1. 7 3 SAFETY 10 Ere Lern 10 4 OPERATION 14 Installing the Mattress Replacement 5 4 4 22 14 Installing the Side EUNT CER HUE a a l6 Installing the Power Unit 16 Connecting the lose ataca e n E aa IN pee a 16 Connecting the Power Cord ce ee asa aedi ER DR M e URN cel ne uae P De diee b ua ada 18 Using the Front Panel eo repeteret RR ge RUD ERI QE ov sette Ree Mr vaipundeldsdecnahsas EEE EEE Oa 19 ds Ca tee E eR ae 26 pi da 27 Transferring Patient From To 8 8 28 Transferring Patient From Toia Wheelchair ttt hte det aia li bb Hd ER de p actae a ol 28 RUE 29 5 MAINTENANCE AND TROUBLESHOOTING 3l Cleaning the System esc pO UU Ee aa a da ee a idee dedere t RM de
2. Lock Mode Pressing this button locks out all control unit functions including the On button to prevent any tampering with the settings Press the button for approximately 3 5 seconds for the Lock LED to activate The Lock LED illuminates when in this mode Alarm Silence mode Pressing this button silences the alarm that sounds in the event of power failure or when the hose is disconnected from the control unit The Alarm LED illuminates when in this mode Part No 1148139 25 Invacare MA90Z Series Mattresses 4 OPERATION Power Fail LED In the event of power outage an alarm sounds and the Power Fail LED flashes AMBER The control unit has internal memory and retains the previous settings during the power outage During a power outage the MA90Z MA90ZB42 and MA90ZB48 mattress retains the air as long as the mattress is connected to the control unit Low Pressure LED In the event that the mattress hose disconnects an alarm sounds and Low Pressure LED flashes AMBER Once the low pressure problem is fixed the control unit resumes operation in the previously set mode Displaying the Total Run Time MA95Z Power Unit Every time the control unit is plugged in the LCD screen displays the Run Time in hours and minutes For example 2 584 hours and 25 minutes displays as 002584 Hours 000025 Minutes To display the Run Time during use Ensure the control unit is plugged in and the Standby LED is On 2 Pre
3. fw FIGURE 2 Installing the Power Unit 4 4 Connecting the Hose CAUTION Ensure that the hose connecting the control unit to the mattress is routed such that it cannot be stepped on kinked squeezed or otherwise damaged Invacare MA90Z Series Mattresses 16 Part No 1148139 Locate the hose at the foot end of the mattress 2 Locate the control unit connector on the right side of the control unit 3 Perform one of the following 907 Power Unit Perform the following steps Detail i Squeeze and hold the tabs on the hose connectors ii Insert the hose connectors into the control unit connector 4 OPERATION iii Push the hose connectors onto the control connector until an audible click is heard MA95Z Power Unit Perform the following steps Detail i Lift and hold the connector cover up ii Push the hose connector onto the control connector iii Ensure the hose connector is flush with the control unit connector DETAIL MA90Z The audible click indicates that the hose connector s are properly engaged with the control connectors DETAIL MA95Z CONNECTOR CONNECTOR MA95Z MA95ZB42 MA9OZ MA95ZB48 Hose MA90ZB42 Connector and MA90ZB48 Hose Connector Connector Cover ontrol Unit Control Unit Connectors FIGURE 3 Connecting the Hose Part No 1148139 17 Invacare MA90Z Series Mattresses 4 4 5 Connecting the Powe
4. User Manual Gay Invacare 44902 Series Mattresses MA90Z Lateral Rotation with Alternating Pressure and On Demand Low Air Loss System MA90ZB42 Lateral Rotation with Alternating Pressure and On Demand Low Air Loss Bariatric 42 MA90ZB48 Lateral Rotation with Alternating Pressure and On Demand Low Air Loss Bariatric 48 Invacare MA95Z Series Mattresses MA95Z Lateral Rotation with Alternating Pressure and True Low Air Loss System MA95ZB42 Lateral Rotation with Alternating Pressure and True Low Air Loss Bariatric 42 MA95ZB48 Lateral Rotation with Alternating Pressure and True Low Air Loss Bariatric 48 This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future reference Yes you can 2011 Invacare Corporation All rights reserved Republication duplication or modification in whole or part is prohibited without prior written permission from Invacare Trademarks are identified by and 9 All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted Invacare MA90Z Series Mattresses 2 Part No 1148139 5 GENERAL 4 MALUI ee T oa 4 Limited Acura D aa ada cai d t a o a oa a aa 5 2 OVERVIEW 6 Mabel iL OCation 6 Typical Product Parameters peu
5. Press and hold the Turn Angle and Inflate buttons The Fowler LED is lit and L displays in the LCD Display ii Raise the bed to 45 The control unit will beep and will return to Standby Part No 1148139 23 Invacare MA90Z Series Mattresses 4 iii Perform one of the following Cancel Press the Turn Angle button to exit Fowler Transmitter Setup without linking to a transmitter Save Wait 60 seconds without pressing a key MA95Z Power Unit Perform the following steps i Press and hold the Hold and Inflate buttons ii AUTO FOWLER LEARN MODE will display PLACE BED IN FLAT POSITION will display iv CONTINUE gt SELECT will display v Press the Select button vi PLACE BED IN FOWLER UPRIGHT POSITION will display vii TRANSMITTER DETECTED will display viii The control unit will beep ix STAND BY will display Mode Button MA90Z Power Unit Press the Mode button t to select the Max Inflate Mode or the Low Air Loss mode Inflate Mode In this mode the mattress inflates rapidly to maximum firmness pressurized to 35 6 mmHg A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active After 30 minutes the Max Inflate mode deactivates and the control unit defaults to the previous setting It takes 5 10 minutes for the MA90Z mattress to inflate fully inflation time depends on size of mattress Max Inflate
6. cycle duration can be set from to 99 minutes In this setting the mattress will remain static at the current pres sure settings for the time selected Then every two minutes for the time selected the pressure in the mattress will decrease by 20 for five seconds This pulse cycle repeats itself for the selected time i e if 10 minutes is selected the mattress will initially remain static for 10 minutes Then every two minutes the pressure the mattress will decrease by 20 for five seconds the mattress will then inflate back to the set pressure setting and remain static for the time selected and the cycle is repeated Part No 1148139 21 Invacare MA90Z Series Mattresses 4 FUNCTION DESCRIPTION FOWLER UPRIGHT Fowler Mode The pressure will increase to prevent the patient from bottoming out 3 Use the Select button to move the pointer to the times or attributes available for the selected function 4 Use the Firmer or Softer buttons to change the times or attributes Hold Button MA95Z Power Unit Press the Hold button 0 to pause the mattress in the current position The mattress will remain in this position until the Hold button is pressed again Firm Soft Buttons l Select comfort pressure settings by pressing Firm A or Soft M buttons Soft Button Pressing this button reduces the pressure the mattress Firm Button Pressing this button increases the pressure in the mattr
7. filter cover to the back of the control unit using the two screws Part No 1148139 33 Invacare MA90Z Series Mattresses 5 MAINTENANCE AND TROUBLESHOOTING Filter Cover Screws FIGURE Changing the Filter Invacare MA90Z Series Mattresses 34 Part 1148139 5 3 Troubleshooting PROBLEM Mattress not inflating Not alternating properly 5 MAINTENANCE AND TROUBLESHOOTING Mattress hose disconnected SOLUTION Connect hose connectors and lock them in place Air hose kinked or split Unkink hose or replace split hose Major leak in the air cushions or overlay pad Replace leaking air cushions or overlay pad Kinked or split manifold Unkink manifold or replace split manifold Has power and fuse is good control unit does not come on Send unit back for repair Not alternating solenoid or valve mal function Send control unit back to factory for repair No air pump or blower malfunction Send unit for repair Part 1148139 Control unit off Check power source and turn on unit Power cord disconnected Connect power cord to the power source No power in the power source Check power source has power power and turn it On Power outage Wait till the power source has power Blown fuse 35 Replace blown fuse with an equivalent fuse Invacare MA90Z Series Mattresses Part 1148139 Rev D 1 11 Invacare Corporation US
8. invacare com or your dealer p n 1150708 Rev A This label is on the foot end of the mattress cover not shown Invacare MA90Z Series Mattresses 6 Part No 1148139 22 Typical Product Parameters 2 OVERVIEW MA90Z power unit is also used MA90ZB42 MA90ZB48 systems MA95Z power unit is also used on MA95ZB42 and MA95ZB48 systems Electrical Parameters INPUT VOLTAGE AC INPUT FREQUENCY CURRENT 2 MAXIMUM POWER CONSUMPTION 30 10 W 150 10 W CIRCUIT PROTECTION Dual fused 250 V A fast blow fuses Dual fused 250 V 5 A fast blow fuses MODE OF OPERATION Performance Parameters WEIGHT CAPACITY STANDARD MATTRESS BARIATRIC MATTRESS 42 inch 48 inch Part No 1148139 Continuous 350 Ibs 650 Ibs 1000 Ibs Invacare MA90Z Series Mattresses 2 OVERVIEW PRESSURE ZONE MAXIMUM FLOW 50 15 LPM 1275 100 LPM MAXIMUM PRESSURE 35 5 mmHg MAXIMUM FLOW TIMER 30 minutes 15 minutes SUPPORT SURFACE INFLATION TIME 5 10 minutes 25 60 seconds PATIENT COMFORT CONTROL PRESSURES SOFT PRESSURE FIRM PRESSURE 8 4 mmHg 32 4 mmHg ROTATION TIME 5 minutes 4 hrs 10 20 30 60 minutes ROTATION ANGLE 0 30 5 0 40 5 Invacare MA90Z Series Mattresses Control unit and mattress have Latex F
9. s Perform one of the following fblood is present decontaminate the product i Remove all visible soil with disposable paper towels ii Scrub the area with freshly prepared effective phenolic detergent disinfectant solution Part No 1148139 3l Invacare MA90Z Series Mattresses 5 MAINTENANCE AND TROUBLESHOOTING If blood is not present remove any soil from the cover with paper towels If soiled the cover should be removed cleaned and decontaminated 8 Perform the following steps to clean the mattress cover Using a clean sponge or paper towel wipe down the cover surface with a dilute detergent solution of quaternary cleaner disinfectant or other germicidal detergent solution B Remove the cover and launder it using the following method A CAUTION Laundry workers should always wash their hands before working with clean bedding DO NOT overload the machine DO NOT use chlorine bleach because it may damage the fabric coating High air temperatures will damage the fabrics and void the Invacare warranty i Place the cover in a washing machine ii Wash with warm water below 120 F iii Add detergent and disinfectant according to the manufacturer s instructions iv Remove excess water v Set the dryer to the lowest setting below 120 F vi Dry the cover until it is completely dry 9 Perform the following steps to clean the control unit and hose fittings Wipe all controls chassis and hose fittings
10. A One Invacare Way Elyria Ohio USA 44036 2125 800 333 6900 www invacare com Canada 570 Matheson Blvd E Unit 8 Mississauga Ontario L4Z 4G4 Canada 800 668 5324 Yes you can
11. D FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Part No 1148139 5 Invacare MA90Z Series Mattresses 2 OVERVIEW 2 Overview 2 1 Label Location DANGER EXPLOSION HAZARD DO NOT use in the presence of flammable anesthetics CAUTION Equipment should be connected to a properly grounded receptacle 3 prong Risk of Electrical shock DO NOT remove back Disconnect air hose before administering CPR This label is on the back of the unit and also contains the serial number ATTENTION Clean Filter Every 5 months and Whenever Dirty Remove Filter Wash Dry amp Replace MA95Z MA95ZB42 and MA95ZB48 only WARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s Operator s Manual prior to using this equipment Invacare product manuals are available at www
12. R Pe 95 PANEL 0 O O STANDBY Alternating Rotation and Pre Turn Time Minutes Center Left Power Button Right Left e FUNCTION Down e ROTATION Lock Alarm Button Rotation Button Down Soft Button LCD Up Firm Button FIGURE 4 Using the Front Panel Max Inflate Button Part No 1148139 19 Invacare MA90Z Series Mattresses 4 OPERATION Power Button l To turn the control unit on or off press and release the Power button O Once the unit is plugged in an AMBER LED on the control unit is lit indicating that the system is in STAND BY mode Once the Power button is pressed and released a GREEN LED illuminates indicating that the control unit is on If the power comes on after a power outage the control unit will go through its system initialization routine for a few seconds and then resume the desired function Turn Button MA90Z Power Unit Press the Turn button to select one of three rotation modes or to set the unit to No Turn Static mode The rotation modes are right left or both and all three modes include alternating pressure therapy When No Turn is selected all of the air cushions are maintained at a constant pressure Turn Time Button MA90Z Power Unit Press the Turn Time button amp g to select one of four rotation times The rotation times are 10 20 30 and 60 minutes Turn Angle Button MA90
13. SORIES WARNINGS Invacare products are specifically designed and manufactured for use conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products NOTICE THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE Check all parts for shipping damage and test before using In case of damage DO NOT use Contact Invacare Carrier for further instruction Contraindications A WARNING ALWAYS consult the patient s physician before using the MA90Z and MA95Z systems Invacare MA90Z Series Mattresses 10 Part No 1148139 3 SAFETY Installation A WARNING The MA90Z MA95Z series including bariatric is recommended to be installed on medical bed frames with bed side or assist rails It is preferred that the rails be in the raised position whenever a patient is on the bed Health care professionals assigned to each case should make the final determination whether side or assist rails are warranted after assessing patient risks of entrapment and falls in accordance with State patient restraint legislation or facility interpretation of such legislation Controls on the footboard may be obstructed by the power unit on a few bed frames It may be necessary to relocate the power unit Refer to Installing the Power Unit on page 16 Check that air hoses and power cord are clear of moving bed components bef
14. Z Power Unit I Press the Turn Angle button p to select one of four rotation angles The rotation angles are 1 4 gt 3 4 and full rotation 45 Invacare MA90Z Series Mattresses 20 Part No 1148139 4 Rotation Button MA95Z Power Unit l Press the Rotation button to select Static mode Alternating Rotation mode When the LED is on Alternating Rotation mode has been selected In Static mode the LED is off and all air cushions are maintained at a constant pressure Select Button MA95Z Power Unit A Press the Select button 3 move the pointer on the LCD to Function 2 Press the Firm or Soft button until the desired function displays Available functions are FUNCTION DESCRIPTION LEFT Rotation to the Left side The time indicates the rotation duration CENTER Rotation to Center The time indicates the rotation duration RIGHT Rotation to the Right side The time indicates the rotation duration STATIC Static mode All air cushions maintain constant pressure PULSE 10 minute pulse time cycle All of the air cushions in the mattress will remain static at the current pressure settings for 10 minutes then the pressure in the mattress will decrease by 20 for five seconds After five seconds the mat tress will once again inflate to the set pressure settings for 10 minutes then deflate by 20 for five seconds and the cycle is repeated MULTI PULSE The
15. eel locks of the wheelchair before transferring the patient between the bed and the wheelchair Engage the wheel locks of the bed Refer to the owner s manual provided with the bed Engage the wheel locks of the wheelchair if applicable Refer to the owner s manual provided with the wheelchair Press the Max Inflate button to achieve maximum mattress pressure Raise or lower the bed to match the wheelchair height Refer to the owner s manual provided with the bed When the mattress has reached maximum firmness perform one of the following Bedto Wheelchair Transfer Slide the patient onto the wheelchair Wheelchair to Bed Transfer Slide the patient onto the bed Invacare MA90Z Series Mattresses 28 Part No 1148139 4 4 11 Preparing for CPR Procedure For this procedure refer to FIGURE 5 on page 30 Perform one of the following MA90Z Power Unit Press and hold the tabs the hose connector Detail B MA95Z Power Unit Lift the connector cover Detail A 2 Pull the hose connector Details B and C from the control unit connectors MA95Z Only If the hose connector is difficult to release pull the hose connector down and then away 3 907 Only Disconnect the RED CPR connector located the hose Detail D Part 1148139 29 Invacare MA90Z Series Mattresses 4 OPERATION DETAIL A DETAIL B Connector 907 Co
16. ess The patient comfort pressure ranges from Soft 6 4 mmHg to Firm 32 6 mmHg The Comfort Control LED displays the patient comfort pressure levels and provides a guide to the caregiver to set approximate comfort pressure level depending on the patient weight If the patient s weight to height ratio is above average increase the pressure setting by approximately 20 MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from 6 to 32 Invacare MA90Z Series Mattresses 22 Part No 1148139 4 OPERATION Automatic Wireless Fowler Manual Operation of the Fowler Press the Select button 3 to move the pointer on the LCD to Function 2 Use the Firm and Soft buttons until Fowler displays When this mode is activated the control unit increases the pressure in the mattress to prevent the patient from bottoming out About Automatic Wireless Mode When the bed articulates to 45 the transmitter in the head of the mattress signals the control unit to add 25 more air into the mattress to prevent the patient from bottoming out Programming the Automatic Wireless Fowler Transmitter Ensure the transmitter is inside the mattress pocket located on the right side of the mattress at the head end Ensure the mattress is in a flat position Ensure the control unit is plugged in and the LCD displays STAND BY Perform one of the following 907 Power Unit Perform the following steps i
17. h bed rail product entry to access this bed safety guide Refer to the owner s manuals for beds and rails for additional product and safety information After any adjustments repair or service and before use make sure all attaching hardware is tightened securely Assist rails with dimensions different from the original equipment supplied or specified by the bed manufacturer may not be interchangeable and may result in entrapment or other injury Invacare MA90Z Series Mattresses 12 Part No 1148139 Fire Hazard Electrical A 3 SAFETY DANGER Smoking DO NOT SMOKE while using this device This system uses room air for circulation through the mattress A cigarette can burn a hole in the bed surface and cause damage to the mattress Also patient clothing bed sheets etc may be combustible and cause a fire Failure to observe this warning can result in severe fire property damage and cause physical injury or death Smoking by visitors in the room will contaminate the system Therefore visitor smoking is NOT permitted Anesthesia Equipment There is an explosion risk if used with flammable anesthetics Oxygen There is a possible fire hazard when used with oxygen administering equipment other than nasal mask or half bed tent type The oxygen tent should NOT extend below mattress support level DANGER Electrical shock hazard DO NOT remove cover Refer to qualified service personnel Before performing any maintenance to
18. mode can be manually disengaged by pressing the Max Inflate button This will deactivate the Max Inflate LED It is recommended that Max Inflate setting be used during patient ingress egress patient wound care patient turning or patient cleaning Low Air Loss Mode In this mode the mattress goes into on demand low air loss relief mode Invacare MA90Z Series Mattresses 24 Part No 1148139 4 Max Inflate Button MA95Z Power Unit o l Press the Max Inflate button to select the max inflate mode and inflate the mattress rapidly to maximum firmness pressurized to 35 6 mmHg A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active After 15 minutes the Max Inflate mode deactivates and the control unit defaults to the previous setting It takes 25 60 seconds for the MA95Z mattress to inflate fully inflation time depends on size of mattress It is recommended that Max Inflate setting be used during patient ingress egress patient wound care patient turning or patient cleaning 2 Perform one of the following e MA90Z Power Unit Press the Max Inflate button again to manually disengage Max Inflate mode 957 Power Unit Select a different setting to exit Max Inflate mode or press the Max Inflate button again This will deactivate the Max Inflate LED Lock Alarm Silence Button Press the Lock Alarm Silence button A to select one of the following modes
19. model number and the date of purchase indicate nature of the defect and if the product is serialized indicate the serial number Invacare Corporation will issue a return authorization The defective unit or parts must be returned for warranty inspection using the serial number when applicable as identification within thirty days of return authorization date DO NOT return products to our factory without our prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS IN ADDITION THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE OR PRODUCTS MODIFIED WITHOUT INVACARE S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE S CONTROL PRODUCTS REPAIRED BY ANYONE OTHER THAN AN INVACARE DEALER SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER EXPRESS WARRANTIES WHATSOEVER WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AN
20. om the bed 2 If necessary lower the side rails to facilitate installation of the mattress 3 Unroll the powered mattress and place it on the bed frame Ensure that the hose is towards the foot end of the bed 4 Use the buckle straps to secure the powered mattress to the bed deck in the following locations Head End Head End Bed Deck Center Center of the Bed Frame Foot End Foot End Bed Deck Part 1148139 Powered Mattress Head End Mattress cover not shown for clarity Buckle Straps Foot End Hose FIGURE Installing the Mattress Replacement System 15 Invacare MA90Z Series Mattresses 4 4 2 Installing the Side Rails A WARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s Operator s Manual prior to using this equipment Invacare product manuals are available at www invacare com or your dealer Refer to the instructions provided with the side rails for the installation procedure 4 3 Installing the Power Unit Bed Hook Pull out the bed hook on the back of the control unit 2 Place the control unit on the footboard Y 3 If the bed does not have a footboard place the control unit on a flat surface leaving room for the hose to hang down
21. ore placing a patient on the bed Operate all bed frame motorized functions through their full range of motion to be certain that there is no pulling interference or pinching DO NOT strap the mattress to the bed frame at the head and foot ends Secure ALL mattress straps Secure the straps to the bed deck at the head and foot ends and to the frame at the center of the bed Otherwise damage to the mattress will occur when the head and foot ends are raised Refer to Installing the Mattress Replacement System on page 14 Part No 1148139 Invacare MA90Z Series Mattresses 3 SAFETY Entrapment May Occur A WARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s Operator s Manual prior to using this equipment Invacare product manuals are available at www invacare com or your dealer Proper patient assessment and monitoring and proper maintenance and use of equipment is required to reduce the risk of entrapment Variations in bed rail dimensions and mattress thickness size or density could increase the risk of entrapment Visit the FDA website at http www fda gov to learn about the risks of entrapment Review A Guide to Bed Safety published by the Hospital Bed Safety Workgroup located at www invacare com Use the link located under eac
22. patient s body Ensure that there is 3 to 4 finger width clearance between the bottom of the patient and the bed frame Adjust the comfort setting if needed mon gt Wait five minutes for the mattress pressure to stabilize Repeat STEPS A E until patient has not bottomed out and patient comfort is achieved If the patient feels that the bed is too soft hard press the Soft or Firm button to adjust the comfort settings Use a regular pillow to help support and stabilize the patient s head Part No 1148139 27 Invacare MA90Z Series Mattresses 4 4 9 Transferring Patient From To Gurney A WARNING ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney before transferring the patient between the bed and the gurney Engage the wheel locks of the bed Refer to the owner s manual provided with the bed Engage the wheel locks of the gurney Refer to the owner s manual provided with the gurney Press the Max Inflate button to achieve maximum mattress pressure Raise or lower the bed to match the gurney height Refer to the owner s manual provided with the bed che ONS When the mattress has reached maximum firmness perform one of the following Bed to Gurney Transfer Slide the patient onto the gurney Gurney to Bed Transfer Slide the patient onto the bed 4 10 Transferring Patient From To a Wheelchair A WARNING ALWAYS engage the wheel locks of the bed and the wh
23. r Cord A WARNING DO NOT alter plug to fit a non conforming outlet Instead have an electrician install a properly grounded outlet Failure to use the correct plug and outlet can result in a potential safety hazard CAUTION Ensure that the power cord of the control unit is not pinched or has any objects placed on it and also ensure it is not located where it can be stepped on or tripped over Examine the hospital grade power cord supplied with the control unit 2 Perform one of the following Ifthe plug is damaged Call your supplier for a replacement hospital grade cord Ifthe plug is not damaged Plug the end of the supplied hospital grade power cord into the power outlet on the side of the control unit 3 Plug the other end of the plug into a properly grounded outlet on the wall Once the unit is plugged in an AMBER LED on the control unit is lit indicating that the system is in STAND BY mode Invacare MA90Z Series Mattresses 18 Part 1148139 4 OPERATION 4 6 Using the Front Panel DETAIL A MA90Z POWER UNIT FRONT PANEL Turn Angle Button Power Fail LED Low Pressure LED microA 90 POWER Low Pressure Power Fail Both 20 Minutes Power Button e Full Turn TURN ANGLE Turn Button Lock Alarm ode Max Inflate Button Firm Button Low Air Loss Button Turn Time Button Soft Button DETAIL Select Button Hold Button Power Fail LED MA95Z POWE
24. ree components Part No 1148139 Mechanical Parameters 2 OVERVIEW CONTROL UNIT DIMENSIONS L X W X H I5 x 65 x II CONTROL UNIT WEIGHT 15 Ibs CONTROL UNIT POWER CORD 10 14 Feet Long Hospital Grade CONTROL UNIT CONNECTION Triple Flow Two gt and Quad Flow Four 4 One 4 Single Magnetic Single Quick Coupling Quick Coupling CONTROL UNIT PACKAGING piece box AIR FILTER Environmental Parameters Charcoal Air Filter with None Fire Retardant MA90Z MA95Z OPERATING CONDITIONS AMBIENT TEMPERATURE RELATIVE HUMIDITY ATMOSPHERIC PRESSURE 50 95 F 30 75 Non Condensing 70 106 kPa STORAGE AND SHIPPING CONDITIONS AMBIENT TEMPERATURE RELATIVE HUMIDITY ATMOSPHERIC PRESSURE Part 1148139 40 158 F 10 100 50 106 kPa Invacare MA90Z Series Mattresses 3 SAFETY 3 Safety 3 1 General Guidelines A WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner s manuals service manuals or instruction sheets supplied with this product or optional equipment If you are unable to understand the warnings cautions or instructions contact a healthcare professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur A ACCES
25. se d 3l Changing the EID RONDE E DD DD Sl 33 Een i M M 35 Part No 1148139 3 Invacare MA90Z Series Mattresses GENERAL General Warnings Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage See the information below for definitions of the signal words A DANGER Danger indicates an imminently hazardous situation which if not avoided will result in death or serious injury A WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury A CAUTION Caution indicates a potentially hazardous situation which if not avoided may result in property damage or minor injury or both IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided Gives useful tips recommendations and information for efficient trouble free use Invacare MA90Z Series Mattresses 4 Part 1148139 GENERAL 1 2 Limited Warranty PLEASE NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare or a dealer This warranty is not extended to any other person or entity and is not transferable or assignable to any
26. ss and hold the Soft and Firm Buttons at the same time 3 LCD screen displays the Run Time as described above 47 Powering Up the System For this procedure refer to FIGURE 4 on page 19 Turn on the power to the system by pressing the Power button on the control unit Once the button is released a green led illuminates when the unit is on Invacare MA90Z Series Mattresses 26 Part 1148139 4 8 x t 9 10 4 Placing the Patient on the Mattress Press the Mode button or Max Inflate button to turn on the control unit to maximum flow In this mode the Max Inflate Led lights up After the mattress is fully inflated place the patient on the bed Ensure that the patient s feet are towards the foot end of the mattress the end with the hose Center the patient on the bed from side to side and head to foot Special positioning may be required with contracted patients to provide comfortable positions After placing the patient make certain no objects will fall under the patient such as feeding tubes IV s etc Wait five minutes for the mattress pressure to stabilize Set the comfort pressure to the desired comfort level Verify that the patient has not bottomed out by performing the following steps Ensure that the patient is lying flat on his her back in the middle of the mattress Place four fingers between the air cushions directly underneath the sacral region of the
27. subsequent purchaser or owner Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state Invacare warrants the mattress and cover when purchased new and unused to be free from defects in materials and workmanship for a period of one year from the date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty Invacare warrants the control unit when purchased new and unused to be free from defects in materials and workmanship for a period of two years from the date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty If within such warranty period any such product shall be proven to be defective such product shall be repaired or replaced at Invacare s option This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product Invacare s sole obligation and your exclusive remedy under this warranty shall be limited to such repair and or replacement For warranty service please contact the dealer from whom you purchased your Invacare product In the event you do not receive satisfactory warranty service please write directly to Invacare at the address on the back cover Provide dealer s name address
28. the power unit disconnect the power cord from the wall outlet DO NOT insert items into any openings of the control unit Doing so may cause fire or electric shock by shorting the internal components The control unit must be kept away from all heat sources and radiators during operation Connect the equipment to properly grounded three prong wall outlet using 10 14 ft hospital grade power cord provided with the product Grounding reliability depends upon a properly grounded receptacle 3 prong Part No 1148139 13 Invacare MA90Z Series Mattresses 4 OPERATION 4 Operation A CAUTION The control unit and mattress the MA90Z series are designed to be used as a system DO NOT replace mattresses or control units with other models or other brands Otherwise damage to the system may occur Contact your supplier to get the correct replacement if needed 41 Installing the Mattress Replacement System CAUTION DO NOT strap the mattress to the bed frame at the head and foot ends Secure ALL mattress straps Secure the straps to the bed deck at the head and foot ends and to the frame at the center of the bed Otherwise damage to the mattress will occur when the head and foot ends are raised Head End Bed Deck Foot End Bed Deck Invacare MA90Z Series Mattresses 14 1148139 4 The powered mattress comes with ten nylon buckle straps Remove the original foam mattress fr
29. ver MA90ZB42 Tab and MA90ZB48 Hose Connector Control Unit Control Unit Connectors DETAIL D MA95Z Mattress MA95ZB42 and MA95ZB48 Hose Connector DETAIL RED CPR Connector MA90Z MA90ZB42 and MA90ZB48 Only FIGURE 5 Preparing for CPR Procedure Invacare MA90Z Series Mattresses 30 Part No 1148139 5 MAINTENANCE AND TROUBLESHOOTING 5 Maintenance and Troubleshooting 5 1 Cleaning the System A WARNING Before cleaning or disassembling the 90 series check the underside of the mattress folds for sharp objects such as scissors needles etc These objects should be removed and discarded before proceeding with further cleaning or disassembly otherwise injury or damage to the product may occur Because of potential risk of infectious exposure cleaning with the patient on the bed is not recommended equipment should be inspected Any item that is visibly soiled with the patient s blood or other body fluids should be properly cleaned or removed Laundry workers should treat all soiled bedding as if it were contaminated with pathogenic microorganisms Remove the bedding If necessary inflate the mattress Ensure that the control unit is off Unplug the power cord from the wall outlet Ensure that the underside of the mattress is clear of all sharp objects Examine the surface of the control unit and mattress assembly components for visible blood or body fluids SO iin br E
30. with a quaternary disinfectant solution B Using a nylon brush gently clean all crevices as they can harbor microorganisms C Air dry all treated surfaces Invacare MA90Z Series Mattresses 32 Part No 1148139 5 MAINTENANCE AND TROUBLESHOOTING 10 Perform the following steps to clean the mattress components air cushions mattress base etc A Using a clean sponge or paper towel wipe down the mattress components with a dilute detergent solution of quaternary cleaner disinfectant or other germicidal detergent solution B Wipe the mattress components with a clean dry cloth Storing the System l Ensure that the control unit is off and disconnect the power cord from the wall outlet Clean the system Refer to Cleaning the System on page 3l Disconnect the air hose connector from the control unit and allow air to vent from the mattress Gently roll up the mattress with minimal handling and agitation Ensure the cover surface is inside the roll DY Ur duco m Store the unit keeping the mattress with the control unit 5 2 Changing the Filter For this procedure refer to FIGURE on page 34 Clean the filter every 5 months or whenever it is dirty l Remove the two screws securing the filter cover to the back of the control unit Remove the filter Wash the filter with water and a mild detergent Dry the filter completely Install che filter into the back of the control unit Qr Secure the