Home
Istruzioni d`uso: Jar Test FC6S Instruction for
Contents
1. Apparecchiature Fisher diga LI asso isa per laboratorio Bioblock Scientific fax 03 88 67 11 68 fax 056 260 270 fax 915159235 email infos bioblock fr email belgium bioblock com email ventas bioblock com Parc d innovation BP 50111 F67403 illkirch cedex www bioblock com www bioblock be www es fishersci com Laboratory Equipment CERTIFICATIONS ISO 9001 ISO 14001 OHSAS 18001 Istruzioni d uso Jar Test FC6S Instruction for use Jar Test FC6S cod code F10500112 220 240 V 50 60 Hz cod code F10510112 110 120 V 60 50 Hz IMPORTANTE LEGGERE LE INFORMAZIONI CONTENUTE NEL PRESENTE MANUALE PRIMA DELLA MESSA IN FUNZIONE LA SOCIETA DECLINA OGNI RESPONSABILITA SULL IMPIEGO NON CONFORME ALLE ISTRUZIONI DELLO STRUMENTO IMPORTANT READ THE INFORMATION CONTAINED IN THE PRESENT MANUAL BEFORE USING THE UNIT THE MANUFACTURER DOES NOT ACKNOWLEDGE ANY RESPONSIBILITY FOR AN IMPROPER USE OF THE EQUIPMENT NOT RESPONDING TO DIRECTIONS FOR USE Quality and satety 10001268 A7 Le targhette applicate sull apparecchio segnalano all Utente i pericoli ai quali esposto durante l uso o la manutenzione dello stesso Devono restare sull apparecchio e se illeggibili sostituite Avviso di pericolo A prevenzione di infortuni che potrebbero O essere causati dall apparecchio Messa in funzione Non mettere in funzione l apparecchio prima di aver letto le istruzioni per l uso e la manutenzione N
2. of diameter 600 ml wastewater samples stirring paddles 25 mm high 75 mm wide with a thickness of 1 mm stirring height of paddles roughly at middle height of wastewater sample quick turbolent stirring after chemical addition 120 rpm for 120 seconds slow speed stirring during flocculation 30 rom for 25 minutes first evaluation of results after 5 minutes of sedimentation after the end of stirring The results can be evaluated on the basis of different criteria a evaluation of floc dimensions with a numerical degree of merit 0 no flocs 2 very small hardly visible 4 small flocs 6 medium size flocs 8 good size flocs 10 very large flocs b time from the addition of chemicals to first appearing of flocs c evaluation of residual turbidity of supernatant after a determined sedimentation time by turbidimeter d measurement of electrocynetic potential of suspended particles on a sample taken immediately after the addition and mixing of chemicals Coagulations should be easier when zeta potential of particles is very low e evaluation of filterability of clarified water by standardized membrane filters under pressure The reduction of water flow with time is related to the degree of clogging of filters due to residual unsettled suspended matter 10001268 A7 Other measurements relate to the pH of water after chemical treatment lime raises the value while iron and aluminium salts lower it a
3. Dimensions 935 x 347 x 260 mm W x H x D Weight 18 kg Stirrable places N 6 Stirrable volume 11 for each place Single or multiple control Multiple control settable speed for each place Settable speeds 9 speeds 10 15 30 45 60 90 120 150 200 r p m Speed precision 2r p m Timer No Fluorescent lamp Yes Safety class Class 2 IEC 1010 Protection degree IP 30 IEC 529 10001268 A7 11 Analytical methods Coagulation and flocculation of wastewaters Chemical coagulants are added to wastewaters for the primary purpose of removing suspended solids and also phosphorus The addition of chemicals reduces also heavy metal concentrations and improves disinfection efficiency Chemical coagulation can be obtained by adding to wastewaters defined amounts of lime calcium hydroxide alum aluminium sulfate or iron salts ferric or ferrous The coagulation sedimentation process typically involves a injection and mixing of the coagulant that neutralize the electric charge prevalently negative on suspended particles Good and rapid mixing of the coagulant and wastewater is important to ensure efficient use of the chemical Typical detention times in plant basins for turbolent mixing are 15 to 120 seconds b agglomeration of the coagulated particles into large settleable flocs This is accomplished by stirring slowly the water to allow the solids formed by the addition of chemicals to grow
4. dopo aver estratto i contenitori del campione asciugare con un panno o della carta assorbente le aste e le pale da eventuali residui di campione per evitare sgocciolamenti sulla struttura della macchina 8 Accessori Beaker 1000 ml in vetro A00001001 Cono Imhoff in plastica A00001002 Cono Imhoff in vetro graduato A00001003 Supporto per coni Imhoff a 2 posti A00001004 9 Parti di ricambio 230V 50Hz 115V 60Hz F10500112 F10510112 Motoriduttore 10001042 10001042 Fusibile 0 5A 10000668 10000672 5x20 mm Lampada fluorescente 10000687 10000687 30W Scheda elettronica 10000540 10000540 Modulo elettronico 10000519 10000519 Cinghia di trasmissione 10000068 10000068 160 mm O ring 2015 10000213 10000213 Guarnizione per steli 10000123 10000123 Interruttore 10000681 10000681 Trasformatore Il ELTROO5 10 Caratteristiche tecniche Alimentazione 230V 50Hz F10500112 115V 60Hz F10510112 Potenza complessiva 80 W Dimensioni B x H x P 935 x 347 x 260 mm Peso 18 kg Punti di agitazione N 6 Volume agitabile 11 per ogni punto Mono Pluricomando Pluricomando Velocit programmabile per ogni punto Velocit selezionabili 9 velocit 10 15 30 45 60 90 120 150 200 giri 1 Precisione della 2 giri 1 velocit Temporizzatore No Lampada di contrasto Si Classe di sicurezza Classe 2 IEC 1010 Grado di protezione IP30 IEC 529
5. flocculato ad opera della forza di gravit solidi vengono lasciati depositare dall acqua a riposo dopo il trattamento chimico L effluente dalla vasca di sedimentazione molto limpido e non richiede normalmente ulteriore filtrazione La quantit di coagulante chimico richiesta per ottenere una buona chiariflocculazione varia nel tempo e secondo il tipo di acqua di scarico Dosaggi tipici di coagulanti sono 75 250 g m per il solfato di alluminio 45 90 g m per il cloruro ferrico 200 400 g m per la calce Valutazione in laboratorio mediante il jar test La scelta ed il dosaggio del coagulante chimico da adottare per la rimozione dei solidi sospesi dalle acque di scarico vengono stabiliti sulla base dei risultati di valutazioni di laboratorio mediante il cosidetto Jar Test Agitatori multipli con velocit di agitazione riproducibili permettono l adozione di condizioni standard per il test esigenza fondamentale per l ottenimento di risultati riproducibili Le condizioni standard si riferiscono a volume di acqua di scarico in ciascuna posizione dell agitatore multiplo dimensioni e forma dei recipienti e delle palette dell agitatore durata e velocit dell agitazione durante la fase di miscelazione turbolenta durata e velocit dell agitazione durante la fase di flocculazione criteri di valutazione per i risultati ottenuti Le condizioni pi diffusamente adottate sono bechers in vetro da 1000 ml
6. jars in inglese forma alta diametro 105 mm campioni di acqua di scarico da 600 ml palette degli agitatori alte 25 mm larghe 75 mm con uno spessore di 1 mm altezza delle palette circa a met altezza del campione di acqua agitazione turbolenta veloce dopo l aggiunta del coagulante 120 giri al minuto per 120 secondi agitazione a bassa velocit durante la fase di flocculazione 30 giri al minuto per 25 minuti prima valutazione dei risultati dopo 5 minuti di sedimentazione dopo l arresto dell agitazione risultati possono essere valutati in base a differenti criteri a valutazione delle dimensioni dei fiocchi con attribuzione di un punteggio di merito 0 nessun fiocco 2 fiocchi molto piccoli appena visibili 4 piccoli fiocchi 6 fiocchi di densioni medie 8 fiocchi abbastanza grandi 10 fiocchi molto grandi b tempo trascorso dall aggiunta del coagulante fino alla prima comparsa di fiocchi c valutazione della torbidit residua del supernatante dopo un determinato tempo di sedimentazione mediante torbidimetro d misura del potenziale elettrocinetico delle particelle sospese su un campione prelevato immediatamente dopo l aggiunta e la miscelazione del coagulante La coagulazione dovrebbe avvenire pi agevolmente quando il potenziale zeta delle particelle molto basso 10001268 A7 e valutazione della filtrabilit dell acqua chiarificata mediante filtri a membrana sotto pressione L
7. 10001268 A7 11 Metodi analitici Coagulazione e flocculazione di acque di scarico Coagulanti chimici vengono aggiunti alle acque di scarico con lo scopo principale di rimuovere i solidi sospesi in qualche caso il fosforo L aggiunta di coagulanti riduce anche le concentrazioni di metalli pesanti e migliora l efficacia della disinfezione La coagulazione chimica pu essere ottenuta aggiungendo alle acque di scarico quantitativi definiti di calce idrossido di calcio allume solfato di alluminio o sali di ferro ferrosi o ferrici Il processo di coagulazione sedimentazione comprende tipicamente a l introduzione e la miscelazione del coagulante che neutralizza le cariche elettriche prevalentemente negative presenti sulle particelle sospese Una buona e veloce miscelazione del coagulante con l acqua di scarico importante per garantire una efficiente utilizzazione del prodotto Tempi di ritenzione idraulica nelle vasche dove avviene la miscelazione turbolenta sono compresi tra 15 e 120 secondi b agglomerazione delle particelle coagulate in grossi fiocchi in grado di sedimentare Questa viene ottenuta agitando a basso numero di giri per permettere ai solidi formati dall aggiunta del coagulante di crescere di dimensioni in modo da poter essere rimossi per sedimentazione indotta dalla forza di gravit L agitazione viene normalmente ottenuta con pale rotanti lentamente durante 10 30 minuti c sedimentazione del materiale
8. FAT Pumps PERISTALTIC PUMPS Dichiarazione di conformita Noi casa costruttrice VELP SCIENTIFICA s r l indirizzo Via Stazione 16 USMATE MI Italy Dichiariamo sotto la ns responsabilit che il prodotto conforme alle seguenti norme EN 61010 1 2001 EN 61326 1 2000 E soddisfa i requisiti essenziali delle direttive Macchine 98 37 CEE Bassa tensione 73 23 CEE Compatibilit elettromagnetica 89 336 CEE Pi successive modifiche e che sono presenti presso la ns sede i documenti richiesti nell allegato V della direttiva macchine Declaration of compliance We manufacturer VELP SCIENTIFICA s r l address Via Stazione 16 USMATE MI Italy Under our responsability declare that the product is in according with the following standards EN 61010 1 2001 EN 61326 1 2000 And satisfies the essential requirements of directives Machines directive 98 37 CEE Low voltage directive 73 23 CEE Electromagnetic compatibility directive 89 336 CEE Plus modifications and that the documents listed in the annex V are available at Velp s offices as per the machine directive Velp Scientifica srl Via Stazione 16 20040 Usmate Milano Italy Tel 39 039 62 88 11 Fax 39 039 62 88 120 e mail inse velp it internet http www velp com 10001268 A7
9. ILLATORI IN CORRENTE DI VAPORE SCRUBBER PASTIGLIE CATALIZZATRICI NAVICELLE DI PESATA ESTRATTORI PER FIBRA GREZZA ESTRATTORI PER FIBRA DIETETICA ESTRATTORI PER ESTRAZIONE SELETTIVA DI PRODOTTI SOLUBILI MEDIANTE SOLVENTI PER GRASSI Pompe POMPE PERISTALTICHE distribuiti solo in Italia distributed only in Italy 10001268 A7 Thanks for choosing a VELP s product Since 1983 Velp offers to the professional in the field a range of sophisticated and reliable equipment which make available high levels of know how and operative capacity at competitive price Velp works according to ISO 9001 ISO14001 and OHSAS 18001 Quality System Certification Instruments are built according to the International norms IEC 1010 1 and to the rules of CE mark We present you our product Lines Environment C O D B O D B O D Sensor REFRIGERATED THERMOSTATS INCUBATORS JAR TEST METALS IN SLUDGES AND SEDIMENTS NEPHELOMETRIC TURBIDITY NITROGEN IN DIFFERENT FORMS PERFORMANCE OF ANAEROBIC DIGESTERS PHENOLS LEACHABILITY TESTS Stirring MAGNETIC STIRRERS HEATING MAGNETIC STIRRERS MULTIPLACE MAGNETIC STIRRERS LIGHTED MAGNETIC STIRRERS HEATING PLATES VORTEX STIRRERS STIRRERS WITHOUT MOTOR Food and feed DIGESTERS STEAM DISTILLING UNITS SCRUBBER CATALYST TABLETS NITROGEN FREE WEIGHING BOATS EXTRACTORS FOR RAW FIBER EXTRACTORS FOR DIETARY FIBER EXTRACTORS FOR SELECTIVE EXTRACTION OF SOLUBLE PRODUCTS BY SOLVENTS FOR
10. a riduzione della portata dell acqua nel tempo in relazione con l intasamento dei filtri dovuto al materiale sospeso residuo Altre misure si riferiscono al pH dell acqua dopo il trattamento chimico la calce alza il valore mentre i sali di ferro e di alluminio lo abbassano ed alla temperatura dell acqua alla quale vengono condotti gli esperimenti Bibliografia American Society for Testing Materials Norma ASTM D 2035 G4 T Passino R e M Beccari 1970 Standardizzazione del jar test nei processi di coagulazione flocculazione per la rimozione di torbidit inorganiche Acqua e Aria Milano 1 10 Ottobre 1970 Degremont 1978 Etude de la coagulation et de la floculation des eaux Methode n 703 948 950 Memento technique de l eau 1200 pagine Parigi U S Association of Environmental Engineering Professors 1972 Environmental Engineering Unit Operations and Unit Processes Laboratory Manual J T O Connor Ed 350 pagine 1 Forewords During chemical physical treatments of waters aimed to coagulate and precipitate pollutants chemical coagulants are evaluated and their dosing optimized by jar tests In relation to the variable composition of wastewaters the chemical dosing and test frequency are to be adapted to local conditions The jar test equipment manufactured by Velp is characterized by an electronic speed control with an independent speed settable for each place aimed to optimize the result of settlin
11. g reducing chemical consumption 2 Instrument description The FC6S equipment is manufactured in a strong metallic frame with a purposely studied epoxy painting giving a high degree of protection against chemical and mechanical corrosion The examined samples can be lighted by a back fluorescent lamp controlled by a switch located on front panel The stainless steel stirring shafts are settable in height by a self blocking clutch Rotation speed is controlled by six gear motors powered by direct current allowing a good reproducibility also at low speeds The FC6S equipment allows to choose for each place among 9 speeds by the speed selectors located on front panel 10 15 30 45 60 90 120 150 200 r p m 3 Checking the instrument as received When received after packing removal check the integrity of unit The delivery is composed by 1 Jar Test model FC6S 1 Connecting electric cable 4 Assembling and installing Verify that electric power voltage corresponds to the voltage shown on the label The equipment is to be placed on a bench in a stable position 5 Operation controls Place in each stirring position an empty beaker centred in corrispondance of each stirring blade The beakers can be placed raising the shafts and lowering them again to the bottom Connect the equipment to electric power by the supplied cable Turn on the equipment by the main switch that lights showing that power is
12. in size so that they can be removed by gravity settling The stirring is usually performed by slowly rotating paddles during 10 to 30 minutes c sedimentation of the flocculated material by gravity separation The solids are allowed to settle by gravity from the chemically treated standing wastewater The effluent from the chemical settling tank is very clear and does not usually require further filtration The quantity of chemical coagulant required to achieve good coagulation varies with time and from wastewater to wastewater Typical coagulant doses are 75 to 250 g m for alum 45 to 90 g m for ferric chloride 200 to 400 g m for lime Laboratory evaluation by jar test The choice and dosing of chemical coagulant to be adopted for the removal of suspended solids from wastewaters are derived from the results of laboratory evaluations by the so called jar test Multiple stirrers with reproducible stirring speeds allow to adopt standard conditions for the test that are the basic requirement for reproducible results The standard conditions relate to volume of wastewater in each stirring position dimensions and shape of containing vessel and stirring paddle time and speed of stirring during turbulent mixing phase time and speed of stirring during flocculation phase evaluation criteria for the obtained results The most diffused adopted conditions are 1000 ml glass beakers jars tall form 105 mm
13. io per l esecuzione di Jar Test caratterizzato da controllo elettronico della velocit e con impostazione indipendente dei giri per ciascuna pala allo scopo di ottimizzare la chiaroflocculazione e quindi di diminuire la quantit dei reagenti da impiegare 2 Descrizione dello strumento FC6S realizzato con una struttura metallica con verniciatura epossidica appositamente studiata per rendere lo strumento altamente resistente all aggressione chimica meccanica e corrosiva in genere Permette di retro illuminare il campione in esame mediante apposito interruttore accessibile sul pannello frontale Le aste di agitazione realizzate in acciaio inossidabile sono regolabili in altezza e sono dotate di un dispositivo auto bloccante con frizione La trasmissione del moto viene assicurata da sei motoriduttori in corrente continua che consentono di ottenere ottime prestazioni e riproducibilit anche ai bassi regimi FC6S permette di preselezionare per ogni punto di agitazione 9 diversi regimi di velocit tramite i relativi selettori posti sul pannello frontale 10 15 30 45 60 90 120 150 200 giri 1 3 Controllo dello strumento al ricevimento Al ricevimento e dopo aver rimosso l imballaggio controllare l integrit dello strumento La fornitura comprende Jar Test Tipo FC6S Cavo elettrico di alimentazione 4 Montaggio e installazione Verificare che la tensione di alimentazione dell apparecchi
14. ion Do not start to use the unit before reading directions for use and maintenance Do not dispose this equipment like an urban waste SAFETY RULES 1 Do not turn on the instrument before placing into position the containers protecting the blades 2 If a fuse or a lamp is to be replaced first of all unplug the unit CLEANING Always unplug the unit before cleaning Use a damp cloth and not flammable not aggressive detergents PERSONAL PROTECTION EQUIPMENT The equipment used for personal protection must be compatible with the dangers due to the working materials and from the glass vessel MAINTENANCE According to the law on products guarantee our instrument must be repaired at our factory save different agreements with our agents GUARANTEE Starts from the date of delivery note and is referred to the number of register of the single unit The manufacturer engaged in a continuous improvement of his products reserves the right to modify any of their characteristics without notice 1 Introduzione Nei trattamenti chimico fisici delle acque idonei alla coagulazione e precipitazione degli inquinanti le sostanze chimiche per precipitazione vengono confrontate per la loro efficacia ed i loro dosaggi ottimizzati per mezzo di prove Jar Test Date le caratteristiche variabili dei liquami i dosaggi chimici e la frequenza del test devono essere adattati alle condizioni locali La Velp ha realizzato questo apparecch
15. nd the temperature of water at which the experiments are performed References American Society for Testing Materials Norm ASTM D 2035 G4 T Passino R and M Beccari 1970 Standardization of jar test in coagulation flocculation processes for the removal of inorganic turbidities in Italian Acqua e Aria Milano 1 10 October 1970 Degremont 1978 Etude de la coagulation et de la floculation des eaux Methode n 703 948 950 Memento technique de l eau 1200 pages Paris U S Association of Environmental Engineering Professors 1972 Environmental Engineering Unit Operations and Unit Processes Laboratory Manual J T O Connor Ed 350 ppg Schema elettrico 230V 50Hz 1 Fusibile ritardato 5x20mm 0 5 A 2 Presa a vaschetta 3 Interruttore generale illuminato 4 Portalampada sinistro 5 Starter 6 Portalampada destro 7 Reattore 8 Trasformatore 9 Scheda elettronica 10 Motoriduttore 11 Selettori di giri 12 Interruttore lampada 10001268 A7 Electric scheme 230V 50Hz 1 Retarded fuse 5x20 mm 0 5A 2 Power inlet with fuses 3 Lighted main switch 4 Lamp holder left 5 Starter 6 Lamp holder right 7 Reactor 8 Transformer 9 Electronic card 10 Gear motor 11 Speed selector 12 Lamp switch Schema elettrico 115V 60 Hz 1 Fusibile ritardato 5x20mm 3 15 A 2 Presa a vaschetta 3 Interruttore generale illuminato 4 Portalampada sinistro 5 Starter 6 Portalampada destro 7 Reattore 8 Trasformat
16. o sia uguale alla tensione di esercizio Posizionare lo strumento sul piano di lavoro assicurandosi della sua stabilit 10001268 A7 5 Controlli di funzionamento Posizionare per ogni punto di agitazione un contenitore vuoto assicurandosi che siano centrati rispetto alla pala di agitazione Per l inserimento dei contenitori sufficiente alzare verso l alto le aste servendosi del relativo pomolo e successivamente spingendolo verso il basso fino al fine corsa Collegare l apparecchio alla rete di alimentazione mediante l apposito cavo in dotazione Alimentare l apparecchio mediante l interruttore generale l avvenuta alimentazione della macchina visualizzata dall accensione del relativo interruttore L accensione della lampada avviene mediante l apposito interruttore luminoso posto sul pannello frontale Impostare mediante gli appositi selettori la velocit desiderata per ogni punto di agitazione 6 Avvio Dopo aver posizionato i contenitori con il campione in esame ed acceso l interruttore generale illuminato impostare la velocit desiderata per ogni punto di agitazione ruotando la manopola del rispettivo selettore Le velocit programmabili sono le seguenti 10 15 30 45 60 90 120 150 200 giri min E possibile in ogni momento retro illuminare i contenitori del campione tramite l apposito interruttore posto sul pannello frontale 7 Operazioni a fine lavoro A fine lavoro
17. on The back light is operated turning on the switch located on front panel The choosen speed is setted by the corresponding selector for each stirring place 10001268 A7 6 Start After placing the sample containing beakers in each stirring position and turning on the main switch each speed of rotation is setted by the speed selectors located on front panel Settable speeds are 10 15 30 45 60 90 120 150 200 r p m The lighting of sample beakers is possible at any moment by operating the light switch 7 Operations at the end of work Remove the sample beakers and wipe off or dry by filter paper the shafts and blades from sample residues in order to avoide drippings on the equipment 8 Accessories 1000 ml glass beaker A00001001 Imhoff cone plastics A00001002 Graduated Imhoff cone glass A00001003 2 place Imhoff cone stand A00001004 9 Replacement parts 230V 50Hz 115V 60Hz F10500112 F10510112 Gear motor 10001042 10001042 Fuse 0 5A 5x20 mm 10000668 10000672 Fluorescent lamp 30 W 10000687 10000687 Electronic card 10000540 10000540 Electronic module 10000519 10000519 Driving belt 160 mm 10000068 10000068 O ring 2015 10000213 10000213 Shaft gasket 10000123 10000123 Switch 10000681 10000681 Transformer 77 ELTR005 10 Technical features Power supply 230V 50Hz F10500112 115V 60Hz F10510112 Total power 80 W
18. on smaltire apparecchiatura come rifiuto urbano NORME DI SICUREZZA 1 Non mettere in funzione lo strumento prima di avere collocato i contenitori a protezione del movimento delle pale 2 L eventuale sostituzione del fusibile e o delle lampade di contrasto deve essere eseguita dopo aver staccato l alimentazione di rete PULIZIA La pulizia dello strumento deve essere eseguita dopo aver staccato l alimentazione di rete con un panno inumidito con detergenti non infiammabili e non aggressivi MEZZI DI PROTEZIONE PERSONALE mezzi di protezione personale devono essere compatibili con i rischi derivanti dal materiale in lavorazione e dai contenitori in vetro MANUTENZIONE In conformit alla legge sulla garanzia dei prodotti le riparazioni dei nostri strumenti devono essere eseguite presso la ns Sede salvo accordi diversi con i distributori locali GARANZIA Decorre dalla data della bolla di consegna e dal numero di matricola del singolo strumento Il costruttore nell impegno di migliorare costantemente la qualit dei propri prodotti si riserva la facolt di variarne le caratteristiche senza preavviso 10001268 A7 The labels applied to the unit warn the user on the dangers to which he is exposed during the use or the maintenance The labels must be left on the unit and substituted if they are no more readable A Warn of danger To prevent accidents which could be caused by the unit Start the operat
19. ore 9 Scheda elettronica 10 Motoriduttore 11 Selettori di giri 12 Interruttore lampada 13 Trasformatore 115V 10001268 A7 Electric scheme 115V 60Hz 1 Retarded fuse 5x20 mm 3 15 A 2 Power inlet with fuses 3 Lighted main switch 4 Lamp holder left 5 Starter 6 Lamp holder right 7 Reactor 8 Transformer 9 Electronic card 10 Gear motor 11 Speed selector 12 Lamp switch 13 Transformer 115V Grazie per aver scelto un prodotto VELP Dal 1983 Velp offre agli operatori del settore una vasta gamma di strumenti sofisticati ed affidabili che rendono disponibili alti livelli di know how e capacit operative a prezzi competitivi Velp opera secondo le norme della Certificazione del Sistema Qualit ISO 9001 Gli strumenti vengono costruiti in conformit alle norme internazionali IEC 1010 1 e alle regole della marcatura CE Vi presentiamo le nostre Linee di prodotti Ambiente C O D B O D B O D Sensor FRIGOTERMOSTATI INCUBATORI PROVE DI JAR TEST METALLI IN FANGHI E SEDIMENTI TORBIDITA NEFELOMETRICA AZOTO NELLE VARIE FORME EFFICIENZA DI DIGESTORE ANAEROBICO FENOLI TEST DI CESSIONE KITS PER ANALISI DELLE ACQUE FOTOMETRI A FILTRI ANALISI DEI GAS Agitazione AGITATORI MAGNETICI AGITATORI MAGNETICI RISCALDATI AGITATORI MAGNETICI MULTIPOSTO AGITATORI MAGNETICI ILLUMINATI PIASTRE RISCALDANTI AGITATORI A VIBRAZIONE AGITATORI AD ASTA AGITATORI SENZA MOTORE Alimentare DIGESTORI DIST