TC-9D Temperature Controller Operator`s Manual Issue 1 10/20/2011
Contents
1. TC 9D OPERATOR S MANUAL Specification Temperature range Temperature Stability Method of control Display Resolution Temperature Sensor Electrical supply Overall Width Depth Height Thermocouple Length Diameter Lead Length 0 to 620 C or 32 to 1148 F 0 2 at 100C SBL2D 0 4 at 100C SBS4 3 Term PID 1 digital Type K Any voltage between 85V and 250V 203mm 8 305mm 12 191mm 7 5 152mm 6 5mm 3 16 2 1m 7ft 13 TC 9D OPERATOR S MANUAL Installation 1 The fluidised bath obtains its power and control through the TC 9D controller The Fludised bath has therefore to be connected through the TC 9D to the mains or line supply Ifthe fluidised bath has had a plug connected to the lead this needs to be removed Also remove the Twist lock plug from the rear ofthe TC 9D Connectthe mains lead from the fluidised bath to this plug The connections to the plug are as follows Line Lead Plug letter LIVE Black Y NEUTRAL White X GROUND Green G Plug the twist lock plug back into the socket on the TC 9D 2 Plugthe thermocouple into the back of the TC 9D 3 Remove the packing bar from the thermostat sheath on the fluidised bath see the booklet for the bath It is held in with a nut and olive a new olive is supplied with the TC 9D Fit the new olive as appropriate to the thermocouple and push the thermocouple fully intothe thermostat sheath Tighten the nut so thatthe o
2. Raccordez le c ble d alimentation la prise situ e l arri re de l appareil 4 Placez l appareil sur un plan de travail ou surface plane ou le cas ch ant dans une hotte d aspiration en s assurant que les trous d a ration situ s sous l appareil ne soient pas obstru s TC 9D OPERATOR S MANUAL Apr s utilisation Lorsque vous avez fini de chauffer les chantillons n oubliez pas que certaines parties de l appareil les prouvettes leurs supports et autres accessoires risquent d tre tr s chaudes est donc recommand de toujours prendre les pr cautions cit es plus haut L appareil est garanti contre tout d faut ou visde fabrication pour la dur e figurant sur la carte de garantie compter de la date d achat de l appareil Au cours de cette p riode toutes les pi ces d fectueuses seront remplac es gratuitement dans la mesure o la d faillance n est pas due une mauvaise utilisation un accident ou une n gligence Toute r paration sous garantie sera effectu e par le fournisseur Malgr la description et les sp cifications de l appareil donn es dans le manuel de l utilisateur Techne se r serve le droit d effectuer les changements n cessaires l appareil ou tout l ment qui entre dans sa composition Ce manuel a t exclusivement r dig l attention des clients de Techne et aucun l ment de ce guide d instructions ne peut tre utilis comme garantie condition ou repr sentat
3. de desconexi n viene ya ajustado para evitar da os en el equipo En caso de que surgiera un problema de seguridad desconecte el equipo de la red nstalaci n 1 Todos los aparatos Techne se suministran con un cable de alimentaci n Puede ser fijo o independiente del aparato 2 Antes de conectarlo compruebe que el voltaie corresponde al de la placa indicadora Conecte el cable de alimentaci n a un enchufe adecuado seg n la tabla expuesta a continuaci n El equipo debe estar conectado a tierra para garantizar la seguridad el ctrica Conexiones 220V 240V 110V 120V Linea Marr n Negro Neutro Azul Blanco Tierra Verde amarillo Verde Aseg rese de que los equipos marcados 230V en la placa indicadora funcionan a 220V y de que los equipos marcados 120V funcionan a 110V No obstante en ambos casos la velocidad de calentamiento se ver reducida en un 896 aproximadamente La placa indicadora est situada en la parte posterior del equipo 3 Conecte el cable a la toma de tensi n en la parte posterior del equipo 4 Sit e el aparato en un lugar apropiado tal como una superficie de trabajo plana o si fuera necesario incluso en una campana con extractor de humos asegur ndose de que las entradas de aire en la parte inferior no queden obstruidas TC 9D OPERATOR S MANUAL Despu s de su uso Cuando haya finalizado el calentamiento de muestras recuerde que las piezas del equipo tales como tubos bloques y demas accesorios pued
4. n Le rogamos lea cuidadosamente la informaci n contenida en este folleto antes de manipular el aparato Aviso LAS TEMPERATURAS ELEVADAS SON PELIGROSAS pueden causarle graves quemaduras y provocar fuego en materiales combustibles Techne ha puesto gran cuidado en el dise o de estos aparatos para proteger al usuario de cualquier peligro aun asi se debera prestar atenci n a los siguientes puntos EXTREME LAS PRECAUCIONES Y UTILICE GUANTES PARA PROTEGERSE LAS MANOS NO coloque objetos calientes encima o cerca de objetos combustibles NO maneje el aparato cerca de l quidos inflamables o gases NO introduzca ning n l quido directamente en el aparato UTILICE EL SENTIDO COMUN en todo momento Seguridad del usuario Todos los usuarios de equipos Techne deben disponer de la informaci n necesaria para asegurar su seguridad De acuerdo con las instrucciones contenidas en este manual y con las normas y procedimientos generales de seguridad es muy importante que s lo personal debidamente capacitado opere estos aparatos De no ser asi la protecci n que el equipo le proporciona al usuario puede verse reducida Todos los equipos Techne han sido dise ados para cumplir con los requisitos internacionales de seguridad y traen incorporados un sistema de desconexi n en caso de sobretemperatura En algunos modelos el sistema de desconexi n es variable lo que le permite elegir la temperatura seg n sus necesidades En otros el sistema
5. 2 HEATSINK 1 13 7032434 SOLID STATE RELAY 1 14 7002862 AIR VALVE 1 15 7002919 FLOW METER SBS4 SBL1 1 1 1 ON O QO CO N gt 16 7002920 FLOW METER SBL2 SBL2D 17 7002947 THERMOCOUPLE TC 9D OPERATOR S MANUAL 1000101 3170 AUS murg wea4beip GuliiM 0631 1 SG6ETO00Z 8 Sv6coO0Z p3jeaipur 5 un pauuo 3 u p Saui 5uij 3s433u 2J0N weibeig Bul 4621 19
6. 2571 e mail labproducts techneusa com Web site www techneusa com TC 9D OPERATOR S MANUAL Introduction Veuillez lire attentivement toutes les instructions de ce document avant d utiliser l appareil Avertissement DANGER DE TEMPERATURES ELEVEES les op rateurs peuvent subir de graves br lures et les mat riaux combustibles risquent de prendre feu Techne a apport un soin tout particulier la conception de ces appareils de fa on assurer une protection maximale des op rateurs mais il est recommand aux utilisateurs de porter une attention sp ciale aux points suivants PROCEDER AVEC SOIN ET PORTER DES GANTS POUR SE PROTEGER LES MAINS NE PAS poser d objets chauds sur ou pres de mat riaux combustibles NE PAS utiliser l appareil proximit de liquides ou de gaz inflammables NE PAS verser de liquide directement dans l appareil FAIRE TOUJOURS PREUVE DE BON SENS S curit de l op rateur Tous les utilisateurs de produits Techne doivent avoir pris connaissance des manuels et instructions n cessaires la garantie de leur s curit Important cet appareil doit imp rativement tre manipul par un personnel qualifi et utilis selon les instructions donn es dans ce document en accord avec les normes et proc dures de s curit g n rales Dans le cas o cet appareil ne serait pas utilis selon les consignes pr cis es par Techne la protection pour l utilisateur ne serait alors plus gara
7. TOR MAINTAINABLE PARTS WITHIN THE EQUIPMENT In the unlikely event that you experience any problems with your unit which cannot easily be remedied you should contact your supplier and return the unit if necessary Please include any details of the fault observed and remember to return the unit in its original packing Techne accept no responsibility for damage to units which are not properly packed for shipping if in doubt contact your supplier See the Decontamination Certificate supplied with your unit 1 Cleaning Before cleaning your unit ALWAYS disconnect it from the power supply and allow it to cool below 50 C Your unit can be cleaned by wiping with a damp soapy cloth Care should be exercised to prevent water from running inside the unit Do not use abrasive cleaners 2 Fuses Your unit is protected by one or two fuses These should only be changed by suitably qualified personnel If the fuses blow persistently a serious fault is indicated and you may need to return the unit to your supplier for repair Contact nformation For technical sales or servicing information contact your local Techne dealer or Bibby Scientific Ltd Beacon Road Stone Staffordshire ST15 OSA United Kingdom Telephone 44 0 1785 812121 Fax 44 0 1785 813748 e mail equipment barloworld scientific com or Techne Inc 3 Terri Lane Suite 10 Burlington New Jersey 08016 USA Telephone 609 589 2560 Toll free 800 225 9243 ext 306 Fax 609 589
8. ZTECHNE TC 9D Temperature Controller Operator s Manual Issue 1 10 20 2011 TC 9D OPERATOR S MANUAL CONTENTS TC 9D Operator s Manual page SAFETY and INSTALATION English Fran ais Deutsch Espafol 10 INTRODUCTION 12 Packing 12 Description 12 Specification 13 Installation 14 OPERATION 15 TECHNICAL INFORMATION 16 Operator maintenance 16 Servicing 17 Replacement parts 18 Wiring Diagram 19 TC 9D OPERATOR S MANUAL TC 9D OPERATOR S MANUAL NOTES TC 9D OPERATOR S MANUAL Introduction Please read all the information in this booklet before using the unit HIGH TEMPERATURES ARE DANGEROUS they can cause serious burns to operators and ignite combustible material Techne have taken great care in the design of these units to protect operators from hazards but Operators should pay attention to the following points USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS DO NOT put hot objects on or near combustible objects DO NOT operate the unit close to inflammable liquids or gases DO NOT place any liquid directly in your unit At all times USE COMMON SENSE Operator Safety All Operators of Techne equipment must have available the relevant literature needed to ensure their safety It is important that only suitably trained personnel operate this equipment in accordance with the instructions contained in this manual and with general safety standards and
9. che Sicherheit des Anwenders beeintr chtigen Die Ger te von Techne entsprechen den internationalen Sicherheitsbestimmungen und sind mit einem automatischen bertemperaturabschalter ausgestattet Bei einigen Modellen ist der bertemperaturabschalter verstellbar und sollte je nach Anwendung entsprechend eingestellt werden Bei allen anderen Modellen ist der Temperaturschutz voreingestellt um Sch den am Gerat zu vermeiden Wenn ein Sicherheitsproblem auftreten sollte muf das Ger t ausgeschaltet und vom Stromnetz getrennt werden Installation 1 Alle Techne Gerate werden mit einem StromanschluRkabel geliefert Dieses ist entweder fest mit dem Gerat verbunden oder zum Einstecken 2 Vergleichen Sie ob die Spannung Ihrer Stromversorgung mit den Angaben auf dem Typenschild des Ger te bereinstimmen Verbinden Sie das Stromanschlu kabel mit einer geeigneten Stromversorgung gem der n chstehenden Tabelle Achtung Das Ger t mu geerdet sein um die elektrische Sicherheit zu gewahrleisten Verbindungen 220V 240V 110V 120V Stromf hrend Braun Schwarz Neutral Blau Wei Erde Gr n Gelb Gr n Ger te die f r 230 Volt ausgelegt sind k nnen auch bei 220 Volt arbeiten Ger te f r 120 Volt auch bei 110 Volt In beiden F llen verringert sich die Aufheizrate um ca 8 Das Typenschild befindet sich hinten am Ger t 3 Stecken Sie das Stromkabel in die vorgesehene Buchse hinten am Ger t 4 Stellen Sie das Ger t auf eine ebene Arbeit
10. dised bath The TC 9D provides the following features not provided by the energy regulator 1 Precise temperature setting 2 Improved temperature stability Control is achieved by monitoring the signal from the thermocouple sensor By this means the controller will react directly to correct temperature stability 3 Stable temperature under varying loads With the energy regulator the dial setting mustbe manually adjusted to correctfor changes in heat requirements due to a variation in bath loading or ambienttemperature The TC 9D will automatically adjust the heater output to take account of these varying conditions The TC 9D consists of a temperature controller mounted in a case with solid state relay switching and is capable of handling up to 20 amps Connections to the fluidised bath and to the thermocouple are conveniently made with sockets on the back of the instrument In use the TC 9D by passes completely the energy regulator control on the fluidised bath and cycles all the heating elements on and off at the same time The unit is mounted on four rubber feet and provided with a three wire cord and thermocouple When you receive your unit make sure you keep the original packing in case you ever need to return it for service or repair The unit must be transported in the original packing to avoid damage Techne accepts no responsibility for damage incurred unless the unit is correctly packed and transported in this way
11. ement ne peut tre effectu que par un personnel qualifi Si les fusibles sautent sans arr t il s agit d un probl me s rieux Nous vous conseillons dans ce cas de prendre contact avec votre fournisseur pour r paration TC 9D OPERATOR S MANUAL Bitte lesen Sie diese Bedienungsanleitung komplett bevor Sie dieses Gerat benutzen HOHE TEMPERATUREN SIND GEFAHRLICH sie k nnen dem Bediener ernsthafte Verletzungen zuf gen und brennbare Materialien k nnen sich leicht entz nden Techne hat bei der Konstruktion dieses Ger tes sehr darauf geachtet da der Bediener vor Gefahren gesch tzt ist Dennoch sollten Sie auf die folgenden Punkte achten SEIEN SIE VORSICHTIG UND TRAGEN SIE SCHUTZHANDSCHUHE Legen Sie hei e Gegenst nde NICHT auf oder in die N he von leicht brennbaren Materialien vermeiden Sie Arbeiten in der N he von leicht entz ndbaren Fl ssigkeiten oder Gasen Bringen sie KEINE Fl ssigkeiten direkt in Ihr Ger t Benutzen Sie immer den normalen Menschenverstand Sicherheit des Anwenders Alle Benutzer von Techne Ger ten m ssen Zugang zu der entsprechenden Literatur haben um ihre Sicherheit zu gew hren Es ist wichtig da diese Ger te nur von entsprechend geschultem Personal betrieben werden das die in dieser Gebrauchsanweisung enthaltenen Ma nahmen und allgemeine Sicherheitsbestimmungen und vorkehrungen beachtet Wenn das Ger t anders eingesetzt wird als vom Hersteller empfohlen kann dies die pers nli
12. en estar muy calientes Tome las precauciones mencionadas anteriormente Este aparato esta garantizado contra cualquier defecto material o de fabricaci n durante el periodo especificado en la tarjeta de garantia adjunta Este plazo inicia a partir de la fecha de compra y dentro de este periodo todas las piezas defectuosas seran reemplazadas gratuitamente siempre que el defecto no sea resultado de un uso incorrecto accidente o negligencia Mientras se encuentre bajo garantia las revisiones las debe llevar a cabo el proveedor A pesar de la descripci n y las especificaciones de los aparatos contenidas en el Manual del Usuario Techne se reserva por medio de este documento el derecho a efectuar los cambios que estime oportunos tanto en los aparatos como en cualquier componente de los mismos Este manual ha sido preparado exclusivamente para los clientes de Techne y nada de lo especificado en este folleto de instrucciones se tomara como una garantia condici n o aseveraci n de la descripci n comerciabilidad o adecuaci n para cualquler fin especifico de los aparatos o sus componentes Mantenimiento ESTE APARATO DEBE SER DESMONTADO SOLO Y EXCLUSIVAMENTE POR PERSONAL DEBIDAMENTE CAPACITADO EL RETIRAR LOS PANELES LATERALES FRONTALES O TRASEROS SUPONE DEJAR AL DESCUBIERTO TENSION DE LA RED PELIGROSA EL EQUIPO NO CONSTA DE NINGUNA PIEZA DE CUYO MANTENIMIENTO SE PUEDA ENCARGAR EL USUARIO En el caso improbable de que experimentara algun proble
13. er BEACHTEN SIE DASS DIESES GERAT NUR VON TECHNISCHEN FACHKRAFTEN GEOFFNET UND DEMONTIERT WERDEN DARF DURCH ENTFERNEN DES GERAUSES ODER GEHAUSETEILEN SIND BAUTEILE MIT LEBENGEFAHRLICHEN SPANNUNGEN FREI ZUGANGLICH IM INNERN DES GERATES BEFINDEN SICH KEINE TEILE DIE VOM ANWENDER GEWARTET WERDEN MUSSEN Falls Ihr Ger t nicht ordnungsgem arbeitet wenden Sie sich an Ihren Lieferanten oder senden Sie das Gerat wenn n tig zur ck F gen Sie eine genaue Beschreibung des Defektes bei Verpacken Sie das Ger t m glichst im Originalkarton Bitte beachten Sie da Techne keine Haftung bei Transportsch den aufgrund unzureichender Verpackung bernehmen Setzen Sie sich im Zweifelsfall mit Ihrem Lieferanten in Verbindung Bitte beachten Sie die Entgiftungsbescheinigung die Sie mit dem Ger t erhalten haben 1 Reinigen Bevor Sie Ihr Ger t reinigen sollten Sie zuerst den Netzstecker ziehen das Ger t unter 50 C abk hlen lassen Ein feuchtes Tuch mit Seifenl sung reinigt Ihr Ger t am besten Achten Sie darauf daR kein Wasser in das Ger t gelangt Verwenden Sie keine Scheuermittel 2 Sicherungen Die Stromzuleitung ist durch ein oder zwei Sicherungen gesch tzt Diese sollten nur durch qualifiziertes Fachpersonal ausgetauscht werden Wenn die Sicherung wiederholt durchbrennt liegt ein gr erer Defekt vor Das Ger t mu zur Reparatur an Ihren Lieferanten eingesandt werden TC 9D OPERATOR S MANUAL ntroducci
14. ier for repair Cleaning your TC 9D Before cleaning your unit ALWAYS disconnect from the power supply and allow to cool You can clean the case of the TC 9D with a cloth dampened in water or ethanol No part of the unit should be immersed in solvents Do notuse acetone or abrasive cleaners 16 TC 9D OPERATOR S MANUAL Before any servicing of the TC 9D is attempted the unit should be disconnected from the mains Inthe interests of safety both of personal and of the instrument maintenance and repair work should be carried out only by trained electronic or electro mechanical technicians Untrained personnel should not attempt to dismantle the instrument CONTROLLER To remove the controller push the side tabs out to unlock the controller and slide it out of the sleeve Replacement is the reverse of this For further details on the controller refer to the Eurotherm operating instructions supplied SOLID STATE RELAY Accessis gained to the solid state relay by removing the two screws securing it to the heatsink Be sure to apply a thin layer of thermal grease when replacing the relay 17 TC 9D OPERATOR S MANUAL Item Number Description Qty Notes 1 7002942 COVER 1 7002704 FOOT 7002694 TEMPERATURE CONT 7002939 TOP LABEL 7002941 CHASSIS 7002917 THERMOCOUPLE SOCKET 7001029 SOCKETMAINS 7002915 CABLE CLAMP 7002918 MAINS CABLE 2m 10 7002945 FUSEHOLDER 1 11 7001395 FUSE 20 amp 1 12 610047
15. ion concernant la description commercialisation adaptation aux conditions d utilisation ou autre des appareils ou leurs composants Entretien utilisateur IMPORTANT CET APPAREIL NE PEUT ETRE DEMONTE QUE PAR DU PERSONNEL QUALIFIE LORSQUE LES PANNEAUX AVANT ARRIERE ET LATERAUX SONT DEMONTES L OPERATEUR EST EXPOSE A DES TENSIONS QUI PEUVENT ETRE MORTELLES CET APPAREIL NE CONTIENT AUCUN ELEMENT QUI DEMANDE UN ENTRETIEN DE LA PART DE L UTILISATEUR Dans le cas peu probable o votre appareil pr sente un d faut de fonctionnement auquel il est difficile de rem dier il est alors pr f rable de contacter votre fournisseur et le cas ch ant de renvoyer le mat riel Veuillez inclure une description d taill e du probl me constat et retourner l appareil dans son emballage d origine Techne ne sera pas tenu responsable des dommages subis par tout appareil dont l emballage est inad quat pour le transport Pour plus de s ret contactez votre fournisseur Voir le certificat de d contamination livr avec le produit 1 Nettoyage Avant de nettoyer l appareil assurez vous TOUJOURS que le c ble d alimentation est d connect et laissez la temp rature redescendre en dessous de 50 C Utilisez un chiffon impr gn d eau savonneuse pour nettoyer l appareil Veillez ne pas introduire d eau dans l appareil N utilisez pas de produits abrasifs 2 Fusibles La protection de l appareil est assur e par un ou deux fusibles dont le remplac
16. lat workspace or in a fume cupboard if required ensuring that the air inlet vents on the underside are free from obstruction After use When you have finished heating samples remember that parts of the unit the tubes blocks and associated accessories may be very hot Take the precautions listed earlier EN TC 9D OPERATOR S MANUAL Guarantee The unit is guaranteed against any defect in material or workmanship for the period specified on the enclosed guarantee card This period is from the date of purchase and within this period all defective parts will be replaced free of charge provided that the defect is not the result of misuse accident or negligence Servicing under this guarantee should be obtained from the supplier Notwithstanding the description and specification s of the units contained in the Operator s Manual Techne hereby reserves the right to make such changes as it sees fit to the units or to any component of the units This Manual has been prepared solely for the convenience of Techne customers and nothing in this Instruction Book shall be taken as a warranty condition or representation concerning the description merchantability fitness for purpose or otherwise of the units or components Operator maintenance NOTE THAT THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL REMOVING THE SIDE FRONT OR REAR PANELS EXPOSES POTENTIALLY LETHAL MAINS VOLTAGES THERE ARE NO OPERA
17. live holds the thermocouple firmly but do not over tighten 4 Connect a clean dry air supply regulated to 30 PSI via 3 8 OD or 9 5mm OD tubing This tubing should be plugged into the quick disconnect fitting on back of the TC9D labeled air supply Connect 3 8 or 9 5mm OD tubing between the Fluidized bath and the quick disconnect fitting on back labeled Fluidized bath 5 Before connecting the mains supply check the voltage with that marked on the rating plate Connect the mains cable on the TC 9D to a suitable supply as follows CONNECTIONS 230Vor110 120V Live Brown or Black Neutral Blue or White Earth Green yellow The rating plate is on the rear of the unit 6 Setthe energy regulator on the fluidised bath to 10 7 Svitch all the fluidised bath heaters on as the TC 9D will control the temperature by controlling the power input TC 9D OPERATOR S MANUAL OPERATION TC 9D 1 Use the up and down arrows to enter a setpoint temperature The bath will begin to heat immediately The top display is the measured thermocouple temperature and bottom is the setpoint 2 To change the display from OC to oF and vice versa press the 2nd button from the left until the lower display shows UNITS Next press Up arrow to reach preferred setting Press 1st putton on the left to exit after changes are made 3 Two setpoint programming can be used by entering values at menu option SP1 and SP2 See step 2 to access options and exit ou
18. ma con su aparato que no pudiera resolver con facilidad deber a ponerse en contacto con su proveedor y devolverlo si fuera necesario Indique de forma detallada todos los defectos que haya notado y devuelva el equipo en su embalaje original Techne no aceptar responsabilidad alguna por da os causados en equipos que no estuvieran debidamente embalados para su env o si tuviera alguna duda p ngase en contacto con su proveedor S rvase consultar el Certificado de Descontaminaci n suministrado con su aparato 1 Limpieza Antes de limpiar su aparato descon ctelo SIEMPRE de la fuente de alimentaci n y permita que se enfr e por debajo de los 50 C Este aparato se puede limpiar pas ndole un pa o h medo enjabonado H galo con cuidado parae evitar que caiga agua dentro del mismo No utilice limpiadores abrasivos 2 Fusibles Su aparato est protegido por uno o dos fusibles S lo deben cambiarlos personal debidamente capacitado Si los fusibles se fundieran repetidamente esto indicaria una aver a grave y puede que tuviera que devolverle el aparato a su proveedor para su reparaci n 11 TC 9D OPERATOR S MANUAL INTRODUCTION Read the whole of this book before commencing work with the unit The TC 9D Temperature Controller is designed as an alternative to the energy regulator control provided in the laboratory fluidised baths It is supplied with a Chromel Alumel thermocouple which fits into the sheath built into the flui
19. ntie Tous les appareils Techne sont concus pour r pondre aux normes de s curit internationales et sont dot s d un coupe circuit en cas d exc s de temp rature Sur certains mod les ce coupe circuit est r glable pour s adapter l application d sir e Sur d autres mod les il est pr r gl en usine pour assurer la protection de l appareil Dans le cas d un probl me de s curit coupez l alimentation lectrique au niveau de la prise murale et enlevez la prise connect e l appareil Installation 1 Tous les appareils Techne sont livr s avec un c ble d alimentation qui peut tre int gr a l appareil ou raccorder 2 Avant de brancher l appareil v rifiez la tension requise indiqu e sur la plaque d identification Raccordez le c ble lectrique la prise appropri e en vous reportant au tableau ci dessous Il est important que l appareil soit reli la terre pour assurer la protection lectrique requise Connexions 220V 240 V 110V 120 V Phase marron noir Neutre bleu blanc Terre vert jaune vert Le fusible l int rieur de l appareil est destin assurer la protection de l appareil et de l op rateur Remarque les appareils dont la plaque indique 230 V peuvent fonctionner sur 220 V et ceux dont la plaque indique 120 V peuvent fonctionner sur 110 V Dans les deux cas cependant la capacit de chauffage diminuera d environ 8 96 La plaque d identification se trouve l arri re de l appareil 3
20. procedures If the equipment is used in a manner not specified by Techne the protection provided by the equipment to the Operator may be impaired All Techne units have been designed to conform to international safety requirements and are fitted with an overtemperature cutout On some models the cutout is adjustable and should be set to suit the application On all other models the cutout is preset to protect the unit If a safety problem should be encountered switch off at the mains socket and remove the plug from the supply Installation 1 All Techne units are supplied with a power cable This may be integral or plug in 2 Before connecting the mains supply check the voltage against the rating plate The rating plate is on the rear of the unit Connect the mains cable to a suitable plug according to the table below Note that the unit must be earthed to ensure proper electrical safety Connections 220V 240V 110V 120V Live Brown Black Neutral Blue White Earth Green yellow Green The fused plug supplied with the mains lead for use in the UK is fitted with the following value fuse to protect the cable 13AMP The fuse in the unit protects the unit and the operator Note that units marked 230V on the rating plate work at 220V units marked 120V work at 110V In both cases however the heating rate will degrade by approximately 8 3 Plug the mains cable into the socket on the rear of the unit 4 Place the unit on a suitable bench or f
21. sfl che bzw falls erforderlich unter einen Laborabzug Beachten Sie da die Entl ftungsrippen an der Ger teunterseite immer frei zug nglich sind TC 9D OPERATOR S MANUAL Nach dem Gebrauch Vergessen Sie nicht da Teile des Ger tes die Gef e die Bl cke und andere Zubeh rteile nach dem Erhitzen von Proben noch sehr heif sein k nnen Bitte beachten Sie die oben genannten Vorsichtsma nahmen Die Garantiedauer des Ger tes ist auf der beiliegenden Garantiekarte angegeben und schlie t Fehler im Material oder der Verarbeitung ein Die Garantiedauer beginnt am Tag des Einkaufs S mtliche defekte Teile werden innerhalb dieses Zeitraumes kostenlos ersetzt unter der Voraussetzung da dem Defekt keine unsachgem e Handhabung Fahrl ssigkeit oder ein Unfall zugrundeliegt Der unter diese Garantie fallende Service wird vom Lieferanten geleistet Ungeachtet der in dieser Gebrauchsanweisung enthaltenen Beschreibungen und Spezifikationen beh lt sich Techne hiermit das Recht vor nderungen an den Ger ten bzw an einzelnen Ger teteilen durchzuf hren Diese Gebrauchsanleitung wurde ausschlie lich dazu erstellt um Kunden die Handhabung der Techne Ger te zu erleichtern Nichts in dieser Gebrauchsanleitung darf als Garantie Bedingung oder Voraussetzung verstanden werden sei es die Beschreibung Marktg ngigkeit Zweckdienlichkeit oder sonstiges bez glich der Ger te oder deren Bestandteile Wartung durch den Bedien
22. t Menu option SP SEL tells the controller which to use 4 Asingle point calibration offset can be applied by accessing option PV OFS See step 2 to access options and exit out Set a correction value in the opposite direction of your external reference reading i e reference shows the bath to be 3 0 degrees high then set PV OFS to 3 0 15 TC 9D OPERATOR S MANUAL TECHNICAL INFORMATION Brief fault finding notes are given in this section NOTE THAT ELECTRICAL EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL REMOVING THE CASE EXPOSES POTENTIALLY LETHAL MAINS VOLTAGES THERE ARE NO USER MAINTAINABLE PARTS WITHIN THIS EQUIPMENT User Maintenance General In the unlikely event that you experience any problems with your TC 9D which cannot be easily remedied you should contact your supplier and return the unitif necessary Please include details of the faultobserved and remember to return the unit in its original packing Techne accept no responsibility for damage to units which are not properly packed for shipping if in doubt contact your supplier Fuses Ifthe temperature controller fails to light when power is supplied to the unit the fuses may have blown Check that there is no external cause such as a faulty plug These should only be changed by a suitably qualified person Ifthe fuses blow persistently a serious fault is indicated and you may need to return the unit to your suppl