lamp board A 34 Error messages D 5 T D 11 TTTT444 CHECK ELECTRODES D 9 CHECK REFERENCE HOUSING D 7 CHECK SAMPLE CONTACTS 9110 D 6 CHECK TEMPERATURE 9140 9180 9181 only D 12 CLEANING FLUID NOT DETECTED D 8 CLOG IN SAMPLE PATH 9110 only D 11 CONDITIONING FLUID NOT DETECTED except 9110 D 8 ERR D 12 ERROR LOWER NEEDLE SENSOR D 13 UPPER NEEDLE SENSOR D 13 INTERFACE ERROR D 9 SAMPLE D 8 NOT CALD D 9 JAM OR PRINTER DEFECT D 9 PERFORM DAILY MAINTENANCE except 9110 D 12 PLEASE CLOSE SAMPLE DOOR except 9181 D 7 REPLACE FLUID PACK D 12 SAMPLER JAMMED OR DEFECTIVE D 14 STANDARD A NOT DETECTED D 5 STANDARD NOT DETECTED STANDARD C NOT DETECTED 9180 9181 only D 6 STATUS NOT CALIBRATED D 5 VALVE OVERTEMP CHECK VALVES D 13 WHEEL MISSING OR SAMPLER DEFECT D 13 ESD protection A 8 Conclusion 9 Explanation of the phenomenon 8 How can ESD protection be guaranteed A 9 Influence of electrostatic charges on components A 8 Why is ESD protection so important today A 9
159. ut Tray moves to cup position Probe moves to lower position Test Sampler NO Test Pump YES Test Pump 2s 2s 2s Test Pump Very Slow Spegd Fast Speed NO 2s Test Valve NO Test Valve V NO NO YES NO Jump Pins 2 amp 3 Interface YES NO ON OFF NO NO Test Amplifier NO Enter Service Code NO Enter Other NO Service Code Enter Code Is Code M valid 9 Code Error Retry in i NO YES Calibrate Figure B 42 9181 service functions menu Code Accepted NAK CA READY Roche Diagnostics April 2007 Service manual Version 2 0 B 59 9 Software operation 9181 91xx pH Electrolyte Analyzers Software operation 9181 Calibration sequence 1 example Na K Li calibration CALIBRATION 5 Asp Std A into Meas Chamber Standard A detected Obtain Electrode Voltage Std A STANDARD A NOT DETECTED CHECK SAMPLE SENSOR Sample Sensor Voltage ok Are Values reproducible Standard B detected STANDARD B CHECK NOT DETECTED ELECTRODES Calc of Std A Std B slope CHECK ELECTRODES Pressurize Ref Aspl Std A into Meas Chamber Reference 1st t i Figure B 43 9181 calibration sequence 1 example Na K Li calibration Roche Diagnostics April 2007 B 60 Service manual Version 2 0 91xx pH Electrolyte Analyzer
160. xpose the rear panel The power supply module located on the rear panel assembly can now be removed by removing the three screws securing the circuit board and by disconnecting the wiring April 2007 Service manual Version 2 0 A 25 4 Description of modules 91xx pH Electrolyte Analyzers Description of modules SBC board P P 9 5 Roche Diagnostics interconnections The main power receptacle can also be replaced by removing the two screws holding this assembly Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new power supply module POWER SUPPLY 9180 gt SN 12001 is available and compatible with 9180 Electrolyte Analyzers serial number gt 12001 only The former power supply MODULE POWER SUPPLY ISE is still available and has to be used for 9180 Electrolyte Analyzers SN lt 11999 for 9110 pH Analyzers and for 9120 30 40 and 81 Electrolyte Analyzers 8 Further information see the 91xx Spare Parts List After changing the main board or the power supply always perform a measurement of protective ground resistance to ensure a safe connection of the system ground Always measure with the power cord attached to the analyzer The resistance measurement has to be done between the ground pin of the power plug and several ground connectors placed on each of the analyzer s electronic components such as power supply or main board and shielding The total resistanc
161. yte Analyzers Contents Version History Edition notice Intended use Copyrights Trademarks Contact addresses Contents Preface How to use this manual Where to Find Information Conventions used in this manual System Description N NO XO WW W 5 3 EJ 1 Safety information Important information Operating safety information Important notes and warnings Disinfectants Deproteinizer Other disinfectants ESD protection measures Explanation of the phenomenon 5 A 6 A 6 A 7 A 7 A 7 A 8 Influence of electrostatic charges on components A 8 Why is ESD protection so important today How can ESD protection be guaranteed Conclusion 2 Specifications Specifications Specifications Operating Parameters Classifications 3 Interface specifications Interface specifications Interface information Example data string information Data Link Information 9180 9181 only 4 Description of modules Description of modules Mechanical assemblies Fluidic Module Electronics 5 Electronic diagrams Electronic diagrams 9110 analyzer block diagram 9120 9130 9140 9180 analyzer block diagram 9181 analyzer block diagram Roche Diagnostics A 9 A 9 A 9 A 13 A 13 A 13 A 15 A 19 A 19 A 20 A 20 A 23 A 23 A 27 A 33 A 39 A 39 A 40 A 41 System interconnect 9110 9120 9130 9140 A 42 SBC PCB power supply 9110 9120 9130 9140 A 43 SBC PCB valve drivers 9110 9120
162. zer 2 second display prompt Decision making by analyzer QUQL OPERATOR FUNCTIONS NO PROGRAM INSTRUMENT NO SERVICE FUNCTIONS NO CALIBRATION Figure B 22 9180 main menu Roche Diagnostics April 2007 Service manual Version 2 0 B 35 8 Software operation 9180 91xx pH Electrolyte Analyzers Software operation 9180 Measurement sequence K CL READY PRINT FUNCTIONS Open Sample Door Introduce Sample lt air voltage lt air voltage 2s 2s Please Close Please Close Sample Door Sample Door FLAP CLOSED FLAP CLOSED Serum Sample Blood Sample in Progress 49 in Progress 49 Na K Na K Cl 151 5 1 114 151 5 1 114 YES NO YES NO 1 min 1 min Figure B 23 9180 measurement sequence Roche Diagnostics April 2007 B 36 Service manual Version 2 0 91xx pH Electrolyte Analyzers 8 Software operation 9180 Print functions menu PRINT FUNCTIONS NO Print Last Sample Report Print Cal Report NO Print QC Values YES and Statistics NO Remain in Print NA K CL READY Functions QC STD DIALYSATE URINE SAMPLE NA K CL READY Software operation 9180 Figure B 24 9180 print functions menu Roche Diagnostics April 2007 Service manual Version 2 0 B 37 8 Software operation 9180 91xx pH Electrolyte Analyzers Software operation 9180 QC standard dialysate urine sample menu QC STD DIALYSATE NO DAILY URINE SAMPLE
Download Pdf Manuals
Related Search
Related Contents
here - Alcor SG20KTL User Manual-EN ajustes - Challenger 添付文書等記載事項の届出等に関するQ&Aについて Viewsonic ViewPad 10s 0.5GB Black Le paiement électronique sur Internet Meditec Produktkatalog 2013 als PDF GL GL800 Copyright © All rights reserved.
Failed to retrieve file