User Manual and Technical Description

image

Contents

1. NZS 3200 2 38 1997 28
2. disinfection 16 1 Basic instructions before cleaning In order to disinfect the bed use detergents designed only for cleaning the equipment for standard use in healthcare Do not use abrasives scouring powder scourers and other materials that could amage the coating Do not use corrosives caustics or strong acids Do not use detergents that could alter the structure or behaviour of the plastics petrol toluene acetone etc Clean by wiping with damp and properly wrung cloth The beds are not designed for maintenance in automatic bed washers The beds are not designed for cleaning with a current of pressure water spraying showering or steam cleaning We cannot be liable for any damage or risk of damage if inappropriate cleaning or disinfectant detergents are used 16 2 Cleaning procedure 1 Adjust the mattress platform to a higher the highest position and also adjust the position of the back and thigh rest so it provides access for cleaning the back part rests 2 Block the adjustment function on the nurse control panel including the foot controls to prevent accidental bed movement and injury during cleaning 3 Disconnect the bed from the mains 4 Move the bed to a place where cleaning will take place and then lock the bed brakes 5 Clean and disinfect the bed to the appropriate extent The extent of the cleaning and disinfection should depend on the intended purpose and level of bed contamination see the fo
3. tightened Other special maintenance procedures are qualified work which we recommend to be carried out by a professional service The user is obliged to ensure a preventive bed inspection after a 12 month period which will be performed by the manufacturer s service or the service firm authorised to perform this activity The manufacturer issues the relevant certificate to these service organisations and the inspection is confirmed by the relevant service worker in the list of performed service activities If you discover irremovable faults or inconsistencies during maintenance the bed may not be used further In such a case immediately contact the manufacturer or seller 20 Environmental protection CareWell is aware of the necessity to protect the environment for future generations and therefore devotes great attention to the development innovation planning production and use of such technology and materials that are environmentally friendly 24 This product is constructed from materials that are environmentally friendly The product does not contain dangerous substances based on cadmium mercury asbestos PCB or CFC The operational noise level of the product meets the requirements for the protection of public health against undesirable noise and vibration in protected indoor building premises Wooden parts used in the product are not made from exotic wood such as mahogany jacaranda ebony teak grenadil or santo and do no
4. to the maximum and minimum position Check the faultless function of the castors and central control extension of the mattress platform removal of the head and foot ends Check all the functions of the control panel Make sure the siderail function works without any problems locking in the upper position A Check for obstructions both above and below the bed frame and position the bed so that it can operate through the full height range without any chance of collision Head end of the bed must be a minimum of 300mm from the wall initially Always engage the brakes when the bed is positioned or unattended 7 1 Connecting the bed to the mains The mains cable for feeding the bed forms an integral part of the bed so the bed is connected by inserting the plug on the mains cable into a 240V socket 7 2 Siderail installation We recommend that the bed be installed by two people 11 1 After removing the packaging of both beds place the right and left siderails including the corner segments along the length of the bed as in figure At the same time the siderails must be positioned so that the foot end of the siderail faces the foot end of the bed The foot ends of the siderail can be easily recognised because they can be extended telescopically Both bed ends remain on the mattress platform if this is an ordered bed component Figure No 6 The bed looking from head to foot end with the siderail attached in the collapse positi
5. 7 Protection against mechanical hazards Protection against excessive temperatures Constructional Requirements o TGA Medical Device Class 1 This declaration of conformity becomes invalid if the products are altered without the written agreement of the manufacturer Product description Hospital Care beds Croft Oden and Contempora Signed Jason Falinski Managing Director Contents Rec ic i it ee ee ee ee ee 6 VAIN Giri sna nse naan EE EE E EE E uni bingeeeeeninneremaan 6 1 1 Means of te S12 A en en ern rn ee 6 3 Conditions OF USC ingacdimscomsatdnesdancatanieccevsnsivetd vags dunevescaisqecetarcsiatisacanveseacds 8 4 Technical parameters sesssssssersrrrerrererrrrerrerrrrsrrrrsrrnrerrrrerrererrerrrrererrerrre 8 TANA EIEE E T A T E EA 9 a Varme SPENGO ceire an E E E EEE EE NEA EEEE 9 Ada 0ns 10 Standard DE wacscaecsetaeds reetecneesy ein a e aE eee aeei 9 Head boards and foot boards in a range of colours and finishes 9 YOUr ChOlGCE OF DEG COlOUF siicucnntaseaennanctadwasncensnedaGsenapanict useetennmneniecavanne 9 0 Bed SS OE Oi N aeu S E culmea 10 Figure No 3 Bed description sessssesssrenssrerrersrrererrersrrnrerrererrerrrren gt 10 7 Bed installation and putting into the OperatiOon cccecececseseeeeeeveveueeeeneneees 11 7 1 Connecting the bed to the MainS ssesssssersersnrsnrsrrerrerrrrnrrrrrerrrrrrrrrrrre 11 k2 9106rall INSCAN ATION asserere renin ed en Aai aarnaiwredebeanuved aS
6. CareWell amp User Manual and Technical Description Care Bed Version 01 220 11 January 2008 SE x amp oO BA gt gt 4 gia ecan Dear Customer Thank you for buying CareWell health care equipment and furniture We feel sure that our equipment will meet and exceed your expectations We are working hard to change the way that people see and feel about health care to make it a nurturing comforting and reassuring The following manual is important for getting the full potential out of our equipment and furniture please read it before use If you need further information or would like to order more of our equipment and furniture please contact us on the following coordinates www carewell com au l info carewell com au l Toll Free 1300 30 32 88 Telephone 61 2 9564 0684 Facsimile 61 2 564 0940 a rn CareWell AS NZS 3200 2 38 1997 GAT eWe EU Directive equivalent 93 42 EEC The signatory CareWell Health Pty Limited 78 90 Old Canterbury Road Lewisham NSW 2049 The products described below are specifically developed and intended for the aged care nursing industry and have been derived in consultation with operators in that industry to suit their specific needs This letter serves to confirm that the referenced products in the form distributed by ourselves meet the safety and stability requirements of AS NZS 3200 The following Standards apply o AS NZS 3200 2 38 199
7. The height of the mattress platform must be adjusted to the height and condition of the patient No other people may sit on the bed When handling the mobile parts of the bed care must be taken that the patient other people or objects do not get trapped during its movement If a lifting pole or infusion stand is fixed to the bed during movement lifting or tipping increased attention must be taken to the space around the lifting pole and infusion stand so the equipment does not get damaged Before cleaning or maintaining the bed the supply cable must be disconnected from the mains The bed may not be used where there is a danger of explosion or in the presence of inflammable liquids When repairing the bed only original materials and components may be used when replacing damaged parts with inappropriate materials and components the manufacturer does not guarantee against any damage that can occur When maintaining and cleaning the bed in the space between the undercarriage and mattress platform the adjustment functions on the central control panel must be blocked so there is no accident due to any unintentional activation of the foot switches and subsequent downward movement of the mattress platform Only CareWell mattresses and accessories may be used on the bed when using incompatible mattresses the siderails may function badly and endanger the safety of patients When transporting the bed the mains cable must be suspended with t
8. a 11 Figure No 6 The bed looking from head to foot end with the siderail attached in the collapse POSITION essssessessersererrerrerrerrrrrrrerrrrrerrerrerrrrns 12 ONE NO Soar E E E E E E 12 B Bed PO IONIN O arnasa E E EE EE ET 13 8 1 Supervisory control panel ssssssesssrsrsrrererrsrrrrerrrrerrrrerrererrererrerrrrsrn 13 0 2 Aanadset Opera tO orreee rnrn Rn NEOA AAEE RTTE 13 Figure No 12 Hands Eiir E A Ea 13 9 Bed GIGS sane cotansesttanaus aceacanseieetarssesraceeasgssthas seawecodamitersseauestixtdestiepiersaedeise 14 TO SIOC GUNG erropea rE E E A E E E ER 15 Figure No 20 Zero position of siderailS essessesserssrerrerrerrrrrrrrrrrrrerrerns 15 Figure No 23 Full position of siderallS cc ccececececucuceeeeeeeeeeuseneneneneneees 16 10 Manipulation with siderailS ccccccecscseseeeeeeeueeeeeeteeereteeseeeeeereearaeananennegs 17 Figure No 25 Manipulation with siderailS sessessssesseriereerrrrsrrrrrerrerns 17 UAE AS CON COM Ol E ST EA E EE he tee atta hate bok ae ant tanta TA 18 Figure No 28 Castor control in unlock and lock position seee 19 12 FOG GOWA TOO DOA O esccaine ia e a E aA EAA 19 13 ACCESSOMNGS oniinn ea E EA A AE 20 ISL SeErHeED and lV DOE ana A EE 20 14 Sal ely NOME ION ternera a a ania eeaniebes eee 20 15 Mattress platform cover removal cccccececeeeeeeeeeeeeeretereeeeaeaeaeaeaeeeenennenans 20 16 Cleaning ANA disinfection cecececececeeeeseseeaeae
9. auauaueeeueueueueetstetetsneteteretes 22 16 1 Basic instructions before cleaning cccscsesecececeeeeeeererereeseeaeaeaeaeaeerans 22 16 2 Cleaning WhO CCOUI Ce iiiscrstacivniinnaitientaviaaniaaniialdieyreiaietnindiaetigende 22 1623 EXCENE OF CIS GING onere su haiadnntevinds awa nceerea cera mateo ganatsuaineue 22 1624 Cleaning te Sideral witsascevsprarecehasabiaeniaarsieidea nicole eterna aiade ss 23 E TEROUDIES NOOUING sicd seen ctaciin tcaeteapta a a a in ena 23 TG SCO RAG C oawtctieiet ates irsovauantaesacuai ans dacsuad uct ols sanaes nce a ence ienentaad eis 24 t9 MaM enan E sictsest pansies O N 24 20 Environmental DrotecCtHONr enaissi a a 24 ZU Guarantee and SCLVIGC sar see aura a a a a a 25 22 Record of performed service procedures and regular service maintenance 26 Z3 BeG COMPICTION COMIC ALC oirrese iain anaieadide ahora 27 p E AE lata eves E E E E A A E EE N E T 27 2I CONACS unea EEEE AEE EEATT 28 1 Introduction The adjustable care bed is designed for the hospitalisation of patients particularly on standard aged care facilities both low and high care environments The manufacturer does exclude installation in hospitals or intensive care units but this can be considered only on the basis of the responsible selection of the health care provider The bed is produced in accordance with international norms in particular Australian Standards AS 3200 3 28 1997 safety and stability provisions The manufacturer works acc
10. d care must be taken that the bed be disconnected from the mains and the bed does not run over the mains cable The mains cable must be suspended when the bed is moved using the plastic hook placed in an appropriately safe place on the bed 12 Fold down foot board If necessary the bed mattress platform can be extended by up to 17 cm at the foot end using the following procedure 1 release the extension by lifting the foot board until it clicks in to position 2 Once it is clicked in to position it is possible to lower the footboard to a 90 angle 3 to restore the foot board to its original position requires the same process 3 1 pull the foot board until it clicks into position unlocking the board 3 2 at this point return the board to its vertical position and lower it until it 1s locked in position 19 13 Accessories It is possible to use accessories for the bed designed by the manufacturer The use of different accessories not compatible with the bed may cause a safety risk or damage the bed respectively In such cases the manufacturer will not be liable for the consequences 13 1 Self Help and IV pole There are 2 sockets to store the lifting or IV pole that are found in the corners at the head end The max Load of the IV stand is 8 kg the load of one plastic hook is 2 kg The infusion stand is not designed for the attachment of infusion pumps A The self help pole is placed in the socket so the pin falls into the slot T
11. e highest position 17 Troubleshooting One of the components of the electrical installations is faulty control controlling unit motor Try controlling the function in question with the aid of another control If only the control in question is not working order a new control from the manufacturer s service organisation If the other controls do not work as well order the servicing of the bed Do not repair the electric motor power box and other electrical parts if a fault occurs Do not open the protective electric motor or power box cover 23 18 Storage For problem free storage we recommend Disconnecting the bed from the electric mains feed Dismantle the accessories lifting pole and so on Wrap up the bed and accessories or cover so that the coating and plastic parts are not damaged The same conditions apply to the storage space areas as for the work environment 19 Maintenance Because some parts of the bed may suffer wear and tear due to use cleaning and disinfection inspect the following places on the bed at least once every 3 months 1 side rail wipe and lubricate with silicon oil 2 joints and pins of the adjustable mechanisms backrest thigh rest and calf rest wipe and lubricate with silicon oil 3 motor pistons wipe and lubricate with silicon oil 4 upper columns fixation check 1f the screws and nuts providing connection between column stirrup and mattress platform are well
12. e products 24 Serial label When communicating with the manufacturer supplier or service organisations the bed must be identified with the aid of the model and serial number stated on the serial label on the outer side of the mattress platform frame 2 25 Contacts Date of issue 10 February 2007 Conformity declaration issued by CareWell Health Pty Limited 6 Durdans Avenue Rosebery NSW 2018 www carewell com au info carewell com au Toll Free 1300 30 32 88 Telephone 61 2 9667 8770 Facsimile 61 2 9669 4830 As the producer of the product name brand Oden Variants of the product The variants are specified in the technical documentation Description and function designation Positionable bed for standard care Clssification of the product as the medical device Class I nonsterile without measuring function A Declaration I declare that the said product is designed for an aged care environment and meets the safety and stability requirements under the AS NZS 3200 2 38 1997 and that measures have been taken to ensure the conformity of all the products brought to market with basic requirements B Fulfilled technical requirements This product s characteristics comply with the safety and stability technical parameters related to it and stated in AS NZS 320 2 38 1997 which stipulates the technical parameters for healthcare products C Means of assessing conformity Conformity was assessed by the procedure stated in AS
13. g staff should explain to the patient the manner and form in which the bed can be positioned If the medical staff admits that the patient s condition is inappropriate for the patient to be able to adjust the bed on his own the bed s position adjustment should be locked in the above described manner with the aid of the nurse lock out 9 Bed ends The beds are equipped with two removable high pressure laminate and aluminium profiled ends Before starting to use the bed the ends must be fixed to it Both the head and foot ends of the beds are removable by simply pulling them up as shown When removing the ends the end must be gripped with both hands and carefully pulled up in the direction of the arrows to remove them When placing the ends back on the bed the reverse procedure 1s used 14 10 Siderails The beds are always equipped with low profile siderails The siderails consist of three parallel horizontal bars mutually connected above each other and are accessible on both sides of the bed The siderails can be adjusted to three height levels and each has its specific use Figure No 20 Zero position of siderails Zero position This level is adjusted when the siderails are completely slid under the mattress platform The siderails do not extend from the bed and enable the nurse to gain access right to the edge of the mattress The bed can be placed right next to a stretcher or another bed so the patient can be mo
14. he aid of a plastic hook on the bed end or other safe place to prevent the cable from being run over and damaged An inspection should be made at least once a month for wear of all mobile parts The doctor treating the patient or another responsible person providing treatment decides about the use of the handrails for supporting the patient when leaving the bed 3 Conditions of use The bed may be used in an indoor environment where the Range of the surrounding temperature is from 10 C to 40 C Range of relative humidity is from 30 to 75 Range of atmospheric pressure is from 700 hPa to 1060 hPa It 1s essential to consult the manufacturer in advance when wanting to use the bed for different purposes The bed is designed for use in rooms for medical purposes Electrical installations must therefore meet local norms laying down the necessary conditions for electrical installations In spite of this it is recommended to disconnect the bed from the mains during exceptional cases i e a storm 4 Technical parameters Description Value External Dimensions 970mm wide x 2060mm long Mattress Platform Dimensions 910mm wide x 2000mm long Castor Wheels Multi directional tandem set dual caster with central locking feature Height Adjustable From 185mm to 790mm Weight 85kg nett Backrest positioning J15 Siderail height above mattress platform 390mm without mattress Safe working load 185kg Max Self Help Pole Lifti
15. he lifting pole must be equipped with a triangular plastic handrail with an adjustable band By shortening this band it is possible to adjust the appropriate height for the patient A When turning the lifting pole in the socket away from the bed and using the lifting pole in this position there is a serious risk of turning the bed over That is why the lifting pole cannot be fixed this way The only two correct positions for the lifting pole in the socket are by the defined locking of the lifting pole pin in one of the slots in the socket 14 Safety night light The bed can be equipped as requested with undercarriage lighting that helps the patient or operator to find their way around the room with switched off or dimmed lights The lighting is constantly switched on when connected to the mains and is automatically switched off when operating on a backup battery 15 Mattress platform cover removal The Image bed has a four section mattress platform a backrest pelvic rest thigh rest and calf rest and its surface consists of removable metal covers By simply removing the infection pads the inner bed construction can be fully cleaned and disinfected After removing the covers it is thereby possible to clean all the metal 20 parts and mechanisms of the mattress platform and all the parts of the electrical installations backrest and thigh rest motors mains box battery box control unit box cables 21 16 Cleaning and
16. llowing chapter 16 3 Extent of cleaning 1 Daily cleaning Surfaces exposed to daily contact with the patient and operator All control elements for bed adjustment All handling features back and thigh rest handrails CPR handrail The entire area of the ends including the side columns and parts beneath both ends The entire siderail area at the highest adjusted position 22 Freely accessible mattress surface Supporting handrail Side accessory bars 2 Cleaning with a new patient Bed parts stated in point and also All plastic mattress platform covers Plastic undercarriage cover Column unit surface The mattress from all sides Freely accessible metal parts on the mattress platform Tubes for feeding cables between the undercarriage and the mattress platform Surroundings of the lifting pole infusion stand sockets Protective corner bumpers Central castor control castors and levers U Complete cleaning and disinfection Bed parts stated in point 1 and 2 and also All the inner construction parts accessible after the removal of the mattress platform plastic covers as mentioned in the chapter Removal of the Mattress Platform Covers Inner parts of the corner segments of the siderails see the next subchapter 16 4 Cleaning the siderails The siderails which consist of 3 profiles are good to decontaminate during daily cleaning Clean the siderails daily by adjusting them to th
17. nal Electric back rest panel removable Remote control pendant Wall protector Central Locking Castor wheel X lifer function CONNDNBRWN 10 7 Bed installation and putting into the operation 1 Remove the packaging from the bed and dispose of it in an environmentally 2 According to the delivery note check that the delivery is complete check whether any visible damage has occurred to the bed during transport 3 Carefully read the user manual 4 Remove safety ties from siderails and electrical wiring 5 If the bed 1s delivered with dismantled ends and siderails two beds stacked above each other these parts of the bed must be installed according to the instructions in the chapters Siderail installation and Bed ends 6 Should the bed be equipped with a satellite control panel it should be installed according to the information in the chapter Satellite control panel installation 7 If the temperature during bed transport or storage 1s different from the operating temperature it is necessary to leave the bed without connection to the net for 24 hours to adapt it to internal conditions 8 Connect the bed to the mains including its connection for balancing the potential see Subchapter 7 1 7 3 and charge the battery back up see Sub chapter 7 2 9 Check the functionality Adjust the individual parts of the mattress platform to the maximum position the mattress platform height and tilting
18. ng Load 75kg Siderail height above the mattress platform without the mattress Safe Working Load SWL 185kg Type of protection against electrical current Class 1 Level of protection against injury from Applied part type B electrical current Electric motor operation mode load factor 10 per cent max 2min 1 8min Warning The bed is powered from the mains meaning sensitive instruments may be disrupted In order to reduce the risk of undesirable electrical magnetic effects as much as possible the bed 1s manufactured in accordance with the safety and stability requirements of AS NZS 3200 2 38 1997 In order to avoid these problems the bed must be used in accordance with this manual 5 Variants specification Optional features according to the customer s wish Siderails 3 segmented 1500mm length siderails adjustable up to 45 cm Castors ProCare tandem dual castors 75mm ProCare dual castor 75mm Control elements Handset Handset with an adapter for easy connection plug and play Other features Add ons to standard bed Head boards and foot boards in a range of colours and finishes Timber trim bed frame with matching head and foot boards Extended length Your choice of bed colour Your choice of bed length 6 Bed description TT oO _ gt A ES Figure No 3 Bed description Foot board optional Foot panel removal Manual knee break electric knee break optio
19. on Figure No 7 The bed head to foot end with the siderail attached and at full height 2 Secure all 4 corner segments of the siderails with the supplied screws A and B The joining section that needs to be tightened to fasten siderails to the bed pictured with the securing device that holds the siderails in place 12 8 Bed positioning The various controls stated in the following chapters can be used for the electrical adjustment of the bed The selection of the controls corresponds to the bed equipment ordered by the customer 8 1 Supervisory control panel A nurse lock is optional and can be located at the end of the bed This gives carers control over the bed position and locks out patients from adjusting the bed on their own 8 2 Handset Operation The handset can be used for positioning the sections and for height adjustment of the mattress platform of the bed by either the patient or the carer The use of the handset is the same as the supervisor panel A Always check for obstruction before raising and lowering Figure No 12 Handset 1 To position the backrest use controls labeled A 1 1 to raise the backrest press down on the up arrow 1 2 to lower the backrest press down on the down arrow 2 To position the platform mattress 2 2 to raise the platform mattress press down on the up arrow 2 3 to lower the platform mattress press down on the down arrow 13 Before using the control the operatin
20. ording to the certified quality management system under the norms EN ISO 9001 2000 and EN ISO 13458 2000 Before handling the bed you must become acquainted in detail with this manual All actions and handling of the bed must be performed in accordance with the manual Any other actions that are inconsistent with the manual or for the purpose of use of the bed are performed at your own risk and the manufacturer shall not be liable for any damage Therefore it is important that the manual be available to the user throughout the bed s service life WARNING The manufacturer shall not be held liable for any damage injuries or accidents arising from inattentiveness bad negligent or erroneous handling The manufacturer seller or trained person shall perform the basic instructions for handling the bed The safety features for operating the bed and the instructions concerning the bed handling must be strictly observed 1 1 Means of delivery The bed is supplied either completely assembled or with foot and head ends package separately requiring insertion on site In the case of the second means of delivery all that is required is for the bed to be removed from the box and the head and foot boards inserted According to the delivery note an inspection must take place on site to make sure the entire delivery is complete Any faults or damage must be immediately reported to the carrier and the supplier in writing When traveling
21. t come from the Amazonian rainforests or other virgin forests All packaging is marked in accordance with the packaging regulations The product contains recyclable steel plastic and electronic components to optimise recycling once the operation of the product ends separate the individual parts so that the raw materials from which the product is made can be put to further use 21 Guarantee and service There is a 36 month guarantee on the bed commencing the date of delivery unless the purchase contract states otherwise The manufacturer or authorised service organisation performs guarantee and post guarantee repairs based on a written qualification certificate to perform these activities The manufacturer issues the relevant certificate 25 22 Record of performed service procedures and regular service maintenance Purpose and Description Performed By I confirm that technical staff have undertaken the repairs has described above and the equipment is now functioning normally CUSTOMER SIGNATURE 26 23 Bed completion certificate Order Number Customer Model Number Supplier Acceptance Date Serial Number Guarantee Period I confirm that the operating staff has been trained for the purpose of becoming acquainted with the use of the bed CUSTOMER SIGNATURE This completion certificate also serves as a guarantee certificate When in contact with the seller the customer shall state the number of the order or number of th
22. tions by smoothly sliding them downwards Raising the siderail upwards Grip the upper siderail with both hands not the locking handgrip and smoothly slide upwards adjusting the required siderail height to one of the locking positions the siderails lock themselves in one of these positions 17 11 Castor control Central Locking Castor All four castors can be locked by pressing down on the locking bar located at the end of the bed a The castors can then be unlocked by raising the bar with the operator s foot Fig 28 0 Single Locking Castor The beds are equipped with castor control The Oden bed comes standard with central locking feature The controlling brake levers are located on each of the wheels Fig 28 1 Unlock 18 Fig 28 2 Lock Figure No 28 Castor control in unlock and lock position 1 Each castor can be locked with the foot 2 A castor is locked by pushing down on the larger lever that extends further and is at the bottom of the two lever system marked A 3 Locking the wheel pushes the smaller lever higher as shown in fig 28 2 4 Unlocking the wheel requires the user to push down on the smaller top lever which resets the levers back to the formation in fig 28 1 readying it to be locked once more Care must be taken to make sure the castor brakes are always activated when the bed 1s standing is being assembled or dismantled so that the bed does not move accidentally When moving the be
23. ved safely Semi position The siderails are 28 cm in height above the mattress platform This position can be used to prevent patients from falling off the bed who due to their mental condition are not expected to try to leave the bed spontaneously In no case 1s this position designed for a patient in shock restless patients or patients with mental disorders 15 Figure No 23 Full position of siderails Full position The siderails are 45 cm in height above the mattress platform Siderails at this height ensure safety even when active antidecubitus mattresses up to 23 cm thick are used Extra high position option The siderails are 59 cm in height above the mattress platform Siderails at this height ensure safety even when active antidecubitus mattresses up to 37 cm thick are used A siderail adjusted only to the High and Extra High position can be considered a full siderail in accordance with the technical norms that apply to an adjustable bed These two positions guarantee the norm prescribed height of 22 cm between the upper edge of the siderail and standard used mattresses 16 10 Manipulation with siderails Figure No 25 Manipulation with siderails Lowering the siderail downwards 1 ensure that there is nothing leaning or resting on the siderail prior to operation 2 pull the locking handgrip outward away from the bed A 3 adjust the siderails to any of the selected lower posi
24. with the bed loading and unloading it care must be taken that the bed brakes are unlocked The bed castors are designed for use in an indoor environment and for travel on even smooth and clean floors ceramic floor tiles linoleum cast floors and the like The castors may become damaged when traveling with the bed along a rough uneven or dirty surface Safety instructions Before using the bed you must read the manual and handle the bed in accordance with it The bed may not be used if faults have been detected on it that may injure the patient staff or a third person the bed or the surrounding interior The bed may only be handled by persons who by their knowledge or practical experience provide a guarantee that the bed shall be handled without any problems The operating staff should acquaint the patient with the control functions that are designed for the patient control panel supporting handrail for leaving the bed The user operator should before using the bed be comfortable with the functionality and faultlessness of the bed The bed may be used only on even and solid floors and under the conditions set forth in the paragraph Conditions of use The bed may not be overloaded even for a brief period If there 1s a patient on the bed the castors must be locked with the exception of when traveling with the bed an unlocked bed poses a risk of injury to the patient when leaving the bed or when supporting himself

Download Pdf Manuals

image

Related Search

Related Contents

Fluoroskan Ascent FL User Manual  Electro-Sentry 16 Operating Manual - Electro  USER MANUAL  1746-6.7, SLC 500™ RTD/Resistance Input Module, User Manual  COMBATTRE LA MORTALITÉ  Manual de Usuario - AIRIS TM475  RMF-RS1/UP - Rohrmotor24  CAN SPIDER - RTD Embedded Technologies, Inc.  Orçamento Posto de Saúde Fronteira  