Home
NL100 Nitrogen Laser - Stanford Research Systems
Contents
1. trigger pulse high voltage laser output Figure 12 Timing diagram of AUTO CHARGE mode Note that time is not shown to scale and the diagram does not reflect triggering delays etc In COMMAND CHARGE mode high voltage charging circuitry becomes active only when a TTL low appears on the trigger input as depicted in Figure 13 To allow for complete high voltage charging In this case the dwell time of the trigger pulse must be gt 30 ms In other words the trigger must be TTL low for gt 30 ms in COMMAND CHARGE mode SRS NL100 Nitrogen Laser 6 Operation trigger pulse high voltage laser output Figure 13 Timing diagram of COMMAND CHARGE mode Note that time is not shown to scale and the diagram does not reflect triggering delays etc Where possible we recommend using the NL100 in the COMMAND CHARGE mode This mode limits the amount of time high voltage resides on the laser components and therefore improves laser performance and lifetime As shown in Figure 14 the trigger pulse is optoisolated from the rest of the laser electronics This allows triggering of the laser without permitting electrical noise from the laser discharge from impacting the user s electronics TRIG IN gt gt TRIGGER CIRCUIT Figure 14 Schematic diagram of optoisolated trigger input Rate When the RATE knob is in the EXT TRIG position i
2. tJ O oO00000000000000 0000 O000000000000 O O oO00000000000000000 O000000000000 O DODOOOOOODODOOOOOODODOO DOODOOOOODODODOO oO ODOOOOOODOOOOODODO ODDOOOCOOODDOOO oO 0000000 OO000000000000 oO 0000000000 000000000 O O O0O0000000000000000 00000000000000 Q OQ oO00000000000000 0000 O00000000000000 O oO 0000 0000000000000040 oO DODOOOOOODOOOOOOOOO DODOOOOOODDOOOOOO O 9 Q Q Q Oo Lon D oO oO oO A cl oO OD LI jama oO LOS D uy oO NW a a D vo oO ce Li H N IN 8 32 UNC 2B 2 PLACES
3. human eyes and skin Therefore all persons operating the NL100 laser and all persons in the vicinity of the NL100 laser when it is in operation must be aware of the hazards of UV laser beams Basic Safe Laser Practices e NEVER look directly into the laser beam e Only qualified personnel should operate the laser system e Always wear laser safety glasses or eyewear that protects against UV laser light whenever the POWER LED emission indicator is lit See below for a partial vendor list of Do not expose skin to the laser beam because burns may result Chronic exposure of skin to UV light may increase the risk of skin cancer e Do not expose eyes or skin to specular reflections generated by the laser beam and an optical surface e Securely mount the NL100 laser system and all optical elements in the beam path Trace all specular reflections of the laser beam and block eye or skin access to these specular reflections e NEVER operate the laser with the protective housing off The laser power supply generates high voltages and currents that have the potential to produce a lethal electrical shock e If the NLI00 is used in a manner not specified by the Stanford Research Systems the protection provided by the NL100 may be impaired Stanford Research Systems recommends that users obtain the American National Standards ANSI Z136 1 2000 Safe Use of Lasers available from the Laser Institute of America on the Internet at www laserinstitute
4. that when the NL100 performance has degraded below the specifications listed in this manual or 2 years have elapsed since the unit was purchased or last serviced that the unit be returned to SRS for cartridge replacement test and calibration Please contact SRS at 408 744 9040 to obtain a return merchandise authorization RMA number For the technically sophisticated NL100 user who wishes to replace the laser cartridge themselves first obtain a replacement laser cartridge from SRS and then carefully follow the instructions and illustrations given below We strongly recommend that only technically sophisticated users who are familiar with lasers and high voltage circuits perform this service operation All other customers should return the unit to SRS for service If you are in doubt call SRS Laser Cartridge Replacement Tools Required 0 Replacement laser cartridge a PHI Phillips screwdriver OQ Nut Driver 4 Procedure 1 Turn the NL100 laser key switch off and disconnect all cables 2 Remove the Phillips screws that secure the protective housing to the laser frame and remove the protective housing SRS NL100 Nitrogen Laser 12 Maintenance and Service Figure 18 Remove the screws on the protective housing WARNING Never provide power to the NL100 laser or attempt to operate the NL100 laser with the protective cover removed Doing so may expose the user to lethal voltages 3 Disconnect the trigger and high vo
5. Operation and Service Manual NL100 Nitrogen Laser 5 RS Stanford Research Systems Revision 1 12 e January 24 2014 Certification Stanford Research Systems certifies that this product is in compliance with IEC 60825 and U S Code of Federal Regulations Title 21 Part 1040 10 and that it met its published specifications when shipped Warranty This product is warranted against defects and materials for a period of one 1 year after shipment The laser cartridge is warranted for a period of one 1 year or 20 000 000 laser shots whichever comes first Service For warranty service or repair this product must be returned to a Stanford Research Systems authorized service facility Contact Stanford Research Systems or an authorized representative before returning this product for repair Information in this document is subject to change without notice Copyright Stanford Research Systems Inc 2008 2010 2014 All rights reserved Stanford Research Systems Inc 1290 C Reamwood Avenue Sunnyvale California 94089 www thinksrs com Printed in U S A SRS NL100 Nitrogen Laser Safety and Preparation For Use iii Safety and Preparation for Use CAREFULLY READ THE SAFETY INFORMATION IN THIS SECTION BEFORE USING THE NL100 NITROGEN LASER LASER SAFETY PAYS The NL100 Nitrogen Laser is a Class 3B that emits invisible ultraviolet UV radiation at 337 1 nm with sufficient intensity to be considered an acute hazard to
6. d connectors to the appropriate receptacle as shown in Figure 25 Be sure to tighten the knurled nuts for each connector to the receptacle Figure 25 Connect the trigger and HV connector plugs to their respective receptacles 10 Make sure the HV lead is clear of all components attached to either printed circuit board as shown in Figure 26 It may touch the trigger wire lead Figure 26 Make sure HV lead is not near board components WARNING Improper trigger and HV connections will result in the unit not operating permanent damage to the unit and expose the user to unsafe conditions Contact SRS if you are unsure if you performed this step correctly 11 Secure new cartridge to bottom plate with four 4 40 nuts 12 Carefully reposition the protective cover and replace the 4 40 Phillips screws that secure the cover ASRS NL100 Nitrogen Laser 16 Mechanical Drawings Mechanical Drawings FRONT VIEW 4 1 938 1 500 p _ 5 1 531 000 000 375 1 630 REAR VIEW D a DI Co 1e 10 32 UNC 2B 1 875 000 1 4 20 UNC 2B 3 PLACES SRS NL100 Nitrogen Laser 1 025 TOP VIEW 8 32 UNC 2B 2 PLACES 4 40 UNC 2B 2 PLACES 150 000 1 875 875 000 O Ob O
7. e connection for externally triggering the laser Note that because the EXTERNAL TRIGGER is optoisolated the reference return for the trigger current Pin 4 Trigger is isolated from GND See TRIGGER IN above for additional information Pin 9 Interlock Return for Pin 1 Interlock LED Indicators Power LED This LED is illuminated whenever the NL100 laser system is powered A three second delay circuit prevents the laser from firing when electrical power is first applied to the laser system to allow time for the laser beam path to be cleared Charge LED This LED is lit when the high voltage charging circuit is active n AUTO CHARGE mode the Charge LED is on whenever the laser is powered In COMMAND CHARGE mode the Charge LED flashes on while the trigger pulse is high Trig LED The Trig LED flashes when the NL100 laser system either receives a valid external trigger or generates a trigger internally SRS NL100 Nitrogen Laser Operation 9 Beam Shutter The beam shutter is located at the front of the NL100 laser system as shown in Figure 15 and Figure 16 Placing the beam shutter in the closed position mechanically blocks the beam aperture preventing laser emission from exiting the laser housing ASRS Stanford Research Systems Model NL100 Figure 15 Front of the NL100 laser system showing the beam shutter in the closed position SRS Stanford Research Systems Model MLIOO Figure 16 Front of the NL100 laser syste
8. ed low the NL100 laser system will be in COMMAND CHARGE mode With COMMAND CHARGE the high voltage charging begins only when a TTL low level is received at one of the TRIGGER inputs either the TRIG IN BNC or Pin 4 of the DB 9 Since high voltage charging requires 30 ms the Trigger pulse width must be greater than 30 ms when COMMAND CHARGE is used Because COMMAND CHARGE reduces the amount of time key laser components are at high voltage especially at lower repetition rates we recommend use of this mode where possible Pin 4 Trigger Current return for opto isolated external trigger pulse Note that because the EXTERNAL TRIGGER is optoisolated the reference for the trigger Pin 8 Trigger is isolated from GND See Pin 8 Trigger below and TRIGGER IN above for more information Pin 5 Interlock The laser system provides 24V 10 mA max level that must be terminated into Pin 9 Interlock or GND for the laser system to operate The NL100 laser system is shipped with one male DB 9 connector where this connection is provided NOTE The laser will not operate if the provided male DB 9 connector is not in place and the Pin 5 Interlock is not otherwise terminated This interlock is provided so that the user can control the laser system externally with a door switch for example Pin 6 and Pin 7 GND Ground and electrical power return SRS NL100 Nitrogen Laser 8 Operation Pin 8 Trigger Pin 8 provides an alternat
9. he box and inspect all components of the NL100 laser system Report any damage to Stanford Research Systems immediately Compare the contents of the shipping box against your original order and the checklist below Report any discrepancies immediately Standard Equipment and Supplies Q Q Q E NL100 laser system unit One Hypertronics DO1PB 306 MST plug Two keys for the key switch One DB 9 interlock defeat connector Optional Equipment E E Optical synchronous detection circuitry installed in factory Sync out BNC connector back panel Additional Necessary Equipment Q 24 VDC 50 W 3 A peak power supply SRS NL100 Nitrogen Laser 2 Quick Start Quick Start WARNING UV light can be dangerous Be familiar with the safety information in this manual before proceeding 1 Supply 24V power See specifications on Page vii for power requirements 2 Defeat Interlock Insert the DB 9 plug supplied with your laser into the DB 9 connector on the back panel of the laser or hook up your own laser interlock connector to the DB 9 on the laser See Page 7 for more details The laser will not operate if the interlock is not defeated Figure 8 Supplied interlock defeat connector 3 Turn on Internal Trigger Turn RATE switch from OFF to somewhere between Hz and 20 Hz Alternatively you can supply an external trigger See Page 5 6 for more information on the requirements for external triggering 4 T
10. located on the bottom right of the back panel A Hypertronics connector is supplied with your laser Alternatively power can be supplied through the DB 9 connector see DB 9 CONNECTOR below Figure 10 Hypertronics power receptacle as viewed from the back panel of the NL100 Supply 24V DC on Pin 1 Pin 2 is the supply return and pin 3 is the case ground Sync Out Optional This BNC connector outputs a 5V pulse into 50 Ohms synchronously with the laser emission from the front aperture Figure 11 shows a single shot from the sync out BNC captured with an oscilloscope 8 00 Gass M re Oe M zit i EH Figure 11 Single shot oscilloscope trace of the sync out of the NL100 ERS NL100 Nitrogen Laser Operation 5 Trigger In When the RATE switch is in the OFF position see RATE below the NL100 is triggered by a TTL pulse at this BNC connector The laser will fire the laser on the rising edge of the TTL pulse The width required of the TTL pulse depends on whether the laser is being operated in COMMAND CHARGE or AUTO CHARGE mode See DB 9 CONNECTOR below for information on how to configure the laser in either COMMAND CHARGE or AUTO CHARGE modes In AUTO CHARGE mode the falling edge of the trigger pulse has no effect on the operation of the NL100 laser system as shown in Figure 12 The trigger pulse width should be gt 1 us The high voltage charging circuitry becomes active immediately after the laser has fired
11. ltage connectors on the laser cartridge as shown in Figure 19 Figure 19 Disconnect high voltage connectors to the laser cartridge ASRS NL100 Nitrogen Laser Maintenance and Service 13 4 Disconnect the ground and monitor line connector as shown in Figure 20 Figure 20 Disconnect the ground and monitor connector 5 Remove the four 4 40 nuts that secure the laser cartridge to the bottom plate Figure 21 Remove laser cartridge mounting nuts 6 Remove the laser cartridge Figure 22 Remove laser cartridge ASRS NL100 Nitrogen Laser 14 Maintenance and Service 7 Connect ground and monitor line connector plug to the power supply board as shown in Figure 23 Notice that both the cartridge connector and the circuit board header are polarized Take care to ensure proper orientation of the cartridge connector WARNING Be certain this connection is made is correctly before proceeding If the unit is powered on after this connection is improperly made the cartridge will be irreversibly damaged Figure 23 Reconnect ground and monitor line connector 8 Identify the trigger and HV leads coming from the board transformers As shown in Figure 24 the trigger lead is the shorter of the two and connects to the upper female high voltage connector on the laser cartridge Figure 24 Identify the trigger and HV leads ASRS NL100 Nitrogen Laser Maintenance and Service 15 9 Connect the trigger and HV lea
12. m showing the beam shutter in the open position SRS NL100 Nitrogen Laser 10 Operation High Voltage Adjust The energy output of the NL100 laser system s roughly proportional to the high voltage charge The high voltage can be changed by adjusting the ten turn potentiometer that can be accessed through the hole labeled HV ADJUST located n the left side of the protective housing The high voltage of each unit is set during testing such that the unit meets specifications As the pulse energy drops as the laser tube ages through use increasing the high voltage will help offset the drop in pulse energy We do not recommend running the NL100 laser system with the high voltage adjusted to the maximum for long periods because this will shorten the life of laser cartridge If your application does not require full pulse energy of the NL100 reducing the high voltage may significantly increase cartridge life Figure 17 Changing the high voltage adjustment potentiometer ASRS NL100 Nitrogen Laser Maintenance and Service 11 Maintenance and Service Maintenance The NL100 laser system with a sealed laser tube within a modular laser cartridge is designed to be completely maintenance free Service The laser cartridge which contains the laser tube and other high voltage components that are expected to degrade with use can be replaced thereby restoring the NLIOO laser system performance to that achieved when new We recommend
13. ndicator pointing down the laser system is in EXTERNAL TRIGGER mode and will fire only when an appropriate pulse is applied either to the TRIG IN BNC or through the DB 9 connector When the RATE knob indicator is between and 20 the laser system is in INTERNAL RATE mode and will generate laser pulses between 1 20 Hz depending on the positioning of the RATE knob indicator SRS NL100 Nitrogen Laser Operation 7 NOTE When the laser system is in INTERNAL RATE mode it s also in AUTO CHARGE mode regardless of the configuration of the DB 9 connector When the laser is in INTERNAL RATE mode trigger pulses applied at either the TRIG IN BNC or the EXT TRIG pin of the DB 9 connector have no effect on the operation of the laser DB 9 Connector The female DB 9 connector provides a convenient way to interface to the NL100 laser s inputs and outputs Below is a description of the function of each pin Pin 1 and Pin 2 24V DC In An alternate connection for providing electrical power to the NL100 Pins and 2 are tied together Pin 3 CHARGE CONTROL The NL100 laser system provides a 5V 0 1 mA max level at Pin 3 If left high the laser system will be in AUTO CHARGE mode In AUTO CHARGE the high voltage charging system will begin to recharge the storage capacitor and spark gap immediately after the laser fires Use of the DB 9 connector provided with the NL100 will put the laser system in the AUTO CHARGE mode If Pin 3 is pull
14. org or the IEC 60825 1 Safety of Laser Products Part 1 Equipment Classification Requirements and Users Guide prior to operating the NL100 nitrogen laser system CAUTION USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPOSURE SRS NL100 Nitrogen Laser iv Safety and Preparation For Use Safety Features The NL100 laser system has been designed with the required safety features for compliance with IEC 60825 and 21 CFR 1040 10 l Protective housing 4 Emission indicator 2 Beam shutter 5 Key switch 3 Remote interlock connector 6 Three second delay not shown Figure 1 Front and back views of the NL100 Warning Labels The NL100 laser system has been designed with the required warning labels for compliance with IEC 60825 and 21 CFR 1040 10 A Warning logotype C Protective housing warning B Aperture warning D Certification Identification label On bottom plate Figure 2 Location of warning labels on the NL100 ASRS NL100 Nitrogen Laser Safety and Preparation For Use V The warning labels that are attached to the NL100 or are painted onto the case are reproduced below These warning labels are a requirement for compliance with 21 CFR A 1040 10 and IEC 60825 Figure 3 Laser hazard label located on the front panel near output aperture Figure 4 Invisible laser radiation label located on the side of the pr
15. otective cover Figure 5 Aperture label located on the front panel near output aperture Figure 6 Caution label located on both sides of the protective cover THIS LASER PRODUCT COMPLIES WITH 21 CFR 1040 10 MODEL no NNNNNNNNNNNNN SERIAL NO SRS stanford Research Systems Inc Sunnyvale CA Figure 7 Identification and certification label located on bottom plate SRS NL100 Nitrogen Laser vi Safety and Preparation For Use Vendors of Laser Safety Eyewear Below is a partial list of vendors of laser safety eyewear SRS does not endorse these vendors or their products over other available safety eyewear vendors and products and provides this vendor list only for reference Kentek Corporation 1 Elm Street Pittsfield NH 03263 800 432 2323 www kentek laser com Rockwell Laser Industries P O Box 43010 Cincinnati OH 45243 800 945 2737 www rli com Lase R Shield 900 Douglas Pike Smithfield RI 02917 1 800 288 1164 www lase rshield com GPT Glendale Inc Glendale 10 Thurber Boulevard Smithfield RI 02917 1896 800 500 4739 www glendale laser com Laser Vision USA 595 Phalen Boulevard St Paul MN 55101 800 393 5565 www lasersafety com SRS NL100 Nitrogen Laser NL100 Nitrogen Optical Laser wavelength Optical bandwidth Pulse repetition rate Pulse width Pulse energy Peak power Average power Beam size Beam divergence Shot to shot energy stability Electrical Trigger mode Power
16. requirements Mechanical Dimensions Weight Specifications vii Laser Specifications 337 1 nm 0 1 nm 1 20 Hz internal trigger 0 20 Hz external trigger lt 3 5 ns FWHM 170 uJ at 10 Hz 45 kW 3 4 mW 3x 7mm 5 x 8 mrad lt 3 standard deviation internal or external 24 IVDC 3 A peak 1 5 A average 20 Hz 11 L x 3 75 W x 3 757H 27 9 x 9 5 x 9 5 cm 7 5 lbs 3 4 kg SRS NL100 Nitrogen Laser viii Specifications SRS NL100 Nitrogen Laser Contents ix Contents Certification 1 Warranty ii Service 1 Safety and Preparation for Use iii Basic Safe Laser Practices lil Safety Features iv Warning Labels Iv Vendors of Laser Safety Eyewear vi NL100 Nitrogen Laser Specifications vii Contents Ix Unpacking 1 Checklist 1 Standard Equipment and Supplies 1 Optional Equipment 1 Additional Necessary Equipment 1 Quick Start 2 Operating Instructions 3 Operating Controls 3 Key Switch 3 Power Input 4 Sync Out Optional 4 Trigger In 5 Rate 6 DB 9 Connector 7 LED Indicators 8 Beam Shutter 9 High Voltage Adjust 10 Maintenance and Service 11 Maintenance 11 Service 11 Laser Cartridge Replacement 11 Tools Required 11 Procedure 11 Mechanical Drawings 16 SRS NL100 Nitrogen Laser X Contents SRS NL100 Nitrogen Laser Unpacking 1 Unpacking WARNING Read the safety information and these operating instructions carefully before operating this laser system Checklist E E E Open t
17. urn Key Switch to On Position Turning the key switch to the on position allows power to flow to the laser system 5 Open Shutter After a 3 second delay after turning the key switch you will hear the laser firing and can observe the laser beam by placing a fluorescing material e g a business card in the path of the beam SRS NL100 Nitrogen Laser Operation 3 Operating Instructions WARNING Read the safety information and these operating instructions carefully before operating this laser system Operating Controls The operating controls for the laser system with the exception of the beam shutter are located on the back panel as shown in Figure 9 POWER CHARGE L t INT RATE Hz 1 20 ext trig SYNC OUT A Figure 9 Back panel and operating controls of the NL100 Key Switch Turning the key switch to the ON position connects the 24V DC input power to the laser system the POWER LED illuminates and the laser system begins a 3 second delay timer Following the 3 second delay the laser is armed and ready to fire either from an SRS NL100 Nitrogen Laser 4 Operation internal trigger see RATE below or from an external trigger The key cannot be removed from the key switch when the switch is in the ON position Power Input The NL100 requires 24V DC 3A peak 1 5 A average input power Power may be applied through a Hypertronics DO1PB 306 MST plug connected to the Hypertronics receptacle