Home

image-vet 70 plus - at www.ImageWorksCorporation.com.

image

Contents

1. and covers pins 1 amp 2 J1 location determines whether the Hand Held device is used or whether an external device is used J1 is located on the opposite side to the Power Switch Setting the System in Remote Control Mode Move J1 to the right so it covers the left and middle pins orient the control box so you would be able to read the label next to the Power Switch Machine Just Clicks but doesn t operate Check that K2 on the PC board is in place located on the upper right corner when you orient the control box so you are able to read the label next to the Power Switch Check that the connections on the hand held control are in place as well For duel tube head operations place a jumper between 14 amp 17 on bar M2 1 The one closest to the Power Switch In addition connect Number 13 on bar M2 1 to tube head B and to wall mount Number 11 on bar M2 1 to tube head B and to wall mount Number 10 on bar M2 1 to tube head A and to wall mount The micro switches provided are to be connected as follows Tube A is connected between terminals 8 amp 9 on the M1 bar in the control box and the lower terminal 8 amp 9 on the steel mounting plate Tube B is connected between terminals 8 amp 10 on the M1 bar in the control box and the lower terminal 8 amp 9 on the steel mounting plate
2. IMAGE VET 70 PLUS Troubleshooting Reference Trained Service Personal ONLY Observe ALL safety Procedures in the service manual Reference Installation Operation Service amp parts Manual P N 897 000002 Table of Contents 1 Wiring Connections 2 No X Ray or Light X Ray__ I 3 Error Codes LL A3 4 Miscellaneous o S 4 Key Greater Than Lower Than Equal To About the Same DI Wiring Connections Three wiring are coming out of the x ray tube and are connected as follows a The black wire L is connected to number 10 on the bar closest to the switch b The white wire N is connected to number 16 on the bar closest to the switch c Green yellow is connected to ground metal screw second on the left from the power switch d Place a jumper between 8 amp 9 on the bar located further away from the switch on the other side of the board e For Hand Held usage make sure the jumper J1 next to the above bar is covering middle and right pins f For Remote Expose Switch make sure the jumper J1 next to the above bar is covering middle and left pins II Switch Connections Position the control box so you can read the label next to the power switch a The upper left and upper right metal stripes of the power switch are connected to number 5 amp 6 on the bar closest to the switch respectively b The lower left and lower right metal stripes of the power switch are connected
3. eplace the Arm Replace the Tube P N 9992700105 Folding Arm P N 9992700130 115V P N 9992700108 Standard Extension Arm 35 4 P N 9992700330 230V g E010 Check if wiring is accurate Wiring Good Wiring Bed Make sure the J6 on the control Fix connection according box is open i e it doesn t cover to the instruction both pins cover only the left one J6 is located on the left edge of the PC Board when facing the Power switch J6 Covers One Pin J6 Covers both Pins Check that the voltage between 10 amp Remove it and place it so 16 on the bar closest to the Power it covers only the left pin switch is 115V during an exposure and that the resistance is between 1 2 1 6Q Shut down the machine before checking the resistance Voltage amp Resistance Good Voltage Good Resistance Bad No Voltage Machine Check the resistance of the tube head Replace PC Board should work Should be between 1 2 1 6Q P N 897 ZZ1 145021 115V P N 897 ZZ 1145051 230V Resistance Good Resistance Bad Replace the Arm Replace the Tube P N 9992700105 Folding Arm P N 9992700130 115V P N 9992700108 Standard Extension Arm 35 4 P N 9992700330 230V 4 Miscellaneous a b c d e No Ready Light Make sure that the jumper between 8 amp 9 on the bar furthest from the Power Switch is in place Machine Doesn t Work but Everything Else is Fine Check that J1 is to the right
4. is good Try to use another jumper and if it doesn t work call electrician The frequency problem is coming from the wall If the frequency coming out from the wall is good replace the board P N 897ZZ1145021 115V 897 ZZ1145051 230V c E003 Turn the machine off and then on If the problem persists replace the PC Board d E004 E005 The operational voltage range is 108 V 126 If the error message is E004 E005 than the voltage supply is above or below this range respectively Call an electrician to fix the problem e E006 Two or more X ray tubes may have been pulled out Make sure that all tube heads besides one are in Position Equipment selector switch may be faulty Check wiring f E009 Check Fuse F1 located in a black Replace fuse F1 tube on the PC Fuse Bad P N 897FU0001013 Board Turn Off the machine and then turn it On That will enable you to take one additional exposure Check that the voltage between 10 amp Fuse Goody 16 on the bar closest to the Power switch during an exposure is about line voltage 115V Check the resistance between 10 amp 16 on the bar closest to the Power switch when the machine is shut down 21 20 1 60 oltage amp Resistan ood oltae ood R nce Bad No Voltag Machine Check the resistance of the tube head Replace PC Board should work Should be between 1 2Q 1 6Q P N 897 ZZ1145021 115V P N 897 ZZ1145051 230V Resistance Good Resistance Bad R
5. to number 4 amp 7 on the bar closest to the switch respectively No X Ray Make sure that the wire connections above are accurate and that you have power from the wall 2 No X Ray or Light X Ray Connections are accurate Connections are not accurate Check Fuse F1 on the PC Board Located in a black cylinder that has to be rotated to be opened Fuse is Good Fuse is Bad Check the voltage between 10 amp 16 on the Replace Fuse F1 bar closest to the Power Switch during an P N 897 FU0001007 Vet exposure should be line voltage Voltage No Voltage Turn down the machine and check Replace Board the resistance between 10 amp 16 on P N 897 771145021 115V the bar closest to the Power Switch P N 897 771145051 230V should be between 1 2 1 6 2 Resistance Good Resistance Bad oo Machine should work Check the resistance of the tube head should be between 1 2 1 6 Q Resistance Good Resistance Bad Replace the Tube Head P N 9992700130 115V Replace the Arm P N 9992700105 Folding Arm 3 Error Codes a E001 x ray exposure button was released too fast b E002 E002 Frequency is incorrect Turn the machine off and remove J3 from the PC board it is located on the left edge when facing the power switch on the control box make sure it is clean and put it back in place covers both pins Turn the machine on Problem Solved Problem wasn t solved Machine