Contents
1. Change the batteries Respect the polarity when you insert battery in battery holder Place the battery holder in the battery cover by pushing it gently Install the battery cover back in the housing Tighten the two 2 screws Only one type of battery is approved Duracell MN1500 Procell PC1500 27 10 5 Display unit replacement THE REPLACEMENT OF THE DISPLAY UNIT REQUIRES TO SETUP THE CONFIGURATION ACCORDING TO YOUR REQUIREMENTS CUSTOMER SPECIFIC SETTINGS FOR MEASUREMENT SCALE AND PRESET 1 AND 2 10 5 1 Disconnecting the Display unit Warning This operation must be done only in a non hazardous area e Switch off the unit e Install the Quick Cap e Unscrew the two 2 screws A of the battery holder using the 2 5mm Hex Allen key located in the carrying case See Figure 10 1 Figure 10 2 Pull it gently out Do not use knife or screwdriver to remove the battery cover in order to avoid damaging the gasket e Unscrew with the 2 5mm Hex Allen key the five 5 screws B of the display unit and pull it gently out of the main housing as shown on Figure 10 2 Disconnect the tape plug Disconnect the motor cable plug Disconnect the coder cable plug by pressing on the locking lever 10 5 2 Connecting the Display unit Connect the coder cable plug Connect the motor cable plug Connect the tape plug Put back the display unit in the main housing tighten the five 5 screws B of Figure 10 2 Reinstall th
2. ISBN It is anticipated that the user will have specific 1 85609 291 7 Fifth Edition 2006 or consult the operating methods laid down to ensure safety when appropriate Legislative Authority for the installation using this type of apparatus In this case the user s by means of the grounding cable and clamp instructions shall be strictly observed 7 Warning change of battery must be carried out in safe area only non flammable atmosphere Use only the approved model and brand Refer to the Care and Maintenance section of this manual 8 This product and his use is may be related to international national local or company regulations or standards It is the customer user responsibility to ensure that the way to use the device complies with such applicable regulations or standards 9 his device is a portable product It must not be permanently installed on the tank and must be disconnected after use and stored in a safe and dry area www tanksystem com 6 Functions Key Features This HERMetic instrument is a gas tight portable digital automatic gauging system designed to perform under closed conditions ullage monitoring and ullage measurements a Topping off monitoring The HERMetic TOMSYS monitors continuously the liquid level over the last three 3 meters from his reference height The liquid level is monitored continuously on the electronic display without any manual operation b Ullage monitoring measurement
3. Distance measurement range 3 6 m up to 12 feet approx 0 1 mm 0 5 mm 1mm or 5 mm 1 64 1 32 1716 1 8 selectable Distance resolution Distance measurement accuracy 2 mm 0 08 approx Hazardous areas under normal atmospheric conditions 80 kPa to 110 kPa 0 1 to 0 2 bar 1 5 psi to 2 9 psi Maximum Tank Over pressure Ambient temperature range 20 C to 50 C 4 F to 122 F Process Fluid temperature range 10 C to 70 C 14 F to 158 F Gauging is done under closed gas tight conditions therefore maintaining over or under pressure in the tank The device is approved for atmospheric pressure conditions Honeywell TT Z Ai ST ER N Tank top NV 74 E b rn ooi i 77177171 VAASZE ESEDSESESESESESESESEDZAESAASASEZASZSAAVAAAAAASAAMMM Ullage level Product Figure 6 1 11 7 Description 7 1 General Each HERMetic instrument is individually identified with a 6 digits serial number starting with the letter T example 110025 This serial number is printed on the identification plate as shown on Figure 7 1 Identification plate Figure 7 1 The unit displays distance Ullage measurement combined with sounds according to level measured value The unit emits and displays Control beep e Liquid level is outside measurement range Waiting RC QUIT Between gt 4 523 and gt 4 534 depending on the settings Double beep ee os ee Liquid is detected distance
4. oo ISSUE 1 03 08 2011 re ro 6 30 100 500 1090 Eed 9 386 7 Eff Em MPSA Drawn Control 1 1 Replacement for Replaced by CPI 03082011 Valves TS 10414 Blind cover assy This drawing is our property and must not without our Enraf Tanksystem SA RUE DE L INDUSTRIE 2 CH 1630 BULLE permission be copied or made available to others The receiver is responsible for every misuse Tel 41 26 91 91 500 Fax 41 26 91 91 505 www tanksystem com 15 Declaration of conformity Honeywell Honeywell Declaration of Conformity Issue 1 TSB_7028_E doc Enraf Tanksystem SA Author QD C E JEGEN Apparatus Identification HERMetic TOMSYS May 25 2011 1of1 Apparatus Classification Measurement Equipment Statement of Conformity Based on sample product test results using appropriate standards industrial environment and in accordance with the following EC Directives amp IEC Standards we Enraf Tanksystem SA hereby declare under our sole responsibility that the HERMetic TOMSYS is in conformity with EC ATEX Directive 94 9 EC Equipment and protective systems intended for use in potentially explosive atmospheres ATEX EC Type Examination Certificate KEMA 10ATEX0091X I 1 G Ex ia IIB T4 Ga 20 C lt T lt 50 C maximum fluid process temperature 70 C EC Directive 89 336 EEC Electromagnetic Compatibility EMC IEC 60079 Electrical apparatus for explosive gas atmospheres IECEx Type Examination Certif
5. 13 7 2 TOMSYS sensing probe The TOMSYS sensing probe consists of a stainless steel tube terminated by a high tech plastic concentrator The sensing probe includes an ultrasonic liquid level transducer The Probe body sensitivity for ullage measurement is not adjustable Concentrator Figure 7 3 7 3 Tape e t contains 2 wires for transmitting the signal and the power between the display unit and the probe The steel tape itself The ETFE TEFZEL coated tape provides 2 main functions is used as a grounding wire between the sensing probe tube e t holds the sensing probe and the display unit TEFZEL STEEL TAPE WIRE Figure 7 4 14 www tanksystem com 7 4 Distance measurement HERMetic UTImeter Zero ullage Liquid level ri Heference level of tank Honeywell HERMetic TOMSYS G CO ER 3 6 m max distance Heaction point Figure 7 5 The distance reading at the height of the reading index of the instrument is indicating the distance between the reaction point and the virtual index If the instrument is installed in such a way that the virtual index is at the same level as the zero ullage reference level the distance reading corresponds to the ullage providing the reaction point of the sensing probe is positioned at the liquid level The distance reading corresponds to the portable HERMetic UTlmeter tape reading If the virtual reading index is positione
6. 13 2 List of parts descriptions TS number 15004 15021 15024 15025 15028 15033 15034 15044 15047 15050 15059 15063 15070 15075 15085 20537 20541 37155 31373 rara 40004 40301 40302 40308 40309 40754 40777 50714 Magnet holder Battery cover assy Reference washer Display unit assy e O Alkaline Battery AA MN1500 Socket head cap screw M4x 8 A Socket head cap screw M4x10 Socket head cap screw M4x20 Socket head cap screw M4x14 Socket button head cap screw M3x10 o MES S c S Socket button head cap screw M3x6 Carrying case TOMSYS 13 3 Spare parts drawings The next pages show the following drawings e general assembly list of the main spare parts e battery holder assembly details e display unit assembly details e OMSYS sensor assembly TS 15050 details e Cap G1 4 details 33 TS 15034 TS 15033 TS 1 TS 15047 TS 15075 S 15063 gt e TASOS n TS 15044 e TS 40309 co 4x co co TS 40302 TS 40004 6x 2X TS 40308 2x TS 15021 TS 15025 TS 37155 TS 37155 d TS 15025 v TS 40308 cati TS 20541 a 71840004 2x TS 15085 Figure 13 1 general assembly list of the main spare parts 34 www tanksystem com Honeywell TS 37374 o Cam P l 9 TS 40801 2X Figure 13 2 battery holder assembly details iB EN HO S Honeywell N m Er TS 15059 TS 40777 D 8 4x D F
7. DW9y Z a dl o po 19T L6 0 ON CEO AJDA j20duo 31190 ke s iiv mm LI 9 001 SI sanoa ES LE ees oes LE saroy L vOEISIV 0 SO v iv WE v661 v0 LC dd estoy Jeu sa dz se x amp tz 1 Bupped ways TG fui fe a DD DI ug mi add KES pes 0 z jz mee emer q paso day 101 JUaWwaoD day C D EF Olly 3903 dNVHS ONY SYYNG T1V 3A0N33 95922 8077 awaz z poglo Ji i veal SEKR EIER EAR ENT RR EES a poto si POW EE EE m 49 06 EURO poe e 6661 9 GG JNSSI ooo oc o oc 9 rro ezis woo 0311193dS JSIMYSHLO SS31N S39NV33101 Rat 346 19M vor EN EE de gt sauran uum sano foes n Fa Fs een IER Mesa 9 usn cl M UC Ge E SS TA D SR 4 IW NN lt SN DET C E KE IFGGI NS N S d KR T d d pl d AA IV www tanksystem com 44 Honeywell GOS 16 16 92 Ly XJ 006 16 16 92 Lp 191 ggnsiu 1949 10 a qISUO saJ SI 49419991 ay la fr 96607 8922 70 SIV SET uo sige Joy vaysem O IL eb 31108 OS91 H0 Z 3IMISDONI 30 INA te fenor ire did azar aseo o 1 a Ge ee S E suo ove zia SZ x 06 98 yse9 0 Lb t WS 19 S SYUD oJU Ino noy im jou jsnu puo ysadoJd sno s BuIMDAp eu ed e le suo ez SEI EI t ET ZT oor sz serrr lo t fe am 9 DW9y cke gov puyfo t 8 Gaor seez 3a v
8. For battery replacement please refer to section 10 3 9 4 2 Level monitor and Ullage detector Install the unit on a table desk or workbench according to fig 9 3 hereunder Make sure the target surface is of solid flat and smooth material floor hardpaper steel plate Table workbench Target surface Figure 9 3 Refer to section 9 2 1 or 9 2 2 accordingly for function tests The Display shows you the distance H between table plate and the target 162 5 mm 2 mm 6 Inch 25 64 5 64 corresponding to the distance between the Virtual internal reading index and the target 26 www tanksystem com 9 5 Installation of the unit This HERMetic equipment must be coupled to a certified HERMetic valve Before starting please read carefully the chapter Recommendation for safe use and follow your company s safety instructions Before removing the valve cover check that the HERMetic valve is closed Remove the valve cover weather cap blind cover security cover of the HERMetic valve Clean the seal surfaces of the nipple of the valve and of the coupler of the instrument from dust or grease Note Cleaning of the mating surfaces is very important for earth grounding purpose and for good accuracy on zero reference level Install the HERMetic TOMSYS on top of the valve by means of the quick coupler Ensure that the equipment is properly earthed If not ground it with the grounding cable before oper
9. If gaskets are used dimension HX is reduced by thickness of gasket 19 9 Operation 9 1 Basic rules concerning the 5 key control pad Those three keys are menu oriented and have functions depending on which menu they are applied Before operating the unit please read carefully this Operation ice M l and Service Manua Soft keys main use Apart from the ONT OFF keys that are self explanatory there ap key I moves the cursor DOWN th th ft are three 3 other soft keys I Il and Ill that help in managing all ket ad Boa thetighlighted function the unit functions key IIl return to previous menu or QUIT selected function HERMetic TOMSYS Portable Digital Topping off Monitoring System 1629 2 t arias EE ER RE Use only approved battery type Duracell Procell MN1500 p Cursor DOWN Select Return or Quit Figure 9 1 20 www tanksystem com 9 2 Quick start 9 2 1 Level monitoring a Press ON key to switch on the unit After displaying some information about software version and setup display shows Level Monitor Ullage Detect Preset Disable service Menu DOWN See b If necessary move cursor by pressing soft key I to Level Monitor C Then press soft key Il Select to activate the function Unit starts by resetting the distance measurement Then sensor moves down and stop when liquid is detected or moves to lower position waiting
10. The storage tube is equipped with a quick connector which fits on the HERMetic valves The sensor goes down or up until the floor detection in our exemple up to 2 152 3mm B Press Key Up 1st mode By pressing Key the sensor lifts up one step by about 50mm The measure displayed is still the same the system compensates automatically the sensor displacement 4 Press Key Il Auto mode By pressing Key Il the sensor lifts up by step of about 50mm with 15 20 seconds break until the target is no more detected The Auto mode allows to simulate the raising of the liquid level This mode simulates the actual operating of TOMSYS during the Topping Off monitoring process oy Press Key IIl Quit Return Mode By pressing Key Ill the sensor lift up up to its park position in the storage tube Note Make sure the target surface is of solid flat and smooth material floor rigid paper sheet steel plate but not a carpet or liquid Honeywell 8 Examples of installation of the gauging system 8 1 General The gauging system consists of the HERMetic instrument and the associated HERMetic valve The TOMSYS is fitted with a 2 connector as shown on Figure 8 1 Q2 2 with Quick Cap Figure 8 1 The following sections describe 2 examples for installing the valves and adjusting the height of the gauging system The valves should be installed in such a way that the zero ullage level coi
11. This function allows you to check the remaining battery capacity at any time The battery capacity is expressed in of capacity and voltage From Main Menu follow the instructions according to flowchart hereunder Level Monitor Ullage Detect Preset DOWN Select EL Sensor Parking Settings Battery Status Position DOWN Honeywell Display shows the remaining battery capacity in 95 40 Val 8 90V ADC 3 29V 9 3 3 Sensor Parking This function forces sensor to be lifted up to park position From Main Menu follows the instructions according to following flowchart Level Monitor Ullage Detect Preset ee DOWN Select ettings Battery Status Position DOWN Selecting sensor parking will return the sensor into the park position and display the distance ascending values during the sensor return Display automatically switches to main menu after park position is reached Ullage Detect Preset Disable Service Menu DOWN Select 25 9 4 Checking the functions before using the unit Before installing the HERMetic TOMSYS unit as described in section 9 5 the following tests are recommended to ensure that the unit is ready to work 9 4 1 Battery Refer to section 9 3 2 Battery status If battery capacity is lower than 20 please replace the battery if you plan to use the unit for several hours without disruption
12. c wje oo pon CN 8 5r CN SE 7 co CL GY Q D E RW WY Y S Dees SE gt S gt NI Kai GA VAN LEE AS y N u E 46 www tanksystem com Honeywell O o 3 WA WV VA m LE Ee LA gt Weight Description 222 4 Gasket holder Cover Lock G3 e Slotted pan head screw 5 1 53 Gasket CS 1 44 S Holder L7 1 08 Short cable assy TOLERANCES UNLESS OTHERWISE SPECIFIED Weight Norm Sizel Over 30 100 300 1000 emer Sef EE pa ISSUE 3 03 08 2011 Fine Fine Drawn Control 1 1 Replacement for Replaced by CPi 03 08 2011 OOOO Valves TS 10408 security cover assembly ND 40495 This drawing is our property and must not without our Enraf Tanksystem SA RUE DE L INDUSTRIE 2 CH 1630 BULLE permission be copied or made available to others The receiver is responsible for every misuse Tel 41 26 91 91 500 Fax 41 26 91 91 505 47 48 j URINE ee SUN ER Iw Nano Kass E EH li S NES Fa D Description Connector 1 2 Weather cap assy 3 1 44lS Holder 4 1 0 8 Short cable assy C2 22863 41032 22608 40543 TOLERANCES UNLESS OTHERWISE SPECIFIED Weight FNomswe Ove 6 3o 100 soo ooo P are ISSUE 1 03 08 2011 a Eq Te Ee zoo re 586 7 Eff REMOVE ALL BURRS AND SHARP EDGES Drawn Control 1 1 CPI 03 08
13. o bo 7 7 HOUSING AND UD mr A A E oo 1 9 DEMO MODE FUNCTION 8 EXAMPLES OF INSTALLATION OF THE GAUGING SYSTEM css e seene RE mI E 8 2 EXAMPLE OF INSTALLATION ON A PIPE CONNECTOR 107 8 38 EXAMPLE OF INSTALLATION ON THE DECK OGONNEC TOR ce 9 XOPERATIOD zi sese s seene 9 1 BASIC RULES CONCERNING THE 5 KEY CONTROL PAD E 92 QUICK SIAR EE 9 2 1 Level monitoring esee 9 2 2 Ullage detection sees Her AE CS RS ibi do SERVIOE MENO E AA 9 3 1 1 Distance measurement scale 9 3 1 2 Distance measurement resolution 11 12 13 14 15 Honeywell 9 3 1 0 Menu Language s 25 dos Jte satus i eset dattia n Oa HERES 29 SEES NEC ears t 25 9 4 CHECKING THE FUNCTIONS BEFORE USING pis r casetineoratetutcent 26 SE CHE a aE ATAA EA AA EO 26 9 4 2 Level monitor and Ullage detector 26 9 6 INSTALLATION OF THE UNIT ce orta etas 26 9 6 LEVEL MONITORING serra eos 26 9f JULEAGE DETECTION censos 26 9 3 TRANSPORTATION scream 26 CARE AND MAINTENANCE eene nnne 27 10 1 SAFETY WARININ E 2f 10 CARE ce 27 10 3 CHECKING THE BATTERY EE 2f 10 4 BATTERY REPLACEMENT emm 2f 10 5 DISPLAY UNIT REPLACEMENT genge nn 28 10 5 1 Disconnecting the display unit 28 10 5 2 Connecting the display unit 28 10 6 DISTANC
14. on liquid level detection Display shows following for a liquid detected at ullage level 215283 m 2152 3 Battery 19 e QUIT Or in Inch scale 7 0 Battery BB 19 Meng QUIT Honeywell 9 2 2 Ullage detection a Press ON key to switch on the unit After displaying some information about software version and setup Display shows Ullage Detect Preset Disable Service Menu DOWN Select b If necessary move cursor by pressing soft key I to Ullage Detector L evel Monitor Ullage Detect Preset Disable Service Menu DOWN Select C Then press soft key II Select to activate the function Unit starts by resetting the distance measurement Then sensor moves down and stops when liquid is detected The displayed distance corresponds to ULLAGE LEVEL Display shows Stand By Meng QUIT 21 This measured value has the same value as ULLAGE LEVEL measured by UTI or UTImeter If there is no detectable liquid sensor moves to lower position waiting on liquid level detection Display shows 24 52X Waiting Menu LT 9 2 3 Preset TOMSYS allows you the possibility to setup two 2 preset liquid levels Those two levels can be used to draw the user s attention when liquid level reaches one of those levels Preset level values are ULLAGE level values in mm When liquid level reaches PRESET level 1 buzzer sounds a
15. on the display Accelerated double beep ee ee ee oo oo Liquid level reaches Preset Level 1 distance on display with Preset 1 message 12 www tanksystem com Fast double beep ee ee ee ee ee ee ee Liquid level reaches Preset Level 2 or Ullage level detected in Ullage mode distance on display with Preset 2 or Ullage message Continuous ton Error or malfunctioning unit combined with error message No message No reference Too close Low battery Tilt error Sensor jam Control beep o o e Double beep ee T ee Accelerated double beep ee ee oo oo oo Fast double beep ee oo oo oo oe ee oo Continuous beep 0000000000000000 Continuous tone An automatic backlight manages LCD Display contrast according to ambient light The HERMetic TOMSYS instrument is powered by six 6 AA cells stored in a battery holder located in the main housing Battery status is displayed on the screen When battery voltage is too low buzzer emits continuous tone and display shows a battery error message Every time it is possible to check battery status Battery must be replaced by approved battery in safe area only Maintenance is easy because design is modular and allows quick exchange of parts See Figure 7 2 to get to know the equipment Honeywell Mechanical housing Display unit LCD Display LED Buzzer Handle keys MI Storage tube SS2 Li Quick connector Q2 Tape Sensing probe TOMSYS Figure 7 2
16. or workmanship The Vendor s obligation is limited to the repair or replacement of such defective parts by his own plant or one of his authorized service stations The Purchaser shall bear the cost and risk of transportation of defective parts and repaired parts supplied in replacement of such defective parts 6 www tanksystem com When returned to Enraf Tanksystem SA or any of its agreed Service Stations equipment must be contamination free If it is determined that the Purchasers equipment is contaminated it will be returned to the Purchaser at the Purchasers expense Contaminated equipment will not be repaired replaced or covered under any warranty until such time that the said equipment is decontaminated by the Purchaser The Purchaser shall notify by fax email or in writing of any defect immediately upon discovery specifying the nature of the defect and or the extend of the damage caused thereby Where no other conditions have been negotiated between the Vendor and the Purchaser General Conditions 188 of United Nations shall apply This instrument has been certified as Intrinsically Safe Instrumentation for only those classes or categories of hazardous areas stated on the instrument label bearing the mark of the applicable approval authority No other usage is authorized Unauthorized repair or component replacement by the Purchaser will void this guarantee and may impair the intrinsic safety of the instrument In par
17. 2011 HERMetic Deck Valves C2SS Cover with weather cap This drawing is our property and must not without our permission be copied or made available to others The receiver is responsible for every misuse www tanksystem com 9 MPSA YYYN Replacement for Replaced by TS 10415 ND 41040 REF ND 30391 20291 Enraf Tanksystem SA RUE DE L INDUSTRIE 2 CH 1630 BULLE Tel 41 26 91 91 500 Fax 41 26 91 91 505 Honeywell COUPE A A lea A Description Matera TS ND CT 1 1605 apfornippel 2 1 38 Clp AISI 301 14310 40762 40542 3 T1 10 Ring e 29 7 x 35 FKM 1132 4 1 1 Cable assy TOLERANCES UNLESS OTHERWISE SPECIFIED Weight Norm Size Over 30 100 300 1000 Exa Ee E 05 2006 Fine 0 05 01 015 02 03 05 01 KT REMOVE ALL BURRS AND SHARP EDGES E Drawn Control Replacement for Le CPI 19 05 2009 mem Valves TS 22609 Weather cap assy ND 40543 This drawing is our property and must not without our Enraf Tanksystem SA permission be copied or made available to others eee l RUE DE L INDUSTRIE 2 CH 1630 BULLE The receiver is responsible for every misuse Tel 41 26 91 91 500 Fax 41 26 91 91 505 49 90 Verst Weight Description Mana TS ND 1 1 3762 Bind cover Gasket 3 1 44 S Holder L4 4 TD 8 Short cable assy TOLERANCES UNLESS OTHERWISE SPECIFIED Weight FNomswe ove 6 30 100 soo
18. E V RIFICATION she 28 IU E He PAR 0 28 10 8 CALIBRATION RECOMMENDATIONS 28 TROUBLE SHOOTINO iii nn nn 29 11 4 SAFETY WARNING cesiones Dorf REI a uet 29 11 2 EMERGENCY CRANK eee 29 11 3 POWER SUPPLY TROUBLES 29 11 4 TRANSMISSION TROUBLES ern 29 11 5 SENSOR MOVING TROUBLES 30 11 5 1 Moving trouble on the way down 30 11 5 2 Moving trouble on the way up 30 11 6 DISTANCE MEASUREMENT TROUBLES 30 11 7 SENSOR ERRORS an szg a rene 30 ME lt Tee lt EN 11 9 VISUAL INSPECTION FOR DAMAGED OR MISSING Wisdom 31 SPECIFICATIONS siestas 32 SPARE PARTS m neac 33 131 HOW TO PROCEED e Se tnm 33 13 2 UST OF PARTS DESCRIPTIONS 33 13 3 SPARE PARTS DRAWINGS nn 33 VALVES DRAWINGS 0occcocccccococccocacononanononanononanenananenanans 39 14 1 VALVES DRAWINGS LIST siens 39 ERT EE 39 DECLARATION OF CONFORMIT Y 51 2 Identification of your equipment 2 1 Serial number TE D Identification plate Each HERMetic instrument is individually identified with a 6 digits serial number starting with the letter T example 110025 This serial number is printed on the identification plate that is located on top of the device See Figure 2 1 2 2 Abbreviations Some abbreviations are used to define the equipment Refer to following table and to Figure 2 2 Abbreviation ETCE Special PTFE used for ext
19. Honeywell Operation and Service Manual for HERMetic TOMSYS for use in non corrosive liquids Portable digital automatic gas tight Topping off monitoring system HERMetic TOMSYS Portable Digital Topping off Monitoring System E 9 Certified ISO 9001 Company aqwrsai Global Certified System Note 1 to identify the unit refer to section 2 Note 2 before using the instrument please read this book Note 3 this document is subject to changes without notice Check updates on www tanksystem com or contact us at tanksystemOhoneywell com This page intentionally left blank www tanksystem com 1 Table of contents T TABLE OF CONTENTS 2 2 IDENTIFICATION OF YOUR EQUIPMENT 2 1 SERIAL NUMBER seiten dote bea Spa tek pu teneros 22 JqQABBHEVIATIONS us ecpun otra su simon tees rA Enni 3 GENERAL INFORMATION creen Sl el Cu CA NOTE sersan a Ou INMALINSPESTION EE 3 3 DOCUMENTATION DISCREPANCIES ANR es So DERF TON baratas EX COMICS armada 3 5 2 Classification society certificate o O rania Of SERVICE AND REPAIR casas od 4 WORLDWIDE SERVICE STATIONS NETWORK 5 RECOMMENDATION FOR SAFE USE 6 FUNCTIONS KEY FEATURES 7 DESCRIPTION me e Ee Sv EE 2 TOMSYS SENSING PROBE hs Co erc 7 4 DISTANCE MEASUREMENT eee 5 9 o Ee H Se ere rene Ter errr eee LO C
20. N K LYUL SI 13 BAJDA FFPEPPEPPE e L subiam 40 Wal CSO a lt P SS KY AN SR Si ES EH EE A ZA ZE OS 3x 1 lt lt www tanksystem com 40 Honeywell BAJEA UO 3199 JOY JauseA 0 LX 6L l 9 494SeD 0 GZ X 06 98 Y J9NSeD SOS 16 16 92 Lt x03 005 16 16 92 Lt 151 asnsiu ag 101 ajqisuodsas SI 19419291 ayy 31108 O 91 H9 Z 3IYISNONI 1 30 208 S134 0 0 2 Q0 DAD apow Jo pardos aq uoissiuJad YS uajs syup D1U Ino jnogj 4 jou jsnu puo ijedojd ino s Buimoip SIY NID neg 3d 70 lS 3 pueH YOE IS Jaysem Buluds 18207 0N 939 SS Z 9 DA J90dUO 0119020 WEE 7 2800 Sl SOA DA TA sa dz amp 8 X 6 EZ LL GEES ON ON 1661 10 90 Idd 9661 11 LC dd j A C A MT 9 X 99 ES Jeas EF 011 3903 d4VHS ANY Sauna 11V 3A0M38 l FAO pod ean 0 cof eo zo sio io soto Uondised s P 3 330 0007 0001 00 1001 DC oj EI 6661791 z anssil es Lin Dos IA psa eim 0314193dS 3SIMAJHLO SSIINN S39NV33101 10911 1 POH ajo Si uo s6704 Jonn OA sao Amas ase L fz ESCH Jero TI 00 23 amen peduos josy i i Fu Fer ATC MOTE ba a ues amp SGIOGL ISNY 0S MOL SIf 0g HS SIN OSNG OrNd NIG OSNG SZNd NIQ OSNG 91Nd NIG S OSNG OLNd NIG de u
21. PL ee eee er gered A e daO E T 08407 sv9zz_ serri z nmeelo v el Josz s80vv1 o 1 1 Gee o rc in Ip da 0tlSIV apeo v SL ve9zz_ mvotlsiV 5 5 eusen BUS 0 v TI ee EE HNDL svo fesz oe isivf ve CS L NEU VOGL YO LL YdN Z eo ieszz 3414 se dz S SE E L ARR LE HR E iiis ose E TE gx sizes s i8 9 1 8 LLY 3903 duvHS ONY Sun 11V 3A0N3J doaz To eue L APO fi i GA TION GT 035 PONT si o eaen uondrJosau iube 10 wa 133 00 0002 f ooo oos foor fos 9f of 31 6661 9 S5 JNASSI u ooi oc oor oc 9 fo fers un 46194 0111193dS JSIMYSHLO SSIINA S39NV33101 sa 9 024 7 gt 18 SS 7 A1DA 120du0 21 15h21 SOA DA HE S x e m N N re ee Ki cn c Y LJ TU a a Wa NV 0y aan HT Asse 39803 puna ose fL b ON CSL HSE ER BAL 40 MEE 6l r 0c ELI Ss Ly a 01 y CL vl ya Qu B M 43 GOS 16 16 92 Let X04 006 16 16 92 Let 191 asnsiu Alana 10 ajqisuodsay S 1219991 34 la fr 96607 89922 oe Isiw ig ZO Se ede qm o fti FE T et 3liishONi 30 INA en onu IN Des au Lx en a sejo fu a S134 0 0 9 QD IDAD app Jo pardos aq uoissiuJad ols YS Wo S SYUD ID uy JNO NOY IM you Oh pun uadoid ino SI But apap SIU 84409 forz dd S7 x 06 98 p yolo k ph EEN sjajno fez oes is fo OL osor Je serri fo Te T6 Pl 9
22. ating optional TS 13051 grounding cable Equipment must be installed in vertical position A tilt detector prevent any tape movement if unit is not in vertical position 9 6 Level monitoring Refer to section 9 2 1 9 7 Ullage detection Refer to section 9 2 2 9 8 Transportation The TOMSYS is equipped with a specific handle enabling its transportation in horizontal position The transportation of the unit must be made in Off mode only Always move the device with the sensor parked in its storage tube and with its cap properly in place This is critical to avoid damaging the motorisation system 10 Care and Maintenance 10 1 Safety warning As this equipment is designed and approved for use in an explosive area intrinsic safe equipment only authorized service stations and the factory are allowed to repair electronic circuits However the customer can exchange parts and modules if the following points are observed 1 Never open the instrument nor carry out any repair or trouble shooting in an hazardous area 2 Use only original spare parts 3 Work shall be done only by maintenance personnel who has an experience with intrinsically safe equipment 10 2 Care Clean the instrument of any excess of liquid after each use Make sure that the sensing probe is completely stored in the storage tube after use Check that sensor is clean and dry for storage Store the instrument in a dry location Check periodical
23. ccelerated double beep Control beep o o e Double beep oo ee ee Accelerated double beep ee ee oo oo oo Fast double beep ee oo oo oo oo oo oo Continuous beep 0000000000000000 When liquid level reaches PRESET level 2 buzzer sounds fast double beep PRESET level function can be enabled or disabled 22 www tanksystem com To setup PRESET level do the following a Press ON key to switeh on the unit After displaying some information about software version and setup displays shows Jllage Detect Preset Disable Service Menu DOWN Select b Move the cursor by pressing soft key I to Preset Level Monitor Ullage Detect Disable E Jj ervice Menu DOWN Select c Then press soft key Il Select to activate the function Preset Menu Enable Deere Preset 2 0850 Preset 1 1500 DOWN Select EE d Enable or disable the Preset function by moving the cursor Down by pressing soft key I Press soft key II to Select Preset Menu Enable DOWN Select The setup for the two level values is described in the flowchart on Figure 9 2 Honeywell Flowchart to set up the two Preset level values Preset 1 is the lower ULLAGE level value Preset 2 is the upper ULLAGE level value Enable or Disable the PRESET function Preset Menu Enable ES Preset 2 0850 Preset 1 1500 DOWN Select T Preset M
24. d below or above the reference level a positive or negative correction of the tape reading is necessary oee also chapter 8 Examples of installation of the gauging system 15 7 5 Gas tightness All parts are assembled together with either gaskets or O rings that makes the device completely tight The sealing of the axle is ensured by a special V shape gasket 7 6 Gaskets Gaskets are made in FKM material 7 9 Demo Mode function The DEMO Mode is specifically dedicated to explain the operation of TOMSYS at end user office To test the device a horizontal flat and stable surface to place TOMSYS and a target surface made of solid flat and smooth material are required only a desk on a flat and smooth concrete tiles or linoleum for instance Demonstration Process La Place TOMSYS vertically on a desk put on its handle SA Switch on TOMSYS and select the menus below Level Monitor Ullage Detect Preset Disable DOWN Select ervice Menu Sensor Parking Settings 4 Battery Status Position DOWN D emo Control Menu DEMO Mode Control Mode DOWN 2152 3 Battery AN 19 tf LT 16 www tanksystem com 7 7 Housing and lid These parts are made in high tech antistatic plastic 7 8 Others The tape is coiled on a drum driven by an electrical motor or by the emergency manual crank The storage tube is screwed to the frame
25. e battery holder with the two 2 screws A of Figure 10 1 Tighten the screws with the 2 5mm Hex Allen key Refer to Figure 10 1 on page 27x Check that the unit is working properly as describe in 9 4 28 www tanksystem com 10 6 Distance verification The distance measured by the unit is a combination of two electronic measurements One measurement is done by the sensor and the other one by the main unit Refer to 9 4 2 check the distance measurement Se Table workbench Target surface H 162 5 mm 2mm Or H 6 Inch 25 64 5 64 Display shows Note Make sure the target surface is of solid flat and smooth material floor hardpaper steel plate 10 7 Plastic parts Use only light solvent to clean plastic parts Do not use acid to clean plastic parts 10 8 Calibration recommendation As for any measurment tools it is recommended to check and calibrate the HERMetic Tomsys against master periodically The period recommended should not exceed 1 year Honeywell 11 Trouble shooting 11 1 Safety warning 3 Work shall be done only by maintenance personnel who has an experience with intrinsically safe equipment As this equipment is designed and approved for use in an explosive area intrinsic safe equipment only authorized The design of the equipment is modular i e in case of service stations and the factory are allowed to repair electronic breakdown the customer can find out which modul
26. emperature range Product Range Display Reading LCD display Mechanical connection Probe diameter Weight Battery Hazardous environments approvals ECEx ATEX Factory Mutual FM Approval Specifications subject to change without notice 32 www tanksystem com 2 mm 0 08 approx 3 6 m up to 12 feet approx 0 1 0 5 1 0 5 mm selectable 1 64 1 32 1 16 1 8 selectable 80 kPa to 110 kPa 0 1 to 0 2 bar 1 5 psi to 2 9 psi 20 C to 50 C 4 F to 122 F 10 C to 70 C 14 F to 158 F Crude oil non corrosive petroleum products mm or Inch selectable Graphic 128 x 64 dot with automatic backlight 02127 43 mm 1 77 8 1 kg Only one type of battery is approved Duracell MN1500 Procell PC1500 Ex ia IIB T4 Ga 20 C lt Ta lt 50 C Il 1 G Ex ia IIB T4 Ga 20 lt Ia 500 CLI DIV 1 GP C amp D Tamo 50 C CLI ZN O AEx ia IIB T4 Tamb 50 C Honeywell 13 Spare parts 13 1 How to proceed 2 Note the item TS number ex TS 20541 3 With the assistance of the below table identify its description Each spare part is identified by the letters TS followed by a 5 ex O Ring 0 56 74 x 3 58 digits number as for instance For each order please note the item number its description and Proceed as follows to identify the part you need to order the required quantity 1 Find the adequate drawing on the next pages Example TS 20541 O Ring o 56 74 x 3 53 1x
27. enu Enable Disable Preset 2 0850 Preset 1 1500 DONT Selec Uc Preset 1 Preset 1 ce AND 2 Preset 2 8 50 LLL Preset 2 OH AND EA Preset 1 and 2 position Preset 2 Preset 1 Figure 9 2 23 9 3 Service Menu 9 3 1 Settings This function allows you setting Up Distance Measurement Scale Distance Measurement Resolution Menu Language 9 3 1 1 Distance Measurement Scale To setup measurement scale in Metric or Imperial units follow the instructions according to the flowchart hereunder Level Monitor Ullage Detect Preset EZ DOWN Select Sensor Parkin 0 Battery SEE Position DOWN Se Devices DOWN inch DOWN 24 www tanksystem com The distance will be displayed accordingly In Metric scale 2152 3 Battery MM 19 Meng QUIT Or in Inch scale 7 0 Battery MM 19 Meng QUIT 9 3 1 2 Distance Measurement Resolution To setup measurement unit resolution follow the instructions according to flowchart hereunder Level Monitor Ullage Detect Preset DOWN NN Sensor Parkin 0 E as Position DOWN Resolution Dist Scale Language Devices DOWN SETTINGS Resolution 0 1 mm 1 mm 5 mm 10 mm DOWN Or in Inch resolution Resolution DOWN 9 3 1 3 Menu Language Only English language is available 9 3 2 Battery status
28. es have to be circuits replaced The instrument consists of the following modules However the customer can exchange parts and modules if the e Mechanical parts following points are observed e Sensing probe e Display unit 1 Never open the instrument nor carry out any repair or trouble shooting in an hazardous area 2 Use only original spare parts The following sections should allow to identify the defective module and to replace it 11 2 Emergency crank The emergency crank aims to rewind the tape manually Turn clockwise to rewind the tape and lift up the sensor Never force on the crank if you feel some resistance Use emergency crank only if equipment is installed in a vertical position or on a 2 valve or 2 coupling device The emergency crank must be fully inserted till the end in order to come up against the guide entry Doing so will permit proper function of the crank See below section where appropriate 11 3 Power supply troubles mer 9m an See Battery too low Change the battery The unit switches on but stops on the message battery the buzzer tones continuously 11 4 Transmission troubles Sensor out of work or Send the unit to factory Ke No Msg is displayed mut Tape out of work Send the unit to factory DE S E N SO R N O MSG Error from Sensor Sensor Parking 29 11 5 Sensor moving troubles 11 5 1 Moving trouble on the way down The sensor can get blocked during d
29. escent at valve crossing due to a mechanical obstacle If this happens the sensor will automatically get back in the top parking position and following message will be displayed SENSOR Valve Closed Error from Sensor Sensor Parking Check if the connection valve is fully open and no obstacle reduces the 2 Inch full bore opening 11 5 2 Moving trouble on the way up In case of sensor movement trouble low battery or even when sensor is blocked into the tank use Emergency Crank to move down or to lift up the sensor to release it from its blocking position This emergency crank must only be used or activated if the equipment is installed in vertical position Sensor BLOCKED Turn OFF equipment and move MANUALY 30 www tanksystem com 11 6 Distance measurement troubles Distance verification out of tolerance Refer to section 10 5 Check the handle position The handle must be installed according to below figure and distance A verified Handle position 53 mm 1mm 2 Inch 5 64 3 64 Distance 11 7 Sensor errors Sensor No reference SENSOR NO REF Error from Sensor Sensor Parking a Check the inside of the concentrator end If necessary clean and dry it b If concentrator end is damaged it must be replaced Inside concentrator end Honeywell Visual inspection for damaged or missing parts Sensor Too Close 11 9 General condition missing parts SENSOR Too Close Disp
30. i obuD 4 UO S I SAJDA T C Y NS EN 6 LIZ SZ N RAN Y 9 re GY AN 7 AN S 4 y O NS 8606 UN FLVOL SI 41 BAJEA UO 3199 JOY JauseA 0 LX 6L l 9 494SeD 0 RC X 06 98 Y J9NseD 0 SOG 16 16 92 Lyt X53 008 16 16 92 Lt 181 asnsiw Alana 10 a qisuodsaJ SI 19419991 31114 0 91 H9 Z 31Y1SNONI 1 30 3n4 d yaa U S134 0 0 9 qD 1DAD opbu Jo pardos aq uoissiuJad VS ua s syup JD1U Ino noy im jou jsnu pup Kjjedojd ino st bunn styl fn abuni 7 88702 ON 1g SS 79 PAJDA 9DdW09 911943 sa2dz 58 x 6EZ L Bulyzed wa 9 X 99 S Jeas rng Apog 00110911 IPO aq Si uoijdiJ2sa veoly Jan egzoz fervor fna z2 amen pedwo 0955 T fi f 0N Zeil NPW lubam 10 ua C SQIOSL ISNY 06 AOL SIP 06 AG Sif OGNd O7Nd NIC OSNA GONd NIC OGNd 9INd NIC ns OSNG OINd NIG ar OH NZ S a 7 LAA Y IVA ZA ZA j SS 8606 UN K LYUL Sl www tanksystem com 42 R aR TT RTT m i EREECHEN 31108 0 9 H9 Z 31YISNONI 1 30 INA Wanni A n n EI 6109 TO 331d Dx 6174 pese 0 T f 4194 0 0 2 qD IDAD apow Jo pardos aq uoissiuJad eos SSC SCH Sz x 06798 s 1452910
31. icate IECEx KEM 10 0042X Ex ia IIB T4 Ga Sample Product Testing for ATEX IECEx Tested by Kema Quality B V Utrechtseweg 310 P O Box 5185 6812 AR Arnhem The Netherlands Notified Body Kema Quality B V Utrechtseweg 310 P O Box 5185 6812 AR Arnhem The Netherlands Notified Body Number 0344 ATEX Standards Used EN 60079 0 2009 Explosive atmospheres Part 0 Equipment General requirements EN 60079 11 2007 Explosive atmospheres Part 11 Equipment protection by intrinsic safety i EN 60079 26 2007 Explosive atmospheres Part 26 Equipment with equipment protection Report ID NL KEM EXTR10 0040 01 IECEx Standards Used IEC 60079 0 ed 5 0 2007 10 Explosive atmospheres Part 0 Equipment General requirements IEC 60079 11 ed 5 0 2006 07 Explosive atmospheres Part 11 Equipment protection by intrinsic safety i IEC 60079 26 ed 2 0 2006 08 Explosive atmospheres Part 26 Equipment with equipment protection Report ID NL KEM EXTR10 0040 01 Quality Assurance notification amp IECEx Quality Assessment Report Baseefa ATEX 1536 Notified Body Baseefa Rockhead Business Park Staden Lane Buxton Derbyshire SK17 9RZ United Kingdom Notified Body Number 1180 Sample Product Testing for EMC Tested by Montena EMC SA Z ne industrielle 1728 Rossens Switzerland Standards Used EN61326 1 2006 05 Electrical equipment for measurement control and laboratory use EMC requirements General requirements Germanischer L
32. igure 13 4 TOMSYS sensor assembly TS 15050 details 37 www tanksystem com Figure 13 5 Cap G1 4 details TS 15004 Honeywell 14 Valves drawings 14 1 Valves drawings list Refer to the table and find the drawing in next section Seege AI A Deck valve A 4 2 1 SS 14 2 Drawings See next pages 39 GOS 16 16 92 Let XDJ 00G 16 16 92 Lt lal 31108 0 91 H9 Z 31Y1SNONI 1 30 3NY esnsiu Asana 10 ajqisuodsay SI 19419991 ayy 184 0 0 9 QD IDAD apow Jo pardos aq 0015510190 Jno noy im jou jsnu pup AyJadodd sno s App su ena abuo 7 ET ON M SS 79 AJDA 90dw09 2119020 8001 Sl UN QN 4q papi da J0j JUuawaod day ys uajs syUD JDJUJ cl EIE OLLE VSdh 6661291 z 30SSI 0 sol eo zo cio o coo eun E 0001 ooe 00 Oc 9 190 ezis ujoN 031419345 ISIMYFHLO SSI1NN S39NV33101 xor sor dey sayjesa uyy sano 065 L z eeo eo rn z3 anen peduoz fosy e 1 CF sin REECH ora a usa BAJEA UO 3199 JOY JIYSEM L X 61 11 49ASEL SZ X 06 98 199560 Ways TIN NIQ neg de pug dd 70EIS ajpuen YOE IS Jaysem Bulsds jerJa 148p G 541081 06 AO 0G YS 9 x 99 ES Je2S rng Apog U0 4d11383 SNV SIP SIP OSNA OvNd NIG OSNA SCNa NIG OSNO 9 Nd NIC OGNd OLNd NIC abunit uo S 8606 U
33. laging is permitted provided that the conductivity of 50 picoSiemens metre pS m or less instrument is properly grounded and earthed before the probe is inserted into the tank and It is very important that the instrument is grounded to SE 9 remains earthed until the probe has been the tank before the probe is introduced into the tank removed from the tank and remains grounded until after complete withdrawal from the tank 6 2 2 Ifthe cargo is a static accumulator liquid i e its conductivity is less than 50 pS m then ullaging 4 1 If the instrument is installed with the genuine is permitted provided that Tanksystem quick connect coupler grounding is 6 2 2 1 The instrument is properly grounded and effected through the quick connect coupler and the earthed before the probe is inserted into mating nipple of the valve provided that these parts EE EN are kept clean and free from corrosion in order to probe has been removed from the tank guarantee electrical conductivity If a grease is used 6 2 2 2 The apparatus is not introduced into a tank for this purpose it must be one which contains until at least 30 minutes have elapsed after graphite completion of any loading operation or 4 2 lf the instrument is not connected to the mating stopping the injection of inert gas deck valve the instrument has to be also earthed 6 3 For further guidance refer to International Safety Guide for Oil Tankers and Terminals ISGOT T
34. lay unit 5 key control pad buzzer front face Error from Sensor screen Sensor Pa rking Sensing probe sensors broken smashed or damaged Concentrator Clean and dry not damaged Mechanical parts check housing handle storage tube quick Target must be foam free liquid or horizontal unruffled flat couplerr surface Sensor blocked Sensor BLOCKED Turn OFF equipment and move MANUALY 11 10 OFF message PLEASE Close the Valve Block Sensor Use Emergency Crank to move down and to lift up the sensor in with the CAP order to release it from its blocking position The sensor must be fully lift up manually in its parking position This emergency crank must only be used or activated if equipment is installed in vertical position 11 8 Tilt error Please install the Quick Cap when you Install the unit remove the equipment from the connection to switch OFF the equipment completely on a valve or in vertical position The display shows this message if equipment is not installed in a fully vertical position A tilt detector prevents any tape movement in order to avoid any damage Install the equipment in a vertical position as described in section 8 31 12 Specifications General Specifications Accuracy of distance measurement Distance measurement range Distance resolution Hazardous areas under normal atmospheric conditions Maximum Tank Over pressure Ambient temperature range Process Fluid t
35. loyd VI 7 2 Section 3 B 15 22 2003 Guidelines for the Performance of Type Approvals Test Requirements for Electrical Electronic Equipment and Systems Report ID 16036 Manufacturer ENRAF TANKSYSTEM SA Rue de l Industrie 2 1630 BULLE Switzerland Philippe Despagne General Manager Created modified Remarks 2011 05 25 2011 05 25 2011 05 25 Creation TL _ Wee See The prints of this document are not controlled under the quality management system unless printed on ORIGINAL paper KAQUALITYVSO_9001 7000_Checking_Lists TS8_7028_E doc Printed 21 06 11 13 51 51 Honeywell Enraf Tanksystem SA Rue de l Industrie 2 1630 Bulle Switzerland Phone 41 0 26 91 91 500 Fax 41 0 26 91 91 505 tanksystem honeywell com 50474 TOMSYS 1207 July 2012 www tanksystem com 2012 Honeywell International Inc Honeywell
36. ly at least every 6 months the continuity of grounding by measuring the electrical resistance between the tape adaptor or the sensing probe tube and the quick connect coupler Resistance should not exceed 10 2 Periodically clean carefully the sensor probe and the mechanical parts as storage tube main housing with an appropriate solvent 10 3 Checking the battery Please note that in case you have to change the battery it must be done only in a safe area Refer to section 10 3 Battery replacement Before starting gauging Switch on the unit The buzzer tones every 2 seconds if the battery is not too low If the power left is less than 50 we recommend to have a spare battery ready for exchange See also 10 3 Battery replacement If the power left is less than 20 battery status appears on the display and battery should be exchanged for longer use 2152 3 Battery MM 19 Meng QUIT Honeywell Or in Inch scale 7 0 47 64 Battery Bl 19 HET QUT 10 4 Battery replacement Warning change the battery only in a non hazardous area owitch off the unit Install the Quick Cap Unscrew the 2 screws of the battery holder using the 2 5 mm Hex Allen key which is located on the carrying case See Figure 10 1 Figure 10 1 Pull it gently out Do not use knife or screwdriver to remove the battery cover to avoid damaging the gasket Remove the battery holder form the battery cover
37. ncides with the virtual reading index level so that no correction would be necessary For achieving this it may be necessary to install an adjusting pipe between the deck and the valve If the valves are installed directly on deck or if for any reason the level of the virtual reading index is below or above the zero ullage level then a correction table should be used There should be no internal tank structure between the valve outlet and the 4 metre below such that will impede the path of the equipment into the tank All valves shall be installed at the same level There is no way to correct systematic level errors When designing the gauging port and to avoid damaging the tape during rewinding it is advised to chamfer or to grind all sharp edges on pipes flanges etc that could damage the tape when operating the gauge 8 2 Example of installation on a pipe connector Q2 Tank Zero Ullage level E Reading index 474mm V ETS supply Customer supply HT Tank Deck Figure 8 2 Valve designation C 2 SS C 2 SS W C 2 SS BL C 2 SS SEC Dimension HV is without gasket If gaskets are used dimension HT is reduced by thickness of gasket 18 www tanksystem com Honeywell Example of installation on the deck 8 3 connector Q2 Tank Zero Ullage level HX H Heading index 474 mm HV Tank Deck Valve designation C 2 SS C 2 SS W C 2 SS BL C 2 SS SEC Dimension HV is without gasket
38. odsv O OOS H t yLE0 ON DIS SS Z 9 DA 1700007 2 19mJ3Hlz TE ess oe is om o fu fe 8 001 SI saA DA 1 E GA NI 1 Ls Ll Sabu lesze COSA pue t a ON um 661 0 LZ UdN Im e asia Sand sexeee n g bumped ueis o Lt e EE Y PEA ET MS ENT 194109 wi E Dim fosse 3ud a es lo 2 z E Olly 53903 duvHS ONY SUUNE TIV 3AOMH o pa eom ea zz fa o L1 ri VSdN Caro solo oso ro so su ON fsi 1emajen_ UO 4011959 Daun to wa De 313 9bLp 0002 o0 ooe foor oc 9 oaf 113 6661 9 656 JASSI ooi occ oor oc 9 9o zis uen 461 M 0311191dS JSIMYSHLO SSI1NN S39NYM3101 ero Jeo 017m de nes i56 L ar mt Fa era FEIER bra 0 ws CI NT 2m NAL o o y A KE lvl a 01 LI CL vl 45 pls len EE CIE le e un SS Aag 7 CD CH De z mem CO ex as S Nix cos srl O ex el ie H lois N LS iA En h NM 5x E a SE nn Ne 1 eo A o es Rew C e SE c y I a7 23 e A 12 O uy a l uw 2o LJ ee E This drawing is our property and must not without our permission be copied or made available to others The receiver is responsible for every misuse Speciality HERMetic Deck Valve Item Ot Weight 1 1 0 Valve 4 without cover ss 0 ith weat o a
39. ruding FFKM Perfluoro rubber special for chemical applications 4 www tanksystem com Honeywell Mechanical housing Display unit pu WITH Storage tube SS2 Quick connector Q2 ima Sensing probe TOMSYS Figure 2 2 3 General information 3 1 Shipment note The following parts should be included in the shipment e 1 instrument fitted out with six 6 batteries in the display 1 Allen keys 3 mm e 1 emergency crank e 1 Operation and Service Manual 3 2 Initial inspection Check the contents of the shipment for completeness and note whether any damage has occurred during transport Carry out the Initial test refer to 9 4 Checking the functions before using the unit to verify the good functioning If the contents are incomplete or if there is a damage do not use the device a claim should be filled with the carrier immediately and Enraf Tanksystem SA Sales or Service organization should be notified in order to facilitate the repair or replacement of the instrument 3 3 Documentation discrepancies The design of the instrument is subject to continuous development and improvement Consequently the instrument may incorporate minor changes in detail from the information contained in the manual 3 4 Warranty One 1 year after installation but max 18 months after delivery ex works except batteries The Vendor undertakes to remedy any defect resulting from faulty design materials
40. rvice and Repair Standards used IEC 60079 0 ed 5 0 2007 10 IEC 60079 11 ed 5 0 2006 07 IEC 60079 26 ed 2 0 2006 08 The customer is responsible for any freight and customs clearance charges If units are sent on a freight collect the charges will be invoiced to the customer When returning units or parts for repair to the factory please fill ATEX 111 GExiallB T4 Ga out a service request form see next page The serial number 20 C Ta 50 C letter T followed by 5 digits is printed on the identification Standards used plate as shown on the Figure 2 1 EN 60079 0 2009 EN 60079 11 2007 When returned to Enraf Tanksystem SA equipment EN 60079 26 2007 must be contamination free If it is determined that the customers equipment is contaminated it will be Factory Mutual returned to the customer at the customers expense FM Approvals CL I DIV 1 GP C 4D Tamb 50 C Contaminated equipment will not be repaired until CL I ZN O AEx ia IIB T4 Tamb 50 C such time that the customer decontaminates the said Standards used equipment 3600 1998 General 3610 2010 Instrinsic Safety Regarding product compliance against standards updates or new standards please refer to the Declaration of conformity If you need a copy of any of these certificates please contact Enraf Tanksystem SA Rue de l industrie 2 1630 Bulle SWITZERLAND Telephone 41 26 91 91 500 Telefax 41 26 91 91 505 Web site www tanksys
41. tem com E mail tanksystemOhoneywell com Service Request Customer s address EE RS ene D ane A e O a DOC ge EE Serial number a ao Short description of Trouble 8 www tanksystem com 4 Service station network Service station network list can be found on our website at www tanksystem com For repairs please contact the after sales department at tanksystem honeywell com Or Send your device to Enraf Tanksystem SA Rue de l Industrie 2 1630 Bulle Switzerland Honeywell 5 Recommendation for safe use 10 This Operation and Service Manual is a guide in order 6 Inthe absence of such instructions the following should to help the user to operate the instrument to our best be noted knowledge hee es 6 1 Ifa metal sounding pipe is fitted beneath the Nevertheless the maker disclaims all responsibility and deck valve or tank is inerted then ullaging etc is liability for damage resulting from the use of the equipment permissible at any time with no restriction regardless of the cause of the damage 6 2 If there is no sounding tube or tank is not inerted the Attention is drawn to the possible hazard due to following precautions shall be taken electrostatic charges which may be present in the 6 2 1 Ifthe cargo is not a static accumulator liquid tank This may happen in particular with static i e its conductivity is more than 50 pS m accumulator liquids i e liquids which have low then ul
42. ticular it is not allowed to repair electronic circuits In no event shall Enraf Tanksystem SA be liable for indirect incidental or consequential loss or damage or failure of any kind connected with the use if its products or failure of its products to function or operate properly Enraf Tanksystem SA do not assume the indemnification for any accident or damage caused by the operation of its product and the warranty is limited to the replacement of parts or complete goods Honeywell 3 5 Certification 3 5 2 Classification society certiifcates The HERMetic TOMSYS is certified by following classification societies 3 5 1 Ex certificates ABS Certificate of design assessment MED D Enraf Tanksystem SA is an ISO 9001 certified company by OM 3 6 Spare parts Ki 3 g ANG MEDO Corteo Company r When ordering spares identify the spare part by TS number and p by Det Norske Veritas DNY Wb DN c3 d description Refer to section Drawings Certification GmbH Certified ISO 9001 Company Some spares might be repairable in this case send the part s QMI SAI Global to any authorised service center or to the factory TRE In case of urgency complete replacement units can be made The equipment has been approved for the electrical intrinsic available Contact the factory or nearest Service Station for safety by the following authorities details IECEx ExiallB T4 Ga 20 C Ta lt 50 C 3 7 Se
Download Pdf Manuals
Related Search
Related Contents
SARL JEAN BUSER DISTRIBUTEUR DE PIÈCES TECHNIQUES ET Cámara de imágenes en color del sistema solar Orion StarShoot™ IV DVI-D and USB 2.0 Extender 19” LED Tube Controller MediTouch Brochure FR Sony MZ-R4ST User's Manual Controller User`s Manual PDF - Library Automation & Management Artwizz SeeJacket Leather for Samsung Galaxy S III, black (GT-i9300) FX-07 Mode d`emploi Copyright © All rights reserved.