1. 35 36 ZERO m EES 36 36 tee onte torte tots vo 36 NBET GROSS 5 36 USER 5 ide v eg 36 UNIES ERO 37 37 VIEW TOTAL VIEW 37 Clear Last Total 37 Clear Totals 38 Keyboard Tate ue ente 38 Display Key tro te b neta 39 SETPOINT S eeN Kaa 39 DT ER 40 PRIN eterne teer 40 SETUP 5 5 inei 40 Troubleshooting 4 042222 42 Service Counter Warning 402 4 2 44 The MSI Limited Warranty 2 22 2 1 45 Proper Loading Procedures 46 MSI 3360 Challenger 2 User Guide 3 He A 1 Nm INTRODUCTION The MSI 3360 Challenger 2 is a combination of the sound and proven me chanical design of the original Challenger with today s most advanced electron ics to provide a superb feature set unmatched by any scale in its class or price range The multi purpose hanging scale is ideal for situations in which head room is at a minimum The Challenger 2 is versatile reliable accurate and easy to operate SPECIFICATIONS Accuracy 0 1 of applied load one disp
2. A B C A A B C gow SY Sr EIM I Possible Cause Filter set too low Faulty electronic circuit board Corroded battery connections Defective charger Defective battery Service Counter exceeded Broken wire or improper connection Unit mis calibrated Load cell damaged A button is being depressed or button is stuck Current weight is greater than rated capacity Load cell damaged Defective electronics Broken wire or improper load cell wiring A The desired set of keys is not active The access codes are incorrect The Challenger 2 is not set up to receive infrared commands The batteries in the IR Remote controller are dead IR Remote is in wrong mode Remote not programmed Mow te ik WN Ao IONAL Solution Change filter setting Requires authorized service Clean connections Requires authorized service Replace Inspect load train and contact factory Check load cell wiring for connectivity Re calibrate with Initial Calibration Replace load cell Release button or contact qualified service technician Remove excess weight Replace load cell Contact qualified service technician Check Load Cell connection wiring Press 2ND or SCALE to activate the correct set of IR Remote keys See THE SCALE AND 2ND KEYS See TO SET REMOTE CONTROLLER ACCESS CODES See TO ENABLE REMOTE CONTROLLER Replace 2 batteries with new See THE SCALE A
3. Action 1 Press to scroll to the CAL r message Press 2 Display flashes UnLd Unload the scale and press The scale will display CALC for calculating 3 Display shows LOAd for 2 seconds then flashes last set CAL r value Input the CAL r value printed on the load cell Cal Sheet After the last number has been entered press The display will show CAL d 4 Display flashes UnLd Unload the scale and press Afterwards CAL r will be displayed until the value is determined Then the CAL r value is displayed for 10 seconds or until any button is pressed Final Calibration is complete and the display shows StorE to signify that the calibration parameters are being saved in backup memory The calibration Page 30 MSI 3360 Challenger 2 User Guide SYSTEMS WAT menu returns Press to return to normal operation or press EISENACH to scroll to a different Calibration Setup operation Note When Calibration is finished seal the Calibration Port on the front panel of the scale STANDARDS StAnd Function Facilitates calibration to your required standard IndUS Industrial nISt National Institute of Science and Technology NTEP setting EuroP Metric OIML 1Unit One Unit only for custom scales Action 1 Press 0101 StAnd is displayed 2 Press and the currently selected standard will flash 3 Pres
4. StAnd Standards Industry National Institute of Standards amp Technology Metric kg only OIML One Unit Only Remote Control Automatic ZERO Tracking CALIBRATION CAL Function Used to Calibrate the scale load cell There are two types of calibration 1 Initial Calibration and 2 Subsequent Calibration INITIAL CALIBRATION Function Initial Calibration is used at the factory when Calibrating for the first time when the circuit boards have been replaced or after the scale has gone through System Initialize Note Do not use this procedure for routine Calibration Action 1 Press until CAL appears then press The following steps will occur 2 Display flashes right hand Ib indicator while displaying number of Ibs if MSI 3360 Challenger 2 User Guide 27 He A S WR Y WAT 1 Ww AL calibrating in lbs press or to select kgs left hand kg indicator flashes press then Note This step will not occur in scales equipped to measure in only one unit 3 Display shows CAP for 2 seconds then flashes 0 Input the capacity using and 4 Display shows d for 2 seconds then flashes the default Count By division Press to accept or press to pick another Count By then press 5 Display flashes GAin4 Press to accept or press EINE to pick
5. GROSS TOTAL 1000 1 Press USER button must be pro all segments amp indicators are displayed for 3 seconds grammed as test see Scale LOBATT PEAK MOTION Setup 71 2 Battery condition is dis H Lb played in Percent of battery NUES RIT life remaining with b XXX 5000 x 2 kg 10000 x 5 Ib figure from 0 to 100 on the CAL number is displayed digits in 5 increments 3 All segments and indicators are displayed for 3 seconds 4 The CAL number is dis 5 5 5 5 5 LO BATT PEAK MOTION 30 GROSS TOTAL 1000 5000 x 2 kg 10000 x 5 Ib played next To ensure that the CAL number is accurate the scale must be unloaded Cal number shifts of up to 10 counts are insignificant 5 digits 7 segment and 16 segment count once from 0 to 9 indicators are tested 6 Internal tests are performed to further ensure scale integrity Final The reading returns to the last condition or an Error message is displayed and all digits count once from 0 to 9 MSI 3360 Challenger 2 User Guide 13 HE A SUREM SYSTEMS ACCESSING THE SERVICE COUNTERS LOBATT PEAK MOTION Function Lint Challenger 2 maintains two service See e TU enn afik counters for safety and two service 5000x2 kg 10000 x5 b counters that monitor internal LCntl will be displayed for 1 second followed electronics These Service Counters are
6. SYSTEMS TWTeER WAT SUBSEQUENT CALIBRATION Function Subsequent Calibration is used for routine annual Calibration or when the scale is not weighing accurately Action 1 To set ZERO the display flashes UnLd Unload the scale and press ENER The scale will display CALC for calculating 2 To input the Calibration weight value the display will show LOAd for 2 seconds then flash the capacity Load the scale with this weight or input another weight After the last number of the weight has been entered and the motion indicator is off press The display will show CAL d if successful or CErr if not successful An unsuccessful calibration can be due to an improper gain setting or use of a calibration weight that is too light 3 To set the CAL r value the display flashes UnLd Unload the scale and press button Afterwards CAL r will be displayed until the value is determined If successful then the CAL r value is displayed for 10 seconds or until any button is pressed If unsuccessful the display will show r Err for 2 seconds Note It may be desirable to bypass a calibration step To do so press The step bypassed will retain the default or last set value However bypassing a calibration step may result in an incomplete or inaccurate calibration Final Calibration is complete and the display shows StorE to signify that the calibration parameters are being saved in back
7. indicat ing that the setpoint is now set to respond when the weight is less than the value you put in Note Ifthe value entered exceeds capacity the message 2 biG appears on the display The TOTAL key functions as a decimal point if needed To enter a Setpoint value less than 1 start with the key fol lowed by the Decimal Point e g 0 5 Use the key to delete incorrect entries ID Not supported by the Challenger 2 at this time PRINT Not supported by the Challenger 2 at this time SETUP Function Setup functions are the same as when accessed by the Challenger Front panel The keys used for entering and maneuvering through menus are simplified with the IR Remote For a detailed description of the Setup functions see Scale Setup page 19 of this manual To Enable the Setup Menus 1 Aim the IR Remote at the Challenger Front Panel Page 40 MSI 3360 Challenger 2 User Guide SYSTEM S 1N WAT 1 NM A OL 2 Press the key twice The first menu Func User Key function will appear Key Functions While In Setup Menus USER Use these keys to scroll through the various menus and to scroll or UNITS through the menu selections once in a menu ENTER key is used to enter into menus and to store the selection in memory once in a menu CLR The or keys are used to exit from menus or EXIT without changing the previously selected item When in the top
8. 6 Setting or changing TARE has no effect on the ZERO setting 7 Tareing will reduce the apparent overrange of the scale For example if a 1000 Ib container is tared and the scale capacity is 5000 Ib the scale will overload at a new weight of 4000 Ib 5000 1000 plus any addi tional allowed overload usually 9 divisions 8 TARE value is stored in backup memory and the value will be restored the next time the scale is turned on TARE OUT 1 The scale must be in NET Mode 2 Press to clear the value 3 Returns to GROSS Mode Final TARE IN TARE is set displaying NET weight TARE OUT TARE weight will be cleared and the scale will revert to GROSS Mode Page 12 MSI 3360 Challenger 2 User Guide SYSTEMS TWTeER WAT USER 52 TEST tESt Function USER can be programmed to any of five functions TEST tESt NET GROSS nEtGr TOTAL totAL UNIT kg Ib and PEAK Phold See Scale Setup for instructions on programming the USER button Action Press WAAN to initiate the preprogrammed function LO BATT PEAK MOTION Function h Provides a functional system 30 NET GROSS TOTAL 1000 test and an on demand display S000 x2 kg TIRE check without disturbing the current weighment Also provides calibration verification in the form of a load cell CAL Battery condition in percent remaining Action e20c NET
9. Ib Function To change unit between kg Ib Action 1 Press to select kg or lb Mode USER must be programmed as Unit see USER Setup The kg and Ib indicators are located on the front panel adjacent to their relative capacity labels Final The correct unit of measurement weight is displayed Alternate Method Use the Setup UNITS Mode to change the UNITS without programming USER to UNITS PEAK Phold Function Allows the capture of a PEAK weight Useful for stress testing and related operations where the maximum load measured is stored Action 1 Press 3 8 USER button must be programmed as PEAK Phold see USER Setup 2 To enable PEAK Phold the PEAK indicator must be off Press 3 To disable PEAK Phold the PEAK indicator must be on Press Final 1 When enabled the PEAK Phold value is reset to 0 and any subsequent displayed value will correspond to the greatest positive weight detected 2 When PEAK Phold is disabled the scale will perform normally Page 18 MSI 3360 Challenger 2 User Guide SYSTEMS TWTeER WAT NET GROSS nEtGr SCALE SETUP Function Switches the display between NET and GROSS Modes NET Weight is defined as GROSS Weight minus a TARE Weight Use 1 There must be a TARE weight established to switch from GROSS Mode to NET Mode see TARE 2 NET GROSS nEtGr will work even when the scale is in Motion Action 1 Press 99 8 USER button must
10. TOTAL indicator will flash again indicating the previous settled reading has been replaced 4 When a weight is totaled the TOTAL indicator will flash for a steady 4 seconds indicating the weight has been totaled and the TOTAL value will be displayed 5 The displayed weight is added to the accumulated value in the TOTAL Register only when the weight returns to ZERO 1 of Full Scale 6 The last settled reading is what will be used for totalizing when the scale returns to ZERO WARNING The operator must take caution that when removing the load the scale does not go out of Motion or an erroneous reading could occur The last settled weight is actually added to the TOTAL when the scale returns to less than 1 of full Scale capacity Action 1 The scale cannot be in motion An Auto TOTAL acceptable reading is indicated by 3 or more short flashes of the TOTAL indicator 2 Weight readings must be greater than 2 of full scale relative to NET ZERO or GROSS ZERO 3 Each reading added to TOTAL must be preceded by a return to ZERO NET or GROSS 1 of full scale Totalization of the last settled weight is indicated by a 4 second flash of the TOTAL indicator and a Page 16 51 3360 Challenger 2 User Guide SYSTEMS WAT NM AL display of the current TOTAL Final After totaling normal weighing resumes TO CLEAR THE LAST TOTALED WEIGHT Function If the last totaled automatic or
11. be programmed as NET GROSS nEtGr see USER Setup 2 Nocurrent TARE is stored TARE 0 No action display continues to read the Gross weight only 3 TARE value is stored Toggles between NET and GROSS indicators Final NET or GROSS indicator will illuminate Function Scale Setup facilitates changes to various internal settings of the scale Setup is initiated by holding in the USER button while turning POWER on Action When in SETUP the following buttons are used for selecting and changing scale settings ZERO functions as CLEAR EXIT Used to exit from a Setup menu without affecting the previous setting and for exiting Setup Mode completely Used also to clear the current digit when entering numeric values i e Set Points TARE functions as ENTER Used to enter into menus and choose menu items USER functions as SELECT Used to scroll through menu selections If held down SELECT will automatically repeat POWER functions as Decimal Point Used for entering a decimal point in displayed values i e Tare or Setpoint Final Pressing will skip all remaining parameters and return the scale to normal operation Pressing will also exit the Setup Mode and turn Power off MSI 3360 Challenger 2 User Guide 19 QUICK USER SETUP GUIDE Note Display items in Top Menu bold are defaults Func S Ilis I Selection Display OFF tESt totAl
12. clevis assembly 5 Clevis pin MSI 3360 Challenger 2 User Guide Page 7 VE A EM y S amp S 1 Ww AL MSI 3360 Assembly Parts drawing 1 1 Slide hook clevis over load cell with open end of hook toward front of scale 2 Align holes of clevis and load cell 3 Slide the clevis pin through the clevis and load cell holes 4 Lock clevis pin in place with cotter pin Bend cotter pin 5 Slide battery pack into battery compartment The battery will automatically engage with its connectors 6 Secure battery pack by turning the two locking fasteners on access door clockwise 1 4 turn 7 The scale is now ready for use BATTERY PACK The Challenger 2 is powered by a six volt rechargeable battery The battery is permanently attached to the battery door To remove the battery pack turn the two fasteners on the access door counter clockwise 1 4 turn then pull the battery pack straight out The battery will automatically disengage from its connectors Page 8 MSI 3360 Challenger 2 User Guide SYSTEMS TWN TER WAT NM AL Battery pack installation and removal The battery will operate for up to 100 hours depending on LED brightness setting before requiring recharging In order to conserve battery life the scale includes an Automatic Power Off Mode which senses operational status for no activity after 12 minutes or one hour and turns the scale o
13. level of the Setup Menu use the key to return to normal scale operation Note The Challenger 2 can also be controlled by the IR Remote during calibration It is still necessary to enable calibration through the seal port but from that point on the Menu Enter Numeric and Clr keys all simplify the calibration process MSI 3360 Challenger 2 User Guide 41 MS te sy I NTERNAT TROUBLESHOOTING GUIDE Problem 1 Display is blank when POWER button is depressed 2 Display does not function properly or front panel buttons do not function normally or scale will not turn off 3 Display does not respond to weight changes 4 Display over ranges below 100 of capacity 5 Display experiences excessive Zero drift between weighments 6 Display shows large number after Power up sequence with AZt feature Off and no applied weight 7 Displayed weight shows large error Page 42 MSI 3360 Challenger 2 gom IU OV User Guide Possible Cause Discharged battery Defective battery Corroded battery Defective button or electronic circuit board Power button not properly depressed Computer lock up Faulty electronic circuit board Faulty front button assembly Same as 2A B above Out of calibration Faulty load cell or electronic circuit board Load cell disconnected from printed circuit board Tared out weight is added to load when overload c
14. manual weighment was a mistake it can be erased with the following procedure This erases only the last totaled value Action 1 Press 53 8 then ENAR The display will read the revised TOTAL for 2 seconds This procedure assumes that you have not modified the USER function Final After displaying the update TOTAL normal weighing resumes TO CLEAR THE TOTAL VALUE VIEWING TOTAL Function To clear the TOTAL register in order to start a new series of totals Action 1 Press Seam then KANARA This procedure assumes that you have not modified the USER function 2 The display reads 0 and the TOTAL indicator will illuminate for 2 seconds Final After clearing the TOTAL normal weighing resumes Function After a weight has been totaled or before the 1 threshold has been ex ceeded the current TOTAL can be Viewed Action 1 Make sure there is no load on the scale Press USER button must be programmed as TOTAL 2 The current TOTAL will be displayed for 3 seconds The TOTAL indicator will illuminate Final Normal weighing resumes MSI 3360 Challenger 2 User Guide 17 He A Teh Caution If in Manual TOTAL and you push the TOTAL button to view the TOTAL value be certain that you are not weighing more than 2 of capacity If you press USER to view the TOTAL and the weight is above 2 of capac ity you will add to the TOTAL rather than view the TOTAL UNIT kg
15. release the button Press and release the button Press and release the HQ button The indicator light will go out The IR Remote is now ready for use WARNING If the access codes are not properly set the IR Remote might fail to function or will function erratically Always reset the access codes after changing batteries THE SCALE AND 2ND KEYS The Challenger IR Remote has two modes of operation 1 Pressing the key defines the IR Remote keys to the light grey commands For example after pressing the key the VIEW TOTAL and RECALL TARE functions are unavailable Active PWR USER ZERO TARE UNITS ENTER TOTAL 0 9 DISP NET GROSS SETUP SETPT1 SETPT2 2 Pressing the key defines the IR Remote keys to the green commands For example after pressing the key the VIEW TOTAL and RECALL TARE keys are active while many other remote controller keys are inactive Active PWR USER RECALL TARE UNITS ENTER RECALL TOTAL MSI 3360 Challenger 2 User Guide 35 He A ulis ZERO TARE RECALL TARE NET GROSS USER Page 36 MSI 3360 Challenger 2 To turn on the Scale 1 Aim the IR Remote at the Challenger front panel 2 Press and hold down the ROMAH key until the Challenger turns on Once the display lights up release the button This usually takes 2 to 4 sec onds To turn off the Scale 1 Aim the IR Remote at the Ch
16. the key 4 Immediately press the key 5 Immediately press the 35 key KEYBOARD TARE Function A tare value can be entered using the numeric keypad Once the value is entered the scale will change into the NET mode Operation 1 Aim the IR Remote at the Challenger Front Panel 2 Press the numeric keys for the Tare value desired Verify as you enter the number that each number appears on the Challenger display 3 Press the key to input the number as a Tare value Page 38 MSI 3360 Challenger 2 User Guide SYSTEMS TWTeER WAT NM AL Note Ifthe value entered exceeds capacity the message 2 biG appears on the display The TOTAL key functions as a decimal point if needed To enter a Tare value less than start with the key followed by the Decimal Point e g 0 5 Use the key to delete incorrect entries Final 1 The scale mode changes to NET 2 The tare value is subtracted from all readings DISP KEY Function The DISPLAY key gives control of the brightness of the LED digits and annunciators In low ambient light it is advantageous to dim the LEDs since dimmer digits require less energy and extend battery life Use the DISP key to adjust the display for the minimum brightness that is easily legible at the distances and light conditions you use the most Note The AUTO mode for the display brightness see SETUP LEDS will automatically dim the display when no a
17. used to judge when the load bearing elements or electronic components should be inspected Action 1 To access the Service Counters program the USER LO BATT PEAK MOTION Bild GROSS TOTAL 1000 5000 x 2 kg 10000 x 5 Ib a flashing total of over 25 weighments button to TEST tESt LOBATT PEAK MOTION 2 Press and hold until the display reads LCnt1 P LLN L Load Counter 1 This will TEE occur after the RCAL value 5000 x 2 kg 10000 x 5 Ib has been displayed LCntl LCnt2 will be displayed for 1 second followed by will be displayed for 1 second followed by a flashing numeric value This value signifies the number of weighments that exceeded 2596 of capacity 3 Press ES LCnt2 Load Counter 2 will be displayed for one second followed by a flashing numeric value This value signifies the number of weighments that surpassed capacity 4 Press BERG A2DEr A2D Errors will be displayed for one second followed by a flashing numeric value This value signifies the number of times the internal microprocessor has reset the analog to digital con verter 5 Press EE Er EEPROM Errors will be displayed for one second followed by a flashing numeric value This value signifies the number of retries needed to successfully read from or write to the long term memory module 6 Press again to complete the TEST sequence Final TEST sequence continues LO BATT PEAK MOTION 30 G
18. MSI CHALLENGER 2 CRANE SCALES User Guide Quality Industrial Weighing and Force Measurement Equipment Measurement Systems International Ve A SUREMENT NAT IMPORTANT Please read this manual carefully before using the MSI 3360 Challenger 2 Program Number Capacity and Resolution Serial Number Cal Number Page 2 MSI 3360 Challenger 2 User Guide 4 B RS UR s Sf S ve il S I N Pe RM AT 1 TABLE OF CONTENTS Introduction e 4 m 4 lun ITI pm 6 d 7 npacking erre tea 7 horum 7 Battery Pack ttt 8 Battery Charger 9 Operation Guide 2 tenter reri 10 erred re eet ete o rte p eee 11 rr RERUM 11 12 USER 13 TEST tESE concetto terre eter tee dete rte ett 13 Accessing the Service Counters 14 To Reset the Service Counter 15 TOTAL totAL 15 Manual 15 Auto rtr tin 16 To Clear the Last Totaled Weight 17 To Clear the TOTAL Value 17 Viewing TOTAL eerte ette teer ertt 17 UNIT RSID uiii entree tr err t
19. ND 2ND KEYS Reset remote access codes See TO SET REMOTE CONTROLLER ACCESS CODES MSI 3360 Challenger 2 User Guide 43 He A ulis SERVICE COUNTER WARNING Page 44 MSI 3360 Challenger 2 Function A Warning is reported at Power On if Capacity of the scale has been exceeded more than 1 000 times or a weight in excess of 25 of capacity has been detected more than 100 000 times This warning informs the operator that the load train should be inspected by a qualified technician Action If a weight in excess of 25 of capacity has been detected more than 100 000 times then LCnt2 Load Counter 2 is displayed followed by the flashing numeric value To continue with Power On press any button Final Power On sequence continues Note If Service Counter Warning has not been reset it will be displayed the next time POWER is turned on with the updated numeric values User Guide SYSTEMS WAT THE MSI LIMITED WARRANTY MEASUREMENT SYSTEMS INTERNATIONAL INC WARRANTS load sensing elements and meters against defects in workmanship and materials for a period of one year from date of purchase and warrants electrical cables and batteries against the same defects for a period of ninety 90 days from date of purchase Any device which proves defective during the warranty period will be replaced or repaired at no charge provided th
20. NOT PULL LATERALLY ON POINT ATTACHMENTS AND ALLOWS THE HOOK WITH SCALE LOADED SCALE FREEDOM OF ALIGNMENT PUB PLP 01 95 Page 46 MSI 3360 Challenger 2 User Guide SYSTEMS TWTeER WAT MW AL NOTES MSI 3360 Challenger 2 User Guide 47 CC 2000 by Measurement Systems International Printed in U S A PUB 230 07 93D
21. ROSS TOTAL 1000 5000 x 2 kg 10000 x 5 Ib a flashing total of over capacity weighments Page 14 MSI 3360 Challenger 2 User Guide SYSTEMS WAT NM AL TO RESET THE SERVICE COUNTER WARNING 1 After inspecting the load train remove the calibration seal screw 2 Push the Calibration button using a blunt screwdriver 3 Push to return to normal operation Note This procedure does not reset the service counter TOTAL TOTAL Function For accumulation of multiple weighments The accumulator always uses the displayed weight so GROSS and NET readings can be added into the same TOTAL There are two Modes of totalizing Manual and Auto The Manual Mode requires the TOTAL button be pressed with the weight on the scale The weight will be added to the previously accumulated value The Auto Mode will automatically add the last settled and stable value to the TOTAL Both Modes require that the weight on the scale return within 1 relative to full scale of GROSS ZERO or NET ZERO before the next weighment can be added This assures that a weight on the scale is only added to the total once Applied weight must be 2 of full scale above GROSS ZERO or NET ZERO before it can be totaled MANUAL TOTAL Function Allows for accumulating weight manually by pressing a button Use 1 The Motion indicator must be off the scale must be stable 1 division If there is considerable motion on the sca
22. S Mode The scale will not ZERO in NET Mode 3 The scale will accept a ZERO setting over the full range of the scale Legal For Trade models might have a limited ZERO range ZERO settings greater than 5 of capacity will subtract from the overall capacity of the scale Action 1 Press 1 The weight reading must be stable within 1 division If this condition is met the digits display 0 or 0 0 or 0 00 depending on resolution MSI 3360 Challenger 2 User Guide 11 VE A S WR EM S amp S NWA dT 1 2 The ZERO setting is stored in backup memory and the setting will be restored the next time the scale is turned on Final GROSS indicator is illuminated and display will be 0 or 0 0 or 0 00 depending on resolution TARE TARE Function TARE IN Stores a TARE weight when in GROSS Mode and displays all subsequent readings in NET Mode or TARE OUT Clears the TARE weight when in NET Mode and returns the scale to GROSS Mode Action TARE IN 1 Press BAG 2 The scale must be in GROSS Mode The entire range of the scale can be tared 3 The Motion indicator must be off and the weight reading must be stable The weight reading must be stable within division 4 Ifthe Motion condition is met the NET indicator lights and the weight registers 0 All subsequent readings are deviations from the set TARE value 5 Only positive weight readings can be tared
23. To maximize battery life set the LED to the Automatic Sleep Mode UNIT kg Ib Function Allows Units to alternate between kg and Ib without the use of the USER button Action 1 Hold in the button while turning on the power 2 Press HEJNA until Unit is displayed 3 Press ENAU The display will show Unit while the indicator which correlates to the currently set Unit kg or Ib will flash 4 Press until desired Unit kg or Ib is displayed 5 Press to store or press to exit without changing Final Setup menu is displayed Press to return to normal operation Or press to scroll to a different USER Setup operation Alternate Method Program the USER button as UNITS SETPOINTS StPt1 StPt2 Function Set Points provide warnings and indications of weighments When the weight is above greater than or below less than a set value the Challenger 2 can respond by turning on the Set Point indicator light SET POINT ENTRY Function To enter a Set Point Action 1 Hold in the 58 button while turning on the power MSI 3360 Challenger 2 e User Guide 23 He A S y 2 Press to choose the Set Point you wish programmed Set Point 1 will be used for this example 3 When the display reads StPt1 press 4 Determine if the Set Point will be active when the weight is above GrEAt or below LESS the entered value Press to choose the mode Fo
24. Unit Phold nEtGr NTERNAT Sub Menu Display Definition Function Off Test Total Unit Peak Hold Net Gross OFF 2hr 1 hour Automatic Power Off Off 12 minutes 1 hour Lo HI 1 HI 2 Filter Low Medium High Auto LO HI 1 HI 2 Light Emitting Diodes Automatic Low Medium High Unit kg Unit Ib Units Kilograms Pounds OFF GrEAt LESS LESS GrEAt followed by value Set Point 1 Off Greater than Less Than OFF GrEAt LESS LESS GrEAt followed by value Set Point 2 Off Greater Than Less Than Page 20 MSI 3360 Challenger 2 User Guide PrESS Auto Total Off Press User Button Automatic Total ONAL SYSTEMS WAT N FUNCTION Func Function Allows the USER button to be programmed to TEST TOTAL UNIT NET GROSS or PEAK Action 1 Hold in the 3 8 button while turning on the power 2 Press HEA until the display reads Func 3 Press and currently selected function will flash 4 Press until desired function is displayed 5 Press to store or press exit without changing Final Setup menu is displayed Press to return to normal operation press to scroll to a different USER Setup operation AUTOMATIC POWER OFF A OFF Function The A OFF feature when enabled prolongs the battery life of the scale by turning POWER off when the scale is not in use Any time a button is depre
25. actory settings There are two types of calibration 1 Initial Calibration and 2 Subsequent Calibration Calibration must be accomplished by a qualified Scale Technician trained in Certified Calibration Standards To initiate calibration of a Challenger 2 the Technician is required to have an accurate test weight system of adequate capacity and in the case of a Legal For Trade Challenger 2 the test weight system must be certified by the appropriate regulatory agency Important Note Atany point except when entering calibration weight values the calibration procedure can be halted by turning POWER off Action 1 Remove the seal port on the left front panel of the scale Insert a probe such as a blunt tip screwdriver to depress the underlying button The display will show C SEt for 1 second 2 The Calibration Setup Menu appears Press the 999 button to scroll through the various Calibration Setup menu choices 3 Once the desired operation is displayed press Final To return to normal operation either scroll with the 390993 button to exit and press 8 134071133 or press 9831743549 Page 26 MSI 3360 Challenger 2 User Guide SYSTEMS TNT ERNAI QUICK CALIBRATION SETUP GUIDE Top Menu Selection Display Display Definition CAL Calibration Note Display items in bold are defaults F CAL Fine Calibration CAL r Calibration with Offset Calibration Resistor
26. allenger Front Panel 2 Press the key The Challenger will turn off immediately See User Guide page 11 for more POWER key information See User Guide page 11 See User Guide page 12 To view the currently stored tare value press the button and then press the Eie RENI button Don t forget to press the 97183 button to reactivate the main IR Remote key functions The NET GROSS function is available on the Challenger 2 as a USER key function The Remote provides this function full time through the use of the NET GROSS key See User Guide page 19 for more detail on NET GROSS The USER key is programmable to any of five functions TEST NET GROSS TOTAL UNIT and PEAK If you are making regular use of the Remote there User Guide SYSTEMS WAT is no need to set the USER key to NET GROSS UNITS or TOTAL as these keys are available full time on the Remote unit However there is no harm in having duplicated functions on the USER key if you so desire See Scale Setup for instructions on programming the USER button UNITS The UNITS function is available at power up or by programming the USER key to change between pounds and kilograms The IR Remote provides this function full time by pressing the key See User Guide page 23 for more information on UNITS TOTAL The Total function is available full time on the Remote rather than just a USER key function See User Guide page 15 for mo
27. an be disabled internally Auto Off Mode Prolongs battery life by turning POWER off after 12 minutes or one hour operator determined of no scale activity Auto Sleep Mode Prolongs battery life by dimming LED display after one minute of no activity Unit kg Ib other Units available with custom calibrations Filtering Selectable Low LO Medium HI 1 High HI 2 Totalization Standard Press button or Automatic TOTAL weight up to 99999 X 1000 kg or Ib Peak Uses unfiltered faster reading of A D Set Points Two internal standard Set Points and two ultra bright LEDs on indicator panel Service Counter Two independent 24 bit registers Register 1 updated each time weight exceeds 25 of capacity Register 2 updated each time weight exceeds overload when register 1 exceeds 99 999 or register 2 exceeds 1000 display reads L Cnt for load cell counter Test function shows the two readings in order Connectors Connections are made with PCB mounted terminal strips allowing easy upgrading Cables are brought out through water tight fittings Infrared Remote Controller Available by adding infrared transmitter no modifications are required to the crane unit of these features are housed in a single low profile cast aluminum housing consisting of three sections 1 The front of the scale houses the display controls and all necessary electronics 2 The center section contains the load cell lifting eye and hook 3 The rear of the scale f
28. another gain then press ERG See Gain Table Capacity of Scale i Capacity of Scale 250 Ib 125 kg 5 000 10 2 500 kg 500 Ib 250 kg 10 000 Ib 5 000 kg 1 000 Ib 500 kg 15 000 Ib 7 500 kg 2 000 Ib 1 000 kg Gain Table 6 Display flashes UnLd Unload the scale and press ENERG The scale will display CALC for calculating 7 Display shows LOAd for 2 seconds then flashes the capacity Load the scale with a test weight that is equal to capacity weight or if the test weight is less than capacity enter the actual weight of the test weight by pressing then After the last number of the weight has been entered and the motion indicator is off press The display will show CAL d if successful or CErr if not successful An unsuccessful calibration can be due to an improper gain setting or use of a calibration weight that is too light 8 Display flashes UnLd Unload the scale and press the button Afterwards r CAL will be displayed until the value is determined If successful then the CAL r value is displayed for 10 seconds or until any button is pressed If unsuccessful the display will show r Err for 2 seconds Final Display shows StorE to signify that the calibration parameters are being saved in backup memory When complete the display will show F CAL for the next menu item Press to return to normal scale operation Page 28 MSI 3360 Challenger 2 User Guide
29. at the defective device is returned to the Company freight prepaid In no event shall the Company be liable for the cost of any repairs or alterations made by others except those repairs or alterations made with its specific written consent nor shall the Company be liable for any damages or delays whether caused by defective workmanship materials or otherwise The Company shall not be liable for any personal injury or property damage resulting from the handling possession or use of the equipment by the cus tomer The warranty set forth herein is exclusive and is expressly in lieu of all other warranties express or implied including without limitation any implied warranties of merchantability or fitness or of any other obligations or liability on the part of the Company The liability of the Company under this warranty is limited solely to repairing or replacing its products during the warranty periods and the final judgement and disposition of all claims will be made by MEASUREMENT SYSTEMS INTERNATIONAL INC MSI 3360 Challenger 2 User Guide 45 Te M STEMS INTERN AT N E PROPER LOADING PROCEDURES DO NOT INTERFACE HARDWARE DO NOT USE MULTIPLE ATTACHMENTS SHACKLES HOOKS ETC THAT ARE OVERSIZED AND RESTRICT SINGLE POINT LOADING AND SELF ALIGNMENT WHICH CAN RESULT IN OFF AXIS LOADING 1 DO NOT PUSH OR PULL ON LOADED USE HARDWARE THAT CREATES SINGLE SCALE 2 DO
30. ation for loading operations Easily customized for special applications PEAK Mode for stress analysis Two Set Points can be set for any in range weight for operator alerts or process control Set Point Low Battery Peak Annunciators Warning Hold Motion Detect LOBATT MOTION 5 Digit Weight Indication Calibration Screw Access GROSS TOTAL X1000 Center of Net Gross Totalizing 6th Digit Zero Mode Mode Substitution Light Emitting Diode Display User Guide SYSTEMS WAT I NM AL LO BATT GROSS PEAK MOTION LO BATT PEAK MOTION GROSS Overload Display Error flashes alternately with LOAd In the event of an overload condition the display will flash a warning to the operator TOTAL 1000 TOTAL 1000 5 Options which you may have ordered with your Challenger 2 may include the following Infrared Remote Controller 115 or 230 VAC input power UNPACKING When unpacking the scale from the shipping container ensure that all assembly parts are accounted for Check the scale for any visible damage and immedi ately report any damage to your shipper It is advisable to use the original shipping container when shipping or transporting the Challenger 2 ASSEMBLY Identify and locate the following see MSI 3360 assembly parts drawing 1 Battery pack 2 Cotter pin 3 Load cell 4 Hook
31. ctivity is detected Push any key on the IR Remote to wake the Challenger up This activity is independent of the display key Operation 1 Aim the IR Remote at the Challenger Front Panel 2 The DISP key will cycle the display brightness from lowest to highest and then back to lowest SETPOINTS Function Setpoint values are entered directly from the numeric keypad Operation For a Setpoint that responds when the weight is greater than or equal to the Setpoint Value 1 Aim the IR Remote at the Challenger Front Panel 2 Press the numeric keys for the Setpoint value desired Verify as you enter the number that each number appears on the Challenger display 3 Press the SETPT1 SETPT2 key to input the number as MSI 3360 Challenger 2 User Guide 39 Ve A SUREM EM SYSTEMS a Setpoint value message StPt 1 or StPt 2 appears for 2 seconds verifying the scale has stored the setpoint value in memory For a Setpoint that responds when the weight is less than or equal to the Setpoint Value 1 Aim the Remote controller at the Challenger Front Panel 2 Press the numeric keys for the Setpoint value desired Verify as you enter the number that each number appears on the display 3 Press the or key to input the number as a Setpoint value 4 While the message StPt 1 or StPt 2 appears immediately push the key The display will change from StPt 1 to LESS
32. e scales Action To change the Filter settings 1 Hold in the button while turning on the power 2 Press 1 the message reads Filtr 3 Press The currently selected filter will flash 4 Press to change from LO low HI 1 medium or HI 2 high settings When the desired Filter setting is displayed on the message display press Or press to exit without changing the setting Final Setup menu is displayed Press to return to normal operation Or press to scroll to a different USER Setup operation LED BRIGHTNESS LEdS Page 22 MSI 3360 Challenger 2 Function Sets the LED brightness to one of three fixed settings or to an automatic setting that adjusts brightness according to ambient light and also adjusts to the lowest level of brightness when the scale is inactive for 1 minute Auto matic Sleep Mode Automatic Sleep Mode is recommended for minimizing battery consumption Action 1 Hold in the 9958 button while turning on the power 2 Press until LEdS is displayed 3 Press zi n3 and currently selected brightness will flash 4 Press change from AUTO LO HI 1 or HI 2 settings 5 Press xum ul3ll to store or press AGS to exit without User Guide SYSTEMS INTERNATI changing Final Setup menu is displayed Press to return to normal operation or press to scroll to a different USER Setup operation Note
33. eatures a quick access battery compartment WARNING The scale has a safe mechanical overload of 20096 and an ultimate overload of 50096 Overloads greater than 50096 may result in struc tural failure and dropped loads Dropped loads may cause serious personal injury or death MSI 3360 Challenger 2 User Guide 5 He A S WR Sxs ulis 1 Ww AL FEATURES The LED display provides excellent readability from full sunlight to total dark ness Page 6 MSI 3360 Challenger 2 Designed to meet or exceed US and International Standards Up to 100 hours of weighing time utilizing Automatic Sleep Mode Automatic Power Off conserves battery life by sensing no activity after 12 minutes or one hour determined by operator and turns Power off Automatic Sleep Mode preserves battery life by dimming the LED display after one minute of no activity Rugged construction throughout Buttons are sealed and rated for over 1 million operations Precise high resolution 2500 division standard and up to 10 000 possible 20 bit A D conversion coupled with advanced 16 bit micro controller provides world class features and accuracy Five large 1 2 inch LED digits for clear weight readings from a distance Easy to maintain Full digital calibration assures reliable repeatable measure ments Selectable for kgs Ibs unless prohibited by regulations Automatic or manual weight totaliz
34. ff An additional battery saving feature is the Automatic Sleep Mode This feature preserves battery life by dimming the display after 1 minute of no scale activity Charging time for a completely discharged battery is approximately 16 hours A spare battery pack is recommended to keep the Challenger 2 in continuous operation Note To obtain maximum service life from your batteries they should be stored between 4 F and 122 F 20 C and 50 C Stored batteries should be recharged every three months Battery is fully charged when the status indicator is flashing BATTERY CHARGER 1 Remove the battery from the scale 2 Plug battery charger into an AC power receptacle of the same voltage and frequency stamped on the plug in module 3 Slide the battery charger connector plate over the top of the battery until the battery terminals mate with the charger connectors as seen through the two observation holes 4 Approximately 16 hours is required to recharge a fully discharged battery MSI 3360 Challenger 2 User Guide 9 VE A EM y S amp S 1 Ww AL Shown is the charger and battery assemblies Note that a second battery is recommended to enable you to use your scale continuously Keep one on the charger while the other is in service Note To obtain maximum service life from your batteries the manufacturer suggests recharging after each 20 hours of use Continuous deep discharging will i
35. ft front panel of the scale Insert a probe such as a blunt tip screwdriver to depress the underlying button The display will show C SEt for 1 second 2 Press r Ctl is displayed 3 Press xui The currently selected Remote Control Mode On Off will flash 4 Press to choose ON 5 Press to store The menu will change to AZt 6 Press to return to normal operation Final The IR Remote is now enabled Caution Do Not enable the Remote Control if the option was not purchased with the scale BATTERIES The IR Remote requires two standard AAA cells Battery life will depend on the frequency of use but could last up to a year Anytime the batteries are changed the IR Remote Access Codes must be programmed Page 34 MSI 3360 Challenger 2 User Guide SYSTEMS TWTeER WAT SET REMOTE CONTROLLER ACCESS CODES The IR Remote contains internal access codes that must be programmed to function with the Challenger 2 This procedure must be followed after changing batteries 1 Press and hold the button until the indicator light goes on then release the button Press and release the button Press and release the HJ button Press and release the QJ button Press and release the button The indicator light will go out 2 Press and hold the button until the indicator light goes on then release the button Press and release the button Press and
36. ien 18 Phold 18 NET GROSS 19 Scale Setup 19 Quick USER Setup Guide 20 Punctiom PUn 21 Automatic Power Off A OFF 21 Filter xiii terret terere tec ge 22 LED Brightness LEAS 2020422221 22 Unit ko Ib 23 Setpoints 51 StPt2 23 Set Point Entty neret tette 23 To Disable a Set Point eese 24 TOTAL etn 25 System Imtialize 22 2 temen 25 Calibration Setup rettet ters 26 Quick Calibration Setup Guide 27 Calibration CAT iei erede nidos 27 io rre mte nn 27 Subsequent Calibration sess 29 Fine Calibration F CAL 29 Calibration with Calibration Offset Resistor TED 30 Standards SEAnd 3 2 ERE Reus 31 Remote Control 31 Auto ZERO Tracking 10 32 INFRARED REMOTE OPTION 33 Introduction E 33 Remote Controller Operation 34 To Enable Remote Controller 34 Batteries ducas de E Are E 34 To Set Remote Controller Access Codes 35 The Scale and 2nd Keys
37. layable increment Resolution Standard displayed resolution 2500 counts Internal A D resolution 1 048 576 counts 20 bits Capacities Ib 250 500 1000 2000 5000 10 000 15 000 kg 125 250 500 1000 2500 5000 7500 Power Battery operated rechargeable sealed lead acid battery pack standard Challenger Charger up to 100 hours of battery life with Automatic Sleep Mode and Automatic Power Off Display 5 digit large 1 2 in 30 mm high numeric red GaAIAs Light Emitting Diode LED Operating Temperature 4 F to 122 F 20 C to 50 C NTEP range 10 C to 40 C Operating Time 50 hours min 100 hours max depends on operating mode Enclosure NEMA 12 IP54 powder coated alodined cast aluminum Load Cell Standard 350 Q Bridge Indicators LO BATT PEAK MOTION Center of ZERO NET GROSS TOTAL X1000 kg Ib StPt1 StPt2 Function buttons POWER Turns scale on and off ZERO Zeros applied load up to 100 of capacity TARE Tare Instores currentGROSS weight as TARE All subse quent readings are Net Weight Tare Out returns the weight display to GROSS Mode USER Programmable multifunction button for use as TEST TOTAL UNIT PEAK NET GROSS Front panel calibration switch located behind seal screw Initiates full digital calibration procedure User button options TEST default TOTAL UNIT PEAK NET GROSS OFF Page 4 MSI 3360 Challenger 2 User Guide SYSTEMS NTERNA NM AL Auto Zero Tracking Standard c
38. le use the Medium or High Filter setting see Filter Setup The scale will accept the TOTAL button when in motion but will not TOTAL until the Motion indicator is off 2 Only positive readings can be accumulated 3 After a weight is totaled the weight on the scale must return below 1 of full scale capacity relative to GROSS ZERO or NET ZERO before another weight can be added to the TOTAL This assures that a weight on the scale is only added to the scale once 4 Press WAAN USER button must be programmed as TOTAL see Scale Setup 5 If the Motion indicator is the TOTAL indicator will begin blinking MSI 3360 Challenger 2 User Guide 15 He A S WR Sf S S amp S and continue blinking until motion stops and the Motion indicator turns off The current weight is added to the TOTAL register and the TOTAL weight is displayed 6 The TOTAL indicator center right of display will also illuminate for 4 seconds providing reinforcement that the TOTAL command was accepted Final After totaling normal weighing resumes AUTO TOTAL Function Allows for Auto accumulate Use 1 When a weight threshold of 2 of Full Scale FS above NET ZERO or GROSS ZERO is exceeded the accumulate function operates automati cally 2 When a weight that meets the minimum acceptable limit settles no Motion the TOTAL indicator will flash 3 times 3 Ifthe weight changes to a new settled value the
39. nhibit maximum battery life cycle estimated at 2000 cycles OPERATION GUIDE Challenger 2 operation is controlled by four push buttons located below the display window GROSS TOTAL 5000 x 2 kg 10000 x 5 Ib POWER ZERO TARE USER Decimal Pt Clear Exit Enter Select Activating the push buttons with a sharp object such as a screwdriver may cause permanent damage or broken seals Page 10 MSI 3360 Challenger 2 User Guide SYSTEMS WAT N The four push button functions are as follows POWER POWER Function Turns the scale on off toggle action Use 1 The battery pack must have enough charge to ensure accurate operation gt 5 4 2 Ambient temperature must be greater than 30 C and less than 60 C 22 F to 140 F Action 1 Press and hold for one second 2 Display check All segment and indicator LED s are illuminated for one second 3 Numerics display the software version number for one second 4 Any weight on the scale up to 5 of capacity will be automatically zeroed 5 During display test all internal operations are checked and any non conformance causes an Error message display Final Scale reads the current weight in the last set Mode NET GROSS PEAK NET PEAK GROSS ZERO ZERO Function Sets scale to ZERO Use 1 The scale must be stable The scale will not ZERO if the motion detect indicator is on 2 The scale must be in the GROS
40. of Tare anes PRINT ID Set Points and calibration constants lt D gt Adds full time NET GROSS key SETUP SETPTI SETPT2 Adds full time TOTAL and VIEW TOTAL 5 keys for easy access to the EXIT SETPOINT FUNCTION totalize functions Adds full time UNITS key Allows full time TARE and adds full time IR REMOTE RECALL TARE keys Adds full time SETPOINT 1 and SET Measurement Systems POINT 2 keys Display Brightness control RIN 501546 Easy access to Setup functions MSI 3360 Challenger 2 User Guide 33 HE A SUREM SYSTEMS REMOTE CONTROLLER OPERATION In general the IR Remote keys function the same as the front panel switches on the crane scale The ZERO TARE and USER keys function the same as the front panel keys The enhanced functions or any functional differences are detailed in the following sections The Infrared Sensor is located to the left side of the front panel In general the operator must aim the IR Remote unit directly at the front panel of the scale As distance increases accurate aiming becomes more critical The IR Remote will not function in direct sunlight and will have very limited range outdoors TO ENABLE REMOTE CONTROLLER Function Used to enable the Remote Controller This procedure is performed directly on the scale not on the IR Remote Procedure 1 Remove the seal port on the le
41. of the Challenger 2 Disabling the AZt is only intended for certification testing Under no other circumstances should the Scale be used with AZt disabled Page 32 MSI 3360 Chalenger2 User Guide SYSTEMS TWTeER WAT INFRARED REMOTE OPTION INTRODUCTION The MSI IR Remote Controller option provides the user of the Challenger 2 Crane Scale complete remote control of all scale operation functions In addition a numeric keypad allows the operator easy entry of tare values and set point values Several features which are only available on the Challenger s USER key are directly accessible on the IR Remote These include NET GROSS VIEW PWR 3 TOTAL and SETUP Also maneuvering in menus is simplified by the use of a SCALE SETUP key combined with MENU UP USER MENU DOWN UNITS ENTER and EXIT SETUP keys USER ZERO TARE The IR Remote is available for any Chal lenger 2 Crane Scale The necessary Infrared Sensor and supporting firmware is standard and therefore no modification of the scale is 1589 PHP necessary other than enabling the remote function 5MNO gPGR Features 750 Reliable Infrared remote control of Power Zero Tare etc All scale functions are controllable from a DISPALPHA distance of up to 25 feet 8 meters e Adds easy numeric keypad entry
42. ondition is determined Zero requires adjustment Out of Zero range AZt turned off Scale electronics do not stabilize after turning on probably due to rapid temp erature change ZERO requires adjustment Defective load cell or electronic circuit board Scale not Zeroed before applying weight Requires recalibration Ib kg in wrong selection Solution Recharge Replace Clean connections Requires authorized service Press POWER firmly and hold until Power turns on Remove and re insert battery If problem persists authorized service is required Requires authorized service Requires authorized service See 2A B above Check calibration Requires authorized service Plug in Normal See Operation Guide Recalibrate Remove dead load and re zero Go into Cal menu and turn AZt on Warm up scale for 5 minutes then re Zero Follow Calibration Section Requires authorized service Depress ZERO before applying weight See Calibration Set to correct selection ONAL 4 S UR s Wl N 10 1 12 18 14 15 Problem Display reading not stable Battery charger indicator does not come on when discharged battery is inserted Display is LCnt1 and or LCnt2 at Power on Display toggles between Error and A2dLo or A2dHi Display toggles between Error and buttn Display toggles between Error and LOAd IR Remote does not work Some IR Remote keys do not work
43. r an overload alarm for example you would choose greater Press f 5 The current Set Point value will flash 6 Press to enter the first digit of the desired Set Point weight value Value cannot exceed scale capacity Press Note Press Gea to enter a decimal point 7 The next significant digit flashes Press to enter the next digit Press twice after the last character to complete the Set Point entry Final Setup menu is displayed Press to return to normal operation Or press to scroll to a different USER Setup operation TO DISABLE A SET POINT Function To turn off a Set Point Action 1 Hold in the 998 button while turning on the power 2 Press to identify the Set Point you wish to disable Set Point 1 will be used for this example 3 When the display reads StPt1 press murum 4 Press until the message OFF appears Press Ez GI3 This disables the Set Point and returns you to the Set Point menu Final Setup menu is displayed Press to return to normal operation Or press to scroll to a different USER Setup operation Page 24 MSI 3360 Challenger 2 User Guide SYSTEMS WAT TOTAL totAL Function Allows the TOTAL function to be set or disabled The USER button must be programmed to TOTAL in order to view the current TOTAL value or to use Manual TOTAL Action 1 Hold in the WAAN button while turning on the power 2 Press MZA until
44. re details Note When the Challenger 2 is set up for AUTO total the TOTAL key has no function VIEW TOTAL VIEW gt Function The VIEW key gives the user immediate access to the Total Register Operation 1 Aim the IR Remote at the Challenger Front Panel 2 Press key 3 Press the key The TOTAL LED will light and the Total weight will appear on the numeric digits for 4 seconds 4 Don t forget to press the key to reactivate the main IR Remote key functions CLEAR LAST TOTAL Function The user may clear the last totaled weight from the Total Register Operation If the USER key is programmed to Total then MSI 3360 Challenger 2 User Guide 37 He A S WR WAT 1 Ww AL 1 Aim the IR Remote at the Challenger Front Panel 2 Press the SUSE key 3 Press the key If the USER key is not programmed to Total then 1 Aim the IR Remote at the Challenger Front Panel 2 Press the PITE key 3 Press the key 4 Immediately press the key 5 Immediately press the key CLEAR ALL TOTALS Function The user may clear all totaled weights from the Total Register Operation If the USER key is programmed to Total then 1 Aim the IR Remote at the Challenger Front Panel 2 Press WAAN key 3 Press the 3598 key If the USER key is not programmed to Total then 1 Aim the IR Remote at the Challenger Front Panel 2 Press the MXP key 3 Press
45. s to choose desired standard 4 Press to store or press to exit without changing Final The calibration menu is displayed Press to return to normal operation or press to scroll to a different Calibration Setup operation REMOTE CONTROL r Ctl Function Used to enable or disable the use of a Remote Control Action 1 Press Eater Era until r Ctl is displayed 2 Press and currently selected Remote Control Mode On Off will flash 3 Press to choose ON or OFF 4 Press ELI to store or press guit to exit without changing MSI 3360 Challenger 2 User Guide 31 He A SUREM y ulis WAT Ww AL Final The Calibration menu is displayed Press XO to return to normal operation or press to scroll to a different Calibration Setup operation Caution Do Not enable the Remote Control if the option was not purchased with the scale AUTO ZERO TRACKING AZT Function To enable disable AZt for certification testing Action 1 Press ZAC until AZt is displayed 2 Press and the currently selected AZt Mode On Off will flash 3 Press to choose the desired AZt Mode On Off 4 Press to store or press to exit without changing Final The Calibration menu is displayed Press to return to normal operation or press to scroll to a different Calibration Setup operation Caution Disabling Auto Zero Tracking will degrade temperature and drift performance
46. ssed any button or the detected weight is in Motion the time limit is reset Therefore the scale will stay on indefinitely if the weight is changing or any button is pressed at least once With A OFF disabled the scale will remain on only pressing POWER will turn it off Action To Change the A OFF Mode 1 Hold in the WAAN button while turning on the power 2 Press until the display reads A OFF 3 Press Exi uil You are now in the A OFF menu 4 Press to choose OFF 2 hours 12 minutes or 1 hour When the desired Mode is displayed press or press to exit without changing Final USER Setup menu is displayed Press to return to normal operation or press to scroll to a different USER Setup operation MSI 3360 Challenger 2 User Guide 21 He A S WR FILTER Filtr Function Allows three levels of filtering to stabilize readability of the weight display Use 1 Use LO low setting for most scale applications It settles fastest and is intended for general use 2 Use HI 1 medium setting when the scale is being used under condi tions that cause light to medium swinging 3 Use HI 2 high setting when there is significant scale motion There is a time penalty when using the HI 2 setting The operator should wait at least 5 15 seconds to ensure that the final reading has settled motion indicator off Not available in some Legal For Trad
47. totAL is displayed 3 Press and the currently selected total mode will flash 4 Press until the desired TOTAL Mode OFF PrESS manual or AUTO automatic is displayed 5 Press to store or press to exit without changing Final Setup menu is displayed Press to return to normal operation Or press to scroll to a different USER Setup operation SYSTEM INITIALIZE Caution Do Not Initialize the scale for routine calibration System initialize is only needed when the circuit boards are replaced This procedure should be initiated only by the factory or any authorized service representative Function Clears the internal Calibration settings Usually used for board replacement troubleshooting or load cell replacement This procedure does not alter factory feature settings Action 1 Turn off the scale by pressing the key 2 Remove the seal port on the left front panel of the scale Press the switch button in the hole with a probe such as a blunt tip screwdriver and simultaneously press the key 3 The scale will prompt the user for confirmation by displaying SurE 4 Pressing will continue the system initialization Pressing will abort the system initialization without resetting any parameters MSI 3360 Challenger 2 User Guide 25 HE A S WR EM NW AT Ww AL CALIBRATION SETUP Function Calibration Setup is used to initiate calibration or to set seldom changed f
48. up memory When complete the display will show F CAL for the next menu item Press to return to normal scale operation FINE CALIBRATION F CAL Function Fine Calibration F CAL is for minor adjustments to calibration and is usually not necessary It is useful however in hydraulic calibration fixtures for fine adjustments MSI 3360 Challenger 2 User Guide 29 He A S WR Y SYSTEMS NW AT Ww AL Action 1 Press to scroll to the F CAL message Press 2 Pick up a test weight of or set the hydraulic tension to at least 25 of capacity The weight is indicated on the numeric digits The display will flash LOAd if the weight is under 25 WARNING Fine Calibration will not function unless the scale is loaded at 25 of capacity or more 3 Press to cause the displayed reading to move up slightly Press the button to cause the reading to move down Each press of the 38591009959 or feas button causes the calibra tion to shift by 1 4 displayed count When the displayed reading is acceptable press Final Fine Calibration is complete press 104359 to return to normal scale operation or press to choose another Calibration Setup function CALIBRATION WITH CALIBRATION OFFSET RESISTOR CAL r Function Calibration with the Cal Resistor CAL r is for emergency calibration only when test weights or a calibration system are not available