1. If audial signal is heard twice the Floor Lock is probably deactivated Only Trendelenburg and backrest adjustments are possible when Floor Lock is deactivated Correct with activating the Floor Lock The chosen adjustment movement continues as long as the button is pressed or till the end position of the movement is reached The movement can also stop before the desired position because some combinations of ad justments are resitricted with table electronics Audial signal informs this situation Preset positions ReFlex Flex Beach Chair and 0 preset Press relevant function button and the table starts the automatic positioning in phases First side tilt to full horizontal then Trendelenburg Slide back rest and leg section in phases This order can vary depending on the starting position of the table Preset position is ready when audial signal is heard Search procedure can be stopped by releasing the button Use extreme caution when using preset positions with patient on the table Adjusting the divided leg section Press first L or R button and after that simultaneously Leg section Up or Leg section Down button Devided leg section can be adjusted parallel by pressing L and R buttons simultaneously Use extreme caution when adjusting divided leg section with patient on the table Back Light button Back Light illuminates the hand control unit for 5 seconds when activated or for 5 seconds
2. 4 pcs carefully without damaging the castors Place the top of the package as shown in below to form a ramp for the table and transfer the table on to the floor WARNING Be careful table weights about 300 kg 660 Ibs If the table has been in cold temperatures during transport or storage allow it to warm up to room tempera ture at least 6 hours before recharging the battery or connecting power This ensures that the possible con densed humidity has time to evaporate N Recharge the battery prior to use See the starting instructions attached to table base panel Cardboard packing material should be recycled Wood and plastics are energy waste e 14 3 2 Construction 3 2 1 Structural elements Head section 25 section Seat section 40 section 25 section 49 gyn section Main switch 4 and appliance inlet 3 2 2 Product symbol and label markings Mains power connection and fuses Voltage range 100 240 V Alternating Current AC AC frequency range 47 63 Hz Mains fuses T6 3AL 250 V 15 Equipotential connector Main switch O OFF All table functions will be ceased when toggle is turned to OFF position Protective grounding Placed inside the table base Foot switch PC and hand control unit connections Foot Switch USB Hand Control Unit Foot control unit USB connection Remote control Warning label for
3. Audial signal during override adjustment FUNCTION BUTTONS AND INDICATORS 5th wheel ON Floor Lock ON Reverse mode Activate button Floor Lock OFF 5th wheel OFF Normal mode ADJUSTING THE TABLE WITH OVERRIDE PANEL Press and keep pressed some of the function buttons Press the Activate button movement starts Stop the movement by releasing both buttons Corresponding adjustment button symbols are presented on the page 23 28 AN AN AN AN AN 3 2 11 Foot control unit When adjusting the operating table take care that the fingers hands or other parts of the body of the patient are not placed between edges of back leg or seat section frames and pivoting points Ensure before adjusting the operating table that table top will not hit other obstacles Ensure before adjusting the operating table that NORMAL REVERSE mode has been chosen correctly and is the same as patient s orientation on the table top After use of main switch NORMAL mode is set as default Lors du r glage de la table d op ration viter de coincer les doigts les mains ou toute autre partie du corps du patient entre les bords des cadres du dossier support pour membres inf rieurs ou du si ge ou aux endroits de rotation Avant de r gler la table d op ration v rifier que le plateau de la table ne risque pas de heurter d autres obstacles Avant de r gler la table d op ration v rifier que le mode NORMAL
4. from the release of last used function button Memory buttons M1 and M2 Adjust the table to the desired position Press R button and M1 or M2 button simultaneously for 2 seconds The desired position is saved to table s MPC audial signal when OK When needed saved position can be recalled by pressing M1 or M2 button Functioning similar to preset positions 26 D pp gt ADJUSTING THE TABLE WITH IR HAND CONTROL UNIT IR control units are individually coded to operate with one specific table Using of IR hand control unit is similar to normal hand control unit Indicator lights for battery condition Floor Lock 5th wheel and NORMAL REVERSE mode are not available in IR control unit Indicator lights for these are on override panel if needed Aim the IR hand control unit towards the black round IR receivers on the table column casings and press needed function button When pressing function button the IR indicator light will flash showing the IR battery condition Green Battery OK Red Battery capacity lowered replace with DURACELL ULTRA M3 LR6 AA 1 5V 4 pcs 3 2 9 Software controlled limits It is possible to adjust the SLIDE adjustment only when back or leg section are in horizontal or above hori zontal position Slide adjustment will stop automatically before collision if back or leg section is under the horizontal position SLIDE adjustment is only possible when table top is in horizontal p
5. 31 o o gt gt gt Batteries contain lead and if replaced they need to be recycled in accordance with local environmental regu lations It s recommended to recharge the battery overnight after day s use Table will be always ready for use and the battery will have a longer service life 3 2 14 Equipotential connector The connector for potential equalization is placed at the table base It s recommended to utilize potential equalization always when using patient monitoring equipment Consult manufacturers instructions Use potential equalization conductor with patient monitoring equipment Connector placed at the table base Utiliser la barre d quipotentiel lorsque l on se sert d un dispositif de monitorage du patient Connecteur plac sur le socle de la table Conductor for equalization available on request Code 7134261 cable 5 m 3 2 15 Manual use of floor lock Knob for floor lock manual release is placed at the rear end of the table base Manual release is needed if floor lock can t be deactivated with hand control unit or override panel Turn the knob 90 counterclockwise and the table will descend on to its castors Time to descend is about 5 seconds Turn the knob back to its normal position after releasing procedure Operating table can be nor mally transported to service or battery recharging 3 2 15 1 Location of the releasing knob Releasing knob 32 o
6. 400 Ibs n mode INVERSE le poids max du patient est de 180 kg 400 Ib In NORMAL mode max patient weight with sliding seat section is 275 kg 600 Ibs En mode NORMAL le poids max du patient avec le si ge coulissant est de 275 kg 600 Ib In NORMAL mode max patient weight with fixed seat section is 325 kg 700 Ibs En mode NORMAL le poids max du patient avec le section fixe de si ge est de 325 kg 700 Ib Adjust the table top to the horizontal position 0 position with the hand control unit before transporting a patient on the operating table R gler le plateau de la table en position horizontale position 0 avec la commande manuelle pour trans porter un patient sur la table d op ration Transporting of patients weighing up to 180 kg 400 Ibs is only allowed when table top is in horizontal posi tion 0 position and the height of the table is max 900 mm 3 ft Le transport de patients de 180 kg 400 Ib au maximum est seulement autoris lorsque le plateau de la table est en position horizontale position 0 et la hauteur de la table est de 900 mm 3 ft au maximum Use extreme caution when transporting the table with a patient on Transporting of the table with a patient on requires two persons To maximize patient safety utilize proper restraint methods during transport Faire preuve d une extr me prudence lors du transport d un patient sur la table Le transport d un patient sur la table exige la
7. CSA C22 2 No 601 1 M90 and EN 60601 1 2 EMC The table complies with directive 93 42 EEC MDD product class I 2 1 Intended use Operating table is intended for general surgical use It is also well suited for Day surgery ENT surgery Plastic surgery Orthopedics Arthroscopy Gynecology Urology Pediatrics Neurosurgery Expertise is essential The patient is the most important part of treatment This is precisely why the equipment used in treatment must be absolutely safe and convenient to use As a health care professional you deserve the very best tools allowing you to concentrate on your own field of expertise We are experts in providing hospital equipment Products have been designed to function efficiently and flexibly during the various stages of treatment They assist you in the performance of your work without distracting you from the task at hand Our modular equipment system includes state of the art equipment for hospital procedures and hospital room environments as well as for nursing homes and home care applications For more information about the products contact our Sales Office For matters related to equipment servicing contact Technical Service 12 2 2 Operating characteristics This medical device is transportable electro hydraulic battery and mains operated operating table It is micro processor controlled and its function parameters are reprogrammable with special PC o
8. THE HAND CONTROL UNIT Open the dust cap of the receptacle and insert the connector carefully Note the red dots on the connector and the receptacle They must be aligned before inserting DISCONNECTING THE HAND CONTROL UNIT Support the receptacle and pull downwards from the connector cover FUNCTION BUTTONS AND SIGNAL LIGHTS ON OFF Battery condition indicators IR control unit signal lights Floor Lock OFF Floor Lock ON 5th wheel OFF 5th wheel ON Reverse mode Normal mode ReFlex preset Flex preset 0 position Beach Chair preset Rev Trendelenburg Trendelenburg Height adjustment UP Height adjustment DOWN Side tilt Left Side tilt Right Back section UP Back section DOWN Leg section UP Leg section Down Left leg section Right leg section separately separately Forward Slide Reverse Slide Memory 1 Memory 2 Back Light 24 ADJUSTING THE TABLE WITH CABLED HAND CONTROL UNIT Hand control unit is normally in stand by mode to save the battery Function buttons of the hand control unit are not active to protect the table against accidental actuating No movement of the table occurs if buttons are pressed Activate the table by pressing the ON OFF button Battery condition indicator lights up green yellow or red Hand control unit is ready for use If the battery indicator does not light up the main switch of the table is in OFF position Check and try again Deactivating
9. patient transport WARNING Consult user manual for right transporting of patient AVERTISSEMENT Consulter le manuel d utilisation pour un transport correct du patient VARNING Konsultera anv ndarmanualen f r korrekt transport av patienter Pinch point warning label Keep hands and fingers clear during the head section adjustment Etiquette de point de pincement Garder vos mains loing de cette surface 16 3 2 3 Permitted table top configurations Normal Reverse With sliding seat section Max 275 kg With sliding seat section Max 180 kg With fixed seat section Max 325 kg With fixed seat section Max 180 kg gt Head section Normal With sliding seat section Max 225 kg Head section attached to 40 section With fixed seat section Max 325 kg Head section Gynecologic With sliding seat section Max 180 kg With fixed seat section Max 180 kg Reverse With sliding seat section Max 225 kg Head section attached to 40 section With fixed seat section Max 325 kg Table top can be assembled in several different configurations Figures above show the permitted con figurations and relevant max patient load gt PPP PP DP DP gt gt 3 2 4 Transporting a patient on the table Place the patient in the longitudinal center line of the table top Placer le patient sur la ligne m diane longitudinale du plateau de la table In REVERSE mode max patient weight is 180 kg
10. pr sence de deux personnes Pour optimiser la s curit du patient utiliser des m thodes de retenue appropri es pendant le transport 18 Transporting of patients weighing 180 225 kg 400 500 Ibs is only allowed when the table is in Normal mode in horizontal position O position and in the lowest position Le transport de patients de 180 225 kg 400 500 Ib est seulement autoris lorsque la table est en mode nor mal en position horizontale position O et dans la position la plus basse Transporting of patients weighing 225 325 kg 500 700 Ibs is not allowed Le transport de patients de 225 325 kg 500 700 Ib n est pas autoris gt gt 3 2 5 Attaching and removing sections All the table sections except seat section in operating table are detachable and removable You can arrange the sections as you wish depending on the surgical concept and operation Normally it s not needed to remove the 40 sections Check that table top sections latches lock correctly Incorrect attachment of table sections is dangerous and can cause personal injury or equipment damage V rifier que les verrous des sections du plateau de la table se ferment correctement La fixation incorrecte des sections de la table est dangereuse et risque de provoquer des blessures corporelles ou des dommages mat riels Use only permitted table top configurations illustrated on page 17 in this manual N utiliser que les configurations
11. syst mes d immobilisation et de r glage des accessoires fonctionnent correctement Lors du r glage de la table d op ration viter de coincer les doigts les mains ou toute autre partie du corps du patient entre les bords des cadres du dossier du support pour membres inf rieurs ou du si ge ou aux endroits de rotation N utiliser que les configurations autoris es du plateau de la table illustr es dans ce manuel Avant de r gler la table d op ration v rifier que le plateau de la table ne risque pas de heurter des obstacles externes Avant de r gler la table d op ration v rifier que le mode NORMAL INVERS a t correctement choisi et qu il correspond l orientation du patient sur le plateau de la table Apr s l utilisation de l interrupteur principal la table est en mode NORMAL par d faut Toujours suivre les instructions du fabricant lors de l utilisation d quipement de diathermie ou de d fibrillation AVERTISSEMENTS MISES EN GARDE ET REMARQUES Lors de l utilisation d un g n rateur chirurgical a haute fr quence viter que le patient n entre en contact avec des pi ces m talliques de la table d op ration ou des accessoires Ne pas placer le patient sur une surface mouill e ou humide ou sur des rembourrages conducteurs d lectricit Risque de br lures Ne pas utiliser de gaz anesth sique combustible avec la table d op ration Utiliser la barre d quipotentiel lorsque l on se sert d
12. un dispositif de monitorage du patient Connecteur plac sur le socle de la table Pour recharger la batterie d abord brancher le cordon d alimentation secteur sur la prise de la table puis sur la prise murale L interrupteur peut servir de syst me d arr t d urgence exig par les normes Normalement le levier de l interrupteur est en position haute Lorsque le levier est abaiss en position d arr t toutes les fonctions de la table sont annul es Le mouvement de la table cesse imm diatement Lorsque la batterie est enti rement recharg e d abord d brancher le cordon d alimentation secteur de la prise murale puis de celle de la table D brancher le cordon d alimentation secteur de la table et mettre l interrupteur principal en position d arr t avant de nettoyer la table D brancher le cordon d alimentation secteur de la table avant les proc dures d entretien Tension de 100 240 V utilis e sur le bloc d alimentation plac dans le socle de la table Risque de choc lectrique La table d op ration a t test e conform ment la norme EN60601 1 2 afin d en v rifier la compatibilit lectromagn tique Les autres produits utilis s proximit de la table d op ration doivent aussi tre conformes cette norme S ils ne sont pas conformes des interf rences risquent de se produire entre les produits et de provoquer des r actions inattendues Contacter le fabricant concern en cas de probl me L
13. 4 TECHNICAL DATA 4 1 Identification plate Identification plate is located on the bottom surface of the left side seat section tube Consult accompanying documents MANUFACTURED IN FINLAND FOR 5368 Kuhl Road Erie PA 16510 ep Tel 800 663 1152 MUVO Fax 814 899 1410 Maximum power NUVO ING www nuvosurgical com PRODUCT NUMBER 100060000 MAX POWER 700 VA Primary voltage and frequency SERIAL NUMBER 111011 12345 INPUT VOLTAGE 100 240 V MFG DATE 11 10 11 YY MM DD FREQUENCY 47 63 Hz P T NAME NUVO V7 DUTY CYCLE ED 20 Manufacturing date Ter cuassiriED PROTECTION CLASS IP X4 pA WEIGHT 325 kg e X DR 6001bs 275k9 Ingress Protection Splash proof Intertek 3196120 Safe max working load SWL B type applied part ETL mark IEEE Before use check the actual product information printed on the installed identification plate Consult accompanying documents We FOR 5358 Kits Road Gite PA 10510 IL panying iid scd aue rt Cli Maximum power PRODUCT NUMBER 100060000 MAX POWER 700 VA 4 Primary voltage and frequency SERIAL NUMBER 111011 12345 INPUT VOLTAGE 100 240 uM MFG DATE 11 10 11 YY MM DD FREQUENCY 47 63 Hz NUVO V7 B3 DUTY CYCLE ED 20 i PROTECTION CLASS IP X4 Manufacturing date ET CLASSIFIED OTECTION CLASS x 4 AT i ae Ingress Protection e us a A Splash proof 3196120 3196120 24 Safe max working load SWL Mark of ETL certifica
14. BLEM Back section moves when leg section button is pressed or vice versa PROCEDURE CAUSE Normal Reverse mode wrongly selected Check the mode PROBLEM Table not responding to IR hand control unit commands PROCEDURE CAUSE Control unit wrongly aimed Aim towards the table Distance between table and control Check the distance IR receiver blocked Check IR control unit battery depleted Replace IR control unit wrongly coded Set code again PROBLEM One piece leg section can not be adjusted PROCEDURE CAUSE One piece leg sections not attached to seat section correctly Check the latches Faulty inductive sensor Contact service PROBLEM Hydraulic unit starts but no movement occurs PROCEDURE CAUSE Valve block solenoid defective or loose Contact service Table adjustment in its end position Check the table posi tioning CONTACT Further information of operating table service advice and spare parts available from your local dealer or from Nuvo inc 800 663 1152 814 899 4220 Fax 814 899 1410 Technical service E mail Internet service nuvosurgical com www nuvosurgical com 800 663 1152 ext 60 814 899 4220 44 9 USER GUIDANCE FOR EMC 9 1 Guidance and manufacturer s declaration Operating table has been tested to IEC EN 60601 1 2 to ensure proper electromagnetic compatibility Porta ble and mobile RF communication
15. Electromagnetic environment guid ance Electrostatic dis charge ESD IEC EN 61000 4 2 Radiated radio fre quency electro magnetic field immunity test IEC EN 61000 4 3 6 kV contact 8 kV air 3 V m 80 2500 MHz 6 kV contact 8 kV air 3 V m 80 2500 MHz The antistatic properties of the table are depending on the use of original mattress set and the use of conduc tive floor material Portable and mobile RF communica tions equipment can affect opera ting table and should be used no closer to any part of operating table including cables than the recom mended separation distance calcu lated from the equation applicable to the frequency of the transmitter Recommended separation distance 1 2 P 2 3 JP where P is the maximum output power rating of the transmitter in watts W according to the transmit ter manufacturer and d is the recom mended separation distance in metres m d 80 MHz 800 MHZ d 800 MHz 2 5 GHz Fields strengths from fixed RF trans mitters as determined by an electro magnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with symbol t Table 2 Electromagnetic immunity part 1 46 Electrical fast transient burst immunity test IEC EN 61000 4 4 2 kV for power supply lines 2 kV for power supply lines Mains pow
16. INVERS a t correctement choisi et qu il correspond l orientation du patient sur le plateau de la table Apr s l utilisation de l interrupteur princi pal la table est en mode NORMAL par d faut Foot switch is connected to the receptacle placed on the OR panel on the left side of the column CONNECTING THE FOOT SWITCH Open the dust cap of the receptacle and insert the connector carefully Note the red dots on the connector and the OR panel receptacle They must be aligned before inserting DISCONNECTING THE FOOT SWITCH Support the OR panel receptacle and pull downwards to the connector cover Model 60850 Connector Side tilt Slide or back section adjustment 29 gt gt gt gt ADJUSTING THE TABLE WITH FOOT SWITCH Activate the table by pressing ON OFF button at hand control unit Press the desired movement pedal twice Movement starts only after second pressing Movement stops when pedal is released Single pressing will start the movement if the switch is pressed again in 5 seconds 3 2 12 Main switch Main switch can be used as an emergency stopping device demanded by the standards Normally the Switch lever is in up position When the toggle is turned down in OFF position all table functions are ceased Table movement stops immediately L interrupteur principal peut servir de syst me d arr t d urgence exig par les normes Normalement le levier de l interrupteur est en positio
17. Ne pas utiliser d accessoires us s ou endommag s Lors de l utilisation de la table s assurer que tous les accessoires sont correctement mont s et que les syst mes d immobilisation et de r glage des accessoires fonctionnent correctement Les accessoires de la table sont pr sent s dans le manuel Compatibilit garantie 37 gt gt gt p gt e 6 CLEANING 6 1 Cleaning instructions Disconnect the mains cord from the table and turn the main switch to OFF position before cleaning of the table Operating table has been classified as a splash proof equipment Use of pressurized water is not allowed Cleaning and disinfecting should be made according to this manual Do not pull or push the shields of back and leg section joints Check during adjustment that there is no foreign obstacles between the shields Surface of the operating table mattresses is electrically conductive Cleaning of the mattresses should be done with neutral detergent pH 7 8 and warm water applied with a single use wipe 6 2 Cleaning Remove accessories and mattresses Clean by wiping down with a mild alkaline detergent pH 7 8 6 3 Disinfection Remove accessories and mattresses Disinfect only when necessary Wipe down the equipment with the surface disinfectant used at the facility in accordance with manu facturer instructions unless the surface disinfectant contains phenols and alcohol which can corrode plas
18. USER MANUAL Nuvo V7 operating table Sif Document DO1123 2 0 en Version DRAFT 2 0 11 10 2011 muvo A member of the Medical Illumination International Group 5368 Kuhl Road Erie PA 16510 4703 800 663 1152 814 899 4220 Fax 814 899 1410 www nuvosurgical com WARNINGS CAUTIONS AND NOTES GENERAL 2 1 Intended use 2 2 Operating characteristics PRODUCT USE 3 1 Unpacking 3 2 Construction 3 2 1 Structural elements 3 2 2 Product symbol and label markings 3 2 3 Permitted table top configurations 3 2 4 Transporting a patient on the table 3 2 5 Attaching and removing sections 3 2 6 Integrated kidney bridge back section 3 2 7 Mattresses 3 2 8 Hand control units 3 2 9 Software controlled limits 3 2 10 Override panel 3 2 11 Foot control unit 3 2 12 Main switch 3 2 13 Recharging of battery and mains use 3 2 14 Equipotential connector 3 2 15 Manual use of floor lock 3 2 15 1 Location of the releasing knob TECHNICAL DATA 4 1 Identification plate 4 1 1 Labeling and symbols 4 1 2 Surface materials 4 2 Specifications 4 2 1 Environmental specifications 4 2 2 Electrical specifications 4 2 3 Classification data 4 2 4 Adjustment ranges 4 2 5 Weights and dimensions 12 12 13 14 14 15 15 15 17 18 19 21 22 23 27 28 29 30 31 32 32 32 33 33 34 34 35 35 35 35 36 36 ACCESSORIES 5 1 Operating table accessories CLEANING 6 1 Cleaning instr
19. allowed only to persons specialized to operating table service work Seules les personnes sp cialis es dans l entretien de la table sont autoris es en effectuer l entretien Disconnect the mains cord from the table before any service procedures 100 240 V is used power unit placed in table base Hazard of electric shock Turn the main switch to OFF position before connecting or disconnecting any of electronic connections or components in order to protect them against voltage surges Static charges can cause sparks harming sensitive electronic components Ground yourself to metallic parts of table before touching electronic components Disconnecting any of the hydraulic hoses can cause uncontrolled and dangerous movement of the table Support the table parts before disconnecting any hydraulic hoses When mains cord is connected to operating table the cooling fan of the power unit is activated causing a humming sound After disconnecting the mains cord the cooling fan still runs for few seconds D brancher le cordon d alimentation secteur de la table avant les proc dures d entretien Une tension de 100 240 V est utilis e sur le bloc d alimentation plac dans le socle de la table Risque de choc lectrique Mettre l interrupteur principal en position d arr t avant de brancher ou de d brancher des composants lectroniques afin de les prot ger des surtensions Les charges lectrostatiques peuvent provoquer des tincelle
20. autoris es du plateau de la table illustr es la page 17 de ce manuel gt gt gt gt 19 A ATTACHING Max 3 sections on the one side of the column head section included First attach the 40 sections to both sides of the seat section Insert the guide pins of the 40 sections to the holes of seat section joints When the attached section is fully inserted the locking latches will snap into place 25 sections or head section will be attached to the holes of 40 sections For safety reasons it is important to check that both latches are properly locked A REMOVING Pull locking latches outwards on both sides Hold the section which is to be removed with both hands and pull section towards yourself 40 section 25 section Locking latch 20 3 2 6 Integrated kidney bridge back section Factory assembled kidney bridge back section adds more versatility to use the operating table on thorax area procedures with simple adjustment of an adjustment crank handle Become familiar using Kidney bridge and body elevator before operation Table top can be constructed normally with desired configura tion Crank handle fixing Kidney bridge and body elevator system ER Crank handle for kidney bridge adjustment Insert the crank handle to the kidney bridge adjusting pin located either side of the back section Check that handle is fully inserted before adjusting Start adjusting slowly counterclockwis
21. cording to this manual Do not pull or push the shields of back and leg section joints Check during adjustment that there is no fo reign obstacles between shields It is possible to adjust the SLIDE adjustment only when back or leg section are in horizontal or above hori zontal position Slide adjustment will stop automatically before collision if back or leg section in under the horizontal position SLIDE adjustment is only possible when table top is in horizontal position Use of SLIDE adjustment is restricted with tilted table top Mattresses should be cleaned only with neutral detergent ph 7 8 The antistatic properties of the operating table require the use of original brand mattresses and antistatic flooring 10 9 WARNINGS CAUTIONS AND NOTES Simultaneous max angles of Trendelenburg reversed Trendelenburg and side tilt are limited with software Turn the main switch to OFF position before connecting disconnecting any electronical components in order to protect them from voltage surges NOTES It s recommended to recharge the battery overnight after day s use Table will be always ready for use and the battery will have longer service life If the main switch is turned to OFF position downwards all table functions are blocked including recharg ing of the battery Check that the green led above the charging inlet lights up when you connect the mains cord to the table When mains cord is conn
22. e to rising the angle Adjust until the proper angle is reached Maximum height is about 150 mm Back section angle can be adjusted normally from the hand control unit or back up control panel Long com bined seat and back section mattress is used with the kidney bridge back section system A WARNING Notice labeling of pinching risk which may occur If adjustment is made carelessly Keep the crank handle safe place and remove before operation and when operating table is transported 21 3 2 7 Mattresses Surface of the mattresses is electrically conductive Cleaning of the mattresses should be done with neutral detergent pH 7 8 and warm water applied with a single use wipe Moulded mattresses are fixed to the table top with pins Place the mattress on the top plate so that the pins and holes are contacting Press mattress slightly to lock it in place To remove the mattress lift it upwards from the edge Other mattresses are fixed with Velcro tape to table top plates Velcro tapes on table top plates can be cleaned with pressurized air Velcro tape can also be replaced as spare parts Instructions available from Technical Service To remove the mattress lift it upwards from the edge Press mattress slightly to lock it in place Fixing pin The antistatic properties of the table are depending on the use of the original mattress set and the use of con ductive floor coating Velcro tapes on table top plates ca
23. eau de la table est en position horizontale position 0 et la hauteur de la table est de 900 mm 3 au maximum Le transport de patients de 180 225 kg 400 500 Ib est seulement autoris lorsque la table est en mode NORMAL en position horizontale position O et dans la position la plus basse Le transport de patients de 225 325 kg 500 700 Ib n est pas autoris Faire preuve d une extr me prudence lors du transport d un patient sur la table Le transport d un patient sur la table exige la pr sence de deux personnes Pour optimiser la s curit du patient utiliser des m thodes de retenue appropri es pendant le transport AVERTISSEMENTS MISES EN GARDE ET REMARQUES Afin d viter d endommager l appareil respiratoire le syst me nerveux et le syst me circulatoire du patient le patient doit tre correctement positionn et maintenu sous observation V rifier que les verrous des sections du plateau de la table se ferment correctement La fixation incorrecte des sections de la table est dangereuse et risque de provoquer des blessures corporelles ou des dommages mat riels Utiliser uniquement les accessoires recommand s par le fabricant de la table Lors du r glage de la table d op ration viter les collisions entre les accessoires et la table d op ration Ne pas utiliser d accessoires us s ou endommag s Lors de l utilisation de la table s assurer que tous les accessoires sont correctement mont s et que les
24. ected to operating table the cooling fan of the power unit is activated causing a humming sound After disconnecting the mains cord the cooling fan still runs for few seconds Turn the main switch to OFF position downwards when the table is not in use or stored for longer period of time This eliminates the possibility of unintended use and is also saving the battery capacity Activate the floor lock before adjusting the table Only trendelenburg and back rest adjustments are function ing if floor lock is not activated Cardboard packing material should be recycled Wood and plastics are energy waste Batteries contain lead and if replaced need to be recycled in accordance with local environmental regula tions Velcro tapes on table top plates can be cleaned with pressurized air Velcro tape can also be replaced spare parts and instructions available from your sales representative Floor lock can be deactivated with a release knob placed at the table base if normal deactivating with hand control unit or override panel can t be done Turn the knob 90 clockwise and the table will descend on to its castors Turn the knob back to its original position after releasing 11 2 GENERAL Operating table is designed to conform ISO 9001 Quality Management and ISO 14001 Environmental Man agement System Standards In addition the operating table meets the following standards for medical devices EN 60601 1 EN 60601 2 46 UL 60601 1 CAN
25. er quality should be that of a typical hospital environment Table is mainly operated with inter nal battery Surge immunity test IEC EN 61000 4 5 Line to line 1kV Line to ground 2kV Line to line 1kV Line to ground 2kV Mains power quality should be that of a typical hospital environment Table is mainly operated with inter nal battery Immunity to conducted disturbances induced by radio frequency fields IEC EN 61000 4 6 0 15 80 MHz 3 Vrms 0 15 80 MHz 3 Vrms Portable and mobile RF communi cations equipment can affect ope rating table and should be used no closer to any part of operating table including cables than the recommended separation distance calculated from the equation appli cable to the frequency of the trans mitter Recommended separation dis tance d 1 24P where P is the maximum output power rating of the transmitter in watts W according to the trans mitter manufacturer and d is the recommended separation distance in metres m Fields strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compli ance level in each frequency range Interference may occur in the vicin ity of equipment marked with sym bol Ca Power frequency mag netic field immunity test IEC EN 61000 4 8 3A m 3A m Operating table can be used together with medical devices complying IEC EN 60601 1 2 Voltage dips
26. es quipements de communication RF portables et mobiles peuvent provoquer des interf rences avec la table d op ration WARNINGS CAUTIONS AND NOTES N CAUTIONS If the operating table has been in the cold allow it to warm up at room temperature for at least 6 hours before recharging the battery or switching on to allow any condensation formed to evaporate Recharge the battery prior to use See the starting instructions attached to table base panel OVERRIDE panel is only for returning the table top to the horizontal position in the case of electronic mal function Override makes it possible to adjust table without the controlling processor and any of the pro grammed restrictions are not functioning Downwards adjusted leg or back section may hit the table base or column casings depending on the used height or tilt angle adjustments Can cause damage to the operating table and creating pinching hazard Do not place any objects on the base of the operating table Danger of damage during adjustments The best place to store the hand control unit especially during transport is the lever under the head section Head section gas spring can be disposed of as metal waste after nitrogen gas and oil has been removed Instructions for gas releasing available from your sales representative Operating table has been classified as splash proof equipment Use of pressurized water is not allowed Cleaning and disinfecting should be made ac
27. est Std 4 2 5 Weights and dimensions SLIDING SEAT FIXED SEAT 650 1110 mm 610 1070 mm 410 mm 25 20 45 80 45 80 105 70 95 70 35 35 45 45 45 45 TOLERANCE 0 10 mm 5 mm 0 10 mm 2 5 4 1 1 4 4 1 1 4 2 5 0 5 0 5 SLIDING SEAT SECTION FIXED SEAT SECTION Table top 6 sections 6 sections Weight of table 325 kg 300 kg Length A 2100 mm 2100 mm Width B 610 mm 610 mm Height C 650 1110 mm 610 1070 mm Lenght of the base D 1200 mm 1200 mm Table top width without rails E 540 mm 540 mm Table top width with rails F 594 mm 594 mm Floor clearance G 25mm 0 3mm 25mm 0 3mm Castors 125 mm 125 mm Longitudinal Shift 410 mm Distance between the castors H 1100 mm 1100 mm 36 gt gt gt gt gt 5 ACCESSORIES 5 1 Operating table accessories Use only accessories recommended by the table manufacturer When adjusting the operating table take care to avoid collision between accessories and the operating table Do not use worn or damaged accessories When using the table ensure that all accessories are properly mounted to it and check the function of accessory lockings and adjustments Accessories for operating table listed in separate accessory brochure Ensure compatibility
28. l equipment take care to prevent the patient coming into contact with metal parts of the operating table or accessories Do not place the patient on wet or damp surfaces or electri cally conductive pads Hazard of burn injuries Combustible anaesthetic gases must not be used with the operating table Use potential equalization conductor with patient monitoring equipment Connector placed at the table base When recharging the battery first plug the mains cord to table appliance inlet after that to wall socket Mains switch can be used as an emergency stopping device demanded by the standards Normally the switch lever is in up position When the lever is turned down in OFF position all table functions are cancelled Table movement stops immediately When recharging is completed Disconnect the mains cord first from the wall socket after that from the table appliance inlet Disconnect the mains cord from the table and turn the main switch to OFF position before cleaning of the table Disconnect the mains cord from the table before any service procedures 100 240 V used in power unit placed in table base Hazard of electric shock Operating table has been tested to EN60601 1 2 to ensure proper electromagnetic compatibility Other prod ucts used in the vicinity of operating table should also comply with this standard If they do not comply inter ference between products in unintended responses may occur Please contact the appropriate ma
29. ld be done according to the work instruction T401932 Releasing nitrogen gas is strictly prohibited without following the proper instructions Gas spring dismantling instructions are available from your sales representative AVERTISSEMENT Le d gagement d azote sans se conformer aux instructions correspondantes est strictement interdit Les ressorts pneumatiques contiennent de l azote sous pression qui doit tre vacu conform ment la consigne de travail T401932 Votre repr sentant commercial vous transmettra les instructions relatives l limination correcte des ressorts pneumatiques NOTE Maintenance free sealed lead acid batteries used in operating table are environmentally hazardous and should be disposed of according to local waste regulations 7 1 2 Oil Hydraulic oil used in operating table ARAL Vitam DE 32 or BP Energol HLP HM 32 Hydraulic oil should be disposed of according to local waste regulations Avoid unnecessary skin contact with oil 39 7 1 3 Electronic waste and batteries Electronic components and devices should be disposed of according to local waste regulations Symbol informs that the product contains electronic devices and cannot be disposed with general waste If so the product must be recycled separa tely and cannot be disposed of along with general waste 40 gt gt gt PR gt gt EEE 8 MAINTENANCE AND REPAIR 8 1 Safety during maintenance procedures Maintenance
30. level in oil tank All electrical repairs must be performed by a licenced electrician Toutes les r parations lectriques doivent obligatoirement tre r alis es par un lectricien agr 42 8 2 Troubleshooting POSSIBLE SITUATIONS PROBLEM Table will not activate when pressing ON OFF button PROCEDURE CAUSE Main Switch in OFF position Turn to ON position Hand Control Unit not connected properly Check the connection Faulty Hand Control Unit Replace the hand cont rol unit Broken Hand Control Unit cable Replace the cable Table battery not charged Connect mains cable Battery fuse blown Check and replace Faulty MPC Replace MPC PROBLEM Table adjustment not functioning when pressing function PROCEDURE button table activated CAUSE Floor lock not activated Press activation button Making adjustment table program restricts Check the table posi tioning Adjustment already in its end position Check the table posi tioning PROBLEM Table movement won t stop although function button is PROCEDURE released CAUSE Severe functioning failure of electronics MAIN SWITCH to OFF position CAUSE Grave dysfonctionnement lectronique INTERRUPTEUR PRIN CIPAL en position d arr t Faulty hand control unit MAIN SWITCH to OFF position Commande manuelle d fectueuse INTERRUPTEUR PRIN CIPAL en position d arr t 43 PRO
31. lowing environmental conditions Ambient temperature 10 40 C Ambient pressure 700 1060 mbar Relative humidity 30 96 75 96 Transport temperature 10 40 C Storage temperature 10 40 C 4 2 2 Electrical specifications Primary voltage 100 240 V Frequency 47 63 Hz Secondary voltage Nom 24V DC Maintenance free lead acid bat 24V 18 Ah tery Max power consumption 700 VA Main fuses T6 3AL 5x20 2 pcs Normal recharging time 5 10h depending on battery charge level Operation time with full battery approx 180 min Max uninterrupted operating time 2 min 10 min ED 20 If you would like more detailed information please contact the Technical Service 4 2 3 Classification data Operating table is classified according to EN 60601 1 as follows Electric shock protection Degree of electric protection Protection against liquids Cleaning and disinfecting Operating rate Protection against flammable anaesthetic gases Safe working load SWL Class equipment internally powered B type IP X4 Splash proof equipment According to instructions on page 37 Intermittent operation ED 20 96 Do not use with combustible gases 325 275 225 180 kg incl patient mattresses and accessories 35 4 2 4 Adjustment ranges Height Longitudinal shift SLIDE Lateral tilt Back section Leg section Trendelenburg and antitrendelenburg Headr
32. ly when the mains cord is disconnected Connect the mains cord only to earthed power supply Operating table shall be operated from it s internal battery if the integrity of the protective earth conductor arrangement is in doubt Activate the floor lock before adjusting the table Only trendelenburg and back rest adjustments are functio ning if floor lock is not activated Place the patient in the longitudinal center line of the table top In NORMAL mode max patient weight with sliding seat section is 275 kg 600 Ibs In NORMAL mode max patient weight with fixed seat section is 325 kg 700 Ibs In REVERSE mode max patient weight is 180 kg 400 Ibs WARNINGS CAUTIONS AND NOTES Adjust the table top to the horizontal position O position with the hand control unit before transporting a patient on the operating table Transporting of patients weighing up to 180 kg 400 Ibs is only allowed when table top is in horizontal posi tion O position and the height of the table is max 900 mm 3 ft Transporting of patients weighing 180 225 kg 400 500 Ibs is only allowed when the table is in NORMAL mode in horizontal position O position and in the lowest position Transporting of patients weighing 225 325 kg 500 700 Ibs is not allowed Use extreme caution when transporting the table with a patient on Transporting of the table with a patient on requires two persons To maximize patient
33. mum output power rating of the transmitter in watts W according to the transmitter manufacturer Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Table 4 Recommended separation distances 48 ORDER FORM Orderer Invoicing address Delivery address Mark Reference Part name Information Dealer NOTES
34. n be cleaned with pressurized air 22 gt gt gt gt gt gt 3 2 8 Hand control units When adjusting the operating table take care that the fingers hands or other parts of the body of the patient are not placed between edges of back and leg or seat section frames and pivoting points Ensure before adjusting the operating table that table top will not hit other obstacles Ensure before adjusting the operating table that NORMAL REVERSE mode has been chosen correctly and is the same as patient s orientation on the table top After use of main switch NORMAL mode is set as default Lors du r glage de la table d op ration viter de coincer les doigts les mains ou toute autre partie du corps du patient entre les bords des cadres du dossier et du support pour membres inf rieurs ou du si ge ou aux endroits de rotation Avant de r gler la table d op ration v rifier que le plateau de la table ne risque pas de heurter d autres obstacles Avant de r gler la table d op ration v rifier que le mode NORMAL INVERSE a t correctement choisi et qu il correspond l orientation du patient sur le plateau de la table Apr s l utilisation de l interrupteur princi pal la table est en mode NORMAL par d faut HAND CONTROL UNITS Remote control with cable IR remote control unit Infra red Hand control unit with cable is connected to the receptacle placed on the OR panel on the left side of the column CONNECTING
35. n haute Lorsque le bouton est abaiss en position d arr t toutes les fonctions de la table sont interrompues Le mouvement de la table cesse imm diatement Turn the main switch to OFF position before connecting disconnecting any electronical components in order to protect them from voltage surges Mettre l interrupteur principal en position d arr t avant de brancher d brancher des composants lectron iques afin de les prot ger des surtensions Turn the main switch to OFF position downwards when the table is not in use or stored for longer period of time This eliminates the possibility of unintended use and is also saving the battery capacity If the main switch is turned to OFF position downwards all table functions are blocked including recharging of the battery Check that the green led above the charging inlet lights up when you connect the mains cord to table Main switch uipotential connector Z Appliance inlet with protective cover 30 gt gt gt gt gt D gt D 3 2 13 Recharging of battery and mains use When recharging the battery first plug the mains cord to table appliance inlet after that to wall socket Connect the mains cord only to earthed power supply Raccorder le cordon d alimentation secteur uniquement une alimentation lectrique mise la terre Charge condition of the table battery is indicated with the lights on the cabled hand control unit and
36. nufacturer if any problems arise Portable and mobile RF communications equipment can affect operating table AVERTISSEMENTS MISES EN GARDE ET REMARQUES A AVERTISSEMENTS La table d op ration doit tre utilis e uniquement dans les installations con ues pour un usage m dical Aucune modification de cet quipement n est autoris e La classification IP X4 est valable uniquement lorsque le cordon d alimentation secteur est d branch Raccorder le cordon d alimentation secteur uniquement une alimentation lectrique mise la terre La table doit tre aliment e par sa batterie interne si l int grit de la mise la terre est mise en doute Actionner le syst me de fixation au sol avant de r gler la table Seuls les r glages de la position de Trendelenburg et du dossier fonctionnent si le syst me de fixation au sol n est pas actionn Placer le patient sur la ligne m diane longitudinale du plateau de la table En mode NORMAL le poids max du patient avec le si ge coulissant est de 275 kg 600 Ib En mode NORMAL le poids max du patient avec le section fixe de si ge est de 325 kg 700 Ib En mode INVERS le poids max du patient est de 180 kg 400 Ib R gler le plateau de la table en position horizontale position 0 avec la commande manuelle pour transporter un patient sur la table d op ration Le transport de patients de 180 kg 400 Ib au maximum est seulement autoris lorsque le plat
37. osition Use of SLIDE adjustment is restricted with tilted table top Simultaneous max angles of Trendelenburg Reversed Trendelenburg and Side tilt are limited with software OVERRIDE panel is only for returning the the table top to the horizontal position in the case of electronical malfunction With OVERRIDE it is possible to adjust table without the controlling processor and any of the programmed restrictions are not functioning 27 3 2 10 Override panel When adjusting the operating table take care that the fingers hands or other parts of the body of the patient are not placed between edges of back leg or seat section frames and pivoting points Ensure before adjusting the operating table that table top will not hit other obstacles Lors du r glage de la table d op ration viter de coincer les doigts les mains ou toute autre partie du corps du patient entre les bords des cadres du dossier support pour membres inf rieurs ou du si ge ou aux endroits de rotation Avant de r gler la table d op ration v rifier que le plateau de la table ne risque pas de heurter d autres obstacles OVERRIDE panel is only for returning the the table top to the horizontal position in the case of malfunc tion Malfunction of control electronics Broken or not functioning control device Override makes it possible to adjust table without the controlling processor and any of the program med restrictions are not functioning
38. override panel Green Good Yellow Lowered Red Low start recharging When the mains cord is connected to the table the charging sequence starts automatically Yellow flashing Charging in progress Green flashing Charging completed Adjustments can be used normally when recharging the table battery When recharging is completed disconnect the mains cord first from the wall socket after that from the table appliance inlet Disconnect the mains cord from the table before any service procedures 100 240 V used in power unit placed in table base Hazard of electric shock IP X4 classification is valid only when the mains cord is disconnected La classification d IP X4 est valide seulement quand la cordon d alimentation secteur est disconnected When mains cord is connected to operating table the cooling fan of the power unit is activated causing a humming sound After disconnecting the mains cord the cooling fan still runs for few seconds If the Main switch is turned to OFF position downwards all table functions are blocked including recharging of the battery Check that the green led above the appliance inlet lights up when you connect the mains cord to table If the table has been in cold temperature during transport or storage allow it to warm up to room temperature at least 6 hours before recharging the battery or connecting power This ensures that the possible con densed humidity has time to evaporate
39. perated service pro gram Table can be connected to 100 240 VAC mains power and its secondary voltage is 24 VDC The table electronics are safely earthed and it has equipotential connection Table top consists of fixed or sliding seat section and detachable 25 and 40 sections You can arrange the sections as you wish depending on the surgical concept and operation Look at the page 17 The width of the table top is 540 mm 21 1 4 inch without rails and length 2100 mm 82 inch Table top plates are x ray translucent material Operating table can be adjusted with cable connected hand control unit IR hand control unit or foot switch It can also be adjusted with override panel in case of electronical malfunction hand control unit or table microprocessor defective Safe and fault free use and maintenance of the equipment requires careful aquainting to these instructions When mounting accessories to the equipment the instructions provided with them must be followed closely Always keep the instructions for accessories together with this manual 13 3 PRODUCT USE 3 1 Unpacking Check the package for possible transport damages Remove strappings around the package and lift the sides and top of package by lifting them upwards Remove covering plastic bag other packing material and possible accessories e Remove M10 nuts 1 8 pcs M10 bolts and fixing parts 2 4 pcs on the castors e Remove wooden fixing parts 3
40. s equipment can affect operating table Other products used in the vici nity of operating table should also comply with this standard The operating table intended for use in the electromagnetic environment specified in this quidance The cus tomer or the user of the operating table should assure that these are used in such an environment Guidance and manufacturer s declaration Electromagnetic emissions Emissions test Compliance Electromagnetic environment guidance RF emissions Group 1 Operating table does not create inten IEC EN 55011 tionally RF energy Therefore it s RF emissions are low and are not likely to cause any interference in nearby elec tronic equipment RF emissions Class A Operating table is suitable for use in IEC EN 55011 all establishment other than domestic and those directly connected to the public low voltage power supply net Harmonic current emissions Class A work that supplies buildings used for IEC EN 61000 3 2 domestic purposes Voltage changes fluctuations Complies and flicker IEC EN 61000 3 3 Table 1 Electromagnetic emissions 45 Guidance and manufacturer s declaration Electromagnetic immunity Operating table is intended for use in the electromagnetic environment specified below The customer or the user of operating table should assure that it is used in such an environment Immunity test IEC EN 60101 1 2 test level Compliance level
41. s et endommager les composants lectroniques sensibles Se mettre la terre en touchant des pi ces m talliques de la table avant de toucher des composants lectroniques Le d branchement d un flexible hydraulique peut provoquer le mouvement incontr l et dangereux de la table Soutenir les parties de la table avant de d brancher des flexibles hydrauliques 41 AN 8 1 1 8 1 2 8 1 3 Daily service During normal cleaning a general visual examination should be made checking for table condition Contact service personnel if needed We recommend that you charge the table battery every day This is to ensure sufficient battery charge level and long service life Check the mains cord condition Operating table must be cleaned after each surgical procedure Monthly service Clean and lubricate guide pins latches spring loaded pins and pivoting points with light machine oil or spray vaceline Check that castors roll smoothly and floor lock is functioning correctly Test table functions by adjusting them to extreme position and back Check hand control unit and foot switch cables connector and receptacle Annual service Perform all daily and monthly service actions Check the functioning of table section s latches Visual check at the hydraulic hoses connections and piston rods regarding external leakage or wear marks If visible wearing is noticed maintenance is needed immediately Check the oil
42. safety utilize proper restraint methods during transport To avoid endangering the patient s respiratory system nerve pathways and circulatory system the patient should be positioned properly and kept under observation Check that table top sections latches lock correctly Incorrect attachment of table sections is dangerous and can cause personal injury or equipment damage Use only accessories recommended by the table manufacturer When adjusting the operating table take care to avoid collision between accessories and the operating table Do not use worn or damaged accessories When using the table ensure that all accessories are properly mounted and check the function of the accessory locking and adjustments When adjusting the operating table take care that the fingers hands or other parts of the body of the patient are not placed between edges of back leg or seat section frames and pivoting points Use only permitted table top configurations illustrated in this manual Ensure before adjusting the operating table that table top will not hit external obstacles Ensure before adjusting the operating table that NORMAL REVERSE mode has been chosen correctly and is the same as patient s orientation on the table top After use of main switch NORMAL mode is set as default WARNINGS CAUTIONS AND NOTES Always follow manufacturer instructions when using diathermy or defibrillation equipment When using high frequency surgica
43. short inter ruptions and voltage variations immunity test IEC EN 61000 4 11 30 500 ms 60 100 ms 100 10 ms 100 5000 ms 30 500 ms 60 100 ms 100 10 ms 100 5000 ms Mains power quality should be that of a typical hospital environment Table is mainly operated with inter nal battery Table 3 Electromagnetic immunity part 2 47 Recommended separation distances between portable and mobile RF communications equipment and operating table patient bed Operating table is intended for use in the electromagnetic environment in which radiated RF dis turbances are controlled The customer or the user of operating table can help prevent electro magnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and operating table as recommended below accord ing to the maximum output power of the communications equipment Rated maximum output Separation distance according to frequency of transmitter power of transmitter m n 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 24P d 1 24P d 2 34P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maxi
44. the table Press ON OFF button again The table returns automatically to stand by mode if the buttons are not pressed in 4 minutes Activating or deactivating Floor Lock Next to function buttons there are indicator lights showing the present Floor Lock mode To activate Floor Lock press relevant button at least 2 seconds Activating time is about 5 seconds and table base and castors will be raised automatically off the floor with 4 hydraulic support feet Floor Lock activating deactivating can be cancelled by pressing the opposite button Deactivating takes about 5 seconds Only trendelenburg antitrendelenburg and backrest adjustments are possible when Floor Lock is deactivated 5th wheel factory assembled option Active mode is shown with indicator light on the OR panel and hand control unit To activate 5th wheel press relevant button steadily at least 2 second Activating time is about 3 seconds 5th wheel activating deactivating can be cancelled by pressing the opposite button Deactivating takes about 3 seconds Choose NORMAL or REVERSE mode according to patient orientation on the table Active mode is shown with indicator light next to button To change patient orient mode press relevant button for 2 seconds Orient mode change is informed with audial signal After using Main Switch Normal mode is set as default 25 To adjust the table press relevant button steadily
45. tic parts and mattresses Dry the operating table carefully immediately after cleaning or disinfecting 6 4 Mattresses and pads Read the care instructions for mattresses and pads Disinfection e g chloramine based diluted disinfectant Cleaned as stated on the mattress or described in section 6 2 38 b 7 RECYCLING 7 1 Metals and plastics When disposing of a product or replacing any of its parts check the recyclability of each item A majority of the metal used in operating table is stainless steel or epoxy zinc coated steel Other metals used are anod ized aluminium and aluminium bronze When recycling plastic parts determine the material type Refer to the surface materials table which is found in this user manual page 33 to confirm whether or not recycling is possible For more information about recycling contact your local waste management facility or visit related sites on the internet Below are recyc ling symbols which are marked on parts made of plastic Products marked with these symbols can be used as energy waste x AN fx x S oN N A A BN PET PE HD PE LD PP PS 0 7 1 1 Hydraulic cylinders and gas springs Hydraulic cylinders and gas springs can be disposed of as metal waste when all oil has been removed Gas springs contains pressurized nitrogen gas which should be removed WARNING Head section gas spring can be disposed as metal waste after nitrogen gas and oil has been removed which shou
46. tion B type applied part Recycling NOTE Safe working load vary depending on operating table seat section construction SLIDING SEAT SECTION TYPE SWL 275 kg 600 Ibs FIXED SEAT SECTION B3 SWL 325 kg 700 Ibs P 275 kg Maximum load with sliding seat section including patient mattresses and accessories p E 325 kg Maximum load with fixed seat section including patient mattresses and accessories gt 33 4 1 1 Labeling and symbols Alternating current B type applied part Attention consult accompanying documents RECYCLING Product must be recycled separately N IX D gt Marking in accordance with European Council Directive 93 42 EEC Marking of ETL certification 9 3196120 4 1 2 Surface materials Surface materials used in operating table with possibility to skin contact Epoxy coated steel Zinc plated steel Stainless steel Aluminium bronze High pressure laminate PUR Anodized aluminium Polycarbonate PC Silicon EPDM rubber ABS PMMA PC ABS composite ABS Frame constructions Frame constructions Accessory rails column casings frame constructions Pivot pins Table top plates Mattresses Table top section sides Hand cntrol unit B cover and battery cover Hand control unit A cover Bellows Base and castor covers Override panel IR receivers 4 2 Specifications 4 2 4 Environmental specifications Operating table is designed to be used in fol
47. uctions 6 2 Cleaning 6 3 Disinfection 6 4 Mattresses and pads RECYCLING 7 1 Metals and plastics 7 1 1 Hydraulic cylinders and gas springs 7 1 2 Oil 7 1 8 Electronic waste and batteries MAINTENANCE AND REPAIR 8 1 Safety during maintenance procedures 8 1 1 Daily service 8 1 2 Monthly service 8 1 3 Annual service 8 2 Troubleshooting EMC GUIDANCE ORDER FORM NOTES 37 37 38 38 38 38 38 39 39 39 39 40 41 41 42 42 42 43 45 49 50 e gt gt 1 WARNINGS CAUTIONS AND NOTES In order to ensure optimal surgical safety all operating table users should carefully read the operating table and maintenance instructions before using the table The entire surgical ward staff should be familiar with the correct use of the operating table as well as all warnings cautions and notes concerning it Warnings and notes found in this user manual are indicated as follows WARNING Please observe to ensure user maintenance personnel and patient safety AVERTISSEMENT Doit tre respect pour garantir la s curit de l utilisateur du personnel d entretien et du patient CAUTION Please observe in order to avoid causing damage to the equipment or its parts NOTE Please observe in order to improve operating table properties WARNINGS Operating table should only be used in facilities made for medical purposes No modification of this equipment is allowed IP X4 classification is valid on