OPERATING INSTRUCTIONS COATING THICKNESS GAUGE ET
Contents
1. CAL button over 2 sec to enter calibration mode LCD will display CAL blinking 3 In calibration mode use A or V button to adjust readings until it matches the standard s thickness 4 Press CAL button to exit two point calibration and return to measuring mode CAL off Calibrating Point Clearance In measuring mode press 0 button over 2 seconds LCD will display oooo and start cleaning and zeroing sub strate one point and two point calibration When calibra tion is not operated properly the clearance function helps users to start calibrating again MAINTENANCE Installing and Replacing Battery 1 Power is supplied by 2pcs 1 5V AAA SIZE 2 The 4 appears in the display when battery replace ment is needed 3 Remove the battery cover by gently sliding it onwards the bottom of the meter 4 Remove the batteries from battery compartment 5 Replace with 2 new AAA batteries with polarity as indi cated on the bottom of Battery Compartment 6 Replace the Battery Cover CAUTION When not in use for long periods remove battery Do not store in locations with high temperatures or high humidity Cleaning Periodically wipe the case with a damp cloth and deter gent do not use abrasives or solvents Ver 1 1 02 09 2013
2. Standby Consuming Current lt 6uA Battery 1 5V AAA size x 2pcs Battery Life 17 hours continuity typical with alkaline battery Low Battery Indication The G is displayed when the battery voltage drops below the operating level Dimensions 120mm H x 40 4mm W x 29 2mm D Weight Approx 100g including battery DEFINITION Zeroing Plate Non ferrous is Aluminum plate Ferrous is steel plate Ferrous Non Ferrous Peel off the other side protection film before use Peel off the other side protection film before use Standard Coating Plate Standard Thickness 4 0 0 1 mils 102 2 micron Peel off the both side of protection film before use e Sensing Tip UV OO Lighting Lighting 2 J C E Funetian Key lt i Peel off the protection films from foil before first use FUNCTION BUTTON Sp Press 02 button each time the display changes in the following sequence Lighting UV light OFF it can be operated in ON or Off mode osk 1 Press X button to turn on or off backlight function It can benefit users for reviewing display in dark environ ment 2 Press X button over 2 seconds to switch between mils and um 1 mils 25 4 um F IN 99 Press F N button each time the display changes in the following sequence Auto Ferrous Non Ferrous Auto Mode and Ferrous Fixed Mode and Non Ferrous Fixed Mode
3. lt Q Quickly press 0 button for substrate zeroing calibration Press Q button over 2 seconds to clear Calibrating Point C AL 1 When power is on press CAL button over 2 seconds to start calibration 2 In calibration mode when you finish calibrating press CAL button to complete calibrating procedure A 1 In calibration mode press A button to increase Display s values y 1 In calibration mode press W button to decrease Dis play s values INSTRUCTION Power on and off 1 Keep the sensing tip of the meter away from any sub strate or any magnetic field 2 Gage automatically powers up and Measuring when probe is pressed 3 Auto Power Off APO Leave the gauge without operation for 1 minute power turns off automatically Measuring 1 Press the sensing tip of the gauge to contact coated surface tightly Wait for the reading to appear and measurement is completed One Beep sound an nounced 2 If the coating thickness is out of range the meter shows Tt z ag CALIBRATION During calibration Auto Power Off function will be inactivated If the LCD display it can not zero the substrate one point calibration or two point calibration When it is calibrated by user its max calibrated value is 43 3 mils 1100um Substrate Zeroing Calibration 1 Press the sensing tip of the probe to contact Uncoated surf
4. OPERATING INSTRUCTIONS MODEL ET 111 C 2 IN 1 COATING THICKNESS GAUGE INTRODUCTION This instrument is a 4 digit portable easy to use and compact sized digital ferrous or non ferrous coating designed for simply one hand operation Meter comes with backlight LCD display and Auto Power Off 60 seconds approx to extend battery life SAFETY INFORMATION It is recommended that you read the safety and operation instructions before using the coating thickness gauge DANGER When the UV LED function is turned on it will radiate intense UV light during operation Since UV light can be harmful to eyes do NOT look directly into the UV light even through an optical instrument In case of the light reflection UV protective glasses are required to use in order to avoid damage by the light CAUTION e Do not use the unit near any device which generates strong electromagnetic radia tion or near a static electrical charge as these may cause er rors e Do not use the unit where it be exposed to corrosive or ex plosive gases The unit may be damaged or explosion may occur Do not keep or use this unit in an environment where it will be directly illuminated by sunshine or where it condensa tion If you do it may be deformed its insulation may damaged or it may no longer function according to specification e Do not place the meter on or around hot objects 70 C 158 F It may cause damage to the ca
5. TERFERENCE This instrument uses magnetic field method to measure the coating thickness on ferrous metal base If this meter was placed in the environment with 20mG mini Gauss or above the accuracy would be affected Suggest that the meter should to put far away from the interfered source at least 30cm Electromagnetic field strength unit mini Gauss 10 100 CD Display Fan 100 1000 Reading Lamp 400 4000 SPECIFICATION ELECTRICAL Detectable Substrate Material Ferrous metal iron steel and Non Ferrous metal copper aluminum zinc bronze brass etc Ferrous Thickness Range 0 to 80 0mils 0 to 2000um Non Ferrous Thickness Range 0 to 40 0mils 0 to 1000um Display Resolution 0 1 mils lum Ferrous Accuracy 4dgts on 0 to 7 8mils 3 4dgts on 7 9mils to 39 0mils 5 4dgts on 39 1mils to 80 0mils 10dgts on 0 to 199um 3 10dgts on 200um to 1000um 5 10dgts on 1001um to 1999um Non Ferrous Accuracy 4dgts on 0 to 7 8mils 3 4dgts on 7 9mils to 40mils 10dgts on 0 to 199um 3 10dgts on 200um to 1000um Response Time second GENERAL Operating Environment 13 F to 122 F 25 C to 50 C at lt 75 R H Storage Temperature 13 F to 140 F 25 C to 60 C 0 to 80 R H with battery removed from meter Temperature Coefficient 0 1x specified accuracy C lt 18 C or gt 28 C Auto Power Off 1 minute
6. ace tightly Wait for the reading to appear and meas urement is completed one Beep sound announced then you can move the Gauge away Press 0 button no longer than 2 seconds to calibrate substrate material or foil LCD displays 0 um One Point Calibration 1 Press the sensing tip of the probe to contact coated sur face tightly Wait for the reading to appear and meas urement is completed one Beep sound announced then you can move the Gauge away 2 Press CAL button over 2 sec into calibration mode LCD will blink CAL icon 3 In calibration mode use A or V button to adjust readings until it matches the standard s thickness 4 Press CAL button to exit one point calibration and return to measuring mode CAL off Two Point Calibration During two point calibration the foil and standard coating plate 4 0 mils 102um can be replaced by uncoated substrate and a standard coating plate with known thickness 1 Press the sensing tip of the probe to contact Uncoated surface tightly Wait for the reading to appear and meas urement is completed one Beep sound announced Press Q button no longer than 2 seconds to calibrate substrate material or foil LCD displays 0 um 2 Press the sensing tip of the probe to contact coated sur face tightly Wait for the reading to appear and measure ment is completed one Beep sound announced then you can move the Gauge away Press
7. se e If the meter is exposed to significant changes in ambient temperature allow 30 minutes for temperature stabiliza tion before taking measurement e Condensation may form on the sensor when going from cold to hot environment Wait for 10 minutes for condensation to dissipate before taking measurements e This unit is not constructed to waterproof and dustproof Do not use it in a wet or very dusty environment e In order to take accurate meas urement make sure the sensing tip contacts the coated surface tightly without tilting Please make sure there is no air bubbles between substrate and coating e Substrate Zeroing Calibration Must be implemented for each use e Two point calibration Must implement for frequent testing points to increase measuring accuracy e The enclosed zeroing plates are only suitable for the use of calibration of coating thickness meter itself Apart from that the meter should be performed two point cal ibration methods to get accurate readings before use The zeroing on specific material substrate still needs to be done before taking formal measurements such as Iron Steel Bronze Copper Nickel Zinc and SUS304 and so on which is to avoid the measuring errors that cause by the difference of individual substrates The end users can get much more accurate measuring readings on the specific metal under test by doing two calibration methods WARNING ELECTROMAGNETIC FIELD IN
Download Pdf Manuals
Related Search
Related Contents
Longshine LCS-VE684 network media converter Nortel Networks BayStack 893-862-B User's Manual LOEWE Individual 40 Compose 3d Weekender 50 manual.cdr lbi-38985b - The Repeater Builder`s Technical Information Page Targus 12" Messenger 北上川上流流域下水道 北上浄化センター 3-1系水処理施設 Audiovox RCR3273R remote control Intempco LTX01 Transmitter Manual Copyright © All rights reserved.