Home

Operation and Service Manual

image

Contents

1. Figure 3 2 Appearance 2 Output Resistor Cover Output resistor of the 16339A can be changed for measuring high capacitance DUTs For details about the output resistor refer to To Measure A High Capacitance DUT High terminal This terminal provides a source voltage of up to 1000 V This terminal is connected to the center conductor of the BNC connector Ground terminal This terminal is connected to ground which is the outer shield of the triaxial and BNC connectors Guard terminal This terminal is connected to the guard which is the inner shield of the triaxial connector Low terminal This terminal is the measured current path and is connected to center conductor of the triaxial connector Alligator clip This is used to measure components which can not be measured using axial lead module or SMD module Miniature banana plug cable This cable interconnects the component socket module and the 163394 SMD module This is used to measure chip components Axial Lead module This is used to measure components with leads Flat table This is used when using alligator clips This table is made of Teflon 3 2 Operation A2003009 Figure 3 3 Appearance 3 15 Short Bar Normally this shortins bar is connected in output resistor cover 16 Output Resistor Th
2. L5006002 Figure 2 2 Connecting the Adapter Preparation for Use 2 3 Exchanging the Component Module The 16339A has three types of component modules The following procedure describes how to set and remove a component module Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339A s output is a high voltage of up to 1000 Vdc maximum You must operate after turning off the voltage source output and you have confirmed the high voltage indicator is turned off Component Module Setting Procedure Figure 2 3 Setting Component Module 1 Pull the clamps up 2 Insert the component module 3 Push the clamps down 4 Connect the miniature banana cables Component Module Removal Procedure Figure 2 4 Removing the Component Module 1 Disconnect the miniature banana cables 2 Pull the clamps up 3 Remove the component module 2 4 Preparation for Use Configuring the Component Module The 16339A has three types of the component modules The configuration for each module is shown in Figure 2 5 Chip Component Soket Module Lead Component Soket Module Alligator Clip and Flat Table Figure 2 5 Component Module Configurations Repackaging the Adapter If shipment to a Ag
3. LK on non 4 4 Assembly Replacement Axial Lead Module 2 2 2 2 LE En ne 4 8 Replaceable Parts on non 4 8 Assembly Replacement SMD Module 0 0 0 008684 4 9 Disassembly and Assembly Procedure 2 2 e 4 9 Replaceable Parts on non 4 10 Troubleshooting 4 11 Cable Connection Check nn 4 11 BNC Cable and Triaxial Cable Connection Check 2 2 2 2 2 2 4 11 Interlock Circuit Connection Check 2 2 cler 4 12 Contents 1 Contents 2 Schematic Diagram Figures 2 1 Product Overview 2 2 2 2 Connecting the Adapter 2 3 2 3 Setting Component Module 2 4 2 4 Removing the Component Module MN 2 4 2 5 Component Module Configurations 08004 2 b 3 1 Appearance 1 2 2 2 2 2 2 2 2 2 2 252 2 2 225 2 5 25 5 3 1 3 2 Appearance 2 2 2 Co Coon 3 2 3 3 Appearance 3 2m moon nn 3 3 3 4 OPEN compensation for the SMD Module 2 3 4 3 5 OPEN Compensation for Alligator Clips rrr 35 3 6 Capcitance Measurement Configuration 0 2 2 08 3 7 3 7 Output Resistor Changing Procedure 2 2 2 2 2 2 3 8 4 1 Micro Switch Adjustment Screw KL m m rn ren 4 3 4 2 Main Body Replaceable Parts 3 of 3 4 6 4 8 Cable Connection Check m nn en 4 11 4 4 Schematic Diagram of the 16339 A nn ren 4 13 Tables 1 1 Furnished Acces
4. is turned off Open Correction Procedure 1 Remove the alligator clips and banana cables from the INPUT terminal as shown in Figure 3 5 Figure 3 5 OPEN Compensation for Alligator Clips Close the top cover Select the current measurement mode at the 4339A Set and apply source voltage required for your measurement at the 4339A Wait until the current has stabled to within 0 5 pA Press keys of the 4339A to perform the OPEN correction DME wD Operation 3 5 To Measure A High Capacitance DUT When the 4339A measures a high capacitance DUT AC noise creates major instability factors in the measurement When you measure a high capacitance DUT measurement result will not stable This due to the capacitance is sensitive for AC noise If you want to measure resistance of capacitor with keeping more than 10 of S N ratio the maximum measurable capacitance of the DUT for each measurement range are listed in Table 3 1 Table 3 1 Maximum Measurable Capacitance Measurement Measurement Time Range Short Medium Long 100 pA 1 nF 10 nF nA nF 10nF 100nF 10 nA 10nF 100nF 1 pF 100 nA 100nF 1uF 10 pF l pA lF 10uF 100 pF 10 pA 10uF 1004F 1 mF 100 pA 100 pF To measure a higher capacitance connecting a resistor in series with DUT is effective in reducing the measurement flicker The 16339A s output resistor connected in series with the DUT is exchangable The 16339A
5. 16088 40001 1 Knob 0535 0031 2 Nut for Knob 2 16339 60011 1 Top Cover Assembly 3 0515 0907 2 Screw 4 16339 60017 1 Panel Assembly 5 16008 40051 1 Trim 0515 1550 4 Screw for Trim 3050 0891 4 Washer for Trim 6 16339 60103 1 Flat Table 16339 00207 2 Plate Teflon 1390 0365 2 Snap In Plunger 0390 0366 2 Snap In Grommet 7 16339 61653 1 Cable Assembly Interlock 8 16339 61611 1 Cable Assembly Triaxial 9 16339 61612 1 Cable Assembly BNC 10 1400 0249 3 Cable Tie 11 16117 40001 1 Holder Gable 12 16117 40002 1 Holder Gable 13 0515 1552 2 Screw 14 16339 00213 1 Lid Output Resistor 15 0515 0914 2 Screw Flat Head M3 L6 16 16339 00651 1 Short Bar 17 0515 0924 2 Screw Pan Head M3 L6 1 Hidden by Panel Assembly 2 included in Flat Table Service 4 5 SIS SS SS Tas Figure 4 2 Main Body Replaceable Parts 3 of 3 4 6 Service Table 4 4 Main Body Replaceable Parts 3 of 3 Reference Agilent Part Designator Number Qty Description 1 16008 60053 1 Bottom Cover 2 0515 0914 2 Screw Flat Head M3 L6 3 0403 0424 4 Cushion 20 6x7 6 4 16339 00101 1 Chassis Assembly 5 16339 68022 2 J W L 6 0515 1087 1 Screw Pan Head M3 L14 for micro switch 7 0535 0031 1 Nut 8 3101 0302 1 Micro Switch 9 0515 0920 2 Screw Pan Head M3 2190 0584 2 Washer 3050 0891 2 Washer 10 16339 01211 1 Angle Bracket 0515 0914 2 Screw Flat Head M3 L6 11 1400 0054 3 Clamp Cable 12 1400 0015 2 Clamp Cable 13 1
6. 3 Set source voltage to 42 V 4 Press the V output key of the 4339A Confirm that the V output indicator and the High Voltage indicator turn on 5 Open the top cover of the 16339A Confirm that the High Voltage indicator turns off immediately 6 Close the top cover again Confirm that the High Voltage indicator still turns off If you encountered any errors in checking procedure contact your nearest Agilent Technologies Office Operation 3 9 Service Introduction This chapter gives the service information for the 16339A Component Test Fixture This chapter consists of assembly replacement information for each of the main body and socket modules and troubleshooting information assembly replacement sections include disassembly and assembly procedures and replaceable parts lists The troubleshooting section includes cable connection information and the schematic diagram Figure 2 1 shows all items included with the 16339A Table 4 2 Table 4 3 and Table 4 4 identify the Main Body replaceable parts Table 4 5 identifies the Axial Lead Module replaceable parts Table 4 6 identifies the SMD Module replaceable parts To order the parts listed in this manual quote the Agilent part number indicate the quantity required and address the order to the nearest Agilent Technologies office Warning These servicing instructions are for use by qualified personnel only Do NOT perform any servicing other than that contained in t
7. has four different resistors for output resistor Ideally higher series resistance enables more stable mesurement However there is limitation for improvement of S N ratio by connecting higher resistance actually Following table lists upper limit of the series resistance for each measurement range Table 3 2 Limit Output Resistance for Each Range Measurement Range Output Resistor 100 pA 100 MQ 1nA 10 MO 10nA 1 MQ gt 100nA 100 kQ Table 3 3 shows an appropriate resistor for each range when you measure higher capacitance than listed in Table 3 1 with keeping more than 10 S N ratio 3 8 Operation Table 3 3 Appropriate Output Resistance for Range Measurement Measurement Time Range Short Medium Long 100 pA 10 MQ 100 MO 10 MQ 100 MA 1nA 10 MQ 1 MQ 1 MQ 10nA 1 MQ 100 kQ 100 kQ 100 nA 100 kQ 100 k0 100 kQ l yA 100 k9 Short Bar Short Bar 10 yA Short Bar Short Bar Short Bar 100 pA Short Bar 1 100 MQ resistor is effective for measuring current less than 10 pA in 100 pA range 2 Using 100 kQ resistor is recommended but the response will become slow 3 4339A s internal input output resistance works well enough so an extra resistor in series is not necessary Using a high resistance causes the charging time constant of capacitive DUT to increase thereby increasing the necessary measurement time Figure 3 6 shows simplified measurement configuration w
8. 400 0493 2 Cable Tie 14 0515 2079 5 Screw Pan Head M4 15 16339 68023 1 J W L 16 16339 60012 2 Terminal Assembly 2950 0043 2 Nut 17 0764 0045 2 Resistor 22 K 2 W 18 16339 25010 1 Stud 19 16339 00611 1 Terminal Plate 0515 0924 1 Screw Pan Head M3 L6 3050 0891 1 Washer 20 1901 0880 4 Diode Service 4 7 Assembly Replacement Axial Lead Module This section includes the Replaceable Parts list for the Axial Lead Module Replaceable Parts Table 4 5 identifies the replaceable parts for the Axial Lead module To order the parts listed on this manual quote the Agilent part number indicate the quantity required and address the order to the nearest Agilent Technologies office Table 4 5 Axial Lead Module Replaceable Parts e NISUS 1 OD ES N O ESO Reference Agilent Part Designator Number Qty Description 16339 60102 1 Axial Lead Module 1 16339 60013 1 Plate Assembly 2 16339 68004 1 J W L 3 16339 68003 1 J W L 4 0890 0720 20 mm x2 Tube Heat Shrinkable 5 1200 1422 1 Socket 6 0515 2058 2 Screw Plastic 7 16339 00605 1 Shield Cover 0515 0914 3 Screw Flat Head M3 L6 for Shield Cover 8 1390 0366 2 Snap In Grommet 9 1390 0365 2 Snap In Plunger 1 The whole unit 2 includes connectors 4 8 Service Assembly Replacement SMD Module This section includes Disassembly and As
9. Agilent 16339A Component Test Fixture Operation and Service Manual ta e eto ll Agilent Technologies Agilent Part No 16339 90010 Printed in JAPAN January 2000 Third Edition Notice The information contained in this document is subject to change without notice This document contains proprietary information which is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without the prior written consent of the Agilent Technologies Agilent Technologies Japan Ltd Component Test PGU Kobe 1 3 2 Murotani Nishi ku Kobe shi Hyogo 651 2241 Japan Warranty This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of one year from the date of shipment except that in the case of certain components listed in this manual the warranty shall be for the specified period During the warranty period Agilent Technologies will at its option either repair or replace products which prove to be defective For warranty service or repair this product must be returned to a service facility designed by Agilent Technologies The Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to the Buyer However the Buyer shall pay all shipping charges duties and taxes for products returned to Agilent Technologies from another co
10. cial incidental or consequential damages whether based on contract tort or any other legal theory Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products If you need assistance contact your nearest Agilent Technologies Sales and Service Office Addresses are provided at the back of this manual Copyright 1991 1999 2000 Agilent Technologies Japan Ltd Manual Printing History The manual printing date and part number indicate its current edition The printing date changes when a new edition is printed Minor corrections and updates which are incorporated at reprint do not cause the date to change The manual part number changes when extensive technical changes are incorporated December 1991 s sss First Edition part number 16339 90000 July 1999 clic Second Edition part number 16339 90010 January 2000 sssssssssssssss ee Third Edition part number 16339 90010 Safety Summary The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment In addition it violates safety standards of design manufacture and intended use of the instrument The Agilent Technologies assumes no liability for the customer s failu
11. connector or the Triaxial connector Table 4 7 Interlock Circuit Connection Pin Top Cover Number Open Close 1 NC GND 2 GND 3 GND 4 GND 5 NC 6 GND 1 Not Connected 2 Connected to the chassis 4 12 Service Schematic Diagram Figure 4 4 is the schematic diagram of the 163394 Figure 4 4 Schematic Diagram of the 16339 A Service 4 13
12. duct Description The 16339A has been designed to operate specifically with the 4339A High Resistance Meter The 16339A is used to measure chip lead or other types of components The 16339A has the following features m Reduce electrical noise effects by using a shielded case m Chip and lead components can be measured in a component socket m High voltage safety designed using an interlock circuit General Information 1 1 Accessories Supplied The accessories listed in Table 1 1 are supplied with the 16339A Table 1 1 Furnished Accessories Description Part Number Quantity Operation and Service Manual P N 16339 90010 1 Operating and Safety Precautions Service The voltage levels up to 1000 V in this adapter warrants extreme care for operator safety Service must be performed only by qualified personnel Specifications This section lists the complete 16339A specifications These specifications are the performance standards and limits against which the 16339A is tested When shipped from the factory the 16339A meets the specifications listed in this section Measurement Parameter sssssssee ees Direct Insulation Resistance Applicable Test Voltage oooooooooocoooorrrr eee ee 1000 V maximum Applicable Test Current isses e eee 10 mA maximum Applicable Instrument 0 0 e n 4339A Interlock Circuito furnished Cable Length ooo 0 8 m Operating Temperature o 0 to 55 C Opera
13. e POTENTIALLY DANGEROUS PROCEDURES throughout this manual Instructions contained in the warnings must be followed Warning Dangerous voltages capable of causing death are present in this instrument Use extreme caution when handling testing and adjusting this instrument Safety Symbols General definitions of safety symbols used on equipment or in manuals are listed below A IS O Warning Caution Y Note vi A Instruction manual symbol the product is marked with this symbol when it is necessary for the user to refer to the instruction manual Alternating current Direct current On Supply Off Supply This Warning sign denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could result in injury or death to personnel This Caution sign denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could result in damage to or destruction of part or all of the product Note denotes important information It calls attention to a procedure practice condition or the like which is essential to highlight Affixed to product containing static sensitive devices use anti static handling procedures to prevent electrostatic discharge damage to component Caution risk of electric shock Terminals which may be supplied from the interior of
14. ere are four exchangable output resistors for the 16339A Indicated number shows its resistance 100 K means 100 kQ 1 M means 1 MQ 10 M means 10 MQ m 100 M means 100 MO For details about output resistors refer to To Measure A High Capacitance DUT OPEN Correction OPEN correction cancels the residual inductance for resistance measurements The procedure for each component socket module is as follows Axial Lead Module When using the axial lead module the OPEN correction procedure is Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339A s output is a high voltage of up to 1000 Vdc maximum You must operate after turning off the voltage source output and you have confirmed the high voltage indicator is turned off OPEN Correction Procedure Remove DUT if connected Close the top cover Select current measurement mode at 4339A Apply the source voltage required for your measurement at the 4339A Wait until the current has stabled to within 0 5 pA Press keys of the 4339A to perform the OPEN correction SOR ON Operation 3 3 SMD Module When using the SMD module the OPEN correction procedure is Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339A s output is a high voltage of up to 1000 Vdc maximum You must operate after turning off the voltage sou
15. es 4 16339 60015 1 Electrode Assembly Fixed 5 16339 25002 1 Block 6 0515 1552 2 Screw Pan Head M3 L12 for block 7 16339 00601 1 Shield Cover 0515 0914 3 Screw Flat Head M3 L6 for Shield Cover 8 16339 60016 1 Electrode Assembly Movable 9 16339 23004 1 Screw 10 16339 05001 1 Lever 11 1390 0366 2 Snap In Grommet 12 1390 0365 2 Snap In Plunger 1 The whole unit 2 includes connectors 4 10 Service Troubleshooting This section gives cable connections and the schematic diagram of the 16339A Cable Connection Check After repairing check the cable connections listed in lt xref CONNECTION undefined and Table 4 7 using a multimeter which has a low resistance range 1 or 2 Q The information in these lists can be also used for troubleshooting the 16339A BNC Cable and Triaxial Cable Connection Check Check the connections between the connectors and the terminals as shown in Figure 4 3 Note Before checking the connections the shorting bar should be attached to the i Output Resistor terminals Y Shield Center Conductor BNC Connector Outer Shield Inner Shield Center Conductor Terminals Triaxial Connector Figure 4 3 Cable Connection Check Service 4 11 Interlock Circuit Connection Check Interlock connector connections and its pin locations are shown in Table 4 7 Check the connections between each of the 6 pins and the outer conductor of the BNC
16. he operating section unless you are qualified to do so Service 4 1 General Information This section includes Figure 2 1 which shows all items included with the 16339A Table 4 1 16339A Product Overview L Reference Agilent Part Designator Number Qty Description 1 16339 60101 1 SMD Module 2 16339 60102 1 Axial Lead Module 3 Not Assigned 1 Main Body 4 8710 1984 2 Alligator Clip 5 16339 61001 1 100 kQ Output Resistor 16339 61002 1 1 M9 Output Resistor 16339 61003 1 10 MQ Output Resistor 16339 61004 1 100 MQ Output Resistor 6 16339 61621 2 Miniature Banana Plug Cable 7 16339 60103 1 Flat Table 16339 60011 1 Carrying Case 16339 90020 1 Operation and Service Manual 1 Not shown in this figure 4 2 Service Assembly Replacement Main Body This section includes Disassembly and Assembly Procedures and Replaceable Parts for the Main Body Disassemble and Assemble Procedure This section explains complex and significant points of disassembly and assembly of the main body Top Cover Removal Before removing the top cover the screw which pushes the micro switch closed should be removed from the top cover The screw is accessed by removing the bottom cover it is mounted on the portion of the top cover which extends down into the box See Figure 4 1 Figure 4 1 Micro Switch Adjustment Screw Micro Switch Adjustment Note When the top cover or the micro switch is reassemb
17. hen measuring capacitance with series resistance HP 16339A HP 4339A Figure 3 6 Capcitance Measurement Configuration Where Rs Output Resistance of the 4339A Rx Parallel resistance of the DUT Cx Parallel capacitance of the DUT Ri Input resistance of the 43394 Ro Additional Series Output Resistance of the 16339A Operation 3 7 Neccessary charging time is approximately led by following equation X x Rs Ri Roy 100 Re te T x In Where te charge time s X Measurement Error T Cp X R Ri Ro Generally when measuring capacitor with as described configuration in this section you should wait charging time for more than 10 or 20 times 7 Note When the additional series resistor value becomes high enough its resistance i can not be disregarded and the resistance of the DUT causes additional Y measurement error Changes output resistor of the 16339A as shown in Figure 3 7 Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339A s output is a high voltage of up to 1000 Vdc maximum You must operate after turning off the voltage source output and you have confirmed the high voltage indicator is turned off To Exchange Output Resistor DW Screw x2 Resistor Cover Screw x2 Output Resistor A9003007 Figure 3 7 Outp
18. ilent Technologies service center is required each adapter should be repackaged using the original factory packaging materials Alteratively comparable packaging materials may be used Wrap the adapter in heavy paper and pack in anti static plastic packing material Use sufficient shock absorbing material on all sides of the 16339 A to provide a thick firm cushion and to prevent movement Seal the shipping container securely and mark it FRAGILE Preparation for Use 2 5 Operation Introduction This chapter describes the features of the 16339A Figure 3 1 and connecting a DUT to the 43394 D A9003001 Figure 3 1 Appearance 1 1 Top cover This shields against external electrical noise and provides an interlock to enable and disable the test voltage when CLOSED and OPEN respectively 2 BNC connector This connector provides the source voltage to the 16339A 3 Triaxial connector The measured signal is carried on the center conductor of this connector 4 Interlock connector This connector enables the interlock function which enables and disables applying the source voltage from the 4339A when the top cover is CLOSED and OPEN respectively When the shield case is OPEN the source voltage is turned OFF inside the 43394 Operation 3 1 10 11 12 13 14
19. led the length of the screw N which pushes the micro switch closed must be adjusted so that it will cause the Y micro switeh to open at the proper top cover opening angle The micro switch adjustment procedure is as follows Micro switch adjustment procedure 1 Put a drop of Lock Tite PN 0470 0013 on the screw s threads at the nut 2 Rotate the screw clockwise until you hear the micro switch click closed and can verify using a low resistance range of a multimeter and see the resistance change when the switch closes 3 Rotate the screw in one more turn clockwise 4 Tighten the nut to fix the screw while holding the screw in position with a screw driver Service 4 3 Replaceable Parts Table 4 2 Table 4 3 and Table 4 4 identify the replaceable parts for the main body To order the parts listed in this manual quote the Agilent part number indicate the quantity required and address the order to the nearest Agilent Technologies office Table 4 2 Main Body Replaceable Parts 1 of 3 Reference Agilent Part Designator Number Qty Description 1 0515 0914 3 Screw Flat Head M3 L6 2 16008 09051 1 Hinge 3 0403 0316 2 Bumper FT PRS IN 4 1 Bushing 5 0515 0914 1 Screw Flat Head M3 L6 6 0515 0907 3 Screw Flat Head M3 L8 7 16008 00251 1 Panel Rear 4 4 Service Table 4 3 Main Body Replaceable Parts 2 of 3 Reference Agilent Part Designator Number Qty Description 1
20. rce output and you have confirmed the high voltage indicator is turned off Open Correction Procedure 1 Remove DUT if connected and leave the SMD module electrode contact spacing the same as the DUT s spacing as shown in Figure 3 4 2 Tighten the screw to hold the interval between the electrodes Same as the DUT s width rat 1 t 7 re Figure 3 4 OPEN compensation for the SMD Module 3 Close the top cover 4 Select the current measurement mode at the 4339A 5 Set and apply the source voltage required for your measurement at the 4339A 6 Wait until the current has stabled to within 0 5 pA 7 Press keys of the 4339A to perform the OPEN correction Note Keep the chip contact area clean because any contamination will degrade i measurement accuracy To keep the chip contact area clean avoid touching Y this area with your bare hands and use clean insulated tweezers when inserting and removing chip components 3 4 Operation Alligator Clip When using the alligator clips the recommended OPEN correction procedure is as follows Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339A s output is a high voltage of up to 1000 Vdc maximum You must operate after turning off the voltage source output and you have confirmed the high voltage indicator
21. re to comply with these requirements Note 16339A is designed for use in INSTALLATION CATEGORY I according to IEC i 61010 1 and POLLUTION DEGREE 1 according to IEC 61010 1 and IEC 60664 1 Y 16339A is an INDOOR USE product Do NOT operate in an Explosive Atmosphere Do not operate the instrument in the presence of flammable gasses or fumes Operation of any electrical instrument in such an environment constitutes a safety hazard Keep Away from Live Circuits Operating personnel must not remove instrument covers Component replacement and internal adjustments must be made only by qualified maintenance personnel Do not replace components with the power cable connected Under certain conditions dangerous voltages may exist even with the power cable removed To avoid injury always disconnect power and discharge circuits before touching them Do NOT Service or Adjust While Alone Do not attempt internal service or adjustment unless another person capable of turning off power and capable of rendering first aid and resuscitation is present Do NOT Substitute Parts or Modify Instrument Because of the danger of introducing additional hazards do not substitute parts or perform unauthorized modifications to the instrument Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure the safety features are maintained Dangerous Procedure Warnings Warnings such as the example below preced
22. s how to install the Agilent 16339A Component Test Fixture The topics covered include initial inspection ambient environmental considerations connecting the adapter for use and repackaging the adapter for shipment Initial Inspection The adapter has been carefully inspected electrically and mechanically before being shipped from the factory It should be in perfect physical condition no scratches dents or the like and it should be in perfect electrical condition Verify this by carefully performing an incoming inspection to check the adapter for signs of physical damage and missing contents If any discrepancy is found notify the carrier and Agilent Technologies Your Agilent Technologies sales office will arrange for repair and replacement without waiting for the claim to be settled 1 Inspect the shipping container for damage and keep the shipping materials until the incoming inspection is completed 2 Verify that the shipping container contains everything shown in Figure 2 1 and listed in Table 2 1 3 Inspect the exterior of the 16339A for any signs of damage Preparation for Use 2 1 Figure 2 1 Product Overview Table 2 1 Contents Description Agilent Part Number Quantity D Component Test Fixture 163394 1 Axial Lead Module 16339 60102 1 SMD Module 16339 60101 1 Fiat Table 16339 60108 1 Miniature Banana Plug Cable 16339 61621 2 amp 100 kQ Output Resistor 16339 61001 1 1 MQ Output Resis
23. sembly Procedures and the Replaceable Parts List for the SMD Module Disassembly and Assembly Procedure SMD Module Disassemble Procedure 1 Loosen the three screws on the cover and remove the cover 2 Unsolder the wires from the terminals 3 Remove the fixed electrode assembly and the block by loosening the inside screws 4 5 Remove the movable electrode assembly Remove the screw which attaches the spring lever SMD Module Assemble Procedure aon A QN Attach the spring lever Set the movable electrode assembly in the gap of the spring lever Tighten the screws holding the movable electrode assembly Attach the block Attach the fixed electrode assembly Attach the cover Service 4 9 Replaceable Parts Table 4 6 identifies the replaceable parts for the SMD Module To order the parts listed in this manual quote the Agilent part number indicate the quantity required and address the order to the nearest Agilent Technologies office Table 4 6 SMD Module Replaceable Parts i c S NN u a Y 0000 Reference Agilent Part Designator Number Qty Description 16339 60101 1 SMD Modulet 1 16339 60014 1 Plate Assembly 2 0515 0924 1 Screw Pan Head M3 L6 3050 0891 1 Washer 16339 00607 1 Plate 42851 24023 3 Washer 42851 29021 1 Spring 3 0515 1550 4 Screw Pan Head for electrod
24. sories Con nn nen 1 2 2 1 Contents onen 2 2 3 1 Maximum Measurable Capacitance oaoa a a a 3 6 3 2 Limit Output Resistance for Each Range a a a 3 6 3 3 Appropriate Output Resistance for Range 2 2 2 2 En nn l l ln 3 7 4 1 16339 A Product Overview Coon on nen 4 2 4 2 Main Body Replaceable Parts 1 of B 4 4 4 3 Main Body Replaceable Parts 2 of 3 4 5 4 4 Main Body Replaceable Parts 3 of 3 4 7 4 5 Axial Lead Module Replaceable Parts nn nn leen 4 8 4 6 SMD Module Replaceable Parts 2 2 KL LE nr ren 4 10 4 7 Interlock Circuit Connection 2 2 nen 4 12 Contents 3 General Information Introduction The purpose of this manual is to enable you to use your Agilent 16339A Component Test Fixture efficiently and confidently This manual contains both general and specific information To use the 16339A to perform a specific function without having to read the entire manual follow the directions in Using the 16339A Using the 16339A The 16339A has been designed to operate specifically with the 4339A High Resistance Meter m To install the 16339A turn to Chapter 2 m To operate the 16339A turn to Chapter 3 m To order replaceable parts for the 16339A turn to Assembly Replacement Main Body in Chapter 4 m To troubleshoot the 16339A turn to Troubleshooting in Chapter 4 Pro
25. the equipment at a voltage exceeding 1 kV or allow connection to a voltage exceeding 1 kV are marked with this symbol Contents 1 General Information Introduction 2 4 2 2 nn 1 1 Using the 16339A oo on nen 1 1 Product Description 2 2 2 on nn 1 1 Accessories Supplied 2 2 2 2 m m nn l l es 1 2 Operating and Safety Precautions 1 2 Service 2 lll Ce 1 2 Specifications Ce a 12 Supplemental Performance Characteristics MM 1 3 2 Preparation for Use Introduction Ko 4 2 nn 2 1 Initial Inspection 4 on on nn 2 1 Ambient Environmental Considerations 2 2 En m rn nn 2 3 Operating and Storage 2 3 Connecting the Adapter for Use 2 MM 2 3 Exchanging the Component Module 2 4 Configuring the Component Module e 2 5 Repackaging the Adapter 2 5 3 Operation Introduction 2 2 2 2 CL m on nn 3 1 OPEN Correction 3 3 Axial Lead Module ll ll ll a 3 3 SMD Module 3 4 Alligator Clip Co on 35 To Measure A High Capacitance DUT CL En nn 3 6 Checking Procedure 3 9 4 Service Introduction 2 4 4 2 4 4 e 4 4 As 4 1 General Information Cu m m m nn 4 2 Assembly Replacement Main Body 2 2 22 lr 4 3 Disassemble and Assemble Procedure 2 2 l l len 4 3 Top Cover Removal 2 2 2 2 2525252525525202D25 5 254 4 8 Micro Switch Adjustment 4 8 Replaceable Parts
26. ting Humidity ssssssssssssssssss eee as lt 70 RH 40 C Non operating Temperature ssssssssssss s n tent eens 40 to 70 C Non operating Humidity 0 0 0 0 00 n ees lt 95 RH 40 C Weight ssssssssssllssee e e e e e e er e 2 2 kg 1 Maximum measurable current of the 4339A is 100 pA 1 2 General Information Supplemental Performance Characteristics This section gives supplemental performance characteristics Supplemental performance characteristics are not specifications but are typical characteristics included as additional information for the operator Supplemental performance characteristics are not guaranteed Measurement Range Leakage Current When Opened ooo ooo ccoo lt 1pA Stability of Leakage Current When Opened isses 0 5 pA Measurable Component Parts 0 0 00 cece eee eee Radial leaded Axial Leaded Chip Measurable DUT Size with Alligator Clip 2222c22ccseenen lt 45 mm diameter AN 1x 10 to 2 x 1015 Q Measurable Chip Size Width 0 5 to 10 mm Height 0 5 to 10 mm Diameter 0 5 to 3 0 mm o Length 0 1 to 8 mm Dimensions ssssssssssssss e ese 200 W x 140 H x 230 D mm 1 After compensation measurement time is LONG 23 5 C lt 50 RH 2 After 1000 V has been applied for 1 minute and under the same conditions as 1 General Information 13 Preparation for Use Introduction This chapter explain
27. tor 16339 61002 1 10 MQ Output Resistor 16339 61003 1 100 MQ Output Resistor 16339 61004 1 D Alligator Clip 8710 1984 2 Carrying Case 16339 60011 1 O Operation and Service Manual 16339 90020 1 1 Carrying Case and Operation and Service Manual is not shown in Figure 2 1 2 2 Preparation for Use Ambient Environmental Considerations Operating and Storage The 16339A must be operated within an ambient temperature range of 0 C to 55 C and relative humidity up to 70 RH at 40 C non condensing The 16339A may be stored within a temperature range of 40 C to 70 and at a relative humidity up to 95 at 40 C non condensing Connecting the Adapter for Use Connect the 16339A to the 4339A as shown in Figure 2 2 You must connect the BNC triaxial and interlock connectors If the interlock connector is not connected the 4339A will not output the source voltage Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339A s output is a high voltage of up to 1000 Vdc maximum You must operate after turning off the voltage source output and you have confirmed the high voltage indicator is turned off DOOS ARE COMPONENT TEET FRTURE
28. untry Agilent Technologies warrants that its software and firmware designed by Agilent Technologies for use with an instrument will execute its programming instruction when property installed on that instrument Agilent Technologies does not warrant that the operation of the instrument or software or firmware will be uninterrupted or error free Limitation of Warranty The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer Buyer supplied software or interfacing unauthorized modification or misuse operation outside of the environmental specifications for the product or improper site preparation or maintenance No other warranty is expressed or implied Agilent Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose Certification The Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology to the extent allowed by the Institute s calibration facility or to the calibration facilities of other International Standards Organization members Exclusive Remedies The remedies provided herein are the buyer s sole and exclusive remedies Agilent Technologies shall not be liable for any direct indirect spe
29. ut Resistor Changing Procedure pa Turn OFF the voltage output of the 43394 if it is turned ON Open the top cover of the 16339A 3 Remove the two screws holding the output resistor cover then remove the output resistor cover 4 Remove the two screws holding the output resistor DO 3 8 Operation 5 Remove the internal resistor 6 Select an appropriate resistor for the measurement range you want to select by using Table 3 3 7 Place the new output resistor so that the two holes are aligned with the two electrodes 8 Insert and tighten the two screws to hold the resistor 9 Close the output resistor cover and refasten with two screws Warning Do NOT use an output resistor other than those furnished with the 16339A Checking Procedure The 16339A and the 4339A is operated with high voltages of up to 1000 V These products are designed so that the operator can make safe measurements To maintain this safe condition you must periodically perform the following safety verification procedure Warning Do NOT touch the electrode and UNKNOWN connector while the High Voltage indicator is lit which shows the 4339A s output is a high voltage of up to 1000 Vdc maximum You must operate after turning off the voltage source output and you have confirmed the high voltage indicator is turned off Daily Safety Verification Procedure 1 Connect the 16339A to the 4339A 2 Close the top cover of the 16339A