StatSpin Express 4
Contents
1. and Closing the Cover An electronic operated cover interlock mechanism prevents operation if the cover is not completely closed and locked and prevents the cover from opening when the rotor is turning An operating cycle can be initiated only if the cover latch is engaged Cover Interlock By pass In case of power failure or malfunction the Stop Open button may not release the cover To recover samples in the rotor the cover interlock mechanism can be released manually by inserting the manual latch release tool supplied with the instrument into the small hole located in the front of the housing Push the tool inward about one inch 25 mm to release the latch CAUTION The cover interlock by pass is for emergency use only Misuse may expose the operator to unsafe conditions Removing and Installing the Rotor The StatSpin Express 4 is shipped with the rotor installed and supported by a shipping retainer To prevent possible damage to the centrifuge s motor mount it is important to remove the rotor whenever the centrifuge is moved or shipped to another location CAUTION The rotor RTH8 has a finite lifespan of 50 000 cycles Discontinue use of rotor after this period Replace with new rotor for continued use of centrifuge Before operating the centrifuge ensure that the rotor is fully seated onto the motor shaft and the hex screw is fully tightened To remove the rotor 1 Open cover and remove shield Grasp shield at the center o
2. key Install o Place the centrifuge on a solid level surface suitable for operation of laboratory instrumentation o Position the centrifuge out of direct sunlight and away from sources of intense heat or cold Improper conditions can cause damage to the housing rendering unit unsafe to operate o Seethe Appendix for the acceptable range of operating temperature and humidity o Remove the foam shipping retainer around the rotor before operating the centrifuge WARNING Maintain a 300 mm 11 8 inches clearance around the centrifuge Do not deposit any potentially hazardous materials within the CLEARANCE ENVELOPE Connect Power o Plug the centrifuge into a grounded outlet supplying 100 240 V 50 60 Hz o Turnon the power switch located on the rear panel above the line cord connector o When the instrument completes the power on check it will give a two beep ready signal The three minute green indicator will illuminate Press the stop button to release the latch and open cover WARNING Do not use the supplied line cord outside of North America Use a power cord with a grounded male connector suitable for the power outlet being used which is rated at 3 0 Amps or more and has an IEC320 CEE22 female connector to attach to the centrifuge Principle and Intended Use IVD tubes Section 2 System Overview For in vitro diagnostic use to produce the rapid separation of whole blood contained in original collection The
3. provided with eight tube inserts for 13 x 100 mm tubes Product No SV09 and eight tube inserts for 16 x 100 mm tubes Product No SV07 Other sizes are available See accessories The rotor must be properly balanced to ensure smooth operation DO NOT spin a single tube Use the same size and style tubes in opposite positions Balance liquid in tubes to within 0 5 mL Close and latch the centrifuge cover Select the spin time by pressing the appropriate cycle time button Press Start button CAUTION Replace Tube Inserts immediately if a tube breaks during centrifugation CAUTION Heplace glass tubes used for rotor balancing after 10 spin cycles to reduce chances of breakage due to the repeated stress of centrifugation Cycle Settings The operator should select spin time settings to achieve optimum results for specific applications The following are general guidelines Setting Speed RCF 3 min Produces high quality plasma serum from whole 5100 rpm 4000 xg blood for chemistry and cardiac testing 5 min Produces high quality plasma serum from whole 5100 rpm 4000 xg blood for chemistry coagulation and cardiac testing Recommended for most Greiner gel tubes 10 min Slower traditional spin 4500 rpm 3100 xg Centrifugal forces are decreased to provide a traditional spin Unloading 1 Upon completion of the cycle the rotor decelerates to a complete stop in 30 seconds or less a signal beep will sound and the cover lat
4. ONLY 3 MINUTES The StatSpin Express 4 high speed horizontal centrifuge provides faster turnaround time and greater productivity for your lean laboratory Place it next to any STAT analyzer and serum or plasma gel tubes are ready in just minutes With only 3 settings it is simple to use Just select the 3 5 or 10 minute button and press start No fumbling with settings or the need to readjust the speed For safety the lid remains locked until the rotor comes to a complete stop An audible alarm sounds when the cycle is completed Samples can be processed as they arrive in the lab no need for batching no unproductive time spent waiting for the centrifuge you ll have results in minutes Now you can process as many as 8 samples at once for higher throughput Horizontal spin provides a flat gel barrier Stat 5pin Express 4 HIGH SPEED HORIZONTAL CENTRIFUGE ACHIEVE GEL SEPARATION IN Gel tubes are separated in just 3 minutes StatSpin Express 4 is Ideal for e Rapid separation of gel tubes e Flat gel barrier e Sample prep for STAT chemistry e Single piece flow for rapid sample processing e Anywhere rapid serum or plasma is needed High Quality Gel Separation In just 3 minutes the Express 4 can produce serum or plasma equal to tubes spun for 10 or more minutes in conventional horizontal clinical centrifuges C Iris Sample Processing StatSpin Express 4 Modern Fa
5. StatSpin Express 4 is a high speed horizontal bench top centrifuge used to rapidly separate blood components in the original sample collection tubes Do Not centrifuge substances with a density greater than 1 2 kg dm The horizontal rotor included with the centrifuge accepts up to 8 primary 16x100mm primary blood tubes with a maximum volume of 10 mL per tube The operator selects a fixed spin time of 3 5 or 10 minutes The rotation speed is 5100 rpm for the 3 and 5 minute cycles producing a centrifugal force of 4000 xg at a rotor radius of 13 8 cm The 10 minute cycle generates 3100 xg at 4500 rpm This setting provides lower g forces for procedures which require a traditional spin Symbols and Definitions IV HEP B B ePi CE cUL UL gt Product Reference Number Caution Temperature limitation EC Representative For in vitro diagnostic use Non sterile Serial Number Consult Instructions European Community Agency Fuse AC Indicates the Iris Sample Processing product catalog number Statement of caution warning Read carefully Indicates storage or operating temperature restrictions European Community Authorized Representative Indicates for use only as in vitro diagnostic device Indicates a non sterile product Indicates instrument serial number Refers operator to the instruction manual for additional information Compliance to IVD Directive Listed device with Underwriter
6. ally in plasma no more than inch from top surface of plasma and away from the tube walls Centrifuge not providing platelet poor plasma Contact customer service for guidelines for obtaining platelet poor plasma Ensure tubes are properly mixed immediately following collection Refer to tube manufacturer for recommended procedure mar Pie E MOSE PATAN REGAR Check rotor speed and cycle times Allow serum tubes to clot prior to centrifugation E E E Replace tube insert Do not attempt to clean glass from inserts g g J Balance tubes should be replaced after 10 cycles Confirm unit has power press the Stop Open button Cover does not open In the event of a power failure or malfunction insert latch release tool into the small hole on front panel Cover will release and samples can be removed 3 min 5 min 10 min flashing unit will not spin Confirm shield is securely locked in place counter clockwise Service Refer all service to Iris Sample Processing or an Authorized Repair Facility Instrument fuse s are located on the rear panel Remove the line cord to access the fuse holder Replace with fuses of the same type and rating 1 asacsove 5x20mm Be sure to complete the Warranty Registration Card as directed Decontamination before returning for service Any instrument or accessory that has been exposed to blood or other biological materials must be cleaned prior to shipment to the manufacturer or distri
7. butor for service This decontamination is required by Federal Law Title 48 and 49 of the Federal Regulations and in accordance with the Environmental Protection Agency s Regulations for Biohazard Waste Management Iris Sample Processing personnel cannot perform this decontamination Refer to maintenance section for cleaning instructions 10 Section 5 Appendix Specifications Product Number SSH4 Model Number M510 Spin Parameters 3 min 180 seconds 5100 rpm 4000 xg 5 min 300 seconds 5100 rpm 4000 xg 10 min 600 seconds 4500 rpm 3100 xg Acceleration Time lt 30 seconds Deceleration Time lt 30 seconds Electrical 100 240 VAC 50 60Hz 2 5 Amps Dimensions Height 8 0 20 3 cm Width 13 0 33 0 cm Depth 16 0 40 6 cm Weight 20 Ibs 9 Kg Environmental For indoor use only Ambient temperature 2 C to 40 C Maximum 80 RH between 2 C and 30 C decreasing to 50 RH at 40 C Maximum altitude 2000 m Main supply voltage fluctuations should not exceed 10 Transient over voltages according to installation category II Pollution degree 2 Specifications are subject to change without notice WARNING Do not operate or store the centrifuge or its rotor below the 2 C minimum operating temperature At extreme low temperature below 18 C the rotor material is subject to cracking under impact stress such as dropping onto a hard surface 11 References 1 Iris Sample Processing A Divi
8. ch automatically releases 2 Lift the cover and carefully remove the tubes Section 4 Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenance on all instruments Contact customer service or your distributor if the instrument is not functioning properly CAUTION Unplug the StatSpin Express 4 from the wall outlet before cleaning or performing maintenance WARNING Inspect instrument for cracks or any physical damage to housing cover and rotor Damage may result in unsafe operation discontinue use until repairs have been performed WARNING Do not expose the StatSpin Express 4 or its rotor to strong or concentrated acids bases or oxidizing agents aromatic or halogenated hydrocarbons esters or ketones Cleaning The StatSpin Express 4 is equipped with a stainless steel bowl for easy cleaning Clean the outside surfaces and switch overlay panel with a water dampened cloth and mild detergent Clean the inner surface or bowl with a mild detergent and if necessary a disinfectant wiping surfaces with a dampened cloth using 7096 alcohol or 1096 bleach solution Do not sterilize rotor CAUTION Do not spray cleaning solutions directly onto the centrifuge bowl or housing overspray could reach the motor bearings or internal circuitry Subsequent damage may not be covered by warranty Before using any cleaning or decontamination methods except t
9. cit teen el edes e ota uoc oass tectae ented a 7 DECIO MM eee eee ees tees Renee eles 8 MaInt NaNnCe ET tS t tmm 8 OVER SW ote destine a opem s ie p M P E eT Mp EU D EUR 8 RNS ANN ls RE ERE 8 Checking the ROTO SO CCC a ir eter ttai tet aet o pU rmt do hot eren 8 TTOBDIGSITOODFIC acci 2 5 ales E bet sette od agate tte eis ur LLL 9 lt hhc tse M ML DNI EMI AU IU EEUU E 10 fedes Tc ane eed anaes eee 11 721 0 01 1 0 dE RR cR 11 SPECIGI airar ea nee ae tame duct Ai M NL aet LL aes mete 11 B21 KS 5 9 O ee one ee ne TTL 12 MLN I SPERO E E 13 This manual along with information contained on product labels and in package inserts should provide you with all the information you need to operate and maintain the StatSpin Express 4 When the information requires special attention a caution symbol N appears next to the italicized text Please pay close attention to the instructions that accompany the notes and symbols as well as the standard laboratory practices outlined by your facility and local regulatory agencies The table below lists all the CAUTIONS and WARNINGS for the StatSpin Express 4 WARNING Do not expose the centrifuge or its rotor to strong or concentrated acids bases or oxidizing agents aromatic or halogenated hydrocarbons esters or ketones WARNING Do not operate or store centrifuge or its rotor below the 2 C minimum operating temperature At extreme low temperature below 18 C the rotor material
10. ential or incidental damages resulting from the sale or use of the products The purchaser shall be deemed liable for any and all claims losses or damages incurred by the use or misuse of the Iris Sample Processing instrument by the purchaser its employees or others following receipt of the instrument or other items 13 EXPRESS 4 CENTRIFUGE MODEL M510 REPLACEMENT FEET 01 005838 001 Iri S 54 005875 001
11. flashing unit will not spin Refer to Error Code Table of manual to identify and correct problem T Be sure the correct tube inserts are in use Check insert listing Tubes are loose and do not fit into buckets l l available under accessories to match with appropriate tube sizes Main power switch is located on rear panel along with fuse holder Check switch position and if necessary condition of fuses No LEDs are on and unit will not spin Confirm line cord is plugged into a powered outlet This instrument detected an imbalance and stopped operation Clear Unit vibrated and shuts down l the error via keypad and check tube balance There is a crack in the top cover or housing Discontinue operation Have unit serviced Remove screw and apply straight upward pressure at rotor center Do Rotor will not come off motor shaft l l l not bend wires Contact Customer Service for additional information Your rotor has reached 50 000 cycles and requires replacement Red Error LED is on and will not turn off Discontinue operation Please contact distributor or customer service for assistance with rotor purchase E E EE E E Be sure the correct tube inserts are in use Check insert listing p popping 9 g available under accessories to match with appropriate tube sizes Check rotor speed and ensure tubes are balanced Check to make sure you are using the 5 minute setting Ensure probe of the analyzer is positioned vertic
12. hose recommended operators should check with the manufacturer that the proposed method will not damage the equipment General Cleaning Open centrifuge cover Unplug centrifuge from the electrical outlet Remove the shield Remove the rotor see Removing and Installing the Rotor Clean bowl area with mild detergent Re install rotor see Removing and Installing the Rotor Plug into electrical outlet Pee ol I Rotor Disinfecting Open centrifuge cover Unplug from the electrical outlet Remove shield to allow access to rotor Remove the rotor The rotor and inserts may be soaked in detergent and warm water or a 10 bleach solution DO NOT attempt to clean an insert that has a broken tube Dispose of insert immediately Dry the rotor and inserts with a clean absorbent paper towel or allow to air dry Re install the rotor see Removing and Installing the Rotor Plug into electrical outlet ee Ol OU s Checking the Rotor Speed The rated speeds 5 can be checked with a stroboscope or photoelectric tachometer Point tachometer thru transparent cover at reflective patch located near rotor center Do not defeat any safety interlocks while performing test If the centrifuge fails to achieve an operating speed of 5100 rpm 5 in the 3 min and 5 min settings or 4500 rpm 5 in the 10 min setting contact your distributor or the Iris Sample Processing customer service department Troubleshooting Red Error Service indicator
13. is subject to cracking under impact stress such as dropping onto a hard surface WARNING Do not use the supplied line cord outside of North America Use a power cord with a grounded male connector suitable for the power outlet being used which is rated at 3 0 Amps or more and has an IEC320 CEE22 female connector to attach to the centrifuge CAUTION Maintain a 300 mm 11 8 inches clearance envelope around centrifuge during operation Do not deposit any potentially hazardous materials within the CLEARANCE ENVELOPE CAUTION Never operate the centrifuge without the rotor locked securely in place Failure to properly install and secure the rotor may result in damage to the centrifuge not covered by warranty WARNING Inspect instrument for cracks or any physical damage to housing cover and rotor Damage may result in unsafe operation discontinue use until repairs have been performed CAUTION Replace glass tubes used for rotor balancing after 10 spin cycles to reduce chances of breakage due to the repeated stress of centrifugation Replace Tube Inserts immediately if a tube breaks during centrifugation CAUTION The rotor RTH8 has a finite lifespan of 50 000 cycles Discontinue use after this period CAUTION The cover interlock by pass is for emergency use only Misuse may expose the operator to unsafe conditions CAUTION Do not spray cleaning solutions directly onto the centrifuge bowl or housing overspray could reach
14. pening and rotate clockwise Simply lift straight up and out of centrifuge housing Locate 1 8 inch hex key supplied and remove center screw of rotor hub 3 Remove rotor lifting straight up and out of centrifuge To re install rotor Note that the motor shaft has a D shape at the tip This mates with the rotor and must be aligned for the rotor to fully seat onto the motor shaft Do not force rotors onto shaft Damage could result to motor and affect warranty 1 Align D shape on shaft with rotor hub 2 Slide the rotor onto the shaft and push down gently to seat 3 4 Reinstall center screw and tighten with supplied hex key Reinstall shield into housing Align shield hooks with cut outs in housing Grasp shield at center and rotate counter clockwise If shield is not locked into place instrument will not operate Spinning the Sample A CAUTION Never operate the StatSpin Express 4 without the rotor and shield securely in place Do not attempt to by pass any interlock switches located around bowl and shield area Failure to properly install and secure the rotor may result in damage to the centrifuge not covered by warranty CAUTION Maintain a 300 mm 11 8 inches clearance envelope around centrifuge during operation Do not deposit any potentially hazardous materials within the CLEARANCE ENVELOPE Loading 1 ee dl euo qe Ensure the appropriate tube inserts are installed The StatSpin Express 4 centrifuge is
15. rd Iris Sample Processing will at its discretion repair or replace any unit or part covered under this warranty returned to Iris Sample Processing with shipping costs prepaid Repaired or replaced instruments supplied under this warranty carry only the remaining portion of the original warranty and repairs shall not interrupt or prolong this warranty For warranty terms and conditions outside the United States contact your Authorized lris Sample Processing Distributor No warranty extended by Iris Sample Processing shall apply to any instrument that has been damaged due to misuse negligence accident or damage resulting from unauthorized repairs alterations or improper installation Iris Sample Processing makes no warranty other than the one set forth herein This warranty is given expressly in lieu of all other warranties expressed or implied The purchaser agrees that there is no warranty of merchantability or of fitness for any intended purpose and that there are no other remedies or warranties expressed or implied which extend beyond the description on the face of the agreement No agent or employee of Iris Sample Processing is authorized to extend any other warranty or assume for Iris Sample Processing any liability except as set forth above This warranty is only applicable to the original purchaser Limitation of Liability Iris Sample Processing shall not be liable for any loss of use revenue or anticipated profits or for any consequ
16. s Laboratory Rating 2 5A 250V F Single Phase Alternating Current Operator Controls Start Stop Open 3 Minute Cycle 5 Minute Cycle 10 Minute Cycle iJ i el Error Service Q e Error Indicators The Start button initiates the selected cycle The Stop Open button interrupts the cycle and releases the cover when the rotor has stopped The 3 min button selects a cycle of three minutes at 5100 rpm The 5 min button selects a cycle of five minutes at 5100 rpm The 10 min button selects a cycle of ten minutes at 4500 rpm The red indicator with an adjacent wrench symbol flashes to signal an error condition or remains continuously illuminated once the rotor reaches 50 000 cycles StatSpin Express 4 error conditions are signaled by a flashing red Error Service indicator on the control panel The cause of the error condition can be identified by the combination of green cycle time indicators that are also flashing See Error Codes Table To clear the error condition correct the problem and press the Stop Open button Error Service light flashing Error Codes Table Continuously illuminated O Error Service indicator Accessories Product No SV02 SV05 SV06 SV07 SV08 SV09 SV10 SV11 RTH8 Failure to reach speed in 30 seconds Confirm no obstruction in bowl area contact customer service Latch is not locked Confirm cover is fully seated in closed position Rotor exceeded
17. s Manual StatSpin Express 4 Horizontal Centrifuge Model Number M510 For in vitro diagnostic use Product Number SSH4 StatSpin Express 4 for 100 240 VAC 50 60 Hz StatSpin and Iris are registered trademarks of IRIS International Inc Copyright 2008 Printed in U S A Table of Contents HOW fo se this manual icen ssp csreieescersesetcenenescaszeenceecessceeateessteaceconegaee ii SOCOM e Metern 1 Unpacking and Installation usse 1 IIS DSC ASK AGMAG RR NR E OM 1 MES oc ERU T TU 1 luci PO OSSD OW Ci S uut ER IM IEEE LUE 1 dol pa 2 SYSTEM OVER VICW cte 2 PRNGIDIC and Interided USB soqenii ectetur a eter tub m Deest 2 SYMDOIS and DeHnillohs 333 m nee pa e Rata ette a sperata 2 Operator Oo aiio RE NEST 3 EEFODIDIdICALOES s eek ed D IUS SUETON DSL D cd eee TII LCD M eect DITE 3 Eror codes EaD IGu eniti rox uta lead ao tiep den a etas eas dq pu eme EE 4 ACCS SONOS eu ene eee eee ee a VO VM EA VI ee ee vU MEUM CEN 4 StatSpin Express 4 Tube Insert Guide Lseesseeeeeeuueese 5 SECUOM qc P EEURLMUM MEMMMMBBSMM 6 Operating Instr cliohs iode ht rano deed area oneni neresen iebsa oec 6 Opening and Closing the Cover sssssssssR es 6 Removing and Installing the Rotor sssssssse 6 Spinning the Sample eie
18. set speed Remove and inspect rotor reinstall Cover is not closed Confirm cover is fully seated in closed position Out of balance Confirm tubes are balanced in rotor and centrifuge is on a level surface Rotor cannot maintain speed Contact customer service Shield is not installed Confirm shield is installed and in the locked position Cycle Counter 50 000 spins have been reached contact distributor or customer service to purchase replacement rotor Q Description Inserts for 3 mL tubes 10 25 x 64 mm 4 pk Inserts for 2 mL tubes 10 25 x 47 mm or BD Microtainers 4 pk Inserts for 7 mL tubes 16 x 75 mm 8 pk Inserts for 10 mL tubes 16 x 100 mm 8 pk Included with centrifuge Inserts for 5 mL BD tubes 13 x 75 mm 8 pk Inserts for 7 mL tubes 13 x 100 mm 8 pk Included with centrifuge Inserts for 5 mL Greiner tubes 13 x 75 mm 8 pk Inserts for 4 mL BD RST tubes 13 x 100 mm 8 pk 8 place horizontal rotor Included with centrifuge StatSpin Express 4 Tube Insert Guide SV02 SV05 SVO6 SV07 10 25 x 64mm 1 5 2mL microtubes 16 x Thmm 16 x 100mm 28 x 47mm and BD Microtainera LSVOS iube insects replace SYOJ and SVOM iube insecis 2 SY07 aud S YOO tube iosecis included with Express 4 J SY 10 and SV 11 ace new tube iasecis foc Exprezz 4 svog 13 x 100mm 10 13 x 74mm Greiner tubes 3 SV11 13 x 100mm BD RST tubes Section 3 Operating Instructions Opening
19. sion of IRIS International Inc Westwood MA 02090 USA Phone 800 782 8774 or 781 551 0100 Fax 781 551 0036 www proiris com ECT REP mdi Europa GmbH Wittekamp 30 D 30163 Hanover Germany 49 511 39089530 Made in USA CLSI Protection of Laboratory Workers from Occupationally Acquired Infections Approved Guideline Second Edition CLSI document M29 A2 ISBN 1 56238 453 8 CLSI 940 West Valley Rd Suite 1400 Wayne Pennsylvania 19087 1898 USA 2001 CDC Recommendations for Prevention of HIV Transmission in Health Care Settings MMWR Suppl No 2S 2S 18S 1987 CDC Updated US Public Health Service Guidelines for the Management of Occupational Exposures to HBV HCV and HIV and Recommendations for Post Exposure Prophylaxis Appendix A and B MMWR 50 RR 11 1 42 June 29 2001 55 005712 001 Rev F 12 Iris Sample Processing Warranty Iris Sample Processing a Division of IRIS International Inc warrants that the centrifuge shall be free from defects in material and or workmanship under normal use and service for the period expiring twenty four 24 months from the date of installation Accessories including rotor shall be free from defects in material and or workmanship under normal use and service for the period expiring twenty four 24 months from the date of installation or 50 000 cycles whichever occurs first The purchaser must have completed and forwarded to Iris Sample Processing the Warranty Registration Ca
20. st Powerful The StatSpin Express 4 is UL and CE marked Universal power input works off any AC electrical power from 100 240V Suitable for worldwide locations The StatSpin Express 4 is powerful The rugged brushless motor is maintenance free StatSpin provides a comprehensive 2 year warranty Adaptors for 7 and 10 mL tubes come with the unit It can accomodate tubes from 1 5mL 10mL Adaptors sold separately System Specifications Dimensions i 8 0 inches 20 3 cm 13 0 inches 33 0 cm 16 0 inches 40 6 cm 22 0 Ibs 10 0 kg 3 minutes 5100 rpm 4000 xg Cycle Times Speeds 5 minutes 5100 rpm 4000 xg 10 minutes 4500 rpm 3100 xg Tube Sizes Accepts blood tubes 1 5 10mL Power Requirements 100 to 240V 50 60 Hz 2 5 amps Specifications subject to change without notice Product Description StatSpin Express 4 for 100 240V 50 60 Hz SSH4 Includes 8 place rotor eight 7 mL inserts eight 10 mL inserts operator s manual manual latch release tool one hex key tool and power cord With 3 settings the unit is simple to use Iris Sample Processing 60 Glacier Drive Westwood MA 02090 U Tel 781 551 0100 Fax 781 551 0036 C L US C 800 782 8774 A Division of IRIS Intemational Inc WWwW proiris c om 64 005754 001 Rev B StatSpin Express 4 Horizontal Centrifuge Operator s Manual C i s Iris Sample Processing A Division of IRIS International Inc Operator
21. the motor bearings or internal circuitry Subsequent damage may not be covered by warranty Before using any cleaning or decontamination methods except those recommended operators should check with the manufacturer that the proposed method will not damage the equipment CAUTION Unplug the centrifuge from the wall outlet before cleaning or performing maintenance CAUTION Follow Universal Precautions with all biological specimens regardless of whether the specimen is known to contain an infectious agent See references Please use the system as intended Improper use of the StatSpin Express 4 Horizontal Centrifuge and or its accessories may cause damage to the system inaccurate results or potentially nullify warranties Section 4 Unpacking and Installation Inspect Packaging The StatSpin Express 4 and its accessories are delivered in a single carton If the centrifuge or accessories have been damaged in transport please inform your carrier immediately NOTE Save the shipping carton and packing materials to facilitate return should service be required Verify Contents The package contains One StatSpin Express 4 horizontal centrifuge Product No SSH4 One centrifuge rotor Product No RTH8 Eight rotor inserts for 13x100 mm tubes Product No SV09 Eight rotor inserts for 16x100 mm tubes Product No SV07 One line cord for use in North America only One Operator s Manual One manual latch release tool One 1 8 inch hex