1. This section of the manual contains information on diagnosing and repairing converter performance problems This Section provides troubleshooting procedures that address problems to the board level For component level troubleshooting consult the schematics for the appropriate board in the Appendix 6 0 1 General Troubleshooting Hints 1 Think of the M501Y as two sections A Section 1 Pneumatics Over 50 of all instrument problems are traced to leaks in the M200EU pump assembly pole mounted converter assembly sample filter or zero span gas delivery system calibrator Suspect a leak first B Section 2 Electronics converter temp control board or DC power supply for the converter temperature readout 2 Incorrect span gas concentration this could come either from the calibrator or entering the expected span gas concentration in the M501Y incorrectly 3 If the instrument does not respond to span gas check troubleshooting section in the M200E EU manual The above should get you started in diagnosing and repairing the most common faults If these reasons have been eliminated the next thing to do is a Factory Calibration covered in M200E EU manual If difficulties persist contact our service department The 800 telephone number is on the cover page of this manual 6 1 Operation Verification M501Y Diagnostic Techniques 6 1 1 Pneumatic System The pneumatic system is diagramed in Figure 5 3 6 1 2 Leak Check R
2. consequences 05386 Rev A 2 Model 200EU NOy Option Manual Addendum Table of Contents SAFETY MESSAGES Table of Contents List of Figures List of Tables 1 0 GETTING STARTED 1 1 Unpacking 1 2 Electrical and Pneumatic Connections 1 3 Initial Operation 2 0 SPECIFICATIONS WARRANTY 2 1 Specifications 2 2 Warranty 3 0 THE M501Y NOY CONVERTER 3 1 Principle of Operation 4 0 CALIBRATION AND ZERO SPAN CHECKS 4 1 Calibration or Cal Check Procedure 5 0 MAINTENANCE 5 1 Maintenance Schedule 5 2 Replacing the Sample Particulate Filter 5 3 Check Bypass Flow Rates 5 4 Replacing the Converter 5 5 Pneumatic Line Inspection 5 6 Leak Check Procedure 6 0 TROUBLESHOOTING ADJUSTMENTS 6 0 1 General Troubleshooting Hints 05386 Rev A Model 200EU NOy Option Manual Addendum 6 1 Operation Verification M501Y Diagnostic Techniques 6 1 1 Pneumatic System 6 1 2 Leak Check 6 1 3 Bypass Pump Diagnostic Procedures 6 1 4 Electrical Fault Isolation 6 2 Setting the Converter Temperature 6 1 4 Temperature Controller Setup 7 0 M501Y SPARE PARTS LIST 05386 Rev A Model 200EU NOy Option Manual Addendum List of Figures FIGURE 1 1 REAR PANEL PNEUMATIC CONNECTIONS se ee ee ee se ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 8 FIGURE 5 1 REPLACING THE PARTICULATE FILTER se ee ee se ee ee se ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 21 FIGURE 5 2 CONV
3. it must be reprogrammed to suit the thermal characteristics of the instrument The controller includes an Autotuning feature which can be used to properly set the PID factors To start Autotune procedure e Press Sel until A7 AT is displayed e Press Data until value 1 enable Autotune e While Autotuning the decimal point will blink e NOTE During the procedure it is normal for the temp to overshoot up to 100 C e When the decimal point LED turns off Autotune is complete No further adjustment should be necessary Table 6 1 Temperature Controller Programming Guide PRESS DISPLAY ACTION SEL X2 P PROP BAND UP DOWN SET TO 5 0 ENTER SPIE sii N MERRIE AR OO a et SEL i INTEGRAL SET TO 50 NE om OE UDS SEL d DERIVATIVE TIME SET TO 5 ABER N cre men OOO AA OE LE EER OT ON SEL 7C CYCLE TIME SET TO 15 FOR FAST SYSTEM DAL e a DER IE a ne aa ne tn et SEL_______HYSHYSTERESIS EE SEL Sm RAMP RATE SE SEL A7 AUTOTUNE SET TO1 ON Ho Bs AU EE OON RE OTE SEL LOC LOCK 0 OPEN 1 LOCKED TE EE t a 2 SVONLYOPEN ___________ SEL END RETURNS TO SV SET TO 315 315 C SET VALUE NOTE DO NOT SET THE TEMPERATURE SV HIGHER THAN 320 C 05386 Rev A 30 Model 200EU NOy Option Manual Addendum SECONDARY MENU The fixed characteristics of the system PRESS DISPLAY ACTION SEL_______pdF DIGITAL FILTER SEIT SEL PSL LOWER LIMIT ____ DIE iii SEL P SU UPPER LIMIT _____ SET EO EO na eee SEL FAB ALARM i
4. NONE ww na DINAN SEL_______PrAnHYTERESIS ______ SO aa SEL PoP PROG DECIMAL SETTOUO nn SEL P 48 COLD JUNCTION SET TO 2 TYPE J ee SEL GEAME TO SET POINT LEAVE ATO EE ee ee SEL _______PVOF PROCESS OFFSET LEAVE ATO EE EE Ee SEL SVOF SET POINT OFFS LEAVE ATO oe SEL san ER La wei EE EE EE EN ERE HOLDER TEEN SEL SCROLLTOAT 3 SET TO 1 AUTOTUNE ___________ _ SV SET VALUE SHOULD BE 315 C 05386 Rev A 31 Model 200EU NOy Option Manual Addendum ac E o co gt zo D 05 SE NE m Conad a gt rm SE N xs De lt N gt ojl Ee a gt z a lt C EA a GE x Rei O lt o a 28 in 2 amp Bei i lt SS a ao Figure 6 1 Electrical Diagram 05386 Rev A 32 Model 200EU NOy Option Manual Addendum 7 0 M501Y Spare Parts List Note Use of replacement parts other than those supplied by API may result in non compliance with European Standard EN 61010 1 PART NO 00094 4 00094 10 01630 04 02315 01872 02293A 01357 02317 02318 00969 00969 1 02230 PS004 01086 00728 01198 01236 01314 FLOO1 FL003 01068 HW020 HW036 HW037 OR058 01631 OR034 PU028 PU030 05386 Rev A DESCRIPTION ORIFICE 4 MIL 80 CC 03 FLOW Rx CELL ORIFICE 13 MIL 1100 CC Rx CELL SAMPLE FILTER ASSY SENSOR ASSY M501Y EXCHANGE WELDED CONVERTER W 03 KILLER OPERATORS MANUAL FOR M501Y THERMOCOUPLE ASSY M501Y EXPENDABLES KIT M501Y LEVEL 1 SPARES KIT FILTER
5. on Ul is 1 100 of the temperature In other words 3 15 VDC represents 315 C 6 2 Setting the Converter Temperature A front panel mounted programmable controller maintains the temperature of the remote molybdenum converter By pressing the PV SV button in the lower left corner of the controller you can see the Present Value actual temperature or the Set Value temperature set point The temperature and PID control parameters have been set at the factory and no further adjustment should be necessary Should you need to adjust the temperature follow these steps Select SV with the PV SV button Press the up arrow under the digit you want to change the digit will flash Press the up arrow under the digit or the down arrow at the left to scroll the digit to the desired value 5 Repeat for the other two digits 6 Press the ENTER button Select PV with the PV SV button to observe the actual temperature EPM 05386 Rev A 29 Model 200EU NOy Option Manual Addendum 6 1 4 Temperature Controller Setup The temperature controller has been programmed at the factory and should not need to be altered under normal usage Indications that would require re programming are e Failure replacement of controller e Replacement of thermocouple element e Controller s loss of PID parameters leading to instability of the temp setpoint In the event that the control parameters are changed or in the event that a new controller is installed
6. CONDITIONS All units or components returned to API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center After the repair the equipment will be returned freight prepaid 05386 Rev A 11 Model 200EU NOy Option Manual Addendum 3 0 The M501Y NOy Converter 3 1 Principle of Operation The API Model 200EU with NOy Option is designed to measure the concentration of NO NO2 and other compounds that are too unstable to be measured when taken in through the normal ambient air sample inlet system Please refer to the M200EU manual supplied with this system for a general discussion of the operation of a NOx analyzer The suite of compounds known collectively as NOy is composed of roughly 30 compounds There is some disagreement over whether certain compounds should be on the list The NOy measurement is generally done in conjunction with a standard NOx measurement with the difference between the two being the concentration of the unstable compounds sometimes referred to as NOz The system is composed of 3 modules e The M2OOEU without a moly converter e A Bypass pump chassis containing e Bypass pump e Flow control e Sample filtration e Moly temperature control e Pneumatic provisions for calibration e An externally mounted molybdenum converter The M200EU w NOy Option system allows the converter to be mounted at the sample inlet point This configuration provides minimal time delay
7. ERTER ASSEMBLY ee se ee ee se se ee ee se ee ee ee ee see ee ee ee ee ee ee ee ee ee ee ee ee ede ee EE ee ee Ee Re ee ee ee ee Re ee ee RA 23 FIGURE 5 3 PNEUMATICS DIAGRAM STANDARD AND Z S ORIION se se ee ee ee ee ee ee ee ee ee ee ee ee 25 FIGURE 5 4 PNEUMATIC DIAGRAM OPT 86 OPTION se ee ee ee ee ee ee ee ees ee ee ee ee ee ee ee ee ee ee ee ee Re ee ee ena 26 FIGURE 6 1 ELECTRICAL DIAGRAM iss Ee Es ss See aiar 32 05386 Rev A 5 Model 200EU NOy Option Manual Addendum List of Tables TABLE 2 1 FINAL TEST AND CALIBRATION VALUES TABLE 4 1 MANUAL ZERO CALIBRATION PROCEDURE ZERO GAS THRU SAMPLE PORT TABLE 4 2 ENTER EXPECTED SPAN GAS CONCENTRATIONS PROCEDURE TABLE 4 3 SPAN CALIBRATION PROCEDURE TABLE 5 1 PREVENTATIVE MAINTENANCE CALENDAR TABLE 6 1 TEMPERATURE CONTROLLER PROGRAMMING GUIDE 05386 Rev A Model 200EU NOy Option Manual Addendum 1 0 Getting Started The M501Y has been designed to operate with the M200EU low level NOx analyzer 1 1 Unpacking 1 The M200EU with NOy Option comes in three boxes e The M200EU analyzer e The Bypass Pump Chassis e The External Converter and umbilical cable assembly 2 Verify that there is no apparent shipping damage If damage has occurred please advise shipper first then API CAUTION AN To avoid personal injury always use two persons to lift and 7 carry the Model 501 Pump Pack Chassis i 3 Before operation of the M200E
8. MANUAL ADDENDUM M200EU NOy CONVERTER OPTION Teledyne Advanced Pollution Instrumentation 9480 Carroll Park Dr San Diego CA 92121 5201 TOLL FREE 800 324 5190 FAX 858 657 9818 TEL 858 657 9800 E MAIL api sales Oteledyne com WEB SITE www teledyne api com 05386 REV A Copyright 2006 API Inc 10 October 2006 Model 200EU NOy Option Manual Addendum SAFETY MESSAGES Your safety and the safety of others is very important We have provided many important safety messages in this manual Please read these messages carefully A safety message alerts you to potential hazards that could hurt you or others Each safety message is associated with a safety alert symbol These symbols are found in the manual and inside the instrument The definition of these symbols is described below i IN General Warning Caution Refer to the instructions for details on the specific danger 9 AN N Hot Surface Warning Electrical Shock Hazard IN Technician Symbol All operations marked with this symbol are to be performed by L ad qualified maintenance personnel only Electrical Ground This symbol inside the instrument marks the central safety grounding point for the instrument CAUTION gt The analyzer should only be used for the purpose and in the manner described in this manual If you use the analyzer in a manner other than that for which it was intended unpredictable behavior could ensue with possibly hazardous
9. NOx bypass and 1 LPM for the M200EU Calibration gas flow must be in sufficient excess 4 LPM to prevent any ambient air from entering the sample port on the pole The maximum 50 ft length of tubing between the Bypass Pump Module and the remote moly converter has been proven to generate too much backpressure for some brands of calibrators Experience to date is that while NO calibration is correct GPT can be incorrect due to errors in dilution of the ozone We strongly recommend that SPAN CALIBRATION be done with NO span gas SPAN CHECKS can be done with either NO only NO2 only or a mixture of NO and NO2 GPT Zero air used for all calibration procedures including GPT should have lt 1 ppt NO and NO2 less than 1 ppt of major interferents such as SO2 NH3 hydrocarbons and a dew point of 5 C or less The calibration gasses should be from a reliable supplier since the quality of the tank concentration values ultimately determines the accuracy of the analyzer NOTE The M200EU with the NOy Option does NOT have equivalency approval and may not be used for EPA monitoring 05386 Rev A 13 Model 200EU NOy Option Manual Addendum 4 1 Calibration or Cal Check Procedure The calibration of the instrument can be checked or adjusted using gas introduced through the calibration port on the Bypass Pump Module The calibration gas is routed to a TEE near the sample inlet port on the remote converter assembly Since the zero gas concentra
10. TFE 47 mm Quantity 100 FILTER TFE 47 mm Quantity 25 DC POWER SUPPLY BOARD 5 15 VDC SWITCHING POWER SUPPLY STATUS TEMP BOARD DVM DISPLAY ASSY THERMOCOUPLE FAN BYPASS PUMP CHASSIS FAN PMT COOLER SINTERED FILTER FILTER DFU HEATER CONVERTER SPRING FLOW CONTROL TFE THREAD TAPE 48 FT TIE CABLE O RING SAMPLE FILTER RETAINING RING SAMPLE FILTER O RING FITTING SAMPLE PUMP M501Y SAMPLE PUMP REBUILD KIT 33
11. U remove the shipping screws as shown in the M200E EU Manual 4 Please check the voltage and frequency label on the rear panel of the instrument for compatibility with the local power before plugging in the M501Y 1 2 Electrical and Pneumatic Connections 1 Mount the converter on a suitable mast outside of the enclosure 2 Route the umbilical cable assembly into the shelter 3 Locate the M200EU and M501Y Bypass Pump Chassis in close proximity preferably mounting one over the other in a 19 rack 4 Connect the electrical pneumatic cable fittings to the rear of the Bypass Pump Chassis Use the tags on each tube and Figure 1 1 to match up the correct tube with the rear panel bulkhead fitting Please note that the dashed lines show optional Zero Span Valve configuration connections Connect the 7 pin power and signal cable to the connector on the rear of the Bypass Pump Chassis 6 Connect the power cord to the correct line voltage SI WARNING Lethal voltages are present inside case Do not operate with cover off during normal operation AN Before operation check for correct input voltage and freguency Do not operate without proper chassis grounding Do not defeat the ground wire on power plug Turn off power before disconnecting electrical subassemblies 05386 Rev A 7 Model 200EU NOy Option Manual Addendum dansen NO OUT CONVERTER CAL OUT HE Semple in Figure 1 1 Rear Panel Pneumatic Connectio
12. between the sample inlet port and the remotely mounted molybdenum converter Minimizing the transit time between the sample inlet and converter enables the conversion of labile components of NOy The equation for the conversion is 315 C 3 NOy Mo gt 3 NO MoO 05386 Rev A 12 Model 200EU NOy Option Manual Addendum 4 0 CALIBRATION AND ZERO SPAN CHECKS Unlike most NOx analyzers the M200EU with the NOy option does not have a sample inlet port on the rear panel of the 200EU The sample port is located on the External Converter therefore zero span calibration is different than a normal NOx instrument Follow the steps in the M200E EU manual to a Enter the expected NO and NOx span gas concentrations in the M200EU b Do the zero calibration procedure c Do the span calibration procedure The analyzer always routes the zero span gas through the external converter In standard configuration the calibration gas needs to be input at the Cal in port of the M501 Pump pack When Zero Span valve option is installed zero and span gases are to be provided at the 200EU zero and span ports In all cases the calibration gas delivered must be under a small amount of pressure 2 5 psig to overcome the resistance of the hoses Also there must be no venting of the zero or span gas to atmosphere Calibration gas must be provided for both the bypass system and the 200EU analyzer Flow requirements are 1 1 LPM for NO bypass 1 1 LPM for
13. d at the rear of the instrument 4 Turn on the sample pump and set the TEST function to RCEL which measures the reaction cell pressure Close the shutoff valve and monitor the cell pressure The pressure should not drop more than 1 Hg 5psi in 5 minutes If there is a leak it is not possible by this method to tell where it is located You can locate the leak by using a pressure leak checker described below If you have a leak checker 1 Turn off instrument power and pump power 2 Disconnect pump at rear panel Cap the sample inlet port ozone generator air inlet and zero air inlet if Z S valve option present and connect the leak checker to the exhaust port CAUTION PRESSURE MUST BE LESS THAN 15 PSI 3 Pressurize system and check for leaks by watching overall pressure The pressure should not drop more than 1 Hg 5psi in 5 minutes If the instrument fails the pressure test each fitting needs to be leak checked to find the location Be careful that the system is always pressurized so as not to draw soap solution into the plumbing system Make sure you dry off any accumulated bubble solution The Sensor module can be leak checked as a unit using a 1 8 tubing fitting on top of the assembly The same rules as above apply 05386 Rev A 27 Model 200EU NOy Option Manual Addendum 6 0 Troubleshooting Adjustments NOTE THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY
14. efer to Section 5 6 for the leak check procedure 05386 Rev A 28 Model 200EU NOy Option Manual Addendum 6 1 3 Bypass Pump Diagnostic Procedures The bypass pump is capable of maintaining a total bypass flowrate of 4 Ipm If flow checks indicate a fall off in bypass flow the pump may need servicing Check the pump inlet fittings and analyzer for leaks first If other causes have been eliminated rebuild the pump 6 1 4 Electrical Fault Isolation 115 VAC is supplied to 1 The bypass pump 2 The 5 volt switcher supply 3 The input power to the converter temperature controller Use Figure 6 1 to check the 115 VAC supply voltage to each of these components The switcher supply produces 5 VDC to power the front panel display Use Figure 6 1 to check the 5 VDC supply voltage to the display The converter temperature is controlled by the 002560100 assembly in the NEMA chassis The 00257 schematic is located at the back of this manual The converter temperature controller produces 15 VDC to power the controller logic The OPTO ISOLATOR component switches the 115VAC to the heater Check the heater element for continuity The DS1 LED is on when power is applied to the heater Check for appropriate voltages The temperature is sensed by a thermocouple attached to TB1 Typical thermocouple voltage at the 315 C setpoint is about 16 mV Ul provides cold junction compensation and linearizes the output of the thermocouple The output of pin 8
15. ns 05386 Rev A 8 Model 200EU NOy Option Manual Addendum 1 3 Initial Operation 1 After confirming proper supply voltage turn on the instrument power If you are unfamiliar with the M200EU with NOy Option we recommend that you read the overview in Section 3 before proceeding 2 The power indicator light and display should immediately light in addition the bypass pump should start up The M501Y requires about 30 minutes for the converter to come up to temperature 3 While the converter is coming up to temperature the M200EU should be powered up also Follow the instructions in the M200E EU Manual to confirm proper operation of that instrument Note It will take 30 minutes for the ozone generator to start up in the M200EU During that time the instrument will not respond to span gas 4 After 30 min the display on the Bypass Pump Chassis should read 315 C indicating that the converter is up to temperature 5 Proceed to Section 4 of this manual to do a zero span check Table 2 1 Final Test and Calibration Values Test Values Observed Value Nominal Range BYPASS FLOW NO CHANNEL eemm 1100 200 ve EE Converter Serial Date Technician 05386 Rev A 9 Model 200EU NOy Option Manual Addendum 2 0 Specifications Warranty 2 1 Specifications Converter Remotely mounted molybdenum converter with temperature controller in a NEMA 4 enclosure Converter Temperature 315 C 7 C with read o
16. pass Flow Rates The external sample pump is capable of maintaining the cell pressure at less than 5 0 Hg A Ifa higher pressure is noted the pump may need servicing Check the pump and pneumatic system for leaks or rebuild pump The ozone scrubber is integrated into the converter case inside the instrument See Figure 1 1 for component locations 5 4 Replacing the Converter The heater thermocouple and converter assembly is designed to be replaced as a single unit Check Figure 5 2 for component location 1 Turn off the power to the M501Y at the chassis front panel 2 Allow the converter to cool CAUTION The converter operates at 315 C Severe burns can result if not enough time is allowed for the assembly to cool Do not handle assembly until it is at room temperature a Disconnect the gas fittings and power cable from the can 4 Remove the entire assembly from the NEMA chassis A Remove the pneumatic fittings B Remove the converter assembly from the chassis by loosening the 4 captive screws that secures the assembly to the chassis 5 Remove bottom bracket and re attach it to the replacement converter assembly 6 Install the assembly back into the analyzer T Re attach the electrical and pneumatic fittings Leak check the assembly when completed 8 Turn the power back on The insulation can emit a burnt odor for the first 24 hours this is normal Allow the converter to burn in for 24 hours then re calibrate the in
17. red to the extent and according to the current terms and conditions of the respective equipment manufacturers warranty GENERAL API warrants each Product manufactured by API to be free from defects in material and workmanship under normal use and service for a period of one year from the date of delivery All replacement parts and repairs are warranted for 90 days after the purchase If a Product fails to conform to its specifications within the warranty period API shall correct such defect by in API s discretion repairing or replacing such defective Product or refunding the purchase price of such Product The warranties set forth in this section shall be of no force or effect with respect to any Product i that has been altered or subjected to misuse negligence or accident or ii that has been used in any manner other than in accordance with the instruction provided by API or iii not properly maintained THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY WHETHER EXPRESSED OR IMPLIED THE REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONTAINED HEREIN API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF API S PERFORMANCE HEREUNDER WHETHER FOR BREACH OF WARRANTY OR OTHERWISE TERMS AND
18. sired compensation for converter efficiency CE can be included in the NOx concentration calculation Prior to calibration the converter efficiency should either be set to 1 0 or determined per the procedure in the M200E EU manual 05386 Rev A 15 Model 200EU NOy Option Manual Addendum Table 4 3 Span Calibration Procedure Step Action Comment Number l Press CAL The M200EU enters the calibrate mode and the adaptive filter is activated This allows the instrument to respond apidly to concentration changes regardless of their magnitude ait for reading to stabilize at span value 3 Press SPAN If you change your mind after pressing SPAN you can still press EXIT here without spanning the instrument d Press ENTR Pressing ENTR actually changes the calculation equations and causes the instrument to read the NO and INOx span concentrations 5 Press EXIT M200EU returns to sampling Immediately after calibration data is not added to the DAS averages 05386 Rev A 16 Model 200EU NOy Option Manual Addendum 5 0 Maintenance 5 1 Maintenance Schedule See Table 5 1 below 05386 Rev A 17 Model 200EU NOy Option Manual Addendum Table 5 1 Preventative Maintenance Calendar Zero Span Calibration Annually or after maintenance or repairs Zero Span Checks Daily Particulate Filter Weekly as needed Ozone Flow Check every year and replace o ring and orifice in reaction cell OzoneDryer Filter DFU Replace every
19. strument 05386 Rev A 22 Model 200EU NOy Option Manual Addendum Figure 5 2 Converter Assembly 05386 Rev A 23 Model 200EU NOy Option Manual Addendum 5 5 Pneumatic Line Inspection Particulate matter in the pneumatic lines will affect both flow rate and response time It is important that the pneumatic system be periodically inspected and thoroughly cleaned if necessary Clean by disassembling and passing methanol through three times Dry with nitrogen or clean zero air Also inspect all pneumatic lines for cracks and abrasion on a regular basis Replace as necessary 05386 Rev A 24 Model 200EU NOy Option Manual Addendum ET N da iN A i rl ia ST N i kl I I 13 31 Ly i Figure 5 3 Pneumatics diagram standard and Z S option 05386 Rev A 25 Model 200EU NOy Option Manual Addendum Ir Figure 5 4 Pneumatic diagram OPT 86 option 5 6 Leak Check Procedure 05386 Rev A 26 Model 200EU NOy Option Manual Addendum If a leak checker is not available it is possible to leak check the instrument using the M501Y s pump plus a shut off valve 1 Turn off instrument power and pump power 2 Cap the sample inlet port ozone generator air inlet and zero span inlets if Z S valve option present 3 Insert a shut off valve between the sample pump and the vacuum manifol
20. tion is defined as 0 ppb it is not necessary to enter the expected zero value Table 4 1 details the zero calibration procedure Table 4 1 Manual Zero Calibration Procedure Zero Gas Thru Sample Port Step Action Comment Number 1 Press CAL The M200EU enters the calibrate mode and the adaptive filter is activated This allows the instrument to respond apidly to concentration changes regardless of their magnitude ait for reading to stabilize at zero value 3 Press ZERO If you change your mind after pressing ZERO you can still press EXIT here without zeroing the instrument 4 Press ENTR Pressing ENTR actually changes the calculation equations Press EXIT M200EU returns to sampling Immediately after calibration data is not added to the DAS averages 05386 Rev A 14 Model 200EU NOy Option Manual Addendum Enter the expected NOx and NO span gas concentrations Table 4 2 Enter Expected Span Gas Concentrations Procedure Step Action Comment Number Press This key sequence causes the M200EU to prompt for the CAL CONC NOX expected NOx concentration Enter the NOx span concentration value by pressing the key under each digit until the expected value is set ENTR ENTR stores the expected NOx span value Now enter the expected NO span concentration as in step Press ENTR Pressing ENTR stores the NO span value and returns the prompt to the CONC menu i Press Returns instrument to SAMPLE mode EXIT If de
21. ut on front panel of bypass pump chassis Tube Electrical Cable Up to 50 feet 15 m maximum length Assembly Power Bypass Pump 120V 60 Hz 230V 50 Hz 360 watts Assembly Power Converter Assembl 120V 50 60 Hz 60 watts Converter Efficiency gt 96 for NO2 Weight Bypass Pump 35 Ibs 16 kg Assembly Weight Remote 30 Ibs 11 kg Converter Bypass Pump Chassis 9 x17 x23 6 23cm x 43cm x 61cm Dimensions HxWxD Remote Converter Assy 12 x19 x7 31cm x 48cm x 18cm Dimensions HxWxD Environmental Installation Category Over voltage Category II Pollution Degree 2 Bypass Flowrate 1100 cc min 20 each channel Cal gas Flowrate 4000 cc min approx Analog output Converter temp analog output 3 15 VDC 315 C 05386 Rev A 10 Model 200EU NOy Option Manual Addendum 2 2 Warranty ADVANCED POLLUTION INSTRUMENTATION INC 02024c Prior to shipment API equipment is thoroughly inspected and tested Should equipment failure occur API assures its customers that prompt service and support will be available COVERAGE After the warranty period and throughout the equipment lifetime API stands ready to provide on site or in plant service at reasonable rates similar to those of other manufacturers in the industry All maintenance and the first level of field troubleshooting are to be performed by the customer NON API MANUFACTURED EQUIPMENT Equipment provided but not manufactured by API is warranted and will be repai
22. year Sample Flow Check every 2 months Check efficiency every 2 months Factory Calibration Calibrate each year or after repairs Leak Check Check every 6 months Replace bypass flow orifices Every year as needed Rebuildy bypass pump Every year 05386 Rev A Model 200EU NOy Option Manual Addendum 5 2 Replacing the Sample Particulate Filter The particulate filter should be inspected often for signs of plugging or contamination It is also common for dirt particles to cause instrument drift and affect accuracy To check and change the filter 1 2 3 4 Fold down the Bypass Pump Module front panel Locate the filter on the left side of the analyzer front panel See Figure 5 1 for an exploded view of the filter assembly Visually inspect the filter through the glass window If the filter appears dirty unscrew the hold down ring remove the Teflon o ring and then the filter Replace the filter being careful that the element is fully seated in the bottom of the holder Replace the Teflon o ring then screw on the hold down ring and hand tighten NOTE Handle the wetted surfaces as little as possible and clean with cloth prior to assembly 05386 Rev A 20 Model 200EU NOy Option Manual Addendum K 7 7 N 4 N J lt N N pr n 7 i i N N p Figure 5 1 Replacing the Particulate Filter 05386 Rev A 21 Model 200EU NOy Option Manual Addendum 5 3 Check By