Home
Service manual (English)
Contents
1. Measuring instrument incl flow meter ON D750 The Measuring instrument provides an easy way to measure and detect if there is something wrong with ECG function of Decapus Ill If the resistance deviates much from the default value for a patient lead this may indicate a fault condition With the flow meter it is easy to measure the system s airflow to find if any leakage has arisen which affects the vacuum Salt bridges can occur both in the patient cable and signal module e The resistance between the shield and the patient lead may indicate the existence of salt bridges between the system s shield and the chosen lead Keep in mind that the chance of detecting salt bridges is best if the system is tested when a problem has just occurred If moisture has time to dry off the possibility of detecting salt bridges is lowered 1 Resistance measurement a Connect the signal module A without Quickels electrode to respective electrode lead position Plug the recorder cable into the D SUB of the measuring instrument Decapus lll does not need to be connected to the mains b Set the multimeter to Q and connect the multimeter s test leads in B ECG End and C Patient Cable End c Turn the knob E to the selected patient lead and record the value If the value of the resistance of the lead deviates much from the default value this lead has to be investigated in more detail see section Troubleshooting Resistance of the in
2. 31 O O Interference on the ECG trace First thing to do when encountering a disturbed ECG signal is to determine wether the disturbed ECG signal arises from the Decapus Ill or the external ECG printer and not a bad electrical contact to between the electrode and patient With normal troubleshooting refers to 1 Check the Quickels electrode electrolyte and placement of the Quickels electrodes 2 Check by using the printer switch which channel is incorrect ECG measurement can be affected by many factors in the measuring environment It could for example be any of these factors e The static electricity in the room is high e Mains cables 50 60 Hz in the patient environment or in close proximity to the recorder cable e Mains cables in a wall near the patient 50 60 Hz e It is also assumed that the six initial assessments described in the introduction of section Troubleshooting guide has been completed IRREGULAR BASE LINE Typical curve with irregular base line is probably caused by fluctuating vacuum level This may in turn depend on leakage in the vacuum system The stabilization of the skin electrode interface may affect the base line This might occur due to excessive movement of the patient which may cause leakage of the electrodes The base line normally stabilizes within a few seconds but stabilization in some cases may take longer e g if the skin is tight tibia Attach if possible the electrode
3. FET STATUS error recurs If it recurs the battery must be battery ERROR replaced Decapus III can be used without the SAFETY STATUS battery with power supply meanwhile the new ERROR battery is to be delivered SWITCH OFF THE SERVICE NEEDED INDICATOR LIGHT When the error indicated by the service needed indicator light on the patient module s control panel is resolved the system must be put into energy saving mode see Technical description Energy saving mode This is done by disconnecting the power supply from the mains and the USB cable from the pump module If the system is equipped with a battery the on off button must then be pressed until the system is put into an idle mode Thereafter connect the power supply into the mains and press any button on the control panel to start the system SERVICE MANUAL 22 DECAPUS III Annual service Arrangements of a recurring nature that should be performed to ensure continuous and trouble free use of Decapus Ill INSPECTION e Ensure that no part of the system is missing e Check that the arm joints do not fit loosley and that the table fixation if one is used are steady e Check that the patient module casing is intact especially where the arm is attached e Check that the pump module casing is intact e Check that the control cable vacuum hose and recorder cable do not show visible damage burn marks or cracks especially where they are bent and connected to
4. 2 7 m Schiller Recorder cable 2 7 m Siemens Reversed recorder cable 2 7 m Siemens Recorder cable 2 7 m Welch Allyn Recorder cable 2 7 m TEPA Recorder cable 4 0 m Schiller Changes may occur without prior notice SERVICE MANUAL 42 43 DECAPUSS III o O 000 Spare parts and accessories contd Spare parts and accessories contd QN NR PRODUCT NAME QN NR PRODUCT NAME ON D40SI Recorder cable 4 0 m Siemens ON D796 Arm attachment Schiller Kangaroo ON D310 1 Control cable 2 8 m ON 500 Quickels electrodes ON D310 2 Control cable 3 1 m QN 509 Signal module 10 pcs QN D311 1 Vacuum hose 2 8 m ON 510 Signal module 1 pc QN D311 2 Vacuum hose 3 1 m QN 511 Marking rings 10 pcs ON D111 Power mains cable QN 512 Marking labels 10 pes ON D101 Power supply XP Power Model AFM 45US15 ON D107 USB protection lid ON 410 Patient cable 1 0 m ON D129 Vacuum connector ON 420 Patient cable 1 2 m ON D133 Rubber foot 4 pcs ON 440 Patient cable 1 4 m ON 400 1 Precordial leads 6x1 0 m Limb leads 4x1 2 m ON D701 Pump attachment ON 400 2 Precordial leads 6x1 0 m Limb leads 4x1 4 m ON D750 Measuring instrument incl flow meter QN 400 3 Precordial leads 6x1 2 m Limb leads 4x1 4 m ON D305 H QN D752 ECG adapter assembly Ask Quickels Systems AB for compatible ECG printers QN 600 Arm assembly ON 601 Table fixation QN 602 Patient module holder QN 603 Cable clip kit QN 604 Electrolyte spray assembly ON 606 Signal cable retainer ON 726
5. Provided that everything is properly installed 1 Check that the signal modules are firmly connected to the patient cables This is easiest done by listening after leakage sounds 2 Control the system s airflow with the flow meter in the Measuring instrument QN D750 If the airflow is low a Block the exhaust nipple or squeeze the vacuum hose near the exhaust nipple while the vacuum is on Listen for leakage in the pump unit b If not there may be leakage in the vacuum hose patient module or one or more of the signal modules e Leakage in the vacuum hose Replace the vacuum hose ON D106 e Leakage in the patient module Replace the vacuum manifold amp ECG circuit board QN D302 1 e Leakage in the pump module Replace the Silencer assembly QN D104 29 O Troubleshooting guide contd ERROR THE SYSTEM WILL NOT START AND NO INDICATOR LEDS LIGHT UP WHEN THE PATIENT MODULE S BUTTONS ARE PRESSED THE SYSTEM S BATTERY WILL NOT CHARGE CAUSE Connect the power supply see No vacuum above The battery is in sleep mode Faulty adapter circuit board in the patient module Faulty motherboard System without battery The mains power indicator light is off power supply is faulty Communication error The power indicator light on the power supply is off power supply is faulty SOLUTION Connect the power supply the battery shall wake up and start charging Connect another
6. Wall attachment with plate QN 793 Arm attachment Changes may occur without prior notice Changes may occur without prior notice
7. cables signal modules Quickels electrodes table fixation and arm SIGNAL MODULE D Signal module with filter that absorbs 1 5 ml of fluid QUICKELS ELECTRODES E The easily replaceable Quickels electrodes contain a primary filter that can absorb 1 ml of fluid and prevents fluid borne contamination from entering the system The electrode itself is coated with silver silver chloride for high signal quality ARM Q The easily adjustable arm supports the patient module and simplifies patient movement from rest to stress position Arm movements are adjusted using the three friction screws POWER SUPPLY O Connects to an earthed mains supply Pump fastening other arm fastenings and ECG adapters for ECG printers are available as accessories SERVICE MANUAL 10 DECAPUS III o Technical description THE SYSTEM S MAIN COMPONENTS The three main functions works by a combinations of the three individual Vacuum function components e Vacuum function to fixate the Quickels electrodes on the patient by regulating the ECG function vacuum O e ECG function to detect and convey ECG signals from the patient to an external ECG printer Control and monitor function Patient e Control and monitor function e al to monitor and control the Decapus Mains vacuum function and the input ees from the control panel Kar Vac Pump module Service in
8. complies with the EMC requirements of IEC 60 601 1 2 Radio transmission equipment etcetera must not be used in the vicinity of the apparatus since this could affect its function Special care must be taken when using powerful interference sources such as for diathermy and similar and also ensure that diathermy cables etcetera are not located on or near the Decapus Ill In case of doubt please consult responsible personnel or the supplier CONNECTION TO ECG PRINTERS AND MONITORS The Decapus lll printer cable may only be connected to e ECG printers and monitors that comply with Medical Device Directive 93 42 EEG and are defibrillation protected type CF according to IEC 60 601 1 e Decapus lll must not be connected to ECG printers and monitors of Safety Classification B according to IEC 60 601 1 e Decapus Ill may only be connected to ECG printers and monitors which are defibrillation protected with either lt 10 kOhm or lt 5 kV and 400 J SERVICE MANUAL 6 DECAPUS III O O O Important safety information contd SYSTEM CONSTRUCTION Decapus IIl is a system made up of modules What is stated in this User manual about safe use assumes that Decapus III is used in its entirety i e that no part of Decapus III is replaced with other equipment The same applies to all the system s cables Decapus Ill must only be used as a vacuum applicator for ECG electrodes in the way described in this user manual Quicke
9. damage Take any necessary action if e Fluid has been spilled onto the system e The cover or casing is split damaged or missing e Cables or connectors show signs of damage e The system has suffered possible mechanical damage e g fallen on the floor Safety precautions should be taken when ECG measurements are to be taken from patients fitted with a pacemaker or other electric stimulator SAFETY CLASSIFICATION Decapus IIl complies with the safety requirements of the International Electrotechnical Commission publication IEC 60 601 1 Safety of Medical Electrical Equipment Safety Classification Group type CF with defibrillation protection and Classification Group lla in accordance with the Medical Device Directive 93 42 EEC WARNING OF POWERFUL ARM SPRING Check that the friction screw in the middle of the arm is tight when the arm is unpacked otherwise the arm may spring up In the interests of safety hold the arm together with a firm grip when the tape is removed The arm is locked with a friction screw CONNECTION TO AN EARTHED ELECTRICAL SOCKET The safety classification states that Decapus III may only be connected to an earthed socket Note Connection must not be via a multi socket extension to which other electrical equipment that does not have the same safety classification as specified above is connected IMPORTANT INFORMATION ABOUT RADIO TRANSMISSION EQUIPMENT ETCETERA Decapus IIl
10. in the casing and do not get pinched when the bottom casing is mounted 4 Replacing the motherboard a See above 1 Open the pump module casing then b Lift the silencer assembly K and disconnect the hose L from the pressure sensor M on the motherboard and hose N from vacuum nipple O Cable tie and aluminum support retain mounted on the large plastic jar c Disconnect the battery cable B and the vacuum pump cable F from connectors C and G on the motherboard d Unscrew the three screws P and remove the motherboard e Place the new motherboard in the casing and secure it with the three screws P f Connect the battery cable B and the vacuum pump cable F to connectors C and G on the new motherboard Connect the hose L to the pressure sensor M on the new motherboard and hose N in the vacuum nipple O Place the silencer assembly K in the casing g Check that the vacuum hoses smoothly go through the notch J in the casing and do not get pinched when the bottom casing is mounted When replacing the motherboard the new motherboard must be loaded with the system s existing serial number This is done with the Service application SERVICE MANUAL 36 DECAPUS III Ra RR O O O O Repairs and parts replacement conta 5 Replacing the silencer assembly a See above 7 Open the pump module casing then b Lift the support E c Lift the silencer assembly and disconnect the hose L from pressure switch M on the motherb
11. on less tight skin Some notches in the base line are in most cases due to the cause that the skin adjust to a new position with the electrode If the irregular curvature occurs for a single ECG channel examine and possibly change the signal module and or the patient cable If the irregular curvature occurs for all ECG signals it is likely leakage in the patient module vacuum hose or pump module see section Troubleshooting guide Leakage NO ECG SIGNAL Typical curve when the ECG signal is missing If a single ECG channel lacks signal do the following 1 Make sure the patient cable is properly connected to the signal module 2 Replace the Quickels electrodes and try again 3 Replace the signal module and try again 4 Measure the resistance of the ECG signal path for the particular ECG trace SERVICE MANUAL 32 DECAPUSS III O O Interference on the ECG trace conta Locate the fault of the high resistance replace the defective part patient cable recorder cable or ECG manifold board see section Measuring instruments QN D750 If the measure shows 0 Q it is a break in the signal path Troubleshoot and replace the defective part 5 Examine the ECG recorder If all ECG signals are missing on the ECG trace you should do the following 1 Switch Quickels electrodes for R L and F 2 Switch signal modules for R L and F and troubleshoot as in point 4 above 3 Measure the resistance of R
12. patient module if the system starts up replace the User interface adapter board in the patient module QN D300 Replace the motherboard ON D140 To restart Decapus lll do the following 1 Disconnect the system from the mains 2 Disconnect the USB cable otherwise the system is programmed not go into sleep alas can not restart 3 Press the on off button for 3 seconds 4 Wait 5 seconds 5 Press any button on the Patient module to restart the system Replace the power supply QN D101 If Decapus Ill is connected to the mains and the indicator for mains power is lit on both the power supply and the patient module but the battery will not charge indicator on the patient module is not flashing unplug the power mains cable from the wall socket and re plug it again Replace the power supply QN D101 SERVICE MANUAL 30 DECAPUSS III Troubleshooting LOCATING LEAKAGE IN THE VACUUM SYSTEM The audible Pump motor speed in vacuum function can be used to determine if there is a leak in the vacuum system A high pump motor speed is an indication that the system is leaking Leakage in the vacuum system can be located by incrementally remove more and more parts of the vacuum system closer to the vacuum pump and for each time clog the vacuum hose with a finger while listening to the pump motor speed The Run pump test in the Service application can preferably be used to detect leakage see section Service application
13. side of the motherboard e Check that the battery cables smoothly goes through notch D in the casing and do not get pinched when the bottom casing is mounted f Connect the plug power supply to Decapus III Charge the battery fully 3 h before the system is operational f tim yy ji Tey A Charge the battery before it is operational The battery is put in a sleep mode when delivered amp Never use other than intended battery AN r LS LS VY WJ Repairs and parts replacement conta orientation of the pump LN EO 3 Replacing the vacuum pump a See above 1 Open the pump module casing then b Lift the support E c Disconnect the vacuum pump cable F from connector G left side of the motherboard d Disconnect the two vacuum hoses H on the silencer assembly J from the Y nipples wherein the vacuum pump with support can be lifted e The new vacuum pump is pressed into the supplied support and placed in the casing f Connect the pump cable F in connector G left side of the motherboard g Connect the two vacuum hoses H in into the vacuum pump s Y nipples l h Place the support E on top of the small plastic jar in the silencer assembly next to the battery i Check that the vacuum hoses H smoothly go through the notch J
14. L and F leads see section Measuring instrument QN D750 and point 4 above 4 Examine the ECG recorder DISTURBANCES CAUSED BY THE MAINS SUPPLY Typical curve with superimposed 50 or 60 Hz disturbance from the No hi mains supply If interference occurs only on one ECG channel you should do the following 1 Replace Quickels electrode and try again 2 Replace signal module and try again 3 Examine the measuring environment 4 Look for salt bridges isolation measurement 5 Measure resistance of specific ECG trace see section Measuring instrument QN D750 If interference occurs on all ECG channels you should do the following 1 Check for 50 60 Hz interference sources nearby 2 Switch Quickels electrodes of the N path R L and F 3 Switch signal modules for the N path R L and F 4 Measure resistance of the ECG signal of the N signal R L and F Locate the fault of a high resistance replace the defective part patient cable ECG manifold board or recorder cable see section Measuring instruments QN D750 5 Conduct isolation measurement between patient leads and shield N path R L and F see section Measuring instruments QN D750 The resistance between the lead and each ECG channel should be gt 30 MQ 6 Examine the ECG recorde
15. PC Service Tests In case of leakage see section Troubleshooting Guide Leakage RESISTANCE OF THE INDIVIDUAL COMPONENTS OF THE SYSTEM One way to find errors in the ECG function of the system is to measure the resistance from the electrode to the ECG recorder connector In order to easily measure the resistance of Decapus Ill see section Measuring instrument QN D750 The resistance of the defibrillation protection contributes X kO Default for your model code of Decapus III see section Technical data the rubber signal transmittor on the ECG circuit board contributes with a resistance less than 100 Q and the connector into the signal module contributes with a resistance less than 100 Q Other parts of the signaling pathway contributes with a negligible resistance If the resistance differ from the standard value see section Technical data with more than 500 Q something is wrong in some part of the signal path Note that this only applies to the ECG signal paths The resistance of the shield shall be less than 100 Q between the D SUB and the shield connector of the patient module rubber shield transmittor The shield between the D SUB and the ECG manifold board shall be less than 3 0 The shield is pin 6 on the D SUB connector on the patient module D sub Signalmodul EKG kretskort Elektrod l Patientkabel a Skrivarkabel l l l l l Standardvardet 500 Q lt 1000 lt 100 0 E n
16. SERVICE MANUAL Decapus IIl Qyickels Contents About the Service manual The Service manual is part of the documentation available for Decapus III and is intended for those who work as service technicians or similar at health care centres hospitals or private clinics and is responsible for servicing and About the Service manual 3 maintenance of the system The manual provides information on what you Terminology 3 need to know about safety installation and how to use the system This is Decapus III 4 Other available documents concerning Decapus III is the User manual Important safety information 5 Decapus III and QuickGuide Decapus lll Key to symbols 8 Decapus Ill system design Technical description 10 Sean i Terminology Signal modules and electrodes 17 Service application PC 18 AUSCULATION Monitoring of body sounds for medical diagnosis Annual service 22 Measuring instrument incl flow meter 23 PATIENT LEAD The signal is derived by measuring the voltage between two points on Installation 25 So Troubleshooting guide 2 MOTHER BOARD Central control unit Troubleshooting 30 Interference on the ECG trace 31 DEFIBRILLATION A method where a patient s heart is exposed to electric shock with a excitement around 5 kV Repairs and parts replacements 34 ECG Electrocardiography ESD ECD Electrostatic Discharge Technical data ay DIATHERMY A technology to burn away tissue by using a high frequency voltage of about
17. Spare parts and accessories 41 7 5 kV PID Proportional Integral and Derivative regulator PWM Pulse width modulation pulse width modulation a square wave of fixed Quickels Systems AB frequency where the part of the period of time the signal is high and low can be varied Bjornnasvagen 21 SE 114 19 Stockholm l2C Inter Integrated Circuit a synchronous serial multi master bus to couple low SWEDEN frequency units to the motherboard Telephone 46 8 709 49 00 Fax 46 8 709 00 12 E mail info quickels com SERVICE MANUAL DECAPUS III This is Decapus II Decapus Ill is hygienic easy to use and time saving The system has all the advantages that are normally associated with vacuum application but lacks the disadvantages Vacuum ECG with flow compensated negative pressure prevents the negative pressure from significantly exceeding the level selected on the patient module Double filters efficiently prevent fluid borne contaminants from being sucked into the system and causing an obstruction The easily exchangeable Quickels electrodes can be replaced as necessary which means that from a hygienic point of view Decapus III is comparable to adhesive electrodes In special situations the signal modules can easily be replaced with adhesive electrodes UNIQUE SIGNAL MODULE AND FILTER What makes Decapus III unique are the patented signal modules and Quickels electrodes Quickels electrodes are easy to attach to the
18. arged and Decapus lll can be used again In case of malfunction of the battery the service lamp lights up on the patient module see section Service application PC If Decapus IIl is connected to the mains and the indicator for mains power is lit on both the power supply and the patient module but the battery will not charge indicator on the patient module is not flashing unplug the power mains cable from the wall socket and re plug it again For more information see section Troubleshooting guide 17 Signal modules and electrodes SIGNAL MODULES AND QUICKELS ELECTRODES The air sucked into the system may contain fluid borne pollutants for this not be sucked into the system the Quickels electrode consists of a barrier the primary filter that may absorb up to 1 ml of liquid When the primary filter has absorbed 1 ml of liquid the airflow is stopped because the primary filter blocks the vacuum and the Quickels electrode can not be attached to the patient The Quickels electrode contains an electrode sensor coated with silver silver chloride to permit high signal quality The secondary filter in the signal module provides additional protection against moisture liquid and impurities sucked into the rest of the vacuum system It can absorb up to 1 5 ml of liquid SIGNAL MODULES The signal module consists of a valve housing of silicon rubber a holder with a cover a connector of electrically conductive silicone and a secon
19. ay be due to old rubber bushings see section Troubleshooting Rubber bushings ECG shield distributor Vacuum manifold e SERVICE MANUAL DECAPUS III Pump module The parts The pump module includes vacuum pump silencer pulsation dampers pressure sensor and control electronics battery as an option Control cable input from patient module USB connection gt Never connect to anything other than the intended power supply gt Only connect with Decapus III patient module gt The USB input must only be connected when Decapus III is being serviced i e never when the system is near or adhered to a patient gt Li ion battery Fire and explosion hazard Never use any other kind of battery The battery may only be replaced by a qualified service technician Power supply input Vacuum hose input from patient module Pump module contd VACUUM FUNCTION Decapus Ill is equipped with a high performance vacuum pump with dual heads Maximum vacuum is gt 700 mbar Free flow rate is gt 8 min The main task of the vacuum function is to provide a controllable pressure in the signal module and Quickels electrodes so that these can be attached to the patient and provide a good electrical contact without dirt and sweat to be taken into the system For more information see section Patient module Control signal ECG FUNCTION The main task of th
20. between the parts of the system Medical Electrical System Medical Electrical Equipment Patient Environment Applied Part _ Control signals Vacuum amp ECG Medical device insulation 1 ECG printer j 7 7 TTT TTT TTT SERVICE MANUAL 12 DECAPUS III Patient module The parts The patient module consists of a control panel control electronics and a defibrillation protector The control panel has touch buttons and is easy to use and clean CY 4 buttons for setting desired negative pressure Indicator for electrode change after 10 measurements Button for acknowledgement of replacement Qyickels DECAPUS III Ei JORIE 46 CO Serre DB xn oosen electrodes after UW TO measurement O Oe ua lt lt Battery operation and charging status indicator lt lt lt c c lt LI Marking panel for leads Mains power indicator Service needed indicator Contact authorised service technician CONTROL SIGNAL When a certain vacuum level is selected on the control panel the motherboard initiates a control signal to the vaccum pump the pump creates a vacuum The mother board also contains a pressure gauge which measures the pressure in a hose branched from the vacuum pulsation damper The signal from the pressure gauge moves on to the processor block comparing the pressure with that from
21. bleshooting guide THE FOLLOWING TROUBLESHOOTING GUIDE REQUIRES THAT YOU COMPLETE THE FOLLOWING INITIAL CONTROLS 1 Decapus Ill is connected to mains turned 4 ECG printer and Decapus III mains cables on and energized from the power supply are plugged into the same earthened socket mains fuse in the building controlled 5 Patient cables are properly connected to the 2 The ECG printer is plugged in turned on signal modules and energized by the mains mains fuse in the building controlled 6 Saline solution is used as electrolyte not water 3 The recorder cable is properly connected to the ECG printer ERROR CAUSE SOLUTION NO VACUUM Decapus IIl not connected to mains Connect the power supply to mains The mains power indicator light is Replace the power supply QN D101 off power supply is faulty The power indicator on the power Replace the motherboard ON D140 supply is lit but not the mains operation indicator on the patient module THE QUICKELS Vacuum pump faulty Replace the vacuum pump ON D100 ELEKTRODES ADHERE Vacuum hose or patient cable Check and repair POORLY OR pinched NOT AT ALL The primary filter clogged in the Replace the Quickels elektrode Quickels elektrode The signal module clogged Replace the signal module Leakage in pump or patient module See section Locating leakage in the vacuum system below ONE OR MORE Poor electrical contact between Remove the electrodes and add mo
22. dary filter The patient cable and the electrode are connected to the contact Note that the function of the valve housing is dependent on the use of the correct Quickels marking disk It opens by pressing lightly on the signal module E y e Cover plate with marking disk 2 fe e lid Electrodes should be w replaced after every 10 5 e Secondary filter p E 2 measurements which is ER Contact indicated by a diode on the front of the patient le a e Holder modula S v T Valve disk p e Primary filter v TIS e Cuff 6 al Electrode All 10 electrodes in the system must be changed at the same time SERVICE MANUAL 18 DECAPUSS III 7 O O O Service application PC Decapus III is equipped with a Service application supplied USB stick to facilitate service preventive maintenance and possible troubleshooting of the system The Service application stores test data and by using the associated ServiceGuide a preventive maintenance report in pdf format can easily be printed On the short side of the pump module an USB port is located 2 to which a PC can be connected Windows operating Systems XP 7 8 enabling direct troubleshooting of the system via the Service application thus obtaining recommendations on the system s fault status The USB cap is removed by the use of a small screwdriver 1 Do the following e Patient module shall be c
23. detail see section Troubleshooting guide Leakage and Technical data 25 Installation 1 Patient module and arm a Attach the arm in the trolley s or the supplied fixation If Quickels Systems AB s table fixation is used se mounting instructions below b Assemble the patient module onto the patient module holder with screw H PN c Attach the control cable recorder cable and vacuum hose to the arm with the 6 cable clips J d Install the cabling according to K so the arm can be moved freely without the cables being stretched or pinched Loosen screws A to uppermost position without detaching Adjust position of counter arm B to the thickness of the table If necessary turn up side down Ensure that studs C fit into hole D Tighten fixing screws E Tighten screws A so that the bench clamp is properly anchored to the table Press the shaft stubb for the arm F all the way down into the foot Friction screw G retains adjusts Check that the installation tolerates loading in all directions d Connect the recorder cable DA 15 into the external ECG printer For ECG printers with other input than D SUB see 5 Adapter For more information see the ECG printer manual Make sure that the recorder cable DA 15 is not in direct contact with the energized mains cable Also remove as far as possible energized cablings in the patient environment see section Technical descri
24. disconnect the control cable U from the adapter circuit board and the pump module see section Installing 2 Pump module and replace with new control cable Mount a new cable tie around the control cable and recorder cable Cut the cable ties Z and unscrew the recorder cable X from the ECG circuit board and the external ECG printer and replace with the new recorder cable and a new cable tie around the control cable and recorder cable Cut the cable tie Z and pull the vacuum hose V from the vacuum nipple on the vacuum manifold W and cut it loose from the pump 37 OO08O Repairs and parts replacement conta module nipple Replace with a new vacuum hose and cable tie For mounting on the pump module nipple see section Installation 2 Pump module c Check that the control cable recorder cable and vacuum hose smoothly go through the notch Y in the casing and do not get pinched when the back casing is mounted 3 Replace adapter circuit board a See above 7 Open the patient module casing then b Disconnect the control cable U and the flexible flat cable from the adapter circuit board c Unscrew the two screws Aa and remove the adapter circuit board d Place the new adapter circuit board in the casing and fasten it with the two screws Aa f Connect the control cable U and the flexible flat cable to the new adapter curcuit board g Check that the control cable recorder cable and vacuum hose smoothly go throu
25. dividual ECG components of the system and Technical data The error may exist in both signal modules and or patient cable where for example salt bridges affects the contact and thus the signal quality 2 Measurement of salt bridges between patient leads and shield Isolation measurement a See above 1 Resistance measurement a then b Set the multimeter to Q and connect the multimeter s test leads in B ECG End and D Shield c Turn the knob E to the intended patient lead and record the value The value should be infinitely large A measured value of 30 MO or less indicates formation of salt bridges Replace the relevant patient cable however the recommendation is to change all patient cables All patient cables are probably worn out and need to be replaced Replace the Quickels electrodes after every 10 measurements and do not spray salt solution directly onto the Quickels electrode By following these recommendation no salt bridges will form SERVICE MANUAL 24 DECAPUS III O Measuring instrument incl flow meter ON D750 contd 3 ECG simulation a Connect Decapus III to the mains Note that the system does not need to be energized to make an ECG simulation test but it can be useful in order to detect possible sources of noise etcetera on the ECG signal b All signal modules should be connected to respective electrode lead position A without Quickels electrode None of the multimeter test cab
26. e ECG function is to detect ECG signals from the patient and convey these to an externally connected ECG printer The ECG function is defibrillation proof Ag AgCl sensor element in the Quickels electrodes are attached to the patient via the negative pressure The signals transduced by the electrodes is transferred by the shielded patient cables to the ECG manifold board in the Patient Module This is the protection against defibrillation and diathermy Both the ECG printer and the recorder cable must be protected from the high voltages which occur during defibrillation 5000 V and diathermy 7500 V The protection consists of serial resistors to limit current and gas discharge tubes that limits voltage The signals from the electrodes are transferred via the shielded patient cables to the ECG manifold board from the ECG manifold board the signals are then transferred via a shielded low noise trunk cable 10 core recorder cable to a 15 pin D sub connector that connects to the ECG printer Patient cables Recorder cable ECG manifold board O Ag AgCl electrodes SERVICE MANUAL 16 DECAPUS III OO Pump module contd BATTERY AND CHARGING The battery power optional means the Decapus Ill can function as a completely mobile system Charging the battery is done automatically when the system is connected to the mains The life time is around 400 charging cycles The battery charging time is about 2 hours and the runtim
27. e is about 3 5 hours which corresponds approximately to e 50 resting ECGs 5 min per round 10 ECG stress tests 20 min per round The battery status is indicated on the control panel by four modes Green continuous light in fields 1 4 indicates the charging level of the battery e Orange flashing light in the lower battery field indicates low battery The system should be connected to a grounded mains outlet for recharging after use e When the battery power reaches the minimum level the system shuts down and flashed with an orange light when the operator presses any button on the control panel e A flashing green light indicates that the battery is charging OEE When the Decapus III is used for the first time after delivery the system shall be connected to the mains in order to ensure that the battery is charging indicated by green flashing light on the control panel This is because the battery on delivery is set in a sleep mode and need to be activated The same applies when changing the battery When the system runs on battery without mains connected the pump starts when the vacuum level is selected with the buttons 1 4 on Va the control panel If the vacuum pump on battery power is left on by mistake the system automatically shuts down after 1 hour If the a7 battery should become completely discharged the system provides no feedback at all By connecting the system to the mains the battery can be rech
28. eart rhythms The system is intended for recording rest and stress ECG although not for longer than 40 minutes NORMAL USE Normal use means about 25 resting ECGs per day 5 min per round PATIENT GROUP From 5 years Not intended for patients with open wounds or other major skin damage in the regions where Quickels electrodes are applied SYSTEM LIFETIME 10 years OTHER Materials in the Quickels electrodes and signal modules do not contain latex The material in the patient module and pump module covers is made of ABS PC plastic Changes may occur without prior notice 41 O O Spare parts and accessories QN NR QN QN QN QN QN QN QN QN QN QN QN QN QN QN QN QN QN QN QN QN QN PUM A B D100 D140 D103 D104 D105 A D106 PAM A C D E F D300 D302 1 D301 A D303 D13SC D13SI D27SC D27SC_R D27Sl D27SI_R D27WA D27TP D40SC PRODUCT NAME Pump module D3 CPL Vacuum pump assembly Motherboard circuit board Battery assembly Silencer assembly Pump module casing top Pump module casing bottom Patient module D3 CPL User interface adapter board circuit board ECG manifold board circuit board amp Vacuum manifold assembly Patient module casing front Patient module casing back including hatch Recorder cable 1 3 m Schiller Recorder cable 1 3 m Siemens Recorder cable 2 7 m Schiller Reversed recorder cable
29. f test you wish to perform Troubleshooting or Preventive Maintenance Tools Button Test Follow the instructions in the information box to check each individual Data Collections Location e g emergency ward ward 22 button The ServiceGuide via the Service femne Serial No application returns OK Not OK bed Test Date 2014 03 11 Button Test Pump Test Checks the function of the vacuum Pump Test Last check date pump The Service application tests each Function Check vacuum level every four second and returns Electrical Safety Finalizing OK Not OK as well as the present vacuum evel Perform the flow test with external eqiupment be sure to click OK Not OK as well as fill in the results Approved if the flow is gt 6 itres min Function Check Perform the tests with external eqiupment e g Quickels Systems AB s Measuring instrument QN D750 Be sure to click OK Not OK as well as fill in the results See section Measruing instrument QN D750 Electrical Safety Perform the tests with external eqiupment e g Quickels Systems AB s Measuring instrument QN D750 Be sure to click OK Not OK as well as fill in the results Finalizing Fill in any final comments and who performed the test Click Save PDF Ext 1 Chec bleshooti Report Type Trou ing report 21 O O O Service application PC contd POSSIBLE ERROR MESSAGES IN THE SERVICE APPLICATION MOTOR PUMP Indicates that the vacuum pump i
30. g State errors recorded by the system SERVICE MANUAL 20 DECAPUSS III O O O Service application PC contd SERVICEGUIDE The ServiceGuide is a step by step guide intended to ease possible troubleshooting and the yearly maintenance see section Annual service For other matters such as installation repair and more specific troubleshooting see the relevant section in this service manual By using the ServiceGuide two different types of tests can be performed e Troubleshooting gt Troubleshooting report e Preventive Maintenance PM gt PM report This is selected in the drop down list under the tab General Information The ServiceGuide is operated by clicking on the icons in the lower right corner You can at any time go back to a previously answered page by clicking on its tab in the upper left corner Tools Fill in used tools and equipment Extrernal Checks Inspection of the system click OK Adjusted N A Data collection The ServiceGuide retrieves data from the Service application Wait until the collection is complete System Test Retrieved data from previous Report type Shows the report step is shown here type to be generated NOTE If the working sensor value differs from 2 83 0 05 V the system must be calibrated see section Service settings above Led Test The led test lights all LEDs on the patient module Be sure to click OK Not OK Ee nen Please choose the type o
31. gh the notch Y in the casing and do not get pinched when the back casing is mounted 4 Replace ECG circuit board and vacuum manifold a See above 1 Open the patient module casing then b Cut the cable ties Z and Z and disconnect the control cable U and the recorder cable X from the ECG circuit board and vacuum hose V from the vacuum manifold c Detach the patient cables Dd see below 5 Replace patient cables d Remove the ECG circuit board and vacuum manifold Bb e Place the new ECG circuit board and vacuum manifold in the casing and connect the recorder cable X and vacuum hose V Fasten the cable ties see 2 Replacing control cable recorder cable vacuum hose f Check that the control cable recorder cable and vacuum hose smoothly go through the notch Y in the casing and do not get pinched when the back casing is mounted 5 Replace patient cables b Detach the patient cables Dd by first gently lever 1 with a small J screwdriver c Install the new patient cables Note that the marking on the patient cable s signal module end shall be consistent with the position displayed by the marking panel on the patient module see Patient module d Screw on the hatch SERVICE MANUAL 38 DECAPUSS III O O O O Repairs and parts replacement conta 6 Replace signal modules a See section Installation 4 Signalmodules and electrodes 7 Replace Quickels electrodes a See section Installation 4 Signalm
32. les should be connected to the measuring instrument B C or D c Connect the recorder cable from the patient module to the ECG printer d Connect the ten laboratory cables can be purchased with ON D750 from the ECG simulation box or similar in the correct socket position on the measuring instrument s left short side F e Start the simulation and check the ECG signal In case of deviations see section Troubleshooting Guide and Interference on the ECG trace 4 Flow measurement The system s airflow can be controlled by using the flow meter a Connect Decapus III to the mains and ensure that the flow meter is stable on a flat surface b Make sure that the signal module is properly connected to each patient cable no gap except for one The end of the patient cable lacking the signal module is pressed into the tube G on the flow meter The measurement can be done with or without Quickels electrodes c Start the vacuum by pressing the vacuum level 4 button on the patient module s control panel and read the value bottom of the bulb with the help of the grading scale on the flow meter H Note that you can listen at the connection between the signal module and patient cable to detect small leakages If you hear any leakage sound push to the signal module tighter onto the patient cable aw L If the value of the airflow deviate much from the default value it has to be investigated in more
33. ls Systems AB will not guarantee the equipment s function if it is exposed to higher levels of electromagnetic interference than it has been tested for Quickels electrodes must be replaced at least once per day and at least after every tenth patient to maintain the system s function and lifetime When Decapus lll is equipped with a lithium ion li ion battery this must not be thrown away when reached the end of their useful life but sent for recycling DEFIBRILLATION PROTECTION The symbol on the back of the patient module means that Decapus lIl has built in protection against defibrillation voltages and that the system has been tested and approved for use in connection with defibrillation The system has been designed so that it cannot be damaged even if the defibrillator electrodes come into contact with the Quickels electrodes During defibrillation never touch the patient or any other conductive item that is in contact with the patient EXPLOSION RISK Decapus Ill is not designed for use in locations where there is an explosion risk FIRE AND EXPLOSION HAZARD Decapus Ill contains a li ion battery as an option Only replace the battery with the type specified WARNING You must read the user manual before using the system Conductive parts of patient cables Quickels electrodes and signal modules including the neutral leads and the electrode must not come into contact with other conductive parts including earth pro
34. n Troubleshooting Resistance of the individual components of the system If the disturbance is dominant on an ECG channel but present on the others the cause could be that there is something wrong with the shield of the signal path used as reference Try the following 1 If you are using the Wilson method change over to the CR method If the disturbance is unaffected you should measure the resistance of the signal path for which the disturbance is dominating If the disturbance changes such that it is only visible on one of the ECG channels you should measure the resistance of the shield of this signal path If the disturbance is present on all ECG channels it might be caused by electrostatic discharge Try and change to less isolated foot ware and the problem might resolve IRREGULAR DISTURBANCES OF THE ECG SIGNAL Uneven disturbances can be caused by a high static charge in the room and too much fluid in or around the electrode Wipe off the excess liquid on the patient Switch Quickels electrode check that the signal module is not full of liquid and that no fluid is absorbed into the patient cables Also note that all motion affects the ECG signal If individuals or objects are moving in the room the electric charge is altered which affect the ECG signal therefore the operator should not move whilst recording the ECG SERVICE MANUAL DECAPUS III 34 ON ON AN N WH NA NA Repairs and
35. n each corner of the lower part of the pump module there is a pre drilled hole where screw and spacer P shall be fasten See supplied installing instructions on the USB stick 4 Signal modules and electrodes a If not done fit each supplied signal modules to the patient cable with the corresponding marking b Ensure that the patient cable is properly connected to the signal module with a tight fit and not being loose Electrodes should be replaced after every 10 measurements which a is indicated by a diode on the front of the patient module When replacing all 10 electrodes in the system must be changed at the same time The acknowledgement button on the control panel of the patient module should then be pressed in 5 Adapter QN D305 H ON D752 Quickels Systems AB s adapter is used to connect the system to an ECG printer with an input other than the D SUB input Quickels Systems AB is using Contact Quickels Systems AB for more information A The adapter may only be used together with Quickels Systems AB s products and approved ECG printers 6 Transport and packaging When packaging the system primarily original packaging should be used If this is not available special care should be taken when packaging the pump module as this contains the vacuum pump and the motherboard These can be damaged during shipment if the pump module is packed inadequate 27 r f O O Trou
36. n the pump See section Pump module ERROR module may be faulty Restart Decapus Ill and activate vacuum level 4 If the error recurs When replacing se section Repairs the pump motor and its cablings must be and parts replacement Pump inspected and replace if necessary module 4 Replacing the circuit board motherboard PRESSURE Indicates the pressure sensor on the pump See section Pump module SENSOR module motherboard may be faulty Restart ERROR Decapus Ill and activate vacuum level 4 If When replacing see section Repairs the error recurs the motherboard needs to be and parts replacement Pump replaced module 3 Replacing the vacuum pump MAINS ERROR Indicates that the power supply that is When replacing see section Repairs connected to the power supply input on the and parts replacement pump module deliver incorrect output voltage Verity that the correct power supply is used Disconnect the power supply and restart Decapus III if it has battery power Plug in the power supply and wait 1 minute to see if the error recurs If the error recurs check the output voltage of the power supply which shall be 15 0 5 VDC If it deviates the power supply must be replaced BATTERY Indicates that the battery s built in security When replacing see section Repairs PF STATUS features are activated to avoid the risk of and parts replacement Pump ERROR damage Restart Decapus III to see if the module 2 Installing replacing the
37. nformation Key to symbols CE label Recycling Follow the manufacturer s instructions Protection for defibrillation voltage Caution Warning Off on Service needed Connected to the mains Battery status 4 levels Serial number Batch code E E A L4 l gt Non sterile Dispose of as electronic waste Keep dry Catalogue number Date of manufacture Manufacturer Decapus lll system design Friction hand I i a qi Spray bottle Na i i x PATIENT MODULE Decapus Ill is controlled from the control panel on the patient module The control cable and vacuum hose connect the patient module with the pump module B The recorder cable connects to an ECG printer PUMP MODULE O The pump module includes vacuum pump silencer vacuum and pulsation dampers control electronics and as an option a rechargeable battery PATIENT CABLES amp The ten shielded interference free patient cables are low friction type and are highly flexible Like the signal modules they are clearly marked with identification complying with IEC standards A connector 4 mm in diameter allows adhesive electrodes to be used Z N The recorder cable must only be connected to MDD 93 42 EEG approved ECG printers of type CF according to IEC 60 601 1 Decapus III consists of pump module patient module power supply control cable vacuum hose recorder cable patient
38. oaded with the system s existing serial number See section Repairs and parts replacements 4 Replacing the motherboard j Bootloader Error Log Serial No Bootloader Shows the version of Checkveron dan a a EE a particular part of the software in Pumprunningtime Decapus Ill r n n Battery status Check status Pump running time Shows how Hi Motor current test long the system has been running pas since start previous reset The time is EA displayed on the upper left corner a o Watchdog Battery status Parameters to Check watchdog consider in the information box is battery voltage number of cycles and absolute charge which indicates the health status of the battery If the cycle count is more than 350 cycles the battery should be replaced Article No 721001_1 Version 1 5 Developed by Quickels Systems AB Motor current test Measurement of the system s power consumption The AC adapter should be connected Set to 100 press the PWM speed button The current consumption of the pump is shown in the information box Is the result over 3 5A it indicates that the vacuum pump is worn out and need to be replaced Watchdog Monitors how many times the processor has been restarted many reboots may indicate errors in the system Contact Quickels Systems and inform about this Article No Shows the article number of the Service application Version Shows the version of the Service application Error lo
39. oard and hose N from the vacuum nipple O dJ LE d Disconnect the two vacuum hoses H from the Y nipples on the vacuum pump wherein the silencer assembly can be lifted see section 4 Replacing the motherboard Om mi gt e Cut the cable tie Q Fixate a new cable tie around the aluminum I as J a O support R and the large plastic jar S in the new silencer assembly Lage f Connect the new hose L to pressure switch M on the motherboard and the new hose N in the vacuum nipple O Connect the two new vacuum hoses H in the vacuum pump s Y nipples Place the new silencer assembly in the casing Pe g Place the support E on top of the small plastic jar in the silencer assembly next to the battery i Check that the vacuum hoses smoothly go through the notch J in the casing and do not get pinched when the bottom casing is mounted PATIENT MODULE 1 Open the patient module casing a Turn of Decapus III Make sure to work with the patient module without weighting the system cablings The patient module may have to be dismantled from the arm by removing the cable clips fixating the cablings b Turn the patient module upside down and unscrew the two screws T as well as the four screws T Then carefully lift up the back casing c Assemble in reverse order 2 Replace control cable recorder cable vacuum hose a See above 1 Open the patient module casing then b Cut the cable tie Z and
40. odules and electrodes 8 Replace the arm a See section Installation 1 Patient module and arm Qualification to perform service Service must only be performed by qualified personnel A Caution Decapus IIl must not be sterilised or immersed in fluids Neither should fluids be allowed to be sucked through the system Annual service If the hospital or institution do not adopt an acceptable schedule for cleaning and inspection of this equipment it may result in equipment failure and health hazards Annual maintenance should be performed by qualified personnel 39 Technical data GENERAL ENVIRONMENT TAL REQUIREMENTS PUMP MODULE PATIENT MODULE POWER SUPPLY CABLES PATIENT CABLE QUICKELS ELECTRODE BATTERY SIGNAL IN OUT PUMP MODULE STAND SERVICE APPLICATION Safety Ambient temperature Relative humidity Air pressure Power consumption Negative pressure Airflow Measurement Weight Defibrillation protection Rapid negative pressure selection Measurement Weight Model Mains voltage and frequency Current rating AC input Classification Printer cable Control cable Vacuum hose Length Tensile resistance Material Battery pack option Manufacturer Lifetime Charging Control cable input USB input DC input Length of folding arm Table fixation Software MDD Classification Class lla IEC 60 601 1 class I type CF with defibrillation pr
41. onnected e Decapus Ill must be connected to the mains The Service application can be used without using the ServiceGuide to obtain test values and information about the system but by using the ServiceGuide a printable pdf report can be generated The Service application and software firmware can also be downloaded from the Website www quickels com but the files must be saved to your computer before they can be used SERVICE TESTS Open up the Service application exe program on your computer Run led test The led test lights all LEDs on the patient module enabling an ocular inspection Firmware version States the firmware version used in your Decapus III Run button test Follow the instructions in the informations box to check the function on each individual button If you wait more than 10 seconds before you press the button it will be recorded as an error Settings Status Device connected io Pump running time Ohour s 00minute s 00second s UIC kels Firmware version 1 50 English Service tests Swedish Run led test Run pump test Checks the vacuum function of the system The Service application tests each vacuum level every four seconds and returns the measured vacuum levels in the information box Run button test A The USB port must only be connected to when Decapus lll is being serviced that is never once the system is near or connected to a pa
42. otection In operation 10 to 40 C In storage 10 to 40 C 25 95 no condensation 700 1060 mbar Max 35 W 0 600 mbar 7 5 litres min at the patient end H 100 mm L 240 mm W 180 mm 1 1 kg 10 kOhm with surge protection 140 V 120 200 240 280 mbar H 30 mm L 110 mm W 135 mm 1 1 kg XP Power Model AFM 45US15 100 240 V 50 or 60 Hz Max 1 1 A IEC 320 C14 Class according to IEC 60 601 1 2 7 m DA 15 plug socket 2 8 m DE 9 plug socket 2 8m Extremities 1 2 m Chest 1 0 m Minimum 50 N Ag AgCl sensor fulfils recommendations of AAMI Biocompatibility and toxicology studies in accordance with biocompatibility In vitro cytotoxicity ISO 10993 5 Skin irritation ISO 10993 10 and Allergy ISO 10993 10 Rechargeable Li ion battery 11 1 V capacity 2 6 Ah Quickels Systems AB Catalogue number QN D103 Approximately 400 charging cycles Charging time approximately 2 hours DE 9 socket plug USB B socket plug DC plug 570 650 mm Clamp measurement 10 130 mm Compatible with Windows XP 7 and 8 Changes may occur without prior notice SERVICE MANUAL 40 DECAPUSS III O Tecnnical data cna OPERATOR PROFILE Clinics with ECG training health centre workers hospitals or private care centres Must understand the language used in the user manual INTENDED USE Record ECG signals from the patient and transfer these to an ECG printer for the purpose of discovering checking various types of deviant h
43. parts replacement Before you start work ensure that the entire Decapus lll system is de energized by 1 Unplug the power supply from the pump module 2 Unplug the recorder cable from the ECG printer Note that some parts the motherboard and the adapter circuit board still are energized if Decapus Ill is equipped with a battery Be sure not to short circuit boards with metal objects alternatively disconnect the battery as well The circuit boards are sensitive to electrostatic discharge ESD which means you must ensure that you have adequate protection against such discharges before handling circuit boards PUMP MODULE 1 Open the pump module casing a Turn of Decapus Ill Make sure to work with the pump module without weighting the system cablings The control cable may have to be dismantled from the pump module b Turn the pump module upside down and unscrew the four screws A Then carefully lift up the bottom casing c Assemble in reverse order 2 Installing replacing the battery a See above 1 Open the pump module casing then b When replacing the battery the battery cable B is to be disconnected from connector C right side of the motherboard the battery and support can then be lifted See picture on next page c Press the new battery into the supplied support with battery cable B downwards facing to the right and then place into the casing d Connect the battery cable B in the connector C right
44. patient s skin Decapus Ill automatically maintain the chosen negative pressure level even if leakage occurs and air gets in This means that Quickels electrodes stay fastened and give a good contact with little risk of the patient feeling uncomfortable or getting suction marks The primary filter in Quickels electrodes is a super absorbent and can absorb up to 1 ml of fluid The secondary filter which is in the signal module provides further protection and never needs changing Quickels Systems AB s innovations in the Quickels electrodes and signal modules make Decapus III a hygienic and reliable system that seldom needs to be shut off to clean blocked signal modules or patient cables The reinforced patient cables with high tensile strength to ensure good signal quality over long periods The system is defibrillation protected BATTERY OPERATION MEANS FREEDOM Decapus Ill has a battery operation with built in charger as an optional extra which makes it completely mobile The battery is a lithium ion type giving a long operating period 4 hours and a fast charging cycle 2 hours 5 O O O Important safety information Decapus III is marked with this symbol This is the international warning symbol to indicate that use of the equipment may be dangerous It is therefore important that you read this user manual carefully Before any equipment is used the operator must ensure that there is no sign of
45. ption 2 Pump module a Install the pump module on the correct position If Quickels Systems AB s pump attachment is used see installation instructions for 3 Pump attachment QN D701 On the short side of the pump module b Connect the control cable to the D SUB input L Make sure that no energized cable comes in direct contact with the control cable c Attach the vacuum hose to the nipple M Make sure that no cables or hose can get caught if the system needs to be moved d Mount the power supply to the trolley or pump module bracket f Connect the mains cable to an earthed socket The same earthed socket multiple socket as the ECG printer g Connect the DC plug of the power supply AFM45US15 to the pump module s DC input N 1 and tighten the screws 2 SERVICE MANUAL 26 DECAPUS III Installation contd i Check that the works by assuring that the mains operation indicator on the control panel of the patient module lights up If the system is equipped with battery the battery operation and charging status indicator lights up see section Pump module Battery and charging 3 Pump attachment QN D701 QN D702 a Quickels Systems AB s pump module bracket is screwed to the trolley by the pre drilled holes O Note that the hole pattern of the pump module bracket is designed so that it can be attached to different types of trolleys hence use the correct holes and screws for mounting b I
46. r 33 O O Interference on the ECG trace contd TO 50 HZ These disturbances are probably caused by the movement of the patient s muscles which can indicate that the electrodes are placed in a less favourable position Relocate the electrodes SUPERIMPOSED DISTURBANCES WITH A FREQUENCY OF 1 HZ DRIFTING BASE LINE WHEN RECORDING ECG STRESS TEST Typical curve where the base line wander during the measurement This form of disturbance is normal when performing a stress test When the patient starts to sweat there is a dramatic and usually much uneven change in skin impedance This is normal interference and can only be remedied by puncturing the upper level of the skin Usually these disturbances subsides and disappearance when the perspiration becomes constant and widespread ECG SIGNAL IS SENSITIVE TO BODY MOVEMENTS Typical recording when the ECG signal is sensitive to movements both patient movement and movements of people in the room This type of problem suggests that the electrical shielding is faulty in some way If the disturbance is present on a single ECG signal try the following 1 Measure the resistance of the shield from the recorder cable s D connector to the shield of the patient cable See sectio
47. re salt solution ECG TRACES elctrodes and the patient ARE MISSING Check with the Measuring instrument Replace the Faulty cable patient cable 50 60 HZ NOISE Poor electrical contact between Remove the electrodes and add salt solution ON ONE ECG elctrodes and the patient See section Interference on the ECG trace TRACE Salt bridges Check with the Measuring instrument Replace the patient cable SERVICE MANUAL DECAPUS III 28 0 O Troubleshooting guide contd ERROR 50 60 HZ NOISE ON SEVERAL ECG TRACES MUSCLE INTERFERENCE ON THE ECG DRIFTING BASE LINE ECG TRACE SENSITIVE TO BODY MOVEMENTS LEAKAGE CAUSE Poor electrical contact with the N electrode EMC disturbing environment Anxious patient Uneven vacuum poor skin contact High static electricity Damaged shield on patient cable or recorder cable SOLUTION Remove the N electrode and add salt solution Make sure no mains connected cables is within the patient environment or in direct contact with the recorder cable and or control cable See section Interference on the ECG trace Make sure the patient can relax See section Interference on the ECG trace Check that the Quickels electrodes are replaced and that the signal modules are not leaking If possible change shoes that are less electrically insulating Troubleshoot the cable shield See section Resistance of the individual ECG components of the system
48. tection The USB input must only be connected when Decapus III is being serviced i e never when the system is near or connected to a patient CURRENT LEAKAGE The patient module cables are galvanically insulated from the pump module and from earth guaranteeing that electric current through 7 O O O Important safety information contd the patient is kept at a safe level If a Quickels electrode the ECG connector or the patient cables come into contact with other conductors there is a risk of patient shock This applies whether the object is earthed or not No equipment not approved by Quickels Systems AB may be connected or there is a danger of current leakage occurring If other electrical equipment which is not classified according to the above is used in the same premises or in the vicinity of the patient appropriate precautions must be taken CLEANING Cleaning should be made daily e Cleaning should be carried out only with a damp cloth and mild detergent soap solution or washing up liquid Wipe dry afterwards e If necessary disinfect with a cloth soaked with 70 alcohol or surface disinfectant Wipe dry afterwards e None of the equipment should ever be immersed into fluids of any kind Nor should liquid be sucked through the system e No part of the system should ever be hot sterilised with water steam or hot air Nor should ether be used If it is decided to use the same Quickels electrodes on more
49. terface THE SYSTEM S MAIN FUNCTIONS Motherboard The three main functions are almost Pump module completely separated The ECG function have no other contact with the two other functions other than being mechanically packed together The interface between the vacuum function and control function consists only in that the control function controls the pump motor in the vacuum function and measure the pressure in a hose coming from the vacuum function Patient module The external boundaries of the system are confined to a power supply a number of electrode signal inputs and an output to an ECG printer The ECG printer is not a part of the Decapus III system Control panel Adapterboard ECG circuit board User interface User interface Vacuum manifold Electrodes 11 O Technical description contd ENERGY SAVING MODE Decapus Ill has a microprocessor and to save energy the system is equipped with an energy saving mode The energy saving mode is activated only when the system runs on battery not when connected to the mains If Decapus lll does not get any signal input from the control panel during a 60 minutes period the system is idle To wake the system press any button on the patient module s control panel If the control cable has been unplugged from the pump module it can take up to 20 seconds before booting the system from its idle mode REGULATORY BOUNDERIES The regulatory bounderies
50. than one patient clean and disinfect by wiping with a cloth dampened with 70 alcohol PREVENTIVE MAINTENANCE For safe and trouble free use of Decapus III preventive maintenance is required at least once a year as follows e Inspection for visible damage e Inspection of the arm e Inspection of cables e Check on power supply and functional earthing e Check for current leakage e Checking the pump module s vacuum functions e Checking the signal modules valve and filter functions e If the system is equipped with a battery Check battery and charger Rectify any faults If in any doubt consult authorised service personnel For more information please see the Service manual Decapus III SERVICE MANUAL DECAPUS III Important safety information contd SERVICE Repairs to Decapus lll or service measures if Decapus III is not working must be performed by Quickels Systems AB or our appointed and authorised agents see www quickels com If any parts are replaced original spares must be used Measures carried out must bedocumented showing what has been done when where and by whom company and signature WASTE HANDLING At the end of their useful lifetime the product and any replaced parts must be treated as electronic waste WEEE in accordance with the local system for recycling Other materials such as paper and plastic are recycled as such Contact your local authority or the supplier for more i
51. the pump module or to the external ECG printer Also check that there are sufficient loops at the arm joints so that the cables are not strained when the arm is bent e Check that the power supply and mains cable do not show visible damage burn marks or cracks e Make sure the mains cable is plugged into the same earthed socket multiple socket as the ECG printer e Check the resistance of the system See section Measuring instrument QN D750 e Check the insultaion of the system to shield See section Measuring instrument QN D750 e Check the buttons on the patient module s control panel by running the Run button test See section Service application PC e Make sure that the patient cables do not show e Check the the vacuum function of the pump visible damage especially where the signal module by running the Run pump test module is connected See section Service application PC e Check that the signal modules do not show If Not OK check for leaks in the system Are all visible damages signal modules tightly fixed Is the motor sound e Check the inside of the patient cable connector high etc See section Troubleshooting so that no salt bridges are visible replace if so e Check the battery status by running the test Battery status e Update the pump module s motherboard with See section Service application PC the latest firmware See section Service application PC Correct any faults 23
52. the selected vaccum level on the control panel The processor block regulates the control signal to 13 Patient module contd the vacuum pump so that the selected underpressure is obtained The controller in the processor block is of the PID type FLOW COMPENSATION The flow compensation is executed by measuring deviations from the nominal pwm signal for a certain selected vacuum level of processor block and interpreted as there is a leak somewhere The processor block increases the PWM signal to compensate for the pressure drop that occurs from the Quickels electrode to the pressure gauge The pump motor is stopped when the on off button on the control panel is pressed or when one hour has elapsed since the last time any button on the control panel was pressed down VACUUM MANIFOLD amp ECG CURCUIT BOARD The negative pressure that is selected on the control panel are distributed through rubber bushings to the patient cables and on to the electrodes The ECG signal detected from the patient is transmitted back through the patient cables and linked to the ECG circuit board via rubber bushings The shield on the patient cables is connected via the ECG shield distributor to the ECG circuit board The ECG circuit board has built in protection for defibrillator voltages and it is constructed so that it can not be damaged even if the defibrillator pads are in contact with the Quickels electrodes Leakage in the patient module m
53. tient Check errors If the service needed indicator lamp on Decapus Il is lit the error is shown in the information box See section Possible error messages in the Service application 19 O O O Service applicatio SERVICE SETTINGS n P contd In order to access service settings and related features you must be logged into the Service application Setingg Update firmware USER NAME service TT PASSWORD qsdiii2014 Update firmware Note that the system must be powered by the power adapter during the upgrade Do not unplug the USB connector until completed This process takes 3 4 minutes Update firmware You need a hex file to update Open the systems firmware Then follow the instructions Update in the information box The software can be found on the supplied USB stick or www quickels com Note that if you download the files via the Website these must be saved to your computer before use Working sensor value at normal air pressure 1 0 Bar Working sensor value is the ameded pressure sensor value e The pump function should be turned off e The value must be 2 83 0 05 V If the value differ the system must be calibrated by changing Zero offset and run calibration so that the working sensor value becomes 2 83 Service settings Serial No State the serial numbers of the system Note This may only be changed if the system is replaced with a new motherboard that needs to be l
Download Pdf Manuals
Related Search
Related Contents
Catalogue VCF 2014 Galactic RGB-600 Value Line MANUEL DE L`OPÉRATEUR & MANUEL DE PIÈCES Mélangeur Samsung HT-C460 User Manual C628 User Manual Samsung RZ28H6150WW freezer Graef Euro 2720-3310 Bedienungsanleitung Operation & Installation Manual 09/01/2005 ALFA LUZ - Alliance Pastorale Copyright © All rights reserved.
Failed to retrieve file