Home
SERVICE MANUAL Bel-Cat TAKARA BELMONT CORPORATION.
Contents
1. Bel Cat 68 502 180 Collimator Mask ASSY gi Dm sy Bel Cat Bel Cat PA Bel Cat CM Bel Cat 69 502 180 X ray Generator No Drawing Number Drawing Name Qty Note 0501 108 01262 01 Inside Case ASSY 1 0502 308 04033 Frame 1 0503 408 10265 01 Lead Plate ASSY 1 0504 408 11644 Attaching Plate for Cover 4 1 0505 408 11643 01 Attaching Plate for Cover 3 1 0506 408 11643 02 Attaching Plate for Cover 3 1 0507 408 07586 02 BEAM MOTOR BOARD ASSY 1 0508 408 11641 Attaching Plate for Cover 1 1 0509 408 11642 Attaching Plate for Cover 2 1 0510 108 01263 Tube Head Cover Front 1 0511 108 01264 Tube Head Cover Back 1 0512 408 11645 Fixing Plate for Cover 5 2 0513 962 01016 01 X ray tube D 052SB 1 Collimator Mask ASSY 0520 978 60004 13 Photo micro Sensor EE SX674 3 0521 408 11635 Filter A 1 0522 408 11630 Filter B 1 0553 408 11918 Collimator Mask B CCD P 1 0554 408 11632 Collimator Mask CCD C 1 0555 408 11633 Collimator
2. 5 7 111 ala Hom 2 O SIE 2 Car Drawing Number 308 04014 Drawing Name Fixing Metal Plate for parts 408 10817 00 Cable clamp 408 11583 Fixing Plate for Fuse 965 85020 Circuit Protector IN 1 A8E 19 1B 975 00032 01 Noise Filter NAH 20 000 408 11588 Fixing Metal Plate for LAN Adaptor 408 11586 Shield Plate 988 50047 Relay Adaptor for STP NWSI C5E SH 408 11589 Name Plate for Board 408 11590 Cover 308 04016 01 Cover 408 11585 Hook Plate 937 38040 04 Steel Spacer CF 302 5E 58 502 180 3 Rotating Unit ASSY M 84 Gang 59 502 180 Rotating Unit ASSY Drawing Number 108 01260 Drawing Name Main Body of Rotating Unit ASSY 408 04713 Hook 931 70029 01 LM Guide 2RSR15VMuu 270LM 408 11605 Y Axis Table 978 60004 13 Photo micro Sensor EE SX674
3. mI cop Sensor DC nat mmm EET Pra CNYU BOARD Flat 30 31 cach one cable eee Cephalometric Arm 1 T cop sensor Origin Sensor for Cephalometric Drive Bel Cat POWER BOARD 2244 5 TE E Y axis Motor X axis Origin Senso MOTOR CPU BOARD Rotation Motor CNB BOARD r3 R axis R axis 360 Origin CSK543 Sensor 5 502 180 able Switching Power Supply FPD Sensor INV BOARD PFC FILTER Rotation Unit 21120 ip e HV BOARD Rotation Axis a lus aysa aus an po en Terminal Table oni X RAY TUBE 5 Terminal Table DI was NT FE FT pre Suitching Power Supply Ecit 60 E
4. 0742 308 04057 02 Ear Rod TMJ2 ASSY Left 0743 408 11704 Rest ASSY for Bite 0744 408 11705 Rest ASSY for MS Maxillary 4 Sinus 0745 308 04067 02 Chinrest ASSY for CT 1 0746 408 11401 Chinrest 1 0747 408 11402 Chinrest for TMJ 1 0745 308 04068 01 Ear Rod CT TMJ ASSY Right 1 308 04068 02 Ear Rod CT TMJ ASSY Left 1 0749 408 11819 Rod Adaptor 1 Bel Cat does not have 0742 and 0744 Bel Cat 78 502 180 7 4 Head Support Bel Cat 79 502 180 Head Support Drawing Number Drawing Name 0760 937 03042 26 KVDF 25 M5 BK 1 0761 408 11697 Screw Bar 1 0762 308 04054 Rotation Body 1 0763 408 11689 Slider 1 0764 036 10213 O Ring 1A P12 2 0765 408 11690 Slid Shaft 1 0766 936 12002 03 O Ring 4C P5 1 0767 Nylon Nut 1 0768 408 11691 Block 1 0769 308 02467 Forehead Rest 1 0770 036 10202 O Ring 1A P4 1 0771 408 11693 Supporting Body 1 0772 408 11695 Shaft D 1 0773 408 11696 Helical Knob 1 0774 036 10216 O Ring 1A P15 1 0775 408 09149 01 Belt 1 0776 408 11657 Occipital Pad 1 0777 408 11794 Tape Table 2 Bel Cat 80 502 180 8 Cephalometric Device ASSY 0801
5. Bel Cat 81 502 180 Cephalometric Device ASSY Drawing Number Drawing Name 0801 408 11534 Cephalometric Fixing Plate 2 2 0802 408 11533 Cephalometric Fixing Plate 1 1 0803 208 02167 Cephalometric ARM 1 0804 408 11459 Membrane Sheet CM 1 0805 408 10162 CM SW BOARD ASSY 1 0806 208 02168 Cephalometric Driving ASSY 1 0807 308 03996 01 Secondary Slit B ASSY 1 0808 408 11547 Ear Rod ARM 2 0809 408 05028 Ear Rod 2 0810 308 03627 01 Forehead Rest 2 ASSY 1 0811 308 03937 Sensor Hook ASSY 1 0812 408 10450 CNP BOARD ASSY 1 Bel Cat 82 502 180 8 1 806 Cephalometric Driving ASSY 0818 0819 0815 0816 Bel Cat 83 502 180 806 Cephalometric Driving ASSY Drawing Number Drawing Name 5 phase Stepping Motor 0813 977 79029 03 ae ee 1 0814 932 21025 01 Timing Belt 152 2GT 6 1 0815 932 21025 02 Timing Belt 800 2GT 6 1 0816 932 21023 01 Timing Belt 1482 2GT 6 1 0817 408 10631 00 Driver ASSY 1 0818 408 11384 CNM BOARD ASSY 1 Relay Adaptor for STP 0819 988 50047 1 0820 978 60004 13 Photo micro Sensor EE SX674 1 0821 108 01255 Cephalometric Cover 1 Bel Cat 84 5
6. Drawing Number Drawing Name 0720 408 11681 Holder 1 0721 962 05006 Laser Marking Projector LDV167LS 1 0722 931 70026 01 LM Guide RSR9KMuu 75LM 1 0723 978 60004 13 Photo micro Sensor 5 674 1 0724 932 21002 80 Timing 6052 292 1 0725 977 79007 07 2 phase stepping Motor PK244 04A 1 Bel Cat 75 502 180 7 2 0713 Operation Panel ASSY Drawing Number Drawing Name 0730 408 11379 PANEL R BOARD ASSY 1 0731 408 11378 PANEL L BOARD ASSY 1 0732 408 11655 Membrane Sheet R 1 0733 408 11656 Membrane Sheet L 1 0734 408 07144 Spacer 1 0735 962 05008 Laser Marking Projector LDV167LA 1 Bel Cat 76 502 180 7 3 Head Holding Rod Ear Rod and each Chinrest 0745 E 0743 77 502 180 HEAD HOLDIONG ROD EAR ROD each CHINREST 0740 Drawing Number 308 04056 01 308 04056 02 Drawing Name Head Holding Rod ASSY Right Head Holding Rod ASSY Left 0741 308 03889 01 308 03889 02 Ear Rod ASSY Right Ear Rod ASSY Left 308 04057 01 Ear Rod TMJ2 ASSY Right ll
7. Error Number Error Message 223 Now interlock works Interlocking is fuctioning Problem The interlock circuit is open Corresponding Confirm whether the LED D28 on MOTOR CPU BOARD lights LED does not light Check electric conversion between MOTOR CPU BOARD and CNYU BOARD and Power Supply Box Repair the harness of CNINT Replace MOTOR CPU BOARD LED lights Error Number Error Message 224 X Ray switch is error Release X ray Exposure switch and click READY Problem READY ON button is pushed while X ray switch is pushed Corresponding Bel Cat Confirm whether the LED D10 on MOTOR CPU BOARD lights LED does not light Check electric conversion of the harness between MOTOR CPU BOARD and CNYU BOARD and external X RAY SW NG LED lights Repair the harness between MOTOR CPU BOARD and CNYU BOARD and X RAY SW Replace MOTOR CPU BOARD 33 502 180 The message Pull out the CCD sensor comes out and CT exposure cannot be executed Confirm the LED D24 on CONTROL BOARD does not light without CCD sensor in CCD sensor holder LED does not light Replace CONTROL BOARD LED lights Check whether the LED of the sensor for existence of CCD sensor located on CCD sensor holder is lightening or not Confirm both sensor for existence of NG Repair both sensor for CCD sensor and CNS2 harness existence of CCD
8. rn naa Lisk CRIS Pleora bonversion Baars 50 TES DC DC awar CONVERTER fenor 1 BOARD Pleora Em var a CONTROL BOARD koo moving switch 2 for Goll instar BEAM BOARD FAN ND stor Signal from R Motor Signal from R Sensor 100 sensor E st Panoramic side Primary ollimator Motol Head Swing Motor PK266 ENE 1 OND kantor for ST Panoramic position sensor ic Head Origin Sensor Origin Sensor Panoramic lead rigin Sensor SUD fot oo ansar bolder P an tar lotor Drive R axis Motor Drive X axis Lag PEN 23 eno BOARD EE STR sLIDING UNIT BODY exposure CONTROL UNIT 05 543 Sensor __ 5 sea rest unit Y RAY TUBE Q rotation axis Moto PK543 Bel Cat 101 502 180 12 Revision data of this manual This manual was created in March 2010 Revised on August 2011 Document number S02 180E Bel Cat 102
9. 421 Head Swing ASSY 0428 Bel Cat 66 502 180 402 Head Swing ASSY 421 Drawing Number Drawing Name 0428 208 02156 Base 1 0429 408 11485 Motor Bracket 1 0430 977 79031 05 1 With Cephalometric 0431 408 10414 redis Processing Drawing 2 4 with Cephalometric or Timing Pulley 0432 932 21005 18 Timing Belt B 100 S3M 252 1 0433 408 10415 Additional Processing Drawing 1 1 For Timing Pulley 0434 408 11499 Angle Plate 1 0435 408 11500 Sensor Bracket 2 0436 978 60004 13 Photo micro Sensor EE SX674 2 0437 408 11486 Sensor Fin 1 0438 408 10902 Plunger Holder 1 0439 937 50019 26 Plunger PBF 10KN 1 With Cephalometric 0440 Hexagon Socket Screw 4 truncated cone point M10x25 0441 408 11502 Stopper Table 2 0442 937 50031 21 Small Plunger PCF 6 BS 2 0443 920 01010 05 Club Screw SGP M6 PL L12 2 0444 408 10420 Fixing Plate 1 with Cephalometric 0445 408 07586 03 BEAM MOTOR BOARD ASSY 1 with Cephalometric 0446 031 60006 Rock Nut ANO6 1 0447 031 60306 Rock Washer AW06 1 0448 930 00012 07 Bearing 690677 2 0449 078 60099 Connector EE 1001 2 Bel Cat 67 502 180 5 X ray Generator 0512 Collimator Mask ASSY
10. X ray Exposure Switch 44 502 180 1 Sliding Unit ASSY 45 Bel Cat 502 180 Sliding Unit ASSY Drawing Number Drawing Name 108 01257 Sliding Unit ASSY 308 04040 01 Stand Pipe 408 11647 03 Rail 408 11647 04 Rail 408 10304 Shading Plate 308 04008 01 Counter Weight ASSY 308 03988 Driving Pulley ASSY 408 11648 Cable Guide 408 11649 Frange 408 11650 Stand Cover 308 04010 Pulley Cover 308 04011 01 Wall Mounting Bracket 208 02174 Top Cover 108 01265 Base C 308 02548 Joint Block 937 30005 Foot TTM 00159 208 01776 03 Stand Lower Cover 408 09633 Flange 46 5 AASA 502 180 1 1 0101 Sliding Unit ASSY Bel Cat 47 502 180 0101 Sliding Unit ASSY Drawing Number 108 01258 Drawing Name Sliding Unit ASSY 308 04006 Upper Roller ASSY 308 04007 Lower Roller ASSY 308 04005 Frankfort Plane Beam ASSY 408 11572 Switch ASSY 408 11558 Up amp Down Movement for Sensor ASSY 308 04029 Fixing Plate for Rel
11. 12 6 0Vdc 0 30V Table 2 mAref Voltage between CP 15 and CP5 If the value is different from the above value replace CONTROL BOARD Error Number Error Message 205 Tube temperature is error Turn the power of X ray equipment off and wait for 30 minutes Problem X ray exposure is stopped for protection of tube head or wire for detection is cut Corresponding Confirm whether the LED D31 on CONTROL BOARD lights or not NG Repair the harness Replace CONTROL BOARD OK Tube head generates heat Please wait for a while OK Bel Cat 22 502 180 Error Number Error Message 206 Communication between Motor CPU and Control Board is error Turn on the power of the X ray equipment again Problem Communication error Corresponding Check electric conversion of CN3 3 pin 4 pin and 5pin on the harness between MOTOR CPU BOARD and CONTROL BOARD NG OK Check the voltage of CONTROL BOARD Measurement point Voltage CP4 CP5 5 0 40 1 CP6 CP5 15 0 30 5 7 5 15 0Vdc 30 75V CP8 CP5 5 0 40 1 Replace CONTROL BOARD OK Measure the voltage between CPU BOARD check pin CP9 and CONTROL BOARD check pin CP5 and confirm whether sometimes the value is under 5Vdc Sometimes under 5V Replace CPU BOARD 5V stable Replace CONTROL BOARD Error Number Error Message 207 Communication between Motor CPU and Cons
12. 9924519 40 105498 Josuog sod 811 105 Josuag 10111804 INI 4030N 12104 peratu sixe X 04 08 ANO 10 SIXB X 10 4010 10 11 204 104 401 0 SIXe 401 10104 096 y 096 y 018140 SINE 04 04 YOLON 10194 Buing 0 14 paemiog 099 9 1 10 1 edo 10104 105055 118120 antag 21 eydeg pis 91 1qauo eydag uo Josuas 009 T070 1234071 10H 058496 3111 weeg 123021 40H 09008 013110 21102 Joy sanag 2030 Hz tun Suip is 51 q uo udao u Odd bi q uo eudeg opis lo iue joue uo oluesoueg josues 00 inje3 Jo 1204 ean 104840 99 1 O 10 00172193 0 08 10410 yor As unog dn 3180 Supi s 01 4osuag 4035 105095 015 49080 0809 weag 21107 Jo Z w
13. GGDSensorID Glear IP Address Setting Method 1 Drag the file from the windows explorer and drop it to the calibration software Name Type 960798 File Folder cephalo lan File Folder File Folder DAT File DAT File CalibP dat DAT File SettingCalib exe Application SettingCalib exe nihon File 3 SettingCalib ini Configuration Sett _ ST6UNST 000 000 File L ST6UNST LOG Text Document Method 2 Click the Sensor Folder button and Find the Damage Correction file Open ay SettingCalib Auge UY ingCalib Auge Ej Recent Places 960798 dat Computer More Folders ii 21 Program Files 21 ProgramData RadiographyAgent 21 ScanXInterFace d script SettingCalib Auge cephalo lan J film DAT File DAT File DAT File 6 10 2009 12 14 PM 1 25 2010 9 24 AM 1 25 2010 9 33 AM Auna File name m J gt Damage Correction File dat w At this moment the Damage Correction file will be copied automatically to the C_ Sensor ID folder in the SettingCalib Auge folder Bel Cat 89 502 180 4 f the damage correction file is copied correctly the buttons as shown below will be available SettingCa b AUG Drag and Drop correction file to this window GGDSensorID 960798 Clear Panoramic Calibration Button Panoramic Calibration
14. See Section 4 Panoramic Calibration Panoramic Radiograph OD Cephalometric Calibration Button Cephalometric Calibration See Section 5 Cephalometric Calibration Cephalometric Radiograph IP Address Setting CCD Sensor Folder Bel Cat 90 502 180 4 Panoramic Calibration E Calibration PICT_CCDCNTL 1 2 9 4 205 16 14 12 CUT Parameter With CUT Up 02 Down 0 Exposure Conditions T Tube Voltage Current Setting 642 125 Create Calibration File Result Target Judgement Average Limit Exceeded Calibration Close When Panoramic Calibration button clicked the window above will be displayed Please follow the instructions below to create calibration data 1 Open CXProgram Files AsahiRoentgen AUGE_LINK AUGE_CCD ini file and please remember the following values File C Program Files AsahiRoentgen AUGE_LINK AUGE_CCD ini Values to be IMAGE CUT noted Cut Cut Bottom Cut_Left 0 Cut Ri ght 0 2 Put a check in the With CUT checkbox in CUT Parameter option insert the Cut Top value the Up and Cut Bottom value in the Down entry field Bel Cat 91 502 180 3 56 the Arm Unit position and then set the 1 1 CCD Sensor Calibration ig vertically on the Rest Unit Operation Panel Base of the panoramic side as shown below Be sure to co
15. air containing chlorine or sulfur compounds and other such causes 3 Ensure safety against tilting vibration or impact including during transport 4 Do not install in a location where chemical agents are stored or gases are produced 5 Pay attention to the supplied power frequency and voltage and allowable current or power consumption 6 Check the power status of batteries e g discharge state polarity 7 Connect grounds correctly 3 Observe the following before using this equipment 1 Inspect switch connections polarity dial settings meters and other elements to ensure proper operation of the equipment 2 Check that grounds are fully connected 3 Check that all cords are connected properly and safely 4 Take special care with joint use of other equipment which can lead to improper diagnosis and cause hazards 5 Re inspect external circuits connected directly to the patient 6 Check battery power 4 Observe the following during use of this equipment 1 Use care not to exceed the duration or amount of exposure needed for diagnosis and treatment 2 Monitor all equipment and the patient continually for problem free status 3 If any problems in the equipment or patient are discovered discontinue operation of the equipment with the patient in a safe status or take other appropriate measures 4 Make sure that the patient is not touching the equipment 5 Observe the following after use of this equipment 1
16. button 6 Put a check in the With CUT checkbox in CUT Parameter option insert the Cut Top value the Up and Cut Bottom value in the Down section 7 Repeat the similar procedure as shown in the section 4 However set the exposure condition to 95kv for the Voltage and 12mA for the Current If the calibration succeeds similar to the panoramic this time calibC dat file will be created in the Sensor 10 folder 8 R e attach the Guard Rod which was removed in step 5 1 6 Radiation Field Width Distribution Calibration If the X ray radiation field s right and left distribution is not adjusted properly equally calibrate the distribution by adjusting following item By inserting number in pixel unit to DEFAULTOFFSET the acquired image on screen will be shifted to the right Negative value cannot be applied File C YSettingCalib AugeYcephalo lanYcephalo prm ini CORRINGSIZE Item to be is adjusted pEFAULTOFFSETAZ Bel Cat 95 502 180 7 Final Verification 1 Copy all of the calibP dat calibC dat and Damage Correction file which is created in the C Sensor ID folder to the C Program FilesYAsahiRoentgenYAUGE LINK folder 2 Open the AUGE CCD ini in CXProgram FilesYAsahiRoentgenYAUGE LINK folder and then complete following items CORRINJ LINE 1 7 9 27 1 Data File Name for damage correction CORRLINEDATA C Program Files AsahiRoent
17. 180 02 Warnings Warning Read this Service Manual at time of repair and follow the instructions If the repair will fail it will cause the damage of the equipment and cause of the trouble Pay special attention to all Dangers Warnings and Cautions as described Note Cautionary terms in this Service Manual are defined as follows Danger Used to describe a direct danger which if not avoided might foreseeable lead to occurrence of death or serious injury total loss of equipment fire or other severe accidents or damage to property Warning Used to describe an indirect danger which if not avoided might foreseeably lead to occurrence of death or serious injury total loss of equipment fire or other severe accidents or damage to property AN Caution Used to describe a potential danger which if not avoided might cause mild or moderate injury partial damage to equipment or loss of electronic data Note Used to indicate helpful information for use of this equipment Bel Cat 5 502 180 Warning Cautions for use of medical electrical equipment hazard prevention and safety 1 This equipment is for use only by trained personnel 2 Observe the following cautions when installing this equipment 1 Install in a dry location 2 Install in a location free from potential negative effects due to atmospheric pressure temperature humidity wind sunlight dust
18. Inverter board lights Confirm the output of equipment specification voltage 120 10 between CN1 1 and 2 on Inverter board NG Confirm whether LED 060 and D 61 on CONTROL BOARD lights when READY D60 NG Replace CPU BOARD CONTROL BOARD D61 NG Check electric conversion of harness CN3 between CONTROL BOARD and MOTOR CPU BOARD Repair the harness CN3 between CONTROL BOARD and MOTOR CPU BOARD OK Replace MOTOR CPU BOARD OK Replace K C relay K Replace Inverter BOARD When LED DL2 on Inverter board lights Confirm the voltage between check pin CP11 and CP5 kVref on CONTROL BOARD Setting Tube Measurement Voltage Voltage CP11 CP5 60kV 12mA 3 19 0 05 70kV 12mA 3 63Vdc 0 10 80kV 12mA 4 09Vdc 0 12 90kV 12mA 4 60 0 13 95kV 12mA 4 70Vdc 0 13 Table 1 kVref Voltage betweenCP11 and CP5 If the value is different from the above value replace CONTROL BOARD Bel Cat 21 502 180 When LED DL3 on Inverter board lights Confirm the voltage between check pin CP15 and CP5 mAref on CONTROL BOARD Setting Tube Measurement Voltage current Point 2mA CP15 CP5 1 1 0 0 05 3mA 1 5Vdc 0 08V 4mA 2 0Vdc 0 10V 5mA 2 5Vdc 0 13V 3 0Vdc 0 15V 7mA 3 5Vdc 0 18V 8mA 4 0Vdc 0 20V 9mA 4 5 0 23 10mA 5 0 0 25 11mA 5 5Vdc 0 28V
19. The abnormal sound comes out while movement of R axis motor Confirm whether the red LED on sensor does not light and the LED D58 on CONTROL BOARD lights when shading the R axis initial position sensor by scale compulsory The red LED on sensor does not stop lightning Replace R axis initial position sensor Check electric conduction of harness between R axis initial position sensor and CNR LED 058 does not light NG Repair the harness between R axis initial position sensor and CNR OK Both are OK CONTROL BOARD Error Number Error Message 204 Inverter is error Turn the power of X ray equipment off and then contact your agent Problem Inverter error Corresponding 1 Error comes out when READY is ON Confirm the LED number which lights on Inverter board on Arm unit when the power of equipment is ON DL2 only NG Check electric conduction Repair the harness between CN4 4 and Head EP between CN4 4 and Head EP OK Replace Head DL3 only Check electric conduction between CN4 1 and Head IP NG Repair the harness between CN4 1 and Head IP OK Replace Head DL2 3 green LED lights Repair the harness etween CN4 and Head OK Replace Head DL1 only Check again after 30 minutes passed from turning power OFF Check electric conduction between NG CN4 and each pin of Head Bel Cat 20 502 180 2 Error comes out when X ray is ON When LED on
20. completing the repair work sterilize equipment portions touched by service personnel WARNING While repairing do not place anything that can be an obstacle within the range of equipment WARNING To avoid damages to equipment measuring instruments etc and electric shock service personnel must not remove covers of equipment except when necessary WARNING To avoid damages to equipment measuring instruments etc and electric shock turn off power of equipment and take extra care not to short circuit with other circuit when connecting lead wire of measuring instrument to a circuit within the equipment during repair work A WARNING To avoid damages to equipment measuring instruments etc and electric shock be sure to turn off power of equipment and use the parts specified by us when replacing machine parts and electric parts of the equipment WARNING To avoid X ray exposure due to carelessness be sure to install lead plate for X ray protection on X radiation aperture of X ray generator when radiating X rays during repair work A WARNING Be sure to use positioning attachment specified for each exposure mode when positioning patient A WARNING Be sure to make patient and assistants to wear X ray protective clothing Assistants in this sentence means a person allowed by doctor Bel Cat 8 502 180 WARNING Be sure to operate X r
21. not stop lightning Replace panorama initial position sensor LED D3does not light Check electric conversion of the harness between CNS1 and CNS3 on each initial position sensor NG Repair or replace the harness between CNS1 and CNS3 on each initial position sensor OK Replace panorama initial position sensor Replace CONTROL BOARD Both are OK Bel Cat 25 502 180 Error Number Error Message 215 Movement of cephalometric motor is error Click RESET button Problem Cephalometric can not go back to the initial position Cephalometric motor does not move at all or the abnormal sound comes out while movement Corresponding 1 The abnormal sound comes out while movement Check alignment of secondary slit and CCD sensor holder NG Adjust secondary slit and CCD sensor holder OK Confirm whether the red LED on sensor does not light and the LED D11 on MOTOR CPU BOARD lights when shading the initial position sensor by scale compulsory The red LED D3does not stop lightening Replace initial position sensor LED on sensor is OK LED 011 does not light Check electric conversion between CNM BOARD and MOTOR CPU BOARD on initial position sensor NG Repair the harness between initial position sensor and CNM BOARD and MOTOR CPU BOARD OK Replace initial position sensor Replace MOTOR CPU BOARD Both are OK Bel Cat 26 502 180 2 Motor does move at all Check the harness betw
22. the tip of the earth Bel Cat 98 502 180 Check Item Working accuracy of the equipment Indication during X ray exposure X ray Generator Check Item Exposure functions Mechanism to up and down movement X ray output Bel Cat Purpose of Check Movement check of the electric power circuit Movement check of the exposure circuit Movement check of the operation circuit Accuracy of the Positioning mechanism Checking of The protector circuit Checking of the movement Indicator ASSY Checking of synchronization with X ray exposure X ray Tube Head Lower voltage cable Exposure cone Filter Slit Collimator Purpose of Check Rotation and movement of X ray Tube Head Movement of ARM Mechanism of patient positioning Beams for patient positioning Wire rope Locking Condition of up and Down movement X ray Tube Voltage X ray Tube Current Exposure Time Contents of Check Checking of the voltage value of the electric power source circuit Checking of the movement ware of the control circuit and setting value Checking of the movement of the operational Sequence Checking of deterioration in the mechanism portion Checking of the setting value and movement Condition of the protector circuit Checking of the circuit function to indicate The movement conditions Checking of the movement for Indicator ASSY during usage Checking of any leakage for insulation oil Checkin
23. 02 180 9 Anchor Bolt for fixing For concrete wall For wooden wall Bel Cat For wooden floor 85 502 180 Anchor Bolt for fixing Drawing Number Drawing Name 0901 926 29003 11 High Anchor 8 50 2 Salat One 0902 926 29005 12 Coach Screw 98x40 2 0903 926 29003 13 High Anchor C8 80 2 0904 929 29005 16 Coach Screw 98x65 2 0905 972 40002 17 Hall Plug 2 Bel Cat 86 502 180 08 CCD Sensor Replacement Manual 1 Required items 1 CCD Sensor Calibration J ig 582 00979 2 Sensor Calibration J ig Installation Base 582 00968 3 Sensor Damage Correction file Attached in the Sensor CD 4 AUGE CCD Sensor Calibration Software I nstaller 2 Port Setting Windows Vista and Windows 7 1 Open Device Manager from the Control Panel 2 Expand Ports COM amp LPT right click on the Communications Port select Properties from the right click menu 5 84 ENGINEER PC ik Computer Disk drives Display adapters t DVD CD ROM drives Ethernet Bus Ethernet Bus Services Ports COM 2 Floppy drive controllers 2 Communications Pork 5 5 Human Interface Devices E ECP Printer Port LPT1 Q IEEE 1394 Bus host controllers J Keyboards JA Mice and other pointing devices 58 Monitors Network adapters Portable Devices Ports COM amp LPT Process
24. 1 54 502 180 1 6 0138 Switching HUB ASSY Drawing Number 408 11579 Drawing Name Base Metal Plate 408 10578 HUB ASSY Bel Cat 408 11580 Presser Metal Plate 55 502 180 1 7 0107 Driving Pulley ASSY L Bel Cat 56 502 180 0107 Driving Pulley ASSY Drawing Number Drawing Name 0178 308 04000 Frame 1 1 0179 408 04717 00 Pulley 1 0180 030 03005 00 Bearing 6005ZZ 2 0181 408 04718 Axis 2 0182 308 04001 Frame 2 1 0183 408 04717 03 Pulley 1 0184 308 04002 Gear Large 1 0185 937 38038 09 Steel Spacer ASF 535 3 0186 408 11553 Motor Bracket 1 DC MOTOR Geard Type 0187 977 79039 DCK31 404 864 1 0188 408 11554 Gear Small 1 0189 408 11555 Adjuster 1 0190 408 11408 DC MOTOR BOARD ASSY 1 0191 408 11548 Insurance Sheet 1 0192 308 04003 Fixing Metal Plate for Board 4 Pulley 0193 408 11557 Wire Guard 2 0194 408 05741 04 Rubber Indicator Axis 1 0195 408 06714 Stopper 1 0196 915 30031 Curl Cord 1 Bel Cat 57 502 180 2 Switching Box ASSY
25. 408 11244 Stopper 408 10225 axisX Table 308 04023 Axis Motor ASSY 308 04018 ARM Driving ASSY 208 02171 R Axis Motor ASSY 408 11246 Rotation Sensor B ASSY 408 11612 Sensor Fin 1 408 07569 Presser Plate L2 408 11613 Fixing Plate for Y Axis Sensor 974 80068 01 Switching Power Supply ECM60 lt N Motor Driver CRD5107P N Supplied with No 0341 0331 408 11380 MOTOR CPU BOARD ASSY 408 11610 Metal Stopper for X Axis 2 408 11611 Rubber Stopper 2 408 11617 Sensor Fin D 408 07101 Sensor Bracket C 978 60004 07 Photo micro Sensor EE SX672 408 07109 01 Sensor Bracket 978 60004 10 Photo micro Sensor EE SX67 1 408 11614 Description Plate 408 11917 Description Name Plate 108 01261 Rotation ASSY Cover 937 38035 29 Spacer CF 420E h 931 00045 02 Terminal Table ML 20 ML 20 3P 408 11381 POWER BOARD ASSY 977 79029 03 5 phase stepping Motor CFK544AP2 308 02561 Motor Bracket 932 41025 01 Coupling ALS 014 Y 5 6 930 00021 07 Miniature Ball Bearing 63622 931 50022 00 Ball Screw 930 00021 05 Miniature Ball Bearing 60622 408 11905 Block C 408 11596 Bearing C Motor Driver CR
26. 9 control Assy is improve the communication environment close the exposure panel and turn the initialized under r lv of th ioment ON in Ga ie power supply of the equipment ON agai initialization 510 There is a possibility that Film Sending Speed Data is incorrect Replace with the correct file Excess of Exposure ihili 511 Range of Input Film There is a possibility that Film Sending Speed Data File is incorrect Speed Data Replace with the correct file CCD Error Message Error Number Error Message Corresponding Excess of Total of Exposure Range by Input I NE p aU 512 Film Speed Data and There isa possibility that Film Sending Speed Data File is incorrect Exposure Range by Time Replace with the correct file Lag Incorrect Proofreading Data ihili i la 513 Elle SE top Coffee rar There is a possibility that Calibration Data File is broken Incorrect Sensitivity Replace with the correct file 514 Incorrect Data File Size for There is a possibility that Damage Correction Data File is broken Damage Correction Replace with the correct file 1 Check the connection with Sensor 2 Check the power source for the equipment 518 Bawa ereac 13 Check the communication environment 4 Close Operation Panel and turn the power of the equipment OFF and ON Communication Control written AUGE_CCD ini o
27. CP7 CPGG rises to 4Vdc Check electric conversion of harness 2 Repair the harness Does not rise to 4 CN2 OK Rises to 4 Replace DC MOTOR BOARD 31 502 180 Sliding Unit does not move up down key cancel sound comes out Confirm whether the LED on main body does not light and the LED D6 on MOTOR CPU BOARD lights when shading the Stop Sensor Replace Stop Sensor OK Replace MOTOR CPU BOARD Sliding Unit does not move up down key cancel sound does not come out Confirm whether the LED D3 and D4 on DC MOTOR BOARD lights when elevation switch is pushed The LED on sensor does not stop lightning LED D6 does not light OK Check electric conversion of harness between DC MOTOR BOARD and CNYU BOARD Repair the harness between DC MOTOR BOARD and CNYU BOARD LED does not light OK LED lights Replace DC motor Confirm the voltage more than 24Vdc between CN3 1 pin and 2 pin on DC MOTOR BOARD in Pulley assy about 4 seconds after elevation button is pushed Check electric conversion of harness CN3 on DC NG More than 20 Replace DC motor OK Confirm whether the voltage between MOTOR CPU BOARD CP7 CPGG rises to 4Vdc Check electric conversion of harness CN2 Repair the harness CN2 OK NG Less than 20Vdc Does not rise to 4Vdc Replace DC MOTOR BOARD Rises to 4Vdc Bel Cat 32 502 180
28. D5107P A pair with No 0366 937 38034 06 Bel Cat Duracon Spacer C 607 60 502 180 3 1 0308 axis Motor ASSY 0347 I 1 LL l d 20 all CNB BOARD Oe Detail Drawing of Wring between CNB and CNC BOARD Drawing Number Drawing Name 5 phase stepping Motor 0341 977 79029 03 1 0342 308 03303 Motor Bracket B 1 0343 932 41025 01 Coupling ALS 014 Y 5 6 1 0344 930 00021 07 Miniature Ball Bearing 636ZZ 2 0345 408 07102 Bearing B 1 0346 931 50022 00 Ball Screw 1 0347 408 11903 00 Block A 2 0348 408 09261 Cable guide 1 0349 937 38019 03 Duracon Spacer C 407 2 0350 408 07263 CNC Board ASSY 1 0351 408 09414 Cable ASSY 1 0352 937 38034 06 Duracon Spacer C 607 1 Bel Cat 61 502 180 3 2 0309 Arm Driving Axis ASSY Drawing Number Drawing Name 0353 408 11592 Shield Plate E 1 0354 408 09259 Cable Guide 1 0355 308 02560 Gear
29. Follow the prescribed order to return operating switches dials and other components to their pre use status and then switch off power 2 Do not use excessive force such as yanking the cord when disconnecting cords 3 Observe the following with regard to storage locations 1 Store in a dry location Il Store in a location free from potential negative effects due to atmospheric pressure temperature humidity wind sunlight air containing chlorine or sulfur compounds and other such causes Ensure stability against tilting vibration or impact including during transport IV Do not store in a location where chemical agents are stored or gases are produced 4 Clean accessories cords conductors and other such items before putting them away b Always leave equipment in a clean state to avoid malfunction during the next use 6 If malfunction occurs post appropriate signage to prevent use and contact trained personnel for repair 7 Do not modify the equipment 8 Periodic Maintenance 1 Perform periodic maintenance of all equipment and components 2 Before resuming use of any equipment not used recently be certain to check that the equipment operates normally and safely Bel Cat 6 502 180 DANGER Liquids on this equipment will cause electric shock accident or equipment damage This equipment is electric equipment Keep liquids away from this equipment DANGER This equipment is electric equ
30. Large 1 0356 308 03302 Boss B 1 0357 408 04746 Key 1 0358 308 03553 Boss C 1 0359 408 07099 Bearing Presser A 1 0360 923 92004 08 Fine U Nut FU07SC 1 Bearing for Combine speculate 0361 931 20001 08 nic POS P 2 0362 308 04019 ARM Suspension Axis B 1 Bel Cat 62 502 180 3 3 0310 R axis Motor ASSY Drawing Number Drawing Name 0364 408 06554 CNB Board ASSY 1 0365 408 11239 Motor Base 1 1 0366 977 79037 05 Harmonics Geared Motor CRK543AP H50 1 0367 408 11326 Gear Small 1 0368 408 11242 Spacer and Guide 2 0369 408 11243 Slide Stopper 1 0370 934 00004 12 Coil Spring 1 0371 037 38707 Support 4 Bel Cat 63 502 180 4 Arm Unit ASSY 0425 0423 e e o pe eJ 22 m o Bel Cat 64 502 180 Arm Unit ASSY Drawing Number Drawing Name 0401 108 01250 1 0402 308 03973 01 Head Swi
31. Mask FPD 3D 1 0526 408 11919 Collimator Mask B FPD P 1 2 2 2 phase Stepping Motor 0527 977 79010 05 PX243 03A 1 Bel Cat 70 502 180 6 Image Receptor ASSY FPD Image Receptor ASSY Bel Cat 71 502 180 Image Receptor ASSY FPD Image Receptor ASSY Drawing Number Drawing Name 0601 208 02162 Base 1 0602 208 02163 Frame of Image Receptor 1 0603 931 70028 02 LM Guide RSR12VMuu 120LM 1 0604 932 21001 05 Timing Belt B40S2M100 1 0605 978 60004 13 Photo micro Sensor EE SX674 2 0606 977 60002 01 Geared Motor TG 0301A SG 1 5 1 0607 408 09220 20 CONVERTER BOARD 4 0608 942 60022 07 Fan KDE2406PHV2 2 Conversion Adaptor for Image BOOS Obtaining PT1000 CL 1 0610 988 50131 FPD VALIAN1313 1 0611 0612 988 50054 01 Digital Sensor CCD Sensor 1 PANO PA Alternative 0613 988 50123 Digital Sensor CCD Sensor p BANO Gian gital Sensor ensor CM 0614 964 50060 02 Push Button Switch MB 2511G 1 0615 964 50031 01 Color Cap AT 442 W 1 0616 208 02165 Cover of Image Receiver Back 1 0617 208 02164 Cover of Image Receiver Front 1 0618 408 11490 00 Hook ASSY 1 0619 408 11507 Bli
32. SERVICE MANUAL Bel Cat To service personnel Read this Service Manual thoroughly before repair working Bring this Service Manual at the time of repair TAKARA BELMONT CORPORATION 502 180 Before Starting Repair Work Please be sure to read this Service Manual before starting repair work and fully understand the contents Some repair works involve risks in adjustment confirmation etc So be very careful about the safety when performing such works The repair works shall be performed by a qualified personnel or a person who completed the training specialized in repair at our company This service manual is the document which describes the proper repair method and items necessary to maintain quality effectiveness and safety of this medical equipment Be sure to follow the procedure and the contents written in the manual Bel Cat 502 180 Contents 01 Introduction NNNM 4 02 Warnings 5 11 Preparation for Repair Work o LM LEA 12 04 Specifications and configuration for the equipment 13 15 AA 16 40 06 Configuration for Printed Circuit Board 41 42 07 Parts List 43 86 08 CCD Sensor Replacement aaa aaa aaa aa aaa aaa 87 96 09 Contact Information 97 10 Service Maintenance ee 98 99 11 Wiring Diagram 100 101 12 Revision Data 102 Bel Cat 3 502 180 01 Introducti
33. agnification ratio Exposure time TMJ 4 section lateral FPD height Magnification ratio Imaging region Voxel size Exposure time mode FOV Imaging region Voxel size Exposure time 60 95 kV 1 kV increment 2 12 mA 1 mA increment 0 5 mm Total filtration 2 5 mmAl min Acquisition modes 138mm H Adult Child Ortho 1 25x Short time 1 9 sec Normal 12 sec Lateral Normal 250 mm W x 220 mm H Short time 180 mm W x 220 mm H Frontal 200 mm W x 220 mm H 1 1x Lateral Short time 2 9 sec Normal 4sec X Frontal 4 sec Option 200 mm W x 220 mm H 51mm x 51mm H L8 R8 L TMJ R TMJ 0 1 mm Short time 1802 8 5 sec Normal 360 2 17 sec 79 mm x 71 mm Dentition L TMJ R TMJ 0 155 mm Short time 1809 8 5 sec Normal 360 2 17 sec Dimensions w 1010 1010 1900 D 1300 1300 1300 2360 2360 2360 X1 Fast is displayed on Operation panel instead of short time 2 Imaging may be 270 depending on patient position Bel Cat 13 502 180 Operational environment Temperature 5 to 35 C Humidity 30 to 85 Atmospheric pressure 700 to 1060 hpa Storage environment Temperature 10 to 60 C Humidity 10 to 95 Atmospheric pressure 700 to 1060 hpa Transportation environment Temperature 10 to 60 C Humidity 10 to 95 Atmospheric pressure 700 to 1060 hpa Bel Cat 14 502 180 Bel Cat series Block Diagram 108498 90 jo
34. annotMakeCalib in Evaluation gt gt gt gt gt Calibration PICT CCDCNTL Reconfigure the CCDCNTL value 2 3 4 Repeat from step 4 4 20 2 16 14 12 again CUT Parameter 6 1f the calibration succeeds the message below will be displayed The calibration file calibP dat will be created in the C Sensor ID folder the Average value exceeds Target value by 50 or more decrease the CCDCNTL1 value and the repeat the calibration from step 4 4 Target 646 Create Calib Data Finished gt gt gt Bel Cat 93 502 180 5 Cephalometric Calibration 1 Remove the Guard Rod 2 M ove the Ear Rods and the Forehead Rest as shown below thus the Ear Rods and the Forehead Read would not block the X rays 3 Attach the sensor to the cephalometric side set up the 1 2 Sensor Calibration J ig Installation Base and 1 1 CCD Sensor Calibration J ig as shown below Sensor Calibration J ig Installation Base Sensor Calibration J ig 4 CXProgram Files AsahiRoentgen AUGE_LINK AUGE_CCD ini file and please remember Bel Cat 94 502 180 the following values File C YProgram FilesYAsahiRoentgenYAUGE LINKYAUGE CCD ini Values to be noted Cut 40 Cut Bottom Cut_Left 0 Cut Right 0 Be sure to check the cephalometric values not the panoramic ones 5 Click the Cephalometric Calibration
35. ay Board 408 11382 CNYU BOARD ASSY 408 11615 Insurance Sheet CNYU 408 11640 Hook Bar 308 04030 L Metal Fitting for Chinrest ASSY 408 11571 Cable Pressor 108 01259 Sliding Unit Cover With Filter A 408 11607 Cable Guide 1 408 11608 Cable Guide 2 408 11609 Cable Guide 3 308 04031 01 Back Cover 308 04031 02 Back Cover 937 38021 07 Spacer BSF 310 408 11577 LAN Adaptor Fixing Plate 4 4 4 1 41 41 41 41 41 41 41 41 41 4 988 50047 Relay Adaptor STP NWSI C5E SH 408 11578 Switching Hab ASSY Bel Cat 975 00028 01 Law Frequency Filter T 72SN13 026K 48 502 180 1 2 0118 Upper Roller ASSY 7171 49 502 180 0118 Upper Roller ASSY 0119 Lower Roller ASSY Drawing Number Drawing Name 408 11575 Roller Bracket 408 05662 Roller 408 05326 Roller Axis 030 03000 00 Bearing 6000ZZ 408 05331 Axis 408 05332 03 Wire Stopper A B Bel Cat 50 502 180 1 3 0120 Frankfort Plane Beam ASSY Bel Cat 99 D N 52 D i 2 HOOD OO Of 51 502 180 0120 Frankfort Plane Beam ASSY Drawing Number 408 11560 Drawing Name Fixing Table 408 11561 Bearing 1 930 00021 03 Bea
36. ay exposure switch from outside of X ray room WARNING Operator must instruct patient not to move while X ray exposure WARNING Watch patient assistants and equipment constantly while X ray exposure and release X ray exposure switch immediately if you find something abnormal WARNING When disposing this equipment contact the dealer you purchased WARNING After completing repair work be sure to turn OFF power for safety CAUTION When X ray exposure to the patient after completing repair work take extra care for patient safety when positioning the patient A CAUTION Damage inside X ray generator etc cannot be repaired on site Depending on service personnel s judgment the equipment will be returned to factory for repair or replacement A WARNING Responsibility for managing the use and maintenance of medical equipments lies with the user hospital or clinic This equipment must be used by doctor or qualified person only As repair or check inside equipment involves risks contact the dealer you purchased from A WARNING When earthquake warning is issued do not use this equipment After an earthquake be sure to conduct maintenance check of the equipment and confirm no abnormality before use Default of the check and or confirmation can harm patient Bel Cat 9 502 180 WARNING Be sure to set up X ray examination ro
37. ctric conversion between Indicator beam initial position sensor and CNINDO Repair the harness between Indicator beam initial position sensor and CNINDO Replace Indicator beam initial position sensor Replace MOTOR CPU BOARD NG OK Both are OK Bel Cat 24 502 180 Error Number Error Message 212 Collimator motor motion is error Turn on the power of the X ray equipment again Problem The abnormal sound comes out from collimator when power is ON The error message is displayed when exposure mode is changed Collimator motor does not move at all or the abnormal sound comes out while movement Corresponding 1 Collimator motor does not move at all Confirm the voltage 24Vdc 1 0V between CP1 CP2 on BEAM MOTOR BOARD NG Check electric conversion of harness CNMB CN21H between BEAM MOTOR BOARD and CNYU BOARD NG OK Confirm the voltage 24Vdc 31 0V between and CP2 Replace fuse F1 Confirm the resistance value approximately 362 between CNM 1 2 3 4 on Motor harness Replace motor Check electric conversion of harness CNMB CN21H between BEAM MOTOR BOARD and CONTROL BOARD NG OK NG OK Re Replace BEAM MOTOR BOARD NG K 2 The abnormal sound comes out while movement Confirm whether the red LED on sensor does not light and the LED D21 D26 D28 on CONTROL BOARD lights when shading the panorama initial position sensor by scale compulsory The red LED on sensor does
38. data Location of the defect which was registered at this time and location of the defect that occurred Image data for check of the defect on the image that was taken at last time 1 Select Save As with Drop Down Menu by right click of the mouse on the image and save the image as file format VIVA Defect Map File refer below Gain Calibration File refer below Offset Calibration File refer below 1 Start VIVA software 2 Set the Acquisition mode on left of VIVA software as 2x2 G1 30fps 3 Select Tools Menu Command Processor and Retriever files 4 Retrieve Files Dialogue is displayed Put check mark to Base Defect File Gain Calibration File and Offset Calibration File then push Browse button to decide the path and the file name 5 Set all the modes under these file names as 2x2 G1 30 1 5 6 Push Retrieve button 7 After capturing click OK button Dark Image Data refer below 1 Start VIVA software 2 Set the Acquisition mode on left of VIVA software to 2x2 G1 30fps 3 Select Acquisition and System Settings 4 System Settings dialogue is displayed Remove the check mark from Offset Corrections Gain Corrections And Pixel Defect Map on right of the dialogue and push Apply button and push OK button 5 Push the green round button on right side of Tool Bar in the screen of VIVA Main Window 6 Allocate Buffer
39. e exposure status as READY Click OK button OK The exposure is ready Push IX RAY SW ON Tube voltage Tube current Exposure Time 60kV 2mA 8 5sec Exposure Make the mode mode CT Tube voltage Tube current Exposure Time 60kV 2mA 8 5sec Exposure OK Make the exposure status as Cancel Exposure mode Half orbit until the equipment stops 1 Exposure mode Test mode exposure status as READY Click OK button TOK The exposure is ready Push the icon of on vertical tool bar on right side of the screen while X ray is emitted using X RAY SW ON gt The image of the chinrest is displayed Push cancel on the displayed window TOK CCD Error Message Error Number Error Message Corresponding Status of Circuit Setting Confirm the sensor connection status confirm the power supply of the device 501 at Initialization Time improve the communication environment close the exposure panel and turn the was changed power supply of the equipment ON again x5 Confirm the sensor connection status confirm the power supply of the device 502 amag e reading improve the communication environment close the exposure panel and turn the power supply of the equipment ON again Ununsuitable Confirm whether there is any dif
40. e than 2 lines Make the following treatment according to FPD Calibration Manual No D01 0138 and FPD Defect Registration Manual No D01 0145 1 Energize FPD more than 1 hour 2 Make Gain Calibration 3 Register Defect Map Remarks If the defect increases make the following treatment in Note 1 2 Note 1 2 In case that it is more than 6 months after the installation or the after maintenance of the equipment Or in case that it is more than 6 months after the installation from the completion of the assembling and the adjustment Make the following treatment according to AUGE system adjustment manual 001 0134 1 Energize FPD more than 3 hours 2 Make Gain Calibration 3 Check of the defect Adjust the brightness value of captured image to change the Tube Voltage and Tube Current 1 Check the image with brightness value of 400 to 500 Or check the image taken by lowest Tube Voltage and lowest Tube Current under test mode 2 Check the image with brightness value of 4 000 to 5 000 3 Check image with brightness value of around 10 000 without saturation 1 When there is any defect in checking of 1 to 3 register the defect 2 When Defect was registered check the defect by checking of 1 to 3 Bel Cat 39 502 180 3 When any defect occurs at 4 checking after occurrence and the registration of the defect three times take the below data and repair FPD together with the below
41. eag 21102 Joy ueeg 21021 40H 40138 91801 dng 401483X3 snouisey zm insodx3 sagnduog Qquvog 4ALYFANO 9d 9d 3 3 davoga AH pisog 1013494100 K ddng samog ddng 188 4211 Kidang 301423115 Adang samog BUI YO IMS Jojo weng Quvo8 LN TaNvd MS ues pu 13100 mm E 9 11991044 1 044 L 4941134 OSTON uo1 e nsu 199 30143185 yo Ing NG NG sang 91884 c LA007 A001 OV 401340 U91108 ajgnog 48405 10119143 snou I say Aouad 15 Bel Cat 502 180 05 Trouble shooting Note 1 To avoid an electric shock be sure to unplug power cable from electrical outlet before maintenance Also confirm whether it is shorted out or not before plug power cable to electrical outlet Note 2 Be sure to check that measuring devices and other devices do not interfere with X ray equipment before the X ray equipment is activated Note 3 For checking of wire harnesses be sure to ch
42. eck all of the electric conduction the wirings connectors and connector pins Error Number Error Message 201 Rotation of X motor is error Click RESET button Problem 1 X axis motor does not move at all 2 The abnormal sound comes out while movement of X axis motor Corresponding 1 X axis motor does not move at all Check harness between X axis motor and the connector CN2 on motor driver NG Repair the harness of CN2 OK Confirm the resistance value approximately 5 3Q between pin 1 2 2 3 3 4 4 5 5 1 in CN20X on Motor harness NG Replace X axis motor OK Confirm the voltage 24 1 0 between pin1 2 in CN1 on motor driver NG Repair the harness between CN1and CN21X OK Replace the harness CN3X and on X axis driver and axis driver And execute RESET movement Y axis moves and X axis does not move Replace X axis motor X axis moves and Y axis does not move Check the harness CN3X between motor driver and MOTOR CPU BOARD Repair the harness CN3X Replace MOTOR CPU BOARD Bel Cat 16 502 180 2 The abnormal sound comes out while movement of axis motor Confirm whether the red LED on sensor does not light and the LED D3 on MOTOR CPU BOARD lights when shading the X axis initial position sensor by scale compulsory The red LED on sensor does not stop lightning Replace X axis initial position sensor LED D3does not light Check harnes
43. een Cephalo motor and connector CN2 on motor driver NG Repair the harness of CN2 OK Confirm the resistance value approximately 5 3Q between CN2 1 2 2 3 3 4 4 5 5 1 on Motor harness NG Replace cephalo motor Confirm the voltage 24Vdc 1 0V between CN1 1 pin and 2 pin on motor driver NG Repair the harness between CN1 and CNF OK Check electric conversion of the harness CN3 between motor driver and MOTOR CPU BOARD NG Repair the harness of CN3 Replace MOTOR CPU BOARD Error Number Error Message 216 Movement of head swing motor is error Turn on the power of the X ray equipment again Problem Head rotation motor movement error Head moves in order to check the panoramic initial position but can not detect the initial position Corresponding 1 Head can rotates The error occurs when the position of the head is on panoramic side Confirm whether the LED on initial position sensor and LED D27 on CONTROL BOARD lights when shading the initial position sensor by scale compulsory The LED on sensor does not stop lightening Replace panoramic initial position sensor Check electric conversion between Repair the harness between panoramic panoramic initial position sensor and CNS1 initial position sensor and CNS1 Replace CONTROL BOARD LED D27 does not light Both are OK Bel Cat 27 502 180 The error occurs when the position of the head is on cephalometric side Co
44. exposure panel and turn the power of the equipment ON again 593 BMP 8bit File Writing Error Restart PC and turn the power of the equipment ON again 594 Error of Image Data 125 Restart and turn the power of the equipment ON again 595 Memory Obtaining Error Restart PC and turn the power of the equipment ON again 596 Image Data File Writing Error Restart PC and turn the power of the equipment ON again 597 Parameter Error Restart PC and turn the power of the equipment ON again 598 Memory Obtaining Error Restart PC and turn the power of the equipment ON again 599 Film Sending Speed Data File Check that FILM FOLDER or Necessary Film Sending Speed Reading Error Data File in FILM Folder exist AG_ dat Bel Cat 38 502 180 The ring shaped artifact occurs on CT axial image This problem might happen suddenly in rare case due to the characteristic of FPD Make the treatment in lt Note 1 gt Ring artifact Black color artifact occurs sometime lt Note 1 gt The action method when ring artifact occurs on the image 1 1 In case that it is within 6 months after the installation or after maintenance of the equipment 1 1 1 When the number of defect is 1 pixel or 1 line Register the Defect Map according to FPD Defect Registration Manual No D01 0145 Remarks If the defect increases obey the instruction of below Note 1 2 1 1 2 When the number of defects is more than 2 pixels or mor
45. ference in the value of the parameter set to 503 Parameter Value that AUGE CCD ini between the status when it is shipped and the status when it is transferred by factor installed s Dn Confirm the sensor connection status confirm the power supply of the device 504 Baie Soin ving improve the communication environment close the exposure panel and turn the power supply of the equipment ON again 505 Failure of securing The memory of is insufficient Restart PC If the error cannot be resolved memory in function increase the memory Access faifure to E 506 Communication Control Confirm the file exists in the path of the program file for communication control Program Load File that default is C NeoPremium panorama043 bin filled in AUGE CCD ini was set by factor Accesstalureto Confirm the file exists in the path of the program file for exposure control default 507 Exposure Control is C YNeoPremiumYce pano004t mcs for Panoramic and Program Load Filethat C YNeoPremiumYce cephalo003t mcs for cephalometric filled in was set by factor AUGE CCD ini 508 Access failure to file Confirm the file exists in the path of the wound correction data file and the that was set by factor calibration data file filled in AUGE_CCD ini Bel Cat 36 502 180 Communication Control Assembly was not initialized or exposure Confirm the sensor connection status confirm the power supply of the device 50
46. g of wearing wound tension and twist Checking of looseness deformation and Damage Checking of twist and damage Gap of exposure field and exposure width Contents of Check Checking of slip abnormal noise and stopping accuracy Checking of looseness operativeness And stability Checking of brightness and positioning accuracy Checking of cutting of the wires and terminals Checking of the movement Checking of the smooth movement Checking of X ray Tube Voltage Checking of X ray Tube Current Checking of exposure time 99 Symbol 502 180 Wiring Diagram Nga Emergency Stop Option Outside m ense Power Supply Switching Box Aa mmm mema on tne option is unused ONLA 12P NG 5 6 s mj LE C Pues C pP P Rotation Unit Terminal Table Rota nm Ed Adapter for STP pesi deer EPD LAN LAN 852320 X RAY SW m een Map d is shart circui ahon unused apo
47. gen AUGE_LINK CORRIMGSIZE IFA Insert Damage P arameter for changing size Correction File s name DEFAU LTOFFSET T ua Insert adjustment value Bel Cat 96 502 180 09 Contact Information Please contact our sales office or a distributor near you Belmont Equipment A Division of TAKARA BELMONT USA Inc 101 Belmont Drive Somerset 08873 Toll Free 800 223 1192 Toll Free Fax 800 280 7504 www belmontequip com TAKARA COMPANY CANADA LTD 2706 South Sheridan Way Mississauga Ontario Canada L5J 2M4 Toll Free 800 268 5351 Fax 905 822 6203 www takarabelmont ca Bel Cat 97 502 180 10 Service Maintenance N Warning It is danger to open the covers of the equipment at time of the maintenance Do not open any covers of the equipment excepting the maintenance dealer Specialized knowledge experience and special measuring instruments are required to check this equipment To maintain the performance of equipment please perform daily check with eyes and implement periodic maintenance by dealer service personnel Maintenance List for Dental X ray Equipment Maintenace by maintenance dealer once a year Symbol Checking Method O Check with eyes Check by using measuring instruments etc Check by operating the equipment Check Item Purpose of Check Contents of Check Symbol Check the conformity of Electric power voltage connec
48. ightning Replace Y axis initial position sensor LED D4 does not light Check electric conduction of harness between Y axis initial position sensor and CNSY Repair the harness between Y axis initial position sensor and CNSY Replace Y axis initial position sensor Replace MOTOR CPU BOARD Both are OK Bel Cat 18 502 180 Error Number Error Message 203 Rotation of R motor is error Click RESET button Problem R axis motor does not move at all or the abnormal sound comes out while movement Corresponding 1 R axis motor does not move at all Check harness between R axis motor and the connector CNe on motor driver Repair the harness of CN3 Confirm the resistance value approximately 5 3Q between pin 1 2 2 3 3 4 4 5 5 1 in CN20X on Motor harness NG OK NG Replace X axis motor OK Confirm the voltage 24Vdc 1 0V between 1 2 in CN1 on motor driver Repair the harness between CN1and CN20R NG OK Check electric conduction of harness CN2 and CN20R between motor driver and MOTOR CPU BOARD Repair the harness CN2 and CN20R Check electric conduction of harness CN21R between MOTOR CPU BOARD and CONTOL BOARD NG OK Repair the harness CN21R between MOTOR CPU BOARD and CONTROL BOARD Confirm whether the LED D18 20 on CONTROL BOARD lights when RESET button is pushed NG Replace MOTOR CPU BOARD K Replace CONTROL BOARD NG OK 19 502 180 2
49. ipment and has some high voltage portions inside Turn off the power of equipment and unplug power cable from electrical outlet before opening cover of equipment for repair work DANGER Some repair works involve risks The repair works shall be performed by a qualified personnel or a person who completed the training specialized in repair at our company A DANGER Some repair works involve risks Service personnel must give instructions to outsiders to stay away from repair work area WARNING Do not modify medical electronic devices Regulations such as the following apply to manufacturers based on the Pharmaceutical Affairs Law Medical devices require item specific manufacturing approval and an Application for Approval of Partial Changes is mandatory for any functional modifications of medical devices unauthorized modifications are prohibited WARNING Radiation Protection in Dentistry Comply with the contents of each clause regarding protection against radiation exposure prescribed in relevant medical regulations when installing and using dental X ray equipments WARNING This equipment for radiograph can cause hazard to service personnel if safe exposure conditions and how to use are not complied with Bel Cat 7 502 180 A WARNING Be sure to sterilize equipment portions to be touched by patient or operator before starting repair work After
50. m line to the patient s eyes Bel Cat 11 502 180 03 Preparation for Repair Work 1 Manuals Bel Cat Service Manual Bel Cat Operation Manual 2 Measuring Instruments 1 Digital multi meter 2 Oscilloscope 3 Lead Wires for measurement Insulation resistance meter X ray detecting paper lt gt wT DH Phillips screwdriver Large size small size Flat braded screwdrivers small size Insulated type Box drivers for M3 4 5 6 screws Hexagonal wrench a set of wrenches Nipper Long nose pliers Electric soldering iron insulated type solder Taps for thread Tap handle 0 Electric drill 1 Drill 2 Crimp tool for crimping terminal 4 Jigs for Adjustment 1 Test piece 2 Lead plate with thickness of 3mm or more for X ray protection Brass plate filter with thickness of 2mm Aluminum filter with thickness of 35mm X ray detecting paper 5 Repair and Service Parts 1 Depending the contents of repair please prepare necessary boards parts wire harness etc 6 Other 1 Depending the contents of repair please prepare screws nuts crimping terminals grease etc 2 Alcohol for disinfection 3 Cloth 4 Clearing solvent Bel Cat 12 502 180 04 Specifications and configuration for the equipment lateral frontal ion ratio time Acauisition area Magnification ratio bone Acquisition area Magnification ratio M
51. mmunication Control 519 DLL File cannot be read Confirm that there is the file in Pass of DLL File Default is C NeoPremium panLan dll Check the program file for exposure control written in AUGE_CCD ini and Error of Exposure Control contin 520 Program Loard Fil the following situation 1 Panorama is included in YNeoPremiumYce pano004t mcs 2 Cephalo is included in C YNeoPremiumYce cephalo003t mcs Data Error by T s 521 comunicato Check the communication environment 522 Invalid image data for There is almost no possibility that this error happens making correction data When this errors happen contact us No density difference by different value of CCD hili 523 driving data in Image data ats ji almost no possibility that this error happens for completing proofreading en this error happens contact us data 524 Communication Control There is possibility that Panorama043 bin is broken Program Load File Error Replace with the correct file Bel Cat 37 502 180 CCD Error Message Error Message Corresponding Initialization error occurred There might be possibility that plural applications for control of 551 Communication Error f sensor are trying to start Then finish one of the applications Communication is not established or re start up PC KU Confirm the sensor connection status confirm the power supply of 552 UNE the device improve
52. n CNM when the CCD holder drive switch is pushed 24Vdc 41 0V can not be detected Replace CONTROL BOARD 2 When the Motor is working Confirm whether the LED on sensor and D25 and D29 on CONTROL BOARD lights when shading the panoramic initial position sensor by scale compulsory Replace panoramic initial position sensor Check electric conversion between Repair the harness between panoramic panoramic initial position sensor and initial position sensor and CNS2 CNS2 The LED on sensor does not stop lightning LED D25 and 29 does not light Replace CONTROL BOARD Both are OK Bel Cat 29 502 180 Error Number Error Message 218 Movement of flat beam motor is error Turn on the power of the X ray equipment again Problem Frankfort beam does not move at all Corresponding Motor does not move at all Confirm the resistance value between 1 pin and 2 pin on CNYU board connector on CNHMO harness side is approximately 30Q NG Check electric conversion between motor and CNHMO NG Repair or replace the harness between motor and CNHMO OK Replace motor OK Confirm the voltage between CNHMO 1 and 2 on CNYU BOARD side is 24Vdc 1 0V when up and down key of Frankfort beam is pushed NG Check electric conversion between motor and CNHMO NG Repair or replace the harness between motor and CNHMO OK Replace MOTOR CPU BOARD OK Error Number Error Message 219 Something is wro
53. nd Board ASSY 1 0620 978 60007 01 Photo micro Sensor EE SY671 1 0621 408 10450 CNP BOARD ASSY 1 Relay Adopter for STP 0622 988 50047 Welcome 1 Bel Cat Bel Cat does not have Digital Sensor 72 502 180 7 Rest Unit ASSY 0702 0703 0701 0706 0704 73 502 180 Rest Unit ASSY Drawing Number Drawing Name 0701 108 01266 Chinrest Main Body 1 0702 108 01267 Chinrest ASSY Cover 1 0703 408 11699 Cover 1 0704 308 04055 Grip Handle 1 0705 974 80072 DC Power Supply CME150 24 1 0706 408 10584 01 DC POWER 1 ASSY 1 0707 408 07586 01 BEAM MOTOR BOARD ASSY 1 Rela HE2aN S DC24 HAE2232 0709 942 60022 07 Fan KDE2406PHV2 1 0710 408 11383 CNKI BOARD ASSY 1 0711 308 04049 Indicator ASSY 1 0712 932 21015 01 Power Grip GT Belt 495 3GT 9 1 0713 308 04042 Operation Panel ASSY 1 0714 978 60005 02 Photo micro Sensor EE SPX403 1 0715 965 85012 03 Circuit Protector NRF110 1A 1 Bel Cat 74 502 180 7 1 0711 Indicator ASSY mae KO i LANG
54. nfirm whether the LED on cephalometric initial position sensor and LED D30 on CONTROL BOARD lights when shading the initial position sensor by scale compulsory The LED on sensor does not stop lightening Replace cephalometric initial position sensor Check electric conversion between cephalometric NG Repair the harness between cephalometric initial position sensor and CNS1 initial position sensor and CNS1 LED 030 does not light Replace CONTROL BOARD Both are OK 2 Head does not rotate Confirm the voltage 24Vdc 41 0V between 1 and CP2 on BEAM MOTOR BOARD Cable check of harness between BEAM MOTOR BOARD and CONTROL BOARD NG Repair the harness between BEAM MOTOR BOARD and CONTROL BOARD OK Replace CONTROL BOARD OK Confirm the voltage 24Vdc 41 0V between and CP2 Replace fuse F1 Confirm that the resistance value between CNM 1 2 3 4 on motor harness is 42 NG Replace motor Replace BEAM MOTOR BOARD NG NG K OK Bel Cat 28 502 180 Error Number Error Message 217 Movement of CCD placement motor is error Push Drive Switch of CCD holder Problem CCD storage motor does not move at all Corresponding 1 When the Motor does not move at all Check electric conversion between CCD storage motor and CNM on CONTROL BOARD NG Repair the harness between CCD storage motor and CONTROL BOARD Confirm the voltage 24 0 between pin1 2 i
55. ng ASSY 1 With Cephalometric 0403 308 03972 Fixing Plate 1 for P C Board 1 Power Supply of 1 0kW High HOS 2 8 800000 Tension Voltage for pantalla 1 0405 408 10743 00 PFC AC FILTER ASSY 1 0406 408 11758 Fixing Plate for Counter Weight 1 0407 408 11481 Fixing Plate for Fan 1 0408 942 60011 05 DC Axis Fan MDS510 24 1 0409 965 60057 11 JM Relay JM1aN ZTM DC24V 1 0410 408 11478 Balance Weight C 1 0411 408 11759 Balance Weight D 1 0412 408 11482 Fixing Plate 2 for P C Board 1 0413 408 11473 CONTROL ASSY 1 0414 975 00029 02 Ferrite Core TFT 152613N 03 2 0415 308 03970 Frame 1 0416 308 03971 Boss D 1 0417 408 11475 Pressing Plate 4 0418 408 11476 Fixing Plate for Terminal Table 1 Terminal Table ML 20 0419 971 00045 02 ee 1 0420 408 11477 Cover 1 0421 308 03973 02 Head Swing ASSY 1 2 ephalometric 0422 923 92004 05 Fine U Nut 045 1 0423 108 01251 ARM Cover 1 1 0424 108 01252 ARM Cover 2 1 0425 408 11479 01 ARM Fixing Metal Plate 1 1 0426 408 11479 02 ARM Fixing Metal Plate 1 1 0427 408 11480 ARM Fixing Metal Plate 2 2 Bel Cat 65 502 180 4 1 402 Head Swing 55 0431 2 eri 2 0 Tool NG
56. ng with EEPROM Turn on the power of the X ray equipment again Procedure If the problem can not be improved replace CONTROL BOARD Bel Cat 30 502 180 Error Number Error Message 220 Take head support off Problem The error of Head support detection sensor Corresponding Insert the driver into the shaft hall for fixing head support located on beam side And confirm whether the LED D33 on MOTOR CPU BOARD lights when driver is inserted and whether it does not light when driver is pulled out Both are OK The LED on sensor does not stop lightening LED D33 does not light gt Replace Head support detection sensor Check electric conversion Repair harness between between sensor and CNS2 sensor and CNS2 Replace CONTROL BOARD Error Number Error Message 222 Sliding unit is error Turn on the power of the X ray equipment again Problem The sensor cannot be detected even if it moves during the fixed time Corresponding Sliding unit moves up down Confirm that the voltage is more than 24Vdc between CN3 1 pin and 2 pin on DC MOTOR BOARD in Pulley assy about 4 seconds after up down button is pushed More than 20Vdc Less than 20Vdc Bel Cat Check electric conversion of harness CN3 on DC motor Repair the harness NG CN3 OK Replace DC motor Confirm whether the voltage between MOTOR CPU BOARD
57. ole computer is error Turn on the power of the X ray equipment again Problem Communication error Corresponding Check electric conduction of harness CNO between MOTOR CPU BOARD and PC NG Repair the harness between MOTOR CPU BOARD and PC Replace MOTOR CPU BOARD Bel Cat 23 502 180 Error Number Error Message 211 Rotation of indicator beam motor is error Turn on the power of the X ray equipment again Problem Indicator beam motor does not move or the abnormal sound comes out while movement Corresponding 1 Indicator beam motor does not move Confirm the voltage between 1 and CP2 on BEAM MOTOR BOARD as 24 1 Check electric conduction of the harness CNMB CNBMOT between BEAM MOTOR BOARD and CNYU BOARD Repair the harness CNMB CNBMOT Confirm the resistance value as 362 between CNM 1 and 2 3 and 4 on motor harness NG Replace motor Check electric conversion of harness CNMB CNBMOT between BEAM MOTOR BOARD and CNYU BOARD Repair the harness CNMB CNBMOT Replace BEAM MOTOR BOARD Confirm whether the red LED on initial position sensor does not light and the LED D12 on MOTOR CPU BOARD lights when shading the indicator beam initial position sensor by scale compulsory 2 The abnormal sound comes out while movement The red LED on sensor does not stop lightning Replace the Indicator beam initial position sensor LED D12 does not light Check ele
58. om and install the equipment body in the X ray examination room WARNING Do not place anything that can be an obstacle within the range of equipment movement WARNING X ray exposure and approach to equipment must be done under the responsibility of user when repairer patient or nursing attendant allowed by doctor has a pacemaker etc WARNING Be sure to use positioning attachment specified for each exposure mode when positioning patient WARNING Be sure to make patient and assistants to wear X ray protective clothing Assistants in this sentence means a person allowed by doctor WARNING Operator must instruct patient not to move while X raying A WARNING Be sure to sterilize and disinfect equipment portions touched by patient or operator after X raying and at daily closing time A WARNING To avoid equipment failure do not rotate arm by hand WARNING Keep patient unmoved until arm reset operation completed after X ray exposure A CAUTION Take extra care for patient safety when positioning patient Bel Cat 10 502 180 A CAUTION Take extra care for patient safety when moving sliding unit up and down A CAUTION The laser beam line uses a safe laser beam Do not look straight at positioning laser beam for your safety Also give this caution to patient and nursing attendant A CAUTION Irradiate bea
59. on 1 Read the Operation Manual of the equipment thoroughly before maintenance 2 The operating procedure of panoramic exposure forms the basis of all operations So become fully familiar with its procedure before taking other exposure mode 3 Be sure to observe the warnings and prohibited matters in the body of this Service Manual strictly 4 Read this Service manual in order from the beginning Reading only portions of the manual may cause accidents or breakage of the equipment from improper maintenance 5 If you find any unclear point during a repair work read this Service Manual again to check about it and restart the repair work 6 Disclaimer Follow all information in this Service Manual Takara Belmont USA Inc is not responsible for malfunction accidents or other such incidents resulting from improper maintenance 7 Warranty period This equipment is warranted for a period of one year from purchase Repair will be provided free of charge only in cases of malfunction or damage to equipment in the course of ordinary use 8 Serviceability of Equipment and Period of Spare Parts Availability Service and spare parts will be available for ten years from discontinued production of this equipment 9 When disposed this equipment constitutes industrial waste Please contact the agent or contract for disposal with an industrial waste handler licensed as legislated and regulated by the local governing authority Bel Cat 4 502
60. ors hg Sound video and game controllers lt gt Storage controllers System devices Universal Serial Bus controllers Bel Cat 87 502 180 3 Select Port Setting tab and click Advanced button Communications Port COM1 Properties General Port Settings Driver Details Resources Bits per second 9600 Data bits 8 Parity None Stop bits Flow control Advanced Restore Defaults 4 The window as shown below will be displayed Change the Receive Buffer and Transmit Buffer to Low For the COM Port Number if the selected number is not COM 1 change the number to COM 1 Advanced Settings for iw 2 1 Select lower settings to correct connection problems Select higher settings for faster performance it Receive Buffer Low 1 High 14 14 Transmit Buffer Low 1 High 16 16 Bot Number Bel Cat 88 502 180 3 Copying Damage Correction File 1 item 1 4 and install AUGE CCD Sensor Calibration Software 2 E xecute C SettingCalib_Auge SettingCalib exe 3 Specify the Damage Correction file dat from the Sensor CD There are 2 methods to specify the file please follow the instruction below The in the file name is the Sensor Number Drag and Drop correction file to this window
61. rd 1 DC Motor Board 2 CNYU Board 3 MOTOR CPU Board 4 POWER Board 5 Rotation Motor Driver 6 Y Axis Motor Driver 7 X Axis Motor Driver 8 Switching Power Supply 9 CNP board 10 CNC Board 11 CNB Board 12 Head Sliding Motor Driver 13 High Frequency Inverter Board 14 PFC AC Filter Board 15 CONTROL Board 16 Collimator Motor Board 17 High tension generator board 18 Power supply for FPD 19 DC DC CONVERTER board 20 CNP board 21 Switching power supply 22 Beam motor board 23 DC POWER 1 board 24 PANEL R board 25 PANEL L board 26 CNK board 27 CNM board 28 Cephalometric motor board 29 CM SW board 42 502 180 07 Parts List Bel Cat Bel Cat PA Emergency Stop Switch Sliding Unit Rotating Unit Head Support Arm Unit a lt p Y X ray Generator POWER Switch Switching Box Rest Unit amy Image Receptor Grip CCD Sensor Panoramic X ray Exposure Switch Head Holding Knob Bel Cat does not include CCD sensor Bel Cat 43 Operation Panel Stand 502 180 Sliding Unit Emergency Stop Switch Rotating Unit Generator POWER Switch Switching Box Bel Cat CCD Sensor Panoramic side Head holding knob Arm Unit Head Support Cephalometric Operation Arm Panel Rest Unit Grip Stand CCD Sensor Cephalometric side
62. ring 68677 408 11562 Shaft 408 11563 Screw Bar 930 00011 63 Bearing F686ZZ 408 11564 Sensor Fin 978 60004 13 Photo micro Sensor EE SX674 408 11565 Slider AN ARA Ss NS 962 05006 Laser Marking Projector LDV167LS 408 11568 Holder 408 11567 01 Base 408 11567 02 Base 408 11569 Beam Fixing Plate 932 41025 02 Coupling ALS 014 Y 6 6 408 11570 Motor Fixing Plate 934 30005 01 ISODUMP Gromet G 410 1 977 60001 07 DC Geared Motor TG 38E LG66 937 38008 40 Brass Space CB 304 078 60099 Connector EE 1001 Bel Cat 937 38041 01 Steel Spacer CF 602E 52 1 1 10021 41 4 4 1 403 C2 502 180 1 4 0121 Switch ASSY 0168 Drawing Number Drawing Name 408 11573 Plate Emergency Stop Button Switch 7649006 964 28001 02 Switch A8G 107 1G 24 614 01445 Stop Seal 408 11574 Cable Guide Bel Cat 53 502 180 1 5 0122 Up Down Movement Sensor ASSY Drawing Number 408 11559 4 p Drawing Name Sensor Fixing Metal Plate 978 60007 02 Photo micro Sensor EE YS672 Bel Cat 078 60099 Connector EE 100
63. s dialogue is displayed Default value is already set as 10 in Number of Frame Buffers in the dialogue Click OK button 7 After a few seconds push red square button on right of Tool Bar in the screen of VIVA Main Window 8 Push Retrieve Image button on the top of the screen of VIVA Main Window to capture a image 9 Select Save As from drop down menu by right click of the mouse on the image and save the image by file format as VIVA 10 Select Acquisition Menu and System Settings 11 System Settings dialogue is displayed Put check mark in Offset Corrections Gain Corrections and Pixel Defect on right side of the dialogue push Apply button and push OK button Then shut down the VIVA software Reference 1 If any defect will occur on the next line or on the pixel after the registration of the Defect there is possibility that the hardware might be defect Reference 2 If any defect will occur intensively in the range of ASIC unit Strip of paper of vertical direction in every 64 pixel there is possibility that the hardware might be defect Bel Cat 40 502 180 06 Configuration for Printed Circuit Board 1 Overall Drawing Printed Circuit Boards are located as shown in the below drawing Bel Cat 41 502 180 Name of Printed Circuit Boa
64. s between X axis initial position sensor and CNSX Repair the harness between X axis initial position sensor and CNSX Replace MOTOR CPU BOARD Error Number Error Message Both are OK 202 Rotation of Y motor is error Click RESET button Problem 1 Y axis motor does not move at all 2 The abnormal sound comes out while movement of Y axis motor Corresponding 1 Y axis motor does not move at all Check harness between Y axis motor and the connector CN2 on motor driver NG Repair the harness of CN2 OK Confirm the resistance value approximately 5 3Q between pin 1 2 2 3 3 4 4 5 5 1 in CN2 on Motor harness NG Replace Y axis motor OK Confirm the voltage 24 1 0 between pin1 2 in CN1 on motor driver Repair the harness between CN1and CN21Y Replace the harness CN3Y and CN3X on Y axis driver and X axis driver And execute RESET movement NG OK X axis moves and Y axis does not move Replace Y axis motor Check the harness CN3Y between motor driver and MOTOR CPU BOARD Y axis moves and X axis does not move Repair the harness Replace MOTOR CPU BOARD Bel Cat 17 502 180 2 The abnormal sound comes out while movement of axis motor Confirm whether the red LED on sensor does not light and the LED D4 on MOTOR CPU BOARD lights when shading the Y axis initial position sensor by scale compulsory The red LED on sensor does not stop l
65. sensor and wire harness of CNS2 LED does not light OK Replace CONTROL BOARD Replace sensor for existence of CCD sensor LED lights The vertical line appears on Cephalometric image 1 Confirm whether the exposure area is aligned on the center of cephalometric CCD when X ray exposure If the exposure area is not aligned on the center adjust the secondary slit in accordance with adjustment manual Bel Cat 34 502 180 FPD Error Message Error Message Corresponding 301 Communication Error 302 State Error 303 Initialization error OR Device not open 1 NC the power of the equipment OFF an 319 Video Not Initialized 2 Check whether the communication 330 Timed out between FPD and the equipment can be XE made and the images can be read with 4 83 vr handisedxcaphann dr ViVa Refer to the below procedure 383 Pleora Error 3 If the image cannot be obtained with the 384 Frame Grabber Initialization Error working in the above item 2 385 Frame Grabber Error Check the connection between the equipment and the PC 386 Error Starting Grabber Where to check for the connection Is shown as below 387 Error Stopping Grabber Check one by one and make the working in Item 2 388 Error Starting Record Sequence 1 Pull out and inspect LAN cable FPDPA FPDCE behind of PC 389 Error Stopping Record Sequence 2 Pull out and in
66. spect LAN cable E and on switching box behind of the 390 Error Communicating With Frame Grabber equipment 405 Receptor Not Ready 3 Remove the back cover of image detector and pull out and inspect LAN cable 462 set user sync error for Pleora on the center 466 4 Remove the Rest unit cover and Sliding IS I ONE unit cover and pull out and inspect LAN 467 LAN link time out error connector FPDPA and at the right of Sliding unit internal 468 RS232C link time out error 470 Sync check error Bel Cat 35 502 180 The image capture procedure with VIVA 1 Attach Chinrest for CT to the rest unit at the highest position 2 Capture the image with VIVA Start menu P C gt Program gt Varian gt PaxscanL05 gt Open VIVA with Push Offset Calib on tool bar It is necessary when VIVA started Push the icon on vertical tool bar on right side of the screen Set the number of capture image sheets as 10 from Video menu Drag the thumbnail image on the upper column of the screen to the image display frame on lower column with a mouse Open TAUGE TEST MODE SOFT from C NEOPREMIUM Calibration Take off the check of OkV area Median Exposure The equipment goes to the initial position READY Click OK button The equipment goes to the initial position Make th
67. the communication environment close the Communication reino established exposure panel and turn the power supply of the equipment ON again nu Confirm the sensor connection status confirm the power supply of 553 the device improve the communication environment close the Communication is not e tabli hea exposure panel and turn the power supply of the equipment ON again EMI Confirm the sensor connection status confirm the power supply of 554 the device improve the communication environment close the Communication iS not established exposure panel and turn the power supply of the equipment ON again Confirm the sensor connection status confirm the power supply of Initialization error occurred the device improve the communication environment close the 555 Communication Error exposure panel and turn the power supply of the equipment ON Communication is not established again Also check the setting of the IP Address too Confirm the sensor connection status confirm the power supply of Initialization error occurred the device improve the communication environment close the 560 Communication Error exposure panel and turn the power supply of the equipment ON Communication is not established again There is possibility that the sensor is defective s Check the power of the equipment check the connection of 570 re indie sI asi RS232C Cable check the communication environment close not established The
68. ting with the equipment and Environment of Electric Power Checking of the range of electric power voltage Source Specifications of the equipment 1 Appearance of a Deformation Damage Description Plate Appearance and The equipment b Note indication 2 Inside of the equipment Siam dust Temperature Checking of conformity of the environmeni n humidity and gass Atmosphere where the equipment is used Levelness of the Check of the influence to the equipment equipment Floor and fixing GHING situation of the Vibration and working stability of the equipment equipment equipment Checking of Checking of the safety within the working intervention Area of the equipment es Checking of the condition of rusting and safety Rusting situation Of the function O Working test before Working of the Checking of the working condition of the equipment Working test equipment oring tas alok Checking of the working condition of the equipment The maintenance Checking of the resistance values between Insulation resistance The electric line and earth Leakage current of Checking of current value that leaks from Safety test of The exterior The exterior of the equipment to the earth Electric shock Resistance value between the metal portion of the 2 outside of the equipment and tip of the earth Resistance of earth Resistance value between the earth terminal 2 Wire Of the equipment and
69. ver the sensor completely with the Calibration J ig set the exposure condition to 64kV for the Voltage and 12mA for the Current 4 Click the Calibration button messages as shown below will be displayed press the X ray Exposure Switch to irradiate the X ray C SettingCalib_Auge cephalo_lan Ceph_smpl exe KO Exposure Conditions AG 2 A 64kU 12 HHH Required X ray Exposure Time 5 079040 1 seconds HAH Calibration Mode for Panorama Cooling Finished 2011712720 16 38 24 Starting Image Acquisition Please Apply the X Rays Starting Image Acquisition Standing By ww 5 When the message below is displayed release the switch to stop the X ray immediately wait for the result r C SettingCalib_Auge cephalo_lan Ceph_smpl exe Exposure Conditions AG_P2_A 64 12 HHH Required X ray Exposure Time 5 079040 1 seconds HAH Calibration Mode for Panorama Cooling Finished 2011712720 16 38 24 Starting Image Acquisition Please Apply the X Rays Starting Image Acquisition Standing By ee Image Acquisition Started 2611 12 26 16 38 46 Acruinineg Image ot seeeee Acquisition Finished 2611 12 28 10 38 51 Bel Cat 92 502 180 5 If the Average value result is below the Target the calibration failed Increase the PICCDCT CNTL1 value and the repeat the calibration from step 4 4 Average 685 Target 640 Limit Exceeded ErrMsg 22 ErrC