INSTRUCTIONS FOR USE SERVICE MANUAL
Contents
1. Description Order Date Linak Control Box and Clip Standard 3 Prong Power Cable PENDANT HAND CONTROLLER Description Order Date 999 0808 305 Non Backlit Pendant Std Model 999 0791 000 Pendant Holster Std for all pendants 1 Holster 999 0778 204 Part Number 999 0779 300 999 0775 208 Part Number 19 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 VY 2 BASIC AMERICAN LIBERTY SERIES MEDICAL PRODUCTS AUT SERVICE REPLACEMENT PART RECORD WITH LINAK ELECTRONICS MISCELLANEOUS SERVICE BED PARTS Description QTY Order Date 1 Plate Part Number 999 0822 012 999 0822 031 1 Unit 999 0822 036 1 Rod 100 8050 008 1 2 E Style Retaining Rings for Leveler Rods 1 Ring 999 0822 001 1 Unit 999 0822 180 1 Unit 999 0822 083 1 Wire 100 5710 019 100 1100 018 100 1100 060 Spanner Plates Pivot Channels Leveler Rods Mattress Retainer Wireform Wallsaver 18GA Ground Wire 10 24 375 HWHTC Ground Screws 23 Standard Cable Ties Wing Tip Cable Ties CASTER CARRIAGE Part Number Description QTY Order Date 999 0822 905G 1 Unit 100 6330 062 3 Twin Locking Caster with 7 16 14 Stems LBT7174 1 Caster 100 6330 061 3 Twin Non Locking Caster with 7 16 14 Stems LBTL7174 1 Caster 399 0832 980G One Step Locking Mechanism LBTL7174 aw 100 6744 001 7 16 14 Nylon Lock Nuts secure caster stems to frame CAPS amp
2. VY BASIC AMERICAN LIBERTY SERIES a MEDICAL PRODUCTS A ex y a vs LINAK ELECTRONIC TROUBLESHOOTING Head Motor HEAD OR FOOT MOTOR MALFUNCTIONING 999 0822 052 DECK DOES NOT GO UP AND OR DOWN 1 Check All Motor Cable Connections 2 Check Motor Condition Cracked Housing Scratches Bent Shaft etc 3 Hi Lo amp Foot Works But Head Does Not a Unplug Power Cord b Remove Cord Guard at Control Box c Switch Head amp Foot Motor Cables at the Control Box d Replug Power Cord and Test Foot Motor 4 Hi Lo amp Head Works But Foot Does Not 999 0822 053 a Unplug Power Cord b Remove Cord Guard at Control Box c Switch Foot amp Head Motor Cables at the Control Box d Replug Power Cord and Test 5 Motor Grinding Loudly Replace Motor HI LO MOTOR MALFUNCTIONING ENTIRE BED DOES NOT GO UP AND OR DOWN 1 Check All Motor Cable Connections 2 Check Motor Condition Cracked Housing Scratches Bent Shaft etc 3 Head amp Foot Motor Works But Hi Lo Does Not a Unplug Power Cord b Remove Cord Guard at Control Box c Unplug Hi Lo Cable at the Control Box and plug into empty port d Replug Power Cord and Test If Hi Lo Works Replace Control Box If Hi Lo Doesn t Work has a Grinding Noise Replace the Hi Lo Motor Hi Lo Motor 999 0778 051 A CONTROL BOX MALFUNCTIONING 1 Check All Motor Cables and Power Cord Connections 2 Hi Lo Motor Works But Head or Foot Do
3. SA gt lt BASIC AMERICAN MEDICAL PRODUCTS INSTRUCTIONS FOR USE SERVICE MANUAL To avoid personal injury or damage to bed please read all sections pertaining to your bed model before use ETL UL CSA APPROVED UL 60601 1 IEC 60601 2 38 CSA 22 2 601 1 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 This service manual covers the following Liberty Models Liberty 76 amp 80 Full Electric with Standard Caster System amp Grid Decks LBT7174 76 amp 80 Full Electric with Locking Caster System amp Grid Decks LBTL7174 WY gt WT BASIC AMERICAN MEDICAL PRODUCTS Basic American Medical Products 336 Trowbridge Drive Fond du Lac WI 54937 For Liberty Bed Service Parts please contact our Customer Service Department at 1 800 365 2338 GF Health Products Inc 2935A Northwest Parkway Atlanta GA 30360 www grahamfield com To order a Liberty Bed please contact a Graham Field Sales Representative at 1 800 554 9215 IMPORTANT NOTICE GF Health Products Inc is not responsible for typographical errors All illustrations specifications packaging and warranties contained in this Service Manual are based on the latest product information available at the time of printing The most current product information can be found online at www grahamfield com Please check all parts for shipping damage and test before using In
4. bracket From the inside secure with the Hair Pin Clips on the other side of the Lanyard Assembly 306 5 For Fixed Position Option 2 insert the 5 8 11 x 4 50 Cap Screws 4 from the outside through the holes in the rail and bracket From the inside secure the screws with 5 8 11 Nylon Lock Nuts 5 A WARNING When assessing the Risk of Entrapment you need to consider your bed mattress headboard and footboard assist devices 1 6 rails and assist bars and other accessories as an entire system bed systems are evaluated for full compliance to the FDA CDRH Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment guidelines It is extremely important to review the resident patient s physical and mental conditions and initiate an appropriate individual care plan to address any entrapment risks Please see page 7 for further information GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 E udin cado LIBERTY SERIES a MEDICAL PRODUCTS A ex y a vs OPTIONAL HALF RAIL INSTALLATION FOOT END KIT ZA82900 Includes Two Rails Two Cross Tubes Two Mounting Brackets and Hardware Hdwr Bag 999 0820 905 3 IMPORTANT Lubricate all pivot joints as needed The manufacturer pes Quick Release recommends Aerosol White 277 Hardware Lithium Grease INSTALLATION 1 All Half Rails LH amp RH mount to the bed in a simil
5. HEALTH identifying your 5 make and model PRODUCTS INC ees when ordering replacement parts The ATLANTA GA 30360 087177 Serial Number is essential if you are EE Era EUM claiming parts or service under warranty MODE OF OPERATION Intermittent e 120 5060 4 AC These helpful labels can be located on _ 450 the main frame rails immediately below N SAFE WORKING LOAD IPX4 000000000000 the sleep decks on either side of the bed Please have this IMPORTANT infor LIBERTY 701 SERIAL LBT7 L 1 14 PENEINGD mation ready when calling our customer LIBERTY 7174 service or technical support staff at MODEL 701 0000 000000 DATE 08 31 11 800 365 2338 it will allow us to better ELTE HI ET 000000000000 assist you and quickly answer your questions and concerns LBT7174 TYPICAL LIBERTY BED IDENTIFICATION LABELS with Ilcon Electronics LBT7174 or LBTL7174 Models Manufactured After Sept 2011 Serial Numbers z to 12020004096 Bed labels are an important part of GF HEALTH identifying your bed s make and model PRODUCTS INC VOS when ordering replacement parts The ATLANTA GA 30360 09011 Serial Number is essential if you are claiming parts or service under warranty DUTY CYCLE 2 min Continuous 18 min VOLT HZ AMP TYPE 9 MODE OF OPERATION Intermittent e 120 5060 2 AC These helpful labels can be located on _ 450 the main fra
6. Locking Mechanism SECTION D Electronic Troubleshooting SECTION E Framing Service Parts Liberty Service Manual 999 0822 190D DEC 2012 LIBERTY SERIES z BASIC AMERICAN MEDICAL PRODUCTS IMPORTANT SAFETY AND WARNING INFORMATION This product isa variable height adjustable mattress The bed should ALWAYS be left in its lowest position platform which will provide comfort and convenience when unattended to reduce the risk of injury while getting for residents patients and caregivers in long term in or out of the bed care settings A Keep all moving parts free of obstructions 1 6 blankets The MAXIMUM SAFE WORKING LOAD for the sheets heating blankets pads wiring etc Liberty Bed with weight evenly distributed including bedding resident patient support surface and all A DO NOT use the assist devices as push handles for moving accessories is 450 18 the bed Assist devices can be deformed or broken if excessive side pressure is exerted Assist devices are not meant for patients considered as high risks for entrapment 1 e patients with pre existing conditions such as confusion restlessness lack of muscle control altered mental status either organic or medicinal or a combi nation thereof Additional safety measures should be considered for such high risk patients NEVER operate the bed if a Power Cord or Plug is damaged or not working properly Contact qualified Service Per
7. PLUGS NOTE CAPS amp PLUGS CAN ONLY BE PURCHASED IN SETS OF 12 100 4700 015PK 1 00 x 1 00 Square Caps Pack pkg fof 100 4700 017PK 1 00 x 2 00 Rectangular Caps Pack 12 pkg 999 0822 041 Black Tube Pivot Caps 12 pkg FOR GENERAL DIAGRAMS PLEASE SEE PAGES 44 amp 45 1 Screw 1 Tie 1 Tie Caster Carriage Weldment 20 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 z gt BASIC AMERICAN LIBERTY SERIES MEDICAL PRODUCTS SERVICE REPLACEMENT PART DIAGRAM STANDARD BED WITH LINAK ELECTRONICS HEAD BOARD HEAD POWER CABLE SECURED TO FRAME MOTOR WITH CABLE TIES CONTROL BOX dieing ATTACHED TO END HI LO MOTOR FU st uw uo pap 225 MOTOR A UP cwm Po tape P SPECIAL C BOARD PANEL mv o Wr uu MOUNTING M v z BRACKETS WIREFORM WALLSAVER FOOTBOARD PENDANT CABLE PLUGGED DIRECTLY INTO CONTROL BOX AND CABLE TIED TO FOOT EITHER SIDE OF SEAT PAN END STANDARD CASTER CARRIAGE MATTRESS WITH FOUR LOCKING CASTERS RETAINER 1 TO ORDER A REPLACEMENT 2 TO ORDER A REPLACEMENT 3 TO ORDER A REPLACEMENT HI LO MOTOR USE SERVICE PACK HEAD MOTOR USE SERVICE PACK FOOT MOTOR USE SERVICE PACK 999 0778 051SP Includes motor 999 0822 052SP Includes motor 999 0822 053SP Includes motor mounting hardware amp installation mounting hardware amp installation
8. and the side without any bends is ori bed and have purchased the 4 00 Extension Kit begin ented toward the head end as shown by first removing your existing footboard assembly from the bed 6 Position the two oval holes at the end of the pan so that they align with the last holes in the foot deck nearest the 2 Remove the two hair pin clips one on each side from curved ends the clevis pins that are holding the brackets into the main frame tubes Detach the existing brackets from the 7 Insert a Washer 9 onto each of the two Phillips Pan Head footboard by removing the four 1 4 20 x 20mm bolts from Machine Screws 10 From the top insert the screws the inside of the brackets Retain the bolts for the next with washers through the holes From the bottom screw step Discard the old brackets if you do not need them on two Flange Lock Nuts 11 to secure Tighten the nuts for future use with a 7 8 open wrench or socket wrench 3 Attach both the left hand bracket 3 and right hand bracket 4 to the inside of the footboard reusing the four bolts 5 you removed in step 2 0 4 Slide the new footboard assembly into the main frame tubes so that the second holes in the 22 brackets align with the first holes in the tubes You will use four clevis pins 6 and four hair Use last hole on top af foot pin clips 7 to secure dack te mouit the brackets to the tubes extension pan eith
9. mounting hardware amp installation instruction sheet instruction sheet instruction sheet 4 TO ORDER A REPLACEMENT 5 TO ORDER A REPLACEMENT 6 FOR NEW WALLSAVER PLEASE CONTROL BOX PLEASE USE MATTRESS RETAINER PLEASE USE 999 0822 180 999 0779 300 Includes Control USE 999 0822 001 Box and mounting clip For new Power Cord use 999 0775 208 FOR ALL OTHER SERVICE PARTS PLEASE REFER TO PAGES 19 20 WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE PLEASE HAVE YOUR BED S SERIAL NUMBER AVAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY SEE PAGE 9 FOR LOCATION OF SERIAL NUMBER ID LABEL GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY BASIC AMERICAN SSS LIBERT Y SERIES 2 EE M AA b 7 MEDICAL PRODUCTS SECTION B Liberty Beds with Electronics Models Manufactured After 09 01 2011 All Beds with Serial Numbers equal to or after SN12020004096 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY 2 lt BASIC AMERICAN Ay GF Health Products Inc www grahamfield com MEDICAL PRODUCTS LIBERTY BED MECHANICS All Models NOTE All dimensions are in a range of 25 inches e Overall Bed Length with brds amp wallsaver 83786 e Overall Bed Width with boards 36 00 Exc EX eee co ot 24 50 S
10. that they roll properly and are unobstructed Electrical Inspection of Control Box every 3 months LJ Check the external power cord that plugs into the control box for any chafing cuts or wear Make sure all attaching hardware is securely tightened Check all electrical connections for any wear or fractures Check your external or mounted battery see page 27 Replace if needed Electrical Inspection of Pendant and Staff Control every 3 months Check the pendant cord for any chafing cuts or wear Check all pendant functions Head raises and lowers properly V Foot raises and lowers properly V Entire bed raises and lowers properly v moves to Chair position properly v moves to Trendelenberg and Reverse Trendelenberg positions properly Check to make sure each button and associated function work properly i e the head section raises when the Head Up button is activated Electrical Inspection of Actuators Motors every 3 months LI Check the actuator motor cords for any chafing cuts or wear LJ Check the range of movement on all motors to ensure they do not bind in the Full Up and Full Down positions WARRANTY 10 Years on Frame 2 Years on Motors and Control Box 1 Year on Other Electronics The warranties contained herein contain all the representation and warranties with respect to the subject matter of this document and supersede all prior negotiations
11. 2 3 Twin Locking Caster with 7 16 14 Stems LBT7174 1 Caster 100 6330 061 3 Twin Non Locking Caster with 7 16 14 Stems LBTL7174 1 Caster 399 0832 980G One Step Locking Mechanism LBTL7174 aw 100 6744 001 7 16 14 Nylon Lock Nuts secure caster stems to frame CAPS amp PLUGS NOTE CAPS amp PLUGS CAN ONLY BE PURCHASED IN SETS OF 12 100 4700 015PK 1 00 x 1 00 Square Caps Pack pkg fof 100 4700 017PK 1 00 x 2 00 Rectangular Caps Pack 12 pkg 999 0822 041 Black Tube Pivot Caps 12 pkg FOR GENERAL DIAGRAMS PLEASE SEE PAGES 44 amp 45 1 Screw 1 Tie 1 Tie Caster Carriage Weldment 26 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 z gt BASIC AMERICAN LIBERTY SERIES MEDICAL PRODUCTS SERVICE REPLACEMENT PART DIAGRAM STANDARD BED WITH LINAK ELECTRONICS HEAD BOARD HEAD END POWER CABLE SECURED TO FRAME LEO MOTOR WITH CABLE TIES CONTROL BOX dieing ATTACHED TO END HI LO MOTOR FU st V 2 e gt 2 cwm Po tape P SPECIAL C BOARD PANEL mv o Wr uu MOUNTING M v z BRACKETS WIREFORM WALLSAVER FOOTBOARD PENDANT CABLE PLUGGED DIRECTLY INTO CONTROL BOX AND CABLE TIED TO FOOT EITHER SIDE OF SEAT PAN END STANDARD CASTER CARRIAGE MATTRESS WITH FOUR LOCKING CASTERS RETAINER 1 TO ORDER A REPLACEMENT 2 TO ORDER A REPLA
12. CEMENT 3 TO ORDER A REPLACEMENT HI LO MOTOR USE SERVICE PACK HEAD MOTOR USE SERVICE PACK FOOT MOTOR USE SERVICE PACK 999 0848 051SP Includes motor 999 0844 052SP Includes motor 999 0844 053SP Includes motor mounting hardware amp installation mounting hardware amp installation mounting hardware amp installation instruction sheet instruction sheet instruction sheet 4 TO ORDER A REPLACEMENT 5 TO ORDER A REPLACEMENT 6 FOR NEW WALLSAVER PLEASE CONTROL BOX PLEASE USE MATTRESS RETAINER PLEASE USE 999 0822 180 999 0848 300 Includes Control USE 999 0822 001 Box and mounting clip FOR ALL OTHER SERVICE PARTS PLEASE REFER TO PAGES 25 26 WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE PLEASE HAVE YOUR BED S SERIAL NUMBER AVAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY SEE PAGE 9 FOR LOCATION OF SERIAL NUMBER ID LABEL GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 117 lt BASIC AMERICAN SERIES 2 E X EDICAL PRODUCTS SECTION C Liberty Bed Accessories All Models GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 E udin cadi LIBERTY SERIES a MEDICAL PRODUCTS A ex y a vs OPTIONAL FIXED ASSIST BAR INSTALLATION HEAD END KIT ZA80300 Includes One Assist Bar and Hardware Bag INSTA
13. CK USE SERVICE PACK 999 0822 920SP includes deck SERVICE REPLACEMENT PART DIAGRAM DECKING E hardware HEAD BACK DECK Channel amp installation 999 0822 920G instructions DETAIL A TO ORDER A PIVOT CHANNELS KN DECK USE viewed from bottom SERVICE PACK 999 0822 940SP includes deck KNEE DECK hardware 242 0208 2 amp installation MAIN E instructions Not a Service Part TO ORDER A REPLACEMENT FT DECK USE SERVICE PACK 999 0822 930SP includes deck s hardware amp installation instructions MISC PARTS DETAIL a Leveler Rod HEAD amp KMEE DECK MAIN FRAME at Deok 999 0822 036 SEAT PAN RS eae b Retaining Rin Main Frame Mounting Bracket DETAIL D LEVELER ROD c Mach Screw FOOT DECK 77 MAIN FRAME RAILS 100 5431 002 KNEE DECK viewed from bottom d Self Nut Insert E Style 10 From the outside insert Main Frame Seat Pan 999 0032 010 onto the end of Leveler Rod ater Deck FI rod bad od amp foot deck mounting 999 0808 455 nee Deck Flange inserted into frame f Circle Cotter Foot Deck Flange 100 7518 001 5 16 18 Machine Screws s perennes _ 8 Ls 5596 Rona Res WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE PLEASE HAVE YOUR BED S SERIAL NUMBER A
14. E amp FOOTDECKANGLE DOWN BUTTON 2 OPERATION 1 To angle the Head Deck upward PRESS 1 2 To angle the Head Deck downward PRESS 2 3 To raise the Bed Up horizontally PRESS 3 4 To lower the Bed Down horizontally PRESS 4 5 To angle the Knee Foot Decks upward PRESS 5 6 To angle the Knee Foot Decks downward PRESS 6 Male End Connector of Pendant Cable 1 a NON BACKLIT PENDANT Pendant Holster Model HB00 999 0791 000 Part 999 0808 305 NOTE The easiest method for deactivating the Pendant function is to simply unplug the power cord from the wall socket or other power supply when the patient is left alone in the bed When you need to use the Pendant function to raise or lower the bed or decks replug the power cord into the power outlet GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 E udin cadi LIBERTY SERIES MEDICAL PRODUCTS LIBERTY BED HI LO amp DECK POSITIONS The Liberty bed has a variety of positions that serve to help the patient be more comfortable and gives the practitioner caregiver easy access during procedures and transfers The correct placement of the bed is essential for the patient s safety The bed should always be set to its lowest position when the resident patient is left unattended The head back and foot knee functions allow the FOOT resident patient to comfortably raise their head or legs In the event o
15. LLATION 1 With the Assist Bar 1 bracket facing inward toward Insert the Hair Pins on the opposite end of the Lanyard the bed as shown align the two large holes on the Assemblies 2b into the small holes at the ends of the bracket with the two holes in the head deck s welded clevis pins mounting bracket 3 For the Fixed Bar Option 2 bottom of page insert the 2 For the Quick Release Option 1 insert the two Clevis two cap screws 3 from the outside through the Assist Pins of the two Lanyard Assemblies 2a from the Bar bracket and mounting bracket holes Secure screws outside through the Assist Bar bracket and mounting by attaching two Thin Nylon Lock Nuts 4 and tighten bracket holes using two 13 16 wrenches ASSIST BAR 999 0803 902G SHOWN AT HEAD END ON RIGHT SIDE OF BED Hardware Bag 999 0803 905 includes both quick release and fixed hardware options Main Frame E 2 d 2 A BE QUICK RELEASE Mounting Bracket FIXED ATTACHMENT 65 Pre welded on deck OPTION 2 HEAD END WARNING When assessing the Risk of Entrapment you need to consider your bed mattress headboard and footboard assist devices 1 rails and assist bars and other accessories as an entire system All bed systems are evaluated for full compliance to the FDA CDRH Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrap ment guidelines It is extremely im
16. LOW Height i 9 95 Maximum Head Back Deck Angle 70 Maximum Knee Foot Deck Angle 307 e Maximum Safe Working Load 450 Ibs WITH WEIGHT EVENLY DISTRIBUTED includes bedding resident support surface and all accessories Mass of bed without assist devices or boards 182 Ibs Bed height calculated from floor surface to top of sleep deck LIBERTY ILCON ELECTRONICS After 09 01 2011 Mfg Date All Beds with Serial Numbers equal to or after SN12020004096 Power Frequency 120 Volt 50 60 Hz e Output Rating 24V IPX4 Overall Movement Draw 2 00 Amps e Classification Class I Type B e Electrical 18 AWG 3 Conductor Type SJT Mode of Operation 10 Max Duty Cycle 2 minutes on 18 minutes off LIBERTY SERIES Liberty Service Manual 999 0822 190D DEC 2012 LIBERTY SERIES z BASIC AMERICAN MEDICAL PRODUCTS VENDOR MODEL Control Junction Box 00 Part 999 0848 300 PENDANT Quantity 1 Part 999 0808 305 Quantity 1 Power Cable is not detachable hard wired into the side of the Control Box Bed also includes Pendant Holster Part 999 0791 000 Quantity 1 o Head Motor Actuator Part 999 0844 052 ee 20 E E Pendant Cable to Control Box Part 999 0848 200 Quantity 1 Knee Foot Motor A
17. TIONAL PIVOTING ASSIST BAR INSTALLATION HEAD END ONLY KIT ZA85500 Includes One Pivoting Assist Bar and Hardware INSTALLATION 1 On the head end align the two holes in the Pivoting Assist Bar 2 bracket with the two holes in the mounting bracket on the underside of the Head Deck 1 2 From the outside insert the two 4 00 Clevis Pins 3 through the holes in the Pivoting Assist Bar bracket and Head Deck mounting bracket Head End Mounting Bracket 3 From the inside insert a 1 pre welded to Hea Deck 3 2 375 Hair Pin Clip through X the small holes at the ends of each of the Clevis Pins to secure HEAD END Recommend Toward Headboard Black Knob to Lock or Release FOR YOUR SAFETY MAKE SURE YOUR FINGERS ARE CLEAR OF THE SIDES OR UNDERNEATH THE OPERATION 1 To release the Pivoting Assist Bar from its vertical lock position hold onto the top of the bar with one hand and slightly pull out the black knob on the outside of the assembly with the other hand to release the locking mechanism For your safety make sure your fingers are not near the pivot assembly when pivoting the rail For ease of patient access it is recommended that you pivot the rail toward the head board until it stops and rests in place To set the assist bar back to its upright locked position grab the top of the assist bar with one hand and pivot the assembly upward until the black knob mechanism snaps in
18. VAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY SEE PAGE 9 FOR LOCATION OF SERIAL NUMBER ID LABEL GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 Av LIBERTY SERIES BASIC AMERICAN MEDICAL PRODUCT SERVICE REPLACEMENT PART DIAGRAM MAIN FRAME CAPPING amp WING STYLE CABLE TIES 1 00 Square Tube End Cap 100 4700 015 1 00 x 2 00 f Rectangular a HEAD END Tube End Cap e B 100 4700 017 m E Right side of Bed b 1 00 Square 1 00 Square M Tube End Cap Tube End Cap SU let 100 4700 015 100 4700 015 2 E 3 100 4700 015 2 1 00 Square Tube End Cap Wing Style Cable Ties for Power Cord 100 4700 060 1 00 x 2 00 Rectangular 1 00 Square Tube End Cap Tube End Cap 100 4700 017 m 100 4700 015 ain Frame Not a Service Part 1 00 Square Tube End Cap 100 4700 015 FOOT END Right side of Bed CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS PLUGS ON YOUR BED SEE BOTTOM OF PAGE 26 FOR ORDER NUMBERS GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 NY LIBERTY SERIES BASIC AMERICAN MEDICAL PRODUCT SERVICE REPLACEMENT PART DIAGRAM CAPPING AND CASTER PARTS ON STANDARD CASTER CARRIAGE BED MODEL LBT7174 7 16 14 Nylon Lock Nu
19. afe Working Load A Consult Accompanying Documents Double Insulated GF Health Products Inc www grahamfield com Protected Grounded Device Type B Equipment amp Applied Parts Liberty Service Manual 999 0822 190D DEC 2012 WY gt BASIC AMERICAN MEDICAL PRODUCTS 14 LIBERTY SERIES RECOMMENDED MAINTENANCE AND INSPECTION Maintenance Inspection of All Components Receipt of shipment LJ Make sure all parts components are included Please see Unpacking Your Bed Page 15 Check all bed components for obvious damage Inspect the Power Supply Cord for any cuts and or damage Check to see all actuator motor cords are routed and connected properly to the control box Mechanical Inspection of Assemblies every 6 months Inspect all welds on the sleeping surface frame and base assemblies for stress fractures 1 Inspect all fasteners for wear and looseness LJ IMPORTANT Lubricate all pivot points actuator motor clevis pins and control arm clevis pins as needed White Lithium Grease 1s recommended Mechanical Inspection of Casters amp Pedal Locking Mechanism every 3 months L Check the pedal locking mechanism to make sure it engages and disengages properly Check the bottom foot pads on the foot end casters making sure they are clean of debris and undamaged Replace if needed Check all casters to ensure
20. agreements and understanding with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any representation assertion guarantee warranty collateral contract or other assurance except those set out in this document GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 z lt BASIC AMERICAN MEDICAL PRODUCTS LIBERTY SERIES ENTRAPMENT amp COMPLIANCE INFORMATION On March 10 2006 the FDA U S Food and Drug Adminis tration released long awaited guidelines for reducing the risk of bed entrapment Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment The new Guidance identifies potential entrapment areas and those body parts most at risk for entrapment provides design criteria for manufacturers of new hospital convalescent beds recommends particular test methods to assess the conformance of existing hospital conva lescent bed systems and answers frequently asked questions about entrapment issues The new Guidance was a result of a long standing collaboration between the FDA and the Hospital Bed Safety Workgroup HBSW formed in 1999 GF Health Products Inc s Long Term Care Bed division Basic American Medical Products 1s an HBSW charter member As a result of our commitment to product safety all our current long term care beds have been strictly tested and conform to the new FDA Guida
21. ain frame rails See Detail A 4 From the outside of the main frame rails insert a 5 16 x 1 50 Clevis Pin 6 through the holes in the frame rails and brackets on both sides of the bed 5 Secure the clevis pins by inserting a 1 9 16 Hair Pin Clip 7 through each of the holes in the ends of the four pins HEADBOARD M MOUNTING SP inserts are pre LE ina When DETAIL A 80 Board Bracket Position FOR 80 BED 2 ordering boards 76 Board Bracket 4 MOUNTING oard Br puo POSITIONS HEAD END FOOT END FOOTBOARD sen teal HEADBOARD HEAD END T 4F a insets pre inserted when ordering boards Adi Li he 7 putsade 21 8 DETAILB faders MOUNTING FOOTBOARD MOUNTING FOR 80 BED END inane male NOTE To order a new Panel Mounting Kit please use part 999 0822 000 Kit includes a right hand and left hand Mounting Bracket 8 Hex Drive Bolts 4 Clevis Pins 4 Hair Pin Clips and 1 Hex Allen Wrench Both Panels Head board and Footboard are sold separately 2121 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 z gt BASIC AMERICAN T MEDICAL PRODUCTS LIBERTY SERIES STANDARD FULL WIDTH MATTRESS RETAINER INSTALLATION The Liberty Bed s mattress retainer was designed to span the width of the sleeping surface One retainer is shipped
22. al 999 0822 190D DEC 2012 WY 2 5 BASIC AMERICAN a MEDICAL PRODUCTS A ex y a vs LIBERTY SERIES TYPICAL SAFETY WARNING BED LABELS The following warning labels have been placed on your bed to help protect you and your bed from damage Please do not remove any labels from your bed WARNING DO NOT LOWER BED WHEN OBJECTS ARE BENEATH BED FAILURE TO INSPECT UNDER BED CAN RESULT IN DAMAGE TO PROPERTY OR PERSONAL INJURY ATTENTION S assurer de ne pas faire descendre le lit lorsque des objets se trouvent sous le lit Ne pas inspecter le dessous du lit pourrait entrainer des dommages materiels et des risques de blessures WARNING DO NOT LOWER BED WHEN OBJECTS ARE BENEATH BED FAILURE TO INSPECT UNDER BED CAN RESULT IN DAMAGE TO PROPERTY OR PERSONAL INJURY CAUTION THIS BED IS SUITABLE FOR USE ONLY WITH OXYGEN ADMINISTERING EQUIPMENT OF THE NASAL MASK TYPE TENT COVERING ONLY THE UPPER HALF HEAD END OF THE BED OXYGEN TENT CANOPIES SHOULD NOT EXTEND BELOW BED SPRING LEVEL LOCK HAND CONTROL AT FOOT OF BED WHEN USING OXYGEN ADMINISTERING EQUIPMENT ATTENTION CE LIT PEUT ETRE UTILISE UNIQUEMENT AVEC UN EQUIPMENT DESTINE A L ADMINISTRATION D OXYGENE DE TYPE NASAL OU MASQUE OU AVEC UNE TENTE RECOUVRANT SEULEMENT LA MOTTIE AVENT DU LIT LES COTES DE LAS TENTE OXYGENE NE DOIVENT PAS SE PROLONGER PLUS DAS QUE LA SOMMIER DU LIT LABEL SYMBOL DEFINITIONS LN AN S
23. ar fashion Left or right designations are based on the patient s LATCH FOR orientation as he she RAISING AND lies in the bed LOWERING RAIL 2 To mount the half rails to the foot end Foot Deck of the bed you will also need to attach two rail brackets 4 and two cross tubes 5 to the foot deck Please see installation instructions on page 30 of this service manual Align the holes in the rail s pivoting arms with the holes in the mounting brackets as shown 4 For Quick Release Option 1 insert the Clevis Pin ends of the Lanyard assembly 323a from the outside through the holes in the pivot arms and mounting bracket From the inside secure with the Hair Pin Clips on the other side of the Lanyard Assembly 36 5 For Fixed Position Option 2 insert the 5 8 11 x 4 50 Cap Screws 6 from the outside through the holes in OPTION 2 the rail and bracket From the inside secure the screws with 5 8 11 Nylon Lock Nuts 7 WARNING When assessing the Risk of Entrapment you need to consider your bed mattress headboard and footboard assist devices 1 e rails and assist bars and other accessories as an entire system All bed systems are evaluated for full compliance to the FDA CDRH Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment guidelines It is extremely important to review the resident patient s physical and mental conditions and initiate an a
24. case of damage DO NOT USE WY 22 lt AMERICAN Ay GF Health Products Inc www grahamfield com MEDICAL PRODUCTS LIBERTY SERIES TABLE OF CONTENTS Safety amp Warnings Warning Labels 2 RACE Soe ee ae ae Recommended Maintenance Liberty Warranty SOR OC orans Entrapment Compliance Unpacking Your Bed Liberty ID Serial Liberty Headboard and Footboard Installation Standard Mattress Retainer Installation Standard Wallsaver Installation Pendant Hand Controller Operation Liberty Bed Positions and Deck Elevations One Step Locking Mechanism Operation SECTION A Liberty with LINAK Electronics Mechanical amp Electronic Specifications LINAK Electrical Components LINAK Service Part Record Bed Diagram with LINAK Electronics SECTION B Liberty with ILCON Electronics Mechanical amp Electronic Specifications Electrical Components Service Part Record Bed Diagram with ILCON Electronics SECTION C Assist Devices Accessories Fixed and Pivoting Assist Bar Installation Foot end Mounting Bracket and Cross Tubes Metal Half Rail Installation Trapeze Support Installation Foot end Extension Pan 4 007 One Step
25. ctuator NOTE Part 999 0844 053 MOTOR CABLES AND Quantity 1 POWER CORD ARE NOT DETACHABLE SEPARATE ITEMS CABLES ARE HARD WIRED DIRECTLY INTO MOTORS AND CONTROL BOX Hi Lo Motor Actuator 4 Part 999 0848 051 Quantity 1 A DO NOT use unauthorized parts accessories or adaptors other than those specified authorized by GF Health Products Inc DO NOT open assemblies such as the Actuators Hand Control or Control Box If unauthorized personnel perform work on these components the manufacturer s warranty becomes void NEVER operate the bed if a Power Cord or Plug is damaged or not working properly Contact qualified Service Personnel for examination and repair Always unplug the Power Cord when performing any maintenance on the bed GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 P BASIC AMERICAN LIBERTY SERIES MEDICAL PRODUCTS SERVICE REPLACEMENT PART RECORD WITH ILCON ELECTRONICS DECKING Partnumber OrderDato Proe _ e _ _ _ vem 8 ek _ _ 1 Deck Hardware Pack INCLUDES 1 Pkg 5 16 Retaining Rings Deck Brackets MOTORS ACTUATORS Part Number Deeipon OrderDate _ Hi Lo Motor Cable hard wired amp Hardware Pack tMotor fof OR Head Motor Cable hard wired amp Hardware Pack fof OR Foot M
26. ducts Inc All Rights Reserved Basic American Medical Products and Zenith are registered trademarks of GF Health Products Inc GF121699RevA12
27. e and can skip steps 5 8 5 Remove the two locking casters at the foot end of the carriage and replace with two non locking casters 9 from your kit Use the same Locking Nuts to secure 6 Remove the two locking casters at the head end of the carriage Discard the nuts and locking casters From the bottom insert the stems from the last two non locking casters from your kit into the carriage weldment holes GF Health Products Inc www grahamfield com Re use existing Lock Muts _ fo attach non lacking casters 7 at foot end of caster carriage View of Pedal Lock Attached to the Foot End of Liberty Bed View of Caster Keeper Assembly attached to the Head End of Liberty Bed Caster Keepers come pre assembled in kit 7 Position the pre assembled Caster Keepers bent arm outward and toward the foot end onto the exposed ends of each of the caster stems The brackets should rest securely on the carriage cross brace Secure the brackets to the carriage using two 2 new 7 16 14 Thin Nylon Lock Nuts from your kit NOTE A Second Locking Mechanism can be attached to the opposite side Order part 999 0832 980G Liberty Service Manual 999 0822 190D DEC 2012 VY P BASIC AMERICAN uu LIBERTY SERIES 2 E X EDICAL PRODUCTS SECTION D TROUBLESHOOTING All Liberty Models GF Health Products Inc www grahamfield com Liberty Service Manua
28. e issues The APS 9000 bed and accessories listed in this manual are in full compliance with FDA guidelines for reducing the risk of bed entrapment Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment Details can be found at www fda gov Liberty Service Manual 999 0822 190D DEC 2012 z gt BASIC AMERICAN 2 MEDICAL PRODUCTS LIBERTY SERIES UNPACKING YOUR BED DISCARD KEEP 1 Cardboard Block 5 Pendant Holster Pendant e Make sure all parts components are included 2 Notched Leg Foam s Power Cord Coiled iar Check all bed components for obvious damage 3 carge Soipping and tied To 4 Large Cable Tie CUT Grid Wire on Head Deck Inspect the Power Supply Cord for any cuts and 6 One Mattress Retainer Or damage 7 One Wireform Wallsaver 8 Panel Mounting Brackets Check to see all actuator motor cables are routed amp Hardware for attaching and connected properly to the control box 9 Service Manual Documents ALSO DISCARD TOP AND BOTTOM CARTONS mem GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY 5 BASIC AMERICAN LIBERTY SERIES MEDICAL PRODUCTS TYPICAL LIBERTY BED IDENTIFICATION LABELS with Linak Electronics LBT7174 or LBTL7174 Models Manufactured Prior to Sept 2011 Serial Numbers to 12020004096 Bed labels are an important part of GF
29. ed Failure to inspect under the bed can result in damage to property or personal injury Operating Conditions Operation of the bed is based on the following conditions Ambient Temperature of 10 C to 40 C Relative Humidity of 30 to 75 Non condensing Atmospheric Pressure of 700hPa to 1060hPa and a Splash Protection of IEC 60529 The bed s Pendant Cord MUST BE ROUTED AND SECURED PROPERLY to ensure it does not become entangled and eventually severed during use Also make sure all electrical cords DO NOT get tangled gt PP PPP around the bed side rails or legs during transport or normal operation of the bed When using nasal type or masked type administering equipment all oxygen or air tubing MUST BE ROUTED AND SECURED PROPERLY to ensure that the tubing does not become entangled and eventually severed during the normal operation of the bed GF Health Products Inc www grahamfield com Storage Storage of the bed is based on the following conditions Ambient Temperature of 10 C to 70 C Relative Humidity of 10 to 100 and an Atmospheric Pressure of 500hPa to 1060hPa Radio Frequency Interference RFI RFI influences most electronic equipment Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment If RFI causes erratic behavior shut the bed off immediately Leave it off while the transmission is in progress Liberty Service Manu
30. er reuse the two 2 Foot Deck sets you removed in Main Frame step 1 and two more FOOT Tube from your kit or all END four from your kit Both bracket holes are SS _ used to mount the board an 3 assembly to the bed Insert the brackets so that the second holes align with the first holes in the main frame tubes Second hole ar First hole used used for mounting for mounting GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY 2 gt BASIC AMERICAN E S BASIC AMERIC LIBERTY SERIES ONE STEP LOCKING MECHANISM INSTALLATION MODEL LBTL7174 Please Order Kit Number LBTL83200 Includes Pedal Lock Caster Keepers 4 Non Locking Caster amp Hdwr Make sure Footboard is positioned in 1st hale m main frame rails for BO bed 4 Position the Locking Pedal Assembly 2 so that the mounting bracket is on top and the red pedal is facing outward away from the carriage weldment 2 Align the holes in the bracket with the pre drilled holes in the cross brace From the outside insert a 5 16 18 x 2 50 Hex Head Bolt 3 through both holes in the bracket and cross brace 4 From the inside secure the bolt to the assembly using a 5 16 18 Thin Nylon Lock Nut 4 Tighten with two 1 2 wrenches If you already have a lock mechanism system on your Liberty bed and are attaching a second Pedal Lock you are don
31. es Not See Head amp Foot Motor Troubleshooting above If determined that port is bad replace Control Box 3 Head amp Foot Motor Works But Hi Lo Does Not See Hi Lo Motor Troubleshooting above If determined that port is bad replace Control Box Control Box 999 0779 300 Pendant Cable Port HB 4 This Port Not Used zm Foot Motor Cable Port 3 P 2d NS Hi Lo Motor Cable Port 2 Cable Bond GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 P E eun LIBERTY SERIES MEDICAL PRODUCTS P ux ILCON ELECTRONIC TROUBLESHOOTING Head Motor 999 0844 052 HEAD OR FOOT MOTOR MALFUNCTIONING DECK DOES NOT GO UP AND OR DOWN 1 Check All Motor Cable Connections 2 Check Motor Condition Cracked Housing Scratches Bent Shaft etc 3 Hi Lo amp Foot Works But Head Does Not a Unplug Power Cord b Remove Cord Guard at Control Box c Switch Head amp Foot Motor Cables at the Control Box d Replug Power Cord and Test 4 Hi Lo amp Head Works But Foot Does Not a Unplug Power Cord b Remove Cord Guard at Control Box c Switch Foot amp Head Motor Cables at the Control Box d Replug Power Cord and Test 5 Motor Grinding Loudly Replace Motor Foot Motor 999 0844 053 HI LO MOTOR MALFUNCTIONING ENTIRE BED DOES NOT GO UP AND OR DOWN 1 Check All Motor Cable Connections 2 Check Motor Condition Cracked Housing Scra
32. f a power HEAD outage it is preferred that the knee and head back sections be lowered END END using an emergency battery pack i P _ T ER 0 ae E To W 8 f 24 50 QUICK REFERENCE Bed Height floor to sleep deck to MAXIMUM High Position 24 50 MINIMUM Low Position 9 95 Head Deck to Sleep Surface MAXIMUM Head Angle 70 Knee Deck to Sleep Surface MAXIMUM Knee Angle 30 Knee Deck to Foot Deck MAXIMUM Knee Foot Angle 45 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 z gt BASIC AMERICAN m MEDICAL PRODUCTS LIBERTY BED ONE STEP LOCKING MECHANISM OPERATION INSTRUCTIONS Step on Red Pedal Lock the Bed _ Ec a Step an the Green Release Lever to Unlock the Bed _ Ec GF Health Products Inc www grahamfield com Locked Position on Your Liberty Bed To lock the foot end of the Liberty bed step on the red pedal until the lock engages The green release lever will show through the slot in the top of the locking mechanism To fully lock the bed make sure the caster keepers at the head of the bed are both in their down positions Unlocked Position on Your Liberty Bed To unlock the foot end of the Liberty bed step on the green release pedal until the lock disengages The green release lever will disappear through the sl
33. for Trapeze Support installation in the Accessories section of this manual To order additional wireform wallsavers please use part 999 0822 180 WALLSAVER ATTACHING POSITION FINAL WALLSAVER RESTING POSITION FLAT TAB IS INSERTED SO THAT IT IS PARALLEL WITH THE SLOT IN THE CASTER CARRIAGE RAIL FLAT TAB LOCKS INTO THE CARRIAGE RAIL BY TURNING IT IN SIDE HOLE UNTIL THE WALLSAVER RESTS ON CROSSBRACE DETAIL B illustrates final angle of flat insert tab rests on caster carriage cross brace DETAIL A Illustrates starting angle of flat insert tab moved to left for clarity WALLSAVER POSITIONING CHART Wallsaver hole for 80 Bed with Trapeze Wallsaver hole for 76 Bed with Trapeze cc 5 Wallsaver hole for 80 Bed without Trapeze 1 Wallsaver hole for 76 Bed without Trapeze PART NUMBER DESCRIPTION STANDARD WIREFORM WALLSAVER 1 999 0822 180 76 AND 80 Also has position for Trapeze Accessory GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 LIBERTY SERIES gt 55 BASIC AMERICAN MEDICAL PRODUCTS STANDARD PENDANT HAND CONTROLLER OPERATION Each of the Liberty Beds are equipped with a pendant type hand control standard Every control function for up and M The 2 ja deactivate the patient control functions To operate simply push the appropriate button to activate the desired function see 6 KNE
34. four holes on the wood grain side 7 Secure the Trapeze Support to the rails using the Clevis of the board Pins 6 and Hair Pin Clips 7 you removed in step 1 3 Position the headboard so that the wood grain finish is facing inward toward the head deck Align the holes in 2 the board with the bracket holes in the Trapeze Support 2 as shown 4 From the back insert the four JCB Bolts 5 you removed in step 2 through the Trapeze Support brackets into the head board They will thread into the 1 4 20 Boards typically are shipped with Inserts pre assembled inta boards To remove the inserts insert the end of the 4mm Hex Allen Wrench into the insert hole from the back of the board and lightly tap with a hammer DETAIL A TRAPEZE SUPPORT TO MAIN FRAME RAIL HOLE POSITIONS Head Headboard End Hole for 76 Bed Hole for 80 Bed GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 E udin cani LIBERTY SERIES a MEDICAL PRODUCTS A ex y a vs OPTIONAL 4 LIBERTY EXTENSION PAN INSTALLATION Please Order Kit Number ZA83000 Includes Hdwr Bag 999 0833 995 1 The Liberty Extension Pan kit includes two footboard 5 Now you can attach your extension pan to your foot deck mounting brackets that are used to extend the footboard Lay your pan on top of the foot deck so that the three bent to accomodate the pan If you already have a Liberty sides face upward
35. ior to 09 01 2011 Mfg Date All Beds with Serial Numbers prior to SN12020004096 Power Frequency 120 Volt 50 60 Hz e Output Rating 12 33V IPX4 Overall Movement Draw 4 00 Amps e Classification Class B Electrical 18 AWG 3 Conductor Type SJT Mode of Operation 10 Max Duty Cycle 2 minutes on 18 minutes off LIBERTY SERIES Liberty Service Manual 999 0822 190D DEC 2012 z lt BASIC AMERICAN MEDICAL PRODUCTS Control Junction Box Part 999 0779 300 Quantity 1 Power Cable plugs into side port part listed at bottom Head Motor Actuator Part 999 0822 052 Quantity 1 Knee Foot Motor Actuator Part 999 0822 053 Quantity 1 Hi Lo Motor Actuator Part 999 0778 051 Quantity 1 STANDARD 3 Prong Power Cable Part 999 0775 208 Quantity 1 LIBERTY SERIES VENDOR MODEL 00 PENDANT Part 999 0808 305 Quantity 1 Bed also includes Pendant Holster Part 999 0791 000 Quantity 1 25 50 Motor Cable Part 999 0808 204 Quantity 1 20 00 Motor Cable Ft Part 999 0822 204 Quantity 1 8 00 Motor Cable Hi Lo Part 999 0778 204 Quantity 1 A DO NOT use unauthorized parts accessories or adaptors other than those specified authorized by GF Health Products Inc DO NOT open assemblies such as the Actuators Ha
36. k tubes four 5 16 18 Thin Nylon Lock Nuts 6 Hand tighten at this time 6 From the bottom secure the screws to the tubes using four 5 16 18 Thin Nylon Lock Nuts 6 Tighten all hardware using a 1 2 crescent socket wrench and 3 16 allen wrench Toward Toward Head End RAIL BRACKET ORIENTATION GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 E udin cadi LIBERTY SERIES a MEDICAL PRODUCTS A ex y a vs OPTIONAL FIXED ASSIST BAR INSTALLATION FOOT END KIT ZA80350 Includes One Assist Bar and Hardware Bag INSTALLATION 1 With the Assist Bar 1 bracket facing inward toward IMPORTANT the bed as shown align the two large holes on the THE MOUNTING BRACKET AND TWO CROSS bar s bracket with the two holes in the pre assembled BRACES MUST BE ASSEMBLED ONTO THE FOOT mounting bracket on the Foot Deck see page 30 DECK PRIOR TO INSTALLING ANY ASSIST DEVICE TO THE FOOT END OF THE BED PLEASE ORDER 2 For the Quick Release Option 1 insert the two Clevis THE SEPARATE KIT IN ADDITION TO THE ASSIST Pins of the two Lanyard Assemblies 222a from the BAR KIT SEE PAGE 30 FOR INSTALLATION outside through the Assist Bar bracket and mounting bracket holes Insert the Hair Pins on the opposite end of the Lanyard Assemblies 2b into the small holes at the ends of the clevis pins 3 For the Fixed Bar Option 2 bottom of page in
37. l 999 0822 190D DEC 2012 E udin cani LIBERTY SERIES a MEDICAL PRODUCTS A ex Ade LIBERTY QUICK REFERENCE ELECTRONIC TROUBLESHOOTING The Liberty beds use a 3 motor system with either LINAK electronics prior to 9 1 2010 or ILCON electronics after 9 1 2010 otaff Control is not offered on the Liberty Bed NO POWER 1 Check Outlet 2 Unplug Power Cord and Check the following a Check Power Cord for Damage or Pinching b Check Short Black Control Box Cable for Damage or Pinching c If any Cables Damaged Replace Immediately PENDANT NOT WORKING 1 Check All Pendant and Cable Connections and retest 2 Unplug Power Cord 3 LINAK Electronics Unplug the Pendant from Control Box Use a new Pendant you know is working from another bed Plug in the power cord and test a If Pendant works Replace old pendant b If Pendant Does Not Work Replace Control Box 4 ILCON Electronics Unplug the Pendant from the Adaptor Cable Use a new adaptor cable you know works from another bed Plug in the power cord and test a If Pendant works Replace Adaptor Cable b If Pendant Does Not Work Unplug the Pendant from Control Box Use a new Pendant you know is working from another bed Plug in the power cord and test i If Pendant works Replace old pendant If Pendant Does Not Work Replace Control Box GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012
38. me rails immediately below N SAFE WORKING LOAD 000000000000 the sleep decks on either side of the bed Please have this IMPORTANT infor LIBERTY 701 SERIAL LBT7 L 1 14 PENDING mation ready when calling our customer LIBERTY 7174 service or technical support staff at 800 365 2338 it will allow us to better MODEL NO 701 0000 000000 MFG DATE 09 01 11 BALA TET 000000000000 assist you and quickly answer your questions and concerns LBT777174 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 E C Sy ere LIBERTY SERIES a MEDICAL PRODUCTS A ex y de HEAD AND FOOTBOARD PANEL ASSEMBLY 1 Assemble four inserts 4 into the outside of both the headboard and footboard if not already pre assembled 2 Assemble a right hand and a left hand Board Mounting Bracket 2 amp 3 to the insides of both the headboard and footboard using four 1 4 20 x 20mm Hex Drive JCB B Bolts 5 for each board The bolts will go through the four pre drilled holes on each board into the inserts 3 Slide the headboard and footboard with mounting brackets into the main frame rails a For a 76 bed align the holes in the brackets with the second holes from the end in the sides of the main frame rails See Detail A b For an 80 bed align the holes in the brackets with the first holes from the end in the sides of the m
39. nce The guidelines set forth by the FDA Guidance layout specific dimensional limitations on potentially injury threatening gaps and spaces that can occur between bed system components such as rails when not properly installed GF Health Products Inc and Basic American Medical Products have conformed to these guidelines from a manufacturing aspect However entrapment issues can often arise when a healthcare provider facility has not correctly assembled the components on a bed It is essential that the provider facility fully understand their responsibility in complying to the guidelines set forth by the FDA in order to avoid injury To that end we have provided the FDA s web address at right as a resource for understanding and following these guidelines for the safety of patients residents GF Health Products Inc www grahamfield com It is also essential to have the correct bed components acces sories that correspond with the needs of your patient resident and the particular bed you have purchased Matching the correct bed component that correlates with the regulatory guidelines can be a daunting task Our sales team at GF Health Products Inc and our friendly Customer Service Representatives at Basic American Medical Products can help you sift through the wide array of compliance and bed options We will help you determine which bed bed part is best for your patient s resident s particular needs and help you with your complianc
40. nd Control or Control Box If unauthorized personnel perform work on these components the manufacturer s warranty becomes void NEVER operate the bed if a Power Cord or Plug is damaged or not working properly Contact qualified Service Personnel for examination and repair Always unplug the Power Cord when performing any maintenance on the bed GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 BASIC AMERICAN L E RIY S E RI ES MEDICAL PRODUCTS AUT SERVICE REPLACEMENT PART RECORD WITH LINAK ELECTRONICS DECKING Description Order Date 999 0822 920SP Head Back Deck amp Hardware Pack 999 0822 930SP OR Foot Deck amp Hardware Pack Deck 0 999 0822 940SP OR Knee Deck amp Hardware Pack 0 Part Number MOTORS ACTUATORS Part Number Description Order Date Price _ Hi Lo Motor Cable not included amp Hardware Pack fof 999 0822 052SP OR Head Motor Cable not included amp Hardware Pack 1 Motor 999 0822 053SP OR Foot Motor Cable not included amp Hardware Pack 1 Motor Motor Hardware Pack INCLUDES 1 Pkg 12mm Retaining Rings Hair Pin Clips Snap Rings Standard Cable Ties Installation Instruction Sheet 1 Cable 999 0808 204 Head Motor Cable 999 0822 204 Foot Motor Cable 1 Cable Hi Lo Motor Cable 1 Cable 0101 115101 amp
41. ot in the top of the pedal locking mechanism and the red pedal will raise To fully unlock the bed make sure the caster keepers at the head of the bed are both in their up positions LIBERTY SERIES Liberty Service Manual 999 0822 190D DEC 2012 WY BASIC AMERICAN SSS LIBERT Y SERIES 2 EE M AA b 7 MEDICAL PRODUCTS SECTION A Liberty Beds with Linak Electronics Models Manufactured Prior to 09 01 2011 All Beds with Serial Numbers prior to 5 12020004096 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 2 gt BASIC AMERICAN a MEDICAL PRODUCTS A ex y a vs GF Health Products Inc www grahamfield com LIBERTY BED MECHANICS All Models NOTE All dimensions are in a range of 25 inches e Overall Bed Length with brds amp wallsaver 83786 e Overall Bed Width with boards 36 00 Exc EX eee co ot 24 50 SLOW Height i 9 95 Maximum Head Back Deck Angle 70 Maximum Knee Foot Deck Angle 307 e Maximum Safe Working Load 450 Ibs WITH WEIGHT EVENLY DISTRIBUTED includes bedding resident support surface and all accessories Mass of bed without assist devices or boards 182 Ibs Bed height calculated from floor surface to top of sleep deck LIBERTY LINAK ELECTRONICS Pr
42. otor Cable hard wired amp Hardware Pack tMotor fof Motor Hardware Pack INCLUDES 12mm Retaining Rings Hair Pin Clips Snap Rings Standard Cable Ties Installation Instruction Sheet CONTROL BOX amp POWER CABLES No Detachable Power Cord Available Cord hard wired to CB PENDANT HAND CONTROLLER Description Order Date 999 0808 305 Non Backlit Pendant Std Model Pendant Adaptor Cable 1 Cable Pendant Holster Std for all pendants 1 Holster Part Number Order Date 999 0848 300 Description Linak Control Box and Clip Part Number 999 0848 200 999 0791 000 25 GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 VY 2 BASIC AMERICAN LIBERTY SERIES MEDICAL PRODUCTS AUT SERVICE REPLACEMENT PART RECORD WITH LINAK ELECTRONICS MISCELLANEOUS SERVICE BED PARTS Description QTY Order Date 1 Plate Part Number 999 0822 012 999 0822 031 1 Unit 999 0822 036 1 Rod 100 8050 008 1 2 E Style Retaining Rings for Leveler Rods 1 Ring 999 0822 001 1 Unit 999 0822 180 1 Unit 999 0822 083 1 Wire 100 5710 019 100 1100 018 100 1100 060 Spanner Plates Pivot Channels Leveler Rods Mattress Retainer Wireform Wallsaver 18GA Ground Wire 10 24 375 HWHTC Ground Screws 23 Standard Cable Ties Wing Tip Cable Ties CASTER CARRIAGE Part Number Description QTY Order Date 999 0822 905G 1 Unit 100 6330 06
43. portant to review the resident patient s physical and mental conditions and initiate an appropriate individual care plan to address any entrapment risks Please see page 7 for further information GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 VA 2 gt 5 BASIC AMERICAN a MEDICAL PRODUCTS A ex y a vs LIBERTY SERIES FOOT END MOUNTING BRACKET AND CROSS TUBE INSTALLATION FOR OPTIONAL FOOT END ASSIST DEVICES KIT ZA82900 Includes Two Cross Tubes Two Mounting Brackets and Hardware Bag 999 0829 905 INSTALLATION 1 Locate the two Cross Deck Tubes 3 in your kit Align 4 Locate the two long U shaped Rail Mounting Brackets the outside oblong holes with the appropriate holes in the 4 in your kit The brackets should be oriented so that foot deck fourth and seventh holes from the foot end of the top hole closest to the edge of the bracket is furthest the foot deck away from the foot end of the bed With the open end facing downward as shown align the top holes with the 2 From the top insert a 5 16 18 x 1 75 Button Head Machine inside oblong holes on the underside of the cross tubes Screw 5 from your kit through each of the holes in both the deck and the deck tubes 5 From the top insert a 5 16 18 x 1 75 Button Head Machine Screw 5 from your kit through each of the holes in both 3 From the bottom secure the screws to the tubes using the deck and the dec
44. ppropriate individual care plan to address any entrapment risks Please see page 7 for further information GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 E udin cadi LIBERTY SERIES a MEDICAL PRODUCTS A ex y a vs OPTIONAL TRAPEZE SUPPORT INSTALLATION The Trapeze Support is attached to the head section of the bed Please Order Kit Number ZA83000 Includes Hdwr Bag 999 0830 905 1 Remove the headboard panel assembly if previously inserts Tighten all JCB bolts with the hex allen wrench assembled to the bed by pulling out the Clevis Pins DO NOT OVERTIGHTEN 6 and Hair Pin Clips 7 that attach the main frame rails to the headboard s mounting brackets Set hard 5 Assemble the Trapeze Support with headboard attached ware aside to the main frame rails by inserting the support s tubes into the rails The vertical stop at the ends of the support 2 Detach the headboard panel from the mounting tubes tubes will help keep the trapeze support level by removing the four 1 4 20 JCB Bolts 5 from the brackets using the 4mm Hex Allen Wrench included in 6 your kit Remove the four 1 4 20 Inserts 4 from the board by inserting the opposite end of the allen wrench into the hole and lightly tap with a hammer Reinstall The Liberty Bed features two positions for the Trapeze Support depending on the length of bed you have please see Detail A below the inserts into the
45. sert the two cap screws 3 from the outside through the Assist Bar bracket and mounting bracket holes Secure screws by attaching two Thin Nylon Lock Nuts 4 and tighten using two 13 16 wrenches ASSIST BAR 999 0803 902G SHOWN AT HEAD END ON RIGHT SIDE OF BED Hardware Bag 999 0803 905 includes both quick release and fixed hardware options QUICK Jr RELEASE Foot Deck CROSS TUBES SOLD with Cross Braces FOOT END SEPARATELY See foot bracket brace assembly amp installation instructions WARNING When assessing the Risk of Entrapment you A need to consider your bed mattress headboard and footboard 7 assist devices 1 rails and assist bars and other accessories 29 as entire system All bed systems are evaluated for full m P compliance to the FDA CDRH Hospital Bed System X rm M ue Dimensional and Assessment Guidance to Reduce Entrap ment guidelines 21 FIXED It i ly important to review the resident pati Me om OPTION2 is extremely important to review the resident patient s 4 gt lt A physical and mental conditions and initiate an appropriate HEAD END individual care plan to address any entrapment risks Please see page 7 for further information GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY 2 BASIC AMERICAN a MEDICAL PRODUCTS A ex a vs LIBERTY SERIES OP
46. sonnel for examination and repair Always unplug the Power Cord when performing any mainte nance on the bed DO NOT open assemblies such as the Actuators Hand Control or Control Box If unauthorized personnel perform work on these components the manufacturer s warranty becomes void NEVER permit more than one 1 person on in the bed at any time Body weight should be evenly distributed over the sleeping surface of the bed DO NOT allow the patient to lay sit or lean in such a way that their entire body weight is placed only on the raised head or foot sections of the bed This especially applies when repositioning or transferring a patient in or out of the bed Increased risk may occur when the patient s size and or weight are inappropriate for the bed s dimensions or weight capacity gt gt DO use unauthorized parts accessories or adaptors other than those specified authorized by GF Health Products Inc When operating the High Low Knee or Back Functions of the bed ALWAYS ensure that the confined individual is positioned properly within the confines of the bed DO NOT let any extremities protrude over the side or A The bed is intended for use storage and transport within between the bed rails when performing these functions a temperature range of 40 C to 60 C It has a water resistance rating of IPX4 and IS NOT to be power DO NOT lower the bed when objects are beneath it washed or submers
47. t 100 6744 001 1 00 x 2 00 Rectangular Tube End Cap 100 4700 017 7 16 14 HEAD END Nylon Lock Nut mE 100 6744 001 p 1 00 x 2 00 Rectangular Eu Tube End Cap 100 4700 017 pm 3 00 Twin 7116 14 Locking Caster Nylon Lock Nut g 7 16 14 Stem 4 100 6744 001 L7 100 6330 062 7 16 14 M e 3 00 Twin 5 T 2 Nylon Lock Nut Locking Caster i op 100 6744 001 p 7 16 14 Stem E 100 6330 062 3 00 Twin Locking Caster 7 16 14 Stem 100 6330 062 1 00 x 2 00 Rectangular Tube End Cap 100 4700 017 FOOT END 3 00 Twin Locking Caster 7 16 14 Stem 100 6330 062 CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS PLUGS ON YOUR BED SEE BOTTOM OF PAGE 26 FOR ORDER NUMBERS GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY gt lt BASIC AMERICAN MEDICAL PRODUCTS NOTES LIBERTY SERIES GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY gt lt BASIC AMERICAN MEDICAL PRODUCTS NOTES LIBERTY SERIES GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 GRAHAM FIELD 336 Trowbridge Dr Fond du Lac WI 54937 Customer Service 1 800 365 2338 Fax 920 929 8213 www grahamfield com O 2012 GF Health Pro
48. tches Bent Shaft etc 3 Head amp Foot Motor Works But Hi Lo Does Not a Unplug Power Cord b Remove Cord Guard at Control Box c Unplug Hi Lo Cable at the Control Box and plug into empty port d Replug Power Cord and Test 4 If Hi Lo Works Replace Control Box 5 If Hi Lo Doesn t Work or has a Grinding Noise Replace the 4 Hi Lo Motor Hi Lo Motor 999 0848 051 CONTROL BOX MALFUNCTIONING 1 Check All Motor Cables and Power Cord Connections 2 Hi Lo Motor Works But Head or Foot Does Not See Head amp Foot Motor Troubleshooting above If determined that port is bad replace Control Box 3 Head amp Foot Motor Works But Hi Lo Does Not See Hi Lo Motor Troubleshooting above If determined that port is bad replace Control Box Control Box 999 0848 300 This Port Not Used 72 Foot Motor Cable Port Head Motor Cable Port Hi Lo Motor Cable Port Control Cable Port 999 0848 200 connects to Pendant Cable GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 VY P BASIC AMERICAN uu LIBERTY SERIES 2 E X EDICAL PRODUCTS SECTION E FRAMING SERVICE PARTS All Liberty Models GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 z gt BASIC AMERICAN T MEDICAL PRODUCTS LIBERTY SERIES 1 TO ORDER A REPLACEMENT HD DE
49. to place locking the assist bar vertically PLEASE MAKE SURE THE PIVOT ASSIST BAR IS ALWAYS IN THE FULL VERTICAL LOCKED PIVOTING ASSEMBLY WHEN PIVOTING THE POSITION WHEN RESIDENT PATIENT IS LEFT ASSIST BAR UP OR DOWN UNATTENDED GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 THIS PAGE REVISED 2 6 2013 RAIL KIT NUMBER CORRECTED e RS LIBERTY SERIES a MEDICAL PRODUCTS ex y de OPTIONAL HALF RAIL INSTALLATION HEAD END 76 amp 80 ZA82800 Includes Rails and Hardware Bag OPTION 1 IMPORTANT Head Rail with Lubricate all pivot joints as Quick Release Hardware needed The manufacturer recommends Aerosol White EA Head Deck Lithium Grease LATCH FOR RAISING AND Hardware Bag is Common for Both Kits 999 0785 905 INSTALLATION 1 All Half Rails LH amp RH mount to the bed in a similar fashion Left or right designations are based on the patient s orientation as he she lies in the bed 2 Insert the two Clevis Pins 3a with Circle Cotter and Lanyard Cable into the two Half Rail arms 3 Attach the Half Rail to the Head Deck s welded mount OPTION 2 ing bracket by aligning the arms with Clevis Pins Head Rail with to the holes in the bracket Fixed Hardware 4 For Quick Release Option 1 insert the Clevis Pin ends of the Lanyard assembly from the outside through the holes in the pivot arms and mounting
50. with your bed and is typically placed at the foot end although it will fit on either end It is easily attached to the decks by inserting it into the pre drilled holes for 76 80 positions see insert below To order an additional mattress retainer please use part 999 0822 001 DETAIL B Final Retainer Resting Position start Installation in vertical Closeup of Retainer Insertion Angle A Be sure to use a mattress that is properly sized to fit A Use a properly sized mattress in order to minimize the sleep deck which will remain centered on the the gap between the side of the mattress and assist deck relative to State and Federal Guidelines Use devices This gap must be small enough to prevent of an improperly fitted mattress could result in injury resident patient from getting his her head or neck or death caught in this location GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190D DEC 2012 WY 2 BASIC AMERICAN LIBERTY SERIES 24 QUICK INSTALL VERSATILE HEAD END WALLSAVER The Liberty Bed is shipped with a durable wireform wallsaver for the head end of the bed No hardware is necessary for attaching this multi positional standard accessory Simply bend in the tabbed ends into the slots in the carriage rails and lock down into place There is even a position if you choose to install an optional trapeze support at the head end of the bed See page 30
Download Pdf Manuals
Related Search
Related Contents
Prothèse individualisée avec NobelProcera™ Manual do Utilizador do Nokia E72 Furuno PM30 Computer Monitor User Manual Yves Grenier - Transactiv.exe 健康展等で血糖検査を受検された皆様へ - 一般社団法人 神奈川県臨床 Electrolux EU8216C User's Manual Pentair LX-100EZ User's Manual Operating Instructions Copyright © All rights reserved.