service manual for ILR

image

Contents

1. Access cover Defrosting drain component structure Electronic Thermostat Function ie Power switch Control keys Box structure Rodent control Power Cord arilla Cabin Grille Condenser KUE Electronic Thermostat compressor 10 Specification Model HBC 70 HBC 200 Power V Hz 220 240V 50Hz 220 240V 50Hz Net weight Kg 68 112 Total packaging weight 69 114 Kg Capacity L 70 200 Dimension of vaccine 670x630x847 1240 x630x847 refrigerator WidthxDepthxHeihgt Dimension of package 755x760x900 1325x760x900 WidthxDepthxHeight 11 Product nameplate Ha ier ICELINED REFRIGERATOR CIRCUIT DIAGRAM Anti shock safety Model HBC 200 Jassification Fa Power 1 6 KWeh 24h Effective volume 200L consumption Climate Power input 194W classifica Refrigerant R134a185g Net Weight 108kg XP PLUG SAT THERMOSTATE a KASTARTER Hye YELLOW SWITCH LAMP C CAPACITOR MI COMPRESSOR Number M2 COOLING FAN Haier ICELINED REFRIGERATOR CIRCUIT DIAGRAM Saanen 1 Power 1 2 KWh 24h consumption Power input 155 W tion Rated Voltage 220 240v Refrigerant R134a 140g Rated Frequency 50Hz Net Weight 87 kg XP PLUG SAT THERMOSTATE Hye YELLOW SWITCH LAMP C CAPACITOR MI COMPRESSOR bd Haier Medical and Laboratory Products Co Ltd pe Haier Industrial Park Economic Technology Development M2 COOLING FAN Zone Qingd
2. 62 Haier Inspired living Haier Medical amp Laboratory Products Co Ltd Room 403D Brand Building Haier Industrial Park No 1 Haier Road Qingdao China Website www haiermedical com 63
3. Place the vaccine refrigerator in a lab with a room temperature of 32 C and wait until balance a temperature difference of 1K is achieved between the temperature inside the vaccine refrigerator and the room temperature Close the door of the vaccine refrigerator and keep the refrigerator in continuous operation Record the time needed for the storage temperature tm to reach tm max With reference to table 13 3 the cooling temperature tm to reach 14 C and the freezing temperature to reach its correspondent star level 60 respectively According to related standards the time for vaccine refrigerators with a capacity of below 250L shall not exceed 2h while for those with a capacity of 250L 500L the time shall not exceed 3h 8 Freezing capacity The freezing capacity of the vaccine refrigerator shall not less than 90 of the rated capacity 9 Power consumption Place the vaccine refrigerator in the lab Room temperature for labs in different climates are respectively as follows M mid temperate zone 27 C T temperate zone 32 C H tropical zone 43 C Duration of the test is 24h Power consumption of the cold storage and refrigerator shall be less than 115 of the rated value power consumption of the freezer shall be less than 110 of the rated value Other tests to be carried out measurement of storage temperature 10 Cold storage and freezer test of storage temperature and ice making capacity etc Ta
4. 0 A2 Alarm delay 0 30 Min 15 dFO Defrost cycle 2 24 Hour 8 dF1 Defrost time 0 240 Min 0 ct0 Start up time with failure of sensor 0 60 Min 15 ct1 Downtime with failure of sensor 1 60 Min 8 AH Temperature of high temperature alarm St AO St A0 8 AL Temperature of low temperature alarm St AO St A0 2 SP1 A eF a forced operation 10 10 C 10 SP2 Parameter for restarting forced operation 10 10 C 25 after power failure 23 rO for controlling temperature difference A parameter to control the start up and stop of the compressor Refer to description of the parameter St r1 for setting maximum temperature To keep the setting range of control temperature St not above the parameter value of r1 Caution minimum of r1 cannot be set to be below parameter r2 and it will vary with the variation of r2 r2 for setting the minimum temperature To keep the setting range of control temperature St not below the parameter value of r2 Caution maximum of r2 cannot be set to be above the parameter r1 and it will vary with the variation of r1 CAL calibrated temperature of the sensor Calibrate the error of the sensor The temperature displayed is equal to the actual value of the measured temperature plus the value of CAL t1 delay in start up When the compressor is started for the first time it shall be started with a delay time of t1 t2 shortest downtime of the compressor After operation of the compre
5. 7 Increase or decrease the value with the buttons or Y l 20 8 Save temporarily the modified parameter value by pressing the button S and return to the parameter displayed 9 If values of other parameters are to modified repeat steps 5 8 10 Hold the button for more than 3s to save the modified parameter and quit the parameter setting program Caution if there is no button operation within 3min the setting shall be automatically quitted and the controller is locked Meanwhile execution shall be implemented with the parameter memory last set 21 III Menu description Parameters of the controller are divided into general menu and administration menu General menu Parameter Description Min Max Unit Default St Control temperature r1 r2 C 5 0 PS Password to access administration menu 50 100 0 Ab Select mode A b b St for controlling temperature With St gt 2 C when the temperature measured PV gt St r0 the machine is started and when PV lt St it is stopped With St lt 2 C when the temperature measured PV gt St r0 the machine is started and when PV lt St rO it is stopped Setting range of St is determined by the parameters r1 and r2 with St lt r1 and St gt r2 PS the password to access the administration menu When accessing the administration parameters only with the correct password entered can the parameters of the administr
6. a new compressor and filter dryer and fax them by welding The system shall be open for duration not over 10min Connect the R134a recovery device to the pipeline on lower side of the compressor before evacuating The evacuating duration shall not be less than 45min Pour R134a refrigerant into the system according to the former refrigerant charge Inspect whether there is leakage at the welded point with soapy water with foam 44 9 Wipe off welding powder and welding flux with clean gauze 10 Connect the electrical control device of the compressor 11 Install and adjust the thermostat 12 Turn on the power for commissioning and check whether the unit is in normal operation 12 3 5 Procedures for installation and brazing of compressor 1 Install a new compressor 2 No direct inhalation of air into the compressor is allowed for inspection or other tests 3 When replacing with a new compressor during maintenance the sealing plug shall be opened at last and brazing and evacuating shall be implemented immediately 4 When taking off the sealing plug of the compressor and if there is no sound of air flow no pressure the compressor shall not be put into application 5 If no sealing plug is installed in the compressor the compressor shall not be put into application 6 Hands of the maintenance operator shall be free of oil 7 Only before application shall the filter dryer of the vaccine refrigerator be ope
7. a small amount of refrigeration oil has good permeability If case of occurence of leakage oil stain will exist on the leakage spot Troubleshooting by sound When the vaccine refrigerator is in normal operation high pressure liquid refrigerant flows into the low pressure evaporator through the capillary and sound of air flow and water flow can be heard When opening the door of the vaccine refrigerator and getting close to the evaporator the sound aboved mentioned can be heard No sound or some occassional sound means leakage or insufficient filling of refrigerant in the system After power is connected listen to the sound of the operation of the compressor to make sure whether the sound is normal A low pitched buzzing means abnormal starting of the motor A rattling noise means damage of the suspended spring in the compressor or tilt operation of the compressor In case of occurence of the above faults replace the compressor Troubleshooting by touch Touch the cover of the condenser and after continuous operation the temperature on top of the cover is the highest and the temperature of the bottom is almost the same as the indoor temperature The temperature is related to the ambient temperature In winter the temperature is lower so is the temperature of the condenser and only a small part feels warm In summer the temperature is higher so is the temperature of the condenser cover and the whole condenser feels warm The high
8. allow for good ventilation in it so that better refrigeration effects can be achieved Electrical requirements Make sure there exists suitable power socket 220V with correct earthing near the vaccine refrigerator Avoid cutting the third earthing pin Otherwise it shall be very dangerous because 17 effective earthing will no longer be supplied to the vaccine refrigerator which may cause electric shock Installation restrictions e Don t install the vaccine refrigerator in any place with thermal insulation or improper heating conditions such as garage etc e The vaccine refrigerator shall be installed in suitable place with solid and smooth surface and it shall be kept from direct sunlight or heat sources such as electric radiator baseboard heater and cooker etc Installation and accessories After the vaccine refrigerator is installed in suitable place plug into the 220V socket The vaccine refrigerator is equipped with a thermostat for controlling temperature of the vaccine chamber The temperature can be set between 2 C 8 Cwith the button A or V When the vaccine refrigerator is powered the yellow light will on After the vaccine refrigerator has been working for 48h and its temperature has been set as required vaccine can be placed in the vaccine refrigerator In case of removal of the plug of vaccine refrigerator or power failure it can only be plugged in after 5min Caution if the vacci
9. compressor is in automatic operation controller by the thermostat The temperature becomes lower and lower over time When it reaches the set value the compessor stops working With the working cycle above the internal temperature can be controlled to be around the set temperature 32 System troubleshooting System troubleshooting Maintenance of vaccine refrigerator Inspect the whole circuit of the vaccine refrigerator to make sure whether there is any falling off of wire connector damage to the wire or shortcircuit Inspect the insulating layer of the vaccine refrigerator Under normal circumstances the resistance is over 2 megaohm in case of a resistance less than 2 megaohm further inspection of the resistance of the vaccine refrigerator is required Set the voltmeter to R1 ohm shift and plug the two pen voltage testers of the voltmeter into terminal L and N respectively to check the indicated value The normal value shall be several ohm to tens of ohms A zero ohm means shortcircuit of the vaccine refrigerator or shortcircuit brought by improper operation of the compressor fan motor Repair it after finding out the cause A too large or infinite resistance means a broken circuit and please inspect it to have the problem settled After implementation of the above inspection the vaccine refrigerator will be free of electrical failure and you can have it connected to power for further inspection 1 2 Cause of electric leakage
10. delay time of t1 t2 shortest downtime of the compressor After operation of the compressor is stopped it can be started again after a delay of time t2 If t2 0 the delay is cancelled not related to delay in start up t3 forced start up time after energized after a power failure of 6h In case of a power failure over 6h and when energized again the compressor is required to be in continuous operation for a time of t3 t3 0 indicates no forced start up when energized Caution if the control temperature is above 0 C when the measured temperature PV lt 2 C the machined is forced to be stopped 28 AO setting range of high and low temperature alarm Setting range of parameters AH and AL shall be between St AO and St AO If St 5 and A0 6 then parameters AH and AL can only be adjusted between 1 and 11 A1 power on alarm and delay in defrosting alarm Valid only in mode B When energized for the first time and defrosting the temperature of the cabinet will probably higher than that of the alarm setting value When A1 1 no alarm occurs during energizing for the first time and defrosting and only after the temperature is below than the control temperature St will alarm be started When A1 0 once the temperature is above AH or below AL alarm occurs immediately A2 alarm delay Only valid in mode B When the temperature is above AH or below AL no immediate alarm occurs but it occurs when the tempera
11. energized after 0 8 0 Hour 0 AO ee of high and low 1 20 C 6 A1 P oa alarm and delay in defrosting 0 1 _ 1 A2 Alarm delay 0 30 Min 1 dFO Defrost cycle 2 24 Hour 8 dF1 Defrost time 0 240 Min 0 ct0 Start up time with failure of sensor 0 60 Min 15 ct1 Downtime with failure of sensor 1 60 Min 8 AH Temperature of high temperature alarm St AO St AO 1e 8 AL Temperature of low temperature alarm St AO St AO 1e 2 SP1 eae forced operation 40 10 C 10 SP2 Parameter for restarting forced operation 40 10 C 25 after power failure 27 rO for controlling temperature difference A parameter to control the start up and stop of the compressor Refer to description of the parameter St r1 for setting maximum temperature To keep the setting range of control temperature St not above the parameter value of r1 Caution minimum of r1 cannot be set to be below parameter r2 and it will vary with the variation of r2 r2 for setting the minimum temperature To keep the setting range of control temperature St not below the parameter value of r2 Caution maximum of r2 cannot be set to be above the parameter r1 and it will vary with the variation of r1 CAL calibrated temperature of the sensor Calibrate the error of the sensor The temperature displayed is equal to the actual value of the measured temperature plus the value of CAL t1 delay in start up When the compressor is started for the first time it shall be started with a
12. pressure exhaust pipe has higher temperature so in summer when it is toughed with hand it feels hot and so does it in winter The low pressure inlet pipe has lower temperature In summer dewdrops appear on the pipe wall and when it is touched it feels cool it feels even colder when touched in winter 34 1 6 1 2 1 8 Troubleshooting by position If the vaccine refrigerator is not level or the ground is soft adjust the foot or make the ground solid In case of noise caused by loose bolt between the foot and the cabinet of the vaccine refrigerator fasten the bolt In case of noise caused by aged rubber washer or loose fixing clamp of the compressor replace the rubber washer or the fixing clamp If the pipe collide with the refrigerator cabinet or they collide with each other related adjustment shall be implemented to get rid of the noise In case of noise caused by stuck fan blades get rid of the barrier around the fan or adjust the fan blade In case of noise caused by loose fan support fasten it Analysis of poor refrigerating effects and solution Inspect the set value of the thermostat and if it is not set within the range of 2 C 5 C set it to be 3C Inspect whether back and top of the vaccine refrigerator are too close to the wall whether it has direct sunlight or is close to heat sources whether the indoor temperature is too high whether the condenser is blocked by cloth paper or dirt Any of
13. the room During the test the vaccine refrigerator shall be kept empty Set the thermostat to be low and close the door Run the vaccine refrigerator for more than 30min before implementation of the test Requirements on the sound power level of the noise from the vaccine refrigerator are as follows those with a capacity of 250L or below lt 52dB A those with a capacity over 250L lt 55dB A 8 2 7 Methods for testing refrigerating property 1 4 or 6 vaccine refrigerators can be tested simultaneously in a lab Polyurethane or polystyrene materials which have good thermal insulation property are applied in the enclosure structure of the lab The lab is equipped with air retreating unit Air with uniform pressure enter the lab from the perforated ceiling to ensure that the speed of the indoor air flow is not over 0 25m s so that the room temperature is kept uniform The temperature in the lab can be adjusted within the range of 10 43 C 2 For requirements on temperature for application of vaccine refrigerators in different climate areas refer to table 6 1 Specified ambient temperature and a fluctuation range of 0 5K shall be kept at each test spot Within the range of 2m above the test platform vertical gradient of the temperature shall not exceed 2K m 59 Table 6 1 Vaccine refrigerators for different climate areas Climate type Label Suitable Measured temperature C temperature C Mid temperate zone
14. will cause vibration of the compressor and thus sound is produced 2 When the vaccine refrigerator is stopped the compressor will produce vibrating sound due to inertial rotation 3 When the vaccine refrigerator is stopped the compressor speeds down and balance is achieved between the high pressure and low pressure at the exhaust pipe and return pipe so sound is produced by the compressor 4 When the compressor is started great sound will be produced by the hammer type starter relay in case of low voltage Even with a voltage stabilizer installed timely measurement of the voltage is still required If requirement of stable voltage cannot 54 be achieved there must be problem in quality of the voltage stabilizer In case of low output voltage don t use the vaccine refrigerator For causes above mentioned sound is produced when the vaccine refrigerator is started and stopped 6 15 Why sometimes the noise heard from a longer distance is greater than that heard from a shorter distance Answer transmission of noise from the vaccine refrigerator is related to factors including the position of the vaccine refrigerator and the indoor space size etc the sound wave produced by the operation of the vaccine refrigerator will reflect when bumping onto walls or furniture Therefore the noise heard from a place with longer distance may sometime greater than the one heard from a nearer place For this reason it is incorrect to come t
15. Implementation of noise test is shown in figure 8 Place the sound transmission devices at measurement points 1 2 3 and 4 respectively and measure the sound level with a sound level meter A weighted The larger reading shall be taken for calculation of the average value The arithmetic mean at the above 4 measurement points shall be the average sound pressure level of the noise Sound power level of vaccine refrigerator for the noise measurement can be calculated with the test results Lw Lpa 2 10LgS Herein Lw is for the sound power level in dB with a reference value of 1PW Lpa is for the average sound pressure level of the surface measured in dB with a reference value of 20uPa S is for the envelope area of the surface measured in m Suppose 11 12 and 13 are respectively the length width and height of the vaccine refrigerator and they are in meter a 1 2 d b 2 2 d c 13 2 d If d 1m is taken 56 Then S 4 ab bc ac Positions of the measurement points are shown in table 4 57 8 Technology for measuring performance parameter 8 1 Measurement content Measurement technologies for vaccine refrigerators include the measurement of refrigerating performance and safety requirements on electrical parts Before leaving the factory the following tests shall be implemented 1 appearance requirements 2 cooling speed 3 electric leakage 4 insulating resistance in cold state 5 electric field stre
16. M 10 27 27 Temperate zone T 15 32 32 Tropical zone H 25 43 43 3 Ambient humidity the relative humidity is generally 15 19 4 Test pack a rectangular test pack with dimensions shown in the following table is used as load for the vaccine refrigerator Table 6 2 Test pack for testing property of the vaccine refrigerator Dimension mm Weight g Remarks Small S 100x100x50 200 Big B 100x100x100 400 5 Thermometer platinum resistance thermometers are generaly used for measurement of temperature They include the following two types a When measuring the temperature place the platinum resistance directly into the test pack b Insert the platinum resistance into the middle of 4 tinned copper posts which have a diameter and a height of both 15 2mm Bridges for measuring temperature with an precision of 0 3K are used as measurement equipment 6 Instruments for measuring humidity relative humidity shall be measured with a wet bulb thermometer and a dry bulb thermometer Place the dry bulb and wet bulb thermometers 1 5 2 layers of absorbent gauze is wrapped on the temperature sensing parts of the thermometer and bottom of the gauze is placed into distilled water in the wind with a speed of 2 5 3m s Record temperature on the dry bulb thermometer and calculate the relative humidity with the reference table of relative humidity for environmental tests in GB6999 86 7 Cooling speed
17. Service Manual FILE No HMRSM 002 02 Ice Lined Refrigerator HBC 70 HBC 110 HBC 200 HBD 116 HBD 286 se Haier Medical amp Laboratory Products Co Ltd Effective models This service manual is effective for following models Model name Product code Pa aad ENDS type Frequency Hz HBC 70 ST 220V 50Hz British round plug HBC 110 ST 220V 50Hz European plug HBC 200 ST 220V 50Hz British round plug Content OCTA EE 4 Safe Caution ntm Arteertee bete ee 5 EPr ductappeoranee Dennen nde nende water tee 6 Parts and fun tions I snee en edet a an 8 Le mpanentstruct re 1e minute 9 ESDOORN zanden nt anunt tante nel 11 C hrod ctameniat dns RS RU nanas 12 Wiring dB an da ea te inrn 13 System flowchart ES EE A Ron 15 ENetdimens n denten eea 16 installation and accessories xi arsecesstsdsstaventasensetdavenicedescdesetesstonsanvdtonesbidennatans 17 Adjustment of thermostat en moenrensersr sehen tandsteen dea dode 19 l Introduction of display panel and buttons onsen nenene sneren enn 19 Il Parameters and operation of controller naan ennn neerenensensenn 20 llMenu descriptions a a a Tnt teint Gr 22 Function procedure laure lim tnietaa tante 31 System troubleshooting u mertenteseren area aheashstaaboneds hankaasdenke worstelt 33 1 System troubleshooting siss iirin E aeaa EE T E E 33 2 Common failures in vaccine refrigerator annen ennen en ennnennr eeen sensenn 42 3 Maintena
18. and filter 3 Replace the compressor 1 Dry the vaccine after it is cleaned before placing it into the refrigerator 2 Only vaccine with suitable amount is stored Electric leakage on body of the vaccine refrigerator and a tingling when touched with hand 1 The body of the vaccine refrigerator has no earthing 2 Binding post of the compressor contacts the shell of the cabinet thus causing short circuit 3 Damping of component of the electrical system thus causing reduced insulating 1 Earthing as specified 2 Replace the compressor 3 Inspect carefully step by step If the insulating capacity is seriously damaged replace the component remove the damp component and place it into a drying cabinet for 39 capacity and electric leakage drying Electric leakage of the fan 1 Too high ambient temperature or reduced insulating capacity of the insulator 2 Short circuit between fan coil or burned fan coil 1 Replace the fan 2 Replace the fan Too loud noise in the fan 1 Fan blade is stuck or fan support is loose 2 Too large space between fan shafts 1 Clear barrier around the fan or fasten the support 2 Replace the fan Fan is not running 1 Inspect whether the connector is falling off 2 Inspect whether the fan has been burned 1 Replace the fan 2 Replace the fan Electric leakage of thermostat 1 Too longer operation ti
19. and technical requirements 3 1 Requirements on refrigeration equipment The compressor condenser evaporator and dry filter shall be the special type for R134a vaccine refrigerator and they cannot be mixed up Requirements on repairing equipment Evacuator and filler shall be special for R134a No special requirement on other equipment such as pressure gauge however if they are applied in R134a vaccine refrigerator they are not allowed to be applied for other purposes Maintenance procedures Operation method for repairing equipment refer to the operating instruction Special requirements Interval between turning on the compressor and beginning evacuating shall not exceed 10min The evacuating duration shall not be less than 45min If leakage occurs on high pressure end of the vaccine refrigerator it can be repaired Before welding blow in N2 nitrogen with high purity If leakage occurs on the low pressure end inspect the pressure of the system if the system has no pressure it shall be discarded and a new vaccine refrigerator shall be bought if there is pressure it can be repaired Procedure for replacement of R134a compressor Disconnect the power and open the protection cover with a Phillips screwdriver or a universal screwdriver Remove electrical components of the compressor Unscrew the 4 assembling bolts of the compressor with a straddle wrench or a universal tool Take off the compressor Install
20. ao 266510 P R China Haief IcELNED REFRIGERATOR CIRCUIT DIAGRAM 5 Anti shock safety Model HBC 70 classification WF i Power 10 KWeh 24h Effective volume 7OL annuler kumala T Power input 150W classification Rated Voltage 220 240v Refrigerant R134a110g Rated Frequency 50Hz Net Weight 58kg XP PLUG SAT THERMOSTATE so CE KA STARTER HyE YELLOW SWITCH LAMP date C CAPACITOR MI COMPRESSOR 12 Wiring diagram HBC 70 XP Power Plug SAT Electronic Thermostat KA Starter M1 compressor Hye switch light C Start capacitor HBC 110 200 XP Power Plug SAT Electronic Thermostat KA Starter M1 compressor M2 fan motor Hye switch light C Start capacitor System flowchart 1 Compressor 2 Condenser 3 Dry filter 4 Capillary 5 Evaporator Cooling system of the vaccine refrigerator by Haier is composed of a type of direct cooling system The cooling effects are achieved by operation of the compressor while operation of the compressor is controlled by the thermostat Detailed information is as follows operation of the compressor forces refrigerant R134a to gradually flow through the condenser dry filter and evaporator and finally return to the compressor This is a complete cooling cycle 15 Net dimension D Model HBC 70 HBC 110 HBC 200 Dimension of vaccine refrigerator width xdepth 670x630x820 1240x630x847 xheight mm Dimension wi
21. ar in the following cases 1 For a new vaccine refrigerator the internal stress of its materials exists The material stress will disappear over time and less sound will be produced 2 Uneven temperature in the vaccine refrigerator after defrosting 3 When the door is opened for taking out the vaccine the evaporator is heated suddenly 4 In several minutes after the vaccine refrigerator is started or stopped significant changes occur in the temperature of the condenser and the evaporator Almost all things expand with heat and contract with cold so the sound is normal 6 6 Why the noise becomes greater after the vaccine refrigerator has been in service for some time Answer after being in service for half a year or a year the space by abrasive wear between internal parts will be increased and the noise from the compressor will also be correspondently increased After that during the whole service life of the vaccine refrigerator the space between parts maintains stable 6 7 Why the noise becomes greater after the vaccine refrigerator has out of service for some time Answer after the vaccine refrigerator has been out of service for a long time the crankshaft of the compressor may be stuck by the dirt brought by the lubricant oil or by the lubricant oil frozen because of low temperature Therefore the compressor cannot be started or after it is started it gives out greater noise and the abrasive wear of the 52 cranks
22. ation menu be modified If a wrong password is entered the parameter setting will be quitted Caution the correct password is 15 Ab for selecting mode When mode A is selected it is special for ice lining projects mode b is a general mode and mode A is default When mode b is selected the part of displayed temperature is specially treated to agree with the temperature display of the cabinet When the measured temperature is between 2 5 7 5 C the measured temperature is displayed with a refreshing frequency of 15s If the measured temperature is above 10 0 Cor below 1 C the actual temperature is displayed If the measured temperature is above 7 5 C and below 10 0 C 7 5 is displayed and if it has been continuously above 7 5 C for over 4min the actual temperature is displayed If the measured temperature is below 2 5 C and above 1 0 C 2 5 is displayed and if it has been continuously below 2 5 C for over 4min the actual temperature is displayed 22 Administration menu in mode A Parameter Description Min Max Unit Default rO Control temperature difference 0 5 5 0 1 r1 Max temperature can be set r2 50 C 8 r2 Min temperature can be set 50 r1 C 2 CAL Calibrated temperature of sensor 3 0 3 0 G 0 t1 Delay in start up 0 12 Min 0 t2 Shortest downtime of compressor 0 60 Min 0 AO Eon of high and low 1 20 C 5 M oe alarm and delay in defrosting 0 1 _
23. ble 6 3 Storage temperature of vaccine refrigerator Climate type Climate type Ambient Storage temperature C t4 to ts t4 ts te t7 te to tro Mid M 10 2 Csti to t3 ta ts te tz tg to troS8 C temperate zone 27 Temperate T 15 2 CSti to ta t4 to te t7 tg to troSB C zone 32 Tropical zone H 25 2Cst to t3 t4 ts te t7 tg to t10S8 C 43 61 Details of wearing parts 20G 2 tii Model SN SPARE PART Chinese name code 1 Compressor FFT NL6 1FT 0070701331 Thermostat FRE SES 0074091494 Power Cord FE Wht Ae 0070402374 HBC 70 4 Combination cu f HH Ee PAN EX 0070402373 ine 5 Cabin guard HEP E 0071210004 6 Filter F ES 0074180002A 1 Compressor FFT NL6 1FT 0070701331 2 Thermostat FREE PEER 0074091494 3 XL Fan motor EBM220V 60HZ 25W 0075030020 dane 4 Combination line Ya PS AE 0070402373 5 Cabin guard PLACE E 0071210004 6 Power Cord HIE ZE 0070402374 7 FED Hr ct HE as 0074180002A Filter zo 20G 4 fii 1 Compressor FHR NL7FT 0070701332 2 Thermostat Me Fy EE SES 0074091494 3 XL Fan motor EBM220V 60HZ 25W 0075030020 eon 4 Combination line YATES BE LEZ 0070402373 gt Cabin guard WL Aga E 0071210004 6 Power Cord Ey 0070402374 7 Hia FED Hr HE as 0074180002A
24. cable fuse 2 Inspect and fix the connection or replace it 3 Keep operation until the voltage is raised to the rated value or add a voltage stabilizer 4 Reset the thermostat 5 Replace the thermostat 6 Replace the starter relay or thermal protection device 7 Measure with the Rx1 shift of a multimeter and in case of broken wire replace the compressor 8 Replace the compressor 9 Replace the compressor Too frequent starting of the compressor and too long operation time but the temperature inside the vaccine refrigerator drops at a too slow speed 1 Too large space between the thermostat probe and the evaporator 2 Failure in the thermostat 3 Too much vaccine stored in the vaccine refrigerator too frequent open and close of the door and the door is open for too long 4 Too high ambient temperature too high humidity and blocked air flow 5 Too thick frost layer on the evaporator 6 Insufficient refrigerant or 1 Adjust the position of the probe to make it close to the evaporator cover 2 Replace the thermostat 3 Reduce the amount of vaccine stored and the time to open the door and shorten the time when the door is open 4 Install the vaccine refrigerator at a suitable position with good ventilation 5 Regularly defrost 6 Inspect or add refrigerant 37 leakage 7 Reduced efficiency of the compressor 8 The dry filter is blocked 7 Re
25. chine to the connector of the system Connect the power Start the motor Open all related valves When the pointer of the vacuum gauge is gradually close to the graduation 1 observe the pointer Evacuating duration is required not to be less than 45min 49 Close all valves after the completion of evacuating Stop the motor Filling First determine the filling volume according to the state of the vaccine refrigerator being repaired each small graduation on the liquid line means 25g while each large graduation means 100g Open all related valves Close each valve when the level is raised to the filling volume Commissioning After filling connect the vaccine refrigerator to power for commissioning Observe whether the vaccine refrigerator is in normal operation Disconnect the power 50 6 Noise from the vaccine refrigerator 6 1 Noise sources of the vaccine refrigerator Noise from the vaccine refrigerator comes from thermal expansion and contraction of the compressor evaporator or condenser from contact between pipe or their resonance from the flow of refrigerant or from shaking of the vaccine refrigerator caused by uneven floor etc 6 2 Compressor Noise from the vaccine refrigerator mainly comes from the compressor The compressor gives out various sounds including continuous discontinuous sound periodic non periodic sound The sounds have various frequencies ranging from tens of Hz to thousands of Hz Caus
26. controller In case of no any button operation the controller will be automatically locked within 3min while the buzzer will sound for about 0 5s Hold simultaneously buttons Stand Fn for 1s and the locked controller will be unlocked or the unlocked controller will be lock while the buzzer will sound for about 0 5s When the controller is locked no operation can be performed on it and only after it is unlocked can operation on it be performed 30 Function procedure Before vaccine is put into the vaccine refrigerator the ice lining on four sides shall be frozen To ensure freezing of ice lining on four sides carry out the following operation 1 Set the external thermostat to be 3 C 2 Inspect the temperature indicated by the thermometer it must be between 2 C 5C Temperature inside the vaccine chamber must be continuously monitored with a thermometer and the temperature must be kept within the range of 2 C 8 C The temperature shall be always be set between 2 C 5 C due to tolerance of the thermostat Be sure to use the ice box inside the vaccine box In case of removal of the plug of the vaccine refrigerator or power failure wait 5min before plugging in again Caution if the vaccine refrigerator has been placed flat or tilted for some time it shall be kept upright for 24h before plugging in again 31 Brief description of thermostat principle When the internal temperature is above the set value the
27. disconnected wait at least 5min before the refrigerator is started again otherwise the compressor or the refrigerating system will be damaged Explosive materials inflammables volatile matters glass container filled with liquid alkali or corrosive acid are not allowed to be stored in the vaccine refrigerator Children are not allowed to play inside the refrigerator cabinet to avoid choke or damage Maintenance Defrosting after being put into application for some time a layer of frost appears in the inner tank If the layer of frost is too thick the refrigerating effects will be affected and the power consumption will be increased 1 2 4 5 Regularly when the frost layer is 5mm thick get rid of the frost with a plastic shovel When defrosting disconnect the power take out the vaccine stored in the vaccine refrigerator and placed it in a place with low temperature To accelerate defrosting a container containing warm water can be placed inside the cabinet After the layer of frost melts pull out the drain plug to empty the water and then dry the vaccine refrigerator by wiping Cleaning for sake of safety disconnect the power before cleaning to avoid electric shock or damage to the vaccine refrigerator When cleaning the vaccine refrigerator clean it with a piece of soft cloth or sponge dipped with clean water or soapy water or non corrosive neutral detergent After cleaning dry the vaccine refrigerator by wi
28. e shell When the compressor is started or stopped the force applied on the compressor greatly changes and it sometimes makes the body contact the shell thus giving out sound The case continues for a short time 6 The electromagnetic sound produced by hermetic motor is comparatively smaller 51 6 3 How the sound of water flow is produced Answer when the liquid refrigerant is flowing especially flowing through the evaporator it will produce a sound similar to that of water flow The sound may vary because of the different technologies applied in the structure of the evaporator If when the vaccine refrigerator is running the sound is not produced with the evaporator not refrigerating and the condenser not hot the refrigerant is restricted or there is leakage of the refrigerant 6 4 Sound of air flow When the compressor is running gas is continuously sucked in and discharged When the liquid refrigerant flows through the capillary gas injection occurs and the sound of air flow is produced This kind of sound is normal but it may vary in different vaccine refrigerators for the compressor and the evaporator used are different 6 5 Why clucks or clicking sounds are suddenly produced by the vaccine refrigerator Answer the sound is produced by the thermal expansion and contraction of the condenser and the evaporator or it is produced by the ice on the surface of the evaporator shock heated or cooled suddenly The sound may appe
29. erican GE products H 25 leak detector and HLD4000 leak detector manufactured by a German company 8 2 2 Insulating resistance The test is to inspect the insulating property of the vaccine refrigerator Apply 500V DC voltage on live parts and accessible metal parts and measure the resistance The resistance value shall not be less than 2 million ohm 8 2 3 Electric field strength The test is to inspect the electric field strength of live parts and all accessible metal parts Flashover and breakdown devices are used as test equipment and a voltage of 1650V 50Hz is applied for 2s During the test no flashover or breakdown shall appear 58 84 2 Earthing resistance Inspect the voltage drop of between earthing point and accessible metal parts of the vaccine refrigerator with AC power supply which has a no load voltage of not over 12V and a current of 10A Calculate the resistance with the current and voltage drop The earthing resistance shall not exceed 0 1 ohm 8 2 5 Leakage current During test apply a voltage 1 06 times of the rated one on the vaccine refrigerator and maintain the test under operating conditions The test is specified in GB4706 1 84 and the leakage current meter is manufactured according to regulations The leakage current measured shall not exceed 1 5mA 8 2 6 Noise test The test is implemented in a semi anechoic room Place the vaccine refrigerator onto an elastic base 5 10mm thick in the geometric center of
30. es for the noise in the compressor include 1 There are two pieces of flexible steel sheet air inlet valve and exhaust valve and when the vaccine refrigerator is being started each steel sheet will open and close for 47 times every second thus giving out the vibrating sound Meanwhile the open and closing action of the steel sheet makes the sheet contact other parts which also gives out sound The continuous buzz heard belong to this type of sound 2 When the compressor is being manufactured space inevitably exists between parts During the operation of the vaccine refrigerator the existence of the spaces make some parts contact with each other thus giving out sound Especially when it is started and stopped the force on parts of the compressor greatly changes so that greater sound is produced 3 Noise of liquid flow The operation process of the compressor is a process of continuous air intake and exhaust If sudden changes in the intake or exhaust pressure of refrigerant gas occur the vibrating sound will be produced 4 Sound of resonance between parts of the compressor Sometimes resonance may be caused between parts of the compressor including the compressor body the exhaust pipe the silencing device inside the compressor and supporting spring etc and sound is given out 5 Collision between the compressor body and the shell The compressor body is suspended by three springs or supported by the spring inside th
31. etermine whether the failure is caused by internal or external factors Including Blown fuse poor contact between the plug and the socket or too low voltage External factors 2 Internal factors Electrical failures in the vaccine refrigerator usually caused by falling off of wire terminal or poor contact Thermostat with burned contact or damaged switching elements etc Poor starting capacitor Poor starter relay or overload protector Poor starter is usually caused by burned contact or falling off of weld poor thermal protection device is usually caused by burned contact or poor heating wire Winding of the motor is burned The shaft of the compressor is damaged and block the cylinder 2 2 Improper working of the vaccine refrigerator With power on the vaccine refrigerator is started but after several seconds or 1 2 minutes it automatically stops before restarting after several minutes It stops works within a short time after starting Repetition goes on and it is impossible for normal operation The following several causes may cause the problem 1 Large voltage fluctuation A voltage of 220V is required for normal operation If the voltage is too high the current of the starter will correspondently increase When the voltage exceeds the allowable value the starting contact will keep connected for a long time and it cannot be released under normal circumstances the starting coi
32. f and release it and the current control temperature is displayed on the display screen Increase or decrease the value with the buttons amp or Vv until the desired value is obtained Confirm the new value by pressing the button Sf again and save the new valuebefore returning to normal operating mode Setting parameters of other general menus 1 Hold the button SF for more than 3s and the parameter code PS password is displayed on the digital tube 2 Roll parameters with the buttons Al and Vv PS AB 3 Confirm the parameter by pressing the button SE and value of the parameter is displayer 4 Increase or decrease the value of the parameter with the buttons A or V 5 Save temporarily the modified parameter value by pressing the button SE and return to the parameter displayed 6 If values of other parameters are to modified repeat steps 3 5 7 Hold the button SE for more than 3s to save the modified parameter value and return to normal circumstances Setting parameters of administration menu 1 Hold the button SF for more than 3s and the parameter code PS password is displayed on the display screen 2 Set the password by pressing the button S 3 Roll the number to 15 with the buttons and Vv 4 Confirm the password by pressing the button SF 5 Roll the parameter name with the buttons A andl F 6 Display the correspondent parameter value by pressing the button SE
33. f any abnormal case found inspect the circuit and each component In case of damage to the trademark or loss replace with a new trademark of Haier Clean the inside and outside of the vaccine refrigerator If the vaccine refrigerator is out of operation for a long time dry off the water on it put it into a plastic bag and place it in a dry storehouse with good ventilation Remember to place it on flat floor 48 Evacuating and filling procedures and tools needed Tools and equipment for repairing vaccine refrigerator Recycling machine Nitrogen gas N2 container oxygen O2 container acetylene container and torch Thermometer Voltmeter Pen voltage tester Manifold pressure gauge Flathead screwdriver Phillips screwdriver Needle nose pliers sharp nose pliers diagonal pliers and sealing pliers 10 Hansen valve 11 Pipe cutter 5 2 Notice for use R134a filling equipment is only special for R134a refrigerant 1 2 It is not allowed to fill R134a refrigerant with R12 filling equipment Please keep the pliers pipe cutting device and Hansen valve clean and make them free of oil stain 5 3 Procedures for filling equipment 1 N Preparation before application Visually inspect whether there is anything abnormal in the recycling machine etc Keep all valves connected to the equipment closed until operation is really started Evacuate Connect hose of the recycling ma
34. for purposes mentioned in this Instruction and Maintenance Guide The vaccine refrigerator shall be installed correctly according to instruction in the chapter of installation For description before application refer to earthing instruction in the chapter of installation Don t remove plug of the vaccine refrigerator by pulling the power line but clutch the plug and remove it from the electric socket Immediately repair or replace all worn or damaged power lines Don t use any power line with crack or scratch in the line plug or connector end Remember to remove the plug of the vaccine refrigerator before cleaning or any repair Caution in case of demand on repair of the vaccine refrigerator it shall be repaired by professional technical personnel If the old vaccine refrigerator is no longer in application it is recommended from us that the vaccine basket be placed in a suitable place to reduce possibility of hazard to child The vaccine refrigerator is forbidden to be placed or embedded into a closed cupboard It can only be independently installed The vaccine refrigerator is not allowed to put into application in places with explosive fume Don t use the vaccine refrigerator for cooling foods Product appearance HBC 70 HBC 110 HBC 200 Inner side EL Parts and functions Handle Lock oc Inner lining Cover Tray Vaccine basket Control panel
35. haft is correspondently increased It is recommended not to stop operation of the vaccine refrigerator Sometimes a new vaccine refrigerator also has such a characteristic After being in service for some time the vaccine refrigerator will become normal 6 8 How to reduce noise from the vaccine refrigerator Currently all types of vaccine refrigerators have problem in noise Scientists are always trying to find a solution to it but little success has been achieved The final solution depends on new technical improvement and development 6 9 Why the noise from the vaccine refrigerator becomes greater at night Answer the impact of the noise from the vaccine refrigerator on people is related to the ambient noise level In a crowded city or shop noise from the vaccine refrigerator will be found to be little while in a village or a quiet sitting room the noise will be greater That is why people find a vaccine refrigerator at home has greater noise than the one in the shop 6 10 What regulations on noise from vaccine refrigerator have been made by the state Answer on side effects on people s health by noise and the technical level which the current vaccine refrigerator industry can achieve regulations on noise from vaccine refrigerators are provided by the national standard GB8059 1 87 vaccine refrigerator with a capacity of less than 250L shall have a noise level of not more than 52dB A The office or other public places we live in eve
36. l of the motor is disconnected within 1s after starting This will rapidly increase the coil temperature and then the thermal protection device is disconnected and it stops running When power is off the temperature of the thermal protection device rapidly falls and when it reaches the reset temperature power is connected At this time if the contact is not released the starting contact may have close contact and mistaken connection which may cause vibration of the compressor the thermal protection device is disconnected after a short time 2 Too high operating current intensity A vaccine refrigerator with too large operating current cannot be in normal operation It may caused by the following two causes First the vaccine refrigerator automatically stops after several seconds after it is started and then it is restarted again after several minutes before it stops again If repetition goes 42 on and after a long time the motor or the starter and the thermal protection device will be burned The main cause is the short circuit in some part of the starting coil Second after running for a certain period of time after the vaccine refrigerator is started it automatically stops before the temperatures falls to the temperature needed After a few minutes the vaccine refrigerator is started again and power consumption is greatly increased The cause is the poor insulation around the motor windings 43 3 Maintenance refrigerator
37. leakage occurs to the thermostat indeed replace the thermostat Electric leakage of starter relay during installation and maintenance bolt of the relay conductor easily contacts the shell thus causing electric leakage If any improper relay operation is found inspect it In case of leakage connect again and place a piece of insulating paper under the bolt 33 5 1 3 1 4 1 5 Electric leakage of the fan too high ambient humidity or improper operation of the control device for the fan during the application of the vaccine refrigerator will burn the fan after it has been working for a long time thus causing electric leakage Measure the insulating resistance between each wiring terminal and cabinet of the vaccine refrigerator with the voltammeter Too small resistance means a electric leakage caused by too high humidity a zero resistance means a burned or shortcircuited fan coil Please replace the fan Troubleshooting by observation Under normal circumstances temperatures around surface of the evaporator of the vaccine refrigerator are almost the same and the frost layer has similar thickness If there is no frost on surface of the evaporator or the whole internal wall of a bed type vaccine refrigerator or on a small part on its top the cooling system may have a fault including insufficient refrigerant or leakage Inspect the system piping to make sure there is no leakage especially on the connector Refrigerant with
38. me 2 Too dirty environment 3 Too high humidity 1 Replace the thermostat 2 Clear dust with a brush 3 Keep the thermostat dry and control the ambient humidity Thermostat failure 1 Leakage of temperature sensitive medium Replace the thermostat Poor contact between the temperature sensing probe and the evaporator Improper installation position of the probe or change in the position during transportation Reinstallation 40 Fan replacement procedure Disconnect the power and open the fan protection cover Unscrew the bolt of the fan with a straddle wrench or a universal tool Loosen the fan blades and take them out Take out the fan from between the compressor and the pipe Install a new fan from between the compressor and the pipe and install the fan blades in position before fixing them with a straddle wrench or a universal tool Trial operation inspect whether the fan is in normal operation and whether the blades are blocked During the removal of the fan remember not the bend or damage the pipe Thermostat replacement procedure Disconnect the power Remove the protection cover on the side Loosen the clamp nut Hold the bottom of the temperature sensing probe and pull it out downwards Replace with a new thermostat 41 2 Common failures in vaccine refrigerator 2 1 The vaccine refrigerator cannot be started With power on the vaccine refrigerator cannot be started First d
39. nce refrigerator and technical requirements ee 44 4 Application and maintenance of vaccine refrigerator eneen 46 5 Evacuating and filling procedures and tools needed 49 6 Noise from the vaccine refrigerator 51 7 National noise Standards nee names 56 8 Technology for measuring performance parameter nanne 58 Details of wearing parts nb A ir 62 Designation Regulations for type naming Design No is expressed with character sequence Rated volume value is shown with L Ice Lined Refrigerator a first letter of Pinyin of Haier name of the group b first letter of pinyin of bingxiang meaning vaccine refrigerator c first letter of pinyin of cang meaning refrigeration Note rated volume can be the gross volume or effective volume the manufacturer can decide it by him according to the actual situation The effective volume value must be marked on the nameplate whether effective volume or gross volume is marked in the product name Examples HBC 200 means that the Ice Lined Refrigerator with the temperature at characteristic temperature 2 8 C horizontal and rated effective volume 200 L Safe Caution Please read carefully all these notes before application of the vaccine refrigerator During application of the vaccine refrigerator remember to comply with the following basic safety notes 1 The vaccine refrigerator can only be used
40. ne refrigerator has been placed flat or tilted for some time it shall be kept upright for 24h before power is connected again for operation 18 Adjustment of thermostat I Introduction of display panel and buttons Three digits a minus sign and 5 symbols refrigeration power supply lock alarm and defrosting can be displayed on the display panel 1 Refrigeration symbol When the compressor is in operation the symbol is displayed 2 Power supply symbol When external power supply is connected to the controller the symbol is displayed 3 Lock symbol When the controller is locked the symbol is displayed 4 Alarm symbol When the controller is in state of alarming the symbol flashes to alert the user POWER SUPPLY En Z 1 2er 3la Isle l N mi Sensor L N 5 Defrosting symbol When the controller is in state of defrosting the symbol is displayed These four buttons S 4 AI and Fn exist on the instrument for operation of the controller 19 II Parameters and operation of controller Under normal circumstances the measured temperature is displayed by the controller and in case of failure of the sensor Err is displayed When the sensor is above 85 C EHi is displayed while it is below 50 C ELo is displayed 1 Parameter setting method Setting control temperature Under normal circumstances press the button S
41. ned 45 4 Application and maintenance of vaccine refrigerator 4 1 Product characteristics 1 The vaccine refrigerator uses refrigerant R134a completely free of CFC chlorofluorocarbon with an ozone depetion coefficient of ODP ozone depletion potential 0 Design with new climate type the vaccine refrigerator has a design with wide temperature range At temperature below 45 C 100 T it can be freely started and stopped 2 Design with a new refrigerating system a two way refrigerating system is applied to improve the refrigerating speed and extend the service life of the compressor 4 2 Notice for use 1 The vaccine refrigerator must be kept level and when it is moved the angle of inclination shall not be over 45 so that the sliding door will not slide out of the guide and cause accidents 2 The space between the vaccine refrigerator and other items shall be over 30cm to allow for ventilation and heat emission 3 Don t place the vaccine refrigerator at damp places or places with corrosive gases 4 The vaccine refrigerator shall be placed at places with good ventilation away from heat sources and without direct sunlight 5 Before application for the first time the vaccine refrigerator shall be in operation for a certain period of time 6h in summer and 4h in winter When the temperature inside the vaccine refrigerator reaches 2 Cor below the vaccine can be placed into it If the vaccine is placed i
42. ngth 6 starting performance 7 earthing resistance 8 sealing performance of refrigerating system 9 documents and attachments Ex factory field tests include 10 storage temperature 11 insulating resistance in wet state and in hot state 12 electric field strength in wet state and in hot state 13 power consumption 14 noise and vibration 15 galvanized parts 16 painted cover 17 protection against electric shock 18 cable 19 door opening force of the vaccine refrigerator 20 freezing capacity 21 increase speed of the load temperature Type tests include items 1 2 8 9 10 13 17 18 20 and 21 and the following items 22 total effective volume 23 ice making capacity 24 defrosting property 25 thermal insulating property and anti condensation 26 air tightness of the door seal 27 durability of latch and handle 28 strength of support or similar parts 29 materials and odor inside the vaccine refrigerator 30 firmness of the vaccine refrigerator and transportation test 8 2 Test methods and instruments 8 2 1 Sealing performance test of the refrigerating system During the test place the vaccine refrigerator in a positive pressure room with a temperature between 16 32 C The leak detector is set to be with an annual leakage of 0 5g and each position in the refrigerating system is inspected Leakage at any point shall not be over 0 5g Common leak detectors include Am
43. noise at the 4 positions is measured The value is the average sound pressure level of the noise and the formula is as follows LW Lpa 2 10 LgS Wherein Lw sound power level in dB Lpa average sound pressure level of the measured surface in dB S envelope area of the measured surface in m The standard 42dB by the Ministry of Light Industry is a sound pressure level measured 1m from the door of the vaccine refrigerator The value plus 10 shall be the approximate value of the sound power level The restriction on noise in national standards is stricter and more scientific Currently it is required to applied the sound power level specified in national standards for noise from vaccine refrigerators 6 13 Why noise varies from different types of vaccine refrigerators Answer the noise of each qualified vaccine refrigerator when leaving the factory is within the range specified in national standards Different types of vaccine refrigerators have different power and different compressors are used in them so the noise produced varies Even in vaccine refrigerators of the same type their compressor may produce different noise and the noise may have different differences between the national standard 52dB A so the noise levels vary 6 14 Why greater noise when the vaccine refrigerator is started and stopped Answer 1 When the vaccine refrigerator is being started the starter relay gives out sound Acceleration of the motor
44. not be used at low temperature When the ambient temperature is below zero lubricating ability of the refrigeration oil in the compressor of the vaccine refrigerator is reduced and it is difficult or even impossible to start the compressor or the compressor may be damaged 1 10How to know about failure in motor of compressor It is very easy to find out motor failure by inspecting resistance of the coil By connecting the ohmmeter to terminals of common and run resistance of the coil is measured An infinite resistance value means a broken circuit of the coil If the value of the resistance of the coil is much smaller than that of the motor the motor must be short circuited or the winding have been burned 36 1 11 Fault analysis of vaccine refrigerator and maintenance Fault Cause Solution The compressor cannot be started 1 Blown fuse 2 Poor contact between the plug socket and cable Broken cable 3 Too low voltage 10 15 below the rated voltage 4 Mistaken setting of the thermostat 5 Failure in the thermostat 6 Failure in or damage to the starter relay or thermal protection device 7 Broken starting coil or operating coil in the motor 8 Burned motor or blocked rotor and starter in the compressor 9 Leakage in the high pressure air valve or the shaft is locked 1 Inspect whether it is short circuited in case of short circuit correct the error if not replace with appli
45. nto the vaccine refrigerator too early it may be damaged because the temperature inside the refrigerator is still higher 6 Don t overstuff the vaccine refrigerator 7 Place reasonably the vaccine by category and try to control the frequency and time of opening door to reduce refrigerating loss and power consumption 8 There are two feet at the bottom of the vaccine refrigerator to make adjustment on uneven floor If frequent moving is required place the vaccine refrigerator on a base with trolley This type of vaccine refrigerator is usually used outdoors so make sure the plug is fixed to avoid necessary loss caused by unfixed plug 9 Itis not allowed to store food beverage or any other thing in the vaccine refrigerator 10 When the vaccine refrigerator is used outdoors don t place it at places with direct sunlight Regularly clean the inside and outside of the vaccine refrigerator especially the part of the engine room including the condenser and the compressor etc 4 3 Safety notice of the product 1 The vaccine refrigerator requires an independent earthing circuit with a minimum 46 4 4 current carrying capacity of 5A If the cable is to be extended the extended line shall have a area of not less than 1 5mm for the cross section Reliable earthing is required for the vaccine refrigerator the earthing wire is not allowed to be connected to water pipe and gas pipe If power of the vaccine refrigerator is
46. o a conclusion that the sound is great because it is heard several meters away 6 16 How to determine that the sound is noise or it is normal Answer generally noise from vaccine refrigerator within the range specified in national standards is normal However if a continuous sound of metal colliding with each other is produced by the compressor inspect it in time to have it settled 55 7 National noise standards There are related regulations in WHO 7 1 No obvious noise when the vaccine refrigerator is running For vaccine refrigerators with a capacity of 250L or below the sound power level measured by WHO shall not over 52dB A For vaccine refrigerators with a fan the sound power levels of the noise shall be as follows Below 300L lt 55dB A 300 400L incl 400L lt 60dB A 401 500L incl 500L lt 63dB A 7 2 Noise test In geometric center of the measurement site place the empty vaccine refrigerator onto an elastic base elastic rubber pad 5 6mm thick Set the thermostat to 4 and close the door or cover Carry out the measurement after the vaccine refrigerator has been in operation for 30min During test if the vaccine refrigerator reaches the temperature and stops the measurement shall be interrupted and it will not continue until 3min after the vaccine refrigerator is restarted The test shall be implemented in semi anechoic environment Other test sites shall satisfy each regulation specified in WHO
47. of vaccine refrigerator 1 Electric leakage of the compressor main due to violent vibration Motor coil of the compressor contacts the shell which causes falling off of coating of the plastic wires on some parts thus contacting the shell Other causes may include too high voltage too low voltage adverse cases for vaccine refrigerator too high temperature too poor ventilation and heat emission and failure of the thermal protection device etc All these may cause overheating of the motor coil loose insulating materials or damage to insulating layer so that electric leakage occurs Electric leakage of wiring terminals of the motor wiring terminals of hermetic compressor are made of ceramic fired at high temperature from soda glass and high polymer is filled around the wiring terminal Linear expansion coeffiecients of the terminal and the insulating material shall be the same and the insulating resistance between the terminal and the shell is required to be over 0 5million ohms Cause for electric leakage on the terminal is damage to the soda glass or that there exists conductive materials such as water and dirt etc In case of external electric leakage of the compressor clean the compressor with CCL or alcohol In case of internal electric leakage of the compressor replace it 3 Electric leakage of the thermostat too long operation time too high temperature too much dirt or high humidity may cause electric leakage If
48. ping with cloth to avoid rust To ensure good refrigerating effects please regulars clean the condenser the compressor and dust proof components inside the engine room with a soft brush clean the dust proof components every 2 3 days Downtime if the vaccine refrigerator is out of operation for a long time disconnect the power clean it with methods above mentioned and open its door until it is dry before it is finally in safe keeping Standard maintenance procedures Clean cabinet of the vaccine refrigerator remove the board on the cabinet and the protection cover on the back with a Phillips screwdriver Clear dusts from the condenser fan compressor and thermostat with a brush Clean it by wiping and avoid damage to the pipe and electrical parts 47 Connect the power and inspect whether the vaccine refrigerator is in normal operation In case of any abnormal case found inspect it and repair it Inspect all brazed joints In case of falling off of black paint or rust polish the surface slightly with polishing paper 0 When getting rid of rust clean with a piece of clean towel Then apply acid soluble paint on the welded joint and pipes with a brush Inspect the foot in case of any damage replace it Inspect the power line in case of any abrasive wear or damage replace it Inspect whether the resistance of the earthing resistor is in normal operation with a multimeter Inspect the insulation with a multimeter in case o
49. place the compressor 8 Replace with a new dry filter The compressor keeps running and the temperature inside the vaccine refrigerator is too low 1 Temperature of the thermostat is set to be too low 2 The starter contact is stuck 3 Improper position of the thermostat probe thus causing maladjustment 4 Reduced efficiency of the compressor 1 Adjust to a suitable temperature 2 Repair or replace it 3 Adjust to a suitable position 4 Replace the compressor The compressor is buzzing and cannot be started with repeated tripping of the thermal protection device After operation for a short time the overload protection device is disconnected Too loud noise during operation of compressor 1 Too low voltage 2 Failure in starter relay 3 Broken starting coil in the starting motor 4 The compressor is out of operation 1 Too high voltage 2 Improper overload protection device and early tripping 3 The starter contact is stuck 4 Short circuit of the compressor 5 Mechanical failure in the compressor 6 Too high temperature around the compressor 1 Soft ground 2 Unstable and not flat cabinet of the vaccine 38 1 Install a voltage stabilizer and adjust it to the rated value 2 Replace the starter relay 3 Replace the compressor 4 Axial piston in the compressor is blocked and it shall be replaced 1 Install a voltage stabilizer and adjust i
50. ry day have much greater noise than vaccine refrigerators 6 11 Why the noise level of vaccine refrigerator measured at home is not correct Answer environmental conditions for measuring noise from the vaccine refrigerator are not available at home so the value measured is invalid Conditions for noise measurement specified in national standards are as follows 1 Measurement shall be carried out in semi anechoic or anechoic rooms without reflection of sound wave 2 Place the vaccine refrigerator on a elastic cushion 5 10mm thick 3 Before measurement have the vaccine refrigerator running idle for at least 30min In case of suddenly stop restart it and carry out the measurement after 3min 4 4 measurements shall be carried out which are implemented respectively 1m from the front the back the left front and the right front of the vaccine refrigerator with a measurement height of 1 2 of the height of the vaccine refrigerator accurate sound level meter is applied A weighted 53 6 12 What are the differences between the national noise standard 52dB A and the standard 42dB by Ministry of Light Industry The 52dB A in national standard is a sound power level Under measurement conditions specified in national standards the average of readings of the sound lever meter under circumstances with greater noise is taken among the noise from vaccine refrigerator measured at the 4 positions so that the arithmetical mean of the
51. s above AH high temperature alarm occurs Or when the measured temperature is above 10 C the mode of high temperature alarm occurs AL temperature of low temperature alarm when the measured temperature is below the set value of AL and after 1h it still remains below AL low temperature alarm occurs Or when the temperature is below 0 5 C low temperature alarm occurs SP1 SP2 parameters for restarting operation after the controller has power failure 25 After the controller has a power failure of 12h and when it is started again the program for controlling is forced to be run as follows 1 Forced start up for 3h first 2 Running 2h with SP1 as the control temperature 3 Running 2 5h with SP2 as the control temperature 4 Forced shutdown for 0 5h at last After completion in the above orders return to the normal operating mode In case of temperature below 1 0 C during forced running of the program return to the normal operating mode and forced running of the program ended in advance The mode is only valid in mode A 26 Administration menu in mode B Parameter Description Min Max Unit Default ro Control temperature difference 0 5 5 0 G 2 r1 Max temperature can be set r2 50 1e 8 r2 Min temperature can be set 50 r1 C 2 CAL Calibrated temperature of sensor 3 0 3 0 0 t1 Delay in start up 0 12 Min 0 t2 Shortest downtime of compressor 0 60 Min 7 3 en
52. ssor is stopped it can be started again after a delay of time t2 If t2 0 the delay is cancelled not related to delay in start up t3 forced start up time after energized after a power failure of 15h In case of a power failure over 15h and when energized again the compressor is required to be in continuous operation for a time of t3 before it has a downtown of 2h t3 0 indicates no forced start up when energized AO setting range of high and low temperature alarm Setting range of parameters AH and AL shall be between St AO and St AO 24 If St 5 and A0 6 then parameters AH and AL can only be adjusted between 1 and 11 A1 power on alarm and delay in defrosting alarm The parameter is invalid in mode A A2 alarm delay The parameter is invalid in mode A dFO defrost cycle When the compressor has a total running time of over dFO defrosting is started dF 1 defrost time After defrosting is started and the defrosting time reaches dF1 the defrosting is stopped If dF 1 0 the defrosting function is invalid ct0 start up time with failure of sensor In case of failure of the sensor run automatically a start up time of ctO and a downtime of ct1 and repeat the cycle When ct0 0 the downtime continues ct1 downtime with failure of the sensor Refer to ct0 AH temperature of high temperature alarm When the measured temperature is above the set value of AH and after 10h it still remain
53. t to the rated value 2 Replace the overload protection device 3 Replace the starter relay 4 Replace the compressor 5 Replace the compressor 6 Increase space for heat emission 1 Reinforce the ground 2 Stabilize body of the vaccine refrigerator refrigerator 3 Resonance caused by the friction between piping and body of the vaccine refrigerator 4 Loose fixing screw of the compressor 5 Too tight too loose or aged shock absorber for fixing the compressor 6 Too loud noise in the compressor or broken suspended spring for reducing shock 3 Remove the pipe to avoid friction 4 Suitably fasten the screw 5 Adjust tightness of the shock absorber or replace the shock absorber 6 Replace the compressor No frost on the evaporator 1 Serious leakage of refrigerant 2 The capillary or the system is blocked with dirt 1 Inspect for leakage repair welding and add refrigerant 2 Clean the capillary or replace the filter Insufficient frost on evaporator Too rapid frosting too thick frost layer or icing of the evaporator 1 Little leakage of the refrigerant 2 Blocked with dirt but not serious 3 Reduced refrigeration capacity of the compressor 1 Vaccine stored in the vaccine refrigerator is with water not dry 2 Too much vaccine stored in the vaccine refrigerator 1 Inspect for leakage repair welding and add refrigerant 2 Clean the capillary
54. th the 79 630 1435 1240x630x1462 door open 16 installation and accessories Removing package of ice lining vaccine refrigerator 1 Remove all packaging materials including the foam base and all tape for fixing accessories inside and outside the vaccine refrigerator Before power is connected to the vaccine refrigerator inspect and remove any remaining packaging material tape or printed material Adjusting vaccine refrigerator 1 The vaccine refrigerator is specially designed for independent installation and it is not allowed to be placed or embedded into cupboard The floor for placement of the vaccine refrigerator shall be solid enough for bearing weight of the vaccine refrigerator stuffed with articles When removing the vaccine refrigerator make sure the angle of inclination shall not be above 45 degrees Otherwise the compressor and the sealing system will be damaged If the vaccine refrigerator has been tilted it shall be placed upright for at least 24h before it is plugged in Good ventilation To ensure that the vaccine refrigerator reaches the designed maximum operating efficiency it shall be installed in placed with good ventilation Piping and electrical connection Recommended space around the vaccine refrigerator is as follows dn 2 50mm Top 2 50mm Back 2 50mm Installation and accessories Don t overstuff the vaccine refrigerator to
55. the above problems will have impact on heat emission of the condenser to reduce its refrigerating capacity Inspect whether whether too much vaccine is placed in the vaccine refrigerator The placement of hot or too much vaccine is bad for the circulation of the cold air Only after the hot vaccine is cooled can it be placed into the vaccine refrigerator Inspect whether the door of the vaccine refrigerator is closed the door seal is too dirty or the door has bad sealing Remove the dirt and add a small amount of talc powder Frequent open and closing of the door of the vaccine refrigerator and open of the door for a long time will reduce the refrigerating capacity How to settle the problem of bad sealing by seal In case of too many dirts or bad sealing clean the seal with cloth dipped in warm water After the cleaning a small amount of talc powder shall be added If there is still gap blow the external surface of the seal with a hair dryer to heat it and keep the temperature between 50 C 60 C Press the seal to recover it and by pressing space around the seal with the hand the seal can be recovered If the method is invalid stuff a piece of sponge around 10mm wide 5mm thick and a bit longer than the gap into the seal that is teh position between the seal bottom and the door washer The sponge will raise the gap position and it will adhere to the vaccine refrigerator body like at other positions 35 1 9 Why vaccine refrigerator can
56. ture has been continuously above AH or below AL for duration of A2 When A2 0 alarm delay is cancelled dFO defrost cycle When the compressor has a total running time of over dFO defrosting is started dF 1 defrost time After defrosting is started and the defrosting time reaches dF1 the defrosting is stopped If dF1 0 the defrosting function is invalid ct0 start up time with failure of sensor In case of failure of the sensor run automatically a start up time of ctO and a downtime of ct1 and repeat the cycle When ct0 0 the downtime continues ct1 downtime with failure of the sensor Refer to ct0 29 AH temperature of high temperature alarm When the measured temperature is above the set value of AH high temperature alarm occurs AL temperature of low temperature alarm When the measured temperature is below the set value of AL low temperature alarm occurs Parameters SP1 and SP2 are invalid in mode B Inquiry to the maximum and minimum temperature Hold the button amp for 3s and the maximum temperature begins to be displayed by flashing for 3s Hold the button Y for 3s and the minimum temperature begins to be displayed by flashing for 3s Clear maximum and minimum temperature records Hold simultaneously buttons 4 and Y for 3s and the maximum and minimum temperature records are cleared and CLr is displayed at the same time on the display screen for 2s Lock and unlock the