Dual Cushion User Manual

image

Contents

1. 100 nylon with PU backing Foam base PU foam Power Input AC100 120V 50 60Hz AC220 240V 50 Hz 9 MOBILE AIR ALTERNATING PRESSURE CUSHION SYSTEM Power Consumption Normal Operation Max 6W Power Control Unit Weight 2 5 kg Power Cord HOSVVH2 F2x0 75mm2 Fuse Rating TTA 250V Operating amp Transportation amp Storage Conditions Temperature 5 C 45 C Humidity 15 60 Electrical Classification Class II Type BF double insulated with or without functional ground wire IPXO do not immerse power unit in any liquid or spray any liquids directly on the power unit This system is not AP APG protected Continuous operation Agency Approval EN60601 1 EN60601 1 2 EN61000 3 2 EN61000 3 3 UL2601 1 Electromagnetic Compatibility EN60601 1 2 1993 Maximum Weight Capacity 100 kg Shelden Healthcare reserves the right to change prices or design parameters and specifications without incurring any liability in connection with such modifications DUAL CUSHION USER MANUAL 10 Shelden Healthcare Ltd Unit 1A The Wharf Stretton Under Fosse Rugby Warwickshire CV23 OPR Telephone 01788 833 989 Fax 0845 127 50 50 Email sales shelden healthcare co uk Web www pressurerelieving co uk
2. shelden HEALTHCARE Dual Cushion User Manual 1 Thank you for purchasing this Shelden Healthcare pressure relieving cushion system PLEASE READ THIS USER MANUAL CAREFULLY BEFORE SETTING UP AND USING THE DEVICE Pay special attention to the warnings and other safety information Use genuine Shelden Healthcare components to maintain the products optimal performance If you do not fully understand all the instructions safety precautions and warnings do not use this device In case you have questions please contact Shelden Healthcare on 01788 833 989 Section 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 Description Page Symbols and Indications Safety Precautions Warnings System Package Features Mattress and Power Unit Installations Program Settings Patient Transfer and Transport 03 03 04 04 04 05 06 06 Section 9 0 10 0 12 0 13 0 14 0 shelden HEALTHCARE Description Emergency CPR Deflation Cleaning Instruction Routine Maintenance Troubleshooting Returns for Service Warranty Page 06 07 08 09 09 09 1 0 Indications Attention Please read enclosed document thoroughly Double insulated Class II Declaration of Conformity to Medical C Device Directive International Electronics Approval IPXO Do not immerse power unit in liquid or spray liquids directly on power unit EN Type B Equipment Indications T
3. soft or firm to suit each patient s comfort Caregivers should always perform a hand check by placing their hands underneath patient s pelvis area to check if there is sufficient air support to ensure the patient is not bottoming out Control Panel Features See Figure 5b on page 5 Power Receptacle 1 Make sure power cord is inserted securely into the wall outlet Couplers 2 Quick release couplers are used to secure cushion s air hoses to power unit Control Panel Features See Figure 5c on page 5 Hanging Hooks 1 Hanging hooks are designed to hang the power unit on almost any foot board DUAL CUSHION USER MANUAL 4 Side Features Figure 5b Rear Panel Features Figure 5c Air Filter and Filter Cap 2 Carilex recommends that the filter to be clean or replace once a month to ensure optimal performance of the power unit Fuse 3 Fuse holder can be opened for maintenance Control Panel Features Figure 5a 6 0 Cushion and Power Unit Installations The following power unit installation is for the standard power unit SR303 For other compatible Carilex power units except power unit SR37 1 please refer to the installation section in each power unit s operation manual 1 Place Carilex mobile cushion with foam base on a sitting platform ex wheelchair with the logo facing the front If this system is used with a wheelchair please verify all wheelchair functions are working properly
4. without interference before proceeding to the next step 2 Secure the power unit onto the sitting platform by using hanging hooks or other means 3 Firmly connect the cushion s air hose couplings to the quick release couplers on the power unit s air outlet 4 Plug the power unit into a wall outlet and turn on the power rocker switch on the control panel to inflate the mobile cushion system 5 Turn the comfort knob on the power unit s control panel to maximum setting 150 kg to shorten the initial inflation time if needed BI MOBILE AIR ALTERNATING PRESSURE CUSHION SYSTEM 6 Wait for approx 10 minutes for the cushion to inflate fully 7 0 Program Settings The following program settings are for the standard power unit SR303 For other compatible Carilex power units except power unit SR371 please refer to the program settings section in each power unit s operation manual 1 Position the patient on the cushion Adjust the cushion s internal pressure according to the patient weight by using the comfort knob on the control panel of the power unit If the patient feels the cushion is too soft or firm increases or decreases the cushion s internal pressure by fine turning the comfort knob one increment at a time and wait for the system to stabilize before making another change until a comfortable state is achieved 2 Caregivers should always perform a hand check by placing their hands undernea
5. his air alternating mattress system is designed for patients who endure pressure ulcer and potential patients who wish to reduce the likelihood of pressure ulcer This device is intended to treat and prevent pressure ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area Always consult a physician or health professional before using this mattress system This anti decubitus mattress system is mainly for high risk patient group Contraindications Certain patient conditions are not suitable for using this type of device such as fracture of instable vertebrae and illness of instable vertebrae Always consult a physician or health professional before using this device The use of this system does not replace the regular repositioning monitoring and nursing of the patient 2 0 Safety Precautions 1 To ensure proper operation please inspect and verity all parts are setup properly and anchor securely Do not place anything on top of the power unit Make sure power cord is underneath bed frame and free from hazard 2 Avoid using this device near open flames lighters or cigarettes Flammability hazard exists This device draws air from the surrounding Thus cigarette smoking may damage internal component 3 This system should disinfect thoroughly between patients to avoid cross contamination 4 Verity patient weight does not excess the sitting object and this mobile cushio
6. n system s weight capacity 3 0 Warnings 1 DO NOT disassemble the power unit if you are not a qualified technician Please contact your local distributor for all service 3 MOBILE AIR ALTERNATING PRESSURE CUSHION SYSTEM 2 This product is NOT AP APG protected 3 Operating transportation and storage conditions Temperature 5 C 45 C Humidity 15 60 CAUTION ENSURE THAT THERE ARE NO PROTRUDING OBJECTS SHARP POINTS OR BED SPRINGS UNDER THE MATTRESS AS THESE COULD PUNCTURE THE AIR CELLS 4 0 System Package Power Unit Package Standard Power unit x 1 Power cord x 1 Operation manual x 1 Standard power unit package is only available when purchasing a complete mobile air alternating pressure cushion system Mobile Cushion Package Mobile cushion with coverlet x 1 Foam base with coverlet x 1 5 0 Features The following features are for the standard power unit SR303 For other compatible Carilex power units except power unit SR371 please refer to the features section in each power unit s operation manual Control Panel Features See Figure 5a on page 5 power on off Once the power rocker switch is turned on the green switch will light to indicate the system is on Comfort Knob Simply turn the knob from 30 kg to 150 kg to adjust the cushion internal pressure for patient The scale is only an approximation Please adjust the pressure setting if the cushion is too
7. ructed positions 3 Patient is not getting pressure Contact your physician and or relief due to system failure nursing service immediately reddening of skin Air tube is referring to the mean of directing air flow from the power unit to each individual air cell 9 9 p DUAL CUSHION USER MANUAL 18 13 0 Returns for Service This device is not self serviceable Service and repair must be performed by an authorized technician or representative All returned device must be cleaned and disinfected prior to shipping Unsanitary or soiled systems will be returned without servicing 14 0 Warranty Shelden Healthcare warrants the product to be free from defects from the date of purchase Please inspect all accessories when you purchase our product If there is any damage or missing accessories when you receive the product please ask for a replacement from your local distributor within three days of purchase The warranty periods for Shelden Healthcare products are according to the regulations in your country the minimum period is 1 year from date of purchase for the power unit and 1 year for the mattress and the coverlet The warranty coverage of any Shelden Healthcare product is contingent up on its purchase from an authorized dealer Warranty for any product in the Shelden Healthcare product line will be honored by the official distributor in your country Warranty coverage will not be extended to any product on which
8. t and water or mild neutral detergent Never spray liquids directly on the unit itself DUAL CUSHION USER MANUAL 16 11 0 Routine Maintenance Remove air filter from the real panel of the power unit by opening up the filter cap to clean or replace the air filter Inspect the filter for dirt or dust and clean it with mild soap and water Reinsert the dried air filter after cleaning and ensure the cap is secure Replace with genuine air filter once a year is recommended and compressor every two year is strongly suggested to prolong system lifetime The mobile cushion system must be disinfected and dried before storage 7 MOBILE AIR ALTERNATING PRESSURE CUSHION SYSTEM 12 0 Troubleshooting 1 Power unit does not function Check if power cord is firmly Secure power cord into wall plugged into wall outlet outlet Check if power surge shuts down A power surge may overload the power unit circuitry temporarily Turn the unit off and check the fuse for damage Turn the unit ON again with normal procedure Power unit does not respond to Please contact Shelden possible solutions Healthcare on 01788 833 989 Check if mattress air hose couplers Secure air hose couplers firmly are properly connected to power into place unit s quick release couplers Inspect air filter for dust or dirt Clean or replace air filter Lift mattress coverlet up to check if Check and adjust air tubes air tubes are kinked or obst
9. th patient s pelvis area to check if there is sufficient air support to ensure the patient is not bottoming out 8 0 Patient Transfer and Transport The following patient transfer instruction is for the standard power unit SR303 For other compatible Carilex power units except power unit SR371 please refer to the patient transfer and transport section in each power unit s operation manual Transter It is recommended to have the cushion system at maximum weight setting during transfer Make sure the cushion is secured before proceeding 9 0 Cleaning Instruction Mobile air alternating cushion and power unit must be cleaned thoroughly between patients to avoid cross contamination The following is a suggested guideline but local infection control policies should be followed as well Regular cleaning can be performed with disinfectant and water followed by drying with a clean dry cloth Use only mild detergents and water to clean the coverlet and the cushion Any appropriate NON PHENOLIC cleaning agent may be used for heavy soiling with urine blood or other body fluids Please ensure cushion and coverlet are completely dry before letting the patient laying on the surface again Machine wash warm water at maximum 40 C Do not use electric or tumble dryers Do not iron WARNING Always unplug the power unit before cleaning Routine cleaning of power unit can be done by wiping down with damp cloth using disinfectan
10. the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by Shelden Healthcare or if in the sole opinion of Shelden Healthcare that the system shows evidence of tampering abnormal or unreasonable abuse negligence accident or operation withoutregard for the restrictions specified in the instructions which accompany the system This warranty does not cover normal maintenance such as cleaning adjustment lubrication and updating of equipment or parts If the damage is result from improper operation a reasonable service fee and part cost will be charged The warranty stated above is the only warranty made and is in lieu of all other warranties whether expressed or implied including any warranty of merchantability or fitness for a particular reason Shelden Healthcare will not be liable for consequential or incidental damages or any kind 15 0 Product Specifications The following power unit related specifications are for the standard power unit SR303 For other compatible Shelden Healthcare power units except power unit SR371 please refer to product specifications section in each power unit s operation manual Cushion Dimension MSR213 45x45x9cm Number of Air Cells 10 air cells Power Control Unit Dimension SR303 26x22x9cm Material Power Unit ABS fire retardant UL Mobile Cushion Cell 100 nylon with TPU lamination Coverlet

Download Pdf Manuals

image

Related Search

Related Contents

Original Operating Instructions F-RER / F-REL  Mode d`emploi  Notifier - AFP 300 & 400 Installation, Programming & User Manual  Patriot Memory DDR3 2GB CL9 PC3-10600 (1333MHz) 2 Rank DIMM  4輪ドリーカメラスタンドセット  SillaCocheBebe.com  Sony VAIO VGNB100 User's Manual  Philips HD camcorder CAM101BL  User`s Manual of WLAN Broadband Router (1T1R)  Bedienungsanleitung  