1. MODE The SET MODE indicator should flash This allows you to change from the default power regulation mode to current regulation mode If the mode is correct and no changes are desired pressing DISPLAY SELECT will move on to SET LIMIT To change the mode press the up or down arrow This will alternately illuminate the CURRENT and POWER indicators Press SET ENTER to enter the selection and move on to the next item SET LIMIT The SET LIMIT indicator should flash This allows you to change the power limit or current limit within the range of 40 300 WATTS and 1 5 12 0 AMPS dependant upon the mode of operation The display shows the limit in AMPS if in current mode and WATTS if in power mode with the least significant digit blinking If the limit is correct and no changes are desired pressing DISPLAY SELECT will move on to MEMORY SAVE To change the limit press the up or down arrow Pressing the left or right arrow button will flash the digit to the right or left of presently blinking digit and allow changing that digit by using the up or down arrows Pressing SET ENTER will lock this value in as the new limit and advance the setup to memory MEMORY The MEMORY indicator will flash The SAVE led will be illuminated allowing you to save up to five front panel setups If you do not wish to save any parameters pressing DISPLAY SELECT will move on to MEMORY RECALL The display shows 1
2. section 3 and adjust the power limit to a value 10 higher then the desired operating point Press the LAMP START button The igniter will fire for approximately 5 seconds or until the lamp starts at the preset level Once the preset level is reached the LAMP ON indicator will illuminate You can use DISPLAY SELECT to display current voltage watts or lamp operating time To shut off the lamp press the LAMP OFF button Once the output reaches 0 the LAMP ON indicator will extinguish Should the lamp fail to ignite refer to the troubleshooting section 111 6 NOTE If 68950 is being used the 68950 must be off until lamp is stabilized by power supply Once control is shifted to 68950 all front panel controls become inactive 14 M 69907 ARC LAMP POWER SUPPLY 11 6 TROUBLESHOOTING This chart provides the basic troubleshooting information for the Newport Model 69907 power supply when used with Newport 669XX series arc lamp housing Contact a Newport sales engineer or your local representative if more information is required e Fan not turning e Display does not turn on Table 4 Symptom Action Power supply does not turn on e g Check e Mains cord connection e Front panel circuit breaker on e AC power at wall outlet Fault light e Display shows Turn off AC power and Check e Cables connected to lamp housing e Door to lamp housing is closed e f lamp was running before fault was fa
3. which means save to memory location 1 Pressing the up or down arrows will increment the number through the 5 possible locations Pressing SET ENTER will save the active parameters to this location for future use by recalling the parameters from memory and advance the setup to memory recall NOTE Be sure to record what is saved to each location as that is the only way you will know what it contains without recalling each location and checking the parameters The MEMORY indicator will flash The RECALL led will be illuminated allowing you to recall up to five front panel setups that you had previously saved If you do not wish to recall any parameters pressing DISPLAY SELECT will move on to HRS RESET The display shows 1 which means recall from memory location 1 Pressing the up or down arrows will increment the number through the 5 possible locations Pressing SET ENTER will recall the saved parameters from this location to the active parameters and advance the setup to HRS RESET HRS RESET The HRS RESET indicator should flash The display will indicate accumulated lamp hours since last reset If you do not wish to reset the lamp hours pressing DISPLAY SELECT will move on to SET MODE Pressing and holding SET ENTER until the display goes blank will reset lamp hours and advance setup to SET MODE 12 4 ILS M 69907 ARC LAMP POWER SUPPLY LAMP OPERATING TIME FUNCTIO
4. ATING THE LAMPo cite teh i ce tere aee ee ree Ir guile bea area 13 IEE TROUBLESHOOTING aea thee esi tee re ate Eee des 15 ILZ RS 232 COMMUNICATIONS sr 16 APPEICATIONS RAE 18 REMOTE CONNEGTOR ne dede de Hd sede t veel ia dert aes Da 18 SPECIFICATIONS nan an ant an a Le e t e and venden e dta edid eed uo 19 DECLARATION OF CONFORMITY rss 20 WARRANTY AND RETURNS raa 21 M 69907 ARC LAMP POWER SUPPLY INTRODUCTION DESCRIPTION FEATURES The Newport model 69907 power supply was designed to meet the needs of a regulated source of power or current for proper operation of high power ARC light sources The 69907 provides constant power current operation of these sources of radiation which is usually required whenever a radiometric measurement is being made or whenever highly stable light output is needed Constant current operation is also available Features include Adjustable output with preset so that the output can be set before running the lamp Digital display is included for precise monitoring of current voltage power and lamp running time LED indicators show the status of important power supply functions Start stop control ignites lamp at preset value to minimize overshoot undershoot of power to lamp Stop allows shut down of lamp with continued cooling until lamp reaches room temperature Safety interlock connector provides a way of safeguarding against accidental exposu
5. CONDITIONS OR LIABILITIES EXPRESSED OR IMPLIED ARRISING BY LAW OR OTHERWISE WHETHER OR NOT OCCASIONED BY NEWPORT S NEGLIGENCE This warranty shall not be extended altered or varied except by a written document signed by both parties If any 21 portion of this agreement is invalidated the remainder of the agreement shall remain in full force and effect CONSEQUENTIAL DAMAGES Newport shall not be responsible for consequential damages resulting from misfunctions or malfunctions of the goods described in this manual Newport s total responsibility is limited to repairing or replacing the misfunctioning or malfunctioning goods under the terms and conditions of the above described warranty INSURANCE Persons receiving goods for demonstrations demo loan temporary use or in any manner in which title is not transferred from Newport shall assume full responsibility for any and all damage while in their care custody and control If damage occurs unrelated to the proper and warranted use and performance of the goods recipient of the goods accepts full responsibility for restoring the goods to their condition upon original delivery and for assuming all costs and charges RETURNS Before returning equipment to Newport for repair please call the Customer Service Department at 203 377 8282 Have your purchase order number available before calling Newport The Customer Service Representative will give you a Return Material Aut
6. M 69907 AD NNewvvport ORIEL PRODUCT LINE 150 Long Beach Boulevard Stratford CT 06615 Phone 203 377 8282 800 714 5393 Fax 203 378 2457 E MAIL oriel sales newport com ARC LAMP POWER SUPPLY MODEL 69907 USER MANUAL Please read these instructions completely before operating this equipment The specification and operating instructions apply only to the model s covered by this manual If there are any questions or problems regarding the use of this equipment please contact Newport or the representative from whom this equipment was purchased Rev 10 06 08 M 69907 ARC LAMP POWER SUPPLY TABLE OF CONTENTS INTRODUCTION ene itr i ed a teh dei dedi ete teet 1 1 DESCRIPTION FEATURES uir A e er iet eroi Oai 1 Pie Rte TR 1 l 3 ORDERING INFORMATION arroztegi 2 1 4 69907 REPLACEMENT ITEMS iiti etit eet eer ec enti ie OI ER Rd erai 3 atazetan uo ta aon a tata aerea 4 IK TRUE 4 1 2 ELECTRICAL HAZARDSO fii aN ntt aaa da teg ae ha sap lo Ee Rd MR rrea 4 bei FIRE HAZARDS ite de aa ite tl m heces item iieri ias 4 1 4 KEEPING THE LAMP IN GOOD CONDITION rrura 4 USING THE POWER SUPPLY ka Eo redet tet arke PERRA nad 5 1 1 REAR PANEL CONNECTIONS rrura 5 ILZ FRONT PANEL CONTROLS AND DISPLAYS sse eene 10 3 SETUP MODE aka ert etie er pa ce ee qi a eee es RR eder nra arto 12 1 4 LAMP OPERATING TIME EUN TION rss 13 IES OPER
7. N The 69907 keeps track of operating time whenever the lamp is running The digital display shows operating time from 0 9999 hours This function may help you determine when to replace the lamp and monitor the performance over time Refer to Table 1 for average lifetimes of compatible arc lamps Toreset the operating time to O hours shut off the lamp if itis presently running and enter the setup mode and follow procedures in section III 3 OPERATING THE LAMP Please refer to Figure 4 CURRENT MODE If you are not using a NEWPORT light source use the manufacturers specifications If the lamp operating power and voltage are known but the operating current is not then determine the current setting by using Ohm s Law amperes power volts Turn on the power supply The CURRENT MODE LED should illuminate If not switch to CURRENT see section 111 3 Set the display to read current AMPS Press and hold the SET ENTER button in until the display reads the preset current with the least significant digit blinking Using the up down arrows you can change the blinking digit use the left right arrows to select another digit Once the display is setfor the desired current press the SET ENTER button The display will now show 0 AMPS You push the SET ENTER button at any time to view the preset current NOTE If you cannot set the desired preset current press SET ENTER and then enter setup per section ILO a
8. SUPPLY Figure 4 Front Panel Display POWER This is the AC mains power switch and circuit breaker In the ON position ac power will be switched into the main circuitry of the power supply A in Figure 4 There is no output until the LAMP ON button is pressed with the interlock satisfied The circuit breaker will turn off the power supply in the event of an electrical overloaded DISPLAY SECTION The display section B in Figure 4 consists of a multi function LED display and several LED indicators It is divided into three areas meter center status indicators left and setup indicators right The METER AREA contains the display and units indicators It normally displays one of four parameters current AMPS voltage VOLTS power WATTS and Lamp operating time LAMP HOURS It is also used when setting current power preset current power limit user memory load save and to display type of fault The STATUS AREA contains seven indicators LAMP ON Flashes while ramping up to and down from the preset current or power level and illuminates continuously when the output is at the preset value LIMIT Illuminates when the output current power reaches the limit POWER MODE Illuminates when the power supply is set to regulate output power CURRENT MODE Illuminates when the power supply is set to regulate output current FAULT Illuminates when the safety interlock loop is open Power supply failure and la
9. Y Sent command PS Response Notes STB See table 5A for ver 7 and earlier HEX Bit 7 Lamp On Bit 6 Ext Bit 5 Power Current Mode Bit 4 Cal Mode Send status of led s lit on front panel 1 LED ON BIT 17 power mode BIT 07 current mode Bit 3 Fault Bit 2 Comm Bit 1 Limit Bit 0 Interlock EGE ESRXXA HEX Send error register Bit 7 Power On Bit 6 User Request Bit 5 Command Error Bit 4 Execution Error Bit 3 Device Dependant Error Bit 2 Query Error Bit 1 Request Control Bit 0 Operation Complete AMPS 1 XX X Send amps as displayed on front panel VOLTS BKK Send volts as displayed on front panel WATTS I XXXX 1 Send watts as displayed on front panel LAMP HRS 1 XXXX 1 Send lamp hrs as displayed on front panel A PRESET 1 XX X 1 Send preset value P PRESET 1 XXXX 4 Send preset value A LIM XX X 1 Send current limit P LIM 1 XXXX 1 Send power limit IDN 1 XXXXX Send power supply model number START I ESRXX I See ESR command for Hex Start lamp update front panel value definition STOP I ESRXX See ESR command for Stop lamp update front panel value definition RST ESRXX See ESR command for Hex Reset Power Supply to Factory value definition Defaults 16 M 69907 ARC LAMP POWER SUPPLY RSTHRSJ ESRXX 1 See ESR command for Hex Reset lamp ho
10. ditions when using the power supply in these systems M 69907 ARC LAMP POWER SUPPLY LA 69907 REPLACEMENT ITEMS International Model Item Description 70010 Line Cord International Color Code 88010720 Plug Great Britain Ireland 13A 250VAC 188010801 Plug Continental Europe 16A 250 VAC 188010732 Plug Switzerland 10A 250VAC 70050 Xe HgXe Lamp interconnect cable 70051 Hg Lamp interconnect cable Table 2 1 Contact PANEL COMPONENTS CORPORATION P O Box 115 Oskaloosa IA 52577 USA 515 673 5000 M 69907 ARC LAMP POWER SUPPLY SAFETY NOTES 11 1 1 2 1 3 LA UV HAZARDS Our ARC Lamps produce considerable ultraviolet and infrared radiation Avoid excessive exposure of the eyes or skin to radiation from these lamps Protective eyewear gloves and UV Warning Signs are available from Newport 49125 UV Safety Spectacles 49126 UV Safety Goggles 49121 Protective Gloves 79004 Lighted Warning Signs 115 VAC 250 mA 50 60 Hz 79005 Lighted Warning Signs 230 VAC 125 mA 50 60 Hz ELECTRICAL HAZARDS Make all connections to or from the power supply with the power off There may be up to 200 volts present at the output terminals this could be dangerous if care is not exercised when the power supply is on Do not use the power supply without its cover in place Lethal voltages are present inside FIRE HAZARDS Arc lamps are extremely hot during operation and for several minutes after being shu
11. djustment repair or replacement Newport shall provide the labor required during the above listed warranty period to adjust repair and or replace the defective goods at no cost to the buyer ONLY IF the defective goods are returned freight prepaid to a Newport designated facility If goods are not returned to Newport and the user chooses to have repairs made at their premises Newport shall provide labor for field adjustment repair and or replacement at prevailing rates for field service on a portal to portal basis Newport shall be relieved of all obligations and liability under this warranty of 1 The user operates the device with any accessory equipment or part not specifically approved or manufactured or specified by Newport unless buyer furnishes reasonable evidence that such installations were not the cause of the defect This provision shall not apply to any accessory equipment or part which does not affect the safe operation of the device 2 The goods are not operated or maintained in accordance with Newport s instructions specifications 3 The goods have been repaired altered or modified by other than authorized Newport personnel 4 Buyer does not return the defective goods freight prepaid to a Newport facility within the applicable warranty period ITIS EXPRESSLY AGREED THAT THIS WARRANTY SHALL REPLACE ALL WARRANTIES OF FITNESS AND MERCHANTABILITY BUYER HEREBY WAIVES ALL OTHER WARRANTIES GUARANTEES
12. et Limit Table 5a STB 1 ESRXX 1 See ESR command for Hex value definition Set Power limit if power preset gt limit Preset Limit MODE X 1 ESRXX 1 See ESR command for Hex value definition Set desired mode if lamp is off use last settings of that mode Else return ESR error bit A PRESET XX X1 With power supply in current mode lamp ON or OFF sets current to PRESET value if current limit A LIM else returns ESR error bit 5 Command Error P LIM XXXX 1 ESRXX 1 See ESR command for Hex value definition With power supply in power mode lamp ON or OFF sets current to PRESET value if power limit P LIM else returns ESR error bit 5 Command Error 17 M 69907 ARC LAMP POWER SUPPLY APPLICATIONS IV 1 REMOTE CONNECTOR The signals which are available at the remote connector allow you to monitor the current voltage and watts output of the supply from a remote location via a meter or an A D converter and a computer There is also a remote start stop input so that the lamp can be started stopped with a simple momentary contact closure at a remote location A control input is also included at the remote connector which is intended for use with the Newport Model 68950 Intensity Controller When the 68950 is connected a sample of the light at all or selected wavelengths is compared to a reference Any difference between the two is sent into the power supply to co
13. following Directives the EMC Directive the Low Voltage Directive and accordingly carries the CE mark Oriel Instruments 150 Long Beach Boulevard Stratford CT 06497 USA Arc Lamps Power Supply 69907 69910 69911 69920 69922 69907 xxxx 6991 10 xxxx 6991 1 xxxx 69920 xxxx 69921 xxxx EN 61010 1 1993 A2 IEC 1010 1 1990 A1 A2 EN 50081 1 1992 EN 55022 1994 EN 55011 1993 Class B EN 50082 1 1992 IEC 801 2 1991 IEC 1000 4 2 1995 ENV 50140 1993 EC 1000 4 3 1995 EN 61000 4 3 1995 IEC 801 4 1988 IEC 1000 4 4 1995 89 336 EEC 73 23 EEC rts D a dd K Ge yos Ga Signature George Buzel Name Director of Engineering Title 20 VII WARRANTY AND RETURNS M 69907 ARC LAMP POWER SUPPLY Newport warrants that all goods described in this manual except consumables such as lamps bulbs filters ellipses etc shall be free from defects in material and workmanship Such defects become apparent within the following period 1 All products described here except spare parts one 1 year or 3000 hours of operation whichever comes first after delivery of the goods to the buyer 2 Spare parts ninety 90 days after delivery of goods to the buyer Newport s liability under this warranty is limited to the adjustment repair and or replacement of the defective part s During the above listed warranty period Newport shall provide all materials to accomplish the repaired a
14. horization number RMA Having an RMA will shorten the time required for repair because it ensures that your equipment will be properly processed Write the RMA on the returned equipment s box Equipment returned without a RMA may be rejected by the Newport Receiving Department Equipment returned under warranty will be returned with no charge for the repair or shipping Newport will notify you of any repairs not covered by the warranty with the cost of the repair before starting the work Please return equipment in the original or equivalent packaging You will be responsible for damage incurred from inadequate packaging if the original packaging is not used Include the cables connector caps and antistatic materials sent and or used with the equipment so that Newport can verify correct operation of these accessories
15. i 1 Connection to lamp negative terminal 2 jarra Connection to lamp positive terminal EN EEE MATING CONNECTOR e Body ITT DAM 7W2P K87 includes pins 1 5 e Pins ITT DM 53745 1 requires 2 per connector A1 and A2 e Backshell Standard 15 pin D SUB M 69907 ARC LAMP POWER SUPPLY INTERLOCK The 69907 has a safety interlock feature which must be satisfied before the power supply output will activate and which if broken during operation will disable the power supply When connected to a Newport Research housing cables must be attached the housing must be closed properly and any over temperature sensor satisfied for the power supply to drive the lamp The cable used for the other applications includes 2 pairs of interlock wires in each cable which are shorted together to represent a satisfied interlock condition If an interlock is desired with these configurations separate the brown and blue interlock wires at the lamp element end of the cable and tie them into the interlock system A contact closure is required to satisfy the interlock M 69907 ARC LAMP POWER SUPPLY REMOTE CONNECTOR Access to the internal metering and control signals is provided through this connector C Figure 1 It is a High Density 15 pin D SUB connector with the following pin assignments Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9 Pin 10 Pin 11 Pin 12 Pin 13 Pin 14 Pin 15 External co
16. l allow changes to this digit Pressing the left or right arrow button will flash the digit to the right or left of presently blinking digit and allow changing that digit by using the up or down arrows Pressing SET ENTER will lock this value in as the new preset NOTE If the LAMP ON indicator is on any change to preset is immediate to the output during adjustment without pressing SET ENTER This allows fine tuning of output light intensity The factory default preset for current is 7 5A and the factory default preset for power is 150W Factory Restore The power supply can have all parameters restored to factory setting by applying AC power while holding down the LAMP OFF button This will set the Power limit to 300 Watts Current limit to 12 Amps mode to power and preset to 150Watts 11 69907 LAMP POWER SUPPLY SETUP MODE Depressing the SETUP button G in Figure 4 when the LAMP ON indicator is off enters setup mode If LAMP ON indicator is illuminated the SETUP button is inactive Once in setup mode there are four main items that may be set up SET MODE SET LIMIT MEMORY and HRS RESET Pressing SETUP at any time will exit the setup mode Newport recommends scrolling through the setup parameters a second time before exiting setup to verify all parameters CAUTION Be sure to press SET ENTER after each adjustment or the selection will not be entered and system will revert to previous setting SET
17. mp housing failure 10 M 69907 ARC LAMP POWER SUPPLY EXT Illuminates when the remote input is active typically when the NEWPORT 68950 Intensity Controller is connected COMM Illuminates when the power supply front panel is locked out via RS232 command Control is through RS232 only The SETUP AREA is only active in setup mode Each of the four indicators flashes while advancing through the parameters See section III 3 for a detailed description of setup mode LAMP START This button C in Figure 4 starts the lamp at the preset power or current level provided there are no fault conditions LAMP OFF If the lamp is on pressing this button D in Figure 4 removes power to the lamp The housing will still be powered allowing cooling fans to continue running until the lamp has reached a safe handling temperature DISPLAY SELECT Each time you press and release this button E in Figure 4 the digital meter and associated units indicators switch between one of four functions current AMPS voltage VOLTS power WATTS and lamp operating time LAMP HOURS SET ENTER Depressing the SET ENTER button F in Figure 4 at any time displays the preset power or current level Holding the SET ENTER button for seconds allows the preset value to be changed to any value within valid limits upper limit set in setup mode The display will show the present value with one digit blinking Pressing the up or down arrow wil
18. mpensate for this change The result will be improved stability in light output over time When the control input is active the front panel EXT LED will illuminate At this time the preset value of current power becomes a maximum setting and is otherwise overridden by the control input The LAMP ON LAMP OFF SET ENTER and DISPLAY SELECT functions are always active 18 M 69907 ARC LAMP POWER SUPPLY V SPECIFICATIONS Output Ratings Power Current Voltage open circuit Voltage loaded Operating Modes Voltage ripple into ohmic Load load power Voltage ripple true RMS Light ripple true RMS Meter accuracy full scale Digital meter resolution Voltage Power Current Safety interlock voltage Line Regulation Input Ratings Voltage Current max Power Factor Frequency Circuit Breaker Ambient operating Temperature Weight 40 300 W 1 5 12A 200 VDC 0 70 VDC Constant power Constant current 3 4 200 W lt 0 1 lt 0 5 lt 0 05 0 1 VDC 1W 0 1A 12 VDC GND 0196 95 264 VAC 4 5A gt 0 99 47 63 Hz 5 A 2 POLE 10 40 20 Ibs 9 kg 19 VI M 69907 ARC LAMP POWER SUPPLY DECLARATION OF CONFORMITY DECLARATION OF CONFORMITY Manufacturer s name Manufacturer s address declares that the product Product Name Model Number conforms to the following Product Specifications Safety EMC complies with the
19. n on housing operational This indicates over temperature in the lamp housing Ensure no blockage of cooling air on housing Fault light e Display shows S This message is displayed usually when trying to ignite and ignition fails It is a result of no open circuit voltage or open circuit voltage too low Cycle power If fault repeats contact Newport for RMA information Fault light e Display shows L P This message is displayed usually when trying to ignite and ignition fails It is a result of no current flow from supply after 5 second of trying to ignite the lamp e Ticking sound heard from top of lamp housing during ignition cycle approximately once per second e Check lamp connection and polarity lamp installed upside down or with reversed polarity should be removed and replaced immediately e Cycle power and try ignition again e Check hours on lamp It may be at end of life and typically will be difficult to start e Try anew lamp e f fault repeats contact Newport for RMA information Lamp doesn t start upon press of start button If 68950 is being used ensure it is turned off until lamp is up to operating point 15 RS 232 COMMUNICATIONS The following table list the commands used for communications with the power supply Please see section 111 1 for cable information The baud rate is 9600 with 8 data bits 1 stop bit and no parity Table 5 M 69907 ARC LAMP POWER SUPPL
20. nd adjust the current limit to a value 10 higher then the desired operating point Press the LAMP START button The igniter will fire for approximately 5 seconds or until the lamp starts at the preset level Once the preset level is reached the LAMP ON indicator will illuminate You can use DISPLAY SELECT to display current voltage watts or lamp operating time To shut off the lamp press the LAMP OFF button Once the output reaches 0 the LAMP ON indicator will extinguish Should the lamp fail to ignite refer to the troubleshooting section 111 6 NOTE If 68950 is being used the 68950 must be off until lamp is stabilized by power supply Once control is shifted to 68950 all front panel controls become inactive 13 M 69907 ARC LAMP POWER SUPPLY POWER MODE Turn on the power supply The POWER MODE LED should illuminate If not switch to POWER MODE see section III 3 Set the display to read power WATTS Press and hold the SET ENTER button in until the display reads the preset watts with the least significant digit blinking Using the up down arrows you can change the blinking digit use the left right arrows to select another digit Once the display is set for the desired wattage press the SET ENTER button The display will now show WATTS You can push the SET ENTER button at any time to view the preset watts NOTE If you cannot set the desired preset watts press SET ENTER and then enter setup per
21. ntrol input A 2 5 5 volt DC signal will decrease the output approximately 2096 maximum 5 Vdc represents maximum turn down Not used Input control common Not used Remote start common Not used Remote start stop Momentary contact with remote start common will start lamp if lamp is off When lamp is on this action will stop the lamp Not used Remote meter output Power 0 2 0V indicates 0 300W Remote meter output Current 0 2 0V indicates 0 12A Remote meter output Voltage 0 2 5 indicates 0 100V Not used Not used Not used Not used Figure 2 Remote Connector 15 Pin D SUB Assignment M 69907 ARC LAMP POWER SUPPLY RS 232 CONNECTOR Access to all of the power supply operating functions can be controlled via a RS 232 communications link to a PC D Figuref It is a 9 pin D SUB connector with the following pin assignments Pin 1 Not used Pin 2 TX Pin 3 RX Pin 4 Not used Pin 5 GND Pin 6 Not used Pin 7 Not used Pin 8 Not used Pin 9 Not used Figure 3 RS 232 Connector 9 Pin D SUB Assignment The cable for the RS 232 connection can be ordered from Newport as follows Computer Power Supply Newport Part Number Female DB9 Male DB9 10 60 018 Female DB25 Male DB9 107695019 Please contact your Newport sales representative to order the cable M 69907 ARC LAMP POWER SUPPLY 11 2 FRONT PANEL CONTROLS AND DISPLAYS OU Thermo Oriel 69935 POWER
22. re to UV light when used with a NEWPORT lamp housing Remote I O connector on the rear panel of the 69907 provides remote metering capability and direct connection to the NEWPORT 68950 Light Intensity Controller The 68950 is typically used when a high level of long term stability is required RS 232 Communication allows remote operation and monitoring of the power supply Optional IEEE Communication allows remote operation and monitoring of the power supply DIMENSIONS M 69907 ARC LAMP POWER SUPPLY 1 3 ORDERING INFORMATION Arc Lamp Power Typical Typical Average Lamp Lamp Lamp Description Range Voltage current Life Hours Housing socket Watts VDC ADC Adapter MERCURY LAMPS 50W 500W 6281 100W Hg 80 110 Research 66150 Housing 6283 200W Hg 160 220 1000 66144 XENON LAMPS 75W Xe Ozone Free 100W Xe Ozone Free 50W 500W Research 150W Xe Housing 1 150W Xe UV 150W Xe 120 165 Ozone Free 150W Xe Compact MERCURY XENON LAMPS 200W HgXe 50W 500W 200W HgXe 160 220 Research 1 Ozone Free Housing Table 1 Arc lamps for Model 69907 300W power supply 1 Models 66901 66905 66912 67001 67003 67005 67006 The difference between these housings is the condenser lens Once you have established which condenser is most suitable for your applications contact a Newport sales representative engineer Please refer to the solar simulator manual for lamp adaptors and operation con
23. t off Keep flammable objects away from the lamp and lamp housing Newport Research Fan Cooled Housings are equipped with a condenser lens The re focused output of this lens can cause ignition of flammable targets ex wooden walls certain chemicals KEEPING THE LAMP IN GOOD CONDITION Never touch the lamp envelope or element with uncovered fingers even during installation or its lifetime and performance can be negatively affected Do not run the lamp more than 1096 above its current or power rating Source lifetime will decrease dramatically at higher operating point M 69907 ARC LAMP POWER SUPPLY I USING THE POWER SUPPLY Figure 1 IL REAR PANEL CONNECTIONS AC MAINS CONNECTION NOTE The Recommended Line Voltage For the 69907 is 95 264 VAC 50 60 Hz Verify that the front panel power switch is in the off position and then connect the provided AC cable between the IEC style socket A in Figure 1 on the rear panel and the wall outlet M 69907 ARC LAMP POWER SUPPLY LAMP CONNECTIONS With the power supply off connect the cables from the Newport lamp housing igniter to the OUTPUT connector B in Figure 1 which provides all the signals necessary to interface this power supply to Newport lamp housing and igniter Table 3 gives signal information at the rear terminals of the power supply Please note that the 70051 cable has A1 and A2 pins reversed in the cable assembly Table 3 pew GER a
24. urs to 0 value definition MODE XA ESRXX See ESR command for Hex Set desired mode if lamp is off use See table 5A for ver 7 value definition last settings of that mode Else return and earlier ESR error bit 1 See ESR command for Hex Lockout unlock front panel keys X 1 for panel lockout value definition X 0 for panel unlock SAVE 2X I ESRXX See ESR command for Hex Save operation parameters to location X Memory location 1 5 value definition specified RECALL X ESRXX See ESR command for Hex If lamp off set operation parameters to X Memory location 1 5 value definition those of memory location Else return ESR error bit A PRESETZXX X I ESRXX 1 See ESR command for Hex With power supply in current See table 5A for ver 7 value definition mode lamp ON or OFF sets and earlier current to PRESET value if current limit A LIM else returns ESR error bit 5 Command Error P PRESET XXXX 1 ESRXX 1 See ESR command for Hex With power supply in power mode See table 5A for ver 7 value definition lamp ON or OFF sets current to and earlier PRESET value if power limit P LIM else returns ESR error bit 5 Command Error A LIMZXX X 1 ESRXX See ESR command for Hex Set Current limit if current preset gt value definition limit Preset Limit P LIM XXXX 1 ESRXX 1 See ESR command for Hex Set Power limit if power preset gt limit value definition Pres
Download Pdf Manuals
Related Search
Related Contents
スパフレックス カップリ ング AL/CLモデル MANUAL DE INSTRUÇÕES ML67Q5260 Preliminary ICC ICCMSCMA62 1.化学物質等及び会社情報 2.危険有害性の要約 Betriebsanleitung A-040 Eglo FORNES Harman-Kardon AVR 5000 Stereo Receiver User Manual 12月号 - 朝日村 Yamaha NS M325 Owner's Manual Copyright © All rights reserved.
Failed to retrieve file