1. SuperCheck2 Blood Glucose Monitoring System Rev date 06 2011 6277AP MO1A2 Explanation of symbols Do not reuse Use by Expiry date Eee Lot number IVD In vitro diagnostic medical device 86 F ac ee ni Temperature limitation Caution refer to accompanying documents AN Please refer to safety related notes in the manual accompanying this instrument 1 Consult instructions for use Keep away from sunlight Pn AN Authorised representive in the european community Manufacturer Important Notice to Reader of this Manual This manual is written for the Btm SuperCheck2 meter model 6277 Voice Meter The Btm SuperCheck2 Model 6277 contains a speaking function that provides audible messages to help the user for using the meter and hearing the test results In the manual you will read many audible sentences which is only for the model 6277 If you don t want to hear any sentences from meter please refer the section 8 Settings and you can adjust the meter to Voice Off 1 Contents Section Topic Page Table of Contents Introduction 95 Appearance and Key Functions of the meter Alternate Site Testing 5 Safety Information Quick Testing Instructions 1 DAnbRWNe CUOMO RWHE Preparing for Testing 7 Installing the Battery 11 8 Settings Date Time Alarm 12 16 Memory Deletion Voice volume 9 Running a Control Solution Test 17 1
2. January 2007 volume 30 Suppl 1 S42 S47 2 American Diabetes Association Diabetes Forecast website information http www forecast diabetes org diabetes 101 hyperglycemia http www forecast diabetes org diabetes 101 hypoglycemia 31 nal BiotesT Medical Corporation No 3 2 Chien kuo Road TEPZ Tantzu 427 Taichung Taiwan ROC Tel 886 4 25326668 Fax 886 4 25326528 Obelis s a Bd G n ral Wahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 732 60 03 E mail mail obelis net C o537
3. The glucose level is number unit and Goodbye when meter turns off Important 1 Never reuse a test strip or a lancet An E_5 error message will be displayed in the center of screen if a used strip is inserted into the meter You will hear The strip is wet or used 2 Discard the test strip and lancet immediately ina puncture proof container with a lid after use 21 11 Recalling the Memory and Viewing the Average A Using the Meter Memory Your blood sample test result will be automatically stored in the meter s memory The meter can store up to 500 of the most recent test results You can also view the average of test results for various periods of 7 14 28 60 and 90 days The control solution test results will not be stored in the memory Note Please make sure that you press C key when you run a Control Solution test so that the control test result will not be added to the memory B Recalling the Test Results from the Meter Memory 1 With meter off press M key to turn on the meter After a brief full screen display and you will hear Recalling the memory the meter will display the total number of test data stored in memory Then the screen will display 001 as the sequential number of the most recent test result in memory The memorized test data will be displayed next If there is no memorized result in the meter is displayed and you will hear Zero memory The meter
4. key You will hear The average for 7 day is number unit 3 Press the M key and hold for only 2 seconds to exit the recall mode and turn off the meter or the meter will automatically power off after 60 seconds You will hear Goodbye 4 When is displayed showing that there are no test results in the memory You will hear Zero memory Calculations The 7 days average is the average of the last 7 day test results The 14 days average is the average of last 14 day test results The 28 days average is the average of last 28 day test results The 60 days average is the average of last 60 day test results The 90 days average is the average of last 90 day test results Note 1 HI LO results are not stored in the memory 2 Do not insert the strip into the meter when you want to recall the test results 23 12 Expected Values Reference Blood glucose levels normally will vary from time to time depending on food intake medication dosages health stress or exercise Consult your physician or healthcare professional for the target glucose value appropriate for you Expected plasma blood glucose values for normal nondiabetic adults are as follows Before eating lt 5 6 mmol L Two hours after meals lt 7 8 mmol L BTM SuperCheck2 meter gives plasma equivalent results 13 Transmitting Results Note Btm SuperCheck2 Glucose Meter allows you to transfer the test results stored in its memory to your pe
5. meter will turn off automatically and you will hear Goodbye and the LCD screen will display OFF 5 Safety Information Please use this device only for the intended use described in this user guide Before using this system to test your blood glucose please read the instructions carefully and practice the test when you first use this system Do a quality check on the system following the instructions and consult with your healthcare professionals for questions or problems Be aware of the safety of young children or handicapped persons near you when you conduct a glucose test using this system Please keep the test strip vial away from children The test strips and vial cap can present a choking hazard Please be cautious when removing the lancet Take the lancet out carefully Always place the protective cap back on the exposed tip or if using a twist lancet place the cap ona hard surface and push the exposed tip into the protective cap and discard of properly Never try to disassemble the meter in any circumstance If your meter is not working properly or if you need the Technical Support please contact your local distributor for help 6 Quick Testing Instructions For detailed instructions start on page 18 for Control Solutions and page 20 for Blood Testing 1 With the meter turned off insert a new test strip into the meter and you will hear Hello the Meter is ready A blinking blood
6. press M key to confirm the minute setting If 25 minute is selected and you will hear The minute is twenty five You will see the measuring unit flashing on the upper left corner of the LCD display Each time you press C key to adjust the hour and you will hear The hour is xxx Each time you press the C key to adjust the minute and you will hear The minute is xxx At the end of this process and you will hear Alarm Reminder Set 14 7 Delete Memory Press C key to toggle between a flashing dEL for memory deletion or a non flashing dEL If you Mem SET 7 want to delete all memory press M key while AN amp z 4 the dEL is flashing If you do not want to delete all memory press M key when the dEL is not flashing Once the M key is pressed the LCD display will move to the setting of voice volume setting mode You a qi will see the display of VoL in the center of screen When the dEL is flashing on the screen and you will hear Ready to delete memory records 7 8 Set the Voice Volume Press C key to toggle between the selection of speaking mode and non speaking mode If you want to turn off the speaker press M key when the screen does not show the small speaker icon You will hear Voice Off If you see a small speaker icon appears below the battery icon on the right side of LCD screen you can press M key to turn on the speake
7. Wash your hands in warm soapy water Be sure to rinse and dry well 19 6 Open the test strip vial Take one new test strip out of the vial and recap the vial quickly and firmly 7 While the meter is off insert anew test strip into the meter s strip slot with the contact bars facing upwards The meter will turn on with a full screen display and you will hear Hello the meter is ready A blinking blood iconwill be displayed on the screen You will hear Apply blood now If no action is taken in 2 minutes the meter will turn off automatically You will hear Goodbye 8 Use the adjustable lancing device or the disposable lancet to take a blood sample from the fingertip and let it form a small round blood drop To avoid body fluid contamination it is recommended that you do not use the first blood drop for glucose testing 20 9 Apply a small blood drop to the 9 front edge of the test strip and the blood should be pulled into the confirmation window before the meter begins to count down Do not push your finger against the test strip or try to apply a smeared blood sample You will hear Testing in process 10 The meter will start counting down for 6 seconds and your test result will be displayed on the screen Remove the used test strip by hand or by pushing the ejector button The meter will turn off and the test result is stored auto matically You will hear
8. affect blood glucose levels Capillary whole blood at the fingertip can reflect test changes faster than capillary blood at other sites of body If you test your blood glucose level during or immediately after a meal physical exercise or stressful events take the other sites of the body blood sample from your fingertip instead of from other sites Use AST only 1 2 hours or more after taking insulin 2 2 hours or more after a meal 3 2 hours or more after exercise Do not use AST if you are pregnant or if you are aware that your glucose level is not as stable as usual or if you think you have hypoglycemia low blood sugar or hyperglycemia high blood sugar Do not use AST if you think your blood glucose is low and if your AST results do not match the way you feel 5 Testing Procedure for AST 1 Please unscrew the lancing device by turning the end cap counter clockwise A clear cap will come with the package of a meter kit or a lancing device The clear cap on the lancing device will make it easy for you to get a drop of blood for AST 2 Insert a new lancet firmly into the lancet holder 3 Twist off the protective top of the lancet 4 Close the end cap of the lancing device and grasp the lancing device as shown Slide into locking position If necessary please set the lancing device for a deeper puncture 5 Select a soft fleshy area on your forearm that is clean and dry away fr
9. icon will be displayed on the screen You will hear Apply blood now 2 Lance the finger and let a blood drop form 3 Apply the blood drop to the front edge of the test strip when the blood drop icon is still blinking on the meter Wait for 6 seconds and the meter will display the test result You will hear Testing in process and The glucose level is number unit 4 Remove the used strips by hand or by pushing the ejector and the meter will power off with a display of OFF on the screen You will hear Goodbye 10 7 Installing the battery 1 Turn off the meter by pressing the M button and hold for 2 seconds before replacing the batteries Slide down the battery cover on the back of the meter by pushing the cover in the direction of the arrow and put the cover aside 2 Remove the two used batteries 3 Insert 2 new batteries with correct direction Battery AAA 1 5V 2 Alkaline 24A LRO3 4 Put the battery cover back in place by pushing the cover in the opposite direction of the arrow to close the cover into position Note 1 It is suggested that batteries need to be replaced when the battery icon appears empty If you keep using the meter the battery icon will start flashing and an error message E_3 will appear on the screen 2 Note that replacing the battery will not reset the stored testing results 3 You need to reset the time and date after
10. if there is no strip inserted Traceability of glucose monitoring system The accuracy of Btm SuperCheck2 Blood Glucose System was assessed by comparing blood results obtained by patients with those obtained using a YSI Model 2300 Glucose Analyzer which is calibrated by YSI calibrator solution refer to NIST SRM standard reference material 917A Clinical Dextrose The device has certified to meet the following standard 98 79 EC EN60601 1 EN 60601 1 1 ISO 15197 P S Please refer to the package insert that came with your test strips and you will also find information on how the system works and on the performance characteristic reference 19 Ordering Supplies To order supplies please contact the Manufacturer Distributed By Apollo Medical Technologies Ltd Tele 01636831201 Email apollomedicalltd1 btinternet com Website www apollomedicaltechnologies com 39 20 Warranty Information Your Btm SuperCheck2 Glucose Meter is warranted to be free of defects in materials and workmanship for one year from the date of the original purchase This warranty does not cover device failure due to owner misuse or negligence or normal wear and tear If you have a question about your Btm SuperCheck2 Glucose Meter or this warranty please contact Biotest Medical Corporation No 3 2 Chien kuo Rd TEPZ Tantzu Taichung 427 Taiwan R O C E mail info mail biotestsystems com Reference 1 American Diabetes Association Diabetes Care
11. the batteries are replaced 4 If there is any unexpected symbol on the display please follow the above procedure to replace the batteries again 5 Please dispose of batteries according to your local ordinances 8 Settings Date Time Alarm Memory Deletion Voice Volume The LCD screen will show as below 1 Measurement Units 2 Memory mmol L item SET_ 3 Setting jal ri ears flag 5 Blood drop icon eb 6 Battery Symbol ALI AA 8 Day Average ___ __ 7 Voice Symbol 9 Alarm AVG nd l pees icon Reminder 19 9 68 9 P 10 Month 13 Minut 11 Day U a pic 1 Appears with the test result or in mmol L 2 Appears when you recall the memory 3 Appears when you are in setting mode 4 Appears for control solution test flag c 5 Indicates the meter is ready to take the blood sample when it flashes 6 Indicates the battery status 7 Indicates if audio function is on or off mmol L 8 Indicates current displayed result is an average x 9 Appears when alarm is on L m e 5 10 Month q 11 Day 3 23 0845 12 Hour i The LCD screen for Power Off When the meter is off the date time 13 Minute measurement unit battery icon and OFF will be still on the screen If the voice is on or the alarm reminder is on the speaker icon and the alarm icon will be displayed as well 1 Set the Year Press C key to adjust the year until the desired year is displayed and then press th
12. 8 Blood Testing 10 Running a Blood Glucose Test 19 21 11 Recalling the Memory and Viewing the Average 22 23 12 Expected Values 24 13 Transmitting Results 24 Maintenance 14 Maintenance of Your System 25 15 Troubleshooting 26 27 16 Customer Service 28 17 Limitations 29 Important Additional Information 18 Specifications 30 19 Ordering Supplies 30 20 Warranty Information 31 NOTE For information about the name of the manufacturer of the lancing device and the lancets please refer to the package 1 inserts that came with your starter kits 2 Introduction to the System Intended Use Indications for Use The Btm SuperCheck2 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the forearm It is intended for use as an aid by either a healthcare professional or a patient diagnosed with diabetes mellitus for monitoring their blood glucose levels when self testing The Btm SuperCheck2 Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus nor for use with neonates The alternative site testing forearm in this system can only be used during steady state blood glucose conditions Contents of the System This User Guide has been prepared to describe the proper use of the Btm SuperCheck2 Blood Glucose Monitoring System Please read this User Guide carefully including the package inserts before test
13. ause incorrect results Uric acid gt 0 65 mmol L acetaminophen gt 0 41mmol L ascorbic acid gt 0 26 mmol L ibuprofen gt 1 82 mmol L 6 The test strips may be used at altitudes up to 5 280 feet 1 609 m without an effect on test results 7 Persons suffering from severe dehydration should not be tested using a capillary whole blood sample 8 Test results below 3 9 mmol L indicate low blood glucose hypoglycemia Test results greater than 13 3 mmol L indicate high blood glucose hyperglycemia If you get results below 3 9 mmol L or above 13 3 mmol L repeat the test and if the results are still below 3 9 mmol L or above 13 3 mmol L please consult your healthcare professional immediately 9 Inaccurate results may occur for individuals experiencing a hyperglycemic hyperosmolar state with or without ketosis Critically ill patients should not be tested with a blood glucose meter 29 18 Specifications Test Glucose Sample Whole Blood Principle of the test method Amperometric glucose oxidase Test sites Fingertip forearm Measurement time 6 seconds Code required No code device Measurement Range 1 1 33 3mmol L Batteries 2x AAA 1 5 V Alkaline 24A LRO3 Operating Temperature 10 40 C 50 104 F Humidity 20 80 RH Width 48mm Length 99mm Thickness 14mm Weight 72g Memory 500 measurement results with date and time Auto power off after 2 minutes without action whena strip is inserted or 60 seconds
14. e M key to confirm the year setting When pressing the C key the year will be adjusted in the range of 2009 to 2029 After the year setting is confirmed you will see the month segment flashing on the screen Each time you press the C key to adjust the year you will hear The year is 20xx If year 09 is selected and you will hear The 1 year is 2009 2 Set the Month Press C key to adjust the month until the desired month is displayed and then press the M key to confirm the month setting You will then see the day segment is flashing on the screen Each time you press the C key SET to adjust the Month you will hear The month is xxx If December is selected and D you will hear The month is December E 3 Set the Day Press C key to adjust the day 59 until the desired day appears and then press wu theM key to confirm the day setting You will 2 0 ml D AM LCD screen Each time you press the C key to adjust the day you will hear The day is xxx If 18 is selected and you will hear The day then see the hour segment is flashing on the NGANG a na 2 SET is the eighteenth E wu NE BAKO an 3 13 SET 4 Set the Hour Press C key to adjust the hour x until the desired hour appears on the screen and then press the M key to confirm the setting You will then see the minute segment is flashing on the screen Each time you press the C key to adjust the Hour y
15. em to another test system which is well maintained and monitored by a healthcare provider 25 15 Troubleshooting The following is a summary of all the Error Messages These messages help to identify certain problems but do not appear in all cases when a problem has occurred Improper use may cause an inaccurate result without producing an error message or a symbol In the event of a problem refer to the information in the table under Solution Error messages E_1 The temperature is too low E_2 The temperature is too high E_3 Battery Low E_4 Memory damaged E_5 The strip is wet or used E_6 Error in meter or strip E_7 The blood sample is not sufficient HI The glucose level is too high LO The glucose level is too low 26 The meter displays an E_2 error message Problem Possible Cause Solution 1 A The meter is operating Repeat the test after the The meter displays temperature below meter and strip are placed in an E_1 error 10 C or 50 F which is 4 Warmer environment and message the lowest range of allow the meter to warm up operating temperature for a while before retesting 2 The meter is operating Repeat the test after the temperature above 40 C or 104 F which ig the cap of operating temperature meter and strip are placed in a cooler environment and allow the meter to cool down for a period before retesting an E_3 error message 3 The meter display3 The bat
16. er and you will hear Hello the Meter is ready A blinking blood icon will be displayed on the screen You will hear Apply blood now 2 Press the C key and the meter will display a letter C on the upper right corner of screen indicating the control test mode You will hear Apply Control Solution now The test result will not be stored in the memory 3 Squeeze a small amount of Control Solution ona flat surface and apply to the front edge of the test strip and you will hear Testing in process 4 Wait for 6 seconds and the meter will display the result You will hear The glucose level is number unit 5 Remove the used strips by hand orby 5 pushing the ejectorand the meter will O power off with a display of OFF on the 4 screen You will hear Goodbye 17 IMPORTANT 1 You must press the C key to distinguish the control solution test from the blood glucose test Do not perform the blood test in Control Solution mode If you perform the blood test in Control Solution mode the test result will not be stored in the memory 2 Use only Btm SuperCheck2 Control Solution with Btm SuperCheck2 Glucose Meter 3 Btm SuperCheck2 Control Solutions are used to check that the meter and the test strips are working together as a system and that you are performing the test correctly 4 Btm SuperCheck2 Control Solution contains a known amount of gluco
17. g mode please refer to manual for detailed function description 5 Speaker Where voice comes out 6 RS232 port Cable connection and data transmitting 7 Battery Compartment Where batteries are located 8 Ejector Remove used strip If your meter has the back light feature this will come on automatically during every operation to help user readability and to assist with the smooth operation of your meter The principle of the method When glucose reacts with the reagents on the test strips an electrical current is produced which is proportional to the glucose concentration in the blood sample The glucose concentration is calculated by the meter and based on the current measured 4 Alternate Site Testing AST What is AST AST is the sampling from anatomical sites parts of the body other than the fingertip i e forearm to check the blood glucose levels This system allows you to test on the forearm with the equivalent results to fingertip testing There are important limitations to AST Please consult your healthcare professional before you use AST What is the advantage Some patients may encounter discomfort when taking a blood sample from the fingertip as the finger tips have a high concentration of nerve endings Other body sites do not have as many nerve endings so you will not feel as much pain as at the fingertip When to use AST Medication stress illness food and exercise can
18. ing The system is available as either a meter or as a complete starter kit If you have a meter only you can purchase any additional spares from your system provider The Btm SuperCheck 2 Blood Glucose Monitoring System contains the following items 1 SuperCheck2 Glucose Meter with 2 AAA batteries 2 User Guide 3 Log Book 4 Carrying Case 5 SuperCheck2 Glucose Test Strips 6 SuperCheck2 Control Solution Medium 7 Sterile Lancets 8 Lancing Device 9 Package Inserts for Test Strips Control Solution and Lancets Lancing Device A single level control solution Medium is provided in the Starter Kit You may purchase any control solutions that you need from your system provider The Btm SuperCheck2 Glucose Meter uses Btm SuperCheck2 Glucose Test Strips Neither the meter nor the test strips will work when used with any other brand of glucose products 3 Appearance and Key functions of the meter GO 6 1 mmol L Mem SET 2 4 fe Oo 3 T 7 R 5 Ld 1 Test strip slot When the strip is inserted into the 8 slot the meter will automatically turn on 2 LCD Display Guide you through the test using symbols and simple messages 3 M Key Power ON OFF also for memory recalling mode please refer to manual for detailed function description 4 C Key Settin
19. om bone and free of visible veins UA and hair 6 Massage the selected area gently to increase blood flow to the puncture site Clean the test location with an alcohol wipe or with soap and water 7 Open the test strip vial Take one test strip out of the vial and quickly cover the vial to keep air out 8 Insert a test strip into the meter while the meter is off The meter will turn on automatically and you will hear Hello the Meter is ready A blinking blood icon will be displayed on the screen You will hear Apply blood now If no action is taken in 2 minutes the meter will turn off automatically You will hear Goodbye 9 Press and hold the loaded lancing device against the forearm for a few seconds then press and release the button 10 Wait for a few seconds until the blood drop forms Make sure you have sufficient blood to fill the small window of the test strip For individuals who experience difficulty in getting sufficient blood for a test it may be helpful to rub the puncture site a little longer before using the lancing device 11 Apply your blood to the front edge the tip of the test strip and the meter will start counting down for 6 seconds then display the test results on the screen You will hear Testing in process and The glucose level is number unit 12 Remove the used strip by hand or by pushing the ejector button The
20. or for advice immediately 9 The meter displays LO on the screen The test result is lower than the measuring range of 1 1 mmol L Test again following the user guide for correct glucose measurement process If you see LO again please call your medical doctor for advice immediately 27 16 Customer Service If you need assistance with your Btm SuperCheck2 Glucose Monitoring System please contact Distributed By Apollo Medical Technologies Ltd Tele 01636831201 Email apollomedicalltd1 btinternet com Website www apollomedicaltechnologies com For questions related to your health condition please call your doctor or healthcare professional 28 17 Limitations 1 No Neonatal use Do not use for neonatal blood glucose testing 2 Hematocrit range Hematocrit in the range of 32 60 has been shown not to affect the glucose results If you do not know your hematocrit level consult your healthcare professional 3 Hemoglobin range Hemoglobin levels of 0 078 mmol L did not interfere with blood glucose test results 4 Elevated Cholesterol and Triglycerides Cholesterol levels up to 12 9 mmol L and Triglycerides up to 33 1 mmol L has been shown not to affect glucose results 5 Medications Interference was observed for therapeutic levels of L DOPA No interference was shown for uric acid acetaminophen ascorbic acid and ibuprofen in normal therapeutic levels However higher concentrations in blood may c
21. ou will hear The hour is xxx If 11 pm is selected and you will hear The hour is eleven PM 5 Set the Minute Press C key to adjust the minute until the desired minute appears on the screen and then press the M key to confirm the setting You will then see an AL displayed on the screen center Each time you press the C key to adjust the Minute and you will hear The minute is xxx If 25 minute is selected and you will hear The minute is twenty five 6 Set the Alarm Reminder Press C key to toggle between the alarm reminder setting or no setting mode If setting alarm is selected and you will hear Ready to set Alarm Reminder If you see the LCD display a small flashing icon of alarm clock in the lower left corner with a time segment display it means that you are in the alarm setting mode If you see only AL without the icon of the small alarm clock it means that you are not in the alarm setting mode Press M key to confirm the alarm setting mode If you select the alarm setting mode alarm clock icon will stop flashing and the hour of the time segment will start flashing You will hear Ready to set Alarm Reminder Press C key until the desired alarm hour appears and then press M key to confirm the hour setting If 8 AM is selected and you will hear The hour is eight AM Then the minute segment will start flashing Press C key until the desired minute appears on the screen and then
22. r Then you will see the existing voice volume level at the lower right corner of the screen There are 7 levels of voice volume You can press C key to adjust the voice level and then press the M key to confirm the setting There will be a beep sound to indicate the level of voice volume adjusted The meter will turn off automatically after 2 seconds with a display of OFF and you will hear Goodbye 15 9 Power off Screen When the meter is off the time date measurement unit and battery icon will be still on the screen as Figure 11 If the alarm reminder has been set or the voice has been turned on then the alarm or the speaker icon will appear on the power off screen as well The LCD screen will display OFF Note 1 When you change the date and time backward you will not change the test results in the meter memory 2 You need to move through the language year month day hour minutes alarm measurement unit dEL and VoL to save the meter settings and turn off the meter 3 The average readings in the meter memory are calculated from the results obtained during the 7 14 28 60 and 90 calendar days preceding the current date and time settings 16 9 Running a Control Solution Test IMPORTANT Always make sure you press C key for Control Solution test otherwise the control solution test result will be stored in the memory and affect your AVG results 1 Insert a strip into the met
23. rsonal computer However you will need to order software and a Data Download Cable separately from your distributor The software and its data download procedure will come as a CD in the download cable box You can also check with your distributor for the software download option The meter still keeps the results in the memory after transmitting With the meter off hold the M and C keys down at the same time for 5 seconds until PC is displayed You will hear Ready to transmit data Press the C key to start transmitting The meter will display a flashing PC during the transmission process When the PC on the screen stops flashing it indicates the transmission is completed The meter will be power off if no action is taken in 60 seconds or M key is pressed for 2 seconds You will hear Goodbye N 24 14 Maintenance of your system Please treat this meter with proper care and keep it in good condition 1 Store your meter in the protective case ina clean dry place at 46 86 F 8 30 0 2 Always clean your meter after use Wipe and clean the surface of the meter with a soft cloth that has been slightly dampened with mild detergent 3 Please handle with care and do not drop the meter 4 If this meter is used by healthcare professional infection control policies should be strictly followed 5 The meter has a shelf life of 2 years We suggest you should periodically compare the test syst
24. se that reacts with test strips The Medium and High level Control Solution are intended to check the monitoring system in different measurement ranges 5 Shake the vial discard the first drop of control solution and wipe off the dispenser tip to ensure a good sample and an accurate result 6 Use only for 3 months after first opening Record the open date on the Control Solution vial Discard after 3 months 7 Compare your Control Solution test results with the expected range printed on the test strip vial label If your glucose control results fall outside the expected range repeat the test Results that repeatedly fall outside the expected range may indicate 1 You may not be doing the test correctly repeat the test by shaking the control solution vial well and carefully following instructions Please make sure the test is done within the temperature range 2 The Control Solution is expired or contaminated 3 The test strips are damaged or outdated 4 You may have applied the Control Solution before the blood drop appears on the screen This will cause incorrect glucose measurement 5 Meter malfunction 18 10 Running a Blood Glucose Test 1 Please unscrew the lancing device by turning the end cap counter clockwise 2 Insert a new lancet firmly into the lancet holder 3 Twist off the protective tip of the lancet 4 Close the end cap of the lancing device as shown Slide into locking position 5
25. tery is too low to operate the meter Please replace the two AAA batteries The meter can perform the The meter displays an E 5 error 4 5 The memory chip glucose measuring without The meter displays of the meter could storing the test results into an E_4 error be damaged or the meter s memory Please message malfunctioning write down the test results into the data logger before you call a customer service representative 5 1 If you apply the Please check the test strip to sample before blood drop icon appears on the screen see if it is damaged or used In either case please discard the strip and repeat the test HI on screen message h A i 8 2 The inserted test strip Using a new strip has been wet or used 6 The meter is not Repeat the test with anew The meter displays working properly test stip IfE_6 continues an E_6 error defective meter or a aa R er a message a gt call a customer service a defective strip representative I J The blood sample is Please re test by inserting a e meter displays ici new test strip ensuring an Th ter displays not sufficient test strip 8 an E_7 error adequate amount of blood message comes into contact with the test strip 8 The test result is Test again following the user The meter displays higher than the guide for correct glucose measuring range of 33 3 mmol L measurement process If you see HI again please call your medical doct
26. will turn off after 60 seconds if no action is taken When the meter displays the first memorized data you will hear The glucose level is number unit 2 The meter will audibly read out the last 10 stored test results with date The screen will display each of the memorized test data in sequence Press M and C key to review your test results in the memory forwards and backwards 3 To turn off the meter press the M key for 2 seconds or the meter will turn off automatically after 60 seconds if no activity and you will hear Goodbye 22 C Viewing the Average AVG of Test Data in Memory 1 To display the average AVG test result please turn the meter off Press C and M keys at the same time and hold for about 2 seconds wait for the AVG to be displayed on the screen then release the two buttons The meter will be in the average mode with AVG font displayed 2 When the AVG is flashing at the lower left corner of the screen the number 7 will be displayed under AVG and the average of the last 7 days test results will be displayed in the center of the screen The number of tests completed in the last 7 days will also be displayed in the lower right hand corner of the screen If you want to see the average of 14 28 60 and 90 days keep pressing the M key to move the LCD display forward from 7 day average to 90 day average The average display will move from 90 days backward to 7 days if you continue to press C