RESmart Fixed CPAP User Manual

image

Contents

1. DO NOT use the RESmart if the display is erratic Contact your home care provider for further instructions If at any time you wish to exit the setup menu press the Pressure Start Stop Button and the display will go back to the system standby screen P11 RESmart User Manual V2 4 First Time Setup cont Setting the Humidifier a When the optional InH2 Heated Humidifier is used this setting can change the IMPORTANT Setting the Altitude IMPORTANT Setting Auto ON Setting Auto Off temperature and humidity of airflow between level 0 to 5 Default setting is 3 Press the Ramp Button to enter change status and press User Buttons to change the setting Press Ramp Button again to confirm the setting Press User Buttons to access the next setting Heater 3 b The altitude setting can be changed between level 0 to 2 Default setting is 0 0 less than 2 460 ft lt 750 m 1 2 460 to 4 921 ft 750 m to 1500 m 2 4 924 to 8 202 ft 1501 m to 2500 m over 8 202 ft The airflow pressure may not be accurate Contact your home care provider to have your pressure setting adjusted Altitude 0 RESmuart is already with automatic altitude compensation Even if without setting the altitude RESmart still can deliver proper pressure in any altitude worldwide c When the RESmart is standby and wearing mask your deep breath will start airflow automatically This function is NOT available on cer
2. RESmart CPAP System User Manual CE 0123 RESmart User Manual V2 4 Table of Contents MD CCIIICALIONNS yenu ran k sek y bee ae kk yee x Saad Sba W Baw a Ga n kw diy Kas wa We w ky Warnings amp Call O 1S assis yesnieak ki kendin anada din na k ka W kwa sw k d a Ua kd w k Unpacking the System sk iyilik n k kulk na ka kk kuh a n aiak System Features sasyyya durin tul kina k ly klk e xes wa xwi enka kada wan aw warek br ano aa FESI ji ja j SE al la SEN N N E a O Installing the Filter lkA k lLLLLLLLLLLLL LL kk kk kk kk kk kk kk kk kk RR RR RR K KK KAKA O Device SENINGS sa suisskn yen n zkuan nen d yd n keya k kan k kk kw b kwa kedl w ya k wa kaka EE O Assembling the Circuit XkK K EKSLLLL LL kk kk kk kk kk kk kK kk kK K AKA KK KK K KA AGCOSS OTTO S j a s n nina nak na ne nenena nan n ken n k n b k E n RE R KA se H NARO RAR E KEND AS n SR W RA kun O Adding a HUI dI ss y s salnsn nsan sana kala kan kall u nak ah RN Ea E benu Rek U rak na O Adding Oxygen to the System ccccccccseecseeeeeeeeeeeeeeseseseeeeeeees ROUNE USE ssil n kak danye ka a we Aka kw ek sl asan a k dik e Alb ue ye tainsuneventecerenneedaamenian O Connecting the CirCU kl kk kk kk kk kk kk kK KK A KK K KK K KAK KA A O Adjusting the Circuit kKkK K C LLLL LEK kk kk kk kK kK KAKA KK KA KK KK inane O Ramp BUHO sarrin nnn ns anek e
3. The RESmart uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Group 1 CISPR 11 Class B Harmonic emissions The RESmart is suitable for use in all establishments IEC 61000 3 2 Class A including domestic establishments and those directly connected to the public low voltage power supply network Voltage fluctuations that supplies buildings used for domestic purposes flicker emissions Class D IEC 61000 3 3 RF emissions Group 1 CISPR 11 Class B P22 RESmart User Manual V2 4 Guidance and manufacturer s declaration electromagnetic immunity The RESmart is intended for use in the electromagnetic environment specified below The user of the RESmart should make sure that it is used in such an environment Immunity test Electrostatic discharge ESD IEC 61000 4 2 Electrical fast transient burst IEC 61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 IEC 60601 test level 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 Ur gt 95 dip in U for 0 5 cycle 40 Ur 60 dip in U for 5 cycles 70 Ur 30 dip in U for 25 cycl
4. Airway Pressure Caution should be used when prescribing CPAP for susceptible patients such as those with cerebral spinal fluid leaks abnormalities of the cribriform plate prior history of head trauma and or pneumocephalus Chest 1989 96 1425 1426 The use of positive airway pressure therapy may be temporarily contraindicated if you exhibit signs of a sinus or middle ear infection Contact your health care professional if you have any questions concerning your therapy P3 RESmart User Manual V2 4 Specifications Device Size Dimensions 220 x 194 x 112mm 313 x 194 x 112 mm with InH2 heated humidifier Weight 1 6kg 2 4 kg with InH2 heated humidifier Product Use Transport and Storage Operation Transport and Storage Temperature 5 to 30 C 20 to 55 C Humidity lt 80 Non condensing lt 93 Non condensing Atmospheric Pressure 860 to 1060 hPa 500 to 1060 hPa Mode of Operation Continuous AC Power Consumption 100 240V AC 50 60Hz 1 0A max Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Shock Type BF Applied Part Degree of Protection Against Ingress of Water IPX1 Drip Proof Vertical Pressure Range 4 to 20 hPa in 0 5 hPa increments Pressure Stability 4 to 20 hPa 1 hPa Altitude Compensation Manually setting level 0 1 2 Automatic altitude compensation Sound Pressure Level lt 30 dB when the device is working at the pressure of 10 hPa Ma
5. Pressures must be verified by your home care provider when alternate or optional accessories are in place a Connect the flexible tubing to the air outlet on the front of the RESmart b If you are using a mask with a built in exhalation port connect the mask s connector to the flexible tubing If you are using a mask with a separate exhalation port connect the flexible tubing to the exhalation port Position the exhalation port so that the vented air is blowing away from your face Connect the mask s connector to the exhalation port Do not block or otherwise try to seal the air openings vent holes on the exhalation port Explanation of the warning The RESmart is intended to be used with masks and circuits that have an exhalation port designed to exhaust CO from the circuit When the RESmart is turned on and functioning properly new air from the RESmart flushes the exhaled air out through the exhalation port When the RESmart is turned off enough fresh air will not be provided through the mask and exhaled air may be rebreathed Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suffocation This warning applies to most CPAP devices If you are using a full face mask i e a mask covering both your mouth and your nose the mask must be equipped with a safety entrainment valve P15 Adding a Humidifier WARNING Adding Oxygen WARNING WARNING WARNING RE
6. cora often Jor any signs of damage Keplace a damaged cor WARNING I h d often fi j f d Repl d d cord immediately WARNING The RESmart is powered on for use when the power cord is connected The pressure start stop button turns the blower on off 3 Place the RESmart on a firm flat surface CAUTION Make sure the RESmart is away from any heating or cooling equipment e g forced air vents radiators air conditioners Also make sure that bedding curtains or other items are not blocking the filter or vents of the device Air must flow freely around the device for the system to work properly P9 Device Settings IMPORTANT RESmart User Manual V2 4 First Time Setup cont Display Screen All device settings will appear here Pressure Start Stop Button Use this button to start stop the airflow DO NOT start the airflow until the circuit tubing is connected Heated Humidifier Button Use this button when the optional InH2 Heated Humidifier has been prescribed This button will control the optional heated humidifier s output Follow the instructions of the humidifier Ramp Button When the airflow is turned on use this button to restart the ramp cycle which lowers the airflow pressure and then gradually increases it This will allow you to fall asleep more comfortably When the airflow is turned off use this button to access the patient menu Note The ramp feature is not prescribed for all users User Butto
7. ken dek n k Weka RO E Wa Aa 4 Wa RA C E Va he Gm A O H KEKE KO KIR M O Turning the System Off KK k kk kk kk kk kk KAK KAK KAK KAK KA O PI DIG NINIS sj sinin saarenney nanna naden n k kekEan ban k hU kan AE l Ll an n d nn ka Ra k n RARA Routine Maintenance ssssssssnnnsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nnmnnn nnmnnn O Cleaning the Filter asannsnnsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn O Cleaning the System sssssssssnnsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nnna CO FREON OL TIL O orsoni REEE EEE O SOVICE sorrera mena kid pa a award a gk n ew awar aa olk e w ne de kawana a w e n Traveling with the System sescseesecsiecscwwsncenes ceivetemencsienwseasnewstwunwantencnuc Tr OUDIOSTIOOT T su sas use nnn cisps cree nein ts yk dik danka saxan n n san k dalal gula dan ks daka EMC BQ UITOTI OM si xa sual y na ua kayan n kek n dak aka Ay dal kwa kala ek u uk kan YAT AIT V aya ey kan kina nena n kak e banka kele deal ke kl Nal k k Wel Si kak NE OS k r k A NEBA k P1 RESmart User Manual V2 4 Symbols Control Buttons h p ss El Pressure Start Stop Ramp Start Humidifier On Off E3 User Button User Button Device A Attention Consult Type BF Applied part m Class Double Insulated Accompanying Documents NN AC Power IPX1 Drip Proof Vertical SN Serial Number
8. of the product u4 Manufacturer Authorised Reprisentative in the European Community European CE Declaration of Conformity RESmart InH2 are registered trademarks of BMC MEDICAL CO Ltd BMC MEDICAL CO Ltd 2011 P2 Intended Use Intended Use IMPORTANT Contraindications RESmart User Manual V2 4 The RESmart system is a CPAP Continuous Positive Airway Pressure device designed for the treatment of adult Obstructive Sleep Apnea OSA only The RESmart is to be used only on the instruction of a licensed health care professional Your home care provider will make the correct pressure settings according to your health care professional s prescription Several accessories are available to make your OSA treatment with the RESmart CPAP system as convenient and comfortable as possible To ensure that you receive the safe effective therapy prescribed for you use only BMC accessories Read and understand the entire user manual before operating this system If you have any questions concerning the use of this system contact your home care provider or health care professional Studies have shown that the following pre existing conditions may contraindicate the use of positive airway pressure therapy for some patients e Bullous Lung Disease Bypassed Upper Airway e Pneumothorax e Pathologically Low Blood Pressure e Pneumocephalus has been reported in a patient using nasal Continuous Positive
9. or harsh sounds if it has been dropped or mishandled if the enclosure is broken or if water has entered the enclosure disconnect the power cord and discontinue use Contact your home care provider e To avoid electrical shock disconnect the power cord before cleaning DO NOT immerse the device in any fluids o Contact your health care professional if symptoms of sleep apnea recur e Using the RESmart at an incorrect elevation setting could result in airflow pressures higher than the prescribed setting Always verify the elevation setting when traveling or relocating P5 RESmart User Manual V2 4 Warnings amp Cautions Unpacking the System CAUTION e Tobacco smoke may cause tar build up within the RESmart Indicates the that may result in the RESmart malfunctioning possibility of damage o The use of accessories other than those specified with the exception of to the device cables sold by the manufacturer of the equipment or system as replacement parts for internal components may result in increased emissions or decreased immunity of the equipment or system Additional warnings and cautions are located throughout this manual as they apply After unpacking the system make sure you have everything shown here RESmart CPAP Carrying Case PL 4 User Manual 1 IMPORTANT If any of the above parts are missing contact your home care provider P6 RESmart User Manual V2 4 System Features Handle Displ
10. MEDICAL CO LTD DISCLAIMS ALL LIABILITY FOR ECONOMIC LOSS LOSS OF PROFITS OVERHEAD OR CONSEQUENTIAL DAMAGES WHICH MAY BE CLAIMED TO ARISE FROM ANY SALE OR USE OF THIS PRODUCT SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU To exercise your rights under this warranty contact your local authorized dealers or MANUFACTURER BMC MEDICAL CO LTD 5 F North No 2 Building No 9 Shixing Street Shijingshan District Beijing 100041 P R China Tel 86 10 51663880 Fax 86 10 51663880 Ext 810 EU AUTHORISED REPRESENTATIVE Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Tel 0049 40 2513175 Fax 0049 40 255726 P25
11. RESmart User Manual V2 4 First Time Setup Do not use the RESmart system until an appropriate professional adjusts the settings To order any accessories not included with this system contact your home care provider Note to home care provider Before beginning setup be sure that you have available the RESmart Home Care Provider Setup Instructions Setup instructions are not provided in this manual DO NOT connect any equipment to the RESmart unless recommended by BMC or your health care provider If the RESmart has been exposed to either very hot or very cold temperatures allow it to adjust to room temperature approximately 2 hours before beginning setup P6 RESmart User Manual V2 4 First Time Setup cont Installing the Filter 1 Install the filters and filter cap CAUTION The foam filter must be in place at all times when the RESmart is operating a Place the foam filter into the filter area on the back of the RESmart b Attach the filter cap Position the cap so that the small opening on the cap is facing down Insert the cap s tabs into the filter area opening 2 Connect the power cord Plug the socket end of the power cord into the AC inlet on Power Cord the back of the RESmart Plug the pronged end of the power cord into an electrical outlet System status will appear on the RESmart screen To remove AC power disconnect the power cord from the electrical outlet IMPORTANT a nspect the power
12. Smart User Manual V2 4 Accessories Contact your home care provider for additional information on the accessories available for the RESmart system When using optional accessories always follow the instructions enclosed with the accessories The InH2 Heated Humidifier is available from your home care provider The humidifier may reduce nasal dryness and irritation by adding moisture and heat if applicable to the airflow The RESmart unit should only be connected to the humidifiers or cacessories specified in this Manual Connection of other Items may result in injury or damage to the RESmart unit Oxygen may be added at the mask connection Please note the warnings listed below when using oxygen with the RESmart The oxygen supply must comply with the local regulations for medical oxygen Turn the RESmart on before turning the oxygen on Turn the oxygen off before turning the RESmart off Explanation of Warning When the RESmart is turned off but the oxygen flow is still turned on oxygen may accumulate within the RESmart enclosure and create a fire risk Turning the oxygen off before turning the RESmart off will prevent oxygen accumulation in the RESmart and will reduce the risk of fire This warning applies to most CPAP devices Oxygen accelerates fires Keep the RESmart and the oxygen container away from heat open flames any oily substance or other sources of ignition DO NOT smoke in the area near the RESmart
13. The ramp feature is not prescribed for all users 3 Press the Ramp button on the top of the RESmart You can use the Ramp button as often as you wish during the night P17 RESmart User Manual V2 4 Routine Use cont Tu rning the System OFF 4 Remove the mask and headgear Press the pressure start stop button on the top of the IMPORTANT Helpful Hints RESmart to stop the airflow Or if the Auto Off setting has been enable the airflow will automatically turn off The humidifier button is active only when an InH2 Heated Humidifier is connected or when the RESmart is in the Setup Menu Refer to the InH2 Heated Humidifier s instructions for additional information If the alert tone sounds press any button on the RESmart to silence the alert tone Refer to the Troubleshooting Section of the manual for further instructions e Make sure that bedding curtains or other items are not blocking the filter or vents of the RESmart Air must flow freely around the RESmart for the system to work properly If the airflow from the RESmart feels cold reposition the circuit tubing so that it runs under your bed covers to reduce heat loss while you sleep If interruption of the power supply occurs during system working the audible alert a beeping sound will start and continue Restoration of the power supply will stop the audible alert After interruption and restoration of the power supply the RESmart will resume w
14. ay Medical Screen Product Note User Buttons bottom Humidifier gt AC Inlet Controller meas J Communications Connector Air Outlet Filter Cap Humidifier els Power AC TMC sssrinin Connect disconnect the AC power cord here Air Outlet Connect the flexible tubing coaxial 22 mm here Communications Connector For clinical use with the Data Management Software Connect the communications cable here Display Screen All system settings total operating time and therapy hours will appear here Filter Cap amp Filters The foam filter screens out normal household dust and pollens The filter cap is designed to reduce the noise from the RESmart Humidifier Controller This controller turns the InH2 heated humidifier on off and allows the heat setting to be adjusted The humidifier is an optional accessory that may be purchased separately Humidifier Power This socket provides the power to the InH2 heated humidifier Medical Product Note For ease at airport security stations there is a note on the bottom of the RESmart stating that it is medical equipment It may help if you also take this manual with you when you travel User Buttons These buttons can be used to turn on off the system and change some of the system settings Handie spisser This handle is for lifting up the system P7 WARNING WARNING CAUTION
15. es lt 5 Ur gt 95 dip in U for5s Compliance level 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 Ur gt 95 dip in U7 for 0 5 cycle 40 Ur 60 dip in U7 for 5 cycles 70 Ur 30 dip in U for 25 cycles lt 5 Ur gt 95 dip in Urs for5s Electromagnetic environment guidance Floor should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical home or hospital Mains power quality should be that of a typical home or hospital Mains power quality should be that of a typical commercial or hospital environment If the user of the RESmart requires continued operation during power mains interruptions it is recommended that the RESmart be powered from an uninterruptible power supply or from a battery If the pressure deviates more than is indicated in the device specification it may be necessary to position the RESmart further from sources of power frequency magnetic fields The power frequency magnetic field should be measured in the intended installation location to ensure that it is sufficiently low NOTE Uris the AC mains voltage prior to application of the test level P23 RESmart User Manual V2 4 Guidance and manufacturer s declaration electromagnetic immuni
16. ightly dampened with water and a mild detergent Let the RESmart dry before plugging in the power cord 3 Inspect the RESmart and all circuit parts for any damage after cleaning Replace any damaged parts 4 For details on cleaning your mask and accessories refer to the cleaning instructions packaged with the accessories P19 Reordering Service WARNING WARNING Traveling with RESmart Security Stations RESmart User Manual V2 4 Reordering Service Traveling with the System Contact your home care provider to order accessories or Replacement filters The RESmart system does not require routine servicing The RESmart unit contains electrical and or electronic equipment When necessary dispose of the RESmart and accessories in accordance with local regulations If you notice any unexplained changes in the performance of the RESmart if it is making unusual or harsh sounds if it has been dropped or mishandled if the enclosure is broken or if water has entered the enclosure discontinue use Contact your home care provider If the RESmart malfunctions contact your home care provider immediately Never attempt to open the RESmart s enclosure Repairs and adjustments must be performed by BMC authorized service personnel only Unauthorized service could cause injury invalidate the warranty or result in costly damage If necessary contact your local authorized dealer or BMC MEDICAL CO Ltd for technical s
17. ith the working status before interruption automatically P18 RESmart User Manual V2 4 Routine Maintenance Cleaning the Filter CAUTION CAUTION Cleaning the System WARNING The foam filter should be cleaned at least once every two weeks under normal usage and replaced with a new one every six months Operating the RESmart with a dirty filter may keep the system from working properly and may cause damage to the device 1 Remove the filter cap by gently pressing on its bottom 2 Change the filter a Remove the foam filter by gently pulling around the edges of the filter Rinse the filter in a steady stream of running water Squeeze out the water and repeat Air dry the filter on a rack for 8 to 12 hours or in a clothes dryer for 15 to 20 minutes b Insert the filter into the filter area on the back of the RESmart Replace the filter cap Never install a wet filter into the RESmart We recommend that you clean the filter in the morning and alternate using the filter provided with the system to ensure enough drying time for the cleaned filter Clean the mask and tubing daily To avoid electrical shock unplug the RESmart before cleaning Do not immerse the RESmart in any fluids 1 Disconnect the flexible tubing from the RESmart Gently wash the flexible tubing in a solution of warm water and a mild detergent Rinse the tubing thoroughly and air dry 2 Wipe the outside of the RESmart with a cloth sl
18. ns These buttons can be used when entering various menus to change some of the RESmart settings pe Display Screen User Buttons Ew gt User Buttons Heated Humidifier cm gt Ramp Button Button Pressure Start Stop Button In the Setup Menu the user buttons are used to go to the previous next question or setting and operate as up and down keys to change the settings the ramp button is used to confirm the setting change and the start stop button will allow you to exit the Setup Menu without saving P10 RESmart User Manual V2 4 First Time Setup cont Standby Device Settings IMPORTANT WARNING IMPORTANT 4 After power on system status will appear on display screen as below Ramp Timer Ramp 15 04 08 Ramp Starting Treatment Pressure Pressure Ramp Starting Pressure The starting pressure after blower is turned on Default setting is 4hPa Treatment Pressure The prescription pressure after ramp time Default setting is ShPa Ramp Timer The ramp time from starting pressure to prescription pressure Default setting is 15 minutes 5 Verify Change the RESmart settings To start the setup menu press and hold the ramp button until the setup menu appear about 5 seconds when system standby The airflow must be turned OFF Pressing the ramp button when the airflow is turned on will lower the airflow pressure if prescribed and then will gradually increase it
19. or the oxygen container P16 Routine Use Connecting the Circuit IMPORTANT Adjusting the Circuit Using the Ramp Button RESmart User Manual V2 4 1 Connect the circuit Before each use examine the flexible tubing for any damage or debris If necessary clean the tubing to remove the debris Replace any damaged tubing a Connect the mask to the headgear following the instructions included with the headgear b Put on the mask and headgear and breathe normally through your nose The airflow should automatically start when you begin breathing through the circuit If the airflow does not start within four breaths press the pressure start stop button on the top of the RESmart When operating the system with some mask types or some circuit configurations the airflow may NOT automatically start 2 Adjust the circuit Lie down on your bed and adjust the flexible tubing so it is free to move if you turn in your sleep Adjust the mask and headgear until you have a comfortable fit and there are no airflow leaks into your eyes Pressing the ramp button will reduce the air pressure when you are trying to fall asleep The air pressure will gradually increase until your prescription pressure is reached If your physician prescribed ramp for you pressing the button will reduce the pressure and then gradually increase ramp the pressure to the therapeutic pressure setting so that you can fall asleep more comfortably Note
20. phones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the RESmart is used exceeds the applicable RF compliance level above the RESmart should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the RESmart P Over the frequency range 150 kHz to 80 MHz the field strengths should be less than 3 V m P24 RESmart User Manual V2 4 Limited Warranty BMC MEDICAL CO Ltd warrants that the RESmart CPAP device shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of one 1 year for main unit and three 3 months for all accessories from the date of sale by BMC MEDICAL CO Ltd to the dealer If the product fails to perform in accordance with the product specifications BMC MEDICAL CO Ltd will repair or replace at its option the defective material or part BMC MEDICAL CO Ltd will pay customary freight charges from BMC MEDICAL CO Ltd to the dealer location only This warranty does not cover damage caused by accident misuse abuse alteration and other defects not related to material or workmanship BMC
21. ressure relief change the setting to 2 This function is NOT available on certain model Default setting is 0 disable Reslex 0 P13 RESmart User Manual V2 4 First Time Setup cont Setting the Delay Off h When the optional InH2 Heated Humidifier is used this setting allows the air flow Use Days Exit the Settings continue for about 15 minutes at very low pressure about 2hPa after RESmart is turned off This will blow off the left vapor and protect the RESmart Default setting is Enable DelayOff Enable 1 The number of nights the RESmart was used for therapy for more than 4 hours will appear This screen is only for reference Your home care provider may periodically ask you for this information Use Days 0 The settings are complete Press the user button to access the SAVE setting and Ramp Button to save all changes and exit the settings menu P14 RESmart User Manual V2 4 First Time Setup cont Assembling the Circuit WARNING WARNING WARNING 6 Assemble the circuit To use the system you will need the following accessories in order to assemble the recommended circuit Nasal Mask with integrated exhalation port e 6 ft 1 83 m flexible Tubing e Headgear for the mask If multiple persons are going to use the RESmart e g rental devices a low resistance main flow bacteria filter should be installed in line between the RESmart and the circuit tubing
22. s set to your altitude Change the altitude setting if necessary If the altitude setting is correct contact your home care provider for directions on having the RESmart serviced Please have the device s serial number ready when you call Replace or clean the filter Make sure the RESmart is away from bedding or curtains that could block the flow of air around the RESmart Make sure the RESmart is away from heating equipment e g forced air vents radiators Contact your home care provider for directions on having the RESmart serviced Please have the device s serial number ready when you call Make sure that the RESmart is plugged into a working outlet Contact your home care provider for directions on having the RESmart serviced Please have the device s serial number ready when you call Contact your home care provider Ramp may not have been prescribed for you Discontinue use Disconnect the power cord from the AC wall outlet Contact your home care provider for directions on having the RESmart serviced Please have the device s serial number ready when you call P21 RESmart User Manual V2 4 EMC Requirements Guidance and manufacturer s declaration electromagnetic emissions The RESmart is intended for use in the electromagnetic environment specified below The user of the RESmart should ensure that it is used in such an environment Emissions Test Compliance Electromagnetic environment guidance
23. tain model Default setting is Enable Auto On Enable d When the mask is removed the RESmart will stop airflow automatically This function is NOT available on certain model Default setting is Enable P12 RESmart User Manual V2 4 First Time Setup cont Auto Off Enable Setting the LCD Light e This setting allows you to adjust the LCD light among Auto On and Off Default setting Setting the Alert Setting the Reslex is Auto Auto LCD light will be turned on when any button pressing and turned off after few seconds On LCD light will be turned ON always Off LCD light will be turned OFF always Light Auto f This setting allows you to turn on off the patient disconnect Alert When a large continuous air leak such as mask removal has been detected in the circuit this setting enables disables the audible alert a beeping sound This setting is only effective when the Auto Off setting is disabled This function is NOT available on certain model Default setting 1s Disable Alert Disable g The Reslex setting displays the Reslex level set by your home care provider This setting allows you to adjust the level of air pressure relief that you feel when you exhale during therapy Your home care provider will have already set a Reslex level If this is not comfortable for you you can increase or decrease the setting The setting of 1 provides a small amount of pressure relief To increase the p
24. ty The RESmart is intended for use in the electromagnetic environment specified below The user of the RESmart should make sure that it is used in such an environment Immunity test aa a died Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the RESmart including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2VP d 1 2VP 80 MHz to 800 MHz d 2 3VP 800 MHz to 2 5 GHz Conducted RF 3 Vrms 3 Vrms IEC 61000 4 6 150 kHz to 80 MHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the Radiated RF 3 V m 3 V m recommended separation distance in meters m IEC 61000 4 3 80 MHz to 2 5 GHz Field strengths from fixed RF transmitter as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Kn NOTE 1 At 80 MHz and 800 MHz the higher frequency range applied NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless tele
25. upport and documents When you are traveling the carrying case is for carry on luggage only The carrying case will not protect the system if it is stored as checked baggage For convenience at security stations there is a note on the bottom of the RESmart stating that it is medical equipment It may be helpful to bring this manual along with you to help security personnel understand the RESmart Checking the Power Cord Ifyou are traveling to a country with a line voltage different than the one you are currently using a different power cord or an international plug adapter may be required to make your power cord compatible with the power outlets of the country to which you are traveling Contact your home care provider for additional information P20 Troubleshooting RESmart User Manual V2 4 The table below lists common problems you may have with the RESmart system and possible solutions to those problems If none of the corrective actions solve the problem contact your home care provider Problem The pressure being delivered feels different The airflow from the RESmart seems warm The noise level of the RESmart has changed to include unusual or harsh sounds during operation The RESmart will not turn on Pressing the ramp button does not reduce the air pressure The RESmart has been dropped into water or fluids have gotten into the enclosure Solution s Check the altitude setting to be sure it i
26. ximum Flow Test Preanure hPe EE KERE E22 Average flow at the patient connection port L min P4 RESmart User Manual V2 4 Warning amp Cautions CAUTION WARNING Indicates the possibility for injury to the user or the operator o This device is restricted to sale by or on the order of a physician o The instructions in this manual are not intended to supersede established medical protocols o This device is intended for adult use only o This device is not intended for life support e CPAP devices have the potential to allow rebreathing of exhaled air To reduce this potential observe the following Use BMC circuit accessories Do not wear the mask and headgear for more than a few minutes while the device is not operating Do not block or try to seal the vent holes in the exhalation port As with most CPAP devices At low CPAP pressures some exhaled gas CO2 may remain in the mask and be rebreathed e Do not use this device if the room temperature is warmer than 86 F 30 O If this device is used at room temperatures warmer than 86 F 30 C the temperature of the airflow may exceed 106 F 41 C This could cause irritation to your airway o This equipment is not suitable for use in the presence of a flammable anaesthetic mixture in combination with oxygen or air or in the presence of nitrous oxide o If you notice any unexplained changes in the performance of the RESmart if it is making unusual