Contents
1. temperature gradient along the probe should be avoided horizontal mounting Vv 33 roof mounting Model C model H seperated sensor probe For mounting template for metal and polycarbonate housing see page 17 Transmitter EE23 xCx and EE23 xHx series are with seperated sensor probe working range sensor probe model C 40 120 C 40 248 F model H 40 80 C 40 176 F working range electronics 40 60 C 40 140 F Mounting of the sensor probe model C 12mm Using the stainless steel mounting flange refer to accessories it is pos sible to mount the probe on the outer wall of the measuring chamber The depth of immersion is adjustable For roof installations use the drip water protection refer to accessories to protect the sensor head and elements against condensed water The transmitter probe must point horizontal or downwards The cable should hang loosely bend for draining water of condensation Mounting of the sensing probe model H 5mm Incl mounting fixture for concealed mounting or in tight spaces e g in museums door frame etc bore diameter 13mm 0 5 material thickness min 3mm 0 1 Model F wall mounting type with rear cable outlet For mounting template for metal and polycarbonate housing see page 17 Positioning of bore holes top left to center line of the probe metal housing x 28 5mm 1 1 y 37 5mm 1 57 polyca
2. s2 6i DI Point de calibration sup rieur D1 e s26 Les calibrations du point de ros e et du point de gel e correspondent la proc dure de calibration de l Humidit Deux m thodes de calibration sont disponibles Calibration d Humidit Temp rature 1 point calibration rapide et simple un point d Humidit Temp rature d fini point de fonctionnement Calibration d Humidit Temp rature 2 points calibration simple pour des r sultats de mesure pr cis sur toute la plage de mesure d Humidit Temp rature Avant la calibration il est recommand de stocker le transmetteur et le dispositif de calibration HUMOR 20 pendant au moins 4 heures dans une m me pi ce a temp rature stable Une temp rature constante devra tre maintenue durant tout le processus Pour une calibration correcte le capteur devra tre stabilis pendant au moins 30 min la valeur d humidit de r f rence Les filtres encrass s devront tre remplac s avant toute re calibration Calibration d Humidit Temp rature 2 points Une calibration d Humidit Temp rature 2 points est recommand e pour obtenir une configuration pr cise sur toute la plage d humidit temp rature La calibration devra tre initi e au point de calibration bas Pour la calibration d humidit 2 points les points de calibration devront tre pris distants et sup rieurs 30 d HR gt 30 C Proc du
3. 28 5mm y 37 5mm boitier polycarbonate x 20 5mm y 25 4mm Les transmetteurs S rie EE23 xFx sont con us pour un montage mural L entr e de c ble s effectue partir de l arri re pour des applica tions de salle blanche par exemple etc Gamme de fonctionnement de la sonde 40 60 C Gamme de fonctionnement de l lectronique 40 60 C Le transmetteur doit tre mont avec la sonde de mesure orient e vers le bas Montage du mod le G Applications m t orologiques Pour le percage et les dimensions des boitiers m tallique ou polycarbo nate voir page 7 Les transmetteurs S rie EE23 xGx sont concus pour des applications en ext rieur Gamme de fonctionnement de la sonde 40 60 C Gamme de fonctionnement de l lectronique 40 60 C Pour des applications en ext rieur le transmetteur doit tre quip d une protection anti rayonnement voir Accessoires Elle permet une circula tion de l air et vite que les valeurs de mesure soient fauss es Le transmetteur doit tre mont avec la sonde de mesure orient e vers le haut et avec une protection anti rayonnement voir Accessoires 28 Connextions lectriques Avec sortie temp rature active NC 8 lt of NC 6 Q T 54 Sorties RF 4 analogiques LLL GND 3 Sorties analogiques Alimentation eg vo D 0 MARCHE c EN tC Raccordement de la partie inf rieure
4. 9888 D gt o 1 CAL s affiche durant la routine de calibration 2 gt s affiche en activant le bouton S1 voir chapitre Composants de fonctionnement Circuit imprim s affiche lorsque le seuil est d pass seulement avec le EE23 avec option alarme lt s affiche en activant le bouton S2 voir chapitre Composants de fonctionnement Circuit imprim s affiche lorsque le seuil est d pass seulement avec le EE23 avec option alarme 3 SET s affiche avec le EE23 avec l option alarme 4 C F Unit pour la Temp rature en C F 5 Td Tf Td Temp rature de point de ros e en C F Tf Temp rature de point de gel e en C F 6 RH Unit pour l Humidit Relative RH 25 Circuit imprim 7 Current Voltage e eooo e eee e 1 Cavalier 2 Afficheur 3 LED D1 rouge D2 vert 4 Bouton poussoir S1 S2 S1 S2 5 Seuil potentiometre 6 Hyst r sis 1 Cavalier 2 Afficheur o Se O10 5 EB Eie sd S z o BE 3 LED i 4 Boutons poussoirs 010V 0 5V OAV CALRH LI CAL M 4 20mA 0 20mA ES 2 5 Seuil potentiom tre Hysteresis 6 Hysteresis Potentiometre Selection du signal de sortie Signal de sortie tension Signal de sortie courant olo d E ara Current P curent I 44222rIF Voltage lolo
5. Horizontale Montage Deckenmontage Montage Modell C Modell H abgesetzter F hler Montagebohrungen und Abmessungen f r Metall bzw Polycarbonatge h use siehe Seite 7 Messumformer der Serie EE23 xCx bzw EE23 xHx sind Messumformer mit einem abgesetzten Messf hler Einsatzbereich F hler Modell C 40 120 C Modell H 40 80 C Einsatzbereich Elektronik 40 60 C Montage des Messf hlers Modell C 12mm Der Montageflansch aus Edelstahl siehe Zubeh r erm glicht eine einfa che Montage an der Messraum Aussenwand und eine stufenlose Verstellung der Eintauchtiefe Bei der Deckenmontage wird ein Tropfwasserschutz siehe Zubeh r zum Schutz des F hlers und der Sensorelemente empfohlen Der Messf hler muss horizontal oder vertikal nach unten montiert sein Wenn m glich sollte bei jeder Montage ein Abtropfbogen gelegt werden Montage des Messf hlers Modell H 5mm Inkl Montagevorrichtung zur versteckten Montage des Miniaturf hlers z B in Museen T rzargen usw Bohrungsdurchmesser 13mm Materialstarke min 3mm Montage Modell F Wandmontage mit r ckseitiger Kabeleinf hrung Montagebohrungen und Abmessungen f r Metall bzw Polycarbonatge h use siehe Seite 7 Positionierung der Montagebohrung oben links zu Bohrung F hlerrohr Metallgeh use x 28 5mm y 37 5mm Polycarbonatgeh use x 20 5mm y 25 4mm Messumformer der Serie EE23 xFx sind f r eine Wandmontage ausge legt
6. lectrique distinct de celui du r cepteur Contacter le revendeur ou faire appel un technicien sp cialis radio TV Attention Toute modification apport e l appareil sans l aval pr alable d un d l gu CEM peut entrainer l interdiction d exploiter l appareil CANADA Conformit ICES 003 Cet appareil de cat gorie B correspond la norme canadienne ICES 003 INHALTSVERZEICHNIS ALLGEMEIN 4 Sicherheitshinweise 4 PRODUKTBESCHREIBUNG 5 BEDIENUNGSELEMENTE 5 Display 5 Platine 6 INSTALLATION 7 Montage Metall bzw Polycarbonatgeh use 7 Montage Modell A 7 Montage Modell B 7 Montage Modell C Modell H 8 Montage Modell F 8 Montage Modell G 8 Elektrische Anschl sse 9 FEUCHTE TEMPERATUR TAUPUNKT FROSTPUNKTKALIBRATION 10 2 Punkt Feuchte Temperaturkalibration 10 1 Punkt Feuchte Temperaturkalibration 11 R cksetzen auf Werkseinstellung 11 INSTANDHALTUNG WARTUNG 11 Sensortausch 11 Sicherungstausch 12 Selbstdiagnose durch LED auf der Platine 12 Selbsthilfe bei St rungen 12 ERSATZTEILE ZUBEH R 12 TECHNISCHE DATEN 13 TABLE OF CONTENTS GENERAL 14 Safety instructions 14 PRODUCT DESCRIPTION 15 OPERATING ELEMENTS 15 Display 15 Circuit board 16 INSTALLATION 17 Mounting of metal and polycarbonate housing 17 Mounting Model A 17 Mounting Model B 17 Mounting Model C Model H 18 Mounting Model F 18 Mounting Model G 18 Electrical Connections 19 HUMIDITY TEMPERATURE DEW POINT FROST POINT CALIBRATION 20 2 p
7. 40 120 C 1 3 0 3 MW rF 2 3 rF 1 4 1 MW rF 1 5 1 5 MW rF Temperaturabh ngigkeit der Elektronik typ 0 015 rF C Ansprechzeit mit Metallgitterfilter bei 20 C too 15s Temperatur Temperatur Sensor EE23 xA B C F Gx Pt1000 Toleranzklasse A DIN EN 60751 EE23 xHx Pt1000 Toleranzklasse B DIN EN 60751 Einsatzbereich Messf hler EE23 xAx 40 60 C EE23 xBx 40 80 C EE23 xCx 40 120 C EE23 xFx 40 60 C EE23 xGx 40 60 C EE23 xHx _ 40 80 C Genauigkeit typ AC 05 EE23 xA B C FIGx AC 05 04 03 4 0 2 LI JL N 2 ee m m m 40 30 20 10 0 aa A 05 J I 10 20 30 40 50 60 70 80 90 100 110 en 02 03 04 EE23 xHx 04 4 0 3 02 4 159983 ES E Ent 0 CH ji meme C I ae lj i Iesel 01 NC 02 oa au os 1 H 70 8 c Ausg nge Allgemeines Temperaturabh ngigkeit der Elektronik typ 0 005 C C 0 1V 0 5 mA lt I 0 5 mA 0 100 rF xx yy C 0 5V 1 mA l 1 mA Temperaturabbildungsbereich wird von E E eingestellt 0 10V 1 mA l 1mA 0 20 mA Hu lt 470 Ohm 4 20 mA R 470 Ohm Max einstellbarer Ausgangsbereich von bis Einheit EE23 AEG EE23 B H EE23 C Feuchte rF 0 100 100 100 rF Temperatur T 40 60 80 120 C Taupunkttemperatur Td 80 60 80 100 C Frostpunkttemperatur Tf 80 0 0 0 C Versorgungsspan
8. General Supply voltage for 0 5 V outputs for O 10 V 0 20 mA and 4 20 mA outputs 10 5 28V DC or 12 28V AC 15 0 28V DC or 15 28V AC optional 100 240V AC 50 60Hz Current consumption for voltage output for DC supply x 25 mA with alarm module for DC supply x 35 mA for AC supply lt 35 mA ff for AC supply 60 mAeg Current consumption for current output for DC supply lt 50 mA with alarm module for DC supply x 60 mA for AC supply 90 mA eff for AC supply x 110 mA ff Housing protection class PC or AI Si 9 Cu 37 IP65 Nema 4 Cable gland M16x1 5 cable 4 5 10 mm 0 18 0 39 Electrical connection screw terminals max 1 5 mm AwG 16 Working temperature range of electronics 40 60 C 40 140 F Working temperature range with display 30 60 C 22 140 F Storage temperature range 40 60 C 40 140 F CE compatibility according EN61000 6 2 FCC Part15 ClassB C EN50081 1 EN61010 1 ICES 003 ClassB Alarm Module optional Output SPDT Switch up to 250V AC 8A or 28V DC 8A threshold hysteresis Setting range 10 95 RH 3 15 RH Setting accuracy 396 RH ij ES 1 Referto the working range of the humidity sensor 8 owning of S Or 2 The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k 2 2 times standard deviation The accuracy was calculated in accordance
9. ckzusetzen muss vorher die 2 INSTANDHALTUNG WARTUNG Sensortausch Um die angegebene Genauigkeit zu erreichen muss nach Sensortausch eine 2 Pkt D Feuchtekalibration durchgef hrt werden Die G ltigkeit der Werkskalibration erlischt mit einem Sensortausch Feuchtesensor nur an den Anschlussdr hten ber hren 1 Versorgungsspannung auschalten 2 Filterkappe abschrauben 3 den defekten Feuchtesensor mit Pinzette entfernen 4 den neuen Feuchtesensor einstecken aktive Seite mit Sensorpads muss nach innen zeigen 5 Filterkappe aufschrauben bei Verschmutzung durch einen neuen Filter ersetzen aktive Seite Versorgungsspannung einschalten active Side 7 Durchf hrung einer Feuchtekalibration siehe 2 Punkt Feuchtekalibration 11 Selbstdiagnose durch LED auf der Platine D1 rot leuchtet w hrend der Kalibrations Routine andauernd leuchtet bei R cksetzung auf Werkskalibration kurz auf D2 gr n blinkt w hrend des Betriebs leuchtet bei defektem Feuchte Temperatur Sensor Selbsthilfe bei St rungen FEHLER il m gliche Ursache gt Ma nahme FALSCHE WERTE Fehler bei Rekalibration des Messumformers R cksetzen auf Werkskalibration und wiederholen der Kalibrationsroutine Filter verschmutzt gt Filtertausch Sensor defekt gt Sensortausch LANGE ANSPRECHZEIT Filter verschmutzt gt Filtertausch falscher Filtertyp gt Filtertyp ist an die Applikation anzupassen
10. r Bezeichnung Buchsenbelegung Schaltausgang E Ansicht von vorne D NO y Euro Norm Anschlussbelegung Geh useunterteil mit integriertem Versorgungsnetzteil 100 240V AC Option V01 Stecker f r ORME Bezeichnung Buchsenbelegung Analogausg nge 8 GND 3 Da OUT1 2 Ansicht von vorne OUT2 1 Furo NONM ec Stecker f r 100 240V Bezeichnung Buchsenbelegung Metallgeh use a m Ansicht von vorne Ed Nulleiter N 3 Stecker f r 100 240V O Bezeichnung Buchsenbelegung Kunststoffgeh use Phase L1 1 Ansicht von vorne 0 Nulleiter N 3 Aussendurchmesser Versorgungsleitung 10 12mm Maximaler Leiterquerschnitt 1 5mm AWG 16 A Die berstrom und Kurzschlussschutzeinrichtung der Versorgungsleitung muss den nationalen Vorschriften entsprechen Geh useunterteil und Gehausemittelteil m ssen im Betriebszustand geerdet sein 9 FEUCHTE TEMPERATUR TAUPUNKT FROSTPUNKT KALIBRATION A CALRH O CAL T unterer Kalibrationspunkt D1 e s2 65 DIO oberer Kalibrationspunkt D1 s26 Taupunkt Frostpunktkalibrationen entsprechen dem Ablauf einer Feuchtekalibration Grunds tzlich stehen zwei Kalibrationsverfahren zur Verf gung 1 Punkt Feuchte Temperaturkalibration schnelle und einfache Kalibration bei einge schr nktem Feuchte Temperaturarbeitspunkt 2 Punkt Feuchte Temperaturkalibration einfache Kalibration f r genaue Messergebnisse ber den gesamten Feucht
11. HIGH permanent condensation on the sensor head dry sensor probe and check the mounting manner of the transmitter wrong filter type gt adjust filter type to application Service stations refer to back cover side Fuse replacement If the green LED on the PCB is not flashing with the supply voltage Switched on check the fuse and replace if required Fuse secondary 250mA T UL248 14 Nominal voltage 250V Replacement types Series MSTU 250 Manufacturer Schurter Order No 0034 7109 Series 374 Manufacturer Littelfuse Order No 374 0250 REPLACEMENT PARTS ACCESSORIES Name order code Name order code extension cable for field HA010302 drip water protection HA010503 bracket for installation onto mounting rails HA010203 radiation shield HA010504 external power supply unit V02 filter caps HA0101xx replacement sensor FEO9 display cover in metal DO3M mounting flange 12mm HA010201 display cover in PC DO3P mounting flange 5mm for model H only HA010208 calibration set HA0104xx 1 calibration EE90 3H 22 TECHNICAL DATA Measured quantities Relative humidity Humidity sensor EE23 xA B C F Gx HC1000 200 EE23 xHx HC105 Working range 0 100 RH Accuracy including hysteresis non linearity and repeatability traceable to intem standards administrated by NIST PTB BEV 15 40 C 5 104 F 15 40 C 5 104 F 25 70 C 13 158 F 40 120 C 40 248 F lt 9
12. against harmful interference in a residential installation This equipment generates uses and can radi ate radio frequency energy and if not installed and used in accordance with the installation manual may cause harmful interference to radio com munications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interfe rence to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the inter ference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Caution Any changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate this device CANADIAN ICES 003 notification This Device B digital apparatus complies with Canadian ICES 003 Cet appareil num rique de la classe B est conforme la norme NMB 003 du Canada E E Elektronik GmbH ne porte aucune garantie et se d gage de toute responsabilit quant cette publication ou un traitement inappropri des produits d crits Le document peut contenir des impr cisions techniques ainsi que des e
13. d Humidit de Temp rature 2 point de calibration haut et stabiliser pendant au moins 30 min 7 BOUTON S1 Presser ce bouton pendant au moins 3 secondes pour initier la routi ne relative au point de calibration haut La LED D1 s allume et le symbole CAL gt s affiche pour indiquer le mode de calibration 8 BOUTON S1 haut et S2 bas Presser ces boutons pour faire correspondre la valeur mesur e avec la valeur de r f rence par tape de 0 196 0 1 C La valeur mesur e peut tre soit lue partir de l affichage optionnel soit mesur e au niveau de la sortie 9 BOUTON S1 Presser ce bouton pendant au moins 3 secondes pour sauvegarder les valeurs de calibration et quitter la routine La LED D1 s teint et le symbole CAL gt s affiche pour indiquer que l on quitte le mode de calibration BOUTON S2 Appuyer sur ce bouton pendant au moins 3 secondes pour quitter la rou tine sans sauvegarder les valeurs de calibration La LED D1 s teint et le symbole CAL s affiche pour indiquer que l on quitte le mode de calibration 30 CALRH O O CALT O O D brasure Cot actif Calibration d Humidit Temp rature 1 point Lorsque la plage d utilisation est restreinte une Humidit Temp rature sp cifique alors une calibration 1 point ce point d Humidit Temp rature sera suffisante Conform ment la plage d utilisation un point de calibration haut ou bas doit tre s lectionn KP gt
14. die Kabeleinf hrung erfolgt von der R ckseite z B Reinraumanwendung Einsatzbereich F hler 40 60 C Einsatzbereich Elektronik 40 60 C Der Messumformer muss so montiert sein dass der Messf hler nach unten zeigt Montage Modell G Meteorologief hler Montagebohrungen und Abmessungen f r Metall bzw Polycarbonatge h use siehe Seite 7 Messumformer der Serie EE23 xGx sind f r Aussenanwendungen ausge legt Einsatzbereich F hler 40 60 C Einsatzbereich Elektronik 40 60 C F r Anwendungen im Freien ist der Messumformer mit einem Strahlungsschutz auszur sten siehe Zubeh r Dieser bewirkt eine Zwangsbel ftung und verhindert eine Verf lschung der Messwerte Der Messumformer muss so montiert sein dass der Messf hler nach oben zeigt und sollte nur mit Strahlungsschutz montiert wer den siehe Zubeh r 8 Elektrische Anschl sse Ausg nge und Stromversorgung Schaltausgang mit aktiven Temperaturausgang 8 nc LUN a Schaar C N 12 nicht erreicht 6 to M analoge 4 Ausgangs signale 3 2 1 mit passivem Temperaturausgang 8 7 E T Sensor passiv 6 5 4 analoges 3 Eu Speisung gd 2 1 Anschlussbelegung Geh useunterteil mit Steckanschl ssen 8 35V DC 12 30V AC Option C03 C07 Stecker f r Versorgung oe van ean und Analogausg nge 5 GND 4 Ansicht von vorne 2 GND 3 OUT1 2 Euro Norm OUT2 1 Stecker f
15. du boitier avec des connecteurs 8 35V DC 12 30V AC Option C03 C07 Connecteur pour ali 4 Gi mentation et sorties analogiques vue de face m c Connecteur pour sortie relais vue de face Norme EU D signation Assignation des connexions Vi 5 GND 4 GND 3 OUT1 2 OUT2 1 D signation Assignation des connexions NC 1 COM 3 NO 4 Raccordement de la partie inf rieure du boitier avec alimentation int gr e 100 240V AC Option V01 Connecteur pour sortie analogique vue de face Connecteur pour boitier m tallique avec alimen tation 100 240V vue de face Connecteur pour boitier polycarbonate avec ali mentation 100 240V vue de face D signation Assignation des connexions GND 3 OUT1 2 OUT2 1 D signation Assignation des connexions Terre PE Phase L1 1 E Non assign N 3 D signation Assignation des connexions Phase L1 1 Non assign N 3 Section c ble d alimentation 10 12mm Section maximum par conducteur 1 5mm AWG 16 v rifi e 29 Les protections contres les surtensions et les court circuits sur les c ble d alimenta tion doivent imp rativement respecter les normes nationales en vigueurs La mise la terre entre la partie inf rieure et la partie interm diaire du boitier doit tre CALIBRATION D HUMIDITE TEMPERATURE POINT DE ROS E ET DE GEL E A CALRH CALT Point de calibration inf rieur S2 D1 D
16. glicht eine einfache Installation und im Servicefall einen raschen Wechsel der Messeinheit Geh usemittelteil Feldjustage und Vor Ort Kalibration der gesamten Messkette werden dadurch auf einfachste Weise erm glicht Durch Wahl der passenden Bauformen kann der EE23 f r praktische jede Anwendung eingesetzt werden Modell A f r Wandmontage Modell B f r Kanalmontage Modell C H mit abgesetztem F hler Modell F Wandmontage mit r ckw rtiger Kabeleinf hrung Modell G f r Ausseneins tze und meteorologischen Anwendungen Verschiedene Optionen erweitern die Einsatzm glichkeiten BEDIENUNGSELEMENTE Display CAL gt lt SET Td Tf 009 D gt WIN 1 CAL wird w hrend der Kalibrations Routine angezeigt 2 gt wird bei Bet tigung des Tasters S1 angezeigt siehe Kapitel Bedienungselemente Platine wird bei berschreitung des Schaltpunktes angezeigt nur bei EE23 mit Schaltmodul lt wird bei Bet tigung des Tasters S2 angezeigt siehe Kapitel Bedienungselemente Platine wird bei Unterschreitung des Schaltpunktes angezeigt nur bei EE23 mit Schaltmodul 3 SET wird bei EE23 mit optionalem Schaltausgang angezeigt 4 C F Einheit f r Temperatur in C F 5 Td Tf Td Taupunkttemperatur in C F Tf Frostpunkttemperatur in C F 6 RH Einheit f r rel Feuchte RH Platine 1 Jumper 2 Display a kel O C
17. of bottom part of the housing with plug connections 8 35V DC 12 30V AC option C03 C07 Description plug for supply and 4 Gi V analogue output 5 GND front view DT GND OUT1 Euro Standard OUT2 plug for alarm output GO Gi coo front view D COM NO Euro Standard Connection assignment Connection assignment Connection configuration of bottom part of the housing with integrated power supply 100 240V AC option V01 plug for GO Gi STEEN analogue outputs 5 OUT1 front view OUT2 EE c o EN plug for 100 240V S EIN i grounding metal housing Sage ER front view neutral wire N plug for 100 240V O Description polycarbonate housing phase L1 front view 0 O neutral wire N External diameter of supply cable 10 12mm 0 39 0 47 Connection assignment Connection assignment PE 1 3 Connection assignment 1 3 The protection of the supply cable against excess current and short circuit shall be in N Maximum wire cross section 1 5mm AWG 16 accordance with national and local codes Bottom and centrepiece of the housing shall be grounded 19 HUMIDITY TEMPERATURE DEW POINT FROST POINT CALIBRATION A CALRH O O CALT amp low calibration point D1 2 6 high calibration point Die s2 65i The procedure of dew point and frost point calibrations are equal with the procedure of humidity calibrations The EE23 transmitter series can be
18. ou 50 HR KP gt ou 5096 de la gamme de Temp rature Ce type de calibration r sulte en une certaine impr cision dans la plage d Humidit de Temp rature restante Proc dure de calibration d Humidit de Temp rature 1 point 1 Avant de d marrer la calibration positionner le cavalier sur CAL RH pour une cali bration Humidit sur CAL T pour une calibration Temp rature 2 Positionner le capteur dans l talon r f rence d Humidit de Temp rature point de calibration et stabiliser pendant 30 minutes 3 BOUTON S1 point de calibration gt 50 HR gt 50 de la gamme de Temp rature Presser ce bouton pendant au moins 3 secondes pour initier la routine La LED D1 s allume et le symbole CAL gt s affiche pour indiquer le mode de calibration ou BOUTON 82 point de calibration 50 HR 50 de la gamme de Temp rature Presser ce bouton pendant au moins 3 secondes pour initier la routine La LED D1 s allume et le symbole CAL gt s affiche pour indiquer le mode de calibration 4 BOUTON S1 haut et S2 bas Appuyer sur ces boutons pour faire correspondre la valeur mesur e avec la valeur de r f rence par tapes de 0 1 0 1 C La valeur mesur e peut tre soit lue partir de l affichage optionnel soit mesur e au niveau de la sortie 5 BOUTON S1 Appuyer sur ce bouton pendant au moins 3 secondes pour sauve garder la valeur de calibration et quitter la routine La LED D1 s tein
19. 0 RH gt 90 RH 1 3 0 3 mv RH 2 390 RH 1 4 1 mv RH 1 5 1 5 mv RH Temperature dependence electronics typ 0 015 RH C Response time with metal grid filter at 20 C too lt 15 sec Temperature Temperature sensor element EE23 xA B C F Gx Pt1000 class A DIN EN 60751 EE23 xHx Pt1000 class B DIN EN 60751 Working range sensing head EE23 xAx 40 60 C 40 140 F EE23 xBx 40 80 C 40 176 F EE23 xCx 40 120 C 40 248 F EE23 xFx 40 60 C 40 140 F EE23 xGx 40 60 C 40 140 F EE23 xHx 40 80 C _ 40 176 F Accuracy typ AU EU EE23 xA B C F Gx Ae eng EE23 xHx 04 04 4 03 4 03 02 0 2 40 30 20 10 0 01 H ds 10 20 30 40 50 60 70 80 90 100 110 120 0 2 03 04 Temperature dependence of electronics typ 0 005 C C Outputs 0 1V 0 5 mA lt l lt 0 5 mA 0 100 RH xx yy C 0 5V 1 mA l 1 mA temperature output scale adjustable by E E or 0 10V 1 mA l lt 1 mA with configuration kit 0 20mA Hu lt 350 Ohm 4 20 mA R 350 Ohm Max adjustable output scaling from up to units EE23 AEG EE23 B H EE23 C Humidity RH 0 100 100 100 RH Temperature T 40 40 60 140 80 176 120 248 C F Dew point temperature Td 80 112 60 140 80 176 100 212 C F Frost point temperature Tf 80 112 0 32 0 32 0 32 C F
20. FEUCHTE TEMPERATUR MESSUMFORMER HUMIDITY TEMPERATURE TRANSMITTER TRANSMETTEUR HUMIDIT TEMP RATURE MANUAL uw YOUR PARTNER IN SENSOR TECHNOLOGY E u EN ELEKTRONIK Ges m b H BA EE23 B 06 Technische nderungen vorbehalten 192075 E E Elektronik Ges m b H bernimmt f r diese Publikation keinerlei Garantie und bei unsachgem er Handhabung der beschriebenen Produkte keinerlei Haftung Diese Publikation kann technische Ungenauigkeiten oder typographische Fehler enthalten Die enthaltenen Informationen werden regelm ig ber arbeitet und unterliegen nicht dem nderungsdienst Der Hersteller beh lt sich das Recht vor die beschriebenen Produkte jederzeit zu modifizieren bzw abzu ndern Copyright E E Elektronik Ges m b H Alle Rechte vorbehalten USA FCC Hinweis Dieses Ger t ist gepr ft worden und stimmt mit den Bedingungen f r ein Ger t der Kategorie B gem Teil 15 der FCC Richtlinien berein Diese Bedingungen sind erstellt worden um einen angemessenen Schutz gegen EMV St rungen in einem Wohnbereich sicherzustellen Dieses Ger t erzeugt verbraucht und kann Hochfrequenzenergie ausstrahlen Wenn es nicht in bereinstimmung mit der Bedienungsanleitung installiert und ver wendet wird k nnen EMV St rungen zu den Funkverbindungen verursacht werden Jedoch gibt es keine Garantie dass EM St rungen nicht in einer bestimmten Installation auftreten k nnen Wenn das Ger t EMV St rungen zum
21. GER TEAUSFALL keine Versorgung gt Zuleitung und Versorgungsspannung berpr fen ZU HOHE FEUCHTEWERTE Betauung im F hlerkopf F hlerkopf trocknen und Montage des Messf hlers berpr fen falsche Filtertype gt Filtertype ist an die Applikation anzupassen Servicestellen siehe hintere Umschlagseite Sicherungstausch Blinkt die gr ne LED auf der Platine bei angelegter Versorgungsspannung nicht ist die Sicherung zu berpr fen und bei Bedarf auszutauschen Sicherung sekund r 250mA T UL248 14 Nennspannung 250V Ersatztypen Serie MSTU 250 Hersteller Schurter BestNr 0034 7109 Serie 374 Hersteller Littelfuse BestNr 374 0250 ERSATZTEILE ZUBEHOR Bezeichnung Bestellcode Bezeichnung Bestellcode Filterkappen HA0101xx Versorgungsnetzteil V02 Strahlungsschutz HA010504 Aufsatz fur Hutschienenmontage HA010203 Display Deckel in Metall DO3M F Austauschsensoren FEO9 Display Deckel in PC DO3P Tropfwasserschutz HA010503 Montageflansch 12mm HA010201 Kalibrierset HA0104xx Montageflansch 5mm HA010208 Zwischenkabel f r Kalibrierung vor Ort A HA010302 nur f r Modell H 1 Kalibration EE90 3H 12 TECHNISCHE DATEN Messwerte Relative Feuchte Feuchtesensor EE23 xA B C F Gx HC1000 200 EE23 xHx HC105 Einsatzbereich 0 100 rF Genauigkeit inkl Hysterese Nichtlinearitat und Wiederholgenauigkeit 15 40 C lt 90 rF 15 40 C gt 90 rF 25 70 C
22. Radio oder Fernsehempfang verursacht das kann festgestellt wer den indem man das Ger t ein und ausschaltet wird dem Benutzer empfohlen die EMV St rungen durch folgende Ma nahmen zu beheben Stellen Sie die Antenne neu ein oder verlagern Sie die empfangende Antenne Erh hen Sie den Abstand zwischen dem Ger t und dem Empf nger Schlie en Sie das Ger t an einem anderen Stromkreis als den Empf nger an Fragen Sie den H ndler oder einen erfahrenen Radio TV Techniker Vorsicht nderungen am Ger t die nicht ausdr cklich durch einen EMV Beauftragten genehmigt sind k nnen dazu f hren dass der Betreiber das Ger t nicht mehr gebrauchen darf KANADA ICES 003 Bescheid Dieses Ger t der Kategorie B entspricht der kanadischen Norm ICES 003 E E Elektronik Ges m b H doesn t accept warranty and liability claims neither upon this publication nor in case of improper treatment of the described products The document may contain technical inaccuracies and typographical errors The content will be revised on a regular basis These changes will be implemented in later versions The described products can be improved and changed at any time without prior notice Copyright E E Elektronik Ges m b H All rights reserved USA FCC notice This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection
23. SS SSSSSSSS avec sondes de mesure d port es GIN Gamme de fonctionnement de la sonde Modele C 40 120 C Modele H 40 80 C Bride de S Gamme de fonctionnement de l lectronique 40 60 C tage en 1305 Z protection anti goutte E SIS Montage de la sonde de mesure Mod le C 12mm Z x m La bride de montage en acier inoxydable voir Accessoires permet le ee doye Ski ea montage de la sonde sur une des parois murs exterieures de la cham bre de mesure ll est galement possible d ajuster en continu la profon Montage horizontal Sal deur d immersion Accessoires est utilis e pour prot ger la sonde et l l ment Dm Durant le montage du couvercle la protection anti goutte voir de d tection contre la condensation La sonde de mesure doit tre positionn e horizontalement ou verticalement dirig e vers le bas Il est recommand si possible d utiliser une plaque d gout tement pour chaque montage Montage de la sonde de mesure Mod le H 5mm Inclus accessoire pour dissimuler le montage de la sonde miniature ex mus e cadre de porte Diam tre de percage 13mm Epaisseur mati re min 3mm Montage du mod le F Montage mural avec passe c ble arri re Pour le per age et les dimensions des bo tiers m tallique ou polycarbo nate voir page 7 Positionnement du per age pour le passage de sonde position de d part en haut gauche bo tier m tallique x
24. allique 20 C too 15s Temp rature Capteur de temp rature EE23 xA B C F Gx Pt1000 classe A DIN EN 60751 EE23 xHx Pt1000 classe B DIN EN 60751 Gamme de mesure de la sondeEE23 xAx 40 60 C EE23 xBx 40 80 C EE23 xCx 40 120 C EE23 xFx 40 60 C EE23 xGx 40 60 C EE23 xHx 40 80 C Erreur de ZE aen EE23 xA B C F Gx SEE EE23 xHx justesse typ 04 u 04 in expen ET E ey Eh En eres ve Tere e 8 Le es mE ta Va EE C STI E RE gereest C 40 30 20 10 0 10 20 30 40 50 60 70 80 90 100 110 20 4p 30 20 10 0 10 20 30 40 50 60 70 8i 0 1 0 2 02 4 03 03 4 04 04 4 05 05 3 Influence de la temp rature sur l lectronique typ 0 005 C C Sorties 0 1V 0 5 mA Il 0 5 mA 0 100 HR xx yy C 0 5V 1 mA l 1 mA chelle de sortie en temp rature r gl e par E E 0 10V 1 mA l lt 1mA 0 20 mA R 470 Ohm 4 20 mA R 470 Ohm Plage de mesure max ajustable de Unit EE23 A RG EE23 B H EE23 C Humidite rF 0 100 100 100 HR Temp rature T 40 60 80 120 C Temp rature de ros e Td 80 60 80 100 C Temp rature de gel e Tf 80 0 0 0 C G n ralit s Tension d alimentation pour sorties 0O 1V 0 5V 10 5 28V DC oder 12 28V AC pour 0 10 V 0 20 mA und 4 20 mA 15 0 28V DC oder 15 28V AC optionel 100 240V AC 50 60Hz Consommation pour sortie
25. ar le biais de quatre vis non fournies lt 4 2mm 3 Connexion du transmetteur voir Connexions lectriques 4 Montage de la demi partie et du capot en utilisation quatre vis inclues Montage du mod le A Montage mural Pour le per age et les dimensions des bo tiers m tallique ou polycarbo nate voir ci dessus Les transmetteurs S rie EE23 xAx sont con us pour un montage mural Gamme de fonctionnement de la sonde 40 60 C Gamme de fonctionnement de l lectronique 40 60 C il Le transmetteur doit tre mont avec la sonde orient e vers le bas Montage du modele B Montage sur gaine Pour le per age et les dimensions des bo tiers m tallique ou polycarbo nate voir ci dessus Positionnement du pergage pour le passage de sonde position de d part en haut gauche boitier m tallique x 28 5mm y 37 5mm boitier polycarbonate x 20 5mm y 25 4mm Les transmetteurs Serie EE23 xBx sont congus pour un montage sur gaine Gamme de fonctionnement de la sonde 40 80 C il Gamme de fonctionnement de l lectronique 40 60 C Le transmetteur doit tre mont avec la sonde dispos e horizontalement ou orient e vers le bas sur la gaine 27 Montage du mod le C mod le H Sonde de mesure d port e Pour le percage et les dimensions des boitiers m tallique ou polycarbo nate voir page 7 ib Cor Les transmetteurs S rie EE23 xCx et EE23 xHx sont des transmetteurs V phe SSSS
26. calibrated in two ways 1 point humidity temperature calibration quick and simple calibration on a defined humidity temperature point working point 2 point humidity temperature calibration simple calibration for accurate measuring results over the whole humidity temperature working range To reach a temperature balance it is recommended to keep the transmitter and the reference chamber e g HUMOR 20 for minimum 4 hours in the same room During calibration procedure it is important to keep the temperature constant For calibration the humidity sensor probe must be stabilised at least 30 minutes in the reference chamber Replace a used dirty filter cap before calibration 2 point humidity calibration 2 point temperature calibration For accurate adjustment over the whole working range a two point calibration is recommended Start calibration at the low calibration point The difference between the two calibration points should be gt 30 RH gt 30 degC 86 F 2 point humidity calibration procedure temperature calibration procedure Start at low calibration point 1 At the beginning of a humidity calibration set the jumper to CAL RH for a temperature calibration set the jumper to CAL T 2 Insert the sensor probe into the reference humidity reference temperature 1 low calibration point and stabilise for at least 30 minutes 3 BUTTON S2 Pressing the button for 3 seconds starts the proce
27. cated by the lit LED D1 and by the symbol CAL on the LC display or BUTTON S2 Calibration point 50 RH 50 of measurement value Pressing the button for 3 seconds starts the procedure The calibration mode is indicated by the lit LED D1 and by the symbol CAL on the LC display 4 BUTTON S1 up and S2 down Pressing the two buttons will adjust the measu ring value in steps of 0 196 0 1 degC up or down to the reference value The actual measuring value is indicated on the display or can be measured with the analogue output 5 BUTTON S1 Pressing the button for 3 seconds the calibration value is stored and the procedure is ended Exiting the calibration mode is indicated by deactivation of the LED D1 and the symbol CAL on the LC display BUTTON S2 Pressing the button for 3 seconds the calibration procedure will be ended without storing the calibration values Exiting the calibration mode is indicated by deactivation of the LED D1 and the symbol CAL on the LC display Resetting the customer calibration to the factory calibration BUTTON S1 and S2 To reset the RH or T values the jumper has to be set on the respective location before starting the factory calibration In neutral mode pressing both buttons simultaneously for 5 seconds customer calibration settings are reset to factory calibration A short flash of the LED D1 indicates the reset Maintenance sensor exchange After changing the sensor it
28. chaltmoduls von den Anschlussklemmen muss die daf r vorgesehene Trennwand im Unterteil montiert sein W hrend des Betriebes muss das Ger t geschlossen sein Die Schutzart des ge ffneten Geh uses ist IPOO Bauteile mit gef hrlichen Spannungen k nnen direkt ber hrt werden Arbeiten an spannungsf hrenden Teilen sind grunds tzlich zu unterlassen und d rfen nur von geschultem Personal durchgef hrt werden Sicherheitshinweise f r Verwendung des integrierten Netzteils Option V01 Das Geh use des Ger ts muss w hrend des Betriebs vollst ndig geschlossen sein Die Schutzklasse eines ge ffneten Geh uses entspricht IPOO Arbeiten an spannungsf hrenden Teilen sind grunds tzlich zu unterlassen und d rfen nur von ausgebildeten Fachleuten durchgef hrt werden Umweltaspekte Die Produkte von E E Elektronik werden unter Ber cksichtigung aller wichtigen Umweltaspekte entwickelt Aus diesem Grund sollte auch bei der Entsorgung auf Vermeidung von Umweltverschmutzung geachtet werden Bei Entsorgung des Messumformers muss auf die sortenreine Trennung der einzelnen Komponeten geachtet werden Das Geh use besteht aus recyclingf higem Polycarbonat bzw Metall Aluminium Al Si 9 Cu 3 Die Elektronik muss im Elektronikschrott gesammelt und fach gerecht entsorgt werden PRODUKTBESCHREIBUNG Die EE23 Messumformer Serie steht f r Multifunktionalit t h chste Genauigkeit einfache Montage und Wartung Das 3 teilige Geh use erm
29. chement switching on 75 d so 0 e re cv NO ru Sch gt 62 70 Seuil switching point 33 34 35 my YOUR PARTNER IN SENSOR TECHNOLOGY jz D BE ELEKTRONIK Ges m b H HEAD OFFICE E E ELEKTRONIK Ges m b H Langwiesen 7 A 4209 Engerwitzdorf Austria Tel 43 7235 605 0 Fax 43 7235 605 8 info epluse com www epluse com TECHNICAL OFFICES E E CHINA BEIJING Tel 86 10 84992361 info epluse cn www epluse cn E E CHINA SHANGHAI Tel 86 21 61176129 info epluse cn www epluse cn E E GERMANY Tel 49 6172 13881 0 info epluse de www epluse de E E FRANCE Tel 33 4 7472 35 82 info epluse fr www epluse fr E E ITALY Tel 39 0331 177 31 02 info epluse it www epluse it E E KOREA Tel 82 31 728 2223 info epluse co kr www epluse co kr
30. d Tf Td dew point temperature C F Tf frost point temperature C F 6 RH unit for relative humidity RH 15 Circuit board 1 Jumper 2 Display a kel O Current o 2 HE WB E e e zm 3 LED 4 Push Button 5 Level Potentiometer Hysteresis 6 Hysteresis Potentiometer 1 Jumper Selection of output range voltage output signal current output signal 5 GI PTET Current G 2222238 gt LH oltage 2222325 Voltage lolo HELF SS 882255238 S i o o o 2 Display place for optional display 3 LED D1red lit permanently during the calibration routine lit up shortly when reset to factory calibration settings D2 green flashes during operation lit up if humidity temperature sensor is damaged 4 Push Button S1 for 1 point humidity temperature calibration humidity gt 50 RH for 2 point humidity temperature calibration high calibration point to store the calibration settings S2 for 1 point humidity temperature calibration humidity 50 RH for 2 point humidity temperature calibration low calibration point to exit the calibration procedure without storing the calibration values S1 S2 to reset to factory calibration settings 5 Level Potentiometer set the threshold model EE23 with alarm module only 6 Hysteresis Potentiometer set the hysteresis model EE23 with alarm module o
31. dure for the low calibration point The calibration mode is indicated by the lit LED D1 and by the symbol CAL on the LC display 4 BUTTON S1 up and S2 down Pressing the two buttons will adjust the measu ring value in steps of 0 196 0 1 degC up or down to the reference value The actual measuring value is indicated on the display or can be measured with the analogue output 5 BUTTON S1 Pressing the button for 3 seconds the calibration value is stored and the procedure is ended Exiting the calibration mode is indicated by deactivation of the LED D1 and the symbol CAL on the LC display BUTTON S2 Pressing the button for 3 seconds the calibration procedure will be ended without storing the calibration values Exiting the calibration mode is indicated by deactivation of the LED D1 and the symbol CAL on the LC display 6 Insert the sensor probe into reference humidity reference temperature 2 high calibration point and stabilise for minimum 30 min 7 BUTTON S1 Pressing the button for 3 seconds starts the procedure for the high calibration point The calibration mode is indicated by the lit LED D1 and by the symbol CAL on the LC display 8 BUTTON S1 up and S2 down Pressing the two buttons will adjust the measu ring value in steps of 0 196 0 1 degC up or down to the reference value The actual measuring value is indicated on the display or can be measured with the analogue output 9 BUTTON S1 Pressin
32. e Temperaturarbeitsbereich Vor Kalibration empfiehlt es sich den Messumformer und die Kalibrationsvorrichtung z B HUMOR 20 mindestens 4h im selben temperaturstabilen Raum zu lagern um Temperaturunterschiede zu vermeiden Wahrend der gesamten Kalibration ist auf konstante Temperatur zu achten F r eine Kalibration muss der F hler min 30 min in der Referenzfeuchte stabilisiert werden Vor einer Kalibration sollten verschmutze Filterkappen getauscht werden 2 Punkt Feuchtekalibration 2 Punkt Temperaturkalibration F r genaue Einstellungen ber den gesamten Feuchtebereich Temperaturmessbereich wird eine 2 Punkt Feuchtekalibration 2 Punkt Temperaturkalibration empfohlen Die Kalibration sollte beim unteren Kalibrationspunkt gestartet werden Abstand der 2 Kalibrationspunkte soll gt 30 r F gt 30 C aufweisen Ablauf der 2 Punkt Feuchtekalibration Temperaturkalibration Start beim unteren Kalibrationspunkt 1 Vor Beginn der Kalibration den Jumper f r eine Feuchtekalibration auf CAL RH setzen f r eine Temperaturkalibration auf CAL T setzen 2 Positionierung des F hlers in der Referenzfeuchte Referenztemperatur 1 unterer Kalibrationspunkt und Stabilisierung f r mindestens 30 min 3 TASTER S2 Durch mind 3 sek langes Dr cken wird die Routine f r den unteren Kalibrationspunkt gestartet Der Kalibrationsmode wird durch das Leuchten der LED D1 und durch das Symbol CAL im LCD Display an
33. easurement NC De HYSTERESIS EIN ji i 3 Siehe Bestellcode 4 Genauigkeit der Rechenfunktionen beachten 5 Stecker siehe Zubeh r 13 Einschaltpunkt Switching On 6 50 gt 62 70 Schaltpunkt switching point gt GENERAL The manual is a part of the scope of supply and serves to ensure proper handling and optimum functioning of the instrument For this reason the manual must be read before start up In addition the manual is for all personnel who require knowledge concerning transport setup operation maintenance and repair The manual must not be used for the purpose of competition without a written consent from E E Elektronik and must also not be forwarded to third parties Copies for personal use are permitted All information technical data and illustrations contained in these instructions are based on information available at the time of publication Symbol assertion This symbol indicates a safety instruction These safety instructions should always be followed carefully By not following these instructions injuries of persons or material damage could happen Therefore E E Elektronik does not accept liability This symbol indicates a note These notes should be followed to achieve optimum functioning of the equipment General safety instructions Extreme mechanical stress and improper use must be avoided Be careful when removing the filter cap to avoid damage o
34. em Betrieb der Wartung und Reparatur befasst sind in Kenntnis zu bringen Diese Bedienungsanleitung darf nicht ohne das schriftliche Einverst ndnis von E E Elektronik zu Zwecken des Wettbewerbes verwendet und auch nicht an Dritte weitergegeben werden Kopien f r den Eigenbedarf sind erlaubt S mtliche in dieser Anleitung enthaltene Angaben technische Daten und Darstellungen basieren auf zum Zeitpunkt der Erstellung verf gbaren Informationen Symbolerkl rung Dieses Zeichen zeigt Sicherheitshinweise an Sicherheitshinweise sind unbedingt zu befolgen Bei Nichtbeachtung k nnen Verletzungen von Personen oder Sachsch den entstehen E E Elektronik bernimmt daf r keine Haftung Dieses Zeichen zeigt Hinweise an Um die optimale Funktion des Ger tes zu gew hrleisten sind Hinweise zu befolgen Allgemeine Sicherheitshinweise e berm ige mechanische und unsachgem e Beanspruchungen sind unbedingt zu vermeiden Vorsicht beim Abschrauben der Filterkappe da das Sensorelement besch digt werden kann Beim Sensorelement handelt es sich um ein ESD gef hrdetes Bauteil d h beim Ber hren des Sensorelementes sind ESD Schutzma nahmen einzuhalten e Sensoren nur an den Anschlussdr hten ber hren Montage elektrischer Anschluss Wartung und Inbetriebnahme d rfen nur von dazu ausgebildetem Fachpersonal durchgef hrt werden Sicherheitshinweise f r Verwendung des Schaltmoduls mit Spannungen gt 50V Zur Trennung des S
35. f the sensor element The sensor is an Electro Static Discharge sensitive component ESD When touching the sensor element ESD protective measures should be followed Hold the sensor on the connection wires only nstallation electrical connection maintenance and start up procedures should be executed by qualified technical personel only Safety instructions for using the alarm output module with voltages gt 50V For the separation of the alarm output module from the connecting terminals the the refor designated partition must be mounted in the bottom During operation the housing of transmitter must be closed Work on live parts is to be omitted basically and may only be executed by trained personnel The protection class of the opened housing is IP00 Components with dangerous voltages can be touched directly Safety instructions for use of the integrated power supply option V01 During operation the housing of transmitter must be closed Work on live parts is to be omitted basically and may only be executed by trained personnel The protection class of the opened housing is IPOO Environmental aspects Equipment from E E Elektronik9 is developed with due consideration to all resultant environmental issues When you dispose the equipment you should avoid environmental pollution For disposal of the transmitter the individual components must be sorted with care The housing consists of recyclable polycarbonate or meta
36. g the button for 3 seconds the calibration value is stored and the procedure is ended Exiting the calibration mode is indicated by deactivation of the LED D1 and the symbol CAL gt on the LC display BUTTON S2 Pressing the button for 3 seconds the calibration procedure will be ended without storing the calibration values Exiting the calibration mode is indicated by deactivation of the LED D1 and the symbol CAL on the LC display 20 CALRH O CALT 5 D aktive Seite active Side 1 point humidity calibration 1 point temperature calibration When the working range is limited to a certain more narrow range a calibration at one humidity point one temperature point is sufficient n accordance with the working range either the high or low calibration point should be selected CP gt or 50 RH CP gt or 50 of measurement value This calibration causes an extra inaccuracy for the rest of the working range 1 point humidity calibration procedure temperature calibration procedure 1 At the beginning of a humidity calibration set the jumper to CAL RH for a temperature calibration set the jumper to CAL T 2 Insert the sensor probe into the reference humidity reference temperature calibration point and stabilise for at least 30 minutes 3 BUTTON S1 Calibration point gt 50 RH gt 50 of measurement value Pressing the button for 3 seconds starts the procedure The calibration mode is indi
37. gebohrungen It Bohrplan siehe Zeichnungen ausf hren Fur das Polycarbonatgeh use steht zus tzlich ein Montageset f r Hutschienenmontage zur Verf gung siehe Zubeh r 2 Der Geh useunterteil wird mit 4 Schrauben montiert Schraubendurchmesser lt 4 2mm nicht im Lieferumfang enthalten 3 Anschluss des Messumformers siehe Elektrische Anschl sse 4 Montage von Mittelteil und Deckel mittels 4 Schrauben im Lieferumfang enthalten Modell A Wandmontage Montagebohrungen und Abmessungen f r Metall bzw Polycarbonatge h use siehe oben Messumformer der Serie EE23 xAx sind f r die Wandmontage ausgelegt Einsatzbereich F hler 40 60 C Einsatzbereich Elektronik 40 60 C Der Messumformer muss so montiert sein dass der Messf hler nach unten zeigt Modell B Kanalmontage Montagebohrungen und Abmessungen f r Metall bzw Polycarbonatge h use siehe oben Positionierung der Montagebohrung oben links zu Bohrung F hlerrohr Metallgeh use x 28 5mm y 37 5mm Polycarbonatgeh use x 20 5mm y 25 4mm Messumformer der Serie EE23 xBx sind f r die Kanalmontage ausgelegt Einsatzbereich F hler 40 80 C Einsatzbereich Elektronik 40 60 C Der Messumformer muss so montiert sein dass der Messf hler waagrecht oder nach unten in den Kanal ragt I 2 Tropfwasser c 2 x schutz E Vermeiden von Temperaturgefalle Abtropfbogen 8 entlang des Messf hlers
38. gezeigt 4 TASTER 81 up und S2 down Durch Dr cken der beiden Taster wird der Messwert in 0 1 0 1 C Schritten auf den Referenzwert abgeglichen Der Messwert kann entweder am Display abgelesen oder am Ausgang gemessen werden 5 TASTER S1 Durch mind 3 sek langes Dr cken wird der Kalibrationswert gespeichert und die Routine verlassen Das Verlassen des Kalibrationsmodes wird durch das Erl schen der LED D1 und dem Symbol CAL im LCD Display angezeigt TASTER S2 Durch mind 3 sek langes Dr cken wird die Routine verlassen ohne die Kalibrationswerte zu speichern Das Verlassen des Kalibrationsmodes wird durch das Erl schen der LED D1 und dem Symbol CAL im LCD Display angezeigt 6 Positionierung des F hlers in der Referenzfeuchte Referenztemperatur 2 oberer Kalibrationspunkt und Stabilisierung f r mindestens 30 min 7 TASTER 1 Durch mind 3 sek langes Dr cken wird die Routine f r den oberen Kalibrationspunkt gestartet Der Kalibrationsmodus wird durch das Leuchten der LED D1 und durch das Symbol CAL gt im LCD Display angezeigt 8 TASTER S1 up und S2 down Durch Dr cken der beiden Taster wird der Messwert in 0 1 0 1 C Schritten auf den Referenzwert abgeglichen Der Messwert kann entweder am Display abgelesen oder am Ausgang gemessen werden 9 TASTER S1 Durch mind 3 sek langes Dr cken wird der Kalibrationswert gespeichert und die Routine verlassen Das Verlassen des Kalibrationsmodus wird du
39. is necessary to perform a two point calibration to reach the specified accuracy again The factory calibration is no longer valid after performing a sensor exchange Touch the sensor elements on the connection wires only 1 Switch off the supply voltage 2 Unscrew the filter cap 3 Pull out the humidity sensor element with a tweezer 4 Put in the new humidity sensor the active side side with the sensor pads has to face the inside 5 Screw the filter cap on again in case of pollution replace it by a new filter cap 6 Switch on the supply voltage 7 Perform a humidity calibration refer to 2 point humidity calibration 21 Self diagnosis LED D1 red lit permanently during the calibration routine lit up shortly when reset to factory calibration settings LED D2 green flashes during operation lit up if humidity temperature sensor is damaged Self help in case of errors ERROR Li possible cause steps against it WRONG VALUE error during or after recalibration of transmitter Reset to factory calibration and recalibrate according to calibration procedure filter cap polluted change filter cap Sensor damaged change humidity sensor element LONG RESPONSE TIME filter cap polluted change filter cap wrong filter type gt adjust filter type to application INSTRUMENT BREAKDOWN no supply voltage check the connection line and voltage supply HUMIDITY VALUES TOO
40. l aluminium Al Si 9 Cu 3 The electronics must be collected as electronic scrap and disposed of according to the regulations in force 14 PRODUCT DESRCRIPTION The EE23 transmitter series stands for multifunctionality highest accuracy easy mounting and service The construction of the transmitter enables field calibration and local loop calibration in a simple way The EE23 can be employed in all common applications by choosing the appropriate housing combination e A and B versions versions are designed for wall or duct mounting C H versions with remote sensing probe F wall mounting version with rear cable outlets is designed for clean room applications G version with optional radiation shield is designed for meteorological applications Various options extend the large number of applications OPERATING ELEMENTS Display CAL gt lt SET Td Tf TERS D gt o m 1 CAL is indicated during the calibration procedure 2 gt is indicated when pushing button S1 refer to Opterating elements Circuit board is indicated when overstepping the threshold model EE23 with alarm module only lt is indicated when pushing button S2 refer to Opterating elements Circuit board is indicated when dropping below the threshold model EE23 with alarm module only 3 SET indication for EE23 with optional alarm output 4 C F unit for temperature C F 5 T
41. nes exp riment es Aspects environnementaux L quipement provenant de E E Elektronik GmbH est d velopp en totale consid ration de toutes les contraintes environnementales r sultantes la mise au rebut de l quipe ment devra ainsi respecter ces contraintes environnementales Il est important de trier chaque composant individuel avec pr caution lors de la mise au rebut du transmetteur Le boitier est constitu de m tal Al Si 9 Cu 3 recyclable Les composants lectroniques doivent tre collect s comme fragments lectroniques et mis au rebut conform ment la r glementation en vigueur 24 DESCRIPTION DU PRODUIT Les s ries EE23 sont des transmetteurs multifonctions d une grande pr cision faciles installer et entretenir gr ce au nouveau concept de boitier compos de 3 modules La conception innovante du transmetteur permet une calibration sur site ais e Le EE23 peut tre utilis dans la plupart des applications en choisissant la combinaison de bo tier appropri e Mod le A pour un montage mural Mod le B pour un montage en gaine Mod le C H pour un montage avec sonde d port e Mod le F pour un montage mural avec les sorties de cables par l arri re Modele G pour les applications en ext rieur et m t orologiques Les diff rentes options permettent diverses possibilit s de montage COMPOSANTS DE FONCTIONNEMENT Module d affichage CAL gt lt SET Td Tf
42. nly 16 all dimensions in mm housing INSTALLATION Mounting of metal and polycarbonate housing 1 Drill the mounting holes according to the mounting template For the polycarbonate housing a special mounting set for installation onto mounting rails is available refer to Accessories 2 The bottom part of the housing is mounted with 4 screws screw diameter 4 2mm 0 2 not included in the scope of supply 3 Connection of the transmitter refer to electrical connection 4 Mounting of the middle part and cover with 4 screws included in the scope of supply Model A wall mounting For mounting template for metal and polycarbonate housing see above Transmitter EE23 xAx series are designed for wall mounting working range sensor probe 40 60 C 40 140 F working range electronics 40 60 C 40 140 F D The sensor probe must point downwards Model B duct mounting For mounting template for metal and polycarbonate housing see above Positioning of bore holes top left to center line of the probe metal housing x 28 5mm 1 1 y 37 5mm 1 5 polycarbonate housing x 20 5mm 0 8 y 25 4mm 17 Transmitter EE23 xAx series are designed for duct mounting working range sensor probe 40 80 C 40 176 F working range electronics 40 60 C 40 140 F The sensor probe must point horizontal or downwards in the duct channel 17
43. nung f r 0 1 V 0 5 V Ausgangssignal f r O 10 V 0 20 mA und 4 20 mA 10 5 28V DC oder 12 28V AC 15 0 28V DC oder 15 28V AC optional 100 240V AC 50 60Hz Stromverbrauch bei Ausgabe eines Spannungssignals bei DC Versorgung lt 25 mA mit Alarmmodul bei DC lt 35 mA bei AC Versorgung lt 35 mA bei AC lt 60 mA Stromverbrauch bei Ausgabe eines Stromsignals bei DC Versorgung lt 50 mA mit Alarmmodul bei DC lt 60 mA bei AC Versorgung lt 90 mA bei AC 110 mA Geh use Schutzklasse PC bzw AI Si 9 Cu 3 IP65 Verschraubung des Anschlusskabels M16x1 5 Elektrische Anschl sse 9 Schraubklemmen f r max 1 5 mm Betriebstemperaturbereich der Elektronik 40 60 C Betriebstemperaturbereich mit Displays 30 60 C Lagertemperaturbereich 40 60 C erf llt CE Richtlinie gem EN61000 6 2 EN61000 6 3 C EN61010 1 Alarm Modul optional Ausgang Wechselkontakt f r max 250VAC 8A oder 28VDC 8A Schaltpunkt Hysterese m glicher Einstellbereich 10 95 rF 3 15 rF Einstellgenauigkeit 3 rF 1 Bezogen auf den Arbeitsbereich des Feuchtesensors E 2 15 3 2 Die Toleranzangaben beinhalten die Unsicherheit der Werkskalibration mit einem Erweiterungsfaktor 9 3 4 V N La gt J k 2 2 fache Standardabweichung Die Berechnung der Toleranz erfolgte nach EA 4 02 unter 257 en TA Ber cksichtigung des GUM Guide to the Expression of Uncertainty in M
44. oint humidity temperature calibration 20 1 point humidity temperature calibration 21 Reset to factory calibration 21 MAINTENANCE 21 Sensor Exchange 21 Fuse replacement 22 Self diagnosis 22 Self help in case of errors 22 REPLACEMENT PARTS ACCESSORIES 22 TECHNICAL DATA 23 TABLE DES MATIERES GENERALITES 24 Conseils de s curit 24 DESCRIPTION DU PRODUIT 25 COMPOSANTS DE FONCTIONNEMENT 25 Module d affichage 25 Circuit imprim 26 INSTALLATION 27 Montage boitier M tallique ou Polycarbonate 27 Montage Modele A 27 Montage Modele B 27 Montage Modele C Mod le H 28 Montage Modele F 28 Montage Mod le G 28 Raccordement lectrique 29 CALIBRATION HUMIDITE TEMPERATURE POINT DE GIVRE ET DE GEL E 30 Calibration d humidit Temp rature 2 points 30 Calibration d humidit Temp rature 1 point 31 Recouvrement des r glages usine 31 ENTRETIEN MAINTENANCE 31 Remplacement du capteur 31 Remplacement du fusible 32 Message d auto diagnostic avec la LED du circuit imprim 32 Aide en cas de messages d erreur 32 PIECES DE RECHANGE ACCESSOIRES 32 CARACTERISTIQUES TECHNIQUES 33 gt ALLGEMEIN Die Bedienungsanleitung ist Bestandteil des Lieferumfanges und dient der Sicherstellung einer sachgem en Handhabung und optimalen Funktion des Ger tes Aus diesem Grund muss die Bedienungsanleitung unbedingt vor Inbetriebnahme gelesen werden Dar ber hinaus ist die Bedienungsanleitung jeglichen Personen welche mit dem Transport der Aufstellung d
45. rbonate housing x 20 5mm 0 8 y 25 4mm 1 Transmitter EE23 xFx series are designed for wall mounting in clean room applications Connection cables do not disturb the cleaning process working range sensor probe 40 60 C 40 140 F working range electronics 40 60 C 40 140 F The transmitter must be mounted with the sensor probe pointed downwards Model G wall mounting for meteorology For mounting template for metal and polycarbonate housing see page 17 Transmitter EE23 xGx series are designed for outdoor applications working range sensor probe 40 60 C 40 140 F working range electronics 40 60 C 40 140 F For outdoor applications the transmitters must be equipped with a radiation shield refer to accessories The shield causes forced ventilation which prevents overheating of the sensor probe in direct sunlight It also protects the sensing probe against water of direct rainfall The sensor probe must point upwards and should be used in combination with the radiation shield refer to accessories 18 Electrical connections output and supply with active temperature output analogue outputs T Sensor passive es 4 ER our DES analogue output 3 Power supply Gi F alarm output nc PEINE OFF switching point c N 12 not exceeded no EN Connection configuration
46. rch das Erl schen der LED D1 und dem Symbol CAL im LCD Display angezeigt TASTER S2 Durch mind 3 sek langes Dr cken wird die Routine verlassen ohne die Kalibrationswerte zu speichern Das Verlassen des Kalibrationsmodes wird durch das Erl schen der LED D1 und dem Symbol CAL im LCD Display angezeigt 10 1 Punkt Feuchtekalibration 1 Punkt Temperaturkalibration Ist der Arbeitsbereich auf einen bestimmten Feuchte Temperaturbereich beschr nkt so ist eine 1 Punkt Feuchtekalibration 1 Punkt Temperaturkalibration f r diesen Feuchtepunkt ausreichend Es sollte entsprechend dem Arbeitsbereich entweder der obere oder untere Kalibrationspunkt gew hlt werden KP gt oder 50 r F KP gt oder lt 50 vom Temperaturmessbereich Durch diese Art der Kalibration ergibt sich eine gewisse Ungenauigkeit im brigen Feuchtebereich Temperaturbereich Ablauf der 1 Punkt Feuchtekalibration 1 Punkt Temperaturkalibration 1 Vor Beginn der Kalibration den Jumper f r eine Feuchtekalibration auf CAL RH setzen f r eine Temperaturkalibration auf CAL T setzen CALRH CALT 2 Positionierung des Fuhlers in der Referenzfeuchte Referenztempertur Kalibrationspunkt und Stabilisierung fur mindestens 30 min 3 TASTER S1 Kalibrationspunkt gt 50 r F gt 50 vom Temperaturmessbereich Durch mind 3 sek langes Drucken wird die Routine gestartet Der Kalibrationsmode D1 D m durch das Leuchten der LED D1
47. re de calibration d Humidit de Temp rature 2 points D marrer avec le point de calibration inf rieur 1 Avant de d marrer la calibration positionner le cavalier sur CAL RH pour une calibration Humidit sur CAL T pour une calibration Temp rature 2 Positionner le capteur dans l talon r f rence d Humidit 1 de Temp rature 1 point de calibration bas et stabiliser pendant au moins 30 min 3 BOUTON S2 Appuyer sur ce bouton pendant au moins 3 secondes pour activer la routine de calibration du point inf rieur La LED D1 de la carte de circuits imprim s s illumine pour indiquer le mode de calibration et le symbole CAL s affiche 4 BOUTON S1 haut et S2 bas Appuyer sur ces boutons pour faire correspondre la valeur mesur e avec la valeur de r f rence par tapes de 0 1 0 1 C La valeur mesur e peut tre soit lue partir de l affichage optionnel soit mesur e au niveau de la sortie 5 BOUTON S1 Appuyer sur ce bouton pendant au moins 3 secondes pour sauve garder la valeur de calibration et quitter la routine La LED D1 s teint et le symbole CAL s affiche pour indiquer que l on sort du mode de calibration BOUTON S2 Appuyer sur ce bouton pendant au moins 3 secondes pour quitter la rou tine sans sauvegarder les valeurs de calibration La LED D1 s teint et le symbole CAL s affiche pour indiquer que l on sort du mode de calibration 6 Positionner le capteur dans l talon r f rence
48. rreurs typographiques Les informations contenues seront r guli rement r vis es Ces modifications seront apport es dans des versions ult rieures Les produits d crits peuvent tre am lior s et modifi s tout moment Les caract ristiques techniques sont sujettes modification Copyright E E Elektronik Ges m b H Tous droits r serv s USA Consigne FCC Cet appareil a t contr l et r pond aux exigences relatives aux appareils de cat gorie B conform ment la partie 15 des directives FCC Ces exigences ont t tablies afin d assurer une protection raisonnable CEM contre les perturbations lectromagn tiques dans les habitations Cet appareil g n re consomme et peut diffuser de l nergie haute fr quence Les liaisons radio peuvent subir des perturbations lectromagn tiques d s lors qu il n est pas install et exploit conform ment au manuel d utilisation N anmoins des perturbations lectromagn tiques peuvent appa raitre dans une installation donn e Si l appareil cause des perturbations lectromagn tiques la r ception radio ou TV ceci peut tre v rifi en mettant l appareil en marche puis l arr t il est conseill l utilisateur d liminer les perturbations lectromagn tiques en prenant les mesures sui vantes Revoir le r glage de l antenne de r ception ou d placer l antenne Augmenter la distance s parant l appareil du r cepteur Brancher l appareil un circuit
49. s 6 Etablir la connexion 7 Effectuer une calibration d Humidit voir calibration d Humidit 2 points 31 Auto diagnostics via les LED de la carte de circuits imprim s D1 rouge s allume durant toute la routine de calibration s allume bri vement lors du recouvrement des param tres usines D2 vert clignote pendant la mise en fonctionnement s allume lorsque le capteur Humidit Temp rature est d fectueux Auto diagnostics et messages d erreur ERREUR D cause possible gt Mesures Aide AFFICHAGE DE VALEURS INCORRECTES Erreur durant la recalibration du transmetteur R initialiser la calibration usine et r p ter la routine de calibration Filtre encrass Remplacer le filtre Capteur d fecteux Remplacer le capteur TEMPS DE REPONSE LONG Filtre encrass Remplacer le filtre Type de filtre incorrect Le type de filtre devra correspondre l application DEFAILLANCE DU TRANSMETTEUR pas de tension lectrique gt V rifier le cable et l alimentation lectrique VALEURS D HUMIDITE ELEVEES Ros e condensation sur la t te S cher la t te de la sonde et v rifier le de la sonde de mesure type de montage de cette derni re Type de filtre incorret gt Le type de filtre doit correspondre l application Remplacement du fusible Si la LED verte du circuit imprim ne clignote pas durant la mise sous ten Zg sion le fusible doit tre contr l et a
50. t et le symbole CAL s affiche pour indiquer que l on sort du mode de calibration BOUTON S2 Appuyer sur ce bouton pendant au moins 3 secondes pour quitter la rou tine sans sauvegarder les valeurs de calibration La LED D1 s teint et le symbole CAL s affiche pour indiquer que l on sort du mode de calibration R initialisation de la calibration personnalis e par la calibration usine BOUTON S1 et S2 Les cavaliers doivent d abord tre s lectionn s pour r inialiser les valeurs HR ou T Appuyer simultan ment sur les deux boutons pendant au moins 5 secondes pour r i nitialiser la calibration personnalis e avec la calibration usine La LED D1 s illumine bri vement pour indiquer l initialisation avec la calibration usine ENTRETIEN MAINTENANCE Remplacement du capteur Pour atteindre la pr cision indiqu e une calibration 2 points doit tre r alis e apr s le remplacement du capteur acalibration usine r alis e apr s le remplacement du capteur est valid e par la LED qui s allume Le capteur d humidit ne doit tre touch qu au niveau des fils de raccordement 1 Mettre l alimentation lectrique hors tension 2 D visser pr cautionneusement le capuchon filtre 3 Retirer le capteur d humidit endommag avec des pincettes 4 Ins rer le capteur de rechange avec le c t actif orient vers l int rieur voir le dia gramme 5 Visser pr cautionneusement le capuchon filtre le remplacer si encras
51. t852E a e oe Per ss3eeas CEPS EF C oo o 4o o TO o Emplacement pour afficheur optionnel s allume durant toute la routine de calibration s allume bri vement durant le recouvrement aux param tres usine clignote pendant la mise en marche s allume en cas de capteur Humidit Temp rature d fectueux Bouton poussoir pour calibration Humidit Temp rature 1 point Humidit gt 50 HR Bouton poussoir pour calibration Humidite Temperature 2 points point de calibration sup rieure Bouton poussoir pour la sauvegarde des valeurs de calibration Bouton poussoir pour calibration Humidit Temp rature 1 point Humidit 50 HR Bouton poussoir pour calibration Humidit Temp rature 2 points point de calibration inf rieure Bouton poussoir pour la proc dure de calibration sans sauvegarde des valeurs de mesure pour le recouvrement des param tres usine R glage du seuil seulement avec le EE23 avec une sortie relais R glage de l hyst r sis seulement avec le EE23 avec une sortie relais 26 Dimensions en mm V FJ D 150 Bo tier m tallique Bo tier polycarbonate INSTALLATION Montage pour bo tier m tallique ou polycarbonate 1 Per age des trous de montage selon le plan Un set de montage est disponible pour un montage encliquetable sur rail uniquement pour bo tier polycarbonate voir Accessoires 2 La partie basse du bo tier est mont e p
52. tension pour alimentation lt 25 mA pour module pour DC 35 mA pour alimentation lt 35 mAeg d alarme pour AC lt 60 mA Consommation pour sortie courant pour alimentation DC lt 50 mA pour module pour DC lt 60 mA pour alimentation AC 90 mA d alarme pour AC 110 mA Boitier Indice de protection PC ou Al Si 9 Cu 37 IP65 Presse toupe M16x1 5 Connexion lectrique Bornes visser max 1 5 mm Gamme de fonctionnement en temp rature de l lectronique 40 60 C Gamme de fonctionnement en temperature avec affichage 30 60 C Gamme de temp rature pour le stockage 40 60 C Compatibilit lectromagn tique EN61000 6 2 EN61000 6 3 C EN61010 1 Module alarme optionnel Sortie Relais inverseur coupure max 250VAC 8A ou 28VDC 8A Seuil Hyst r sis Gamme de r glage 10 95 HR 3 15 HR Incertitude de r glage 3 HR 1 Se r f rer la gamme de fonctionnement du capteur d humidit ag s ulfde 2 Les incertitudes d talonnage avec un coefficient d largissement k 2 2 fois l erreur standard sont inclus dans 2 A d clenchement Va y i l erreur de justesse Les incertitudes sont calcul es selon EA 4 02 en tenant compte du GUM Guide to the E switching off g L 5 expression of Uncertainty in Measurement Eni M HYSTERESIS 3 Se r f rer la r f rence de commande ON VON os 4 Voir la pr cision des fonctions de calcul 5 Connecteurs voir Accessoires seuil de o a L4o d enclen
53. u besoin remplac f Fusible secondaire 250mA T UL248 14 Tension nominale 250V Types de remplacement gar Serie MSTU 250 Fabricant Schurter Nr de commande 0034 7109 5 Serie 374 Fabricant Littelfuse Nr de commande 374 0250 N US PIECES DE RECHANGE ACCESSOIRES Description Code de commande Description Code de commande Filtre HA0101xx Alimentation pour module externe V02 Protection anti rayonnement HA010504 Support de fixation pour Afficheur couvercle de boitier rails de montage HA010203 metallique DO3M Capteur de remplacement HR FEO9 Afficheur couvercle de boitier Protection anti ruissellement HA010503 polycarbonate DO3P Set de calibration HA0104xx Bride de montage 12mm HA010201 C ble de connection pour calibration Bride de montage 5mm HA010208 en ligne HA010302 Seulement pour mod le H Etalon de calibration 1 EE90 3H 32 CARACTERISTIQUES TECHNIQUES Donn es mesur es Humidit relative Capteur d humidit EE23 xA B C F Gx HC1000 200 EE23 xHx HC105 Gamme de mesure 0 10096 HR Erreur incluant hyst r sis non lin arit et reproductibilit 15 40 C lt 90 HR 1 3 0 3 valeur mesur e HR 15 400C gt 90 HR 2 3 HR 25 70 C 1 4 1 valeur mesur e HR 40 120 C 1 5 1 5 valeur mesur e HR Influence de la temp rature sur l lectronique typ 0 015 HR C Temps de r ponse avec filtre en grille m t
54. und durch das Symbol CAL im LCD Display angezeigt TN oder TASTER S2 Kalibrationspunkt 5096 r F 5096 vom Temperaturmessbereich Durch mind 3 sek langes Dr cken wird die Routine gestartet Der Kalibrationsmode wird durch das Leuchten der LED D1 und durch das Symbol CAL lt im LCD Display angezeigt 4 TASTER 1 up und S2 down Durch Dr cken der beiden Taster wird der s2 8 Messwert in 0 1 0 1 C Schritten auf den Referenzwert abgeglichen Der Messwert kann entweder am Display abgelesen oder am Ausgang gemessen werden 5 TASTER S1 Durch mind 3 sek langes Dr cken wird der Kalibrationswert gespeichert und die Routine verlassen Das Verlassen des Kalibrationsmodes wird D1 Ka durch das Deaktivieren der LED D1 und dem Symbol CAL im LCD Display angezeigt TASTER S2 Durch mind 3 sek langes Dr cken wird die Routine verlassen ohne die Kalibrationswerte zu speichern Das Verlassen des Kalibrationsmodes wird durch das Deaktivieren der LED D1 und dem Symbol CAL im LCD Display angezeigt R cksetzten der Kundenkalibration auf die Werkskalibration jeweilige Jumpereinstellung gew hlt werden Werden ausserhalb des Kalibrationmodus beide Tasten gemeinsam mindestens 5 sek lange gedr ckt wird die Kundenkalibration wieder auf die Werkskalibration zur ckge setzt Optisch wird die Zur ckstellung auf die Werkskalibration durch ein kurzes Aufleuchten der LED D1 angezeigt TASTER S1 und S2 Um die RH oder T Werte r
55. une remarque Ces consignes doivent tre respect es afin d obtenir un fonctionnement optimal de l appareil Consignes de s curit Toute sollicitation m canique extr me et incorrecte doit tre vit e Lors du d vissage du filtre veiller ne pas endommager le capteur de mesure es mesures de protection li es aux risques d lectricit statique doivent tre respect es lors de tout contact avec le capteur de mesure Connecter le capteur aux fils de raccordement uniquement Le montage le raccordement lectrique la maintenance et la mise en service ne doivent tre effectu s que par du personnel qualifi Consignes de s curit pour l utilisation du module relais avec une alimentation gt 50V Pour isoler le module d alarme optionnel du c t basse tension du transmetteur la cloison pr vue cet effet doit tre install e en partie basse Le boitier de l appareil doit tre enti rement ferm lors du fonctionnement e L indice de protection d un boitier ouvert correspond un IPOO Tout travail sur des pi ces conductrices doit tre effectu par des personnes exp riment es Consignes de s curit pour l utilisation de l alimentation int gr e option V01 Le boitier de l appareil doit tre enti rement ferm lors du fonctionnement de l ap pareil Lindice de protection d un boitier ouvert correspond un IPOO Tout travail sur des pi ces conductrices doit tre effectu par des person
56. urrent o 2 HE WB E e e zm 3 LED 4 Taster 5 Level Potentiometer Hysteresis 6 Hysteresis Potentiometer 1 Jumper Auswahl Ausgangsbereich Spannungsausgangssignale Stromausgangssignale oobocooo Current 5 200000 Current o e e o o olo Voltage oo KEE Voltage lolo ERSTE oo 2 Display Steckplatz f r optionales Display 3 LED D1rot leuchtet w hrend der Kalibrations Routine andauernd leuchtet bei R cksetzung auf Werkskalibration kurz auf D2 gr n blinkt w hrend des Betriebs leuchtet bei defektem Feuchte Temperatur Sensor 4 Taster S1 Taster f r 1 Pkt Feuchte Temperaturkalibration Feuchte gt 50 r F Taster f r 2 Pkt Feuchte Temperaturkalibration oberer Kalibrationspunkt Taster zum Speichern der Kalibrations Werte S2 Taster f r 1 Pkt Feuchte Temperaturkalibration Feuchte 50 r F Taster f r 2 Pkt Feuchte Temperaturkalibration unterer Kalibrationspunkt Taster zum Verlassen der Kalibrations Routine ohne die Werte zu speichern S1 S2 f r R cksetzung auf Werkskalibration 5 Level Potentiometer Einstellen des Schaltpunktes nur bei EE23 mit Schaltmodul 6 Hysteresis Potentiometer Einstellen der Hysterese nur bei EE23 mit Schaltmodul Angaben in mm geh use Polycarbonat SNS INSTALLATION Montage fur Metall bzw Polycarbonatgehause 1 Monta
57. with EA 4 02 and with regard to GUM Guide to the Expression of Uncertainty in Measurement 3 Refer to ordering guide 4 Refer to accuracies of calculated values 5 Connection plugs refer to ordering guide 5 N N f HYSTERESIS Einschaltpunkt Switching on 7 do 50 gt Schaltpunkt switching point 62 70 23 gt GENERALITES Le manuel d utilisation est un l ment de la livraison standard et assure une manipulation appropri e et un fonctionnement optimal de l appareil Pour cette raison le manuel d utilisation doit tre consult avant toute mise en service De plus ce manuel d utilisation doit tre transmis chaque personne intervenant pour le transport la mise en service l utilisation la maintenance et la r paration Ce manuel d utilisation ne peut tre utilis des fins concurrencielles ou transmis un tiers sans accord crit pr alable de E E Elektronik Toute copie pour les besoins de l utilisateur est autoris e Les informations contenues dans ce manuel les donn es et sch mas techniques sont bas s sur les informations disponibles un temps donn Explication des symboles Ce symbole indique une consigne de s curit Les consignes de s curit doivent absolument tre respect es Leur non respect peut provoquer des blessures aux utilisateurs ou l endommagement des mat riels E E Elektronik n assume alors aucune responsabilit Ce symbole indique