Dual Low Air Loss User Manual

image

Contents

1. can be opened for maintenance Air Filter and Filter Cap 3 Shelden Healthcare recommends that the filter to be clean or replace once a month to ensure optimal performance of the power unit 6 0 Mattress and Power Unit Installations 1 Remove existing mattress from bed frame 2 Place Shelden Healthcare mattress replacement on the bed frame with the logo at foot end Secure the mattress at each side using anchor strap Please verify all bed functions are working properly without interference before proceeding to the next step 3 Secure the power unit onto the foot board using hanging hooks 4 Firmly connect the mattress air hose couplings to the CPC quick release couplers on the power unit s air outlet 5 Plug the power unit into a wall outlet and turn on the main power rocker switch on the left side panel The STANDBY CPR orange LED on control panel will light to acknowledge the power unit is in standby mode 6 Press the POWER button on the control panel to inflate the mattress system MAX INFLATE may be used to shorten the initial inflation time if needed 7 0 Program Settings 1 Center the patient on the mattress Adjust the mattress internal pressure according to the patient weight by using the weight kg button on the control panel of the power unit If the patient feels the mattress is too soft or firm increases or decreases the mattress internal pressure one increment at a time a
2. into wall outlet outlet Check if power surge shuts down A power surge may overload the power unit circuitry temporarily Turn the unit off and check the fuse for damage Turn the unit ON again with normal procedure Power unit does not respond to Please contact Shelden possible solutions Healthcare on 01788 833 989 Check if there is leakage in air tubes or air cells Please contact Shelden Healthcare on 01788 833 989 3 alert LED is lit during operation Check if mattress air hose couplers Secure air hose couplers firmly are properly connected to power unit s quick release couplers into place Inspect air filter for dust or dirt Clean or replace air filter Lift mattress coverlet up to check if Check and adjust air tubes air tubes are kinked or obstructed positions Please contact Shelden Healthcare on 01788 833 989 5 Alternating or static setting is Possible control failure not available Air tube is referring to the mean of directing air flow from the power unit to each individual air cell DUAL LOW AIR LOSS USER MANUAL 8 13 0 Returns for Service This device is not self serviceable Service and repair must be performed by an authorized technician or representative All returned device must be cleaned and disinfected prior to shipping Unsanitary or soiled systems will be returned without servicing 14 0 Warranty Shelden Healthcare warrants the product to
3. mode UPRIGHT mode is used to prevent patient from bottoming out in sit up position Make sure UPRIGHT mode DUAL LOW AIR LOSS USER MANUAL 4 wearttaraer Rear Panel Features Figure 5c mode is switch off when the patient is laying down to avoid incorrect pressure setting Press PANEL LOCK button to lock out function buttons besides max inflate and power functions from unauthorized changes If you wish to cancel panel lock simply press and hold the PANEL LOCK button for 3 seconds Control Panel Features Figure 5a max inflate Press the MAX INFLATE button to inflate the mattress rapidly to maximum pressure Press the button again to restore to the original setting or this function will disable automatically after 20 minutes if unattended Right Side Panel Features See Figure 5b Couplers 1 Quick release CPC couplers are used to secure mattress air hoses to power unit Left Side Panel Features See Figure 5b Rocker Switch 1 Power unit s main power switch The power unit automatically goes into STANDBY CPR mode after the switch is turned on Power Receptacle 2 Insert power cord firmly into receptacle Left Side Panel Features See Figure 5c Hanging Hooks 1 Hanging hooks are designed to hang thepower unit on almost any foot board Hooks are spring loaded and will fold against the power unit when not in use ad LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM Fuse 2 Fuse holder
4. state and or facility requirements for disposal of the unit 7 The use of this system does not replace the regular repositioning monitoring and nursing of the patient 8 Operating transportation and storage conditions Temperature 5 C 45 C Humidity 15 60 CAUTION ENSURE THAT THERE ARE NO PROTRUDING OBJECTS SHARP POINTS OR BED SPRINGS UNDER THE MATTRESS AS THESE COULD PUNCTURE THE AIR CELLS 4 0 System Package Power Unit Package Power unit x 1 Power cord x 1 Operation manual x 1 Mattress Package Mattress replacement unit with coverlet x 1 5 0 Features Control Panel Features See Figure 5a on page 5 power on led standby cpr led The system is in STANDBY CPR mode and the orange LED will light when the main power rocker switch on the left side panel is turned on Press the POWER button to turn on the system and the green LED will light Please refer to 9 0 Emergency CPR Deflation for CPR function patient weight Simply press the or button to adjust the patient weight according to each individual The scale is only an approximation Please adjust the weight setting if the mattress is too soft or firm to suit each patient mode STANDARD mode is for patient who weights in the range of 35 kg 140 kg BARIATRIC mode is for patient who weights in the range of 160 kg 450 kg Press and hold the desired mode button for 10 seconds to switch between Standard and Bariatric
5. 601 1 2 1993 Type BF Applied Part Maximum Weight Capacity 450 kg Shelden Healthcare reserves the right to change prices or design parameters and specifications without incurring any liability in connection with such modifications DUAL LOW AIR LOSS USER MANUAL 10 Shelden Healthcare Ltd Unit 1A The Wharf Stretton Under Fosse Rugby Warwickshire CV23 OPR Telephone 01788 833 989 Fax 0845 127 50 50 Email sales shelden healthcare co uk Web www pressurerelieving co uk
6. be free from defects from the date of purchase Please inspect all accessories when you purchase our product If there is any damage or missing accessories when you receive the product please ask for a replacement from your local distributor within three days of purchase The warranty periods for Shelden Healthcare products are according to the regulations in your country the minimum period is 1 year from date of purchase for the power unit and 1 year for the mattress and the coverlet The warranty coverage of any Shelden Healthcare product is contingent up on its purchase from an authorized dealer Warranty for any product in the Shelden Healthcare product line will be honored by the official distributor in your country Warranty coverage will not be extended to any product on which the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by Shelden Healthcare or if in the sole opinion of Shelden Healthcare that the system shows evidence of tampering abnormal or unreasonable abuse negligence accident or operation withoutregard for the restrictions specified in the instructions which accompany the system This warranty does not cover normal maintenance such as cleaning adjustment lubrication and updating of equipment or parts If the damage is result from improper operation a reasonable service fee and part cost will be charged The warranty state
7. bed linens to a single layer in order to allow the moisture escape efficiently through the coverlet Only breathable incontinent pads are recommended for use with this mattress system 3 Avoid using this device near open flames lighters or cigarettes Flammability hazard exists This device draws air from the surrounding Thus cigarette smoking may damage internal component 4 This system should disinfect thoroughly between patients to avoid cross contamination 5 Verify patient weight does not exceed bed frame bed rails and this mattress system s weight capacity 3 0 Warnings 1 Use this mattress with the correct side rails to ensure the gap between the top of the mattress and the top of the bed rails meets the EN60601 2 3 LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM 52 Failure to do so could result in serious patient Injury 2 DO NOT disassemble the power unit if you are not a qualitied technician Please contact your local distributor for all service 3 This product is NOT AP APG protected 4 Re position the patient once awhile is still necessary when using this mattress 5 Incorrect battery replacement may result in risk of explosion Replace only with same or equivalent battery type recommended by the manufacturer 6 If the equipment needs maintenance contact Shelden Healthcare as soon as possible For equipment that is no longer functional make sure to follow national
8. d above is the only warranty made and is in lieu of all other warranties whether expressed or implied including any warranty of merchantability or fitness for a particular reason Shelden Healthcare will not be liable for consequential or incidental damages or any kind 15 0 Product Specifications Mattress Dimension MSR208 36 200x90x20cm 39 200x99x20cm 42 200x107x20cm 48 200x122x20cm 54 200x137x20cm 60 200x153x20cm Number of Air Cells 20 air cells Power Control Unit Dimension SR363 29 5 x 22 5 x 12 0 cm Material Power Unit ABS fire retardant UL 9 LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM Mattress Cell 100 nylon with TPU lamination Coverlet 100 nylon with PU backing Power Input AC100 120V 50 60Hz5A AC220 240V 50 Hz 5A Power Consumption Normal Operation Max 350W Power Control Unit Weight 4 6 kg Power Cord HO5VV F3x1 00mm2 Fuse Rating T1A 250V Operating amp Transportation amp Storage Conditions Temperature 5 C 45 C Humidity 15 60 Atmospheric pressure 700 1060 hPA Electrical Classification Class Il Type B double insulated with or without functional ground wire IPXO do not immerse power unit in any liquid or spray any liquids directly on the power unit This system is not AP APG protected Continuous operation Agency Approval EN60601 1 EN60601 1 2 EN61000 3 2 EN61000 3 3 UL2601 1 Electromagnetic Compatibility EN60
9. nd wait for the system to stabilize before making another change until a comfortable state is achieved 2 Caregivers should always perform a hand check by placing their hands underneath patient s pelvis area to check if there is sufficient air support to ensure the patient is not bottoming out 3 IMPORTANT TUCKING THE SHEET IN TIGHTLY REDUCES THE EFFECTIVENESS OF THE SYSTEM 8 0 Patient Transfer and Transport Transfer It is recommended to have the mattress system in MAX INFLATE mode during transfer Make sure the bed is secured before proceeding Transport If a transport occasion comes up detach mattress air hose couplings from the power unit s CPC quick release couplers and connect the two air hose couplings together Shelden Healthcare Theraflo system has a safety foam base underneath the air cells to support the patient for a short period of time To resume normal operation please follow 7 0 Program Setting DUAL LOW AIR LOSS USER MANUAL l6 9 0 Emergency CPR Deflation In the case of emergency press the POWER button on the control panel to turn the system into STANDBY CPR mode to deflate the mattress rapidly in order to perform CPR Press the POWER button again and reset the patient weight to resume therapy 10 0 Cleaning Instruction Air mattress and power unit must be cleaned thoroughly between patients to avoid cross contamination The following is a suggested guideline but local infection con
10. shelden HEALTHCARE Dual Low Air Loss User Manual Specialist Suppliers to the Healthcare Industry 1 NV AT N DI TACCO NAAN VV A J f YK A VAAN IM WV _ n AN LUVV A BS _ VIA TRESS REPLACEMEN Thank you for purchasing this Shelden Healthcare pressure relieving mattress system PLEASE READ THIS USER MANUAL CAREFULLY BEFORE SETTING UP AND USING THE DEVICE Pay special attention to the warnings and other safety information Use genuine Shelden Healthcare components to maintain the products optimal performance If you do not fully understand all the instructions safety precautions and warnings do not use this device In case you have questions please contact Shelden Healthcare on 01788 833 989 SYSTEM Section 1 0 2 0 3 0 4 0 DU 6 0 7 0 8 0 Description Page Symbols and Indications Safety Precautions Warnings System Package Features Mattress and Power Unit Installations Program Settings Patient Transfer and Transport 03 03 03 04 04 06 06 06 Section 9 0 10 0 12 0 13 0 14 0 15 0 Description Emergency CPR Deflation Cleaning Instruction Routine Maintenance Troubleshooting Returns for Service Warranty Product Specifications YALL LOW AIR OSS LISER MANIJAI JUFAL LVVV AIK ZIJ UOLIA IV AIN AI shelden HEALTHCARE Page 07 07 07 08 09 09 09 1 0 Indications AY document thoroughly 0 Double ins
11. trol policies should be followed as well Mattress cleaning with hand hot water with a PH neutral detergent Decontamination by wiping with a Sodium Hypochlorite solution of 1000ppm in line with HSG9518 and BHTA guidelines Please ensure air mattress and cover are completely dry before patient use Machine wash warm water at maximum 71 C Dry mattress cover on a cool tumble setting Do not iron WARNING Always unplug the power unit before cleaning Routine cleaning of power unit can be done by wiping down with damp cloth using disinfectant and water or mild neutral detergent Never spray liquids directly on the unit itself 11 0 Routine Maintenance Remove air filter from the rear panel of the power unit by opening up the filter cap to clean or replace the air filter Inspect the filter for dirt or dust and clean it with mild soap and water Reinsert the dried air filter after cleaning and ensure the cap is secure Replace with genuine air filter once a year is recommended and compressor every two year is strongly suggested to prolong system lifetime Only disinfected and dry systems are to be stored Disconnect the air hoses from the power unit Roll up the mattress starting from head end and working down toward the foot end Use the straps to fix it 7 LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM 12 0 Troubleshooting 1 Power unit does not function Check if power cord is firmly Secure power cord into wall plugged
12. ulated Class ll C Declaration of Conformity to Medical Attention Please read enclosed Device Directive Afe International Electronics Approval IPX0 Do not immerse power unit in liquid or spray liquids directly on power unit Type B Equipment Indications This low air loss mattress system is designed for patients who are at high risk of pressure of ulcers or require treatment for existing pressure ulcers This device is intended to treat and prevent pressure ulcers by facilitating blood circulation reactive hyperaemia and decreasing interface pressures Always consult a physician or health professional before using this mattress system This pressure relieving mattress system is suitable for patients assessed to be high risk and up to catagory 3 pressure ulcers Contraindications Certain patient conditions are not suitable for using this type of device Contraindications such as unstable fractures of vertebrae and conditions of the vertebrae Always consult a physician or health professional before using this device The use of this system does not replace the regular repositioning monitoring and nursing of the patient 2 0 Safety Precautions 1 To ensure proper operation please inspect and verify all parts are setup properly and anchor securely Do not place anything on top of the power unit Make sure power cord is underneath bed frame and free from hazard 2 It is recommended to limit