Model 9307 pico-TIMING™Discriminator Operating and

image

Contents

1. Monochromator gt PMT Sample 556 High Voltage Supply 9306 or VT120 Preamplifier Photodiode 9307 pico TIMING Discriminator 9306 1 GHz Preamplifier 921 or ACE Multichannel Pulse Height Analyzer Time to Amplitude Converter TAC Delay 90 of TAC Time Range Em 9307 pico TIMING Discriminator 425A Nanosecond Delay Personal Computer Cavity Dumped Pulsed Laser A ed Beam Splitter Fig 1 2 Typical Block Diagram for Fluorescence Lifetime Spectrometer 2 SPECIFICATIONS 2 1 PERFORMANCE TIME SLEWING Walk lt 20 ps shift in the timing output for input signal amplitudes from 150 mV to 1 5 V Typically lt 50 ps for signal amplitudes from 50 mV to 5 V Measured using a 1 ns wide input pulse with 350 ps rise and fall times PULSE PAIR RESOLUTION 10 ns at the fast negative NIM outputs MAXIMUM INPUT OUTPUT RATE Accepts burst rates up to 100 MHZ OPERATING TEMPERATURE RANGE 0 TO 50 C THRESHOLD TEMPERATURE SENSITIVITY 30 1 mV C 0 to 50 C TRANSMISSION DELAY TEMPERATURE SENSITIVITY lt 10 ps C 0 to 50 C 2 2 CONTROLS AND INDICATORS THRESHOLD Front panel 10 turn potentiometer with locking dial allows adjustment of the input discriminator threshold from 25 mV to 1 V SLEWING COMPENSATION Front panel 20 turn screwdriver fi
2. Pin gt oo 11 12 18 14 15 16 17 18 19 20 21 22 Function 3 V 3V Spare Bus Reserved Bus Coaxial Coaxial Coaxial 200 V dc Spare 6 V 6V Reserved Bus Spare Spare Reserved 12 V 12 V Spare Bus Reserved Bus Spare Spare Reserved Pin 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 Function Reserved Reserved Reserved Spare Spare 24 V 24 V Spare Bus Spare Spare 117 V ac Hot Power Return Ground Reset Scaler Gate Reset Auxiliary Coaxial Coaxial Coaxial 117 V ac Neutral High Quality Ground Ground Guide Pin Pins marked are installed and wired in ORTEC s Model 4001A 4001C Modular System Bins
3. et eot E tutae utet ue uo eo eec 2 24 OUTPUTS v sse EE eu xeu eue pd Wege decur DE US 2 2 5 ELECTRICAL AND MECHANICAL 3 2 6 ORDERING INFORMATION 2 3 3 INSTALLATION AND OPERATION 3 BAT GENERAL eec erect e nete eth enl e i e e t e er e v C 3 3 2 CONNECTION TO POWER cer xr PRU ROUX Ra WE BE HS 3 3 3 INPUT CONNECTIONS ais et nece bbe bbe 4 3 4 OUTPUT GONMEGTIGNS 52 6 4 3 5 ADJUSTING THE INPUT SIGNAL AMPLITUDE 5 3 6 THRESHOLD ADJUSTMENT er rper erar arena ur e e a uf a a a Cal dunt Au e 6 3 7 ADJUSTING THE SLEWING COMPENSATION 6 3 8 LIMITING THE COUNTING RATE TO AVOID SPECTRUM DISTORTION 7 3 9 THE BENEFIT OF REVERSED 7 4 MAINTENANCE AND 8 SAFETY INSTRUCTIONS AND SYMBOLS This manual contains up to three levels of safety instructions that must be observed in order to avoid personal injury and or damage to equipment or other property These are DANGER Indicates a hazard that coul
4. it may be desirable to raise the Threshold above the value set in the previous paragraph The smallest pulse amplitudes from the detector have inherently greater timing jitter Consequently the time resolution with any detector can be improved by rejecting the lower amplitude pulses The best way to make this additional adjustment of the THRESHOLD is to monitor the counting rate at the Model 9307 TTL output on a counting ratemeter or counter timer Connect the TTL output of the Model 9307 to the input of the ratemeter or counter timer Note the counting rate produced by the setting in the previous paragraph Now raise the THRESHOLD to a higher value Record the THRESHOLD dial reading the counting rate and the time resolution in the time spectrum on the multichannel analyzer Fig 1 2 Repeat this process for a number of threshold setting and plot the results on graph paper From the graph choose a THRESHOLD setting that provides the best compromise between better time resolution and higher counting rate Lock the THRESHOLD dial at the desired setting 3 7 ADJUSTING THE SLEWING COMPENSATION Function and Effect When the SLEWING COMPENSATION is properly adjusted to match the characteristics of the signal from the detector and preamplifier there will be minimal systematic bias in marking the arrival time of smaller pulses versus larger pulses If the SLEWING COMPENSATION is adjusted too far in the negative voltage direction the lower
5. pulse rate when the reversed start stop logic is used as described in the next section 3 9 THE BENEFIT OF REVERSED START STOP LOGIC When operating with a pulsed excitation source that has a high repetition rate the efficiency of data collection can be improved drastically by using reversed start and stop logic Instead of starting the TAC with the signal from the pulsed source and stopping the TAC with the detected fluoresced photon the TAC is started with the fluoresced photon and stopped by a delayed source pulse This is the scheme actually depicted in Fig 1 2 signal from the pulsed excitation source is delayed by approximately 9096 of the TAC time range and used as the stop pulse The fluoresced photon detector provides the start pulse This results in a time spectrum that is reversed with shorter times to the right and longer times to the left If reversed start stop logic is NOT used the TAC spends most of its time converting start pulses without stop pulses This results in excessive dead time in the TAC and MCA The TAC and MCA spend most of the time processing signals with no recordable data The reversed start stop logic solves this problem The TAC only starts a conversion on the low rate pulses from the fluoresced photon detector For each of these start pulses a delayed stop pulse from the pulsed excitation source is guaranteed to occur With reversed start stop logic the busy time or dead time of the TA
6. C and ADC is determined by the collection of useful data at a counting rate set by the fluoresced photon detector In general itis best to use the signal with the higher counting rate as the stop input to the TAC 4 MAINTENANCE AND SERVICE No regular maintenance is required for the Model 9307 except to remove any collection of dust that might interfere with good electrical insulation and adequate heat dissipation In the unlikely event of instrument failure this unit can be returned to the ORTEC factory for service and repair at a nominal cost The ORTEC standard procedure for repair ensures the same quality control and checkout that are used for a new instrument Always contact Customer Service at ORTEC before sending an instrument for repair to obtain shipping required Return Authorization Number can be assigned to the unit The Return Authorization Number should be written on the address label the outside of the package and on the explanatory letter enclosed with the unit A multitude of packages flow into the Receiving Department at ORTEC each day Consequently the Return Authorization Number is critical to ensure that your package will be promptly routed to the Customer Service Department After inspecting your unit the Customer Service Department will contact you regarding the necessary repairs and the expected cost Table 1 Bin Module Connector Pin Assignments for Standard Nuclear Instrument Modules per DOE ER 0457T
7. Printed U S A Model 9307 pico TIMING Discriminator Operating and Service Manual ORTEC Part No 764020 Manual Revision C 1202 Advanced Measurement Technology Inc a k a ORTEC a subsidiary of AMETEK Inc WARRANTY ORTEC warrants that the items will be delivered free from defects in material or workmanship ORTEC makes no other warranties express or implied and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ORTEC s exclusive liability is limited to repairing or replacing at ORTEC s option items found by ORTEC to be defective in workmanship or materials within one year from the date of delivery ORTEC s liability on any claim of any kind including negligence loss or damages arising out of connected with or from the performance or breach thereof or from the manufacture sale delivery resale repair or use of any item or services covered by this agreement or purchase order shall in no case exceed the price allocable to the item or service furnished or any part thereof that gives rise to the claim In the event ORTEC fails to manufacture or deliver items called for in this agreement or purchase order ORTEC s exclusive liability and buyer s exclusive remedy shall be release of the buyer from the obligation to pay the purchase price In no event shall ORTEC be liable for special or consequential damages Quality Control Before being approved for shipment each ORTEC instrument m
8. ading edge of the output pulse at the input to the pico TIMING Discriminator The FAST NIM outputs are intended to drive a 50 Q cable terminated in 50 at the remote end of the cable Thus these two outputs produce pulses of nominally 800 mV amplitude on a terminated 50 Q cable The quiescent level between pulses is nominally 0 V Most TACs provide the 50 O termination internally at their inputs A 50 O cable with BNC connectors can be used to connect the FAST NIM outputs to either the START or STOP input on the TAC Normally the length of this cable is not critical When driven by a noise free pulser with a rise time 1 ns at the input of the Model 9307 the pico TIMING discriminator is capable of atime resolution 15 ps FWHM Full Width at Half Maximum peak height Consequently most of the contribution to the time resolution comes from timing jitter or noise in the detector connected to the Model 9307 input TTL Output The third output provides a positive TTL logic pulse with a width of approximately 500 ns This output is useful for driving slower instruments such as counters and ratemeters Either 50 O or 93 O cables with BNC connectors can be used with the TTL output For cable lengths 1 meter it is not necessary to terminate this cable in its characteristic impedance The TTL output can drive either TTL inputs or Slow Positive NIM Logic inputs The TTL output is triggered from the FAST NIM outputs and it flashes the adjace
9. amplitiude pulses will produce a timing output that arrives too early compared to the arrival of the timing output from the higher amplitude pulses If the SLEWING COMPENSATION is adjusted too far in the positive voltage direction the smaller pulses will generate a timing output that arrives too late compared to the arrival time of the timing output from the larger pulses Setting up for the Adjustment The easiest way to optimize the SLEWING COMPENSATION is to set up the complete timing system e g Fig 1 2 Choose a pulsed source of photons or ions that should generate a timing peak having a FWHM of lt 100 ps In other words the pulsed source of ions or photons must have negligible intrinsic time spread Determining the Initial Conditions Record the THRESHOLD setting that will be used for normal operation Next set the THRESHOLD control on the Model 9307 as low as possible without triggering on noise see section 3 6 Measure the SLEWING COMPENSATION setting at the front panel test points with a voltmeter The reading should be measured and recorded to the nearest Disconnect the voltmeter and record a time spectrum on the MCA with at least 10 000 counts in the peak height Observe the shape of the peak on a logarithmic vertical scale Note the width of the peak and the symmetry of the peak shape particularly near the base of the peak If the base of the time peak is skewed towards earlier time the SLEWING COMPENSATION shoul
10. ator Allow the instrument to dry completely before reconnecting it to the power source Pico TIMING DISCRIMINATOR THRESHOLD 2591 TO 1 X 5 m i 22 SLEWING COMPENSATION NI M E TTL Jt FAST NIM 6 RREAFP LEF ORTEC Model 9307 pico TIMING Discriminator 1 DESCRIPTION The ORTEC Model 9307 pico TIMING Discriminator defines the arrival time of analog pulses from ultra fast detectors with picosecond precision Moreover this superb performance is delivered over an extremely wide range of pulse heights with negligible influence of pulse amplitude on the timing output With the Model 9307 the difficult task of adjusting pulse shaping cables has been eliminated The internal pulse shaping in the pico TIMING Discriminator is optimum for single photon or single ion time measurements with microchannel plate detectors microchannel plate photmultiplier tubes and fast silicon photodiodes 10 Photons m 1 10 200 400 Fig 1 1 The Fluorescence Lifetime Instrument Response Function Recorded with a Model 9306 1 GHz Preamplifier and the Model 9307 pico TIMING Discriminator Time resolutions from 30 to 60 ps FWHM are possible with the system shown in Fig 1 2 The pico TIMING Discriminator accepts analog input pulses with amplitudes ranging from 50 mV to 5 V and pulse widths from 400 ps to 5 ns FWHM The amplitude threshold for
11. d be adjusted towards a more positive voltage If the base of the peak is skewed towards later time the SLEWING COMPENSATION should be adjusted towards a more negative voltage Making the Necessary Adjustments Make the adjustment of the SLEWING COMPENSATION indicated by the results in the previous paragraph Use the voltmeter to monitor the change Remove the voltmeter and record another time spectrum Observe the effect of the adjustment on the symmetry of the time peak at its base Make further adjustments in the same manner until satisfactory symmetry and minimum peak width and achieved Typically a setting of 12 10 mV will be measured when optimum SLEWING COMPENSATION has been achieved Asymmetry Intrinsic to the Signal Source Note that the detector or the photon source may have an inherent asymmetry that shows up in the time spectrum In such cases it will not be possible to achieve perfect peak symmetry with the SLEWING COMPENSATION see Fig 1 1 for example When adjusting the SLEWING COMPENSATION on the Model 9307 it will become obvious which asymmetries in the time peak are affected by the SLEWING COMPENSATION adjustment and which asymmetries are intrinsic to the detector and pulsed photon source Returning to the Operating Conditions Once the desired SLEWING COMPENSATION adjustment has been achieved record the final voltmeter reading for future reference and return the THRESHOLD adjustment to the desired operating s
12. d result in death or serious bodily harm if the safety instruction is not observed WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product In addition the following symbol may appear on the product ATTENTION Refer to Manual DANGER High Voltage Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product SAFETY WARNINGS AND CLEANING INSTRUCTIONS DANGER Opening the cover of this instrument is likely to expose dangerous voltages Disconnect the instrument from all voltage sources while it is being opened WARNING Using this instrument in a manner not specified by the manufacturer may impair the protection provided by the instrument Cleaning Instructions To clean the instrument exterior Unplug the instrument from the ac power supply Remove loose dust on the outside of the instrument with a lint free cloth Remove remaining dirt with a lint free cloth dampened in a general purpose detergent and water solution Do not use abrasive cleaners CAUTION To prevent moisture inside of the instrument during external cleaning use only enough liquid to dampen the cloth or applic
13. e section 3 7 This distortion is caused by saturation of the pulses at the preamplifier output Gradually reduce the pulse amplitudes until the distortion is eliminated 3 6 THRESHOLD ADJUSTMENT Function and Range The 10 turn locking dial on the front panel controls the amplitude threshold for accepting input pulses that are allowed to produce a timing output This threshold is adjustable over the nominal range of 25 mV to 1 0 V The dial reading is moderately accurate on pulse widths near the 5 ns FWHM limit On pulse widths approaching the 400 ps limit the actual threshold will be higher than the dial indication due to bandpass limitations in comparator circuits Initial Adjustment to Reject Noise THRESHOLD control is typically used to prevent the Model 9307 from triggering on low level noise from the signal source The simplest way to adjust this control is to turn off the supply of photons or ions to the detector while leaving the detector in its normal operating condition Subsequently the THRESHOLD dial is turned counter clockwise until the output LED glows brightly Under this condition the Model 9307 is triggering on noise from the detector and preamplifier Next the THRESHOLD dial is slowly rotated clockwise until the LED turns off Locking the dial at this point will set the threshold just above the detector noise Raising the Threshold for Improved Time Resolution When the source of photons or ions is turned on
14. e signal polarity This is the normal pulse polarity provided by the anode of a photomultiplier tube PMT a microchannel plate or a microchannel plate PMT For detectors which have a positive signal polarity a fast inverting transformer such as the Model IT 100 or the Model VT 120 Fast Timing Preamplifier can be used to invert the signal The Proper Input Pulse Width For best timing performance the pulse supplied to the input of the Model 9307 should have a width at half its amplitude that falls in the range of 400 ps to 5 ns This is the typical pulse width for single photon or timing with micorchannel plates microchannel plate PMTs and photmultiplier tubes The pico TIMING Discriminator can be used with fast scintillators on PMTs but the resulting time resolution will be 10 to 30 worse than can be obtained with a conventional Constant Fraction Discriminator Conversely the Model 9307 is the optimum solution for single photon or single ion timing 3 4 OUTPUT CONNECTIONS There are three outputs on the Model 9307 two fast negative NIM logic outputs and one TTL output FAST NIM Outputs The two FAST NIM outputs are intended for precise timing applications with a Time to Amplitude Converter TAC such as the ORTEC Models 457 566 or 567 see Fig 1 2 Each of these two outputs generates a current pulse which has a rise time of 400 ps a width of 2 5 ns and an amplitude of approximately 16 mA The le
15. e the efficiency for the longer time intervals will be much less than 100 This causes a distortion of the recorded time spectrum and leads to an error in the measurement of the fluorescence decay time constant The distortion of the time spectrum can be made negligible by limiting the counting rate of the detected fluoresced photons For example the optical intensity can be restricted so that the probability of detecting a single fluoresced photon after each laser pulse is 0 01 or lt 1 Under these circumstances the probability of observing 2 photons at the detector output for each laser pulse will be 0 01 which is 196 of the single photon rate This will render the distortion of the time spectrum insignificant To turn this guideline into a limit on the detected counting rate of photons it is sufficient to know the repetition rate of the pulsed laser at the sample For example a pulsed laser repetition rate of 1 MHZ requires an upper limit on the fluoresced photon counting rate of 10 KHz i e 0 01 x 1 MHZ to achieve less than 1 distortion of the time spectrum from the lost second stop pulses In general the limit on the detected photon rate to avoid significant spectrum distortion from lost second stop pulses with a pulsed excitation source is Max Counting Rate 0 01 x Source Repetition Rate Operating at counting rates below this limit will further reduce the distortion Note that this limit applies to the start
16. etting determined in Section 3 6 The Model 9307 is now ready for use 3 8 LIMITING THE COUNTING RATE TO AVOID SPECTRUM DISTORTION Typically the Model 9307 pico TIMING Discriminator will be used in conjunction with a TIME to Amplitude Converter TAC as illustrated in Fig 1 2 Time to Amplitude Converters enable the measurement of nanosecond time intervals with picosecond time resolution To achieve such high resolution one must accept a compromise because a Time to Amplitude Converter can only measure the time interval between pairs of start and stop pulses In other words the TAC measures the time interval from the first accepted start pulse to the next stop pulse It ignores all subsequent start pulses and any additional stop pulses until it has finished converting the first pair of start and stop pulses Consider the obviously logical arrangement where the laser pulse in Fig 1 2 is the start pulse for the TAC and the detected fluoresced photon generates the stop pulse The benefit of reversed start stop logic will be discussed in the next section If the probability of observing a fluoresced photon after each laser pulse is very high then fluoresced photons emitted late in the decay time scale will often find that the has already responded to a stop pulse from a photon emitted earlier in the decay process Consequently the efficiency for recording the shortest decay time intervals will be approximately 10096 whil
17. generating a timing output is adjustable from 25 mV to 1 V with a 10 turn locking dial Ultra fast circuits are incorporated in the pico TIMING Discriminator to minimize time slewing As aresult input amplitudes can vary over as much as a 100 1 range with negligible shift in the timing output This ensures excellenttime resolution even when the signal source produces a wide range of randomly varying signal amplitudes A front panel screwdriver adjustment permits fine tuning the slewing compensation to match the characteristics of a particular detector An adjacent test point makes it easy to monitor the adjustment with a voltmeter Two fast negative NIM outputs provide the flexibility to trigger a time to amplitude converter TAC while simultaneously driving other instruments The 500 ns wide TTL output can be used with instruments that require a positive logic signal such as counters and ratemeters A front panel LED flashes with each output pulse to indicate triggering For detectors having rise times less than 1 ns the ORTEC Model 9306 1 GHz Preamplifier should be used to amplify small signals before presentation to the input of the Model 9307 pico TIMING Discriminator For rise times gt 1 ns the Model VT120 Fast Timing Preamplifier should be substituted for the Model 9306 The Model 9307 incorporates a compatible 9 pin D connector on its rear panel to supply power to either preamplifier Microchannel Plate
18. he limit of the power supply or it indicates a possible fault in the power supply 3 3 INPUT CONNECTIONS For Detector Rise Times 1 ns The Model 9307 pico TIMING Discriminator has been designed for use with ultra fast detectors which have pulse rise times as short as 350 ps In order to obtain the best time resolution special care must be exercised to preserve the fast rise time Normally this requires the use of short 50 O coaxial cables with SMA connectors Typically a fast preamplifier such as the Model 9306 1 GHz Preamplifier is required between the detector and the input of the Model 9307 to increase the amplitude of the detector output pulses see Fig 1 2 The total length of the coaxial cables from the detector to the preamplifier and from the preamplifier to the pico TIMING Discriminator should be kept less than 1 7 metersto avoid degradation of the signal rise time The coaxial cable between the detector and the Model 9306 preamplifier also should be as short as possible to minimize electrical interference from neighboring instruments The inputs of both the Model 9306 and the Model 9307 include internal 51 resistors for proper termination of the 50 O cable For Detector rise Times gt 1 ns If a detector having an output rise time 1 ns is employed a longer length of 50 Q coaxial cable with BNC connectors can be used between the preamplifier and the Model 9307 See the Ordering Information at the end of Sectio
19. ircuits can exceed the maximum power available from the NIM bin power supply Before installing the modules in the NIM bin add the current requirements specified for each module and check that the total current required at each voltage does not exceed the maximum output of the power supply Also check the total power required against the maximum power specification for the power supply If the suite of modules demands excessive power some of the modules must be moved to a second NIM bin power supply CAUTION Do not insert the Model 9307 in a NIM bin without first turning off the NIM bin power Inserting the module in the NIM bin with power on may damage the ultra fast comparators in the Model 9307 The NIM bin power should be turned on only after the Model 9307 has been securely installed in the NIM bin Before installing the Model 9307 in the NIM bin turn off the NIM bin power then slide the 9307 into aconvenient slot and tighten the retaining screws at the top and bottom of the module The Model 9307 should be installed in the NIM bin at a location that is close to the module that utilizes the output of the Model 9307 After all the modules have been mounted in the NIM bin turn on the NIM bin power Using a voltmeter and the test points at the right side of the NIM bin check that the power supply voltages are operating at their correct values A sagging power voltage indicates that the current demand from the modules exceeds t
20. n 2 or consult the ORTEC catalog for the proper coaxial cables and the BNC to SMA connector adapters If the detector produces negative pulses with amplitudes that cover a substantial portion of the range from 50 mV to 5 V the detector can be connected directly to the input of the Model 9307 Note that the input of the Model 9307 is protected only to 5 V If the detector produces signal amplitudes beyond 5 V diodes will have to be installed at the signal source to limit the amplitude to within 5 V If signal amplification is required for a detector having a rise time gt 1 ns the Model VT120 Fast Timing Preamplifier should be used in place of the Model 9306 Preamplifier Power Connection Either the Model 9306 or the Model VT120 can derive its power from the preamplifier power plug on the Model 9307 rear panel Avoiding Ground Loop Noise To avoid noise induced by ground loops the detector and preamplifier should be isolated from any grounds in the vicinity of the detector The only ground supplied to the detector and preamplifier should come directly from the NIM bin housing the Model 9307 This can be accomplished by mounting the detector high voltage supply in the NIM bin next to the Model 9307 and by relying on the high voltage cable the preamplifier power cable and the coaxial signal cables to provide the ground to the preamplifier and detector Getting the Right Input Signal Polarity The Model 9307 input requires a negativ
21. ne tuning to minimize time slewing as a function of input pulse amplitude Adjustable over arange of approximately 30 mV A front panel test point located next to the potentiometer facilitates monitoring the actual setting Test point output impedance 100 Q OUTPUT LED Front panel LED flashes on each output pulse to indicate active triggering 2 3 INPUTS INPUT Front panel SMA connector accepts unipolar input signals with amplitudes in the range of 50 mV to 5 V Minimum input pulse width 400 ps FWHM Maximum input pulse width 5 ns FWHM Input impedance 50 Q The input is protected to 5 V 2 4 OUTPUTS FAST NEGATIVE NIM OUTPUTS Front panel BNC connectors provide two independent fast negative NIM output logic pulses Output amplitude is nominally 800 mV into a 50 Q load Pulse width is nominally 2 5 ns TTL OUTPUT Front panel BNC connector provides a positive TTL pulse triggered by the fast negative NIM output The 500 ns width of the TTL pulse is non updating Output impedance 1 short circuit protected PRAMP POWER Rear panel 9 pin D connector provides 12 V and 24 V power for the ORTEC Model 9306 1 GHz Preamplifier the Model VT 120 fast Timing Preamplifier or other preamplifiers utilizing the industry standard preamplifier power plug 2 5 ELECTRICAL AND MECHANICAL POWER REQUIRED The Model 9307 derives its power from a NIM bin power supply such as the ORTEC Model 4001 4002 Required dc
22. nt LED on every output pulse 3 5 ADJUSTING THE INPUT SIGNAL AMPLITUDE Large Pulse Amplitudes from a PMT Anode If the amplitude of the output signal from the anode of a photmultiplier tube is large enough to use directly with the input to the pico TIMING Discriminator the range of signal amplitudes can be selected by adjusting the high voltage applied to the photomultiplier tube An oscilloscope with a 500 MHZ bandwidth and a 50 Q input impedance is adequate to measure the PMT pulses The high voltage should be increased until the signals cover most of the 0 to 5 V operating range The PMT manufacturer s instructions regarding operating voltage and maximum anode signal should be consulted to select the optimum voltage range and to avoid distortion of the signals by anode saturation Do not exceed the 5 V limits on the input to the Model 9307 Amplified Pulses with gt 1 ns Rise Times For single photon counting with a photomultiplier tube a fast amplifier is typically needed to increase the amplitude of the pulses from the PMT anode An oscilloscope with a 500 MHZ bandwidth and a 50 Q input impedance can be used to observe the amplified pulses The easiest way to select the pulse amplitude range is to adjust the high voltage on the PMT Consult the PMT manufacturer s specifications for the allowable voltage on the photmultiplier tube The high voltage should be adjusted until the largest pulse amplitudes lie just within the linea
23. r range of the amplifier output If the linear range of the amplifier output is exceeded the pulses will be distorted and the time resolution will be compromised Do not exceed the 5 V limit of the Model 9307 input Amplified Pulses with 1 ns Rise Times For microchannel plate PMTs a fast preamplifier with an output rise time of about 350 ps is typically used to prepare the pulses for the Model 9307 input To observe these pulses an oscilloscope with a bandwidth 4 GHz and a 50 input impedance is required The simplest method for selecting the range of signal amplitudes is to adjust the high voltage applied to the microchannel plate PMT Consult the manufacturer s specifications regarding the allowable range of anode voltage If it is not possible to adjust the anode voltage high bandwidth attenuators such as the Hewlett Packard Models HP 8494B and HP 8495B can be inserted between the microchannelplate PMT output and the input to the fast preamplifier These attenuators can be adjusted to select the desired range of pulse amplitudes at the preamplifier output One of the above two methods should be used to adjust the gain so that the largest pulse amplitudes are just less than the linear limit of the preamplifier output Do not exceed the 5 V limit of the Model 9307 input If a 4 GHz oscilloscope is not available the pulse amplitude can be increased until a noticeable distortion of the peak in the time spectrum is observed se
24. rranty should follow the same procedure and ORTEC will provide a quotation Damage in Transit Shipments should be examined immediately upon receipt for evidence of external or concealed damage The carrier making delivery should be notified immediately of any such damage since the carrier is normally liable for damage in shipment Packing materials waybills and other such documentation should be preserved in order to establish claims After such notification to the carrier please notify ORTEC of the circumstances so that assistance can be provided in making damage claims and in providing replacement equipment if necessary Copyright 2002 Advanced Measurement Technology Inc rights reserved is a registered trademark of Advanced Measurement Technology Inc All other trademarks used herein are the property of their respective owners CONTENTS WARRANTY E EXE EE o E e EXE ii SAFETY INSTRUCTIONS AND SYMBOLS iv SAFETY WARNINGS AND CLEANING INSTRUCTIONS 1 DESCRIPTION d RO HERR CHER CREER 1 2 SPEGIFIGATIONS ita tati oat et 2 2 1 PERFORMANCE ee ss ee E ER C E E eee ene 2 2 2 CONTROLS AND INDICATORS 1 4 2 2 39 INPUTS CO
25. ust pass a stringent set of quality control tests designed to expose any flaws in materials or workmanship Permanent records of these tests are maintained for use in warranty repair and as a source of statistical information for design improvements Repair Service If it becomes necessary to return this instrument for repair it is essential that Customer Services be contacted in advance of its return so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause unnecessary delays in getting the unit repaired The ORTEC standard procedure requires that instruments returned for repair pass the same quality control tests that are used for new production instruments Instruments that are returned should be packed so that they will withstand normal transit handling and must be shipped PREPAID via Air Parcel Post or United Parcel Service to the designated ORTEC repair center The address label and the package should include the Return Authorization Number assigned Instruments being returned that are damaged in transit due to inadequate packing will be repaired at the sender s expense and it will be the sender s responsibility to make claim with the shipper Instruments not in wa
26. voltages and currents are 12 V and 35 mA 6 V at 70 mA 6 V at 360 mA and 12 V at 100 mA WEIGHT Net 0 9 kg 2 0 Ib Shipping 1 8 kg 4 0 Ib DIMENSIONS NIM standard single width module 3 43 X 22 13 cm 1 35 X 8 714 in Front panel per DOE ER 0457T 2 6 ORDERING INFORMATION To order specify Mode Description 000002 9307 pico TIMING Discriminator SMA RG 58U 50 Coaxial Cable with SMA 58 0 15 Connectors 0 15 m length SMA RG 58U 50 Coaxial Cable with SMA 58 0 5 Connectors 0 5 m length RG 58U 50 Q Coaxial Cable with SMA Connectors 1 5 m length SMA BNC SMA to BNC Adapter with male SMA and female BNC BNC SMA BNC to SMA Adapter with male BNC and female SMA 3 INSTALLATION AND OPERATION 3 1 GENERAL The Model 9307 power requirements must be furnished from a NIM standard bin power supply that includes 6 V power distribution such as the 4001A 4002D 4001C 4002D or 4001C 4002E NIM Bins Power Supplies The NIM bin power supply in which the 9307 normally will be operated is designed for relay rack mounting If the equipment is rack mounted be sure that there is adequate ventilation to prevent excessive localized heating in the Model 9307 The temperature equipment mounted in racks can easily exceed the maximum limit of 50 C unless precautions are taken 3 2 CONNECTION TO POWER The power demanded by a NIM bin full of NIM modules that incorporate high density electronic c