Home

User's Manual

image

Contents

1. Meter Display What It Means What To Do Important Testing Notes Severe high blood glucose Repeat the test with a new test strip If the emos Use the meter within the test strip operating range Appears when result is or result is HI contact your healthcare professional S e Use the test strip immediately after removing it from the foil packet Higher than There may be a problem immediately E g Expiry date Use each test strip only once D7 me with the test strip Check blood B Ketone if checking ketones is part March 31 2012 Higher than of your diabetes management programme Do not use expired test strips they may produce inaccurate results 500 mg dL The expiry date is the last day of the month if only the month and year are shown Do not put urine on the test strip Do not use a wet bent scratched or damaged test strip Do not use the test strip if its foil packet has a puncture or tear in it Note If you see the error messages E 3 or E 4 consult the Error Messages section in this user s manual IMPORTANT Contact your healthcare professional if you have Note See test strip instructions for use for operating range and more information on symptoms that do not match your test result and you have followed limitations of use the instructions in this user s manual Check your ketones When you have an illness When your blood glucose is above 16 7 mmol L 300 mg dL When you are experiencing
2. os ni ph I Ir ERE E k YES 6 Testing Your Blood 7 Important Testing Notes ss oe tannini n sored ce 9000 7 Ir erc crema 8 Alternative Sites Sa ener nnn neyene ba e Ra WIRE R Da KAN D UN NE T REAT 9 Performing a Blood Glucose Test y y W yena nea DET ak n 10 Understanding Blood Glucose Test lt 13 Low and High Blood Glucose 13 Testing Your Blood 16 Important Testing lNOtes rh a a biadan s k 53 FRE ES RE cash te 16 Se oni en ASA 17 Performing a Blood B Ketone 17 Understanding Blood B Ketone Test 21 High Blood B Ketone Results 4 21 MediSense or Optium Glucose and Ketone Control Solutions 23 Important Control Solution Testing Notes 24 Performing a Control Solution 25 Managing Your Results 29 Viewing Test Results in Memory heh ph Oh eR q adana kudi 29 Viewing Blood Glucose
3. 31 Transferring Results to a 33 Error Messages iussisse ko n bay e RSEN RO GEAR IURE RU d s 33 E 1 to E 9 What It Means What To Do 33 Troubleshooting nee i USER REA ead 36 Maintaining the Meter 000 39 Replacirigtlie Battety sd sa K cenan mnn a p pure bo beu dede E PE UE dp 39 Cl aning the Meter oed da ripe ume prs pg iex am 41 Storing the Meter see bka A HR RU MEE 42 Specifications oe Uber e EE w r rues 42 References 4 s 3k50 340054 8340A 0 ki R Kin da h 4 W W n kn e rae 45 3 813 11 39 AM Intended Use IMPORTANT See test strip instructions for use for more information about sample 77 2 types Use outside the body only in vitro diagnostic use for self testing or professional use in the yp management of diabetes Potential infection risk Healthcare professionals performing blood tests with this system on multiple patients must always wear gloves and should follow the infection Use for measuring blood glucose from samples taken from the forearm upper arm base of the controllpolicies and procedures approved by facility thumb or fingers Test blood glucose in fresh whole blood samples Read the instructions in this user s manual Failure to follow instructions will cause incorrect results Practise the testing procedures before using the meter Follow yo
4. Repeat the previous testing steps 3 If the error reappears contact Customer Service 33 34 ART20287 Rev Bindd 33 34 e 8 8 13 11 39 AM Message What It Means E B Meter error E 1 No coding required or Test strip may be damaged used or the meter does not recognise it F f Meter error E 35 ART20287_Rev B indd 35 36 WhatTo Do Checkthat you are using the correct strip for this meter See test strip instructions for use to verify your strip is compatible with this meter Repeat the test using a test strip for use with your meter If the error reappears contact Customer Service Checkthat you are using the correct test strip for this meter See test strip instructions for use to verify your strip is compatible with this meter Repeat the test using a test strip for use with your meter If the error reappears contact Customer Service Turn off the meter Repeat the previous testing steps If the error reappears contact Customer Service e What It Means The meter Test strip is not inserted doesnotenter correctly or fully into the test mode meter after inserting a test strip No battery is installed Battery is installed incorrectly Dead battery Troubleshooting WhatTo Do With the contact bars 3 black lines facing up insert the test strip into the meter until it stops This turns on the meter If the meter still does not enter test mode contact
5. You must set the time and 10 1 The day flashes Press or P until the correct day appears Press to save the date to review averages Month Day day and to move to set time format Year Measurement Units Time Format mmol L The unit of measurement set by the factory appears You cannot change this mg dL setting Contact Customer Service for more information RE D Note lf using the 12 hour time format AM and PM appear Press to return to set beeper or press and hold 1 to turn off the meter 24 mir sr end The time format flashes Press or P to change the time format Press to save the time format and to move to set date format 5 6 ART20287_Rev B indd 5 6 e 8 8 13 11 39 AM Testing Your Blood Glucose pmi c ug ExP 2012103 Do not use expired test strips they may produce inaccurate results The expiry date is the last day of the month if only the month and year E g Expiry date are shown March 31 2012 Important Testing Notes Do not use a wet bent scratched or damaged test strip LOTI 9 This display always appears when the meter is turned on CODE Do not use the test strip if its foil packet has a puncture or tear in it i You should check that your meter matches the example you mg dL G 3 mmol L QE see here exactly every time your meter turns on Note See test strip instructions for use for operating range and more information 6 JJ Do not use the meter
6. of inactivity 32 3 813 11 39 AM Transferring Results to a Computer Message What It Means What To Do Transferring test results to a computer requires a compatible data management system For more The blood glucose level may 1 Review the testing instructions information please go to our website or contact Customer Service ETI be too low to be read by the 2 Repeat the test using a new HAR SUD system 3 If the error reappears contact your healthcare Error Messages or professional immediately There may be a problem with the test strip Message What It Means What To Do E The temperature is too hot 1 Move the meter and test strips to a location or too cold for the meter to where the temperature is within the test strip work correctly operating range See test strip instructions for The blood glucose level may 1 Repeat the test using a new test strip use for the appropriate range be too high to be read by the 2 If the error reappears contact your healthcare 2 Wait for the meter and test strips to adjust to system professional immediately the new temperature e 3 Repeat the test using a new test strip 4 If the error reappears contact Customer Service There may be a problem with the test strip 5 Blood was applied thetest 1 Review the testing instructions strip too soon 2 Repeat the test using a new test strip 0 MERE GIU d Tem ef The meter 3 If the error reappears contact Customer Service 2
7. press and hold example Contact Customer Service KI LI See Testing Your Blood Glucose section for more information A 56 58 88 98 Appear on the display when there are no previous results to view lt lt gt gt Press gt to return to a previously reviewed result Repeat this for each result you wish to review Note To exit meter memory at any time press and hold 29 30 ART20287_Rev B indd 29 30 e 8 8 13 11 39 AM Viewing Blood Glucose Averages IMPORTANT Set the date and time before testing to be able to review averages e Averages do not include glucose and B Ketone control solution results blood B Ketone results and results that do not show time and date Control solution results not marked as control solution tests may cause averages to be inaccurate e appear on the display when there no current averages to view e L blood glucose test results appear in averages as 1 1 mmol L 20 mg dL e H blood glucose test results appear in averages as 27 8 mmol L 500 mg dL 31 ART20287 Rev B indd 31 32 While in memory press Q Your 7 day average appears Note n number of results included in your 7 14 or 30 day average Press to view the 14 day average Press to view the 30 day average Press to return to the 7 day average or Press 1 to return to the memory screen Press and hold CD to turn off the meter The meter turns off after 30 seconds
8. a sample to the test strip Obtain a Blood Sample Select a test site Use the lancing device to obtain a blood sample See lancing device instructions for use for more information Apply Blood to the Test Strip Bring the blood drop to the white area at the end of the test strip The blood is drawn into the test strip IMPORTANT If the countdown does not start you may not have applied enough blood to the test strip See test strip instructions for use for re application instructions If the countdown still does not start remove the used strip and discard it correctly Start a new test with a new test strip mmol L ALI 949 10 10 Hold the blood drop the white area until the meter beeps if sound is on and the status bar appears on the display This indicates the test strip has obtained enough blood The countdown appears on the display as the meter checks your glucose level Note Do not remove the test strip from the meter or disturb the test strip during the countdown View the Result The meter beeps if sound is on when the result appears on the display The test is complete example shown The result is stored in memory or you can write it in a logbook Turn Off the Meter Use the opened foil packet to remove the used test strip This turns off the meter Discard the used test strip correctly Note You can also press and hold Q for at least 2 seconds to turn off the meter The meter also turns
9. amp x appears you may still use the meter and the results will be accurate The display light will not work Replace the battery immediately When amp x appears alone the meter is not usable The meter turns off automatically Replace the battery immediately 1 Gently push the battery cover in and up to remove it If the meter includes a plastic tab pull on the tab to remove the battery If the meter does not include a plastic tab push the battery upward to remove it Insert a new CR 2032 lithium coin cell battery 4 facing up Slide the battery cover into place until it clicks Note Dispose of used batteries correctly see that the battery was installed correctly If meter turns on the meter may prompt you to reset the time and date See Setting Up the Meter in this user s manual Note Test results will be not be lost even if the meter loses its time and date settings Press and hold to turn the meter on If the meter does not turn on check to 40 3 8 13 11 39 AM Cleaning the Meter Avoid getting dirt dust blood control solution or liquid in the meter test strip port Clean the outside of the meter using a damp cloth and mild soap Healthcare Professionals Acceptable cleaning solutions include 7096 isopropyl alcohol or A mixture of 1 part ammonia 9 parts water or A mixture of 1 part household bleach 9 parts water IMPORTANT Do not try to clean the test st
10. if the display check screen does not O ran O OL use exactly match the example j lt lt gt Il au If the screen does not match the meter may show an incorrect JESUS Pisasa contaci Cusiomer Seisa Test Site IMPORTANT Wash your hands and the test site with warm soapy A Ag AM H B H H water to ensure accurate results Thoroughly dry your hands and the F test site To warm the site apply a warm dry pad or rub vigorously for a few seconds Note Avoid areas near bones areas with lots of hair and squeezing the puncture site Bruising may occur at alternate sites but will disappear quickly If bruising occurs consider selecting another site mo Use the meter within the test strip operating range C Grexe xum _ ONT Finger tip testing is different to alternative site testing Adjust the lancing device accordingly for testing site See lancing device instructions for use for more E g Expiry date Use each test strip only once Teyan i March 31 2012 ART20287 Rev Bindd 7 8 e 8 8 13 11 39 AM Alternative Sites Performing a Blood Glucose Test Obtaining a blood sample from alternative sites forearm upper arm and base of the thumb is subject to limitation Contact your healthcare professional before beginning alternative site testing SS Open the foil test strip packet at the notch and tear down to remove the ull
11. off after 60 seconds of inactivity 12 3 8 13 11 39 AM Understanding Blood Glucose Test Results Low and High Blood Glucose Results The meter displays results in mmol L or mg dL The unit of measurement is preset You cannot change this setting IMPORTANT The meter displays results from 1 1 27 8 mmol L 20 500 mg dL Low or high blood glucose results can indicate a potentially serious medical condition General Result What It Means What To Do Lower than 2 8 mmol L Lower than 50 mg dL Low hypoglycaemic Repeat the test with a new test strip even if you do not feel that you have low blood glucose Higher than 16 7 mmol L Higher than 300 mg dL High hyperglycaemic Repeat the test with a new test strip even if you do not feel that you have high blood glucose 13 ART20287 Rev B indd 13 14 e Meter Display in LU Appears when result is Lower than 1 1 mmol L Lower than 20 mg dL Appears with result which is Higher than or equal to 16 7 mmol L Higher than or equal to 300 mg dL What It Means Severe low blood glucose or There may be a problem with the test strip Blood glucose level is high What To Do Repeat the test with a new test strip If the result is LO contact your healthcare professional immediately Check blood B Ketone if checking ketones is part of your diabetes management programme 14 3 8 13 11 39 AM e Testing Your Blood B Ketone
12. test strip Alternative site results may differ from fingertip results when blood glucose levels change rapidly e g after eating a meal or taking insulin or during or after exercise Use alternative sites before or more than two hours after you eat a meal take insulin or exercise Insert the Test Strip With the contact bars 3 black lines facing up insert the test strip into the meter A until it stops This turns on the meter Do not use blood samples from alternative sites when Note The meter turns off after 3 minutes of inactivity Remove and reinsert the unused You think your blood glucose is low or changing rapidly j test strip to restart the meter e You have been diagnosed with hypoglycaemic unawareness Alternative site results do not match the way you feel You are within two hours of eating a meal taking insulin or exercising mU This display always appears when the meter is turned on mg dL EE OD mmol L N Forearm Do not use the meter if the display check screen does not exactly match the FI example Contact Customer Service KI Hb See Testing Your Blood Glucose section for more information 68 6gw88 98 10 30 10 5 Time and date if set appear next 9 10 ART20287_Rev B indd 9 10 e 8 8 13 11 39 AM qe n 11 ART20287_Rev B indd 11 12 The Apply Sample symbols 6 and Fl appear next indicating the meter is ready for you to apply
13. therapy on glucose metabolism in diabetic ketoacidosis Diabetes 1988 37 1470 1477 6 Hale PJ Crase J Nattrass M Metabolic effects of bicarbonate in the treatment of diabetic ketoacidosis Br Med J 1984 289 1035 1038 7 Directive 1999 45 EC of the European Parliament and of the Council of 31 May 1999 concerning the approximation of the laws regulations and administrative provisions of the Member States relating to the classification packaging and labelling of dangerous preparations 45 ART20287 Rev B indd 45 46 3 8 13 11 39 AM ART20287_Rev B indd 47 48 e 8 8 13 11 39 AM ART20287_Rev B indd 49 50 e 8 8 13 11 39 AM ART20287 Rev B indd 51 Abbott Diabetes Care 666 Doncaster Road Doncaster 3108 Victoria Australia ABN 95000 180 389 1800 801 478 Abbott Laboratories M Sdn Bhd Abbott Diabetes Care No 22 Jalan Pemaju U1 15 HICOM Glenmarie Ind Park 40150 Shah Alam Selangor Malaysia 03 5569 3388 Abbott Laboratories N Z Ltd Ground Floor Bldg D 4 Pacific Rise Mount Wellington 0 Box 22 801 Otahuhu Auckland New Zealand 0800 106 100 Abbott Laboratories Pakistan Ltd Abbott Diabetes Care Opposite Radio Pakistan Transmission Hyderabad Road Landhi Karachi Pakistan 021 35100321 Abbott Laboratories Singapore Pte Ltd Abbott Diabetes Care 1 Maritime Square 12 09 Harbour Front Centre Singapore 099253 1800 272 2881 National Phytopharma Joint Stock Company 24 Nguyen Thi Ng
14. 39 AM Understanding Blood B Ketone Test Results High Blood B Ketone Results The meter displays results in mmol L The unit of measurement is preset You cannot change this setting Blood B Ketone is expected to be below 0 6 mmol L High blood B Ketone may be caused by illness fasting vigorous exercise or uncontrolled blood glucose levels Repeat a blood B Ketone test using a new blood B Ketone test strip when Hl appears on the display Your result is unusually high You question your result Your blood P Ketone result is 0 0 mmol L but your blood glucose is higher than 16 7 mmol L 300 mg dL 21 ART20287 Rev B indd 21 22 IMPORTANT The meter displays results from 0 0 8 0 mmol L Consult your healthcare professional before changing your diabetes medication programme Meter Display Result is between 0 6 1 5 mmol L and your blood glucose result is Higher than 16 7 mmol L Higher than 300 mg dL Result remains 0 6 1 5 mmol L or becomes Higher than 1 5 mmol L Appears when result is Higher than 8 0 mmol L What It Means High blood 8 Ketone A problem requiring medical assistance may be occurring You may be at risk of developing diabetic ketoacidosis Very high blood B Ketone or There may be a problem with the test strip What To Do Contact your healthcare professional for sick day management instructions Contact your healthcare professional im
15. Apply Sample symbols pon appear indicating the meter is 18 3 8 13 11 39 AM 21 Apply Blood to the Test Strip EN View the Result Bring the blood drop to the white area at the end of the test strip The blood is The meter beeps if sound is on when the result and i appears on the drawn into the test strip display The test is complete example shown The result is stored in memory or you can write it in a logbook IMPORTANT If the countdown does not start you may not have Turn Off the Meter 7 applied enough blood to the test strip See test strip instructions for Dbm 1025 Use the opened foil packet to remove the used test strip This turns off the meter use for re application instructions If the countdown still does not Discard the used test strip correctly start remove the used strip and discard it correctly Start a new test 8 Note You also press and hold Q for at least 2 seconds to turn off the meter The with a new test strip meter also turns off after 60 seconds of inactivity Hold the blood drop to the white area until the meter beeps if sound is on and the status bar appears on the display This indicates the test strip obtained enough blood The countdown appears on the display as the meter checks your B Ketone level Note Do not remove the test strip from the meter or disturb the test strip during the countdown 19 20 ART20287_Rev B indd 19 20 e 8 8 13 11
16. Customer Service Install battery with facing up Replace battery Reset date and time if necessary 3 8 13 11 39 AM 36 What It Means The test does Blood sample is too small not start after applying the blood sample Sample applied after meter turns off 37 ART20287 Rev B indd 37 38 What To Do See test strip instructions for use for re application instructions If the countdown still does not start remove the used strip and discard it correctly Repeat the test using a new test strip If the test still does not start contact Customer Service Remove the used strip and discard it correctly Review the testing instructions Repeat the test using a new test strip If the test still does not start contact Customer Service The test does not start after applying the blood sample What It Means Defective meter or test strip What To Do Remove the used strip and discard it correctly Repeat the test using a new test strip If the test still does not start contact Customer Service 38 3 8 13 11 39 AM Maintaining the Meter Replacing the Battery mmol L EAD dn g uq it eb Erz 39 ART20287_Rev B indd 39 40 The meter comes with a CR 2032 lithium coin cell battery installed It provides power for about 1 000 tests E Appears the display when the battery is low Causes the display light to fail to turn on When
17. DO NOT PRINT FROM THIS FILE FOR REFERENCE ONLY e User s Manual c Abbott ART20287_Rev B indd 1 3 8 13 11 39 AM ART20287 Rev B indd 2 3 e Definition of Symbols These symbols are associated with your meter LOT El amp E BPE Batch code Catalogue number Caution Consult instructions for use Date of manufacture Do not drink Do not reuse In vitro diagnostic medical device STERILE R g lt The European Battery Directive requires separate collection of spent batteries aiming to facilitate recycling and to protect the environment The batteries in this product should be removed and disposed in accordance with local regulations for separate collection of spent batteries This packaging is capable of being recycled Manufacturer Sterilised using irradiation lancets only Temperature limitation Use by Serial number 3 8 13 11 39 AM Table of Contents ART20287 Rev B indd 4 5 Intended Use RC nin ed 1 Monitoring System Kit 3 Setting Up the Meter 0 kanala yel ka n ene een en 4 Enter Setup a xuya bib W 4 Io I ET 4 mv 5 Date me en a 5 Tin pormmat ei a a BIS SR EDUC DEE KA i 5 Date Fornmati Wy W j 6 Measurement Units
18. e match Repeat the test if control solution results are outside this range Stop using the meter if control solution results are continually outside the range printed on the test strip instructions for use Contact Customer Service IMPORTANT Control solution results do not reflect your blood glucose or blood Ketone level 24 3 8 13 11 39 AM Performing a Control Solution Test 5 4 The Apply Sample symbols 6 and Al also appear 2 0 se foil test strip packet at the notch and tear down to remove the 9 and A will appear if performing a B Ketone control solution i test lt J Press to mark the test as a control solution test 4 The X B appears The meter is now ready for you to apply control solution to the Insert the Test Strip test strip With the contact bars 3 black lines facing up insert the test strip into the meter r until it stops This turns on the meter IMPORTANT The test result will be saved to memory as a blood glucose or blood B Ketone result if not marked as a control solution Note The meter turns off after 3 minutes of inactivity Remove and reinsert the test This may affect your blood glucose averages unused test strip to restart the meter js Apply Control Solution to the Test Strip This display always appears when the meter is turned on man Shake the control solution bottle to mix the solution Apply a drop of control solution to the white area at the end of the tes
19. est Strips Lancets Ji Jj Press and hold until and JJ appear This indicates the beeper is turned on Lancing Device X and JJ indicate the beeper is turned off Items not included Beeper Blood Ketone Test Strips Data Management System The beeper sounds when Control Solutions Press or gt to change the beeper setting The test countdown starts Press to save the beeper setting and move The test countdown ends to set time ART20287 Rev Bindd 3 4 e 8 8 13 11 39 AM Time Date Format 2 0 man The hour flashes Press or P until the correct hour appears Press to save Im 15 The date format flashes Press to change the date format Press T to the hour and to move to set minutes save the date format and to move to the measurement units The minutes flash Press or P until the correct minutes appear Press T to save the minutes and to move to set date Month Day in i5 IU Day Month Date 2 0 0 The year flashes Press or gt until the correct year appears Press 1 to save the Me et Co i Co cn IMPORTANT Please check that the correct time and date is set before you use the meter for the first time This will help you keep records of The month flashes Press or gt until the correct month appears Press to when you monitor and will help you and your healthcare professional int save the month and to move to set day make informed decisions about your care
20. hia District 1 Ho Chi Minh City Vietnam 910 6640 ext 858 Abbott Laboratories Abbott Diabetes Care 102 E delos Santos Ave Madison Street 1554 Mandaluyong Metro Manila Philippines 02 7028577 02 7028578 Abbott Diabetes Care Ltd Range Road Witney Oxon 0X29 OYL UK FreeStyle and related brand marks are trademarks of Abbott Diabetes Care Inc in various jurisdictions 2010 Abbott ART20287 Rev 02 13 3 813 11 39 AM
21. mediately Repeat the test with a new test strip If the result is HI contact your healthcare professional immediately 22 3 813 11 39 AM MediSense or Optium Glucose and Ketone Control Solutions 23 ART20287_Rev B indd 23 24 IMPORTANT When you open a control solution bottle for the first time count forward 90 days and write this date on the control solution bottle using a pen that won t smear or wipe off Throw away any remaining solution after this date Control solutions may be used with either glucose or B Ketone test strips Use control solution To practise testing without using your own blood To confirm that the meter and test strips work together correctly When you are unsure of your blood glucose or blood B Ketone test results Contact Customer Service for information on how to obtain control solutions e Important Control Solution Testing Notes Use only MediSense or Optium Glucose and Ketone control solutions with the meter Replace the cap securely on the bottle immediately after use Do not use control solution past the expiry date Do not add water or other liquid to control solution Do not swallow or inject the control solution Do not use the control solution as eye drops IMPORTANT Control solution results should fall within the control solution range printed on the test strip instructions for use Check that the lot number printed on the test strip packet and instructions for us
22. n condensing Blood g Ketone See blood Q Ketone test strip instructions for use 7 my N Functions System relative humidity 10 to 90 non condensing Blood glucose testing Clinical testing demonstrates that altitudes up to 2 195 metres System altitude 7 200 feet above sea level do not affect results Blood Ketone testing Memory 450 events Glucose averaging 7 14 and 30 day Control solution marking and testing 43 44 ART20287_Rev B indd 43 44 e 3 8 13 11 39 AM References 1 Schade DS Eaton RP Metabolic and clinical significance of ketosis Special Topics in Endocrinology and Metabolism 1982 4 1 27 2 Wiggam MI O Kane MJ Harper R Atkinson AB Hadden DR Trimble ER Bell PM Treatment of diabetic ketoacidosis using normalization of blood 3 hydroxybutyrate concentration as the endpoint of emergency management Diabetes Care 1997 20 1347 1352 3 Harano Y Kosugi Hyosu T Suzuki M Hidaka H Kashiwagi A Uno S Shigeta Y Ketone bodies as markers for Type 1 insulin dependent diabetes and their value in the monitoring of diabetes control Diabetologia 1984 26 343 348 4 Ubukata E Diurnal variation of blood ketone bodies in insulin dependent diabetes mellitus and non insulin dependent diabetes mellitus patients The relationship to serum C peptide immunoreactivity and free insulin Ann Nutr Metab 1990 34 333 342 5 Luzi L Barrett EJ Groop LC Ferrannini E DeFronzo RA Metabolic effects of low dose insulin
23. range Causes of out of range results include The meter and test strips may not be working correctly Repeat the test with a new test strip Follow the testing instructions carefully Contact Customer Service if the result is still outside the printed range Turn Off the Meter Use the opened foil packet to remove the used test strip This turns off the meter Discard the used test strip correctly Note You can also press and hold D for at least 2 seconds to turn off the meter The meter also turns off after 60 seconds of inactivity 28 3 8 13 11 39 AM Managing Your Results Meter memory stores up to 450 events control solution blood glucose and blood Ketone results and other meter information The most recent events appear first Your most recent result with time and date if set it was obtained appears next b oH The following symbols indicate the type of result H Blood glucose result Viewing Test Results in Memory Blood B Ketone result Start with the meter off no test strip inserted Press Dae n n ia mazar en xw Glucose control solution result eg Ketone control solution result Press to review the previous result Repeat this for each result you wish to review This display always appears when the meter is turned on CODE mg dL E EB J Do not use the meter if the display check screen does not exactly match the Note To exit meter memory at any time
24. rip port Do not pour liquid into the test strip port or onto the buttons Do not immerse the meter in water or other liquid 41 ART20287 Rev B indd 41 42 Storing the Meter Store the meter in the carrying case provided Specifcations Size Length Width Weight Power Source Battery Life Memory Storage Temperature Meter 7 47 cm Top 5 33 cm Bottom 4 32 cm Depth 1 63 cm 40 to 46 grams 1 CR 2032 Lithium coin cell battery Approximately 1 000 tests Up to 450 events including control solution blood glucose and blood Ketone results and other meter information 25 to 55 13 to 131 F 42 3 813 11 39 AM Test strips See test strip instructions for use Beeper On When countdown starts and ends Control solution See control solution instructions for use Display Light On Stays on during countdown turns off 30 seconds after results Pay 9 appear Stays on while transferring results to a computer Assay Method Amperometric electrochemistry Data Port Yes Assay Range Operating Range Blood glucose See blood glucose test strip instructions for use Meter temperature l0 to 50sGI bOStO225E Blood 8 Ketone See blood Q Ketone test strip instructions for use System temperature See operating range of the test strip you are using Control Solution Range See test strip instructions for use Blood glucose See blood glucose test strip instructions for use Meter relative humidity 1096 to 9096 no
25. t strip in the area shown The control Do not use the meter if the display check screen does not exactly match the races solution is drawn into the test strip example Contact Customer Service See Testing Your Blood Glucose section for more information Hold the control solution drop to the white area until the meter beeps if sound is on and the status bar appears on the display This indicates the test strip has obtained enough control solution nam 10 5 Time and date if set appear next H 26 ART20287_Rev B indd 25 26 e 8 8 13 11 39 AM lt J r 349 12 10 27 ART20287_Rev B indd 27 28 The countdown appears on the display as the meter checks the glucose or B Ketone level Note Do not remove the test strip from the meter or disturb the test strip during the countdown If the countdown does not start remove and discard the used test strip turn off the meter and try again with a new strip View the Result The meter beeps if sound is on when the result appears on the display The test is complete examples shown The result is stored in memory as a control solution result or you can write it in a logbook as a control result appears with the result if performing a B Ketone control solution test X B mmol Compare the control solution result to the range printed on the blood glucose or blood Q Ketone test strip instructions for use The result should fall within the
26. unusual blood glucose results When you and your healthcare professional feel it is necessary 15 16 ART20287 Rev Bindd 15 16 e 8 8 13 11 39 AM Test Site Fingertips IMPORTANT Wash your hands with warm soapy water to ensure accurate results Thoroughly dry your hands To warm the site apply a warm dry pad or rub vigorously for a few seconds Note Avoid squeezing the puncture site Use only fingertip blood samples for blood B Ketone testing Performing a Blood G Ketone Test 17 ART20287_Rev B indd 17 18 Open the foil test strip packet at the notch and tear down to remove the test strip Insert the Test Strip With the contact bars 3 black lines facing up insert the test strip into the meter until it stops This turns on the meter Note The meter turns off after 3 minutes of inactivity Remove and reinsert the unused test strip to restart the meter mod EE CB mmo BEI BB 8B BB BB 190 10 5 This display always appears when the meter is turned on Do not use the meter if the display check screen does not exactly match the example Contact Customer Service See Testing Your Blood Glucose section for more information Time and date if set appear next mn for you to apply a sample to the test strip Obtain a Blood Sample Select a test site Use the lancing device to obtain a blood sample See lancing device instructions for use for more information and the
27. ur healthcare professional s advice when testing blood glucose levels and blood p Ketone levels Test blood B Ketone B hydroxybutyrate in fresh whole blood samples from fingers only IMPORTANT Use only FreeStyle Optium blood glucose test strips Observe caution when using around children Small parts may constitute a choking and FreeStyle Optium blood Ketone test strips Other test strips hazard may produce inaccurate results How Your FreeStyle Optium Blood Glucose Monitoring System Works When you insert a test strip into your meter the Apply Sample message shows on your meter s display window When a blood sample or control solution sample is applied to the test strip the glucose or B Ketone reacts with the chemicals on the test strip This reaction produces a small electrical current that is measured The result shows on your meter s display window IMPORTANT Avoid getting dust dirt blood control solution water or any other substance in the meter s test strip port ART20287 Rev Bindd 1 2 e 8 8 13 11 39 AM Monitoring System Kit Contents Setting Up the Meter Monitoring System Kit contains Enter Setup Mode Commands SO logbook Start with the meter off no test Press to move between setup displays in j Press B to move forward User s Manual Warranty Card po up Meanie e Carrying Case 4 e Press 4 to go back NL Presse to turn display light On and Off Your Kit may also contain Blood Glucose T

Download Pdf Manuals

image

Related Search

Related Contents

Ergopower uktra shift 10 et 11V  Mode d`emploi, Instruction JURA F9  5532 Glovebox Chamber User Manual  Benutzerhandbuch Nord Lead 4 Nord Lead 4R  H.264 Digital Video/Audio Recorders User`s Manual 9-CH  MANUAL DEL USUARIO  flo-dynamics brakemate brake flush machine  Samsung French Door 543L Inox Look 127V manual do usuário  Fractal Design Tesla 450W  Dose journalière recommandée et mode d`emploi  