orthophos xg - Sirona Support
Contents
1. e To determine power supply adequacy the line volt age drop during exposure must be measured 1 Be sure power is disconnected at the central distribu tion panel e Remove front cover for details see Service Manual 2 Select 300VAC line voltage range on multimeter Connect measuring leads to terminal K1 L and N 3 Connect power and switch unit ON Wait 1 min for self adjustment of the unit Press key R to return X ray tube head into the initial position 4 Establish exposure readiness via SIDEXIS 5 Select highest exposure level e g 90kV 12mA A CAUTION RADIATION Depress the exposure key until meter reading is obtained 59 38 373 D3352 D3352 103 01 06 02 4 Power supply adequacy 300 VAC Line voltage no load Max permissible line voltage drop 180 208V 9V 208 230V 8V 230 240V 7 5V 240 264V 7V Record reading e Turn unit OFF e Remove meter leads and refit cover NOTE If the voltage drop is not within the specified range ad vise the customer that an adequate power supply must be installed Refer to Pre Installation Instructions Disconnect unit and do not release for use 5 Tube Current Verification Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 5 Tube Current Verification A L WAR
2. Ate ATTENTION Electrical Shock Hazard Always turn unit OFF before connecting and disconnecting the test leads to the test points ON OFF 2 AA WARNING Life threatening voltage on DX6 2 Visual Check Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 2 Visual Check SIDEXIS Software Logiciel This product complies with DHHS This product complies with DHHS This product complies with OHHS This product complies with OHHS regulations 21 CFR Subchapter J regulations 21 CFR Subchapter J regulations 21 CFR Subchapter J regulations 21CFR Subchapter J appicable at date of manufacture applicable at date of manufacture applicable at date of manufacture apricable at date of manufacture Date of manufacture oes Date of manufacture ses Date of manufacture be eS Date of manufacture fg Sirona Dental Systems Gat lH Sirona Dental Systenes fmt Sirona Dental Systems GatH Sirona Dental Systems GatH Farish pi LAAB Besen Farish i use Danem Fabekste 21 D 44425 Berchein Fabrixstr 3 D 44625 Bershein Cernay Cernay Cermay Made in Germany Made in Germany Made in Germany Made in Germany e Look for mechanical damage possibly affecting radia tion safety e Verify that all labels are affixed and legible Defaced labels must be replaced Order same from Sirona address see rear in writing stating e Customer Name e Customer Address All Model Numbers with Serial Numbers still legible on the unit
3. ORTHOPHOS XG 5 3 ORTHOPHOS XG 3D ady e Make a exposure e Interrupt an exposure deadman feature X ray head must be in the initial position If not Observe a cool off time of max 5 mins between press return key R exposures automatic exposure blockage Establish exposure readiness via SIDEXIS Setting same as above For more details and possible error messages see A Operating Instructions CAUTION RADIATION Set the P1 exposure program using the menu keys Depress the release button until X ray lights up and XG 5 3 or keys XG 3D 9Y subsequently release the exposure must Select 68kV 8mA using submenu keys XG 5 3 ee immediately The radiation indicator or submenu keys XG 3D 29Y IS Up Z ATTENTION A CAUTION RADIATION Defective light indicators constitute a safety hazard to the Depress the exposure button and hold until the patient as well as to the operator l l l exposure terminates automatically The user is not permitted to use the unit until repairs The exposure ends when the rotation and radiation are made automatically switch off The radiation indication must light up during the exposure period Simultaneously an audible beep must sound at the unit 59 38 373 D3352 8 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 4 Power supply adequac mul 2
4. Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373
5. The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 010 1 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears ORTHOPHOS XG 3D 2 The exposure dialog box showing the exposure status appears in Sidexis Service routine 010 1 is displayed on the Easypad touchscreen 59 38 373 D3352 D3352 103 01 06 02 Sirona Dental Systems GmbH 8 Checking the X Ray Beam for panorama exposure Maintenance Instructions ORTHOPHOS XG 8 2 Checking the pan sensor adjustment vvv o OOO XG 3p eady XG5 3 Adjustment ok Oo NOTE 3 Take an exposure 60 kV 3 mA If the criteria specified above are not fulfilled the sensor Press the R key to move the unit back to the must be adjusted The procedure for adjusting the unit is de starting position scribed in the service manual Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 4 Evaluate the exposed image The three needle images must lie in the center of the exposed areas and inside the auxiliary lines see image 4 59 38 373 D3352 D3352 103 01 06 02 19 8 Checking the X Ray Beam for panorama exposure 8 3 Checking the p
6. Save values 10 00 mm Fx gt lt gt ONLA 0 20 8 4 Checking the pan filter adjustment Not necessary for ORTHOPHOS XG 3 1 Change to the PAN FILTER submenu 2 Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION ORTHOPHOS XG 5 22 The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine S030 3 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG in SIDEXIS Exposure Preview Follow a control tt 60 kV 3 mA 0 20 s ORTHOPHOS XG 3p a9y The exposure dialog box showing the exposure status appears in Sidexis Service routine 030 3 is displayed on the Easypad touchscreen 59 38 373 D3352 D3352 103 01 06 02 Sirona Dental Systems GmbH 8 Checking the X Ray Beam for panorama exposure Maintenance Instructions ORTHOPHOS XG 8 4 Checking the pan filter adjustment aq a OA 28 m 6a JERAI XG 3p eady g XG5 09 OOO oe Adjustment ok voer e e ee NOTE 3 Take am exposure 60 Ky S MAJ If the current image differs from the ideal image see image Press the R key to move the unit back to the 4 the diaphragm must be adjusted The pro
7. in selection field 1 to select the kV mA level 62kV 8mA 2 Switch the unit on via the switch A see also Operating Instructions The X Ray radiation indicator B lights up briefly 6 Use the arrow keys A in selection field 2 to select the radiation time 0 5 s After approx 2 seconds the green LED C in the Performing measurements upper part of the control panel lights up This LED 7 Initiate the radiation Hold the release button pressed remains lit as long as the unit is on until the set radiation time has expired CAUTION Activating the release button triggers The start screen appears on the touchscreen of X rays g gg the Easypad and the system s self adjustment routine starts running for approx 1 minute The rotating element rotates briefly clockwise and counterclockwise The diaphragm moves into position The forehead and temple supports on the panoramic unit open and close and then stop moving in fully opened position On completion of the self adjustment routine the main menu appears on the touchscreen Help message H301 prompts you to move the unit into the starting position 3 Press the R key The unit moves to its starting position 59 38 373 D3352 D3352 103 01 06 02 13 6 kV verification Exposure Time Verification Sirona Dental Systems GmbH 6 1 kV verification Maintenance Instructions ORTHOPHOS XG kV verification Exposure Time Verification 6 1 kV verification Analy
8. test block and needle phantom delivered with the unit cus tomer s property 59 38 373 D3352 D3352 103 01 06 02 ATTENTION RADIATION Observe the radiation protection guide lines as outlined in the Operating Instructions manual X radiation is emitted as long as the release button is de pressed The X ray indicator must light up on the Easypad during ra diation An acoustic signal must also be heard Power Supply Adequacy To assure that the ORTHOPHOS XG system performance is in accordance with Sirona specifications an adequate power supply for permanently installation is essential The Federal Performance Standard for diagnostic X ray units code of Federal Regulations Title 21 CFR Subchap ter J mandates an adequate power supply Duty Cycle Between exposures maintain at least a cool off time auto matic exposure blockage see Operating Instructions man ual Operating Instructions During measurements and controls it is necessary to ener gize or de energize the unit For all operating steps please refer to the Operating Instructions manual CAUTION with PC Boards All PC boards are fitted with electronic components sensi tive to electrostatic discharge ESD In an environment of moving people electrostatic charges are unavoidable due to friction caused by clothing carpeting etc ATTENTION To prevent damage of electronic chips do not touch same Always handle circuit boards by their edges
9. D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified lim
10. D3352 D3352 103 01 06 02 in SIDEXIS Preview 80 kV 14 mA 14 9 s ORTHOPHOS XG 5 The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine S010 3 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears ORTHOPHOS XG 3p ady The exposure dialog box showing the exposure status appears in Sidexis Service routine 010 3 is displayed on the Easypad touchscreen 27 9 Checking the X Ray Beam for cephalometer exposure Sirona Dental Systems GmbH 9 2 Checking the Ceph primary diaphragm adjustment Maintenance Instructions ORTHOPHOS XG XG 3p eady XG5 3 Take an exposure 80 kV 14 mA Press the R key to move the unit back to the starting position Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 59 38 373 D3352 28 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG Adjustment ok i NOTE If S3 is significantly larger than 60 mm check whether light barrier V61_ 1 LB on the horizontal diaphragm slider is properly installed If necessary correct the alignment of the light barrier and then repeat the entire pan and ceph adjust men
11. General control of the unit by means of the Multipad as in the user mode is not possible in this mode 15 7 Checking the X Ray Beam Sirona Dental Systems GmbH 7 1 Diaphragm system adjustment menu Maintenance Instructions ORTHOPHOS XG The DIAPHRAGM SYSTEM ADJUSTMENT menu has 11 submenus Pan Sensor adjustment Pan Diaphragm Pan Filter Pan Symmetry Ceph Primary diaphragm Ceph Fixed point of rotation Ceph Main X ray beam direction Ceph Quickshot full version only Pan Reset adjustment Ceph Reset adjustment Sirona Service Sheet internal SIRONA information You can change between the individual submenus by click ing the tabs with the mouse To quit the DIAPHRAGM SYSTEM ADJUSTMENT menu click CANCEL 59 38 373 D3352 16 D3352 103 01 06 02 Sirona Dental Systems GmbH 8 Checking the X Ray Beam for panorama exposure Maintenance Instructions ORTHOPHOS XG 8 1 Preparing the Pan adjustment 8 Checking the X Ray Beam for panorama exposure Needles on top side Preparing the Pan adjustment In order to perform the PAN sensor adjustment and the symmetry adjustment you must insert needle phantom A in the bite block holder of the panoramic X ray unit The needle phantom must be removed from the bite block holder for the pan diaphragm adjustment M ATTENTION It is essential that the needle phantom is removed from bite block holder of the panoramic X ray unit again before a
12. for identification purposes For serial numbers see also Installation Report Warranty Passport 59 38 373 D3352 6 D3352 103 01 06 02 Sirona Dental Systems GmbH 3 Indicator Lights and audible Signal Maintenance Instructions ORTHOPHOS XG 3 indicator Lights and audible Signal Radiation indicator Unit ON LED sirona XG 3p eady Radiation indicator Unit ON LED e Unit ON LED Depress the main switch into the I position to turn unit ON The unit adjusts itself automatically wait about 1 minute The Unit ON LED in the upper right corner of the user interface will then indicate that the unit is ON e Testing the keys and the functions on the user in terface When the height adjustment is made the loud speaker should hum If the forehead support functions correctly the max holding torque is still present If remote is installed test the remote keys and dis play in addition The release button must not be defective or broken See also Operating Instructions under Preparing the Exposure subchapter Switching the system on 59 38 373 D3352 D3352 103 01 06 02 7 3 Indicator Lights and audible Signal Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG CAUTION RADIATION Observe Radiation Protection Guide Lines see Operating Instructions Radiation indicator Unit ON LED Radiation indicator Unit ON LED subme nukeys Return key R Return key R keys Submenu
13. Ceph exposure is taken otherwise the phantom may collide with the sensor O NOTE When fitting the needle phantom make sure that it is correctly oriented For the adjustment of the X ray unit the phantom must be fitted in such a way that the needles point upward 2 59 38 373 D3352 D3352 103 01 06 02 17 8 Ch ecking the X Ray Beam for panorama exposure 8 2 Checking the pan sensor adjustment Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 1 Pan 5i 52 53 IV on the Multipad XG 5 3 60 3 8 2 o The Diaphragm system adjustment Kammer4 Image acquisition Save values Sensor adjustment Pan Diaphragm Pan Symmetry Pan Reset adjustment 0 00 mm Precision adjustment t t 4 gt j Ste gt j 82 gt 33 i _ gt 0 60 Checking the pan sensor adjustment Plug the sensor into the sensor slot on the panoramic X ray unit Insert the needle phantom in the bite block holder of the panoramic X ray unit See page 17 Open the PAN SENSOR ADJUSTMENT submenu NOTE menu provides a precision adjustment and a coarse ad justment precision adjustment is preset Perform a preci sion adjustment first In most cases previous coarse adjust ment is not necessary 2 18 Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION in SIDEXIS Preview 60 kV 3 mA 0 68 ORTHOPHOS XG 5 3
14. D3352 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 1 General To stay in compliance with the DHHS requirements the ORTHOPHOS XG must be maintained annually following date of installation It is the responsibility of the user to insure that the equip ment is maintained with the manufacturer s recommended Maintenance Instructions to insure compliance with the Federal Performance Standard The manufacturer and the assembler installer are relieved from responsibility in those cases where noncompliance with the standard results from the user s failure to have the manufacturer s recommended maintenance performed The actual maintenance inspection and consequent ser vice must be accomplished by a trained serviceman Nei ther the inspection nor service is part of the equipment war ranty Technical instructions required Operating Instructions Service Manual Instruments and adjustment tools required All measuring devices must be suitable for measurements on the mains potential This includes the pulse counter This means A voltage of 1000 V between a connection and ground must be possible 1 Digital multimeter Fluke 87 III or equivalent Accuracy DC voltage 0 1 of reading plus 0 02 of range DC current 0 4 of reading plus 0 1 of range 2 A dose measurement device e g Mult O Meter type 510L is required for dosimetry 3 Adjustment set with alignment tool for X ray beam
15. NING The electronics of the X ray tube assembly are always 1 Remove cover for details see Service Manual connected to line voltage 2 Loosen the four screws A and remove cover plate B of Always switch the X ray unit off and wait until V203 is the electronics box no longer illuminated before contacting the test leads Z WARNING The test leads and measuring instruments used must have a dielectric strength of at least 1000V Be sure to use a battery powered measuring instru ment with shock hazard protected sockets Use only test leads with shock protection ATTENTION The X ray tube assembly moves during the measurement Therefore be sure to use test leads of sufficient length and place the measuring instrument in a location where it is firm ly seated so that it doesn t fall down 59 38 373 D3352 10 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG PaE PC board DX6 p 9 ey ie co OCce j A WARNING Be sure to switch the X ray unit off before removing the jumper for the mA measuring jack 3 Remove jumper A from MA MA test points on PC board DX6 Connect digital ammeter to MA and MA and select range 20 mADC 4 Reattach cover C and tighten screw D securely Switch unit ON Wait 1 min for self adjustment of the unit Press key R to return X ray tube head into the initial position 6 Select 66kV 8mA Establish exposure readin
16. New as of 09 2012 golir The Dental Company ORTHOPHOS XG 3D 29Y Ceph ORTHOPHOS XG 5 Ceph ORTHOPHOS XG 3 Maintenance Instructions ORTHOPHOS XG 3D Version 6 0 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG General information ATTENTION Proper shielding of room and operator position is essential Since these requirements vary from state to state it is the assembler s installer s responsibility that all local radiation safety requirements are met New as of 09 2012 Changes since the last version 12 2010 59 38 373 D3352 D3352 103 01 06 02 N Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG AGING ied aes sree ea a owe anos a e a a O 5 VISUAL COCK nn i a a a a 6 Indicator Lights and audible Signal sccccscseeseeseeeeeeeseeseeeeseeeeeseeeeseeeeseseeseeeeeseeeseenens 7 POWEr SUDDIY AGCOUACY encra a aa a eres 9 TUDES Current Verifica on sssaaa aa E E a a aF 10 kV verification Exposure Time Verification ccccccssesseeessseseeeeseenseeseeesseeeseeenneeeaes 13 6 1 VS VSI ATI Osteen sas dyademtesn yacanesoventcaes a 14 6 2 Exposure Time VerifiCation cccccssecccccceeeeeeeseeeeeeseeseeeeeseneeeeeesaneeeeeas 14 CHECKING the X Ray CAIN smsa a a a aaa a a aa 15 7 1 Diaphragm system adjustment MENU ccecccesseeeeceeeeeeeeeeseeeeeensaees 15 Checking the X Ray Beam for panoram
17. a eXpoSuUre s sssssssnnsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nnne 17 8 1 Preparing the Pan adjustment ccccccsseeeeeeeeeeeeeeeeeeeeeeeseeeeeeneeeesaanees 17 8 2 Checking the pan sensor adjustment ceccecceseeeeeeeeeeeeeeeeseeeeeeeeaees 18 8 3 Checking the pan diaphragm adjustment ccccceeeseseeeeeeeeeeeeeeneees 20 8 4 Checking the pan filter adjustment cece ceseeeeeeeeeeeeeeeeeeseeeessaeeeeeeas 22 8 5 Checking ihe pan symmetry scx tencd cei ec eave ee a 24 Checking the X Ray Beam for cephalometer CXPOSULE scccceeseeeeeeeeeeseeeeeneeeeeeeees 26 9 1 Preparing the Ceph adjUStMent cccccccsseeeeeceeeeeeeeceeeeeessenseeeesaeees 26 9 2 Checking the Ceph primary diaphragm adjustment ccccseeeeeeeeees 27 9 3 Checking the ceph rotation fixpoint adjuStMent ccccssseeeeeeeeeeeeeees 30 9 4 Checking the ceph main X ray beam direction adjustment 32 Checking the alignment of the ear PIUGS cceccceeeseeeeeneeeeseeeeaseeeeseeeeaseeeeaseeoeneennsenes 34 10 1 Preparing the ear OLIVES cc ccecccesececeseeecesseeeeeueeeeeeeeeseseeeseueesseeesseaes 34 10 2 Checking the ear plug alignment cccccsseeeeeeeeeeeeeeceeeeeeeeseeeeeesaaees 35 Checking TING TAS Claes ects asain ee aa a tow tacice 38 59 38 373 D3352 D3352 103 01 06 02 3 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 59 38 373
18. an diaphragm adjustment Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG ORTHOPHOS XG 3D Y Ceph ORTHOPHOS XG 5 Ceph Diaphragm system adjustment Kammer4 Pan Sensor adjustment Pan Diaphragm Pan Filter Pan Symmetry Ceph Primary 4 gt Sx 0 00 mm Sy 000mm IV Precision adjustment Image acquisition Save values 2 zel on the Multipad XG 5 60 3 0 20 8 3 Checking the pan diaphragm adjustment e Remove the needle phantom from the bite block holder of the panoramic X ray unit 1 Change to the PAN DIAPHRAGM submenu id NOTE For XG 3D 8 Y XG 5 The menu provides a precision adjustment and a coarse adjustment Always try to use precision adjust ment first when adjusting the unit In most cases it is not necessary to perform a coarse adjustment prior to precision adjustment 2 Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION 20 in SIDEXIS Preview 60 kV 3 mA 0 20 s ORTHOPHOS XG 5 3 The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 030 2 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears ORTHOPHOS XG 3D 2 The exposure dialog box showing the exposure status appears in S
19. cedure for ad starting position justing the unit is described in the service manual Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 4 Evaluate the exposed image The superimposed filter must cover one half of the diaphragm see image 4 59 38 373 D3352 D3352 103 01 06 02 23 8 Checking the X Ray Beam for panorama exposure 8 5 Checking the pan symmetry Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 1 Diaphragm system adjustment Kammer4 Image acquisition Save values Siig 0 00 mm H E 52 0 00 mm 53 0 00 mm E 2 5 on the Multipad XG 5 3 8 5 Checking the pan symmetry 24 Insert the needle phantom in the bite block holder of the panoramic X ray unit See page 17 Change to the PAN SYMMETRY submenu Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION See aN a STNG in SIDEXIS Preview g for exposure SIT ON a i ere on the touchscreen XG 3D 2 7 60 kV 3 mA 14 15 ORTHOPHOS XG 5 3 The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 010 2 60 kV 3 mA 0 6 s are di
20. de the superimposed auxiliary lines see image 4 59 38 373 D3352 D3352 103 01 06 02 9 Checking the X Ray Beam for cephalometer exposure 9 3 Checking the ceph rotation fixpoint adjustment Adjustment ok i N A white border surrounding the image on all sides must be visible O NOTE If the criteria specified above are not fulfilled the ceph fixed point of rotation must be adjusted The procedure for adjust ing the unit is described in the service manual 31 9 Checking the X Ray Beam for cephalometer exposure 9 4 Checking the ceph main X ray beam direction adjustment 1 Ceph Fixed point of rotation Ceph Main x ray beam direction Ceph Quickshot TSA o4 gt 51 on the Multipad XG 5 8 9 4 32 Diaphragm system adjustment Kammer4 0 00 mm Image acquisition Save values E Cancel 0 14 14 9 Prog Checking the ceph main X ray beam direction adjustment Insert the test phantom in the sensor slot on the pan oramic X ray unit see page 26 Swing the ear plug holders out of the beam direction Change to the CEPH MAIN X RAY BEAM DIRECTION submenu Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG in SIDEXIS Preview 80 kV 14 mA 14 9 s ORTHOPHOS XG 5 The exposure dialog box showing the exposure status appears in Sidexis The initialization status i
21. e Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ra
22. ess via SIDEXIS 7 IfP1 program and 66kV 8mA are selected The unit must be ready for radiation 59 38 373 D3352 D3352 103 01 06 02 5 Tube Current Verification Measurement A CAUTION RADIATION Depress the exposure key and hold depressed until meter reading is obtained The ammeter shall indicate 8mA 1 6mA Record reading i NOTE Readings 1mA corresponds to a tube current of 1MA per missible tolerance 20 i NOTE If specified values cannot be obtained see Service Man ual chapter Tube Current Verification 11 5 Tube Current Verification Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG ba i ee ff Nas eth IZ ay J T i f y I Nj e f specified value is obtained switch unit OFF 8 Remove upper cover and meter leads Replace jumper wo e Screw the cover plate back on to the electronics box 10 Reattach the housing covers 59 38 373 D3352 12 D3352 103 01 06 02 Sirona Dental Systems GmbH 6 kV verification Exposure Time Verification Maintenance Instructions ORTHOPHOS XG 6 kV verification Exposure Time Verification A X RAY 60kV 8mA gt oD Sirona Preparing the measurement 4 Call the Service menu and the Service routine S002 3 see Service Manual 1 Attach the Mult O Meter sensor in the middle of the du al sensor 5 Use the arrow keys A
23. idexis Service routine 030 2 is displayed on the Easypad touchscreen 59 38 373 D3352 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG o o OOOO XG 3p ady 3 Take an exposure 60 kV 3 mA Press the R key to move the unit back to the starting position Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 4 Evaluate the exposed image The exposed diaphragm area must lie horizon tally centered in the image field as well as inside the superimposed auxiliary lines A white border surrounding the image on all 59 38 373 D3352 D3352 103 01 06 02 8 Checking the X Ray Beam for panorama exposure 8 3 Checking the pan diaphragm adjustment Adjustment ok sides must be visible The maximum density must lie in the center of the diaphragm area i NOTE If the current image differs from the ideal image see image 4 the diaphragm must be adjusted The procedure for ad Justing the unit is described in the service manual 21 8 Checking the X Ray Beam for panorama exposure 8 4 Checking the pan filter adjustment T Diaphragm system adjustment Kammer4 Pan Sensor adjustment Pan Diaphragm Pan Filter Pan Symmetry Ceph Primary 4 gt on the Multipad XG 5 60 3 Fx Image acquisition X
24. its Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 We reserve the right to make any alterations which may be required due to technical improvements Sprache englisch Printed in Germany D3352 103 01 06 02 09 2012 A Nr 116382 Imprim en Allemagne aaa rey Sirona Dental Systems GmbH in the USA FabrikstraBe 31 Sirona Dental Systems LLC Order No 59 38 373 D3352 64625 Bensheim 4835 Sirona Drive Suite 100 Germany Charlotte NC 28273 WWW sirona com USA
25. log box for selecting the test type appears on the screen 4 Select confirm the test type Click SERVICE EXPOSURE The dialog box for selecting the service exposure ap pears on the screen 59 38 373 D3352 D3352 103 01 06 02 10 Checking the alignment of the ear plugs 10 2 Checking the ear plug alignment ST x Acceptance test Constancy test Cancel 35 10 Checking the alignment of the ear plugs Sirona Dental Systems GmbH 10 2 Checking the ear plug alignment Maintenance Instructions ORTHOPHOS XG Select Service exposure Digital test pattern Follow and control the Factory Service ae Diaphragm test exposure gt Diaphragrn system adjustment a Quality test exposure S fON a 5 mer Cancel i NOTE The ceph mode must be activated on the Multipad XG 5 or Easypad XG 3D for the ceph quality test exposure 5 Select confirm the service exposure Click QUALITY TEST EXPOSURE i NOTE If necessary select the X ray component The exposure dialog box showing the exposure status appears in Sidexis 59 38 373 D3352 36 D3352 103 01 06 02 Sirona Dental Systems GmbH 10 Checking the alignment of the ear plugs Maintenance Instructions ORTHOPHOS XG 10 2 Checking the ear plug alignment XG 3p eady XG5 6 Take an exposure 80 kV 14 mA Press the R key to move the unit back to the starting position Press the release b
26. must be ad justed The procedure for adjusting the unit is described in the service manual 25 9 Checking the X Ray Beam for cephalometer exposure Sirona Dental Systems GmbH 9 1 Preparing the Ceph adjustment Maintenance Instructions ORTHOPHOS XG 9 Checking the X Ray Beam for cephalometer __ 100 i 9 1 Preparing the Ceph adjustment You must insert test phantom B in the sensor slot on the panoramic X ray unit in order to perform the adjustment of the ceph primary diaphragm and the adjustment of the ceph main X ray beam direction The test phantom must be removed from the sensor slot on the panoramic X ray unit for the adjustment of the ceph fixed point of rotation and for the ceph quickshot 59 38 373 D3352 26 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 9 Checking the X Ray Beam for cephalometer exposure 9 2 Checking the Ceph primary diaphragm adjustment 2 a Cancel on the Multipad XG 5 9 2 Checking the Ceph primary diaphragm adjustment Move the ear plug holders on the cephalometer completely apart and swing them out of the beam direc tion ap Insert the test phantom in the sensor slot on the pan oramic X ray unit See page 26 Plug the sensor into the sensor slot on the cephalome ter Open the DIAPHRAGM SYSTEM ADJUSTMENT menu Select the CEPH PRIMARY DIAPHRAGM submenu Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION 59 38 373
27. nufactured in Bensheim SAG laser vertical line The light beam must strike the center of the head support in a vertical direction FH laser horizontal line The light beam must strike the FH line of the sheet of card board attached to the head support It must be possible to displace the light beam vertically using slide C Unwanted movement of the FH light localizer is not permitted i NOTE To adjust the lasers see Maintenance Instructions No controls are available to adjust the laser power 59 38 373 D3352 D3352 103 01 06 02 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenanc
28. re for adjusting the unit is described in the service manual 33 10 Checking the alignment of the ear plugs Sirona Dental Systems GmbH 10 1 Preparing the ear olives Maintenance Instructions ORTHOPHOS XG 10 1 Preparing the ear olives 1 Move the ear plug holders completely apart and swing them into the beam direction i NOTE 2 Fitadjusting caps D onto the ear plugs and secure them The sensor must be plugged into the sensor slot on the with adhesive tape l l cephalometer No sensor may be plugged into the sensor Black adjusting cap on the outside sensor side slot on the panoramic X ray unit Transparent adjusting cap on the inside tube as sembly side e Select the CEPH mode on the Multipad XG 5 or 3 Unscrew the cover from the cephalometer Easypad XG 3D 9Y 59 38 373 D3352 34 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG Plugins b Macros b Configure system 3 Select X ray device xj Existing devices al OF Cancel i 10 2 Checking the ear plug alignment 1 In Sidexis XG select the constancy test EXTRAS gt CONSTANCY TEST The typical SIDEXIS user interface is started The con stancy test is preset 2 Start the exposure mode Click XCXP The dialog box for selecting the X ray device appears on the screen 3 Select confirm the X ray device Select e g Room2 Kammer2 and click OK The dia
29. s visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine S010 6 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears ORTHOPHOS XG 3D 2 The exposure dialog box showing the exposure status appears in Sidexis Service routine 010 6 is displayed on the Easypad touchscreen 59 38 373 D3352 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG Ka a 9 Q 3 3 maniy 7 A A amp o A ap ab WwW Wy aE OO ahoh XG 3p ady XG5 9 Checking the X Ray Beam for cephalometer exposure 9 4 Checking the ceph main X ray beam direction adjustment Adjustment ok 3 Take an exposure 80 kV 14 mA Press the R key to move the unit back to the starting position Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 59 38 373 D3352 D3352 103 01 06 02 4 Evaluate the exposed image A horizontal bar must be visible in the center of the image see image 4 If this bar is visible the exposure is OK and the two beams imaged are within the tolerance band of 10 mm see image 4 O NOTE If the criteria specified above are not fulfilled the ceph main X ray beam direction must be adjusted The procedu
30. splayed and the progress in dicator disappears ORTHOPHOS XG 3D 2 y The exposure dialog box showing the exposure status appears in Sidexis Service routine 010 2 is displayed on the Easypad touchscreen 59 38 373 D3352 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG Al o o o ke 1 OOO w X XG 3pready LJ XG 5 3 0 75 mm lt gt A1 88 6 mm 1 mm 8 Checking the X Ray Beam for panorama exposure 8 5 Checking the pan symmetry 0 75 mm gt A2 44 3 0 5 mm le gig y A2 44 3 0 5 mm Adjustment ok 3 Take an exposure 60 kV 3 mA Press the R key to move the unit back to the starting position Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 4 Evaluate the captured image The shadow of the center needle the needle im age and the auxiliary line must be congruent and located behind each other A tolerance offset of needle from the central auxiliary line of 0 75 mm is admissible 59 38 373 D3352 D3352 103 01 06 02 Distance A1 must be 88 6 1 mm Distances A2 must be identical each being 44 3 0 5 mm A white border surrounding the image on all sides must be visible when zoomed out fully i NOTE If these criteria are not fulfilled the symmetry
31. t Da Image acquisition Save values Cancel n ET ae gt i Ste gt j S2 je gt j S3 je lt gt XG 3 pread 7 1 Diaphragm system adjustment menu The DIAPHRAGM SYSTEM ADJUSTMENT menu guides you through the adjustment of the panoramic unit and the cephalometer This service routine is started from SIDEXIS XG EXTRAS gt CONSTANCY TEST gt XCXP gt SELECT X RAY DEVICE gt SERVICE EXPOSURE gt DIAPHRAGM SYSTEM ADJUSTMENT i NOTE The DIAPHRAGM SYSTEM ADJUSTMENT menu is pass word protected As password enter the first four digits of the current system date PC in reverse order Example On 05 30 2004 the service password is 5003 i NOTE When you open the DIAPHRAGM SYSTEM ADJUSTMENT menu the unit switches from the user mode to the PC ser 1S9 lagged by the PC D3352 103 01 06 02 Diaphragm system adjustment OP_121 xj a AN a ORTHOPHOS XG 3D Y Ceph This mode is indicated by the PC service screen 2 dis played on the Easypad In PC service mode the Easypad control options are determined by SIDEXIS and the current ly selected service routine General control of the unit by means of the Easypad as in the user mode is not possible in this mode ORTHOPHOS XG 5 Ceph ORTHOPHOS XG 3 This mode is indicated by the message SERVICE 3 on the Multipad In PC service mode the Multipad control op tions are determined by SIDEXIS and the currently selected service routine
32. t procedure must be installed vertically 59 38 373 D3352 D3352 103 01 06 02 S3 approx 60 mm 9 Checking the X Ray Beam for cephalometer exposure 9 2 Checking the Ceph primary diaphragm adjustment 4 Evaluate the captured image The vertical pin must be horizontally centered in the exposed image area A uniform white border surrounding the image on all sides must be visible Distance S3 must be approx 60 mm i NOTE A slight vertical offset of the grid is permissible O NOTE If the image differs from the ideal image 4 the ceph primary diaphragm must be readjusted The procedure for adjusting the unit is described in the service manual 29 9 Checking the X Ray Beam for cephalometer exposure 9 3 Checking the ceph rotation fixpoint adjustment Diaphragm system adjustment Kammer4 I Pan Filter Pan Symmetry Ceph Primary diaphragm Ceph Fixed point of rotation Ce4l gt S1 b 00 mm m 52 0 00 mm S3 0 00 mm IV Precision adjustment Image acquisition ie Save values g 2 iel on the Multipad XG 5 a AN a 80 14 0 60 9 3 Checking the ceph rotation fixpoint adjustment e Remove the ceph test phantom from the sensor slot of the panoramic X ray unit 1 Change to the CEPH FIXED POINT OF ROTATION submenu i NOTE The menu provides a precision adjustment and a coarse adjustment Always try to use precision adjust ment firs
33. t when adjusting the unit In most cases it is not necessary to perform a coarse adjustment prior to precision adjustment 2 Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION 30 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG in SIDEXIS Preview 80 kV 14 mA 0 60 s ORTHOPHOS XG 5 The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 010 5 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears ORTHOPHOS XG 3D 2 The exposure dialog box showing the exposure status appears in Sidexis Service routine 010 5 is displayed on the Easypad touchscreen 59 38 373 D3352 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG GOP A un y A aa AA ADU MT 1 00 al o o XG 3p eady XG5 3 Take an exposure 80 kV 14 mA Press the R key to move the unit back to the starting position Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 4 Evaluate the exposed image The exposed diaphragm area must lie centered and straight in the image field as well as insi
34. tion indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks
35. utton Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box e The lead balls in the adjusting caps appear as dots on the image The two dots must be coincident i NOTE If the criteria specified above is not fulfilled the ear olives must be adjusted The procedure for adjusting the ear olives is described in the service manual 59 38 373 D3352 D3352 103 01 06 02 37 11 Checking the laser 10 2 Checking the ear plug alignment Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 11 Checking the laser eS ae a a ae re a ee ee ee ee 1 Insert the forehead and temple supports 2 Check the laser e Fasten a piece of white cardboard between the temple supports e Swivel the mirror by pressing the left recess A on the touch bar e Switch on the light localizer with key B on the Easy pad Multipad The Lasers generate a red line Z WARNING Class 1 radiation is emitted during installation Always keep eyes a minimum distance of 100mm away from the laser Do not stare into the beam 38 Horizontal beam Vertical beam LASER KLASSE 1 CLASS 1 LASER PRODUCT APPAREIL A LASER DE CLASSE 1 Date of manufacture August 2007 Complies with 21 CFR 1040 10 and 1040 11 except for deviations pursant to Laser Notice No 50 dated June 24 2007 Sirona Dental Systems GmbH Fabrikstr 31 D 64625 Bensheim Germany Label Ma
36. y beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radiation indicator X ray lights up audible buzzer OK Deadman feature OK Power supply adequate kV Verification is OK Tube current is within specified limits Specified exposure time OK X ray beam position panoramic OK X ray beam position ceph OK PTO UU The unit is in compliance with MFG specified tests and safety Technician Dealer D3352 103 01 06 02 59 38 373 D3352 rona The Dental Company ORTHOPHOS XG Yearly Maintenance Checklist Customer Address Dealer Address Date of original installation Date of inspection Report of Assembly FD 2579 SCHEDULE Yes No Remarks Test instruments as required All manuals are present Manufacturer Model Accuracy Last calibrated Voltmeter Any mechanical damage noticed All labels are present and legible All indicator lights are OK Radia
37. zing measurements gt Read the voltage values on the Mult O Meter The value for the voltage displayed on the Mult O Meter must correspond to the 62 kV selected in the service routine The permissible tolerance is 10 Ifthe measured values do not fall within the permissible tolerance range replace the tube assembly see Service Manual Ifthe measured values fall within the permissible tolerance range finalize the measurement 6 2 Exposure Time Verification Analyzing measurements gt Read the radiation time on the Mult O Meter The value for the radiation time displayed on the Mult O Meter must correspond to the radiation time of 0 5s selected in the service routine The permissible tolerance is 10 Ifthe measured radiation time does not fall within the permissible tolerance replace the tube assembly see Service Manual Ifthe measured radiation time falls within the permissible tolerance finalize the measurement Finalizing the measurement gt Exit the service routine gt Switch off the unit see Operating Instructions 59 38 373 D3352 14 D3352 103 01 06 02 Sirona Dental Systems GmbH Maintenance Instructions ORTHOPHOS XG 7 Checking the X Ra 7 Checking the X Ray Beam 7 1 Diaphragm system adjustment menu Pan Sensor adjustment Pan Diaphragm Pan Filter Pan Symmetry Ceph Primary KIL s MM x 52 0 00 mm Dx s3 0 00mm Dy N I Precision adjustmen