Troubleshooting a Pumping Latch Closed alarm
Contents
1. reserved Alaris and MedSystem Ill are trademarks or registered trademarks of CareFusion Corporation or one of its subsidiaries IF2296 01 0511 PDF
2. Troubleshooting a Pumping Latch Closed alarm MedSystem Ill infusion pump Pumping Latch Closed alarms can be reduced with the following practices STANDARD MORE START ON OFF DISPLAY OPTIONS STOP e Turn the pump on before inserting the cassette into the pump e With tubing down be sure to use a 45 degree upward motion to insert cassette into channel e Stop the channel before removing the cassette from the pump e Fully extend the cassette slide clamp when removing the cassette from the pump Figure A eho To HE mm Air in line fl detector Pump latch mechanism To address a Pumping Latch Closed alarm Figure 1 e Use only your finger to push down the closed pumping latch jaw until it snaps open Figure 2 e If the pumping latch jaw is visibly broken the channel may be disabled by pressing the Service key e Do not press the Service key unless you wish to disable the channel Pumping latch open Note Illustrations are shown with protective base cover removed For product support please contact Customer Advocacy at 888 812 3266 or email customerfeedback carefusion com For technical support please contact Instrument Technical Support at 888 812 3229 For product orders please contact Customer Order Management at 800 482 4822 CareFusion San Diego CA carefusion com CareFusion 2011 CareFusion Corporation or one of its subsidiaries All rights