1. Objectives After training students should have the minimum competencies Successfully simulate a patient rescue using the AED Training Device Understand the operation of rescue AEDs Understand differences between the AED Training Device and rescue AEDs 70 01124 01 D 4 Powerheart G5 AED Training Device User s Guide AED Training Device Overview The AED Training Device is a simulated AED for teaching basic rescue skills and correct defibrillation procedure to first responders who use Cardiac Science AEDs The AED Training Device can Simulate each rescue stage Change rescue scenarios three rescue scenarios are available Create simulated cardiac rescue scenarios both shockable and non shockable Instructors operate the AED Training Device by remote control AED Training Device Parts The AED Training Device has the same general external features as a rescue AED Note The AED Training Device uses special training pads only to demonstrate proper pad placement Training pads cannot deliver a shock or provide feedback The AED Training Device may be configured for semi automatic or automatic operation Refer to the appropriate rescue AED User s Guide for detailed descriptions of parts of a rescue AED 70 01124 01 D 5 AED Training Device Overview AED Training Device Remote Control Functions The remote control has the following functions Table 1 AED Traini
2. Pads in the same area with other AED equipment To avoid confusion when referring to the AED Training Device and other devices in this manual the following terms are used Rescue any device or accessory intended to deliver a therapeutic shock includes the word rescue for example the Powerheart G5 Automatic is a rescue AED Training any device or accessory used to demonstrate AED functionality includes the words training or Training Device for example AED Training Pads cannot deliver a therapeutic shock 70 01124 01 D 3 Supported Models Supported Models This AED Training Device approximates the function these Cardiac Science Corporation AED models Powerheart G5 Automatic Powerheart G5 Semi automatic Additionally the Training Device supports branded and other models that have the same functionality as the listed models Intended Use The AED Training Device and this manual are intended for use by qualified instructors with the following competencies and experience American Heart Association or European Resuscitation Council or their equivalents cardio pulmonary resuscitation CPR certification and first responder defibrillation training A thorough understanding of operation and use of supported rescue AED models and accessories For detailed operating instructions for a specific rescue AED model refer to the User s Guide for that model Learning
3. and Compressions Only 3 Press the Shock button to select and save the CPR prompt style 70 01124 01 D 9 Rescue Scenarios Metronome Option 1 Select the CPR prompt style The Training Device displays the Metronome option 2 Press the left soft key to scroll through the Metronome option ON or OFF Note If the Prompt Level is set as Basic or Standard and the CPR Prompt style is Compressions Only the metronome is set to ON 3 Press the Shock button to select and save the Metronome option CPR Device Prompt Option 1 Select the Metronome option The Training Device displays whether or not the CPR Device prompt option is ON or OFF 2 Press the left soft key to change the CPR Device prompt option 3 Press the Shock button to select and save the CPR Device prompt option 4 The Training Device restarts with the new setting Rescue Scenarios This section describes a typical rescue scenario and then lists the pre set rescue scenarios To change scenarios refer to AED Training Device Remote Control Functions on page 5 Note Unlike a rescue AED the AED Training Device does not store rescue information 70 01124 01 D 10 Powerheart G5 AED Training Device User s Guide Typical Rescue Scenario The following steps demonstrate a typical rescue scenario Once the AED Training Device is set and the lid is opened the AED Training Device advances through the complete scenario The instructor may speed u
4. device prompts A shock is not delivered and enters CPR mode If the scenario defines a non shockable rhythm the AED Training Device prompts If needed perform CPR as instructed Recommended Rescue Scenarios Note If simulating an automatic AED instead of prompting for shock the AED Training Device counts down and then automatically delivers a shock Table 2 Rescue scenarios Step Scenario 1 Scenario 2 Scenario 3 1 Open the Training Device lid 2 Place pads on manikin 3 The Training Device performs analysis 4 No shock advised Shock advised No shock advised 5 30 chest compressions and 2 breaths Prompt to shock Semi auto or shock delivered Auto 30 chest compressions and 2 breaths 6 Second analysis 30 chest compressions and 2 breaths Second analysis 7 No shock advised Second analysis Shock advised 8 30 chest compressions and 2 breaths Shock advised Prompt to shock Semi auto or shock delivered Auto 9 Third analysis Prompt to shock Semi auto or shock delivered Auto 30 chest compressions and 2 breaths 10 No shock advised 30 chest compressions and 2 breaths Third analysis 70 01124 01 D 12 Powerheart G5 AED Training Device User s Guide Maintenance The AED Training Device has no user serviceable parts The only maintenance required is to Clean external surfaces with a damp cloth Replace batteries in the AED Training Device
5. or remote as needed The AED Training Device requires three C cell alkaline batteries The remote control requires two AAA alkaline batteries When replacing batteries ensure that the polarity is correct For any other maintenance or repair problem call Cardiac Science Technical Support or outside the United States your local representative 11 Third analysis Shock advised 12 Shock advised Prompt to shock Semi auto or shock delivered Auto 13 Prompt to shock Semi auto or shock delivered Auto 30 chest compressions and 2 breaths 14 30 chest compressions and 2 breaths Table 2 Rescue scenarios continued Step Scenario 1 Scenario 2 Scenario 3 70 01124 01 D 13 FCC Notice to United States Users FCC Notice to United States Users This device complies with part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept any interference received including interference that may cause undesired operation See 47CFR Sec 15 19 3 Specifications The following table lists AED Training Device specifications Table 3 Specifications Parameter Detail Dimensions Height 8 cm 3 3 in Width 24 cm 9 5 in Depth 32 cm 12 5 in Operating and Standby Conditions Temperature 0 C to 50 C 32 F to 122 F Humidity 5 to 95 non condensing Shipment Conditions Up to
6. 1 Powerheart G5 AED Training Device User s Guide 70 01124 01 D COPYRIGHT 2014 CARDIAC SCIENCE CORPORATION ALL RIGHTS RESERVED Use the Automated External Defibrillator AED Training Device for teaching basic AED rescue skills and correct defibrillation procedures Table of Contents Package Contents 2 Safety 2 Supported Models 3 Intended Use 3 Learning Objectives 3 AED Training Device Overview 4 Training Device Setup 7 Rescue Scenarios 9 Maintenance 12 FCC Notice to United States Users 13 Specifications 13 70 01124 01 D 2 Powerheart G5 AED Training Device User s Guide Package Contents In the AED Training Device box you should find the following Powerheart G5 Training Device 1 Three C cell batteries are required not included Adult trainer pads 1 pair Infra red remote control 1 Two AAA batteries are required not included Language Packs and User s Guide CD Overlay for an automatic version Language configuration instructions Safety Before using the Training Device be aware of the following WARNING Patient injury or death The AED Training Device is not a rescue device and cannot deliver defibrillation shock therapy The AED Training Device is for training or demonstration only To avoid confusion do not store the AED Training Device or AED Training
7. d on the screen 2 Press the left soft key on the Training Device to scroll through the language selections default or alternate 3 When the desired language is displayed press the Shock button 10 CPR feedback Press 10 then 1 to prompt Press harder Press 10 then 2 to prompt Press softer Press 10 then 3 to prompt Press faster Press 10 then 4 to prompt Press slower Note Turn ON CPRD option on the configuration menu before performing this operation Table 1 AED Training Device Remote Buttons continued Button Use 70 01124 01 D 8 Powerheart G5 AED Training Device User s Guide Operating Mode To set the mode 1 Select the language The Training Device displays the current operating mode 2 Press the left soft key to display the mode selection Automatic or Semi automatic 3 Press the Shock button to select and save the mode Prompt Level To set the prompt level 1 Select the operating mode The Training Device displays the current prompt level 2 Press the left soft key to scroll through the prompt levels Basic Standard and Advanced 3 Press the Shock button to select and save the prompt level CPR Prompt Style To set the CPR prompt 1 Select the prompt level The Training Device displays the current CPR prompt style 2 Press the left soft key to scroll through the CPR prompt styles Traditional 30 2
8. ng Device Remote Buttons Button Use 1 Shock cancelled Press while the trainer prompts Shock Advised to change the prompt to Rhythm changed Shock cancelled Press during Place Pads prompts to forward to the Analysis prompts The next sequence Shock or CPR is determined by the selected scenario Press during a CPR sequence to forward to the Analysis prompts followed by another CPR sequence 2 Shock advised Press during Place Pads prompts to forward to the Analysis prompts The next sequence Shock or CPR is determined by the selected scenario Press during a CPR sequence to forward to the Analysis prompts followed by a Shock sequence 3 Check pads Press during analysis to prompt Press pads firmly to patient s bare chest Press again to continue the scenario 4 Place pads After the prompt Remove white square package from lid of AED press to prompt Peel one of the white pads completely from blue plastic Firmly place the pad on the patient Next peel second white pad from the blue plastic Firmly place the second pad on other location Note Press to advance through the prompts more quickly 70 01124 01 D 6 Powerheart G5 AED Training Device User s Guide 5 Analysis interrupted Press during Analysis prompts to prompt Analysis interrup
9. one week Temperature 40 C to 65 C 40 F to 149 F Humidity 5 to 95 non condensing Battery AED Training Device Three C cell alkaline batteries Remote control Two AAA alkaline batteries Emissions Radiated E M EN 55011 Group 1 Category B with Radio frequency CISPR 11 Class B Group 1 Immunity Radiated E M EN 61000 4 3 Radiated M EN 61000 4 8 ESD EN 61000 4 2 70 01124 01 Cardiac Science Corporation N7 W22025 Johnson Drive Waukesha WI53186 1856 USA 1 262 953 3500 US toll free 1 800 426 0337 Fax 1 262 953 3499 Email care cardiacscience com Orders and Customer Care US and International 1 262 953 3500 US toll free 1 800 426 0337 Fax 1 262 798 8292 care cardiacscience com Technical Support US toll free 1 800 426 0337 Fax 1 262 798 5236 techsupport cardiacscience com International internationalservice cardiacscience com Cardiac Science the Shielded Heart logo Powerheart and Rescue Ready are trademarks of Cardiac Science Corporation Copyright 2014Cardiac Science Corporation All Rights Reserved 14
10. p the scenario if required 1 Open the lid of the AED Training Device The AED Training Device prompts Stay calm and follow these instructions and ensure that Emergency services are called now 2 Press 4 on the remote to advance through the pad placement prompts Note The prompts below are provided when the Standard Prompt level is selected Tear open white package across dotted line and remove pads Peel one of the white pads completely from the blue plastic Firmly place the pads without the blue plastic on patients bare chest exactly as shown on the pads Next peel the second white pad from the blue plastic Firmly place the second pad on the other location exactly as shown on the pads 3 Press 4 again to indicate pad placement 4 The AED Training Device advances to the analysis stage of the rescue sequence and prompts Do not touch patient Analyzing rhythm 5 If a shockable rhythm is selected the AED Training Device prompts Shock Advised 6 Next follow these steps to deliver a simulated shock Note Following steps are applicable only for Semi automatic mode a The AED Training Device prompts Press red flashing button to deliver Shock b The student presses the Shock button to deliver the first simulated defibrillation 70 01124 01 D 11 Rescue Scenarios If the Shock button is not pressed after 30 seconds the
11. ted Stop patient motion Press again to continue 6 Battery low Press during a rescue scenario to prompt Battery low the battery LEDs will also show low battery Press again to turn off the low battery LED and turn on two green battery LEDs 7 Service required Press to prompt Service required Press again to continue 8 Pause Resume Press at any time to pause the AED Training Device beeps while paused Press again to continue Note If the device is already in Pause mode because a scenario sequence finished the AED Training Device restarts 9 Change scenario Change volume Press then 1 to start Scenario 1 Press then 2 to start Scenario 2 Press then 3 to start Scenario 3 Press then 7 to decrease volume Press then 8 to increase volume Table 1 AED Training Device Remote Buttons continued Button Use 70 01124 01 D 7 Training Device Setup Training Device Setup The Training Device has several features that can be tailored to the needs of the instructor Language Selection Note This is only for devices with dual languages installed To change the language of the voice and text prompts 1 Open the Training Device lid and immediately press and hold the Shock button for 5 seconds The Shock button remains lit and the current language is displaye