Home
GE Medical Systems
Contents
1. 7 L GE HEALTHCARE DIRECTION 5176235 100 Rev 1 PET vqc z axis roll PET ctac conv scale PET image set id PET constrast route PET image one loc PET image index loc PET num of slices PET rest stress B 2 5 Discovery PET Frame Module PET num of rr interval PET num of time slots PET num of time slices DiscovERY DICOM CONFORMANCE STATEMENT 0009 10D0 0009 10D1 0009 10D2 0009 10D3 0009 10D6 0009 10D7 0009 10DD 0009 10DE 0009 10DF 0009 10E0 0009 10E2 Private Creator Identification GEMS PETD 01 3 1 1 1 1 L L 71 71 US US US 0 Attribute Tag Type VR VM PET frame_id PET scan_id PET exam_id PET patient_id PET compatible_version PET software_version PET where is frame PET frame_size PET file exists PET patient entry PET table height PET table z position PET landmark datetime PET slice count PET start location PET acq delay PET acq start PET acq duration 0009 105C 3 Ul 1 154 of 169 xww ea ey x 3 SH SH ST SL SL SL FL FL D
2. B 2 3 Discovery PET Exam Module sse einen ennt either nnns 14 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT B 2 4 Discovery PET Scan Module 151 2 5 Discovery PET Frame sss sessi nnne 154 B 2 6 Discovery PET ImageSet Module isses 155 2 7 Discovery PET Image Module 157 2 8 Private Creator Identification GEMS 01 158 BASIC DIRECTORY IOD 1 444 001 159 C 2 DIRECTORY INFORMATION MODULE 159 DIRECTORY RECORD SELECTION KEYS eene entere 160 C 3 1 2 160 C 3 2 Study Kays soni En dod at ls e e nent abs tines 160 C 3 3 MIN um ie 160 C 3 4 uo NC 161 0 1 GE PRIVATE PET RAW DATA INFORMATION OBJECT 162 D 1 1 Discovery PET RAW Data Patient Entity 16
3. 56 3 5 1 2 SOP Specific Conformance Statement for MR SOP 56 3 5 2 Private Transfer Syntax Specification sse 56 3 6 gt CONFIGURATION wien Heer idet e d ota He eL auda He eo e in es Geek 58 3 7 SUPPORT OF EXTENDED CHARACTER SETS csccceeceeceeeeeeeaeeeeneeceeeeecaaeeeeaaeeeeaeeseaeeesaeeesaeeeeneesaas 58 4 PRINT SCU 58 4 1 EAE 58 4 2 IMPLEMENTATION MODEL PRINT entente 59 4 2 1 Application Data Flow Diagram 59 4 2 2 Functional Definition of AE S ois eret aer ED e LR D 59 4 2 3 Sequencing of Real World Activities esses 59 4 3 HIER ERR RE antes 61 4 3 1 DICOM Print SCU AE 61 4 3 1 1 Association Establishment 61 4 3 1 1 1 GOH CT alt 61 4 3 1 1 2 Number of ASSOCIALIONS 2 noire cR HRS 61 4 3 1 1 3 Asynchronous Nature ger ei pere te ere EN LO ea nt 61 4 3 1 1 4 Implementation identifying 61 4 3 1 2 Association Establishment 62 4 3 1 2 1 Real World
4. 20 1 5 IMPORTANT REMARKS s tetti eee e e tdt ie e A ds 20 1 6 REFERENCES tee settore iem ene odd oet dete etd oa to 21 1 7 kekscetoe 21 1 8 SYMBOLS AND ABBREVIATIONS sccccccccsesssesceececeesessnneaeeeseseeseseaeaeeececeeseneaeaeeececseesesssesaeaeeas 21 2 SECTION 2 NETWORK CONFORMANCE STATEMENT 1111 1 21 2 1 INTRODUCTION 22554 ae ec ee el eo oed es yeux 21 2 2 IMPLEMENTATION MODEL ni nsns ri sans n nnne nennen nnns 22 2 2 1 Application Data Flow Diagram esiste nnns 23 2 2 2 Functional Definition 6 24 2 2 3 Sequencing of Real World Activities esses 25 2 3 CAE SPECIFICATIONS 25 2 3 1 DICOM Server 5 25 2 3 1 1 Association Establishment 26 2 3 1 1 1 General iis PE 26 2 3 1 1 2 Number of ASSOCIATIONS i erret en rix 26 2 3 1 1 3 Asynchronous eee erf ret ette tee ecce i et wi aer cae 27 2 3 1 1 4 Implementation Identifying 27 2 3 1 2 Association Initiation by Real World 27 2 3 1 2 1 Push Image s to Re
5. 62 4 3 1 2 1 1 Associated Real World Activity 62 4 3 1 2 1 2 Proposed Presentation Contexts ccceccceeeeeeeeeeeeneeeeeeceeeseaeeseaeeseaeeseeeeeaeeteaeeeeaeene 62 4 3 1 2 2 SOP Specific Conformance 62 4 3 1 2 2 1 Basic Film Session SOP 63 4 3 1 2 2 2 Basic Film box SOP 63 4 3 1 2 2 3 Basic Grayscale Image SOP 2 2444400 400 0 65 4 31 2 2 4 Printer SOP Class eite eise ee RO e esposa a 65 431225 PrintJob SOP Class iustitie e 66 4 3 1 3 Association Acceptance Policy 4 4 4 0 4444 000 nnnm inen 67 4 4 COMMUNICATION 888888 67 4 4 1 Supported Communication Stacks parts 8 9 sse 67 4431 TGP IP Stack RUE ERR 67 4 4 1 1 1 Dir E 67 4 4 1 1 2 Physical M dia iei 67 4 5 STANDARD EXTENDED SPECIALIZED PRIVATE 67 4 6 GONFIGURATION tete ete PO POE Celo t decet tA Ue tento ok AEn 67 4 6 1 AE Title Presentation Address Mapping sse 67 4 6 2 Configurable Parameters esses aia iiaae 67 4 7 SUPPORT OF EXTENDED CHARACTER 68 5 DICOM STORAGE
6. Only the ype 125 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Lo SC Image Not sent consists entirely of M type 3 element Modality LUT Only the type 1C elements are sent Common 126 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT CT SC IOD Common Module Definitions A 3 1 Patient Module Attribute Name 744 Type Notes ______ Patient s Name 0010 0010 2 As entered at user interface or from worklist Supports 5 different components delimited by Supports a maximum length of 32 characters including the delimiter Patient ID 0010 0020 2 As entered at user interface or from worklist Patient s Birth Date 0010 0030 Always sent zero length Patient s Sex 0010 0040 Always sent A 3 2 General Study Module Attribute Name Type Nots 0 00 Study Instance UID 0020 000D 1 Generated uniquely for each exam Value from worklist if present is not filled in here Study Date 0008 0020 2 Generated for each exam and always sent Study Time 0008 0030 Generated for each exam and always sent 0008 0050 Value from user interface or worklist sent Referring Physician s 0008 0090 Value from user interface sent Name Study ID 0020 0010 Value from user interface or worklist sent Study Description 0008 1
7. at ae ees ee ae ice DET 48 3 4 2 1 3 Bile Set Updater iiem sep hea eee 48 3 4 3 Standard Application Profiles esses siens entente nnns 49 3 4 3 1 STD CTMR MOD12 Class Profile essssesssseeeeeeeneenneen nene nennen nennt 49 3 4 3 1 1 SOP Classes and Transfer 2 1 49 3 4 3 1 2 Physical Media and Media 50 3 4 3 1 3 Eogical Format dme Rae ARR HORN ERR eR pee aden ted 50 3 4 3 1 4 Directory Information in 51 34 31 41 Additional Keys cede ette tet eR er he E e rue Heide 52 3 4 3 1 5 Other Parameters om eerie deere ae asain ai eo deve leven 52 3 4 3 1 6 Image Attribute 004 52 3 4 3 1 6 1 Attribute Value 2 52 3 4 3 2 STD CTMR MOD293 Class Profile 0 0 2100000 0000 52 3 4 3 2 1 SOP Classes and Transfer Syntax s sse 52 3 4 3 2 2 Physical Media and Media 53 3 4 3 2 3 Logical Format eie ped guest BRI ERI 53 3 4 3 2 4 Directory Information in 53 34 32 41 Additional Keys nein eie tnn Phe er a Re nne 54 3 4 3 2
8. A43 Sois GewmSwes A44 Frame of Reference Frame of Reference aaia Equipment General Equipment A45 M lt mage _______ Ju maePae A amp TF image Pixel sor common A 1 2 CT Image Module Ta Notes lt lt lt lt lt Reconstruction Diameter 0018 1100 3 Value sent for all images except scouts Distance Source to Detector 0018 1110 Value always sent Distance Source to Patient 0018 1111 Value always sent Gantry Detector Tilt 0018 1120 Value always sent 123 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Filter 0018 1160 3 Sent Defined terms BODY FILTER MEDIUM FILTER HEAD FILTER Focal Spot 0018 1190 Sent Fixed value of 0 7 or 1 2 Convolution Kernel 0018 1210 3 Not sent for scouts Defined terms SMOOTH SOFT STANDARD DETAIL BONE BONEPLUS CHST EDGE SHARP LUNG Table Feed per Rotation 0018 9310 Sent for spiral scan only CT Pitch Factor 0018 9311 Sent for spiral scan only It is possible for the operator of CT LightSpeed system to change the table height while scanning a series of images Therefore implementations must use the Frame of
9. Worklist Server AE SCU will initiate only one DICOM association at a time to perform a modality worklist query of a single remote AE 6 3 1 1 3 Asynchronous Nature Asynchronous mode is not supported All operations are performed synchronously 77 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 3 1 1 4 Implementation Identifying Information The Implementation UID for this DICOM Implementation is ModalityWorklist for Discovery implementation UID 1 2 840 113619 6 50 6 3 1 2 Association Initiation Policy The Worklist Server AE initiates a new association due to an update operation being initiated from the Discovery user interface 6 3 1 2 1 Real World Activity Worklist Query 6 3 1 2 1 1 Associated Real World Activity The operator of the system initiates a query for a modality worklist by either opening the Schedule screen or by opening the Schedule screen and pressing the Update button The choice of which of these two behaviors occurs is user configurable The Worklist Server will then initiate an association with the remote AE in order to query for the worklist A user can configure a number of parameters that directly control the worklist query request The user can request worklist items that are intended for the scanner the user is working at all items that apply to the modality of the scanner the user is working at or all worklist items available These selections an
10. To REO P E AHS NSA R IER 5 BES AD ROR LIL ER LD ABT TRED 9 AMEE EWA dE GEMS EP EAA ACTF AAA SC BU ANETTA AE SS HEE PREDA Bk ft gt REVISON HISTORY REV DATE REASON FOR CHANGE 1 3 26 2007 Based on PF2CardiacDicomCS 3 26 2007 5 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT REV DATE REASON FOR CHANGE 1 continued 3 26 2007 6 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Page Left Intentionally Blank GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 1 TABLE OF CONTENTS 1 TABLE OF CONTENTS te eee ees 8 1 5 1 16 nep Max S iL 16 1 2 OVERALL CONFORMANCE STATEMENT DOCUMENTATION 18 1 3 INTENDED AUDIENGE Bo ce 19 1 4 SCOPE AND FIELD OF 1 nannten ennt asinis ast
11. APPENDIX D DICOMDIR DIRECTORY INFORMATION D 1 GE Private PET Raw Data Information Object Implementation This section specifies the implementation of GE private defined Information Object namely GE Private PET Raw Data IOD used in the GE PET Raw Data Storage SOP Class 1 2 840 113619 4 30 The implementation is detailed to the extent that vendor applications can recognize an instance of the object Storage devices that supports full fidelity storage of this IOD can use the data detailed here to optimize the storage and retrieval of the data or display useful information to the user to distinguish among the IOD instances The IOD contains data that are not published in this document No implementation shall assume any meaning to such data and its meaning and definition is subject to change without notice D 1 1 Discovery PET RAW Data Patient Entity Module This section specifies the attributes of the Patient that describe and identify the Patient who is the subject of a diagnostic Study This Module contains Attributes of the patient that are needed for diagnostic interpretation of the Image and are common for all studies performed on the patient Attribute Name Tag Type VR VM Notes Patients Name 0010 0010 2 64 1 Patient ID LO 64 Patient s Birth Date DA 26 Patient s Sex 0010 0040 2 5 16 1 D 1 2 Discovery PET RAW Data Exam Module This section specifies the attributes which
12. ee M pd e ta da 81 6 4 3 2 Physical Media SUppott 2 eem tbe 81 6 4 4 Poinbto PolnE SIaCK i eiat podia nop npn renes 81 6 5 EXTENSIONS SPECIALIZATIONS PRIVATIZATIONS ssescceseecccceeseeaeceseeececseseeaececeeseeseesssaeseeess 81 6 5 1 Standard Extended Specialized Private SOPS sse 81 6 5 2 Private Transfer isses einen 81 6 6 CONFIGURATION 81 6 6 1 AE Title Presentation Address Mapping seen 81 6 6 2 Configurable Parameters 82 11 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 7 SUPPORT OF EXTENDED CHARACTER SETS 110 11 0 82 6 8 MODALITY WORKLIST INFORMATION MODEL 82 6 8 1 T Ie Coro Wieso erated 82 6 8 2 Modality Worklist Information Model Description 83 6 8 3 Modality Worklist Information Model Entity Relationship 83 6 8 4 Entity Descriptions ie rt rta er tetra Deae bere og 84 6 8 4 1 Scheduled Procedure Step 2 oie mre etie p ren es 84 6 8 4 2 Requested Procedure Entity Description
13. or DISCONTINUE based on user selection 7 11 USE OF SPECIFIC DICOM DATA 7 11 1 Patient Level When PPS start NLCREATE message is sent this element will have the Attribute Name Patient Name Patient ID Patient Birthdate Ref SOPclassuid 0081150 RetSOPismmeud 909159 Usage in CT DICOM Image Header Tag 109 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 7 11 2 Study Level Attribute Name Usage in CT DICOM Image Header Study Instance UID 0020 000D Copied from SPS if selected Study ID 0020 0010 Scanner generated study ID Referring Physicians name 0008 0090 Copied from SPS if selected number from the first selection determined by Accession Number 0008 0050 If multiple SPS s are selected then accession the user is used Referenced Study Sequence 0008 1110 Copied from SPS if selected Ref SOP class uid 0008 1850 Ref SOP Instance uid 0 15 7 11 3 Series Level otherwise generated from user input on scanner Procedure Step Sequence HIS RIS is selected for scanning Requested Attribute 0040 0275 Will be present only if SPS obtained form Sequence HIS RIS is selected for scanning If multiple SPS s are selected then this will contain multiple items one for each SPS For all other cases this element will not be present Requested Procedure Id 0040
14. All the operations used by this SOP class support a Session Timer This timer is started when the association is established and stopped when association is ended The default time out value is 3000 seconds If any of the above timers expires the association is aborted and the operation in progress is considered FAILED 8 2 4 3 Association Acceptance Policy The GSPS AE does not respond to attempts by remote AE to open an association 8 3 COMMUNICATION PROFILES 8 3 1 Supported Communication Stacks PS 3 8 DICOM Upper Layer PS 3 8 is supported using TCP IP 114 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 8 3 2 OSI Stack The OSI Communication stack is not supported by this implementation 8 3 3 TCP IP Stack The TCP IP stack is inherited from the operating system 8 3 3 1 API Not Applicable to this product 8 3 3 2 Physical Media Support Ethernet 802 3 provides the physical network layer for this product 8 3 4 Point to Point Stack The Point to Point Stack is not supported by this implementation 8 4 EXTENSIONS SPECIALIZATION PRIVATIZATION 8 4 4 Standard Extended Specialized Private SOPs GSPS for Discovery does not implement any private SOP classes 8 4 2 Private Transfer Syntaxes GSPS for Discovery does not implement any private Transfer Syntaxes 8 5 CONFIGURATION The Exam Split feature is an option The DICOM configuration items below are configurab
15. This timer starts when the association request is sent or received and stops when the association is established The default time out value is 30 seconds Each C FIND operation supports a configurable Session Timer This timer starts when an association is established and stops when the association is ended The default time out value is 3600 seconds If any of the above timers expires the association is aborted A ABORT and the operation in progress is considered to be failed Any previously received worklist items are discarded 6 3 1 2 1 2 2 Record Acceptance Policy The Discovery implementation adheres to strict value checking of incoming query responses from the remote AE Each response received is examined to verify that all Type 1 attributes are present with non zero length that all Type 2 attributes are present possibly with zero length and that the data for all attributes is consistent with respect to the attributes value representation VR 79 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Any inconsistencies in the response data with respect to the categories described above are considered errors Upon detecting any such errors in the response data the Worklist Server AE will issue a C FIND CANCEL and upon receipt of a C FIND RSP or if an applicable timer expires will abort the association All previously received worklist items are retained Note that the absen
16. VR VM PET compatible_version 0009 10A4 3 SH PET software_version 0009 10A5 3 SH SL FL PET slice_number PET total_counts PET other_atts PET other_atts_size PET archived 3 PET bp_center_x 3 157 of 169 x Ke SL SL FL gt GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT PET bp center y 0009 10AC PET trans frame id PET tpluse frame id PET profile spacing PET ovlp frm id PET ovlp trans frm id PET tpulse frm id 0009 10CA PET ir z filter flag PET ir z filter ratio PET left shift PET posterior shift 0009 10E6 PET superior shift 0009 10E7 3D Filter flag 3D Filter cutoff 3D Filter order 0009 10EC Reformat group 0009 10 0 3 Ul B 2 8 Private Creator Identification GEMS PETD 01 158 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT APPENDIX C DICOMDIR DIRECTORY INFORMATION Enclosed here is a listing of only the optional conditional modules and optional attributes used by this implementation in the DICOMDIR definition All standard attributes as defined in Part 3 Addendum Basic Directory Information Object are supported by this implementation but not listed here Basic Directory IOD Definition Module Reference Usage Direc
17. Window Width 0028 1051 BEEN window width Window Center amp Width 1055 1 n Not Used Explanation A 4 21 Discovery PET Image Module Refer to Section B 2 7 for details A 4 22 Discovery PET Frame Module Refer to Section B 2 5 for details A 4 23 SOP Common This section defines the Attributes which are required for proper functioning and identification of the associated SOP Instances They do not specify any semantics about the Real World Object represented by the IOD Attribute Name Tag Type VR VM Notes SOP Class UID 0008 0016 1 Ul SOP Instance UID Specific Character Set 0008 0018 0008 0005 NULL current_date Instance Creation Date 0008 0012 0008 0013 0008 0014 Instance Creation Time current_time mla lalala x Instance Creator UID 3 UI 146 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT APPENDIX B PRIVATE DATA ELEMENTS Enclosed is a listing of private data elements used in this implementation for CT Image IOD definition CT Image IOD Private Data Elements Definition B 1 1 Private Creator Identification GEMS IDEN 01 Attribute Name Tag Product id 00094004 SH 1 Image actual date 0009 1027 BS CES Equipment UID 0009 10E3 Note For all images created by Discovery system 0009 xx01 element will have the value LIGHT
18. 0040 1003 0040 1004 0040 1005 0040 1008 0040 1010 0040 1400 0040 2400 0040 3001 na SH SH SH LO na na SH SH LO LO LO PN LT LT LO na na 1 no value available 0 0 SpecialNeeds no value available 0 0 CurrentPatientLocation value available 0 0 PatientState 92 4 1 ModalityWorklistGroupLength Sequence with explicit Length 1 136 1 ScheduledProcedureStepSequence Item with explicit Length 4 12 128 1 Item no value available 0 O Modality no value available 0 0 RequestedContrastAgent value available 0 0 ScheduledStationAETitle no value available 0 0 ScheduledProcedureStepStartDate no value available 0 0 ScheduledProcedureStepStartTime no value available 0 0 ScheduledPerformingPhysiciansName value available 0 0 ScheduledProcedureStepDescription Seq with explicit Length 1 32 1 ScheduledActionItemCodeSequence Item with explicit Length 3 24 1 Item no value available 0 0 CodeValue value available 0 0 CodingSchemeDesignator no value available 0 0 CodeMeaning ItemDelimitationItem for re encoding 0 1 ItemDelimitationItem SequenceDelimitationItem for 0 1 SequenceDelimitationItem no value available 0 0 ScheduledProcedureStepID value available 0 0 ScheduledStationName no value available 0 0 ScheduledPr
19. 34 2 3 1 3 1 2 Presentation Context Table 34 2 3 1 3 2 Verification Request from Remote 1 35 2 3 1 3 2 1 Associated Real World 35 2 3 1 3 2 2 Presentation Context Table sese nennen 35 2 3 1 3 3 Query Request from Remote 4 2 22 0 4 36 2 3 1 3 3 1 Associated Real World 2 36 2 3 1 3 3 2 Presentation Context Table sees enne nnne nnne 36 2 3 1 3 4 Retrieve Request From Remote 38 2 3 1 3 4 1 Associated Real World 38 2 3 1 3 4 2 Presentation Context Table ssesssseseseeneeeeeeeeenne nennen 39 2 3 1 3 4 3 Presentation Context Acceptance Criteria 2 39 8 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 3 1 3 4 4 Transfer Syntax Selection Policy eceeeceeeeeeeeeeseeeeeeeteeeeseaeeseeeeseaeeseeeeseaeeseeeeeaas 40 2 4 ei sees steaks eur ene Cate Pen o ce duras 40 2 4 1 Supported Communication Stacks parts 8 9 002000 41 0 4 00000000 40 24 2 Sack tet rer t ert wading 40 2 4 2 1 Physical Media eee RU En pie tet e e 40 2 4 3 Point tosPolht ee
20. A 4 18 PET Image Module This section specifies the Attributes that describe the image within a particular PET Series Attribute Name Tag VR Notes 142 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Image Type 0008 0008 3 5 1 ImageSet is_source Samples per Pixel 0028 0002 1 Photometric Interpretation 0028 0004 IERES MONOCHROME2 Bits Allocated 0028 0100 US image ES always Bits Stored 0028 0101 image at always High Bit 0028 0102 Rescale Intercept 0028 1052 Rescale Slope 0028 1053 DS Use ERI ImageSet units Frame Reference Time 0054 1300 DS IF frame_ref_time NULL THEN frame_ref_time ELSE image_time Trigger Time 0018 1060 DS 1 Scan scan_mode gated THEN image bin time ELSE NULL Frame Time 0018 1063 DS 1 Scan scan_mode gated THEN image bin dur ELSE NULL Low R R Value 0018 1081 1C IS 1 Scan lower reject limit High R R Value 0018 1082 1C IS 1 Scan upper reject limit Lossy Image Compression 0028 2110 1C CS 1 IF compression none THEN 00H IF compression lossy THEN 01 H Image Index 0054 1330 Scan scan_mode image_location 143 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT image_time image_bin_time Acquisition D
21. Manufacturer 0 00080070 Returned ____ Images in series 0 00201002 Returned Table 2 3 1 3 3 2 1 3 Supported image level keys Description ____ Type __ Usage Image number 00200013 Matched Image type 0 00080008 Returned Ros __ 00280010 Returned Columns 0 0028001 Returned image position O 00200032 image orientation O 00200037 _____ Slice thickness 00180050 Returned Slice spacing O 00180088 Returned Gantry tit O 00181120 Returned Convolution kernel 00181210 Returned Reconstructiondiameter 0 00181100 Returned _____ Data collection diameter ____ 00180090 Returned Flip angle 0 00181314 Returned Echo number O 00180086 Echo time Oo 00180081 Returned Inversion time O 00180082 Returned Repetition time ___0 00180080 Retumed Trigger time 00181060 Returned Dfov P 0019 001 Returned Midscan Time 00190024 Returned Azimuth P 00190026 Returned Number of Echo P 0019007E Returned 37 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Description 7 Type Tag Usage Scout Ane 0021004A Returned Location RAS P 00270040 Returned Location P 00270041 Returned Center R Coordinate 00270042 Returned CenterA
22. 3 iiid breites n pois teg tto eo dede ntes eve 98 7 2 3 Sequencing of Real World Activities essent 98 7 2 3 1 5 from Acquisition System with MWL data 98 7 2 3 2 PPS from acquisition system without MWL data essen 99 7 2 3 3 PPS from post processing 100 7 3 ABZSPECIFIGATION t 100 7 3 1 PPS Server AE Specification esses sienne nennen tnn en rennen 100 7 3 1 1 X Association Establishment 101 7 3 1 1 1 101 7 3 1 1 2 N mber of Associaltiors ee ceni tenent rene 101 7 3 1 1 3 Asynchronous Nature net GREC RR RR REGINE apean 101 7 3 1 1 4 Implementation Identifying 101 7 3 1 2 Association Initiation 1 enne nne 101 7 3 1 2 1 Real World Activity Performed Procedure Step creation and update 101 7 3 1 2 1 1 Associated Real World 101 7 3 1 2 1 2 Proposed Presentation Context Table 102 7 3 1 3 Association Acceptance 102 12 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATE
23. 56 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 DiscovERY DICOM CONFORMANCE STATEMENT The early release product software had a JPEG encode routine that is called the BAD ENCODER for the following explanation In the BAD ENCODER produced SOS OxFFDA header the entropy coding table selector codes are 0x11 They should be 0x00 This mistake is the key that specifies the BAD ENCODER processing problems The BAD ENCODER computation of the pixel value difference to be entropy encoded is computed as Predictor value when it should be calculated as value Predictor 1 The result is that the decompression via industry standard packages result in a negative of the original image pixel value The BAD ENCODER predictor value used at the beginning of each line uses the last value of the previous line in the image The JPEG standard says the predictor for the first line element of each line should be the first element of the line immediately above the current line The first line first element predictor should be the unsigned value that is half the full scale range for the sample precision The first line first element predictor is correctly determined in the BAD ENCODER Explanation This section is provided to explain what was done during Image encoding so that people who wish and need to decode the images can modify the available JPEG codec software These packages are available from the WEB and were originally developed at Stanfor
24. N CREATE Requests the Print SCP to create an instance of Basic Film Session The following Attribute values are supported by the N CREATE Attribute DICOM Tag Valid Range Default Value Number of Copies 2000 0010 Print Priority 2000 0020 HIGH MED LOW Set in Configuration file Default value is HIGH Medium Type 2000 0030 CLEAR FILM Set in Configuration File PAPER PROCESSOR Note denotes that the attribute is optional for the SCU However we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value If Failure status is returned during N CREATE operation of Film session the following action will be taken by Print SCU 0x213 Resource Limitation message will be sent All other status Failure message will be sent In all the cases the print session will be terminated 4 3 1 2 2 2 Basic Film box SOP Class The Print SCU supports the following DIMSE Service Elements for the Basic Film Box SOP Class e N CREATE Requests the Print SCP to create an instance of Film Box N ACTION Requests the Print SCP to print the Film Box onto Printer e N DELETE Requests the Print SCP to delete the Film Box Instance The Following Attribute values are supported DICOM Tag Valid Range Default Value 63 of 169 GE HEALTHCARE DIRECTION 5176235 100 R
25. The following table 3 3 3 1 3 1 defines the specific values used for this profile 50 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Table 3 3 3 1 3 1 Boot Sector for STD CTMR MOD12 conformant media Bytes s Big As Stored Description Endian 00 02 OxEB3C90 Jump instruction to loader NOPs 03 10 MSDOSA 0 The formatting DOS vendor specific s 2 File Allocation Tables FAT 19 20 0x0000 0x0000 Flag for more than 65536 sector disk Use offset 32 value OxF8 OxF8 Flag for disk type FOH if not otherwise specified 26 27 1 side head per disk 32 35 1163306 0x2AC01100 1163306 Total sector disk Varies from disk to disk ________ Dont care Any contents acceptabie 3 4 3 1 4 Directory Information in DICOMDIR Conformant Application Entities shall include in the DICOMDIR File a Basic Directory IOD containing Directory Records at the Patient and subsidiary levels appropriate to the SOP Classes in the File set All DICOM files in the File set incorporating SOP Instances defined for the specific Application Profile shall be referenced by the Directory Records All DICOM SOP Instance files in the File set are stored in the MSDOS filesystem subdirectory GEMS and are referenced from the DICOMDIR directory records Refer to Appendix C for a complete listing of all the optional modules and optional attributes used in the
26. as entered in user interface Contrast Bolus Route 0018 1040 3 Sent if contrast exam as entered in user interface A 3 9 SOP Common Module Attribute Name Noes Specific Character Set 0008 0005 ISO IR 100 A 3 10 VOI LUT Module Attribute Name Tag Type Notes 0 0 Window Center 0028 1050 Window Center for display Always sent 3 Window Width 0028 1051 1C Window Width for display Always sent A 3 11 Frame of Reference Module Tag Tye Nos CS 129 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Frame of Reference UID 0020 0052 a Position Reference Indicator 0020 1040 2 Value as entered at the user interface It is possible for the operator of CT LightSpeed QX i system to change the table height while scanning a series of images CAUTION Therefore implementations must use the Frame of Reference UID 0020 0052 in conjunction with the Table Height 0018 1130 to determine if two images are spatially related A 3 12 SC Equipment Module Attribute Name Type Modality 0008 0060 Modality of original image CT 130 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT PET Specific IOD Common Module Definitions A 4 1 Image IOD Modules Entity Name Module Name R f rence
27. coeficients 09100 1 PET activity_factor_3d 0019 100B 165 of 169 x 2 we FL FL LO DT SH FL 1 e wm GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 166 of 169 End of Document 167 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 168 of 169 GE Healthcare 3000 Grandview Blvd Waukesha WI 53188 U S A www gehealthcare com Imagination at work
28. on patient data Patient s Size 0010 1020 Patient s Address 0010 1040 Patient s Telephone 0010 2154 Numbers 6 8 13 4 Patient Medical Attribute Name g Expected Expected Mapped Matching Returned into the Table 6 8 13 4 1 Patient Medical Module Attributes Attribute Name Ta Expected Expected Mapped Matching Returned into the Key Type Key Type g Patient State 0038 0500 Pregnancy Status 0010 21 0 Displayed More Info screen Medical Alerts 0010 2000 No Displayed on More Info screen Contrast Allergies 0010 2110 No Displayed on More Info screen Special Needs 0038 0050 No Displayed on More Info screen Additional Patient History 0010 2180 Displayed on More Info screen 6 9 Private Data Dictionary The ModalityWorklist implementation does not define any Private Attributes within the Modality Worklist Information Model 91 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 10 C FIND Request Message This section provides a detailed description of the C FIND request message data that is provided to the remote AE during a worklist query operation The dump in Table 6 10 1 below lists in exact message order the fields transferred as part of the C FIND request message for a typical query In this particular dump no values are specified for the Schedul
29. 1001 Copied from SPS if selected Step Id Step description Scheduled Action Item 0040 0008 Copied from SPS if selected Code Sequence poner Created at the scanner The string PPS ID examnumber is used start date start time Performed Procedure Step 0040 0254 description 110 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 8 Grayscale Presentation State 8 1 8 2 Introduction This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance requirements for the Grayscale Presentation State GSPS referred to as the Exam Split option feature on this GEMS product Note that the format of this section strictly follows the format defined in DICOM Standard PS 3 2 Conformance Please refer to that part of the standard while reading this section The Exam Split option for Discovery allows a GSPS object to be sent to any remote host that supports the display of images using GSPS The Exam Split feature provides the DICOM Modality GSPS service as a service class user SCU This feature works in conjunction with DICOM Modality Work list feature if installed However the conformance of this feature is dependent on Modality work list feature For information on conformance of Modality Work List feature to DICOM standard please refer to the appropriate section or document Implementation Model Exam Split option have the two options i e Virtual
30. 2 1 2 840 113619 5 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1 2 1 2 840 113619 5 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 GE Private PET raw 1 2 840 113619 4 30 Implicit VR Little Endian 1 2 840 10008 1 2 data Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 Note The SCP does not have a default acceptance policy if more than one acceptable transfer syntax is proposed by the SCU It is the responsibility of the SCU to make a selection from more than one transfer sentences accepted 2 31 31 21 SOP Specific Conformance to C_STORE SCP The DICOM Server AE conforms to the SOP s of the Storage Service Class at level 1 base Private elements will be discarded from the image when receiving images containing non GE private data elements All of the standard type elements 1 1 2 2 3 will be retained 33 66 Each C STORE SCP operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 900 seconds 900 seconds and 60 minutes respectively Image Reception 34 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT If the DICOM Server AE returns one of the following status codes then the C STORE operation was unsuccessful and no image will be installed e 0110 Processing Failure Indicates that an internal system
31. 33 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 3 1 3 1 Receive Image s This AE is indefinitely listening for associations No operator action is required to receive an image 2 3 1 3 1 1 Associated Real World Activity The Real World Activity associated with the Receive Image s operation is the storage of the image on the disk drive of the Discovery 2 3 1 3 1 2 Presentation Context Table Table 2 2 1 3 1 2 1 Accepted Presentation Contexts for DICOM Server and Real World Activity Receive Image s Presentation Context Table Accepted Abstract Syntax Transfer Syntax List UID List 1 2 840 10008 1 1 Implicit VR Little Endian 1 2 840 10008 1 2 CT Image Storage 1 2 840 10008 5 1 4 1 1 2 Implicit VR Little Endian 1 2 840 10008 1 2 MR Image Storage Secondary Capture Image Storage PET Image Storage 1 2 840 10008 5 1 4 1 1 128 1 2 840 10008 5 1 4 1 1 4 1 2 840 10008 5 1 4 1 1 7 Implicit VR Big Endian GE Private Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Implicit VR Big Endian GE Private Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Implicit VR Big Endian GE Private Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Explicit VR Little Endian Explicit VR Big Endian 1 2 840 113619 5 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1
32. 5 Other P rameters iiec Eee ede la aep eese do err ene 54 3 4 3 2 5 1 Image Attribute 4 4 54 9 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 4 3 2 5 2 Attribute Value Precedence cccccceeeeceeceeeeeeeeeeceaeeeeeeeeeeeeaneeeeeeeseecueeeeeeeeseeeeaaees 54 3 4 4 Private Application 54 3 4 4 1 12 Class Profile essent 54 3 4 4 1 1 SOP Classes and Transfer Syntax ccceeccceeseeeeseeeeeeeeeseeeeeeeeeeeeesaeeeeeesaaeeeseeesaeeenaeenaas 54 3 4 4 1 2 Physical Media and Media 55 3 4 4 1 3 Logical Eortmat oem eere Ye bre e Ir eir eH EO 55 3 4 4 1 4 Directory Information in DICOMDIR sees 55 93 4 4 1 44 p eet teta etie es 55 3 4 4 1 5 Oilier Parameters iin aque Ea 56 3 4 4 1 5 1 Image Attribute 56 3 4 4 1 5 2 Attribute Value Precedence 56 3 5 EXTENSIONS SPECIALIZATIONS AND PRIVATIZATION OF SOP CLASSES AND TRANSFER SYNTAX 56 3 5 1 Extensions Specialization s and Privatization s Of SOP 56 3 5 1 1 SOP Specific Conformance Statement for CT SOP
33. 84 6 8 4 3 Imaging Service Request Entity 84 6 8 4 4 Visit Entity Description seinio iere Lo ce d steer tit cet e ea cs 84 6 8 4 5 Patient Entity Description siiis n 84 6 8 5 ModalityWorklist Mapping of DICOM Entities 84 6 8 6 Information Model Module Table sessi esent 85 6 8 7 Information Model Keys a a Ld utm ne a E t 85 6 8 8 o pported t ore de gd eet odes 86 6 8 9 Scheduled Procedure Step Entity seen 86 6 8 9 1 SOP Common Mod le azores pessit 0 0 nenne entere 86 6 8 9 2 Scheduled Procedure Step Module sssssssssssseeeeeeeeeen nennen enne 87 6 8 10 Requested Procedure Entily isses 88 6 8 10 1 Requested Procedure Module sse 88 6 8 11 Imaging Service Request Entity esses eene eene 89 6 8 11 1 Imaging Service Request Module sssseeeeeeeenennenneee nennen nnne nnn 89 6 8 12 VIS sis 89 0 9 12 1 Visit IdentifiCatlOTi luce eere eate ett tete e re em tne re d dte gen 89 6 8 12 2 iu eet el In Dedi a utei quist s Americae 89 6 8 12 3 Visit Relationship 4 eie etie eite 90 0 8 12 4 Visit AdmissiOni oerte Pee do
34. 87 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 10 Requested Procedure Entity 6 8 10 1 Requested Procedure Module Table 6 8 10 1 1 Requested Procedure Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type Requested Procedure ID 0040 1001 1 Yes User can enter the value for Requested Procedure Id prior to query If user has entered the value then that value will be sent as part of the query The value returned in the response shall be mapped to Requested Procedure 0032 1060 1C Yes Truncated to 22 characters Description Requested Procedure Code 0032 1064 1C No Sequence E O Coding Scheme Designator 0008 0102 Code Meaning 0008 0104 Study Instance UID 0020 000D User Modality will be able to configure the following If the user has set the option to Use MWL Study UID then study Instance UID will be copied into the final DICOM image header If this option is not set then a new Study instance UID is generated locally on the scanner Referenced Study Sequence 0008 1110 Only 1 Referenced Study Sequence is mapped into the Referenced SOP Class 0008 1150 UID Referenced SOP Instance 0008 1155 UID Arrangements 88 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 11 Imaging Service Request Entity 6 8 11 1 Imaging Service
35. Endian 1 2 840 10008 1 2 2 GE Private PET 1 2 840 113619 4 30 Explicit VR Little Endian 1 2 840 10008 1 2 None Raw data SOP 1 2 840 10008 1 2 1 class 1 2 840 10008 1 2 2 Note Other Abstract Syntax and Transfer Syntax names may be offered when establishing an association These are private syntax that may be ignored See section 2 4 2 3 1 2 1 2 1 SOP Specific Conformance Statement C STORE SCU This implementation can perform multiple C STORE operations over a single association Upon receiving a C STORE confirmation containing a successful status this implementation will perform the next C STORE operation The association will be maintained if possible Upon receiving a C STORE confirmation containing a Refused status this implementation will terminate the association Upon receiving a C STORE confirmation containing any status that is not Success or Refused this implementation will consider the current request to be a failure but will continue to attempt to send the remaining images in the request on the same association Each C STORE operation supports an Association Timer This timer starts when the association request is sent or received and stops when the association is established The time out is 60 seconds Each C STORE operation also supports an Operation Inactivity Timer This time out starts once the first C STORE request has been issued on association or received and is reset each time a C STORE response ha
36. Pixel Module px oo HO nma ERREUR UO PO ER REIR A 4 18 PET Image Module idus eee eade digerere eae ee A 4 19 Overlay Plane Module 22 4 20 VOLLUT Module 4 21 Discovery PET Image Module 4 22 Discovery PET Frame Module 4 23 SOP COMMON 81 CT IMAGE IOD PRIVATE DATA ELEMENTS DEFINITION essen B 1 1 Private Creator Identification GEMS IDEN 01 2 22 2 00000000000000000 B 1 2 Private Creator Identification GEMS ACQU 01 22002222 2 00000000000000 0 B 1 3 Private Creator Identification GEMS RELA 201 sese B 1 4 Private Creator Identification GEMS STDY 01 sess B 1 5 Private Creator Identification GEMS IMAG 01 sese B 1 6 Private Creator Identification GEMS PARM 201 sse B 1 7 Private Creator Identification GEMS HELIOS 01 sss B 1 8 Private Creator Identification GEMS CT CARDIAC 001 B 2 PET IMAGE IOD PRIVATE DATA ELEMENTS DEFINITION 1 B 2 1 Private Creator Identification GEMS 01 essen B 2 2 Discovery PET Patient
37. Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET exam id 0009 1007 3 UI 1 PET compatible version 0009 1008 PET software version 0009 1009 B 2 4 Discovery PET Scan Module Private Creator Identification GEMS PETD 01 Attribute Name Tag VR VM PET scan id 0009 100A UI 1 3 PET compatible_version SH 1 software_version SH 1 scan_datetime 0009 100D 1 PET scan_ready 0009 100E 1 PET scan_description ST 1 PET hospital_name PET scanner_desc PET manufacturer PET for_identifier PET landmark_name PET landmark_abbrev PET patient_position PET scan_perspective PET scan_mode 0009 1019 PET start_condition start cond data PET sel stop cond PET sel stop cond data 151 of 169 G gt I 7 7 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 PET collect_deadtime PET collect_singles PET collect_countrate PET countrate_period PET delayed_events PET delayed_bias PET word_size PET axial_acceptance PET axial angle 3d PET theta compression PET axial compression PET gantry tilt angle PET collimation PET scan fov PET axial fov PET event separation PET mask width PET binning mode PET trig rej method PET number for reject PET lower reject limit PET upper reject limit PET trigge
38. Real World Activities iere eene re Lee de Sones aiid 46 3 3 1 2 1 Real World Activity Create MOD 46 3 3 1 2 1 1 Application Profiles for the RWA Create 2 0040 46 3 3 1 2 2 Real World Activity Display 46 3 3 1 2 2 1 Application Profiles for RWA MOD Directory 46 3 3 1 2 2 2 Options 1e enim tne e ento ode kr eats 46 3 3 1 2 3 Real World Activity Copy to Local 46 3 3 1 2 3 1 Application Profiles for the RWA Copy to Local 47 3 3 1 2 3 2 47 3 3 1 2 4 Real World Activity Update 47 3 3 1 2 4 1 Application Profiles for the RWA Update MOD 47 wp desi eles wht ea ead O eaten cece 47 34 STANDARD AUGMENTED AND PRIVATE APPLICATION 47 3 4 1 Class and Profile Identification esses eene 47 3 4 2 Clinical 48 3 4 2 1 Roles and Service Class Options 2 10004000 nennen enne nennen enne 48 3 4 2 1 1 File Set Grealot Serials 48 3 4 2 1 2 File Set Reader
39. Remote DICOM AE by choosing Get Examination Get Series or Get Images 4 Real World Activity Choose Ping DICOM host Option consists of an operator selecting Ping DICOM host from the Network pull down menu This will cause the DICOM Server AE to initiate a DICOM Verification Request to the remote AE to verify the remote system activeness 5 The Choose storage commitment Option Real World activity consists of an operator configuring the remote node as an archive node Select choose media option under Archive pull down menu Select Remote Node in that menu Select one or more study series or image in the local database manager and choose either Save Exam or Save Series from the Archive pull down menu This will cause the DICOM server AE to send the image s to the selected destination and wait for the storage commitment from the remote node The remote node should be a storage commitment provider There is no Real World activity required for the DICOM Server AE to respond to an incoming DICOM store query or retrieve The DICOM Server AE is always prepared to respond to a DICOM Store Query or Retrieve by any remote DICOM AE The DICOM Server AE will perform the Real World activity Image Installation after the remote AE sends an image to the Discovery product Once a Query request is received the DICOM Server AE will search the local database for all entries that match the keys requested by the Re
40. STATEMENT gt Code Value Not Used gt Coding Scheme Designator Not Used gt Code Meaning Not Used gt Anatomic Region Modifier 0008 2220 3 SQ 1 Not Used gt gt Code Value Not Used gt gt Coding Scheme Designator 0008 0102 1 5 1 Not Used Code Meaning 0008 0104 3 1 Not Used Primary Anatomic Structure 0008 2228 3 5 1 Not Used Sequence gt Code Value 0008 0100 1 5 1 Not Used gt Coding Scheme Designator Not Used gt Code Meaning Not Used gt Primary Anatomic Structure 0008 2230 3 ISQJ 1 Not Used Modifier Sequence gt gt Code Value Not Used gt gt Coding Scheme Designator Not Used gt gt Code Meaning 0008 0104 3 LO 1 Not Used A 4 19 Overlay Plane Module This section contains Attributes that describe characteristics of an Overlay Plane Overlay Planes are not currently used in GE Discovery ST PET AE Attribute Name Tag VR Notes None A 4 20 VOI LUT Module This section specifies the Attributes that describe the VOI LUT Attribute Name Tag Type Notes VOI Lut Sequence 0028 3010 3 SQ Not Used gt LUT Descriptor 0028 3002 1 US US 3 Not Used or SS US gt LUT Explanation 0028 3003 Not Used 145 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 0028 3006 tgp Not Used LUT Data Window Center
41. Server AE title is supplied for matching this is user selectable Scheduled Procedure 0040 0002 Matching is supported as one of the Step Start Date following this is user selectable all days today only today and a number of days before today today and a number of days after today today and a number of days before today and a number of days after today Number of days before after is specified by the user Returned values must be exactly 8 numeric characters in YYYYMMDD format Scheduled Procedure 0040 0003 This attribute is sent with zero length Step Start Time Returned values must be exactly 6 numeric characters in HHMMSS format Modality 0008 0060 Matching is supported as follows either no Modality is supplied universal matching or the scanner s Modality is supplied for matching this is user selectable Scheduled Performing 0040 0006 Physician s Name Scheduled Procedure 0040 0007 Step Description Scheduled Procedure 0040 0011 Step Location Scheduled Protocol Code 0040 0008 Up to 5 Scheduled Action Item Code Sequence Sequence Items are mapped into the image if they system is configured to support Code Value 0008 0100 Coding Scheme 0008 0102 Designator Code Meaning 0008 0104 Pre Medication 0040 0012 Scheduled Procedure 0040 0009 Step ID Requested Contrast 0032 1070 Agent
42. This Server has only one Application Entity the MOD Archive Application The MOD Archive Application can perform these functions Itcaninitialize create DOS filesystem a piece of media writing a new label and DICOM File set onto the media LABEL t can update a piece of media by adding new SOP instances to an already existing DICOM File set from local storage SAVE display a directory listing of the File set on a piece of media QUERY t copy SOP instance from the MOD onto local storage RESTORE Sequencing of Real World Activities The updating function can only be performed on a piece of media that has already had a DICOM File set created The display function can only be performed on a piece of media that has already has a DICOM File set created With no SOP instances having been added the directory will be displayed empty The copy function can only be performed with a piece of media that has been updated and has DICOM SOP instances in the File set There are no other sequencing requirements File Meta Information for Implementation Class and Version The File Meta Information for this implementation is File Meta Information Version 1 Discovery CT Implementation UID 1 2 840 113619 6 55 Discovery PET Implementation UID 1 2 840 113619 6 131 44 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONF
43. This product uses DICOM Print SCU to print the images to DICOM Compliant Printers Section 5 DICOM Storage Commitment SCU is the DICOM Conformance Statement related to DICOM Storage Commitment SCU Application This product uses DICOM Storage Commitment Push Model to store the images using remote DICOM entity which is Storage Commitment SCP Section 6 HIS RIS Modality Worklist The ModalityWorklist option for Discovery allows a user to query for and display DICOM modality worklist information ModalityWorklist is providing the DICOM C FIND service as a service class user SCU Section 7 Performed Procedure Step The PPS option for Discovery allows a Modality Performed Procedure Step to be communicated to the Hospital Radiology information system The PPS feature is providing the DICOM Modality Performed Procedure Step service as a service class user SCU Section 8 Gray Scale Presentation State The Virtual Exam Split Option for Discovery provides the DICOM Modality GSPS service as a service class user SCU Appendix A specifies the CT MR PET IOD information object Appendix B specifies the private data element definition for CT MR PET IOD Appendix C specifies the DICOMDIR directory information Appendix D specifies the GE Private PET Raw IOD 17 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 1 2 Overall Conformance Statement Documentation Structure The documentation structure
44. ____ UID List 27 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Presentation Context Table Proposal Abstract Syntax Transfer Syntax Extended Negotiation UID Lit 1 2 840 10008 5 1 4 1 1 2 Implicit VR Little Endian 1 2 840 10008 1 2 SCU None CT Image Storage Implicit VR Big Endian GE 1 2 840 113619 5 2 Private Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 MR Image Storage 1 2 840 10008 5 1 4 1 1 4 Implicit VR Little Endian 1 2 840 10008 1 2 SCU None Implicit VR Big Endian GE 1 2 840 113619 5 2 Private Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 1 2 840 10008 5 1 4 1 1 7 Implicit VR Little Endian 1 2 840 10008 1 2 None Secondary Capture Implicit VR Big Endian GE 1 2 840 113619 5 2 Image Storage Private Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 Study Root 1 2 840 10008 5 1 4 1 2 2 Implicit VR Little 1 2 840 10008 1 2 None Query Retrieve 1 Endian FIND Study Root 1 2 840 10008 5 1 4 1 2 2 Implicit VR Little 1 2 840 10008 1 2 SCU None Query Retrieve 2 Endian MOVE Verification SOP 1 2 840 10008 1 1 Implicit VR Little 1 2 840 10008 1 2 None Class Endian PET Image 5 2 840 10008 5 1 4 1 1 12 Implicit VR Little Endian 1 2 840 10008 1 2 None Storage Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big
45. also apply to the following CT products based on this software LightSpeed VCT LightSpeed Pro LightSpeed VCT Select BrightSpeed Excel BrightSpeed Edge BrightSpeed Elite BrightSpeed Excel Select BrightSpeed Edge Select BrightSpeed Elite Select LightSpeed Xtra LightSpeed RT 16 LightSpeed RT LightSpeed QX I LightSpeed Plus LightSpeed Ultra LightSpeed 16 HiSpeed QX I LightSpeed Pro 16 Note The Discovery product lines present an Image Works desktop which contains advanced applications offered on the Advantage Workstation The DICOM Conformance Statements related to these applications can be found at the following website http www gehealthcare com usen interoperability dicom products workstation dicom html Note All references to DICOM refer to year 2007 1 1 Overview Section 1 Introduction provides general information about the content and scope of this document Section 2 Network Conformance Statement is the DICOM Conformance Statement related to this product Conformance Statements define the subset of options selected from those offered by the DICOM standard 16 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Section 3 Media Storage Conformance Statement is the DICOM Conformance Statement related to Media Storage Application Profile Section 4 Print SCU Conformance Statement is the DICOM Conformance Statement related to DICOM Print SCU Application
46. and attributes Table 6 8 6 1 Modality Worklist Information Model Modules SewdwedPcedweSep 6892 veSmus 88123 vstMmsso 88194 2 8193 oeoa Information Model Keys Please refer to DICOM Standard PS 3 3 Information Object Definitions and PS 3 4 Service Class Specifications for a description of each of the Entities contained within the Modality Worklist Information Model The following Module descriptions contain the attributes that are present in a C FIND request message sent by the Worklist Server AE to a remote AE It should be noted that they are the same as those defined in the DICOM Standard PS 3 4 Service Class Specifications and include e Name e Tag group and element numbers e Expected Matching Key Type R required O optional e Expected Return Key Type 1 non zero value required 1 conditionally of type 1 2 required to be present possibly with zero length value 85 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 optional e Mapped into The Image whether this data is mapped into subsequently acquired images e Notes clarification of this implementation s use treatment of this attribute All data elements in the following Module descriptions are requested by the Worklist Server AE Values of data elements that are not m
47. describe and identify the Study performed upon the Patient Attribute Name Tag Type Notes Study Instance UID 0020 000D 1 Ul 1 Study Date 0008 0020 2 DA 1 Study Time 0008 0030 2 1 Referring Physician s 0008 0090 2 PN 1 162 of 169 E EXCITE Accession Number COS D 1 3 Discovery PET RAW Data Scan Entity Modules This section specifies the attributes which identify and describe general information about the Scan within a Study Attribute Name Tag Type VR VM Notes Modality 0008 0060 Series Instance UID 0020 000E Series Number 0018 5100 2C CS 1 Operators Name Patient Position D 1 4 Discovery PET RAW Data Equipment Entity Modules This section specifies the attributes which identify and describe the piece of equipment which produced the instance of the Private SOP class Attribute Name Tag Type VR VM Notes Manufacturer 0008 0070 Institution Name Manufacturer s Model Name 0008 1090 Software Versions 0018 1020 D 1 5 Private Creator Identification Information Refer to section B 2 1 of this document as GEMS PETD 01 is used for this private IOD D 1 6 Discovery PET Patient Module Refer to section B 2 2 of this document D 1 7 Discovery PET Exam Module Refer to section B 2 3 of this document 163 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1
48. display browser will not list up the series for contents rendering the retrieval to local storage impossible 3 4 8 2 4 1 Additional Keys None 3 4 3 2 5 Other Parameters This section defines other parameters common to all specific Application Profiles in the PRI CTMR MOD12 class that need to be specified in order to ensure interoperable information interchange 3 4 3 2 5 1 Image Attribute Values The attributes listed in Table 3 3 3 1 5 1 1 used within the CT MR PT SC Standalone Overlay Image files shall take the values specified Overlay data if present shall be encoded in Overlay Data 60XX 3000 3 4 3 2 5 2 Attribute Value Precedence None 3 4 4 Private Application Profiles The MOD Archive Server supports a private Application Profile PRI CTMR MOD12 as FSR or FSU 3 4 4 1 PRI CTMR MOD12 Class Profile 3 4 4 1 1 SOP Classes and Transfer Syntax This class of Application Profiles is based on the Media Storage Service Class See PS3 4 SOP Classes and corresponding Transfer Syntax supported by the PRI CTMR MOD12 Application Profiles is specified in the Table 3 3 3 1 1 1 54 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Note The MOD Archive Application chooses the JPEG Lossless compression selection value 1 as the default transfer syntax for storing images on the media But the transfer syntax is selectable on a per media basis at install time using a method defined by
49. length Code Sequence sequence gt Code Value 0008 0100 1 SH 1 gt Coding Scheme Designator 0008 0102 1 SH 1 gt Code Meaning 0008 0104 3 LO 1 A 4 11 Discovery PET ImageSet Module Refer to Section B 2 6 for details A 4 12 Discovery PET Scan Module Refer to Section B 2 4 for details 138 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT A 4 13 Frame of Reference Module This section specifies the Attributes necessary to uniquely identify a frame of reference which insures the spatial relationship of Images within a Series It also allows Images across multiple Series to share the same Frame Of Reference This Frame Of Reference or coordinate system shall be constant for all Images related to a specific Frame Of Reference Attribute Name Tag Type VR VM Notes Frame of Reference UID 0020 0052 1 Ul 1 IF identifier l NULL THEN ImageSet for identifier ELSE idbMakeld Position Reference Indicator 0020 1040 2 LO 1 Scan landmark_name A 4 14 General Equipment Module This section specifies the Attributes which identify and describe the piece of equipment which produced a Series of Images Attribute Name Tag Type VR VM Notes Manufacturer 0008 0070 2 LO 1 Institution Name 0008 0080 ERR hospital name Institution Address 0008 0081 EXE AES Not Used Station Name 0008 10
50. manager Note Also refer to section 5 3 1 3 2 3 1 2 5 2 Proposed Presentation Context Table Refer to section 5 3 1 4 2 3 1 2 5 3 SOP Specific Conformance Statement Storage Commitment SCU Refer to section 5 3 1 4 2 3 1 3 Association Acceptance Policy The DICOM Server AE places limitations on who may connect to it If the Remote AE needs to Push Images or Query Retrieve Images to the local system then it has to be configured in the Local system to do the same To configure a remote AE choose Select Remote Host in the network pull down menu Select the Remote Host and choose Update option if the Remote AE is already present If the Remote AE is not present choose Add option to add the Remote AE Set Push Images option to Yes if the Remote AE needs to push images to the Local system Set Query Retrieve option to Yes if the Remote AE needs to Query Retrieve the local system When the DICOM Server AE accepts an association for image storage it will receive any images transmitted on that association and store the images on disk It will also respond to queries from Remote AEs by sending matching entries Any Remote AE can request and receive a list of images on the local database The Remote AE must be configured in the local database manager s list of Remote AE for it to be able to retrieve images from DICOM Server AE Any remote AE can open an association to the DICOM Server AE for the purpose of verification
51. panel Once a DICOM host is designated as a remote archive device the host will be available in the Archive device selection menu and can be chosen for any subsequent archive operation Configurable Parameters The following fields are configurable for the DICOM Server AE Local AE Title the machine hostname Local IP Address Local IP Netmask Max PDU length Time outs which are set for all hosts are configurable in dcs cfg Note denotes SOP class time sec Association time out bi_assoc_tio Inactivity time out bi_store_tio Session time out bi_session_tio Note All configurations should be performed by a GE Field Service Engineer 5 7 Support of Extended Character Sets This does not support any Extended Character set 6 Modality Work List Information Model Definition 6 1 Introduction This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance requirements for the relevant Networking features on this GEMS product Note that the format of this section strictly follows the format defined in DICOM Standard PS 3 2 Conformance Please refer to that part of the standard while reading this section 75 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 2 6 2 1 6 2 2 6 2 3 The ModalityWorklist option for Discovery allows a user to query for and display DICOM modality worklist information from a remot
52. respectively If a Cancel or Refused status is returned from the Remote AE the association is closed and the operation terminated The DICOM Server AE will parse each matching C FIND RSP reply and ignore the entries it fails to parse Tables 2 2 1 2 2 2 1 1 2 2 1 2 2 2 1 3 shows the various fields that are requested at the Study Series and Image levels of the C FIND request Query results are filtered based on the Modality field Only CT MR Screen Save images are supported Table 2 2 1 2 2 2 1 1 Requested Study Level Keys Study date 8 7 00080020 Zerolengh Stuyime 8 00080030 Zerolengh Palien sname 10010 0010 Zerolengh Sudyid Study Instance UID Study id R 7 0020010 Zero length ____ Study Instance UID U ______ 0020 000D_ Zero length ____ Study description O 0008 1030 _ Zero length Private Creator Identification _____ ______ 0009 00 GEMS IDEN 01 Suitela 0009 02 Table 2 2 1 2 2 2 1 2 Requested Series Level Keys Modality TR 0008000 Zero length Series number LR 0020001 Zerolength Series description _________ 6____ 0008 103E__ Zero length Manufacturer 0 00080070 Zero length _____ Images in series 10 0020100 Zerolength ____ Table 2 2 1 2 2 2 1 3 Requested Image Level Keys Imagenumbe R 0020 0013 _ Zero length Image type O 0008008 Zero length Rows 0028000 j
53. s O Manufacturer 0008 0070 2 Always sent as GE MEDICAL SYSTEMS me uonName 6008080 3 Semt Value is configurable omen 3 _ Manufacturers Model 0008 1090 Device Serial 0018 1000 Sent if present in image Number Software Versions 0018 1020 3 Always sent as 05 for this version does not distinguish individual software releases Spatial Resolution 0018 1050 Sent only for MR images A 3 6 General Image Module Attribute Name Notes 128 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Image Number 0020 0013 Generated sequentially always sent Patient Orientation 0020 0020 Always sent zero length Image Date 0008 0023 Generated for each image always sent C Image Type 0008 0008 Always sent Acquisition Number 0020 0012 3 Generated for each acquisition always sent Acquisition Date 0008 0022 3 Generated for each acquisition always sent Acquisition Time 0008 0032 Generated for each acquisition always sent A 3 7 Image Plane Module Attribute Name Notes 0 0 Slice Thickness 0018 0050 Value always sent Image Time 0008 0033 Generated for each image always sent Image Slice Location 0020 1041 Value always sent A 3 8 Contrast Bolus Module Contrast Bolus Agent 0018 0010 2 Sent if contrast exam
54. set for all hosts are configurable in dcs cfg Association time out bi assoc tio Session timeout bi session tio Create timeout bi ncreate tio Note All configurations should be performed by a GE Field Service Engineer Note The local port on which the Discovery system receives DICOM incoming TCP connections is port 4006 41 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 7 Support of Extended Character Sets In addition to the DICOM default character set Discovery supports the ISO IR 100 Latin alphabet 1 supplementary set for the purpose of interchange 42 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 MEDIA STORAGE CONFORMANCE STATEMENT 3 1 3 2 3 2 1 Introduction This conformance statement specifies the Discovery conformance to the DICOM Media Interchange It details the DICOM Media Storage Application Profiles and roles that are supported by this product in its version 1 0 Note the format of this section follows the format of the DICOM Standard Part 2 conformance Annex A hence the paragraph numbering scheme Please refer to that part of the standard while reading this section Implementation Model MOD Archive Server All DICOM functionality on the Discovery product is handled by the DICOM Server Application Entity AE The DICOM Server AE is commanded to perform DICOM services through the buttons and men
55. system updates the state and indicates the same on the user interface If the operation is Failed detailed message is logged into system log file and system provides an alternative mechanism to retry the failed operation through the user interface to ensure that transient failures do not affect the feature performance 116 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 8 8 Use of Specific DICOM Data The following table gives specific usage of some of attributes in the GSPS and splitting PPS SOP instance created with reference to each of the real world scenarios mentioned in 8 2 3 of this document Attribute Name Tag GSPS Splitting PPS Comment Group Length 0002 0000 Exam Split Application will calculate File Meta Information 0002 0001 X Exam Split Version Application will calculate Media Storage SOP Class 0002 0002 X Exam Split UID Application will calculate Media Storage SOP 0002 0003 X Exam Split Instance UID Application will calculate Transfer Syntax UID 0002 0010 X X Exam Split Application will calculate Implementation Class UID 0002 0012 X Exam Split Application will calculate Implementation Version 0002 0013 X Exam Split Name Application will calculate Patient s Name 0010 0010 X X Copied from acquired image s Patient ID 0010 0020 Copied from acquired image s Pati
56. the installed base system will be required to learn how to be a FSU or FSR of the STD CTMR MOD 12 23 profile directory sequences 3 3 1 1 File Meta Information for the Application Entity Following are the values set in the File Meta Information for this AE Source Application Entity Title Not used by the MOD Archive Application e Private Information Creator UID Not used by the MOD Archive Application Private Information Not used by the MOD Archive Application 45 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 3 1 2 Real World Activities 3 3 1 2 1 Real World Activity Create MOD Request The MOD Archive Application acts as an FSC when requested to initialize LABEL the media The AE will use the appropriate Application Profile depending on the physical media type that is in the optical disk drive be it 2 3GB 1 2GB media The MOD Archive Application will label the media and take the user provided list of SOP instances eliminating any SOP instances on that list that does not correspond to one of the Application Profiles in Table 3 1 1 These SOP instances are written to the media and a corresponding DICOMDIR sequence record is created The determination of the potentially applicable Application Profile is dependent on the type of media and the associated software on which the AE has been invoked 3 3 1 2 1 1 Application Profiles fo
57. to the relationship boxes Illustration 6 8 3 1 Modality Worklist Information Model E R DIAGRAM Scheduled W orklist Procedure Step Item v 1 Requested Procedure requested for 1 Im aging Service Request done for 1 Patient 83 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 4 Entity Descriptions Please refer to DICOM Standard PS 3 3 Information Object Definitions and PS 3 4 Service Class Specifications for a description of each of the Entities contained within the Modality Worklist Information Model 6 8 4 1 Scheduled Procedure Step A Scheduled Procedure Step is an arbitrarily defined scheduled unit of service that is specified by the Procedure Plan for a Requested Procedure It specifies one or more Action Items events involving equipment i e imaging modality equipment human resources location and time i e start time stop time duration 6 8 4 2 Requested Procedure Entity Description A Requested Procedure is an instance of a Procedure of a given Procedure Type An instance of a Requested Procedure includes all of the items of information that are specified by an instance of a Procedure Plan that is selected for the Requested Procedure by the imaging service provider 6 8 4 3 Imaging Service Request Entity Description An Imaging Service Request is a set of one or more Requested Procedures selected from a list of Pro
58. user The old transaction uid is not applicable for these requests 72 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT The Following DIMSE service Elements are supported for the Storage Commitment request and response processing N ACTION Requests the remote Storage Commitment SCP to commit to storing the image instances N EVENT REPORT The response sent by the remote Storage Commitment SCP The following attributes are sent as part of the DATA Set for the N Action request Attribute 7 Transaction UID 0008 1195 Transaction UID Referenced SOP Sequence 0008 1199 reo SOP Class UID 00084150 ________ SOP Instance UID 00081155 Referenced Study Component sequence attribute is not sent Storage Media File Set ID and Storage Media File Set UID attributes are not supported 5 3 1 6 Response Processing Once the N EVENT REPORT response is received the following actions will be taken depending on the status of response 5 3 1 6 1 Commit response with SUCCESS status The Archive flag information in the browser for all the successful instances will be updated The status will be changed to Y The job queue entry will be removed Note The following attributes are expected as part of DATA Set for N Event Request from SCP Transaction UID 0008 1195 Value received from SCP Referenced SOP Sequence 0008 1199 Value received fr
59. 006 0045 1004 0045 1008 0045 100C 0045 100E 0045 1010 0045 1011 0045 1012 5 5 149 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Attribute Name EkgFullMaStartPhase 0045 1036 CS 1 EkgFullMaEndPhase 0045 1037 CS 1 EkgModulationMaxMa 0045 1038 CS 1 0045 1039 CS 1 0045 1038 LO 1 0045 1044 55 1 B 1 8 Private Creator Identification GEMS CT CARDIAC 001 Note Private Group 49 is present only if the appropriate cardiac option is installed on the scanner 0049 1001 0049 1002 0049 1005 0049 1006 0049 1007 0049 1008 B 2 PET Image IOD Private Data Elements Definition B 2 1 Private Creator Identification GEMS PETD 01 Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM Private Creator Data Element 0009 0010 1 SH 1 GE Discovery PET Implementation 0009 1001 3 LO 2 Version Name B 2 2 Discovery PET Patient Module Private Creator Identification GEMS 01 Attribute Name Tag Type VR VM PET Implementation Name 0009 1001 3 LO 2 PET patient_id 0009 1002 8 1 3 PET compatible_version 0009 1003 3 SH 1 PET patient_datetime 0009 1005 150 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 009 1006 B 2 3 Discovery PET Exam Module
60. 030 Value from user interface sent Name of Physician s 0008 1060 3 Sent if entered at the user interface Reading Study A 3 3 Patient Study Module Attribute Name ep _ Patient s Age 0010 1010 NM from Date of Birth entered at user interface Patient s Weight 0010 1030 au Value from user interface Additional Patient s 2 3 Value from user interface History 127 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT A 3 4 General Series Module Attribute Name Type Notes 0020 0011 Generated sequentially always sent Laterality 0020 0060 Always sent zero length Series Date 0008 0021 3 Generated for each series and always sent Series Time 0008 0031 3 Generated for each series and always sent Performing 0008 1050 3 Physician s Name Protocol Name 0008 1030 Sent if entered at user interface Series Description _ Series Description _ 0008 103E Value from user interface is sent Operators Name 0008 1070 Value from user interface is sent Patient Position 0018 5100 Sent As selected by operator when patient is positioned Defined terms are HFP Head First Prone HFS Head First Supine HFDR Head First Decubitus Right HFDL Head First Decubitus Left FFDR Feet First Decubitus Right FFDL Feet First Decubitus Left FFP Feet First Prone FFS Feet First Supine A 3 5 General Equipment Module wm Tw
61. 10 ESE JE Not Used Institutional Department Name 0008 1040 Not Used Manufacturer s Model Name 0008 1090 EARTE scanner_desc Device Serial Number 0018 1000 3 LO 1 Software Versions 0018 1020 3 LO 1 Image software_ve rsion Frame software_v ersion Spatial Resolution 0018 1050 3 DS 1 Not Used Date of Last Calibration 0018 1200 3 DA 1 Not Used Time of Last Calibration 0018 1201 3 TM 1 Not Used Pixel Padding Value 0028 0120 3 1 Not Used 139 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT A 4 15 General Image Module This section specifies the Attributes which identify and describe an image within a particular series Attribute Name Tag VR VM Notes Image Number 0020 0013 2 IS 1 Scan scan_mode image_location image_time image_bin_time Not Used Extract date from image_datetime Patient Orientation 0020 0020 Image Date 0008 0023 a Image Time 0008 0033 Image 0008 0008 LI Acquisition Number 0020 0012 ERE Extract time from image datetime 1 n Superceded by PET Image Image Type 0x0008 0x0008 Not Used Acquisition Date 0008 0022 DA Superceded by PET Image Acquisition Date 0x0008 0x0022 Acquisition Time 0008 0032 TM 1 Superceded by PET Image Acquisition Time 0x0008 0x0032 Referenced Image Sequence 0008 1140 SQ 1 Not Used Referenced SOP Class
62. 18 1 PatientName 94 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 11 Use of Specific DICOM Data This section details the use of the DICOM data returned by remote AEs during worklist queries The Discovery user interface fields that display the data along with the data s mapping into resulting acquired and transferred DICOM images are presented in following table 6 11 1 Table 6 11 1 Specific Data Usage DICOM Worklist Data Element Patient Schedule Screen Field Discovery DICOM Image Data Element Accession Number 0008 0050 Req Number Accession Number Supports maximum of 16 characters Supports maximum of 16 characters Supports maximum of 32 characters to calculate age Patients Sex 0010 0040 Sex 0008 0090 0032 1060 Scheduled Procedure Step Start Not available Date 0040 0002 Scheduled Procedure Step Start Time Not available Time 0040 0003 Study Instance UID 0020 000d Study instance UID only displayed Study Instance UID if User has set the on the More Info screen preference to Use MWL Study UID Requested Procedure Id Requested Proc ID Requested Procedure Id 0040 1001 Pregnancy Status 0010 21 0 Pregnancy Status only displayed on Not available the More Info screen Medical Alerts 0010 2000 Medical Alerts only displayed on Not available the More Info screen Contrast Allergies 0010 2110 Contrast Allergies on
63. 2 D 1 2 Discovery PET RAW Data Exam Module sse 162 D 1 3 Discovery PET RAW Data Scan Entity Modules sse 163 D 1 4 Discovery PET RAW Data Equipment Entity 163 D 1 5 Private Creator Identification Information sese 163 D 1 6 Discovery PET Patient Module sisse sienne tnnt 163 D 1 7 Discovery PET Exam Module sss einen 163 D 1 8 Discovery PET Scan Module enne nennen 164 D 1 9 Discovery PET Frame Module sss esistente nena 164 D 1 10 esa du seta tees 164 D 1 6 1 SOP Common Module Attributes sessi sienne 164 D 1 11 Discovery PET Raw Data Module sss sisse ennt 164 D 1 12 Discovery PET Correction Cal Module sees enne 164 D 1 13 Discovery PET Well Counter Module sse eene 165 15 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 1 SECTION 1 INTRODUCTION Note Throughout this entire document the term Discovery refers to the following Discovery products Discovery VCT Discovery DST Discovery DSTE The IHE European Connectathon Beta PET CT Software applies to the PET CT products above and the CT aspects of this document
64. 2 STD CTMR MOD23 Class Profile 3 4 3 2 1 SOP Classes and Transfer Syntax s This class of Application Profiles is based on the Media Storage Service Class See PS3 4 SOP Classes and corresponding Transfer Syntax s supported by the STD CTMR MOD23 Application Profiles is also specified in the Table 3 3 3 1 1 1 52 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 4 3 2 2 Physical Media and Media Formats The CT MR PT Application Profiles in the STD CTMR MOD23 require the 130mm 2 3GB 512 Byte per Sector Magneto Optical Rewriteable physical media with the PC File System as defined PS3 12 3 4 3 2 3 Logical Format The STD CTMR MOD23 Application profile media format conforms to the Logical Format specification in PS3 12 The following table 3 3 3 2 3 1 defines the specific values used for this profile Table 3 3 3 2 3 1 Boot Sector for STD CTMR MOD23 conformant media Bytess Big Endian As Stored 00 02 mm OxEB3C90 Jump instruction to loader NOPs 03 10 0x4d53444F53342E30 MSDOS4 0 The formatting DOS vendor specific 2 File Allocation Tables FAT 19 20 0x0000 0 0000 Flag for more than 65536 sector disk Use offset 32 value OxF8 OxF8 Flag for disk type FOH if not otherwise specified 26 27 1 side head per disk 32 35 2244896 0x20412200 2244896 sector disk Varies from disk to disk 36 37 0x0080 0 8000 Physical Drive number 80 8
65. 51 Kbytes SOP class Extended Negotiation is not supported Maximum PDU Length of Zero 0 is not supported The association request will be rejected when a request with Maximum PDU length of zero is received The maximum number of Presentation Context Items that is supported is 60 Note that the same Abstract Syntax may be offered multiple times with different Transfer Syntax The user information items sent by this product are e Maximum PDU Length and e Implementation UID 2 3 1 1 2 Number of Associations DICOM Server AE SCU will initiate only DICOM association at a time to perform an image store to a remote host or retrieve image s from a Remote AE 26 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT The DICOM Server AE SCP can have a maximum of four DICOM associations open simultaneously to receive and store image store or respond to an echo 2 3 1 1 3 Asynchronous Nature Asynchronous mode is not supported All operations will be performed synchronously 2 3 1 1 4 Implementation Identifying Information The Implementation UID allows unique identification of a set of products that share the same implementation The Implementation UID for this ID Net v3 0 Implementation is Discovery CT Implementation UID 1 2 840 113619 6 55 Discovery PET Implementation UID 1 2 840 113619 6 131 2 3 1 2 Association Initiation by Real World Activity This AE attempts to ini
66. 6 6 2 1 Application Data Flow Diagram sessi seen nnne 76 6 2 2 ei et D e eda 76 6 2 3 Sequencing of Real World Activities sisse 76 6 37 SPECIFICATIONS uel A date eee te eiat Ue 77 6 3 1 Worklist Server AE Specification esses 77 6 3 1 1 Association Establishment 77 6 3 1 1 1 ccc 77 6 3 1 1 2 Number of ASSOCIAlO S 35 0 ic aidan e cte re tct dece tede cep cecus rot 77 6 3 1 1 3 Asynchronous Nat re o E a epa e iHe 77 6 3 1 1 4 Implementation Identifying 78 6 3 1 2 Association Initiation 78 6 3 1 2 1 Real World Activity Worklist 78 6 3 1 2 1 1 Associated Real World n 78 6 3 1 2 1 2 Proposed Presentation Context 2 20 44 00 0 79 6 3 1 3 Association Acceptance 80 6 4 COMMUNIGATION PROFILES Ax cie oet idee e ti Pei tad epe i deter recede 81 6 4 1 Supported Communication Stacks PS 3 8 PS 3 9 81 6 4 2 htec d i e eiui e dace bd ee i ee d eei Ye Nan 81 6 4 3 TCP IP SIack edere e ee dare e eee dA ce Reve Bel atte edd 81 6 43 1 ABE IER quee eet
67. 8 5 1 4 1 1 4 Secondary Capture image storage 1 2 840 10008 5 1 4 1 1 7 Standalone Overlay storage 1 2 840 10008 5 1 4 1 1 8 PET Image Information Storage 1 2 840 10008 5 1 4 1 1 128 In addition to the C MOVE response status values defined in DICOM part 4 the following status values will be returned e C000 Indicates that an error occurred while retrieving records from the local database e C001 Indicates all other processing error e C011 If the Destination AE returns a Storage Full condition this status will be returned This status will only be sent if the Destination AE returns a status of A711 and is only applicable if the Destination AE is another GE Medical Systems product 2 3 1 3 4 3 Presentation Context Acceptance Criteria No criterion 39 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 3 1 3 4 4 Transfer Syntax Selection Policy All protocols detailed above in the tables are supported no priority selection is given 2 4 Communication Profiles 2 41 Supported Communication Stacks parts 8 9 DICOM Upper Layer Part 8 is supported using TCP IP 2 4 2 TCP IP Stack The TCP IP stack is inherited from a UNIX Operating System 2 4 2 1 Physical Media Support Ethernet v2 0 IEEE 802 3 Auto senses 10 100 Base T Ethernet connection 2 4 3 Point to Point Stack 50 ACR NEMA connection is not applicable to this product 2 5 Extensions Specializations Priva
68. ALTHCARE DIRECTION 5198351 100 Rev 1 DiscovERY DICOM CONFORMANCE STATEMENT 4 3 AE Specifications 4 3 1 DICOM Print SCU AE Specification Print SCU provides Standard Conformance to the following DICOM 3 0 SOP classes as an SCU SOP CLASS Name SOP CLASS UID Basic grayscale print management meta SOP class 1 2 840 10008 5 1 1 9 Print Job SOP class 1 2 840 10008 5 1 1 14 4 3 1 1 Association Establishment Policy 4 3 1 1 1 General The Film Composer provides options to indicate the printer AE title and Printer host name These parameters are used by the Print SCU to establish the association with the desired printer The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for association initiated by the DICOM Print SCU AE is Maximum Length PDU 4 3 1 1 2 Number of Associations The Print SCU will initiate only one association with printer Multiple associations are not supported 4 3 1 1 3 Asynchronous Nature The print SCU does not support asynchronous operations All operations will be performed synchronously 4 3 1 1 4 Implementation identifying information The Implementation UID allows unique identification of a set of products that share the same implementation The Implementation UID for this ID Net Implementation is Discovery CT I
69. ATE request to the remote AE MPPS SCP in order to create a MPPS SOP instance at the remote AE The MPPS SCP returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user The DICOM association is closed System includes the necessary information related to Scheduled Procedure Steps and the Performed procedure Step in the image instances created 98 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT At the end of image acquisition system initiates a Complete PPS or Discontinue PPS based on the choice selected by the user using the user interface provided The user is also given a choice Defer PPS which is described below PPS Server AE initiates a MPPS N SET request to the remote AE in order to update the MPPS SOP instance that is already created The N SET is sent over a new DICOM association The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user At the end of image acquisition if the user has chosen Defer PPS the user is provided with an interface to Complete PPS or Discontinue PPS at any later time The user might wish to alter the image set generated through acquisition before invoking these operati
70. COMMITMENT SCU CONFORMANCE 69 5 1 INTRODUCTION i5 351 d EE pex i DE ERR 69 5 2 IMPLEMENTATION 41 1 0 tuntut 69 5 2 1 Application data flow diagram 69 5 2 2 Functional definitions eese sese eee nnns en 70 5 2 8 Sequencing of real world activities sss 70 10 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 9 3 AE SPECIFICATIONS Set 70 5 3 1 DICOM Storage Commitment SCU specifications sess 70 5 3 1 1 Association establishment policies 70 5 3 1 1 1 CHI ICM 70 5 3 1 1 2 Number 71 5 3 1 1 3 Asynchronous Nature Ip nla go e ee 71 5 3 1 1 4 Implementation identifying m 71 5 3 1 2 Association initiation poliCy isise ccce ee eden De 71 5 321232 World ACIUVILy aii eiecit teer tee re hb ced ete o et 71 5 3 1 3 1 Associated Real World Activity Save 71 5 3 1 3 2 Associated Real World Activity Auto Archive 71 5 3 1 4 Propos
71. Communication Stacks PS 3 8 114 8 3 2 SIE e aa 115 8 3 3 TOPAP SIaCK aser dais e e t E e EL e 115 8 3 3 1 GAP 115 8 3 3 2 Physical Media SUpport risna duree decet eee tne ire oet neat dee nero ec e 115 8 3 4 Point to Point Stack 115 84 EXTENSIONS SPECIALIZATION PRIVATIZATION eese 115 8 4 1 Standard Extended Specialized Private 5 115 8 4 2 Private Transfer isses seinen nitent 115 8 5 GONEIGUBAT IOI eii dae ied A ra A A 115 8 5 1 AE Title Presentation address Mapping seen 116 8 5 2 Configurable Parameters isses eiie einen tnn insit ennt nnns nnns 116 86 SUPPORT OF EXTENDED CHARACTER 6 116 87 ERROR HANDLING AND RECOVERY sss eene 116 8 8 USEOFSPECIFIGIDICOM DATA etiara tanta da i ni lt e e ERA a 117 8 9 IMAGE HEADER CHANGES SUPPORTING 56 5 122 8 9 1 Request Attributes Sequence sss sienne enne innen 122 13 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEM
72. Coordinate 0270043 Returned Table Start Location 00270050 Returned Table End Location 00270051 Returned RAS Letter for Side of Image 00270052 Returned RAS LetterforAnteror Posterio 00270053 Returned RAS Letter for Scout Start Location ___ _____0027 0054 Returned RAS Letter for Scout End Location 1 00270055 Returned DimensionX 00270060 Returned Image Dimension Y 00270061 Returned Note In the above tables the type field has the following meaning R Required U Unique Optional Private Only keys with Usage type Matched will be matched against values in the database Values in keys of type Returned will be ignored and will be filled in with data from the database If an optional key is requested that does not appear in any of the tables above that key will be ignored and no corresponding element will be returned If the database does not have a value corresponding to any requested optional key a zero length element will be returned Sequence matching is not supported Range matching is supported for attributes of type date and time Only hierarchical query is supported 2 3 1 3 4 Retrieve Request From Remote AE This AE is indefinitely listening for associations No operator action is required to respond to a retrieve request 2 3 1 3 4 1 Associated Real World Activity The Real World Acti
73. DICOMDIR definition Please note that the mandatory attributes in each of the directory records as mentioned in Part 3 Addendum Basic Directory 51 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Information Object Definition are not listed in the appendix but are supported by the implementation Note modality attribute 0008 0060 in the DICOMDIR Series records should be CT MR for Secondary Capture and Standalone Overlay SOP Class Images Otherwise the Discovery DICOM media display browser will not list up the series for contents rendering the retrieval to local storage impossible 3 4 3 1 4 1 Additional Keys None 3 4 3 1 5 Other Parameters This section defines other parameters common to all specific Application Profiles in the PRI CTMR MOD12 class that need to be specified in order to ensure interoperable information interchange 3 4 3 1 6 Image Attribute Values The attributes listed in Table 3 3 3 1 5 1 1 used within the CT MR SC Standalone Overlay Image files shall take the values specified Table 3 3 3 1 5 1 1 CTMR MOD Profiles Required Image Attribute Values for CT PT and MR SC and OV Images Attribute Tag Modality 0008 0060 CT MR PT Photometric Interpretation 0028 0004 MONOCHROME2 Bits Allocated 0028 0100 Bits Stored 0028 0101 Overlay data if present shall be encoded Overlay Data 60XX 3000 3 4 3 1 6 1 Attribute Value Precedence None 3 4 3
74. Data Flow Diagram DICOM Standard Interface Choose Archive SAVE Option Storage Provider Set Archive State DICOM Standard Interface 69 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 5 2 2 Functional definitions 5 2 3 DICOM Server AE initiates the following operations Negotiates and establishes association with remote Storage Commitment Provider Sends the selected images to the remote DICOM AE Closes the association If all the images are transferred without failures the following steps will be executed If there are any failures the job will be marked as failed and the Storage Commitment request will not be sent Establishes a new association for sending the commitment request Here note that a storage commitment request does not imply that the acquisition of images is complete Closes the association for sending the commitment request Receives the response on a different association Updates the archive flag information for successful instances The archive flag is shared with the local archive and the two cannot be distinguished Sequencing of real world activities The user has to select the remote commitment provider and the exams series to be pushed before clicking Save Exam Series button on the Archive menu 5 3 AE Specifications 5 3 1 DICOM Storage Commitment SCU specifications Provides Standard Conformance to the following D
75. Date and Performed Procedure Step End Time respectively Not the actual end date and end time of post processing Procedure Code Sequence This sequence 15 sent with ZERO items in the MPPS message Performed Action Item Code Sequence This sequence is sent with ZERO items in the MPPS message 105 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT e Referenced Standalone SOP Instance Sequence The sequence is sent with ZERO items in the MPPS message 7 9 ERROR HANDLING AND RECOVERY PPS Server AE does not define any extended error codes The standard error codes are handled On a response with status success for the N CREATE or N SET request the system updates the state and indicates the same on the user interface On a response with status other than success the operation is deemed Failed and the system updates the state and indicates the same on the user interface If the request has failed or response is not received before the association timeout the operation is deemed Failed and the system updates the state and indicates the same on the user interface If the operation is Failed detailed message is logged into system log file and system provides an alternative mechanism to retry the failed operation through the user interface to ensure that transient failures do not affect the feature performance 106 of 169 GE HEALTHCARE DIRECTION 5198351 100 R
76. Discovery DICOM CONFORMANCE STATEMENT 8 2 4 GSPS Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU SOP Class Name SOP Class UID Grayscale Softcopy Presentation State 1 2 840 10008 5 1 4 1 1 11 1 8 2 4 1 Association Establishment Policies 8 2 4 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU from an association initiated by the PPS Server AE is Maximum Length PDU 128 Kbytes The SOP Class Extended negotiation is not supported The maximum number of presentation negotiation items that will be proposed is 1 The user information items sent by this AE are e Maximum PDU Length e Implementation UID 8 2 4 1 2 Number of Associations The GSPS application will initiate only one DICOM association at any time to perform a GSPS operation to the remote AE 8 2 4 1 3 Asynchronous Nature Asynchronous mode is not supported All operations are performed synchronously 8 2 4 1 4 Implementation Identifying information The implementation UID for this DICOM Implementation is PPS Feature for Discovery Implementation UID 1 2 840 113619 6 55 8 2 4 2 Association Initiation Policy The GSPS AE initiates a ne
77. Discovery DICOM CONFORMANCE STATEMENT D 1 8 Discovery PET Scan Module Refer to section B 2 4 of this document D 1 9 Discovery PET Frame Module Refer to section B 2 5 of this document D 1 10 General Modules This section defines the attributes which are required for proper functioning and identification of the associated SOP Instances They do not specify any semantics about the Real World Object represented by the IOD D 1 6 1 SOP Common Module Attributes Attribute Name Tag Type VR VM Notes SOP Class UID 0008 0016 1 Ul 1 SOP Instance UID Specific Character Set Instance Creation Date Instance Creation Time 1 Instance Creator UID 0008 0014 3 UI D 1 11 Discovery PET Raw Data Module Private Creator Identification GEMS PETD 01 Attribute Name Tag VR raw data ype A 1001 PET anase 1 PET raw dae bib 00231003 3 1 D 1 12 Discovery PET Correction Cal Module Private Creator Identification GEMS_PETD_01 Attribute Name Tag bi VR VM PET scan type 0009 1018 PET correction cal id 0017 1001 164 of 169 00171003 3 017 1009 017 1004 0009 1016 00171009 0077 1008 0017 1009 D 1 13 Discovery PET Well Counter Module Private Creator Identification GEMS_PETD_01 Attribute Name Tag Type VR VM PET scan_type 0009 1018 3 SL 1
78. ENT 8 9 2 ACCESSION i t n Pr e e a p f e s A1 OT SPECIFIC IOD DEEINITION eee eet eR C Ree EE e P Dade A 1 1 CT Image IOD Modules ete ie e P i m n red A 1 2 CT Image Modules Dag ee an m ae le ae de A 1 2 1 Implementation Specific details ssessseeeeeeeeeeenenen nen nennen 2 5 SPECIFIC IOD 1 0000 enter A 2 1 SC Image 10D Modules 3 nett eT dd Ee dau da CT SC IOD COMMON MODULE 8 2 01 6 66 nennen nnns nnne nene A 3 1 Patient Module a ee e este e er eee iE ver iets A 3 2 General Study Module nennen nnn A 3 3 Patient Study ine ba dt e bere e den Reine dee A 3 4 General Series Module ss eese ee a ero a e leet e Ea A 3 5 General Equipment Module esistente nnne nnn A 3 6 General Image Module A 3 7 Image Plane Module eese seiten A 3 8 Contrast Bolus Mod le morresi enere a ee A 3 9 SOF Common Modi le 2 1 tee atis ero deo eye Pe ue euo Pag A 3 10 VO
79. Exam Split and Hard Exam Split Virtual Exam Split will create Gray Scale Presentation State and Performed Presentation State e In Hard Exam Split application will send the user selected images from grouped Images that acquired during acquisition PPS will not sent along with these images The DICOM Modality Gray Scale Presentation State service is provided by the Exam Split The Remote Host DICOM AE is commanded to perform transmission of Gray Scale Presentation State object through the user interface The Remote Host DICOM AE is commanded to perform transmission of Hard Split Images through the user interface 111 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 8 2 1 Application Data Flow Diagram The basic Application models for the feature are shown in the following illustration O perator Scanner receives MWL Requested Procedures from H IS RIS associates images display settings W indow Level W indow W idth Pan Zoom and Image Orientation with a requested procedure Operator selects multiple requested procedures and scans images as a single exam Y After post processing Exam Split application presents user selected images and Requested procedures for split to create PPS and GSPS objects GSPS Viewer Remote 8 2 2 Functional Definition of AEs 8 2 3 Sequencing of Real World Activities 8 2 3 1 GSPS Acquisition System
80. Extended boot record signature 41 39 42 Date Stamp Volume serial number 43 53 0 474540535 4449434 4 GEMS DICOM D20 54 61 The file system label 62 509 ___ _ Don t care Any contents acceptable 510 511 OxAA55 0x55AA Signature flag 3 4 3 2 4 Directory Information in DICOMDIR Conformant Application Entities shall include in the DICOMDIR File a Basic Directory IOD containing Directory Records at the Patient and subsidiary levels appropriate to the SOP 53 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Classes in the File set All DICOM files in the File set incorporating SOP Instances defined for the specific Application Profile shall be referenced by the Directory Records All DICOM SOP Instance files in the File set are stored in the MSDOS filesystem subdirectory GEMS and are referenced from the DICOMDIR directory records Refer to Appendix C for a complete listing of all the optional modules and optional attributes used in the DICOMDIR definition Please note that the mandatory attributes in each of the directory records as mentioned in Part 3 Addendum Basic Directory Information Object Definition are not listed in the appendix but are supported by the implementation Note The modality attribute 0008 0060 in the DICOMDIR Series records should be CT MR for Secondary Capture and Standalone Overlay SOP Class Images Otherwise the Discovery DICOM media
81. File with all types of Directory Records related to the SOP Classes stored in the File set FSC for the MOD profile has no concept of finalizing the disc at the completion of most recent write session This is a requirement of CD R and does not apply to the MOD profiles 3 4 2 1 2 File Set Reader The role of File Set Reader is used by the Application Entities that receive a transferred File Set Typical entities using this role would include display workstations and archive systems that receive a patient record transferred from another institution File Set Readers shall be able to read all the SOP Classes defined for the specific Application Profile for which a Conformance Statement is made using all the defined Transfer Syntax 3 4 2 1 3 File Set Updater The role of File Set Updater is used by Application Entities that receive a transferred File Set and update it by the addition of information Typical entities using this role would include analytic workstations that for instance may add to the File set an information object containing a processed e g edge enhanced image Stations that update patient information objects would also use this role File set Updaters do not have to read the images File set Updaters shall be able to generate one or more of the SOP Instances defined for the specific Application Profile for which a conformance statement is made and to read and update the DICOMDIR file FSU shall offer the ability to allow multi s
82. GE Healthcare Discovery Dicom Conformance Statement for DICOM 2007 ID Net v3 0 Refer to section 1 for a list of products in the Discovery families to which this Conformance Statement applies 5198351 100 Rev 1 2005 General Electric Company All Rights Reserved GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Page Left Intentionally Blank 2 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT LEGAL NOTES TRADEMARKS All products and their name brands are trademarks of their respective holders OMISSIONS amp ERRORS Customers please contact your GE Sales or Service representatives GE personnel please use the GE Healthcare PQR Process to report all omissions errors and defects in this publication Copyrights All Material Copyright c 2005 by the General Electric Company All rights reserved 3 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Important Precautions WARNING This Service Manual is available in English only f acustomer s service provider requires a language other than English it is the customer s responsibility to provide translation services Do not attempt to service the equipment unless this service manual has been consulted and is understood Failure to heed this Warning may result in injury to the service provider operator or patient from electri
83. HO SCU The Discovery DICOM Server AE provides standard conformance to the DICOM Verification Service Class Each ECHO operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 15 seconds and 60 minutes respectively 2 3 1 2 5 Real World Activity Storage Commitment 2 3 1 2 5 1 Associated Real World Activity The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname Select the Update option if the hostname is already present Set the Archive Node option to Yes If 32 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT the remote host is not present add the remote host in the local database with Archive Node option set to Yes Note The remote node should be a Storage Commitment SCP The operator must then select Archive pull down menu on the local database manager Select Select Media option menu Select Remote Node in the menu Note Only if you configure the remote node as a archive node will you see the remote node option in the archive menu The operator then selects one or more study series or images in the local database manager amp and then chooses Save Examination Save Series or Save Image from the Archive pull down menu on the local database
84. ICOM 3 0 SOP class as an SCU SOP class name SOP class UID Storage Commitment Push model SOP class 1 2 840 10008 1 20 1 5 3 1 1 Association establishment policies 5 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 70 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for association initiated by the DICOM Storage Commitment SCU AE is Maximum Length PDU 10Kbyes ___ 5 3 1 1 2 Number of associations A single association is used to send the commitment request This commitment request association is closed after the request is successfully acknowledged lt accepts an association for receiving the response from commitment provider at any time 5 3 1 1 3 Asynchronous nature Asynchronous window negotiation is not supported 5 3 1 1 4 Implementation identifying information The Implementation UID for this GEMS Implementation is Discovery CT Implementation UID 1 2 840 113619 6 55 Discovery PET Implementation UID 1 2 840 113619 6 131 5 3 1 2 Association initiation policy An association is initiated by the Storage SCU to the provider to send the images to be committed A separate association is established to send the commitment request 5 3 1 3 Real World Activity 5 3 1 3 1 Asso
85. MENT 7 4 COMMUNICATION PROFILES 102 7 4 1 Supported Communication Stacks PS 3 8 02000002200000000000 102 7 4 2 OSI StaGh 102 7 4 3 FG PAP Stach inti andar 102 EASA gt AP tet tein ten elie es die ae nee nee ee Ades eee 103 4 443 2 Support ete crie Te creep toe det gea lec da 103 7 44 Point tosPoInt SLACK os us ti i i e 103 75 EXTENSIONS SPECIALIZATION PRIVATIZATION sese 103 7 5 1 Standard Extended Specialized Private SOPS esee 103 7 5 2 Private Transfer isses eene 103 7 6 en ate RE ee eo ERR Fee enti aa do Pe Ree OT AREE ar us 103 7 6 1 AE Title Presentation address Mapping sees 104 7 6 2 Configurable Parameters seii ia enne enimse enne nennen 104 77 SUPPORT OF EXTENDED CHARACTER 104 7 8 N CREATE REQUEST 5 5 104 7 9 ERROR HANDLING AND 106 7 10 USEOF SPEGIFIC DICOM DATA teet th iae eme ba fe ere te ed 107 7 11 USE OF SPECIFIC DICOM 109 7 11 1 Panoni LOVE ONDE 109 7 11 2 OlUdy Level oe bp E est tee ti els ma 110 7 11 3 TI LEVET MC 110 8 GRAYSCALE PRE
86. OD media compressed uncompressed images CT MR PT Studies on STD CTMR MOD12 Handles single 12 or 16 bit grayscale 1 2GB MOD media compressed uncompressed images CT MR PT Studies on PRI CTMR MOD12 Handles single 12 or 16 bit grayscale 1 2 GB MOD media compressed uncompressed images 47 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 4 2 Clinical Contexts This Application Profile facilitates the interchange of primary CT and MR images as well as processed CT and MR images as Secondary Capture and Standalone Overlay Images with certain defined attributes CT MR PT SC and OV images may coexist within the same File set Typical interchanges would be between acquisition devices archives and workstations within and between institutions 3 4 2 1 Roles and Service Class Options This Application Profile Class uses the Media Storage Service Class defined in PS3 4 The Application Entity shall support one or more roles of the File set Creator File set Reader and File set Updater defined in PS3 10 3 4 2 1 1 File Set Creator The Application Entity acting as a File set Creator generates a File Set under the PRI CTMR MOD12 Application Profile Class Typical entities using this role would include CT or MR equipment and archive systems that generate a patient record to transfer to another institution File Set Creator shall be able to generate the Basic Directory SOP Class in the DICOMDIR
87. OP Class Name SOP Class UID Modality Performed Procedure Step 1 2 840 10008 3 1 2 3 3 100 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 7 3 1 1 Association Establishment Policies 7 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU from an association initiated by the PPS Server AE is Maximum Length PDU 128 Kbytes The SOP Class Extended negotiation is not supported The maximum number of presentation negotiation items that will be proposed is 1 The user information items sent by this AE are e Maximum PDU Length e Implementation UID 7 3 1 1 2 Number of Associations The PPS Server will initiate only one DICOM association at any time to perform a PPS operation to the remote AE 7 3 1 1 3 Asynchronous Nature Asynchronous mode is not supported All operations are performed synchronously 7 3 1 1 4 Implementation Identifying information The implementation UID for this DICOM Implementation is PPS Feature for Discovery Implementation UID 1 2 840 113619 6 55 7 3 1 2 Association Initiation Policy The PPS Server AE initiates a new association for every PPS operation initiated 7 3 1 2 1 Real World Activity Performed Procedur
88. ORMANCE STATEMENT Implementation Version Name Helios Ver1 3 3 AE Specifications 3 3 1 MOD Archive Specification The MOD Archive provides standard conformance to DICOM Media Storage Service Class The Application Profiles and roles are listed in Table 3 1 1 Table 3 1 1 Application Profile Activities and Roles for MOD Update Application Profiles STD CTMR MOD23 Create MOD FSC pod ______ FSU 1 Display Directory FSR Copy To Local Storage FSR px STD CTMR MODi2 CreteMOD FSC pod Update MODE FSU 1 Display Directory _ OopyToLocaStoage FSR PRECTMR MOD12 UpdateMOD FSU 1 Display Directory po Copy Storage FSR The MOD Archive Application will query the user before initializing media when a File set is found on the media and an initialize operation has been requested Depending on the physical media type being 2 3GB or 1 2GB media the system will apply the correct characteristics as an FSC to meet the appropriate Application Profile As an FSU role the system will produce directory sequences for each SOP instance that is added to the File set consistent with the STD CTMR MOD 12 23 profile and it will make the File set directory so that an early installed base system that only understands the PRI CTMR MOD12 profile will not be able to use the media An upgrade to
89. P 1 2 840 10008 5 1 1 14 DICOM Implicit 1 2 840 10008 1 2 SCU Class VR Little Endian Transfer Syntax Note Certain Presentation Contexts may be negotiated that are not used during the association See the following sections for the DICOM Print services performed by the Discovery 4 3 1 2 2 SOP Specific Conformance Statement The Print SCU supports the following mandatory SOP classes which are defined under the Basic Grayscale Print Management Meta SOP Class The Print SOP Classes supported by Print SCU NAME Basic Film Session SOP Class 1 2 840 10008 5 1 1 1 Basic Film Box SOP Class 1 2 840 10008 5 1 1 2 Basic Grayscale Image Box SOP Class 1 2 840 10008 5 1 1 4 Printer SOP Class 1 2 840 10008 5 11 16 62 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT The Print SCU does not support any optional SOP Classes Note The asterisk preceding some of the SOP attributes in the following sections indicates optional SCP attributes The SCU is aware of these optional attributes and will not fail on the receipt of a response without the optional attributes the limitation being that this information is not available for the User Some SCP s will terminate the association with a failure upon the receipt of an optional attribute it does not support 4 3 1 2 2 1 Basic Film Session SOP Class The Print SCU supports the following DIMSE Service Elements for the Basic Film Session SOP Class
90. PEUT Module ize th etim dte en Es A 3 11 Frame of Reference Module sisse esses entree 3 12 SC Equipment 4 PET SPECIFIC IOD COMMON MODULE DEFINITIONS eese 4 1 PET Image IOD 0 0 42222 0004400550060 4 2 Patient Module idend Dee HS ser EE ease A 4 3 General Study Module isses saei A 4 4 Patient Study Module ebrei ae He Hd e i e Poe dendo A 4 5 Discovery PET Exam Module sss einen nnne nnne nennen A 4 6 General Seri s Module ii tr e ria 4 7 PET S ries Module s ibn ee a AR TS RI AR A 4 8 PET Isotope Modu le eei ERR SU DO RR A 4 9 PET Multi gated Acquisition sse nnns A 4 10 NM PET Patient Orientation A 4 11 Discovery PET ImageSet Module sisse eene A 4 12 Discovery PET Scan A 4 13 Frame of Reference Module sess sese enhn A 4 14 General Equipment Module esses eitis A 4 15 General Image Module A 4 16 Image Plane Module esses essent nennen thea nite A 4 17 Image
91. Patient Patient A 4 2 Study General Study A 4 3 Patient Study A 4 4 Discovery PET Exam A 4 5 Series General Series A 4 6 PET Series A 4 7 PET Isotope A 4 8 PET Multi gated Acquisition A 4 9 NM PET Patient Orientation A 4 10 Discovery PET Imageset A 4 11 Discovery PET Scan A 4 12 Frame of Reference Frame of Reference A 4 13 Equipment General Equipment A 4 14 Image General Image A 4 15 Image Plane A 4 16 Image Pixel A 4 17 PET Image A 4 18 Overlay Plane 4 19 VOI LUT A 4 20 Discovery PET Image A 4 21 Discovery PET Frame A 4 22 General Modules SOP Common A 4 23 A 4 2 Patient Module This section specifies the Attributes of the Patient that describe and identify the Patient who is the subject of a diagnostic Study This Module contains Attributes of the patient that are needed for diagnostic interpretation of the Image and are common for all studies performed on the patient Attribute Name Tag Type VR VM Notes Patient s Name 0010 0010 2 64 1 Patient ID 0010 0020 2 LO 64 1 131 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Patient s Birth Date 0010 0030 2 DA 26 1 Patient s Sex 0010 0040 2 5 16 1 Referenced Patient Sequence 0008 1120 3 SQ 1 Not used gt Referenced SOP Class UID 0008 1150 1 Ul 1 Not used g
92. Reference UID 0020 0052 in conjunction with the Table Height 0018 1130 to determine if two images are spatially related 124 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT A 1 2 1 Implementation Specific details e Forall CT Images created by Discovery patient birth date element 0010 0030 is sent if the operator has entered the details If the operator has not entered the details it will be sent as a Zero length element length of Accession number element 0008 0050 is a maximum of 16 characters e Window Center 0028 1050 and Window Width 0028 1051 are sent for all CT Images created by Discovery e Based on whether contrast was used or not the following applies for CT Images created by Discovery 0018 0010 0018 1040 Oral Contrast Agent name The string Oral IV Contrast IV Contrast agent name The String IV Oral and IV Oral contrast agent amp The string Oral amp IV contrast IV Contrast agent e Supported field length for DICOM fields A 2 SCSpecific IOD Definition A 2 1 SC Image IOD Modules Entity Name Module Name Reference Usage Paien Suy A52 M ________ ____ lt Series General Series asa Equipment General Equipment A58 ______ M General mage _ A56
93. Request Module Table 6 8 11 1 1 Imaging Service Request Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type Image Accession Number 0008 0050 Co Requesting Physician 0032 1032 Referring Physician s Name 0008 0090 Requesting Service 0032 1033 6 8 12 Visit Entity 6 8 12 1 Visit Identification 3 No 2 User will be able to enter value for Accession Number prior to query to be sent as part of C FIND request Supports maximum of 16 characters 2 No 2 Truncated to 32 characters Table 6 8 12 1 1 Visit Identification Module Attributes Attribute Name Tag Expected Matching Returned Key Type Admission ID 0038 0010 0006 0080 6 8 12 2 Visit Status Table 6 8 1 12 2 1 Visit Status Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type Image 0300 Current Patient Location 0038 0300 Loe ee Displayed on More Info screen 89 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 12 3 Visit Relationship Table 6 8 12 3 1 Relationship Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type E NE Referenced Patient Sequence 0008 1120 2 M 6 8 12 4 Visit Admission No data elements are requested from the
94. SCU does not support any Extended Character Set 68 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 5 DICOM Storage Commitment SCU Conformance 51 Introduction This section of the DICOM conformance statement specifies the Storage Commitment Push Model SOP class as user the optional attributes and service elements supported 5 2 Implementation Model 5 2 1 Application data flow diagram Illustration 5 1 1 shows the data flow diagram of Storage Commitment SCU There is only one Real World activity that will cause the DICOM Server Application Entity DICOM Server AE to initiate an association to a remote DICOM entity that is a Service Class Provider SCP of the Storage Commitment SOP class The Choose Archive save Option Real World activity consists of an operator choosing a remote DICOM AE that supports Storage Commitment as provider as the archive device and selecting one or more exam or series from the local database and choosing either Save Exam or Save Series from the archive menu The images to be committed are sent to the remote provider entity first The Commitment request for the transferred image instances is sent after the complete image transfer The Commitment response must come on a different association There is no Real World activity required for the DICOM Server AE to respond to an incoming Storage Commitment response from the remote DICOM AE Illustration 5 2 1 Application
95. SENTATION 5 442422 2 222 111 8 1 INI RODUGTIONS osten des dee co Mene a ceto tabo cos db t ed 111 8 2 IMPLEMENTATION MODEL 2 12 00 010000000000000000000000000000 111 8 2 1 Application Data Flow Diagram eese eene 112 8 2 2 Functional Definition of AES sees nnns 112 8 2 3 Sequencing of Real World Activities essen 112 8 2 3 1 GSPS Acquisition System with MWL 112 8 2 4 GSPS Server AE Specification sss eitis 113 8 2 4 1 Association Establishment eee 113 8 2 4 1 1 Gerierallz abiecit MEA 113 8 2 4 1 2 Number of Associations 200 0 4 00010000 113 8 2 4 1 3 Asynchronous Nature initi RE RR EH RRRRU EXER 113 8 2 4 1 4 Implementation Identifying 113 8 2 4 2 Association Initiation 113 8 2 4 2 1 Real World Activity GSPS ted tede e pet deae ciere 114 8 2 4 2 1 1 Associated Real World 114 8 2 4 2 1 2 Proposed Presentation Context 2 2 2 040 00 114 8 2 4 8 Association Acceptance 114 83 COMMUNICATION 2 114 8 3 1 Supported
96. SOP Specific Conformance Statement for SCU When the operator starts a Get operation at any level Study Series Image the DICOM Server AE will initiate a C MOVE RQ request to the Remote AE with the DICOM Server AE as the Destination AE The Storage SCP will handle the incoming images as described in section 2 2 1 3 1 Each C MOVE SCU operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 900 seconds 300 seconds and 60 minutes respectively The DICOM Server AE will send a C CANCEL MOVE RQ to the Remote AE if the operator Pauses or Clears the job from the local database manager Network queue 2 3 1 2 4 Real World Activity Verify DICOM Ping 2 3 1 2 4 4 Associated Real World Activity The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname The operator must then select the Ping DICOM host form Network pull down menu The DICOM server will initiate an association with the remote DICOM AE in order to verify communication at the application level The success or failure of the verification process is displayed to the user 2 3 1 2 4 2 Proposed Presentation Context Table Refer to the Table 2 2 1 2 1 2 1 for the Proposed Presentation Contexts for DICOM Server AE and Real World activity Verification 2 3 1 2 4 21 SOP Specific DICOM Conformance C EC
97. SPEED B 1 2 Private Creator Identification GEMS ACQU 01 0019 1002 0019 1003 Cell spacing 0019 1004 First scan ras 00191018 LO Last scan ras 0019 101A LO Table speed 0011023 DS Mid scan time 0019 1024 DS Degrees of azimuth 0019 1026 51 Gantry period 001910277 DS Number of triggers 0019 102C SL 0019 106A 147 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT B 1 3 Private Creator Identification GEMS RELA_01 Attribute Name Series from which Prescribed 0021 1003 Series Prescribed From 0021 1035 0021 1036 B 1 4 Private Creator Identification GEMS STDY 01 Attribute Name Start time secs in first axial 0023 1070 B 1 5 Private Creator Identification GEMS 01 0027 0027 0027 1044 0027 1045 0043 1016 Corrected after glow terms 0043 1021 Reference channels 0043 1025 0043 101F No views ref chans blocked 0043 1026 148 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Tag 0043 1043 0043 1044 0043 1045 0043 1064 B 1 7 Private Creator Identification GEMS HELIOS 01 Note Dicom elements 0045 1030 1034 and 0045 1036 1039 and 0045 103B are present only if the appropriate cardiac option is installed on the scanner 0045 1001 SS 1 0045 1004 0045 1005 0045 1
98. Split End Date Application will calculate Performed Procedure Step 0040 0251 Exam Split End Time Application will calculate Performed Procedure Step 0040 0252 Set as In Status Progress Performed Procedure Step 0040 0253 X Will be set as ID Requested Procedure ID 0040 1001 from 0040 0275 Performed Procedure Step 0040 0254 X This will be set as Description Requested Procedure Code Meaning If Code Meaning absent it will be Requested Procedure Description Performed Procedure Type Description 0040 0255 Set as empty Performed Protocol Code Sequence 0040 0260 Set as Empty sequence 119 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Performed Series Sequence 0040 0340 X Exam Split Application will calculate and copy from Image s Study and set these sequence values Rescale Type 0028 1054 Exam Split Application will set this value Referenced Performed Procedure Step Sequence 0008 1111 Exam Split Application will calculate and set these sequence values Manufacturer 0008 0070 Exam Split Application will set as GE MEDICAL SYSTEMS Instance Number 0020 0013 Exam Split Application will set as OFFIS_DCMTK_ 341 Presentation label 0070 0080 This will be set as Requested Procedure Code Meaning or Requeste
99. T SL FL SL DT SL 1 1 1 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT acq bin dur 0009 106E PET acq bin start 0009 106F PET actual stop cond 0009 1070 SL WS St 1 PET total_prompts PET total_delays PET frame valid PET validity info PET compression PET uncompressed size PET accum bin dur PET frame number PET list file exists PET where is list frame PET unlisted scan 0009 10E1 PET phase percentage PET acq bin dur percent 0009 10E9 3 FL 1 B 2 6 Discovery PET ImageSet Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET compatible_version 0009 1079 3 ISH 1 PET software_version SH 1 PET is datetime PET is source 0009 107C PET is contents 0009 107D PET is reference PET multi patient PET number of normals 0009 1081 PET color map id 0009 1082 PET window level type 155 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT PET rotate PET zoom FL SL FL SL SL FL FL SL SL FL SL FL SL FL FL SL SL SL UI UI UI UI FL FL UI SL FL SL FL FL FL SL SL SL PET window_level_min PET window_level_max PET recon_method PET attenuation 0009 108C PET atten_coefficient PET bp_filter PET bp filter cutoff PET bp fi
100. UID 0008 1150 1C 1 Not Used gt Referenced SOP Instance UID 0008 1155 Not Used Derivation Description 0008 2111 Truncate ImageSet derivation to 1024A Source Image Sequence 0008 2112 Not Used gt Referenced SOP Class UID 0008 1150 1 Not Used gt Referenced SOP Instance UID 0008 1155 fic ul 1 Not Used Images in Acquisition 0020 1002 3 IS 1 Not Used Image Comments 0020 4000 Not Used Lossy Image Compression 0028 2110 1 Superceded by PETImage Lossy Image Compression 140 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 0x0028 0x2110 A 4 16 Image Plane Module This section specifies the Attributes which define the transmitted pixel array of a two dimensional image plane Attribute Name Tag Type VR VM Notes Pixel Spacing 0028 0030 1 05 2 Value 1 pixel_width Value 2 pixel_height Image Orientation Patient 0020 0037 DS ImageSet patient_row_cos_ ImageSet patient row cos p ImageSet patient row cos s ImageSet patient col cos ImageSet patient col cos p ImageSet patient col cos s 3 Image Position Patient 0020 0032 DS 3 patient 1 patient p patient s ACE image_thickness image_location Slice Thickness 0018 0050 Slice Location 0020 1041 A 4 17 Image Pixel Module This section specifies the Attributes that describe the pixel data of the image A
101. Visit Admission Module 6 8 13 Patient Entity 6 8 13 1 Patient Relationship No data elements are requested from the Patient Relationship Module 6 8 13 2 Patient Identification Table 6 8 13 2 1 Patient Identification Module Attributes Expected Expected Mapped Attribute Name Tag Matching Returned into the Key Type Key Type Image Patient s 0010 0010 1 Yes This attribute is sent with zero length Name Supports up to 32 characters See also Note1 Patient ID 0010 0020 1 Yes This attribute is sent with zero length Supports up to 16 characters e Modality Worklist server supports 32 characters for patient Name Supports DICOM format for patient Name with as delimiters e f patient name worklist has more than 32 characters then Worklist will be accepted by the server Worklist Browser will display only the first 32 characters More Info screen will display the full patient name Only the FIRST 32 characters is copied into the final DICOM image header 90 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 13 3 Patient Demographic Table 6 8 13 3 1 Patient Demographic Module Attributes Patient s Sex 0010 0040 2 Y Y Ta Patient s Birth Date 0010 0030 2 Yes This value is also used to calculate the Patient s Age EE CONUM eee Patient s Weight 0010 1030 Limited to maximum value of 999 kg Confidentiality constraint 0040 3001 2 No
102. Zerolengh Columns O 0028 0011 jZerolengh _Image position 77 0 ____ 0020 0032 Zero length Image orientation 17 0 0020 0037 Zerolength Slice thickness 10 00180050 Zero length 30 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Description Type Tag Value Slice spacing 00180088 Zero length _____ 001 120 Zerlengh Convolution keme 00184210 Zerolengh Reconstruction diameter O 0018 1100 Zerolengh Data collection diameter O 0018 090 Zerolengh Flip angle 00184814 Zero length Echo number 008 086 Zero length Echo time 00180081 Zero length inversion time 0018 0062 Zero length Repetition time 00150080 Zerlengh Private Creator Identification P 001900 GEMS_ACGU_01 Dfov Rect 0019004 Zero Length Dfov Rect 0019 Zero Length Midscan Time P 00199094 Zero Length Azimuth P 00199026 Zero Lengthy Number of Echo 00976 Zero Length Private Creator Identification P 002100 GEMS_RELA 01 ScuAwe 0021 Zero Length Private Creator Identification P 002700 GEMS MAGO Location RAS P 002740 Zero Length Location 002741 Zero Length Center R Coordinate P 0027 42 Zero Length Center A Coordinate P 00753 Zero Length Table Start Location 0027580 Zero Length Table End Lo
103. ality the Modality will be filled in as follows 0008 0060 CS CT 2 1 Modality If the query is for records for this Scanner the Modality will be filled in with CT as above and the Scheduled Station AE Title will be filled in with the value configured for this system For example this station was configured as CTRoom1 0040 0001 AE CTRoom1 8 1 ScheduledStationAETitle User will be able to enter the values for Accession Number prior to the query If value is entered then that value will be sent as part of the query For example if 1234 is entered then 0008 0050 SH 1234 d 4 1 AccessionNumber 93 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT User will be able to enter the values for Requested Procedure Id prior to the query If value is entered then that value will be sent as part of the query For example if 3456 is entered then 0040 1001 SH 3456 4 1 RequestedProcedureID User will be able to enter the values for Patient Id prior to the query If value is entered then that value will be sent as part of the query For example if 6789 is entered then 0010 0020 LO 6789 4 1 PatientID User will be able to enter the values for Patient prior to the query If value is entered then that value will be sent as part of the query For example if Lastname Firstname is entered then 0010 0010 PN Lastname Firstname
104. amera installation utility 4 6 2 Configurable Parameters The following fields are configurable for the DICOM Print SCU AE Note All configurations should be performed by a GE Field Service Engineer Local IP Address Local IP Netmask Max PDU length The following DICOM print parameters are configurable The valid ranges are shown in earlier sections of this conformance statement e Medium Type 2000 0030 67 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Film Destination 2000 0040 Magnification Type 2010 0060 Min Density 2010 0120 Max Density 2010 0130 Empty Image Density 2010 0110 Border Density 2010 0100 Configuration Information 2010 0150 Smoothing Type 2010 0080 e The following network timers are supported and are configurable When these timers expire the association is terminated The default values are e Association Time out 120 seconds Session Time out 1200 seconds total time for open association N SET Time out 300 seconds N ACTION Time out 300 seconds N CREATE Time out 300 seconds N DELETE Time out 300 seconds N GET Time out 100 seconds Time outs which are set for all hosts are configurable in dprint cfg Association time out bi_assoc_tio Session timeout bi_session_tio Create timeout bi_ncreate_tio Note All configurations are performed by a GE Field Service Engineer 4 7 Support of Extended Character Set The Print
105. apped into images and are not otherwise dealt with displayed on the user interface etc are not used and are thus discarded upon receipt See Table B 1 for further information Data elements for which values can be sent for matching purposes are described as such Data elements for which values are not sent are sent with zero length and universal matching will apply This is the default case if no other description to the contrary is provided 6 8 8 Supported Matching The following are the types of matching that can be request by the implementation e Single Value matching e Universal Matching e Range of date time 6 8 9 Scheduled Procedure Step Entity 6 8 9 1 SOP Common Module Table 6 8 9 1 SOP Common Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type Image Specific Character Set 0008 0005 1 Matching for this item is supported only for the character set ISO IR 100 This value is always sent and therefore must be returned 86 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 9 2 Scheduled Procedure Step Module Table 6 8 9 2 Scheduled Procedure Step Module Attributes Attribute Name Scheduled Procedure Step 0040 0100 Sequence Scheduled Station AE 0040 0001 Matching is supported as follows either no Title AE title is supplied universal matching or the scanner s Worklist
106. as part of image header Not sent in MPPS N N Create Create Create Performed Procedure Step 0040 0250 Date when all Date when all images The date Complete End Date images got got installed PPS or installed Discontinue PPS is invoked not the actual end of post processing Performed Procedure Step 0040 0251 Time when all Time when all images The time Complete End Time images got got installed PPS or installed Discontinue PPS is invoked not the actual end of post processing Modality 0008 0060 Value CT is Value CT is stored Value CT is stored stored in image in image header in image header header Study ID 0020 0010 Same as exam Same as exam Copied from source number number image Performed Action Item 0040 0260 Not used Not used Not used Code Sequence Performed Series 0040 0340 One item for One item for each One item for each Sequence each series series created series created with created post processing gt Performing Physician s 0008 1050 Copied from User input on the Not sent as part of Name SPS if selected scanner image header gt Protocol Name 0018 1030 The name of the protocol selected on the scanner The name of the protocol selected on the scanner Copied from source image gt Operator s Name 0008 1070 Copied from SPS if selected and present otherwise User input on the scanner No
107. ase note that the mandatory attributes in each of the directory records as mentioned in Part 3 Addendum Basic Directory Information Object Definition are not listed in the appendix but are supported by the implementation Directory records found in the PRI CTMR MOD12 profile do not contain the Attribute Frame of Reference 0020 0052 This Attribute was specified as Required after the first release of this Application Profile Note The modality attribute 0008 0060 in the DICOMDIR Series records should be CT MR for Secondary Capture and Standalone Overlay SOP Class Images Otherwise the Discovery DICOM media display browser will not list up the series for contents rendering the retrieval to local storage impossible 3 4 4 1 4 1 Additional Keys None 55 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 4 4 1 5 Other Parameters This section defines other parameters common to all specific Application Profiles in the PRI CTMR MOD12 class that need to be specified in order to ensure interoperable information interchange See 3 4 2 for JPEG encoding problem in the PRI CTMR MOD12 profile 3 4 4 1 5 1 Image Attribute Values The attributes listed in Table 3 3 3 1 5 1 1 used within the CT MR PT SC Standalone Overlay Image files shall take the values specified Overlay data if present shall be encoded in Overlay Data 60XX 3000 3 4 4 1 5 2 Attribute Value Precedence None 3 5 Extensi
108. ata elements per the DICOM standard Unspecified private data element information is subject to change without notice If however the device is acting as a full fidelity storage device it should retain and retransmit all of the private data elements that are sent by GE devices 1 5 Important Remarks The use of these Conformance Statements in conjunction with the DICOM Standards is intended to facilitate communication with GE imaging equipment However by itself it is not sufficient to insure that inter operation will be successful The user or user s agent needs to proceed with caution and address at least four issues e Integration The integration of any device into an overall system of interconnected devices goes beyond the scope of standards DICOM and of this introduction and associated Conformance Statements when interoperability with non GE equipment is desired The responsibility to analyze the applications requirements and to design a solution that integrates GE imaging equipment with non GE systems is the user s responsibility and should not be underestimated The user is strongly advised to ensure that such integration analysis is correctly performed e Validation Testing the complete range of possible interactions between any GE device and non GE devices before the connection is declared operational should not be overlooked Therefore the user should ensure that any non GE provider accepts full responsibilit
109. ate 0008 0022 DA extract date from Scan scan_datetime Image image_time Acquisition Time 0008 0032 extract time from Scan scan_datetime Image image_time Actual Frame Duration 0018 1242 1 image_duration Converted from oe in DB to milliseconds Nominal Interval 0018 1062 IS Not Used Primary Counts Prompts Accumulated Intervals Acquired 0018 1083 Scan triggers_ acquired Intervals Rejected 0018 1084 IS Scan triggers_rejected 0054 1310 3 IS 1 total prompts Secondary Counts 0054 1311 3 IS 1 n Value 1 total delays Accumulated Slice Sensitivity Factor 0054 1320 3 DS 1 IF coefficient NULL THEN coefficient ELSE 1 0 Decay Factor 0054 1321 decay_factor Dose Calibration Factor 0054 1322 3 DS 1 activity factor 1 0e 06 convert MBq ml to Bq ml Scatter Fraction Factor 0054 1323 3 DS 1 scatter subtracted total counts Dead Time Factor 0054 1324 3 DS 1 deadtime factor Referenced Overlay Sequence 0008 1130 3 50 1 Not Used gt Referenced SOP Class UID 0008 1150 EEIE Not Used gt Referenced SOP Instance 0008 1155 UI Not Used UID Referenced Curve Sequence 0008 1145 Not Used gt Referenced SOP Class UID COCOS em Not Used SOP Instance Not Used UID Anatomic Region Sequence 0008 2218 Not Used 144 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE
110. ation Context Table Table 2 3 1 3 3 2 1 Acceptable Presentation Contexts for DICOM Server AE and Real World Activity Query Request Presentation Context Table Abstract Syntax Transfer Syntax Extended Negotiation UID List TUE amp Each C FIND SCP operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 300 seconds and 60 minutes respectively 2 3 1 3 3 21 SOP Specific Conformance to C FIND SCP All Required and Unique U study series and image level keys for the Study Root Query Retrieve information model are supported Some optional O keys are also supported as described in the following tables Table 2 3 1 3 3 2 1 1 Supported study level keys Tag Usage Study date R 00080020 Study time 0008 030 Matched Accession number 77 00080050 matched 36 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT ___ Patient s name 8 00100010 __ Matched 17 R 00100020 Matched Studyid 8 00200010 Matched Studydescripion 0 00081030 Retumed Suteld P 00090002 Returned Table 2 3 1 3 3 2 1 2 Supported series level keys Description 7 Type Tag Usage R 00080060 Series number 0020001 Matched _____ Series description 71 0008 103 Returned
111. ation contexts for the Worklist Server AE after real world activity Worklist Query has been initiated Presentation Context Table Proposed Abstract Syntax Transfer Syntax Extended UD Modality Worklist 1 2 840 10008 5 1 4 31 Implicit VR Little Endian 1 2 840 10008 1 2 SCU None Information Model FIND 6 3 1 2 1 21 SOP Specific DICOM Conformance Statement for the Worklist SOP Class If the remote AE does not support the proposed Presentation Context an appropriate error is logged and the operator is notified This implementation can receive multiple C FIND results over a single association Only one association is opened at a time Each C FIND response received from the remote AE is parsed to verify the length type of the items in the response see section 2 3 1 2 1 2 2 for more information Upon detecting any error in the response data the Worklist Server AE will issue a C FIND CANCEL and upon receipt of a C FIND RSP or if an applicable timer expires will abort the association All previously received worklist items are retained On receipt of any error from the remote AE the Worklist Server will issue a C FIND CANCEL and upon receipt of a C FIND RSP or if an applicable timer expires will abort the association All previously received worklist items are retained Warnings received from the remote AE are ignored Each C FIND operation supports a configurable Association Timer
112. c shock mechanical or other hazards AVERTISSEMENT Manuel de maintenance n est disponible qu en anglais Si le technicien du client a besoin de ce manuel dans une autre langue que l anglais c est au client qu il incombe de le faire traduire pas tenter d intervention sur les quipements tant que le manuel Service n a pas t consult et compris Le non respect de cet avertissement peut entra ner chez le technicien l op rateur ou le patient des blessures dues des dangers lectriques m caniques ou autres WARNUNGI Dieses Kundendienst Handbuch existiert nur in englischer Sprache Falls ein fremder Kundendienst eine andere Sprache ben tigt ist es Aufgabe des Kunden f r eine entsprechende Ubersetzung zu sorgen Versuchen Sie nicht das Ger t zu reparieren bevor dieses Kundendienst Handbuch zu Rate gezogen und verstanden wurde Wird diese Warnung nicht beachtet so kann es zu Verletzungen des Kundendiensttechnikers des Bedieners oder des Patienten durch elektrische Schl ge mechanische oder sonstige Gefahren kommen AVISO Este Manual de Servicio s lo existe en ingl s Si algun proveedor de servicios ajeno a GEMS solicita un idioma que no sea el ingl s es responsabilidad del cliente ofrecer un servicio de traducci n No se deber dar servicio t cnico al equipo sin haber consultado y comprendido este manual de servicio La no observancia del presente aviso puede dar lugar a que el pro
113. call has failed while processing an image A711 Out of Resources Indicates that probably there was not enough disk space to store the image The user should attempt recovery by removing some images from the Discovery system A712 Out of Resources Indicates that there was not enough resource such as memory to store the image A800 SOP Class not supported In the event of a successful C STORE operation the image has successfully been written to disk The image will then be accessed in the same manner as any other image by the applications on the Discovery system Images may be deleted when instructed to do so by the user Thus the duration of the storage of the image is determined by the users of the Discovery system Image Installation If the image installation is unsuccessful a message will appear in the Message Log informing the user of the failure and the image will be removed If the image installation process finds that an element is not encoded according to the DICOM standard it will fail to install the image and the file will be removed Image Installation of non GE Created MR CT or PET Images Images received from non GE products are installed as appropriate image object without any of their private data elements Also if some critical fields mandatory are missing then the image will not be installed 2 3 1 3 2 Verification Request from Remote AE This AE is indefinitely listening for associations No
114. cation P 007581 Zero Length RAS Letter for Side of mage P 0027 52 Zero Length TRAS Letter for Anterior Posterior P 007559 Zero Length RAS Letter for Scout Start Location P 00275064 Zero Length RAS Letter for Scout End Location P 0027 85 Zero Length mage DimensionX P 002760 Zero Length mage Dimension Y 002781 ZeroLength Note1 Refer to section 2 3 1 2 2 1 for Custom Search Query option Note2 Type P refers to a private dicom element 2 3 1 2 3 Get Image s from Remote AE 2 3 1 2 3 1 Associated Real World Activity The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname The operator then has to perform the Real World activity Query to get a list of Study Series Image Once the list of Study Series Image is retrieved the operator can invoke the Get operation by choosing Get Exam or Get Series or Get Image from the Network pull down menu 31 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 3 1 2 3 2 Proposed Presentation Contexts When the Real World activity Get is initiated all presentation contexts shown in table 2 2 1 2 1 2 1 are proposed during association establishment but only the Query Retrieve MOVE related contexts are applicable to this activity 2 3 1 2 3 2 1
115. ce of requested Type 3 attributes is not considered an error Fields considered Type 1 by the Worklist Server include 0010 0010 Patient Name 0010 0020 Patient ID 0020 000D Study Instance UID 0040 0001 Scheduled Station AE Title 0040 0002 Scheduled Procedure Step Start Date 0040 0003 Scheduled Procedure Step Start Time 0040 0009 Scheduled Procedure Step ID 0040 1001 Requested Procedure ID Start Date must be of the form YYYYMMDD exactly eight numeric characters and Start Time must be of the form HHMMSS exactly six numeric characters Fields considered Type 2 by Worklist Server include 0008 0050 0008 0060 0008 0090 Accession Number Modality Referring Physician Name 0010 0030 Patient Date of Birth 0010 0040 Patient Sex 0010 1030 Patient Weight in kg 0010 2000 Medical Alerts 0010 2110 Contrast Allergies 0010 21 0 Pregnancy Status 0032 1032 Requesting Physician 0032 1070 Requested Contrast Agent 0038 0010 Admission ID 0038 0050 Special Needs 0038 0300 Current Patient Location 0038 0500 Patient State 0040 0006 Performing Physician 0040 0010 Scheduled Station Name 0040 0011 Scheduled Procedure Step Location 0040 0012 Pre order Medication 0040 1003 Requested Procedure Priority 0040 1004 Patient Transport Arrangements 0040 3001 Confidentiality Constraint Sn ee 6 3 1 3 Association Accepta
116. cedure Types An Imaging Service Request is submitted by one authorized imaging service requester to one authorized imaging service provider in the context of one Service Episode 6 8 4 4 Visit Entity Description A Visit is the context in which the treatment or management of an arbitrary subset of a Patient s medical conditions occurs A Visit is limited to the description of a Patient s activities at a single facility 6 8 4 5 Patient Entity Description A Patient is a person receiving or registered to receive healthcare services 6 8 5 ModalityWorklist Mapping of DICOM Entities Table 6 8 5 1 Mapping of DICOM Entities to ModalityWorklist Entities DICOM ModalityWorklist Entity Scheduled Procedure Step 84 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 6 6 8 7 Information Model Module Table Within an entity of the DICOM Modality Worklist Information Model attributes are grouped together into related set of attributes called modules A module facilitates the understanding of the semantics concerning the attributes and how the attributes relate to one another A module grouping does not infer any encoding of information into datasets Table 6 8 6 1 identifies the defined modules within the entities that comprise the DICOM Modality Worklist Information Model Modules are identified by Module Name See DICOM PS 3 3 and PS 3 4 for a complete definition of the entities modules
117. ciated Real World Activity Save exam series The user selects the exam series to be committed All the images currently in the selected exam series will be sent to the selected remote archive node which is also the Storage commitment SCP using DICOM C STORE operations Once all the images are transferred the commitment request will be sent on a separate association 5 3 1 3 2 Associated Real World Activity Auto Archive exam series The Auto archive API s are used to archive the exams series onto local archive media or the remote archive node which shall be a Storage Commitment SCP without manual interface If the default device selected for Auto Archive is a remote Storage Commitment SCP then all the images currently in the specified exam series will be sent to the selected Storage commitment SCP using C STORE operations On successful transfer of all the images the Storage Commitment request will be sent 71 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 5 3 1 4 Proposed Presentation Contexts The Proposed Presentation Context table for the DICOM Storage Commitment SCU is as shown in following Table Presentation Context Table Abstract Syntax Transfer Syntax Role Extended Storage Commitment 1 2 840 10008 1 20 1 DICOM Implict VR Little 1 2 840 10008 1 2 SCU None Push Model SOP Class Endian Transfer Syntax The Storage Commitment SCU can send the commitment request for follo
118. d 7 7 SUPPORT OF EXTENDED CHARACTER SETS PPS feature for Discovery does not support any extended character set 7 8 amp NSET REQUEST MESSAGE PPS Feature for Discovery supports all named attributes listed in Table F 7 2 1 in PS3 4 of DICOM standard That is attributes that are not explicitly referenced by name in the table are not supported Example is last row in the table reads All other attributes from Radiation Dose Module and Billing and Material Code Module The attributes referenced here are not supported 104 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT For the MPPS associated with an acquisition the following attributes are copied from the Modality Work list SCU into the MPPS request Message if procedure performed corresponds to the SPS information retrieved through the Modality Work list Discovery supports the selection of single or multiple SPS for a scan The following are applicable Single SPS selection results in single PPS message Multiple SPS selection results in single PPS message Multiple SPS selection is allowed only if they all correspond to same patient id A maximum of 15 SPS s can be selected Referenced Study Sequence a maximum of ten Item s is supported This attribute will be present only if SPS information is available from Modality Work list SCU At the end of acquisition the user might choose to Defer PPS and later ch
119. d Procedure Description Presentation Description 0070 0081 Exam Split Application will set as Exam Split Presentation Create Date 0070 0082 Exam Split Application will calculate Presentation Create Time 0070 0083 Exam Split Application will calculate Presentation creators name 0070 0084 Exam Split Application will set this value as null Referenced Series Sequence 0008 1115 Copied from source image Displayed area selection sequence 0070 005A Exam Split Application will calculate values for Window Width Window Level pan and zoom 120 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT eee eee Softcopy VOI LUT Sequence 0028 3110 Exam Split Application will calculate and copy the from acquired Image s Image Horizontal flip 0070 0041 Exam Split Application will calculate from Image Orientation Image Rotation 0070 0042 Exam Split Application will calculate from Image Orientation Presentation LUT Shape 2050 0020 Exam Split Application will set as IDENTITY SOP Class UID 0008 0016 Exam Application will set as 1 2 840 10008 5 1 4 1 1 11 1 SOP Instance UID 0008 0018 Copied from acquired study Specific Character Set 0008 0005 Copied from acquired study Referenced Patie
120. d and Cornell as well as others in the industry What was done in Generation 1 DICOM MOD encoding by GEMS In the JPEG codec output the JPEG SOS header defines the Huffman table selector codes to find the appropriate Huffman table FF DA 00 08 01 00 11 01 00 00 The GEMS DICOM Access Kit coded these as 0x11 They should have been 0x00 Zero was the number assigned in the DHT header This error is a primary key that is used in the GEMS Generation 2 system to determine that the JPEG data was encoded using the GEMS Generation 1 coding technique In Generation 1 images the predictor value used at the beginning of each line is the last pixel value from the previous line in the image Assumes Left to Right Top to Bottom scanning of the pixels The Generation 2 and industry compliant predictor for the first line element of each line should be the first element of the line above the current line For the first line first element predictor should be the unsigned value that is half the full scale range for the sample precision GEMS uses 1 lt lt 15 32768 The GEMS Generation 1 JPEG computation of the difference to be Huffman encoded was computed as Predictor value when it should have been calculated as value Predictor The result is that the decompression via industry standard packages results in a negative of the 1 No specific direct confirmation of the value Predictor policy was found but the graphic on page 185 o
121. d their affects on worklist query parameters are given below This Scanner Modality 0008 0060 set to CT Scheduled Station AE Title 0040 0001 set to local AE title This Modality Modality 0008 0060 set to CT Scheduled Station AE Title 0040 0001 zero length universal matching All Scanners Modality 0008 0060 zero length universal matching Scheduled Station AE Title 0040 0001 zero length universal matching The scheduled dates of procedures of interest can be specified for query by selecting a specific date range The date ranges available are Today Days Before Today Days After Today and All Days These selections and their affects on worklist query parameters are given below Today Scheduled Procedure Step Start Date 0040 0002 set to YYYYMMDD where this date is the current date Days Before Today Scheduled Procedure Step Start Date 0040 0002 set to and Days After Today YYYYMMDD YYYYMMDD where this date range represents the specified number of days before today and or after today Note that number of days both before and after can be specified in the same query and that each always includes today All Days Scheduled Procedure Step Start Date 0040 0002 zero length universal matching 78 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 3 1 2 1 2 Proposed Presentation Context Table The following table shows the proposed present
122. data sets is also being deployed to handle the small number of Multi Fragment images that have reached media 3 6 Configuration The MOD Archive Application may have either a 1 2GB or a 2 3GB drive installed Table 3 6 1 Supported Profiles for various Drive Configurations Disk Drive Installed Profiles Supported 1 2GB Optical Disk Drive PRI CTMR MOD12 fa ee ee STD CTMR MOD12 2 3GB Optical Disk Drive PRI CTMR MOD12 parte STD CTMR MOD12 pus STD CTMR MOD23 3 7 Support of Extended Character Sets The MOD Archive Application will support copy of SOP instances containing the ISO IR 100 Latin alphabet No 1 supplementary set and DICOM default character sets as defined in PS3 5 4 PRINT SCU CONFORMANCE 4 1 Introduction This conformance statement specifies the Discovery conformance to the DICOM Print SCU Application Profile The Discovery product uses DICOM Print SCU to print images on DICOM Compliant Printers 58 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 4 2 4 2 1 4 2 2 4 2 3 Implementation Model PRINT SCU The DICOM Print SCU is a DICOM print filter that provides the capability to print images to DICOM printers The DICOM Print filter acts as an SCU of the DICOM print management SOP class Application Data Flow Diagram DICOM print SCU Implementation model DICOM INTERFACE Film DICOM Composer Printer The Film Composer is the Use
123. e New generated may not be unique 107 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Attribute Name Tag Usage in MPPS Instance Acquisition Acquisition Post Processing with MWL data without MWL data have the the following following PPS ID exam PPS ID exam number number Performed Station AE Title 0040 0241 Local system Local system host Local system host host name name name Performed Station Name 0040 0242 Local system Local system suite id Local system host suite id name Performed Location 0040 0243 Local system Local system suite id Not used suite id Performed Procedure Step 0040 0244 Same as exam Same as exam date Same as exam date Start Date start date Performed Procedure Step 0040 0245 Same as exam Same as exam time Same as exam time Start Time start time Performed Procedure Step 0040 0254 Copied from Not sent as part of Not sent as part of Description SPS if selected image header image header Performed Procedure Step 0040 0252 See Note 1 See Note 1 See Note 1 status Performed Procedure Type Description 0040 0255 Not sent as part of image header Not sent as part of image header Not sent as part of image header Procedure Code Sequence 0008 1032 Not sent as part of image header Not sent in MPPS Not sent as part of image header Not sent in MPPS N Not sent
124. e Step creation and update 7 3 1 2 1 1 Associated Real World Activity The real world activities are mentioned in section 7 2 3 Sequencing of Real World Activities Each of the real world activity results in either creating a new Performed procedure Step SOP instance at the remote SCP or updating an already created Performed Procedure Step SOP instance as per the DICOM standard 101 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 7 3 1 2 1 2 Proposed Presentation Context Table The following table shows the proposed presentation contexts for the PPS Server AE after any of the real world activity listed in section 7 2 3 Sequencing of Real World Activities is initiated Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Modality 1 2 840 10008 3 1 2 3 3 Implicit VR 1 2 840 10008 1 2 SCU None Performed Little Endian Procedure Step 7 31 21 21 SOP Specific DICOM Conformance Statement for MPPS SOP class If the remote AE does not support the proposed Presentation context an appropriate error message logged Only one association is opened at a time All the operations used by this SOP class support an association timer which is configurable The timer is started when a request association request N CREATE request or N SET request is send and stopped when the respective response is
125. e e eae ere 40 2 5 EXTENSIONS SPECIALIZATIONS PRIVATIZATIONS ss0sccesccccecseseeaeceseeececseeeaeceeeeseeeeestsaeeeeess 40 2 5 1 Specialized Information Object Definition essen 40 2 5 2 Private Data o t eta ec eee D etel 41 2 6 A GONFIGUBATIONS costes sp etat neut i Ie e e A 41 2 6 1 AE Title Presentation Address Mapping sse 41 2 6 2 Configurable Parameters einen tnnt 41 2 7 SUPPORT OF EXTENDED CHARACTER SETS 1 00 1000 42 3 MEDIA STORAGE CONFORMANCE STATEMENT 222 422 11 43 3 1 INTRODUCTION 1 2 died 1I et di atte ee eee eae 43 3 2 IMPLEMENTATION MODEL MOD ARCHIVE SERVER nnne 43 3 2 1 Application Data Flow Diagram 43 3 2 2 Functional Definitions Of AE s 44 3 2 3 Sequencing of Real World sse 44 3 2 4 File Meta Information for Implementation Class and 44 3 33 AE SPEGIFICATIONS 3 nire iet et Mee cR ive Pee en dd ede Ye eene aa de 45 3 3 1 MOD Archive Specification 45 3 3 1 1 File Meta Information for the Application 45 3 9 12
126. e eee oca dep de E VUE UY eye essere 90 6 8 13 Patient 90 68134 Patient Relationshlp pie pte eee ee 90 6 9 13 2 Patlentildentificatio mi 55 aru eat eet redi etn ei reet tete 90 6 8 13 3 Patient 91 6 8 13 4 Medical alin beg e e 91 6 9 PRIVATE DATA DICTIONARY tic etn tbt eode ide e eese tad ue de 91 6 10 C FIND REQUEST MESSAGE 92 611 USE OF SPECIFIC DICOM nass 95 6 12 SETTING USER PREFERENCES 1 4 1 24 111 11 00 nasi theses sensn nnns entes assa na ases nnns 96 6 12 1 Setting Use MWL Study UID Option sss 96 6 12 2 Setting Custom Query 96 7 PERFORMED PROCEDURE STEP CONFORMANCE STATEMENT 97 7 1 INTRODUGTION rt 97 7 2 IMPEEMENTATION MQODBEL e ote sorte t ec 97 7 2 1 Application Data Flow Diagram sisse 000 nennen 97 7 2 2 Functional Definition of AES
127. e hospital or radiology department information system computer For example a user may wish to query for all procedures scheduled to be performed on the scanner In this situation ModalityWorklist is providing the DICOM C FIND service as a service class user SCU Implementation Model All DICOM functionality provided by the ModalityWorklist feature is logically provided by the Worklist Server DICOM AE The Worklist Server DICOM AE is commanded to perform DICOM modality worklist query services through the use of the Discovery user interface Application Data Flow Diagram The Basic and Specific Application models for this device are shown in the following illustration Association Initiation ConnectPro Worklist Remote Worklist SCP Modality Worklist SCU Server AE DICOM Standard Interface Functional definitions The ModalityWorklist Worklist Server AE is implemented as an application process on the scanner host computer It runs as a daemon serving requests from the user interface to obtain modality worklists query remote AE s and return the results to the user interface The ModalityWorklist Worklist Server AE initiates the following functions Query Initiates DICOM association in order to query a remote AE If the remote accepts a presentation context applicable to modality worklist the Worklist Server AE will issue a modality worklist query request via the C FIND service Sequenci
128. ed Presentation Contexts 0 cecceeecceceneeeeeeeceeeeeeeeseaeeeeaeeseaeeseeeeseaeeseaeeseaeeseaeeteeeeeaeete 72 5 3 1 5 Request PrOCOSSING otto to bte Deere de 72 5 3 1 6 ERIT CEREREM REM MER tT 73 5 3 1 6 1 Commit response with SUCCESS 73 5 3 1 6 2 Commit response with FAILURE 74 5 3 1 7 Association Acceptance 74 5 4 COMMUNICATION 1 74 5 4 1 Supported Communication Stacks parts 8 9 sse 74 5 4 2 TCP IP SIaCk iiia e dud TAE 74 5 4 2 1 Physical M dia SUppOrt ade teret te ee dece n dede 74 5 4 3 Pointto Polnt SIaCk ii ox t pa Dg RR 74 5 5 5 2 0 75 5 6 75 5 6 1 Title Presentation Address Mapping seen 75 5 6 2 Configurable Parameters 75 5 7 SUPPORT OF EXTENDED CHARACTER 6 5 75 6 MODALITY WORK LIST INFORMATION MODEL 2 24224411 75 6 1 INTRODUCTION 5 e e 75 6 2 IMPLEMENTATION MODEL 2 7
129. ed Procedure ID 0040 1001 Copied from User input on the Copied from source SPS if selected scanner image gt Requested Procedure 0032 1060 Copied from Not used Not sent as part of Description SPS if selected image header gt Placer Order 0040 1006 Not sent as part Not used Not sent as part of Number Procedure of image header image header gt Filler Order 0040 1007 Not sent as part Not used Not sent as part of Number Procedure of image header image header gt Scheduled Procedure ID 0040 0009 Copied from Not used Copied from source SPS if selected image gt Scheduled Procedure 0040 0007 Copied from Not used Copied from source Step Description SPS if selected image gt Scheduled Action Item 0040 0008 Copied from Not used Not sent as part of Code Sequence SPS if selected image header Patient s Name 0010 0010 Copied from User input on the Copied from source SPS if selected scanner image Patient ID 0010 0020 Copied from User input on the Copied from source SPS if selected scanner image Patient s Birth Date 0010 0030 Copied from User input on the Copied from source SPS if selected scanner image Patient s Sex 0010 0040 Copied from User input on the Copied from source SPS if selected scanner image Referenced Patient 0008 1120 Copied from Not used No item Sequence SPS if selected Performed Procedure Step ID 0040 0253 Created at the scanner Will Created at the scanner Will hav
130. ed Procedure Step Start and End Dates the attributes are sent with zero length In DICOM this is interpreted as meaning all dates i e universal matching The Modality is also not specified in this particular dump meaning all modalities Note that the user through the use of the Discovery user interface can submit a worklist query that will cause non zero values to be sent for these attributes Table 6 10 1 C FIND Request Message Dump 0008 0000 0008 0005 0008 0050 0008 0080 0008 0090 0008 1110 ffe e000 fffe e00d fffe e0dd 00008 1120 fffe e000 0008 0008 ffe e00d fe e0dd 10 0000 10 001 10 002 10 003 10 004 10 100 10 102 10 103 10 104 10 200 10 211 10 2154 10 2160 10 21b0 10 21c0 10 4000 0000 0020 000d 0032 0000 0032 1032 0032 1033 0032 1060 0032 1064 fffe e000 c re OS OO 0 0 0 Seta Ste cy COE Sle See ENS 0008 0100 0008 0102 0008 0104 ffe e00d ffe e0dd 0038 0000 0038 0010 0008 1150 0008 1155 1150 1155 UL CS SH LO PN SQ na UI UI na na 50 UI UI na na UL PN LO DA CS LO DS DS LO LO LO SH SH LT US LT UL UI UL PN LO LO SQ na na na UL LO SH SH LO 108 ISO IR 100 no value availab
131. el image Type wo 3 Manufacturer s Id 0008000 3 J Series Number 0020003 1 160 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT C 3 4 Image Keys Attribute Name Tag ______ msecmanwss somm 1 Sequence Variant teen 3 Slice Thiokness 3 _ 001800 3 O Number ofAverages 008083 8 o Spacing Between Sies 10018008 3 Oo Data Collection Diameter 0018009 3 ICowesiBousRowe 014104 3 FTriggerTime 0109 Reconstruction Diameter 0018100 3 FGantryiDetector Tit 3 O S S Convolution onea s Tosso 3 272270 a Image Orientation Patient e Frame of Reference UID 0020 0052 Absent from PRI CTMR MOD12 Seeiwwon 3 Samples Perera R FPhotometic inerpretaion 00280000 R Absent rom allprofies Ros 1 Cms Piel Spacing 0 00 G6 somma R Pets Stored 0026 0101 R MRI Absent rom proies Hgsmi 0028 0102 MRI Absent from profes 161 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT
132. ements is described in the document Introduction to the Integrated DICOM Network ID Net v3 0 Conformance Statements Direction 2118780 This introduction familiarizes the reader with DICOM terminology and general concepts It should be read prior to reading individual products ID Net v3 0 Conformance Statements The ID Net v3 0 Conformance Statement contained in this document also specifies the Lower Layer communications that it supports e g TCP IP However the Technical Specifications are defined in the DICOM Part 8 standard For more information including Network Architecture and basic DICOM concepts please refer to the ntroduction For the convenience of developers there is a collector Direction available By ordering the collector the Introduction described above and all of the currently published ID Net v3 0 Product Conformance Statements will be received The collector Direction is ID Net v3 0 Conformance Statements Direction 2117016 For more information regarding DICOM copies of the Standard may be obtained by written request by contacting ACR NEMA Representative NEMA 1300 N 17th Street Suite 1847 Rosslyn VA 22209 USA 1 3 Intended Audience The reader of this document is concerned with software design and or system integration issues It is assumed that the reader of this document is familiar with the DICOM standards and with the terminology and concepts that are used in those standards If readers a
133. ent s Birth Date 0010 0030 X X Copied from acquired image s Patient s Sex 0010 0040 Copied from acquired image s Instance Creation Date 0008 0012 X Copied from acquired image s Instance Creation Time 0008 0013 X Copied from acquired image s Study Instance UID 0020 000D Copied from acquired image s Study Date 0008 0020 X Copied from acquired image s Study Time 0008 0030 X Copied from acquired image s Accession Number 0008 0050 X Copied from acquired image s 117 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT eee Referring Physician s Name 0008 0090 Copied from acquired image s Study ID 0020 0010 Copied from acquired image s Referenced Study Sequence 0008 1110 Copied from acquired image s Modality 0008 0060 Copied from acquired image s Study Description 0008 1030 X X This will be set as Requested Procedure Code Meaning If Code Meaning absent it will be Requested Procedure Description Procedure Code Sequence 0008 1032 Copied from acquired image s Protocol Name 0018 1030 Set as Virtual Exam Split Series Description 0008 103E This will be set as Requested Procedure Code Meaning If Code Meaning absent it will be Requested Procedure Description Series Instance UID 0020 000E Exam Sp
134. ers that pertain to the remote AE e Remote AE HIS RIS IP address IP address used to contact the remote AE 81 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 6 2 e Remote AE HIS RIS IP port IP port used to contact the remote AE These parameters define where worklist queries will be directed Configuration of these parameters is performed by GEMS Field Service engineers using the ModalityWorklist installation facilities Configurable Parameters The following parameters are configurable for the DICOM Worklist Server AE e Local Worklist Server AE Title automatically set to host name of scanner e Local IP address e Local IP netmask e Local IP gateway The following parameters are configurable by changing their values in the configuration file usr g config WLdcm cfg Note that these parameters typically need not be changed Furthermore no support is provided for retaining changed settings the values will require changing again after a system software upgrade Implementation UID PDU size Association time out period Session time out period C FIND time out period 6 7 Support of Extended Character Sets ModalityWorklist will support only the ISO_IR 100 ISO 8859 1 1987 Latin alphabet N 1 supplementary set as extended character sets 6 8 Modality Worklist Information Model Definition 6 8 1 Introduction This section specifies the use of the DICOM Modality W
135. ession writes additional information may be subsequently added to the volume 48 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 4 3 Standard Application Profiles The MOD Archive Server supports the standard Application Profile CTMR MOD as an FSC FSR or FSU 3 4 3 1 STD CTMR MOD12 Class Profile 3 4 3 1 1 SOP Classes and Transfer Syntax s This class of Application Profiles is based on the Media Storage Service Class See PS3 4 SOP Classes and corresponding Transfer Syntax s supported by the STD CTMR MOD12 Application Profiles is also specified in the Table 3 4 3 1 1 1 Table 3 4 3 1 1 1 100 5 and Transfer Syntax s for CTMR MOD Application Profiles Information FSC FSR Object SOP Class UID Transfer Syntax and UID Req Req Definition Basic 1 2 840 10008 1 3 10 Explicit VR Little Endian Directory Uncompressed 1 2 840 10008 1 2 1 CT Image 1 2 840 10008 5 1 4 Explicit VR Big Endian Storage 1 1 2 Uncompressed 1 2 840 10008 1 2 2 CT Image 1 2 840 10008 5 1 4 JPEG Lossless Process Storage 1 1 2 Selection Value 14 1 2 840 10008 1 2 4 70 MR Image 1 2 840 10008 5 1 4 Explicit VR Big Endian Storage 1 1 4 Uncompressed 1 2 840 10008 1 2 2 MR Image 1 2 840 10008 5 1 4 JPEG Lossless Process Storage 1 1 4 Selection Value 14 1 2 840 10008 1 2 4 70 PET Image 1 2 840 10008 5 1 4 Explicit VR Big Endian Storage 1 1 128 Uncompressed 1 2 840 10008 1 2 2 PET Ima
136. ev 1 Discovery DICOM CONFORMANCE STATEMENT 7 10 Use of Specific DICOM Data The following table gives specific usage of some of attributes in the MPPS SOP instance created with reference to each of the real world scenarios mentioned in 7 2 3 of this document Attribute Name Tag Usage in MPPS Instance Acquisition with MWL data Acquisition without MWL data Post Processing Specific Character Set 0008 0005 Not used but copied into image header Not used Not used but copied into image header Scheduled Step Attribute Sequence 0040 0270 Only one item Only one item gt Study Instance UID 0020 000D Copied from SPS if option to copy is selected or else created at the scanner Created at the scanner Copied from source image gt Referenced Study 0008 1110 Copied from Not sent as part of Not sent as part of Sequence SPS if selected image header image header gt Accession Number 0008 0050 Copied from User input on the Copied from source SPS if selected scanner image gt Placer Order 0040 2016 Not sent as part Not sent as part of Not sent as part of Number Imaging Service of image header image header image header Request gt Filler Order 0040 2017 Not sent as part Not sent as part of Not sent as part of Number Imaging Service Request of image header image header image header gt Request
137. ev 1 Discovery DICOM CONFORMANCE STATEMENT Image Display Format 2010 0010 STANDARD C R Set in User Interface Printer Dependent Reference Film 2010 0500 Session Sequence i PORTRAIT Film Orientation 2010 0040 PORTRAIT Set in Configuration File Film Size ID 2010 0050 Sent zero length Sent zero length Magnification type 2010 0060 BILINEAR Set in Configuration File CUBIC REPLICATE NONE Density 2010 0130 0 4095 Set in Configuration File Configuration 2010 0150 Printer Dependent Set in Configuration File Information Smoothing type 2010 0080 Printer Dependent Set in Configuration File Border density 2010 0100 BLACK Set in Configuration File WHITE Empty image density 2010 0110 BLACK Set in Configuration File WHITE Min density 2010 0120 0 4094 Set in Configuration File Note Attributes sent zero length use the camera default values Note denotes that the attribute is optional for the SCU However we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value If Failure status is returned during N CREATE operation of Film box following action will be taken by Print SCU 0x213 Resource Limitation message will be sent 0x106 Unsupported Film Format message will be sent All
138. ever we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value If Failure status is returned during N SET operation of Image Box following action will be taken by Print SCU 0xC605 Resources temporarily not available message will be sent All other status Failure message will be sent In all the cases the print session will be terminated 4 3 1 2 2 4 Printer SOP Class N GET DIMSE service is supported for the Printer SOP Class If NLEVENT REPORT DIMSE service is received when the association is active Print SCU handles the relevant states but the data received is ignored 65 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Print SCU issues the request to retrieve the following attributes The Print SCU issues the N GET service to retrieve the printer status The status is processed as follows e f Printer status 2110 0010 is NORMAL the film is printed e f Printer status 2110 0010 is FAILURE the print job is terminated The Printer Status Info 2110 0020 attribute is not processed e f Printer status 2110 0010 is WARNING one of three things will happen a Ifthe Printer Status Info 2110 0020 is SUPPLY LOW the status is displayed to the user and the print job continues b If the Printer Sta
139. f the book JPEG still image data compression standard by Pennebaker shows the predictor entering the adder as a negative element and the value entering as positive element Like wise both Cornell and Stanford packages perform the computation as value Predictor 57 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT original image and a breakdown of the pixel basis when the last pixel on a line is different from the first pixel on the next line It was recently Jul 98 learned that the JPEG software distributed by the Independent JPEG Group does not include an important feature found in the GEMS the Stanford and the Cornell JPEG packages with respect to handling 16 bit deep images In the case where the huffDECODE would return a category 16 value indicating that 16 bits should be acquired from the bit stream the correct action is to add 32768 to the current value and to NOT get the number from the bit stream Without this test if 5 16 dx 32768 else the image will exhibit a spatial offset resulting from the incorrect synchronization of the input bit stream This version and all future GEMS Discovery software will read but not write the Installed Base Generation 1 media In addition the creation of Generation 2 media as has been initiated and the new format media will be Industry compatible JPEG and will NOT use the Multi Fragment segmentation The ability to handle Multi Segment
140. ge 1 2 840 10008 5 1 4 JPEG Lossless Process Storage 1 1 128 Selection Value 14 1 2 840 10008 1 2 4 70 GE Private 1 2 840 113619 4 30 Explicit VR Big Endian PET Raw Data Uncompressed Storage 1 2 840 10008 1 2 2 SC Image 1 2 840 10008 5 1 4 Explicit VR Big Endian Storage 1 1 7 Uncompressed 1 2 840 10008 1 2 2 SC Image 1 2 840 10008 5 1 4 JPEG Lossless Process Storage 1 1 7 Selection Value 14 1 2 840 10008 1 2 4 70 49 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Information FSC FSR FSU Definition Standalone 1 2 840 10008 5 1 4 Explicit VR Big Endian Overlay 1 1 8 Uncompressed 1 2 840 10008 1 2 2 Standalone 1 2 840 10008 5 1 4 JPEG Lossless Process Overlay 1 1 8 Selection Value 14 1 2 840 10008 1 2 4 70 M Mandatory O Optional Note The MOD Archive Application chooses the JPEG Lossless compression selection value 1 as the default transfer syntax for storing images on the media But the transfer syntax is selectable on a per media basis at install time using a method defined by the Application 3 4 3 1 2 Physical Media and Media Formats The CT MR PET Application Profiles in the STD CTMR MOD12 require the 130mm 1 2GB 512 Byte per Sector Magneto Optical Rewriteable physical media with the PC File System as defined in PS3 12 3 4 3 1 3 Logical Format The STD CTMR MOD12 Application profile media format conforms to the Logical Format specification in PS3 12
141. ion 0018 0026 3 SQ 1 Not Used Sous gt Intervention Drug Name Not Used gt Intervention Drug Code Not Used Sequence gt gt Code Value 0008 0100 1C SH Not Used gt gt Coding Scheme Designator 00080102 1C SH 1 Not Used gt gt Code Meaning 0080004 3 LO 1 Not Used Intervention Drug Start Time 00180035 3 TM 1 Not Used Intervention Drug Stop Time 0018 0027 ae Not Used gt Intervention Drug Dose 0018 0028 3 Not Used A 4 9 PET Multi gated Acquisition Module Attribute Name Tag Type VR VM Notes Beat Rejection Flag 0018 1080 2 CS 1 IF trig_rej_method none 137 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT THEN ELSE Y Trigger Source or Type Not Used Heart Rate 0018 1088 3 Not Used 3 LO 1 Framing Type 0018 1064 4 10 NM PET Patient Orientation Module Attribute Name Tag Type VR Notes Patient Orientation Code 0054 0410 2 SQ 1 Zero length Sequence sequence gt Code Value 0008 0100 gt Coding Scheme Designator 0008 0102 gt Code Meaning 0008 0104 gt Patient Orientation Modifier 0054 0412 2C SQ 1 Code Sequence gt gt Code Value 0008 0100 gt gt Coding Scheme Designator 0008 0102 gt gt Code Meaning 0008 0104 JH Patient Gantry Relationship 0054 0414 2 SQ 1 Zero
142. le no value available no value available Sequence with explicit Length Item with explicit Length ue available ue available imitationItem for SequenceDelimitationItem Sequence with explicit Length 4 1 Item with explicit Length no val no val ItemDel 4 1 IdentifyingGroupLength 12 1 SpecificCharacterSet 0 O AccessionNumber 0 0 InstitutionName 0 0 ReferringPhysicianName 1 24 1 ReferencedStudySequence 2 16 1 Item 0 0 ReferencedSOPClassUID 0 0 ReferencedSOPInstanceUID re encoding 0 1 ItemDelimitationItem for re enc 1 SequenceDelimitationItem 24 1 ReferencedPatientSequence 2 16 1 Item value available 0 0 ReferencedSOPClassUID value available 0 0 ReferencedSOPInstanceUID ItemDelimitationItem for re encoding 0 1 ItemDelimitationItem SequenceDelimitationItem for re enc 0 1 SequenceDelimitationItem 96 4 1 PatientGroupLength no value available 0 0 PatientName value available 0 0 PatientID value available 0 0 PatientBirthDate value available 0 0 PatientSex no value available 0 0 Other Patient IDs no value available 0 0 PatientSize no value available 0 0 PatientWeight value available 0 0 PatientAddress value available 0 0 MedicalAlerts no value available 0 0 ContrastAllergies no value available 0 0 PatientTele
143. le or re configurable by the Field Service Engineer and are not accessible by users through the Discovery s user interface 115 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 8 5 1 Title Presentation address Mapping Exam Split allows for the configuration of the following parameters that pertain to the remote AE e Remote GSPS AE IP Address IP Address used to contact the remote GSPS AE e Remote GSPS Port IP Port used to contact the remote GSPS AE e Remote GSPS AE Title Application Entity Title used to contact the remote GSPS AE e Remote HES AE IP Address IP Address used to contact the remote HES AE e Remote HES IP Port IP Port used to contact the remote HES AE e Remote HES AE Title Entity Title used to contact the remote 8 5 2 Configurable Parameters The Following are configurable for the GSPS Remote Host IP address Remote Host port Remote AE Title VES_ORIENTATION environment variable for enable disable Flip Rotate functionality in VES e Changesplit_mode lt HES VES gt 8 6 SUPPORT OF EXTENDED CHARACTER SETS GSPS feature for Discovery does not support any extended character set 87 ERROR HANDLING AND RECOVERY Exam Application has an extended error codes and the standard error codes are also handled If the sending GSPS PPS objects request has failed or response is not received before the association timeout the operation is deemed Failed and the
144. left to the user s discretion whether the local sop instances with their archive flag set to Y are to be deleted N Event Response will be sent on the same association itself No DATA Set will be sent along with the response 5 3 1 7 Association Acceptance Policy A single association will be accepted at any time to receive Storage Commitment responses 5 4 Communication Profiles 5 4 1 Supported Communication Stacks parts 8 9 DICOM Upper Layer Part 8 is supported using TCP IP 5 4 2 TCP IP Stack The TCP IP stack is inherited from a UNIX Operating System 5 4 2 1 Physical Media Support Ethernet v2 0 IEEE 802 3 5 4 3 Point to Point Stack A 50 pin ACR NEMA connection is not applicable to this product 74 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 5 5 EXTENSIONS SPECIALIZATIONS PRIVATIZATIONS None 5 6 CONFIGURATION 5 6 1 5 6 2 Ethernet v2 0 IEEE 802 3 AE Title Presentation Address Mapping The system allows the user to add Remove or Update the mapping of remote AE Titles to IP Addresses and Ports These options can be selected from the Remote Host Selection menu displayed by choosing Select Remote Host from the Network pull down menu from the local database manager A remote DICOM host can be designated as a remote archive device by setting the Remote archive node radio button on the remote host configuration
145. lit Application will calculate Series Number 0020 0011 Copied from acquired image s Laterality 0020 0060 Set as 0 zero Requested Attribute Sequence 0040 0275 Copied from source image Originally from Request Attributes Sequence gt Accession Number 0008 0050 gt Referenced Study Sequence 0008 1110 gt Study Instance UID 0020 000D gt Requested Procedure Description 0032 1060 gt Requested Procedure Code Sequence 0032 1064 gt Scheduled Procedure Step Description 0040 0007 gt Scheduled Protocol Code Sequence 0040 0008 118 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT eee E A i i e AA gt Scheduled Procedure ID 0040 0009 gt Requested Procedure ID 0040 1001 Performed Station AE Title 0040 0241 Exam Application will set as Host name Performed Station Name 0040 0242 Exam Application will set as Host name Performed Location 0040 0243 Exam Application will set as Host name Performed Procedure Step 0040 0244 X Exam Split Start Date Application will calculate and same time will use in GSPS Performed Procedure Step 0040 0245 X Exam Split Start Time Application will calculate and same time will use in GSPS Performed Procedure Step 0040 0250 Exam
146. lter order PET bp center PET bp center p PET atten smooth PET atten smooth param PET angle smooth param PET wellcountercal id PET trans scan id PET norm cal id PET blink cal id PET cac edge threshold PET cac skull offset PET emiss sub id PET radial filter 3d PET radial cutoff 3d PET axial filter 3d PET axial cutoff 3d PET axial start PET axial spacing PET axial angles used 3 PET ir num iterations 3 PET ir num subsets 3 156 of 169 j j j j j j j j j GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT PET ir_recon_fov PET ir_corr_model PET ir_loop_filter PET ir pre filt parm FL SL SL FL SL FL SL EL SL FL SL PET ir loop filt parm PET response filt parm PET post filter PET post filt parm PET ir regularize 0009 10BC PET regularize parm PET ac bp filter PET ac bp filt cut off PET ac bp filt order PET ac img smooth PET ac img smooth parm r SL SL FL SL SL FL LO FL ST PET scatter method PET scatter num iter PET scatter parm PET ctac conv scale PET loop filter parm Recon Protocol Co B 2 7 Discovery PET Image Module Private Creator Identification GEMS PETD 01 Attribute Name Tag
147. ly displayed on Not available the More Info screen Special Needs 0038 0050 Special Needs only displayed on Not available the More Info screen Requested Contrast Agent Requested Contrast Agent only Not available 0032 1070 displayed on the More Info screen Current Patient Location Current Patient Location only Not available 0038 0300 displayed on the More Info screen Additional Patient History Additional Patient History only Patient History 0010 21 0 displayed on the More Info screen 95 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Note that the display of a specific data item on the More Info screen is contingent on the item being enabled for display Depending on the preferences of each specific site data can either be displayed or not A GE field service engineer can assist in setting these site preferences 6 12 Setting User Preferences 6 12 1 Setting Use MWL Study UID Option Setting this option to Yes copies the Study Instance UID from work list into the final DICOM Image header If this option is set to No then a new study instance uid will be generated locally 1 Click on Patient Schedule 2 Click on Preferences button 3 Set the option Use Study Instance UID to either Yes No 6 12 2 Setting Custom Query Option This option allows the user to enter val
148. mote 27 2 3 1 2 1 1 Associated Real World 27 2 3 1 2 1 2 Proposed Presentation Contexts ecceeecceeeceeeseeeeeeeeseeeeeeeeeseeeeeeeeeeeneeeseneeeseeeeeaes 27 2 3 1 2 2 Query Remote ABs ao nete Huit Eu a RU 29 2 3 1 2 2 1 Associated Real World 29 2 3 1 2 2 2 Proposed Presentation Contexts essssssssseeeeneeneneenenen nennen 29 2 3 1 2 3 Get Image s from Remote 244 24 204440 0 1 100 00000 nnne enne nnns 31 2 3 1 2 3 1 Associated Real World een 31 2 3 1 2 3 2 Proposed Presentation Contexts 32 2 3 1 2 4 Real World Activity Verify DICOM Ping 32 2 3 1 2 4 1 Associated Real World 2 32 2 3 1 2 4 2 Proposed Presentation Context 2 0200242 400 0 32 2 3 1 2 5 Real World Activity Storage 32 2 3 1 2 5 1 Associated Real World enn 32 2 3 1 2 5 2 Proposed Presentation Context Table 33 2 3 1 2 5 3 SOP Specific Conformance Statement Storage Commitment SCU 33 2 3 1 3 Association Acceptance 33 2 3 1 3 1 Receive Image s ree ee e e PAE r EAEri SEARRE A E 34 2 3 1 3 1 1 Associated Real World
149. mote DICOM AE and send back the list of matches The DICOM Server AE will also respond to an incoming retrieval request from a Remote AE by sending the image s to the Destination AE Functional Definition of AE s DICOM Server Application Entity initiates the following operations e Ping DICOM Host Verification The DICOM Server initiates an association and sends a C ECHO RQ message to the remote DICOM AE the remote DICOM Server will send back a C ECHO RSP message with a status of success Push Initiate an association to a Remote to send image s If the Remote AE accepts the presentation context applicable to the image s being sent the DICOM Server AE will send the image s by invoking C STORE RQ operation for each image on the same association Query Initiate an association with a Remote AE to query for images on the remote host A Study Root Study Level C FIND RQ request will be sent to the Remote AE once an association has been established After all responses are received DICOM Server AE will issue a Series Level C FIND RQ request to get the series for a study in the list An Image Level C FIND RQ will be issued for the first series in the series list 24 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Get Send a C MOVE RQ request to a Remote AE after successful association establishment The DICOM Server AE s Storage SCP will receive the images over a separate a
150. mplementation UID 1 2 840 113619 6 55 Discovery PET Implementation UID 1 2 840 113619 6 131 61 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 4 3 1 2 Association Establishment Policy Print SCU initiates the association with the Printer 4 31 21 Real World Activity 4 3 1 2 1 1 Associated Real World Activity Print The Film Composer allows the user to select printers and it also allows the user to drag and drop the images from the viewer application into the film It also allows the user to manipulate some print parameters like film format and number of copies to print When the user presses the Print Button the Film composer communicates this request to Print SCU which then tries to establish the association with the requested printer and transmits the images for printing 4 3 1 2 1 2 Proposed Presentation Contexts The Proposed Presentation Context Table for the Print SCU is as shown in following Table Presentation Context Table Proposed Abstract Syntax Transfer Syntax Extended Negotiation Basic Grayscale 1 2 840 10008 5 1 1 9 DICOM Implict 1 2 840 10008 1 2 SCU None Print Management VR Little Endian Meta SOP Class Transfer Syntax Verification SOP 1 2 840 10008 1 1 DICOM Implict 1 2 840 10008 1 2 SCU Class VR Little Endian Transfer Syntax Printer SOP Class 1 2 840 10008 5 1 1 16 DICOM Implict 1 2 840 10008 1 2 SCU VR Little Endian Transfer Syntax Print Job SO
151. nce gt Radionuclide Code Sequence 0300 Single item sequence gt gt Code Value 0100 Code value is based on radionuclide_name gt gt Coding Scheme Designator 0008 0102 1 SH 99SDM gt gt Code Meaning 0008 0104 radionuclide_name gt Radiopharmaceutical Route 0018 1070 Not Used gt Administration Route Code 0054 0302 3 SQ 1 Not Used Sequence gt gt Code Value 0008 0100 1C SH Not Used gt gt Coding Scheme Designator 00080102 1C SH Not Used gt gt Code Meaning RU EIRAN Not Used Radiopharmaceutical Volume 00181071 3 DS 1 gt Radiopharmaceutical Start 136 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Time gt Radiopharmaceutical Stop 0018 1073 Time gt Radionuclide Total Dose 0018 1074 gt Radionuclide Half Life 0018 1075 gt Radionuclide Positron Fraction 0018 1076 gt Radiopharmaceutical Specific 0018 1077 3 DS 1 Activity gt Radiopharmaceutical 0018 0031 gt Radiopharmaceutical Code 0054 0304 3 SQ 1 Sequence gt gt Code Value 0008 0100 Not Used tracer_activity post_inj_activity half_life meas datetime admin datetime post inj datetime Not Used tracer name Single item sequence Value is based on tracer name Coding Scheme Designator 99SDM gt gt Code Meaning tracer_name Intervention Drug Informat
152. nce Policy The Worklist Server AE does not respond to attempts by a remote AE to open an association 80 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 4 Communication Profiles 6 4 4 Supported Communication Stacks PS 3 8 PS 3 9 DICOM Upper Layer PS 3 8 is supported using TCP IP 6 4 2 OSI Stack The OSI Communication Stack is not supported by this implementation 6 4 3 TCP IP Stack The TCP IP Communication Stack is inherited from the operating system 6 4 3 1 API Not applicable to this product 6 4 3 2 Physical Media Support Ethernet 802 3 provides the physical network layer for this product 6 4 4 Point to Point Stack The Point to Point Communication Stack is not supported by this implementation 6 5 Extensions Specializations Privatizations 6 5 1 Standard Extended Specialized Private SOPs ModalityWorklist for Discovery does not implement any private transfer SOP classes 6 5 2 Private Transfer Syntaxes ModalityWorklist for Discovery does not implement any private transfer syntaxes 6 6 CONFIGURATION The ModalityWorklist feature is configured by GEMS Field Service engineers The DICOM configuration items below are configurable or re configurable by a Field Service Engineer and are not accessible by users through the Discovery user interface 6 6 1 AE Title Presentation Address Mapping ModalityWorklist allows for the configuration of the following paramet
153. nce that is already created with the remote AE If the remote AE accepts a presentation context applicable to Modality performed Procedure Step the PPS Server AE will issue a request to update the SOP instance in the remote AE via the N SET service The PPS Status is set to COMPLETED Discontinue PPS Initiates a DICOM association in order to update a DICOM Modality Performed Step instance that is already created with the remote AE If the remote AE accepts a presentation context applicable to Modality performed Procedure Step the PPS Server AE will issue a request to update the SOP instance in the remote AE via the N SET service The PPS Status is set to DISCONTINUED 7 2 3 Sequencing of Real World Activities 7 2 3 1 PPS from Acquisition System with MWL data The system has a Modality Work list Server AE installed Work List information is obtained from HIS RIS system through the use of Basic Work list Management Service Use of the information retrieved in the creation of Image SOP instance is described in the Modality Work list Conformance statement Use of the information retrieved in MPPS SOP instances is described later in this document The system initiates a Start PPS before starting a scan i e when the image acquisition is started The system retrieves necessary information related to the Scheduled Procedure Step from Modality Work list Server PPS Server AE initiates a MPPS Modality Performed Procedure Step N CRE
154. ng of Real World Activities 1 The user or the system initiates a modality worklist query as a modality worklist SCU to the modality worklist SCP with a given set of query parameters 76 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 The modality worklist SCP returns responses that match the query parameters Items from the returned worklist responses are presented to the user 4 A subset of the returned worklist responses is included in acquired DICOM images related to the responses 6 3 AESpecifications 6 3 1 Worklist Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU SOP Class Name SOP Class UID Modality Worklist Information Model FIND 1 2 840 10008 5 1 4 31 6 3 1 1 Association Establishment Policies 6 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for an association initiated by the DICOM Worklist Server is Maximum Length PDU 50 Kbytes The SOP Class Extended Negotiation is not supported The maximum number of Presentation Context Items that will be proposed is 1 The user information Items sent by this product are Maximum PDU Length e Implementation UID 6 3 1 1 2 Number of Associations
155. nt Sequence 0008 1120 Scheduled Step Attribute Sequence 0040 0270 Copied from source image Originally from SPS sequence gt Accession Number 0008 0050 gt Referenced Study Sequence 0008 1110 gt Study Instance UID 0020 000D gt Requested Procedure Description 0032 1060 gt Scheduled Procedure Step Description 0040 0007 gt Scheduled Action Item Code Sequence 0040 0008 gt Scheduled Procedure ID 0040 0009 gt Requested Procedure ID 0040 1001 121 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 8 9 Image Header Changes Supporting GSPS 8 9 1 Request Attributes Sequence The 0040 0275 Request Attributes Sequence has been expanded to include dicom tags that the VES application requires to complete the VES GSPS and PPS This sequences now contains the following elements 0040 0275 Request Attributes Sequence gt 0008 0050 Accession Number gt 0008 1110 Reference Study Sequence gt gt 0008 1150 Referenced SOP Class UID gt gt 0008 1155 Referenced SOP Instance UID gt 0020 000D Study Instance UID gt 0032 1060 Requested Procedure Description gt 0032 1064 Requested Procedure Code Sequence gt gt 0008 0100 Code Value gt gt 0008 0102 Coding Scheme Designator gt gt 0008 0104 Code Meaning gt 0040 0007 Scheduled Pr
156. ocedure Description gt 0040 0008 Scheduled Action Item Code Sequence gt gt 0008 0100 Code Value gt gt 0008 0102 Coding Scheme Designator gt gt 0008 0104 Code Meaning gt 0040 0009 Scheduled Procedure Step ID gt 0040 1001 Requested Procedure ID The Requested Attributes Sequence may contain a maximum of 15 sequence items This sequence will only appear in the image header if the VES HES option is installed or PPS is enabled 8 9 2 Accession Number If scheduled procedures are grouped the accession number for each procedure is stored in the Request Attributes Sequence 0040 0275 in the image header If two or more of the accession numbers in the grouped case are different the top level Accession Number 0008 0050 shall contain 1 of the accession numbers 122 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT APPENDIX A OPTIONAL CONDITIONAL MODULES ATTRIBUTES IOD DEFINITIONS Enclosed is a listing of only the Optional Conditional modules attributes used by this implementation for CT SC Note All other modules that are mandatory and attributes that are type 1 amp 2 in nature per the DICOM standard are supported by this implementation but are not explicitly listed here CT Specific IOD Definition A 1 1 CT Image IOD Modules Entity Name Name R f rene usage Patent ______ Suy Study A42 o Ju ______
157. ocedureStepLocation no value available 0 0 PreMedication ItemDelimitationItem for re encoding 0 1 ItemDelimitationItem SequenceDelimitationItem for 0 1 SequenceDelimitationItem value available 0 0 RequestedProcedureID no value available 0 0 RequestedProcedurePriority value available 0 0 PatientTransportArrangements no value available 0 0 RequestedProcedureLocation value available 0 0 ConfidentialityCode value available 0 O Names of Intended Recipients no value available 0 O Requested Procedure Comments no value available 0 O Imaging Service Request Comm no value available 0 0 ConfidentialityConstraintOnPatientData If the query is for a particular date range the ScheduledProcedureStepStartDate will be filled with a valid date range If either the start or end date are left blank by the user they will simply be blank in the query Below is an example of a date range for August 30 1997 through October 12 1997 004 Below is an 0 0002 004 Below is an 0 0002 004 0 0002 DA 19970830 19971012 18 1 ScheduledProcedureStepStartDate example of a date range for August 30 1997 through the end of time DA 19970830 18 1 ScheduledProcedureStepStartDate example of a date range from the beginning of time through August 30 1997 DA 19970830 18 1 ScheduledProcedureStepStartDate If the query is for records for this mod
158. of View Shape 0018 1147 n Field of View Dimensions 0018 1149 Gantry Detector Tilt 0018 1120 3 Gantry Detector Slew 0018 1121 Type of Detector Motion 0054 0202 iii Collimator Type 0018 1181 2 Collimator Grid Name 0018 1180 EAR 135 of 169 IF Scan scan_type emission THEN TR 14 emiss_randoms trans_randoms NULL ELSE TR 14 trans randoms emiss_randoms NULL attenuation atten_coefficient atten_smooth pot 10 0 cm to m IF is type native PET images THEN CYLINDRICAL RING 10 0 Scan scan fov Scan axial fov IF is type native PET images GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Axial Acceptance 0054 1200 Axial Mash 0054 1201 Transverse Mash 0054 1202 1 Scan axial_acceptance acceptance_flag axial_angle_3d aM 2 Scan axial acceptance acceptance flag Detector Element Size 0054 1203 3 3 Not used Coincidence Window Width 0054 1210 DS 1 Scan upper coinc limit Scan lower coinc limit Energy Window Range 0054 0013 SQ 1 Sequence Energy Window Lower Limit 0054 0014 DS 1 Energy Window Upper Limit 0054 0015 DS 1 Secondary Counts Type 0054 1220 3 CS 1 n IF Scan delayed events separate THEN DLYD A 4 8 PET Isotope Module Attribute Name Tag Type VR Notes Radiopharmaceutical 0054 0016 2 SQ 1 Single item Information Sequence seque
159. of the ID Net v3 0 Conformance Statements and their relationship with the DICOM Conformance Statements is shown in Illustration 1 1 Illustration 1 1 ID Net v3 0 Introduction to the Integrated DICOM Network v3 0 ID Net v3 0 Conformance Statement Direction 5116427100 mw APPLICATION ENTITY SPECIFICATION SERVICE CLASSES INFORMATION OBJECTS MESSAGE EXCHANGES ETC MR A Lightspeed Product Conformance Statement Implementation Direction 5168904 100 Conformance Statement Direction DICOM STANDARD DICOM V 3 0 Standard Specification Part 1 DICOM V 3 0 Part 13 The Documentation structure given in Illustration 1 1 shows the overall documentation structure for all of the GE ID Net v3 0 Conformance Statements 18 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT This document specifies the DICOM implementation supported by the Discovery It is entitled Discovery Conformance Statement for DICOM 2007 ID Net v3 0 Direction 5176235 100 This Conformance Statement documents the DICOM Conformance Statement and Technical Specification required to interoperate with the GE ID Net v3 0 network interface Introductory information which is applicable to all GE ID Net v3 0 Conformance Stat
160. om SCP SOP Class UID 0008 1150 Value received from SCP SOP Instance UID 0008 1155 Value received from SCP N Event Response will be sent on the same association itself No DATA Set will be sent along with the response 73 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 5 3 1 6 2 Commit response with FAILURE status The following attributes are expected as part of DATA Set for N Event Request from SCP Attribute Tag Transaction UID 0008 1195 Value received from SCP Failed SOP Sequence 0008 1198 Value received from SCP SOP Class UID 0008 1150 Value received from SCP SOP Instance UID 0008 1155 Value received from SCP 0008 1197 Value received from SCP In case of complete partial failure the user will be notified about the status and the job entry will be paused There is no attempt made to retry automatically the failed sop instances However the user can manually retry the failed jobs Such requests will be treated as new requests This will go through the whole sequence of operations once again The failure reason is ignored Failed SOP instances will have their archive flag information unaltered Note The archive status flag in the browser is a shared flag with local archive When the status is Y it means that the images are archived but doesn t specify whether on local archive device or remote archive device It is
161. on or document Implementation Model The DICOM Performed Procedure Step service is provided by the PPS Server DICOM AE The PPS Server DICOM AE is commanded to perform Performed Procedure Step services either automatically or through the user interface Application Data Flow Diagram The basic Application models for the feature are shown in the following illustration Patient Scan with MWL data automatic Patient Scan without MWL data automatic PPS Server MPPS SCU MPPS SCP Post processing thr user interface DICOM Standard Interface 97 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 7 2 2 Functional Definition of AEs The PPS Server AE is implemented as an application process on the scanner host computer It runs as a daemon serving requests from other applications to send the PPS information to the remote AE and return the results to the requesting application The PPS Server AE initiates the following functions Start PPS Initiates DICOM association in order to create DICOM Modality Performed Procedure Step SOP instance in the remote AE If the remote AE accepts a presentation context applicable to Modality performed Procedure Step the PPS Server AE will issue a request to create the SOP instance in the remote AE via the N CREATE service Complete PPS Initiates a DICOM association in order to update a DICOM Modality Performed Step insta
162. ons Note that the user explicitly uses the user interface provided to invoke this operation as in the case of PPS generated for post processing which is described in the following section PPS messages N CREATE if applicable and N SET will be sent over the same DICOM association The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user 7 2 3 2 PPS from acquisition system without MWL data The system either does not have a Modality Work list Server AE installed or a Modality Work list Server AE installed but no Work List information is obtained from HIS RIS system for the current procedure that is being performed The information required for performing the procedure is supplied through the user interface of the system The use of this information in MPPS SOP instances is described later in this document The system initiates a Start PPS before starting a scan i e when the image acquisition is started PPS Server AE initiates a MPPS Modality Performed Procedure Step N CREATE request to the remote AE MPPS SCP in order to create a MPPS SOP instance at the remote AE The MPPS SCP returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user System includes the necessary info
163. ons Specializations and Privatization of SOP Classes and Transfer Syntax 3 5 1 Extensions Specialization s and Privatization s of SOP Classes The CT MR PET SOP Class Images have definitions extended for Defined Terms and include GE specific Private Data elements The following sections describe the details for these SOP classes 3 5 1 1 SOP Specific Conformance Statement for CT SOP Class Refer to section 2 4 for standard extensions and Appendix B for private data elements 3 5 1 2 SOP Specific Conformance Statement for MR SOP Class Refer to section 2 4 for standard extensions and Appendix B for private data elements 3 5 2 Private Transfer Syntax Specification The SOP instance files the Media File set in the PRI CTMR MOD12 profile uses a non standard JPEG compression process resulting in interoperability problems when decoded using readily available industry standard JPEG decompression routines GEMS PRI CTMR MOD12 DICOM MOD JPEG Encoding Information Note Some earlier versions of that were intended to write JPEG Lossless Process Selection Value 14 1 2 840 10008 1 2 4 70 actually contained errors in byte ordering of encapsulation of fragment item tags selection of entropy coding tables and calculation of difference values from predictors These errors cause the compressed images to be unreadable by non GEMS software All future GEMS Discovery software will read but not write archive media written with these errors
164. oose to Complete PPS or Discontinue PPS from the user interface provided in the system In this case the date and time when user chooses to Complete PPS or Discontinue PPS is taken as the Performed Procedure Step End Date and Performed Procedure Step End Time respectively Not the actual end date and end time of acquisition Mapping of SPS data to MPPS SOP instance is explained in section 7 10 Mapping of specific SPS data to DICOM IMAGE HEADER for PPS is explained in section 7 11 For the MPPS associated with a post processing the following restrictions apply on the attributes listed below Referenced Study Sequence The sequence is not sent in the MPPS message Scheduled Step Attribute Sequence a maximum of ONE item is supported The attribute will be send only if SPS information is available in the image instance Referenced Patient Sequence This sequence is not added Scheduled action Item Code Sequence The sequence is not send in the MPPS message Performed Procedure Step Start date amp Performed Procedure Step start time The exam date and exam time that is the Start date and Start Time of the associated Study Component Exam is used not the actual time when post processing started Performed Procedure Step end date amp Performed Procedure Step end time The date and time when user chooses to Complete PPS or Discontinue PPS is taken as the Performed Procedure Step End
165. operator action is required to respond to a verification message 2 31 3 21 Associated Real World Activity The Real World Activity associated with the verification request is to send a C ECHO response message with a status of success to the requesting AE 2 3 1 3 2 2 Presentation Context Table Table 2 2 1 3 2 2 1 Acceptable Presentation Contexts for DICOM Server AE and Real World Activity Verification Request Presentation Context Table 35 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Abstract Syntax Transfer Syntax Extended Negotiation fee UID Lit 1 2 840 10008 1 1 Implicit VR Little Endian 1 2 840 10008 1 2 2 3 1 3 2 2 1 SOP Specific Conformance to C_ECHO SCP The DICOM Server AE provides standard conformance to the DICOM Verification Service Class Each ECHO operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 900 seconds 15 seconds and 60 minutes respectively 2 3 1 3 3 Query Request from Remote AE This AE is indefinitely listening for associations No operator action is required to respond to a query request 2 3 1 3 3 1 Associated Real World Activity The Real World Activity associated with the query request is to search the local database for entries that match the request and send a C FIND response message with a status of pending for each matching entry 2 3 1 3 3 2 Present
166. orklist Information Model used to organize data and against which a Modality Worklist Query will be performed The contents of this section are 6 8 2 Information Model Description 6 8 3 Information Model Entity Relationship Model 6 8 4 Information Model Module Table 6 8 5 Information Model Keys 82 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 6 8 2 Modality Worklist Information Model Description In order to serve as a Service Class Provider SCP of the Modality Worklist Service Class a DICOM Application Entity AE possesses information about the attributes of a number of managed worklist items These items are organized into Modality Worklist Information Modules In this Service Class the Information Model plays a role similar to an Information Object Definition of most other DICOM Service Classes 6 8 3 Modality Worklist Information Model Entity Relationship Model The Entity Relationship diagram for the Modality Worklist Information Model schema is shown in Illustration 3 3 1 It represents the information that composes a Worklist Item In this figure the following diagrammatic convention is established to represent the information organization Each entity is represented by a rectangular box Each relationship is represented by a diamond shaped box The fact that a relationship exists between two entities is depicted by lines connecting the corresponding entity boxes
167. other status Failure message will be sent In all the cases the print session will be terminated If Failure status is returned during N ACTION operation of Film box following action will be taken by Print SCU 0xC602 Unable to Create Print Job message will be sent All other status Failure message will be sent In all the cases the print session will be terminated 64 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT If Failure status is returned during N DELETE operation of Film box following action will be taken by Print SCU All the return status Failure message will be sent and the print session will be terminated 4 3 1 2 2 3 Basic Grayscale Image Box SOP Class The Print SCU supports the following DIMSE Service Elements for Grayscale Image Box SOP Class N SET Requests the Printer to set the image box attributes The Following Attribute values are supported Attribute DICOM Tag Valid Range Default Value Image Position 2020 0010 Based on Image No Default Value Display Format Pre formatted Grayscale Image 2020 0110 Sequence Samples per pixel 0028 0002 Photometric Interpretation 0028 0004 MONOCHROME2 MONOCHROME2 0028 0010 Image Dependent Image Dependent Allocated 008000 e 8 Bissbed Xon s 8 0 unsigned integer 0 unsigned integer Note denotes that the attribute is optional for the SCU How
168. overy DICOM CONFORMANCE STATEMENT e Based upon the filming mode and the attributes used to acquire images into the film composer films will be automatically printed when the film session is full or the user must press the Print Button to print the images e The Print SCU retrieves the Print SCP status by using the N GET service of Printer SOP Class If the printer returns a FAILURE status the print session will be terminated The printer status is not read e The film session is created using the N CREATE service The print session will be terminated if the printer fails to create the film session e The film box is created using the N CREATE service The print session will be terminated if the printer fails to create the film box e The images are placed on the film box by using the N SET If the printer fails to place the image on the film box the print session will be terminated e The film will be printed using the N ACTION Only film box printing is supported If the printer fails to print the film the print session will be terminated e After the successful N ACTION the film box will be deleted using the N DELETE Any N EVENT REPORTS given to the PRINT SCU will be received but the data is ignored After the film box is deleted the association will be closed e Upon the successful completion of the above sequencing and the association is closed the user will be notified of the successful print session 60 of 169 GE HE
169. phoneNumber value available 0 0 Ethinc Group no value available 0 1 AdditionalPatientHistory value available 0 0 PregnancyStatus no value available 0 O Patient Comments 8 4 1 ImageGroupLength no value available 0 0 StudyInstanceUID 64 4 1 StudyGroupLength no value available 0 0 RequestingPhysician value available 0 0 RequestingService value available 0 0 RequestedProcedureDescription Sequence with explicit 1 32 1 RequestedProcedureCodeSequence Item with explicit Length 3 24 1 Item no value available u 0 CodeValue no value available 0 0 CodingSchemeDesignator no value available d 0 0 CodeMeaning ItemDelimitationItem for re encoding 0 1 ItemDelimitationItem SequenceDelimitationItem for 0 1 SequenceDelimitationItem 32 4 1 VisitGroupLength no value available 0 0 AdmissionID 92 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 i Discovery DICOM CONFORMANCE STATEMENT 0038 0050 0038 0300 0038 0500 0040 0000 0040 0100 ffe e000 0008 0060 0032 1070 0040 0001 0040 0002 0040 0003 0040 0006 0040 0007 0040 0008 ffe e000 LO LO LO UL 50 CS LO AE DA LO 50 0008 0100 0008 0102 0008 0104 ffe e00d fffe e0dd 0040 0009 0040 0010 0040 0011 0040 0012 ffe e00d fffe e0dd 00040 1001
170. post processing system The user initiates post processing on the images generated through acquisition The system creates a Modality performed Procedure Step instance locally in the system If the source image instance has the Scheduled Procedure Step information it is copied into the image instances created Also the system includes the necessary information related to the Modality Performed Procedure Step into the image instance At the end of one or more post processing the user initiates Complete PPS or Discontinue PPS through the user interface provided PPS Server AE initiates a MPPS Modality Performed Procedure Step N CREATE request to the remote AE MPPS SCP in order to create a MPPS SOP instance at the remote AE which is actually a replica of the locally created MPPS SOP instance The remote AE returns response indicating the success failure of the request execution If the response indicates success PPS Server AE initiates a MPPS N SET request to the remote AE in order to update the MPPS SOP instance that is already created with the additional information The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user 7 3 AE Specification 7 3 1 PPS Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU S
171. quencing of Real World Activities Real World Activity Query Remote must be performed before Choose Get Option can be performed 2 3 AE Specifications 2 31 DICOM Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU SOP Class Name SCU SOP Class UID Verification Echo 1 2 840 10008 1 1 CT Image Information Storage 1 2 840 10008 5 1 4 1 1 2 MR Image Information Storage 1 2 840 10008 5 1 4 1 1 4 Secondary Capture Image Storage 1 2 840 10008 5 1 4 1 1 7 Study Root Query Retrieve FIND 1 2 840 10008 5 1 4 1 2 2 1 Study Root Query Retrieve MOVE 1 2 840 10008 5 1 4 1 2 2 2 25 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Storage Commitment Push Model 1 2 840 10008 1 20 1 PET Image Information Storage 1 2 840 10008 5 1 4 1 1 128 GE Private PET Raw data Storage 1 2 840 113619 4 30 This Application Entity provides Standard Conformance to the following DICOM SOP classes as an SCP 1 2 840 10008 5 1 4 1 1 7 1 2 840 10008 5 1 4 1 1 128 2 3 1 1 Association Establishment Policy 2 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for association initiated by the DICOM Server AE is Maximum Length PDU
172. r interface and this is used to initiate the local real world activity The user issues the print request using Film Composer Film composer allows printer selection and it composes the pre formatted film file This film file is interpreted by Print SCU and it sends the appropriate messages to DICOM print SCP running on a DICOM printer Functional Definition of AE s DICOM Print SCU Establishes the Association to the requested printer for printing the composed film Sequencing of Real World Activities e The DICOM printer is installed through a camera installation process The DICOM printer is selected from the Film Composer Interface for Manual Filming Autofilming is enabled using the camera setup via the camera installation process e Images may be acquired for printing in the following manner Images to be printed be manually dropped from Viewing applications into the manual film composer slots Images to be printed be automatically dropped from Viewing applications into the manual film composer slots Images to be printed be manually dropped from Viewing applications into the auto film composer slots Image series to be printed be automatically dropped from the PrintSeries application into the manual film composer slots Images to be printed may be automatically dropped from the Scanning Application into the auto film composer slots 59 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Disc
173. r the RWA Create MOD For the list of Application Profiles that invoke this AE for the Create MOD RWA see Table 3 1 1 For extensions and specialization s see section 3 4 3 3 1 2 2 Real World Activity Display Directory The MOD Archive Application acts as an FSR when requested to provide a directory listing When an MOD update Application is requested to provide a directory listing it will read the File set and display the DICOMDIR directory entries for those SOP instances in the File set that correspond to the user selected Application Profile 3 31 2 21 Application Profiles for the RWA MOD Directory Listing For the list of Application Profiles that invoke this AE for the MOD directory listing RWA see Table 3 1 1 There are no extensions or specializations 3 3 1 2 2 2 Options None applicable 3 3 1 2 3 Real World Activity Copy to Local Storage The MOD Archive Application acts as an FSR when copying from the MOD to local storage The MOD Archive Application will copy any SOP Instance selected from an MOD Directory list from the MOD to the local storage upon request The MOD Directory listing Real World Application will filter out the SOP Instances that do not match the Application Profile 46 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 3 1 2 3 1 Application Profiles for the RWA Copy to Local Storage For the list of Application Profiles that invoke this AE for the MOD Cop
174. re unfamiliar with DICOM terminology they should first refer to the document listed below then read the DICOM Standard itself prior to reading this Conformance Statement document Introduction to the Integrated DICOM Network ID Net v3 0 Conformance Statements Direction 2118780 19 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 1 4 Scope and Field of Application It is the intent of this document in conjunction with the Introduction to the Integrated DICOM Network ID Net v3 0 Conformance Statements Direction 2118780 to provide an unambiguous specification for GE ID Net v3 0 implementations This specification called a Conformance Statement includes a DICOM Conformance Statement and is necessary to insure proper processing and interpretation of GE medical image data exchanged using DICOM The ID Net v3 0 Conformance Statements are available to the public The reader of this conformance statement should be aware that different GE devices are capable of using different Information Object Definitions For example a GE CT scanner may send images using the CT Information Object MR Information Object Secondary Capture Object etc Included in this Conformance Statement are Module Definitions that define all data elements used by the GE ID Net v3 0 implementation If the user encounters unspecified private data elements while parsing a GE Data Set the user is well advised to ignore those d
175. received The default time out value is 300 seconds All the operations used by this SOP class support a Session Timer This timer is started when the association is established and stopped when association is ended The default time out value is 3000 seconds If any of the above timers expires the association is aborted and the operation in progress is considered FAILED In any case an operation N CREATE or N SET fails system updates the state to enable operator to manually invoke the operation at any later time 7 3 1 3 Association Acceptance Policy The PPS Server AE does not respond to attempts by remote AE to open an association 7 4 COMMUNICATION PROFILES 7 4 4 Supported Communication Stacks PS 3 8 DICOM Upper Layer PS 3 8 is supported using TCP IP 7 4 2 OSI Stack The OSI Communication stack is not supported by this implementation 7 4 3 TCP IP Stack The TCP IP stack is inherited from the operating system 102 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 7 4 3 1 API Not Applicable to this product 7 4 3 2 Physical Media Support Ethernet 802 3 provides the physical network layer for this product 7 4 4 Point to Point Stack The Point to Point Stack is not supported by this implementation 7 5 EXTENSIONS SPECIALIZATION PRIVATIZATION 7 5 1 Standard Extended Specialized Private SOPs PPS for Discovery does not implement any private SOP classes 7 5 2 Priva
176. rmation related to Scheduled Procedure Steps and the Performed procedure Step in the image instances created At the end of image acquisition system initiates a Complete PPS or Discontinue PPS based on the choice selected by the user using the user interface provided The user is also given a choice Defer PPS which is described below PPS Server AE initiates a MPPS N SET request to the remote AE in order to update the MPPS SOP instance that is already created 99 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user At the end of image acquisition if the user has chosen Defer PPS the user is provided with an interface to Complete PPS or Discontinue PPS at any later time The user might wish to alter the image set generated through acquisition before invoking these operations Note that the user explicitly uses the user interface provided to invoke this operation as in the case of PPS generated for post processing which is described in the following section The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user 7 2 3 3 PPS from
177. rs acquired PET triggers rejected PET tracer name PET batch description PET tracer activity PET meas datetime PET pre inj volume PET admin datetime PET post inj activity PET post inj datetime PET radionuclide name PET half life 0009 102C 152 of 169 DiscovERY DICOM CONFORMANCE STATEMENT j j j j j j j j j j GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT PET positron_fraction PET source1_holder PET source activity PET source1 meas Gt PET source1_radnuclide PET source1 half life PET source2 holder PET source2 activity PET source2 meas Gt PET source2 radnuclide PET source2 half life PET source speed PET source location PET emission present 0009 104D PET lower axial acc PET upper axial acc PET lower coinc limit PET upper coinc limit PET coinc delay offset PET coinc output mode PET upper energy limit PET lower energy limit PET normal cal id PET normal 2d cal id PET blank cal id PET wc cal id PET derived PET contrast agent PET vqc x axis trans PET vqc x axis tilt PET vqc y axis trans PET vqc y axis swivel vqc 2 axis trans 3 153 of 169 FL SL FL DT SH FL SL FL DT SH FL SL FL SL SL SL SL SL SL SL SL SL UI UI UI UI SL LO FL FL FL FL FL x Ke x gt
178. s Time 0008 0031 1 Extract time from Scan scan_ datetime Counts Source 0054 1002 CS 1 Units 0054 1001 CS 1 Series Type 0054 1000 CS 2 Reprojection Method 0054 1004 CS 1 Number of R R Intervals 0054 0061 US 1 IF Scan scan_mode gated THEN Count the unique image image_bin_time for images with Image image_set_id thisSOPInstance ImageS etID ELSE NULL Number of Time Slots 0054 0071 1 US 1 IF Scan scan_mode gated THEN 1 ELSE NULL Number of Time Slices 0054 0101 1 US 1 Count the unique image image time for images with Image image set id thisSOPInstance ImageS etID 0054 0081 Number of Slices 0054 0081 1 US 1 Count the unique image image_location for images with Image image_set_id thisSOPInstance ImageS etID Corrected Image 0028 0051 1 n 134 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Randoms Correction Method 0054 1100 Attenuation Correction Method 0054 1101 Scatter Correction Method Sea Decay Correction Reconstruction Diameter 0018 1100 DS 1 Convolution Kernel 0018 1210 3 SH 1 n Reconstruction Method ose 1120 Detector Lines of Response 0054 1104 LO Used Acquisition Start Condition 0018 0073 Acquisition Start Condition Data 018 0074 Acquisition Termination 0018 0071 Condition Acquisition Termination 0018 0075 Condition Data Field
179. s and roles that are supported by this product in its version 1 0 21 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT The Discovery product uses DICOM services to import images for possible further analysis and or processing It also uses DICOM services to export images to other DICOM compliant machines Note the format of this section follows the format of the DICOM Standard Part 2 conformance Annex A hence the paragraph numbering scheme Please refer to that part of the standard while reading this section 2 2 Implementation Model All DICOM functionality on the Discovery product is handled by the DICOM Server Application Entity AE The DICOM Server AE is commanded to perform DICOM services through the buttons and menu selections on the main user interface panel The DICOM Server AE is also listening to a pre defined port for incoming connections 22 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 2 1 Application Data Flow Diagram ILLUSTRATION 2 0 IMPLEMENTATION MODEL DATA FLOW DIAGRAM Choose Push Option Image Stored locally DICOM SERVER AE Query Retrieve SCP Choose archive SAVE DICOM Standard Image Remotely Stored Query Get Request emote system requests image storage Retrieve from remote Q R SCP Verify Request 0 Image Remo
180. s been received or when subsequent C STORES are sent This time out is 300 seconds 28 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Each C STORE operation also supports a Session Timer This timer starts when the association is established and stops when the association is ended This time out is 60 minutes If any of the three timers mentioned above expires the connection is closed and the operation in progress is considered failed Note The time outs are configurable When DICOM Server AE initiates an association to issue a C STORE the following will occur 1 The image will be transmitted by the DICOM Server AE with the same elements as was originally received or created locally for the standard elements only 2 3 1 2 2 Query Remote AE 2 3 1 2 2 1 Associated Real World Activity The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname To do custom queries select Yes for Custom search option Note Custom queries can be done on the following fields Last Name contains Exam Number Patient Id Accession Number To initiate a Query select Receive from Network pull down menu Note If Custom Search option is set then a small GUI will show up Operator will be able to enter custom query fields Entering values in Last Name con
181. s from Exam exam datetime Patient patient birthdate Patient s Size 0010 1020 3 DS 1 Patient Height 100 0 Convert cm to m Patient s Weight 0010 1030 3 DS 1 Occupation 0010 2180 3 SH 1 Not used Additional Patient s History 0010 21B0 3 LT 1 A 4 5 Discovery PET Exam Module Refer to Section B 2 3 for details A 4 6 General Series Module Attribute Name Tag Type VR Notes Modality 0008 0060 1 1 Series Instance UID 0020 000E Series Number 0020 0011 Laterality 0020 0060 Not used Series Date 0008 0021 3 DA 1 Superceded by PET Series Series Date 0008 0021 Series Time 0008 0031 3 1 Superceded by PET Series Series Time 0008 0031 Performing Physician s Name Not used Protocol Name Not used Series Description Referenced Study Component 0008 1111 3 SQ 1 Not used Sequence gt Referenced SOP Class UID Not used gt Referenced SOP Instance UID Not used Body Part Examined Not used Patient Position Smallest Pixel Value in Series 0028 0108 Not used 133 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Attribute Name Tag Type VR Notes Largest Pixel Value in Series 0028 0109 05 55 1 Not used 4 7 Series Module Attribute Name Tag VR Notes Series Date 0008 0021 1 DA 1 Extract date from Scan scan_ datetime Serie
182. ssociation The DICOM Server AE waits for association requests from Remote AEs that wish to perform the following operations Verification C ECHO RQ message is received the DICOM Server AE will send back C ECHO RSP message with a status of success Image Storage If a C STORE RQ message is received the DICOM Server AE will receive the image and try to update the local database If the image is stored successfully on storage media and the database updated a status of success will be returned in a C STORE RSP message Query If a C FIND RQ message is received the DICOM Server AE will search the database for the requested attributes and send back a C FIND RSP message containing a match and a status of pending After all matching records have been sent a status of success will be returned in a C FIND RSP message The Remote AE can terminate the query by sending a C CANCEL FIND RQ message Retrieve a C MOVE RQ message is received the DICOM Server AE will lookup its list of configured Remote AEs for the Destination AE If the Destination AE is configured the DICOM Server AE will open a new association to the Destination AE and use C STORE RQ to send the image s The DICOM Server AE will send a C MOVE RSP message with a status of pending after every five images are sent When all images are sent or if DICOM Server AE receives C CANCEL MOVE RQ a final C STORE RSP will be sent back with an appropriate status 2 2 3 Se
183. sult in the loss of function and or connectivity as the DICOM Standard changes and GE products are enhanced to support these changes To be kept informed of the evolution of the implementation described in this document the user should register on the GE Internet server accessible via anonymous ftp by entering his her e mail address GE Internet Server Address ftp med ge com 192 88 230 11 e Interaction It is the sole responsibility of the non GE provider to ensure that communications with the interfaced equipment does not cause degradation of GE imaging equipment performance and or function 1 6 References A list of references that are applicable to all ID Net v3 0 Conformance Statements is included in the ntroduction to the Integrated DICOM Network ID Net v3 0 Conformance Statements Direction 2118780 1 7 Definitions A set of definitions applicable to all ID Net v3 0 Conformance Statements is included in the Introduction to the Integrated DICOM Network ID Net v3 0 Conformance Statements Direction 2118780 1 8 Symbols and Abbreviations A list of symbols and abbreviations that are applicable to all ID Net v3 0 Conformance Statements is included in the ntroduction to the Integrated DICOM Network ID Net v3 0 Conformance Statements Direction 2118780 2 SECTION 2 NETWORK CONFORMANCE STATEMENT 21 Introduction This Conformance Statement CS specifies the Discovery compliance to DICOM It details the DICOM Service Classe
184. t Referenced SOP Instance UID 0008 1155 1 UI 1 Not used Patient s Birth Time 0010 0032 3 TM 1 Not used Other Patient IDs 0010 1000 3 LO 1 n Not used Other Patient Names 0010 1001 3 PN 1 n Not used Ethnic Group 0010 2160 3 SH 1 Not used Patient Comments 0010 4000 3 LT 1 Not used A 4 3 General Study Module Attribute Name Tag Type VR VM Notes Study Instance UID 0020 000D 1 Ul 1 Study Date 0008 0020 2 DA 1 Extract date from exam_datetime Study Time 0008 0030 2 TM 1 Extract time from exam datetime Referring Physician s Name 0008 0090 2 PN 1 Study ID 0020 0010 2 SH 1 Accession Number 0008 0050 2 SH 1 Study Description 0008 1030 3 LO 1 Physician s of Record 0008 1048 3 PN 1 n Not used Name of Physician s Reading 0008 1060 3 PN 1 Diagnostician Study Referenced Study Sequence 0008 1110 3 SQ 1 Not used gt Referenced SOP Class UID 0008 1150 1 Ul 1 Not used gt Referenced SOP Instance UID 0008 1155 1 Ul 1 Not used A 4 4 Patient Study Module This section defines Attributes that provide information about the Patient at the time the Study was performed Attribute Name Tag Type VR VM Notes Admitting Diagnosis Description 0008 1080 LO 1 n Not Used 132 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Patient s Age 0010 1010 3 AS 1 Truncate year
185. t sent as part of image header 108 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT Attribute Name Tag Usage in MPPS Instance Acquisition with MWL data Acquisition without MWL data Post Processing from user input on the scanner gt Retrieve AE Title 0008 0054 Local system Local system host host name of the host name name system gt Referenced Image 0008 1140 One item for One item for each One item for each Sequence each image image created within image generated by created within the series the series post processing gt Referenced Standalone SOP Instance Sequence 0040 0220 Not sent as part of image header Not sent as part of image header Not sent as part of image header gt All other attributes from Performed Series Sequence which Table F 7 2 1 of DICOM standard PS3 4 does not explicitly list Not sent as part of image header Not sent as part of image header Not sent as part of image header All other attributes from Radiation Dose Module and Billing and Material Code Module which Table F 7 2 1 of DICOM standard PS3 4 does not explicitly list Not sent as part of image header Not sent as part of image header Not sent as part of image header Note 1 e value INPROGRESS e When PPS end N SET message is sent this element will have either COMPLETE
186. tains field will initiate wild card query for patient name The Query operation will cause the DICOM Server AE to initiate an association to the selected Remote AE when the Query Remote Host entry is selected from the Network pull down menu Once a list of Study Series Image is retrieved the operator can invoke the Get operation by choosing Get Exam or Get Series or Get Image from the Network pull down menu 2 3 1 2 2 2 Proposed Presentation Contexts When the Real World activity Query or is initiated all presentation contexts shown in table 2 2 1 2 1 2 1 are proposed during association establishment but only the Query Retrieve FIND related contexts are applicable to this activity 2 3 1 2 2 2 1 SOP Specific Conformance Statement for C FIND SCU After the Query operation is initiated the DICOM Server AE will perform a study root C FIND RQ request at each of the three levels Study Series and Image in succession The Initial Study Level request will ask for all studies in the Remote database The C FIND SCU will not perform any extended negotiation and so will only perform hierarchical query Relational Queries are not supported 29 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT amp Each C FIND SCU operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 300 seconds and 60 minutes
187. te stored and committed There are five Real World Activities that will cause the DICOM Server Application Entity DICOM Server AE to initiate a DICOM association to a remote DICOM Application Entity 1 The Choose Push Option Real World activity consists of an operator selecting one or more study series or image in the local database manager and choosing either Push Examination Push Series or Push Image from the Network pull down menu on the local database manager to send the image s to a selected destination 2 Real World Activity Query Remote causes the DICOM Server AE to initiate an association to the Remote DICOM AE and request the list of all studies Once the DICOM Server AE receives the list of studies it will select the first study as determined through the local database manager list sort criterion and request the list of series for that study After receiving the list of series the DICOM Server AE will ask for the list of images for the first series in the list The operator can then select any study in the study list to retrieve the list of series and images 8 Real World Activity Choose Get Option will be available once the Query Remote activity is performed The operator can now select one or more study series or image and ask the 23 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 2 2 2 DICOM Server AE to retrieve the selected image s from the
188. te Transfer Syntaxes PPS for Discovery does not implement any private Transfer Syntaxes 7 6 CONFIGURATION The PPS feature is configured by GEMS Field Service Engineers The DICOM configuration items below are configurable or re configurable by the Field Service Engineer and are not accessible by users through the Discovery s user interface 103 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 7 6 1 7 6 2 AE Title Presentation address Mapping PPS allows for the configuration of the following parameters that pertain to the remote AE e Remote MPPS AE IP Address IP Address used to contact the remote MPPS AE e Remote AE HIS RIS IP Port IP Port used to contact the remote MPPS AE e Remote MPPS AE Title Application Entity Title used to contact the remote MPPS AE These parameters define where the MPPS requests will be directed Configuration of these parameters is performed by GEMS Field Service engineers using the installation facilities Configurable Parameters The following are configurable for the PPS Server AE e Local PPS Server AE AE Title The default is host name of the scanner appended with string _PPS e Local IP Address e Local IP net mask e Local IP gateway e The following parameters are configurable but need change only in case of a system software upgrade e Implementation UID e PDU size e Association time out period e Session time out perio
189. the Application The SOP instance files in the Media File set in the PRI CTMR MOD12 profile uses a non standard JPEG compression process resulting in interoperability problems when decoded using readily available industry decompression routines See section 3 4 2 for details of the encoding problem unique to the PRI CTMR MOD12 profile 3 4 4 1 2 Physical Media and Media Formats The CT MR PT Application Profiles in the PRI CTMR MOD12 require the 130mm 1 268 512 Byte per Sector Magneto Optical Rewriteable physical media with the PC File System as defined in PS3 12 3 4 4 1 3 Logical Format The PRI CTMR MOD12 Application profile media format conforms to the Logical Format specification in PS3 12 The table 3 3 3 1 3 1 defines the specific values used for this profile 3 4 4 1 4 Directory Information in DICOMDIR Conformant Application Entities shall include in the DICOMDIR File a Basic Directory IOD containing Directory Records at the Patient and subsidiary levels appropriate to the SOP Classes in the File set All DICOM files in the File set incorporating SOP Instances defined for the specific Application Profile shall be referenced by the Directory Records All DICOM SOP Instance files in the File set are stored in the MSDOS filesystem subdirectory GEMS and are referenced from the DICOMDIR directory records Refer to Appendix C for a complete listing of all the optional modules and optional attributes used in the DICOMDIR definition Ple
190. tiate a new association due to a Ping DICOM Host or Push operation initiated by the user A new association is also initiated when the user performs a Query Remote operation or issues a retrieve operation by performing a Get operation at Study Series Image level 2 3 1 2 1 Push Image s to Remote AE 2 3 1 2 1 1 Associated Real World Activity The operator must first select a destination by choosing Select Remote Host from the Network pull down menu on the local database manager and then choose a hostname The Push operation will cause the DICOM server AE to initiate an Association when the operator selects one or more study series or images in the local database manager and then chooses either Push Examination Push Series or Push Image from the Network pull down menu on the local database manager Note If multiple study series or images are chosen to be pushed one association will be established for each of the studies series or images 2 3 1 2 1 2 Proposed Presentation Contexts The following table shows the proposed presentation contexts for the DICOM Server AE after Real World Activity Push Operation has been performed Table 2 2 1 2 1 2 1 Proposed Presentation Contexts for DICOM Server AE and Real World activities Push Query Pull Get Images and Verification Dicom Ping SCUs Presentation Context Table Proposal Abstract Syntax Transfer Syntax Extended Negotiation
191. tizations 2 5 1 Specialized Information Object Definition Following is a list of additional private attributes defined for a standard CT IOD 0008 0008 Value 3 CT Image IOD specific specializations AXIAL LOCALIZER SEGMENTED REFORMATTED PROCESSED COMBINED CTINTERVENTION Value 4 CT Image implementation specific MIN IP MIP AVERAGE VOLREN INTEGRAL HD MIP RAYSUM SURFACE MINMAX FLUORO 40 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 0018 0022 Scan Options Defined Terms AXIAL MODE SCOUT MODE AXIAL XRON MODE AXIAL XROFF MODE STATIC XRON MODE STATIC XROFF MODE TUBE HEAT MODE DAS MODE TUBE CAL MODE BIOPSY MODE CINE HELICAL ROTGENCAL MODE FLUORO MODE 2 5 2 Private Data Elements Refer to Appendix B for a complete listing of private data elements used with this implementation 2 6 Configuration 2 6 1 AE Title Presentation Address Mapping The Discovery system allows the user to add or Update the mapping of remote AE Titles to IP Addresses and Ports These options can be selected from the Remote Host Selection menu displayed by choosing Select Remote Host from the Network pull down menu from the local database manager 2 6 2 Configurable Parameters The following fields are configurable for the DICOM Server AE e Local AE Title the machine hostname e Local IP Address e Local IP Netmask e Max PDU length Time outs which are
192. tory Information B X 3 2 1 DICOM PS3 10 C 2 Directory Information Module Attribute Name 1 Offset of the Next Directory Record 0004 1400 Record Flag 0004 1410 Po Offset of Referenced Lower Level Directory 0004 1420 Entity Directory Record Type 0004 1430 PATIENT STUDY SERIES and IMAGE Referenced File ID 0004 1500 Present only in IMAGE Directory Record Referenced SOP Class UID in file 0004 1510 Present only in IMAGE Directory Record Referenced SOP Instance UID in File 0004 1511 Present only in IMAGE Directory Record Referenced Transfer Syntax UID in File 0004 1512 Present only in IMAGE Directory Record 159 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT C 3 Directory Record Selection Keys As indicated in table C 1 2 the PRI CTMR MOD12 application profile will have only the Patient Study Series and Image directory record types Given below are the list of attributes supported under each of these directories C 3 1 Patient Keys Attribute Name Tag Type Notes Specific Character Set 0008 0005 ISO IR 100 C 3 2 Study Keys Type Notes Suybwe ws 1 5 w08 1 Accession Number 009009 2 Study Description 09109 2 Study Instance 16 00200010 d C 3 3 Series Keys Attribute Name Tag Type Notes 00 Specific Character S
193. ttribute Name Tag Type VR VM Notes Samples per Pixel 0028 0002 1 US 1 Superceded by PET Image Samples Per Pixel 0x0028 0x0002 Photometric Interpretation 0028 0004 Superceded by PET Image Photometric Interpretation 0x0028 0x0004 Rows 0010 array_heig ht Columns 0028 0011 image_array_width Bits Allocated 0028 0100 superceded by 141 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT PET Image Bits Allocated 0x0028 0x01 00 Bits Stored 0028 0101 1 US 1 Superceded by PET Image Bits Stored 0x0028 0x0101 High Bit 0028 0102 1 US 1 Superceded by PET Image High Bit 0x0028 0x01 02 Pixel Representation 0028 0103 1 US 1 0001 2 s complement Pixel Data 7 0 0010 1 1 pixel data sent as OW Planar Configuration 0028 0006 Not Used Pixel Aspect Ratio 0028 0034 Not Used Smallest Image Pixel Value 0028 0106 US SS Not Used Largest Image Pixel Value 0028 0107 Not Used Red Palette Color Lookup Table 0028 1101 1C 3 Not Used Descriptor Green Palette Color Lookup 0028 1102 1C 3 Not Used Table Descriptor Blue Palette Color Lookup Table 0028 1103 1C 3 Not Used Descriptor Red Palette Color Lookup Table 0028 1201 1C 1 n Not Used Data Green Palette Color Lookup 0028 1202 1 1 Not Used Table Data Blue Palette Color Lookup Table 0028 1203 1 1 Not Used Data
194. tus Info 2110 0020 is RECEIVER FULL or SUPPLY EMPTY or FILM JAM the status is displayed to the user and the print job is aborted For all other Printer Status Info 2110 0020 values the status is ignored and the print job continues 4 3 1 2 2 5 Print Job SOP Class Print SCU looks for following attributes in N EVENT REPORT data received from Print SCP If Print SCU does not receive N EVENT REPORT it requests the Print SCP to retrieve the following set of attributes using N GET Kiribule Name Originator 66 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 4 3 1 3 Association Acceptance Policy The Print SCU does not accept associations 4 4 Communication Profiles 4 4 1 Supported Communication Stacks parts 8 9 DICOM Upper Layer Part 8 is supported using TCP IP 4 4 1 1 TCP IP Stack The TCP IP stack is inherited from the UNIX Operating System 4 4 1 1 1 API Not Applicable 4 4 1 1 2 Physical Media Support Ethernet v2 0 IEEE 802 3 4 5 Standard Extended Specialized Private SOPs None 4 6 Configuration 4 6 1 AE Title Presentation Address Mapping The Local AE title is hostname where hostname is the unique hostname within the network The Discovery system allows the user to add Remove or Update the mapping of remote DICOM Printer AE Titles to IP Addresses and Ports These options can be selected from the Service Desktop c
195. u selections on the main user interface panel Application Data Flow Diagram The MOD Archive Server creates and or updates the 5 25 inch 130mm MOD media with various DICOM SOP instances It can process CT MR PT Secondary Capture and Overlay IOD s Illustration 3 1 Specific AE Application Model Display Directory Create FSC D MOD DICOM MOD Archive Server AE Media Storage Local Storage Update MOD The MOD Archive Server AE has a local storage that may contain various SOP instances These may have been obtained by original creation network DICOM or proprietary or by 43 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 3 2 2 3 2 3 3 2 4 removable media using other application entities These instances are external to this conformance claim and the origin of SOP instances is outside the scope of this claim The MOD Archive Server AE can initialize Media by acting as an FSC to create a new DICOM File set on 2 3GB MOD media or a 1 2GB MOD media It initializes the DICOM File set and writes the specified SOP instances onto the MOD The SOP instances written will be limited to instances that match the criteria of one of the Application Profiles that is supported When updating media a pre existing File set will be updated with the selected SOP instances that match one of the supported Application Profiles Functional Definitions of AE s
196. ues for Accession Number and Requested Procedure Id which are used for Custom Query 1 Click on Patient Schedule 2 Click on Preferences button Set the option Show Update Parameters to Yes 4 To do a query click on Update button A User Interface appears with provision to enter values for a Accession Number b Requested Procedure Id 96 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 7 PERFORMED PROCEDURE STEP CONFORMANCE STATEMENT 7 1 7 2 7 2 1 Introduction This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance requirements for the Performed Procedure Step feature on this GEMS product Note that the format of this section strictly follows the format defined in DICOM Standard PS 3 2 Conformance Please refer to that part of the standard while reading this section The PPS option for Discovery allows a Modality Performed Procedure Step to be communicated to the Hospital Radiology information system The PPS feature is providing the DICOM Modality Performed Procedure Step service as a service class user SCU This feature works in conjunction with DICOM Modality Work list feature if installed However the conformance of this feature is independent of Modality work list feature For information on conformance of Modality Work List feature to DICOM standard please refer to the appropriate secti
197. veedor de servicios el operador o el paciente sufran lesiones provocadas por causas el ctricas mecanicas de otra naturaleza Este Manual de Assist ncia T cnica s se encontra dispon vel em Ingl s Se qualquer outro servi o de assist ncia t cnica quE nao a GEMS solicitar estes manuais noutro idioma da responsabilidade do cliente fornecer os servicos de traduc o Nao tente reparar equipamento sem ter consultado e compreendido este Manual de Assist ncia T cnica QOn ocumprimento deste aviso pode por em perigo a seguran a do t cnico operador ou paciente devido a choques el tricos mec nicos ou outros 4 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT AVVERTENZA l presente manuale di manutenzione disponibile soltanto in inglese addetto alla manutenzione esterno alla GEMS richiede il manuale in una lingua diversa il cliente tenuto a provvedere direttamente alla traduzione Si proceda alla manutenzione dell apparecchiatura solo dopo aver consultato il presente manuale ed averne compreso il contenuto Non tenere conto della presente avvertenza potrebbe far compiere operazioni da cui derivino lesioni all addetto alla manutenzione all utilizzatore ed al paziente per folgorazione elettrica per urti meccanici od altri rischi mk He LO3R UATE AT VILL 0 dS As G E MS DIITH E EHEH S nA RETRO REE BOR S NARA BERRI CONG OWE CITI BOL
198. vity associated with the Retrieve Request is to send all images corresponding to the C MOVE request to the destination AE through a separate association 38 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 DiscovERY DICOM CONFORMANCE STATEMENT 2 3 1 3 4 2 Presentation Context Table Table 2 3 1 3 4 2 1 Acceptable Presentation Contexts for DICOM Server AE and Real World Activity Retrieve Request Presentation Context Table Abstract Syntax Transfer Syntax Extended Negotiation cuc ta Study Root 1 2 840 10008 5 1 4 1 2 2 2 Implicit VR 1 2 840 10008 1 2 SCP None Query Retrieve Little Endian MOVE 2 3 1 3 4 21 SOP Specific Conformance to C MOVE SCP The DICOM Server AE provides standard conformance to the baseline Study root C MOVE Service Class SCP amp Each C MOVE SCP operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 300 seconds and 60 minutes respectively These time outs are configurable in dcs cfg as bi assoc tio bi move tio and bi session tio respectively All images requested in a C MOVE RQ will be sent over a single association A C MOVE RSP with a pending status will be returned to the requester every five images The C MOVE SCP will invoke C STORE requests for the following SOP classes SOP Class Name SOP Class UID CT Image Information Storage 1 2 840 10008 5 1 4 1 1 2 MR Image Information Storage 1 2 840 1000
199. w association for every GSPS operation initiated 113 of 169 GE HEALTHCARE DIRECTION 5176235 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT 8 2 4 2 1 Real World Activity GSPS 8 2 4 2 1 1 Associated Real World Activity The real world activities are mentioned in section 8 2 3 Sequencing of Real World Activities Each of the real world activity results in either creating a new GSPS SOP instance at the remote SCP 8 2 4 2 1 2 Proposed Presentation Context Table The following table shows the proposed presentation contexts for the GSPS AE after any of the real world activity listed in section 7 2 3 Sequencing of Real World Activities is initiated Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Grayscale 1 2 840 10008 5 1 4 1 1 Explicit VR 1 2 840 10008 1 2 SCU None Presentation 11 1 Little Endian State 8 2 4 2 1 2 1 SOP Specific DICOM Conformance Statement for GSPS SOP class If the remote AE does not support the proposed Presentation context an appropriate error message logged Only one association is opened at a time All the operations used by this SOP class support an association timer which is configurable The timer is started when a request association request N CREATE request or N SET request is send and stopped when the respective response is received The default time out value is 300 seconds
200. wing DICOM 3 0 SOP classes NAME CT Imaae Storaae SOP Class 1 2 840 10008 5 1 4 1 1 2 Imaae Storaae SOP Class 1 2 840 10008 5 1 4 1 1 4 GE Private PET Raw data Storaae 1 2 840 113619 4 30 Secondary Capture Storage SOP Class 1 2 840 10008 5 1 4 1 1 7 Grayscale Softcopy Presentation State Storage SOP Class 1 2 840 10008 5 1 4 1 1 11 1 5 3 1 5 Request Processing The images in the selected exam series are sent to the remote commitment provider entity using DICOM C STORE operations If there are any failures in the image transfers the Storage commitment request will not be sent The corresponding job will be marked as failed and user will be notified of the status of job If all the images are successfully transferred then the commitment request will be sent on a different association with the list of sop instances If the N Action request fails the job will be marked as failed otherwise following sequence of actions will take place The SCU waits for N Action Rsp from provider on the same association for a configurable amount of time If it does not receive N Action Rsp during this time it will close the association It changes the Job state to Wait indicating the job is waiting for the response from commitment provider The job will be marked as failed if the response is not received by stop job time Stop job time is the maximum duration the job can wait for responses A New transaction uid will be created for each retry by
201. with MWL data PPS object PPS Manager The system has a Modality Work list Server AE installed Work List information is obtained from HIS RIS system through the use of Basic Work list Management Service Use of the information retrieved in the creation of Image SOP instance is described in the Modality Work list Conformance statement Use of the information retrieved in MPPS SOP instances is described later in this document After Post processing Exam Split application presents associated requested Procedure s along with selected acquired Images Exam Split application includes the necessary information related to Requested Procedure scheduled Procedure Steps and the Performed Procedure Step of the images acquired during acquisition After User modifications on Image s user has to click on Send button to create Gray Scale Presentation State and Performed Procedure Step objects to transmit to remote host If there is any Image Orientation user will be notified about the orientation before sending to Remote Host Exam Split will provide Host Selection user interface to select remote host to transmit of Gray Scale Presentation State object Performed Procedure Step object will be sent to default Performed Procedure Step host After successful transmission of Gray Scale Presentation State and Performed Procedure Step objects to remote host user will be notified with success 112 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1
202. y for all validation required for their connection with GE devices This includes the accuracy of the image data once it has crossed the interface between the GE imaging equipment and the non GE device and the stability of the image data for the intended applications Such a validation is required before any clinical use diagnosis and or treatment is performed It applies when images acquired on GE imaging equipment are processed displayed on a non GE device as well as when images acquired on non GE equipment is processed displayed on a GE console or workstation 20 of 169 GE HEALTHCARE DIRECTION 5198351 100 Rev 1 Discovery DICOM CONFORMANCE STATEMENT e Future Evolution GE understands that the DICOM Standard will evolve to meet the user s growing requirements GE is actively involved in the development of the DICOM Standard DICOM will incorporate new features and technologies and GE may follow the evolution of the Standard ID Net v3 0 is based on DICOM as specified in each ID Net DICOM Conformance Statement Evolution of the Standard may require changes to devices that have implemented DICOM In addition GE reserves the right to discontinue or make changes to the support of communications features on its products reflected on by these ID Net DICOM Conformance Statements The user should ensure that any non GE provider which connects with GE devices also plans future evolution of the DICOM standard Failure to do so will likely re
203. y to Local Storage RWA see Table 3 1 1 For extensions and specialization s see section 3 4 3 3 1 2 3 2 Options None applicable 3 3 1 2 4 Real World Activity Update MOD The MOD Archive Application acts as an FSU when requested to update an MOD The MOD Archive Application will take the selected list of SOP instances and eliminate any SOP instance that does not correspond to permissible SOP instances listed in the Application Profiles Table 3 1 1 The remaining SOP instances are written to the media that is found in the MOD Archive Application disk drive The determination of the potentially applicable Application Profile is dependent on the type of media and associated software on which the AE has been invoked 3 3 1 2 4 1 Application Profiles for the RWA Update MOD Request For the list of Application Profiles that invoke this AE for the MOD Copy to Local Storage RWA see Table 3 1 1 For extensions and specialization s see section 3 4 3 3 1 2 4 2 Options None applicable 3 4 Standard Augmented and Private Application Profiles 3 4 1 Class and Profile Identification The Class of Application Profiles defined here is for Computed Tomography and Magnetic Resonance Imaging Clinical applications The identifier for this class is CTMR MOD The specific Application Profile in this class is shown in Table 3 3 1 1 Table 3 3 1 1 CTMR MOD Profiles CT MR PT Studies on STD CTMR MOD23 Handles single 12 or 16 bit grayscale 2 3GB M
Download Pdf Manuals