Home
Operating Instructions
Contents
1. d Remove the two springs and two balls from the stop ring e Replace the shaft back into the body shaft fits only one way f Follow the steps in Section 8 2 for reassembly SIX 9 0 OPERATING SUGGESTIONS AND TROUBLESHOOTING 9 1 LEAKAGE If you see liquid between the stator and stator ring tighten the pressure adjusting nut as explained in Section 7 0 Replace the rotor seal and stator face assembly if the leak continues 9 2 USE OF AQUEOUS BUFFERS OR SALT SOLUTIONS To prevent the formation of salt crystals in the valve flush out the passages and vent line s with water after using salt solutions 9 3 ACCURACY OF SAMPLE LOOPS Sample loop sizes are not actual values The actual volume can differ by 10 for a 20 uL loop and 5 for a 10 mL loop There is a greater difference for smaller loops 10 0 RECOMMENDED SPARE PARTS Rheodyne offers a RheBuild Kit including all the necessary parts tools and instructions for Model 9010 and offers genuine Rheodyne replacement parts to maintain the quality performance of your valve 9010 999 RheBuild Kit for Model 9010 7030 015 Rotor Seal for Model 9030 7060 074 Rotor Seal for Model 9060 3710 008 Rotor Seal for Models 3000 and 3000 038 3030 005 Rotor Seal for Models 3030 and 3030 038 8125 094 Stator Face Assembly for Model 9030 9060 015 Stator Face Assembly for Model 9060 GC Shaft End Model 9060 11 0 WARRANTY All Rheodyne products are warr
2. e Fax 707 588 2020 www idex hs com AnalyticalO Column e 20 Detector Detector Fig 7 Five column selection using two Model 9060 valves The tube connecting Port 6 in the valves is used as a flush out line 8 0 MAINTENANCE The only parts that may need eventual replacement are the rotor seal and stator face assembly Abrasive particles in the sample can damage the rotor seal and stator face assembly surfaces Genuine Rheodyne parts are easily replaced by the following instructions 8 1 DISASSEMBLY To disassemble the valve refer to Figure 8 and proceed as follows a Remove the handle assembly by loosening the handle set screw s b Remove the valve if mounted on a panel c Remove the three stator screws d Remove the stator support ring Type 90 Pre Column B S Waste Detector AnalyticalO Column e A 5y Position B valves stator stator face assembly and stator ring from the body e Pull the rotor seal off the pins f Leave the isolation seal and bearing ring in place On Model 9010 remove the isolation seal 8 2 REASSEMBLY To reassemble the valve refer to Figures 8 and 9 and proceed as follows a On Model 9010 put the new isolation seal on the shaft assembly with the open side facing the handle b Orient the new rotor seal as shown in Figure 9 For six position valves orient the rotor seal on the rotor relative to the arrow on t
3. EAKAGE OR HIGHER PRESSURE OPERATION A single pressure adjusting nut at the handle end of the valve is used for pressure adjustment If you need operation up to a higher pressure or if there is a leak between the stator and stator ring remove the handle set screws This gives access to the pressure adjusting nut Using a wrench tighten the adjusting nut about 1 20th turn Use the red dot and punch mark as guides Column A Sample Injector Detector Column B Position A Sample Injector If there is still leakage at this new setting repeat the process Finish by retightening the handle set screw into the hole on the shaft Replace the rotor seal and stator face assembly if the leak continues Note When the valve is not panel mounted the pressure adjusting nut can be hard to turn In this case loosen the three stator screws a 1 4 turn before adjusting the nut Retighten the stator screws before testing for leaks Note Model 9010 Injector only Vent Line 6 should be the same horizontal level as the luer tube to avoid siphoning A siphoning leak will stop when the vent line and sample loop are empty A leak due to a damaged rotor seal will continue Column A Detector Column B Position B Fig 5 Two column selection using Model 3000 3000 038 or 9010 without the loop Page 2 of 4 2320662B 10 08 Rheodyne LLC e A Unit of IDEX Corporation e 600 Park Court e Rohnert Park CA 94928 707 588 2000
4. also includes a 20 uL sample loop in a separate bag e Hex Keys e Mounting Screws e Luer Tube Model 9010 only 3 0 SPECIFICATIONS e Maximum Operating Pressure Models 3000 and 3030 28 MPa 276 bar 4000 psi All other models 34 MPa 345 bar 5000 psi e Maximum Operating Temperature 50 C e Flow Passage Diameters Type 90 0 3 mm 0 013 and 0 5 mm 0 018 Type 3000 1 0 mm 0 040 e Wetted Surfaces Models 9010 9030 9060 PEEK ceramic and an inert polymer Models 3000 038 3030 038 stainless steel and PEEK Models 3000 3030 PEEK Position A Position B Models 3000 and 9010 Switching Valve Two Position Six Port Position B Position A Models 3000 and 9010 with Connecting Loop Four Way Position A Position B Models 3030 and 9030 Double Three Way gt NGas 9 Wires Model 9060 Six Position Seven Port Fig 1 Flow diagrams of Rheodyne switching valves The stainless steel versions have the same flow diagrams 4 0 IMPORTANT SAFETY NOTICES 4 1 Warning Using 9010 as an injector When using sample loops larger than 100 uL shield yourself from mobile phase coming out of the needle port when the valve is turned from INJECT to LOAD Example 1 mL loop ejects 20 uL upon decompression from 19 MPa 200 bar 2898 psi 4 2 Warning Burst pressure of PEEK tubing depends on dimensions solvents temperature and duration of exposure A pinhole rupture in PEEK tubing can emit a high
5. anted against defects in materials and workmanship for a period of one year following the date of shipment by Rheodyne Rheodyne will repair or replace any Rheodyne product that fails during the warranty period due to a defect in materials or workmanship at no charge to the customer The product must be returned to Rheodyne s factory in original packaging or equivalent transportation prepaid Damage occurring in transit is not covered by the warranty This limited warranty is Rheodyne s sole warranty of its products and all other warranties of merchantability or fitness for any particular purpose are hereby disclaimed Under no circumstances will Rheodyne be liable for any consequential or incidental damages attributable to a claimed failure of a Rheodyne product even if Rheodyne has been placed on notice of possibility of such damages Page 4 of 4 2320662B 10 08 Rheodyne LLC e A Unit of IDEX Corporation 600 Park Court Rohnert Park CA 94928 e 707 588 2000 e Fax 707 588 2020 www idex hs com
6. ation 600 Park Court Rohnert Park CA 94928 e 707 588 2000 e Fax 707 588 2020 www idex hs com 6 1 1 COMPLETE LOOP FILLING In complete filling the volume of sample injected is set by the volume of the loop this includes the valve passages This method produces the highest precision Overfill the loop with at least two to five loop volumes of sample Six to ten loop volumes will provide even better precision An excess of sample is needed because mobile phase near the wall of the loop is displaced slowly due to the laminar flow effect shown in Figure 2 To completely fill the loop see Figure 3 a See Warning 4 1 and turn to LOAD b Insert the syringe into the injection port c Load the sample d Leave the syringe in position and turn to INJECT 6 1 2 SUCTION LOADING The preceding method exposes the sample to the metal needle of the loading syringe Metal can be completely avoided by using a syringe to draw sample into the loop The steps are as follows see Figure 4 a In LOAD dip the tube attached to Port 6 into the sample b Insert an empty syringe into the injection port and draw sample into the loop c Leave the syringe in position and turn to INJECT The metal needle of the syringe will contact the sample if an excess of sample is drawn into the syringe but this excess sample is external to the sample loop and will not be injected The syringe can be used many times before it needs to be empti
7. cating Pin Stator Face Assembly Stator Stator Support Ring Type 90 valves only Stator Screws Type 90 valves only Shaft assembly includes shaft rotor and pins Fig 8 Exploded view of Rheodyne switching valves the two handle set screws into the hole on the shaft Page 3 of 4 2320662B 10 08 Rheodyne LLC e A Unit of IDEX Corporation 600 Park Court Rohnert Park CA 94928 e 707 588 2000 e Fax 707 588 2020 www idex hs com Notch Models 3000 038 and 9010 Stop Ring Rotor Pin Models 3030 038 and 9030 Fig 9 Correct rotor seal orientation as viewed from the stator 8 3 POSITION SENSING SWITCH IN MODEL 9010 The switch is a magnetic reed switch actuated by a magnet sealed inside the shaft The switch is rated for 100 V at 200 mA To replace or remove the switch a Remove the three stator screws b Remove the stator support ring stator stator ring and stop ring c Pull the switch out of the stop ring d Replace with new switch or leave the hole empty if the switch is not needed e Follow the steps in Section 8 2 to reassemble 8 4 DETENT MECHANISM POSITION VALVES ONLY If mounting the six position valve on an actuator the detent mechanism must be disabled a Remove the stator screws stator support ring stator stator ring and stop ring b Loosen the handle set screw s on the knob and remove the handle assembly c Push the shaft through the body
8. ed To load the loop with all of the available sample the loop should be at least four times the volume of sample loaded The loop is first filled with mobile phase via the dip tube then the whole sample is drawn into the dip tube and loop followed by more mobile phase The sample is now sandwiched between two zones of mobile phase in the loop Sample Mobile Phase Tube Wall Flow Fig 2 Laminar flow effect 6 2 MODEL 9010 AS A 4 WAY Model 9010 can serve as a switching valve and a 4 Way valve by simply removing or changing the position of the external sample loop see Figure 1 For 4 Way operation reposition the 20 uL sample loop to Ports land 5 Pressure Loading Pump e 3 4 Instrument Panel or Mounting Bracket Loop Filler Port E re Syringe Sample Loop Fig 3 Using a luer tube to fill the loop by pressure loading Suction Loading Dip Tube Pump 4 Column e 3 4 Instrument Panel or Mounting Bracket Loop Filler Sample Port je E rms Syringe Sample Loop Fig 4 Using a luer tube to fill the loop by suction loading In suction loading sample is sucked from a vial into the loop using a dip tube 6 3 SUGGESTED APPLICATIONS FOR SWITCHING VALVES Flush the valve with mobile phase before connecting the valve to system components Suggested applications for switching valve applications are shown in Figures 5 6 and 7 7 0 ADJUSTING FOR L
9. he shaft The rotor seal grooves face the stator c Replace the stator ring so that the pin in the stop ring enters the mating hole in the stator ring For Type 90 valves confirm that the position sensing switch falls into the mating hole in the stator ring d On Type 90 valves place the stator support ring onto the stator Be sure the stator support ring s bumpy inner contour fits into the mating contour of the stator e Place the pins in the new stator face assembly into the mating holes in the stator The three stator face assembly pins in the Type 3000 valves fit into the mating holes in the stator only one way f Replace the stator and stator face assembly on the valve so that the pin in the stator ring enters the mating hole in the stator g Replace the three stator screws Tighten each screw a 1 2 turn past fingertight h If panel mounted remount the valve onto the panel i Replace the handle assembly and tighten Shaft Assembly zii imi Cn Handle Assembly Stator Screws Type 3000 valves only Pressure Adjusting Nut Body Thrust Bearing Spring Washers 4 Body Locating Pin Position Sensing Switch Model 9010 only 60 Stop Ring Seal Pins 4 Bearing Ring Isolation Seal Rotor Seal Stator Ring Stator Lo
10. v w RHEODYNE Operating Instructions Type 90 and Type 3000 Switching Valves 1 0 DESCRIPTION Type 90 and Type 3000 valves are manual switching valves containing six or seven ports for a wide range of applications Type 90 ports accept 1 16 fittings Type 3000 ports accept 1 8 fittings and are for preparative scale applications All Type 90 valves Models 3000 and 3030 are PEEK versions while Models 3000 038 and 3030 038 are stainless steel versions Model 9010 contains a position sensing switch Model 9010 can be used as a manual sample injector with two positions A sample loop attached at Port 1 and Port 4 comes standard When the sample loop is repositioned to Port 1 and Port 5 Model 9010 can be used in a 4 Way configuration Removing the sample loop completely makes Model 9010 into a switching valve Model 3000 or 3000 038 can also be used as a 4 Way valve two position by connecting a loop to Port 1 and Port 5 Models 9030 3030 and 3030 038 are Double 3 Way valves two position and Model 9060 is a Six Position valve Figure 1 shows the flow diagrams of Type 90 and Type 3000 switching valves The circles represent the ports in the valve stator The dark grooves represent the connecting passages in the rotor seal See Figure 3 for the flow diagram of Model 9010 as an injector 2 0 SUPPLIED WITH THE VALVE Supplied in a separate bag are RheFlex Fittings sets for all ports and the following items Model 9010
11. velocity jet of fluid that can penetrate skin and eye tissue 4 3 Caution Do not mount the valve with the ports facing up Leakage due to a damaged rotor seal or loose fittings can cause the bearings to corrode 4 4 Caution Rinse the valve thoroughly after using buffer solutions to prevent salt crystals from forming which can cause damage to the rotor seal and stator face assembly 4 5 Caution When using PEEK valves use only RheFlex PEEK fittings sets in the stator ports Metal ferrules can cause irreparable damage to the PEEK stator 5 0 INSTALLATION All switching valves can be panel mounted Model 9010 as a sample injector connects to different system components See Caution 4 5 a Connect the luer tube to Port 5 and the vent line to Port 6 Place the injection port at the same horizontal level as Vent Line 6 to avoid siphoning b Connect the pump to Port 2 and the column to Port 3 Leave the column disconnected from the valve during initial flushing 6 0 OPERATION 6 1 INJECTIONS MODEL 9010 ONLY Before connecting the column to the injector flush the injector with mobile phase in both the LOAD and INJECT positions After flushing the injector turn to LOAD and connect the column With the accessory injection port P N 9012 or 9013 or the luer tube supplied you can load the injector by two methods Complete Filling and Suction Loading Page 1 of 4 2320662B 10 08 Rheodyne LLC e A Unit of IDEX Corpor
Download Pdf Manuals
Related Search
Related Contents
ECF2: Claims Data Warehouse (CDW) User Guide 取扱説明書 通信用保安器 LP-2111B 取扱説明書 - ご家庭のお客さま/大阪ガス FT KID WAY 02 Approx Cheesecake XL QUAD 8GB Black, Silver Installation and operating instructions for the KB-series ServerView Virtual Gigabyte GA-P43-ES3G Desa 2-Level User's Manual Copyright © All rights reserved.