SS420x Analog to Digital Converter Board User Guide

image

Contents

1. 3 Instrument Method Setup Specifying the Voltage Range and Trigger Type Table 7 Acquisition page parameters Sheet 2 of 2 Parameter Run Until the Device Ends Specify Time min Description Specifies that the SS420x A D converter board acquires data from the specified device until the device is turned off Selections Any any device AS autosampler Detector GC gas chromatograph LC liquid chromatograph MS mass spectrometer or Other device Specifies that the SS420x A D converter board acquires data for the specified time Default 5 0 minutes Range 0 1 to 100 000 0 minutes Specifying the Voltage Range and Trigger Type Use the Configuration page see Figure 9 to specify the voltage range and trigger type for your analog device Figure 9 Configuration page with the default settings Acquisition External Events Configuration Channels ES Trigger A l0 104 Trigger input Trig 1 Trigger on contact Closed v Ip D Io o O O al 5 5 lt Xp lt 4 4 4 Calibration Low offset High offset Low gain High gain Channel 4 0 0000 0 0000 0 0000 0 0000 ChannelB 0 0000 0 0000 0 0000 0 0000 ChannelC 0 0000 0 0000 0 0000 0 0000 ChannelD 0 0000 0 0000 0 0000 0 0000 Base Frequency Hz 10 Version D 16 Agilent SS420x A D Converter Board User Guide Thermo Scientific 3 Instrument Method Setup Specifying the Voltage Range and Trigger Type lt To specify the voltage range of t
2. M Cancel Help 8 Agilent SS420x A D Converter Board User Guide Thermo Scientific 2 Instrument Configuration 4 Select the COM port where the device is attached 5 Click OK to save the changes and close the SS420x Configuration dialog box 6 Click Done to close the Thermo Foundation Instrument Configuration window Note Before you can open the Thermo Xcalibur data system you must close the Thermo Foundation Instrument Configuration window You cannot open the Xcalibur data system while the Foundation application is running Thermo Scientific Agilent SS420x A D Converter Board User Guide 9 ee ie Instrument Method Setup This chapter describes how to set up an instrument method that uses the SS420x analog to digital board to acquire data from an analog device such as a detector or trigger an external device such as a fraction collector See Chapter 1 Hardware Connections for information on connecting the SS420x to an analog device such as a detector or an external device such as a fraction collector See Chapter 2 Instrument Configuration for information on adding the SS420x to your instrument configuration IMPORTANT If you want to use the SS420x to convert an analog signal from an unsupported analog detector the detector must have its own control panel or you must have an appropriate data system and a second data system computer to control the detector For example to download
3. You can connect an external device to one of the eight relay outputs on the back panel of the SS420x A D converter You can program up to 50 events during a run The events can occur at a delay time from 0 0 to 1000 000 seconds into the run lt To trigger an external device at a specified time 1 Click the External Events tab The External Events page appears see Figure 10 Agilent SS420x A D Converter Board User Guide Thermo Scientific Thermo Scientific 3 Instrument Method Setup Triggering an External Device with the SS420x 2 In the Events Description area do the following a In the Number of Events box type or select the number of events you want to control You can control up to 50 events In the Description box type a description of the multi event procedure you want to run The description can consist of up to 83 alpha numeric characters 3 In the Event Setup area set up an event as follows a b d In the Descriptions column type a description of the event in the Descriptions box In the Delay Time column type a delay time for the event The delay time determines when an event occurs The delay time equals zero when the SS420x starts acquisition or the MS detector sends a contact closure signal In the Digital Output in Use column select the digital output terminal that you want to trigger For a Trigger Type Closed Contact device the following occurs When the Digital Output I
4. System Computer lt To connect the SS420x to the Xcalibur data system computer 1 Turn off the data system computer 2 Connect the serial communication cable as follows see Figure 1 a Connect one end of the serial communication cable from the Xcalibur SS420x Interface Kit to the RS 232 port on the back panel of the SS420x A D converter b Connect the other end of the serial cable to the RS 232 port located on the back panel of the data system computer 3 Connect the power cable from the 9 V dc power supply included with the SS420x A D converter to the POWER inlet on the back panel of the SS420x A D converter 4 Restart the data system computer Figure 1 RS 232 connection between the SS420x analog to digital converter board and the data system computer Optiplex GX520 computer subject to change O O oly a o 55420x analog to digital converter USB e ENET RS 232 la fur lO RS 232 DB9 to DB9 ENET serial communication cable d iE O M J m m o O Agilent SS420x A D Converter Board User Guide Thermo Scientific 1 Hardware Connections Connecting the Trigger and Signal Cables Connecting the Trigger and Signal Cables The SS420x A D converter has four analog input channe
5. Trig 1 Trig 2 Trig 3 and Trig 4 For information on connecting the trigger terminal to the analog device see Connecting the Trigger and Signal Cables on page 3 Trigger on Contact Thermo Scientific Specifies the trigger type used to initiate acquisition Selections Closed and Opened Agilent SS420x A D Converter Board User Guide 17 3 Instrument Method Setup Triggering an External Device with the SS420x Table 8 Configuration page parameters Sheet 2 of 2 Parameter Description Calibration Calibration Displays the values set during the factory calibration of this device Base Frequency Hz Displays the base frequency in Hz of the SS420x A D Converter This value corresponds to the selection in the Frequency list on the Acquisition page Version This readback displays the version number of the SS420x A D Converter Triggering an External Device with the SS420x 18 Use the External Events page see Figure 10 to set up the SS420x to control external devices that require a contact closure such as gas sampling or column switching valves Figure 10 External Events page with the default settings Acquisition External Events Configuration Events description Number of Events Description i Description Event Setup Ae Digital Output in Use Event Descriptions Delay Time s 12345678 sandalia JEvent2 0 roten Exp O RARA Evna b FTrrrrry ja ja je Je ja Ja ala
6. computer desktop depending on the Xcalibur version do one of the following e For Xcalibur 2 0 x or lower choose Start gt Programs gt Xcalibur gt Instrument Configuration The Instrument Configuration window appears Note The Instrument Configuration window not shown has the same functionality as the Thermo Foundation Instrument Configuration window e For Xcalibur 2 1 x or higher with Thermo Foundation choose Start gt Programs gt Thermo Foundation gt Instrument Configuration The Thermo Foundation Instrument Configuration window appears see Figure 4 2 Under Available Devices double click the SSI SS420x A D converter button A copy of the SSI SS420x A D converter button appears in the Configured Devices pane see Figure 4 Thermo Scientific Agilent SS420x A D Converter Board User Guide 7 2 Instrument Configuration Figure 4 Thermo Foundation Instrument Configuration window with the A D converter added to the Configured Devices pane Thermo Foundation Instrument Configuration Device Types All Available Devices Configured Devices Accela 45 Accela PDA 551 55420x A D converter Accela 1250 Pump 551 55420x A D converter C oe J Ce 3 Under Configured Devices double click the SSI SS420x A D Converter button The SS420x Configuration dialog box appears see Figure 5 Figure 5 SS420x Configuration dialog box 554 20x Configuration Serial port com
7. e Related Documentation e Safety and Special Notices Contacting Us Related Documentation For the SS420x A D converter the data system provides Help and this manual as a PDF file SS420x Analog to Digital Converter User Guide lt To view product manuals For the Xcalibur 2 1 x data system choose Start gt Programs gt Thermo Instruments gt Manuals gt LC Devices gt SSI gt SS420x lt To open Help e From the SS420x A D converter view in the Instrument Setup window choose Help gt SS420x A D converter Help 551 55420x A D converter Help Instrument Setup Help Help On Current Itern Fi SSI 55420x A D converter Contents and Index Xcalibur Help Glossary How to use Online Help About Instrument Setup e Ifavailable for a specific window or dialog box click Help or press F1 for information about setting parameters Thermo Scientific Agilent SS420x A D Converter Board User Guide v Preface Safety and Special Notices Make sure you follow the precautionary statements presented in this guide The safety and other special notices appear in boxes Safety and special notices include the following CAUTION Highlights hazards to humans property or the environment Each CAUTION notice is accompanied by an appropriate CAUTION symbol IMPORTANT Highlights information necessary to prevent damage to software loss of data or invalid test results or might contain information that is critical for optima
8. time see Figure 8 If the SS420x is the start instrument data acquisition starts when all configured devices report Ready for Run If the SS420x is not the start instrument the channels in use are started by the input trigger line specified on the Configuration page Data acquisition ends at a specific time or when the run time for the specified device ends Figure 8 Acquisition page with the default settings Acquisition Extemal Events Configuration Number of channels in use Frequency Hz il 10 Channel Description Channel 1 Bs Co rm D Im Acquisition Time Bun until the device ends hd Specify Time min 5 0 lt To specify the data acquisition settings for the SS420x converter board 1 In the Number of Channels in Use box type or select the appropriate number of channels 2 In the Frequency Hz list select the data acquisition rate The selections are 10 25 30 50 60 100 and 120 Hz 3 In each available Channel Description box type the name of the analog device 14 Agilent SS420x A D Converter Board User Guide Thermo Scientific 3 Instrument Method Setup Specifying the Data Acquisition Settings 4 To specify the acquisition time do one of the following in the Acquisition Time area e Select the Run Until the Device Ends option to use a device to stop data acquisition Then select the device that controls the acquisition time The selections are Any any device AS autosam
9. DOCUMENTATIO SURVEY 2010 Thermo Fisher Scientific Inc All rights reserved ChromQuest and Xcalibur are registered trademarks of Thermo Fisher Scientific Inc in the United States Agilent is a registered trademark of Agilent Technologies Inc in the United States and other countries Thermo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited except with the written authorization of Thermo Fisher Scientific Inc The contents of this document are subject to change without notice All technical information in this document is for reference purposes only System configurations and specifications in this document supersede all previous information received by the purchaser Thermo Fisher Scientific Inc makes no representations that this document is complete accurate or error free and assumes no responsibility and will not be liable for any errors omissions damage or loss that might result from any use of this document even if the information in the document is followed properly This document is not part of any sales contract between Thermo Fisher Scientific Inc and a purchaser This document shall in no way govern or modify any Terms and Conditions of Sale which Terms and Conditions of Sale shall govern all conflicting information between the t
10. Roadmap view click the Instrument Setup icon The Thermo Xcalibur Instrument Setup window appears see Figure 7 Figure 7 SS420x A D converter view in the Instrument Setup window ES Untitled Thermo Xcalibur Instrument Setup File 55420x Help Acquisition External Events Configuration Number of channels in use Frequency Hz 10 551 55420x A D converter Description Channel 1 Channel 2 Channel 3 Channel 4 Acquisition Time Bun until the device ends MS Specify Time min 5 0 View bar 3 In the view bar click the SSI SS420x A D converter button The SS420x view appears with the Acquisition page displayed see Figure 7 To acquire data from an analog detector follow these procedures e Specifying the Data Acquisition Settings on page 14 e Specifying the Voltage Range and Trigger Type on page 16 To trigger an external device such as a fraction collector follow these procedures e Specifying the Voltage Range and Trigger Type on page 16 e Triggering an External Device with the SS420x on page 18 Agilent SS420x A D Converter Board User Guide 13 3 Instrument Method Setup Specifying the Data Acquisition Settings Specifying the Data Acquisition Settings When you are using the SS420x A D converter board to acquire data from an analog device use the Acquisition page to specify the number of acquisition channels the data rate frequency and the acquisition
11. a method to the Surveyor RI Detector Plus you must have a second data system computer with the ChromQuest data system installed The SpectraSYSTEM UV Vis detectors have front panel controls Contents e Opening the Instrument Setup View for the SS420x Board e Specifying the Data Acquisition Settings e Specifying the Voltage Range and Trigger Type e Triggering an External Device with the SS420x Opening the Instrument Setup View for the SS420x Board lt To open the Instrument Setup view for the SS420x A D converter 1 From the computer desktop open the Xcalibur data system e For Xcalibur 2 0 x or lower choose Start gt Programs gt Xcalibur gt Xcalibur The Home Page appears e For Xcalibur 2 1 x or higher with Foundation choose Start gt Programs gt Thermo Xcalibur gt Xcalibur Thermo Scientific Agilent SS420x A D Converter Board User Guide 11 3 Instrument Method Setup Opening the Instrument Setup View for the SS420x Board The Thermo Xcalibur Roadmap window appears see Figure 6 Figure 6 Thermo Xcalibur Roadmap view Thermo Xcalibur Roadmap EIER Eile Actions View Tools GoTo Help E Run Manager Check Devices Sequence Sample Name Working On Position Raw File Inst Method E 551 55420x A D converter Not Connected 12 Agilent SS420x A D Converter Board User Guide Thermo Scientific 3 Instrument Method Setup Opening the Instrument Setup View for the SS420x Board 2 In the
12. digital outputs Thermo Scientific em BCD INPUTS NO ELEIE lt 2a lt 2 lt EZEZE nonon RS 232 POWER e 9 0 L ANALOG INPUTS L RELAYOUTPUTS L STARTINPUTS 4 Digital outputs The meaning of the check box depends on the Trigger Type you specified for a Trigger Line on the Configuration page Closed Contact or Open Contact Trigger Type Closed Contact e When the Digital Output in Use check box is selected the external device receives a Closed Contact signal at the specified delay time e When the Digital Output in Use check box is not selected the external device receives an Open Contact signal at the specified delay time Trigger Type Open Contact e When the Digital Output in Use check box is not selected the external device receives an Open Contact signal at the specified delay time e When the Digital Output in Use check box is selected the external device receives a Closed Contact signal at the specified delay time Agilent SS420x A D Converter Board User Guide 21 CCC _ ie Index Numerics S 2 wire trigger contact closure connection for serial communication cable 2 terminal panel interface table 4 SS420x A D Converter Setup view opening 11 SS420x Configuration dialog box 8 A start instrument 14 Acquisition page 14 T C Thermo Xcalibur Roadmap view 12 trigger type specifying 16 cables serial communication 2 COM port computer 9 V computer co
13. ector see Table 5 Table 5 Wiring the autosampler and the MS detector for contact closure Autosampler MS detector 1 0 Inject out pin START IN Ground pin START IN ground pin Connecting the Relays to Control an External Device Use the eight digital outputs labeled RLY1 to RLY8 on the back panel of the SS420x A D converter board to control devices such as a fraction collector not controlled by the Xcalibur data system Note The SS40x converter board can trigger external devices through either a closed contact or open contact signal Refer to the reference manual that is supplied with your external device to determine its trigger type lt To connect an external device to the SS420x A D converter Connect a 2 wire trigger cable from the input terminals of the external device to the SS420x A D converter Follow the wiring scheme shown in Table 6 Table 6 Wiring an external device and the SS420x for contact closure External device SS420x A D converter RLY1 to RLY8 Input pin RLYA Ground pin RLY B Thermo Scientific Agilent SS420x A D Converter Board User Guide 5 es _ i Instrument Configuration The SS420x communicates with the Xcalibur data system through a serial communication link To set up this link you must specify the COM port on the computer where the serial communication cable is connected see Figure 1 on page 2 lt To check the data system configuration settings for the SS420x 1 From the
14. els convert the analog data to a digital format that the data system can process Additionally contact closure control for these devices includes four inputs and eight outputs In this chapter contact closure refers to open collector TTL logic or relay closure The Xcalibur data system can support up to four SS420x units however only one unit can be used at a time IMPORTANT If you want to use the SS420x to convert an analog signal from an unsupported analog detector the detector must have its own control panel or you must have an appropriate data system and a second data system computer to control the detector For example to download a method to the Surveyor RI Detector Plus you must have a second data system computer with the ChromQuest data system installed Contents e Connecting the SS420x to the Xcalibur Data System Computer e Connecting the Trigger and Signal Cables e Connecting the Relays to Control an External Device Table 1 lists the items in the Xcalibur SS420x Interface Kit Table 1 Xcalibur kit used with the SS420x Part number Description of kit OPTON 21721 Xcalibur SS420x Interface Kit e SS420x serial cable e 2 wire trigger cable for contact closure e Power supply e Xcalibur Additional 4 Port Serial Kit Thermo Scientific Agilent SS420x A D Converter Board User Guide 1 2 1 Hardware Connections Connecting the SS420x to the Xcalibur Data System Computer Connecting the SS420x to the Xcalibur Data
15. er and Signal Cables Figure 3 Wiring diagram for triggering data acquisition and collecting data from an analog device Tu Signal cable annuun Trigger cable Analog device Xcalibur controlled MS detector Xcalibur controlled autosampler El Y To connect the trigger and signal cables 1 Connect a 2 wire signal cable from the SS420x A D converter to the analog device see Table 2 To connect more than one analog device to the SS420x A D converter use a separate channel Channel A to Channel D for each device Table 2 Wiring an analog device and the SS420x for A D data acquisition SS420x A D converter analog inputs Analog device 0 to 1 V or 0 to 10 V output CH1 Signal output pin CHI Ground pin 2 Connect a 2 wire trigger cable from the analog device to the autosampler see Table 3 Table 3 Wiring the LC autosampler and the analog device for contact closure Analog device Autosampler Start in pin Inject out pin Ground pin Ground pin 3 Connect a 2 wire trigger cable from the autosampler to the SS420x see Table 4 4 Agilent SS420x A D Converter Board User Guide Thermo Scientific 1 Hardware Connections Connecting the Relays to Control an External Device Table 4 Wiring the LC autosampler and the SS420x for contact closure Autosampler SS420x A D converter START1 to START4 Inject out pin STARTI Ground pin GNDI 4 Connect a 2 wire trigger cable from the autosampler to the MS det
16. he analog channels and the trigger type for each channel 1 In the SS420 x A D converter view click the Configuration tab The Configuration page appears see Figure 9 2 In the Channels area select the voltage range for the channel you are using to acquire data e Select 0 1 V if the output signal from the analog device is between 1 and 1 V e Select 0 10 V if the output signal from the analog device is between 10 and 10 V 3 In the Trigger Input list select the appropriate trigger The selections are Trig 1 Trig 2 Trig 3 and Trig 4 If you want to use a device other than the autosampler to start data acquisition select the trigger that is connected to the device 4 In the Trigger on Contact list select Closed Contact or refer to the analog device reference manual to determine the trigger type setting Note To determine whether the device is triggered by a contact closure signal or an open contact signal refer to its reference manual Configuration Page Parameter Descriptions Table 8 describes the instrument method parameters on the Configuration page Table 8 Configuration page parameters Sheet 1 of 2 Parameter Description Channels Range Specifies the range of voltages that are expected from the specified channel The range can either be 0 to 1 VDC or 0 to 10 VDC Trigger Trigger Input Specifies the terminal on the back panel of the SS420x used to initiate acquisition Selections
17. he events list Default Description The user specified description can consist of up to 83 alpha numeric characters Event Setup Specifies the description delay time and the 8 digital outputs for each activated event The following figure and description show an example of how to use this table Event Setup Digital Output in Use Event Descriptions Delay Time s lee cor thea 1 OrenVvevea 5 ROO OrenVaves o RECTO E A A a Een a el e el el el el ol es PFrrrrrr 1 When the acquisition is started the base time is set to zero n m N 2 Valve A and B are set to open at 5 seconds and 30 seconds respectively Valve A and B are connected to the relay output connectors 1 and 2 so that the SS420x A D converter can control them during the run 3 The two valves are turned off after 600 seconds Event Displays the event number When the number of events exceeds 5 you can use the vertical scroll bar to display the hidden events You can specify from 0 to 50 external events 20 Agilent SS420x A D Converter Board User Guide Thermo Scientific 3 Instrument Method Setup Triggering an External Device with the SS420x Table 9 External Events page parameters Sheet 2 of 2 Parameter Description Descriptions Describes a given event Delay Time s Specifies the time in seconds after the start of acquisition that an event is to occur Digital Output in Use Use these check boxes to program the eight external
18. l performance of the system Note Highlights information of general interest Tip Highlights helpful information that can make a task easier vi Agilent SS420x A D Converter Board User Guide Thermo Scientific Preface Contacting Us There are several ways to contact Thermo Fisher Scientific for the information you need lt To contact Technical Support Phone 800 532 4752 Fax 561 688 8736 E mail us techsupport analyze thermofisher com Knowledge base www thermokb com Find software updates and utilities to download at mssupport thermo com lt To contact Customer Service for ordering information Phone 800 532 4752 Fax 561 688 8731 E mail us customer support analyze thermofisher com Web site www thermo com ms lt To copy manuals from the Internet Go to mssupport thermo com and click Customer Manuals in the left margin of the window To suggest changes to documentation or to Help e Fill out a reader survey online at http www surveymonkey com s aspx sm R7gKOvhLXn3NTkpK2BefHQ_3d_3d e Send an e mail message to the Technical Publications Editor at techpubs lcms thermofisher com Thermo Scientific Agilent SS420x A D Converter Board User Guide vii ee _ i Hardware Connections The SS420x A D converter manufactured by Agilent Technologies Inc offers four 20 bit resolution analog to digital channels that you can use to acquire data from analog devices that Xcalibur does not support These chann
19. ls Channel A to Channel D and four start input terminals that you can use to acquire analog data from analog devices not controlled by the Xcalibur data system Figure 2 shows the analog input channels and the start inputs on the back panel of the A D converter Figure 2 Back panel of the SS420x A D converter Gan BCD INPUTS CE om eor lt 2xSx lt EZEZE non on RS 232 POWER L ANALOG INPUTS Ll RELAYOUTPUTS 1 L STARTINPUTS Signal cable terminals Trigger cable terminals Note The following is a general procedure for connecting up to four analog devices to the SS420x A D converter Your particular application might require a different procedure or a different configuration of devices Acquiring data from an analog device requires the following connections e A 2 wire signal cable from the analog device to the SS420x A D converter e A 2 wire trigger cable contact closure from one to four analog devices to the autosampler e A 2 wire trigger cable contact closure from the autosampler to the SS420x A D converter e A 2 wire trigger cable contact closure from the autosampler to the MS detector Figure 3 shows a general wiring diagram for triggering data acquisition and collecting data from an analog device such as a detector not supported by the Xcalibur data system Thermo Scientific Agilent SS420x A D Converter Board User Guide 3 1 Hardware Connections Connecting the Trigg
20. n Use check box is selected the external device receives a closed contact signal at the specified delay time e When the Digital Output In Use check box is not selected the external device receives an open contact signal at the specified delay time For a Trigger Type Open Contact device the following occurs e When the Digital Output In Use check box is selected the external device receives an open contact signal at the specified delay time e When the Digital Output In Use check box is not selected the external device receives a closed contact signal at the specified delay time Repeat steps step 3a through step 3c for each event Agilent SS420x A D Converter Board User Guide 19 3 Instrument Method Setup Triggering an External Device with the SS420x External Events Page Parameter Descriptions Table 9 describes the instrument method parameters on the External Events page Table 9 External Events page parameters Sheet 1 of 2 Parameter Description Events Description Number of Events Specifies the number of external digital events that you want to occur Default 0 Range 0 to 50 The data system activates the number of events numbered event rows that correspond to the number selected For example if you select 3 events the data system activates Event 1 Event 2 and Event 3 so that you can enter an event description and an event delay time and program the 8 digital outputs for each event Description Describes t
21. nnecting the SS420x to 2 voltage range specifying 16 Configuration page 16 connection kit 1 X Xcalibur data system interface kit 1 data acquisition settings 14 digital outputs 5 E external device triggering 18 External Events page 18 F Foundation Instrument Configuration application 7 K kit connection 1 P power inlet SS420x 2 relay contacts 5 RS 232 connections 2 Thermo Scientific Agilent SS420x A D Converter Board User Guide 23
22. pler Detector GC gas chromatograph LC liquid chromatograph MS mass spectrometer or Other device or e Select the Specify Time min option to stop data acquisition after a specified time Then type or select the time in the associated box The range is 0 1 to 100 000 0 minutes To specify the voltage range and trigger type for the analog device go to Specifying the Voltage Range and Trigger Type on page 16 Acquisition Page Parameter Descriptions Table 7 describes the instrument method parameters on the Acquisition page Table 7 Acquisition page parameters Sheet 1 of 2 Parameter General Controls Description Number of Channels in Use Frequency Hz Specifies the number of analog input channels to be monitored Selections 1 2 3 or 4 The data system activates the number of channels A B C and D that correspond to the number selected For example if you select 3 channels the data system activates channels A B and C Specifies the data acquisition rate in Hz data points per second Default 10 Hz Selections 10 25 30 50 60 100 and 120 Hz Channel Description Specifies a name for each active channel Default 1 2 3 and 4 The user specified name can consist of up to 76 alpha numeric characters Acquisition Time Specifies how long analog data is to be acquired by the data system Thermo Scientific Agilent SS420x A D Converter Board User Guide 15
23. wo documents Release history Revision A released January 2007 Revision B released January 2009 Revision C released April 2010 Software version Xcalibur 2 0 x or higher For Research Use Only Not for use in diagnostic procedures A WE Contents Prelate APPO ee Pere rr eee rere re v Related Documentation ra a A ied ooo A v Safety and Special Notices pss i ota es sense vi Contacting US usina ar a a GERE vii Chapter1 Hardware Connections 0 0cce cece ee eee eee eee eee nun nenn 1 Connecting the SS420x to the Xcalibur Data System Computer 2 Connecting the Trigger and Signal Cables nen ne 3 Connecting the Relays to Control an External Device o ooooooooooo 5 Chapter2 Instrument Configuration zures nsenennnnnnnnnnn nennen nun nenn 7 Chapter3 Instrument Method Setup 0c cece eect eee nennen nennen 11 Opening the Instrument Setup View for the SS420x Board 11 Specifying the Data Acquisition Settings ua seb keene eae new ewe 14 Specifying the Voltage Range and Trigger Type 00 0000s eee 16 Triggering an External Device with the SAME 01 00 un deena 18 MER ren armen 23 Thermo Scientific Agilent SS420x A D Converter Board User Guide iii ee Preface This guide describes how to use the SS420x analog to digital A D converter to acquire data from an analog device or to trigger external devices that the data system does not control Contents