Home
User's Guide RIGOL DS1000E, DS1000D Series Digital
Contents
1. Measurement Category The DS1000E DS1000D series Digital Oscilloscope is intended to be used for measurements in Measurement Category I Measurement Category Definitions Measurement Category I is for measurements performed on circuits not directly connected to MAINS Examples are measurements on circuits not derived from MAINS and specially protected internal MAINS derived circuits In the latter case transient stresses are variable for that reason the transient withstand capability of the equipment is made known to the user WARNING IEC Measurement Category I The input terminals may be connected to circuit terminal in IEC Category I installations for voltages up to 300 VAC To avoid the danger of electric shock do not connect the inputs to circuit s voltages above 300 VAC Transient overvoltage is also present on circuits that are isolated from mains The DS1000E DS1000D series Digital Oscilloscopes is designed to safely withstand occasional transient overvoltage up to 1000Vpk Do not use this equipment to measure circuits where transient overvoltage could exceed this level 2008 RIGOL Technologies Inc III User s Guide for DS1000E DS1000D Series RIGOL Safety Terms and Symbols Terms in This Guide These terms may appear in this guide WARNING Warning statements identify conditions or practices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that coul
2. 166 AkHz Time 5 ms Mark of digital filter Figure 2 10 The waveform when turning on digital filter 2008 RIGOL Technologies Inc 2 7 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 11 Table 2 4 The Filter menu ot Menu Turn on the digital filter Turn off the digital filter OF on Setup as LPF Low Pass Filter L Lo n Digital Filter l Setup as HPF High Pass Filter Setup as BPF Band Pass Filter F Filter Type f Setup as BRF Band Reject Filter 4D U limit Turn 4 knob to set high limit Lower limit Turn 4 knob to set low limit lt frequency gt te fo Back to higher level menu The followings are the same 5 Volts Div settings The Volts Div control has Coarse or Fine configuration The Vertical Sensitivity is 2mV div 10V div Coarse It is the default setting of Volts Div in a 1 2 5 step sequence from 2mV div 5mV div 10mV div 20mV div 10V div Fine This setting changes the vertical scale to small steps between the coarse settings It will be helpful to adjust the waveform in smooth steps Fine adjustment Time 16H Aus Fine adjustment data Figure 2 12 Fine configurations 2 8 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Coarse Fine Shortcut key To change Coarse Fine setting not only by menu but also by pressing vertical SCALE knob 6 To invert a waveform Invert turns the displayed waveform
3. Figure 3 7 Waveform display The oscilloscope triggers on the Odd field or Even field To avoid confusion when Odd field and Even field trigger simultaneously choose Odd field or Even field as in step 5 above 3 10 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Triggering on the Video Lines Press the key in the Trigger control area to see the trigger menu Press Mode to select Video Press Source to select CH1 as trigger source Press Polarity to select U Press Sync to select Line Num Turn 4 knob to trigger on a specified line number a ee a E Adjust the to set trigger level at the video sync pulse to get a stable trigger 8 Turn the horizontal knob to observe a complete waveform on the screen ee es Figure 3 8 Waveform display 2008 RIGOL Technologies Inc 3 11 User s Guide for DS1000E DS1000D Series RIGOL Example 8 FFT Cursor measurement FFT measurements include Amplitude measurement Vrms or dBVrms and Frequency measurement Hz Do these steps as follow 1 Press Cursor Manual 2 Press Type to select X or Y 3 Press Source to select FFT 4 Turn 4 4 knob to move the cursor to an interesting point Cursors Mode Cure 626ml ANY 1 22 Type Source Time SHH Aus Figure 3 9 The example of Cursor measuring FFT amplitude por Cursors LUFH i LB kH tui ode Curb 9 LAKH 4476 a BHA KHz 1 Ax 167uUs Type Source
4. Logic low voltage is lower than threshold setting 3 X Ignore Don t care If all the channels are ignored the oscilloscope won t be triggered 4 Rising Edge _ or Falling Edge Set the code as an edge of the channel rising edge or falling edge When the edge is appointed if code settings of other channels are all true the oscilloscope will be triggered on the appointed edge If no edge is appointed the oscilloscope will be triggered on the last edge whose code is true Key Points Edge of appointed code You can only appoint one code as edge If you have appointed an edge then appointed another edge in a different channel and the first appointed edge will be set to X Ignore 2 46 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Duration Trigger Only for DS1000D Series Duration Trigger Mode Trigger in appointed time when code terms are satisfied Press Trigger MENU gt Mode Duration to display the duration trigger menu as follows Figure 2 61 Table 2 37 The Duration Trigger Menu Page 1 2 foe Menu Settings D15 D0 Choose digital channel for Duration Trigger Select De Code Qualifier Set time limit terms Set duration and limit symbol time IE Acquire waveform even no trigger occurred Sweep Auto Acquire waveform when trigger occurred Sweep Normal l i l When trigger occurs acquire one waveform Single Set Up and t
5. in this guide If the contents are incomplete or damaged please notify your RIGOL Sales Representative 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL The Panels and User Interface The first thing to do with a new oscilloscope is to know its front panel This chapter helps to be familiar with the layout of the knobs and buttons and how to use them Read the chapter carefully before further operations 1 Front Panel Figure 1 1 Front Panel the knobs are used most often and are similar to the knobs on other oscilloscopes The buttons allow you to use some of the functions directly but also bring up soft button menus on the screen which enable the access to many measurement features associated with advanced functions mathematics and reference or to run control features The front panel of DS1000E a D a p MENU SAE O16 Sle _ pm VERTICAL Se HORIZONTAL E TRIGGER A o CH2 l gt 6 U J REF 0 i _ _ FORCE lt v mV S nS p CH z FE 0 OO 5 A eA Ss gt ATT J KN Xe J e The front of DS1000D
6. 500MSa s 25GSa s 10GSa s Averages The waveform will be displayed one time while all the channels finish N times Sample N could be selectable from 2 4 8 16 32 64 128 and 256 Input Coupling DC AC GND Input Impedance IMQ 2 The input capacity is 18pF 3pF Probe Attenuation Factors 1X 5X 10X 50X 100X 500X 1000X 400V DC AC Peak 1MQ input impedance Maximum Input Voltage j 9 40V DC AC Peak 2 Time Delay between Channel 500ps typical Sample Rate Real Time 13 65Sa s 1GSa s Range 13 65Sa s 25GSa s oe Simo oe Channel Mode Sample rate Memory Memory Depth Depth normal long record or lower asec woe or lower or lower Scanning Scanning Speed 2ns div 50s div DS1102X 5 2 2008 RIGOL Technologies Inc ser s Guide for A erles RIGOL Sec div 1 2 5 Sequence Sample Rate and Delay Time 50ppm any interval gt 1ms Accuracy Delta Time Single shot 1 sample interval 50ppm x reading 0 6 Measurement ns Accuracy gt 16 averages 1sample interval 50ppm x reading 0 4 Full Bandwidth ns Vertical A D converter 8 bit resolution all channels samples simultaneously Volts div Range 2mV div 10V div at input BNC Maximum Input Maximum input voltage on analog channel CAT I 300Vrms 1000Vpk instantaneous overvoltage 1000Vpk CAT II 100Vrms 1000Vpk RP2200 10 1 CAT II 300Vrms RP3200 10 1 CAT II 300Vrms RP3300 10 1 CAT II 300Vrms Offset Range 40V 250mV div 10
7. Connecting to the Ground 1 12 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL To Display a Signal Automatically The oscilloscope has an automated feature to display the input signal best fit The input signal should be 50Hz or higher and a duty cycle is greater than 1 Use Auto Setting 1 Connect a signal to the oscilloscope as described above 2 Press AUTO The oscilloscope will automatically set up VERTICAL HORIZONTAL and TRIGGER controls to display the input signal Adjust the controls manually to get the best results if necessary 2008 RIGOL Technologies Inc 1 13 User s Guide for DS1000E DS1000D Series RIGOL To Understand the Vertical System Figure 1 18 shows the VERTICAL controls CH1 CH2 MATH and buttons and vertical lt 2POSITION knobs Following the exercise of the buttons knobs and the status bar to be familiar with the vertical parameters settings VERTICAL POSITION Pa e a Po le u e 9 DELEN Figure 1 18 The vertical system 1 Moves the signal vertically on the display with the knob When turning the knob a voltage value is displayed for a short time indicating its value with respect to the ground reference located at the center of the screen Also notice that the ground symbol on the left side of the display moves in conjunction with the POSITION knob Measurement hints If the channel is DC coupled measuring the DC compone
8. Measure overshoot in percentage Overshoot of an edge Measure preshoot in percentage of an edge 2008 RIGOL Technologies Inc 2 89 User s Guide for DS1000E DS1000D Series RIGOL 10 Time Measurement Parameters Figure 2 126 Table 2 81 The Time Measurements menu Page 1 4 Measure Period of a waveform Measure Frequency of a waveform Rise Time a Fall Time r Measure Rise Time of a rising edge Fall time Measure Fall Time of a falling edge Figure 2 127 Table 2 82 The Time Measurements menu Page 2 4 Menu _ as oes 7 Ta Width Measure Pulse Width of a pulse wave Width Measure Pulse Width of a pulse wave Measure Duty Cycle of a pulse wave lt n Measure Duty Cycle of a pulse wave ridth te cele Duty aL Duty Figure 2 128 Table 2 83 The Time Measurements menu Page 3 4 Settings Delay1 2 ra Measure the signals delay between two channels at the rising edge ra Measure the signals delay between two channels at the falling edge Delay1 2 2 90 2008 RIGOL Technologies Inc sers Guide for erles RIGOL Figure 2 129 Table 2 84 The Time Measurements menu Page 4 4 Measure the phase deviation Phas1 2 between two channels at the rising edge Measure the phase deviation between two channels at the falling edge Phas1 2 NOTE The results of the automatic measurements will be displayed on the bottom of the screen Maxim
9. NOTE The volts div and horizontal time base can be adjusted in a fixed limit That is to zoom in out the signal in vertical and horizontal directions 2008 RIGOL Technologies Inc 2 103 User s Guide for DS1000E DS1000D Series RIGOL Chapter 3 Application amp Examples Example 1 Taking Simple Measurements The function is used to observe an unknown signal to display measure frequency and peak to peak amplitude To quickly display a signal please do the steps as follow 1 Set the probe and the channel attenuations to 10X 2 Connect signal to CH1 with the probe 3 Press the button The oscilloscope sets the vertical horizontal and trigger controls at the best status automatically To optimize the waveform display adjust these controls manually to meet the requirements Selecting Automatic Measurements The oscilloscope takes automatic measurements on most signals To measure the frequency and the peak to peak amplitude do these steps as follows 1 Measure peak to peak amplitude Press Measure Source gt CH1 to set measurement source Press Voltage Vpp to select peak to peak measurements and the result will be displayed on the screen 2 Measure frequency Press Measure gt Source CH1 to set measurement source Press Time gt Freq to select frequency measurements and the result will be displayed on the screen NOTE The frequency and peak to peak measurements are shown on the screen and are updated
10. Standard Select Video standard T NTSC Set Lp Set Up To go to set up menu see table 2 39 2 44 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Figure 2 59 Table 2 35 The Alternate menu Trigger Mode Video All Lines Odd Field or Even Filed Page 2 2 ae Odd field Select to trigger on odd field or even an Even field field Standard PAL SECM Standard Select Video standard NTSC set Up Set Up To go to set up menu see table 2 39 ALL lines Trigger on all lines Line Num Trigger on an specified line 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series 2 45 RIGOL Pattern Trigger Only for DS1000D Series Pattern Trigger Mode Pattern trigger identifies trigger terms by checking appointed code The code is logical relationship of all channels with high H low L and ignore X Press Trigger MENU Mode Pattern to display the pattern trigger menu as follows Figure 2 60 Table 2 36 The Pattern Trigger Menu aa cer osor select D15 015 00 Choose digital channel for Pattern trigger Select High ners Low sas Code Ignore eee Rising Edge aie Falling Edge i Acquire waveform even no trigger occurred Normal Acquire waveform when trigger occurred Single When trigger occurs acquire one waveform and then stop Sweep Set Up ene To go to set up menu see table 2 40 1 H High Logic high voltage is higher than threshold setting 2 L Low
11. Time SHH Bus Figure 3 10 The example of Cursor measuring FFT frequency 3 12 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Example 9 Pass Fail Test When running test function the oscilloscope automatically examines the input signal compared to the already built waveform mask If the waveform touches the mask a Fail occurs otherwise the test passes When needed a programmable output can be used for external automatic control applications such as the automatic failure product kick out in a production line The output is built in as a standard feature and is optically isolated Do the steps as following Press Utility Pass Fail Press Enable Test and select ON 1 2 3 Press Mask Setting Load 4 Press Load to recall the saved mask or press X Mask and Y Mask to adjust the horizontal limit and vertical limit then press Create Mask to create a new mask Press Output to select the expected outputting waveforms 6 Press Operate to start the test Total 773 wf Source E Operate Msg Display City Time 2 6ebuUS Figure 3 11 Waveform display 2008 RIGOL Technologies Inc 3 13 User s Guide for DS1000E DS1000D Series RIGOL Example 10 Triggering on a Digital Signal The triggering way especially to the Digital Signal which the Pattern and Duration triggering are use for to carry on when triggering system was being used The two triggering way
12. e tront panel O y na be Measure Acquire E O E em S I pm VERTICAL a HORIZONTAL A 2 amp CH2 a C A scate REF ee ee sox Q 0 N ED C OW IROD eo CH CH 2 EXT TRIG k 4 m OROLI e j Z mans Ne CATI 2 Fi Front Panel Figure of DS1000E DS1000D Series Oscilloscope 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Multi function Common Menu Run Control RUN CONTROL Ns AUTO RUN STOP f Trigger Co Horizontal Control Vertical Co Me G eee i USB Host Logic Analyzer Port Signal Input EXT Trigger Probe Channel Input Compensation Figure 1 2 Front Panel Instruction 2 Rear Panel HAZARDOUS VOLTAGE INSIDE DO NOT REMOVE THE COVER UNLESS BY SPECIFIED PERSONNEL LINE 45 440Hz AC 100 240 S0VA MAX WARNING MAINTAIN GROUND TO AVOID ELECTRIC SHOCK RS232 Pass Fail out f 4 ISOLATEO X ay Figure 1 3 Rear Panel Figure of DS1000E DS1000D Series Oscilloscope 1 4 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL a Le HAZARDOUS VOLTAJE HFID CHJ REIT FI Michie THE COVER INLE ey SCF iRo EP ae Sa Li ks Sa i LOVE dope AD ike S RIG
13. 8 Adjust the vertical and knobs to a desirable waveform display 9 Apply the Ellipse method to observe the phase difference between the two channels See Figure 3 6 3 8 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Signal must be centered in X Figure 3 6 Ellipse method to observe the phase difference Sin A B or C D where 6 phase shift in degrees between the two signals From the formula above you could get 0 arcsine A B or arcsine C D If the main axis of the ellipse is at and IIl quadrant 0 must be in the range of O T 2 or 311 2 2T1 If the main axis is at Il and IV quadrant 6 must be in the range of 11 2 Tr or T1 3T1 2 2008 RIGOL Technologies Inc 3 9 User s Guide for DS1000E DS1000D Series RIGOL Example 7 Triggering on a Video Signal Test a video circuit in the DVD set Use video trigger to obtain a stable display Triggering on Video Fields To trigger on the video fields please do as the following steps 1 Press the MENU key in the Trigger control area to see the Trigger menu 2 Press Mode to select Video mode 3 Press Source to select CH1 as trigger source 4 Press Polarity to select U 5 Press Sync as Odd Field or Even Field 6 Adjust the to set trigger level at the video sync pulse to get stable trigger 7 Turn the horizontal knob to see a complete waveform on the screen Trigger Mode Source Time SHH Ans
14. Figure 2 43 Table 2 21 The Pulse Width Trigger Menu page 1 2 Trigger Mode Selects CH1 as trigger signal Select CH2 as trigger signal Select EXT TRIG as trigger signal Select a digital channel in D15 D0 as trigger source Only Setting for DS1000D series gt Pulse width T less than F gt L Pulse width more than L Pulse width equal to f Pulse width less than E24 Pulse width more than Pulse width equal to Source Source Wher To select pulse condition 2008 RIGOL Technologies Inc 2 37 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 44 Table 2 22 The Pulse Width Trigger Menu page 2 2 Trigger Acquire waveform even no trigger occurred Acquire waveform when trigger occurred When trigger occurs acquire one waveform and then stop NOTE The Pulse width adjust range is 20ns 10s When the condition is met it will trigger and acquire the waveform 2 38 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Slope Trigger Slop Trigger Mode Slope trigger sets the oscilloscope as the positive negative Slope trigger within the specified time Press Trigger IMENU Mode Slope to display the slope trigger menu as follows Figure 2 45 Table 2 23 The cae T aa Menu Page 1 2 Set channel 1 as trigger source Source Set channel 2 as trigger source Set EXT channel as trigger source Trigger Mode Source when T
15. Gl 8 Ol yaa amp position ONTAL TRIGGER lt Position D uve QY CYA ma Wk qd JATH 9 a SCALE D e REF l OA i7 OO CH1 EXT TRIG lel 6 CATI Figure 1 9 Turn on and Check the instrument A WARNING To avoid electric shock be sure the oscilloscope is properly grounded 2 Inputa signal to a channel of the oscilloscope DS1000E series 2 channels input 1 external trigger channel input DS1000D series 2 channels input 1 external trigger channel input 16 channels digital input Do the following steps 1 Set the switch on the probe to 10X and connect the probe to Channel 1 on the oscilloscope 1 8 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Align the slot in the probe connector with the button on the CH 1 BNC Push to connect and twist to the right to lock the probe in place Attach the probe tip and ground lead to the PROBE COMP connector Probe compensator Figure 1 10 Attach the probe Set the probe attenuation to 10X To do this push CH1 Probe 10X CH1 Coupling BY Limit OFF Probe Probe scale Digital Filter S Figure 1 11 Figure 1 12 Set attenuation on the probe Set attenuation in the menu Push the button Within a few seconds a squ
16. Offset The position of the practical trigger point relative to center of the memory Adjust the horizontal POSITION knob to move the trigger point horizontally 2008 RIGOL Technologies Inc 1 17 User s Guide for DS1000E DS1000D Series RIGOL To Understand the Trigger System Figure 1 20 shows the trigger control MENU 50 FORCE and a LEVEL trigger level knob Following the exercise to familiarize with the buttons trigger level knob and status bar TRIGGER fa LEVEL VF Oi MENU 50 FORE E Figure 1 20 The trigger control window 1 Turn the trigger Level knob and notice the changes on the display As you turn the knob or pressing the button two things will happen on the display for a short time First the trigger level value is displayed at the bottom left of the screen Second a line is displayed showing the location of the trigger level as long as AC coupling or low frequency reject are not selected Trigger Level to 0 Shortcut key Turn the knob to change trigger level value and press the LEVEL knob to set trigger level back to 0 as a shortcut key 1 18 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL 2 Change the trigger setup and notice these changes in the status bar Press MENU button in the Trigger control A soft button menu appears on the display showing the trigger setting choices as shown in Figure 1 21
17. User s Guide for DS1000E DS1000D Series RIGOL Figure 2 80 Table 2 48 The Storage menu For factory 2 Storage Recall factory setups toad Recall factory setups or files I Go to disk manage menu see Disk Mana 2008 RIGOL Technologies Inc 2 61 User s Guide for DS1000E DS1000D Series RIGOL Internal Memory Press Storage Internal to go to the following menu Figure 2 81 Table 2 49 The Internal Memory Menu eS Menu Comments Location Settings easy Int_00 S hInt_al ind Set up the location of files in internal memory Int_09 N Recall waveform files and setup files from the internal memory location Delete File Save waveform files and setup files to the internal memory location Delete File Delete the selected location file Folder folder 2 62 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSI000D Series RIGOL External Memory Press Storage External to go to the following menu Figure 2 82 Table 2 50 The Memory menu External Explorer File Switch among Path Directory and Explorer Directory File Mew File File New File Delete Fil Folder To create new file or folder Delete File ra Delete the selected location file Folder folder Recall waveform and setup from USB storage device External Path Explorer Back to previous folder 5r E NewFile_G T B HewFile_i Current directory4 RIGO E HewFile_ ig Mow Fil
18. as REF Import import menu see table 2 14 Load Import Load Press Utility gt Pass Fail Mask Setting Load to go to the following menu Figure 2 99 Table 2 63 The Load menu Load Explorer Boorer Ipirectony Switch among Path Directory and File File load AAA Recall the specified file NOTE Pass Fail function is unavailable in X Y mode 2008 RIGOL Technologies Inc 2 73 User s Guide for DS1000E DS1000D Series RIGOL Pass Fail connection In Pass Fail testing output circuit the Oscilloscope adopts the technique of Optical Isolation It needs to connect to another circuit to fulfill the function Before connecting to an external circuit make sure the maximum voltage current does not exceed 400V 100mA The output device has no polarity limit and can be connect arbitrarily PhotoMOS Relay 1 lt 100mA 100mA z ay to internal circuit V lt 400V BNC socket 9 Optically Isolated Figure 2 100 Schematic Diagram of Pass Fail output 2 74 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Record Waveform recorder records input waveforms from CH1 and CH2 with a maximum record length of 1000 frames This performance can also be activated by the Pass Fail test output which makes this function especially useful to capture abnormal signals in long term without keeping an eye watching it This function contains four modes OFF Record Pl
19. between the coarse settings 3 Channels can be adjusted by the vertical POSITION and SCALE only when they are selected 4 During the vertical position a position message is displayed on the left bottom of the screen in the same color as the corresponding channel The unit is V Volts 2 26 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL To Set up the Horizontal System The oscilloscope shows the time per division in the scale readout Since all active waveforms use the same time base the oscilloscope only displays one value for all the active channels except when using Delayed Scan or Alternate Trigger The horizontal controls can change the horizontal scale and position of waveforms The horizontal center of the screen is the time reference for waveforms Changing the horizontal scale causes the waveform to expand or contract about the screen center Horizontal position changes the displayed waveform position relative to the trigger point Press the horizontal MENU button to display the horizontal menu as follows Figure 2 37 Table 2 19 The Horizontal menu Time Delayed OFF Turn on Delayed Scan mode Turn off the Delayed Scan mode Show the relative relation between Time Base vertical voltage and horizontal time Sa Rate Time Base Show CH value at X axis CH2 value at Y axis Trig Offset In Roll Mode the waveform display updates from right to left Sa Rate i Show system
20. channels only Turn off Turn on Running status digital channels diaital channels LA Dy D0 O715 D8 Channel 1 Current gt 9 iL CMOS Threshold Digital channels Channel coupling Horizontal Trigger and vertical div time base div offset Figure 1 6 User Interface Analog and Digital channels 1 6 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Appearance and Dimensions S03 0mm S 159 0 mm 146 0 mm 00000 Go leew Cpa le 7 N Figure 1 7 Front Elevation 131 0 mm 122 1 mm h H l ata B T 158 1 mm 146 1 mm j K COD _ ______D C_D eee alll i F QJ Figure 1 8 Side Elevation 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL To Perform a Functional Check Perform this quick functional check to verify that the instrument is operating correctly 1 Turn on the instrument Use the power cord designed for the oscilloscope only Use a power source that delivers 100 to 240 VACpms 45Hz to 440Hz Turn on the instruments and wait until the display shows the waveform window Push the button select Storage in the top menu box and push the Factory menu box Power button 4 Storage button f 3 3 Go 4 8
21. components and attenuates the signal below 10Hz DC DC coupling passes both AC and DC components LF Reject LF Reject coupling blocks DC component and attenuates all Signal with a frequency lower than 8 kHz HF Reject HF Reject coupling attenuates all signals with a frequency higher than 150 kHz 4 Pre trigger delayed trigger The data collected before and after trigger The trigger position is typically set at the horizontal center of the screen In the full screen display the 6div data of pre trigger and delayed trigger can be surveyed More data of pre trigger and 1s delayed trigger can be surveyed by adjusting the horizontal knob This feature is very useful to study the events that led up to the trigger point Everything to the right of the trigger point is called post trigger information The delay range pre trigger and post trigger information depends on the sweep speed selected 5 Adjustable trigger sensitivity To avoid the influence of noise from the physical world and get the stable trigger the trigger circuit has adopted Stickiness In DS1000E DS1000D series the stickiness is adjustable from 0 1div 1 0div which means when it sets to 1 0div the trigger circuit will not affect any signal with peak peak amplitude less than 1 0div so as to avoid the influence of the noise 2 52 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSI000D Series RIGOL To Set up the Sampling System shows the menu
22. mode then choose OK to start keyboard locked function as the following figure In this mode except from F1 to F5 and MENU ON OFF other keys are locked If you choose cancel it will cancel this function Keyboard locked Time 1 0 uz Figure 2 118 Interface of keyboard locked 2 Unlock the keyboard Press any button among F1 to F5 to enter the Unlock mode then please input the unlock keys the default is 111111 users can reset new keys with 6 bit length to unlock the keyboard 2 86 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Mode Please enter unlock kewe Time 1 0 us Figure 2 119 Interface of keyboard unlock 3 Reset the keys Press Utility button to choose the Preference menu then choose set keys option in second page of the menu just like the following figure Input the old keys and press Enter input your new code and ensure it again the system will clew that you have finished new keys setting r the old keys Enter Time 1 666us Figure 2 120 Interface of key resetting 2008 RIGOL Technologies Inc 2 87 User s Guide for DS1000E DS1000D Series RIGOL To Measure Automatically The button in the menu area activates the automatic measurement function The instruction below shows the powerful measurement function of the oscilloscopes Auto measurement setup button MENU Figure 2 121 Auto measurement setup button Me
23. pass both AC and DC components of the input signal The waveform is displayed as Figure 2 4 Coupling DC coupling AYN Limit setup Probe DC coupling status symbol Figure 2 4 DC coupling setting Press ICH1 Coupling GND to set GND coupling it disconnects the input signal The screen displays as Figure 2 5 a Coupling GND coupling Bi Limit setup GFF Probe Digital Filter GND coupling status symbol Figure 2 5 GND coupling setting 2 4 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSi000D Series RIGOL 2 Set up the channel bandwidth limit To use Channel 1 as an example input a signal containing high frequency component Press ICH1 BW Limit OFF to set up bandwidth limit to OFF status The oscilloscope is set to full bandwidth and passing the high frequency component in the signal The waveform is displayed as Figure 2 6 Coupling Turn off the BW limit Time 106 8rns Figure 2 6 Turn off the BW limit Press CH1 gt BW Limit ON to set up bandwidth limit to ON status It will reject the frequency component higher than 20MHz Coupling Ay Limit 20M BW limit Probe Digital Filter Mark of Time 188 Gns BW limit Figure 2 7 Turn on the BW limit 2 5 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL 3 Probe Attenuation Setting The oscilloscope allows selecting the attenuation factor for the probe The at
24. periodically 2008 RIGOL Technologies Inc 3 1 User s Guide for DS1000E DS1000D Series RIGOL Example 2 View a Signal Delay Caused by a Circuit This example is to test the input and output signals of a circuit and observe the Signal delay First set the probe and the channel attenuation to 10X and connect CH1 probe to the input CH2 to the output of the circuit Do these steps as follow 1 Display the signals CH1 and CH2 1 Press the button 2 Adjust the vertical and the horizontal scale by turning the knobs to appropriate ranges for display 3 Press the button to select Channel 1 and turn the vertical knob to adjust the vertical position of Channel 1 waveform 4 Press the button to select Channel 2 and turn the vertical knob to adjust the vertical position of Channel 2 waveform 2 Measure the delay time when a signal going through the circuit Auto measuring the delay 1 Press Measure gt Source CH1 to set measurement source 2 Press Time to select the measurement Type 3 Press Delay 1 gt 2 to display the result on the screen You can see the change of the waveform in the following figure Coupling Time 16 JAus Figure 3 1 Waveform distortion 3 2 2008 RIGOL Technologies Inc User s Guide for DSI000E DS1000D Series RIGOL Example 3 Capture a Single Shot Signal To capture a single event it needs to gather some pre test knowledge of the signal in order to set
25. sample rate Trig offset Adjust to the center Reset The parameter status symbol will be displayed on the screen during horizontal system settings which can help users to judge the channel status quickly 2008 RIGOL Technologies Inc 2 27 User s Guide for DS1000E DS1000D Series RIGOL A ee ee e walts Cir Invert Time 1A Aus 4 Figure 2 38 Status bar and mark for Horizontal control Marks Indicator 1 The position of current waveform window in the memory 2 The trigger position in the memory G The trigger position in the current waveform windows 4 The horizontal time base main time base The trigger s horizontal offset according to the center of the window 2 28 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Key Points Y T The conventional oscilloscope display format It shows the voltage of a waveform record on the vertical axis as it varies over time on the horizontal axis X Y XY format displays channel 1 in the horizontal axis and channel 2 in the vertical axis Roll Mode In this mode the waveform display rolls from right to left No trigger or horizontal offset control of waveforms is available during Roll Mode and it s only available when set to 500 ms div or slower Slow Scan Mode This mode is available when the horizontal time base is set to 50ms div or slower In this mode the oscilloscope acquires sufficient data for the
26. signal of rising EXT 6 of setting 200 mV and falling time 220ns Normal mode pre trigger memory depth 2 sample Trigger Offset rate delayed trigger 1s Slow Scan mode pre trigger 6div delayed trigger 6div Trigger Holdoff Range 500ns 1 5s Set Level to 50 Typical Edge Trigger Edge Trigger Rising Falling Rising Falling Slope Trigger Condition gt lt positive pulse width gt lt negative pulse width Pulse Width 20ns 10s Range 5 4 2008 RIGOL Technologies Inc ser s Guide for A erles When input signal frequency gt 50Hz RIGOL Video Trigger Video Standard amp Support for standard NTSC PAL and SECAM broadcast Line Frequency systems Line number range 1 525 NTSC and 1 625 PAL SECAM Slope Trigger gt lt positive slope gt lt negative slope 20ns 10s Alternate Trigger Pattern Trigger Trigger Mode Duration Trigger Trigger Type DO D15 select H L X Qualifier a Time Setup 20ns 10s Measurements Manual Voltage difference between cursors AV Cursor Time difference between cursors AT Reciprocal of AT in Hertz 1 AT Track Voltage value for Y axis waveform Time value for X axis waveform Auto Measure Vpp Vamp Vmax Vmin Vtop Vbase Vavg Vrms Overshoot Preshoot Freg Period Rise Time Fall Time Width Width Duty Duty Delay1 2 Delayl1 2 t Phas 1 2 Phas 1 2 Remarks 1 When sampling is 1GSa s on
27. the channel bandwidth to Digital Filter BW Limit 20MHz to reduce display noise Get full bandwidth i Set this to match your probe attenuation factor to make the vertical scale readout correct Digital filter Soe Setup Setup digital filter See table 2 4 filter See table 2 4 Go to the next menu page The i followings are the same no more explanation 2 2 2008 RIGOL Technologies Inc sers Guide for erles RIGOL Figure 2 2 Table 2 2 The Channel Menu Page 2 2 Menu Settings _ r 272 2 2 Back to the previous menu page ValtsiDiv Tr The followings are the same no more explanation Invert Selects the resolution of the OFF Coarse SCALE knob Volts Div Defines a 1 2 5 sequence Fine To change the resolution to small steps between the coarse settings ON Turn on the invert function Invert OFF Restore original display of the waveform 1 Channel coupling To use Channel 1 as an example input a sine wave signal with DC shift Press CH1 gt Coupling gt AC to set AC coupling It will pass AC component blocks the DC component of the input signal The waveform is displayed as Figure 2 3 CH1 Coupling AC coupling Caen ole CFF Probe Digital Filter AC coupling Time 168 Gus status symbol Figure 2 3 AC coupling setting 2008 RIGOL Technologies Inc 2 3 User s Guide for DS1000E DS1000D Series RIGOL Press CH1 Coupling DC to set DC coupling It will
28. the two cursors When the cursor is in X type a pair of vertical cursors CurA and CurB will be appeared in the screen whose time value could be measured separately as well as the time difference between them and which will be changed with the location of cursor if you twirl the knob Cursors ESE Gms hoe axe 2 24ms l an 446 4H2 Type Source Time 56H 8us Figure 2 135 Manual Measure Mode of Cursor Measurement 2 96 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL When the cursor is in Y type a pair of horizontal cursors CurA and CurB will be appeared in the screen whose voltage value could be measured separately as well as the time difference between them and which will be changed with the location of cursor if you twirl the knob To do manual Cursor Measurements please do as the following steps 1 Select the Manual mode for cursor measurement by pressing soft button as Cursor gt Mode Manual 2 Select channel Source for measurements by pressing soft button as Cursor gt Source CH1 CH2 MATH FFT or LA DS1000D series NOTE While measuring the channel of MATH the results are valued with d division as units 3 Select the cursors type by pressing soft button as Cursor gt Type X or Y 4 Move the cursors to adjust the increment between the cursors Details in the following Table Table 2 86 The Cursor menu Increment Operation T
29. trigger event the range is lt Holdoff Setting gt per Serene from 500ns to 1 5s Holdoff Reset Reset Holdoff time to 500ns 20ans Haldott L Sensitivity Figure 2 64 Table 2 40 The Trigger Set Up Menu Settings for sensitivity and Seu holdoff Eee Sensitivity v Set trigger sensitivity Haldotf lt Sensitivity Setting gt S8Gns Holdoff Set time slot before Holdott lt Holdoff Setting gt another trigger event Holdoff Reset Reset Holdoff time to 500ns t 2 48 2008 RIGOL Technologies Inc ser S Guide for J erles RIGOL Figure 2 65 Table 2 41 The Trigger Set Up Menu Settings only for holdoff Settings Comments Haldatt Menu 4 4 Set time slot before another Soars Holdoff aaa lt Holdoff Setting gt trigger event Holdoff Reset Holdoff time to 500ns 2008 RIGOL Technologies Inc 2 49 User s Guide for DS1000E DS1000D Series RIGOL Trigger Holdoff Trigger Holdoff can stabilize complex waveform such as the pulse range Holdoff time is the oscilloscope s waiting period before starting a new trigger During Holdoff oscilloscope will not trigger until Holdoff ends For instance To trigger on the first pulse on a group of them users can set the holdoff time to Pulse cluster width Set Up Coupling Trigger site soe fm gem poe ge ey A 38diy Haldott Holdott Time i164 Anse Time of holdoff Figure 2 66 Trigger Holdo
30. up the trigger level and slope correctly For example if the event is derived from TTL logic a trigger level of 2 volts should work on a rising edge The following steps show how to use the oscilloscope to capture a single event 1 Set the probe and the channel attenuations to 10X 2 Set up the trigger 1 Press the button in the Trigger control area to display the menu 2 Press Edge to select the trigger mode 3 Press Slope to select 4 Press Source to select CH1 5 Press Sweep to select Single 6 Press Set Up Coupling to select DC 3 Turn the vertical and horizontal knobs to adjust the Volts Div and the Time base in a proper range for the signal 4 Turn the knob to adjust trigger level 5 Press RUN STOP button to start capturing When the trigger conditions are met data appears on the display representing the data points that the oscilloscope obtained with one acquisition This function helps to capture the occurrence easily such as the noisy with large amplitude set the trigger level higher a little above the normal level and press and wait When noise occurs the instrument will record the waveform before and after the trigger Adjust the button in the horizontal control area and change the level of the trigger position will get the inverted delay trigger It is useful to observe the waveform before the occurrence of the noise 2008 RIGOL Technologies Inc 3 3 User s Guide for DS1000E DS1000D Series
31. 0D Series RIGOL To do Track mode Cursor Measurements follow these steps 1 Select the Track mode for cursor measurement by pressing soft button as Cursor Mode Track 2 Select channel Source for Cursor A and Cursor B by pressing soft button as Cursor Cursor A or Cursor B CH1 CH2 or None 3 Move the cursors to adjust the horizontal positions of the cursors Details in the following Table Table 2 88 The Cursor usage Cursor Cperation O Turn the multifunctional knob 4 to move cursor A horizontally Turn the multifunctional knob 4 to move Cursor B horizontally NOTE Moving cursor horizontally is not allowed when other not tracking cursor menu is activated 4 To get measurement value Position of Cursor 1 Time cursor centered on the midpoint of screen Voltage cursor centered on channel ground level Position of Cursor 2 Time cursor centered on the midpoint of screen Voltage cursor centered on channel ground level Read the horizontal space between Cursor 1 and 2 aX Time between cursors units in seconds 1 aX units in Hz KHz MHz GHz Vertical space between cursor 1 and 2 aY Voltage between cursors units in V 2 100 2008 RIGOL Technologies Inc User s Guide for DSI000E DS1000D Series RIGOL Auto Mode Press Cursor Mode Auto to display the auto mode menu Figure 2 138 Table 2 89 The Auto Mode Menu CUFSorsS Mode Display the cursors for the curren
32. 180 degrees as respect to the ground level When the oscilloscope is triggered on the inverted signal the trigger is also inverted Figure 2 13 and Figure 2 14 show the changes after inversion Volts Div Invert C OFF Y Invert OFF volts Div Coarse Invert C ON Invert ON Figure 2 14 The waveform after inversion 2008 RIGOL Technologies Inc 2 9 User s Guide for DS1000E DS1000D Series RIGOL Math Functions The mathematic functions include Add Subtract Multiply and FFT for Channel 1 and Channel 2 The mathematic result can be measured by grid and cursor Press MATH button the system go into the MATH interface as follows hath Operate Source H1 Source B CH Invert OFF MATH scale Time SHE Aus Figure 2 15 The Math function Figure 2 16 Table 2 5 The Math menu SE Add source A and source B Operation Subtract source B from source A Multiply source B by source A Fast Fourier Transform Source 4 Source H he Define CH1 or CH2 as source A CH1 Source B CH2 Define CH1 or CH2 as source B Invert the MATH waveform OFF Restore to original waveform display Invert 2 10 2008 RIGOL Technologies Inc sers Guide for erles RIGOL 1 Using the FFT The FFT Fast Fourier Transform process converts a time domain signal into its frequency components mathematically The horizontal axis shows the frequency and the vertical axis shows the amplitude dBV
33. 30 The Alternate menu Trigger Type Pulse Page 2 2 a rae Setting ee Set the width of the pulse Setting lt pulse width gt EESE Set Up To go to set up menu See table 2 38 Set Wn Figure 2 55 Table 2 31 The Alternate menu Trigger Type Slope Page 1 2 Mode Menu Settings Set trigger mode for Channel 1 Select Select Set trigger mode for Channel 2 eal Set Slope Trigger for the vertical Type mo ae i channel wher 2 When Set trigger condition H Figure 2 56 Table 2 32 The Alternate menu Trigger Type Slope Page 2 2 Trigger a Pada a Time Time Set slope time lt Time ee gt artical Vertical Select the level to be adjusted by LEVEL vee Set Up a i to set up menu See table Set Up 2008 RIGOL Technologies Inc 2 43 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 57 Table 2 33 The Alternate menu Trigger Type Video Page 1 2 des d CH1 Set trigger mode for Channel 1 CH2 Set trigger mode for Channel 2 select Video Trigger for the channel Type a i Polarity ji positive Triggers on positive going sync pulses U negative Triggers on negative going sync pulses Polarity ALL lines Trigger on all lines aaa Sak Line Num Trigger on an specified line i Odd field Select to trigger on odd field or even Even field field 4 l 4 No 1 Line Num Select the specified line number for sync Standard lt Lines Set gt Line Murn NTSC PAL SECM
34. Figure 1 21 Trigger Press the trigger Mode button and choose Edge bode Press the trigger Source button to select CH1 Source Press the trigger Slope button to choose Press the trigger Sweep button to select Auto Press the trigger Set Up button to enter secondary menu NOTE The trigger type slope and source change in conjunction with the status bar on the top right of the screen 3 Press The button sets the trigger level to the center of the signal 4 Press Starting an acquisition regardless of an adequate trigger signal usually used in Normal or Single trigger mode This button has no effect if the acquisition is already stopped Key point Holdoff A time interval before the oscilloscope response to next trigger signal During this holdoff period the trigger system becomes blind to trigger Signals This function helps to view complex signals such as an AM waveform Press Holdoff button to activate 4 knob then turn it to adjust Holdoff time 2008 RIGOL Technologies Inc 1 19 User s Guide for DS1000E DS1000D Series RIGOL Chapter 2 Operating Your Oscilloscope By now a user should understand the VERTICAL HORIZONTAL and TRIGGER control systems and knows how to determine the system setup by status bar of a DS1000E DS1000D series digital oscilloscope This chapter will go through all groups of front panel buttons knobs and menus and further the knowledge of the operation b
35. GL C 1 A eh PERI hal Hcg i MANTAN GR TO oe Ge Tea M E Cy sooo po ae S232 port Figure 1 4 Rear Panel Instruction Rear Panel of DS1000E DS10000D series contains the following parts Pass Fail output port The Pass Fail testing results can be output by the port RS232 port To connect the oscilloscopes and external equipment as serial port USB Device port It is used to transfer data when external USB equipment connects to the oscilloscope regarded as Device equipment For example to use this port when connect PictBridge printer to the instrument by USB Notation definitions in this User s Guide Throughout this manual notation symbols of buttons and knobs are the same to those on front panel A box around the name of the button denotes MENU function buttons on front panel such as Measure 4 denotes the multi function knob denotes the two POSITION knobs denotes the two SCALE knobs denotes the LEVEL knob The name with a drop shadow denotes the menu operating button such as WAVEFORM soft button in STORAGE menu 2008 RIGOL Technologies Inc 1 5 User s Guide for DS1000E DS1000D Series RIGOL 3 User Interface Location of waveform window Trigger point Trigger point in Running status in memory in memory waveform window Coupling Channel 1 e AE AET AA Probe Menu nanne Waveform display window Time 288 6u Figure 1 5 Display screen Analog
36. Menu 1s Set the time before menu fades 2S away The menu will be hidden Display 5s after the set time of last button 10s pressing 20s Infinite 1 Display type Display type includes Vector and Dot In vectors type the oscilloscopes connect dots through digital interpolation including both linearity and sin x x Sin x x interpolation is suitable for Real time sampling and will be more effective at 50ns or faster time base Key points 2 Refresh rate It is an important performance of digital oscilloscopes It means the number of display refreshing per second and it will affect the ability to observe signal Adjusting waveform intensity Default setup of multi function knob 44 is adjusting waveform intensity 2 58 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL To Store and Recall shows the menu button for the storage system on the front panel Storage setup button MENU Figure 2 76 Storage setup button Press the button to show the menu for the settings of the storage system Waveforms and setups can be stored in and recalled from both internal memory and external memory The waveform file setup file bitmap and CSV file in external memory can be created and deleted Note internal file can be deleted or overwritten System supports English Chinese key in Waveform and setup the menu as following Figure 2 77 Table 2 45 The Storage menu For Waveform and Setups S
37. NG sanamini 2 51 O SQUUD rana 2 69 ACCESSOVIES oiran 6 1 6 2 INPUTS aiaa 5 2 ACQUISIEION siisii 5 2 Instant Execute Key s is 2 102 Alternative Trigger cscsceceeees 2 42 Invert a waveform ssesssesssseresen 2 9 ADDCNCIX aridsirminiiiria 6 1 6 4 Language resaka 2 70 AUTO inean 2 102 2 103 EV Eb stohavdinituiimerettdeanastcnten 2 34 Auto Trigger crs 2 51 LFREJOC unissa 2 52 Automatic Measurement 2 88 Matsui tide eda ia 2 10 Average Acquisition 2 56 Me aSUreMeNtS cccceeeceeeeeeeeeeees 5 5 Bandwidth Limit ccccceeeeeees 2 5 Normal Trigger scscscseeseeeeeees 2 51 Blackman WINdOW 0sceceeeeees 2 12 Nyquist Frequency sceceeeees 2 13 Channel Coupling ccccseeeeeees 2 3 CSMIOMAl kdacntaansthesstinudsdemwmastieetidies 6 1 deann carinici 6 3 Pattern Trigger cscscscssererencnes 2 46 CoarsSe FING ccccesesessenevans 1 14 2 9 POW aree aara 5 6 CUNSOM orrasa di 2 95 PROUIGGE arrisera 2 52 Cursor Measurements 2 95 3 6 Probe Attenuation s 2 6 DC COUDIING srrirairean aa 2 4 Probe COMpeNnsatiONn csceceeees 1 9 DISDIAV mrenani 5 6 Probe Compensator sceceeees 5 6 Display System cccceceeeeeeeeeees 2 57 Pulse Trigger imiwinwcsicretidwintvees 2 37 Duration Trigger cccceeeeeeeees 2 47 Random NOISC sceceeeeeeeeeeeees 3 4 Edge TriQQer scsscsesreessesearennes 2 36 Realtime Sampling scsceeee 2 56 ENVEI DE einni 2 56 Rectang
38. OS Threshold ECL Setup threshold 96 Anu Setup data of Time S88 8ns threshold by user Figure 2 33 Set the threshold Threshold explanation LOGIC STANDARD TTL CMOS ECL USER THRESHOULD VLOTAGE 1 4V 2 5V 1 3V 8V to 8V 2 22 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Set up Channel Group Press LA I gt D7 DO or D15 D8 Turn on off the channel single or in a group Also you can change the size of waveforms in 8 bits as a group See table 2 16 and 2 17 Figure 2 34 Table 2 16 The Digital Channel menu Page 1 D7 D0 Turn on or off single channel of D7 DO OF D0 Turn on D7 D0 Turn on or off all 8 channels together Size Give JL Display 8 channels in a single screen m Display 16 channels in a single screen Rest An Reset waveform of channel D7 D0 channel D15 D8 Turn on or off single channel of D15 D8 D15 D4 Turn on D15 D8 Turn on or off 8 channels together Size JL Display 8 channels in a single screen T Display 16 channels in a single screen a Reset waveform of channel D15 D8 L 2008 RIGOL Technologies Inc 2 23 User s Guide for DS1000E DS1000D Series RIGOL 1 Turn on or off a digital channel Press LA gt D7 D0 Channel and choose the wanted channel by turning 4 knob Press No 1 soft button or push down 44 knob to turn on off the channel When the channel is on we can see the mark i When the channel is turned off the mark will display as B
39. RIGOL Example 4 To Reduce the Random Noise on a Signal If the signal is noisy Figure 3 2 set up the oscilloscope to reduce the noise on the waveform and avoid its interference to the signal Acquire Acquisition Sampling i Mem Depth Time 1608 Aus Figure 3 2 Signal with noise 1 Set the probe and the channel attenuations to 10X 2 Connect a signal to the oscilloscope and obtain a stable display 3 Improve the trigger by setting the Coupling 1 Press the MENU in the Trigger control area 2 Press Set Up Coupling LF Reject or HF Reject HF Reject High frequency reject adds a low pass filter with the 3 dB cut off point at 150 kHz Use HF rejects to remove high frequency noise such as AM or FM broadcast stations from the trigger path LF Reject Low frequency reject adds a high pass filter with the 3 dB cut off point at 8 kHz Use LF Reject to remove low frequency signals such as power line noise from the trigger path 4 To reduce the noise by setting the acquisition type and adjust the waveform intensity 3 4 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL 1 If there is noise within the signal and the waveform looks too wide in this case choose average acquisition In this mode the waveform will be thin and easy to observe and measure To use average follow these steps Press soft button as Acquire Acquisition Average Toggle the Averages soft but
40. Trigger Field Synchronization 2008 RIGOL Technologies Inc 2 41 User s Guide for DS1000E DS1000D Series RIGOL Alternate Trigger Alternate Trigger Mode When alternate trigger is on the trigger sources come from two vertical channels This mode can be used to observe two non related Signals You can choose two different trigger modes for the two vertical channels The options are as follows Edge Pulse Slope and video The information of the trigger level of the two channels will be displayed on the upper right of the screen Press Trigger IMENU Mode Alternate to display the alternate trigger menu as follows Figure 2 52 Table 2 28 The Alternate menu Trigger Type Edge Mode Set trigger mode for Channel 1 Select Select Set trigger mode for Channel 2 edge S Edge Trigger as the trigger CH1 Type g rigger rigg Type Slope o Rising Trigger on rising edge i Falling Trigger on falling edge To go to set up menu See table Set Up 3 38 Trigger Mode CH1 Set trigger mode for Select Channel 1 select Set trigger mode for Sul CH2 Channel 2 DEE Set Pulse Trigger for the Type Pulse TE channel iii Hle Pulse width less than F gt 4L Pulse width more than L Pulse width equal to ale Pulse width less than 2a Pulse width more than shee Pulse width equal to To select pulse condition 2 42 2008 RIGOL Technologies Inc sers Guide for 3 eries RIGOL Figure 2 54 Table 2
41. User s Guide RIGOL Publication number UGA07114 1110 Sept 2010 DS1000E DS1000D Series Digital Oscilloscopes DS1102E DS1052E DS1102D DS1052D 2008 RIGOL Technologies Inc All Rights Reserved RIGOL 2008 RIGOL Technologies Inc All Rights Reserved RIGOL products are protected by patent law in and outside of P R China Information in this publication replaces all previously corresponding material RIGOL Technologies Inc reserves the right to modify or change part of or all the specifications and pricing policies at company s sole decision oe ee IS NOTE RIGOL is registered trademark of RIGOL Technologies Inc 2008 RIGOL Technologies Inc I User s Guide for DS1000E DS1000D Series RIGOL Safety Notices Review the following safety precautions carefully before operating the instrument to avoid any personal injuries or damages to the instrument and any products connected to it To avoid potential hazards use the instrument as specified by this user s guide only The instrument should be serviced by qualified personnel only Avoid Fire or Personal Injury Use Proper Power Cord Use the power cord designed for the instrument and authorized in your country only Connect and Disconnect Accessories Do not connect or disconnect probes or test leads while they are connected to a voltage source Ground The Instrument The oscilloscope is grounded through the grounding conductor of the power cord To av
42. V div 2V 2mV div 245mV div Analog Bandwidth 100MHz DS1102D DS1102E 50MHz DS1052D DS1052E Single shot 100MHz DS1102D DS1102E Bandwidth 50MHz DS1052D DS1052E Selectable Analog Bandwidth Limit 20MHz typical Lower Frequency Response AC lt 5Hz at input BNC 3dB Rise Time at BNC lt 3 5ns lt ns typical respectively at 100MHz 50MHz Dynamic range 5div DC Gain Accuracy 2mV div 5mV div 2008 RIGOL Technologies Inc 5 3 User s Guide for DS1000E DS1000D Series RIGOL 4 Normal or Average acquisition mode 10mV div 10V div 3 Normal or Average acquisition mode DC Measurement Average of 216 Waveforms with vertical position at zero Accuracy Average DC Gain Accuracy xreading 0 1div 1mvV Acquisition Mode Average of 216 Waveforms with vertical position not at zero DC Gain Accuracyx reading vertical position 1 of vertical position 0 2div Add 2mV for settings from 2mV div to 245 mV div Add 50mV for settings from 250mV div to 10V div Delta Volts Delta Volts between any two averages of 16 waveforms Measurement acquired under same setup and ambient Accuracy Average conditions DC Gain Accuracyxreading 0 05 div Acquisition Mode Trigger Trigger Sensitivity 0 1div 1 0div adjustable Trigger Level Range 6 divisions from center of screen EXT 1 2V Trigger Level Accuracy 0 3div x V div 4 divisions from center of typical applicable for screen the
43. a on the front panel includes a knob and three buttons LEVEL The knob that set the trigger level press the knob and the level will reset to zero The instant execute button setting the trigger level to the vertical midpoint between the peaks of the trigger signal UI g c FORCE Force to create a trigger signal and the function is mainly used in Normal and Single mode MENU The button that activates the trigger controls menu Press the Trigger MENU key to enter the trigger system setting interface Trigger Mode Source Time SH Abus Figure 2 41 Trigger controls interface 2 34 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Trigger Modes The oscilloscope provides seven trigger modes Edge Pulse Width Slope Video Alternate Pattern only for DS1000D series and Duration trigger only for DS1000D series Edge An edge trigger occurs when the trigger input passes through a specified voltage level at the specified slope direction Pulse Width Use this trigger type to catch pulses with certain pulse width Slope The oscilloscope begins to trigger according to the signal rising or falling speed Video Use video trigger on fields or lines for standard video signals Alternate Trigger on non synchronized signals of the dual channels Pattern To Trigger through detecting a specified code Duration To trigger within a specified time on the conditions of a specified cod
44. a voltage source 1 10 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Digital Leads Only for DS1000D Series Digital leads are provided only for DS1000D series which have Logic Analyzer 1 Switch off power supply of the device under test if necessary to avoid short circuit Since no voltage is applied to the leads at this step you may keep the oscilloscope on 2 Connect one end of the flat cable FC1868 to the Logic Analyzer Input connect the other end to Logic Head LH1116 An identifier is located on each end of the flat cable it can only be connected in one way It is unnecessary to switch off power supply of your oscilloscope when connecting the cable C TOE i fey Figure 1 14 Connect the digital leads CAUTION Use only FC1868 LH1116 TC1100 and LC1150 made by RIGOL for specified DS1000D series 3 Connect a test clip to one lead wire make sure it s connection good f gt Ca se SA Yo Ce eS o g SS Bee RS os S D G PS Oks Ss x gt x Se De X D Test clip Figure 1 15 Input Interface of Digital Channel 2008 RIGOL Technologies Inc 1 11 User s Guide for DS1000E DS1000D Series RIGOL 4 Test your device with the clip Figure 1 16 Digital Channel Measurement 5 Remember to connect Ground Channel to the DUT s ground terminal Figure 1 17
45. ais E ari amuage FYCCEHH i Portugues Espa ol Figure 2 93 The Language menu Please use the fourth menu button or use the knob to select the language you need 2 70 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Pass Fail The Pass Fail function monitors changes of signals and output pass or fail signals by comparing the input signal that is within the pre defined mask The testing results not only can be displayed on screen or output by isolated pass fail port but also can be alarmed if turn on system sound Press Utility Pass Fail to go to the following menu Figure 2 94 Table 2 58 The Pass Fail Menu Page 1 2 1a ae EPET ON Turn on Pass Fail test OFF Turn off Pass Fail test m Sane CH1 Select Pass Fail test on CH1 CH2 Select Pass Fail test on CH2 Operate a P RUN Pass Fail test stopped press to run Operate Mise Display WE STOP Pass Fail test running press to stop OFF ON Turn on Pass Fail information 3 display a Msg displa S ia a OFF Turn off Pass Fail information display Figure 2 95 Table 2 59 The Pass Fail Menu Page 2 2 PassiF ail Output n i Fail 41 detected LIUT AU Output and beep when Fail Fail arena Pass condition detected oton UU F te psoas Pass th Output when Pass _ condition OFF detected when Fail condition Mask Setting Output and beep when Pass condition detected T Stop ON Stop test when output occ
46. al The values are displayed on the boxes below the menu Before using cursors make sure to set the Signal Source as the channel for measuring 2 Track Mode In this mode the screen displays two cross cursors The cross cursor sets the position on the waveform automatically Adjust cursor s horizontal position on the waveform by rotating the multifunctional knob 4 The oscilloscope displays the values of the coordinates on the boxes below the menu 3 Auto Mode This mode will take effect with Automatic Measurements The instruments will display cursors while measuring parameters automatically These cursors demonstrate the electrical meanings of these measurements NOTE The Auto Measure mode for Cursor measuring will take no effect without automatic measurements 2008 RIGOL Technologies Inc 2 95 User s Guide for DS1000E DS1000D Series RIGOL Manual Mode Press Cursor gt Mode Manual to display the manual mode menu Figure 2 134 Table 2 85 The Manual Mode Menu Curenrs blode l Adjust the cursor to measure X Y Mode Manual parameters manually Type Shown as vertical line to measure the X z horizontal parameters Source Type y Shown as horizontal line to measure the vertical parameters EESE CH1 Sanne CH2 Select the measurement signal source MATH LA is only for DS1000D series LA In this mode the oscilloscope measures the Y or X coordinate values of the cursors and the increments between
47. are wave will display Push the button again to turn off Channel 1 Push the button to turn on channel 2 repeat steps 2 and 3 NOTE The signal output from Probe compensator should only be used for probe compensation not for calibration 2008 RIGOL Technologies Inc 1 9 User s Guide for DS1000E DS1000D Series RIGOL To Compensate Probes Perform this adjustment to match the characteristics of the probe and the channel input This should be performed whenever attaching a probe to any input channel for the first time 1 From CH1 menu set the Probe attenuation to 10X press ICH1 Probe 10X Set the switch to 10X on the probe and connect it to CH1 of the oscilloscope When using the probe hook tip inserting the tip onto the probe firmly to ensure a proper connection Attach the probe tip to the Probe compensator connector and the reference lead to the ground pin Select CH1 and then press AUTO 2 Check the shape of the displayed waveform Over compensated Correctly Compensated Under Compensated E C e Si Figure 1 13 Probe Compensation 3 If necessary use a non metallic tool to adjust the trimmer capacitor on the probe for the flattest square wave possible as displayed on the oscilloscope 4 Repeat as necessary WARNING To avoid electric shock while using the probe be sure the perfection of the insulated cable and do not touch the metallic portions of the probe head while it is connected with
48. ay back and Storage More details about the function please refer to the following introduction 1 OFF It will close all record functions 2 Record To record waveform by selected time interval till the last frame Press Utility Record Mode Record to go to the following menu Figure 2 101 Table 2 64 The Record menu Page 1 2 Record Select record mode Play back Select play back mode Storage Select storage mode Turn off all recorder functions OFF End Frame EESE CH1 l Source CH2 Select record source channel Uperate P F OUT mo Mode Source Mode lt 1 1000 gt Record stopped press to Start recording Recording press to stop 2008 RIGOL Technologies Inc 2 75 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 102 Table 2 65 The Record menu Page 2 2 Record 4 eae Interval Set time interval ee lt 1 00ms 1000s gt Interval 4 22 AMS 3 Play back Play back the recorded waveforms Press Utility Record Mode Play back to go to the following menu Figure 2 103 Table 2 66 The Record menu Page 1 2 Record Mode Play stopped press to Start playback Playing press to stop C Set repeat play mode Play Mode Play mode i b E Set single time play mode Operate 4 Interval Interval Set up interval 1 Beams 2 76 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Figure 2 104 Table 2 67 The Record menu Page 2 2 Rec
49. button for the acquire system on the front panel Acquire setup button Figure 2 67 Acquire setup button Press the Acquire button the interface menu as follows Figure 2 68 Table 2 42 The Acquire menu Acquire Acquisition ee Normal Normal Acquisition mode A Acquisition Average Average Acquisition mode ae Peak Detect Peak Detect Acquisition mode Sampling Averages 2 to 256 Step by 2 ames the power of N Set average times from 2 to 256 EET Sampin Real Time Real time sampling mode i Equ Time Equivalent sampling mode aoe en Deas Long Mem Set up memory as 512kpts or 1Mpts i Normal Set up memory as 8kpts or 16kpts j ON Set the insert mode to sinx x Sinx x i OFF Set the insert mode to line NOTE This function is used in Average Acquisition NOTE 2 This function is used in Real Time sampling The waveform display will be different when selecting different acquisition and sampling 2008 RIGOL Technologies Inc 2 53 User s Guide for DS1000E DS1000D Series RIGOL Select Real time acquisition to observe the single shot or pulse signals Select Equ Time to observe high frequency repetitive signals To reduce the displayed random noise select Average Acquisition And this mode would make the screen refresh slower To Avoid signal aliasing select Peak Detect Acquisition The waveform displayed on the screen will change in conjunction with the setting of Acquire men
50. cation Software A Quick Guide Optional Accessories USB Cable BNC Cable RS232 Cable USB GPIB Adapter DS1000E DS1000D Soft Carrying Case All accessories standard and optional are available by contacting your local RIGOL office 2008 RIGOL Technologies Inc 6 1 User s Guide for DS1000E DS1000D Series Appendix B Warranty RIGOL warrants its products mainframe and accessories in materials and workmanship within the warranty period During the period concerned RIGOL guarantees the free replacement or repair of products which are approved defective To get repair service or obtain a copy of the whole warranty statement please contact with your nearest RIGOL sales and service office RIGOL do not provide any other warranty items except the one being provided by this summary and the warranty statement The warranty items include but not being subjected to the hint guarantee items related to tradable characteristic and any particular purpose RIGOL will not take any responsibility in cases regarding to indirect particular and ensuing damage 6 2 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Appendix C Care and Cleaning General Maintenance Do not store or leave the instrument in where the instrument will be exposed to direct sunlight for long periods of time Caution To avoid damages to the instrument or probes do not expose them to liquids which have causticity Clea
51. d result in damage to this product or other property Terms on the Product These terms may appear on the product DANGER indicates an injury hazard may be immediately accessible WARNING indicates a potential injury hazard may be immediately accessible CAUTION indicates that a potential damage to the instrument or other property might occur Symbols on the Product These symbols may appear on the product A A wm t Hazardous Refer to Protective Grounding Test Voltage Instructions Earth Terminal Terminal Grounding of Chassis Terminal IV 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSI000D Series RIGOL General Purpose Oscilloscopes This book covers the following four types of DS1000E DS1000D Series Digital Oscilloscopes DS1102E DS1052E DS1102D DS1052D With Logic Analyzer DS1000E DS1000D series are kinds of economical digital oscilloscope with high performance DS1000E series are designed with dual channels and 1 external trigger channel DS1000D series are designed with dual channels and 1 external trigger channel as well as 16 channels logic analyzer The front panel of DS1000E DS1000D series digital oscilloscopes are clear direct viewing and easy to use which consistent with the traditional instrument use fully To speed up the adjustment and measurement please press button to get suitable waveform display and settings Besides 1GSa s maximum real time sample rate 25GSa s maximum equi
52. e 2008 RIGOL Technologies Inc 2 35 User s Guide for DS1000E DS1000D Series RIGOL Edge Trigger Edge Trigger Mode An edge trigger determines whether the oscilloscope finds the trigger point on the rising or the falling edge of a signal Select Edge trigger Mode to trigger on Rising edge falling edge or rising amp falling edge Press Trigger MENU Mode Edge to display the edge trigger menu as follows Figure 2 42 Table 2 20 The Edge Trigger Menu Trigger Mode Select CH1 as trigger signal Select CH2 as trigger signal Select EXT TRIG as trigger signal AC Line Select power line as trigger signal D15 D0 Select a digital channel in D15 DO as trigger source for DS1000D series Source Source _ Rising Trigger on rising edge sions Falling Trigger on falling edge T Rising amp Trigger on both ring amp falling edge Falling Auto Acquire waveform even no trigger occurred Sweep Normal Acquire waveform when trigger occurred Single When trigger occurs acquire one waveform then stop setup A To go to Set Up menu see table 2 38 2 36 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Pulse Width Trigger Pulse Width Trigger Mode Pulse trigger occurs according to the width of pulse The abnormal signals can be detected through setting up the pulse width condition Press Trigger IMENU Mode Pulse to display the pulse width trigger menu as follows
53. e B HewFile_ii 3 a B HewFfile 2 Files B HewFfile_3 7 ByNewFile_4 Delete File B HewFile_S B NewFile_ EX NewFile_T B HewFile_ amp B NewFile_o t File information Figure 2 83 File system display 2008 RIGOL Technologies Inc 2 63 User s Guide for DS1000E DS1000D Series RIGOL Disk Manage Press Storage Disk Mana to display the following menu Figure 2 84 Table 2 51 The Disk Manage Menu Page 1 2 Disk Mana Explorer Switch among Path Directory and Explorer Directory File Mew Folder File Deets Fil New folder To create new folder same as new reste files see table 2 13 Delete File ae Delete file Delete File waveform Pass Fail file i Figure 2 85 Table 2 52 The Disk Manage Menu Page 2 2 Rename Pa To rename a file see table 2 52 Rename Diskino A Display disk information 2 64 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Rename Press Storage Disk Manage Rename to go to the following menu Figure 2 86 EE 2 53 The Rename Menu Settings Comments Rename Rename F gt SFIGOL HewFile_e bme I F B HewFile_G EB HewFile_i g O File Hame WewFile 6 BICID E F S AH IJKLM HIOIF D E STUY u sY 1ij 2 3 4 5 6 7 Ss 3 _ Ena File Size 146E Figure 2 87 Rename the file Key Points Factory The oscilloscope has default settings and can be recalled at anytime Memor
54. elect to trigger on odd field Figure 2 48 Table 2 26 The Video Trigger menu Page 2 2 Sync is line U negative Trigger a Hine Numi 4 Select the specified line number 272 lt Line sync gt for sync Line Mur PAL SECM Standard 7 a Select Video standard awe Acquire waveform even no trigger Auo occurred Acquire waveform when trigger occurred When trigger occurs acquire one waveform and then stop To go to set up menu see table Set Up 3 39 2 40 2008 RIGOL Technologies Inc sers Guide for erles Normal Single RIGOL NOTE The line number ranges can be set from 1 to 525 under NTSC standard and from 1 to 625 under PAL SECAM Figure 2 49 Table 2 27 The Video Menu When the Sync is set as All lines Odd Trigger field and Even field E 22 PAL SECAM l Standard Standard NTSC Select Video standard MTS Acquire waveform even no trigger CA Auto Bice occurred Acquire waveform when trigger Normal q Jg Set Up occurred When trigger occurs acquire one waveform and then stop Set Up nn To go to set up menu see table 2 39 Single Key points Sync Pulses When positive polarity is selected the trigger always occurs on negative going sync pulses If the video signal has positive going sync pulses use the negative polarity selection hlode Y ode Source aea Dead Source Time SHH Ans Time SHH Ane Figure 2 50 Figure 2 51 Video Trigger Line Synchronization Video
55. esesseeeeeerereneeeeeens 3 8 Example 7 Triggering On a Video SiGnal cccceccecseeeeeeeeeeeeaeeeeeeeeaneeees 3 10 Example 8 FFT Cursor measurement csceceeseseseesetseeeeeeeeeareraneneeeers 3 12 2008 RIGOL Technologies Inc VII User s Guide for DS1000E DS1000D Series RIGOL Example 9 Pass Fail TeSt icicisntcicccietelaeseutadtcweeceineueeticthaadeeievimeesdidanmeeds 3 13 Example 10 Triggering on a Digital Signal cccccsceceeeeeeeeeeeeeseeeeaees 3 14 Chapter 4 Troubleshooting ssssnssnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn 4 1 Chapter 5 Specifications scccscsssecsceseneseeceneseeceneseeceneeseeeseeseeeseesenenens 5 1 SPECINCAONS ariaa iaa vied Ea AAE E a tan ueavenvetetadvauaneauees 5 2 General Specifications nsssenssnsnnrnsnnnnnnurnnnurnnnurnnnurnnnnrnnnurnrnnrnennrnennne 5 6 Chapter 6 APD nNndixX iiidenecavecaieducendsnacscavececwcndsuadenecenacuvavaneducedaueduvavececus 6 1 Appendix A ACCESSOrIES nper ae a ai 6 1 Appendix B Waranty aiaviieicaniteturncannamiaceuanistunans A 6 2 Appendix C Care and Cleaning ccsccsescecsecseseersecsensersesseneersensensersansenss 6 3 Appendix D Contact RIGO wuissesasitdutieianiniua ca teteatwebaieaetdaveincadeusiatenesuent 6 4 MOEN Gausetaiienthe panacea E TINGE ERNE RNE SHER ReEneea RoR 1 VIII 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSI000D Series RIGOL Chapter 1 Quick Start This chapter covers the follo
56. ff display To use trigger Holdoff 1 Press the trigger key to display Trigger Menu 2 Press Set Up key to display trigger set up menu 3 Turn the multi function knob 44 to change Holdoff time until waveform is stable 4 Press Reset can reset the Holdoff time to its default value 2 50 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Trigger Key points 1 Trigger Source Trigger can occur from several sources Input channels CH1 and CH2 AC Line Ext CH1 or CH2 It is the most commonly used trigger source The channel works when selected as a trigger source whatever displayed or not Ext Trig The instrument can be triggered from a third source while acquiring data from CH1 and CH2 For example to trigger from an external clock or with a signal from another part of the test circuit The Ext trigger sources use an external trigger signal connected to the EXT TRIG connector Ext uses the signal directly it has a trigger level range of 1 2V to 1 2V AC Line AC power can be used to display signals related to the power line frequency such as lighting equipment and power supply devices The oscilloscope gets triggered on its AC power input an AC trigger signal is not required When AC Line is selected as trigger source the oscilloscope automatically set coupling to DC set trigger level to OV Sweep Mode The sweep mode determines how the oscilloscope behaves in the absence of a tri
57. following steps 1 Connect printer DS1000E DS1000D series oscilloscopes provide two print modes Normal and PictBridge If use PictBridge printer please select PictBridge mode and then connect the oscilloscopes regarded as Device equipment and printer via USB Device port on the rear panel ee ee En A 5 i s gt Figure 2 108 PictBridge print If use Normal printer please select Normal mode and then connect the oscilloscopes regarded as Host equipment and printer via USB Host port on the front panel pef wom tm _ SEOL O a Oi i Opa TEBE eS e BP of of S O O s I Figure 2 109 Normal print 2008 RIGOL Technologies Inc zal User s Guide for DS1000E DS1000D Series RIGOL 2 To set up print parameters Press Utility Print set to go to the following menu Figure 2 110 Table 2 71 The Print Set Menu Normal Mode Print Mode Normal ee Select the print mode PictBridge Print Execute print function ON OFF Print Set Printhiod Inverted Set up the invert of Inverted UPN printing color on or not Palette SCAG Set up print color Color PP Palette L Figure 2 111 Table 2 72 The Print Set Menu PictBridge Mode Page1 3 Printhtod Normal ictEride Print Mode Select the print
58. gger event The oscilloscope provides three trigger modes Auto Normal and Single Auto This sweep mode allows the oscilloscope to acquire waveforms even when it does not detect a trigger condition If no trigger condition occurs while the oscilloscope is waiting for a specific period as determined by the time base setting it will force itself to trigger When forcing invalid triggers the oscilloscope cannot synchronize the waveform and the waveform seems to roll across the display If valid triggers occur the display becomes stable on the screen Any factor results in the un stability of waveforms can be detected by Auto Trigger such as the output of Power supply NOTE When horizontal control is set under 50 ms div Auto mode allows the oscilloscope not to capture trigger signal Normal The Normal mode allows the oscilloscope to acquire a waveform only when it is triggered If no trigger occurs the oscilloscope keeps waiting and the 2008 RIGOL Technologies Inc 2 51 User s Guide for DS1000E DS1000D Series RIGOL previous waveform if any will remain on the display Single In Single mode after pressing the RUN STOP key the oscilloscope waits for trigger While the trigger occurs the oscilloscope acquires one waveform then stop 3 Coupling Trigger coupling determines which signal component passing to the trigger circuit Coupling types include AC DC LF Reject and HF Reject AC AC coupling blocks DC
59. gure 2 28 Table 2 14 The Import menu Import Explorer Explorer Directory File Switch among Path Directory and File Import the REF file into internal memory D 1 im p T al F RIGOL SREF_ amp REF s Explorer B REF REF EB Cors bmF fy RIGOL E Cofy_1 bme EB Corps_ J bme Cory 3 bme EB Cop y_4 sme B HewFile_ B HewFile_i E HewFile_ie B HewFile_ii E HewFile_2 E HewFile_3 r ize 162M Figure 2 29 Import interface 2008 RIGOL Technologies Inc 2 19 User s Guide for DS1000E DS1000D Series RIGOL 5 Displaying a Reference Waveform Figure 2 30 Reference waveform display 1 Push button to show the reference waveform menu 2 Press soft button No 1 to select the reference channel CH1 CH2 MATH FFT or LA DS1000D series 3 Turn vertical and vertical to adjust the REF waveform to a suitable position 4 Press soft button No 2 to select the save location of REF waveform 5 Press soft button No 3 to save the waveform as REF NOTE The reference function is not available in X Y mode 2 20 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Set up LA Channel Only for DS1000D Series 16 digital channels logical analyzer are provided for DS1000D series which can achieve mixed signal measure in line with the dual analog channels Single channel or group channels can be chosen ON or OFF and also can set the size
60. hen stop Set Up ra To go to set up menu see table 2 40 1 H High Logic high voltage is higher than threshold setting 2 L Low Logic low voltage is lower than threshold setting 3 X Ignore Don t care If all the channels are ignored the oscilloscope won t be triggered 4 Qualifier A timer begins when code terms are satisfied Duration trigger occurred in the time set by the qualifier Key Points 2008 RIGOL Technologies Inc 2 47 User s Guide for DS1000E DS1000D Series RIGOL Trigger Setup Set up different trigger settings according to different trigger modes When choosing source as D15 D0 in the mode of Edge and Pulse DS1000D series only Holdoff is adjustable When source is non digital channel and in slope trigger only trigger coupling trigger sensitivity and Holdoff can be set For video trigger Sensitivity and Holdoff can be set For the pattern trigger and the duration trigger DS1000D series only Holdoff is adjustable Figure 2 63 Table 2 39 The Trigger Set Up Menu Settings for trigger coupling trigger sensitivity and holdoff Menu Settings Comments DC Allow all signals pass AC Block DC signals Coupling HF Reject Reject high frequency signals LF Reject Reject DC and low frequency Set Up Coupling Sensitivity A aadi Holdott Signals 4 Set trigger sensitivity the lt Sensitivity Setting gt range is from 0 1div to idiv Set time slot before another Holdoff
61. i As figure 2 36 shows D DU On off channels Channel list and status of on off Time SHA Bns Figure 2 36 Turn on or off the digital channel 2 Force turn on or off all digital channels Press LA gt D7 DO Turn On Turn Off or D15 D8 Turn On Turn Off will force to turn all the channels on off If you want to turn on off any single channel instead select the Channel by turning 4 knob then press No 1 soft button or 44 knob 3 Set up the viewing size of digital channels Press LA D7 D0 Size or D15 D8 Size to select wave size of logic channels Select tJL to view 8 channels on the screen Select to viewall of the 16 channels on the screen 4 Reset the digital channels display Press LA D7 D0 Reset or D15 D8 Reset to reset the display of digital channels 2 24 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Turn on off Channels The CH1 CH2 Ext Trigger and LA DS1000D series channels are input channels All functionalities applied will be based on operating the instrument with channels So MATH and REF can be regarded as relatively isolated channels To turn on off any one of the channels press the corresponding button on the front panel The key backlight indicates the channel is currently active Press the button again to turn the channel off Or when channel is currently selected press will turn the channel off as well and the key bac
62. ilt in help menu enables information getting more convenient Multiple Language User Interface support Chinese amp English input Support U disk and local files storage Waveform intensity can be adjusted To display a signal automatically by Pop up menu makes it easy to read and use VI 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSI000D Series RIGOL Content Sato NOUICES aaoi datesantsaauoneebanusnaeseetonedeutaGundduiuimooimeumaesiauaanduatunnns II General Purpose OSCIIOSCOPDES ccsceceeeesecesecenereeseseeeeeaeaneeaeaesesenenensnnes V Chapter Quick Start irsana EN 1 1 To Inspect the Instrument wii siccscsictoratceicrtaneicreianieaeeniatie eaten 1 2 The Panels and User Interface cccscsecseseersesseseersessesearseseenentensassenserss 1 3 Appearance and DIMENSIONS cscscsceesecseneeusessenserseseenearseseenearenseseenterss 1 7 TO Performa Functional CHECK iirisra tiniest a 1 8 TO Compensate Probes iirinn 1 10 Digital Leads Only for DS1000D Series s nsnnonennonnnnsnnnnsnnrrsrnrrrnnrrenne 1 11 To Display a Signal Automatically ccccsceceeceeeeeeeeeeeeeeaeeeeeeeeaeseeeeeeanes 1 13 To Understand the Vertical System scseecsesseseareeseesenrenseenenrersensenserss 1 14 To Understand the Horizontal SyStem cccccecseeeseeeeeseaeeeeeeeeaeeeseseeaees 1 16 To Understand the Trigger System essssssssreresnernrrrrnrrernrrernrrernrrennne 1 18 Chapter 2 Operat
63. ime To select the slope condition NOTE Slope time can be set from 20ns to 10s When a signal meets the trigger condition The oscilloscope will execute the acquisition Figure 2 46 Table 2 24 The Slope Trigger Menu Page2 2 Select the level that can be adjusted by LEVEL Acquire waveform even when no trigger condition is met Acquire waveform when trigger condition is met When trigger condition is met acquire one waveform and then stop Set Up p To go to set up menu See table 2 38 NOTE You can adjust LEVEL A LEVEL B or both simultaneous by turning the LEVEL knob 2008 RIGOL Technologies Inc 2 39 User s Guide for DS1000E DS1000D Series Normal Single RIGOL Video Trigger Video Trigger Mode Video trigger is used to capture the complicated waveforms of television The trigger circuitry detects the vertical and horizontal interval of the waveform and produces triggers based on the video trigger settings you have selected Choose video trigger to trigger on fields or lines of NTSC PAL SECAM standard video signals Trigger coupling preset to DC Figure 2 47 Table 2 25 The Video Trigger Menu Page 1 2 Trigger Mode CH Selects CH1 as trigger source Source EXT Select CH2 as trigger source Select EXT TRIG as trigger source Polarity is Triggers on positive going sync pulses ji positive Polari ty Triggers on negative going sync pulses An Lines Tigger on all ines Odd field S
64. ing Your OscilloScCOpe sssssnssnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn 2 1 To Set up the Vertical SYSteM cscssccsecsesrersessensersesseseareeseenentensessenserss 2 2 To Set up the Horizontal System cccceceeeeseceeseeeeeeeeeeeeaeeeeeeetaeseneeeeanes 2 27 To Set Up the Trigger System a vecieiiiccitsen aver atcnied inn voveadoiniiuiiaddaWeneueetees 2 34 To Set up the Sampling System cccecscseersesscsearseseesenreuseesestersersenserss 2 53 To Set up the Display System cccecsesseseercessesearceseeneareuseesensersensenserss 2 57 TO Store and Recall isnin aA 2 59 TO Setup the UUl SYSTEM isiin dinin a 2 67 TO Meastire Automatically ansin aa 2 88 To Measure with CUrSOTS sssnssssenssnnnnnnnnnrnnrunnnnunnnnurnnnurnnnurnrnnrnrnnrnennne 2 95 To Use Run Control Buttons s ssesessssennsssrnrnsrrrnnrnrnnnnrnnrnrnnnnennnnennnnnnn 2 102 Chapter 3 Application amp Examples csscccscssseesessseenesssceseeseneseesenesees 3 1 Example 1 Taking Simple Measurements cccsceeeeeeeeeeeeeeeeeeeeaeeeeeeseaees 3 1 Example 2 View a Signal Delay Caused by a Circuit ccccceeeeeeeeeeeeeaees 3 2 Example 3 Capture a Single Shot Signal cccscseseseceeeeeeeeeeeeeaeeeseeeeaees 3 3 Example 4 To Reduce the Random Noise on a SiQnal ccsccseseeeseeeeeeeees 3 4 Example 5 Making Cursor Measurements cccceceeeeeeeeeeeaeeeeeeeeaeeeeesseaees 3 6 Example 6 The application of the X Y operation scsccec
65. ing the exercise to familiarize with the buttons knobs and status bar HORIZONTAL mm lt POSITION MENU lt SCALE gt Fs Figure 1 19 The horizontal system 1 Turn the knob and notice the change in the status bar The horizontal knob changes the sweep speed in a 1 2 5 step sequence and displays the value in the status bar The time base ranges of the oscilloscope is from 2ns div to 50s div Delayed Scan Shortcut key To press the knob in the horizontal control area on the front panel is another way to enter or exit Delayed Scan mode and it is equal to the following menu operations MENU gt Delayed gt ON NOTE The speed of horizontal scan varies by different models 1 16 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL 2 The horizontal knob moves displayed signal horizontally on waveform window Turn the horizontal knob to adjust trigger offset The waveform will move horizontally at the same time Horizontal offset back to 0 shortcut key Press the POSITION knob to set the horizontal offset to 0 as a shortcut key this is especially helpful when the trigger point is far out of the screen and want it to get back to the screen center immediately 3 Press the MENU button to display the TIME menu To enter or exit the Delayed Scan mode set the display to Y T X Y or ROLL mode or reset the trigger offset Horizontal position control Trig
66. isplay Comments Preference Set Keys ra Set up new code 2 82 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Key Points 1 Screen saver This function extends the life of LCD backlighting system 2 Expand reference When changing the volts div for channels the signal expands or compresses around the signal ground level or the center of the screen When Center is selected the waveform will expand or compress around the center of the display When Ground is selected the channel ground level will remain the same position on the display and waveform will zoom about the ground level 3 Sticky key If sticky feature is turned ON when adjusting positions CH1 CH2 MATH REF Trig Lev and Trig Pos the object will stop at zero position until next adjustment for the ease of getting back to initial positions 4 Key Sets This function help users to set new code to unlock the keyboard 2008 RIGOL Technologies Inc 2 83 User s Guide for DS1000E DS1000D Series RIGOL Self Calibration The Self Calibration adjusts the internal circuitry to get the best accuracy Use these functions to calibrate the vertical and horizontal systems For maximum accuracy at any time run this calibration if the ambient temperature changes by 5 C or more Disconnect any probes or cables from all channel inputs otherwise failure or damage to the oscilloscope may occur Press Utility gt Self Cal
67. klight also goes off The channel status symbol is displayed at the lower left of the screen which can help users to judge the channel status quickly Table 2 18 Status of the channels Channel Mode Settings Status Indicator ON CH1 yellow letter Channel 1 CH1 Selected CH1 black letter OFF No indicator ON blue letter Channel 2 CH2 Selected CH2 black letter OFF No indicator ON purple letter Selected Math black letter OFF No indicator NOTE Pressing LA will turn all the digital channels on off 2008 RIGOL Technologies Inc 2 25 User s Guide for DS1000E DS1000D Series RIGOL Set up Vertical Position and Scale You can use the vertical controls to display waveforms adjust vertical SCALE and 23POSITION and set input parameters 1 Using vertical POSITION knob The verticalPOSITION control changes the position of signal waveforms in all channels including MATH and REF The resolution changes according to the vertical level set Pressing this knob will clear the channel offset to zero The function is available for DS1000D series but not includes digital channel 2 Using vertical SCALE knob The vertical SCALE can change the vertical sensitivity of waveforms in all channels including MATH and REF excluding LA If the Volts Div is set to Coarse the waveform scales in a 1 2 5 step sequence from 2 mV to 5 V If the Volts Div is set to Fine it scales to small steps
68. le WINdOW sceceeees 2 12 Equivalent Sampling 2 56 RE Feer mau aeaentaeyeseae gee 2 14 EXE and EXU Sincssusentsavesimenepenent 2 51 Roll Mode Display sceceeees 2 29 Factory Setup ccccccecesesscnenens 2 65 RUINS TOP imase 2 102 ir E E E E E 2 11 Sampling System cscsceeeeeees 2 53 FORCE auaa 1 18 2 34 Save and ReCal ccsecsseseevevevaees 2 59 Front panel s ccsecsessersessenseres 1 3 Self Calibration cccsceceeeeees 2 84 Functional Check cccceeeeeeeeees 1 7 Single Trigger 2 52 GND COupling ccceceeceeeeeeeeees 2 4 Slope TOG sasieiicavodseetvavarcaien 2 39 Hamming WINdOW 0ccseeeeees 2 12 STORAGE nna vesienaaee 2 59 Hanning Window scsecseeeeeees 2 12 Sync PUISCS etic a eecniierne 2 41 2008 RIGOL Technologies Inc 1 RIGOL User s Guide for DS1000E DS1000D Series TIAJ i 5 4 Vertical System csceceeeeeeeeeaees 2 2 Trigger COntrOl ccccceeeeeeeeees 1 17 Vertical WiNdoOW ssssssessssesesnnn 1 13 Trigger System ccceeeeeeeeeees 2 34 Video TrIQGGer csscseseersecsenseres 2 40 JTI TY cenana 2 67 eV AEE E E TEE 2 29 Venical sree 5 3 NOU E E EE 2 29 2 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series
69. left part to the trigger point then wait for trigger when trigger occurs it continues to draw the rest part from the trigger point to the end of the right side When choosing this mode to view low frequency signals it is recommended that the channel coupling be set as DC Time Div Horizontal scale The Time Div control expands or compresses the waveform 2008 RIGOL Technologies Inc 2 29 User s Guide for DS1000E DS1000D Series RIGOL Delayed Scan The Delayed Scan is a magnified portion of the main waveform window Use Delayed Scan to locate and horizontally expand part of the main waveform window for a more detailed higher horizontal resolution analysis of signal The Delayed Scan time base setting cannot be set slower than the Main time base setting Press horizontal IMENU Delayed ON or press horizontal knob to enter Delayed Scan mode Waveform to be horizontally expanded Time base of Delayed Scan Time 18 88us Expanded waveform in horizontal Main timebase Figure 2 39 Delayed Scan window The screen splits into two parts The upper half displays the main waveform window and the lower half displays an expanded portion of the main waveform window This expanded portion of the main window is called the Delayed Scan window Two blocks shaded at the upper half the un shaded portion is expanded in the lower half The horizontal and knobs control the size and position of the Delayed Scan The value at b
70. lloscopes provide four FFT windows Each window is a trade off between frequency resolution and amplitude accuracy What you want to measure and your source signals characteristics help determine which window to use Use the following guidelines to select the best window Table 2 7 FFT Windows Window Features Best for measuring Best frequency Transients or bursts the signal Resolution and worst levels before and after the event magnitude resolution are nearly equal This is essentially the Equal amplitude sine waves with same as no window fixed frequencies Broadband random noise with a relatively slow varying spectrum Rectangle Better frequency poorer magnitude accuracy than Rectangular Hamming has slightly better frequency resolution than Sine periodic and narrow band random noise Transients or bursts where the Signal levels before and after the events are significantly different Hanning Hamming Hanning Blackman Best magnitude worst Single frequency waveforms to frequency resolution Find higher order harmonics 2 12 2008 RIGOL Technologies Inc ser s Guide for eries RIGOL Key points FFT Resolution the quotient between sampling rate and number of FFT points With a fixed FFT points the lower sampling rate results in better resolution Nyquist Frequency The highest frequency that any real time digitizing oscilloscope can acquire without aliasing It s norma
71. lly half of the sample rate This frequency is called the Nyquist frequency Frequency above the Nyquist frequency will be under Sampled causing a situation known as aliasing 2008 RIGOL Technologies Inc 2 13 User s Guide for DS1000E DS1000D Series RIGOL REF Function Reference Waveforms are saved waveforms to be selected for display The reference function will be available after saving the selected waveform to non volatile memory Press button to display reference waveform menu Figure 2 18 Table 2 8 REF Menu when using internal memory Source Select channell as REF channel Select channel2 as REF channel Location Source MATH FFT Select MATH FFT as REF channel elect LA as REF channel DS1000D series LA S Save Locion Internal Select memory location in scope rnp Eve External Select memory location out scope RH CAp Save __ Save REF waveform Imp Exp Lee Go to import export menu see table 2 10 Preset __ Reset REF waveform Figure 2 19 Table 2 9 REF Menu when using external memory Source Select channell as REF channel Select channel2 as REF channel scaly SOUTCE MATH FET Sel h hannel Select Math FFT as REF channe Select LA as REF channel DS1000D series LA save i Location Internal Select internal memory location Import External Select external memory location TIRI Save Save REF waveform to outer memory location Import Go to import menu see table 2 14 Preset __ Rese
72. ly single channel can be used 2 For DS1000D Series 2008 RIGOL Technologies Inc 5 5 User s Guide for DS1000E DS1000D Series RIGOL General Specifications Backlight 300 nit Brightness typical Probe Compensator Output Output Voltage typical Approximately 3Vpp peak to peak Frequency typical Supply Voltage 100 240 VACpms 45 440Hz CAT II Power Consumption Less than 50W 2A T rating 250 V o Humidity ee E Non operating 15 000 m or below Size Width Height Depth Heavy Without package 2 3 kg Packaged 3 5 kg 5 6 2008 RIGOL Technologies Inc User s Guide for DSI000E DS1000D Series RIGOL IP Protection IP2X Calibration Interval The recommended calibration interval is one year 2008 RIGOL Technologies Inc 5 7 User s Guide for DS1000E DS1000D Series RIGOL Chapter 6 Appendix Appendix A Accessories Standard Accessories Probex2 1 5m 1 1 or 10 1 adjustable Passive Probes The passive probes have a 6MHz bandwidth with a rating of 150V CAT II when the switch is in the 1X position and a Full oscilloscope bandwidth with a rating of 300 V CAT IT when the switch is in the 10X position A Power Cord that fits the standard of destination country A Data Cable only for DS1000D series A Active Logic Head only for DS1000D series 20 Logic Testing Nips only for DS1000D series 20 Logic Testing leads only for DS1000D series A CD ROM including User s Guide and Appli
73. mode ELENEN PictBridge P Execute print function Stop print function Query the status of print job 2 80 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSI000D Series RIGOL Figure 2 112 Table 2 73 The Print Set Menu PictBridge Mode Page2 3 Print Set Default Select the print paper size aa Paper Size A4 A5 and default A2 A3 A4 A5 A6 Paper size so on and B5 Default File Type Bmp Select the print image type Exif Jpeg File Type Default Normal Draft Fine Palette Gray scale Inverted Color Inverted a Turn the invert on or off L Print Quality Print Quality Select the print quality 3 Print Make sure the Printer has been connected and the print settings are all ready before printing Then press Utility gt Print set gt Print to execute print operation 2008 RIGOL Technologies Inc 2 81 User s Guide for DS1000E DS1000D Series RIGOL Preference Press Utility Preference to go to the following menu Figure 2 114 Table 2 75 The Preference menu 1 2 Preference Mem Display Commens _ Screen saver i min Expand Refer Screen saver Set up screen saver timer 5 hour up waveform vertical Expand Refer Center pan reference Set up sticky state of CH1 Sticky key CH2 Math Ref Trig Level and Trig Position Classical Modern Tradition Succinct Set up skin style Figure 2 115 Table 2 76 The Preference menu 2 2 Men D
74. ning Clean the instrument and probes regularly based on its operating conditions requires To clean the exterior surface perform the following steps 1 Disconnect the instrument from all power sources 2 Clean the loose dust on the outside of the instrument and probes with a lint free cloth with a mild detergent and water When clean the LCD take care to avoid scarifying it AN WARNING To avoid injury resulting from short circuit make sure the instrument is completely dry before reconnecting into a power source 2008 RIGOL Technologies Inc 6 3 User s Guide for DS1000E DS1000D Series Appendix D Contact RIGOL If you have any problem or requirement during using our products please contact RIGOL Technologies Inc or the local distributors Domestic Please call Tel 86 10 8070 6688 Fax 86 10 8070 5070 Service amp Support Hotline 800 810 0002 9 00 am 5 00 pm from Monday to Friday Or by e mail service rigol com Or mail to RIGOL Technologies Inc 156 CaiHe Village ShaHe Town ChangPing District Beijing China Post Code 102206 Overseas Contact the local RIGOL distributors or sales office For the latest product information and service visit our website www rigol com 6 4 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series Index O PEA E 1 18 2 34 AFRO ECEnorainie a 2 52 AG COUPINO wiseetsnastuetis cidni 2 3 IGKIZON Al ses Sgaaecniaseneasacdeieaaeatins 5 2 AG TI
75. nt of the signal by simply noting its distance from the ground symbol If the channel is AC coupled the DC component of the signal is blocked allow you to use greater sensitivity to display the AC component of the signal 1 14 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Vertical offset back to 0 shortcut key Turn the knob to change the vertical display position of channel and press the knob to set the vertical display position back to 0 as a shortcut key this is especially helpful when the trace position is far out of the screen and want it to get back to the screen center immediately 2 Change the vertical setup and notice that each change affects the status bar differently View the status bar which is on the bottom of the screen to understand the vertical scale Change the vertical scale by turning the knob and notice the change in the status bar Press CH1 CH2 MATH REF LA only for DS1000D series to display the menu sign waveform of corresponding channel Press OFF button to turn off the channel Coarse Fine Shortcut key The Coarse Fine vertical control can be set by simply pressing the vertical SCALE knob 2008 RIGOL Technologies Inc 1 15 User s Guide for DS1000E DS1000D Series RIGOL To Understand the Horizontal System Figure 1 19 shows the HORIZONTAL controls IMENU button POSITION and knobs of horizontal system Follow
76. nu explanation Press the button to display the menu for the settings of the Automatic Measurements The oscilloscopes provide 22 auto measurements Vpp Vmax Vmin Vtop Vbase Vamp Vavg Vrms Overshoot Preshoot Freg Period Rise Time Fall Time Delay1 gt 2 Delayl 2t Phasi 2 Phasi 2 Width Width Duty Duty 10 voltage and 12 timing measurements Figure 2 122 Table 2 77 The Measure menu Source CH1 Select CH1 or CH2 as source channel CH2 for measurement Voltage aa Select to measure voltage parameter Source CH1 Time A lt Select to measure time parameter clear AA Clear measurement result on screen Display OFF Turn off all measurement result All ON Turn on al measurement result Display All GFF 2 88 2008 RIGOL Technologies Inc sers Guide for erles RIGOL 10 Voltage Measurement Parameters Figure 2 123 Table 2 78 The Voltage Measurements menu Page 1 3 Measure maximum voltage of a waveform Measure minimum voltage of a waveform 4 ohare een Yimin Measure Peak to Peak Voltage Measure a flat top voltage of a square waveform Viton ka Figure 2 124 Table 2 79 The Voltage Measurements menu Page 2 3 Measure a flat base voltage of a square waveform Measure voltage between Vtop and Vbase Measure average voltage of a waveform Measure Root Mean Square Voltage of a waveform Figure 2 125 Table 2 80 The Voltage Measurements menu Page 3 3
77. o see the Cursor menu Press Mode to set Manual mode Press Type to select Y Turn 4 4 knob to place cursor A on the first peak of the wave Turn 4 knob to place cursor B on the second peak of the wave Observe the following measurements in the cursor menu See Figure 3 5 The delta voltage peak to peak voltage of the waveform The voltage at Cursor 1 The voltage at Cursor 2 CurA 1 68 Curb 386ml Time SHH Bus Figure 3 5 Waveform display 2008 RIGOL Technologies Inc 3 7 User s Guide for DS1000E DS1000D Series RIGOL Example 6 The application of the X Y operation Viewing Phase Changes through a Network Theme Connect the oscilloscope to monitor the input and output of the circuit and capture the phase changes To view the input and output of the circuit in an X Y display do these steps 1 From the probe menu set the attenuation to 10X Set the switch to 10X on the probes 2 Connect the CH 1 probe to the input of the network and connect the CH 2 probe to the output 3 If the channels are not displayed press the and CH2 buttons 4 Press the button 5 Adjust the vertical SCALE knob to display approximately the same amplitude Signals on each channel 6 Press the MENU in horizontal control area to display the menu 7 Press the Time Base soft button to select X Y The oscilloscope displays a Lissajous pattern representing the input and output characteristics of the circuit
78. of waveform Change display location of digital channel on screen and select threshold style Press LA function button and go to the following menu Figure 2 31 a 2 15 e LA menu LA 07 0 D7 D0 Settings up channel group D7 DO see table a 015 08 D15 D8 _ set up channel group D15 D8 see 2 17 Current l 9 lt D15 om Select channel by turning 44 knob Threshold Select mode of whole digital channels Threshold The threshold voltage can set by user User when in user defined style a i 1 Ami 43 eer ae Set threshold voltage by turning 42 knob Voltage gt 1 Display and re line up the digital channels 1 Press LA D7 D0 or D15 D8 and go to the group channel setting menu Turn on or turn off the display of the digital channels 2 Press LA current and choose digital channel by turning knob The chosen channel will display in red color 3 Turn vertical POSITION knob to re position the channel in screen The figure of menu shows as follow 2008 RIGOL Technologies Inc 2 21 User s Guide for DS1000E DS1000D Series RIGOL LA D7 D0 4 GE oio BE a gt running vo channel Threshold Running aoe channel Ser Time SHH Ans Figure 2 32 Turn on the digital channel 2 Set threshold mode of digital channels Press LA gt Threshold select logic standard or User to define your own threshold voltage The figure of menu shows as follow LA O7 O0 015 08 Current 4 49 T CM
79. oid electric shock the instrument grounding conductor s must be grounded properly Before making connections to the input or output terminals of the instrument Connect The Probe The probes ground terminals are at the same voltage level of the instrument ground Do not connect the ground terminals to a high voltage Observe All Terminal Ratings To avoid fire or shock hazard observe all ratings and marks on the instrument Follow the user s guide for further ratings information before making connections to the instrument Do Not Operate Without Covers Do not operate the instrument with covers or panels removed Use Proper Fuse Use the fuse of the type voltage and current ratings as specified for the instrument Avoid Circuit or Wire Exposure Do not touch exposed connections and components when power is on Do Not Operate With Suspected Failures If suspected damage occurs with the instrument have it inspected by qualified service personnel before further operations Provide Proper Ventilation Refer to the installation instructions for proper ventilation of the instrument II 2008 RIGOL Technologies Inc User s Guide for DSIO00E DS1000D Series RIGOL Do not operate in wet damp conditions Do not operate in an explosive atmosphere Keep product surfaces clean and dry The disturbance test of all the models meet the limit values of A in the standard of EN 61326 1997 A1 A2 A3 but can t meet the limit values of B
80. ologies Inc 4 1 User s Guide for DS1000E DS1000D Series RIGOL proper range by turning the LEVEL knob or pressing the button Or set the Trigger Mode as AUTO Moreover push button to display the waveform on screen 6 After the Acquisition is set to Averages or Display Persistence is set ON the waveform refreshes slowly It is normal in these settings 7 The signal ts displayed as ladder like waveform 1 The time base setting maybe is too low Turn the horizontal knob to increase horizontal resolution to improve the display 2 Maybe the display Type is set to Vectors set it to Dots mode to improve the display 4 2 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Chapter 5 Specifications All specifications apply to DS1000E DS1000D Series Oscilloscopes unless where noted To come up to these specifications two conditions must be met firstly The instrument must have been operated continuously for 30 minutes under the specified operating temperature Do perform Self Calibration operation through the Utility menu if the range of operating temperature variations up to or more than 5 C NOTE All specifications are guaranteed unless where marked typical 2008 RIGOL Technologies Inc 5 1 User s Guide for DS1000E DS1000D Series RIGOL Specifications Sample Modes Real Time Sample Equivalent Sample Sample Rate 1 DS1102X DS1052X 1GSa s
81. ord erate ba Set start frame e Frame lt 1 1000 gt Start Frare Curreri 4 Le Select current frame to be played ES Frame lt 1 1000 gt LHET Prarie t2 End Frame v Set End frame Fric Rey aga lt 1 1000 gt nd Frame NOTE The RUN STOP button can also replay or continue the waveform display 4 Storage Store recorded waveforms to non volatile memory according to setup frames Press Utility Record Mode Storage to go to the following menu Figure 2 105 Table 2 68 The Storage menu Page 1 2 Record acor Menu Settings Comments 4 Start Frame Set first frame to be saved lt 1 1000 gt Start Frare x9 EESE End Frame Set last frame to be saved End Frame lt 1 1000 gt Internal A Location Set up Store location Location ae i 2008 RIGOL Technologies Inc 2 77 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 106 Table 2 69 The Storage menu Page 1 2 Record When the save to Internal memory Figure 2 107 Table 2 70 The Storage menu when the save as external memory Go to save menu same as REF save menu See table 2 12 Go to load menu see table 2 63 Go to import menu same as REF Import import menu see table 2 14 Load Import tL 2 78 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Print Setting DS1000E DS1000D series oscilloscopes support pint function Please execute print operation according to the
82. ottom of the screen is the main time base and the value on the center bottom means the Delayed Scan time 2 30 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL Use the horizontal knob to change the position of the expanded portion Use the horizontal knob to adjust the Delayed Scan resolution To change the main time base turn off the Delayed Scan mode Since both the main and Delayed Scan are displayed there are half as many vertical divisions so the vertical scaling is doubled Notice the changes in the Status bar Delayed Scan Shortcut Key Delayed Scan function can be activated not only by menu but also by pressing horizontal knob 2008 RIGOL Technologies Inc 2 31 User s Guide for DS1000E DS1000D Series RIGOL X Y Format This format is useful for studying phase relationships between two signals Channel 1 in the horizontal axis X and channel 2 in the vertical axis Y the oscilloscope uses a none trigger acquisition mode data is displayed as dots Press horizontal IMENU Time Base xX Y to enter the X Y mode Time Delayed Trig Offset Figure 2 40 X Y display format NOTE In Y T format all sample rates are available But in X Y format 100 MSa s is not available In common deceasing the sample rate can display the wavefrom better The following modes or functions will not work in X Y format LA Function DS1000D series Automatic Measurements C
83. points Rise Time Fall Time 90 50 10 o W Width lt Width Figure 2 132 Time parameters Rise Time Time that the leading edge of the first pulse in the waveform takes to rise from 10 to 90 of its amplitude Fall Time Time that the falling edge of the first pulse in the waveform takes to fall from 90 to 10 of its amplitude Width The width of the first positive pulse in 50 amplitude points Width The width of the first negative pulse in the 50 amplitude points Delay 1 2 The delay between the two channels at the rising edge Delay 1 2 t The delay between the two channels at the falling edge Phas 1 2 The phase deviation between the two channels at the rising edge Phas 1 2t The phase deviation between the two channels at the falling edge Duty Duty Cycle defined as Width Period Duty Duty Cycle defined as Width Period 2 94 2008 RIGOL Technologies Inc User s Guide for DSI000E DS1000D Series RIGOL To Measure with Cursors Figure displays the button on the front panel 1 MENU mm Cursor measurement setup button Figure 2 133 Cursor measurement setup button The cursor measurement has three modes Manual Track and Auto Measure 1 Manual Mode In this mode the screen displays two parallel cursors Move the cursors to make custom voltage or time measurements of the sign
84. rms or Vrms FFT function can not only to help you find crossfire problem and analog waveform distortion caused by nonlinear amplifier but also can adjust the analog filter FFT waveforms are useful in the following applications Measuring harmonic content and distortion in systems Characterizing noise in DC power supplies Analyzing vibration Press MATH gt Operate FFT to display FFT menu Figure 2 17 Table 2 6 The FFT menu Menu _ Settings ___Comments_ ties Add source A to source B rae Operate Subtract source B from source A ivi j AxB Multiply source B by source A CH1 F Fast Fourier Transform inc Display a oe l Hanning Scale Window Hamming Blackman Split Display FFT waveform on half screen Set Vrms as vertical unit dBVrms Set dBVrms as vertical unit Select window for FFT 2008 RIGOL Technologies Inc 2 11 User s Guide for DS1000E DS1000D Series RIGOL Key points for FFT Signals that have a DC component or offset can cause incorrect FFT waveform component magnitude values To minimize the DC component choose AC Coupling on the source signal To reduce random noise and aliases components in repetitive or single shot events set the oscilloscope acquisition mode to average To display FFT waveforms with a large dynamic range use the dBVrms scale The dBVrms scale displays component magnitudes using a log scale 2 Selecting an FFT Window The osci
85. s are only allowed to a digital signal Pattern Do the steps as following 1 Press the MENU key in the triggering control area to see the trigger menu 2 Press Mode to select Pattern mode 3 Turn 4 knob to select the channel you want to set 4 Press Code to set Code Setting H L X or 5 Press Sweep to select trigger mode to Auto Normal or Single 6 Press Set Up to adjust the Holdoff time Trigger ode Select DT Code Time 3 Figure 3 12 Pattern Trigger on a digital signal 3 14 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSi000D Series RIGOL Duration Do the steps as following 1 Press the MENU key in the triggering control area to see the trigger menu 2 Press Mode to select Duration mode 3 Turn 4 knob to select the channel you want to set 4 Press Code to set Code Setting H L or X 5 Press Qualifier to set time limit terms 6 Press Time to set duration and limit symbol time 7 Press Sweep to select trigger mode to Auto Normal or Single 8 Press Set Up to adjust the Holdoff time Trigger hode Select Dr Code CiUalifier Time 16 86us Figure 3 14 Duration Trigger on a digital signal 2008 RIGOL Technologies Inc 3 15 User s Guide for DS1000E DS1000D Series RIGOL Chapter 4 Troubleshooting 1 After the oscilloscopes is powered on the screen remains dark no display 1 Check the power cable connection 2 Check if the power switch is turned on 3 After
86. st Witilities Record Setup Waveform Recorder Print set Setup printing Pass Fall Record Print set Figure 2 91 Table 2 56 The Utility menu Page 3 3 Preference ee Go to preference menu Preference Self Cal Execute Self calibration System Info ra Display system information Potent Special Mode Lock Unlock Lock the keyboard Spectal hlode NOTE Self Calibration Oscilloscope will calibrate parameter of vertical system CH1 CH2 and Ext horizontal system and trigger system 2 68 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL I O setting Press Utility I O setting to go to the following menu Figure 2 92 Table 2 57 The oe Setup Menu VO Setup ee ee Set RS 232 baud rate as Job Device 300 2400 4800 9600 19200 or 38400 Sela 38400 an pees Set the USB d ae M PictBridge GPB Set the GPIB address NOTE Users must set the USB device type manually first to ensure the equipment connected is the same as the settings when switch the device from computer PictBridge printer to PictBridge printer computer by the USB Device port 2008 RIGOL Technologies Inc 2 69 User s Guide for DS1000E DS1000D Series RIGOL Language The oscilloscopes have multi language user menu choose as your desire Press Utility Language to select the language Litilities a IO Setting HPE ge cound 30 H Counter English Deutsch OFF Fran E
87. t Mode Auto automatic measuring See the following figure There will be no cursor display if no parameters are chosen in Measure menu The oscilloscope could move cursor automatically to measure 22 parameters in Measure menu Time 286 8us Figure 2 139 Auto Measure Mode of Cursor Measurement 2008 RIGOL Technologies Inc 2 101 User s Guide for DS1000E DS1000D Series RIGOL To Use Run Control Buttons The run control buttons include auto setting and RUN STOR Auto The features automatic adjustments to produce a usable display of the input signal Press button the following menu appears Figure 2 140 Table 2 90 The Auto menu Multi cycle JUUUL Press to display multi cycle ur Press to display single cycle Single Cycle waveform on the screen a Press to display the waveform s rising edge and measure its rise time automatically Press to display the waveform s falling edge and measure its fall time automatically Ta Press to cancel all the Auto Set Cancel actions the oscilloscope will recover to its previous status ise Edge 2 102 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Auto set functions After the is pressed the oscilloscope is configured to the following defaults Table 2 91 The Auto menu Menu Settings Vertical V div Horizontal position Horizontal S div Tiggertype Edge RUN STOP Run or Stop waveform acquiring
88. t REF waveform 2 14 2008 RIGOL Technologies Inc sers Guide for erles RIGOL 1 Import and Export Press REF gt Imp Exp and go to the following menu Figure 2 20 Table 2 10 The Imp Exp menu ImpJEXp Menu Settings Explorer Path Explorer Directory Switch to Path directory or file Export File Export the REF file from internal memory to export memory see table 2 11 Delete file ila Import the REF file to internal amore lt Import the REF fle to internal Pai memory g Delete File _ Imp Exp F RIGOL REF_ amp REF lt Explorer I F Bs REF_ REF Files E E HewFile_ A RIGOL B NewFile_i Export B HewF i le_ 1g B NewFile_il EB HewFile_2 B NewFile_Z Import E HewFile 4 EB HewFile_S EY HewFile_G Delete File B HewFile T B WewFile_S File Size 1 82ME6 Figure 2 21 Import export interface 2008 RIGOL Technologies Inc 2 15 User s Guide for DS1000E DS1000D Series RIGOL 2 Export Press REF Imp Exp Export and go to the following menu Figure 2 22 Table 2 11 The Export menu Move the cursor up Move the cursor down xX Export To delete chosen letter Execute the operation Export F RIGOL REF_ amp REF BF Ba REF_ REF fe E Cory bme m mile e ExPOR TE ALEC IPLEJFIG H I J E LIA HPOPPPALRPSPT Upp Wp aly ye 1ij2 3 4 5 6 7 o 9 _ GGE 2126 Figure 2 23 Expor
89. t as function the units are assigned to the function Cursor X Cursor X appears as vertical lines on the display to measure horizontal parameters Usually it indicates the time of trigger excursion When the source is set as FFT X means frequency 2 98 2008 RIGOL Technologies Inc User s Guide for DSI000E DS1000D Series RIGOL Track Mode Press Cursor Mode Track to display the track mode menu Figure 2 136 Table 2 87 The Track Mode Menu Mode Set Track mode in cursor Mode Track measurement Cursor A CH1 Cursor A CH Set Cursor A in conjunction with Cursor B None CH1 CH2 or turn off Cursor A Set Cursor B in conjunction with CH1 CH2 or turn off Cursor B None Turn the multifunctional knob 4 to move cursor A horizontally CurA Cursor A CurB Turn the multifunctional knob 4 Cursor B to move cursor B horizontally In Track mode the cross cursor is displayed on the waveform under measuring If you move the horizontal position of cursor the horizontal coordinate and vertical coordinate of current point as well as the increments of them will be displayed immediately Note horizontal coordinate represents time value vertical coordinate represents voltage value Cursors Mlode Ana g azAmsS C ursor A 1 aa er ae AY Foe Cursor A 4D Time 500 Bus Figure 2 137 Track Measure Mode of Cursor Measurement 2008 RIGOL Technologies Inc 2 99 User s Guide for DS1000E DS100
90. t interface 2 16 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL 3 Save Press RE Save and go to the following menu Figure 2 24 Table 2 12 The Save Menu Explorer Explorer Directory File Switch among Path Directory and File Mew File New File Set up new file in Path and File Folder Set up new folder in directory Delete Delete file Fold File Folder L eener Explorer BS REF_G REF u B Cor bme RIGOL E Cors 1 bmF B HewFile_ B HewFile_i E HewFile_ ig B HewFile_ii B HewFile_2 E HewFile_z B HewFile_4 B HewFile_5 E HewFile_ inal Mew File Delete File Figure 2 25 Save interface 2008 RIGOL Technologies Inc 2 17 User s Guide for DS1000E DS1000D Series RIGOL New File or New Folder Press REF Save New File or New Folder and go to the following menu Figure 2 26 Table 2 13 The New File menu Mew File Move the cursor up Move the cursor down To delete chosen letter Execute the operation M Bay F j l F RIGOL REF_ amp REF a REF_ REF EN Cors bmp File name ee key in HewF ile a Bblcid Feaphyiayapky lim a Ee ee le File Size 1 42Mb6 Switch Chinese English Switch Capital on off Figure 2 27 Chinese Input interface 2 18 2008 RIGOL Technologies Inc User s Guide for DSIOOOE DSi000D Series RIGOL 4 Import Press REF Import and go to the following menu Fi
91. tenuation factor changes the vertical scaling of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip To change or check the probe attenuation setting press the or button according to which channel in using Toggle the Probe soft button to match the attenuation factor of the probe This setting remains in effect until changed again Figure 2 8 shows an example for using a 1000 1 probe and its attenuation factor CH1 Coupling B Limit CFF Probe Probe attenuation Digital Filter Time 168 8us Vertical volt div Figure 2 8 Use the 1000 1 attenuation Table 2 3 Probe setting Probe attenuation factors _ s settings 2 6 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL 4 Digital Filter settings DS1000E DS1000D series offer 4 types of digital filters Low Pass Filter High Pass Filter Band Pass Filter and Band Reject Filter In order to achieve a good filter effect you can filter the special waveform frequency by setting up the range of filter bandwidth Press CH1 Digital filter display the digital filter menu Turn 4 knob to set high and low limit of frequency Filter Digital Filter lt i Turn off digital filter Filter Type Upper Limit 166 AkHz Time 3 6eem Figure 2 9 The waveform when turning off digital filter Filter Digital Filter Turn on digital filter Filter Type Upper Limit
92. the above inspection restart the oscilloscope 4 If the problem still remains please contact RIGOL for help 2 After the signal acquisition the waveform does not appear 1 Check if the probes are correctly connected with the signals 2 Check if the probes are firmly connected to the channels 3 Check if the probes are correctly connected with the object being tested 4 Check if the circuit can generate signal at the test point Connect the probe compensator output to an input channel to confirm whether the testing point has no signal or the input channel is in some troubles 5 Repeat the acquisition 3 The measurement result is 10 times higher or lower than the value expected Check if the probe attenuation is the same as the channel attenuation 4 If the oscilloscope does not get a stable waveform display 1 Check the Trigger Source and notice if it is set to the channel in use 2 Check the Trigger Type Use Edge for normal signals and use Video for VIDEO signals 3 Switch the coupling into HF Rejection or LF Rejection in order to filter the noise which disturbs trigger 4 Adjust the trigger Sensitivity and the hold off time 5 After pressing the button the oscilloscope does not display any waveform on screen Check whether the Trigger Mode is set to Normal or Single and see whether the trigger level is out of the signal range If yes set the trigger level in 2008 RIGOL Techn
93. to enter the self calibration interface as follows Press RUN STOP key to start the operation and press AUTO key to exit the system Disconnect everything fron all inputs CH1 Wertical System Press RUMSSTOP key to start Press ALTO ker t Exit Figure 2 116 Self calibration NOTE The oscilloscope must have been working or warm up at least 30 minutes before running self calibration to get best accuracy 2 84 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL System Information System information can help users to query the instrument model serial number software version installed module and so on Press Utility System Info to enter the system information interface as follows Press RUNSSTOP key to exit Figure 2 117 System information interface 2008 RIGOL Technologies Inc 2 85 User s Guide for DS1000E DS1000D Series RIGOL Special Mode DS1000E DS1000D series oscilloscopes provide a new function Key Lock which can meet the needs of industrial production In this mode except from F1 to F5 and MENU ON OFF other keys are locked You must input keys to unlock the keyboard the initial password is 111111 Further more you can reset new keys with six length by keys setting menu More details for operation steps please refer to the following introductions 1 Lock the keyboard Press Utility button press Special Mode menu to enter the Lock
94. to remove uncorrelated noise and improve measurement accuracy Reduces random or uncorrelated noise in the signal display The averaged waveform is a running average over a specified number of acquisitions from 2 to 256 Peak Detect Acquisition Peak Detect mode captures the maximum and minimum values of a signal Finds highest and lowest record points over many acquisitions 2 56 2008 RIGOL Technologies Inc User s Guide for DSI000E DS1000D Series RIGOL To Set up the Display System shows the menu button for the display system on the front panel MENU Display setup button Figure 2 73 Display setup button Press the button to pop up the menu for the settings of the display system Figure 2 74 Table 2 43 The Display menu Page 1 2 Display Display waveforms as vectors Dots Display waveforms as dots Clear all existing waveforms DEN ea from screen e Infinite The sample points remain OA pee displayed until turn the as persistence OFF a OFF Turn off the persistence function Intensity D Set up waveform intensity the lt percentage gt range is from 0 to 100 2008 RIGOL Technologies Inc 2 57 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 75 Table 2 44 The Display menu Page 2 2 Display grids and coordinates on the screen Grid Turn off the grids ightness Turn off the grids and coordinates eri Digg ty Brightness Set up grid brightness lt percentage gt
95. tomatic measurement of voltage parameters The oscilloscope provides automatic voltage measurements including Vpp Vmax Vmin Vavg Vamp Vrms Vtop Vbase Overshoot and Preshoot Figure below shows a pulse with some of the voltage measurement points Overshoot Vmax AY Vic Vpp Vba Vmin Preshoot Figure 2 131 Voltage parameters Vpp Peak to Peak Voltage Vmax The maximum amplitude The most positive peak voltage measured over the entire waveform Vmin The minimum amplitude The most negative peak voltage measured over the entire waveform Vamp Voltage between Vtop and Vbase of a waveform Vtop Voltage of the waveform s flat top useful for square pulse waveforms Vbase Voltage of the waveform s flat base useful for square pulse waveforms Overshoot Defined as Vmax Vtop Vamp useful for square and pulse waveforms Preshoot Defined as Vmin Vbase Vamp useful for square and pulse waveforms Average The arithmetic mean over the entire waveform Vrms The true Root Mean Square voltage over the entire waveform 2008 RIGOL Technologies Inc 2 93 User s Guide for DS1000E DS1000D Series RIGOL The automatic measurement of time parameters The oscilloscope provides time parameters auto measurements include Frequency Period Rise Time Fall Time Width Width Delay 1 2 Delay 12 Phas 1 gt 2 Phas 1 2 t Duty and Duty The following figure shows a pulse with some of the time measurement
96. ton to select the number of averages that best eliminates the noise from the displayed waveform It can be adjusted from 2 256 See Figure 3 3 Acquire Sampling Ecu Tirne MemDepth Long Mer Time i166 Bus Figure 3 3 Signal without noises 2 To reduce the noise it can also be achieved by reducing the intensity of the display NOTE It is normal that the refresh rate will slow down when the average acquisition mode is ON 2008 RIGOL Technologies Inc 3 5 User s Guide for DS1000E DS1000D Series RIGOL Example 5 Making Cursor Measurements There are 22 build in automatic measurements They can also be conducted using cursors to make time and voltage measurements of a waveform quickly Measure the Peak Frequency of the First Sinc Waveform To measure the ringing frequency at the rising edge of a signal do these steps 1 Press key to see the Cursor menu 2 Press Mode to set Manual mode 3 Press Type to select X 4 Turn 2 knob to place cursor A on the first peak of the wave 5 Turn 42 knob to place cursor B on the second peak of the wave Observe the delta in time and frequency displayed on the screen lran 1 66 rT kHz Time 5600 Bus Figure 3 4 Waveform display 3 6 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Measure the Amplitude of the First Waveform Peak of the Sinc Please follow these steps 1 2 3 4 5 Press key t
97. torage Storage Waveform Store or recall waveform or Setups instrument setups Internal Go to menu for internal memory Internal l operation see table 2 48 External Storage Go to menu for external memory External operation see table 2 49 isk Disk Mana Go to disk manage menu see Disk Wana table 2 50 2008 RIGOL Technologies Inc 2 59 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 78 Table 2 46 The Storage menu For Bitmap storage Create or delete bit map files Bi N Save the current oscilloscope Para Save Para Save settings in different format O OFF with the same file name ane Go to menu for external SOME External memory operation see table 2 49 l Go to disk manage menu Dskmaa _ etabi 230 Disk Wana Figure 2 79 Table 2 47 The Storage menu For CSV atorage Menu Settings Comments 5 Storage Displayed Save currently displayed Data Depth Daa Deni waveform data to CSV file AA E Maximum Save the whole waveform data Para Save in memory to CSV file OFF On Save the current oscilloscope Para Save Off settings in different format with External the same file name Hex Save data in hexadecimal format LA Data Disk hana Type Decimal Save data in decimal format Binary Save data in binary format operation see table 2 49 ees A see table 2 50 Disk Manage NOTE Applicable to DS1000D Series only 2 60 2008 RIGOL Technologies Inc
98. u Acquire Acquisition Sampling WemDepth SNK fx Figure 2 69 Signal that contains noise and without average sampling Acquire Acquisition Sampling Feal Time WemDepth Morral SNK fx Figure 2 70 Display signal after average sampling 2 54 2008 RIGOL Technologies Inc User s Guide for DS1000E DS1000D Series RIGOL Acquire Acquisition Sampling WemDepth Figure 2 71 Signal without Peak Detect Acquisition Acquire Acquisition sampling WMemDepth Figure 2 72 Signal with Peak Detect Acquisition Stop Acquisition When the scope is acquiring waveforms the waveforms is in a live status when acquisition is stopped frozen waveform will be displayed the position and scale can still be adjusted by vertical control and horizontal control 2008 RIGOL Technologies Inc 2 55 User s Guide for DS1000E DS1000D Series RIGOL Key Points Real time Sampling The oscilloscope has Real time sampling rate up to 1GSa s At the time base 50ns or faster the oscilloscopes use the sine x x interpolation to expand the horizontal time base Equivalent Sampling Known as Repetitive sampling to get up to 40ps of horizontal resolution equivalent 25Gsa s This mode is good for observing repetitive signals and it is not recommended for single shot or pulse Normal Acquisition Oscilloscope acquires signal by equal time interval Average Acquisition Apply averaging to your signal
99. um 3 results could be displayed at the same time When there is no room the next new measurement result will make the previous results moving left out of screen 2008 RIGOL Technologies Inc 2 91 User s Guide for DS1000E DS1000D Series RIGOL Using Automatic Measurement 1 Select the signal channel for measuring CH1 or CH2 according to the signal of interest Press soft buttons as follows Measure Source CH1 or CH2 2 To see all measurement values set the Display All to ON 18 measurement parameters will be displayed on the screen except for Delayl 2 and Delay1 2 3 Select parameters page for measuring select voltage or time parameters pages by pressing soft button as follows Measure Voltage or time Vmax 4 To get the measured value on the screen select the parameters of interest by pressing the soft button on the right of the menu and read the data on the bottom of the screen If the data is displayed as it means the parameter cannot be measured in current condition 5 Clear the measure values press Clear and all of the auto measure values would disappear from the screen except for Display all parameters Measure Source CH1 Clear Clear measured data from screen Display All mi Turn on off the mass measurement Figure 2 130 Measurement display 2 92 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL The au
100. ur on Output OFF Continue test when output occur Mask Setting a Go to mask setting menu NOTE The beeper should be opened 2008 RIGOL Technologies Inc 2 71 User s Guide for DS1000E DS1000D Series RIGOL Mask Setting Users can create save load and import export the Pass Fail mask by this function Press Utility Pass Fail Mask Setting to go to the following menu Figure 2 96 Table 2 60 The Mask setting menu Page 1 2 Em Menu Settings Comments __ ESE Mais 1 Set horizontal clearance to the B 20div X Mask waveform Y Wask lt x div gt 4 0 04div 4 00div Be 2bidiv Y Mask 4 Set vertical clearance to the waveform lt y div gt 0 04div 4 00div Create Create a test mask according to the S Mask above clearance l l Internal Set the memory location of the mask lsz Location External files Figure 2 97 Table 2 61 The Mask setting menu Page 2 2 When the save as Internal memory Store created test mask into internal Recall mask setting file from internal memory Go to import export menu same as REF import export menu See table L 2 72 2008 RIGOL Technologies Inc ser S Guide for J erles RIGOL Figure 2 98 Table 2 62 The Mask setting menu Page 2 2 ieee teen the save as external a Sae Setings to save menu ac eee as REF save menu See table 2 carmen 1 Load Go to Go to load menu see table 2 63 menu see table 2 63 Go to import menu Same
101. urn the multifunctional knob 4 to move cursor A horizontally Turn the multifunctional knob 4 to move Cursor A cursor A vertically Turn the multifunctional knob 4 to move cursor B horizontally Turn the multifunctional knob 4 to move cursor B vertically Cursor B NOTE Cursor could be moved only when the curse function menu is displayed 5 To get measurement values Position of Cursor A time cursor centered on the midpoint of screen Voltage cursor centered on channel ground level Position of Cursor B Same as above Horizontal space between cursor A and B aX Time between cursors 1 aX units in Hz KHz MHz GHz 2008 RIGOL Technologies Inc 2 97 User s Guide for DS1000E DS1000D Series RIGOL Vertical space between cursor A and B aY Voltage between cursors If you choose the source as LA DS1000D series the measurement value is as follows Position of Cursor A time cursor centered on the midpoint of screen Position of Cursor B time cursor centered on the midpoint of screen Cursor A value Hex Cursor A value Binary Cursor B value Hex Cursor B value Binary NOTE The values will be automatically displayed on the right upper corner of screen when the cursor function menu is hidden or displaying other menus Key Points Cursor Y Cursors Y appear as horizontal lines on the display to measure vertical parameters Usually it is used to measure the Volts When the source is se
102. ursor Measurements REF and MATH Operations Delayed Scan Mode Vector Display Mode Horizontal knob Trigger Controls 2 32 2008 RIGOL Technologies Inc User s Guide for DSI000E DSIO00D Series RIGOL The Horizontal Knobs POSITION The horizontal knob adjusts the horizontal position of all channels including Math waveforms The resolution of this control varies with the time base Pressing this button clears trigger offset and moves the trigger point to the horizontal center of the screen SCALE Use to select the horizontal time div scale factor for the main or the time base of Delayed Scan When Delayed Scan is enabled it changes the width of the window zone by changing the Delayed Scan time base 2008 RIGOL Technologies Inc 2 33 User s Guide for DS1000E DS1000D Series RIGOL To Set up the Trigger System The trigger determines when the oscilloscope starts to acquire data and display a waveform When a trigger is set up properly it can convert unstable displays or blank screens into meaningful waveforms When the oscilloscope starts to acquire a waveform it collects enough data so that it can draw the waveform to the left of the trigger point The oscilloscope continues to acquire data while waiting for the trigger condition to occur After it detects a trigger the oscilloscope continues to acquire enough data so that it can draw the waveform to the right of the trigger point The trigger control are
103. valent sample rate and the powerful trigger and analyzer functions enable users view or capture the waveform more detailed and faster Main Features Dual analog channels 1GSa s maximum real time sample rate and 25GSa s maximum equivalent sample rate Bandwidth of per channel 100MHz DS1102E DS1102D 50MHz DS1052E DS1052D Optional 16 digital channels DS1000D series each channel can be turned on or off independently or in a 8 bit group 5 6 inch and 64 k TFT LCD Abundant trigger types Edge Pulse Width Video Slope Alternate Pattern and Duration Trigger Only for DS1000D series Unique adjustable trigger sensitivity enables to meet different demands Enable to measure 22 types of wave parameters and track measurements via cursor automatically Unique waveform record and replay function Fine delayed scan function Built in FFT function hold practical digital filters 2008 RIGOL Technologies Inc y User s Guide for DS1000E DS1000D Series RIGOL Provide 4 types of digital filters LPF HPF BPF BRF Pass Fail detection function enables to output testing results Math operations available to multiple waves Powerful PC application software UltraScope Standard configuration interface USB Device USB Host RS 232 and support U disk storage and PictBridge print standards The new function Special Mode can meet the needs of industrial production Support for remote command control Bu
104. wing topics To Inspect the Instrument The Panels and User Interface Appearance and Dimensions To Perform a Functional Check To Compensate Probes Digital Leads Only for DS1000D Series To Display a Signal Automatically To Understand the Vertical System To Understand the Horizontal System To Understand the Trigger System 2008 RIGOL Technologies Inc 1 1 User s Guide for DS1000E DS1000D Series RIGOL To Inspect the Instrument When you get a new DS1000E DS1000D series oscilloscope please inspect the instrument according to the following steps 1 1 2 Inspect the shipping container for damage Keep a damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically Inspect the instrument In case there is any mechanical damage or defect or the instrument does not operate properly or fails performance tests please notify the RIGOL Sales Representative If the shipping container is damaged or the cushioning materials show signs of stress please notify the carrier as well as the RIGOL sales office Keep the shipping materials for the carrier s inspection RIGOL offices will arrange for repair or replacement at RIGOL s option without waiting for claim settlement Check the accessories Accessories supplied with the instrument are listed in Appendix A Accessories
105. y hints in this guide It is recommended to perform all of the following exercises to get the most of the powerful measurement capabilities of the oscilloscope This chapter covers the following topics To Set up the Vertical System CH1 CH2 MATH REF LA for DS1000D series OFF VerticalPOSITION VerticalSCALE To Set up the Horizontal System MENU HorizontalPOSITION Horizontal SCALE To Set up the Trigger System LEVEL MENU 50 FORCE To Set up the Sampling System Acquire To Set up the Display System Display To Store and Recall Storage To Set up the Utility System Utility To Measure Automatically Measure To Measure with Cursors Cursor To Use Run Control Buttons AUTO IRUN STOP Please refer to this chapter more carefully to get more information of the measurement and operation methods of DS1000E and DS1000D 2008 RIGOL Technologies Inc 2 1 User s Guide for DS1000E DS1000D Series RIGOL To Set up the Vertical System Settings of the Channels This series oscilloscopes offer dual channels Each channel has an operation menu and it will pop up after pressing CH1 or button The settings of all items in the menu is shown in the table below Figure 2 1 Table 2 1 The Channel menu Page 1 2 Coupling Blocks the DC component of the input Signal HALAN Coupling Passes both AC and DC OFF components of the input signal Probe Disconnects the input signal Limits
106. y location Specify memory location to save recall the waves and setups Load Recall saved waveforms setups and default settings Save Save waveforms and setups 2008 RIGOL Technologies Inc 2 65 User s Guide for DS1000E DS1000D Series RIGOL NOTE 1 Select Save stores not only the waveforms but also the current settings 2 To ensure the setups being saved properly only after the settings are changed for more than 5 seconds can the user turn off the instrument The oscilloscope can store 10 settings permanently and can restore at anytime 2 66 2008 RIGOL Technologies Inc User s Guide for DSI000E DS1000D Series RIGOL To Set up the Utility System shows the menu button for the Utility on the front panel MENU Utility setup button Figure 2 88 Utility setup button Press the Utility button to pop up the menu for the settings of the Utility system Figure 2 89 Table 2 54 The Utility Menu Page 1 3 lO Setting I O setting ra Setup I O configuration Sound F O ge ya Switch beeper sound on off Counter aa OFF Turn off Frequency Counter OFF Counter ON Turn on Frequency Counter Language a fey ASA SC x 3 nce Select languages Language sie More languages may be added in later firmware versions Francois 2008 RIGOL Technologies Inc 2 67 User s Guide for DS1000E DS1000D Series RIGOL Figure 2 90 Table 2 55 The Utility menu Page 2 3 Pass Fail Setup Pass Fail te
Download Pdf Manuals
Related Search
Related Contents
Manual de instruções do tacômetro digital TC-5030. Samyang 135MM T2.2 VDSLR Nikon F Fiche technique & mode d`emploi Deutsch/Französisch Professor PLUS Professeur d`allemand Advantech MIC-3645 User Manual Canon LV-7555 (Turbo Bright) Copyright © All rights reserved.
Failed to retrieve file