JEOL 1400 TEM User Guide

image

Contents

1. Set the desired image magnification focus the image and spread the beam to achieve a uniformly bright image Insert the digital camera In the Camera View box check that the Camera Inserted box is ticked and that the Auto Exposure box Is ticked 4 Click on Start View If you need to change image magnification click on Stop View and remove the digital camera from the microscope column Reintroduce digital camera and click on Start View again In the Camera Acquire box check that the Auto Exposure box Is ticked Click on the Tools icon in the bottom right corner x Setup Search iw Aubo Exposure Exposure s 1 a Start VIEW Focus Loupe Mw Auto Survey i Camera Inserted i Camera Acquire Setup Record M Auto Exposure Exposure s 5 Stark Acquire y 6 28 2007 9 10 11 Check that Full CCD is selected Ta CoolSNAP1 Setup Check that Auto Exposure is Setup Record gt CCD Area Processing Frame Sum frames 2 selected and that Connections eee IS set to Gain Normalised Top E Corrections Gain Normalized Usually Binning 1 is ok S ae Good resolution can be obtained ieee by setting Frame Sum to 4 The specimen must not be Center 1 4 Center1 2 FullCCD drifting otherwise the 4 frame sane averaged image will be out of n f O C u S Advanced Settings If specimen drift is an issue a reduce the number of frames to be averaged The Gatan image file
2. BRIGHTNESS focus 6 28 2007 Correcting condenser lens astigmatism e Adjust the BRIGHTNESS knob for the smallest electron beam e Press COND STIG e Slowly turn BRIGHTNESS knob back and forth through focus e Use DEF STIG X and Y knobs to make electron beam round rather than elliptical at either side of focus e Press COND STIG again to deactivate Diffraction Mode e Turn the BRIGHTNESS knob to darken the image sufficiently e Press the DIFF switch e Adjust the MAG CAML knob to choose the desired camera length e Adjust the DIFF FOCUS knob to obtain the sharpest diffraction pattern e Press MAG 1 to return to normal imaging I maging difficulties e f the microscope imaging system is misaligned Press IMAGE SHIFT then NTRL Press PLA then NTRL Press COND STIG then NTRL Press OBJ STIG then NTRL Press DARK TILT then NTRL Press BRIGHT TILT then NTRL This reloads the default settings optimised during installation Emergency Shutdown e n an emergency open the narrow panel on the right side under the desktop console e Press the EM STOP switch e Notify a staff member immediately 6 28 2007
3. format is dm3 To save images as tif or jpg go to File Save display as and name the file When prompted choose actual resolution include annotations to save scale bar and save as grey scale Specimen exchange l S Se ee If soecimen has been tilted during imaging open STAGE CONTROL window from CONTROL menu and click STAGE NEUTRAL At top of left hand control panel press BEAM and wait until blinking stops Remove specimen holder from microscope column as described previously Load new grid into specimen cartridge Reload specimen holder into microscope as described previously Press BEAM and wait until blinking stops Begin imaging At end of session l 2 E a a 9 10 11 Minimise microscope magnification If specimen has been tilted during imaging open STAGE CONTROL window from CONTROL menu and click STAGE NEUTRAL Ensure digital camera is retracted and Stop View has been selected At top of left hand control panel press BEAM and wait until blinking stops Remove specimen holder from microscope column as described previously Remove grid from specimen cartridge Reload specimen holder into microscope as described previously To ensure your images are archived copy them to a folder under eol1400 in the Images folder of the EMU samba server M drive It may be worthwhile to also save a copy to your User folder on the M drive Logoff the EMU Server Turn both monitors off 6 28 2007 Advanced Users Notes Optimi
4. J EOL 1400 TEM User Guide Prior to getting started 1 Turn on Monitors 1 and 2 2 Login to EMU server at Monitor 2 3 At Monitor 1 open the TEM Centre software 4 Make sure that HT On amp Beam Ready are displayed in the HT Beam condition window 5 Open Beam Controller window from Control menu e Beam dark current should indicate e 44uA for 80kV e 55uA for 100kV e 65uUA for 120kV 6 If HT is off no beam current is displayed or you want to work at a higher or lower kV than that indicated please seek help from Jenny Sigrid or Katie 6 28 2007 Loading a specimen 1 Remove the specimen holder from the microscope e Pull out in a straight line until it stops e Turn anti clockwise until it stops e Pull out again for a short distance e Turn anti clockwise until it stops e Pull out toggle switch and flip to AIR e Wait 10 sec then remove holder from the microscope Set the switch to Pull out completely _ Wait for 30 seconds and _ remove the holder 2 Remove the specimen cartridge from the specimen holder e Hold specimen cartridge with blunt forceps e Insert levering rod into hole on specimen cartridge fastening jaw e Raise the jaw by gently tilting the levering tool backward 3 Insert grid into specimen cartridge e Use levering rod to hold back end of specimen cartridge down e Use blunt forceps to lever up the brass specimen retainer e Place a grid specimen side up into t
5. ark current should indicate 44uA for 80kV 55uA for 100kV or 65uA for 120kV e If HT is off and no beam current is displayed please seek help from Jenny Sigrid or Katie 2 At top of left hand control panel press BEAM and wait until blinking stops 3 Turn the brightness knob to see if an electron image appears on screen e If no electron image can be seen move the specimen in case you are situated over a grid bar e If you still can t see an electron image seek staff assistance 6 28 2007 4 A good starting point for getting underway is e Condenser aperture 2 second largest dot Press LOW MAG Press MAG 1 e Spot size set to 2 Remove objective aperture by turning anti clockwise to the red dot Locate region of interest on grid and centre to screen Introduce objective aperture 2 second largest dot 5 Centre the beam using the gun shift X and gun shift Y knobs 6 Begin imaging Focusing using the wobbler Roughly focus the image through the binocular head l 2 3 4 Press IMAGE WOBB X or Y Adjust the OBJ FOCUS knob so that the image stops oscillating and the double image is unified Press IMAGE WOBB X or Y again to deactivate Taking digital images using D G TAL MI CROGRAPH software The digital camera can be damaged by the bright flash that occurs during changes to magnification Please only insert the camera to take an image then remove again to move around the specimen or adjust image magnification
6. he retainer e Again use the levering rod to hold the specimen cartridge down e Use blunt forceps to clip the brass specimen retainer closed e The brass specimen retainer should lie completely flat when closed E as Bg Levering Clamping jaw r Guide hole Brass specimen pin retainer lever e7 e a z grid ce close grid retainer 6 28 2007 4 Reattach the specimen cartridge to the specimen holder e Hold the specimen cartridge with the blunt forceps e Insert the levering rod into the hole on specimen cartridge fastening jaw e Raise the jaw by gently tilting the levering tool backward e Align the guide hole over the guide SS pin and close jaw Guide Prea Si retainer Levering Clamping jaw r Guide hole 5 Insert specimen holder into microscope column e Align guide pin with grove e Push specimen holder in until it stops do not turn yet Push in completely e Pull out toggle switch and flip to PUMP orange LED will illuminate e Wait until green LED illuminates 1 min e Turn specimen holder clockwise until it stops small turn Se anih e Guide holder as it is drawn into the to PUMP microscope column e Turn clockwise until it stops larger turn e Very gently guide the holder into its final position Align guide pin and push in Generating an electron beam 1 Confirm that HT On and Beam Ready are displayed in the HT Beam condition window e Beam d
7. sing the electron image Changing the objective aperture The objective aperture has a significant effect on image contrast and resolution The smaller the aperture the higher the image contrast but the darker the image We have 20um 40um 60um and 120um apertures selectable Insert a specimen into the beam path Ensure an objective aperture is inserted Press DIFF on the right hand control panel Using DIFF FOCUS knob focus to create bright spot on screen Centre bright spot using DEF STIG X and Y knobs Press PLA and focus for smallest spot using BRIGHTNESS knob Centre the spot using SHIFT X and Y Select desired objective aperture Focus the aperture image using the DIFF FOCUS knob e Adjust aperture knobs A and B so bright spot is in the centre of the aperture e When done press DIFF and PLA to deactivate and return to normal imaging Changing the condenser aperture The condenser aperture controls the illumination area angle and brightness The smaller the aperture the higher the image quality but the darker the image We have 100um 200um and 300um apertures selectable e Adjust the brightness knob for the smallest beam spot e Centre the beam spot to the screen using SHIFT X and Y _ e Insert the desired condenser aperture e Slowly turn the brightness knob clockwise e f the beam moves off centre re centre it using the condenser aperture knobs A and B e Ensure electron beam converges and diverges centrally through