pump trouBLeshootinG procedures

image

Contents

1. daintreescientific com au E mail OTHER SYMPTOMS If your symptoms differ from these you may have multiple problems and should use the Modified Ramp Test This test is also useful if you want to verify your troubleshooting from the above steps or you are unsure of your diagnosis MODIFIED RAMP TEST Perform steps 1 5 from the standard Ramp Test 1 Disconnect the outlet of the pump from the rest of your HPLC system 2 Connect a short piece of tubing to the pump outlet and put the other end of this tube into a beaker 3 Prime the pump thoroughly with methanol Make sure you have drawn enough methanol through the solvent inlet line to ensure the pump is pumping methanol 4 Remove the tubing at the pump outlet and install a plug fitting to stop flow 5 Set the pressure limit for 4000 5000 psi 6 Disconnect the tube between the outlet check valve and the reference valve or tee fitting at the reference valve or tee on the side of the pump not being tested if you are testing the left side disconnect the right side 7 Put a plug fitting into the reference valve or tee where you removed the tube You may want to put a towel under the open end of the disconnected tube it will drip a little during the following steps Do not cap the dripping tube 8 Set the flow to 1 ml min and observe the psi meter and the indicator rod on the side being tested SYMPTOM 1 The pressure increases steadily while the plunger is c
2. PUMP TROUBLESHOOTING PROCEDURES You can troubleshoot most pumping problems in a short time by observing the pressure meter on your pump The following procedures will show you how RAMP TEST This is the first test to try as it involves the least amount of work and disassembly and it will find most pumping problems 1 Disconnect the outlet of the pump from the rest of your HPLC system 2 Connect a short piece of tubing to the pump outlet and put the other end of this tube into a beaker 3 Prime the pump thoroughly with methanol Make sure you have drawn enough methanol through the solvent inlet line to ensure the pump is pumping methanol 4 Remove the tubing at the pump outlet and install a plug fitting to stop flow 5 Set the pressure limit for 4000 5000 psi 6 Set the flow rate for 0 3 1 0 ml min and observe the psi meter and the indicator rods With a properly operating pump you will see a smooth increase in pressure until the pressure limit is reached Below are two examples of the most common symptoms of Ramp Test failure and how to resolve them The examples use the left and right side to describe the symptoms interchange the left and right in the text if needed to match your symptoms SYMPTOM 1 While watching the left indicator rod coming out the pressure increases When the right indicator rod comes out the pressure stops increasing and stays about the same When the left indicator rod again begins coming ou
3. oming out and holds pressure while the plunger is retracting Analysis This pump head is pumping OK SYMPTOM 2 The pressure never increases above 0 psi Analysis The inlet check valve is not working Reprime this pump head and try again If this fails see Symptom 1 of the Ramp Test to repair SYMPTOM 3 The pressure increases steadily while the plunger is coming out but drops abruptly while the plunger is retracting Analysis The outlet check valve is leaking back See symptom 2 of the Ramp Test to repair SYMPTOM 4 The pressure increases steadily during the first plunger out stroke but increases less during the next out strokes the pressure holds while the plunger is retracting solution The plunger seal is beginning to leak or the inlet check valve is leaking back slightly To repair look at the weep hole at the bottom of the support bushing or between the pump head and the support bushing and see if the mobile phase has been leaking past the plunger seal If so slightly tighten the two allen head screws holding the pump head and retest If the pump head is tight against the support bushing it is time to replace the plunger seal If the inlet check valve is leaking back see symptom 1 of the Ramp Test to repair OTHER SYMPTOMS If you need any further assistance please feel free to call 952 895 8292 or toll free at 800 682 3295 during the hours of 8 a m to 5 p m CST Thanks for calling Fax 03 6376 3334 ww
4. t the pressure again begins increasing ANALYSIS The right inlet check valve is not working This symptom is fairly easy to understand The inlet check valve is either stuck open and the mobile phase is sloshing back and forth as the plunger goes in and out or it is stuck shut never allowing mobile phase into the pump head Possible Problem Solution Check valve dirty Clean and rebuild Seat worn out Clean and rebuild SYMPTOM 2 While watching the left indicator rod coming out the pressure increases When the right indicator rod comes out the pressure drops almost back to zero When the left indicator rod again begins coming out the pressure again increases ANALYSIS The left outlet check valve is leaking back This symptom is more difficult to understand until you think about where the mobile phase is going When the left plunger comes out the pressure increases When the right plunger comes out it pumps mobile phase out into the reference valve or tee fitting and over to the left side The left outlet check valve is leaking backwards allowing what the right side is pumping to enter and fill the left side as the left plunger is going in Possble Problem Solution Clean in ultrasonic cleaner Flush thoroughly with methanol in priming syringe Disassemble and rebuild check valve Check valve dirty Air bubble trapped in check valve Reprime with methanol DAINTREE SCIENTIFIC Tel 03 6376 3335 info
5. w daintreescientific com au

Download Pdf Manuals

image

Related Search

Related Contents

Mise en page 1 - Hôpital Riviera  User's Manual of  Technical Notes - Matan Digital Printers  DeWalt DW321 Instruction Manual  ShelterLogic 90391 Instructions / Assembly  BULLETINDE PROGRAMMES  ATLAS Muon Drift Tube Electronics - BMC Page  S - IFSC - Universidade de São Paulo  - cardi.co  Kase StatWare Manual  

Copyright © All rights reserved.
Failed to retrieve file