Home

Field Rotameters 320 Series Operating Instructions 37505 PDF

image

Contents

1. Operating Instructions 863 Valley View Road Eighty Four PA 15330 USA Tel 724 941 9701 Fax 724 941 1369 e mail skctech skcinc com Field Rotameters This rotameter is packed to protect against damage during shipment If damage occurs notify the carrier immediately Also check that you have received the correct model and flow range required for your application Calibration SKC rotameters are secondary standards and should be calibrated by the user with a primary standard calibrator at regular intervals Since rotameters are affected by temperature and atmospheric pressure rotameter calibration should be performed under the conditions of use or a mathematical correction should be used Operation Operate the rotameter in a vertical position to assure accuracy Use it in a manner that minimizes vibration and flow pulsation both conditions may affect the rotameter Flow rate is indicated by the point on the printed scale at which the center of the ball float comes to rest The field rotameter inlet must always be at atmospheric pressure i e exposed to the atmosphere The rotameter is the last device connected to the sampling train Connect the inlet of the sampling pump to the outlet of the sampling medium Connect the inlet of the sampling medium to the outlet of the rotameter CAUTION This rotameter is designed for use with non hazardous gases at pressures up to 100 psig and temperatures up to 65 C 150 F Do not use
2. hazardous gases or liquids and do not exceed temperature or pressure limits Use with hazardous fluids or beyond pressure and temperature limits may cause failure which could result in personal injury Maintenance Occasionally clean the rotameter if dirt appears in it or if float movement is restricted To clean 1 Remove the top plug float stop and ball float 2 Wash the tapered hole ball float float stop and top plug with a mild liquid detergent and a soft brush Rinse all parts with clean water and dry thoroughly Avoid using solvents 3 Reinstall the ball float float stop and top plug Use a small amount of halocarbon grease or any other compatible O ring lubricant to ease installation of the top plug and prolong O ring life Specifications Meas Range Scale Accuracy Subdivisions Cat No L min inch L min 0 05 0 5 2 5 0 05 320 2A05 0 1 1 0 3 20 320 4A1 0 4 5 0 3 0 2 320 4A5 4 20 3 0 5 320 4A20L 4 50 3 1 0 320 440 5 30 3 1 0 320 530 3 30 3 320 100T Full Scale Fittings adapted for use with HV30 Sample Pump t Fittings adapted for use with QuickTake Sample Pumps Float Black glass or stainless steel ball Body Clear acrylic Seals Buna N O ring Pressure 100 psig maximum Temperature 150 F 65 C maximum Fittings Brass with hose connector Tygon tubing and Luer taper adapter Form 37505 Rev 0907

Download Pdf Manuals

image

Related Search

Related Contents

LG LMV1814SB Specification Sheet  "user manual"  エクト - 厚生労働省  Bedienungsanleitung Digitaler Satelliten Receiver - AHG  Manual del usuario PL260d-PL280d  Manuel d`instructions presse ICLAM  Tripp Lite Dual Full-Motion Desk Mount for 13" to 26" Flat-Screen Displays  

Copyright © All rights reserved.
Failed to retrieve file