1. AY isl gt OMEGA Sensatec Troubleshooting OMMUNICATIONS LTD If Sensatec bed alarm continues to alarm when resident is not applying weight to the bed pad check the following 1 Make sure that all connectors are plugged in power and bed pad 2 Check the led light on the front of the Sensatec controller If flashing red the battery requires replacement If rapidly flashing green the controller is alarming proceed to next step 3 Check the bed pad for the following Proper position on resident s bed place under buttocks and lower back e Expiry date replace if expired Inspect bed pad if folded creased replace pad Check for excessive weight on bed pad multiple catch pads Make sure controller is attached to the bed using the supplied strap If controller continues to alarm after confirming the above steps remove controller from residents room and replace with another unit supplied from spares check with supervisor how to obtain spare e healthcare omegacom ca e http omegacom ca Omega Communications Ltd e 888 860 8016